MAC 600 Resting ECG Analysis System: Quick Reference Guide

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 14

MAC™600 Resting ECG

Analysis System
Quick Reference Guide

Education Services
Clinical Development

QRG MAC 600 R3 062210.indd 1 6/24/10 10:17 AM


QRG MAC 600 R3 062210.indd 2 6/24/10 10:17 AM
Notice
The materials contained in this document are intended for
educational purposes only. This document does not establish
specifications, operating procedures or maintenance methods for any
of the products referenced. Always refer to the official written materials
(labeling) provided with the product for specifications, operating
procedures and maintenance requirements.
Proprietary Training Material, property of GE Healthcare. Use of these
materials is limited to agents and employees of GE Healthcare or other
parties expressly licensed by GE. Unlicensed use is strictly prohibited.

Contents

Skin Preparation and Lead Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1


Acquire an ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Print Continuous Rhythm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
File Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

QRG MAC 600 R3 062210.indd 3 6/24/10 10:17 AM


Skin Preparation and Lead Placement

AHA IEC
V2 C2
AHA IEC AHA IEC
V1 C1 V3 C3
AHA IEC
V4 C4
AHA IEC AHA IEC
Alternative RA R LA L Alternative

AHA IEC
V6 C6

AHA IEC
V5 C5

AHA IEC AHA IEC


RA R LA L

AHA IEC AHA IEC


Alternative RL N LL F Alternative

AHA IEC AHA IEC


RL N LL F

QRG MAC 600 R3 062210.indd 4 6/24/10 10:17 AM


Skin Preparation and Lead Placement

Skin Preparation and Lead Placement


1. Shave any hair from each electrode site and degrease each electrode site with alcohol.
2. Dry skin completely.
3. Apply the leads on the prepared area.
4. Verify the leads are connected and working properly.

Table 1 AHA and IEC Lead Placement

AHA lead IEC lead Electrode Placement


V1 red C1 red Fourth intercostal space at the right sternal border.
V2 yellow C2 yellow Fourth intercostal space at the left sternal border.
V3 green C3 green Midway between C2/V2 and C4/V4.
V4 blue C4 brown Mid-clavicular line in the fifth intercostal space.
V5 orange C5 black Anterior auxiliary line on the same horizontal level as C4/V4.
V6 purple C6 purple Mid-auxiliary line on the same horizontal level as C4/V4 and C5/V5.
LA black L yellow Above left wrist (alternate placement: left deltoid).
LL red F green Above left ankle (alternate placement: upper leg close to torso).
RL green N black Above right ankle (alternate placement: upper leg close to torso).

RA white R red Above right wrist (alternate placement: right deltoid).

QRG MAC 600 R3 062210.indd 5 6/24/10 10:17 AM


Acquire an ECG

Press ‘ECG’ (Green) key to record an ECG.


- + 3a
0:29:40 77 BPM
5-Aug-2009

I aVR
Leads
V1 V4

II aVL V2 V5

III aVF ECG


V3 V6
Patient Data 25 mm/s 10 mm/mV More

4 F1 F2 F3 F4 6
1 2 3 3b
1 4 5 6 7
MAC 600
7 8 9

esc 0

QRG MAC 600 R3 062210.indd 6 6/24/10 10:17 AM


Acquire an ECG

Acquire an ECG
1 Power the system on. 6 Use the F keys to make any
Enter Patient Data.
0:29:40
77 BPM
- +
necessary adjustments to:
2. Prepare the patient. Refer to MAC600
5-Aug-2009

section Skin Preparation and I


ID number
V1 V4
• Speed (F2)
Date
aVRof birth
Lead Placement for details. Gender
II aVL
Return
V2 V5 • Gain (F3)
3. Verify waveform quality by: III aVF V3 V6
• More (F4) > Filter (F1) to
3a
• Checking Hook-up
adjust Waveform Filter.
Advisor. Green (always Color Display is a purchasable option. Press More (F4) to return
right most circle)
Note! Refer to the ECG Screen.
indicates a good quality
waveform. Operator’s Manual for 7 Press the ECG key on the
instruction on how to use keypad to record an ECG.
• Press the Leads key to
3b Secondary ID, age and
scroll through the leads. alfa-numeric input feature.
4 Press F1 to open the Patient 5. Select Return, and press the
Data window. Enter the
Enter Patientcenter
Data. of the arrow pad to
patient ID, DOB and gender close the Patient
0:29:40Data window.
- +
77 BPM
5-Aug-2009
using the numbered MAC600
keyboard and arrow pad. ID number
I Date V1 V4
• Press F1 to insert a dash aVRof birth

while entering the patientII Gender


aVL V2 V5
ID if needed. Return
III aVF V3 V6
• Press F2 to insert a +
symbol while entering the
patient ID if needed.

QRG MAC 600 R3 062210.indd 7 6/24/10 10:17 AM


5-Aug-2009
77 BPM

I V1 V4
aVR

II aVL V2 V5
Acquire an ECG

Continue Rhythm Store Cancel

10 F1 F2 F3 F4

1a 2 c 3 b 9

4
8. Press Analysis (F2) to view 5
9 6
If after reviewing the ECG and 10 Other options from the
the Measurement and Analysis if you wish to stop Analysis window include:
Interpretation information. and acquire another ECG,
7 8 9 (F4) and press a
Press Continue (F1) to
Once you have pressed press Cancel
continue with the
Analysis you see the F keys the ECG key on the keypad to acquired 12 lead ECG.
as displayed in the above record another ECG.
esc 0 6 Press Store (F3) to store
b
illustration. Otherwise continue with
step 10. the ECG to the SD Card.
Note! Analysis and
Measurement are Note! Storing to SD card is
purchasable options. a purchasable option.

c Press Rhythm (F2) to


view the waveform.

QRG MAC 600 R3 062210.indd 8 6/24/10 10:17 AM


0000003 29-Jul-09 03:17:10 No
0000004 29-Jul-09 03:15:36 No
0000005 29-Jul-09 03:16:14 No

Total files: 20 Selected files: 1


Select Select All Report Setup More

Print Continous Rhythm


F1 F2 F3 F4

1 2 3 1

4 5 6

7 8 9 2

esc 0 3

Print Continous Rhythm


1 Press the Leads key to select Note! If your display is 3 Press the Stop Writer key
the desired 3-lead group for showing the 12-lead or on the keypad to stop the
rhythm. 6-lead presentation, printing of the continuous
pressing the Rhythm key rhythm.
2 Press the Rhythm key on the
will result in the display of
Keypad to start the printing
3 leads from the group
of continuous rhythm.
configured as the
autorhythm and will begin
rhythm printing with those
3 leads.

QRG MAC 600 R3 062210.indd 9 6/24/10 10:17 AM


File Manaager (Purchasable option)

Directory of files in SD card

File Manager
- +

Directory of files in SD card


ID number Date/Time Sent
0000001 01- Aug-09 23:35:48 No
0000002 30-Jul-09 23:35:48 No
0000003 29-Jul-09 03:17:10 No
0000004 29-Jul-09 03:15:36 No
0000005 29-Jul-09 03:16:14 No

Total files: 20 Selected files: 1


Select Select All Report Setup More

F1 F2 F3 F4

41 2 5 3 3
File Manager (Option) 4 5 6 Enter key
1. From the startup screen press f. Press Return (F2) to get
More (F4). 7 8 9 back to the main File
Manager Screen.
2. Press File Manager (F3).
esc 0
4 Press Select (F1) to select
3 Press Report Setup (F3) to
Arrow key individual reports. Use the
select desired format for your
arrow pad to navigate
printed reports. c. Enter 1 for each report
through the list and press the
you wish printed. You may
a. Press Report Format (F1) Enter key of the arrow pad to
enter a higher number if
to open the reports highlight the desired report.
you wish more copies.
selection screen. Use the same procedure to
d. Select Return to save and highlight multiple reports
b. Use the arrow pad to go
exit screen. if desired.
to the desired report.
e. Press More (F4). 5 If you wish to highlight all
reports press Select All (F2).

QRG MAC 600 R3 062210.indd 10 6/24/10 10:17 AM


File Manager
Directory of files in SD card
- +
File Manager
Directory of files in SD card
ID number Date/Time Sent
Selected 0000001 01- Aug-09 23:35:48 No
report 0000002 30-Jul-09 23:35:48 No
0000003 29-Jul-09 03:17:10 No
0000004 29-Jul-09 03:15:36 No
0000005 29-Jul-09 03:16:14 No
Total files: 20 Selected files: 1
Print Display Transmit More

F1 F2 F3 F4

1
1 2 2 3 3 4
4 5 6
Note! Your F key order may 2 Display (F2): allows you to 4 Select More (F4) then Return
vary slightly from what is 7 8 the Medians
view 9 (F1), (F2) to get back to File
described based on your Analysis (F2) or Print (F3) of Manager main screen.
purchased options. esc the selected record(s).
0 6
Once you have selected a 3 Transmit (F3): allows you to
report(s) you have the following transmit to a MUSE system or
options: XML output via Serial Line.
1 Print (F1): prints the ECG of Note! Serial Transmission
the selected report(s) is a purchasable option.

QRG MAC 600 R3 062210.indd 11 6/24/10 10:17 AM


File Manager

Directory of files in SD card

File Manager
- +

Directory of files in SD card


ID number Date/Time Sent
0000001 01- Aug-09 23:35:48 No
0000002 30-Jul-09 23:35:48 No
0000003 29-Jul-09 03:17:10 No
0000004 29-Jul-09 03:15:36 No
0000005 29-Jul-09 03:16:14 No
Total files: 20 Selected files: 1
Delete Save XML Save PDF More

F1 F2 F3 F4

1
a 2 b 3 c d
5. More (F4): Gives more 4 c 5
Save PDF6(F3): saves 6. From the main File Manager
selections including: record to SD card in PDF screen press More (F4) then
format press Resting ECG (F3) to
a Delete (F1): press to 7 8 9
Note! Save to PDF is return to the Resting ECG
delete record. User must
purchasable option. screen.
enter password to esc 0 6
delete record(s) from
system d More (F4): returns previous
b Save XML (F2): saves screen. Press Return (F1)
record to SD card in XML to return to File Manager
format main screen.

Note! Save to XML


is purchasable
option.

QRG MAC 600 R3 062210.indd 12 6/24/10 10:17 AM


QRG MAC 600 R3 062210.indd 13 6/24/10 10:17 AM
© 2010 General Electric Company – All rights reserved.

General Electric Company reserves the right to make changes in specifications and features shown herein, or
discontinue the product described at any time without notice or obligation.

GE, GE Monogram, Marquette, 12SL, MAC and MUSE are trademarks of General Electric Company.

GE Medical Systems Information Technology Inc., a General Electric company, doing business as GE Healthcare.

Notice: The materials contained in this document are intended for educational purposes only. This document does
not establish specifications, operating procedures or maintenance methods for any of the products referenced.
Always refer to the official written materials (labeling) provided with the product for specifications, operating
procedures and maintenance requirements.

Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to agents and employees of
GE Healthcare or other parties expressly licensed by GE Healthcare. Unlicensed use is strictly prohibited.