revision date: 2 Mei 2007 expiry date: 1 Mei 2011

PT. Indomobil Suzuki International


Accreditation Number : LK - 050 - IDN
Address : Jl. Raya Penggilingan Cakung - Jakarta Timur
Telephone : 021- 460 9214; 460 9217
Facsimile : 021- 460 9216
Contact Person : Agung Hartono
Length
Best
No. Instrument to be calibrated Measurement Range Measurement Methods/Specifications Notes
Capability *)
1 Outside Micrometer 0 mm ~ 25 mm 1.5 µm JIS B 7502-1994
25 mm ~ 50 mm 2 µm
50 mm ~ 75 mm 4 µm
75 mm ~ 100 mm 5 µm
2 Gear micrometer 0 mm ~ 50 mm 2.4 µm JIS B 7502-1994
50 mm ~ 100 mm 3 µm
3 Inside Micrometer 0 mm ~ 75 mm 1.6 µm JIS B 7502-1994
External Micrometer 0 mm ~ 150 mm 1.6 µm
4 Vernier Caliper 0 mm ~ 300 mm 36 µm JIS B 7507-1993
5 Height Gauge 0 mm ~ 600 mm 36 µm JIS B 7517-1993
6 Dial Gauge 1 mm ~ 3 mm 1.8 µm JIS B 7503-1997
1 mm ~ 10 mm 1.8 µm
1 mm ~ 20 mm 2.4 µm
7 Lever Type Dial Test Indicator 0.5 mm ~ 1 mm 1.75 µm
8 Cylindrical plug gauge 5 mm ~ 200 mm 2.4 µm JIS B 7420-1997
9 Thread Gauge 6 mm ~ 60 mm 2.5 µm JIS B 0255-1998
10 Ring Gauge 15 mm ~ 200 mm 5 µm JIS B 7420-1997
11 Snap Gauge 15 mm ~ 200 mm 5 µm JIS B 7420-1997
12 Plate Plug Gauge 5 mm ~ 200 mm 5 µm JIS B 7420-1997
Notes:
Best Measurement Capability is stated as Expanded Uncertainty at approximately 95% co
coverage factor of k = 2, includes uncertainty contribution due to reference standards used by the laboratory, measurement process
at the laboratory, and contribution due to typical characteristic of nearly ideal device under calibration respective to the laboratory

The actual characteristic of device under calibration shall also be considered for individual calibration certificate issued
by the laboratory