Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Geovista

Sonde Operations Manual

Magnetic Susceptibility Sonde


Magnetic Susceptibility Sonde

Specifications

Recommended mnemonic MSUS

Sonde measurement parameters

Parameter Range Resolution


MSUS 10-5 - 0.5 S.I. units 1 Count

Sonde Dimensions

Sonde length 1.05 m


Sonde diameter 45mm
Sonde weight 5.0 Kg

Communications Parameters

Sonde Protocol Type Geovista

Channel Depth Recommended Recommended Number Timebase Counter Platinum


Offset Mnemonic Unit of bytes channel based Correction
channel
MSUS 33 cm MSUS mSI 3 No No No

Principle Of Operation

This probe is designed to measure the magnetic susceptibility of the formations crossed
by the borehole. It is mainly used for exploration for deposits of Fe minerals – magnetite,
pyrrhotite and haematite. It can identify formations containing magnetite from 0.005 to
100 % concentration. The sonde includes one transmitter coil operating at 2 kHz
frequency, and a receiver coil with 25 cm interspacing.
A calibration jig with two detachable calibration units (0.6 x10-3 & 5 x 10-3 SI units) is
supplied witht he sonde. Follow the standard channel calibration
procedure to calibrate the sonde in the desired range using either of
the 2 units. The calibration jig must be positioned at the measure
point point over the sonde housing, the measure point is marked with
an orange dot on the body of the sonde.

Please note: During calibration, the sonde must be kept well away from any conductive
body.

Channel Description

Channel 1 is Magnetic Susceptibility (calibrated in S.I. units)