revision date: 09 Februari 2007 expiry date: 08 Februari 2011

BALAI PENGUJIAN DAN SERTIFIKASI MUTU BARANG - PROPINSI SULAWESI SELATAN


Accreditation Number : LK-014-IDN
Address : Jl. Andi Pangerang Pettarani Makassar
Telephone : (0411) 441 239;
Facsimile : (0411) 457 368
Contact Person : Ir. Muhdar
Temperature

Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications Notes
Capability *)

1 Liquid in glass thermomter 0 °C ~ 100 °C 0.9 °C ASTM E 77 - 1998


2 Temperature enclosure
Incubator 25 °C ~ 100 °C 1.5 °C AS 2853 : 1986,
Waterbath 25 °C ~ 100 °C 1.6 °C BS 2648 - 2000
Oven 50 °C ~ 300 °C 1.5 °C
Furnace 100 °C ~ 1000 °C 1.5 °C
Mass

Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications Notes
Capability *)

1 Weights 0.001 g ~ 0.005 g 0.015 mg


The Calibration of Balance
(conventional mass) 0.01 g 0.019 mg
0.02 g 0.023 mg NML-CSIRO: 1995
0.05 g 0.03 mg
0.1 g 0.04 mg
0.2 g 0.05 mg
0.5 g 0.06 mg
1g 0.09 mg
2g 0.14 mg
5g 0.19 mg
10 g 0.22 mg
20 g 0.3 mg
50 g 0.4 mg
100 g 0.6 mg
200 g 1.1 mg
2 Balance 0g ~ 100 g 0.6 mg Hoklas
(electronic, mechanic) 100 g ~ 200 g 0.7 mg Calibration of Trades
200 g ~ 500 g 1.8 mg Weight and Balance
500 g ~ 1000 g 7.6 mg
1000 g ~ 2000 g 9 mg
2000 g ~ 6000 g 13 mg
6000 g ~ 10000 g 20.3 mg

1 of 2
revision date: 09 Februari 2007 expiry date: 08 Februari 2011
BALAI PENGUJIAN DAN SERTIFIKASI MUTU BARANG - PROPINSI SULAWESI SELATAN
Accreditation Number : LK-014-IDN
Address : Jl. Andi Pangerang Pettarani Makassar
Telephone : (0411) 441 239;
Facsimile : (0411) 457 368
Contact Person : Ir. Muhdar
Pressure

Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications Notes
Capability *)

1 Pressure Gauge 0 psi ~ 15 psi 0.06 psi


0 psi ~ 600 psi 2.4 psi BS-EN 837 - 1998
0 psi ~ 5000 psi 20 psi
0 psi ~ 10000 psi 54 psi
( 1 psi = 6894,76 Pa )
Force

Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications
Capability *)

1 Tensile testing machine 0.5 kN ~ 5 kN 1% ISO 75001 - 1 : 1999


2 Compression Testing machine 0.5 kN ~ 5 kN 1%

Length

Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications Notes
Capability *)

1 Outside Micrometer 0 mm ~ 25 mm 0.9 µm DIN 863 - 1999


25 mm ~ 50 mm 1.2 µm
50 mm ~ 75 mm 1.5 µm
75 mm ~ 100 mm 2 µm
2 Caliper 0 mm ~ 350 mm 30 µm JIS B 7507 - 1993

3 Steel ruler 0 mm ~ 350 mm 100 µm JIS B 7516 - 1987


Notes:
Best Measurement Capability is stated as Expanded Uncertainty at approximately 95% confidence level, calculated using
coverage factor of k = 2, includes uncertainty contribution due to reference standards used by the laboratory, measurement process
at the laboratory, and contribution due to typical characteristic of nearly ideal device under calibration respective to the laboratory

The actual characteristic of device under calibration shall also be considered for individual calibration certificate issued
by the laboratory