Service Manual
1、General........................................................................................................................... 1
4、Precautions .................................................................................................................... 2
5、Maintenance................................................................................................................... 2
6、Troubleshooting.............................................................................................................. 3
7.9 Separation of the optical detection board pin and the socket ............................. 11
Power switch, power socket, fan, RS-232 interface, bar code scanning interface,
grounding pole, voltage regulator components, LCD display, touch screen, printer, printer
driver, mainboard, optical detection board、step motor.
To ensure injection of proper volume of blood into the ESR tube, there is a
indicating line on the ESR tube. The blood should be around indicating line ±
4、Precautions
1、 Do not use the ESR tube if there is breakage or dust around it.
2、 Check if there is anti-coagulant in the ESR tube before measurement; make sure
3、 The timing from sample collection to measurement should be less than 30 minutes.
4、 The upside-down mixing should be slow. Air bubbles should be avoided; otherwise
5、Maintenance
dry and clean. Keep the ESR tube hole away from the dust. Avoid direct sunlight to
the instrument. Please shut down the instrument when the measurement is
2、Insert and taking out the ESR tube carefully to avoid the tube being damaged which
will lead to the blood flowing into the instrument and the test result uncorrectly.
3、Clean the instrument surface carefully, so that there is no liquid flowing into the
holes.
5、 Clean the outside surrounding of the holes surface with 5% sodium hypochlorite to
Figure 1
Figure 2
b) Then cut off the band that the blue arrowhead pointing to in the figure 3. Four
screws need disassembly. As shown in figure 4:
Figure 4
Figure 6
Figure 7
Figure 8
Figure 9
Figure 11
Figure 12
XC-A30 ESR Analyzer Service Manual 9
7.8 Printer disassembly
Open the cover, pull out the pin of printer data line and power code.
a) Loose the screws as shown in figure 13, then take off the printer components
Figure 13
Figure 14
Figure 15
Figure 16
7805 Voltage
regulator
7812 Voltage
regulator
Figure 17
Voltage regulator
lead wire
J102 printer power code Printer data line LCD data line
J104 main
testing motor
2803 integration
Optical
detection board
2803 integration