Gujarat State Electricity Corporation Limited

Wanakbori Thermal Power Station


TA-Galteshwar Dist. - Kheda 388239

TECHNICAL BID
Reference No. : WP/S/TECH/RFQ-47845E/CHEMICAL LAB /
SUBJECT: - Supply of Multi Parameter Analyzer in ppb level at Chemical Laboratory,
Wanakbori TPS.

Sr. No. Description of Material HSNCODE GST Rate Qty. Reqd Unit
With Cess
If Any IN%
1 MULTI PARAMETER 1 Number
ANALYZER (UP TO PPB) IN
BOILER WATER/FEED
WATER/STEAM/OR ANY KIND
OF WATER SAMPLE.
2 AUTOTITRATOR FOR A) 1 Number
ALUMINIUM ANALYSIS; B)
HARDNESS-CA,MG
ANALYSIS; C) CHLORIDE
ANALYSIS-AS PARALLEL SET
UP

Technical Specification:

ITEM NO.-1 Multi Parameter Analyzer, analysis of Boiler water, Feed water, Steam or any kind of
water sample & result in ppb level.

Dual channel Ion Chromatography system for analysis of Anions : fluoride, chloride, nitrite, nitrate,
phosphate, sulphite, sulphate specification of arsenic, selenium by suppressed conductivity detection
using carbonate / bicarbonate eluent chemistry.

Cations: lithium, sodium, potassium, calcium, total hardness by conductivity detection. Should be able
to use Organic modifiers ( at least up to 50% ) like acetone, acetonitrile and methanol in the eluent. The
system should be able to perform both suppressed and non-suppressed conductivity applications.
System should perform heavy Metal and its specification at ppb - ppm level (As, Pb, Cd, Cr, Co, Fe,
Cu,Ni, pt, Hg, Zn etc) by using electrochemical detector.
The system should be PC based with data acquisition and system control through the same software.
The software should be able to identify columns pump , detector automatically. The system should
have the components with following technical specifications.

*PUMPS:
1) Two numbers of high pressure pumps of serial dual piston type (upgradeable to ternary
compositional gradient) with selectable 0.001 to 15mL / min flow rate with a flow reproducibility of +/-
0.1%.

Page 1 of 5
Bidder's Signature
With date and seal RFQ No: 47845E
Pump should have the following specifications:
- Serial dual pistons with two inert check valves.
- Resolution of flow rate : 0.001mL
- Pulsation : Lower than 1%
- Pressure range : 0 - 5000 PSI
- Should be able to perform high pressure binary gradient applications.

*Conductivity Detector:
1)Two numbers of Conductivity detector for analysis of anion and cation which is micro processor
based with a Thermostated micro-flowcell conductivity block with an accuracy of +/- 0.0010 degree C.
The user should be able to set temperature of the conductivity block between 20 degree C to 50 degree
C.
2)Conductivity measurement range: 0 to 15000 microS/ cm.
3)Electronic noise less than 0.1nS at 1 nS/cm
4)Temperature coefficient range: 0 - 5%

*Electrochemical Detector :
- Voltage Range: +/- 25 V
- Sweep potential range: +/- 5 V
- Current Range: +/- 224 mA
- Resolution of measured current (in the smallest measuring range): 6 fA

Mercury electrodes and system should be fully compatible with rotating disc electrode facility and
various solid-state electrodes like Gold, Silver, etc. Instrument should have measurement technique like
Sampled DC , Differential Pulse (DP) , Direct Current (DC), etc. It should be possible to do both
standard addition and Calibration technique.
Measuring cell for higher volume of samples (10 - 70mL) Stirrer with the speed setting of 200 - 3000
revolution per minute (Speed Stability +/- 5%)

*Column Housing :
Column housing should be able to identify the columns and set the optimal operating conditions for
column operations.

*Chemical Suppressor :
Chemical suppressor should have high back-pressure (at least 350 psi) tolerance with continuous
regeneration. The suppressor should be able to take flow rate up to 10 mL/min. The regeneration of the
suppressor should be by external chemical regeneration mode with the possibility of controlling the
regenerant flow rate. At least
8 year warranty should be provided along with manufacturers certificate indicating the warranty.
Suppressor should have 100% solvent compatibility.

*Injector :
Two numbers of dual position 6-Port injector valve with fast response time and controlled through
software should be quoted.

*IC Columns :
IC columns for analyses of anions, cations , transition metals, phenols should be individually quoted
with respective guard columns. The columns should have electronic chip to store data and history of
column use. It should also be possible to store the data , number of samples injected, working hours
etc.

Page 2 of 5
Bidder's Signature
With date and seal RFQ No: 47845E
Automation :
Auto sampler with minimum 55 position should be quoted along with the vials of minimum of 10mL
capacity and caps for anions , cations , transition / heavy metal analysis. The auto sampler should be
able to take sample volumes from 0.5 - 10mL samples. The complete flow path including the sample
needle should be non-metallic. The auto sampler should be able to communicate with system and
software. Auto sampler with combination of liquid handling device should be able to plot calibration
curve with single standard - inline calibration. No valves should be used for inline calibration.

The software should be able to control the sample processor completely.

*Data Processor :
The system should be PC based with data acquisition and system control through the same software.

Others :
(i)One year warranty on electronic and mechanical components is mandatory.
(ii)Calibration Standards (mixed) for the standard ions shall be provided.
(iii)The complete set up should be provided by single supplier
(iv)Should be supplied with Ultrapure water purification system - DM water to type 1.

ITEM NO.- 2 Specification for Boiler water & oil tester.


Potentiometric Automatic Titrator for Alkalinity, Hardness, TAN, TBN, in Lube Oil & Transformer Oil:

1) The system should be capable of performing all types of titration like Acid-base, non-aqueous, redox,
precipitation etc. It should also have the capability of performing volumetric Karl Fischer titration.
2) The system should be completely software controlled and should have a galvanically isolated
measuring interface with measuring cycle of 100 ms for all measuring modes. It should also have the
facility to upgrade with additional measuring interface for performing parallel titrations.
3)The instrument should have the following measurement range
a) pH : 0.0 to +14.0 pH with Resolution of 0.001 pH or better
b) mV : +/- 1200 with Resolution of 0.1 mV or better
c) Ipol : +/- 120 microA or better
d) Vpol : +/- 1200 mV or better
4) As a security measure the basic instrument and all the peripheral devices connected to it should be
recognized automatically by the software as soon as it is connected to the software.
5) The system should have the possibility to connect up to Four Titrating burettes for different types of
titrations.
6) Burette should have a resolution of 1/10000 steps or better of its burette volume and the flexibility to
handle more than one solvent for liquid handling.
7) The Burette drive should be mounted on top of the Burette so that the electronics in the burette drive
is not damaged due to reagent spill or leak. The Burette piston should move from top to the bottom of
the burette, so that no problems due to the air bubble present in the titrant.
8) Preparation of burette with new titrant and emptying the burette, with a single command should be
possible.
9) Burettes volume of 2, 5, 10, 20 and 50 ml should be available if required.
10) Dosing drives with chip on board technology with the capability of read and write should be
provided. It should detect the burette volume once connected. The chip on the burette should store
the important data like titrant name, concentration, date of preparation, expiry date, etc.
11) The instrument should have the facility to connect intelligent sensors with chip on board technology
to capture all data related to calibration and performance of sensors.
12) The instrument should also have the facility to perform pH measurements with the capability of 9
points calibration. Suitable electrode should be provided with the system.

Page 3 of 5
Bidder's Signature
With date and seal RFQ No: 47845E
13) The instrument should also have the facility to perform ion measurements with suitable ion selective
electrodes.
14) A suitable magnetic stirrer with control on stirring speed and direction should be provided.
15) Electrodes and accessories for the analysis of Alkalinity, Hardness, TAN, TBN, in Lube Oil &
Transformer Oil should be provided along with the system.
16) Auto sampler for existing system need to be quoted for anion & cation analysis, should have 55
position and flexibility.

Note:-
1) Solvent filtration kit,Calibration Standrad (Mixed) for Ion & all other standard will be provided by
Company.
2) For system needs, PC & Printer will be provided by company.
3) Price: FOR WTPS.
4) Bill payment will be made after completion of work satisfactory, as per corporation rules.
5) The instruments will be installed & Commissioned at intender's lab & all features of the instrument
are to be demonstrated through operation by supplier/ their authorized agent at our site along with
training to our personnel for correct operation at free of cost.

BQR:-
1) Bidder / Agency must have minimum 5 yrs. experience of supply of similar instruments in water
treatment plant of semi government / government organization or any thermal power station.
2) Bidder has to provide manufacture Guarantee / Warrantee certificate with instrument.
3) Bidder OEM/OES should have fully quipped technical support office/laboratory for facility of
calibration, servicing & reparing of instrument in Gujarat.

For & on behalf of GSECL,

Chief Engineer (C&O)


GSECL, WTPS

Page 4 of 5
Bidder's Signature
With date and seal RFQ No: 47845E
DECLARATION FORM
ON LETTER HEAD OF SUPPLIER
SUB:

Tender No/ RFQ No:

1. I/We hereby declare that I/We have carefully&thoroughly studied the entire tender placed on the
WEB site and General terms& conditions, technical specification, PQR and other documents of
the tender mentioned under reference along with the Commercial terms and conditions.
2. I/We hereby confirm and certify that the offer submitted by me/us is strictly in accordance with the
GSECL’s Tender specifications, Guaranteed Technical Particulars and drawing as mentioned in
the Tender Specifications. There is no commercial or Technical deviation (except the deviations
shown in respective Annexure 12 of Technical Deviations of this tender document) in the offer
from the Board’s Tender Specification. I undertake to abide by the GSECL’s Technical
specification / Guaranteed Technical Particulars / Drawing, I undertake to supply materials strictly
as per the GSECL’s Technical specification / Guaranteed Technical Particulars / Drawing, even if
any technical deviations are mentioned by me. I also undertake to abide by all commercial
conditions of the GSECL, including delivery schedule. I have submitted all the annexure 1to
15,form of integrity pact etc along with the technical Bid dully filled alongwith all the details”
3. I / We, hereby declare that our Technical Bid is strictly in line with the Tender specifications
(except the deviations shown in Annexure12 of Technical Deviations of this tender document).
4. I / We also agree that additional conditions / deviations, if any, found in the Commercial terms &
conditions (except the deviations shown in Annexure12 of Technical Deviations of this tender
document) our offer shall be out rightly rejected without assigning any reason thereof.
5. I/We hereby abide to execute the contract agreement&submitted the same alongwith the
acceptance of LOI and the same will be acceptable to me/us.
6. I/We hereby accept and confirm that any dispute on this regard shall not be entertained by the
tender inviting authority.
7. The full value of the “Earnest Money Deposit” paid herewith should be absolutely forfeited to the
Corporation, should I/ we do not deposit the specified amount of specified Security Deposit within
10 Days from receipt of the Letter of Intent.
8. If, I fail to submit all required documents with application or tender copy ( in each tender ) then my
tender will be cancelled, which is binding to me.
9. I/We ________________________________________________________________ authorized
signatory of M/s ______________________________________________ here by certify that M/s
_________________________________________________ and their proprietor / any partner /
any directors of the firm is not stop deal and / or banned for business dealing and / or black listed
by GUVNL and / or their any subsidiary company viz. GSECL / GETCO / DGVCL / MGVCL /
UGVCL / PGVCL as well as Government and / or Semi Government company / department.
10. We here by confirm that we havenot stake in one or more other agencies / bidders who are
participating in same tender and if it is found that the submitted information / declaration are
wrong in the tender / enquiry as stated in tender terms , than we will be disqualified automatically
&it will be accepted to us. EMD / SD will be forfeited. More over if GSECL shall take stern action
including stop dealing with our firms which is also accepted to us

Seal of the Firm Signature of the Tenderer


Place:With Designation.
Date

Page 5 of 5