Install Manual
For
FLAATZ 560
FLAATZ 760
Digital
Radiography
Technologies
www.drtech.co.kr
CONTENTS
CHAPTER 1
Preparation for Retrofit
Classification
(Immediate sale, sale after demo, demo)
Demo eriod
DRT-MAN-049
Digital Radiography Technologies
1-2-2. Distance between the wall's FPD and the control box
- This distance should be measured based on the length of the cable to be laid, not the lineal distance
between the wall's flat-panel detector (FPD) and the control box (default value: 10 meters).
IS560 : Available in two versions, the Standard (10 meters) and the Option (7 meters).
IS760 : Available in one version only, the Standard (10 meters).
- A repeater and an extension cable (7 meters in length) are required if the distance between the wall
FPD and the control box is over 10 meters and shorter than or equal to 17 meters.
LVDS communication problem will occur if the distance is longer than 17 meters.
WARNING
1-2-3. Distance between the table FPD and the control box
- This distance should be measured based on the length of the cable to be laid, not the lineal distance
between the table FPD and the control box (default value: 10 meters).
IS560 : Available in two versions, the Standard (10 meters) and the Option(7 meters).
IS760 : Available in one version only, the Standard (10 meters).
- A repeater is required if the distance between the table FPD and the control box is over 10 meters and
shorter than or equal to 17 meters.
LVDS communication problem will occur if the distance is longer than 17 meters.
WARNING
The distance between the FPD and the generator should maintain at least 1.83 meters.
CAUΤΙΟΝ
It should be noted that the user will be held responsible for any damage that may result from
CAUΤΙΟΝ
operating or storing the FPD at temperatures outside the above range
1-2-6. Dehumidifier
- High humidity levels during the rainy season may cause damage to the FPD.
- Operation humidity range : 20 ~ 75% RH (non-condensed)
- Storage humidity : 5 ~ 95% RH (non-condensed)
It should be noted that operating or storing the FPD at humidity levels outside the above ranges
WARNING
may shorten its lifetime and cause failure.
1-2-7. Cleanness
- Dust or foreign substances may cause malfunction of the bucky or artifact defects in the FPD.
- The FPD should be cleaned regularly.
3m
1.8
2.5m
1.83m 1.83m
1-3-1. Grid
- Fixed Bucky : FG14(for F560), FG17(for F760)
- Spring Bucky : It is available from 71 ~ 103 lines per inch, 103 lines per inch recommended
- Stepping motor : 200 or above lines per inch recommended
- CAM motor : Not recommended (Unlike computed radiography images (CR), digital radiography images
(DR) can express elaborate grid patterns. Also, since CAM motors do not move in parallel with the
FPD, it is highly likely that grid patterns will occur
In case of the CAM motor, first, apply the Grid Filter Algorism. When you install the FPD
system, prepare the FG14 or FG17 for eventuality.
NOTE - The SID (focal distance) should be used in accordance with the operating environment.
A shield is required if there are parts that generate magnetic or electric noise from the
CAUΤΙΟΝ
outside.
For example
www.drtech.co.kr 07
08
DRT-MAN-049
Digital Radiography Technologies
Refer to DRTECH Terminal Emulator Manual for adjusting Ready Delay and Exposure window
time.
NOTE
+5V
Prep / Ready
Function
Ready / Prep
Exposure
Function
EXP_
sensor
Exposure
Function
B. DROC PC Information
- Obtain information of DROC PC's IP Address, Subnet Mask, etc. which is to be installed.
- Check whether remote access of DROC PC is possible from outside source.
- If Teamviewer (remote access program) is available, remote support from DRTECH is possible in the
event of a problem.
1-6-2. PACS
- Obtain PACS information before the X-ray Image of the hospital is sent.
A. Storage Server
- IP Address
- Port
- AE Title
- Modality
B. Worklist
- IP Address
- Port
- AE Title
- Modality
- IP Address
- Port
- AE Title
www.drtech.co.kr 13
14
CHAPTER 2
Installation of FLAATZ system
2-1 Tools 16
2-2 Product Identification 18
2-3 Connection of the Product 21
2-4 Connection of Ready, Exposure Input 22
2-5 Ground Connection 25
2-6 Generator Time setting 26
2-7 Driver Installation 28
2-8 Calibration 29
2-9 GRID Guide 31
2-10 F560 & F760 insert to Bucky 40
www.drtech.co.kr 15
16
2-1. Tools
Flathead screwdriver
& In various size
Philips screwdriver
Modular Press
&
RJ 11 Connector
Nippers
&
Stripper
Dust Remover
Tester
Cable Ties
Cable Mount
Cable Mold
Ladder
Styrofoam Cushion
FLAATZ Package
Protection Case
(F560 only)
Upper Cushion
PCM Packs
PCM Packs
Lower Cushion
Protection Case is only available for F560 and it is enclosed when the order is requested.
PCM is enclosed for products that are shipped from Nov. 1 to Mar. 31 of next year.
NOTE
www.drtech.co.kr 19
20
Ethernet Adapter
Interface Cable Software CD
Switch Box Cable Switch Box
Lan Cable Repeater Box (Opt.)
Hand Switch
Accessory Box
Detector Panel
Interface Cable ►
DATA IN 1 DATA IN 2
Syn
Ext Sync.
Hand Switch
Switch Box
LAN Cable ▼
Switch Box Cable ▼
Stand Table
Detector Panel Detector Panel
DATA IN 1 DATA IN 2
Ext Syn
Sync.
Hand Switch
Switch Box
- When there is only one Ready and Exposure Common input, use by tying A and D or change the Dip S/W
Setting within the Switch Box (Refer to Switch Box setting)
DRT-MAN-049
Digital Radiography Technologies
REL1
JRC-23F
2 6 1 8
+ RDY_IN_GEN
3
2 7
1 8 5 4 RDY_IN_GEN_COM
-
3 6
2 7 EXP_OUT_GEN
2 6 4 5
+ EXP_OUT_GEN_COM
3 6 3
5 4 DIP2
4 5 -
RDY_IN_GEN_COM EXP_OUT_GEN_COM SW / SMD_DIP8
10uF_10V
DIP1 C5 REL2
SW / SMD_DIP8 JRC-23F
JRC-23F
2
+
6 Ready_com Common
3
5 4
-
Exposure Exposure (+24V)
JRC-23F
Ready Ready (+24V)
2 6
+
3
5
- 4 Exp_com
JRC-23F
2
+
6 Ready_com Ready_com
3
5 4
-
Exposure Exposure (+24V)
JRC-23F
Ready Ready (+24V)
2 6
+
3
5
- 4 Exp_com Exposure com
0.5A 125VAC
UL/CUR Rating 1A 30 VAC
0.3A 60 VDC
Digital Radiography Technologies
DRT-MAN-049
2-7-3. Precautions
- When using Ethernet Type, firewall must be disabled for stable operation.
- When security programs such as V3, Macfree, etc. are being used, there can be problems with image
capture function.
DRT-MAN-049
Digital Radiography Technologies
2-8. Calibration
2-8-1. MAP, GMP, PMP
- Copy MAP, GMP, PMP file on the enclosed MAP CD to the set folder in the corresponding DROC S/W.
- Correct the file name to the set file name in the corresponding DROC S/W or correct the ini file.
2-8-2. Caliview
Please refer to Caliview manual for details.(Manual file name : Caliview_Manual_Eng.pdf)
- Run setup program to install Caliview software on a computer
- Please refer to Chapter 3. Installation Procedure of Caliview manual.
2-8-6. Caution
When proceeding with Gain Calibration, it is recommended that Gain Calibration is executed
without the Grid or Bucky Cover
CAUΤΙΟΝ
DRT-MAN-049
Digital Radiography Technologies
B. Stepping Motor
- Moving speed and status must be good.
- Grid Pattern must not show on the final X-ray when the Grid Filter is used.
C. CAM Motor
- Moving speed and status must be good.
- Grid Pattern must not show on the final X-ray when the Grid Filter is used.
D. Fixed Type
- Only FG14 and FG17 provided by DRTECH may be used.
- Regarding Filter file, there are current filter file on the C: Caliview Filter folder
www.drtech.co.kr 31
32
DRT-MAN-049
Digital Radiography Technologies
If you use the ‘Spring moving bucky’ or ‘Motor moving bucky’, you should input the Lines/inch and Angle.
Grid ratio and SID is not necessary
DRT-MAN-049
Digital Radiography Technologies
- Parameters
Default
Session Item Range of input value Description
Value
0 = No grid
Specify the type of grid.
1 = Spring Bucky
Parameters are used for mathematical
2 = Fixed Grid(19 degree tilting)
GRID TYPE 0 operation to remove grid patterns.
3 = Motor Bucky
Motor Bucky include the CAM and Stepping
4 = FG14(Fixed Grid for 1417 size)
Motor Bucky
5 = FG17(Fixed Grid for 1717 size)
Default
Session Item Range of input value Description
Value
- If the value of the parameters is set wrong, serious problem can occur in the image.
Please be careful not to make mistakes.
CAUΤΙΟΝ - In case of using earlier version of API, INI file location can be “C: \ DRTECH.INI”
DRT-MAN-049
Digital Radiography Technologies
Model
Angle FLAATZ-500 FLAATZ-560
0 Degree
90 Degree
(The Spring,
Motor bucky
will be
update later.)
Model
Angle FLAATZ-750E FLAATZ-760
0 Degree
If the grid is not parallel with the Detector and the gap is over 2cm, the grid line may appear
on the X-ray Images.
CAUΤΙΟΝ
ȭGrid
ȭDetector
Properly Paralleled
When the gap between the GRID and the Detector are different (Deviation should be less than 2 cm)
ȭGrid
ȭDetector
Improperly Paralleled
www.drtech.co.kr 39
40
CHAPTER 3
DROC Operation
DRT-MAN-049
Digital Radiography Technologies
1 2 3 4
Power On / Off
Brightness, Contrast Automatic-Adjusting
Menu Button Button
/ Reduce (-) Button / Increase (+) Button
with marks, etc.
2. Press - and + button to navigate down to the Color Settings menu icon.
3. Select the appropriate Red Preset for programs that generally require a lot of colors (medical imaging,
photo editing, etc.).
When the Red Preset doesn’t satisfy your requirements, the ‘User Preset Setting’ can be
used to independently increase or reduce the colors (R, G, B).
DRT-MAN-049
NOTE
www.drtech.co.kr 45
46
Acquiring
PACS Viewer Monitor in the
initial images by Diagnostic monitor
Monitor doctor’s office
the first monitor
After obtaining and transmitting the initial set of images, follow the phased verification
process and check the monitor in the doctor’s office. Then, if the images need
modifying, re-verify them by adjusting their brightness and contrast levels. If the
NOTE
images need to be further modified, it is also desirable to use the monitor’s OSD menu.
Acquiring
PACS Viewer Monitor in the
initial images by
Monitor doctor’s offic
the first monitor
After obtaining the initial set of images, transmit them to the PACS viewer’s monitor,
compare the transmitted images. Then, make necessary modification on them.
NOTE
If you have created the same settings between the first monitor, diagnostic monitor and the PACS viewer
monitor but they display images with different qualities, please request a calibration to your PACS manufacturer
Digital Radiography Technologies