Download as pdf or txt
Download as pdf or txt
You are on page 1of 40

Microbiological Incubator

AL01-06 – 53 liters
AL01-07 – 115 liters

Service Manual
Issue AL01-06+07 SM 02-10_en

valid from device serial number 10-02000


EC - Declaration of Conformity
EG – Konformitätserklärung

Manufacturer / Advantage Lab GmbH


Hersteller:
Address / Jachthoornlaan 8
Anschrift: B-2970 Schilde, Belgium
Product / Microbiological incubator
Erzeugnis: Mikrobiologischer Brutschrank
Type / AL01-06- 53 liters
Typenbezeichnung: AL01-07- 115 liters

This unit corresponds to the following product specification:


Dieses Erzeugnis entspricht den folgenden Produktspezifikationen:
Safety / Sicherheit:
IEC/CEI 61010-1:2001 and IEC/CEI 61010-2-010:2003; DIN 12880:2007
EMC / EMV:
IEC/CEI 61326:1997 + A1:1998 + A2:2000

This unit corresponds to the demands of the low voltage directive 2006/95/EC, and to the EMC
directive 89/336/EEC, amended 93/68/EEC, and, corresponding to this, it bears the CE-mark.
Dieses Gerät entspricht den Anforderungen der Niederspannungsrichtlinie 2006/95/EG und der EMV-
Richtlinie 89/336/EWG, Änderung 93/68/EWG, und trägt entsprechend die CE-Kennzeichnung.

Schilde, September 13th, 2007


Advantage Lab GmbH

Paul Gréant
Managing Director / Geschäftsführer
Content

Content
1 Introduction .............................................................................................................................4
1.1 General guidelines ....................................................................................................................4
1.2 Documentation reference..........................................................................................................4
1.3 Target group and minimum qualification requirements .............................................................4
1.4 Safety instructions structure......................................................................................................4

2 General safety instructions....................................................................................................6


2.1 Intended use .............................................................................................................................6
2.2 Dangers.....................................................................................................................................6

3 Troubleshooting ......................................................................................................................8

4 Repair .....................................................................................................................................14
4.1 Replacing a unit foot ...............................................................................................................15
4.2 Replacing the outer door gasket .............................................................................................16
4.3 Replacing the glass door gasket.............................................................................................17
4.4 Replacing the glass door.........................................................................................................18
4.5 Replacing the miniature fuse ..................................................................................................19
4.6 Replacing the controller housing with controller .....................................................................20
4.7 Replacing the main switch ......................................................................................................21
4.8 Removing the rear panel.........................................................................................................22
4.9 Replacing the IEC connector plug ..........................................................................................23
4.10 Replacing the solid state relay ................................................................................................24
4.11 Replacing the temperature fuse..............................................................................................25
4.12 Replacing the temperature safety device class 3.1 ................................................................26
4.13 Checking and replacing the Pt100 temperature sensor..........................................................28

5 Spare parts.............................................................................................................................31

6 Wiring diagrams ....................................................................................................................33

Advantage-Lab, AL01-02+03 SM 02-10_en page 3/34


AL01-06+07 Microbiological Incubator – Service Manual

1 Introduction
1.1 General guidelines
This service manual describes the trouble shooting and the repair of the AL01-06 and AL01-07
incubators. It contains the service information required for the AL01-06/07.
This service manual is valid for the following incubator models:
• AL01-06 (53 liters)
• AL01-07 (115 liters)
Before starting service work pay attention to the change index of the AL01-06/07 service manual und
to service notes available from your distributor account at www.advantage-lab.com.

1.2 Documentation reference


All information about initial operation, normal operation, cleaning, disinfecting, adjusting, alarm and
error messages of the AL01-06/07 can be found in the user manual.
Manual updates and service notes are available from your distributor account at
www.advantage-lab.com.

1.3 Target group and minimum qualification requirements


This service manual is aimed at the service personnel of the unit dealer.
The service personnel must be familiar with unit operating procedures to be able to execute the
service work.
Personnel are sufficiently qualified when they meet the following requirements:
• Electrotechnical training
• Knowledge of the most recent user and service manuals

1.4 Safety instructions structure


These operating instructions employ the terms and symbols below to describe dangerous situations, in
line with the harmonization of ISO 3864-2 and ANSI Z535.6.

1.4.1 Safety instructions hierarchy

WARNING LEVEL
Safety symbol Type of danger
Prohibitory symbol [ Action prohibited
Imperative symbol ¾ Safety recommendation to be followed
Introduction

1.4.2 Warning levels

Dangers are indicated, according to the severity and likelihood of their consequences, by a word to
attract attention, with the associated warning color and any appropriate safety symbol.

DANGER
Indicates an imminently hazardous situation that, if not avoided, will result in death or
serious (irreversible) injury.

WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death
or serious (irreversible) injury

CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in
moderate or minor (reversible) injury.

CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in damage of the product
and/or its functions or of a property in its ambiance.

1.4.3 Safety symbols

Danger of injury.
Follow all the directions indicated by the safety symbol in order to avoid injury or death.

Danger of electric shock

Danger of infection

Danger of poisoning

Danger of injury through an object tipping up

Advantage-Lab, AL01-02+03 SM 02-10_en page 5/34


AL01-06+07 Microbiological Incubator – Service Manual

Danger of injury through lifting heavy objects

1.4.4 Imperative symbols

Pull out the power plug

1.4.5 Information symbols

Important information

2 General safety instructions


2.1 Intended use
The AL01-06/07 microbiological incubator is designed to cultivate microorganisms at defined
temperatures.
It can be used e.g. for determining the CFU through plate count methods or detecting of pathogenic
germs in clinical samples or product samples.
It is intended for use in laboratories in these sectors: life sciences, clinical medicine, veterinary
medicine, the pharmaceuticals industry, the food industry, toxicology, and basic research into biology
and preclinical medicine.
Any other use is not intended, and can cause damage and be harmful to health.
The AL01-06/07 should not, in particular, be used to prepare foods intended for consumption.

2.2 Dangers
If the AL01-06/07 is installed, operated, cleaned, disinfected, adjusted or set up incorrectly, there is a
risk of malfunction which could cause harm to human beings and material damage to the equipment
and samples.
Therefore the AL01-06/07 should only be installed, operated, cleaned, disinfected, adjusted and set up
by qualified persons.
• Persons qualified to install, operate, clean, disinfect, and adjust the AL01-06/07 include anyone
who has acquainted themselves with the contents of the user manual.
• Persons qualified to repair the AL01-06/07 only include service engineers belonging to the
equipment supplier who have undergone appropriate electrical training and who have acquainted
themselves with the contents of the AL01-06/07 service and user manual.
General Safety Instructions

DANGER
Danger of electric shock.
Danger to life.
¾ All work on the device's electrical components should only be carried out by
qualified electricians.
¾ The unit must be live to perform some of the diagnostic tasks described in the
chapter “Troubleshooting”.
Take all precautionary measures not to get in contact with live components of the
unit.
¾ The unit must be switched off prior to performing most of the service tasks
described in the chapter “Repair”.
¾ Disconnect the unit from the facility-side mains supply to make sure that it can not
be switched on again by mistake.

WARNING
Contamination of the unit with toxic, infectious or radioactive substances.
Danger of poisoning.
Danger of infection.
[ Never service a unit with sticking toxic or radioactive substances or sources of
infection.
¾ Prior to service, let the unit free from sticking toxic or radioactive substances by the
customer and disinfect it.
¾ Dispose of a unit which cannot be safely freed from all toxic, toxic, infectious or
radioactive substances as special waste according to national law.

CAUTION
Danger of injury.
Through the device slipping or tipping over.
Damage to the device.
¾ The device should be lifted by 2 people.
¾ The device should only be transported in its original packaging.
¾ The device should be secured with straps when being transported.
¾ The device should only be raised along with the supporting pallet from beneath
using mechanical tools (e.g. fork-lift truck).
[ The device should not be lifted directly from below using mechanical tools (e.g.
fork-lift truck).
[ The device should not be lifted or moved by the door.

Advantage-Lab, AL01-02+03 SM 02-10_en page 7/34


AL01-06+07 Microbiological Incubator – Service Manual

3 Troubleshooting

WARNING
Contamination of the unit with toxic, infectious or radioactive substances.
Danger of poisoning.
Danger of infection.
[ Never service a unit with sticking toxic or radioactive substances or sources of
infection.
¾ Prior to service, free the unit from sticking toxic or radioactive substances by the
customer and disinfect it.
¾ Dispose of a unit which cannot be safely freed from all toxic, infectious or
radioactive substances as special waste according to national law.

DANGER
Danger of electric shock.
Danger to life.
¾ The unit must be live to perform some of the diagnostic tasks described below.
¾ Take all precautionary measures not to get in contact with live components of the
unit.

Alarm and error messages as well as basic measures the operator himself may take to
clear them are described in the chapter “Troubleshooting” of the AL01-06/07 user
manual.
Please make sure these measures have been tried before looking further for any other
possible error causes.

This troubleshooting table below consists of four columns as follows:


• Error pattern 1:
This column describes the main error symptom. The housing of the unit may have to be opened
up or measuring equipment used in order to identify the error.
• Error pattern 2:
This column describes another error symptom or the operating condition of the unit, which will
show up together with error pattern 1.
Please note that the same error pattern 1 may occur along with different error patterns 2.
• Possible error causes / diagnosis:
This column describes possible errors and their diagnosis, which may be the cause for the
simultaneous occurrence of error patterns 1 and 2.
• Remedial measures:
This column lists remedial measures to clear specific error causes.
The sequence of the columns describing potential causes for error, and the remedial actions to clear
the described error patterns, is mainly intended to facilitate basic error diagnosis and clearance. So it
is recommended to work your way from top to bottom through the relevant columns of an error pattern.
Notations between brackets and in italics, e.g. (-1S1), behind a component in the troubleshooting
table, refer to the respective equipment codes for this component, as specified in the wiring diagram
(see chap. 6, page 31), and in labels on some parts inside the unit.
Troubleshooting

Table 1: Troubleshooting

Error pattern 1: Error pattern 2: possible error causes / diagnosis: Remedial action
The housing is damaged or Diagnosis: Straighten the housing if possible and check
deformed by external 1. Check whether the unit is still electrically both electrical safety and fitness for purpose.
mechanical impact (e.g. fall) safe.
2. Check whether the unit is still fit for normal
intended use.

No indications on the Power plug isn’t inserted Connect the power plug to the power supply.
controller displays (-2N1)
Main switch (-1S1) is switched to 0 Switch the main switch to I
Power cord is damaged Replace the power cord
Miniature fuse (-X1) is blown Replace the miniature fuse (chap. 4.5,
Diagnosis: page 19)
• If the mains voltage is not equal to the
Look for possible causes for the fuse blowing
nominal voltage displayed on the type
• Compare the mains voltage to the data plate, the unit cannot be operated with
indicated on the type plate. that mains supply.
• Check visually for mechanical damage to • If contaminated with conductive dust,
the electrical cables and burn marks on the clean the unit carefully
electrical cables or on electrical
• If the cables are mechanically damaged,
components
replace or insulate the cable in a
• Check the mains supply professional manner
• Check the heating (see page 11) • If cable or components are burned,:
replace cable or component if possible
(chap. 4, page 14)
• If the mains supply is defective,
repair the mains supply
• If the heating is defective,
dispose of your unit (see page 11)
The pins of the main switch (-1S1) in the Connect the pins of the main switch to its
controller housing are not connected to its cables (chap. 4.7, page 21).
cables.

Advantage-Lab, AL01-06+07 SM 02-10_en page 9/34


AL01-06+07 Microbiological Incubator – Service manual

Error pattern 1: Error pattern 2: possible error causes / diagnosis: Remedial action
No indications on the Main switch (-1S1) is defective Replace the main switch (chap. 4.7,
controller displays (-2N2) Diagnosis: page 21).
1. Disconnect the cables from the pins of the
main switch (chap. 4.7, page 21).
2. Switch the main switch to I.
3. Test continuity of the main switch using a
multimeter.
If continuity test fails, main switch (-1S1) is
defective.
3-pole plug (-X10) (B, Figure 6, page 20) is not Connect the plug (-X10) (chap. 4.6,
connected to RP2 controller board (-2N1) page 20).
Temperature fuse (-1F2) is blown due to an • Replace the temperature fuse (chap.
error on the controller board and at the 4.11, page 25)
temperature safety device class 3.1 • Replace the controller housing with
Diagnosis: controller (chap. 4.6, page 20).
1. Disconnect the cables from the pins of the • Replace the temperature safety device
temperature fuse. (chap. 4.11, page 25) (chap. 4.12, page 26).
2. Test continuity between these pins using a
multimeter.
If continuity test fails, temperature fuse
(-1F2) is defective.
Controller board (-2N1) is defective Replace the controller housing with controller
(chap. 4.6, page 20).

Advantage-Lab, AL01-06+07 SM 02-10_en page 10/34


Troubleshooting

Error pattern 1: Error pattern 2: possible error causes / diagnosis: Remedial action
Chamber does not heat up • Controller display High voltage part of solid state relay (-2K1) is Replace solid state relay (chap. 4.10,
shows current interior defective page 23).
temperature of Diagnosis:
chamber
1. Disconnect cables from connections T1 and
• LED is lit L1 of solid state relay (-2K1) (chap. 4.10,
• LED of solid state relay page 23).
is lit 2. Switch chamber on and set SP Temperature
set point to maximum value.
3. Test continuity between connections T1 and
L1 using a multimeter.
If continuity test fails, solid state relay
(-2K1) is defective.

Heating is defective Defective heating cannot be repaired.


Diagnosis: Dispose of the chamber (refer to the user
manual).
1. Disconnect first cable for power supply of
the heating (-2E1) from solid state relay
(-2K1) (chap. 4.10, page 23; chap. 6
page 33).
2. Disconnect second cable for power supply
of the heating (-2E1) from temperature
safety device (-1F3) (chap. 4.12, page 26).
3. Test resistance of the chamber heating
between both cables using a multimeter.
If measured resistance is not 265 Ω ±10 %
(AL 01-06) resp. 212 Ω ±10 % (AL 01-07)
resp, heating (-2E1) is defective.

Advantage-Lab, AL01-06+07 SM 02-10_en page 11/34


AL01-06+07 Microbiological Incubator – Service manual

Error pattern 1: Error pattern 2: possible error causes / diagnosis: Remedial action
Chamber does not heat up • Controller display 4-pole plug is not connected to connection Connect the plug (-X11).
shows current interior (-X11) (C, Figure 6, page 20) of controller (chap. 4.6, page 20).
temperature of board (-2N1)
chamber
• LED is lit
• LED of solid state relay
is not lit
Controller board (-2N1) is defective Replace controller housing with controller
Diagnosis: (chap. 4.6, page 20).

1. Disconnect 4-pole plug from connection


(-X11) of controller board (-2N1).
2. Switch chamber on and set SP Temperature
set point to maximum value.
3. Measure voltage between pin 3- and 4+ of
connection (-X11) using a multimeter.
If measured voltage is not approx. 5 V DC,
controller board (-2N1) is defective.
LED of solid state relay is Temperature safety device class 3.1 (-1F3) is Replace temperature safety device (chap.
lit defective 4.12, page 26).
Diagnosis:
1. Demount temperature safety device (chap.
4.12, page 26).
2. Disconnect cables from pins 1, 2 and 4.
3. Turn shaft of the thermostat clockwise to
the limit stop.
4. Test continuity between pins 1 and 2 using
a multimeter.
If continuity test fails, temperature safety
device (-1F3) is defective.

Low voltage part of solid state relay (-2K1) is Replace solid state relay (chap. 4.10,
defective page 23).

Advantage-Lab, AL01-06+07 SM 02-10_en page 12/34


Troubleshooting

Error pattern 1: Error pattern 2: possible error causes / diagnosis: Remedial action
SP Temperature set point Interior temperature Controller board (-2N1) is defective Replace controller housing with controller
being set on the controller is reaches the set point of (chap. 4.6, page 20).
exceeded permanently the temperature safety
device permanently and
the red alarm lamp lights
up
Current interior temperature as Current temperature on Temperature control is not, or not properly, Calibrate temperature control (refer to the
measured with a calibrated the controller display is calibrated user manual)
external measuring instrument equal to the Temperature
in a stable operating phase set point.
deviates by more than ±1 K
from the temperature shown
by the reference measuring
instrument ± the measurement
uncertainty from the
temperature set point.
Alarm “999” Wires of Pt100 temperature sensor (-2B1) are Connect the measuring sensor properly to
Rupture of temperature sensor not or incorrectly connected to clamps on the the clamps (chap. 4.6, page 20).
or exceeding of measuring controller board (-2N1)
range white wire – pin PT 100+
red wire – pin PT 100-
(chap. 4.6, page 20; chap. 6 page 33).
Pt100 temperature sensor (-2B1) is defective Replace Pt100 temperature sensor (chap.
4.13, page 28).
Controller board (-2N1) is defective controller housing with controller (chap. 4.6,
page 20).
Alarm “-199” Pt100 temperature sensor (-2B1) is defective Pt100 temperature sensor (chap. 4.13,
Short circuit of temperature page 28).
sensor, underflow of
measuring range
Controller board (-2N1) is defective Replace controller housing with controller
(chap. 4.6, page 20).

Advantage-Lab, AL01-06+07 SM 02-10_en page 13/34


AL01-06+07 Microbiological Incubator – Service Manual

4 Repair

WARNING
Contamination of the unit with toxic, infectious or radioactive substances.
Danger of poisoning.
Danger of infection.
[ Never service a unit with sticking toxic or radioactive substances or sources of
infection.
¾ Prior to service, let the unit free from sticking toxic or radioactive substances by the
customer and disinfect it.
¾ Dispose of a unit which cannot be safely free from all toxic, infectious or
radioactive substances as special waste according to national law.

The following chapter provides the service technician with all the information required to mount and
dismount components.

DANGER
Danger of electric shock.
Danger to life.
¾ The unit must be switched off prior to performing most of the service tasks
described below.
¾ Disconnect the unit from the facility-side mains supply to make sure that it can not
be switched on again by mistake.

Advantage-Lab, AL01-06+07 SM 02-10_en page 14/34


Repair

4.1 Replacing a unit foot

Spare part See B11.1 to B11.2, chap. 5 page 31


Required Tools Crosstip screwdriver

Put down a blanket to cover the working area. This will protect the powder coated
surfaces from scratching.

5. Disconnect the AL01-06/07 from the facility-side mains supply.


6. Remove everything from the inside of the unit (incl. trays).
7. Place the unit on its side.
8. Unscrew the screw (B).
9. Remove the unit foot (A).
10.Mount the new foot by following steps 1. to 5. in reverse order.

A B

Figure 1: Replacing the unit feet

A Unit foot B Screw

Advantage-Lab, AL01-06+07 SM 02-10_en page 15/34


AL01-06+07 Microbiological Incubator – Service Manual

4.2 Replacing the outer door gasket

Spare part See B2, chap. 5 page 31).


Required Tools Crosstip screwdriver

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Unscrew the 16 screws (C) of the inner door side (A). The screws are hidden by the gasket (B).
3. Remove the inner door side (A) with the gasket (B).
4. Remove the old door gasket (B) from the edges of the inner door side.
5. Mount the new gasket by following steps 1. to 4. in reverse order.

Figure 2: Replacing the outer door gasket

A Outer door gasket C Screws


B Door panel of the outer door D Outer door

Advantage-Lab, AL01-06+07 SM 02-10_en page 16/34


Repair

4.3 Replacing the glass door gasket

Spare part See B3, chap. 5 page 31


Required Tools ---

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Remove the old glass door gasket (A).
3. Position the new glass door gasket on the upper door gasket edges (1) + (2).
4. Press the glass door gasket down onto the horizontal top part of the gasket edge.
5. Position the glass door gasket on the lower right gasket edge (3).
6. Pull the glass door gasket down diagonally (4) until you can press the vertical and horizontal parts
of the gasket onto all gasket edges.

Figure 3: Replacing the glass door gasket (open glass door not shown)

A Door gasket

Advantage-Lab, AL01-06+07 SM 02-10_en page 17/34


AL01-06+07 Microbiological Incubator – Service Manual

4.4 Replacing the glass door

Spare part see A1.1 to A1.2, chap. 5, page 31


Required Tools Crosstip screwdriver

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Open the outer door (A) and the glass door (D).
3. Remove the screws (C) at the upper and then the lower glass door hinges (B).
4. Remove the glass door (D).
5. Mount the new glass door by following steps 1. to 4. in reverse order.

Figure 4: Replacing the glass door

A Outer door C Screws


B Glass door hinge D Glass door

Advantage-Lab, AL01-06+07 SM 02-10_en page 18/34


Repair

4.5 Replacing the miniature fuse

Spare part See B9, chap. 5 page 31).


Required Tools Slotted screwdriver

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Open the fuse holder (A) by pulling it in the direction of the arrow.
3. Pull the fuse (B) out of the holder.
4. Check the fuse type:
5 x 20mm, 250V / medium time-lag M, 10 Amp
5. Insert the new fuse by following steps 1. to 3. in reverse order.

A B

Figure 5: Replacing the miniature fuse

A Fuse holder B Miniature fuse (-X1)

Advantage-Lab, AL01-06+07 SM 02-10_en page 19/34


AL01-06+07 Microbiological Incubator – Service Manual

4.6 Replacing the controller housing with controller

Spare part See B1.1 to B1.2, chap. 5 page 31).


Required Tools Slotted screwdriver

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Use the screwdriver to push both holder clips (H) upwards.
3. Remove the controller housing (B) from the mounting plate (A).
4. Disconnect the main switch contacts (F).
5. Disconnect the controller board connectors (D, E).
6. Remove Pt100 wires from spring terminal (C).
white wire – pin Pt100+
red wire – pin Pt100-
7. Install the new controller housing by following steps 1. to 6. in reverse order.

Figure 6: Replacing the controller housing

A Mounting plate E 4-pin connector (-X11)


B Controller housing F Main switch (-1S1)
C Spring terminal for Pt100 connection G Controller board (-2N1)
D 3-pin connector (-X10) H Holder clip

Advantage-Lab, AL01-06+07 SM 02-10_en page 20/34


Repair

4.7 Replacing the main switch

Spare parts See B1.2, chap. 5 page 31).


Required Tools Slotted screwdriver

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Use the screwdriver to push both holder clips (F, see Figure 6) upwards.
3. Remove the controller housing from the mounting plate.
4. Disconnect the main switch contacts (A).
5. Press the clamps (C) and push the main switch (A) downwards out of the controller housing.
6. Mount the new main switch by following steps 1. to 5. in reverse order.
A B C

Figure 7: Replacing the main switch

A Main switch C Clamps


B Controller board

Advantage-Lab, AL01-06+07 SM 02-10_en page 21/34


AL01-06+07 Microbiological Incubator – Service Manual

4.8 Removing the rear panel

Spare parts ---


Required Tools Crosstip screwdriver

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Remove all the screws (C) (AL01-06:12 pieces, AL01-07:14 pieces).
3. Remove the back panel (B).
4. Mount the rear panel by following steps 1. to 3. in reverse order.

A B C

Figure 8: Removing the rear panel

A Unit housing C Screws


B Rear panel

Advantage-Lab, AL01-06+07 SM 02-10_en page 22/34


Repair

4.9 Replacing the IEC connector plug

Spare parts See B10.1 to B10.2, chap. 5 page 31


Required Tools Crosstip screwdriver

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Remove the rear panel (A) (see chap. 4.8 on page 22).
3. Mark the cables connected to the IEC connector plug (B) with their positions.
4. Disconnect the cables from the IEC connector plug (B).
5. Push the IEC connector plug toward the front out of the rear panel (A).
6. Install a new IEC connector plug by following steps 1. to 5. in reverse order.

A B

Figure 9: Replacing the IEC connector plug

A Rear panel B IEC connector plug (-X2)

Advantage-Lab, AL01-06+07 SM 02-10_en page 23/34


AL01-06+07 Microbiological Incubator – Service Manual

4.10 Replacing the solid state relay

Spare parts See B6.1 to B6.2, chap. 5 page 31


Required Tools Crosstip screwdriver

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Remove the rear panel (see chap. 4.8 on page 22).
3. Mark the cables connected to the relay (B) with their positions at the relay.
4. Disconnect the cables from the connectors of the relay.
5. Remove the both screws (A).
6. Remove the relay (B).
7. Install the new relay by following steps 1. to 5. in reverse order.

Figure 10: Replacing the solid state relay

A Screws B Relay

Advantage-Lab, AL01-06+07 SM 02-10_en page 24/34


Repair

4.11 Replacing the temperature fuse

Spare parts • See B8.1 to B8.2, chap. 5 page 31).


• Aluminum scotch tape
Required Tools • Crosstip screwdriver
• Knife

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Remove the rear panel (see chap. 4.8 on page 22).
3. Cut the rock wool with the knife at location (A).
4. Disconnect the cables from the connectors of the temperature fuse (C).
5. Remove both screws (D).
6. Remove the temperature fuse (B).
7. Install the new temperature fuse by following steps 1. to 6. in reverse order.
8. Seal the cut in the rock wool with aluminum scotch tape.

Figure 11: Replacing the temperature fuse

A Location of temperature fuse C Connector


B Temperature fuse D Screws

Advantage-Lab, AL01-06+07 SM 02-10_en page 25/34


AL01-06+07 Microbiological Incubator – Service Manual

4.12 Replacing the temperature safety device class 3.1

Spare parts • See B12.1 to B12.6, chap. 5 page 31).


• Aluminum scotch tape
Required Tools • Crosstip screwdriver
• Knife

E
A

B
C
D

Figure 12: Localization of the temperature safety device class 3.1

A Alarm lamp D Screws


B Alarm lamp cover (red) E Capillary tube
C Thermostat F Temperature sensor (expansion vessel)

Advantage-Lab, AL01-06+07 SM 02-10_en page 26/34


Repair

1. Disconnect the AL01-06/07 from the facility-side mains supply.


2. Remove the rear panel (see chap. 4.8 on page 22).
3. Remove the black thermostat knob.
4. Remove the screws (D) which fix the thermostat.
5. From the rear, pull the thermostat (C) out of the housing and the alarm lamp (A) out of its cover (B).
6. Mark the cables connected to the thermostat (C).
7. Disconnect the cables from the connectors of the thermostat (C).
8. Cut the rock wool with a knife at the point where the temperature sensor (F) is placed.
9. The temperature sensor (F) is fixed at the rear panel of the inner chamber with aluminum scotch
tape. Unstick the temperature sensor for changing the thermostat (C).
10.Install the new temperature safety device by following steps 1. to 9. in reverse order.
11.Seal the cut in the rock wool with aluminum scotch tape.

B
A B C D

Figure 13: Replacing the temperature safety device class 3.1 (rear view)

A Alarm lamp C Capillary tube


B Thermostat D Temperature sensor (expansion vessel)

Advantage-Lab, AL01-06+07 SM 02-10_en page 27/34


AL01-06+07 Microbiological Incubator – Service Manual

4.13 Checking and replacing the Pt100 temperature sensor

Spare parts • See B7.1 to B7.3, chap. 5 page 31).


• Aluminum scotch tape
Required Tools • Crosstip screwdriver
• Resistance measuring device (Ohm meter)
• Reference temperature measuring device

1. Place the reference instrument’s sensor alongside the incubator’s Pt100 sensor.
2. Disconnect the AL01-06/07 from the facility-side mains supply.
3. Open the controller housing (see chap. 4.6, page 20).
4. Disconnect the Pt100 wires from the spring terminal of the controller board (A, see Figure 6 on
page 20)
5. Connect the Pt100 wires to the resistance measuring device (Ohm meter). Compare the measured
resistance value to the values given in Table 2 (page 30) at the end of this chapter and assess if it
is situated inside the indicated tolerance bandwidth. If not, the Pt100 sensor is defective and must
be replaced. If no replacement is required, follow steps 1. to 5. in reverse order.
6. Remove the upper hinge cover (A).
7. Remove the screws (E) which fix the door panel (D).
8. Lift the door panel (D) out of the door housing (B) and disconnect the ground cable.
9. Remove the rock wool from the outer door.
10.Remove the rear panel (see chap. 4.8 on page 22).
11.Cut the rock wool with the knife at the point (G) where the Pt100 temperature sensor (H) is placed.
12.Pull the Pt100 temperature sensor (H) through the duct (G) out of the inner chamber.
13. Cut all lacings cords which attach the Pt100 cable (I) to other cables.
14.Pull the Pt100 cable (I) carefully out of the unit from the rear.
15.Introduce the new Pt100 temperature sensor with its cable from the front through the hinge until
you can take it from the rear.
16.Mount the new Pt100 temperature sensor by following steps 1. to 10. in reverse order.
17.Seal the cut in the rock wool with aluminum scotch tape.
18.Calibrate and adjust the controller (see description in the user manual)

Advantage-Lab, AL01-06+07 SM 02-10_en page 28/34


Repair

E
F

Figure 14: Replacing the Pt100 temperature sensor

A Hinge Cover F Outer door gasket


B Outer door G Pt100 duct
C Rock wool H Pt100 temperature sensor
D Door panel I Pt100 cable
E Screws

Advantage-Lab, AL01-06+07 SM 02-10_en page 29/34


AL01-06+07 Microbiological Incubator – Service Manual

Table 2: Pt100 resistance and temperature values

Temperature as given Pt100 standard value Temperature as given Pt100 standard value
by the reference according to DIN EN by the reference according to DIN EN
measuring device 60751 (IEC 751) measuring device 60751 (IEC 751)
[°C] / [°F] [Ω] [°C] / [°F] [Ω]
15 / 59 105.85 28 / 82.4 110.9
16 / 60.8 106.2 29 / 84.2 111.3
17 / 62.2 106.6 30 / 86 111.7
18 / 64.4 107.0 31 / 87.8 112.0
19 / 66.2 107.4 32 / 89.6 112.45
20 / 68 107.8 33 / 91.4 112.8
21 / 69.8 108.2 34 / 93.2 113.2
22 / 71.6 108.6 35 / 95 113.6
23 / 73.4 109.0 36 / 96.8 114.0
24 / 75.2 109.35 37 / 98.6 114.4
25 / 77 109.7 38 / 100.4 114.8
26 / 78.8 110.1 39 / 102.2 115.15
27 / 80.6 110.5 40 / 104 115.5

The measured resistance value [Ω] should correspond to the temperature value measured with the
reference instrument, within a tolerance range of +/-3 Ω. This tolerance can be assumed if the
accuracy of the temperature reference measuring device is at least 2 degrees and the accuracy of the
employed resistance testing unit is at least 2 Ω. With less accuracy, consider the respective tolerances
of these devices.
Remark: This is a rough test meant to decide whether to replace the Pt100 sensor or not. If you note a
deviation from the expected value given in Table 2 which is less than +/-3 Ω, this may be due to the
incertitude of the measuring equipment. In order to be sure that there is no deviation, you can calibrate
the controller (see description in the user manual).

Advantage-Lab, AL01-06+07 SM 02-10_en page 30/34


Spare Parts

5 Spare parts

B1.1 B6.3 B7.1


B1.2 B2 A1.1 B3 B6.4 B4 B7.2 B6.1

B11.5 B9.1 B10.1 B5.1 B11 B8.1


B9.2 B10.2 B5.2 B8.2
B10.3
B10.4

Advantage-Lab, AL01-06+07 SM 02-10_en page 31/34


AL01-06+07 Microbiological Incubator – Service Manual

No Description Quantity Order No.


A1.1 Glass door with lock and door hinges AL01-06 1 8010-0082
A1.1 Glass door with lock and door hinges AL01-07 1 8010-0084
Lock block for glass door lock, handle 1 6002-0366
Lock block for glass door lock, holder 1 6002-0365
B1.1 Controller housing with controller 1 5001-0018
B1.2 Main switch 1 5019-0019
B2 Outer door gasket AL01-06 1 6005-0184
B2 Outer door gasket AL01-07 1 6005-0191
B3 Glass door gasket AL01-06 1 6005-0118
B3 Glass door gasket AL01-07 1 6005-0143
B4 Trays AL01-06 2 6004-0077
B4 Trays AL01-07 2 6004-0079
B5.1 Solid state relay 1 5011-0036
B5.2 Screw 3.9 x 9.5 2 3001-0003
B6.1 Pt100 temperature sensor 1 5002-0056
B6.2 Pt100 temperature sensor holder 1 6006-0016
B6.3 Screw M4 x 6 A2 1 3001-0004
Toothed disc for M4 1 3003-0002
B7.1 Temperature fuse 110°C 1 5006-0043
B7.2 Screw 3.9 x 9.5 A2 2 3001-0015
B8.1 IEC connector plug with fuse holder 10A 1 5024-0205
B8.2 Miniature fuse10 Amp medium time lag 1 5006-0012
B8.3 Connection cable EU 1 5023-0174
B8.3 Connection cable England 1 5023-0182
B8.3 Connection cable Switzerland 1 5023-0183
B9.1 Unit foot 5 6002-0006
B9.2 Screw 3.9 x 9.5 5 3001-0003
B10.1 Thermostat 20°C up to 90°C class 3.1 1 5006-0057
B10.2 Turning knob 1 6002-0040
B10.3 Thermostat holder 1 4020-0585
B10.4 Screw M4 x 6 4 3001-0001
B10.5 Red alarm lamp 1 5008-0003
B 11 Interference capacitor with nut 1 5026-0001
--- Flat assembly bar with 2 screws each 2

Advantage-Lab, AL01-06+07 SM 02-10_en page 32/34


Wiring Diagrams

6 Wiring diagrams

Advantage-Lab, AL01-06+07 SM 02-10_en page 33/34


Manufacturer:
Advantage Lab GmbH
Jachthoornlaan 8
2970 Schilde
Belgium
Internet: www.advantage-lab.com

The distribution and duplication of this document, and the use and disclosure of its content, is
prohibited except where explicit permission has been granted. Damages will be sought for violations.
All rights to patents, utility models and registered designs are reserved.
© by Advantage Lab GmbH, 2006

You might also like