Service Report Instalation X-Ray Stationary Banda Aceh

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

PT.

MITRA KARYA ALKESINDO


Medical – Hospital Equipment, Supplier Barang dan Jasa Perumahan
Bulak Indah Blok A No 35, Kalibaru, Cilodong - Depok Hp : 081546214250,
Email : mitrakaryaalkesindo@gmail.com

Service Request Type: Service Request Number:


Service Report Corrective Repair 2022-06-09/08

Customer Details

Customer Name: Location: Contact Banda Aceh-Indonesia


PO #: Mr.Dedi Fahrizal Phone: Email -
Contact Nome: Address:
Initial Problem Description:

Equipment Details

System ID: Initial Equipment Status: Install Unit


Serial Number: 101090 Entitlement: -
Meditronics 300Ma
Product Description: Stationary

Activity Description
Activity #: Activity Close Dote:
Activity Type: Field Support Engineer Name: Sukrisdianto
Tube Slice Reading: 300MA
Problem Description: Install Unit

Couse/ Diagnosis: Jun 06-2022 :


- Installation and functional check x-ray stationary 300Ma

Resolution/ Service Performed:


Jun 06-2022 :
- Survie and chek x-ray room OK
- Prepare tool set and Material OK
- Marking and install all rell for stand tube OK
- Install stand tube and tube assembly OK
- Connec installation control and power cable from HTT tank OK
- Check and inspect suplay electricity from panel lab Room OK
- Disposisi control and connec power electricity 3phase from Lab room OK
- All check connection after installation unit OK
- All check mechanical after installation X-ray unit OK
- All check suplay panel electricity from lab room R.413V,S.423 and T.432 ( Over voltage )
- Trun on x-ray unit and check LCD parameter,roary knop,kv control,Mas control OK
- Test exposure set KV 70 ,Mas 100 from media ip screen results OK
- Trun off unit console,and general cleaning x-ray unit OK
- Trun on unit and all functional check OK
- X-ray stationary ready to use

Page No.1 of 1
Recommendotion :
 Provide stabilizier 20A 3P for supply x-ray stationary

 Repaired supply and installation electricity from

Lab room

 Perform calibration according to schedule

 Maintenance according to schedule

Page No.2 of 2
Service Request Type: Corrective Repair
Service Report Request Number: 22069-10

Date Time Hours Labour Type


7 Jun , 2022 10:00-22:23 Onsite and
install
8 jun ,2022 10:00-00:30 Install
9 Jun ,2022 13:00-17:40 Finishi

Ports Used Details

Port Number 0tY Description Asset Serial No

Expense Tracker

Expense Dote Type 0tY Comments

Engineer Signoture: Mr.Dedi


09-Jun-2022 17:40 Fahrizal
Customer Signoture:
SUKRISDIANTO
Page No.3 of 3