Professional Documents
Culture Documents
Chorus-Trio Expander Service Manual Rev. 2 07.2021 en
Chorus-Trio Expander Service Manual Rev. 2 07.2021 en
FOR CHORUS TRIO (REF. 81200) AND CHORUS TRIO RAS (REF. 81220)
TECHNICAL SUPPORT
Strada dei Laghi 39, 53035 Monteriggioni (SI) Italy
Tel. +39 0577 307109, Fax. + 39 0577 307106
Numero Verde: 800 606932
e-mail: technicalsupport@diesse.it
No page in this manual may be reproduced in any form or by any means, electronic,
mechanical or otherwise, for any use whatsoever without prior written permission
from DIESSE Diagnostica Senese S.p.A.
Manufacturing date
Serial number
Manufacturer
Protective conductor
Attention: please read this manual carefully and comply with the
safety symbols.
Crush hazard: moving parts can crush and cut; keep hands away
Risk of thermal shock due to the warming of the tray for samples
Firmware version 7.00, revision 0 of this manual corresponds to the CHORUS TRIO
model of the instrument with the 7.xx family of firmware installed.
It was drafted and carefully reviewed and this version is closely related to the
instrument model (data can be obtained from the instrument’s ID plate) and the
version of the software that controls it (data can be obtained through a procedure on
the instrument itself).
It must be read carefully before using the instrument, especially the parts relating to
safety.
ATTENTION
Do not use this manual except in its complete form.
If this manual is used in an incomplete form, DIESSE Diagnostica Senese S.p.A.
declines any responsibility for adverse results.
ATTENTION
Any instrument updating done with the customer’s authorization requires that the
user manual be updated in a corresponding manner.
Fig. 1-1
Since the strip has two reaction cuvettes, wash and reading stations are created so
that the operations can be performed on both cuvettes simultaneously.
No. Description
1 Locking ring-nut
2 Upper surface
3 Column
4 Lower surface
5 Strip tray
6 Tray synchronization device (TSD)
7 Absolute tray sensor
8 Tray shaft with pulley
9 Drive belt
10 Drive shaft with pulley
11 Temporary lock nut
12 Tensioning bracket
13 Hole sensor (relative)
No. Description
1 Motor
2 Temporary fastening nut
3 Motor unit movement slide
4 Belt tensioning screw
1. locking ring-nut
2. upper surface
3. column
4. lower surface
1
2
5. strip tray
6. Tray Synchronization
13 Device (TSD)
3
7. absolute tray sensor
12 8. drive shaft with pulley
5
4
9. drive belt
11 10. motor with pulley
9
8 6
11. temporary fastening
10
7
nut
12. tensioning bracket
13. hole sensor
14. (relative)
Fig. 1-4
The rotation control software analyses the time intervals of the passages between the
dark zones and the illuminated zones and checks that the rotor is turning correctly.
Fig. 1-5
1.3.3 SYNCHRONIZATION
Once the rotation is finished, or after the shaking of the plate has ended, the tray is
correctly positioned by a mechanical unit made up of a special device called the Tray
Synchronization Device (TSD) or simply the tray synchronizer, which inserts a cone-
headed cylinder in one of the holes beneath the tray that correspond to each of the
30 strip insertion positions.
1 A
3
1. conical pin of the TSD
4
2. positioning hole
3. relative position sensor
Fig. 1-6
The strip is inserted in the slot and is kept in position by the spring located at the end
of the slot, which keeps the strip locked in position by holding the last well.
Upon insertion the strip must be pushed to the bottom of the seat so it can be held
by the spring.
The insertion of the strip into the slot is aided by the bevelling of the lower edge of
the plate.
There are two holes at the bottom of every slot that allow the optical ray to pass
through. The diagram below shows the top view of a plate with 28 strips fully inserted,
one strip partially inserted and one position empty.
7 6
Fig. 1-7
1. lock pin
2. centering hole
3. solenoid
1
3
Fig. 1-8
The plate alignment is obtained acting on the Tray offset parameter. This parameter
represents the number of steps that the tray must carry out, starting from the
absolute position (which is determined by the absolute sensor Fig. 1.4), to reach the
correct alignment. The following operations can be performed to check the
calibration:
1. Open the Motor window:
Chorus / Utility / Service / Diagnostics / Motors
2. Run the Reset Tray with TSD command.
3. Check that the tray doesn’t move when the TSD enters in the tray cavity.
4. If the TSD drag the tray counterclockwise, decrease the Tray offset parameter
until the desired result is obtained.
If the TSD drag the tray clockwise, increase the Tray offset parameter.
Double click on the row Tray offset and enter the number of correction steps, then
confirm with OK.
2
1. light emitter
2. receiver
Fig. 2-1
1. barcode reader
3
2. reader bracket
4 3. barcode reading beam
4. tray upper surface
5. strip being read
Fig. 2-2
Figure 2-3 shows the connections diagram of the internal barcode reader (front view).
CAB533 connects directly with the 2010 CPU while the LD2 LED represents the power
supply LED.
Fig. 2-3
7. If these conditions are not respected, loosen the 2 screws that lock the reader
support to the upper plane and move the support until reaching the optimal
position, repeating the operations at points 5 and 6, then tighten the screws
again
8. Open the Test Macros window:
Chorus / Utility / Service / Diagnostics / Test Macros
9. Run the Startup Check command and wait the reset of all the motors.
10. Insert others strips in the tray and open the Identification Test window:
Chorus / Utility / Service / Diagnostics / Test Macros / Identification Test
11. Run the Scan command and check that all the strips are read correctly.
Bput
CS sen.
(1-4)
250ul
closure
cap
SV9 SV8
SV3 Aput
ON
ON
closure
cap SV7
disp #1
disp #2
SV4
ON level sensor
ON level sensor
Pp3
SV5 Sput#2 Sput#1
Pasc Pasc
Plav
Fig. 3-1
pp3: the peristaltic pump that takes in wash water from the tank and distributes it
inside the instrument.
SV3: enables the connection of the dispensing needles or the liquid coming from
pump pp3 or the syringes.
SV4: enables the delivery of water for the external washing of the needles
SV5: enables the external wash flow on disp1 or disp2.
SV7: enables the flow selected by ev3 on disp1 or disp2.
SV8: guides the inlet/outlet flow from the 250 ml syringe.
SV9: ---
As can be seen in the figure, each needle has a pair of wells (Plav, Pasc) with the
following characteristics:
▪ The wash wells (Plav) have an internal spout that delivers a jet of water onto
the tip of the needle when it is inserted into the well.
▪ The drying well (Pasc) has a large mesh sponge that retains any drops of wash
water that collect on the outer side of the needle.
Each of these wells has a drain that directs the water to the main waste well.
Every needle has a capacitive level sensor that is able to detect a minimum of 50 µl
and has a sensitivity of 10 µl. The inside and outside of every needle tip is covered with
a ceramic material that reduces carry-over to almost zero.
1 2 3 4
#2
Fig. 3-2
3.3 DISPENSER
The dispenser is a device that is designed to move the dispensing needle up and
down within the wells of the strips, for drawing up or dispensing liquid.
The dispensing needle is a stainless steel tube with its tip cut at 45°, and which has
been sharpened to help cut through the protective film of the wells. The coarseness
of the end of the needle, both internally and externally, is increased by a micrometric
ceramic deposit that helps reduce carry-over to a minimum. The upper part of the
needle is shaped to stop the tube.
The two prongs of the block’s fork are then inserted in the special holes to prevent
the needle from coming out of position. The thrust of the spring ensures needle
stoppage. The stability of the needle is also ensured by the fork which stops the tube.
The needle block is mounted on a bracket that is connected to the slider which moves
up and down along the slide. The bracket is integrated with a rack and therefore the
rotation of the motor, with flush fit shaft pinion, produces the vertical movement of
the needle.
A U-shaped aluminium bracket supports all the parts and its base is fixed to the
sliding base plate of the Dispenser Carriage.
1. bracket
2. tube
3. connecting bracket
sinistra destra 2
4. needle
1
A 14
5. level sensor wire
6. needle cylinder block fork
3
13 7. insulated needle-holding
block
12 4 5 15 8. slider
9. slide
6
10. toothed rack
7
11. pinion
11 8 16
12. motor
17
10
9 18
13. position sensor
A 14. fork
sezione
15. slider for the drum slide
DESTRA FRONTE SINISTRA A-A
16. spring
17. tapering
18. tip
Fig. 3-3
The carriage runs along the slide since it is fastened to the slider (which cannot be
seen in the diagram). The carriage is attached to the toothed belt by a hook that
comes out of the belt.
The toothed belt is driven by a motor through the pulley. The other end of the belt is
kept taut by a base plate on which an idle pulley is fastened.
The motor is fastened to the upper surface by an angular bracket. The position of this
angular bracket can be adjusted along the special slots, so that the belt tension can
be modified.
To allow the dispensing needles to reach the wash well, a special passage area is
opened on the upper surface in the area for accessing the strip area.
3 4 5 6 7 8
9
1. drive hook
2
2. frame’s upper surface
10
11 3. motor
4. angular bracket
5. adjustment slots
12
6. motor pulley
1 7. position sensor
8. sliding carriage
9. threaded holes
10. wash access area
11. slide
12. access to the strip wells
13. toothed belt
14. idle pulley
13
15. base plate pulley
15 14
Fig. 3-4
The unit is made up of a stepper motor connected by a joint to a worm screw. Based
on the motor’s direction of rotation, this unit acts on the sliding block and, therefore,
the slider mounted on its lower part, to provide movement in one direction or the
other.
For positioning, the magnetic sensor mounted on the electronic board and the
corresponding magnet mounted on the slider is used.
Two Teflon pads allow the slider to move fluidly over the guide slot of the base plate.
The thrust plate and the end bracket keep the worm screw in line through the use of
internal bearings.
The upper part of the two syringes are coupled to the Plexiglas syringe unit and the
middle portion of the syringes is coupled to the clamping bracket. The pistons are
mounted on the fixing bracket located on the slider.
Based on the amount of liquid to be processed, the solenoid valve (14) mounted on
the top of the syringe unit opens and closes the syringe lines, thus allowing the
instrument to withdraw or dispense liquids.
1. motor
2. joint
3. thrust plate
4. worm screw
5. sliding block
6. slider
7. end bracket
8. pads
9. guide slot
10. syringe unit
11. syringes
12. clamping bracket
13. fixing bracket
14. solenoid valves
15. position sensor
Fig. 3-5
In particular:
▪ the section of the circuit that goes from the cleaning solution tank to the
syringe
▪ the syringe
▪ the section connecting Sv3 to Sv7
▪ the tubing that connects the two outlets of ev7 with the dispensing needles
▪ the needles (2/3 of their capacity, thus leaving the entire tip dry).
The tubing in which wash water flows must be completely filled so that the system
functions properly for all the tests.
To do this, allow water to flow through the tubes until it exits into the main waste and
can be detected. This procedure is also performed automatically when the
instrument priming is started.
3.6.4 Withdrawal
The removal of liquid from a well, done at the cuvette bottom level, which guarantees
that the entire contents of the well can be drawn up.
Starting cuvette is the starting cuvette (or well) number from which the liquid must
be drawn up.
The dispensing function is organized so that, after the liquid has been withdrawn
from the starting cuvette, the amount equal to the Quantity #1 parameter can be
dispensed into the cuvette defined to be Cuvette #1 and the amount equal to the
Quantity #2 parameter can be dispensed into the cuvette defined to be Cuvette #2.
The amount withdrawn from the Starting cuvette is therefore the total of the
Quantity #1 and Quantity #2 amounts.
Fig. 3-6
The duration of the needle washing is defined in the Wash mode parameter, which
is set whenever a certain transfer procedure is defined.
The final washing of the needle is skipped if the next transfer belongs to the same
family as that being processed. This functional characteristic can save considerable
amounts of wash water.
The pre-boring creates a hole in the membrane where air can safely enter.
This is necessary since the sides of the hole made in the membrane for withdrawing
can seal around the needle itself, thus reducing or preventing air from entering.
The lack of incoming air into the well during the suction phase causes a lesser
amount of liquid to be removed with respect to that programmed, thus leading to
measurement errors.
Fig. 3-7
The dispensing needles are made with oblique tips and with sharp edges. The cutting
of the membrane may be imperfect due to a problem on the needle caused by
wearing or an accidental impact on the tray or if the membrane is too tough due to
an imperfect heat sealing.
A check of the hole and the tip’s cutting ability must therefore be included for every
perforation operation.
If the return to the home vertical level is correct, the difference between the
theoretical number of steps between cuvette bottom level and home vertical level
and that actually performed is calculated.
▪ outside the ± acceptance range: a membrane error is sent and the sample is
not processed.
Note that with this condition the return to the home vertical position cannot be
carried out.
In this case, try vertical repositioning two times. If only the first reset is wrong, the
sample is skipped and the run continues. If the second reset is also wrong, the run is
stopped.
The sensitivity of the sensor is measured as: minimum detectable amount of liquid (≥
50 μl) and min 10 μl discrimination, which indicate the minimum level difference that
can be detected in cuvettes containing at least 50 μl.
The limiting factor is that it cannot be used for membrane covered cuvettes, as this
would cause measurement errors.
3.6.8 Mixing
After having distributed a liquid in a cuvette which already had other liquid (ex: serum
in the diluent cuvette), the solution needs to be mixed in order to make it
homogeneous.
The shaking of the plate is only sufficient for keeping homogeneous solutions in
suspension.
1. The needle gets to the bottom cuvette level, where it dispensed the liquid into
the cuvette and the dosing syringe is completely closed.
Warning: the mixing mode requires that the liquid to be mixed be taken without any
extra amount, in order to avoid bubbles from forming.
2. Now, with the needle still at the cuvette bottom level, the Shaking quantity is
drawn in for the first time and then expelled. This is done the number of times
indicated by the Shaking number.
3. When finished, the needle exits the cuvette and is washed.
#1 #2
1. washing of Disp #1
#2 #1 #2
#1
1. drying of
Disp #1
1 2
2. disp #1 carries out a
vertical reset
#2 #1 #2
#1 1. disp #1 moves and
makes a hole in the
1 2 membrane of Cuvette #1
2. disp #1 carries out a
vertical reset
#2 #1 #2
1. disp #1 dispenses
#1 Quantity #1
1 2
2. disp #1 carries out a
vertical reset
#2
Fig. 3-8
Warning: check that the nozzles that supply wash water are correctly positioned so that the
water hits the needle.
NOTE: since the error check was activated, the warning alarm must activate once the warning
level has been reached and pump PP8 must switch on to empty the waste well.
3.7.2 Dispensers
3.7.2.1 Dispensers Needles Height Calibration
For the calibration of the heights of the dispensers needles, proceed as follow:
Fig. 3-9
6. The difference between the expected and measured volume must be equal to
0±10 µl.
FUNCTIONAL TESTS
Tray Position: The tray position where the strip has been inserted
Dispenser
Needle to use for the transfer. The possible values are 1
(Disp#1=norm,
(standard) or 2 (conjugated).
Disp#2=conj):
If the transfer was successful, the instrument will not print any report.
If, during the test, the instrument finds an error, a report with the difference between
the expected values (Ex) and the measured ones (Ms) will be printed:
Ex. TRASF:Smp# -; Stp# -; Tr# -Ex: 100; Ms:90; diff : -10 ; MaxDiff :50
Fig. 3-11
8. Since two perforations are performed, each well is characterized by two data
lines. The maximum Diff must be lower then 15 steps. If during the test the Diff
measured is higher then 15 steps, it is necessary to replace the needle.
The focused light passes through a monochromatic filter, which is selected based on
the type of test to perform. A heat filter is inserted in order to prevent the filter and
the optical fibre from overheating.
1. optical fibres
2. cylinder
3. monochromatic
filter
4. LED group
5. Power control
6. Electronic control
Fig. 4-1
The electronic circuit for regulating the control voltage is what makes the lighting
device particular. It is interesting to analyze the control circuit starting from the
power circuit.
1. the light that each line can transmit is different from the other, even by 30%,
depending on the light gradient that covers the head of the cylinder and the
lack of radial symmetry of the line terminals collected in the cylinder.
2. If, in the three measuring stations, the same amount of light does not traverse
each cuvette, homogeneity of the measuring points cannot be reached.
In order for all the lines in an optical fibre to be homogeneous from a light emission
point of view, a calibration procedure is required, managed by the central
microprocessor, that guides the lamp’s electronic control, which in turn is able to
generate approximately 800 voltage values within the lamp’s operating range of 5V
DC to 12 V DC.
Therefore, a voltage value is determined for each optical line in order to generate light
beams of the same intensity.
5 7 1. optical fibre
2 1 2. optical fibre terminus
8 ferrule
4 3 6 3. optical channel
4. lenses
5. optical fibre support
6. upper surface
7. strip tray
10 8. strip
9. focalization point
10. lenses
11. optical channel
12. optical receiver
9
13 13. receiver card
14. lower surface
12 11 14
Fig. 4-2
The light ray focuses a 1.5 mm spot about 1 mm from the bottom of the cuvette.
After it has passed through the strip, the light beam is focalized by a pair of dual lenses
on the optical receiver positioned on the electronic receiving board. The diameter is
3 mm and ensures a good signal/disturbance ratio.
The translated voltage signal is sent from the conditioning electronics to the A/D
converter.
When one of these elements is modified, the characteristics of the optical channel
are modified.
The filter block is mounted on the plate which, in turn, is fastened onto a bracket (3)
that connects it to the slider, which is moved up and down along the slide (5). The
bracket has a toothed rack and therefore the rotation of the motor (8), with flush fit
shaft pinion (7), allows the filter block to move.
A U-shaped aluminium bracket supports all the parts and is fixed to the support of
the fiber optic cylinder.
1
1. bracket
left right 2 2. filter block
3
10 3. connecting bracket
11
9 4. slider
5. slide
8
12
6. toothed rack
7. pinion
13
8. motor
7 4
9. position sensor
10. plate
6
5 11. filter #1
12. filter #2
RIGHT FRONT LEFT 13. no filter
Fig. 4-3
The higher the value of the received signal, with regard to emitted light, means more
light has passed through the test solution and therefore less turbidity. Transmittance
is thus the primary measurement that is made in the optical channel.
In the Chorus TRIO, the amount of light received (transmittance) is initially detected
as electrical voltage and then converted with a 12-bit A/D into a number that can
range from 0-1023.
0 corresponds to the absolute lack of transmitted light in the optical channel (dark)
and 1023 is the maximum value of receivable light (channel completely free).
1800
800
1500
1200
600
900
400
600
200
300
1/32
1/128
1/16
Fig. 4-4
Transmittance is however a relative measurement that does not take into account
the characteristics of the optical channel. Therefore, in the presence of more than one
optical channel, a different parameter must be used to have comparable
measurements: absorbance. Absorbance is a measurement of the ability to absorb
(and therefore not allow to pass through) light emitted by the emitter. The higher the
absorbance the lower the turbidity of the solution.
The absorbance (Abs), for an optical channel, in its theoretical formulation, is given by
the formula:
This formulation does not consider the problem of using electronic amplifiers and
analogue/digital converters. In fact, the transmittance voltage reading of a
transmittance signal is such that the darkness value (theoretical transmittance = 0) is
a residual voltage value (Toff.set), which is subtracted from every transmittance reading.
(𝑇𝐻20 – 𝑇𝑜𝑓𝑓−𝑑𝑎𝑟𝑘 )
𝐴𝑏𝑠 = 𝑙𝑜𝑔
(𝑇𝑠𝑜𝑙𝑛 − 𝑇𝑜𝑓𝑓−𝑑𝑎𝑟𝑘 )
Since the optical calibration must be done in a completely automatic manner, thus
without operator intervention, the measurement of transmittance in water was
made the same as that in air. A modest absolute error is therefore added for the
absorbance differences with regard to the instrument, but which does not influence
the final result.
It should be noted that the assessment of the optical values is done in digital terms,
on a numeric scale that theoretically ranges from 0 to 1023.
Given that an optical reading depends primarily on the light emitted by the light
source and then by the upstream amplification device, six optical fibres must emit
the same amount of light, with a precision on the order of 5 ‰.
Fig. 4-5
To obtain the same optical response from displaced optical fibres, the incident light
must be different depending on the optical fibre taken into consideration. This means
that the lamp is powered with a different voltage depending on the optical fibre to
be controlled.
Given that, during a step, all six optical channels may need to be read, the time
available for each channel is slightly higher.
Since the adjustment time for lamp luminosity (from 5V to 12 volt) is less than 2
seconds, emission stability can be obtained for enough time to perform a reading of
an optical channel after this interval. IV° fibra
V° fibra
VI° fibra
I° fibra
II° fibra
III° fibra
vibration time
11 sec
18 sec
0 sec
Fig. 4-6
In each of the time intervals in which the tray waits for the step, it’s therefore possible
to:
▪ modify the lamp’s control voltage to the required value, for example for the 1st
optical fibre
▪ perform the reading on the relative optical channel.
120
100
80
60
40
20
0
0 1 2 3 4 5 6 7 8
Fig. 4-7
In the second case, the digital indication at 0 does not mean that the output from the optical channel’s
amplifier is exactly at 0 analogical volts, but could have a negative value due to the amplifier’s inherent
offset. If we were to accept this value, it would mean losing the evaluation of optical values which, in
analogical terms, come from the negative value of the offset up to zero. A digital value for the dark should
therefore be set that gives an indication > 0, for example 20 digit.
The calibration of the dark offset therefore requires (see diagram) that a voltage ramp
be generated that progresses from -1 Volt to +1 Volt until a value not greater than 20
digit coming out of the A/D converter is obtained.
optical
receiver A/D converter
Fig. 4-8
The potentiometer setting value for each optical channel is stored in the instrument’s
Flash-ROM and is loaded at start-up or recalculated after every optical calibration.
The calibration must be done with the instrument closed, or properly darkened, so
that the interleaver sensors do not receive external light.
The light intensity that allows a digital value of 1000 to be read is therefore set as the
maximum reference intensity for each optical fibre and for each of the available
optical filters.
fixed gain
amplifier
light
power optical
supply receiver A/D converter
digital voltage
ramp ramp
generator generator
Fig. 4-9
Having then set the gain of the amplification and conditioning circuits of the six
optical channels, the only way to obtain a light reception equal to 1000 is to act on
the lamp’s control voltage by aligning the intensities of the light transmitted by the
six optical fibres (as described previously).
The control voltage of each channel is scanned from 5 to 12 Volts in 10 mV steps using
the “bisection” method, until the value that gives a reading of 1000 is determined.
The control of the power occurs digitally and the digital value that corresponds to the
calibration of the optical fibre, for a certain wavelength, is stored in the flash memory
and loaded at instrument start-up. It is modified with each calibration.
One of them functions as the control range selector (lamp voltage range) and varies
from 0 to 7, and the other as a fine selector (lamp volt. lev.) and varies from 0 to 127.
8.0 8.0
4.0 4.0
127
31
41
38
37
43
29
48
0 127
0
a) b)
Fig. 4-10
As can be seen in Fig. 4-10 b), the passage from one characteristic to the next takes
place, increasing the range and restarting the count from 0, when the connection
point with the next characteristic is reached. This gives a lesser number of resolution
points since the characteristics no longer start from 0, but instead from the value
corresponding to the intersection and indicated in the figure.
This provides a true characteristic equal to that in fig. C, where there is a virtual
counter that lays out 749 horizontal points [ 127 + (127-31) + (127 – 41) + (127 – 38) + (127
– 37) + (127 - 43) + (127 – 29) + (127 – 48) for scanning voltages from 5 to 12 Volts.
volt
12.0
8.0
4.0
784
0
Fig. 4-11
The voltage on the lamp can be determined using the following formula:
Fig. 4-13
V. ramp starts the control procedure for the virtual ramp
Dark calibrates the offset values and reports them in the dark column
Light starts the adjustment of the light emission at a value of 1000 dgt for each
channel, using an empty cuvette
The following data are reported:
the transmittance (Air) the virtual ramp value (Vir)
the voltage control value inside the range (Val)
The selected range (Ran)
The dark value (Dark)
Check: performs a complete reading of the optical channels using the selected
parameters
Print: Prints the calibration report displayed on the screen
The check of the optical system is carried out during the testing phase of the
instrument with specific control solutions.
1. Prepare six LAB type strips (1899), with the 2 Labsystem reaction wells (#5 and
#6) correctly inserted and both filled with the following solutions:
Warning: The values can change if the calibration solutions are changed or if they were
improperly stored
Fig. 4-14
1. Prepare six LAB type strips (1899), with the 2 Labsystem reaction wells (#5 and
#6) correctly inserted and both filled with the following solutions:
Fig. 4-15
CS sen.
(1-4) closure
cap Pp7-1
21
SV10
Pp7-2
Pp2
SV2 S1
Pp6
closure
Buff. Sol. cap 25 Pp4-1
Autoimm. SV11
SV1
Pp4-2
Pp1
SV12 S2
B1S sen
(1-4) Pp5
Buff. Sol.
Infective SV6 28
P1
B2S sen
(1-4) Wasting
Fig. 5-1
SV1: solenoid valve for exchange between the autoimmunity washing-buffer solution and the air
SV2: solenoid valve for exchange between the ev12 selection and the cleaning solution
SV6: solenoid valve for exchange between the infective washing-buffer solution and the air
SV10: solenoid valve for stopping the flow of washer #1
SV11: solenoid valve for stopping the flow of washer #2
SV12: solenoid valve for the exchange of the buffer tanks
Pp1: supply pump in the needles of washer #2
Pp2: supply pump in the needles of washer #1
Pp4: aspiration pump from the needles of washer #2
Pp5: aspiration pump from the collection basin of washer #2
Pp6: aspiration pump from the collection basin of washer #1
Pp7: aspiration pump from the needles of washer #1
P1: aspiration pump from the needles of washer #3
▪ a plastic tube with relative connector (6) for the drawing up of the liquid,
highlighted with a blue, green or white band (8)
▪ an electric wire, with relative connector (7), for detecting the liquid level
which are attached to the respective connectors located in the instrument’s tank
chamber
Warning: Each probe must always be used with the same solution in order to prevent cross
contaminations.
Use the probe with the blue band for the tank with the Infective Washing Buffer, that with the
green band for the tank with the Autoimmunity Washing Buffer and that with the white band
for the Cleaning solution, following the color-coding indicated on the instrument’s connectors
Once connected, the probe takes the liquid from the spout and puts it into the
hydraulic circuit through the fitting.
The four sensors, located on the probe rod, report the liquid level in the tank. The
sensor signals are sent to the instrument through the electric connector.
7 1. withdrawal spout
2. 0% level sensor
3. 25% level sensor
1 2 3 4 5 8 4. 50% level sensor
6
5. 75% level sensor
6. hydraulic fitting
7. electric connector
8. band
Fig. 5-2
0%: tank empty – the instrument stops any ongoing cycle and reports a
warning
25%: tank almost empty – if the buffer solution drops below this level, a
warning will be generated during the initial check and the cycle cannot
be started
50%: if the washing solution drops below this level, a warning will be
generated during the initial check and the cycle cannot be started
75%: first control level
1
A sezione
left A-A
right
2
12
11 1. bracket
3 2. connecting bracket
10
3. needles
4. needle-holding block
9
13 5. slider
6. slide
14
7. toothed rack
8. pinion
4
9. motor
8 5 15 10. position sensor
7
11. aspiration tube
6 12. supply tube
A 16
13. tube casing
14. spring
15. axial bearing
RIGHT FRONT LEFT 17 16. tapering
17. Teflon covering
Fig. 5-3
Fig. 5-4
A washing station is a device that can drop a bifilar needle into either of the reaction
cuvettes. The needle is composed of one larger diameter straight needle and one
smaller diameter needle that is tapered at the end. The two tips are Teflon coated in
order to prevent the formation of drops.
The straight needle is used to remove the liquid in the well. The tapered needle is
used to dispense washing buffer.
The two needles move up and down together and the descent into the bottom of the
well is dampened by a spring.
Fig. 5-5
The diagram shows the section below the strip introduction tray.
As can be seen, there is a collection basin. When there are no strips in the tray, the
contents of the needles are unloaded in the well during the priming or cleaning of
the needles.
The well is drained by pump pp6, if we are in position 21 or by pump pp5 if we are in
position 25.
For safety purposes, each well is equipped with a pair of sensors that can report any
overflowing of liquid.
1 2 3 4
5 6 7
Fig. 5-6
▪ one tank with infettivology washing buffer and connect it to the probe marked
with blue colour;
▪ one tank with cleaning solution and connect it to the probe marked with white
colour;
▪ remove the strips from the carousel and reset the tray.
Increase the steps number to lower the volume or decrease the steps number
to increase the volume.
Increase the steps number to lower the volume or decrease the steps number
to increase the volume.
The circuit is designed so that the Chorus TRIO can be connected to a collection tank
or to a central drain.
Fig. 6-1
The following devices intervene:
11
10 3 1. aspiration pump a
2. waste intake (blue)
9 3. liquids intake
4. pump expulsion
2 5. waste outlet (red)
1 6. outlet line
The waste well is the device that regulates the evacuation of the liquids not needed
by the Chorus TRIO.
Peristaltic pump pp8 takes in the liquid collected from the fitting and sends it
towards the red outlet, connected to the well by the junction, through the outlet line;
the connection to the blue outlet is ensured by the fitting.
Besides collecting and expelling waste liquids, the well monitors the proper
functioning of the circuit through the use of an auxiliary waste outlet well, connected
to the outlet line by the fitting and to a series of level sensors.
Since the hydraulic circuit is directly controlled by the waste well, the use of external
level sensors are therefore not necessary.
Fig. 6-3
The liquids to be discarded are sent through one inlet (waste inlet).
When the liquid reaches the level of the warning sensor WS (S4), drain pump pp8 is
immediately activated and the liquid is evacuated through the output indicated with
the colour red on the rear panel.
The liquid flows toward the outlet without entering in the recirculation circuit, as the
diameter of the tube for this circuit is much smaller.
The intervention of the warning sensor is not reported during normal operation.
Fig. 6-4
When the drain is obstructed or cannot be used, for example if the tank is not
connected, the liquid pushed by pump pp8 is sent into the recirculation circuit and
then returns into the waste well, raising the liquid level inside the well.
There is an error sensor ES (S3) in the well, placed at the maximum allowable filling
level.
The instrument stops and a fatal error is generated when the sensor is reached.
If the obstructed drain is resolved before the ES is reached, the liquid is evacuated as
normal by pump pp8.
Fig. 6-5
To summarize:
ES activated = drain problems
Fig. 6-6
When drain pump pp8 has broken down, the liquid continues to flow into the waste
well, initially activating the warning sensor (WS), and then the error sensor (ES) when
the well is full. A fatal error is generated at this point.
First of all, the instrument is considered to be “closed” from a thermal point of view,
since the container that houses it is thermally insulated. It should be noted that there
are heat sources inside the instrument: heater, power supply, lamp, motor drivers,
motors, etc.
The following components are found inside the instrument to regulate the
temperature:
1. fans
2 2. protective housing
3. instrument temperature
sensor
1 4. air vents
Fig. 7-1
Heating occurs by the emission of hot air into the heat chamber (where the tray
rotates) by means of a heater, where a fan directs hot air into the measuring chamber.
The hot air also heats the upper part of the tray for the strips and consequently the
strips themselves.
The input of hot air is regulated by a system that uses the temperature sensor inside
the chamber.
6 1. lower base
2. heater
4 3. strip temperature sensor
4. measuring chamber walls
5. heat chamber (internal)
3 1
5 2 6. tray
Fig. 7-2
7.3 HEATER
1 Fan
2 Housing
3 Heating coil
7
5 6 4 Dust filter
5 Flange
2
6 Fastening holes
4 1
Fig. 7-3
The heater is the device that works to keep the temperature in the heat chamber
stable.
It is composed of a coil wrapped around the heating unit and encased by a PVC cover;
a 50x50 mm wide fan with a 24 V DC power supply is used to transfer the heat into
the chamber. This hot air then passes into the heat chamber through the heater
cylinder.
The impurity filter is positioned between the fan and the protective grill.
The heater is mounted on the upper plate of the Chorus TRIO by means of the 2
fastening holes in the support flange.
The cycle will only restart if there are tests in the identification that have the same
operating temperature.
If the temperature of the tray is not at the required temperature, the cycle will only
restart when the system will reach the cycle temperature.
The stand-by temperature prevents the tray from cooling and reduces the successive
heating time of the cuvettes. This stand-by temperature is normally set at 38°C (380
tenths of a degree) but it can be changed through the General Settings window:
Switch on the instrument and verify in the Check window the rise of the strip
temperature. With the thermometer, check that the temperature in the strips falls
within the range 36÷38 °C within 30 minutes.
Attention: the temperature in the various positions of the plate is not perfectly
homogeneous when the plate is still. Therefore, before changing the correction value,
insert more strips in different positions of the plate and wait 15 minutes. The average
between the measured temperatures in the various positions must be between 36°
and 37°C.
1. Display
2. Touch screen
display area
3. Display lamp cable
4. Display data cable
Fig. 8-1
The Chorus TRIO is equipped with a liquid crystal display touch screen; this allows you
to interact with the instrument by simply touching the screen with your fingers.
1. Roller
2. Trolley opening
lever
Fig. 8-2
The Chorus TRIO is fitted with a thermal printer that reports the results of the various
instrument tests and the results of previous tests onto paper.
Thermal paper must be used for this type of printer and is supplied in rolls. As can be
seen in the figure above, to insert the paper one must raise lever 2, place the edge of
the paper under roller 1, let it pass under the roller until it comes out above, then lower
the lever.
1. Raise the lever, remove the paper and lower the lever again;
2. Check that the signals below are in the following state:
▪ Printer paper Ok sensor must be switch off (grey color);
▪ Printer carriage open sensor must be switch off (grey olor).
3. Manually introduce the paper until the roller begins to drag it, then check that
the signals below are in the following state:
▪ Printer paper Ok sensor must be switch on (yellow color);
▪ Printer carriage open sensor must be switch off (grey color).
4. Raise the lever and check that the signals bolow are in the following state:
▪ Printer paper Ok sensor must be switch on (yellow color);
▪ Printer carriage open sensor must be switch on (yellow color).
The Carrier+IMX8 is the board that manages: all the other boards (i.e. CPU 2010, Low
Power, Connectors, Power Supply), all the motors (i.e. Wash1-3, Syringe, Tray, Disp1-2,
X Carriage, Filter and TSD), the two thermostats (i.e. instrument and chamber), the
optical channels, the LCD display (i.e. touch screen & display), the printer and the
Internal Bar Code Reader.
The Carrier board is located on the left side, looking at the instrument from the back,
under the display/printer assembly.
Carrier+IMX8 Board
Fig. 9-2
Fig. 9-4
All the functions of the Chorus TRIO refer to them, from the moving of the motors, to
the management of the hydraulics, to the acquiring of analogue signals coming from
the optical sensors, etc.
Fig. 9-5
Test
Point Description
Number
Positive of the tester on TP2 and negative on TP15. Measure of the analogue signal
TP2
of CH2. Signal variations from 0 to 2.5V
Positive of the tester on TP3 and negative on TP15. Measure of the analogue signal
TP3
of CH1. Signal variations from 0 to 2.5V
Positive of the tester on TP4 and negative on TP15. Measure of the analogue signal
TP4
of CH3. Signal variations from 0 to 2.5V
Positive of the tester on TP10 and negative on TP15. Measure of the analogue
TP10
signal of CH6. Signal variations from 0 to 2.5V
Positive of the tester on TP11 and negative on TP15. Measure of the analogue signal
TP11
of CH4. Signal variations from 0 to 2.5V
Positive of the tester on TP13 and negative on TP15. Measure of the analogue
TP13
signal of CH5. Signal variations from 0 to 2.5V
Positive of the tester on TP14 and negative on TP24. Check the presence of -12V.
TP14
Also reported by the start-up of DL6.
TP15 Analogical Ground
Positive of the tester on TP16 and negative on TP24. Check the presence of +12V.
TP16
Also reported by the start-up of DL5.
Positive of the tester on TP22 and negative on TP24. Check the presence of +5V.
TP22
Also reported by the start-up of DL1.
TP24 Digital Ground
Positive of the tester on TP26 and negative on TP15. Measure of the analogue
TP26
signal of the lamp voltage. Signal variations from 0 to 2.5V
Positive of the tester on TP27 and negative on TP15. Measure of the analogue
TP27
signal of the accelerometer. Signal variations from 0 to 2.5V
Positive of the tester on TP26 and negative on TP15. Measure of the analogue
TP28
signal of the lamp voltage. Signal variations from 0 to 2.5V
[1-2] (Default). On/off control of the back light of the display. [2-3], lamp display
JP1
always on.
J1 Programming condition for the microprocessor
J2 Reset condition for the microprocessor
J3 Reset condition for the microprocessor
J4 No Maskable Interrupt of the microprocessor
JS1 Closed (default). Programming signal FPGA IC3, IC6
JS2 Open (default). Programming signal FPGA IC3, IC6
JS3 Open (default). Programming signal FPGA IC3, IC6
JS4 Open (default). Programming signal FPGA IC3, IC6
JS5 Open (default). Programming signal FPGA IC3, IC6
9.3.4 Troubleshooting
The CPU 2010-DRIVERS 2010 boards have been completely realized in Surface Mount
Technology (SMT), and therefore integrated circuits cannot be replaced unless SMT
reworking equipment is made available.
The board is found on the right side, looking at the instrument from the rear, and is
fixed to the frame by 4 automatic hooks located in its 4 corners (see Fig. 9-14).
The power supply voltages are +24V and +5V, in addition to the the ground (GND),
and their presence on the board is indicated through two LEDs (DL1, +5V and DL2,
+24V).
There are also two fuses on the socket (F3 at 5V and F4 at +24V) to protect the board
and the instrument.
Listed below are the actuators and the I.C. that control them.
Anomaly Countermeasures
LED DL1 is off
Replace fuse F3
(the +5V are not present)
LED DL2 is off
Replace fuse F4
(the +24V are not present)
Needle1 or 2 sensors does not work Responsible I.C.: IC1, IC2, IC3, IC4,
(also LED DL3 does not turn on) IC5, IC6, IC7
The level sensors do not work
Responsible I.C.: IC15, IC16, IC18
(various wells, tanks)
Its main function is to transform the line voltage into low voltage direct current
needed to power the instrument.
If it needs to be opened, unscrew the 6 screws (V1-V6) that fasten the sheet metal
cover and then remove it by sliding it towards the rear part of the power supply unit.
The boards contained in the power supply unit will be described in the following
chapters.
9.6 DESCRIPTION
The Rev.3 Power Supply board is responsible for different things inside the power
supply unit; from regulating the LED voltage, to the controlling of the heater, to the
switching of the output voltages from the power supply unit, to powering the cooling
fans.
Fig. 9-22 shows the positioning of the board in the case of the power supply unit. Fig.
9-24 and 78 are shown the connections that go in and out of the power supply unit.
To replace the board, disconnect all the wires and unscrew the 4 screws located in its
corners.
Power supply
The Connectors Board is powered at +5V and at +24V through the cabling 30147041
and the cabling 30147050. It is possible to check the presence of the power supplies
by observing the lighting of the LD5 (+ 5V), LD3 (+ 12V) and LD1 (+ 24V) LEDs.
The board is fastened to the chassis with 5 screws located in the corners of the board.
If it must be replaced, just remove the connections, the fastons to the power
connector and unscrew the 5 screws. To make the extraction of the board easier, the
power socket must also be removed by loosening the two screws holding it to the
rear panel of the power supply unit.
Fig. 9-26 shows the layout of the board inside the chassis.
To CHORUS EXPANDER
Code Description EXTERNAL CONNCECTION
BOARD 30146921
30147060 12Vdc J6
30147070 OFF SIGNAL CONTROL CABLE J5
21890410 RJ45 UTP ETHERNET CABLE, CAT 5e 1mt J9
30147090 BC EXTERNAL SERIAL CABLE FOR CHORUS J11
21890360 USB CABLE 2.0 A-B M/M 1,5 Mt J12
30147082 CPU CHORUS SERIAL & HOST RS232 CABLE CN3
9.7.2 Troubleshooting
The board was realized using SMD technology, therefore the board cannot be
repaired in the field.
Then insert a USB key into the port next to the printer and from the service
application click on Settings and then Parameters.
Fig. 10-4
Check in the bar at the top that the proposed path is correct (run/media/sda1).
Confirm with Ok if you want to save the file on the USB key with the name proposed
by the software or modify it by double clicking on the file name (ex. Chorus Trio
followed by serial number).
2. The fault is related to the hardware component; in this case it is not possible to
connect to the CPU 2010 board and it is also not possible to retrieve the
parameter file in any way;
3. The microcontroller present in the CPU 2010 board has lost the information so
to communicate with the outside; in this case it will be necessary to connect
to the CPU with a special device called Loader and with an external PC on
which the Chorus manager service application is loaded. To do this, download
the dedicated procedure from www.diesse.it website.
Fig. 10-6
Press on the Yes button and wait for the firmware upgrade.
If instead you want to force the update, it is possible to access the service from the
following path:
Select Upload Firmware from File to Chorus and select the file related to the firmware
intended to be uploaded.
Attention: if the CPU board has been replaced, it will be necessary to re-enter the
parameters and the serial number of the instrument. If it was possible to download
the parameters before replacing the CPU board, access the Parameters window and
then press Import. Otherwise, it will be necessary to repeat all the mechanical
calibrations and re-insert them manually.
If the CPU board is new, a manual optical calibration procedure must be performed
the first time the instrument is turned on.
Access the Optical Calibration window:
Fig. 10-9
Double click on Filter 450, select the Filter 650 then repeat the above-mentioned
Attention: the operating system, the Chorus Expander software, the methods file, and
the calibration file are saved on a µSD card. In case of Carrier+IMX8 board
replacement, all information can be transferred simply by transferring the µSD to the
new Carrier board. Anyway, if the instrument allows it, before carrying out any
operation, it is advisable to retrieve the methods file (if it is not possible to download
the updated file from the Diesse website) and the calibrations of the tests processed
up to that moment. Starting from software version 1.1.0, it is possible to backup the
archive of sessions, calibrations and methods using a single command.
Fig. 10-10
Then click on Export Methods and the following window will open:
Fig. 10-12
At this point, click on Export and the following window will open:
Fig. 10-13
• Session archive
• Calibration archive
• Method files
• Instrument configuration
Check in the top bar that the proposed path is correct (run/media/sda1). Confirm with
OK if you want to save the file in the USB key with the name proposed by the software,
or modify it by double clicking on the file name (Fig. 10-15).
To restore the information previously saved on the new board, insert the USB key into
the port near the printer, then open the service application again, click on Backup
and then on Restore (Fig. 10-14).
Check in the bar at the top that the proposed path is correct (run/media/sda1). Select
the backup file previously saved on the USB key (file with .cbk extension) and confirm
with OK. Follow the directions on the screen to complete the process.
If the software is not updated to the latest version, go to Diesse website and download
the latest version on a USB key.
Then insert the USB key into the port near the printer and open the Software Update
window:
Verify on the top bar that the proposed path is correct (run/media/sda1).
Select the file containing the update (characterized by the .cpk extension) then click
on OK. If the software update also requires a firmware update, the latter will be
requested automatically.
Once the software is updated, check that the configuration is correct:
Fig. 10-18
Enable the Autosampler (if present) and the RDP connection (if you want to connect
the instrument to the network).
10.2 AUTOSAMPLER
Starting from version 1.1.0 of the software it is possible to execute some service
operations to test the correct functioning of the Chorus + Autosampler system,
without necessarily perform an analysis cycle. To do this, open the service application
and go to Sampler:
Fig. 10-21
Fig. 10-22
10.3.1 Connect the reader to the instrument through the RS232 cable
Insert the RJ connector of the RS232 cable (CAB350) into the socket located at the
base of the reader and push until the tab clicks into place; connect the other end of
the cable to the serial port on the Chorus TRIO labelled Barcode (Fig. 10-25).
Warning: since the serial port also supplies the power, it is best to connect the reader with the
instrument off.
Fig. 10-26
Normally the readers arrive already programmed to work in serial mode. If problems
occur, it is possible to try to reprogram the reader, reading on the Quick guide the
following barcodes sequence:
Start of configuration / RS-232C / End of Configuration
Fig. 10-27
Start of Configuration
PC/AT
USB
WAND EMULATION
SHOW VERSION
End of Configuration
Program Reset
PC/AT End
Program Reset
RS-232C End
End
Any anomalies or errors will appear in the error window and, depending on the type,
may or may not be removed by the user (see Operating Manual).
The washing procedure must be launched at the end of the work day in order to
prevent incrustations from forming on the needles, tubes, etc. (see Operating
Manual).
▪ Check the number of tests performed since the last maintenance intervention;
▪ Start the sanitization procedure;
▪ Disconnect the instrument from the network;
▪ Wear protective glasses and gloves
The maintenance for each device will be described in the following paragraphs.
Fig. 10-28
Dry and reassemble all the parts of the well, making sure that the teflon gaskets have
been replaced. Before remounting it on the platform, clean it.
Fig. 10-29
10.4.6 Washers 21 – 25 – 28
Dirt mainly accumulates on the needles and at the base of the washers
(see Fig. 10-33)
▪ Remove the three washers from the plate;
▪ Remove the needles from the washers;
▪ Check that the needles are in good condition and that the lines are perfectly
open and free of incrustations;
▪ Remove the guide from the bracket for all washers;
▪ Clean the bracket and the guide and any incrustations with WD40 and
lubricate the guide with silicone oil;
▪ Check the needle springs and replace them if rusted or after 5000 tests;
▪ Check the spherical slide couplings of the needles. Replace if worn or
encrusted.
After the platform has been cleaned, the units can be remounted.
In the following figures are depicted the placements inside the instrument for all the
components.
Fig. 10-35
To remove the emitter from the machine, just loosen the two screws fastening it to
the upper platform. When lifting it from the platform, we can now see from below
the lenses that are going to be taken out.
Using a cloth, delicately clean the lenses, being careful not to scratch them.
Warning: when the lenses are remounted, their convexity direction must be
respected.
Fig. 10-36
As mentioned previously, the receivers are mounted below the lower platform. The
figure below shows their layout
Fig. 10-37
Receiver
Fig. 10-38
Fig. 10-39
10.4.10 Carousel
The cleaning of the carousel is done in two phases:
Upper plate of the tray, accessing the slit on the platform in correspondence with the
x-axis shifting (see Fig. 10-40) and clean, sector by sector, both the upper part and
inside the strip positioning chamber.
Be sure not to use abrasive liquids and/or substances that can remove the numbering printed
on the tray.
Fig. 10-40
Insert a strip in each of the thirty strip positions and check that the spring holds it in
place. Replace it if it does not hold.
To remove the springs, take off the sealing ring to which the fastening screws access
from the special opening in the platform.
Remount the head, being careful of its direction of flow in order to avoid switching
the aspiration/expulsion flow (see Fig. 10-42)
Fig. 10-42
Fig. 10-43
Fig. 10-44
Warning: when handling the syringe, do not push the piston so that it exits the side of the ring
nut otherwise the tip will become damaged when it is reinserted.
▪ Check the conditions of the syringe O-rings and replace them if necessary;
▪ Screw the syringe onto the Plexiglas support and fasten the piece;
▪ Check the hydraulic fittings by disassembling them from the Plexiglas
manifold, removing the silicone;
▪ Put new silicone on the fittings and reposition them on the Plexiglas manifold;
▪ Clean the syringe feed screw with WD40 and lubricate with silicone oil;
▪ Check the tightness of the screws, especially those for the joint;
▪ Reposition the syringe on the cross beam (Fig. 10-45) and remount everything
on the instrument’s frame.
10.4.16 Rotation
The rotation requires no special maintenance operations besides cleaning of the dust
and checking the voltage and the condition of the belt.
In case of rust, clean with WD40 and disinfecting/cleaning liquid and/or replace the
probe.
Check the conductivity of the sensors.
10.4.20 Tubing
This check pertains to all the tubes in the instrument. Replace any tube that has
mould or other residues.