Professional Documents
Culture Documents
Atomic Absorption Spectrometry-A Review: RE/, I F/" ,/RTI (,LE
Atomic Absorption Spectrometry-A Review: RE/, I F/" ,/RTI (,LE
Atomic Absorption Spectrometry-A Review: RE/, I F/" ,/RTI (,LE
2, July-Dec 2010
Keywords: Atomic Absorption Spectroscopy; Metals; independent of the wavelength of absorption and the
AAS; Absorption. temperature of the atoms.
Members Copy, Not for Commercial Sale
www.IndianJournals.com
T= I 110
%T= 100xIllo
A=IOO-T
%A= 100-% CONC.ENT~ATJO N
~
There is another term "absorbance" which is purely a
mathematical quantity. Fig 2: Graphical Representation of Beer's Law.
A=log(U1o)
INSTRUMENTATION
Absorbance is the most used term for characterizing light
There are two basic types of atomic absorption
absorption in absorption spectrophotometry, as this
instruments, one is single-beam and other is double-beam.
quantity follows a linear relationship with concentration.
Schematic diagram of single beam atomic absorption
As per Beer's law defines as
spectrophotometer and double-beam atomic absorption
A=abc spectrophotometer are given in Figure 3 and Figure 4
respectively. Atomic Absorption spectroscopy instrument
Where generally contains the following basic components3-4.
A = absorbance, 1. The light source of the element of interest.
2. Chopper to break the steady light from the lamp into
pulsating light.
0-----'
~O\l't..(.t of:
3. An "absorption cell" in which atoms of the sample are
produced (flame, graphite furnace, MRS cell).
Ll.u. .. '\' 4. A monochromator for light dispersion.
5. A detector for measurement of light intensity.
6. A display that shows the reading.
Fig 1 : Atomic Absorption Process.
65
Journal of Forensic Medicine & Toxicology Vol. 27 No.2, July-Dec 2010
maximum sensitivity, it is necessary to use a line source, applied, the coupled energy will vaporize and excite the
which emits the very specific wavelengths that can be atoms inside the bulb into emitting their characteristic
Members Copy, Not for Commercial Sale
absorbed by the atom. The hollow cathode lamp is an spectrum. Electrodeless discharge lamps are available for
www.IndianJournals.com
excellent, light source for most of the elements shown in Aluminium, Antimony, Arsenic, Bismuth, Cadmium, Cesium,
figure 5. The anode and cathode are sealed in a glass Germanium, Mercury, Potassium, Phosphorus, Rubidium,
cylinder filled with either neon or argon. A window Selenium, Tellurium, Thallium, Tin, Titanium etc. Hollow
transparent to the emitted radiation is fused to the end of cathode lamps also exist for these elements.
the cylinder. The anode is in the form of a thick wire,
generally of tungsten or nickel. 2. CHOPPER
The emission process is illustrated in Figure 6. When A rotating wheel is placed in between the hollow
an electrical potential is applied between anode and cathode lamp and the flame. This rotating wheel is called as
cathode, some of the filled gas atoms are ionized. The chopper and is imposed to break the steady light from the
positively charged ions accelerate through the electrical lamp into pulsating light. Chopper gives a pulsating current
in the photocell. There is also a steady current caused by
light which is emitted by a flame, but only the pulsating
current is amplified and recorded and, thus the absorption
oflight will be measured without interference from the light
emitted by the flame itself.
3. ABSORPTION CELL
Burner System: In order to achieve absorption by atoms, it
becomes necessary to bring the sample to the atomic state.
The most common way is to use a flame which is used for
converting the liquid sample into the gaseous state and
also for conversion of the molecular entities into an atomic
Fig 5: Hollow Cathode Lamp vapour. Cutaway view of an atomic absorption burner
system is shown in figure 7. Generally there are two types
1 .JoU1T''''H~
:.. l.MI1o~,eN of burners are in use
I ,:-' -~:--, :
~ ::: ~-:j ":r'" , ,;,4 a. Total consumption burner: Total consumption burner is
.: ...,M·
l-) \'
"~'-
..... '-"I:I: 1/".... )
.~
noisy and hard to use. The efficiency of this burner is not
\! ---..---,.,~~
!, _ _
~ - -'-·-1
_. lL, ...:...-=,=-' very good. In this burner,the sample solution,the fuel and
oxidizing gas are passed through separate passages and
Fig 6 : Emission Process meet at an opening at the base of the flame
66
Journal of Forensic Medicine & Toxicology Vol. 27 No.2, July-Dec 2010
range.
gas is introduced into the mixing chamber through the fuel
inlet, and oxidant enters through the nebulizer sidearm. 2. Acetone passing into the gas control box may damage
valve and tUbing. For these reasons, acetylene
Stainless steel has been the most common material used
cylinders should be replaced when the tank pressure
for construction of the nebulizer, but this has the
falls below 75 psig (520 kPa).
disadvantage of corrosion from samples. For such cases,
nebulizers constructed of a corrosion resistant material, 3. The high solubility of acetylene in acetone enables it
such as an inert plastic or platinum-rhodium alloy can be to be supplied dissolved in acetone.
used. 4. Acetylene cylinders are filled with a porous material,
Generally, the uptake rate of the sample solution for a which carries the acetone.
conventional fixed nebulizer is about 6 mVmin. At higher 5. Cylinders should always be used in a vertical position
uptakes the system tends to saturate, and the additional to minimize liquid acetone entering the gas line.
solution is not effectively atomized by the flame.
6. As the cylinder pressure falls, the acetone entering
Burner Heads: Burners are designed specifically for the the gas stream increases, producing instability of the
various combustion gas mixtures employed. They are all flame and erratic results for flame stoichiometry
made of solid titanium which is corrosion resistant and free sensititive elements such as calcium, tin and others.
of most of the elements commonly determined by atomic
absorption. There are three burner heads available for use 7. Acetylene line pressure from the cylinder to the
with the dual-option burner system. instrument should never be allowed to exceed 100 k Pa
(15 psig). At higher pressures, acetylene can
i. The 10-cm burner head is designed to be used with the spontaneously decompose or explode.
air-acetylene flame. Because of its long burner path
length, it provides the best sensitivity for air-acetylene 8. Acetylene and copper form an explosive combination.
elements. So Copper tubing or fittings for acetylene gas must,
hence be strictly avoided. Both fuel and oxidant
ii. The 5-cm burner head or nitrous oxide burner head is pressure in all gas lines should be relieved at the end
required for nitrous oxide-acetylene operation. It can of the experiment.
also be used with air-acetylene or air-hydrogen. It can
be rotated 90° to provide reduced sensitivity. 9. Compressed air is needed for Air-acetylene flame. The
air supply should provide a minimum of28 liters lmin
iii. The three slot burner head can be used for air-acetylene at a minimum pressure of 275 k Pa. It is essential to
and argon-hydrogen-entrained air flames. It is have a water and oil traps or filter between the
67
Journal of Forensic Medicine & Toxicology Vol. 27 No.2, July-Dec 2010
compressor and the instrument gas control system to the detector through a window strike a photocathode,
keep oil and water droplets out of the flow-metering Electrons are ejected from the photocathode and
stem. accelerated by a voltage potential to the 1st dynode, where
more electrons are released. These are accelerated to the
Nitrous oxide - Acetylene: The nitrous oxide -
2nd dynode, repeating the electron multiplication step. The
acetylene flame has a maximum temperature of about 2900"(: process repeats with each dynode stage until fmally the
and is used for the determination of elements which form stream of electrons reaches the anode, amplified as much
refractory oxides. It is also used to overcome chemical as a hundred million times. The resulting electrical signal
interferences that may be present in flames of lower then goes to the main amplifier8•
temperature. The use of nitrous oxide requires a number of
accessories and precautions.
l. When nitrous oxide is rapidly removed from the
cylinder, the expanding gas causes cooling of the
cylinder pressure regulator and the regulator diaphragm
so that it sometimes freezes. This can create erratic
flame conditions. It is therefore, advisable to heat the
regulator, using either a built in heater or an externally
supplied source of heat such as an electrical resistance
heating tape.
2. Only the nitrous oxide burner head can be used with
the nitrous oxide-acetylene flame.
Downloaded From IP - 112.215.66.76 on dated 5-Oct-2013
Fig 8: Monochromator
4. Care should be exercised with nitrous oxide on cold 6. READOUT SYSTEM
and humid days in that the flow of gas through the
regulator forms ice and under extreme conditions can The readout system of an atomic absorption
freeze up the regulator causing erratic results. spectrophotometer is simply the means whereby amplified
electrical signals from the photomultiplier are translated
4. MONOCHROMATOR into instrumental readings corresponding to the analytical
result. In most of the atomic absorption measurement simple
The function of the monochromator is to isolate that meters, digital displays, printers, chart recorders, and of
resonance line from non absorbing lines situated close to it course computers are used.
in the source system. A general purpose monochromator
should be able to separate two lines 0.1 nm apart or less SENSITIVITY, DETECTION LIMIT AND WORKING
when operating with minimum slit width. The better the
RANGE
monochromator the smaller the slit width 7•
Sensitivity: Sensitivity is the concentration of an element
Pictorial view of monochromator is given in Figure 8.
that will produce a transmittance of99% (absorbance of 0.0044),
Light from the source enters the monochromator at the
also known as reciprocal sensitivity orcharncteristic concentration..
entrance slit and is directed to the grating where dispersion
As long as measurements are made in the linear working
takes place. The diverging wavelengths oflight are directed
range, the sensitivity of an element can be determined by
towards the exit slit. By adjusting the angle of the grating,
a selected emission line from the source can be allowed to reading the absorbance produced by a known concentration
pass through the exit slit and fall onto the detector. All of the element, and solving the proportional equation below
other lines are blocked from exiting. The useful atomic Sensitivity Cone. ofstandard x 0.0044
absorption wavelength ranges from 200--900 nm. Measured Absorbance
5. DETECTORS Sensitivity Check: Different instruments will produce
different values of sensitivity for the same element. The
Photo-multiplier tubes are used for conversion of the
sensitivity check (in mg/L) value is the concentration of
radient energy signal to an electrical one. Photo-multiplier
element that will produce a signal of approximately 0.2
tube consists of a vacuum photo cell with an anode, a
absorbance units under optimum conditions at the
radiation sensitive electrode and a number of emission
wavelength listed. Using the sensitivity check, the operator
cathodes which have increasing positive potential with
can determine if instrumental parameters are optimized and if
respect to the photo cathode. Photo-multipliers with a total
of 10- 15 electrodes are frequently used. Photons entering the instrument is performing up to specifications.
68
Journal of Forensic Medicine & Toxicology Vol. 27 No.2, July-Dec 2010
Detection Limit: The detection limit is defined as the physical characteristics of solutions such as
minimwn concentration of an element which can be detected viscosity, surface tension, density, PH, and solvent
with 95% certainity. This is that quantity of the element vapour pressure, since all these parameters influence
which gives a reading equal to twice the standard deviation nebulization.
of a series of at least 10 determinations at or near blank 3. For accurate results the standards and samples
level. Detection Limit =2 X Std. Dev .x Conc.lMean Abs. should be matched as closely as possible with
Working Range: This is the range of concentration of respect to the above physical parameters.
analyte element which will produce suitable absorbance 4. Other forms of chemical interference are stable
values for practical analysis. Generally, however, the working compound formation and ionization.
range should lie between 0.1- 0.8 absorbance since this is
c. Chemical Interferences
usually the region of optimum optical precision. Specifically
the % relative error is minimum in the absorbance range 1. Stable compound formation is responsible for many
from 0.1 - 0.8. Useful measurements can be made at lower of the depressive effects that are observed, both in
absorbance, and the overall working range can extend to a atomic absorption and flame emission spectrometry.
lower limit that is about ten times the detection limit. 2. It arises because the element forms a stable
The working range can be extended by either diluting compound either with a species in the matrix or with
all the solutions or altering the instrumental sensitivity. In a species in the flame such as aluminium, vanadium,
flame analyses, selecting alternative (less sensitivity) boron etc., which forms stable refractory oxides not
wavelengths, rotating the burner across the optical path, broken down in the cooler air-acetylene flame.
or doing both will allow determinations to be carried out at 3. Using high temperatures in the Nitrous Oxide-
significantly higher concentrations Acetylene flame, dissociates the refractory
Downloaded From IP - 112.215.66.76 on dated 5-Oct-2013
flame conditions. After optimizing the wavelength and 4. Different elements in Agricultural samples
zeroing the absorbance reading, the burner is adjusted by
Besides the elements Na, K, Ca, Mn, the trace elements
using the horizontal vertical rotational controls. The main
Cu, Mn, Zn, Fe, Mo and B are of special interest in fertilizers
aim in adjusting these controls is to locate the optimum
and plant because they play an important role in the cycle
position of the flame in the light beam. Generally, an uptake
of plants. The analysis of plant materials is always done by
of3-6 mllmin is optimum for pneumatic nebulizers. In the
ashing and dissolving in hydrochloric acid.
case of Nitrous Oxide-Acetylene flame an uptake of2-4 mll
min is good. Only nebulization should be done in air- S. Forensic Science / Toxicology
acetylene flame. Optimization ofthe nebulizer is normally Different Forensic samples like Wearing apparels, wood,
carried out with 5 ppm Copper standard solution. Usually
plastic, skin, glass, leather, metal piece etc., are analyzed
three standards and a blank are used to cover the range of for trace elements Cu, Sb, Hg, Pb, Ni, Sn by AAS.
0.1 to 0.8 Abs. The calibration is performed by using the
blank solution to Zero. The standards should be aspirated 6. Determination oftoxic elements in wine analysis
in ascending order and blank run between standards to Metals in wine occur at the mgllievel or less and though
ensure the baseline has not changed. A graph of not directly related to the taste of the final product, their
Absorbance Vs Concentration is plotted for linearity. content should be determined because excess is
undesirable, and in some cases prohibited, due to potential
APPLICATION OF THE ATOMIC ABSORPTION
toxicity. Lead content in wine for example is restricted in
SPECTROSCOPY
several countries. Ten trace elements which may be toxic or
1. Analysis of metallic elements in biological samples to which wine is sensitive, can be determined by AAS.
Downloaded From IP - 112.215.66.76 on dated 5-Oct-2013
The atomic absorption spectroscopy is becoming a very 7. Monitoring of toxic elements in waters and effluents
important instrument for the measurement of trace metals
Members Copy, Not for Commercial Sale
70
Journal of Forensic Medicine & Toxicology Vol. 27 No.2, July-Dec 2010
can be aspirated into the flame of an Atomic Absorption 6. Slavin M; Atomic Absorption Spectroscopy,
Spectrophotometer for determination of Lead (Pb), Barium 2nd edition, John Wiley & Sons, New York, 1978.
(Ba) and Tin (Sn) content. 7. Welz B; Atomic Absorption Spectrometry, 2nd edition.
VCHVerlagrgesellschabt mbH, Weinheim (Germany),
REFERENCES 1985
1. Elwell W. T., et aI., Atomic Absorption 8. Willard H.H., et aI., Instrumental Methods of
Spectrophotometry, 2 nd
edition, pengamon press, Analysis, 6th edition, CBS Publishers & Distributors,
oxford, 1966. Delhi, 1986.
2 Furman N.H., Standard Methods of Chemical 9. JalswalA.K., Moon D.V., Moharana M. and Gupta M,
th
Analysis, Vol. 1, The Elements, 6 Edn., Robert E. Determination oflead in Forensic samples by Atomic
Krieger, Florida, 1985. Absorption Spectrophotometer, Medico legal update,
2006(1): 23-25.
3. Mavrodineanu R., (editor), Analytical Flame
Spectroscopy, Selected topics, 1st edition, Macmillon 10. JalswalA.K., Moon D.V., Moharana M and Gupta M,
& Co Ltd., London, 1970. Determination ofBa in Forensic samples by atomic
absorption spectrophotometer, J. of Forensic
4. Mendham J. et al., VOGEL'S text book of quantitative Medicine and Toxicology, 2006; 23(1): 6-8.
analysis, 6th edition Pearson education (Singapore)
Pte Ltd., Delhi, 2004. 11. Gupta M., Moon D.V., Moharana M., Lal N. and
JaiswalA.K., Determination ofAntimony in Forensic
5. Skoog D.A., Principles of Instrumental Analysis, samples by atomic absorption spectrophotometer,
3rd edition, CBS College publishing, Philadelphia,
International J. of Medical Toxicology and legal
1985.
Downloaded From IP - 112.215.66.76 on dated 5-Oct-2013
71