Professional Documents
Culture Documents
Software Spark Analyzer Pro Maxx Version 2000
Software Spark Analyzer Pro Maxx Version 2000
Software Spark Analyzer Pro Maxx Version 2000
1 INTRODUCTION ................................................................................................. 3
1.1 Conventions .............................................................................................................................. 3
2 DESCRIPTION .................................................................................................... 8
2.1 Settings...................................................................................................................................... 8
2.1.1 Application Profiles ................................................................................................................. 8
2.1.1.1 Creating new Application Profiles .................................................................................. 9
2.1.1.2 Editing existing Application Profiles ............................................................................. 18
2.1.2 Methods ................................................................................................................................ 49
2.1.2.1 Method Selection ......................................................................................................... 49
2.1.2.2 Method Editor .............................................................................................................. 50
2.1.2.3 Element Extensions ..................................................................................................... 51
2.1.3 APF ...................................................................................................................................... 65
2.1.3.1 APF Editor ................................................................................................................... 65
2.1.4 Grades .................................................................................................................................. 67
2.1.4.1 Grades Editor .............................................................................................................. 67
2.1.4.2 Grade Definition ........................................................................................................... 68
2.1.4.3 Creation of a Grade ..................................................................................................... 70
2.1.4.4 Modification of Grade Data .......................................................................................... 76
2.1.4.5 Duplication of a Grade ................................................................................................. 85
2.1.4.6 Deletion of a Grade ..................................................................................................... 86
2.1.4.7 Multi-Deletion of Grades .............................................................................................. 87
2.1.4.8 Import of Grades.......................................................................................................... 88
2.1.4.9 Saving of Grades ......................................................................................................... 91
2.1.4.10 Export of Grades ......................................................................................................... 93
2.1.4.11 XML Format of Grades Import ..................................................................................... 94
2.1.4.12 Example of a Grades Import File ................................................................................. 95
2.1.5 User Management ................................................................................................................ 96
2.1.5.1 User Management View .............................................................................................. 96
2.1.5.2 User Authentication ................................................................................................... 115
2.1.6 Argon Saver ....................................................................................................................... 117
2.1.6.1 Argon Saver Module .................................................................................................. 118
2.1.7 Standards ........................................................................................................................... 122
1 Introduction
The measurement program of the Spark Analyzer Pro MAXx system is used to
perform quantitative element using the SPECTROMAXx instrument.
All relevant settings to support the analysis of samples can also be configured
within the measurement program.
1.1 Conventions
DisplayText Name of an item or control within the user interface of
the program.
Note Remark for additional information regarding a specific
subject.
1.2 Terms
Quantitative Analysis Measurement process, which evaluates the elemental
composition of a sample.
Application Profile Application profiles simplify the measurement process
by defining settings and behaviors of the system for
the measurement.
In general, application profiles include display, export
and print settings to be applied as well as the definition
of sample identification fields, number formatting and
the assembly of the toolbar within the measurement
screen.
Method Methods consist of an analytical section with
definitions of the analytical elements and lines,
excitation and detection parameters, calibration
parameters and calculation settings. The data of the
analytical section are essential for performing
measurements. They are independent on user specific
configuration. The data of the analytical section of a
method cannot be modified within this program.
Besides the analytical section, there is a section for
additional application related data, which depend on
the analytical section of the method. The application
related section consist of the definition of element
extensions, pseudo elements, control standards, type
standards, grade filters, grade relations and automatic
reproducibility test criteria.
APF Abbreviation for Automatic Method Selection (called
1.3.2 Navigation
The home view of the program displays icons for the following functions:
Settings
Application Profiles
Methods
Grades
Standards
APF
Argon Saver
Plasma Assessment
User Management
Change Login
Measurement
iCAL
Analysis
Maintenance and Service
Maintenance Reminders
Service
Service Report
Navigation from a different screen back to the home view can be done by using
the Back button, which is visible in each screen, which has the home view as
origin.
2 Description
2.1 Settings
The Application Profiles view will be displayed. It contains a list of already defined
application profiles with base element specification of the assigned default
analytical method and short description. The currently active application profile is
marked by the green check sign.
Overwrite the default profile name with requested new name and enter the
corresponding description.
Note: Make sure, that the new profile name doesn’t already exist on your
system. It’s not possible to enter the next wizard page until an application
name has been entered.
Click on the button Next to display the next wizard page.
The wizard page for Automatic Method Selection (APF) settings will be
displayed.
Three APF modes are available: Single Measurement, Full Automatic and
Semi Automatic.
Note: The analytical configuration of the APF has to be defined within the
separate APF definition tool.
Select the desired APF Mode and press the Next button to step to the next
wizard page.
In the Methods Overview select the requested base element (left list) and then
choose the method you want to use for your application. You may click on the
corresponding method name or navigate with the cursor keys through the list and
then accept the selection with the Return key. The selected method is marked by
the green check sign and will be loaded automatically, when the application
profile is activated.
For your information the elements of the selected method with their concentration
ranges are displayed in the Method Info table.
2.1.1.1.4 Grades
The wizard page for the selection of the grade identification mode of the profile
will be displayed.
Two grade identification modes are available: Grade Verification and Grade
Search. If not requested, the grade identification may be deactivated (option No).
In this mode the program searches automatically for the grade, which fits the best
to results of a measured sample.
Note: When the Grade Search is selected, the Sample IDs definition is
extended automatically with the Grade Norm field. For the Grade
Verification selection, the Sample IDs definition is extended with the Grade
Norm and Grade Name fields.
Select the grade identification mode you want to use for your application and
confirm your selection with a click on the Next button.
2.1.1.1.5.1 Automatically
In this mode the Sample Identification dialog opens automatically, when the
sample analysis finishes (‘Next Sample’)
2.1.1.1.5.2 Manually
In this mode the Sample Identification dialog opens on user’s request (by a
corresponding function button or key shortcut).
Two automatic printout modes are available: Replicates and Mean on ‘Next
Sample’ and Only Mean on ‘Next Sample’. If not requested, the automatic
printout may be deactivated (option No).
Two automatic export modes are available: On ‘Next Sample’ (Analysis Finish)
and On each Replicate (Measurement Finish). If not requested, the automatic
export may be deactivated (option No).
In this mode the complete analysis results (all replicates and statistics) are
automatically exported, when the analysis finishes, i.e. when the ‘Next Sample’
button is pressed.
At the actual state, all basic information has been entered. The wizard logic
checks all settings for completeness and correctness. Any detected errors are
displayed in a red text area. To correct the errors, navigate to the corresponding
wizard page and correct your settings. Come back to final wizard page. To
finalize the wizard and store the new application profile click on Finish.
The editor consists of following property pages: General, Analytical, Grade ID,
Sample IDs, Display, Print, Export, Formatting and Function Keys.
Each page corresponds to a related parameters group. A click on the desired tab
activates the corresponding parameter view. Parameters can be changed in any
order.
Note: Parameter changes can influence settings of other parameters in the
same or another group.
If any parameter was changed, the buttons Save and Cancel in the toolbar
become active.
A click on the Save button stores the modified application profile on disk. The last
saved version of the profile is then immediately reloaded and the buttons Save
and Cancel become inactive.
The not yet saved changes can be rejected with a click on the Cancel button.
To leave the editor, click on the Back button in the toolbar. If the application
profile was modified and not stored yet, an appropriate warning dialog appears
and you can choose to save or to reject the changes.
2.1.1.2.1 General
This page allows modifying the profile identification settings like the Profile Name
and the Description.
If you modify the Profile Name and save this modification, a new profile with the
specified name will be created. The original profile stays unchanged by this
operation.
Note: An already existing application profile with the same name will be
overwritten!
2.1.1.2.2 Analytical
This page allows modifying the analytical characteristics of the application profile:
You can activate or deactivate the Automatic Method Selection (APF), select
the APF Mode, activate or deactivate the Automatic measurement
reproducibility test and select the start-up method.
2.1.1.2.3 Grade ID
This page allows configuring parameters for the automatic grade identification
(Grade ID).
One of two Grade ID modes can be activated: Grade Verification and Grade
Search. If not requested, the automatic grade identification may stay deactivated.
The Grade ID Mode determines if and how the grade identification is performed.
If the Grade Search is activated, the program searches automatically
for a grade, which fits the best to sample analysis results.
If the Grade Verification is activated, the test of the sample analysis
results against the limits of a specified grade is performed. The grade
specification results in the special field of the sample identification and
can be supported by method dependent grades filtering.
Note: Either Grade Search or Grade Verification may be activate at the
same time.
The option Apply Grade Search in the panel Action if Grade
Verification fails allows activation of the automatic grade search, if the
grade verification fails.
The option Show Message in the panel Action if Grade Verification
fails indicates whether a message box is displayed if grade verification
fails.
The option Show Message in the panel Action if Grade Search fails
indicates whether a message box is displayed if no matching grade has
been found.
Note: When the Grade Search is activated, the Sample IDs definition is
extended automatically with the Grade Norm field. At activation of the
Grade Verification, the Sample IDs definition is extended with the Grade
Norm and Grade Name fields.
The Search Mode specifies the algorithm of results and grade limits comparison
at the grade search:
The Sharp Limits is the classical algorithm. Only grades, which at all
element concentrations match exactly to concentrations of the verified
analysis results are accepted and displayed as the search result.
The Smart Search is a flexible algorithm based on the probability
calculations. It uses tolerant matching criteria at borders of the
concentration limit ranges and evaluates the probability of matching to
concentrations of the verified analysis results. Grades, whose probability
exceeds a predefined threshold, are accepted and displayed as the
search result.
The Common Settings are parameters concerning both grade ID modes.
In the panel Proceed On the type of analysis result, for which the grade ID is to
be performed, is specified.
If the Each Replicate is active (checked), the grade ID operation is
applied for each single measurement result (replicate).
If the Average is active (checked), the grade ID operation is applied for
the average result.
In the panel Limits the kind of grade limits, which should be used at values
comparison, is specified.
If the Check Internal Limits is active (checked), the internal (warning)
limits are used.
If the Check External Limits is active (checked), the external
(acceptance) are used.
The activation of the Grade Filtering option causes that at the grade
verification and search only these grades are considered, which pass a
grades filter of the applied analytical method.
In the section Fields the list of sample ID fields is presented. The fields can be
added, modified and removed.
A click on the button adds an empty sample ID field to the list.
The order of the fields can be modify using the and buttons.
A field can be deleted from the list with a click on the button.
Each sample ID field consists of a set of parameters:
Display Name is the name of the sample ID field. This name identifies
the field and will be displayed together with the corresponding
identification value.
Default Value specifies a value that will be set automatically in this field,
when the sample identification dialog opens. The value can be then
modified by the user.
Type defines the type of the sample ID field. It determines which values
can be entered into the field in the sample identification dialog. The
following field types are available:
Direct text input: entry of any text is possible
Predefined list of labels: entry of only predefined texts is
possible (see List of Labels below for external definition
details).
Predefined editable list of labels: entry of any text is
possible. The entered texts are stored and can be used as
predefined texts (see List of Labels below for external
definition details).
Read Only indicates the user is not able to modify the value of this
sample ID field. This option is helpful if a preset Default Value shouldn’t
be modified during the sample identification input.
Unique Name indicates that the value at this sample ID field will be
used for generation of a unique sample result ID. If this option is
activated for multiple sample ID fields, the result ID is built from all of
their values separated with a / character.
In the section Settings the additional parameters related to the sample
identification can be set.
Display Sample ID Dialog after Next Sample: If this option is activated
the sample identification dialog opens automatically, when the sample
analysis finishes, i.e. when the user clicks on the Finish button in the
measurement screen (‘Next Sample’).
In the panel User Task the facility for external sample identification (see
External Sample Identification below) can be activated and
configured. If the option is activated (the checkbox Enabled is checked),
the path and name of the executable of batch can be specified in the
standard file selection dialog ( button). The checkbox Start minimized
should be checked, if the user task application should start without
appearing on the screen.
When a sample ID filed of the type Predefined list of labels and Predefined
editable list of labels is selected in the fields table, buttons and appear.
Direct choice list clearing:
A click on the button clears (after warning) the choice list of the
selected sample ID field. This option is useful for a Predefined editable
list of labels field, when the user requests quick clearing of labels
accumulated in the choice list.
Status file edition:
The button is provided for opening of the Settings Editor and edition
of the xxxx_SampleIDChoiceLists.xcfg file.
Sub nodes of the root node correspond to names of the sample ID fields of
type Predefined list of labels or Predefined editable list of labels defined
in the profile.
Each sample ID field node contains the item MaxItemsCount, which specifies the
capacity of the corresponding label list.
The label list is then specified as values of items (1 up to the list capacity) in the
sub node Items.
2.1.1.2.5 Display
The settings on this page control the presentation of analysis results at the
measurement screen.
The section Display Result Selection specifies which result details are to be
displayed on the screen.
In the subsection Replicates displaying of single measurement results
(replicates) is determined:
All: results of all single measurements (replicates) will be displayed.
Only Last: results of only last single measurement will be displayed.
None: results of single measurements won’t be displayed.
In the subsection Statistics displaying of statistical measurement results
(average and standard deviation) is determined:
Mean: averages of single measurement values will be displayed.
SD: Standard Deviations of single measurement values will be
displayed.
Note: When this option is activated, displaying of the average is
activated automatically.
2.1.1.2.5.1 Settings
The items in the section Settings provide the control on the displayed elements
or channels and the applied display styles.
In the subsection Element Order Selection the set of presented elements is
specified. Note that these settings are base element specific. The base element,
for which the settings are provided, can be select in the Settings for Base list.
All: all elements of the applied analytical method will be displayed.
All except Base Element: all elements with exception of the base
element of the applied analytical method will be displayed.
Filter: the collection of displayed elements depends on the specified
element list. The list and the elements order can be modified with the
buttons below the list control. The button provides the possibility to
fill the element list with the element list of the used analytical method.
Copy Settings to all Output Devices facilitates copying of the display element
filter to the Print and Export settings.
Fonts and Colors determine font and display colors for the different value and
state types. It is also possible to define the ordering of statistics rows (mean, SD,
RSD, limits).
Template: allows using one of the preinstalled templates. The template
can be selected from the list below.
User Defined: provides the customization of font and color sets. A click
on the button opens the related editor dialog (see the Template
Editor for Fonts and Colors below).
Font Size Analysis Results: the values in this group determine the user
defined sizes of fonts in the table of Quantitative Results.
Font Size Found Grades: the values in this group determine the
independent sizes of fonts in the table of Grade Search Results.
The Initial Rows Count specifies the count of rows initially visible in the
table.
In the subsection Value Control a setup of display options controlled by the
result values is possible.
Sort by Values of first Replicate: the display order of the elements will
be determined dynamically based on the element concentration of the
first replicate measurement. The element with the highest concentration
appears on first position.
Hide Values below DL: elements whose measured concentration is
below their detections limits won’t be shown at the display. This option is
available only in combination with the Replicates setting Only last.
Initial Display Control allows predefining, which panels of the Measurement
Screen (Sample Identification, Sample Properties, Quantitative Results,
Grade Search Results) are initially visible. Note that the panels can be activated
or deactivated automatically (by measurement process) or manually at the
Measurement Screen.
Clicking on one of the buttons on the left hand side within this group will toggle
the display of the corresponding statistics row. When rows will be activated, the
corresponding entry in the ordering list will be added at the end of the list. The
ordering of the active rows can be changed by using the arrow buttons.
Note: Statistics rows are only visible in the measurement screen, if
corresponding values are available within the result. Example: If the result
does not contain warning limits, the warning limits rows W. Min and W. Max
will not be displayed independent of the setting within this definition. On
the other hand, if a given row is deselected within this definition, the row
will not be displayed even if corresponding values exist in the result.
Display Styles
The Display Styles group consists of a navigation tree at the left hand side of the
group and a details section at the right hand side. The navigation tree has three
main sections for General display data and display data for Row Styles and
Value Styles:
General
Extra Font Size: Font size for the last replicate row
Font Name: Font name of the font to be applied for the
measurement screen
Font Size: Font size for all values within the measurement
tables with the exception of the last replicate row, which uses
the Extra Font Size
Table Header Font Size: Font size for the table header
Row Styles: Replicate, Deleted Replicate, Mean, SD, RSD, Reported,
Acceptance (External Limits) Range, Warning (Internal Limits) Range,
Calibration Range
BackColor: Background color
Font Style: One of the following types: Regular, Bold, Italic,
Underline, Strikeout
ForeColor: Foreground color
Value Styles: Deleted Value, Status Analytical Error, Status Math Error,
Status Overflow, Ext. Status Inhomogeneous, Ext. Status Modified, Ext.
Status Outlier, Acceptance InRange, Acceptance OverRange,
Acceptance UnderRange, Acceptance OutOfRange, Warning InRange,
Warning OverRange, Warning UnderRange, Calibration OutOfRange,
Calibration OverRange, Calibration UnderRange
Appendix: Text or character, which will be appended to the
value
BackColor: Background color
However it is possible to preserve the recent analysis results in the Analysis View
up to the start of the next sample analysis or also during the analysis of
successive samples.
To preserve the analysis results in the Analysis View up to the start of the next
sample analysis the flag When next analysis starts needs to be activated. The
SPECTRO Analytical Instruments GmbH 34 of 228
SPECTRO
Analysis View is then cleared only, when the first measurement of the next
sample begins.
If preserving of the analysis results on screen, also during the successive
analyses, is requested, the results pinning mechanism can be activated (the Max.
pinned results value needs to be set greater than 0). The results of the finished
analysis can be then pinned manually (the flag When next analysis starts has
to be active) or automatically (the flag Pin results automatically needs to be
activated). The Max. pinned results value decides, how many pinned results
can be shown at the Analysis View simultaneously. On exceeding this value, the
oldest result is removed from the view automatically.
To pin the analysis results automatically (also when Analysis View cleaning
When current analysis finishes is active), activate the flag Pin results
automatically.
If preserving of the lasts analysis results only up to the start of the next sample
analysis is requested, activate the flag When next analysis starts. Than the
results of the finished analysis are not removed from the Analysis View, when the
analysis finishes, even the results pinning mechanism is deactivated (the Max.
pinned results value is 0).
2.1.1.2.6 Print
The settings on this page configure the printout of analysis results.
The section Print Result Selection defines the printout target and which result
details will be printed out.
In the subsection Printer Target the target for the printout of the analysis results
is determined. The results can be printed automatically at only one target.
Default Printer: The results will be printed out on the default printer of
the operating system (for installation of the default printer, please refer
the help of your Windows© operating system).
PDF File: The results will be stored as the PDF document. Click on the
button to select the desired path and name of the target file. The
activation of the Auto Increment option results with adding of a
numerical suffix to the specified file name, incremented each time the
results are printed to the PDF file. Otherwise a new PDF result file will
overwrite the existing one.
Note: PDF documents can be opened and displayed with the Adobe©
Reader©.
In the subsection Automatic Printout a time point the results are printed out
automatically can be set. The results can be printed automatically at only one
time point.
On each replicate (Measurement Finish): The results will be printed
out automatically directly after each single measurement finishes.
In the subsection Print Header Text an own customer’s text can be entered. It
will appear in header on every printout.
In the subsection Extended Settings font and display colors for the different
value and state types can be determined. It is also possible to define the order of
statistics rows (mean, SD, RSD, limits).
Clicking on one of the buttons on the left hand side within this group will toggle
the display of the corresponding statistics row. When rows will be activated, the
corresponding entry in the ordering list will be added at the end of the list. The
ordering of the active rows can be changed by using the arrow buttons.
Note: Statistics rows are only visible in the measurement screen, if
corresponding values are available within the result. Example: If the result
does not contain warning limits, the warning limits rows W. Min and W. Max
will not be displayed independent of the setting within this definition. On
the other hand, if a given row is deselected within this definition, the row
will not be displayed even if corresponding values exist in the result.
Print Styles
The Print Styles group consists of a navigation tree at the left hand side of
the group and a details section at the right hand side. The navigation tree
has four main sections for General settings, General Styles, Row Styles and
Value Styles:
General
Font Name: Font name of the font to be applied for printing
Font Size: Normal font size for all values to be used, when
no different font size has been specified
Image Header: Picture to be used within the results header.
The image can be one from the existing image files (png
format) in the Template\Print\Results of the software (default:
C:\Spectro Smart Studio\Spark Analyzer Pro
Mobile\Template\Print\Results)
2.1.1.2.7 Export
This page provides the settings for the export of analysis results in a file.
The section Export Result Selection activates and configures the export
functionality.
In the subsection Export Format the used data format for the export is selected.
The results may be exported in different formats at the same time.
XML: activates the export of analysis results in an XML format to the
specified file. A click on the button allows selection of the path and
name of the target file. The activation of the Auto Increment option
results with adding of a numerical suffix to the specified file name. The
suffix is incremented each time the results are exported.
Note: The XML export includes all available result information and thus is
extremely convenient for a further data processing or formatting.
ASCII: activates the export of analysis results into an ASCII file. A click
on the button allows selection of the path and name of the target file.
The activation of the Auto Increment option results with adding of a
numerical suffix to the specified file name. The suffix is incremented
each time the results are exported.
Note: The content of the ASCII export is customizable in the section
Advanced Settings (ASCII Format only) described below.
In the subsection Automatic Export a time point the results are exported
automatically can be set. The results can be exported automatically at only one
time point.
On each replicate (Measurement Finish): The results will be exported
automatically directly after each single measurement finishes.
2.1.1.2.8 Formatting
This page presents rules for formatting of the result values and allows their
adjustement.
With a click on the button can be easily copied the entered formatting
exceptions for the output devices Display to the Print device.
In the section Function Key Assignment all available function keys are listed.
The field Function describes the functionality associated with this function key.
The checkbox Enabled generally activates or deactivates the functionality.
Note: The effective activation or deactivation of functions depends on the
temporary state of the program.
The option Toolbar adds or removes the corresponding function button
(presented in the Icon column) in the toolbar of the measurement screen.
Note: The appearance of the button icon may depend on the temporary
program state.
The Key represents the keyboard short key assigned to the function. It can be
chosen from the list of available keyboard keys. The field Key Extender provides
the possibility to activate the function only in combination with one of the
additional keys Shift, Ctrl or Alt.
In the section Tool Bar the preview of the function buttons in the toolbar of the
measurement screen is presented. The order of the buttons can be modified with
the and buttons.
2.1.2 Methods
Each analytical method consists of analytical section (analytical core) and
additional, application related data.
The analytical core is essential for performing of measurements. It contains
definitions of analytical elements and lines, excitation and detection parameters,
calibration parameters and calculation settings. These data are independent on
the user specific configuration. The data of the method analytical core cannot be
modified within this program.
The additional, application related data depend on the corresponding method
analytical core. They consist of element extensions, pseudo element definitions,
type standard and control standards collections, grade filters and relations as well
as of automatic reproducibility test criteria.
Note: Because of the dependency on the analytical core, the application
related data are not a matter of the Application Profile definition, which can
use different methods.
For loading of analytical methods and managing their additional application
related data click on the icon Method in the home view.
The list on the left hand side shows the available bases. The base element of the
currently loaded method is marked with a green check sign. Methods available
for the currently selected base are shown in the list aside. The loaded method is
marked with a green check sign.
On the right side (the Method Info table) the element concentration ranges of the
selected method are shown.
A click on the desired method name loads the method immediately.
The table Elements presents the element list of edited method. For each element
is its Concentration Range, Unit, Visible and Editable flag shown.
The Concentration Range is provided for information only (the field is read-
only).
The element Unit can be modified by selection from the list of units defined at the
instrument. Note, that the modification of the unit has influence on the values
presentation only. The calculations and values archiving stay unchanged.
The Visible flag provides the opportunity to generally disable the element from
the output.
The flag Editable activates a special functionality of the measurement screen,
which allows an in-site correction of average analysis results (element
concentrations). See Manual Correction of Analysis Result below for details.
Enter the modified value and click on the button Apply. The value is put into the
sample results and the dialog closes. The new value overwrites the
corresponding analysis results element concentration in replicates and in the
statistics (the SD and RSD are removed). The modified value is marked (here by
the prefix M and blue color).
To restore the original element concentration, open the same dialog with the
double click on the mean value of the requested element and push the button
Reset. The original concentration value is put into the sample results and the
dialog closes.
Note: The Manual Correction of Analysis Result is possible only for element
concentrations of not finished analyses.
The tree on the left hand side of the Pseudo Element Editor view provides the
available Data Source for the Pseudo Elements. The list of pseudo elements of
current method is displayed on the right hand side.
To create a new pseudo element, click on the button below the formula list. A
new empty formula of type User def. appears. Enter the Display Name of the
formula. If the formula result is a concentration value then check the option Is
Concentration and select the Display Unit from the list of available units.
Otherwise enter your own unit name.
Select the first variable from the Data Source. Add the selected item with a
double click or press the button below the Data Source. The item will be
added to the formula.
Enter, if necessary, an arithmetic operator. Possible operators are + - * / ().
Repeat this procedure until the formula is completed.
The formula syntax can be checked with a click on the button.
An existing formula can be deleted with the button.
By default the pseudo element symbols are added at the end of the method
elements list. The display order depends on the order in the formula list, which
can be modified with the and buttons.
Data which has been modified since the last storage are marked with a yellow
background color.
SPECTRO Analytical Instruments GmbH 54 of 228
SPECTRO
2.1.2.3.1.2 Impurities
A pseudo element of the type Impurities is a special value calculated in relation
to the grade verification.
To create a pseudo element of type Impurities, click on the button below the
formula list, enter the Display Name of the added pseudo element and change
its Type to User def.. The formula of such pseudo element will be generated
automatically during the analysis and need not to be specified here. It will
consider the elements marked as Impurity in the grade assigned to the analysis
results and calculated as sum of their concentration values. See the paragraph
Grades for specification of the grade impurity marks.
The upper list in the group Global Standards shows all standards of the global
standards library, which belong to the base of the currently loaded method and
are marked as a type standard.
SPECTRO Analytical Instruments GmbH 55 of 228
SPECTRO
The list below displays the element data of the selected global standard.
The upper list in the group Type Standards shows the type standards associated
with the loaded method.
Name: Name of the type standard (read only).
Alloy Group: Name of the alloy group the standard belongs to (read
only).
Used: Indicates whether the type standard can be selected for type
correction.
Valid: Indicates whether the type correction data are valid or not (read
only). Note that a type standard correction data becomes valid after
successful type standardization and a subsequently performed
ICalization invalidates this data.
Min. Measurements: Min. number of measurements required for this
standard at the type standardization.
Modification Date: Date and time of the last modification (read only).
Measure Date: Date and time of the last type standardization with this
standard (read only).
The list below displays the element data of the selected type standard.
Element: Name of the element (read only).
Use: Indicates whether the element should be used for type correction
or not.
Unit: Display unit of the element (read only).
Min. Conc: Lower acceptance limit of the measured element
concentration during type standardization.
Max. Conc: Upper acceptance limit of the measured element
concentration during type standardization.
Max. RSD: Upper acceptance limit of the measured element RSD
during type standardization.
Corr. Value: Correction value that will be applied at the type correction
to the element concentration. The meaning of the value depends on the
Apply Offset setting.
Apply Offset:
Checked: The Corr. Value is a concentration value. This
value will be added to the measured element concentration.
Unchecked: The Corr. Value is a factor. This factor will be
multiplied with the measured element concentration.
A click on the button adds a global standard to the list of available type
standards.
A click on the button removes a type standard from the list of available type
standards.
Data which has been modified since the last storage are marked with a yellow
background.
The upper list in the group Global Standards shows all standards of the global
standards library, which belong to the base of the currently loaded method and
are marked as a control standard.
SPECTRO Analytical Instruments GmbH 57 of 228
SPECTRO
The list below displays the element data of the selected global standard.
The upper list in the group Control Standards shows the control standards
associated with the loaded method.
Name: Name of the type standard (read only).
Alloy Group: Name of the alloy group the standard belongs to (read
only).
Min. Measurements: Min. number of measurements required for this
standard at the control analysis.
Modification Date: Date and time of the last modification (read only).
Measure Date: Date and time of the last control standard analysis (read
only).
The list below displays the element data of the selected control standard.
Element: Name of the element (read only).
Unit: Display unit of the element (read only).
Min. Conc: Lower acceptance limit of the measured element
concentration during a control standard test. The default value is 90% of
the Set Conc. value.
Set Conc: The expected concentration value of the element.
Max. Conc: Upper acceptance limit of the measured element
concentration during a control standard test. The default value is 110%
of the Set Conc. value.
Max. RSD Upper acceptance limit of the measured element RSD during
a control standard test.
A click on the button adds a global standard to the list of available control
standards.
A click on the button removes a type standard from the list of available control
standards.
The button generates a printout of the currently selected control standard.
Data which has been modified since the last storage are marked with a yellow
background.
Using the Grade Filter, only such grades are considered in the Grade ID, which
match the concentration ranges of the elements defined in the filter. Elements,
which are not defined in the filter, are not considered within the filtering process.
The Method Elements tree at the right hand side of the Grade Filter window
shows the elements of the method together with the corresponding concentration
ranges.
A click on the button adds the selected element to the list of Grade Filter
Elements. Each filter element provides the following parameters:
Display Name: The name of the filter element. The name has to match
the display name of the element in the grade’s definition (read only).
Display Unit: The concentration unit of filter range.
Min. Conc.: The lower limit of the filter range. If no value has been
entered the limit won’t be considered for filtering.
Max. Conc.: The upper limit of the filter range. If no value has been
entered the limit won’t be considered for filtering.
A click on the button deletes the selected element from the list.
The buttons and allow to modify the check order of the elements. This can
take influence on the performance of the filtering process.
A click on the button presets the grade filter with the elements used for APF
method selection (see APF Editor).
Note: This button is not active, if the current method doesn’t exist in the
APF configuration.
Data which has been modified since the last storage are marked with a yellow
background.
The list Available Grades contains names of grades, to which the edited method
can theoretically be assigned. If to any presented grade another method than the
edited one has already been assigned, the name of this method, and if applicable
the name of the assigned type standard, is specified. The activation of the option
Use at sample identification below this list assures that the grade selection list
in the Sample Identification dialog will be filled with names of Available Grades.
The list Referencing Grades contains names of grades, to which the edited
method is assigned. If to any presented grade a type standard is assigned, the
type standard is specified. The activation of the option Use at sample
identification below this list assures that the grade selection list in the Sample
Identification dialog will be filled with names of the Referencing Grades names.
The Grade section in the center presents parameters and limits of the currently
selected grade.
Note, that the list of the Available Grades may be limited correspondingly, if
the method Grade Filter is configured and active.
For assigning of the edited method to a grade, select the requested grade in the
list of Available Grades and click on the button below. The selected grade
moves to the list of Referencing Grades.
Note: You can assign only one method and one type standard to a grade. If
you want to assign the just edited method to a grade, which have already
another method assigned, you can simply overwrite the existing
assignment by the above procedure.
For assigning of a type standard to a grade, select the requested grade in the list
of Referencing Grades and in the section Grade, select the name in the field
Type Standard. Note, that type standards have to be preset in the Type
Standards of the edited method.
Note: The same type standard may be assigned to different grades.
For removing of the edited method assignment to a grade, select the requested
grade in the list of Referencing Grades and click on the button below. The
selected grade moves to the list of Available Grades.
Note: This view supports multiple selection and drag and drop mechanism
for grade assignments.
The SD/RSD Test uses the standard deviation of replicate values for
outliers’ detection.
Note: When the Grubbs’ Outlier Test is selected the minimal count of
measurement replicates (Min. Replicates) has to be greater or equal 3.
Outlier Type:
When the Entire Replicate is selected, a single outlier causes the entire
replicate it belongs to, is automatically marked as the outlier. It is the
default selection.
Note: The outlier values are marked at the measure grid by a color, font
style and/or prefix. If the outlier type Entire Replicate is selected, all values
of the outlier replicate are marked too. The marking of the non-outlier
values of the outlier replicate differs from the marking of the outlier values.
The marking style can be customized in the application profile (section
Display – Fonts and Colors – User Defined).
Outlier Handling:
Marking a single value or an entire replicate as an outlier eliminates it
automatically from the statistics calculation of the measurement.
The Conc. Limit value determines the upper concentration limit for a sub-range.
The corresponding Max. Deviation and Deviation Type values describe the
used reproducibility criterion in this sub-range.
The Deviation Type indicates, if the Max. Deviation value is a relative standard
deviation (RSD) or a standard deviation (SD). The Unit indicates, in which unit
the Max. Deviation is expressed.
At the SD/RSD Test the concentration values are reproducible, if their (relative)
standard deviation doesn’t exceed the specified Max. Deviation. In the recursive
steps, the test marks values the most deviated from the average as outliers, until
the above condition is fulfilled or the count of non-outlier values falls below the
Max. Replicates count.
2.1.3 APF
APF (Automatic Method Selection) is a procedure that provides the possibility to
determine the best fitting method based on a measurement with an Orientation
Method. Orientation methods are calibrated over a wide concentration range and
provide approximate analysis. These concentration values are used to determine
be best fitting method within the same base for the given analysis result. If a
fitting method has been found, the software loads this method automatically and
recalculates the analysis with the intensities determined by the orientation
method.
The usage of APF can be configured for Analytical Application Profiles.
Methods
The Filter settings determine the group of methods which should be currently
setup for the APF.
Base: Determines the base element the displayed methods.
Orientation Method: Defines the method which should be used as
startup method for the selected analytical methods in the list below.
Analytical Method data:
Elements
This list contains all analytical elements of the currently selected method. By
default the elements concentration ranges are preset with the measuring ranges
of the method.
Element: The name of the element
Use: indicates whether this element is used for determination of the best
fitting method
Unit: The concentration unit of the element
Min. Conc.: The lower concentration boundary of the element
Max. Conc.: The upper concentration boundary of the element
2.1.4 Grades
Grades describe chemical element composition of alloys.
At the instrument a basis library of common grades (Grades Library) is installed.
This library is used by the Grade ID functionality for the grade search and the
grade verification.
A grades library can be presented, managed and extended by the Grades Editor.
Name
The Name of a grade together with the Basis and Norm identifies the grade.
This identification must be unique within the system.
Basis
The Basis name specifies the base element of the alloy and is part of the grade
identification.
Norm
The Norm name specifies the norm in which the grade is defined and is part of
the grade identification.
Description
The Description of a grade is optional and allows specification of detailed
information, remarks and comments concerning the grade.
Target Conc.
The grade data view is cleared and you can to enter new data.
Note: The navigation in the grade tree is disabled when the Grade Editor is
active.
Note: At the Basis selection only names of the base elements installed at
the instrument are available. When selecting the Norm, you can specify a
new norm name in addition to the already used names available in the
choice list.
Note: The combination of grade Name, Basis and Norm must be unique at
the system. Otherwise the error message occurs and you will need to
correct your entry.
Select the element name from the choice list and enter limit values as well as the
corresponding concentration unit. If requested mark the added element as
impurity.
Note: The Element name choice list contains elements installed at the
instrument for the grade Basis. The element choice list is optional can be
disabled via an item in then configuration file Business_Settings.xcfg.
Repeat the above procedure to insert further element concentration range items.
Note: The Unit indicates how the concentration limits are presented in the
grade data view and has no influence on the stored limit values.
Note: The created grade is not permanently stored until you click on
the Save button in the Grades View.
Clicking the Cancel button rejects the new grade and closes the Grade Editor
(i.e. disables entry in the grade data view).
Note: Before the deletion is completed you will be prompted for the
additional confirmation.
Note: The Element name is selected from a choice list containing elements
installed at the instrument for the specified grade Basis.
The currently modified parameters are marked by the yellow background.
Note: The Unit indicates how the concentration limits are presented in the
grade data view and has no influence on the stored limit values.
Then select the values where the copied value should be inserted and press
Ctrl+V or open the context menu of the selected cells (right mouse button click)
and click on the option Paste.
Note: The grade modifications are not permanently stored until you click on
the Save button in the Grades View.
Clicking the Cancel button rejects the grade modifications, restores the original
grade data and closes the Grade Editor (i.e. disables entry in the grade data
view).
Note: You may set the appropriate name of the duplicated grade using the
Edit function.
The grades data can be imported from XML file, MSAccess file and SQL server.
The import from the XML File is a primary way to import grades from external
equipment or from another SPECTROMAXx instrument.
Note: The grades data in the XML file has to correspond to the specified
XML-format. See the sections “XML format of grades import” and “Example
of a grades import file” for details.
The import from MSAccess File and from SQL Server allows transferring grades
from data collection of the Spark Analyzer Vision software (SPECTROLAB or
SPECTROMAXx).
Select the requested grade data type and source (use Browse button to specify
the file name) and click on the OK button. The grades data are imported and
analyzed. If more than one base element is installed at the instrument, the base
element for the imported grades needs to be specified and the grade base
element identification dialog appears (Grade basis). If necessary, the dialog
prompts also for the grade norm (Grade norm).
Specify the missing attribute(s) and click on the OK button. The grade
identification will be completed and the grade data imported.
If the same identification attribute(s) is missing for multiple or all grades and you
don’t want to specify and confirm the missing attribute for each grade separately,
check the Use the basis for all imported grades and/or Use the norm for all
imported grades options and click on the OK button. The identifications of all
succeeding grades will be completed with the specified attribute(s) and the grade
data imported.
The names of imported grades appear in the grades tree and are yellow
highlighted.
Click the Save button in the toolbar to store the imported grades permanently in
the grades library.
Note: The imported grades would overwrite the grades with the same
identification already existing in the library. In such case you will be asked
for overwriting confirmation and can reject (Skip) the imported grade.
Note: If you attempt to leave the Grades View and any modification of the
grades library has not been stored, a warning message appears and you
can cancel the action or continue it with storing or rejecting the changes.
The standard file selection dialog appears and allows you specify the target file
for grades data.
Enter the file name and click on the Save button. The grades will be stored in the
target file. Then you can transfer the file to another instrument and import the
data.
Note: The grades data in the target file corresponds to the specified XML-
format. See the Appendix “XML format of grades import” and “Example of a
grades import file” for details.
Initially the display shows an overview of the available User accounts and their
relationship to Permission Roles.
The tree on the left hand side is provided for navigation to the requested
components (User, Permission Roles and Settings) of the User Management.
The central view corresponds to the selection in the navigation tree and facilitates
the data creation and modification.
When any data within the User Management was modified, the Save button in
the toolbar becomes available. It has to be pushed for the permanent saving of
the changes.
For cancelling of the performed changes, the User Management view needs to
be left with a click on the Back button in the toolbar. The user is then prompted to
confirm the data saving or rejection.
After pushing the New User button an input form for a new user account is
displayed.
Filling of the Real Name, Login Name and Password fields is mandatory. As
soon as the Password (Confirmed) field has been filled correctly, the Apply
button in the toolbar appears.
The Real Name is the public name of the user account, which is typically the first-
and last name of the user. It will be displayed in the status bar of the program and
used as operator signature in results of sample analysis.
The Login Name is the user alias facilitating the login process and will be
required at the User Authentication (Login) dialog. Its minimum length is
configurable within the settings of the User Management system. The login name
must be unique within the system.
The Permission Role entry specifies one of the existing permission roles and
decides about the user’s rights in the system. Only one permission role can be
assigned to the user.
Note: To the first user’s account the Administrator permission role is
assigned automatically.
The Password will be used in combination with the Login Name at the login
process. When the password is entered for the first time or modified, it has to be
confirmed by entering the same characters combination in the Password
(confirmed) field. The minimum password length is configurable within the
settings of the User Management system.
The field Change Password After (Days) determines the password expiration
time in days. If the password expires, the user will be required to change it, when
he logs-in next time. The empty entry in this field (default) disables the password
expiration.
The option Logoff After (Minutes) determines the user’s inactivity time in
minutes, after which the user is logged-off automatically. The empty entry in this
field (default) disables the automatic logoff.
The flag Have To Change Password enforces a password change at the next
user login process.
The flag User Is Active is used to temporary enable (checked) or disable
(unchecked) the user’s account.
The new user account is finally created by pushing the Apply button or discarded
by pushing the Cancel button in the toolbar.
After pushing the New Role button an input form for a new role is displayed.
After entering the name of the permission role, the requested permissions can be
selected.
A double click on a permission in the Available Permissions list or a click on the
button below the list adds the selected permission to the
Assigned Permissions list.
2.1.5.1.2.4 Permissions
Adding permission to a permission role assigned to the account of a user
authorizes this user to utilize the corresponding software functionalities. The
below list specifies available permissions and their meaning.
The permissions of the section Analysis concern the measurement screen:
Can perform grade Enables or disables the manual activation of the grade
search manually search for the current sample analysis results.
Can call results Enables or disables the start of the Sample Result
overview Manager (context dependent).
Can perform type Enables or disables the start and processing of a type
standardization standardization
Can print manually Enables or disables the manual printout of the current
sample analysis results
Can export results Enables or disables the manual export of the current
manually sample analysis results
Can recall the last Enables or disables recalling of the last measurement
measurement results
Enables or disable the possibility to terminate the
Can terminate the
Argon flush process Argon flush process
Can edit print output Enables or disables changing of the results printout
configuration (Printer page)
Can edit export output Enables or disables changing of the results export
configuration (Export page)
Can edit function keys Enables or disables changing of the function key
assignments (Function Keys page)
Can edit grade ID Enables or disables changing of any properties for the
grade identification (Grade ID page)
Note: Removing of this permission overrides the
permissions Can Choose Grade Search Mode and
Can Choose Grade Search Or Verification.
Can edit profiles’ Enables or disables changing of the setup method and
analytical selections APF mode (Analytical page)
Can import grades Enables or disables import of grades into the Grade
Library
Can export grades Enables or disables export of grades from the Grades
Library.
Can modify air filter Enables or disables modification of the Air Filter
service intervals Service intervals
Can modify argon exit Enables or disables modification of the Argon Exit
service intervals Service intervals
Can edit argon saver Enables or disables modification of the Argon Saver
tasks tasks
2.1.5.1.3 Settings
The component Settings of the User Management defines common defaults
used at the creation of new user accounts.
Within the login process the User Identification dialog is displayed. After
entering of the Login Name and the Password, pushing on the OK button
activates the application.
Note: At start-up time, using the Abort button will prevent the application
start. When pressing the Abort button, when calling the User Identification
dialog manually (Change Login button) the application will not be closed.
Note: The Abort button is not available when the User Identification dialog
is displayed after the user has been automatically logged off.
On pushing the OK button the warning message appears and prompts for
confirmation of the service login request.
After the confirmation by a click on the Yes button the application grants the
appropriate service permissions.
Switching on the plasma generator causes that the Argon flow is reactivated by
flushing the instrument with a high level Argon flow (analytical Argon flow). The
flush process time is variable and depends on the period of time without Argon
flow. The flush time is calculated automatically by the Argon Saver.
During the flush process an info dialog appears, showing the required flush time
and the flushing progress.
The flush process can be terminated by clicking the Abort button. The Close
button provides the opportunity to leave Flush Progress dialog without
terminating the flush process.
Note: It’s highly recommended to pass the complete preset flush time. The
Argon flow state of the instrument has direct influence on the accuracy of
the measure results.
2.1.6.1.1 Scheduler
The Scheduler allows to select time and date when the instrument should be
Ready for Operation. Three scheduler modes are available.
Today: Variable time selection with the fixed preset actual date. This option can
be used for non-working times within the current day.
Selected Date: Variable time and date selection, the date can be selected by a
calendar control. This option is helpful when the instrument is used in irregular,
non-cyclic time intervals.
Scheduler Task: Predefined scheduler tasks, with fixed times applied on specific
week days or week day intervals. This option should be selected, when the
instrument is used in iterative regular intervals (e.g. Monday to Friday, at 7:00
a.m.). The desired Task Name can be selected from a list of stored scheduler
tasks. Scheduler Tasks can be created and modified with the Scheduler Task
Editor. The general operation of the Argon Saver Module can activated or
deactivated by the Enabled control.
Note: Changes in the scheduler setup must be accepted with a click on the
button.
Note: The scheduler only operates if Spark Analyzer Pro MAXx is running
and the spectrometer is connected with the mains supply!
Note: The flush process doesn’t start if the Argon pressure is low. If the
argon supply has been restored before the scheduled time has been
elapsed the system restarts the flush process automatically.
The setup of a new Scheduler Task can be started with a click on the symbol
below the Task List. Enter the Name of the new task. This name is used later on
for the selection of the applied scheduler task. Select the Time when the
instrument should be Ready for Operation. Select the Operating Days. The
Start field defines the first and the End field the last weekday of the operating
interval. If the task should be performed on one weekday only, both fields have to
be set to the same value. If the scheduler task should be performed on every
week day select the option Everyday for the field Start. The field End must be
empty in this case.
The setup of the new scheduler task as well as any modification of an existing
task must to be accepted with a click on the symbol below the Task List. The
currently selected scheduler task can be deleted with a click on the symbol.
The Scheduler Task Editor can be hidden with a click on the symbol in the
Scheduler group.
2.1.7 Standards
The Global Standards Manager provides the opportunity to get an overview to
installed standard samples and to administrate the standard samples used for
control standard test and type standardization. Each standard belongs to a base
element and an alloy group. Alloy groups represent a class of standards with
similar composition, e.g. low alloy steels or Cr/Ni steels. A specific standard
exists only once in the complete Spark Analyzer Pro MAXx environment and can
be referred by several methods. Standards can be used in multiple manners, e.g.
as calibration and control standard.
Note: All standards that are used for calibration, standardization or quality
control can’t be modified or deleted. However it’s possible to use a preset
calibration standard as well as control or type standard.
The filter can be activated by entering the element name and the desired
concentration range. It’s possible to set the lower or upper concentration limit or
both. A click on the button clears element filter and disables the element
filtering.
All properties, except Name and Base, of user defined control or type standards
can be modified. The Element Concentrations of the currently selected
standard can be modified. A click on the button deletes the element. New
elements can be added with a click the button. Changes can be stored
permanently with a click on the symbol in the toolbar.
A new standard can be created with a click on the symbol in the toolbar. A
dialog with the header data of the new standard appears.
Name, Base Element and Alloy Group are essential inputs and must be
entered or selected. Supplier and Description are optional information for the
extended documentation of a standard. A multiple selection of the standard Type
is possible. The definition of the new standard must be finished with a click on the
Ok button. The preselected standard appears in the list of Standards. Now you
SPECTRO Analytical Instruments GmbH 123 of 228
SPECTRO
can enter the elements and their concentration values in the list of Element
Concentrations. Element, Certified Concentration and Unit are essential
inputs; the Concentration RSD value is optional and describes the RSD of the
concentration determination.
The currently selected standard can be copied with a click on the symbol in the
toolbar. A dialog with header data of the selected standard appears. Modify the
preset standard Name and accept the changes with a click on the Ok button.
Note: Standards can be copied only within the same base.
User defined standards can be deleted with a click on the symbol in the
toolbar. After accepting a warning dialog the standard will be deleted. A multiple
selection of standards to delete is possible.
Note: Preset calibration, standardization and QC standards can’t be
deleted!
The changes can be permanently stored on disk with a click on the symbol in
the toolbar.
Note: Modifications of standards which are used by a method may have
unforeseeable effects on the analytical performance!
2.2 Measurement
The Analysis View can be activated also from the Application Profiles view with
a click on the Analysis button in the toolbar or with a double click on an
application profile name.
The default Analysis View consists of two panels: Sample Identification and
Analysis Results.
In dependence on the measurement context and/or user request two further panels
(Sample Properties and Grade Search Results) can be activated. At the bottom a
customizable button toolbar and a status bar are available.
Sample Identification
The Sample Identification panel is placed at the top and presents the
current sample identification (Sample ID). The presented sample ID will
be stored later together with the sample analysis results.#
Sample Properties
If active, the Sample Properties panel is placed directly below the
Sample Identification. It shows the sample properties Unique Name,
Type Standard and Correction Mode are shown. If the automatic
reproducibility test is active, the properties are extended with the Repro
Test.
Button Toolbar
In the Button Toolbar function buttons are displayed. To facilitate performing
of analyses, the function buttons are shown or hidden in dependency on the
availability of their functions and the current analysis stage
Note: The general availability of function buttons in the Button
Toolbar (together with the corresponding keyboard shortcuts) is
controlled by settings in the Function Keys section of the
paragraph Application Profiles.
Status Bar
The Status Bar presents the general status of the instrument.
The text fields specify the names of the currently loaded application profile
and analytical method, the real name of the logged-in user, the type of the
performed analysis (measurement mode) and the current system state. The
buttons allows activation and deactivation of the Analytical View panels (
Sample Properties, Sample Identification, Analysis Results, Grade
Search Results). The / button indicates the current system status ( OK,
failure) and activates the Diagnosis View.
Push the F5 key or click on the Edit Smp toolbar button to open the Sample
Identification dialog.
After entering the required sample identification, accept the entry by a click on the
Accept toolbar button or by pushing the keyboard ENTER key in the last line of
the dialog. The sample identification is completed and the dialog closes. The
entered sample identification appears in the Analysis View.
A click on the Cancel toolbar button would discard the just entered sample
identification and preserve the previous one.
Note: The configuration of the sample identification format is described in
the section Sample IDs of the paragraph Application Profiles.
To perform a measurement, push the F2 key or click on the Start button in the
toolbar. If the option cleaning flush is installed and activated on your instrument
the following dialog could appear on the screen. For setup the cleaning flush
have a look in the config file Instrument_Settings.xcfg section ArFlush.
After the optional cleaning flush has been performed the cleaning flush dialog
disappears automatically and the progress of the measurement sequence is
visualized in the status bar.
A click on the Stop button or pushing the F3 key during the measurement is
processing, would abort the measurement.
Note: If the APF (Automatic Method Selection) is active, the appropriate
method is loaded during or immediately after the measurement.
SPECTRO Analytical Instruments GmbH 133 of 228
SPECTRO
The below illustration shows the Analysis View after three measurements have
been performed.
The presentation of the sample analysis results can be switched between four
modes: Default, Mean, Mean/RSD and Single. Use the corresponding buttons in
the toolbar or the keyboard shortcut keys for temporary selection to the requested
mode.
Mean (key E): simple statistics (averages without standard and relative
standard deviations) is shown.
Default (Escape): The display returns to the default mode preset in the
application profile.
Note: For the configuration of fonts and colors signalizing the states of
displayed values as well as for the composition of the statistics rows
(reported/mean values, limits, SD and RSD) see the Fonts and Colors in the
section Display of the paragraph Application Profiles.
The Analysis View allows limited manipulation of the presented results. It is
possible to delete and undelete single measurements for elimination of outliers
from statistics calculations. In certain circumstances the displayed (reported)
analysis results can be corrected manually.
Single measurement deletion
To delete a single measurement, select the requested measurement
and push the key D or Del at the keyboard (if the Delete/Undelete
button is activated in the application profile, the Delete button appears
on the single measurement selection and can be used alternatively).
The deleted single measurement is not removed from the view but
displayed “strikethrough”.
Note: Even the deleted single measurement results are not
SPECTRO Analytical Instruments GmbH 138 of 228
SPECTRO
To finish the analysis, push the button Finish in the toolbar or push the key F9 at
the keyboard.
Note: If activated in the application profile, the automatic printout and
export of the sample analysis results occurs, when the sample analysis
finishes.
The Analysis View is cleared and another analysis can be performed.
If activated in the loaded application profile, the button Output in the toolbar
provides the opportunity to manipulate the output settings of the application
profile without deletion of the already performed measurements of the current
sample. For the possibilities to modify the output settings please have a look on
the paragraph Edit Existing Application Profile sections Display, Print,
Export, Formatting.
The detailed results of the reproducibility test are presented in the Analytical
Results panel by marking outliers.
During the analysis, the single measurements reproducibility is tested for the first
time, when the count of single measurements reaches the preset minimum.
When the outliers are ignored at the statistics calculations and the reproducibility
test result is Inconclusive or Failed, the statistics is not displayed.
Depending on the configuration of the test, the single element values or entire
single measurements are interpreted and marked as outliers. The outliers can be
considered or ignored at the calculation of the statistics (averages).
Note: See the Fonts and Colors in the section Display of the paragraph
Application Profiles for configuration of fonts, colors and prefixes used to
mark outliers.
As default following font and colors are used for marking the outlier values and
outlier measurements.
Meaning Display
Outlier value. O 0.151
Non-outlier value in an outlier single measurement. 0.151
Note: If at the sample identification the field Grade Norm is used, it should
be filled before selecting the Grade Name.
The grade choice in the Sample Identification dialog is predefined in the Grade
Library of the instrument (see the paragraph Grades). However the currently
loaded analytical method can limit the grade selection. See the sections Grade
Filters and Grade Relations in the paragraph Methods for possible limitations.
After acceptance of the entered sample identification (button Accept in the
toolbar) the measurements can be started with a click on the button Start in the
toolbar or by pushing the key F2 at the keyboard.
Note that when the user attempts to start the first measurement and the required
grade specification in the sample identification is missing, a warning message
appears.
Then it is possible to ignore the message (button Yes) and continue the analysis
without grade verification (the button Start of the key F2 needs to be actuated
once more). If the button No is clicked, the Sample Identification dialog opens
and the user is requested to identify the sample.
After each measurement the grade verification is performed automatically.
Depending on the configuration, the verification concerns the last single
measurement results or the average results.
If any element concentration exceeds the control limits of the grade, the Grade
verification failed message is displayed and has to be confirmed by a click on
the Close button.
As default the limits used for grade verification are displayed above and below
the average (reported) values (<x>).
Depending on settings in the application profile, the Warning/Internal (W. Min and
W. Max) or Acceptance/External (A. Min and A. Max) Limits of the specified
grade are used. However the used limits type can be dynamically modified in the
measurement view. The corresponding option Use Ext/Int Grade Limits may
need to be activated in the section Function Keys of the application profile.
Note: For the configuration of fonts and colors signalizing the states of
displayed values as well as for the composition of the statistics rows
(reported/mean values, limits, SD and RSD) see the Fonts and Colors in the
section Display of the paragraph Application Profiles.
As default following font and colors are used for marking the results of the grade
verification.
Meaning Display
Grade verification succeeded for external limits. The 1.23
concentration value is within the grade limits.
Grade verification failed for external limits. The 1.23 +
concentration value exceeds the grade upper limit.
Grade verification failed for external limits. The 1.23 -
concentration value exceeds the grade lower limit.
SPECTRO Analytical Instruments GmbH 147 of 228
SPECTRO
If the automatic grade search is not activated in the application profile, the grade
search can be started manually with the click on the button Srch Grade in the
toolbar or by pushing the key Ctrl-F at the keyboard. The grade search concerns
then the selected results (single measurement or averages).
Note: You may need to activate the Srch Grade button at the Function Keys
page of the Application Profile Editor.
The found grades are presented in the Grade Search Results panel, which is
activated automatically by the search process. Note that in the Grade Search
Results panel the results corresponding to the results selected in the Analysis
Results panel are displayed. If the grade search was not performed for the
selected analysis results or the search process didn’t find any matching grade,
the list in the Grade Search Results panel is empty.
The result of the grade search may contain multiple grades. Any grade in the list
may be manually assigned to the current analysis results with a click on the
Apply Grade button in the toolbar. If the sample identification of the current result
contains the Grade Name field, the name of the selected grade is also put to the
field and the grade verification performed automatically.
The check sign at the assigned grade name indicates which grade is currently
assigned to the sample analysis results.
Note: You may need to activate the Apply Grade button at the Function
Keys page of the Application Profile Editor.
2.2.1.1.4 APF
The APF (Automatic Method Selection) is used for finding the most optimal
analytical method for analysis of the measured sample and thus, for improvement
of the analysis accuracy.
The first measurement of a sample is performed in an orientation method. Then,
based on the obtained results, a corresponding, more precise sub method is
automatically selected and the results are reevaluated using the measured
intensities.
The analytical configuration of the APF procedure has to be done in the APF
Editor. See the section APF in paragraph Settings for details.
The general activation of the APF analysis is provided in the application profile.
Start a new sample analysis. The first measurement is performed with the preset
orientation method.
Immediately after the measurement has finished, another, better fitting sub
method is loaded automatically. The analysis results are then reevaluated and
presented corresponding to the automatically selected sub method.
Start a new sample analysis. The first measurement is performed in the preset
orientation method.
Immediately after the measurement has finished, another, better fitting sub
method is loaded automatically. The analysis results are then reevaluated and
presented corresponding to the automatically selected sub method.
All following measurements and analyses are then performed in the automatically
selected sub method. Reloading of the startup method of the application profile
(orientation method) has to be performed manually. Only then the next analysis
begins again with the APF procedure.
SPECTRO Analytical Instruments GmbH 156 of 228
SPECTRO
Start a new sample analysis. The single measurement is performed in the preset
orientation method.
Immediately after the measurement has finished, another, better fitting sub
method is loaded automatically. The analysis results are reevaluated and
presented corresponding to the automatically selected sub method. Then the
analysis is finished automatically.
Note: The results presentation of the Single Measurement APF Analysis is
always prolonged up to the start of the next analysis.
The start of the next measurement reloads the startup method of the
application profile (the orientation method) and starts a new analysis. The APF
procedure is repeated again.
After the type correction has been activated, the name of the selected type
standard is displayed in the Sample Properties panel of the Analysis view, in the
field Type Standard. The field Correction Mode switches from None to Type
Corrected.
Note: You may need to activate the Sample Properties panel with a click on
the button in the status bar.
After selection of the type standard, the measurements can be started with a click
on the button Start in the toolbar or by pushing the key F2 at the keyboard.
In the list Select Type Standard the defined type standards (Sample Name) with
the time of their last analysis (Measure Date) are shown. To choose the standard
for the type standardization, double click on the requested standard name in the
list or select it and click on the button Apply below.
The name of the chosen standard appears in the header of the Analysis Result
panel.
To start the measurement, click on the button Start in the toolbar or push the key
F2 at the keyboard.
The limits used for type standardization are displayed above (acceptance
minimum A.Min) and below (acceptance maximum A.Max) the average values
(<x>).
Note: For the configuration of fonts and colors signalizing the states of
displayed values as well as for the composition of the statistics rows
SPECTRO Analytical Instruments GmbH 162 of 228
SPECTRO
(reported/mean values, limits, SD and RSD) see the Fonts and Colors in the
section Display of the paragraph Application Profiles.
As default following font and colors are used for marking the results of the type
standardization analysis.
Meaning Display
Acceptance limit verification succeeded. The 1.23
concentration value is within the type standard limits.
Acceptance limit verification failed. The concentration 1.23+
value exceeds the type standard upper limit.
Acceptance limit verification failed. The concentration 1.23-
value exceeds the type standard lower limit.
When the minimal count of single measurements has been reached, the analysis
of the type standard can be finished with a click on the button Finish or by
pushing the key F9 at the keyboard
After finishing the analysis, the Type Standardization Results dialog appears.
With a click on the button Accept the type standardization finishes and the
analyzed type standard is added to the list of available and valid type corrections.
In the list Select Standard Sample the defined control standards (Sample
Name) with the time of their last analysis (Measure Date) are shown. To choose
the standard for the control standard test, double click on the requested standard
name in the list or select it and click on the button Apply below.
The name of the chosen standard appears in the header of the Analysis Result
panel.
To start the measurement, click on the button Start in the toolbar or push the key
F2 at the keyboard.
The limits used for control standard test are displayed above (acceptance
minimum A.Min) and below (acceptance maximum A.Max) the average values
(<x>).
Note: For the configuration of fonts and colors signalizing the states of
displayed values as well as for the composition of the statistics rows
(reported/mean values, limits, SD and RSD) see the Fonts and Colors in the
section Display of the paragraph Application Profiles.
As default following font and colors are used for marking the results of the control
standard test analysis.
Meaning Display
Acceptance limit verification succeeded. The 1.23
concentration value is within the control standard limits.
Acceptance limit verification failed. The concentration 1.23-
value exceeds the control standard upper limit.
Acceptance limit verification failed. The concentration 1.23+
value exceeds the control standard lower limit.
When the minimal count of single measurements has been reached, the analysis
can be finished with a click on the button Finish or by pushing the key F9 at the
keyboard.
After the analysis has been finished, the next control standard is selected
automatically and the control test can be continued.
Note: If any element concentration exceeds the acceptance limits, when the
analysis is finished, the warning message Control Standard Test failed
appears and has to be confirmed by a click on the Close button.
If the last control standard has been finished, the control standard test finishes
automatically.
Note: Before start of the next sample analysis the control standard test can
be finished manually with a click on the Finish button below the list of the
control standards.
2.2.1.4 Options
2.2.1.4.1 Analysis Calculation Steps
The Analysis View can be used to trace the single calculation steps of the
analysis result calculations, from raw intensities up to concentrations (if applied,
up to type corrected concentrations).
Switching between presentation of the analysis results at different calculation
levels effects by pushing the – or + keys on the keyboard or by using the context
menu on the Analysis Results panel.
Note: To open the context menu hover the pointer over the top left cell of
the Analysis Result panel.
The analysis results on the following calculation levels can be presented:
Type Cor. Conc.: Type corrected element concentrations
Conc.: Element concentrations
Line Conc.: Line concentrations
Uncorr. Conc.: Uncorrected line concentrations
Status Display
Deleted Value 1.23
Analytical Error (the reference line intensity exceeds A 1.23 !
limits)
Math Error (the concentration calculation failed) M 1.23 !
Overflow (any related lines pixel region is saturated) O 1.23 !
Extended Status Inhomogeneous (not used) | 1.23
Extended Status Modified (the concentration was M 1.23
modified manually by the user)
Extended Status Outlier (the concentration is an outlier O 1.23
at the automatic reproducibility test)
Acceptance In Range (the concentration is within limits 1.23
at control standard test, type standardization or
verification of external grade limits)
Acceptance Over Range (the concentration exceeds 1.23 +
the upper limit at control standard test, type
standardization or verification of external grade limits)
Acceptance Under Range (the concentration exceeds 1.23 -
lower limit at control standard test, type standardization
or verification of external grade limits)
Acceptance Inconclusive (the concentration exceeds ? 1.23
the upper calibration limit; the reported concentration is
within limits at verification of external grade limits)
Warning In Range (the concentration is within limits at 1.23
verification of internal grade limits)
In addition to the visualization of the value states by fonts and colors, the
corresponding tooltip information is provided. To display the tooltip hover the
pointer over the value of interest.
Then, after the analysis has been finished, but is still available in the Analysis
View, you can use the function Pin Results and preserve the shown results from
being removed from the view on start of the next analysis.
Note: You may need to activate the results pinning related function buttons
in the section Function Keys of the Application Profile Editor.
The results pinned at the Analysis View are marked by the pin sign and can be
displayed at the usual modes (Single, Mean, Mean/RSD and Default) and
calculation levels (Type Corr. Conc., Conc., Line Conc. etc).
They can also be removed from the view individually (button Remove) or
generally (button Remove All).
If multiple results are pinned in the Analysis View, displaying of the corresponding
sample identifications can be activated (button Show/Hide IDs).
The pinning of results can also occur automatically, either when the analysis
finishes or when the next analysis starts. To activate this function, check the
option Pin results automatically in the section Display of the Application Profile
Editor.
The toolbar at the bottom of the Analysis View provides buttons for activation of
the available system functions. The functions availability depends on the current
analysis context and user’s permissions.
Additionally the functions availability and the corresponding buttons visibility can
be customized in section Function Keys of the Application Profile Editor.
Start a measurement
Gives access to the application profile’s output
parameters to modify the output of the current results.
The status bar at the bottom of the measurement screen informs the user briefly
about the instrument state. It contains following information:
1, 2, 3, 4, 5, 6, 7 , 8,9,10,11,12
2.2.2 iCALization
The iCALization is a procedure to ensure the optimal performance of the optical
detection system.
The iCAL View can also be activated using a function key F7 or a toolbar button
iCAL from within the Analysis View.
The sample to be analyzed will be shown at the top of the view written in red
foreground color.
A click on the Flush toolbar button will start an additional flush time.
A click on the Last Result toolbar button opens a view presenting the current
successful iCALization result.
Start a measurement by a click on the Start (F2) button or the appropriate
accelerator key. The measurement process can be aborted by pushing the Stop
(F3) button.
The origin screen (home screen or analysis view) will be displayed, after closing
the dialog.
The row ChipID of the selected detector gets focused and scrolled in the visual
range. The average chip intensities of previous measurements will be shown for
comparison with the actual pending result.
To display more details, push the More Details toolbar button in the header of
the Pending iCALization Result view.
After closing the dialog, the origin screen (home screen or analysis view) will be
displayed.
If the instrument is ready for use the LED symbol is green. On any device error,
the LED symbol color changes to red.
Note: Depending on the user’s permissions, the measurement start in the
error state of the instrument may be disabled.
Additionally, the LED symbol in the status bar serves as a button opening the
System Diagnosis View.
Note: If on the start of the software any error has been detected, the System
Diagnosis View opens immediately. The encountered errors are listed in the
message table beneath the iconized instrument presentation. Additionally,
an omnibus message box specifying software initialization errors can
appear.
Hovering with the pointer over the graphics of the Readout System opens a
tooltip containing information about the current device state and controllers
connected to the readout system. In case of a failure, the current error, last error
and process error types are specified.
Errors occurring during the measurement process can be temporary and are
latched. They are displayed as an error text at the end of the measurement. By
closing and reopening the System Diagnosis View the temporary errors are reset
and the default status is restored. Because of the nature of temporary errors, they
don’t prevent continuing of measurements.
If errors occur persistently, please contact the SPECTRO service.
Note: The readout system plays a key role in communication between the
PC and the instrument electronics. Thus, if a communication break occurs
(e.g. network cable break), the System Diagnosis obtains no information
about the instrument. In such case the known (presented) component
information is frozen and only the Readout System error is signalized.
2.3.1.2.2 EK_0014
The graphics representing the EK_0014 is shown at the left part of the view,
outside the iconized instrument picture.
Hovering with the pointer over the graphics of the EK_0014 opens a tooltip
containing the information about the current device state and the installed
Firmware version. In case of a failure the current and last error types are
specified.
Hovering with the pointer over the graphics of the Air Optics opens a tooltip
containing the information about the current device state, current and nominal
optics temperatures as well as about the current output of the temperature
stabilization unit.
Note: After the instrument has been switched on, it takes a while before the
air optics has reached its operating temperature.
2.3.1.2.4 UV Optics
The graphics representing the UV Optics is shown in the center of the iconized
instrument picture, behind the spark stand.
Hovering with the pointer over the graphics of the UV Optics opens a tooltip
containing the information about the current device state, current and nominal
optics temperatures as well as about the current output of the temperature
stabilization unit.
Note: After the instrument has been switched on, it takes a while before the
UV optics has reached its operating temperature.
Note: The status of the unit switch is signaled by green (On) or red (Off)
coloring of the Plasma Generator graphics. If the plasma generator is on
but interlocked (e.g. the clamp is up), its graphics becomes yellow colored.
Hovering with the pointer over the graphics of the Plasma Generator opens a
tooltip containing the information about the current device status and the state of
the unit switch. If the unit is On, the status of the interlock circuit, the supply
voltages (Vcc, and Vdd), the heat sink temperature and the unit Firmware version
are displayed. In case of a failure the current error, last error and process error
types are specified.
Note: If the plasma generator is off, neither Clamp nor Power Supply states
can be determined.
Hovering with the pointer over the graphics of the Spark Stand opens a tooltip
containing the information about the stark stand cleaning request (only if the
corresponding data were specified in the Maintenance Reminder system), the
status of the Argon Saver, the shutter position, the mask activity and the state of
the spark stand cover.
Note: The spark stand cleaning request of the Maintenance Reminder
System is signaled by yellow coloring of the Spark Stand graphics.
2.3.1.2.7 Clamp
The graphics representing the Clamp is shown at the iconized instrument picture
above the spark stand.
Hovering with the pointer over the graphics of the Clamp opens a tooltip
indicating the current status of the clamp.
In the normal state of the Clamp (clamp is down) its graphics is not colored.
However, when the clamp safety test (i.e. a short break of the security circuit) is
requested or the clamp is up, its graphics becomes red.
Due to interlinking in the instrument electronics the Plasma Generator is put in
the Interlocked state as soon as the Clamp has been moved up (the security
circuit has been broken).
Note: The clamp status cannot be detected, when the plasma generator is
switched off. An undefined status of the clamp is signaled by yellow
coloring of its graphics.
Hovering with the pointer over the graphics of the Power Supply opens a tooltip
indicating the main supply voltage of the instrument electronics.
Note: The main supply voltage of the instrument electronics cannot be
determined, when the plasma generator is switched off.
Hovering with the pointer over the graphics of the Argon Supply opens a tooltip
containing the current information about the device status, the argon pressure,
the state of the argon flow valves and the status of the Argon Saver.
Note: If the argon pressure is in range the Argon Supply graphics is green,
otherwise it becomes red.
Hovering with the pointer over the graphics of the Argon Exit System opens a
tooltip containing the information about the argon exit maintenance request.
Note: The argon exit maintenance request is signaled by yellow coloring of
the Argon Exit System graphics.
Hovering with the pointer over the graphics of the Air Filter opens a tooltip
containing the information about the filter maintenance request.
Note: The air filter maintenance request is signaled by yellow coloring of
the Air Filter graphics.
2.3.1.2.12 Hydrosorb
The graphics representing the Hydrosorb (gas purifier cartridge) is shown at the
right part of the view, outside the iconized instrument picture.
Hovering with the pointer over the graphics of the Hydrosorb opens a tooltip
containing the information about the replacement request of the Gas Purifier
Cartridge.
Note: The gas purified cartridge replacement request is signaled by yellow
coloring of the Hydrosorb graphics.
The view provides a separate property page for each defined maintenance. Click
on the tab with a maintenance name at the top of the view to access the
requested data.
Each maintenance property page has the same structure.
In the group Maintenance Status the fields Performed on and at
Measurements Counter specify the time and the measurements counter state at
the last maintenance, respectively. The Passed Interval progress bar presents
graphically the maintenance interval expiration. When an interval has expired, the
field Expired appears and indicates the due time or the due measurements
counter state of the maintenance.
If allowed (depending of the maintenance type and your user permissions), you
can confirm the maintenance and reset the interval counters (Set Done button).
In the group Maintenance Intervals the Time and Counter intervals can be
adjusted. Checking or unchecking the Remind checkbox activates or deactivates
the remind message function. Additionally the predicted time or measurements
counter state of the next maintenance reminder is indicated.
Note: If you wish to deactivate one of the criteria set the value to 0. The
software will deactivate the control of this interval criterion.
In the group Current Instrument Status the current time (Date & Time), the
current measurements counter state (Measurements Counter) and the
maintenance age (Age) are shown.
2.3.2.3.1 iCALization
The iCALization automatic reset mechanism synchronizes the last maintenance
time with the last time of the iCALization. Thus the iCALization maintenance
reminder is Set Done automatically.
Check the Set Done automatically checkbox to activate this mechanism.
Note: If the Set Done automatically is active, the maintenance cannot be
reset manually.
analytical method. Thus the Control Standard Test maintenance reminder is Set
Done automatically.
Check the Set Done automatically checkbox to activate this mechanism.
Note: If the Set Done automatically is active, the maintenance cannot be
reset manually.
Note: The time of the last Control Standard Test is specific for the analytical
method (here, the method Fe-10-M).
This confirmation triggers then automatically the Set Done action of the Spark
Stand Cleaning maintenance reminder.
Check the Warn on base change checkbox at the Spark Stand Cleaning page
for activation of this mechanism.
The maintenance reminder message displays the due maintenances and the
expiration time or counter state.
To confirm, that maintenance has been performed, select the corresponding
entry in the message list and click the Done button. The interval counter of the
corresponding maintenance will be reset.
Note: The Done button corresponds to the Set Done button at the
maintenance control settings.
It is possible to postpone a reminder by selection of the corresponding entry in
the maintenance list and specifying the delay time in the combo box at the bottom
of the reminder message box and clicking on the button Remind. After that, the
reminder message will be displayed after the specified time or at the next
software start.
Closing the reminder message box by pushing the Close button will reactivate
the message 10 seconds later.
If multiple maintenance actions are reminded simultaneously you can click on the
button All done to confirm all listed maintenances have been done.
Note: The confirmation option can be unavailable. This situation can arise,
when the synchronization of the reminder with the maintenance action is
active (Native Controller) for actions like iCALization, Control Standard test
or Spark Stand Cleaning, when the software is able to detect the execution
of the maintenance action.
Note: The confirmation option of the General Maintenance reminder is
available for the authorized SPECTRO Service personal only.
If the meassage has been confirmed with Yes the Pollution Units Counter of
the maintenance reminder will be reset automatically. Otherwise the counter
keeps its current state.
Note: The user permissions for the General Maintenance are limited. Its
confirmation and interval modification are reserved for the SPECTRO
Service.
The view presents the configuration (Instrument No., Spark Analyzer Pro MAXx
version, Firmware versions of the hardware components) and the current status
(instrument temperatures, maintenance time etc.) of the instrument.
Select the folder where you want to store the report and click the OK button.
The service report archive with the name ServiceReport_yyyyMMdd_hhmmss
(where the yyyyMMdd_hhmmss is the current date and time) is created in the
selected folder and a confirmation message will appear.
Now you can send this report archive to the SPECTRO Customer Support
Center.
2.3.4 Service
This module provides the possibility to check the state and functionality of the
argon valve instrument components.
2.3.4.1 Instrument
The page Instrument shows the states of the different Argon valves.