#Parameters
Working voltage : 12VDC- 24VDC
Output signal: (0-5V, 0-10V, 1-5V, ± 5V) 4-20MA
Operating temperature range : -20 -- 80
Operating Humidity : 10% RH-90% RH ( non-condensing )
Accuracy : better than 0.1% F.S
Precision is better than 0.05% F.S
Signal outputs
24V: transmitter power +
IO: current signal output
VO: voltage signal output
GND: transmitter Power -
1 zero adjustment
First, the system is in a state of no-load, when the input signal is the current, adjust the mark "RV3" potentiometer to
adjust the deviation of the zero point. When the input signal is voltage, adjust the mark "RV1" potentiometer to adjust
the deviation of the zero point.
To adjust the display value (current or voltage) to zero.
2 gain adjustment
The system is loaded with a certain weight (recommended range: 80%~100%FS), according to the ratio of the weight of
the load and the maximum value of the calculated value of the relationship, the adjustment is marked as "RV2" or "RV4"
potentiometer,adjust amplifier gain,Adjust the amplifier gain, and adjust the display value to the load ratio.
3 Review
The system is in a state of no-load, zero output measurement; to load a certain weight value, measuring the output
voltage is in line with the expected magnification. If different, please repeat steps 1, 2 once again.
Note: after the adjustment is completed, the adjustment end of the potentiometer is required to be fixed with sealant,
Prevent vibration and other factors to change the resistance value of the potentiometer.