Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 4

Re-validation of EO Sterilization – When is re-validation required?

This article reviews some of the factors to consider when you are evaluating the need
for EO sterilization re-validation.
Your cart is empty

ISO 11135-1:2014 is the international standard for sterilization validation for Ethylene
Oxide (EO or EtO) sterilizers. The standard describes multiple methods of sterilization
validation: 1) overkill approach, 2) single lot release, and 3) parametric release. The
overkill approach is the most common method for validation of your EO sterilization
process. If you are using a contract sterilizer, the sterilizer will already have completed
an Installation Qualification (IQ) and an Operational Qualification (OQ). You will need
to complete a Performance Qualification (PQ) for your product. A typical PQ for initial
process validation consists of the following:
1. Process Challenge Device (PCD) validation
2. Bioburden measurement
3. EO residual measurement (as per ISO 10993-7:2008/R2012)
4. Fractional Cycle (at least one)
5. 3 Half Cycles
6. 3 Full Cycles (or 1 Full Cycle, if performed in parallel with the three half-cycles)
Most contract sterilizers already have one or more Process Challenge Devices (PCDs)
that they have developed, and they may evaluate multiple PCDs in your fractional cycle
to determine which PCDs are more challenging to kill than an internal biological
indicator (BI). In addition to the above “typical” PQ, you might also choose to validate
partial loads and/or re-sterilization of product in the case of rework.
Re-validation of EO Sterilization – When is re-validation required?
In the ISO 11135 standard, #4 and #5 are referred to as the microbial performance
qualification (MPQ), while #6 is referred to as the physical performance qualification
(PPQ). For a successful MPQ, at least some of the PCDs must be non-sterile after a
fractional cycle to demonstrate the ability to recover the BI challenge organism. After a
half-cycle, however, all biological indicators should be sterile.

Ethylene oxide sterilization is typically outsourced to a contract sterilizer due to the


environmental and safety requirements of working with EO. Usually, the contract
sterilizer will provide a standard validation protocol for full validation that is compliant
with ISO 11135-1. However, the ISO 11135-1 standard requires that manufacturers
perform annual process reviews to evaluate the need for re-validation of the
sterilization process. Assuming there have been no problems, and no changes to the
product or process, then re-validation is not required at the end of the first year.
However, companies are expected to re-validate the process after two years–even if
there have been no changes. So why do some companies perform re-validation after
three years or more?
Longer Re-Validation Cycles
If there have been no changes to the sterilization process, the product, or the biological
indicators, then the manufacturer can use this as a justification for waiting until two
years have elapsed before re-validating the ethylene oxide sterilization process. Also,
there should be no evidence of sterilization failures or other problems with the validated
process. However, that alone is not necessarily enough to justify extending the
duration between validations beyond two years. Companies that can justify intervals of
three or more years have multiple products that are using the same sterilization
process.
In this case, the manufacturer may alternate annually between three, four, or even five
different product families that are using the same sterilization process. In this case, one
of the product families is being re-validated each year or every two years, but the
interval between validations for any one product family is longer. If the products are
made of similar materials and using the same sterilization process, then this approach
is valid. If you only have one product, then you need to re-validate the sterilization
process once every two years to verify the process remains active.
Minimum Re-Validation Requirements
Re-validation of EO Sterilization – When is re-validation required?
When you determine that it is time to re-validate your ethylene oxide sterilization
process, you need to perform the following tests to meet the minimum requirements of
ISO 11135-1:
1. Re-validation of PCD
2. Bioburden measurement
3. EO residual measurement
4. 1 Half Cycle
5. 1 Full cycle (to verify the EO residuals are acceptable)
The purpose of #1 is to verify that the resistance of internal BIs used in the half-cycle is
more resistant than the product bioburden. The purpose of #2 is to verify that bioburden
levels have not changed, and the type of organisms has not changed. In practice, most
companies monitor bioburden quarterly, and therefore this step should be routine. Step
3, EO residual measurement, should be performed to verify that there have not been
minor changes to the product or process that would increase the concentration of EO,
Ethylene Chlorohydrin (ECH), or Ethylene Glycol (EG) beyond the Tolerable Contact
Limit (TCL). The purpose of this third test is to prevent localized irritation caused by
residual chemicals from the ethylene oxide sterilization process.
Step 4 of the re-validation is intended to verify that a full injection of EO is more than
required to kill the bioburden present for the number of injections required for a half-
cycle.
The final step is to perform a full cycle. The product from the full cycle is typically used
for EO residual testing. Any product from the full cycle that is not used for testing can
be sold after sterility testing is complete.
Partial Loads & Rework
If you occasionally sterilize loads that are less than “full loads,” then you need to
ensure that you have validated a minimum load or a specific partial load (e.g., half-
pallet, instead of a full pallet). In the case of a partial or minimum load, you may identify
different locations in your load that is considered “worst-case.” These are the locations
that had PCDs that were not sterile in a fractional cycle.
Most companies do not have concerns about the cost of the actual sterilization runs
during re-validation, and biological indicators are typically less expensive than boxes of
product. The primary cost concern for re-validation is any product that must be
scrapped. Therefore, many companies will accumulate dunnage (i.e., empty packaging
or scrap product) over time to fill a sterilizer. This dunnage may be used to ensure that
every load is a full load, or it may only be used for re-validation.
Re-validation of EO Sterilization – When is re-validation required?
Another alternative to using dunnage for re-validation is to validate a rework process.
Any product that is exposed to a fractional cycle or half-cycle can be resterilized in a
full cycle. To justify the commercial use of that product, a company needs to validate
that the product will not be damaged by exposure to two full cycles. One of the key
acceptance criteria for rework is the EO residual levels in the product. However, the
manufacturer also needs to determine if there has been any deterioration of the
product by a second exposure to EO that would affect performance.
Other Considerations
Many companies do a poor job of reviewing the potential impact of changes to a
product, packaging, and biological indicators. Ideally, initial validation involves different
lots of product, packaging, and biological indicators to assess lot-to-lot variability.
However, many times, the packaging and biological indicators consist of only one lot
during validation. Minor changes to the tolerances may reduce the amount of ethylene
oxide that is absorbed by the product or change the resistance of the biological
indicator to the sterilization process. Therefore, these minor changes should trigger a
re-validation.
Changes in suppliers with the same specification can also be difficult to evaluate. If a
component is made of a material that absorbs EO, then it may be recommended to re-
validate sterilization for any changes to suppliers of those components. Re-validation in
these cases may consist of only a fractional cycle, half cycle, or a full-cycle to evaluate
risks associated with the change.
Who Should Be Making Evaluations
The evaluation of the need for re-validation should include the input of three types: 1)
microbiological, 2) materials, and 3) performance. To make these assessments,
typically, a cross-functional team is needed. Someone with responsibility for design
and development can assess the performance impact of changes. A materials
engineer is generally needed for assessment of the interaction between components
and EO. Finally, a microbiologist is needed to confirm that there is no impact related to
biological indicators or bioburden.

You might also like