Professional Documents
Culture Documents
Bac Lab Rep 4
Bac Lab Rep 4
EXPERIMENT 4:
CHROMATOGRAPHY
METHODOLOGY
Materials:
1. Panadol soluble
2. 1 sachet 3 in 1 coffee
4. Perfume
6. Micropipette (1mL)
7. Micropipette tips
Method:
3. 1 gram of 3 in 1 coffee was dissolved in 50mL of boiled water and mixed with a stainless-
steel spoon.
4. 1mL of the coffee solution was transferred into a fresh HPLC/GC vial for HPLC analysis
after the coffee solution cool.
1. 950μL of hexane (HPLC grade) was transferred into 2mL HPLC/GC vial.
1. A previous data of serial diluted caffein standard was provided by laboratory staff to
students.
2. A standard curves for caffein was plotted from the data given.
3. The previous HPLC data for Panadol soluble will be provided by laboratory staff. The
amount of caffein in a tablet of Panadol soluble was calculated by comparing to the standard
curve. The result was compared with the original label on the Panadol soluble provided by
the manufacturer.
4. The previous HPLC data for 3 in 1 coffee will be provided by laboratory. The amount of
caffein in the coffee solution was calculated by comparing to the standard curve. The result
was compared with the original label on the 3 in 1 coffee sachet provided by the
manufacturer.
B) Samples Analyzed Using GC
1. The previous data of serial diluted linalool and benzyl acetate standards was provided to
students.
2. A standard curves was plotted from the data given for linalool and benzyl acetate.
3. The previous GC data for perfume was provided by the laboratory staff. The amount of
linalool and benzyl acetate perfume was calculated by comparing to the standard curves.
4. The previous GC-MS data for perfume will be provided by laboratory staff. All the
fragrance compounds in the perfume were identified.
RESULTS
Part A: HPLC Analysis of Panadol and 3 in 1 coffee
Peak Retention time (m) Peak area (mAU*s) Peak Area (%)
1 5.465 2.42885e4 100.0000
Table 1: Peak, retention time, peak area, and peak area percentage of Panadol soluble.
From Figure 1, we can conclude that there is no caffeine present in the Panadol soluble. This
is because there is no peak between retention time of 15 min and 16 min in the
chromatogram. The peak between 4 min and 6 min represent another source of compounds.
Therefore, the caffeine standard curve equation cannot be used to estimate the caffeine in
Panadol soluble since caffeine is absent in the Panadol soluble.
Graph 1: Peak Area against caffeine stock concentrations
Figure 2: HPLC analysis produced a chromatogram of the 3 in 1 coffee.
Peak Retention time (m) Peak area (mAU*s) Peak area (%)
1 2.044 154.07098 25.0802
2 2.552 52.01206 8.4667
3 4.613 14.11189 2.2972
4 5.774 33.54494 5.4605
5 15.734 360.57446 58.6954
Table 2: Peaks, retention times, peak areas, and peak area percentage of 3 in 1 coffee.
The equation y= 48265x + 146.83 was derived from the caffeine calibration curve (Graph 1).
In contrast, with a retention duration of 15.734 minutes and a peak area of 360.57446mAU*s,
the fifth peak in Table 2 showed the presence of caffeine in the 3 in 1 coffee sample.
Substituting y= 360.57446 into the equation y= 48265x + 146.83,
0.004 𝑋 50
Concentration of caffeine in the 3-in-1 coffee sample = 1000
The total mass of 3-in-1 coffee powder in a sachet of 3-in-1 coffee = 19g
0.0002 𝑋 19
Total concentration of caffeine in a sachet of 3-in-1 coffee = 1
D-Limonene
15682 005989-27-5 99
2 5.982 13.43 D-Limonene 15681 005989-27-5 94
Based on the Figure 3, Figure 4 and Table 3, with a retention time of 5.110 minutes,
the first peak was .beta.-Pinene,.beta.-Pinene, and Bicyclo[3.1.1]heptane, 6,6-dimethyl-2-
methylene-, (1S)-. The proportion of peak area was 2.28%, with a quality score of 91 for
each. At the second peak, followed by D-Limonene, D-Limonene, and Limonene,
respectively. They have a total retention period of 5.982 minutes and a peak area of 13.43%.
The first D-limonene has a quality of 99, the second D-limonene has a quality of 94, and
Limonene has a quality of 93. Three 1,6-Octadien-3-ol, 3,7-dimethyl were found on the third
peak, each with a distinct quality value: the first had a quality value of 97, the second had a
quality value of 95, and the third had a quality value of 80. They had a retention time of 7.120
minutes and a peak area of 9.52%. 1,6-Octadien-3-ol, 3,7-dimethyl-, propanoate, 1,5-
Dimethyl-1-vinyl-4-hexenyl butyrate, and Linalyl acetate were detected on the fourth peak.
They had a retention duration of 9.417 minutes and a peak area of 36.82%. 1,6-Octadien-3-ol,
3,7-dimethyl-, propanoate and 1,5-Dimethyl-1-vinyl-4-hexenyl butyrate both had 90 of
quality value while for Linalyl acetate 83 of quality value. On the fifth peak, Isobornyl
acetate, Bicyclo[2.2.1]heptan-2-ol, 1,7,7-trimethyl-, acetate, (1S-endo)-, and Acetic acid,
1,7,7-trimethyl-bicyclo[2.2.1]hept-2-yl ester were found. They had a retention time was 9.799
minutes, and their percentage of peak area was 6.64%. Their respective quality values were
98, 91, and 91. Three Diethyl Phthalate were discovered in the last peak, which was the sixth
peak. They had a retention time of 13.845 minutes and a peak area of 31.30 percent. Diethyl
Phthalate had a quality rating of 97, whereas the other had a quality score of 96. All the
components identified have quality above 90.
Table 3 shows the results of the GC-MS perfume analysis. As was shown, the peak with
the longest retention time was the sixth peak, at 13.845 minutes. The longer retention time
indicates the slower the components will be eluted out. The first peak which has retention
time of 5.110m elute fastest among the components in perfume while peak 6th with retention
time of 13.845m elutes the slowest. The fourth peak had the highest proportion of peak area
at 36.82%, which means its concentration is highest among other components in the perfume
essential oil sample. D-Limonene had the highest quality of any ingredient, with a result of
99.
The precautions that should be taken are membrane filters should be used when
handling HPLC because if they are not used, particles can block the column when the sample
is processed through the HPLC column. If there are many particles, the clogging is almost
always irreversible. The second step is to degas the solvent used in the column run. The
apparatus generates pressure in order to flush the solvent through the column. As a result, if
air bubbles are caught in the solvent, the pressure on the HPLC pumps may rise dramatically.
HPLC operates at constant pressure, flow rate, and temperature, as we all know. The presence
of air bubbles can cause pressure and flow rate to be disrupted. In order to keep the mobile
phase (solvents) free of air bubbles, degassing is required. If any particle escapes into the
flow despite taking safeguards, they will be captured by the guard column. A guard column is
a tiny column that is placed in front of the main HPLC column. We also need to eliminate
particle matter and air bubbles, as we've observed earlier. When employing GC-MS analysis,
take precautions such as performing frequent visual inspections and pressure leak checks of
the sampling system plumbing, fittings, and valves, and installing columns according to the
manufacturer's recommendations. Handle glass or fused capillary columns with care and
wear safety glasses to protect your eyes from falling objects while managing, trimming, or
updating them. Switch off and enable heated areas such as the oven, inlet, and detector, as
well as attached equipment, to decrease the temperature before touching them, and turn off
the hydrogen gas supply at its source when modifying columns or maintenance the
instrument.
CONCLUSION
In this experiment, the calibration curve was determined. The calibration curve of
caffeine was obtained with the equation y= 48265x + 146.83. The graph's regression line was
0.9962, indicating almost a fit line. The fifth peak in Table 2 showed the presence of caffeine
in the 3 in 1 coffee sample with a retention duration of 15.734 minutes and a peak area of
360.57446mAU*s.
In addition, HPLC was used to examine Panadol soluble and 3 in 1 coffee. The retention
time for the only peak shown in chromatogram of Panadol was 5.465 minutes, which
indicates the absence of caffeine. For 3 in 1 coffee, the retention time was 15.734 minutes.
The retention time for caffeine is between 15 min and 16 min. therefore, the 3 in 1 coffee
shows presence of caffeine.
Besides, the data analysis of GC-MS of perfume were obtained and tabulated in Table 3.
The peak with the longest retention time, at 13.845 minutes, was the sixth peak, as illustrated,
which represent the components in peak 6th will be the last component to elute out. With
36.82 percent of peak area, the fourth peak had the highest proportion of peak area, which
indicates the highest concentration of component among other components in perfume. With
a score of 99, D-Limonene has the greatest quality of any component. This indicate that D-
Limonene had less contamination.
REFERENCES
1. Staff, S. A. A. (2019, October 17). What is Chromatography and How Does it Work?
Ask a Scientist. https://www.thermofisher.com/blog/ask-a-scientist/what-is-
chromatography/#:%7E:text=Chromatography%20is%20a%20process%20for%20sep
arating%20components%20of%20a%20mixture.&text=The%20different%20compon
ents%20of%20the,to%20separate%20from%20one%20another.
4. Nugraha, A., & Nandiyanto, A. (2021). How to read and Interpret GC/MS
Spectra. Indonesian Journal of Multidiciplinary Research, 1(2), 171-206. Retrieved
from https://ejournal.upi.edu/index.php/IJOMR/article/view/35191
11. Understanding the Difference Between Retention Time and Relative Retention Time.
(1 August, 2014). Retrieved from Chromatography Today :
https://www.chromatographytoday.com/news/autosamplers/36/breaking-
news/understanding-the-difference-between-retention-time-and-relative-retention-
time/31166
12. What is Retention Time? (31 July, 2014). Retrieved from Chromatography Today:
https://www.chromatographytoday.com/news/gc-mdgc/32/breaking-news/what-is-
retention-time/31159
APPENDICES