C020100 Model LPD • C020200 Model MSD

Intended Use . Unpacking .


This equipment is a dental amalgamator designed for mixing 1. Carefully remove unit from packing material.
dental amalgam. 2. Remove the red packing screws from bottom of the unit. Save the screws;
should it ever become necessary to transport your WIG-L-BUG, the screws
must be reinstalled to prevent damage in shipping.
Technical Specifications .
Speeds: 3800 rpm-4800 rpm.
Capacity: Designed to use amalgam capsules with lengths
between 1.250⬙ and 1.350⬙.
Power Requirements: 115v ±10%, 0.5A, AC, 60 Hz
Fuse Rating: 2 amps, 250 volts

Operation • Model MSD .


1 Plug into grounded outlet. 䢇
5 Load capsule. Insert one end of capsule 䢇
6 Close safety cover H.
into holding arm, then push other end

2 Turn on power switch A. down to lock in place. Use only standard 䢇
7 Press F to start. Unit will automatically
stop when preset time expires. To stop at

3 Set time: Press E and adjust with B & C. amalgam capsules.
any time, press G.
Available range is from 0 to 30 seconds.


4 Set speed: Press D and adjust with B & C. 䢇
8 Turn off unit when not in use.
Four speeds are available: 3800, 4200,
4600 and 4800 RPM.

H 6

D 4
B

C E 3

7 F G
2
A
8
Operation • Model LPD .


1 Plug into grounded outlet. 䢇
5 Load capsule. Insert one end of capsule Use only standard amalgam capsules. The
into holding arm, then push other end WIG-L-BUG LPD does not accept Fuji capsules

2 Turn on power switch A. down to lock in place. or any capsule with a tail or pouring spout.

3 Set time: Press E and adjust with B & C. 䢇
6 Close safety cover H.
Available range is from 0 to 30 seconds.

7 Press F to start. Unit will automatically

4 Set speed: Press D and adjust with B & C. stop when preset time expires. To stop at
Four speeds are available: 3800, 4200, any time, press G.
4600 and 4800 RPM.

8 Turn off unit when not in use.

H 6
5

2
A
8

D 4 B

E 3 C
7 F G

Environmental Conditions . Maintenance . Caution .


For indoor use only, and in temperatures of Unplug from power before cleaning. Use of this equipment in a manner other than
41°F-104°F (5°C-40°C). Clean with a soft, damp cloth. specified by these instructions may impair
Altitude up to 2000 M. Do not use an abrasive cleaner. protection provided by the equipment.
Maximum relative humidity: Contains no user-serviceable parts.
80% for temperatures up to 31°C, Warning .
50% for temperatures up to 40°C.
Warranty . ! This equipment is not designed for use in
Overvoltage Category II hazardous atmospheres or with hazardous
Pollution Degree II The WIG-L-BUG is warranted for parts and materials.
labor for one year after proof of purchase is ! Mercury is a health hazard. Tighten
Ground protective conductor terminal established. capsule completely to prevent mercury
escaping during trituration. Immediately
replace any capsule that becomes worn or
loose-fitting. Close safety cover during use.

LISTED
LABORATORY EQUIPMENT
44TM

A Division of DENTSPLY International, Inc. Customer Service: 800 323 0970


1212 Abbott Drive International: 847 742 1140
Elgin, Illinois 60123-1819 FAX (24 hours a day): 800 544 0787
www.rinncorp.com International: 847 742 1122
847 742 1115
©2006 DENTSPLY Rinn Form 5558 Rev. D 12.06