DRTECH is a global provider of high quality Flat-Panel Digital X-ray

Detectors with its own development and manufacturing facilities.

If you need technical support


or have any difficulty or
problem with our products,
please contact our Customer
Support Team as indicated below.
Web Site
http://www.drtech.co.kr

Customer Support Team


E-mail: drtech@drtech.co.kr
Tel: +82 31 730 6805 / Fax: +82 31 730 6899
DIGITAL RADIOGRAPHY TECHNOLOGIES

Install Manual
For
FLAATZ 560
FLAATZ 760

FLAATZ 560 & 760 X-ray Detector


Flat-Panel Digital X-ray Detectors High Definition X-ray Imaging!
Digital Radiography Technologies

Digital
Radiography
Technologies
www.drtech.co.kr

CONTENTS

Chapter. 1 Preparation for Retrofit


1-1 Survey for Installation site 4
1-2 The X-ray Room Survey 5
1-3 Survey for X-ray Equipment 7
1-4 Survey for Operating Room 9
1-5 Survey for Generator 10
1-6 Survey for PACS or Dicom Printer 13

Chapter. 2 Installation for Retrofit


2-1 Tools 16
2-2 Product Identification 18
2-3 Connection of the Product 21
2-4 Connection of Ready, Exposure Input 22
2-5 Ground Connection 25
2-6 Generator Time setting 26
2-7 Driver Installation 28
2-8 Calibration 29
2-9 GRID Guide 31
2-10 F560 insert to Bucky 40

Chapter. 3 DROC Operation


3-1 PACS, Dicom Printer 42
3-2 Monitor Setting Guide 43
Digital Radiography Technologies

CHAPTER 1
Preparation for Retrofit

1-1 Survey for Installation site 4


1-2 The X-ray Room Survey 5
1-3 Survey for X-ray Equipment 7
1-4 Survey for Operating Room 9
1-5 Survey for Generator 10
1-6 Survey for PACS or Dicom Printer 13
www.drtech.co.kr 03
04

1-1. Survey for Installation site


For a more smooth installation of the instruments, a preliminary survey should be made by visiting the
hospital or the clinic and observing the following checkpoints, before installing these items.

Survey for Installation site


Customer Name

Classification
(Immediate sale, sale after demo, demo)

Demo eriod

Hospital (Clinic) Name

Hospital (Clinic) Address

Hospital (Clinic) Staff

Hospital (Clinic) Contact

X-ray room number for installation

Average number of daily patients

Average number of daily X-rays taken

Any operating X-ray equipment other than


from DRTECH

DRT-MAN-049
Digital Radiography Technologies

1-2. The X-ray Room Survey


For a more smooth installation of the instruments, a preliminary survey should be made by visiting the
hospital or the clinic and observing the following checkpoints, before installing these items.

1-2-1. Dimensions of the X-ray room


- The X-ray room's length, width and height should not inhibit the patient's traffic line.

1-2-2. Distance between the wall's FPD and the control box
- This distance should be measured based on the length of the cable to be laid, not the lineal distance
between the wall's flat-panel detector (FPD) and the control box (default value: 10 meters).
IS560 : Available in two versions, the Standard (10 meters) and the Option (7 meters).
IS760 : Available in one version only, the Standard (10 meters).
- A repeater and an extension cable (7 meters in length) are required if the distance between the wall
FPD and the control box is over 10 meters and shorter than or equal to 17 meters.

LVDS communication problem will occur if the distance is longer than 17 meters.
WARNING

1-2-3. Distance between the table FPD and the control box
- This distance should be measured based on the length of the cable to be laid, not the lineal distance
between the table FPD and the control box (default value: 10 meters).
IS560 : Available in two versions, the Standard (10 meters) and the Option(7 meters).
IS760 : Available in one version only, the Standard (10 meters).
- A repeater is required if the distance between the table FPD and the control box is over 10 meters and
shorter than or equal to 17 meters.

LVDS communication problem will occur if the distance is longer than 17 meters.
WARNING

1-2-4. Distance between the FPD and the generator


- Check whether there is any interference due to high voltage, by checking the distance between the FPD
and the generator.
(Unlike in computed radiography (CR), noise may occur in the images if the generator creates high-
voltage interference in digital radiography (DR). This may break the synchronization.)

The distance between the FPD and the generator should maintain at least 1.83 meters.
CAUΤΙΟΝ

1-2-5. Air conditioner


- High temperatures in summer and low temperatures in winter may cause damage to the FPD.
- The temperature around the FPD must be kept at room temperature.
- Operation temperature range : 5 ~ 35
- Storage temperature range : 0 ~ 40

It should be noted that the user will be held responsible for any damage that may result from
CAUΤΙΟΝ
operating or storing the FPD at temperatures outside the above range

FPD : Flat Panel Detector


www.drtech.co.kr 05
06

1-2-6. Dehumidifier
- High humidity levels during the rainy season may cause damage to the FPD.
- Operation humidity range : 20 ~ 75% RH (non-condensed)
- Storage humidity : 5 ~ 95% RH (non-condensed)

It should be noted that operating or storing the FPD at humidity levels outside the above ranges
WARNING
may shorten its lifetime and cause failure.

1-2-7. Cleanness
- Dust or foreign substances may cause malfunction of the bucky or artifact defects in the FPD.
- The FPD should be cleaned regularly.

1-2-8. Earth connection terminals


- To prevent power noise that may occasionally occur, it is necessary to connect the FPD system with the
earth terminal.

Example of patient surroundings (Recommended minimum distance)

3m
1.8

2.5m

1.83m 1.83m

Check List for X-ray Room


X-ray Room
Width : Length : Height :
Dimension

Distance from Wall FPD


Within 10M 10M ~ 17M More than 17M
to Control Box

Distance from Table


Within 10M 10M ~ 17M More than 17M
FPD to Control Box

Distance from FPD


Within 1.83M More than 1.83M
to Generator

Air conditioner Yes No Dehumidifier Yes No


Operating Date Full Weekday Operating Time 24Hr Only day
DRT-MAN-049

Maintain Temperature 24Hours Only office hours Absence


Cleaning Yes No Earth Ground Yes No
Digital Radiography Technologies

1-3. Survey for X-ray Equipment


For a more smooth installation of the instruments, a preliminary survey should be made by visiting
the hospital or the clinic and observing the following checkpoints, before installing these items.

1-3-1. Grid
- Fixed Bucky : FG14(for F560), FG17(for F760)
- Spring Bucky : It is available from 71 ~ 103 lines per inch, 103 lines per inch recommended
- Stepping motor : 200 or above lines per inch recommended
- CAM motor : Not recommended (Unlike computed radiography images (CR), digital radiography images
(DR) can express elaborate grid patterns. Also, since CAM motors do not move in parallel with the
FPD, it is highly likely that grid patterns will occur

In case of the CAM motor, first, apply the Grid Filter Algorism. When you install the FPD
system, prepare the FG14 or FG17 for eventuality.
NOTE - The SID (focal distance) should be used in accordance with the operating environment.

1-3-2. Bucky Operation


- The operation mechanism should have no problem when using the spring, the CAM motor or the
stepping motor type bucky
- Suitable grids should be used for each operation type
- The speed, movement and vibration types should be checked

- The grid should be moving during X-ray exposure checks


- DR Tech's fixed grid should be used if the current grid does not move in accordance with
CAUΤΙΟΝ the applicable conditions

1-3-3. Bucky circumstance


- Check whether any transformers, etc. are present around the FPD.
- Noise may occur in the images by magnetic or electric fields.

A shield is required if there are parts that generate magnetic or electric noise from the
CAUΤΙΟΝ
outside.

For example
www.drtech.co.kr 07
08

Check List for X-ray Equipment (Wall)


Bucky Types Fixed Stepping motor Spring CAM motor
GRID Manufacture GRID Model No.
Fixed Grid FG14 FG17 None DRTECH supplied product
Line Resolution lines/inch lines/cm
Ratio SID cm
Speed Good Bad
Moving Status Good Bad
Vibration Severe Middle Mild
Transformer No Yes (no shields) Yes (shields)

Check List for X-ray Equipment (Table)


Bucky Types Fixed Stepping motor Spring CAM motor
GRID Manufacture GRID Model No.
Fixed Grid FG14 FG17 None DRTECH supplied product
Line Resolution lines/inch lines/cm
Ratio SID cm
Speed Good Bad
Moving Status Good Bad
Vibration Severe Middle Mild
Transformer No Yes (no shields) Yes (shields)

(1) When using Fixed Type


- Only usable for FG14 (for 560), FG17 (for 760) provided by DRTECH
(2) When using Stepping motor Type Bucky
- When Moving state is good : Grid with 200 or above Lines/inch is recommended
- When Moving state is bad : FG14 or FG17 is recommended
(3) When using Spring type Bucky
- When Moving state is good : 71 to 103 lines/inch GRID can be used, 103 lines/inch is
recommended.
NOTE
- When Moving state is bad : FG14 or FG17 is recommended
(4) When using CAM motor
- When using CAM motor, there is a good chance that a Grid pattern may emerge
because the Grid movement is not parallel to the Detector. When installing, first use
Caliview and after the test using Grid Filter Algorism, prepare FG14 or FG17 in advance
DRT-MAN-049

in case of Grid pattern occurrence.


Digital Radiography Technologies

1-4. Survey for Operating Room

1-4-1. DROC Operating PC


- Check to see if there is additional space to place the PC for installing DROC S/W.
- Make a request to the hospital staff for a LAN line to connect to the network before installation.
- When using a router, image transfer to PACS and Dicom Printer may be slow, so it is preferable to assign
an independent IP.
- Check the position of the multi-tap and grounding for connecting PC, Monitor, etc.

1-4-2. Placing the Control Box, Switch Box,


and other components
- Confirm appropriate space for placing the Control Box (Failure may occur when there is severe bending of
the connecting cables)
- Confirm appropriate space for placing the Switch Box (Place it where the user can easily check the LED
status of the switch)
- Hand switch should be placed where the operator can easily access it.
- Switch Box's Ready and Exposure connection cable should be easily connectable to the OP Console.

Check List for Operating Room


Desk Yes No
Network (LAN Port) Yes Using hub Not connected

PC Power Earth Ground


Connected Not connected
Connection

Control Box Lamp Check Possible Impossible

Control Box Connection


Severe Middle Favorable
Cable Bending

Switch Box Lamp Check Possible Impossible

Cable Length from


Less than 7.6M More than 7.6M
Switch Box to Control Box
www.drtech.co.kr 09
10

1-5. Survey for Generator


1-5-1. Generator Specification
- Regular HFG is recommended for General Human filming. (Due to insufficient capacity, good image
cannot be obtained when using Portable Generator.)
- At least 30kW grade is recommended in case of maximum capacity of L-spine LAT (95kVp 320mA
100mAs, 312msec).
- Generators over 40kW grade is recommended for taking high-voltage Chest PA (120kVp 320mA 10mAs
32msec).

- Irradiation time must be short to reduce lowered sharpness from motion.


- When using a generator that is not fixed, the quality of the image may be poor due to
WARNING lack of examination according to kVp and mAs settings.

Check List for Generator


Manufacturer
Model Name
Output Rating kW
kV Range ~ kVp
mA Range ~ mA
Year in use Year Month
Recent correction date (MM / DD / YYY)

DRT-MAN-049
Digital Radiography Technologies

1-5-2. Prep and Exposure input


1) Input method
- Verify that the Ready (Prep) and Exposure input of the OP Consol or generator is operating in a
contact point format.

Cannot connect when the format is Pulse input.


WARNING

2) Generator Ready Delay time


- Necessary time for Generator warming up after Ready input

3) Generator Exposure Delay time


- Necessary time before the start of actual X-ray examination after Exposure input
- Necessary for the Exposure Window time setting during actual installation later on.

Refer to DRTECH Terminal Emulator Manual for adjusting Ready Delay and Exposure window
time.
NOTE

4) Ready and Exposure Common Separation


- When the Common terminal is connected internally to the generator, it is necessary to tie up our
company’s A and D port of the Switch Box before connecting ( Refer to Switch Box connecting)

A. Contact point method of Hand Switch B. Other method of Hand Switch

+5V

Prep / Ready
Function

Ready / Prep
Exposure
Function
EXP_
sensor
Exposure
Function

Hand Switch Generator or OP Consol Box


Hand Switch Generator or OP Consol Box

Check List for Ready, Exposure Input


Input mode Contact Input Pulse Others
Ready Delay m sec
Exposure Delay m sec
Longest Exposure time m sec
Common Separation Connection Separation
www.drtech.co.kr 11
12

1-5-3. Checking Tube Spec


- Check the bottom check list for any problems with X-ray filming.
- X-ray filming may not work properly if there is no heat dissipation device during continuous high-voltage
filming.
- Adjustment of X-ray Tube is necessary when there is severe left and right variation due the Heel effect.

Check List for Ready, Exposure Input


Manufacturer
Model Name
Installation Date (MM / DD / YYY)

Recent Correction Date (MM / DD / YYY)

Maximum Tube Voltage kV


Focal Size mm

1-5-4. Checking Collimator Spec


- Max field size : When the SID is 100cm, it must be more than 45 x 45 cm (17 x 17 inch) in order for
examination over the entire FPD.
- Light source : When there is light, it helps with the patient's positioning.
- Lamp timer : If it is too fast, it causes difficulty in verifying the shooting angle.
- Pollutant : It occurs when there is too much dust or stain and affects the image.

Check List for X-ray Collimator


Max Field Size cm x cm (at SID 100cm)
Light Source Yes No
Lamp Timer seconds
Pollutant Clean Dirty
DRT-MAN-049
Digital Radiography Technologies

1-6. Survey for PACS or Dicom Printer


1-6-1. Basic Information Survey
A. Contact information of the hospital's data processor
- Obtain contact information of the hospital's data processor beforehand.
- Obtain contact information of the hospital's staff who manages its PACS or Dicom Printer.

B. DROC PC Information
- Obtain information of DROC PC's IP Address, Subnet Mask, etc. which is to be installed.
- Check whether remote access of DROC PC is possible from outside source.
- If Teamviewer (remote access program) is available, remote support from DRTECH is possible in the
event of a problem.

1-6-2. PACS
- Obtain PACS information before the X-ray Image of the hospital is sent.
A. Storage Server

- IP Address

- Port

- AE Title

- Modality

B. Worklist

- IP Address

- Port

- AE Title

- Modality

1-6-3. Dicom Printer server


- Obtain Dicom Printer information before the X-ray Image of the hospital is sent.

- IP Address

- Port

- AE Title
www.drtech.co.kr 13
14

Check List for Hospital information


Hospital Name
Person in Charge
Computer Room
Manager Contact Info
DROC PC
IP Address
Subnet Mask
Gateway
DNS
External Remote contact Available Impossible
Teamviewer ID Teamviewer Password

Reference site : http://www.teamviewer.com

Check List for PACS or Dicom Printer


Manufacture
Manager Contact Info
Storage Server Information
IP Address
Port
AE Title
Modality
Worklist Server Information
IP Address
Port
AE Title
Dicom Printer Information
IP Address
Port
AE Title

Refer to DROC Software Manual


DRT-MAN-049
Digital Radiography Technologies

CHAPTER 2
Installation of FLAATZ system

2-1 Tools 16
2-2 Product Identification 18
2-3 Connection of the Product 21
2-4 Connection of Ready, Exposure Input 22
2-5 Ground Connection 25
2-6 Generator Time setting 26
2-7 Driver Installation 28
2-8 Calibration 29
2-9 GRID Guide 31
2-10 F560 & F760 insert to Bucky 40
www.drtech.co.kr 15
16

2-1. Tools

2-1-1. Basic tools

Photo Tools Note

Flathead screwdriver
& In various size
Philips screwdriver

Hexagon wrench set L-shaped, T-shaped

Soldering Iron and solder For charging

Modular Press
&
RJ 11 Connector

Nippers
&
Stripper

Dust Remover

Tester

Terminal Press Ground Cable Connecting


DRT-MAN-049
Digital Radiography Technologies

2-1-2. Tools and Materials for Cable Arrangement

Photo Tools Note

Cable Ties

Cable Mount

Cable Mold

Ladder

Earth Ground Cable

Earth Ground Terminal


www.drtech.co.kr 17
18

2-2. Product Identification


2-2-1. Unpacking
- Check for tears in the package or signs of damage from impact or heat.
- When there are missing components of the product or other problems, make a photo record and submit to
DRTECH's CDN.

Accessory Box Package

Control Box Package

Detector Panel Package

Styrofoam Cushion

FLAATZ Package

Check List for Package


Product package Normal Defective
Components Normal Defective
Impact Normal Defective
Temperature Normal High Low
DRT-MAN-049
Digital Radiography Technologies

2-2-2. Checking Panel Package Box


- Check for tears in the package or signs of damage from impact or heat.
- When there are missing components of the product or other problems, make a photo record and submit to
DRTECH.

Protection Case
(F560 only)

Upper Cushion

PCM Packs

FLAATZ 560 Detector

PCM Packs

Lower Cushion

Detector Panel Package

Check List for Panel Package


Product package Normal Defective
Components Normal Defective
Impact Normal Defective
Temperature Normal Hight Low
PCM Yes No

Protection Case is only available for F560 and it is enclosed when the order is requested.
PCM is enclosed for products that are shipped from Nov. 1 to Mar. 31 of next year.
NOTE
www.drtech.co.kr 19
20

2-2-3. Checking Control Box & Accessory Package


- When there are missing components of the product or other problems, make a photo record and submit to
DRTECH.

Ethernet Type Accessory Package

Ethernet Adapter
Interface Cable Software CD
Switch Box Cable Switch Box
Lan Cable Repeater Box (Opt.)
Hand Switch

Accessory Box

Control Box Package

The components might be changed by order.

Check List for Control Box and Accessory Package


Control Box Normal Defective
MAP CD Normal Defective
Interface Cable Normal Defective
Switch Box Yes No Switch Box Cable Yes No
Hand Switch Yes No Power Cable Yes No
DROC S/W CD Yes No DROC S/W Dongle Yes No
LAN Card Yes No LAN Cable Yes No
DRT-MAN-049

PCI Frame Grabber Yes No Camera Link Cable Yes No


Interface Box Yes No D-SUB 15P Cable Yes No
Digital Radiography Technologies

2-3. Connection of the Product


2-3-1. Connecting F560, F760 System
F560 and F760 Single Connection

Detector Panel

Interface Cable ►

Control Box Front Control Box Rear

DATA IN 1 DATA IN 2
Syn
Ext Sync.

DATA OUT Ext I/O


Use Only With Fuses, T2.0AL 250V

Hand Switch

Switch Box

LAN Cable ▼
Switch Box Cable ▼

F560 and F760 Dual Connection

Stand Table
Detector Panel Detector Panel

Interface Cable ▼ Interface Cable ▼

Control Box Front Control Box Rear

DATA IN 1 DATA IN 2
Ext Syn
Sync.

DATA OUT Ext I/O


Use Only With Fuses, T2.0AL 250V

Hand Switch

Switch Box

LAN Cable ▼ Switch Box Cable ▼


www.drtech.co.kr 21
22

2-4. Connection of Ready, Exposure Input


2-4-1. Connecting Ready(Prep) and Exposure
- Connect the Gen_con connector terminal of the Switch Box to the Generator's Ready(Prep) and Exposure
input terminal.

2-4-2. DRTECH Ready(Prep), Exposure output pin map

Pin No. Pin Name Pin Description


A Ready Common Ready or PREP signal is sent from hand switch
B Exposure Exposure signal is sent from FPD System
C Ready / PREP Ready or PREP signal is sent from hand switch
D Exposure Common Exposure signal is sent from FPD System.

- When there is only one Ready and Exposure Common input, use by tying A and D or change the Dip S/W
Setting within the Switch Box (Refer to Switch Box setting)
DRT-MAN-049
Digital Radiography Technologies

2-4-3.Example of Generator’s Ready(Prep) and


Exposure input terminal

The input terminal is positioned behind the OP Console.

Connected to the middle part of the Hand switch as t


here is no input terminal

The Ready(Prep) and Exposure input connector is


position within the OP Console.
www.drtech.co.kr 23
24

2-4-4.Connection diagram between switch Box and Generator


P5V

REL1
JRC-23F
2 6 1 8
+ RDY_IN_GEN
3
2 7
1 8 5 4 RDY_IN_GEN_COM
-
3 6
2 7 EXP_OUT_GEN
2 6 4 5
+ EXP_OUT_GEN_COM
3 6 3
5 4 DIP2
4 5 -
RDY_IN_GEN_COM EXP_OUT_GEN_COM SW / SMD_DIP8

10uF_10V
DIP1 C5 REL2
SW / SMD_DIP8 JRC-23F

Switch Box Generator (3pin Input)

JRC-23F
2
+
6 Ready_com Common
3
5 4
-
Exposure Exposure (+24V)

JRC-23F
Ready Ready (+24V)
2 6
+
3
5
- 4 Exp_com

Switch Box Generator (4pin Input)

JRC-23F
2
+
6 Ready_com Ready_com
3
5 4
-
Exposure Exposure (+24V)

JRC-23F
Ready Ready (+24V)
2 6
+
3
5
- 4 Exp_com Exposure com

JRC-23F Contact Data


Initial Contact Resistance
100m
Max. (at 1A 6VDC)

Contact Material Silver Alloy (Au clad)


Contact Rating (Res. Load) 0.5A 125VAC / 1A 30VDC
Max. switching Voltage 125VAC / 60 VDC
Max. switching Current 2A
Max. switching Power 62.5 VA / 30W
DRT-MAN-049

0.5A 125VAC
UL/CUR Rating 1A 30 VAC
0.3A 60 VDC
Digital Radiography Technologies

2-5. Ground Connection


- Connect Ground Connection point on control box to a stable ground point inside the power distribution box
at the Installation site and make sure that the device is properly grounded.
- In order to solve problems caused by unstable power noise from the installation environment,
it is recommended that the earth cable is connected to a stable ground inside the main power distribution
box at the installation site since a more unstable power can be caused when it is connected to the ground
point of the Generator.
www.drtech.co.kr 25
26

2-6. Generator Time setting


2-6-1. Definition of Ready Delay
- Warming up time required for exposure examination after Ready input.
- The required time is different depending on the kind of Generator.
- The default Ready Delay of DRTECH is set at 2 seconds.
- Use of Drtech Terminal Emulator can be changed after checking the Warming up time of the Generator
model you wish to use.
- If a Generator's Warming up time is delayed for more than 2 seconds, the Ready delay time should be
adjusted using the DRTECH Terminal Emulator

2-6-2. How to change Ready Delay through


DRTECH Terminal Emulator
- Refer to DRTECH Terminal Emulator Manual for detailed instructions.

2-6-3. How to check Ready Delay Setting


- It can be checked through the time required for the Switch Box’s Lamp to change from green to red after
pressing the Hand switch’s “Ready” button.

DRT-MAN-049

Terminal Emulator (Ethernet)


Digital Radiography Technologies

2-6-4. Definition of Exposure Window time


- The maximum time the Detector can be exposed to X-ray.
- DRTECH’s default Exposure Window Time is set at 500 msec.
- Set it so that the time is greater than Generator’s Exposure Delay time + Maximum required exposure time
of the corresponding site.
ex> Generator Exposure delay (100msec+a) + Maximum required exposure time (450msec)
= Approximately 600msec

2-6-5. Exposure Window time (Maximum)


- The basic Model provided by DRTECH can have a maximum setting of 2sec.
- If more Window time is required, please contact DRTECH.

2-6-6. How to change Exposure Window time through


DRTECH Terminal Emulator
- Refer to the DRTECH Terminal Emulator Manual for detailed instructions.

2-6-7. How to check for Exposure Window time


- It can be checked through the time for the Switch Box's blue Lamp to turn on.

Terminal Emulator (Ethernet)


www.drtech.co.kr 27
28

2-7. Driver Installation


2-7-1. Ethernet Card Driver Installation
- Driver is installed automatically on most PCs. However, when driver is not automatically installed, install it
manually using the Ethernet Card driver file on the CD enclosed in the product.

2-7-2. Ethernet Card Setting


- Set the IP address as a 192.168.250.XXX and subnet mask as a 255.255.255.0
(IP address range : 1 XXX 254, excluding 200)

2-7-3. Precautions
- When using Ethernet Type, firewall must be disabled for stable operation.
- When security programs such as V3, Macfree, etc. are being used, there can be problems with image
capture function.

2-7-4. Verifying normal operation


- When System is turned on, check whether the Network status has changed.
- Window XP : ‘Unconnected’ is changed to ‘Connected’
- Window 7 : ‘Unconnected’ is changed to ‘Unidentified’ or ‘Connected’.

DRT-MAN-049
Digital Radiography Technologies

2-8. Calibration
2-8-1. MAP, GMP, PMP
- Copy MAP, GMP, PMP file on the enclosed MAP CD to the set folder in the corresponding DROC S/W.
- Correct the file name to the set file name in the corresponding DROC S/W or correct the ini file.

2-8-2. Caliview
Please refer to Caliview manual for details.(Manual file name : Caliview_Manual_Eng.pdf)
- Run setup program to install Caliview software on a computer
- Please refer to Chapter 3. Installation Procedure of Caliview manual.

2-8-3. Setting MAP, GMP, PMP


Check the configuration (Detector model, map file name etc..)
- Please refer to Chapter 4. Instruction of Caliview - Gain Calibration Procedure - Step 1 & Step 2 of
Caliview manual.
www.drtech.co.kr 29
30

2-8-4. Pixel Calibration


- Execute only in the event of Dead pixel or Line defect.
- Please refer to Chapter 4. Instruction of Caliview - Pixel Correction Procedure - Step 1 & Step 2 of
Caliview manual.

2-8-5. Gain Calibration


- Must be executed after product installation for adjusting to X-ray environment such as Heel effect.
- Please refer to Chapter 4. Instruction of Caliview - Gain Calibration Procedure - Step 4 of Caliview
manual.

2-8-6. Caution
When proceeding with Gain Calibration, it is recommended that Gain Calibration is executed
without the Grid or Bucky Cover
CAUΤΙΟΝ

- Refer to the Caliview software Manual

DRT-MAN-049
Digital Radiography Technologies

2-9. GRID Guide


2-9-1. Cases when GRID change is not required
- There must not be any physical damage (scratches, cracks) to the prior GRID being used.
- Moving must be done in parallel to FPD.
- Most importantly, the Grid Pattern must not show on the final X-ray Image when the Grid Filter is used.

A. Spring Type Grid


- Moving speed and status must be good.
- Grid Pattern must not show on the final X-ray when the Grid Filter is used.

B. Stepping Motor
- Moving speed and status must be good.
- Grid Pattern must not show on the final X-ray when the Grid Filter is used.

C. CAM Motor
- Moving speed and status must be good.
- Grid Pattern must not show on the final X-ray when the Grid Filter is used.

2-9-2. Cases when GRID change is necessary


To obtain better image quality, the use of Grid Spec being recommended for the Bucky type being used is
recommended. The use of FG14 and FG17 is recommended when Grid moving is poor or when Grid
Pattern is severe even after the use of Grid Filter. The Grid must be fixed when using FG14 or FG17.

A. When using Grid of Spring Type


- When Grid Pattern occurs even when Moving is good and Grid Filter is applied, change of Grid is
recommended.
- When Bucky does not function occasionally
- When Grid Line Resolution is too low (below 71 Lines/inch).
- When Moving is good : 71~103 Lines/inch may be used and 103 Lines/inch is recommended
- When Moving is poor : FG14 or FG17 is recommended

B. When using Grid of Stepping Motor Type


- When Grid Pattern occurs even when Moving is good and Grid Filter is applied, change of Grid is
recommended.
- When Moving does not function occasionally
- When Grid Line Resolution is too low (below 150 Lines/inch)
- When Moving is good : Grid with 200 or above Lines/inch is recommended
- When Moving is poor : FG14 or FG17 is recommended

C. When using Grid of CAM Motor Type


- Usually change is recommended when Grid does not move in parallel to FPD.
- First, set the Grid by executing Caliview during installation, then check whether Grid Pattern occurs by
getting an X-ray Image to decide if change is necessary.
- When Grid pattern occurs, change to FG14 or FG17.

D. Fixed Type
- Only FG14 and FG17 provided by DRTECH may be used.
- Regarding Filter file, there are current filter file on the C: Caliview Filter folder
www.drtech.co.kr 31
32

2-9-3. Grid Filter Setting Guide


When you have chosen to use the Grid moving method and Grid specification, you need to input the Grid
Filter information by Caliview or DRTECH.ini.

(1) Grid selection by Caliview


Click the Option Configuration

Click the ‘GRID’ section and input the Grid information

DRT-MAN-049
Digital Radiography Technologies

If you use the ‘Spring moving bucky’ or ‘Motor moving bucky’, you should input the Lines/inch and Angle.
Grid ratio and SID is not necessary

(2) Grid selection by DRTECH.ini


A. find the ‘DRTECH.ini’ file
- Press the window key and ‘R’ key simultaneously

- Type ‘ %AppData% DRTECH ’ and click ‘OK’ button


www.drtech.co.kr 33
34

B. Execute the ‘DRTECH.ini’ file

C. Edit for Grid information. Refer to ‘Appendix for DRTECH.ini’

DRT-MAN-049
Digital Radiography Technologies

- Parameters

Default
Session Item Range of input value Description
Value

0 = No grid
Specify the type of grid.
1 = Spring Bucky
Parameters are used for mathematical
2 = Fixed Grid(19 degree tilting)
GRID TYPE 0 operation to remove grid patterns.
3 = Motor Bucky
Motor Bucky include the CAM and Stepping
4 = FG14(Fixed Grid for 1417 size)
Motor Bucky
5 = FG17(Fixed Grid for 1717 size)

Set the value of line pair / inch.


(LP/cm must be converted to LP/in by using
calculator. 1 cm = 0.393701 inch)
GRID DENSITY 0 ~ 999 0 Parameter is used for mathematical
operation to remove grid patterns
But FG14 and FG17 do not use this value.
(It doesn’t matter what this value is)

Set the value of tilting angle of grid.


Parameter is used for mathematical
operation to remove grid patterns
GRID TILTING But FG14 and FG17 do not use this value.
-360 ~ 360 0
ANGLE (It doesn’t matter what this value is)
Grid angle value must be set to
a negative value
Ex) 19 degree = -19

Set this value 1 to save the pre - processed


captured - image at C: \ folder with the date -
0 = Raw image will not be saved. time based file name.
SAVE RAW IMAGE 0
1 = Raw image will be save. This option is needed to be checked for
an accurate analysis of acquired image.

[PANEL 1] Set the maximum value of pixel.


or Each detector models uses different
[PANEL 2] default value
8500
THRESHOLD 0 ~ 16383 or 10000 Model Default Value
or 11000 F500, F560 8,500
F750E, F760 10,000
For the rest 11,000

PANELMAP DLL File


Panel map file path Set the value of file path for panel map.
FILENAME Location

Set the value of range for left side edge-cut.


Each detector models uses different default
value
20 Model Default Value
EDGE RANGE 0 ~ 1280 or FDXD810,
10 20
LEFT F330, F330N
For the rest 10
Please have a prior consultation with
DRTECH before changing this value.

Set the value of range for right side edge


cut.
Each detector models uses different default
value
20 Model Default Value
EDGE RANGE
0 ~ 1280 or
RIGHT FDXD810,
10 20
F330, F330N
For the rest 10
Please have a prior consultation with
DRTECH before changing this value.
www.drtech.co.kr 35
36

Default
Session Item Range of input value Description
Value

Set the value of range for upper side


edge-cut.
Each detector models uses different default
value
20 Model Default Value
EDGE RANGE
0 ~ 1536 or
TOP FDXD810,
10 20
F330, F330N
For the rest 10
Please have a prior consultation with
DRTECH before changing this value.

[PANEL 1] Set the value of range for lower side


or edge-cut.
[PANEL 2] Each detector models uses different default
value
20 Model Default Value
EDGE RANGE
0 ~ 1536 or
BOTTOM FDXD810,
10 20
F330, F330N
For the rest 10
Please have a prior consultation with
DRTECH before changing this value.

Set the value of pixel range for edge-cut.


EDGE VALUE 0 ~ 16383 11,000 Please have a prior consultation with
DRTECH before changing this value.

Set this value 1 to enable GPU


acceleration in case of using NVIDIA’s
0 = Enable CUDA acceleration. CUDA compatible graphic card.
ACC MODE 0
1 = Disable CUDA acceleration. Please visit the following website for
checking CUDA compatible card list.
[API] http://developer.nvidia.com/cuda-gpus

Set Set this value 1 to save log file which


0 = Log file will not be saved. records the API’s error or operating status
SAVE LOG 1
1 = Log file will be saved. Log file will be saved as below.
%APPDATA%\DRTECH\LOG\Log.txt

- If the value of the parameters is set wrong, serious problem can occur in the image.
Please be careful not to make mistakes.
CAUΤΙΟΝ - In case of using earlier version of API, INI file location can be “C: \ DRTECH.INI”
DRT-MAN-049
Digital Radiography Technologies

2-9-4. How to change Tray form Spring GRID


- Grid size should be manufactured to fit the existing Grid Tray
- Add fixing form that fits the existing Grid thickness
- Request DRTECH for special manufacturing when the Grid size isn't normal.

2-9-5. How to change Screw form Spring GRID


- Grid size should be manufactured to fit the existing Grid Tray
- Drill holes that fit the existing Grid's holes and fix them in place with bolts.
www.drtech.co.kr 37
38

2-9-6. How to install Fixed type Grid change


Install by matching the Label direction on the Grid to the installation direction of F760 detector.

Model
Angle FLAATZ-500 FLAATZ-560

0 Degree

90 Degree
(The Spring,
Motor bucky
will be
update later.)

Model
Angle FLAATZ-750E FLAATZ-760

0 Degree

FG17 supports 0 degree only. Please be cautious of the FG17 orientation


CAUΤΙΟΝ
DRT-MAN-049
Digital Radiography Technologies

2-9-7. Grid and Detector alignment


- The grid should be perfectly aligned with the detector.
- If the grid and detector is out of alignment and the deflection angle is over 2 degrees, the grid line may
appear on the X-ray images.

Proper alignment Improper alignment

The grid should be parallel with the Detector

If the grid is not parallel with the Detector and the gap is over 2cm, the grid line may appear
on the X-ray Images.
CAUΤΙΟΝ

ȭGrid

ȭDetector

Properly Paralleled

When the gap between the GRID and the Detector are different (Deviation should be less than 2 cm)

ȭGrid

ȭDetector

Improperly Paralleled
www.drtech.co.kr 39
40

2-10. F560 & F760 insert to Bucky


2-10-1. Wall Bucky
As the Image Plate and thickness are equal, insert by matching the existing Bucky.
Take caution not to scratch or damage the cable or Interface Cable of the Detector when inserting.
Use a Cable Tie to organize the cables as poor movement may occur when the cables are severely bent.

2-10-2. Table Bucky


As the Image Plate and thickness are equal, insert by matching the existing Bucky.
Take caution so that the Bucky does not scratch or damage the cable or Interface Cable of the Detector
when inserting.
Use a Cable Tie to organize the cables as poor movement may occur when the cables are severely bent. DRT-MAN-049
Digital Radiography Technologies

CHAPTER 3
DROC Operation

3-1 PACS, Dicom Printer 42


3-2 Monitor Setting Guide 43
www.drtech.co.kr 41
42

3-1. PACS, Dicom Printer


3-1-1. PACS, Dicom Printer Setting
- Set PACS Server and Worklist Server in accordance with DROC S/W in use.
- For more details, refer to relevant DROC Software manual.

3-1-2. Network Priority


- If you often have the disconnection with PACS or Dicom Printer, please try to change the priority of
Network card and don't use the HUB.

3-1-3. For more detail, refer each DROC Software manual.

DRT-MAN-049
Digital Radiography Technologies

3-2. Monitor Setting Guide


3-2-1. Adjusting and using monitors
This section explains the adjusting and use of monitor through an example (DELL).
The basic setting method may be different according to each manufacturer.

Basic setting method of monitors


The base to optimizing a monitor is in the positioning of the monitor.
It is ideal for the monitor to be looking down. Install the monitor by using the tilt of the screen or the chair so
that the bottom of the screen is at eye level.

Front Panel Controls

1 2 3 4
Power On / Off
Brightness, Contrast Automatic-Adjusting
Menu Button Button
/ Reduce (-) Button / Increase (+) Button
with marks, etc.

When the menu is off, press the button to display


the ‘brightness’ and ‘contrast’ menu.
After adjusting the ‘Brightness’, adjust ‘Contrast’ when
further adjustment is needed.
‘Brightness’ function adjusts the brightness of the
screen.
‘Contrast’ function adjusts the degree of light and dark
of the display screen.
“+” increases the ‘Brightness’ or ‘Contrast’.
“- ” reduces the ‘Brightness’ or ‘Contrast’.
www.drtech.co.kr 43
44

3-2-2. Setting optimal resolution


This section explains the setting of optimal resolution by example (DELL).
The basic setting method may be different according to each manufacturer.

How to set optimal resolution


1. Click on the right mouse button on the desktop and select Properties.
2. Click the Settings tab.
3. Set the screen resolution to the setting recommended by DROC S/W.
4. Click OK.

3-2-3. Setting medical imaging using OSD Manual


Accessing menu system
1. When the menu is turned off, select the 'Menu' button to open the OSD system and display the main
functions.

2. Press - and + button to navigate down to the Color Settings menu icon.
3. Select the appropriate Red Preset for programs that generally require a lot of colors (medical imaging,
photo editing, etc.).

When the Red Preset doesn’t satisfy your requirements, the ‘User Preset Setting’ can be
used to independently increase or reduce the colors (R, G, B).
DRT-MAN-049

NOTE
www.drtech.co.kr 45
46

3-2-4. Configuration methods for different


medical image monitors
This section explains how you can create same settings on different monitors.

Why create same settings on different monitors


After obtaining the initial set of images, they are modified, edited, and transmitted to the picture archiving
and communication system (PACS). Since these images are displayed on different media, this may have
negative effects on accurate diagnosis.

Configuration methods for different monitors


1. Diagnostic monitor (resolution: 3 megapixels or above) - basic monitor
Since diagnostic monitors use their own periodical correction functions, no modification will be required. For
this reason, the brightness, contrast and color of the monitor that has first obtained the images should be
modified in accordance with the diagnostic monitor.

The basic procedure


Modification Verification Verification

Acquiring
PACS Viewer Monitor in the
initial images by Diagnostic monitor
Monitor doctor’s office
the first monitor

After obtaining and transmitting the initial set of images, follow the phased verification
process and check the monitor in the doctor’s office. Then, if the images need
modifying, re-verify them by adjusting their brightness and contrast levels. If the
NOTE
images need to be further modified, it is also desirable to use the monitor’s OSD menu.

2. Ordinary low-cost monitors


It is best to select a standard monitor since various LCD monitors made by different manufacturers are
being used. It is recommended that the images from the monitor first obtaining these images are transmitted
to monitors used as the PACS viewers, and then the differences between the images obtained by all the
monitors should be compared and modified.

The basic procedure


Modification Verification

Acquiring
PACS Viewer Monitor in the
initial images by
Monitor doctor’s offic
the first monitor

After obtaining the initial set of images, transmit them to the PACS viewer’s monitor,
compare the transmitted images. Then, make necessary modification on them.
NOTE

If you have created the same settings between the first monitor, diagnostic monitor and the PACS viewer
monitor but they display images with different qualities, please request a calibration to your PACS manufacturer
Digital Radiography Technologies