Photometric Report 04/03/2020 10:37:11

i to
File Name: C:\UVProbe-Data\Data\NO2.pho

[Wavelengths] Standard Curve


Wavelength Name: WL543.0

Ed
0.450
Wavelength: 543.00 nm

[Calibration Curve]
Column for Cal. Curve: WL543.0
Cal. Curve Type: Multi Point
Cal. Curve Unit: mg/l
0.300
Selected Wavelength: WL543.0
Calibration Equation: Abs = K1*(Conc) + K0
Zero Interception: Not Selected

[Measurement Parameters(Standard)]

DF
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled

[Measurement Parameters(Sample)] 0.200


Abs.
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled

rP
[Equations]

[Pass Fail]

[Method Summary] 0.100


Title:
Date/Time: 13/08/2018 10:09:57
Comments:
Sample Preparations:
te
[Instrument Properties]
Instrument Type: UV-1900 Series
Measuring Mode: Absorbance
Slit Width: 1.0 nm
Light Source Change Wavelength: 340.0 nm -0.000
S/R Exchange: Normal 0.00 0.050 0.100 0.150 0.200
as
Conc. (mg/l)
[Attachment Properties] r2 = 0.99911
......
Standard Table
Sample ID Type Ex Conc WL543.0 Wgt.Factor Comments
1 STD_1 Standard 0.010 0.033 1.000
2 STD_2 Standard 0.050 0.064 1.000
M

3 STD_3 Standard 0.100 0.127 1.000


4 STD_4 Standard 0.200 0.336 1.000
5
in

Sample Table
Sample ID Type Ex Conc WL543.0 Comments
ed

1 Sungai_A Unknown 0.0026 0.056


2 Sungai_B Unknown 0.0021 0.046
3 Sungai_C Unknown 0.0020 0.043
4 Sungai_D Unknown 0.0027 0.057
5 Sungai_E Unknown 0.0035 0.074
6 Sungai_F Unknown 0.0034 0.073
at

7 Sungai_G Unknown 0.0102 0.208


re

Page 1 / 1