GP-200V

PORTABLE HYDROGEN PURITY


ANALYSER

MODEL: GP-200V

Operating Instruction Manual

ENDEE ENGINEERS PVT LTD.


D-122, Ansa Industrial Estate,
Saki Vihar Road, Mumbai – 400 072.

Tel.: 91-22-2847 1776, 2847 0075


Fax:- 91-22-2847 1615

E-mail: sales@endee-engineers.com
works@endee-engineers.com

Website: www.endee-engineers.com

1/13
GP-200V

Endee Engineers Private Limited

Specialists in design,
Development and Manufacturing of
Advanced High Quality
Gas Monitors and Analyzing Instruments.

Dear Customer

Thanks and congratulations to you on the


purchase of this instrument. You are
strongly advised to read this manual
carefully before using the instrument. It is
highly desirable that you are also familiar
with the process being measured.

Endee Engineers Pvt Ltd. shall not be


responsible for any injury to any personnel
and or disturbance or damage caused to
the process and plant due to mishandling
or improper use of the instrument.

2/13
GP-200V

FRONT VIEW OF GP-200V

CHARGER

3/13
GP-200V

Contents

1. Introduction ............................................................................................................ 5
2. Principle of Operation ............................................................................................ 5
3. Specifications.......................................................................................................... 6
4. Know your instrument ............................................................................................ 7
5. Description of various parts and controls ............................................................... 8
6. Some features of the display and control ................................................................ 9
7. How to operate...................................................................................................... 10
8. Functions: ............................................................................................................. 11

4/13
GP-200V

1. Introduction
MODEL GP-200V is a battery-operated portable gas analyzer. It has
been especially designed for analyzing the processes gas Hydrogen. It
gives the concentration as of Hydrogen in %V/V.

Warm up time is 5 minutes

Being a state-of-art microcontroller based instrument, MODEL GP-200V


has in-built intelligence that makes it very easy to use and operate.
Calibration of instrument Zero is achieved with fresh air by opening the
instrument with the help of potentiometer. For span calibration span gas
is required. It also provides an interactive user machine interface which
includes user prompts.

MODEL GP-200V also has in-built battery condition monitoring


function, which gives advance warning of the impending battery-run
down condition. It also includes provision for interfacing with a battery
charger.

2. Principle of Operation
The analyzer is based on a transducer whose signal increases with
hydrogen concentration. The instrument includes flow meter for fixed
flow.

5/13
GP-200V

3. Specifications
Instrument Type : Portable gas Analyzer for Hydrogen

Model Number : GP-200V

Measurement
Measuring Range : 0 – 100 % V/V

Resolution : 0.1%v/v

Measuring Accuracy :  2 % FS

Response Time : Maximum 30 seconds for 95% Variation

Display:
Display Type : 8 Characters x 1 Line Alphanumeric LCD

Sensor:
Sensor Type : Thermal Conductivity Sensor

Power:
Operating Power Supply: 6 V D.C. (Rechargeable Battery pack)
With Battery Charger
Environmental :
Working Temperature : -5C To +55C
Storage Temperature : -20C To +70C

Physical:
Construction:
Dimensions : 100mm (h) * 170mm (w) * 130mm (d)
Weight : 2000 gm Approximately

Accessories:
1) Carrying case
2) Batteries (Installed in the instrument)
3) Operation Manual.
4) Battery Charger.
5) Sampling System

6/13
GP-200V

4. Know your instrument


Given below is a sketch of the instrument showing the layout of various
parts and operating controls.

HYDROGEN PURITY ANALYsER


On/OFF
S/W

OUT LET
GAS INLET

FLOW METER

CHARGER
FN EN

ENDEE GP-200V

7/13
GP-200V

5. Description of various parts and controls

5.1. Display:

It is an 8 Character x 1 line Alphanumeric LCD display which serves as the


readout for the measurements. The default display shows Hydrogen
concentration in % v/v,. It also serves as the visual display terminal for
operator interface for configuring various settings and parameters of the
instrument.

5.2. Key: On/off Switch

It is a slide switch, which turns On the power to the instrument. It is a slide


switch.

5.3. Key “FN”

This key is a push type key switch, which facilitates selection and carrying
out various functions of the instrument. Successive presses of the key select
the next function from a circular list of functions.

5.4. Key “EN ”

This key has functions to enter the data as well as increment key during
setting.

8/13
GP-200V

6. Some features of the display and control


6.1. The key switches provide tactile feedback.

6.2. In order to guard against inadvertent pressing of a key, key press is


accepted only if the same has been kept pressed for a definite time.

6.3. Acceptance of any key press for carrying out the subject function is
accompanied by an audio feed back in the form of a beep.

6.4. POWER ON Self Test


Every time the instrument is switched, it carries out Power On Self
Test (Post), where in it checks for the proper operation of its various
measuring circuits and display and control subsystems. When POST
is in progress, the display shows “Endee” 10 times. It is followed by
displaying of Model number of the instrument, again 10 times. Each
display of “Endee” and Model number is accompanied by an audible
beep.

The instrument is not ready for use for measurements until the POST
has been completed. When the POST is completed successfully and
no malfunctions or faults are observed, the instrument is ready for use
to carry out the measurements. This is indicated by “hydrogen nn.n”
on the display. Here “nn.n” is the instantaneous value of hydrogen
concentration.

6.5. The battery condition is continuously monitored. At any time during


the operation, if battery voltage falls below a set limit, the display will
show “LOWBATT.
In this condition, it is advisable to immediately charge the battery
with proper charger supplied. If the instrument is continued to be
used under “LOWBATT” condition, there is no guarantee that the
measurements will be reliable. If the battery is continued to be
used and drained beyond certain specified limits of the
manufacturer, and if an attempt is made to charge the battery in
this condition, the battery would be permanently damage with
consequences.

9/13
GP-200V

7. How to operate
7.1. Preparation of the Instrument:

Before putting the instrument in use, please ensure that batteries are fully
charged.

7.2. Using the Instrument:

After preparing the instrument as described in 1 above, switch the


instrument on by sliding on/off switch.

Wait for the instrument to complete the POST described in previous chapter.
If the default factory settings are acceptable, the instrument can be directly
used for actual measurements.

When the instrument is carrying out measurements, it is said to be in


“Measuring mode”.

Display Hydrogen
00.0 %v/v

When the function key is pressed, it goes into function mode. In case, the
settings need to be changed, use the Key “EN” in conjunction with the Key
‘FN” key to select the desired function by successively pressing the key till
the desired function is displayed on LCD display.

In the following section, the number of key presses mentioned will be


referred from measuring mode.

10/13
GP-200V

8. Functions:
The various functions provided by the instrument are as follows:

 AUTOZERO
 AUTOSPAN
 RST-CAL

8.1 AUTOZERO

1. The display will show “XXX %V/V”


2. Then press ”FN”
3. Display shall be “FN/EN”
4. To return back to measure display press ”EN”
5. For “CALMODE” press “FN” once
6. The display will now “CALMODE” Blinking
7. To select auto ZERO press “EN”
8. Display shall be “AUTOZERO” Blinking.
9. If you press ”FN” It will switch over to AUTOSPAN
10. For “AUTOZERO “ Press “EN”
11. Now keep the instrument in clean air as it takes clean as
the sample gas for 79.1% and wait for two minutes.
12. Initially “auto ZERO” display shall be steady till auto
ZERO completes, after completion display will show
“AUTOCAL”
13 This is for ZERO calibration.

NOTE: This action shall be completed in clean atmosphere because


here the fresh air is a sample gas

11/13
GP-200V

8.2 AUTOCAL (SPAN)

1. After carrying out AUTOSPAN,


2. Press key “FN”
3. The display now will show blinking “AUTOCAL”.
4. To return back to measure display press ”EN”
5. For “AUTOCAL” press “FN”
6. The display will now “AUTOCAL” Blinking
7. To select auto CAL press “EN”
8. Now Connect the calibration gas (sample gas of 99.9%) at
inlet of instrument with the proper tubing etc. and wait for
two-three minutes.
9. Press Key “EN”
10 After completion of auto cal instrument will show ”auto cal”
means auto cal is completed
11. The completion or aborting of span calibration will be
indicated by the display showing blinking “RST-CAL

8.3 RESET CAL: The instrument has been provided with this function to
accept or reject (suspect) calibration values.

1. While display is showing “RST-CAL”, press


2. “ENTER“. The display will now show “H2”.
3. The display will now “RST-CAL” Blinking

12/13
GP-200V

13/13