Certificate of Performance For Anesthesia Machine: Medicover Hospitals

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

CERTIFICATE OF PERFORMANCE FOR ANESTHESIA MACHINE

Customer : Medicover Hospitals


Sampath Nagar, Kurnool, Andhra Pradesh 518003

Equipment : WATO EX-20

Serial No. : DW-6B005197

Manufacturer : Shenzhen Mindray Biomedical Equipment Co. Ltd

Date of Calibration : 14/07/2022

Valid Up to : 13/07/2023

CALIBRATION STANDARD USED


SL. CAL. CAL DUE
ASSET NO TEST EQUIPMENT MANUFACTURER MODEL NO SERIAL NO
NO. DATE DATE
1 FLOW ANALYZER FLUKE MEDICAL VT 650 4969019 22/10/22 22/10/23
ELECTRICAL FLUKE ESA 615 4610025 22/10/22 22/10/23
2
SAFETY ANALYZER
TEST REPORT
RESPIRATORY MEASURED VALUE ACCEPTABLE TEST
SL.NO UNIT SET VALUE
PARAMETERS VENTILATOR ANALYZER RANGE RESULT
1 RESPIRATORY RATE BPM 12 12 12 +/- 2 PASS
2 OXYGEN % 100 100 99 +/- 5% PASS
3 INSP. TIDAL VOL. mL 550 545 551 +/- 5% PASS
4 PEAK PRESSURE cmH2O NA 20 20.1 +/- 2 PASS
5 MINUTE VOLUME L/min 7.5 7.5 7.5 +/-5% PASS
6 I:E RATIO Ratio 1:2 1:2 1:2 +/-5% PASS
7 PEEP cmH2O 5 5 5 +/-2 PASS

Electrical Safety Test Passed as per IEC Standard 60601.1

Calibration Passed: YES NO

Calibrated & Approved By:______________________

Tel: +91 22 4020 0000


Fax: +91 22 4020 0011