Professional Documents
Culture Documents
Toaz - Info Mek7300k SM A PR
Toaz - Info Mek7300k SM A PR
0634-900772A
Printed: 2013/06/27
Automated
Hematology Analyzer
MEK-7300K
Keep this manual near the instrument or in the reach of the operator and refer to it whenever the operation is unclear.
Copyright Notice
The entire contents of this manual are copyrighted by Nihon Kohden. All rights are reserved. No part of this document
may be reproduced, stored, or transmitted in any form or by any means (electronic, mechanical, photocopied, recorded,
or otherwise) without the prior written permission of Nihon Kohden.
Trademark
The mark printed on the SD card that is used in this instrument is a trademark. The company name and model name are
trademarks and registered trademarks of each company.
Contents 1
6
Section 1 General.................................................................................... 1.1
Introduction....................................................................................................................... 1.2 7
Service Policy................................................................................................................... 1.3
Specifications.................................................................................................................... 1.4
Measured Parameters, Ranges and Reproducibility to Specimen............... 1.4
Detection Method......................................................................................... 1.4
Standardization Analysis Method................................................................. 1.5
Dilution Ratio................................................................................................ 1.5
Counting Time.............................................................................................. 1.5
Display.......................................................................................................... 1.5
Data Storage................................................................................................ 1.5
Environmental Conditions............................................................................. 1.5
Power Requirements.................................................................................... 1.6
Dimensions and Weight................................................................................ 1.6
Electromagnetic Compatibility...................................................................... 1.6
Safety........................................................................................................... 1.6
Bar Code Specifications............................................................................... 1.7
Panel Description.............................................................................................................. 1.8
Front Panel............................................................................................................. 1.8
Right Side Panel..................................................................................................... 1.9
Left Side Panel..................................................................................................... 1.10
Top Panel.............................................................................................................. 1.10
Rear Panel............................................................................................................ 1.11
Caution Labels on the Analyzer...................................................................................... 1.12
Composition.................................................................................................................... 1.14
Standard Composition.......................................................................................... 1.14
The following describes some common interference sources and remedial actions:
1. Strong electromagnetic interference from a nearby emitter source such as an authorized radio station
or cellular phone:
Install the equipment and/or system at another location if it is interfered with by an emitter source
such as an authorized radio station. Keep the emitter source such as cellular phone away from the
equipment and/or system.
2. Radio-frequency interference from other equipment through the AC power supply of the equipment and/
or system:
Identify the cause of this interference and if possible remove this interference source. If this is not
possible, use a different power supply.
4. Electromagnetic interference with any radio wave receiver such as radio or television:
If the equipment and/or system interferes with any radio wave receiver, locate the equipment and/or
system as far as possible from the radio wave receiver.
If the above suggested remedial actions do not solve the problem, consult your Nihon Kohden
representative for additional suggestions.
This equipment complies with International Standard EN 55011: 2007 Group 1, Class B. Class B EQUIPMENT is
equipment suitable for use in domestic establishments and in establishments directly connected to a low voltage
power supply network which supplies buildings used for domestic purposes.
The CE mark is a protected conformity mark of the European Community. Products with the CE mark comply
with the requirements of the Medical Device Directive 93/42/EEC.
NOTE about Waste Electrical and Electronic Equipment (WEEE) directive 2002/96/EC
For the member states of the European Union only:
The purpose of WEEE directive 2002/96/EC is, as a first priority, the prevention of waste electrical and
electronic equipment (WEEE), and in addition, the reuse, recycling and other forms of recovery of such
wastes so as to reduce the disposal of waste.
Contact your Nihon Kohden representative for disposal.
WARNING
A warning alerts the user to the possible injury or death associated with the use or misuse of the instrument.
CAUTION
A caution alerts the user to possible injury or problems with the instrument associated with its use or
misuse such as instrument malfunction, instrument failure, damage to the instrument, or damage to other
property.
NOTE
A note provides specific information, in the form of recommendations, prerequirements, alternative methods or
supplemental information.
Introduction........................................................................................................................................................... 1.2
Service Policy....................................................................................................................................................... 1.3
Specifications....................................................................................................................................................... 1.4
Measured Parameters, Ranges and Reproducibility to Specimen................................................... 1.4
Detection Method............................................................................................................................. 1.4
Standardization Analysis Method..................................................................................................... 1.5
Dilution Ratio.................................................................................................................................... 1.5
Counting Time.................................................................................................................................. 1.5
Display............................................................................................................................................. 1.5
Data Storage.................................................................................................................................... 1.5
Environmental Conditions................................................................................................................ 1.5
Power Requirements........................................................................................................................ 1.6
Dimensions and Weight................................................................................................................... 1.6
Electromagnetic Compatibility.......................................................................................................... 1.6
Safety............................................................................................................................................... 1.6
Bar Code Specifications................................................................................................................... 1.7
Panel Description................................................................................................................................................. 1.8
Front Panel................................................................................................................................................. 1.8
Right Side Panel......................................................................................................................................... 1.9
Left Side Panel......................................................................................................................................... 1.10
Top Panel................................................................................................................................................. 1.10
Rear Panel............................................................................................................................................... 1.11
Caution Labels on the Analyzer.......................................................................................................................... 1.12
Composition........................................................................................................................................................ 1.14
Standard Composition.............................................................................................................................. 1.14
Introduction
CAUTION
To maintain the instrument in normal condition, the user must
perform the periodic maintenance. Refer to “Maintenance” of the
operator’s manual.
NOTE
If the instrument has a problem and there has been no periodic
maintenance, the instrument will usually be normal again by cleaning the
fluid paths or replacing a consumable with a new one.
The information in the operator’s manual is primarily for the user. However, it
is important for service personnel to thoroughly read the operator’s manual and
service manual before starting to troubleshoot, service, maintain or repair this
instrument. This is because service personnel needs to understand the operation
of the instrument in order to effectively use the information in the service
manual.
1
Service Policy
CAUTION CAUTION
Be careful not to directly touch any place where Wear rubber gloves to protect yourself from
blood is or may spread to. infection before doing maintenance.
NOTE
• When ordering parts or accessories from your nearest Nihon Kohden
Corporation’s representative, please quote the NK code number and
part name which is listed in this service manual, and the name or model
of the unit in which the required part is located. This will help us to
promptly attend to your needs.
• Always use parts and accessories recommended or supplied by Nihon
Kohden Corporation to assure maximum performance from your
instrument.
Specifications
Reproducibility to Specimen
Measured Parameters Measuring Range
(CV: Coefficient of Variation)
WBC: White blood cell count 0 to 299 × 103/µL within 2.0%CV
NE%: Neutrophil percent 0 to 99.9% within 5.0%CV
LY%: Lymphocyte percent 0 to 99.9% within 5.0%CV
MO%: Monocyte percent 0 to 99.9% within 12.0%CV
EO%: Eosinophil percent 0 to 99.9% within 20.0%CV
within CV30.0% (>2%) or average value ±1% (0
BA%: Basophil percent 0 to 99.9%
to 2%)
NE: Neutrophil count 0 to 299 × 103/µL
LY: Lymphocyte count 0 to 299 × 103/µL
MO: Monocyte count 0 to 299 × 103/µL —
EO: Eosinophil count 0 to 299 × 103/µL
BA: Basophil count 0 to 299 × 103/µL
RBC: Red blood cell count 0 to 14.9 × 106/µL within 1.5%CV
HGB: Hemoglobin concentration 0 to 29.9 g/dL within 1.5%CV
HCT: Hematocrit 0 to 99.9% —
MCV: Mean cell volume 20 to 199 fL within 1.0%CV
MCH: Mean cell hemoglobin 10 to 50 pg
MCHC: Mean cell hemoglobin concentration 10 to 50 g/dL —
RDW: Red blood cell distribution width 0 to 50%
PLT: Platelet count 0 to 1490 × 103/µL within 4.0%CV
PCT: Platelet crit 0 to 2.9%
MPV: Mean platelet volume 0 to 20.0 fL —
PDW: Platelet distribution width 0 to 50%
Detection Method
Blood cell count: Electrical resistance detection
Hemoglobin: Surfactant method (colorimetric method)
Hematocrit: Histogram calculation
WBC population: Light scatter by laser
Platelet crit: Histogram calculation
RBC distribution width: Histogram calculation
Platelet distribution width: Histogram calculation
Dilution Ratio
• Venous blood
Sample volume: 55 μL (for 23 parameters)/30 μL (for WBC, RBC, HGB, HCT, MCV, MCH, MCHC, and PLT)
WBC/HGB: 200:1
RBC/PLT: 40,000:1
• Pre-dilution blood
Sample volume: 10 μL 20 μL
WBC/HGB: 1200:1 600:1
RBC/PLT: 240,000:1 120,000:1
Counting Time
Open mode: 63 s/sample (from measurement start to data display)
Closed mode: 75 s/sample (from measurement start to data display)
Display
Display: 10.4 inch, LCD with backlight and touch screen keys
Resolution: 800 × 600 dots
Screen size: approx. 211.2 × 158.4 mm
Display contents: Numerical data, scattergrams, histograms, measuring conditions, alarm message and other messages,
touch screen keys
Data Storage
Numerical data for all counted parameters for up to 400 samples and histograms and scattergrams for up to 50 samples
Environmental Conditions
Storage temperature: –20 to +60ºC (–4 to +140°F)
Operating temperature: 15 to 30ºC (59 to 86°F)
Storage humidity: 10 to 95%
Operating humidity: 30 to 85% (noncondensing)
Storage atmospheric pressure: 70 to 106 kPa
Operating atmospheric pressure: 70 to 106 kPa
Operating altitude: less than 3000 m
Power Requirements
Power requirements: 110 to 240 V ± 10% AC, 50/60 Hz
Power consumption: 250 VA
Electromagnetic Compatibility
IEC 61326-1: 2005
IEC 61326-2-6: 2005
EN 61326-1: 2006
EN 61326-2-6: 2006
The power supply short interruption test is performed through a transformer which has at least three times the power
capacity of the instrument.
Safety
Safety standards: IEC 61010-1 2nd Edition: 2001
EN 61010-1 2nd Edition: 2001
CAN/CSA-22.2 No. 61010-1 2nd Edition: 2004
UL 61010-1 2nd Edition: 2004
IEC 61010-2-081: 2001
IEC 61010-2-101: 2002
CAN/CSA-22.2 No. 61010-1-04 + No. 61010-2-101-04
EN 61010-2-101: 2002
Laser: IEC 60825-1: 1993 Amendment 1: 1997, Class 1
EN 60825-1: 1994 Amendment 11: 1996
Type of protection against electrical shock: CLASS I EQUIPMENT
Degree of protection against harmful ingress of water: IPX0 (non-protected)
Degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH
OXYGEN OR NITROUS OXIDE: EQUIPMENT not suitable for use in the presence of FLAMMABLE ANAESTHETIC
MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE
Mode of operation: CONTINUOUS OPERATION
EQUIPMENT types (classification): Indoor stationary EQUIPMENT
Pollution Degree: 2 EQUIPMENT
Panel Description
Front Panel
9
1
2 10
3 11
6
7
13
5
4 14
12
15
16
Top Panel
Rear Panel 1
4 6
7
1
8
2
9
3
NOTE
• Replace the filters periodically.
• When attaching the filter joint assembly, be careful
not to bend or damage the filter packing at the
bottom of the measurement bath.
• When there is a leakage, check that there is no
scratch or damage to the circumference of the filter.
Composition
Standard Composition
UT-7242 POWER BD
UT-7243 DRIVER BD
UT-7244 ANALOG BD
UT-7250 CONNECTION BD
1
MO-820V2 LASER OPTICAL UNIT
UT-7172 MO PREAMP BD
UT-7173 MO FS PREAMP BD
UT-7174 MO FL PREAMP BD
UT-7175 MO SD PREAMP BD
UT-7176 MO LD DRIVER BD
UT-7246 LCD BD
UT-7247 KEY BD
Check Flowchart
2
YES
YES
YES
YES
Measure MEK-5D hematology control • Check the damaged unit from alarm and error
• No alarm occurs • Check accuracy
• Reproducibility is fine NO
• PLT
• HGB
• Parameters related to WBC and PLT
• HGT and MCV
• Data is correct (within the assay value limits)
YES
YES
YES
Finish
Operating environment
Temperature: 15 to 30ºC (59 to 86°F)
Humidity: 30 to 85%
Atmospheric pressure: 70 to 106 kPa
Even when the room temperature is within the above range in the winter, the
diluent temperature may be low because the diluent gets cold at night. To
eliminate cold temperature from the floor, put an insulator such as polystyrene
foam under the diluent or use a pet heater. The diluent temperature must be
managed carefully.
Special Samples
Be careful about “Interference Substances” as described in Section 5 of the
operator’s manual. They may affect the measurement value.
Leak
First, you must find the leak. Find the leak from the flow or crystals of liquid.
Leak may occur in the following places except for the valve itself and connection
joint.
i) Waste tube
ii) WBC measurement bath
iii) RBC measurement bath
iv) Rinse chassis for open mode
v) Rinse chassis for closed mode
i) Waste tube
• When the leak occurs from the waste tube or waste port, a clog may occur
between the waste container and tube. If the analyzer is used continuously,
waste leaks continuously and infection may occur. Remove the cause.
• Check that the tube is not removed, bent or pressed under the analyzer.
Filter
Filter
Case
iv) When the leak occurs from the rinse chassis for open mode, check the
following.
• The drain tube is not pressed or bent.
• The O-ring of the rinse chassis is not deteriorated.
• The right MP-640V rotary pump rotates normally.
• The 2-way valves 22, 23, 24 of the JQ-730V inlet/outlet unit operate
normally (check on the maintenance screen).
v) When the leak occurs from the rinse chassis for closed mode, check the
following.
• The drain tube is not pressed or bent.
• The right MP-640V rotary pump rotates normally.
• The 2-way valves 22, 23 and 24 of the JQ-730V inlet/outlet unit operate
normally (check on the maintenance screen).
• The 2-way valves 5 and 6 of the JQ-732V valve unit operate normally
(check on the maintenance screen).
Error Codes
Following is a list of the error codes and the probable trouble spot.
E001: Blood Pump Initial Error E082: Lyse Pump Movement Error
E003: Blood Pump Movement Error E101: Right Rotary Pump Initial Error
E021: Sampler Initial Error E102: Left Rotary Pump Initial Error
E022: Vertical Sampler Movement Error E122: Check Settings
E023: Horizontal Sampler Movement Error E124: CBC Circuit Error
E041: WBC Subbus Initial Error E125: 5Diff Circuit Error
E042: WBC Subbus Movement Error E127: Watch Dog Timer Error
E051: RBC Subbus Initial Error E128: Mix Chamber Warmer Control Error
E052: RBC Subbus Movement Error E131: Sheath High Pressure Error
E061: Iso Pump Initial Error E132: Sheath Pressure Sensor Error
E062: Iso Pump Movement Error E141: Cap Pierce Initial Error
E071: Sample Pump Initial Error E142: Cap Pierce Movement Error
E072: Sample Pump Movement Error E145: Tube Holder Movement Error
E081: Lyse Pump Initial Error
Detection sensor
Upper position sensor (U014) of the MIXED PUMP BD (UT-7200) on the MP-
730V
Detection method
The sensor board is not detected 15 seconds after moving the motor of the MP-
730V upward.
The MP-730V has two cylinders (diluter and sheath pump) and they operate
separately. The upper position of the diluter cylinder is detected by the sensor
board which moves with the motor. When it reaches the slit part, it cuts the
sensor light. The moving distance of the diluter cylinder is detected by the sensor
disk as it passes the slit of the encoder sensor.
SteppingUT-7200
motor
Cause
位置決め冶具取付位置
i) When the motor does not rotate
位置決め冶具取付位置
• The drive part is caught. → Remove the cause of catch.
• Poor contact or break of the cable → Disconnect and connect the connector.
Check continuity of the cable.
• Drive IC of the motor failure → Replace UT-7243 DRIVER BD.
• Motor failure → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or break of the cable → Disconnect and connect the connector.
Check continuity of the cable.
ENCODE and ORIGIN LEDs light ENCODE LED is off
ENCODE and ORIGIN LEDs light
• Sensor signal receive IC failure. → Replace the UT-7200 MIXED PUMP
ENCODE LED is off
BD.
Detection sensor
Upper position sensor of the sheath pump
Name: KI1234-AALF photosensor (code no.: 676769A)
Pin position
1 +5V
2 Sensor signal
3 GND
Sensor board
The MP-730V has two cylinder (diluter and sheath pump) and they operate
separately. The upper position sensor detects the upper drive for initialization of
the sheath pump cylinder. The drive is detected by the sensor board which moves
with the motor reaches the slit of the sensor and cuts the sensor light.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or break of the cable → Disconnect and connect the
connector. Check continuity of the
cable.
• Failure of the drive IC of the motor → Replace UT-7243 DRIVER BD.
• Motor failure → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Sensor failure. → Replace the sensor.
• Sensor signal receive IC failure. → Replace the UT-72473 DRIVER BD.
Detection sensor
Sheath pump upper position sensor and lower position sensor
Name: KI1234-AALF photosensor (code no.: 676769A)
Pin position
1 +5V 2
2 Sensor signal
3 GND
Pin position
1 +5V
2 Sensor signal
Detection method 3 GND
The upper position sensor does not detect the pass of the sensor board even 15
seconds after the motor of the sheath pump operates from the upper position.
MP-730V has two cylinders (diluter and sheath pump) and they operate
separately. The upper and lower position sensor detect the up and down drive of
the sheath pump cylinder of the MP-730V. The drive is detected by the sensor
board which moves with the motor. When it reaches the slit of the sensor, it cuts
the sensor light.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or break of the cable → Disconnect and connect the
connector. Check continuity of the
cable.
• Failure of the drive IC of the motor → Replace UT-7243 DRIVER BD.
• Motor failure → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or break of the cable → Disconnect and connect the
connector. Check cable continuity.
• Sensor signal receive IC failure → Replace the UT-7200 MIXED PUMP
BD.
Detection sensor
PD0101 upper position sensor on the SAMPLER SENSOR BD
Detection method
The upper position sensor does not detect the sensor board 15 seconds after the Y
axis motor drives to upward.
Sensor board
Middle
中 position sensor
Encoder
sensor
MP-730V moves the sampling nozzle with the two motors (X axis and Y axis). Y
axis position is detected by the sensor board which moves with the motor until it
reaches the slit of the sensor on the UT-7235 SAMPLER SENSOR BD and cuts
the sensor light. The X axis position is detected by the sensor board which moves
with the motor until it reaches the slit of the sensor on the UT-7235 SAMPLER
SENSOR BD. The light of sensor 1 to 3 passes through the holes on the sensor
board.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or break of the cable → Disconnect and connect the
connector. Check continuity of the
cable. Flexible cable is deteriorated.
• Failure of the drive IC of the motor → Replace UT-7243 DRIVER BD.
• Motor failure. → Replace the unit.
ii) When the motor rotates but the sensor cannot detect it
• Poor contact or cable break → Disconnect and connect the connector. Check
continuity of the cable. Replace the cable.
Detection sensor
PD0101 upper position sensor, PD0102 middle position sensor and PD0103
lower position sensor on the UT-7235 SAMPLER SENSOR BD
Detection method
The upper, middle and lower position sensors of the UT-7235 do not detect the
sensor board 15 seconds after the Y axis motor drives. Or the pass of the sensor
board is not detected.
Sensor board
Middle
中 position sensor
Encoder
sensor
MP-730V moves the sampling nozzle to the default position with the two motors
(X axis and Y axis). The Y axis position is detected by the sensor board which
moves with the motor until it reaches the slit of the sensor on the UT-7235
SAMPLER SENSOR BD and cuts the sensor light.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Drive IC of the motor failure → Replace UT-7243 DRIVER BD.
• Motor failure → Replace the unit.
ii) When the motor rotates but the sensor cannot detect it
• Poor contact or cable break → Disconnect and connect the connector.
Check continuity of the cable. Replace
the cable
• Sensor signal receive IC failure → Replace the UT-7234 SAMPLER
JUNCTION BD or UT-7243
DRIVER BD.
Detection sensor
PD0103 to PD0106 position sensor on the UT-7235 SAMPLER SENSOR BD
Detection method
The sensor of the UT-7235 does not detect the slit on the sensor board 15
seconds after the X axis motor drives or the detected position code is not correct.
Sensor board
Middle
中 position sensor
Encoder
sensor
MS-730V moves the sampling nozzle to the default position with the two motors
(X axis and Y axis). The X axis position is detected by the sensor board which
moves with the motor until it reaches the slit of the sensor on the UT-7235
SAMPLER SENSOR BD and the light of sensor 1 to 3 passes through the holes
on the sensor board.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the
connector. Check cable continuity.
The relation of the sampler unit and LED for checking the sensor
5 diff dispense ● ● ○ ○
Cap pierce ● ○ ● ○
Encoder sensor
[P D 0103 to 106 ] RBC aspiration ○ ● ● ○
WBC dispense ● ○ ○ ○
Detection sensor
WBC sub bath sensor
Name: KI1234-AALF photosensor (676769A)
Pin position
1 +5V
2 Sensor signal
3 GND
Sensor board
The drive of the sub bath is detected by the each WBC and RBC sub bath sensor.
The drive is detected by the disk slit on the sensor board which moves with the
motor until it reaches the detection part of the sensor and passes the sensor light.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the
connector. Check cable continuity.
• Failure of the drive IC of the motor → Replace UT-7243 DRIVER BD.
• Motor failure or the connector is deteriorated. → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check continuity of the cable. Replace
the cable.
• Sensor failure → Replace the sensor.
• Sensor signal receive IC failure. → Replace the UT-7243 DRIVER BD.
Detection sensor
RBC sub bath sensor
Name: KI1234-AALF photosensor (676769A)
Pin position
1 +5V 2
2 Sensor signal
3 GND
Pin position
1 +5V
2 Sensor signal
3 GND
Detection method
The sub bath sensor does not detect the slit ofSensor
the sensor board 15 seconds after
board
the motor drives.
Sensor board
The drive of the sub bath is detected by the each WBC and RBC sub bath sensor.
The drive is detected by the disk slit on the sensor board which moves with the
motor until it reaches the detection part of the sensor and passes the sensor light.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Drive IC of the motor failure → Replace UT-7243 DRIVER BD.
• Motor failure or the connector is deteriorated. → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check continuity of the cable. Replace
the cable.
• Sensor failure → Replace the sensor.
• Sensor signal receive IC failure → Replace the UT-7243 DRIVER BD.
Detection sensor
Upper position sensor of the sample pump
Name: KI1234-AALF photosensor (code no.: 676769A)
Pin position
1 +5V
2 Sensor signal
3 GND
Pin position
1 +5V
2 Sensor signal
3 GND
Detection method
The upper position sensor does not detect the sensor board 15 seconds after the
Upper position sensor
motor of the sample pump moves to up.
Sensor board
Sensor board
Lower position sensor
The upper and lower position sensors detect the sample pump cylinder of the
MP-731V up and down drive (the sensors are different from the sensors of the
lysing reagent cylinder pump). The drive is detected by the sensor board which
moves with the motor until it reaches the slit of the sensor and cuts the sensor
light.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the
connector. Check cable continuity.
• Failure of the drive IC of the motor → Replace UT-7243 DRIVER BD.
• Motor failure → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Sensor failure → Replace the sensor.
• Sensor signal receive IC failure → Replace the UT-7243 DRIVER BD.
Detection sensor
Sample pump upper and lower position sensor
2
Detection method
The upper position sensor does not detect the pass of the sensor board 15 seconds
after the motor of the sample pump drives from the upper position. Or the
encoder sensor does not turn on or off for the predetermined number of times.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Drive IC of the motor failure → Replace UT-7243 DRIVER BD.
• Motor failure → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Sensor failure → Replace the sensor.
• Sensor signal receive IC failure → Replace the UT-7200 MIXED PUMP
BD.
Detection sensor
Upper position sensor of the lysing reagent pump
Detection method
The upper position sensor does not detect the sensor board 15 seconds after the
motor of the lysing reagent pump drives to up.
The motor sensor is common for the two lysing reagent pump syringes.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Drive IC of the motor failure → Replace UT-7243 DRIVER BD.
• Motor failure → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
Detection sensor
Lysing reagent pump upper and lower position sensor
Detection method
The upper position sensor does not detect the pass of the sensor board 15 seconds
after the motor of the lysing reagent pump drives from the upper position. Or the
encoder sensor does not turn on or off for the predetermined number of times.
Cause
i) When the motor does not function
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Drive IC of the motor failure → Replace UT-7243 DRIVER BD.
• Motor failure → Replace the unit.
ii) When the motor functions but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Sensor failure → Replace the sensor.
• Sensor signal receive IC failure → Replace the UT-7200 MIXED PUMP
BD.
Detection sensor
Rotary pump sensor
Detection method
The sensor does not detect the slit of the sensor board even when the motor of
the rotary pump drives.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or break of the cable → Disconnect and connect the connector.
Check continuity of the cable. Replace
the cable.
• Sensor failure → Replace the sensor.
• Sensor signal receive IC failure. → Replace the UT-7243 DRIVER BD.
Detection sensor
Rotary pump sensor
Detection method
The sensor does not detect the slit of the sensor board even when the motor of
the rotary pump drives.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or break of the cable → Disconnect and connect the connector.
Check continuity of the cable.
• Drive IC of the motor failure → Replace UT-7243 DRIVER BD.
• Motor failure or deterioration → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check continuity of the cable. Replace
the cable.
• Sensor failure → Replace the sensor.
• Sensor signal receive IC failure → Replace the UT-7243 DRIVER BD.
Detection sensor
Upper position sensor
Detection method
The upper position sensor does not detect the sensor board 15 seconds after the
motor of the MS-731V drives to upward.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Drive IC of the motor failure → Replace UT-7243 DRIVER BD.
• Motor failure → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check continuity of the cable. Replace
the cable.
• Sensor failure → Replace the sensor.
• Sensor signal receive IC failure → Replace the UT-7243 DRIVER BD.
Detection sensor
Upper and lower position sensor
Detection method
• The pass of the sensor board from the lower position sensor is not detected
15 seconds after the motor of the MS-731V drives to upward. Or the upper
position sensor does not detect the sensor board.
• The pass of the sensor board from the upper position sensor is not detected
15 seconds after the motor of the MS-731V drives to downward. Or the lower
position sensor does not detect the sensor board.
Cause
i) When the motor does not rotate
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Drive IC of the motor failure → Replace UT-7243 DRIVER BD.
• Motor failure → Replace the unit.
ii) When the motor rotates but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Sensor failure → Replace the sensor.
• Sensor signal receive IC failure → Replace the UT-7243 DRIVER BD.
NOTE
Pressing the eject button when holding the tube holder or the holder is
caught because the front panel is not closed correctly causes an “E145:
TUBE HOLDER OPERATION ERROR” error and the analyzer must be
restarted. Before pressing the eject button, check the holder condition.
Detection sensor
Holder sensor
Detection method
The holder sensor does not detect the pass of the sensor board one second after
driving the solenoid of the MS-731V and opening the holder.
Cause
i) When the holder does not open
• The drive part is caught. → Remove the cause of catch.
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Solenoid drive IC failure → Replace UT-7243 DRIVER BD.
• Solenoid failure → Replace the unit.
ii) When the holder opens but the sensor cannot detect
• Poor contact or cable break → Disconnect and connect the connector.
Check cable continuity.
• Sensor failure → Replace the sensor.
• Sensor signal receive IC failure → Replace the UT-7243 DRIVER BD.
Alarm
When a problem which affects the measurement result occurs during
measurement, the analyzer displays alarms on the screen. When the Recovery
button which is displayed with alarms is pressed, the analyzer performs the
optimum operation (such as cleaning or refilling the reagents). Also check the
error code and sensor and unit condition. Open the front panel and check the
condition of nozzle, sampling nozzle, measurement bath and sub bath before
troubleshooting.
There are two types of alarms. One is displayed to the right of the measurement
value and another is displayed as messages. The former alarm indicates that the
result must not be used to the diagnosis because the analyzer has problems or
some phenomenon affects the result even though the sample can be measured.
The later alarm can be classified as no reagent, manometer, CBC measurement,
HGB and flow cell alarms. These alarms indicate that the measurement is not
finished.
Following is the list of error code and the troubled place which is assumed from
the each error code.
! (to the right of the HGB measured value) HGB voltage rise (voltage adjustment error)
! (to the right of the MCHC measured value) RBC index error (error in preparing the pre-diluted
samples)
! (to the right of the WBC measured value) WBC counting error (poor hemolyzation)
C (to the right of the WBC or PLT measured value) Platelet coagulated
? (to the right of the WBC, RBC or PLT measured value) Abnormal sample (the sample is out of measuring range)
? (to the right of the HGB measured value) HGB voltage reduction, WBC measurement bath is dirty
? (to the right of the NE, NE%, LY or LY% measured Room temperature rise
value)
? (to the right of the NE, NE%, LY, LY%, MO, MO%, EO, Optical count error
EO%, BA or BA% measured value)
? (to the right of the NE, NE%, LY, LY%, MO or MO% Room temperature fall
measured value)
* (to the right of the HGB measured value) HGB circuit error or interference substances incorporation
No reagent alarms
A001:No isotonac•3!
A005:No cleanac!
A006:No cleanac•3!
A007:No hemolynac•3N!
A008:No hemolynac•5!
Manometer alarms
HGB alarms
Cause
• Poor adjustment of the HGB sensor voltage → Adjust the HGB sensor (refer
to the unit adjustment)
• Poor cable connection → Check the cable connection and
conduction.
! (to the right of the MCHC measured value) RBC index error
General
The MCHC is calculated by HGB/HCT and indicates the concentration average
of the HGB in the one RBC. Normally MCHC is 33.3. If the MCHC is totally
different from this value, it means that the HGB and HCT ratio of the analyzer
is not correct. If the MCHC is ≤28.0 or ≥38.0, “!” appears to the right of the
MCHC value and an RBC index error appears.
When the difference between the threshold high count and threshold low count
is large, the sample is measured again. If the difference is still large after three
measurements, the poor hemolyzation alarm occurs because many unhemolyzed 2
RBC hemolytic ghosts are detected and they may affect the WBC count.
C (to the right of the WBC or PLT measured value) Platelet coagulated
General
The platelets stick together, the apparent number of PLT decreases and
aggregated platelets may be counted as WBC.
? (to the right of the WBC, RBC or PLT measured value) abnormal sample
General
This alarm is displayed when the electrode voltage is more than 32.0 V.
Normally the WBC electrode voltage is about 18 V and the RBC electrode
voltage is about 24 V. When the temperature is low, the electrical resistance of
the diluent increases and the electrode voltage rises. If tap water is used instead
of diluent, the tap water does not pass electrical current because tap water does
not contain salt so the electrode voltage is very high.
? (to the right of the HGB measured value) HGB voltage reduction
General
This alarm is displayed when the HGB sensor voltage is less than 1.6 V. The
HGB sensor voltage is adjusted to 3.0 V when the WBC measurement bath is
filled with diluent. When the WBC measurement bath is dirty with blood protein,
the sensor voltage decreases. The reduced sensor voltage affects the HGB
reproducibility.
? (to the right of the NE, NE%, LY or LY% measured value) room
temperature rise
General 2
This alarm is displayed when the temperature sensor detects that the room
temperature is higher than the specified operating temperature. If the room
temperature is higher than the operating temperature, the lysing reagent response
changes. This affects WBC 5 part differential measurement especially NE% and
LY% and ? is displayed to the right of the NE%, NE, LY% and LY measurement
value.
? (to the right of the WBC 5 part differential value) optical count error
General
This alarm is displayed when there is a big difference between the measurement
result by the WBC electrical resistance method and optical measurement result
by the flow cell.
? (to the right of the NE, NE%, LY, LY%, MO or MO% measured value) room
temperature fall
General
This alarm is displayed when the temperature sensor detects that the room
temperature is lower than the specified operating temperature. If the room
temperature is lower than the operating environment, the lysing reagent response
changes. This affects WBC 5 part differential measurement especially NE%,
LY% and MO% and ? is displayed to the right of the NE%, NE, LY%, LY, MO%
and MO measurement value.
* (to the right of the HGB measured value) HGB circuit error
General
This alarm is displayed when the HGB sensor voltage is 0.5 V or more even
when the LED for the HGB sensor is turned off. When the HGB sensor LED is
turned off, only the dark current flows into the sensor and the current of 0.5 V or
more is not detected.
No Reagent Alarms
A001:No isotonac•3!
A005:No cleanac!
A006:No cleanac•3!
A007:No hemolynac•3N!
A008:No hemolynac•5!
Sensor 2
2
Sensor 1
Sensor 3
Sensor 4
Five types of reagents are connected to the inlet and outlet units. Four optical
sensors judge the reagent connection.
Sensor 1: detergent (CLEANAC, CLEANAC•3) and diluent (ISOTONAC•3)
Sensor 2: diluent (ISOTONAC•3)
Sensor 3: lysing reagent (HEMOLYNAC•3N)
Sensor 4: lysing reagent (HEMOLYNAC•5)
When the sensor voltage is more than 2.5 V, the analyzer judges that the sensor is
in air. When the voltage is less than 2.5 V, the analyzer judges that the sensor is
in liquid. To make a 1.0 V margin, the sensor voltage must be adjusted to judge
that in liquid is 1.5 V or less and not in liquid is 3.5 V or more.
PTr
LED
• The liquid sensor detects the liquid level by using the light refraction.
• When there is liquid, the refraction index of the plastic of quantitative part and
the liquid is almost the same. The LED light goes straight and reaches the light
receiving element (PTr).
• When there is no liquid, the LED light is reflected and does not reach the
light receiving element (PTr) because the refraction index of air and plastic is
different.
PTr
LED
• When there are small bubbles in the light path of the liquid sensor, a no reagent
alarm may occur frequently because the LED light is reflected and does not
reach the light receiving element (PTr).
• The small bubbles stay in the sensor even when liquid flows and cannot be
removed by repeated cleaning.
PTr
LED
• If bubbles get in the sensor, remove the tube and pass air. This removes the
bubbles easily.
• Remove the diluent tube and press the Clean key while holding the Reset key.
The analyzer aspirates air instead of diluent and the air passes through the
liquid sensor. After the operation, reattach the diluent tube and press the Clean
key while holding the Reset key to fill the diluent.
A001: No isotonac•3!
General
This alarm is displayed when the liquid sensor for ISOTONAC•3 of the JQ-730V
detects air when aspirating diluent. The UT-7235 LIQUID SENSOR BD of the
JQ-730V has a sensor for diluent only and a sensor for diluent and detergent. The
A: 001 NO DILUENT alarm occurs when one of the sensors detects air.
A005: No cleanac!
General
This alarm is displayed when the sensor for diluent and detergent of the JQ-730V
detects air while aspirating CLEANAC. The UT-7239 LIQUID SENSOR BD
detects the liquid and sends the information to the UT-7241 MAIN BD.
When the A: 005 NO CLEANAC alarm occurs and A: 001 NO DILUENT and
A: 006 NO CLEANAC•3 alarms do not occur, there is a problem in the flow path
between the CLEANAC tank and 2-way valve for change.
A006: No cleanac•3!
General
The alarm is displayed when the sensor for diluent and detergent of the JQ-730V
detects air while aspirating CLEANAC•3. The UT-7239 LIQUID SENSOR BD
detects the liquid and sends the information to the UT-7241 MAIN BD.
When the A: 006 NO CLEANAC•3 alarm occurs and A: 001 NO DILUENT and
A: 005 NO CLEANAC alarms do not occur, there is a problem in the flow path
between the CLEANAC•3 tank and 2-way valve for change.
A007: No hemolynac•3N!
General
This alarm is displayed when the sensor for HEMOLYNAC•3N of the JQ-730V
detects air while aspirating HEMOLYNAC•3N. The UT-7239 LIQUID SENSOR
BD detects the liquid and sends the information to the UT-7241 MAIN BD.
A008: No hemolynac•5!
General
This alarm is displayed when the sensor for HEMOLYNAC•5 of the JQ-730V
detects the air while aspirating HEMOLYNAC•5. The UT-7239 LIQUID
SENSOR BD detects the liquid and sends the information to the UT-7241 MAIN
BD.
Manometer Alarms
A009:WBC priming error
A010:RBC priming error 2
A071:WBC upper manometer dirty
A072:WBC middle manometer dirty
A073:WBC lower manometer dirty
A074:RBC upper manometer dirty
A075:RBC middle manometer dirty
A076:RBC lower manometer dirty
Up sensor
LED
Low360 sensor
Low500 sensor
• The part which aspirates the diluted blood sample from the aperture and
measures the volume is called “manometer”. The manometer is a transparent
plastic tube which has upper, middle and lower optical sensors that detect the
liquid level in the manometer.
• When the sensor voltage is more than 2.5 V, the analyzer judges that the sensor
is in air. When the voltage is less than 2.5 V, the analyzer judges that the sensor
is in liquid. To get 1.0 V of margin, the sensor voltage must be adjusted to
judge that in liquid is 1.5 V or less and out of liquid is 3.5 V or more.
• Thus even if there is liquid and the sensor voltage is 1.51 V, the analyzer does
not malfunction. But if the manometer sensor voltage is 2.25. to 2.75 V, the
manometer may misjudge the situation so that the “MANO ERROR” alarm
occurs.
General
The “MANO ERROR” alarm is displayed when loading each sensor voltage of
the manometer during measurement and the voltage of the manometer is less
than 2.25 to 2.75 V.
First, MEK-7300 is different from the previous models. It has two lower position
sensors to measure 360 μL or 500 μL quantity.
2
The measurement and alarm process is the same as the previous models.
Normally, the determined quantity is 360 L. If the measurement value is low, the
analyzer increases the quantity and measures the sample again (advanced count
function).
standby
360 uL
Up sensor
Low 360
sensor
Low 500
sensor
Time
500 uL
Up sensor
Low 360
sensor
Low 500
sensor
Time
Time
Time
Detection method
When the quantity is 500 or 360 μL, the upper sensor of the manometer does not
detect fluid level within 8 seconds.
Time
Time
3 6 0 uL
Level 1 Level 2 Level 3 Clog
Detection method
When the manometer decreases the fluid level after measurement starts, level 1 is
passed but the fluid level is not detected within the following times:
When the quantity is 360 μL, middle position sensor of the manometer: 8
seconds
When the quantity is 500 μL, lower position sensor of the manometer: 11.2
seconds
Time
Time
360 uL
Level 1 Level 2 Level 3 Clog
Detection method
After starting a measurement, level 1 and 2 are passed and the manometer
decreases and increases the fluid, but the fluid level is not detected within the
following times:
When the quantity is 360 μL, middle position sensor of the manometer: 11.0
seconds
When the quantity is 500 μL, lower position sensor of the manometer: 11.0
seconds
Time
Time
Detection method
This alarm appears when the upper or lower position sensor of the manometer
detects bubbles when the manometer decreases the fluid after starting
measurement.
Time
Time
Detection method
This alarm appears when the upper sensor detects the fluid level before the
lower sensor of the manometer detects the fluid level (illustration 2) after a
measurement starts, the manometer decreases fluid and the upper sensor of the
manometer detects the fluid level.
Time
500 uL
Level 1 Level 2 Level 3 Clog
Time 360 uL
Level 1 Level 2 Level 3 Clog
Detection method
This alarm appears when the upper sensor detects the fluid level before the
lower sensor of the manometer detects the fluid level (illustration 5) after a
measurement starts, the manometer decreases fluid and upper sensor of the
manometer detects the fluid level.
Time
5 0 0 uL
Level 1 Level 2 Level 3 Clog
Time
3 6 0 uL
Level 1 Level 2 Level 3 Clog
Detection method
This alarm appears when the lower sensor detects bubbles before the upper
sensor of the manometer detects the fluid level (5 of the illustration) after a
measurement starts, the manometer increases the fluid and the lower sensor of
the manometer detects the fluid level.
Time
500 uL
Level 1 Level 2 Level 3 Clog
Time
360 uL
Level 1 Level 2 Level 3 Clog
Detection method
This alarm appears when the time from the pulse count starts to upper sensor of
the manometer detects the fluid level and the measurement finishes is shorter
than the following.
When the quantity is 500 μL: 14.5 seconds
When the quantity is 360 μL: 11.50 seconds
Time
500 uL
Level 1 Level 2 Level 3 Clog
Time 360 uL
Level 1 Level 2 Level 3 Clog
Detection method
The alarm appears when the time from starting a measurement, the manometer
increases the fluid and the lower sensor is turned on to the upper sensor is turned
on (5 of the illustration) is longer than the following.
When the quantity is 500 μL, normal mode: 21.25 seconds
high altitude mode: 28.38 seconds
When the quantity is 360 μL: normal mode: 15.30 seconds
high altitude mode: 20.43 seconds
Filter
Filter
Case
Filter
Filter
Case
Cause
The analyzer was turned on while holding down the reset key.
Countermeasure
• Turn on the analyzer normally (5 minutes).
• Refill the diluent (8 minutes) or refill all reagents (9 minutes).
• Perform cleaning (16 minutes).
• Measure samples until cleaning is performed after 10 measurements.
• Press the HGB BLANK MEAS key on “MC UNITE CHECK” of the
maintenance window.
Particle CV Check
Measuring Particles
Press the System → Adj. Flowcell key to display the Adj. Flowcell (Rough)
window.
Press the Count switch to aspirate and measure the standard particles. After
measurement, CV and peak of FS (forward-scattered light small angle
component) and FL (forward-scattered light large angle component) histogram
appear. Check that these values are in the following range.
If the CV value of FS and FL is not in the above range, press the Clean key
on the lower middle of the screen to clean the flowcell and remove dust and
bubbles in the flowcell then check if the CV is improved. If the CV of particle
measurement is more than 7.0% after cleaning, adjust the flowcell position. Refer
to “Adjusting the Flowcell Position” later in this section.
* The laser is only irradiated during measurement. While the laser is irradiated,
the laser check LED of the front panel lights. You can always irradiate the
laser by displaying the particle measurement screen. When checking the laser
irradiation by opening the cover of the analyzer, make sure that the laser is not
in your eyes. When the cover of the MO-820V2 is opened, the interlock switch
on the UT-7176 and laser are turned off.
ii) When the scattergram and histogram appear but the TOC is low (200 or less).
If the CV of particles is incorrect after cleaning the flowcell, adjust the flowcell
to the proper position by the following procedure.
1. Remove the two flowcell cover fixing screws on the top cover and the
flowcell cover.
2. Loosen but do not remove the two screws that fix the flowcell position adjust
screw by turning the Phillips head screw driver counterclockwise.
NOTE
Do not remove the screws.
Flowcell cover
3. Insert the T box wrench (standard accessory) into the flowcell position adjust
screw.
5. When the particle count starts, turn the T box wrench about 2 to 3 degrees to
the right or left.
T box wrench
2
About 2 to 3 degrees
6. Press the Clear key and check the distribution of the particle scattergram
and histogram. When the Clear key is pressed, the Count is reset and starts
from 0. The scattergram is also cleared and you can check the adjustment
condition. If the Clear key is not pressed, the scattergram is not cleared and
you cannot check the condition.
7. Compare the scattergram distribution of the particles before and after turning
the T box wrench. If the distribution is smaller (the width of the FS and FL
histogram is thinner) after turning the wrench, turn the wrench 2 to 3 degrees
in the same direction. If the distribution is wider, turn the wrench to the
opposite direction.
8. Press the Clear key and check the distribution as described in step 6.
10. Remove the T box wrench from the position adjust screw.
11. Tighten the two screws that fix the flowcell position adjust screw with the
Phillips head screw driver (turn clockwise).
12. Measure standard particles and check the CV and peak value. If the CV of
FS and FL is 7.0% or less, hook the MO blind plate to the analyzer and fix it
in place with the screws.
13. When the particles are measured and the OK key is pressed on the Adj.
Flowcell screen, a confirmation message appears. Press the OK key to
perform post-cleaning of particle measurement (about 6 minutes).
Backwash procedure
1. Fill a syringe with about 10 cc of Cleanac (green detergent).
2. Disconnect the joint from the end of the flowcell and connect the syringe
filled with detergent.
3. Disconnect the valve joint (17) which is connected to the entrance flow path
with the sheath under the flowcell and put the tube into an empty cup.
2
4. Press the syringe to flow detergent from the end of the flowcell several times.
5. Re-connect the joint and press the Clean key to clean the analyzer.
Pressure sensor
Manifold
Background Check
Measure only the diluent in both closed and open mode measurement to check
the noise effect.
Set an empty vacuum sample tube to the tube holder and start measurement in
closed mode.
Press the Count switch to measure only the diluent in open mode. The detergent
and hematology control do not need to be aspirated.
If the measurement values are near the following values, there is no problem.
WBC: 2 (×102/μL)
RBC: 5 (×104/μL)
HGB: 0.1 (g/dL)
PLT: 1.0 (×104/μL)
TOC: 100 count
Perform cleaning and measure the background for any items. If the value is still
out of range, perform strong cleaning, measure the background and check the
value again.
Ready → Operation → Strong cleaning
If the background does not decrease, check the following possible causes.
Reproducibility Check
2
Reading Data
The reproducibility of the data is judged by the “CV value” that is obtained by
dividing the standard deviation with the average. The larger the CV value, the
larger the variation is (the reproducibility is bad).
Example
Item Average CV CV
WBC 68 1.3 2.0
NE% 5.0
LY% 5.0
MO% 12.0
EO% 20.0
BA% 30.0
RBC 443 0.8 1.5
HGB 13.0 0.7 1.5
MCV 91.0 0.4 1.0
PLT 24.2 3.2 4.0
For example, the average is 443 and the CV value is 0.8 for RBC of above list.
0.8% of 443 is 3.5, 443 ± 3.5 is 439.5 to 446.5. When the CV value is 0.8% mean
RBC measured data is within 439.5 to 446.5 six times out of ten.
• Check that the reproducibility CV% for each item is within the standard
values.
Reproducibility Failure
If the reproducibility is poor after “Checking the Reproducibility”, refer to the
following section to find the cause and take proper action.
2
Cause Action
WBC measurement bath • Press the Clean key to clean the analyzer.
is dirty. • Perform strong cleaning from Operations.
• Remove and clean the WBC measurement bath.
Refer to “Checking and Cleaning Measurement
Baths, Sub Baths and MC Tray” in Section 9 of
the operator’s manual.
• Replace the WBC measurement bath.
WBC sub bath is dirty. • Press the Clean key to clean the analyzer.
• Perform strong cleaning from Operations.
• Remove and clean the WBC sub bath. Refer to
“Checking and Cleaning Measurement Baths,
Sub Baths and MC Tray” in Section 9 of the
operator’s manual.
• Replace the WBC sub bath.
WBC aperture is dirty. • Press the Clean key to clean the analyzer.
• Perform strong cleaning from Operations.
• Remove and clean the WBC aperture. Refer
to “Cleaning Apertures” in Section 9 of the
operator’s manual.
• Replace the WBC aperture.
Drain path of the WBC • Replace the filter under the air trap.
measurement bath is dirty. • Clean the air trap.
Rinse chassis is dirty. Clean the rinse chassis. Refer to “Checking,
Cleaning and Replacing the Rinse Unit, Sampling
Nozzles, Cap Pierce Nozzle and Sample Cup” in
Section 9 of the operator’s manual.
Cause Action
Temperature exceeds Use the analyzer in the specified operating
the specified operating environment. When the temperature is low, use a
environment (15 to 30°C). heater.
Hemolynac•5 is connected Perform Drain all, clean the analyzer with distilled
as the lysing reagent by water, drain the analyzer, connect the reagent
mistake. correctly and perform strong cleaning.
Hemolynac•3 is connected Perform Drain all, clean the analyzer with distilled
as the lysing reagent by water, drain the analyzer, connect the reagent
mistake. correctly and perform strong cleaning.
Cause Action
Sample is not agitated Teach the operator to agitate the sample enough.
enough when measured.
Sample is not agitated Teach the operator to agitate the sample enough.
enough when collected.
Cause Action
Specified quantity of • Hemolynac•3 tube assy does not touch the
lysing reagent is not bottom of the bottle.
delivered. • Replace the Hemolynac•3 tube assy.
• Replace the MP-731V triple pump unit.
Circuit failure. When • Replace the MC-730V.
checking the circuit by • Replace the UT-7244 ANALOG BD.
pressing “Operation” →
“Circuit Check” from the
Ready screen, the WBC
exceeds 80 ±5%.
WBC measurement • Replace the MP-731V triple pump unit.
sample failure. • Replace the MS-730V sampler unit.
• Replace the MS-731V cap pierce unit.
i) High background
Cause Action
Diluent is dirty. Replace the diluent.
Unspecified reagents are Use the specified reagents.
used.
Flow path is dirty. • Press the Clean key to clean the analyzer.
• Perform strong cleaning.
Noise from the power Change the AC outlet.
source.
Noise from external • Ground the analyzer.
instruments. • Keep the analyzer away from the external
instruments.
The temperature exceeds Use the reagent within the specified operating
the operating environment environment.
of the reagents.
MO-820V2 failure Replace the MO-820V2.
UT-7244 ANALOG BD Replace the UT-7244 ANALOG BD.
failure.
Hemolynac3N is • Clean the connection tube.
connected to the • Perform Drain all, clean the analyzer with
Hemolynac5 port. distilled water, drain the analyzer again and
perform strong cleaning.
Cause Action
No scattergram Refer to “Particle Reproducibility Failure”.
2
There are each white
blood cell distributions but
they do not come together.
Each white blood cell
distributions come
together but do not in the
areas that is divided by the
classification line.
Cause Action
Pinch valve is clogged. Check and replace the pinch valve tube.
Drain tube is bent. Check and replace the drain tube
MJ-730V failure • Check if there is a leak in the mix chamber.
• Check if there is a leak from joints.
• Check the valve operation on the “MV CHECK”
of the maintenance window.
• Replace the MJ-730V.
MP-730V failure. Replace the MP-730V.
MP-731V failure. Replace the MP-731V.
There is a leak in flow Repair the flow path.
path.
Cause Action
MEK-3D is measured. Measure MEK-5DN.
MEK-5D hematology Use a new MEK-5DN.
control is deteriorated.
A hematology control is Measure the control in the quality control or
measured in the human control mode.
blood mode.
A sample which passed Use a normal sample within 12 hours of collection.
12 hours or more after
collection is measured.
WBC value is low or Use human blood which is within the following
classification value of range.
the sample is outside the WBC 40 to 400 × 102/μL
standard. NE 40 to 70 (%)
LY 20 to 45 (%)
MO 2 to 10 (%)
EO 2 to 10 (%)
BA 0 to 3 (%)
Cause Action
The temperature exceeds Use the analyzer in the operating environment.
the operating environment
(15 to 30°C).
Pinch valve is clogged. Check and replace the pinch valve tube.
Lysing reagent is not put • The lysing reagent tube does not touch the
correctly. bottom of the Hemolynac•5 and lysing reagent is
not aspirated enough.
• Check the MP-730V operation or replace it.
• MP-730V piping is clogged.
Hemolynac•3N is Perform Drain all, clean the analyzer with distilled
connected as lysing water, drain the analyzer again, connect the
reagent by mistake. reagents correctly and perform strong cleaning.
Lysing reagent is diluted Perform Drain all, clean the analyzer with distilled
with water. water, drain the analyzer again, connect the
reagents correctly and perform strong cleaning.
Cause Action
WBC sub bath is dirty. Clean the sub bath by soaking it in Cleanac•3.
WBC measurement bath Clean the measurement bath by soaking it in
is dirty. Cleanac•3.
Obstacle in the HGB • Keep the electrode of the measurement bath
measurement light path. away from the HGB measurement light path.
• Check for crystals in the HGB measurement light
path of the HGB block.
Cause Action
Temperature exceeds the Use the analyzer in the specified operating
operating environment (15 environment. When the temperature is low, use a
to 30°C). heater.
Hemolynac•5 is connected Perform Drain all, clean the analyzer with distilled
as the lysing reagent by water, drain the analyzer, connect the reagent
mistake. correctly and perform strong cleaning.
Hemolynac•3 is connected Perform Drain all, clean the analyzer with distilled
as the lysing reagent by water, drain the analyzer, connect the reagent
mistake. correctly and perform strong cleaning.
Cause Action
Sample is not agitated • If the sample is not agitated enough and RBC
2
enough when measured. thick part is aspirated, RBC, HCT and HGB
become a falsely high value. When the thin part
is aspirated, RBC, HCT and HGB become a
falsely low values.
• Teach the operator to agitate the sample enough.
Refer to “Measurement” in Section 5 of the
operator’s manual.
Cause Action
Standard quantity of • Check that the Hemolynac•3 tube touches the
lysing reagent is not bottom of the tank.
delivered. • Replace the Hemolynac•3 tube assy.
• Replace the MP-731V triple pump unit.
HGB measurement bath • Replace the MP-731V triple pump unit.
is dirty. • Replace the MS-730V sampler unit.
HGB sensor output • Adjust the hemoglobin unit adjustment VR so
failure. HGB LED ON that the HGB LED voltage is 3.0 ±0.5 V.
voltage is outside the 1.5 • If the voltage cannot be adjusted, replace the
to 4.5 V with the Sensor MC-730V or UT-7244 ANALOG BD.
Monitor screen.
Cause Action
RBC measurement bath • Press the Clean key to clean the analyzer.
is dirty • Perform strong cleaning from the Operation
window.
• Remove and clean the RBC measurement bath.
Refer to “Checking and Cleaning Measurement
Baths, Sub Baths and MC Tray” in Section 9 of
the operator’s manual.
• Replace the RBC measurement bath.
RBC sub bath is dirty. • Press the Clean key to clean the analyzer.
• Perform strong cleaning from the Operation
window.
• Remove and clean the RBC sub bath. Refer to
“Checking and Cleaning Measurement Baths,
Sub Baths and MC Tray” in Section 9 of the
operator’s manual.
• Replace the RBC sub bath.
Cause Action
RBC aperture is dirty. • Press the Clean key to clean the analyzer.
• Perform strong cleaning from the Operation
window.
• Remove and clean the RBC aperture. Refer
to “Cleaning Aperture” in Section 9 of the
operator’s manual.
• Replace the RBC aperture.
RBC measurement bath Replace the filter under the RBC measurement
draining path is dirty. bath.
Rinse chassis is dirty. Clean the rinse chassis. Refer to “Checking,
Cleaning and Replacing the Rinse Unit, Sampling
Nozzles, Cap Pierce Nozzle and Sample Cup” in
Section 9 of the operator’s manual.
Cause Action
Sample is not agitated • If the sample is not agitated enough and RBC
enough when measured. thick part is aspirated, RBC, HCT and HGB
become a falsely high value. When the thin part
is aspirated, RBC, HCT and HGB become a
falsely low values.
• Teach the operator to agitate the sample enough.
Refer to “Measurement” in Section 5 of the
operator’s manual.
Sample is not agitated Teach the operator to agitate the sample enough.
enough when collected. Refer to “Measurement” in Section 5 of the
operator’s manual.
Cause Action
When checking the circuit • Replace the MC-730V.
by pressing “Operation” • Replace the UT-7244 ANALOG BD.
→ “Circuit Check”
from the Ready screen, the
RBC exceeds 160 ±5%.
RBC measurement sample • Replace the MP-731V triple pump unit.
failure. • Replace the MS-730V sampler unit.
i) High background
Refer to “Background Check” in this section.
2
ii) Sample is not agitated enough
Cause Action
Sample is not agitated Teach the operator to agitate the sample enough.
enough when measured. Refer to “Measurement” in Section 5 of the
operator’s manual.
Cause Action
Sensitivity and threshold Check that RBC sensitivity is set to 5, RBC
setting is not proper. threshold is set to AUTO and PLT threshold is set
to 5.
Cause Action
MS-730V sampler unit Replace the MS-730V sampler unit.
failure.
MP-730V triple pump unit Replace the MP-730V triple pump unit.
failure.
Cause Action
Aperture is dirty. Clean the aperture. Refer to “Cleaning Aperture” in
Section 9 of the operator’s manual.
Aperture is broken. Replace the aperture.
When checking the circuit • Replace the UT-7147 SLVE CBC BD.
by pressing “Operation” • Replace the MC-730V CBC measuring unit.
→ “Circuit Check” • Replace the UT-7242 POWER BD.
from the Ready screen, the
PLT exceeds 16.0 ±5%.
MC-730V CBC measuring Replace the MC-730V CBC measuring unit.
unit failure.
Cause Action
RBC measurement bath • Press the Clean key to clean the analyzer.
is dirty. • Perform strong cleaning from the Operation
window.
• Remove and clean the RBC measurement bath.
Refer to “Checking and Cleaning Measurement
Baths, Sub Baths and MC Tray” in Section 9 of
the operator’s manual.
• Replace the RBC measurement bath.
RBC sub bath is dirty. • Press the Clean key to clean the analyzer.
• Perform strong cleaning from the Operation
window.
• Remove and clean the RBC sub bath. Refer to
“Checking and Cleaning Measurement Baths,
Sub Baths and MC Tray” in Section 9 of the
operator’s manual.
• Replace the RBC sub bath.
RBC aperture is dirty. • Press the Clean key to clean the analyzer.
• Perform strong cleaning from the Operation
window.
• Remove and clean the RBC aperture. Refer
to “Cleaning Aperture” in Section 9 of the
operator’s manual.
• Replace the RBC aperture.
RBC measurement bath Replace the filter under the RBC measurement
draining path is dirty. bath.
Rinse chassis is dirty. Clean the rinse chassis. Refer to “Checking,
Cleaning and Replacing the Rinse Unit, Sampling
Nozzles, Cap Pierce Nozzle and Sample Cup” in
Section 9 of the operator’s manual.
Cause Action
Sensitivity and threshold Check that RBC sensitivity is set to 5, RBC
setting is not proper. threshold is set to AUTO and PLT threshold is set
to 5.
Cause Action
Aperture is broken. Replace the aperture.
When checking the circuit • Replace the MC-730V.
by pressing “Operation” • Replace the UT-7244 ANALOG BD.
→ “Circuit Check”
from the Ready screen, the
MCV exceeds 118 ±5%.
RBC measurement sample • Replace the MP-731V triple pump unit.
failure. • Replace the MS-730V sampler unit.
NOTE
Always wear rubber gloves to protect yourself from infection when
handling and measuring blood samples.
Measure a sample of normal human blood within 8 hours after collection. Check
that the following clinically important flags do not appear.
If the above flags which indicates increase or decrease of the blood cells
frequently appear despite the reproducibility has no problem, these flags are
detected when the CBC measurement value is high or low and the followings
may be the cause.
If the above flags which indicates increase or decrease of the blood cells
frequently appear despite the reproducibility has no problem, the followings may
be the cause.
• PLT-RBC interference
Cause Action
Background is high. Refer to “Checking Background” in this section.
Sensitivity and threshold Check that RBC sensitivity is set to 5, RBC
setting is not proper. threshold is set to AUTO and PLT threshold is set
to 5.
If the above WBC 5 part differentiation flags appear frequently, the causes are
largely classified as follows.
i) Rough optical adjustment trouble 2
ii) Fine optical adjustment trouble
iii) When increasing or decreasing flags of the WBC 5 part differentiation
appear frequently.
iv) When abnormal cell flags of the WBC 5 part differentiation appear
frequently.
• When there are each WBC distributions of the scattergram but they do not
come together.
The followings may be the reason, other than “Fine Optical Adjustment
Trouble”.
Function
• The CBC measuring unit is a main unit that measures the HGB and CBC
components such as WBC, RBC and PLT.
• The unit catches the resistance change when blood cells pass the aperture cap
by voltage conversion and measures the aspirated sample from the aperture
cap.
• The unit measures the change in absorbance of the diluted or hemolyzed
blood sample and takes out the HGB sensor voltage from the hemoglobin
concentration.
• The unit switches the internal fluid path by the control signal from the
UT-7241 MAIN BD.
• The unit turns the measurement bath and transmits the measurement bath
position sensor signal to the MAIN BD by the control signal from the UT-7242
DRIVER BD.
Composition
1. WBC measurement bath
2. RBC measurement bath
3. Sub bath
4. Aperture cap
5. Electrode room
6. Air trap
7. Manometer
8. Tray
9. HGB block
10. PRE AMP BD (WBC)
11. PRE AMP BD (RBC)
12. CONNECTION BD
13. 2-way valve
Function
• The laser optical unit is a main unit that measures the WBC 5 part differential.
• The unit detects the laser scattered light by flow cytometry and converts the
light to electrical signals.
• The unit consists of optical components such as the laser diode and flow cell.
Composition
1. Flow cell
2. Semiconductor laser
3. LASER DRIVER BD (UT-7176)
4. Inter lock switch
5. FS PREAMP BD (UT-7173)
6. FL PREAMP BD (UT-7174)
7. SD PREAMP BD (UT-7175)
Function
• The sampler unit moves the sampling nozzle to the aspirate and dispense
position.
• The unit dispenses the diluted sample to the measurement baths with the
diluter unit.
• The unit rotates the motor and moves the sampling tube to the predefined
position by the control signal from the UT-7243 DRIVER BD. Also the unit
transmits the position information of the sample tube to the UT-7241 MAIN
BD with the sensor on the UT-7235 sampler sensor BD.
Composition
1. Sampling nozzle (WBC and RBC each 1 nozzle)
2. Motor (X axis)
3. Motor (Y axis)
4. Sensor BD (UT-7235)
5. Junction BD (UT-7236)
6. Rinse chassis
7. Sample cup
Function
• The cappierce unit enables measurement without removing the cap of the
vacuum sample tube so that the operator does not touch blood.
• The cappierce unit has the sampling needle which pierces the rubber cap of the
vacuum sample tube and relieves the pressure in the vacuum sample tube.
• The sampling nozzle for blood sampling has a double tube structure and passes
inside the sampling needle. The nozzle also has a rinse function to rinse inside
the sampling needle.
• The unit rotates the motor and moves the sampling needle up and down by the
control signal from the UT-7241 MAIN BD through the UT-7243 DRIVER
BD and UT-7238 VALVE CONNECTION BD.
• The unit detects the presence or absence of the sampling tube and opening
and closing of the holder with the detection sensor to the UT-7241 MAIN BD
through the UT-7243 DRIVER BD and UT-7238 VALVE CONNECTION BD.
Composition
1. Sampling needle
2. Vacuum sample tube holder
3. Rinse chassis
4. Sensor for detecting the upper position of the sampling needle
5. Sensor for detecting the lower position of the sampling needle
6. Sensor for detecting the opening and closing of the holder
7. Sensor for detecting the presence or absence of the vacuum sample tube
8. Solenoid for opening and closing
9. Motor
Function
• The diluter part aspirates and dilutes a sample by the constant concentration.
• The sheath pump part dispenses a constant amount and speed of diluent.
• The unit rotates the motor and moves the piston up and down by the control
signal from the UT-7243 DRIVER BD.
• The unit aspirates and dispenses a sample and reagents by switching the fluid
path inside the 3-way valve in synchronization with the move of the piston.
• The unit transmits the position information of the piston to the UT-7241 MAIN
BD through UT-7243 DRIVER BD by each position sensor.
Composition
1. MIXED PUMP BD (UT-7200)
2. Diluter cylinder
3. Sheath pump cylinder
4. Sheath pump sensor (upper part)
5. Sheath pump sensor (lower part)
6. 3-way valve (XP-513V)
7. Diluter motor
8. Sheath pump motor
Function
• The unit aspirates, measures and dispenses diluent for WBC 5-part differential
measurement, delivers a sample for WBC 5-part differential measurement and
dispenses diluent for rinse.
• The unit aspirates, measures and dispenses Hemolynac•3N and Hemolynac•5
lysing reagent.
• The unit rotates the motor and moves the piston up and down by the control
signal from the UT-7241 MAIN BD through UT-7243 DRIVER BD.
• The unit aspirates and dispenses the lysing reagent by switching the fluid path
inside the 3-way valve in synchronization with the move of the piston.
• The unit transmits the position information of the piston to the UT-7241 MAIN
BD through UT-7243 DRIVER BD by each position sensor.
Composition
1. Cylinder block for sample pump
2. Cylinder clock for Hemolynac•3N
3. Cylinder clock for Hemolynac•5
4. XP-513V 3-way valve for Hemolynac•3N
5. XP-513V 3-way valve for Hemolynac•3N
6. VALVE CONNECTION BD2 (UT-7239)
7. Sensor for detecting upper position
8. Sensor for detecting lower position
9. Motor for sample pump
10. Motor for lysing reagent pump
Function
• The unit produces pressure for aspirating a sample and draining.
• MEK-7300K has two pump units. One is for blood cell count and another is
for WBC 5-part differential measurement.
• The unit rotates the motor. The rotor on the motor axis strokes the pump tube
and drains inside the measurement baths and instrument after measurement.
The unit also supplies reagents inside the instrument and produces constant
pressure when measuring a sample.
• The unit transmits the rotor information to the UT-7241 MAIN BD through
UT-7243 DRIVER BD by the sensor.
Composition
1. Pump tube
2. Pump rotor
3. Motor
4. Sensor (outside the unit)
Function
• The flowcyto piping unit is connected to other units with piping tubes and
transmits and distributes the reagents.
• The unit has the fluid path for WBC 5-part differential measurement, starts the
lysing reagent and sample and sends the lysed sample to the measurement part.
• The unit has the warmer of the diluent for WBC 5-part differential
measurement and controls the temperature.
• The unit controls 21 2-way valves and switches the fluid path between each
unit by the control signal from the MAIN BD through the DRIVER BD,
VALVE CONNECTION BD1 and 2.
• The unit controls the warmer and adjusts the temperature of diluent for WBC
5-part differential measurement and lysing reagent by the control signal from
the MAIN BD through the DRIVER BD, VALVE CONNECTION BD1 and 2.
• The unit changes the preset temperature of the warmer by the temperature
setting switch on the MAIN BD.
Composition
1. Mixing chamber of the WBC 5-part differential measurement sample
2. Manifold for aspirating a lysed sample
3. Chamber for stabilizing the sheath pressure
4. Warmer for diluent for WBC 5-part differential measurement and lysing
reagent
5. Chamber for waste
6. Pinch valve
and piping tubes between the other units and 2-way valves for switching the fluid
path.
Function
• The unit supplies and drains the diluent, detergent, lysing reagent and waste
from the tubes connected to the reagent port.
• The unit detects the reagent with the liquid sensor on the UT-7236 LIQUID
SENSOR BD and transmits the signals to the UT-7241 MAIN BD.
• The unit detects the sheath pressure with the puressure sensor on the LIQUID
SENSOR BD and transmits the signals to the UT-7241 MAIN BD.
• The unit turns on or off the laser oscillation of the laser optical unit by laser
switch and key.
• The unit switches diluent, Cleanac detergent and Cleanac•3 detergent by
moving the 2-way valves by the control signal from the UT-7241 MAIN BD
through the UT-7243 DRIVER BD and UT-7239 VALVE CONNECTION
BD2.
Composition
1. LIQUID SENSOR BD (UT-7235)
2. Liquid sensor block
3. Pressure sensor
4. Laser switch
5. Laser key
6. 2-way valve
Function
• The unit binds several 2-way valves for easy replacement.
• The unit opens and closes the fluid path by moving the 2-way valves by the
control signal from the UT-7241 MAIN BD through the UT-7243 DRIVER
BD and UT-7240 VALVE CONNECTION BD3.
Composition
1. 2-way valves
Function
• The unit has the key switches and touch panel to operate the instrument and
the color LCD displays measurement data (numeric value and histogram data)
and messages.
• The instrument condition can be checked with the main power, power and auto
print mode LED.
• The measurement data can be printed (option).
• The unit displays LCD, prints measurement data, lights LED and generates
beeps by the control signal from the UT-7241 MAIN BD through the UT-7246
LCD BD.
• The unit transmits the input signal from the touch panel and each switch to the
UT-7241 MAIN BD.
Composition
1. LCD
2. Touch panel
3. Inverter
4. LCD BD (UT-7246)
5. KEY BD (UT-7247)
6. PRINTER KEY BD (UT-7248)
7. SD card slot
UT-7241 MAIN BD
3
UT-7241
MAIN BD
Function
• The MAIN BD controls the analyzer.
• The board controls the actuators, display, each measurement signal and signals
from/to the external instrument by the software program which is installed in
the flash ROM.
• The board has the backup battery and saves the settings and data.
UT-7244 ANALOG BD
UT-7244
ANALOG BD
Function
• The board performs analog measurement.
• The board processes the analog data (CBC data or optical data).
• The board converts the analog data from the MC-730V CBC measuring unit
and MO-820V2 LASER OPTICAL UNIT to the digital data and sends the data
to the UT-7241 MAIN BD.
UT-7243 DRIVER BD
3
UT-7243
DRIVER BD
Function
• The board controls the actuator.
• The board controls the 2-way valves under FPGA on this board, and motor by
the control command from the UT-7241 MAIN BD.
• The board transmits the sensor signal of the each unit to the MAIN BD.
UT-7242 POWER BD
UT-7242
POWER BD
Function
• The board generates each voltage.
• The board converts DC +24 V input from the switching power to the necessary
voltage (+3.3 V, +5 V, +12 V, ±15 V, 24 V, –36 V).
• The board generates high-voltage AC power for removing clogs.
• The board controls the start and stop of the power supply by the control
command from the UT-7241 MAIN BD.
3
UT-7236
LIQUID SENSOR BD
Function
• The board detects whether or not a liquid enters into the analyzer, pressure in
the fluid path and laser key on or off information.
• The board detects whether or not the diluent, two types of lysing reagent and
detergent enter into the analyzer with each opposed sensor and transmits the
information to the UT-7241 MAIN BD.
• The board detects, amplifies and transmits the pressure in the fluid path with
the pressure sensor to the UT-7241 MAIN BD.
• The board creates the conduction pathway of the two external laser keys and
transmits the signal to the UT-7241 MAIN BD.
UT-7246 LCD BD
UT-7246
LCD BD
Function
• The board is attached to the front panel and controls the SD card, buzzer,
interface to the LCD and touch panel.
• The board relays the control of the SD card and buzzer and signal to the LCD
and transmits touch panel information to the UT-7241 MAIN BD.
• The board turns the power supply to the LCD on or off by the control from
the MAIN BD, and supply power voltage (+12 V) to the inverter for LCD
backlight.
UT-7247
KEY BD
UT-7248
PRINTER KEY BD
Function
• The board transmits the pressing information of the switch on the board to the
UT-7241 MAIN BD.
• The board has an LED to indicate the condition.
Function
• The board transmits the sensor signal of the UT-7235 SAMPLER SENSOR
BD to the UT-7241 MAIN BD.
UT-7234
SAMPLER
JUNCTION BD
UT-7235
SAMPLER
SENSOR BD
Function
• The board has the sensor to decide the sampler nozzle position.
• The 3 photo interrupters detect up and down movement of the sampler nozzle
and the 4 opposite sensors detect the cutout on the sampler unit. The board
transmits the position information of the right and left movement of the
sampler nozzle to the UT-7241 MAIN BD.
• The LED on each sensor indicates the sensor condition.
UT-7200
MIXED PUMP BD
Function
• The board has the diluter piston sensor of the mixed pump unit and transmits
the signal to the UT-7243 DRIVER BD.
UT-7238
VALVE
CONNECTION BD1
Function
The board relays the UT-7243 DRIVER BD and following modules.
• MP-640V
• MS-731V
• Thermistor for correcting the hemoglobin temperature
• Analyzer internal temperature thermistor
• Heater temperature thermistor
• Heater
• Measurement SW
• Part of the 2-way valves in the MJ-730V
UT-7239
3
VALVE CONNECTION BD2
Function
The board relays the UT-7243 DRIVER BD and following modules.
• 2-way valves in the JQ-730V
• Thermistor for correcting the hemoglobin temperature
• 3-way valves in the MP-730V
• Part of the 2-way and 3-way valves in the MJ-730V
• Pinch valve of the MJ-730V
UT-7240
VALVE CONNECTION BD3
Function
The board relays the UT-7243 DRIVER BD and following units.
• 2-way valves of the JQ-731V
• 2-way valves of the JQ-732V
Function 3
The board is on the MC-730V CBC measuring unit and transmits the HGB
absorbance signal to the UT-7244 ANALOG BD through the UT-7250
CONNECTION BD.
Function
The board is on the MC-730V CBC measuring unit and lights the LED for HGB
absorbance measurement.
UT-7202
HGB LED BD
UT-7201
HGB AMP BD
4
Before You Begin.................................................................................................................................................. 4.2
Warnings and Cautions.............................................................................................................................. 4.2
Required Tools............................................................................................................................................ 4.2
Caution and Notes Related to Valve Joint, Black Screw and Tube Joint in the Instrument........................ 4.3
Draining the Hematology Analyzer....................................................................................................................... 4.4
Opening the Front Cover...................................................................................................................................... 4.6
Removing the Side Cover..................................................................................................................................... 4.7
Removing the Top Cover...................................................................................................................................... 4.8
Removing the PV-730VK Front Panel Unit........................................................................................................... 4.9
Removing the Rear Cover.................................................................................................................................. 4.11
Removing the MC-730V CBC Measuring Unit................................................................................................... 4.12
Removing the MS-730V Sampler Unit................................................................................................................ 4.13
Replacing the MS-731V Cap Pierce Unit........................................................................................................... 4.15
Removing the MO-820V2 Laser Optical Unit..................................................................................................... 4.16
Removing the JQ-731V Valve Unit..................................................................................................................... 4.17
Removing the JQ-732V Valve Unit..................................................................................................................... 4.18
Removing the MP-640V Rotary Pump Unit........................................................................................................ 4.19
Removing the Left CBC Unit.................................................................................................................... 4.20
Removing the JQ-730V Inlet/Outlet Unit............................................................................................................ 4.21
Removing the MP-731V Triple Pump Unit.......................................................................................................... 4.22
Removing the MP-730V Mixed Pump Unit......................................................................................................... 4.24
Removing the MJ-730V Flow Cell Piping Unit.................................................................................................... 4.26
Removing the Power Supply Part....................................................................................................................... 4.28
Removing the UT-7242 POWER BD.................................................................................................................. 4.29
Removing the UT-7244 ANALOG BD................................................................................................................. 4.30
Removing the UT-7241 MAIN BD....................................................................................................................... 4.31
Replacing the UT-7243 DRIVER Board.............................................................................................................. 4.32
Removing the UT-7200 MIXED PUMP BD......................................................................................................... 4.33
Removing the UT-7235 SAMPLER SENSOR BD.............................................................................................. 4.34
Removing the UT-7236 LIQUID SENSOR BD.................................................................................................... 4.35
Removing the UT-7250 CONNECTION BD........................................................................................................ 4.36
Removing the UT-7025 PRINTER DRIVER BD.................................................................................................. 4.37
Removing the UT-7246 LCD BD......................................................................................................................... 4.38
Removing the UT-7247 KEY BD......................................................................................................................... 4.39
Removing the CXA-0454 Inverter....................................................................................................................... 4.40
The procedures in this section tell how to remove, replace and install major
components in the instrument.
WARNING WARNING
To avoid the possibility of injury to yourself or To avoid accidental discharge of static electricity
damage to the instrument, do not install or which could damage the instrument components,
remove any component or change switch settings use a wrist ground strap when installing or
while the power is on and wait 10 minutes after removing any component of the instrument.
the power is off before installing or removing any
component from the instrument.
WARNING WARNING
When replacing any parts or units in the Wear rubber gloves to prevent infection by blood.
instrument, do not touch any part of the
instrument where blood is or may be.
CAUTION CAUTION
Before connecting or disconnecting any cables, Fuses cut off the power when an abnormality
turn off the instrument and unplug the AC power occurs in the instrument. Eliminate the malfunction
cord from the instrument. before replacing the fuse. Use the correct fuse
only. The fuse rating is shown on the holder.
CAUTION CAUTION
Removal and replacement of any component in Use only parts recommended by Nihon Kohden
the instrument should be done by qualified to assure maximum performance from your
service personnel. instrument.
Required Tools
• Anti-static bench mat
• Wrist ground strap
• Phillips screwdriver (insulated type)
• Flat-blade screwdriver (insulated type)
• Tweezers
Caution and Notes Related to Valve Joint, Black Screw and Tube Joint in the Instrument
Valve joint
CAUTION
When connecting the valve joint to the electromagnetic valve, turn
the valve joint clockwise, using moderate force until the valve joint
4
comes to a stop. Do not use extreme force to tighten the valve joint
further because this will damage the tip of the valve joint. If the valve
joint is loosely connected, it will leak.
NOTE
Black screw
Black screws are used to fasten the individual units to the chassis of the
instrument to enable the quick removal and replacement of these units.
However, to fasten the pump unit to the chassis, normal screws are used.
• There are 2 types of spring tube joints, white (inlet side) and black
(outlet side). Each tube joint and its corresponding port in the
instrument are marked with the same color or number to ensure
matching.
To avoid any trouble, drain all fluid from the hematology analyzer. Some of the
units described in this section require the following procedure.
1. Press the Strong clean key on the Operation screen to perform strong
cleaning.
2. Remove the diluent tube from the ISO3 diluent inlet, the hemolysing reagent
tube from the HEMO3N and HEMO5 inlets, the detergent tube from the
CLN inlet and the cleanac tube 8 from the CLN3 inlet on the right side
panel.
NOTE
Waste
outlet Do not remove the waste tube from the waste outlet.
Waste container
3. Press the “Drain all” key on the Operation window. A confirmation message
appears.
4. Press the Yes key to drain the reagents. A “Draining the instrument” message
appears and the remaining time is displayed during draining.
5. After the draining is finished, press the main power switch on the rear panel
to turn the power off. The power lamp and main power lamp go off.
WARNING
For safety, before disassembling the hematology analyzer, wait
approx. 10 minutes after turning off the main power switch.
1. Remove the four BH4 × 16 screws which secure the front cover to the
hematology analyzer chassis.
2. Slightly pull the front cover toward you and open it as shown below.
1. Remove the three BH3 × 8 screws with TLW3 which secure the side cover.
2. Remove the 4 screws which secure the top cover to the hematology analyzer
chassis on the rear panel.
3. Slide the top cover to the rear and remove it from the chassis.
1. Open the front cover according to the procedure described in “Opening the
Front Cover”.
4
2. Release the nine locking wire saddles which secure the cables connected to
the front panel.
3. Open the SHEET (PV-BD) drip-proof cover in the direction of the arrow and
disconnect the four cable connectors A which are connected to the LCD BD.
LCD BD
4. Remove the two PS3 × 6 screws B which secure the SHEET (PV-BD) drip-
proof cover and open the cover which covers the printer.
5. Remove the four PS3 × 8 screws C which secure the PRINTER DRIVER
BD, lift up the PRINTER DRIVER BD toward the direction of the arrow
and disconnect the cable connector D which is connected to the PRINTER
DRIVER BD.
B B
E
F
3. Remove the 20 BH3 × 8 screws with TLW which secure the rear cover to the
hematology analyzer chassis.
4. Pull the rear cover toward the direction of arrow and remove it.
2. Remove the eight tube joint (1 to 4, 13, 14, 19 and 20) and disconnect the
two cable connectors (a and b) as shown in the illustration.
3. Loosen the four screws which secure the CBC measuring unit until the
screws are removed from the main chassis.
5. Disconnect the four cable connectors from the rear of the unit (refer to A).
A
13
14
19 a
B
20
1
2
3
4
NOTE
When attaching the CBC measuring unit, the screw B is a set screw.
Attach this screw first.
5. Remove the two tube joint (17 and 18) on the rinse chassis.
6. Loosen the screw “a” and remove the 5 diff sample cup and PharMed tube.
7. Remove the two tube joints which are connected to the 2-way valve of the
MJ-730V (refer to illustration A).
8. Disconnect cable connectors “b” and “c” which are connected to the motor
and cable connector “d” which is connected to the SAMPLER JUNCTION
BD.
9. Remove the four screws which secure the sampler unit and remove the
sampler unit.
A
White
Black
d
Motor
16
17
B
a 18
PharMed tube
NOTE
When attaching the sampler unit, attach the screw B first.
4. Remove the four PSW3 × 8 screws (black) A which secure the cap pierce
unit and pull the unit toward the direction of arrow.
5. Disconnect the cable connector B which is connected to the cap pierce unit
from the main board.
15
A C
6
4. Disconnect the cable connector A which is connected to the laser optical unit
from the ANALOG BD.
5. Disconnect the tube joint and PharMed tube as shown in the illustration.
6. Remove the four nuts which secure the laser optical unit.
7. Lift up the laser optical unit toward the direction of arrow and remove the
PharMed tube on the bottom.
NOTE
The laser optical unit has the raised parts on the bottom. So put the unit
on the dedicated stand and do not put a load on the parts.
C 4
PharMed tube
(Under MO-820V2 flow cell) PharMed tube
3. Remove the two PS3 × 6 screws (black) A and pull out the valve unit.
4. Disconnect the cable connector B of the valve unit from the JUNCTION BD.
A
1
2
3
4
7
A
4. Remove the two PS3 × 6 screws (black) A and pull out the valve unit.
5. Disconnect the cable connector B of the valve unit from the JUNCTION BD.
A 6
8
2. Disconnect the cable connector A of the rotary pump unit from the
CONNECTION BD.
3. Remove the three screws which secure the rotary pump unit and pull out the
rotary pump unit. Do not damage the rotary sensor on the rear.
C
D
Sensor plate
C
5. Disconnect the four tube joints (3, 4, 51 and 52) from the inlet/outlet unit
and a tube joint (no number) from the rear of the unit. Remove the two tube
joints (4 and 9) from the MJ-730V flow cyto piping unit.
6. Disconnect the PharMed tube A and two cable connectors B which are
connected to the UT-7235 SAMPLER SENSOR BOARD.
7. Remove the two black joints C which are connected to the two 3-way valves
of the MP-730V mixed pump unit.
8. Disconnect the cable D from the Driver BD and release the cable clamp E.
9. Remove the four PS3 × 8 screws F which secure the JQ-730V inlet/outlet
unit and remove the unit.
4
PharMed tube 3 Rear of the unit
A
(pressure sensor)
F
52
a
B
4
MJ-730V
51
Black joint
D
E
3. Remove the four tube joints A ( two white and two black) from the 3-way
valve A of the MP-731V triple pump unit and remove the tube joint B from
the D of the MP-731V triple pump unit.
4. Remove the four PS3 × 8 screws C (black) which secures the triple pump
unit and pull the unit toward the direction of the arrow.
5. Disconnect the three connectors D from the DRIVER BD, release the tube
from the clamp E and remove the MP-731V triple pump unit.
Black
White
A
A
D
2. Remove the four tube joints (33, 45, 46 and A) as shown in the illustration.
3. Remove the cable connector B from the UT-7243 MIXED PUMP BD.
4. Remove the two PS3 × 8 screws (black) which secure the mixed pump unit
and pull the mixed pump unit.
The numbers in the illustration indicate the numbers of the tube joints.
33
45
A
46
C
45
B
46
5. Disconnect the two cable connectors from the DRIVER BD and remove the
mixed pump unit.
When attaching the unit, attach the
MP-730V to the selspacer of the main
chassis and fix it with the screw.
4. Remove the two tube joints from the 2-way valves and eight tube joints (2,
3, 23, 24, 25, 32, 33 and 38) from the MS-730V sampler unit.
5. Remove the four tube joints (two 45 and two 46) from the 3-way valves of
the MP-731V triple pump unit.
6. Remove the two tube joints from the 2-way valves of the MS-730V, the two
tube joints (45 and 46) from the MS-731V cap pierce unit, and the three tube
joints (16, 17 and 18) from the MJ-730V flow cell piping unit.
PharMed tube
(pressure sensor)
25
2 3
24
32
White
Black
33
45
46
16
23
38
17
18
8. Disconnect the four cable connectors of the flow cell piping unit from the
VALVE CONNECTION BD 1 and 2.
9. Remove the seven screws which secure the flow cell piping unit and pull the
unit out.
B 4
B
B
Heater cable
Measurement SW cable
4. Disconnect the six cable connectors which are connected to the POWER BD
and pull the power supply part with the SC chassis toward the direction of
the arrow.
1. Pull out the power supply part by sliding it according to the procedure
described in “Removing the Power Supply Part”.
4
2. Disconnect the VHR-8N (W120) cable A (black and red) from the power
supply part.
3. Remove the five PS3 × 5 screws B and remove the POWER BD from the SC
chassis ASSY.
PS3 × 5 screws B
Cable A
3. Remove the seven cable connectors which are connected to the ANALOG
BD.
4. Remove the five screws A which secure the ANALOG BD. Pinch the fixture
on the upper center of the board and remove the board.
A
Cable from the Connection BD
A
A
4. Remove the 13 cable connectors which are connected to the MAIN BD.
5. Remove the five screws a which secure the MAIN BD, pinch the fixture b
on the upper center of the board and remove the board. To disconnect the
connectors on the back, pull the board toward the front panel and remove it.
Pinch the
open part and
pull the board.
5. Disconnect the seven cable connectors which are connected to the DRIVER
BD.
6. Remove the five screws which secure the DRIVER BD, pinch the fixture on
the upper center of the board and remove the board.
A
Cable from the Valve
Connection BD 1
Cable from the Main BD
4. After the replacement, press the power key while holding the reset key and
enter the value on the mixed pump to the MD Coefficient column - Unit
Calib. window - Instrument set - Service - System.
1. Set the measurement mode to closed mode and turn the analyzer power off.
NOTE
If the measurement mode is not set to closed mode, the SAMPLER
SENSOR BD may be impossible to remove because of parts in the way.
2. Open the front cover according to the procedures described in “Opening the
Front Cover”.
3. Remove the three PS3 × 6 screws and SAMPLER SENSOR BD from the
MS-730V sampler unit.
SAMPLER SENSOR BD
4. When attaching the board, adjust the hole A on the board and unit. A bar of
the 2 mm diameter enables easy adjustment.
2. Remove all the cable connectors and piping tubes which are connected to the
LIQUID SENSOR BD.
3. Remove the four PS3 × 6 screws which fix the LIQUID SENSOR BD to the
JQ-730V chassis and remove the board.
2. Remove the five screws which fix the rear cover of the CBC measuring unit
and remove the rear cover.
4. Remove the two screws which fix the CONNECTION BD and remove the
board.
NOTE
The screw A is ground. Firmly fix the screw when attaching it.
1. Open the front cover according to the procedures described in “Opening the
Front Cover”.
4
2. Remove the two PSW3 × 6 screws A which fix the drip-proof cover and
open the cover.
3. Remove the four screws which fix the PRINTER DRIVER BD and lift up
the board.
4. Remove all the cable connectors which are connected to the PRINTER
DRIVER BD and remove the board.
1. Open the front cover according to the procedures described in “Opening the
Front Cover”.
3. Remove all the connectors which are connected to the LCD BD.
4. Remove the four PS3 × 6 screws C which secure the LCD BD and the board.
A
A
C
A
C
Inside the
front panel
1. Open the front cover according to the procedures described in “Opening the
Front Cover”.
4
2. Remove the 11 PSW3 × 6 screws A which secure the drip-proof cover,
release the two locking wire saddles B and remove the drip-proof cover
(refer to the illustration).
4. Remove the two PS3 × 6 screws D which secure the KEY BD, and remove
the board.
A
C
A
Inside the
front panel
1. Open the front cover according to the procedures described in “Opening the
Front Cover”.
3. Remove the ISOLATION SHEET which is hooked to the hole of the inverter
cover (73) and two PS3 × 6 screws C.
4. Remove the two PS3 × 6 screws D which secure the inverter and inverter.
A
Inverter
A
Inverter cover
D
ISOLATION SHEET
C (INVERTER73)
Inside the
front panel
To adjust the sensor and check and enter the corrected value, the Service screen
must be displayed. To enter the Service screen, change the operator to “Factory”.
3. The password is requested. Enter “4321” and press the Enter key.
The adjustment is performed on the Sensor Monitor screen. Press the System key
→ Service key → Sensor Monitor key.
The liquid sensor of the manometer detects the fluid level by the refraction of
light. When liquid is in the manometer, the refractive index of the plastic and
liquid is almost the same and the LED light travels straight to the light receiving
element (PTr) without refraction.
With liquid
LED PTr
Without liquid
LED PTr
When there is no liquid, the LED light refracts and does not reach the light
receiving element (PTr) because the refractive index of the air and plastic is
different.
The range of the manometer voltage is 1.5 V or less when there is liquid or 3.5
V or more when there is no liquid. If the voltage is out of the range, press the DP
Calib. key and adjust the sensor voltage.
WBC
Up: WBC manometer upper sensor
Lo360: WBC manometer middle sensor 5
Lo500: WBC manometer lower sensor
RBC
Up: RBC manometer upper sensor
Lo360: RBC manometer middle sensor
Lo500: RBC manometer lower sensor
Change the WBC and RBC manometer voltage. Press the desired item to change
and enter the setting value with the numeric key pad. The setting range is 1
to 255. The higher the value, the more voltage is generated. After the setting,
press the Sensor Moni. key to return to the Sensor Monitor screen and check the
voltage.
Open the Unit Correction Coefficient screen. Press the System key → Service
key → Instrument Set key → Unit Calib. key.
Select the desired item to change. Enter the value with the numeric keypad and
press the Enter key.
When the CBC measuring unit is replaced, the correction coefficient for the new
unit is already set at the factory. So you do not need to set the coefficient.
50 51
52 49
HGB cover
Pull the cover right
and forward.
Adjust the HGB voltage by turning the volume checking the HGB LED ON and
OFF voltage on the Sensor Monitor screen.
Normal value of the HGB voltage (with water in the measurement bath):
HEG LED ON: 3.0 ±0.5 V
HGB LED OFF: 0.5 V or less
The HGB sensor cancels the effect of ambient light and reads the difference of
HGB LED ON and OFF. Do not adjust the HGB LED OFF voltage. Only check
it. If the voltage is out of range the circuit may be damaged.
The diluter part creates the 200 times diluted blood sample by setting the
correction coefficient of XX% for the sample. The correction coefficient is set at
the factory. Normally you only need to check but not change the coefficient.
When the Mixed Pump unit is replaced, the dilution ratio correction coefficient
for the new unit is already set at the factory. So you do not need to set the
coefficient.
Open the Unit Correction Coefficient screen. Press the System key → Service
key → Instrument Set key → Unit Calib. key.
Select the MP Coefficient value, enter the value with the numeric keypad and
press the Enter key.
Sensor 4
Sensor 3
Sensor 2
Sensor 1
The 5 reagents are connected to the inlet/outlet unit. The following 4 optical
sensors are on the unit to judge the presence and absence of the reagents.
Sensor 1
Sensor 3
Sensor 4
When the sensor voltage is over 2.5 V, the sensor judges it is in air. When the
sensor voltage is under 2.5 V, the sensor judges it is in liquid. To leave 1.0 V
margin between the two states, the sensor voltage must be adjusted to be less
than 1.5 V in liquid and more than 3.5 V in air.
With liquid
PTr
LED
Without liquid
PTr
LE
The range of the voltage is 1.5 V or less when there is liquid or 3.5 V or more
when there is no liquid. If the voltage is out of the range, press the DP Calib. key
and adjust the sensor voltage.
Change the setting value of the liquid sensor voltage. Press the desired item to
change and enter the setting value with the numeric key pad. The setting range is
1 to 255 and the more value is entered, the more voltage is generated. After the
setting, press the Sensor Moni. key to return to the Sensor Monitor screen and
check the voltage.
There is a VR to adjust the pressure sensor on the UT-7236 board but a dedicated
tool is needed to apply correct pressure. So do not touch the VR.
Measure particles and perform optical adjustment (rough and fine) on the Adjust
flow cell window. For details, refer to “Checking the Particle Distribution Width”
in Section 9 of the operator’s manual.
2. When the sampling mode is Closed, press the key and change the mode to
Open.
3. Press the count switch to aspirate and count the YZ-0194 normal particles.
5. If the CV is more than 7.0 %, press the Clean key to clean the flow cell. If
the CV is still more than 7.0 % after cleaning, adjust the flow cell position.
NOTE
• Before fine adjustment, do rough optical adjustment. Measure MEK-
CAL 3 to 5 times.
• Use MEK-CAL for fine optical adjustment. You cannot adjust it correctly
with human blood.
• Use MEK-CAL before the expiration date and follow the storage
condition.
2. Press the Ideal Peak key to display the MEK-CAL Ideal Peak window.
Switches between the window to register data
for fine optical adjustment and the window to
enter MEK-CAL optical center assay value.
3. Enter the FS, FL and SD ideal peak and FS threshold. Enter the ideal peak
on the assay sheet of the MEK-CAL.
i) Touch the desired item.
ii) Enter the number with the numeric key pad and press the Enter key.
6. Press the Data key to load the measurement data of the New Sample SEQ#.
The scattergram is displayed.
8. When the sample is optimum, press the Register key to register the
measurement data to the data list. Check that the Average Peak is within ±2
of the assay value.
9. If the Average Peak is not within ±2 of the assay value, adjust the gain.
• Auto adjustment
Press the Auto key. The Target Gain is automatically adjusted and set to Current
Gain.
• Manual adjustment
Enter the value for Edit Gain. Use the arrow keys to move the cursor. Press the
Manual key to set the value entered in Edit Gain to the Current Gain.
When adjusting the vertical axis of the scattergram, adjust the FS Gain. When
adjusting the horizontal axis, adjust the FL Gain and SD Gain.
FS 127
0 127 255
FL
10. Press the Refresh key and check the settings. The scattergrams are displayed
again using the ratio of Edit Gain and Current Gain.
If the Target Gain is out of the 0 to 255 range or the center of the MEK-CAL
is not calculated, an “Out of range” message is displayed, Target Gain is not
displayed and the Auto key is not available. In this case, do rough adjust flow
cell again.
3. Check that there is no difference larger than ±2 between Average Peak and
Ideal Peak.
If there is a difference larger than ±2, delete bad data from the list or adjust
the gain.
Mode
Check that the “CAL” is displayed.
Average Peak
The average value of the
data registered to the list.
Ideal Peak
The peak target value set on the
optical sample setting window.
Deleting Data
• To delete one data, select the desired data to delete and press the Delete key.
• To delete all data, press the Delete key without selecting data. When a
confirmation message appears, press the Yes key to delete all data.
1. Press the Delete key on the History window of the Adj. Flowcell screen. A
confirmation message appears.
Calibrate the touch screen when the pressed position and operating position do
not match.
1. Hold down the [ Reset] key until the emergency stop message disappears.
Then press the [ Eject] key while holding down the [ Reset] key. The
Touch panel calibration screen appears.
NOTE
Do not use a sharp object to press the mark. Use your finger.
Setting Items
UT-7241 MAIN BD
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP0301 Vcc (primary power ED 2.87 to 2.93 V: Low Analyzer cannot start. 4
source) voltage detection > 2.93: High
TP0302 5 V (secondary power ED 4.059 to 4.141 V: Low Analyzer cannot start. 3
source) voltage detection > 4.141: High
TP0303 Address bit (A0) is not ED — — 4
used.
TP0304 CPLD spare terminal is ED — — 4
not used 5
TP0305 SRAM chip select ED Address D80000 to Cannot operate correctly 4
DFFFFF: Low (cannot set).
TP0306 CPU input reset signal ED About 100 ms after Analyzer cannot start. 3
detecting 5 V power
source: cancel reset
TP0401 FPGA JTAG connection ED — — 5
signal (TCK) is not used.
TP0402 FPGA JTAG connection ED — — 5
signal (TDI) is not used.
TP0403 FPGA JTAG connection ED — — 5
signal (TMS) is not used.
TP0404 FPGA JTAG connection ED — — 5
signal (TDO) is not used.
TP0405 FPGA spare terminal is ED — — 4
not used.
TP0406 FPGA internal circuit ED 2.4 to 2.6 V Cannot operate correctly. 5
auxiliary power (2.5 V)
(MAIN BD circuit control
FPGA)
TP0407 FPGA core power (1.2 V) ED 1.19 to 1.25 V Cannot operate correctly. 5
(MAIN BD circuit control
FPGA)
TP0501 FPGA JTAG connection ED — — 2
signal (TCK) is not used.
TP0502 FPGA JTAG connection ED — — 2
signal (TDI) is not used.
TP0503 FPGA JTAG connection ED — — 2
signal (TMS) is not used.
TP0504 FPGA JTAG connection ED — — 2
signal (TDO) is not used.
TP0505 FPGA internal circuit ED 2.4 to 2.6 V Cannot operate correctly. 2
auxiliary power (2.5 V)
(CBC system control
FPGA)
TP0506 FPGA core power (1.2 ED 1.19 to 1.25 V Cannot operate correctly. 2
V) (CBC system control
FPGA)
TP0601 FPGA JTAG connection ED — — 1
signal (TCK) is not used.
TP0602 FPGA JTAG connection ED — — 1
signal (TDI) is not used.
TP0603 FPGA JTAG connection ED — — 1
signal (TMS) is not used.
TP0604 FPGA JTAG connection ED — — 1
signal (TDO) is not used.
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP0605 FPGA internal circuit ED 2.4 to 2.6 V Cannot operate correctly. 2
auxiliary power (2.5 V)
(5 DIFF system control
FPGA)
TP0606 FPGA core power (1.2 V) ED 1.19 to 1.25 V Cannot operate correctly. 1
(5 DIFF system control
FPGA)
TP0801 LCD controller core ED 1.764 to 1.854 V LCD does not display 3
power (1.8 V) correctly.
TP1102 Heater control ED > 40°C: Low Cannot control the heater. 5
TP1401 Vcc voltage ED 3.25 to 3.43 V Analyzer cannot start. 5
TP1402 5 V voltage ED 4.93 to 5.21 V Analyzer cannot start. 5
TP1403 3.3 V voltage ED 3.25 to 3.43 V Analyzer cannot start. 5
TP1404 ED ED — — 5
TP1405 Backup system voltage ED When the power is on: Cannot operate correctly. 5
2.65 to 3.43 V Setting data and clock are
When the power is off: incorrect.
2.4 to 3.1 V
TP1406 Backup system voltage 1 ED When the power is on: Cannot operate correctly. 5
(to check the current) 2.65 to 3.43 V Setting data and clock are
When the power is off: incorrect.
2.4 to 3.1 V
TP1407 Backup system voltage 2 ED When the power is on: Cannot operate correctly. 5
(to check the current) 2.65 to 3.43 V Setting data and clock are
When the power is off: incorrect.
2.4 to 3.1 V
TP1408 Power voltage for the ED 4.93 to 5.21 V Internal printer cannot 5
internal printer operate correctly.
TP1917 Read signal (5 V system: ED At read access: Low Analyzer does not operate 3
CPU output) correctly.
TP1918 Write signal (low byte) (5 ED At write access: Low Analyzer does not operate 3
V system: CPU output) correctly.
TP1919 Write signal (high byte) (5 ED At write access: Low Analyzer does not operate 3
V system: CPU output) correctly.
TP1920 DRAM low column ED When address 400000 to Analyzer does not operate 3
address strobe (5 V 5FFFFF is selected: Low correctly.
system: CPU output)
TP1921 DRAM high column ED When address 400000 to Analyzer does not operate 3
address strobe (5 V 5FFFFF is selected: Low correctly.
system: CPU output)
TP1922 Chip select (area 0) (5 V ED When address 000000 to Analyzer does not operate 3
system: CPU output) 1FFFFF is selected: Low correctly.
TP1923 Chip select (area 1) (5 V ED When address 200000 to Analyzer does not operate 3
system: CPU output) 3FFFFF is selected: Low correctly.
TP1924 DRAM low address strobe ED When address 400000 to Analyzer does not operate 3
(5 V system: CPU output) 5FFFFF is selected: Low correctly.
TP1925 Chip select (area 3) (5 V ED When address 600000 to Analyzer does not operate 3
system: CPU output) 7FFFFF is selected: Low correctly.
TP1926 Chip select (area 4) (5 V ED When address 900000 to Analyzer does not operate 3
system: CPU output) 9FFFFF is selected: Low correctly.
TP1927 Chip select (area 5) (5 V ED When address A00000 to Analyzer does not operate 3
system: CPU output) BFFFFF is selected: Low correctly.
TP1928 Chip select (area 6) (5 V ED When address C00000 to Analyzer does not operate 3
system: CPU output) DFFFFF is selected: Low correctly.
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP1929 Watchdog timer overflow ED At watchdog timer When the CPU runs away, 3
(5 V system: CPU output) overflow: Low the system is not reset.
TP1930 Spare not used — — — 2
TP1931 CPU output reset signal (5 ED At BOOT start: Low Analyzer cannot start. 4
V system: CPU output) At system start: High
TP1932 RS232C (for PRINTER) ED Normal times: High Instrument that is 3
serial send signal When sending data: serial connected external 232C
data connector (for option) 5
cannot operate correctly.
TP1933 RS232C (for PC) serial ED Normal times: High Instrument that is 3
send signal When sending data: serial connected external 232C
data connector (for serial
board connector 2) cannot
operate correctly.
TP1934 RS232C (for barcode ED Normal times: High Instrument that is 2
reader) serial send signal When sending data: serial connected external 232C
data connector (for barcode
reader) cannot operate
correctly.
TP1935 Internal printer serial send ED Normal times: High Internal printer cannot 5
signal When sending data: serial operate correctly.
data
TP1936 RS232C (for PRINTER) ED Normal times: High Instrument that is 3
serial receive signal When receiving data: connected external 232C
serial data connector (for option)
cannot operate correctly.
TP1937 RS232C (for PC) serial ED Normal times: High Instrument that is 3
receive signal When receiving data: connected external 232C
serial data connector (for serial
board connector 2) cannot
operate correctly.
TP1938 RS232C (for barcode ED Normal times: High Instrument that is 2
reader) serial receive When receiving data: connected external 232C
signal serial data connector (for barcode
reader) cannot operate
correctly.
TP1939 Internal printer serial ED Normal times: High Internal printer cannot 5
receive signal When receiving data: operate correctly.
serial data
TP1940 CPU input reset signal ED At reset: Low Analyzer does not operate 3
correctly.
TP1941 CPU standby signal ED At reset: Low Analyzer does not operate 3
correctly.
TP1942 NMI ED At reset: Low Analyzer does not operate 3
correctly.
TP1943 WAIT signal from the ED When accessing the LCD LCD does not display 4
LCD controller controller: Low correctly.
TP1944 Data bus (D0) (5 V ED Normal times: High Analyzer does not operate 2
system: CPU output) Access times: depend on correctly.
the data
TP1945 Data bus (D1) (5 V ED Normal times: High Analyzer does not operate 2
system: CPU output) Access times: depend on correctly.
the data
TP1946 Address bus (A1) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) Access times: depend on correctly.
the data
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP1947 Address bus (A2) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) Access times: depend on correctly.
the data
TP1948 Power off request signal ED Normal times: Low Cannot power off correctly. 3
from CPU When requesting off: High
TP1949 LCD controller memory ED When LCD controller LCD does not display 4
bank change internal register is correctly.
selected: Low
When VRAM is selected:
High
TP1950 Chip select for SD card ED Low at address C40000 to Cannot use SD card IF. 4
controller C7FFFF
TP1951 Chip select for USB ED Low at address C80000 to Cannot use USB IF. 2
controller CBFFFF
TP1952 SRAM chip select 3 for ED Low at address B00000 to Cannot memory scatter. 4
scatter memory B7FFFF
TP1953 SRAM chip select 2 for ED Low at address A80000 to Cannot memory scatter. 4
scatter memory AFFFFF
TP1954 SRAM chip select 1 for ED Low at address A00000 to Cannot memory scatter. 5
scatter memory A7FFFF
TP1955 SRAM chip select for ED Low at address 800000 to Cannot display scatter 1
scatter decompression 97FFFF correctly.
TP1956 Chip select for driver BD ED Low at address 9A0000 to Actuator does not operate 4
(FPGA) 9AFFFF correctly.
TP1957 Chip select for analog ED Low at address 9B0000 to Cannot measure the CBC 2
system FPGA 9BFFFF and 5 DIFF data correctly.
TP1958 Chip select for MAIN BD ED Low at address 990000 to Analyzer does not operate 4
circuit control FPGA 99FFFF correctly.
TP1959 Chip select for LCD ED Low at address C00000 to LCD does not display 4
controller C3FFFF correctly.
TP1960 Chip select for FLASH ED Low at address 000000 to Analyzer does not operate 5
ROM 3FFFFF correctly.
TP1961 Switch config or download ED Config: Low Analyzer does not operate 2
of FPGA data Download: High correctly or cannot down
load FPGA data correctly.
TP1962 Write signal (high byte) ED At write access: Low Analyzer does not operate 4
correctly.
TP1963 Write signal (low byte) ED At write access: Low Analyzer does not operate 4
correctly.
TP1964 Write signal (high and low ED At write access: Low Analyzer does not operate 5
byte) correctly.
TP1965 SRAM high byte select ED At SRAM access: Low Analyzer does not operate 5
signal correctly.
TP1966 SRAM low byte select ED At SRAM access: Low Analyzer does not operate 5
signal correctly.
TP1967 Read signal ED At read access: Low Analyzer does not operate 1
correctly.
TP1968 DRAM low address strobe ED Low when selecting Analyzer does not operate 4
the address 400000 to correctly.
5FFFFF
TP1969 DRAM high column ED Low when selecting Analyzer does not operate 4
address strobe (5 V the address 400000 to correctly.
system: CPU output) 5FFFFF
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP1970 DRAM low column ED Low when selecting Analyzer does not operate 4
address strobe (5 V the address 400000 to correctly.
system: CPU output) 5FFFFF
TP1971 System lock ED 25 MHz Analyzer does not operate 2
correctly.
TP1972 RTC clock enable signal ED 32.768 kHz output from Analyzer does not start. 4
RTC at High
TP1973 System reset ED At rest: Low Analyzer does not start. 1 5
TP1974 Power ON or OFF signal ED ON: High Analyzer does not operate 5
other than 5 or 3.3 V. OFF: Low correctly.
TP1975 5 V power ON or OFF ED ON: High Analyzer does not start. 5
control signal OFF: Low
TP1976 3.3 V power ON or OFF ED ON: High Analyzer does not start. 5
control signal OFF: Low
TP1977 Power OFF request signal ED Normal times: Low Cannot turn the power OFF 4
When requesting OFF: correctly.
High
TP1978 ROM for FPGA → CPLD ED Normal times: Low Cannot download the 2
serial data When receiving data: FPGA data correctly.
serial data
TP1979 Configuration completion ED Normal times: Low Analyzer does not start. 4
signal (MAIN BD circuit At config completion:
control FPGA) High
TP1980 Configuration completion ED Normal times: Low Analyzer does not start. 1
signal (CBC system At config completion:
control FPGA) High
TP1981 Configuration completion ED Normal times: Low Analyzer does not start. 1
signal (5 DIFF system At config completion:
control FPGA) High
TP1982 Configuration completion ED Normal times: Low Analyzer does not start. 5
signal (driver BD control At config completion:
FPGA 1) High
TP1983 Chip select for ROM ED Normal times: High Cannot download the 2
for FPGA data storage When selecting ROM: FPGA data correctly.
download (MAIN BD Low
circuit control FPGA)
TP1984 Chip select for ROM ED Normal times: High Cannot download the 2
for FPGA data storage When selecting ROM: FPGA data correctly.
download (CBC system Low
control FPGA)
TP1985 Chip select for ROM ED Normal times: High Cannot download the 2
for FPGA data storage When selecting ROM: FPGA data correctly.
download (5 DIFF system Low
control FPGA)
TP1986 Chip select for ROM ED Normal times: High Cannot download the 2
for FPGA data storage When selecting ROM: FPGA data correctly.
download (driver BD Low
control FPGA 1)
TP1987 Serial data for ROM for ED Normal times: High Cannot download the 3
CPLD → FPGA download When receiving data: FPGA data correctly.
serial data
TP1988 ROM clock for CPLD → ED When accessing the ROM Cannot download the 3
FPGA for FPGA: 3.125 MHz FPGA data correctly.
TP1989 Configuration enable ED Normal times: High Analyzer does not start. 2
signal (FPGA in MAIN At config enable: Low
BD)
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP1990 Configuration enable ED Normal times: High Analyzer does not start. 5
signal (FPGA in DRIVER At config enable: Low
BD)
TP1991 RTC clock ED 32.768 kHz Analyzer does not start. 3
TP1992 SRAM chip select ED Low at address D80000 to Analyzer does not operate 5
DFFFFF correctly (cannot set the
settings etc.).
TP1993 DRIVER BD → MAIN ED Normal times: High Actuator does not operate 4
BD serial communication When receiving data: correctly.
serial data
TP1994 MAIN BD → DRIVER ED Normal times: High Actuator does not operate 5
BD serial communication When receiving data: correctly.
serial data
TP1995 Touch panel controller ED Normal times: High Touch panel is disabled. 3
→ CPU serial When receiving data:
communication serial data
TP1996 CPU → touch panel ED Normal times: High Touch panel is disabled. 3
controller serial When receiving data:
communication serial data
TP1997 CPU → serial EEPROM ED Normal times: High Cannot set the liquid 1
of LIQUID SENSOR BD When receiving data: sensor.
SPI serial data serial data
TP1998 Serial EEPROM of ED Normal times: High Cannot set the diluter. 5
MIXED PUMP BD → When receiving data:
CPU SPI serial data serial data
TP1999 Chip select for serial ED Normal times: High Cannot set the diluter. 5
EEPROM of MIXED When accessing: Low
PUMP BD
TP2000 Clock for serial EEPROM ED Normal times: Low Cannot set the diluter. 5
of MIXED PUMP BD At clock output: 1 MHz
TP2001 CPU → serial EEPROM ED Normal times: High Cannot set the diluter. 4
of MIXED PUMP BD SPI When receiving data:
serial data serial data
TP2002 Chip select for serial ED Normal times: High Cannot set the wave sensor. 1
EEPROM of LIQUID When accessing: Low
SENSOR BD
TP2003 Clock for serial EEPROM ED Normal times: Low Cannot set the wave sensor. 1
of LIQUID SENSOR BD At clock output: 1 MHz
TP2004 Serial EEPROM of ED Normal times: High Cannot set the wave sensor. 1
LIQUID SENSOR BD → When receiving data:
CPU SPI serial data serial data
TP2005 Clock for buzzer control ED Normal times: High Cannot buzz correctly. 3
When controlling: clock
output
TP2006 RS232C (for CRP) serial ED Normal times: High Instrument that is 5
send signal When sending data: serial connected to the external
data 232C connector (for serial
port connector 1) does not
operate correctly.
TP2007 Configuration completion ED Normal times: Low Analyzer does not start. 5
signal (driver BD control At config completion:
FPGA 2) High
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP2008 RS232C (for CRP) serial ED Normal times: High Instrument that is 5
receive signal When receiving data: connected to the external
serial data 232C connector (for serial
port connector 1) does not
operate correctly.
TP2009 Chip select for ROM ED Normal times: High Cannot download the 2
for FPGA data storage When selecting ROM: FPGA data correctly.
download (driver BD Low 5
control FPGA 2)
TP2010 USB overcurrent detection ED Normal times: High Cannot detect the over 1
flag When detecting current failure of USB IF.
overcurrent: Low
TP2011 Spare of LCD controller ED 4
— —
port, not used
TP2012 Spare of LCD controller ED 4
— —
port, not used
TP2013 Chip select for FPGA ED Normal times: High Cannot perform FPGA 2
data storage ROM When selecting ROM: configuration correctly.
configuration (MAIN BD Low Analyzer does not start.
circuit control FPGA)
TP2014 Clock for FPGA ED Normal times: High Cannot perform FPGA 3
data storage ROM At config: clock output configuration correctly.
configuration (MAIN BD Analyzer does not start.
circuit control FPGA)
TP2015 Serial data for FPGA → ED Normal times: High Cannot perform FPGA 3
FPGA ROM configuration When receiving data: configuration correctly.
(MAIN BD circuit control serial data Analyzer does not start.
FPGA)
TP2016 Chip select for FPGA data ED Normal times: High Cannot perform FPGA 2
storage ROM (MAIN BD When selecting ROM: configuration correctly or
circuit control FPGA) Low cannot download the FPGA
data correctly. Analyzer
does not start.
TP2017 FPGA data storage ROM ED Normal times: High Cannot perform FPGA 3
clock (MAIN BD circuit At config: clock output configuration correctly or
control FPGA) cannot download the FPGA
data correctly. Analyzer
does not start.
TP2018 Input serial data of FPGA ED Normal times: High Cannot perform FPGA 3
ROM When receiving data: configuration correctly or
serial data cannot download the FPGA
data correctly. Analyzer
does not start.
TP2019 Output serial data of ED Normal times: High Cannot perform FPGA 2
FPGA ROM When receiving data: configuration correctly or
serial data cannot download the FPGA
data correctly. Analyzer
does not start.
TP2020 RBC hardware noise flag ED Normal times: Low Cannot detect the RBC 1
When detecting the hardware noise.
hardware noise: High
TP2021 RBC threshold detection ED Normal times: Low Cannot measure the RBC 1
flag When detecting: High and PLT data correctly.
TP2022 RBC peak detection ED Normal times: Low Cannot measure the RBC 1
When detecting: High and PLT data correctly.
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP2023 WBC hardware noise flag ED Normal times: Low Cannot detect the WBC 1
When detecting the hardware noise.
hardware noise: High
TP2024 WBC threshold detection ED Normal times: Low Cannot measure the WBC 1
flag (High side) When detecting: High (CBC) data correctly.
TP2025 WBC threshold detection ED Normal times: Low Cannot measure the WBC 1
flag (Low side) When detecting: High (CBC) data correctly.
TP2026 WBC peak detection ED Normal times: Low Cannot measure the WBC 1
When detecting: High (CBC) data correctly.
TP2027 Liquid sensor detection ED Normal times: Low Cannot measure the RBC 2
signal of RBC manometer When detecting: High and PLT data correctly.
500 μl
TP2028 Liquid sensor detection ED Normal times: Low Cannot measure the RBC 2
signal of RBC manometer When detecting: High and PLT data correctly.
360 μl
TP2029 Liquid sensor detection ED Normal times: Low Cannot measure the RBC 2
signal of RBC manometer When detecting: High and PLT data correctly.
upper part
TP2030 Liquid sensor detection ED Normal times: Low Cannot measure the WBC 2
signal of WBC manometer When detecting: High (CBC) data correctly.
500 μl
TP2031 Liquid sensor detection ED Normal times: Low Cannot measure the WBC 2
signal of WBC manometer When detecting: High (CBC) data correctly.
360 μl
TP2032 Liquid sensor detection ED Normal times: Low Cannot measure the WBC 2
signal of WBC manometer When detecting: High (CBC) data correctly.
upper part
TP2033 Clock for RBC AD ED Normal times: High Cannot measure the RBC 2
converter When receiving data: 1 and PLT data correctly.
MHz: clock output
TP2034 Clock for WBC AD ED Normal times: High Cannot measure the WBC 1
converter When receiving data: 1 (CBC) data correctly.
MHz: clock output
TP2035 Chip select of WBC ED Normal times: High Cannot store the WBC 2
manometer sensor setting When accessing: Low manometer sensor setting
data storage ROM value
TP2036 Chip select of digital ED Normal times: High Cannot set the WBC 2
potentiometers for WBC When accessing: Low manometer sensor. Cannot
manometer sensor measure the WBC (CBC)
data correctly.
TP2037 Spare, not used ED — — 2
TP2038 Spare, not used ED — — 2
TP2039 Chip select of RBC ED Normal times: High Cannot store the setting 2
manometer sensor setting When accessing: Low value of the RBC
data storage ROM manometer sensor.
TP2040 Chip select of digital ED Normal times: High Cannot set the RBC
potentiometers for RBC When accessing: Low manometer sensor. Cannot
manometer sensor measure the RBC data
correctly.
TP2041 RBC/WBC setting data ED Normal times: High Cannot set the RBC/WBC 2
storage ROM, SPI clock When accessing: clock manometer sensor. Cannot
for digital potentiometers output measure the RBC/WBC
(CBC) data correctly.
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP2042 CPU → RBC/WBC ED Normal times: High Cannot set the RBC/WBC 2
setting data storage ROM, When receiving data: manometer sensor. Cannot
SPI serial data for digital serial data measure the RBC/WBC
potentiometers (CBC) data correctly.
TP2043 RBC/WBC setting data ED Normal times: High Cannot set the RBC/WBC 2
storage ROM → CPU, When receiving data: manometer sensor. Cannot
SPI serial data serial data measure the RBC/WBC
(CBC) data correctly. 5
TP2044 Power ON or OFF control ED Power OFF: Low Cannot measure the HGB 2
signal for HGB LED Power ON: High correctly
TP2045 RBC measurement enable ED When measuring: Low Cannot measure the RBC 2
signal (relay ON/OFF When not measuring: and PLT data correctly.
control) High
TP2046 RBC CAL pulse enable ED When outputting the CAL Cannot check the RBC 2
signal (relay ON/OFF pulse: Low circuit.
control) When not outputting:
High
TP2047 RBC remove clog enable ED When remove clog is ON: Cannot remove the RBC 2
(relay ON/OFF control) Low clog.
When remove clog is
OFF: High
TP2048 RBC peak hold reset ED Normal times: High Cannot measure the RBC 1
signal At reset: Low and PLT data correctly.
TP2049 WBC measurement enable ED When measuring: Low Cannot measure the WBC 2
signal (relay ON/OFF When not measuring: (CBC) data correctly.
control) High
TP2050 WBC CAL pulse enable ED When outputting the CAL Cannot check the WBC 2
signal (relay ON/OFF pulse: Low (CBC) circuit.
control) When not outputting:
High
TP2051 WBC remove clog enable ED When remove clog is ON: Cannot remove the WBC 2
(relay ON/OFF control) Low (CBC) clog.
When remove clog is
OFF: High
TP2052 WBC peak hold reset ED Normal times: High Cannot measure the WBC 1
signal At reset: Low (CBC) data correctly.
TP2053 CAL pulse of RBC/WBC ED Normal times: High Cannot check the RBC/ 1
(CBC) When outputting the CAL WBC (CBC) circuit.
pulse: Low/High
TP2054 RBC AD conversion ED Bit 1, 2 (00): measurement Cannot measure the RBC 1
receive data selection bit 1 data and PLT data correctly.
Bit 1, 2 (01): electrode
voltage
Bit 1, 2 (10): ANA data
TP2055 RBC AD conversion ED Bit 1, 2 (00): measurement Cannot measure the RBC 1
receive data selection bit 2 data and PLT data correctly.
Bit 1, 2 (01): electrode
voltage
Bit 1, 2 (10): ANA data
TP2056 WBC AD conversion ED Bit 1, 2, 3 (000): Cannot measure the WBC 1
receive data selection bit 1 measurement data (CBC) data correctly.
Bit 1, 2, 3 (001): electrode
voltage
Bit 1, 2, 3 (010): ANA
data
Bit 1, 2, 3 (011): HGB
data
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP2057 WBC AD conversion ED Bit 1, 2, 3 (000): Cannot measure the WBC 1
receive data selection bit 2 measurement data (CBC) data correctly.
Bit 1, 2, 3 (001): electrode
voltage
Bit 1, 2, 3 (010): ANA
data
Bit 1, 2, 3 (011): HGB
data
TP2058 WBC AD conversion ED Bit 1, 2, 3 (000): Cannot measure the WBC 1
receive data selection bit 3 measurement data (CBC) data correctly.
Bit 1, 2, 3 (001): electrode
voltage
Bit 1, 2, 3 (010): ANA
data
Bit 1, 2, 3 (011): HGB
data
TP2059 Heater control ED At 40°C or more: Low Cannot control the heater. 1
TP2060 Chip select for FPGA ED Normal times: High Cannot perform FPGA 1
data storage ROM When selecting ROM: configuration correctly.
configuration (CBC Low Analyzer does not start.
system control FPGA)
TP2061 Serial data for FPGA → ED Normal times: High Cannot perform FPGA 1
FPGA ROM configuration When receiving data: configuration correctly.
(CBC system control serial data Analyzer does not start.
FPGA)
TP2062 Clock for FPGA ED Normal times: High Cannot perform FPGA 1
data storage ROM At config: clock output configuration correctly.
configuration (CBC Analyzer does not start.
system control FPGA)
TP2063 FPGA data storage ROM ED Normal times: High Cannot perform FPGA 2
chip select (CBC system When selecting ROM: configuration correctly or
control FPGA) Low cannot download the FPGA
data correctly. Analyzer
does not start.
TP2064 FPGA data storage ROM ED Normal times: High Cannot perform FPGA 2
clock (CBC system At config: clock output configuration correctly or
control FPGA) cannot download the FPGA
data correctly. Analyzer
does not start.
TP2065 Input serial data of the ED Normal times: High Cannot perform FPGA 2
FPGA ROM (CBC system When receiving data: configuration correctly or
control FPGA) serial data cannot download the FPGA
data correctly. Analyzer
does not start.
TP2066 Output serial data of the ED Normal times: High Cannot perform FPGA 2
FPGA ROM (CBC system When receiving data: configuration correctly or
control FPGA) serial data cannot download the FPGA
data correctly. Analyzer
does not start.
TP2067 5 DIFF FS threshold level ED Normal times: Low Cannot measure the 5 1
detection flag When detecting: High DIFF data correctly.
TP2068 5 DIFF FS peak detection ED Normal times: Low Cannot measure the 5 1
When detecting: High DIFF data correctly.
TP2069 5 DIFF AD conversion ED Bit 1, 2, 3 (000): FS data Cannot measure the 5 1
receive data selection bit 1 Bit 1, 2, 3 (001): FL data DIFF data correctly.
Bit 1, 2, 3 (010): SD data
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP2070 5 DIFF AD conversion ED Bit 1, 2, 3 (000): FS data Cannot measure the 5 1
receive data selection bit 2 Bit 1, 2, 3 (001): FL data DIFF data correctly.
Bit 1, 2, 3 (010): SD data
TP2071 5 DIFF AD conversion ED Bit 1, 2, 3 (000): FS data Cannot measure the 5 1
receive data selection bit 3 Bit 1, 2, 3 (001): FL data DIFF data correctly.
Bit 1, 2, 3 (010): SD data
TP2072 5 DIFF AD converter ED Normal times: High Cannot measure the 5 1
clock When receiving data; 1 DIFF data correctly. 5
MHz: clock output
TP2073 5 DIFF CAL pulse ED Normal times: High Cannot check the 5 DIFF 1
When outputting the CAL circuit.
pulse: Low/High
TP2074 5 DIFF CAL pulse output ED When outputting the Cannot check the 5 DIFF 1
ON/OFF enable pulse: Low circuit.
When not outputting:
High
TP2075 5 DIFF peak hold reset ED Normal times: High Cannot measure the 5 1
signal At reset: Low DIFF data correctly.
TP2076 Laser key detection output ED Normal times: Low Cannot detect the laser key 1
signal When detecting the laser ON/OFF. Cannot perform
key: High 5 DIFF measurement.
TP2077 Serial data for FPGA → ED Normal times: High Cannot perform FPGA 1
FPGA ROM configuration When receiving data: configuration correctly.
(5 DIFF control FPGA) serial data Analyzer does not start.
TP2078 FPGA data storage ROM ED Normal times: High Cannot perform FPGA 1
chip select (5 DIFF When selecting ROM: configuration correctly.
control FPGA) Low Analyzer does not start.
TP2079 Clock for FPGA data ED Normal times: High Cannot perform FPGA 1
storage ROM (5 DIFF At config: clock output configuration correctly.
control FPGA) Analyzer does not start.
TP2080 FPGA data storage ROM ED Normal times: High Cannot perform FPGA 2
chip select (5 DIFF When selecting ROM: configuration correctly or
control FPGA) Low cannot download the FPGA
data correctly. Analyzer
does not start.
TP2081 FPGA data storage ROM ED Normal times: High Cannot perform FPGA 2
clock (5 DIFF control At config: clock output configuration correctly or
FPGA) cannot download the FPGA
data correctly. Analyzer
does not start.
TP2082 Input serial data of FPGA ED Normal times: High Cannot perform FPGA 2
ROM (5 DIFF control When receiving data: configuration correctly or
FPGA) serial data cannot download the FPGA
data correctly. Analyzer
does not start.
TP2083 Output serial data of ED Normal times: High Cannot perform FPGA 2
FPGA ROM (5 DIFF When receiving data: configuration correctly or
control FPGA) serial data cannot download the FPGA
data correctly. Analyzer
does not start.
TP2084 LCD controller port ED Normal times: High LCD does not display 5
interrupt When an interrupt occurs: correctly.
Low
TP2085 USB controller interrupt ED Normal times: High USB cannot communicate 1
When an interrupt occurs: correctly.
Low
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP2086 SD card controller ED Normal times: High Cannot use the SD card IF. 3
interrupt When an interrupt occurs:
Low
TP2087 RBC, PLT data receive ED Normal times: High Cannot measure the RBC 3
control interrupt When an interrupt occurs: and PLT data correctly.
Low
TP2088 WBC (CBC), 5 DIFF data ED Normal times: High Cannot measure the WBC 3
receive control interrupt When an interrupt occurs: (CBC) and 5 DIFF data
Low correctly.
TP2089 Manometer sensor, ED Normal times: High Analyzer does not operate 3
each position sensor, When an interrupt occurs: correctly.
SW, DRIVER BD Low
communication interrupt
TP2090 USB controller interrupt ED Normal times: High USB cannot communicate 3
When an interrupt occurs: correctly.
Low
TP2091 SD card controller ED Normal times: High Cannot use the SD card IF. 2
interrupt When an interrupt occurs:
Low
TP2092 Touch panel controller, ED Normal times: High LCD does not display 3
RTC, LCD controller When an interrupt occurs: correctly. Touch panel is
interrupt Low disabled. Clock data is
incorrect.
TP2093 RS232C (for CRP) ED Normal times: High Instrument that is 3
interrupt When an interrupt occurs: connected to the external
Low 232C connector (for serial
port connector 1) does not
operate correctly.
TP2094 Reset signal of the ED At reset: Low Analyzer does not start. 1
FLASH ROM, USB USB cannot communicate
controller correctly.
TP2095 Baseline release control ED Normal times: Low Cannot measure the RBC 1
signal when receiving the When detecting the data: and PLT data correctly.
RBC data High
TP2096 RBC pulse width ED Normal times: Low 1
judgement flag, not used When the pulse width is —
abnormal: High
TP2097 Baseline release control ED Normal times: Low Cannot measure the WBC 1
signal when receiving the When detecting the data: (CBC) data correctly.
WBC (CBC) data High
TP2098 WBC (CBC) pulse width ED Normal times: Low 1
judgement flag, not used When the pulse width is —
abnormal: High
TP2099 5 DIFF data receive ED Normal times: High Cannot measure the 5 1
control interrupt When an interrupt occurs: DIFF data correctly.
Low
TP2100 Each position sensor, ED Normal times: High Analyzer does not operate 2
SW, DRIVER BD When an interrupt occurs: correctly.
communication interrupt Low
TP2101 Read signal for USB ED At read access: Low USB cannot communicate 1
controller correctly.
TP2102 Data bus (D0) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP2103 Data bus (D1) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2104 Data bus (D2) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2105 Data bus (D3) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly. 5
on the data
TP2106 Data bus (D4) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2107 Data bus (D5) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2108 Data bus (D6) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2109 Data bus (D7) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2110 Data bus (D8) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2111 Data bus (D9) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2112 Data bus (D10) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2113 Data bus (D11) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2114 Data bus (D12) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2115 Data bus (D13) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2116 Data bus (D14) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2117 Data bus (D15) (5 V ED Normal times: High Analyzer does not operate 3
system: CPU output) When accessing: depend correctly.
on the data
TP2118 Write signal for USB ED At write access: Low USB cannot communicate 1
controller correctly.
TP2119 Data bus (D0) (3 V ED Normal times: High Analyzer does not operate 1
system) When accessing: depend correctly.
on the data
TP2120 Data bus (D1) (3 V ED Normal times: High Analyzer does not operate 1
system) When accessing: depend correctly.
on the data
TP2121 Data bus (D2) (3 V ED Normal times: High Analyzer does not operate 1
system) When accessing: depend correctly.
on the data
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP2122 Data bus (D3) (3 V ED Normal times: High Analyzer does not operate 1
system) When accessing: depend correctly.
on the data
TP2123 Data bus (D4) (3 V ED Normal times: High Analyzer does not operate 1
system) When accessing: depend correctly.
on the data
TP2124 Data bus (D5) (3 V ED Normal times: High Analyzer does not operate 2
system) When accessing: depend correctly.
on the data
TP2125 Data bus (D6) (3 V ED Normal times: High Analyzer does not operate 1
system) When accessing: depend correctly.
on the data
TP2126 Data bus (D7) (3 V ED Normal times: High Analyzer does not operate 2
system) When accessing: depend correctly.
on the data
TP2127 Data bus (D8) (3 V ED Normal times: High Analyzer does not operate 2
system) When accessing: depend correctly.
on the data
TP2128 Data bus (D9) (3 V ED Normal times: High Analyzer does not operate 2
system) When accessing: depend correctly.
on the data
TP2129 Data bus (D10) (3 V ED Normal times: High Analyzer does not operate 2
system) When accessing: depend correctly.
on the data
TP2130 Data bus (D11) (3 V ED Normal times: High Analyzer does not operate 1
system) When accessing: depend correctly.
on the data
TP2131 Data bus (D12) (3 V ED Normal times: High Analyzer does not operate 2
system) When accessing: depend correctly.
on the data
TP2132 Data bus (D13) (3 V ED Normal times: High Analyzer does not operate 2
system) When accessing: depend correctly.
on the data
TP2133 Data bus (D14) (3 V ED Normal times: High Analyzer does not operate 2
system) When accessing: depend correctly.
on the data
TP2134 Data bus (D15) (3 V ED Normal times: High Analyzer does not operate 1
system) When accessing: depend correctly.
on the data
UT-7244 ANALOG BD
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP0101 –36 V (power voltage for E36 –35.801 to –36.493 V Cannot measure the data 1
data receiving part data correctly.
measurement)
TP0102 EA (GND for analog EA 0V 1
—
system circuit)
TP0103 +5 V (digital system ED 4.93 to 5.21 V Analyzer does not start. 1
circuit power voltage)
TP0104 ED (GND for digital ED 0V 1
—
system circuit)
TP0105 +15 V (analog system EA 14.55 to 15.75 V Cannot measure the data 4
circuit power voltage) correctly.
TP0106 –15 V (analog system EA –14.55 to –15.75 V Cannot measure the data 1
circuit power voltage) correctly.
TP0107 E36 (GND for data E36 0V 1
receiving part data —
measurement)
TP0108 +5 V (analog part digital EA 4.8 to 5.2 V Cannot measure the 5 1
system circuit power DIFF data correctly.
voltage) (5 DIFF)
TP0109 +5 V (analog part digital EA 4.8 to 5.2 V Cannot measure the 1
system circuit power RBC/WBC (CBC) data
voltage) (CBC) correctly.
TP0403 RBC measurement data EA Normal times: Low Cannot measure the RBC 6
ANALOG BD input When checking the and PLT data correctly.
signal (R-IN) measurement or circuit:
pulse input
TP0404 RBC electrode voltage EA Normal times: Low Cannot measure the 6
ANALOG BD input When measuring the electrode voltage of the
signal (R-ELE) electrode voltage: 36 V RBC measurement part.
or less
TP0405 Baseline correction signal EA Normal times: Low Cannot measure the RBC 6
of the RBC data receiving When checking the and PLT data correctly.
phase control part measurement or circuit:
(R-ROB) pulse input
TP0406 RBC data receiving phase EA 100 mV Cannot measure the RBC 6
control part threshold and PLT data correctly.
level (RCLLEV)
TP0407 Threshold level detection EA Normal times: Low Cannot measure the RBC 6
signal of RBC data When detected: High and PLT data correctly.
receiving phase control
part (R-RBP)
TP0408 Received data after EA Normal times: Low Cannot measure the RBC 6
passing the RBC Bessel When checking the and PLT data correctly.
filter (R-FIL) measurement or circuit:
pulse input
TP0409 Baseline release control EA Normal times: Low Cannot measure the RBC 6
signal when receiving the When detecting the data: and PLT data correctly.
RBC data (R-GATE) High
TP0410 RBC hardware noise flag EA Normal times: Low Cannot detect the RBC 6
(R-NOISE) When detecting the hardware noise.
hardware noise: High
TP0503 RBC threshold level EA Depend on the setting Cannot measure the RBC 3
(R-THR) and PLT data correctly.
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP0504 Received data after EA Normal times: Low Cannot measure the RBC 3
increasing the RBC setting When checking the and PLT data correctly.
sensitivity (R-ANA) measurement or circuit:
pulse input
TP0505 RBC threshold detection EA Normal times: Low Cannot measure the RBC 3
flag (R-PLS) When detected: High and PLT data correctly.
TP0506 RBC receiving data peak EA Normal times: Low Cannot measure the RBC 3
hold signal (R-PH) When checking the and PLT data correctly. 5
measurement or circuit:
pulse input
TP0507 RBC peak detection EA Normal times: Low Cannot measure the RBC 3
(R-PKD) When detected: High and PLT data correctly.
TP0603 WBC (CBC) measurement EA Normal times: Low Cannot measure the WBC 6
data ANALOG BD input When checking the (CBC) data correctly.
signal (W-IN) measurement or circuit:
pulse input
TP0604 WBC (CBC) electrode EA Normal times: Low Cannot measure the 5
voltage ANALOG BD When measuring the electrode voltage of the
input signal (W-ELE) electrode voltage: 36 V WBC (CBC) measurement
or less part.
TP0605 Baseline correction signal EA Normal times: Low Cannot measure the WBC 5
of the WBC (CBC) data When checking the (CBC) data correctly.
receiving phase control measurement or circuit:
part (W-ROB) pulse input
TP0606 WBC (CBC) data EA 100 mV Cannot measure the WBC 5
receiving phase control (CBC) data correctly.
part threshold level
(WCLLEV)
TP0607 Threshold level detection EA Normal times: Low Cannot measure the WBC 5
signal of WBC (CBC) When detected: High (CBC) data correctly.
data receiving phase
control part (W-RBP)
TP0608 Received data after EA Normal times: Low Cannot measure the WBC 6
passing the WBC (CBC) When checking the (CBC) data correctly.
Bessel filter (W-FIL) measurement or circuit:
pulse input
TP0609 Baseline release control EA Normal times: Low Cannot measure the WBC 6
signal when receiving When detecting the data: (CBC) data correctly.
the WBC (CBC) data High
(W-GATE)
TP0610 WBC (CBC) hardware EA Normal times: Low Cannot detect the WBC 5
noise flag (W-NOISE) When detecting the hardware noise.
hardware noise: High
TP0703 WBC (CBC) High EA Depend on the setting Cannot measure the WBC 3
threshold level (CBC) data correctly.
(W-THR-H)
TP0704 WBC (CBC) Low EA Depend on the setting Cannot measure the WBC 3
threshold level (CBC) data correctly.
(W-THR-L)
TP0705 Received data after EA Normal times: Low Cannot measure the WBC 3
increasing the WBC When checking the (CBC) data correctly.
(CBC) setting sensitivity measurement or circuit:
(W-ANA) pulse input
TP0706 WBC (CBC) Low EA Normal times: Low Cannot measure the WBC 3
threshold detection flag When detected: High (CBC) data correctly.
(W-PLS-L)
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP0707 WBC (CBC) High EA Normal times: Low Cannot measure the WBC 3
threshold detection flag When detected: High (CBC) data correctly.
(W-PLS-H)
TP0708 WBC (CBC) receiving EA Normal times: Low Cannot measure the WBC 3
data peak hold signal When checking the (CBC) data correctly.
(W-PH) measurement or circuit:
pulse input
TP0709 WBC (CBC) peak EA Normal times: Low Cannot measure the WBC 3
detection (W-PKD) When detected: High (CBC) data correctly.
TP0801 HGB measurement data EA Normal times: Low Cannot measure the HGB 6
When detected: several V correctly.
TP0802 WBC (CBC) AD EA Normal times: Low Cannot measure the WBC 2
converter receiving data When detected: several V (CBC) data correctly.
Cannot measure the HGB
correctly.
TP0803 RBC AD converter EA Normal times: Low Cannot measure the RBC 3
receiving data When detected: several V and PLT data correctly.
TP0901 Receiving data before 5 EA Normal times: Low Cannot measure the 5 4
DIFF FS peak hold (CL1) When checking the DIFF data correctly.
measurement or circuit:
pulse input
TP0902 5 DIFF FS measurement EA Normal times: Low Cannot measure the 5 4
data ANALOG BD input When checking the DIFF data correctly.
signal (ANA1) measurement or circuit:
pulse input
TP0903 5 DIFF FS receiving data EA Normal times: Low Cannot measure the 5 1
peak hold signal (PHA1) When checking the DIFF data correctly.
measurement or circuit:
pulse input
TP0904 Baseline correction signal EA Normal times: Low Cannot measure the 5 4
of the 5 DIFF FS data When checking the DIFF data correctly.
receiving phase control measurement or circuit:
part (RB1) pulse input
TP0905 5 DIFF FS data receiving EA 460 mV Cannot measure the 5 4
phase control part DIFF data correctly.
threshold level (RBTHR)
TP0906 Threshold level detection EA Normal times: Low Cannot measure the 5 1
signal of the 5 DIFF FS When detected: High DIFF data correctly.
data receiving phase
control part (RBP)
TP0907 Baseline release control EA Normal times: Low Cannot measure the 5 1
signal when receiving the When detecting the data: DIFF data correctly.
5 DIFF FS data (GATE) High
TP0908 Receiving data after EA Normal times: Low Cannot measure the 5 4
passing the 5 DIFF FS When checking the DIFF data correctly.
Bessel filter (FIL1) measurement or circuit:
pulse input
TP1001 5 DIFF FL measurement EA Normal times: Low Cannot measure the 5 4
data ANALOG BD input When checking the DIFF data correctly.
signal (ANA2) measurement or circuit:
pulse input
TP1002 Receiving data before EA Normal times: Low Cannot measure the 5 4
the 5 DIFF FL peak hold When checking the DIFF data correctly.
(CL2) measurement or circuit:
pulse input
Symbol Corresponding BD
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin Area
TP1003 Receiving data after EA Normal times: Low Cannot measure the 5 4
passing the 5 DIFF FL When checking the DIFF data correctly.
Bessel filter (FIL2) measurement or circuit:
pulse input
TP1004 5 DIFF FL receiving data EA Normal times: Low Cannot measure the 5 1
peak hold signal When checking the DIFF data correctly.
measurement or circuit:
pulse input 5
TP1101 5 DIFF SD measurement EA Normal times: Low Cannot measure the 5 5
data ANALOG BD input When checking the DIFF data correctly.
signal (ANA3) measurement or circuit:
pulse input
TP1102 Receiving data before the EA Normal times: Low Cannot measure the 5 5
5 DIFF SD peak hold When checking the DIFF data correctly.
measurement or circuit:
pulse input
TP1103 Receiving data after EA Normal times: Low Cannot measure the 5 5
passing the 5 DIFF SD When checking the DIFF data correctly.
Bessel filter (FIL3) measurement or circuit:
pulse input
TP1104 5 DIFF FL receiving data EA Normal times: Low Cannot measure the 5 2
peak hold signal (PHA2) When checking the DIFF data correctly.
measurement or circuit:
pulse input
TP1201 5 DIFF FS threshold level EA Depend on the setting Cannot measure the 5 1
(ANA1-THR) DIFF data correctly.
TP1202 Threshold level for 5 EA 460 mV Cannot measure the 5 2
DIFF FS peak detection DIFF data correctly.
TP1203 5 DIFF FS threshold EA Normal times: Low Cannot measure the 5 2
detection flag (THP) When detected: High DIFF data correctly.
TP1204 5 DIFF FS peak detection EA Normal times: Low Cannot measure the 5 2
(PKD) When detected: High DIFF data correctly.
TP1301 5 DIFF AD converter EA Normal times: Low Cannot measure the 5 1
receiving data When detected: several V DIFF data correctly.
TP1401 Not used — — — 5
UT-7243 DRIVER BD
Symbol Corresponding
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin
TP0101 +5 V (digital system circuit ED 4.93 to 5.21 V Analyzer does not start.
power voltage)
TP0102 +3.3 V (digital system circuit ED 3.25 to 3.43 V Analyzer does not start.
power voltage)
TP0103 +12 V (power voltage for E12 11.75 to 12.45 V Three-way valve, buzzer
three-way valve, pinch valve, and fan do not operate. LCD
buzzer, fan and LCD inverter) backlight is dark. Analyzer
does not operate correctly. 5
TP0104 ED (GND for digital system ED 0V
—
circuit)
TP0105 +15 V (power voltage for two- EM15V_2 14.55 to 15.75 V Two-way valve does not
way valve) 2 operate. Analyzer does not
operate correctly.
TP0106 +24 V (power voltage for ED 21.6 to 26.4 V Motor does not start.
motor) Analyzer does not operate
correctly.
TP0107 +15 V (power voltage for two- EM15V_1 14.55 to 15.75 V Two-way valve does not
way valve) 1 operate. Analyzer does not
operate correctly.
TP0108 ED (GND for digital system ED 0V
—
circuit)
TP0109 EDM12V (GND for three-way E12 0V
valve, pinch valve, buzzer, fan —
and LCD inverter)
TP0110 EM15V_2 (GND for two-way EM15V_2 0V
—
valve)
TP0111 EM15V_1 (GND for two-way EM15V_1 0V
—
valve)
TP0112 E24 (GND for motor) E24 0V —
TP0201 Chip select for FPGA data ED Normal times: High Cannot download the FPGA
storage ROM download When selecting ROM: Low data correctly.
(DRIVER BD control FPGA1)
TP0202 ROM clock for CPLD → ED When accessing the FPGA Cannot download the FPGA
FPGA ROM: 3.125 MHz data correctly.
TP0203 ROM download serial data for ED Normal times: High Cannot download the FPGA
CPLD → FPGA When receiving data: serial data correctly.
data
TP0204 Chip select for FPGA data ED Normal times: High Cannot download the FPGA
storage ROM download When selecting ROM: Low data correctly.
(DRIVER BD control FPGA2)
TP0205 Switch configuration or ED Config: Low Cannot download the FPGA
download of FPGA data Download: High data correctly.
TP0301 FPGA internal circuit ED 2.4 to 2.6 V Cannot operate correctly.
supplemental power (2.5 V)
(DRIVER BD control FPGA1)
TP0302 FPGA core power (1.2 V) ED 1.19 to 1.25 V Cannot operate correctly.
(DRIVER BD control FPGA1)
TP0501 FPGA internal circuit ED 2.4 to 2.6 V Cannot operate correctly.
supplemental power (2.5 V)
(DRIVER BD control FPGA2)
TP0502 FPGA core power (1.2 V) EA 1.19 to 1.25 V Cannot operate correctly.
(DRIVER BD control FPGA2)
UT-7242 POWER BD
Symbol Corresponding
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin
TP0101 E24 E24 0V —
TP0102 DC 24 V (switching power E24 21.6 to 26.4 V Analyzer does not start.
output)
TP0103 EDP (GND for internal printer EDP 0V
—
power)
TP0104 +5 VP (power voltage for EDP 4.93 to 5.21 V Internal printer does not
internal printer) operate.
5
TP0105 ED (GND for digital system ED 0V
—
circuit)
TP0106 +3.3 V (digital system circuit ED 3.25 to 3.43 V Analyzer does not start.
power voltage)
TP0107 ED (GND for digital system ED 0V
—
circuit)
TP0108 +5 V (digital system circuit ED 4.93 to 5.21 V Analyzer does not start.
power voltage)
TP0201 +22 V (voltage for generate E24 21.14 to 22.43 V Cannot measure the data
the analog power voltage –15 correctly.
V)
TP0202 E24 E24 0V —
TP0203 +22 V (voltage for generate E24 21.14 to 22.43 V Cannot measure the data
the analog power voltage +15 correctly.
V)
TP0204 E24 E24 0V —
TP0205 +15 V (analog system circuit EA 14.55 to 15.75 V Cannot measure the data
power voltage) correctly.
TP0206 EA (GND for analog system EA 0V
—
circuit)
TP0207 –15 V (analog system circuit EA –14.55 to –15.75 V Cannot measure the data
power voltage) correctly.
TP0208 E36 (GND for data receiving E36 0V
—
part data measurement)
TP0209 –36 V (power voltage for E36 –35.801 to –36.493 V Cannot measure the data
data receiving part data correctly.
measurement)
TP0301 +15 V (power voltage for two- EM15V_1 14.55 to 15.75 V Two-way valve does not
way valve) 1 operate. Analyzer does not
operate correctly.
TP0302 +12 V (power voltage for E12 11.75 to 12.45 V Three-way valve, buzzer
three-way valve, pinch valve, and fan do not operate. LCD
buzzer, fan and LCD inverter) backlight is dirk. Analyzer
does not operate correctly.
TP0303 +15 V (power voltage for two- EM15V_2 14.55 to 15.75 V Two-way valve does not
way valve) 2 operate. Analyzer does not
operate correctly.
TP0601 Voltage for generating the clog ED About 100 to 170 V Cannot generate the clog
removing voltage removing voltage.
Symbol Corresponding
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin
TP0101 –15 V (analog system circuit EA –14.55 to –15.75 V Cannot measure the data
power voltage) correctly.
TP0102 EA (GND for analog system EA 0V
—
circuit)
TP0103 +15 V (analog system circuit EA 14.55 to 15.75 V Cannot measure the data
power voltage) correctly.
TP0104 +5 V (digital system circuit ED 4.93 to 5.21 V Analyzer does not start.
power voltage)
TP0105 ED (GND for digital system ED 0V
—
circuit)
TP0106 Liquid sensor detection signal ED No liquid: 3.5 V or more Analyzer does not operate
(detergent) With liquid: 1.5 V or less correctly.
TP0107 Liquid sensor detection signal ED No liquid: 3.5 V or more Analyzer does not operate
(Isotonac3) With liquid: 1.5 V or less correctly.
TP0108 Liquid sensor detection signal ED No liquid: 3.5 V or more Analyzer does not operate
(Hemoynac3N) With liquid: 1.5 V or less correctly.
TP0109 Liquid sensor detection signal ED No liquid: 3.5 V or more Analyzer does not operate
(Hemolynac5) With liquid: 1.5 V or less correctly.
TP0201 Pressure data EA Normal times: about 3.0 V Analyzer does not operate
Abnormal sheath pressure: correctly.
4.5 V or more
Abnormal pressure sensor:
1.0 V or less
UT-7246 LCD BD
Symbol Corresponding
Check Purpose Judging Standard Extent of the Effect
Number Ground Pin
TP0101 Vcc voltage ED 3.25 to 3.43 V Analyzer does not start.
TP0102 5 V voltage ED 4.93 to 5.21 V Analyzer does not start.
TP0103 3.3 V voltage ED 3.25 to 3.43 V Analyzer does not start.
TP0104 ED ED — —
TP0105 +12 V (power voltage for E12 11.75 to 12.45 V Buzzer does not operate
buzzer, fan and LCD inverter) correctly. LCD backlight is
dirk.
TP0106 E12 (GND for buzzer, fan and E12
— —
LCD inverter)
To be Replaced Periodically
• Sampling nozzle
Replace the sampling nozzle once a year. It may affect the measurement
accuracy.
• Pinch valve tube
Replace the pinch valve tube once a year. It may affect the measurement
accuracy. For replacement, contact your Nihon Kohden representative.
• Rinse O ring
Replace the rinse O ring once a year. It may affect the measurement accuracy.
For replacement, contact your Nihon Kohden representative.
• Pump tube
Replace the pump tube every four months or every 3000 measurements. It may
affect the measurement accuracy.
• Filter
Replace the filter every 1000 measurements. It may affect the measurement
accuracy.
• Solenoid valve (replace some valves according to the status of use)
Replace the solenoid valve every 5 years or 25000 measurements. It may affect
the measurement accuracy. The number of measurement can be checked on
the third page of the User Maintenance window. For replacement, contact your
Nihon Kohden representative.
Refer to the following list for the code number and supply code of the periodic
replacement parts.
Periodic Model/
Name Supply code
replacement parts Code no.
Sampling nozzle (two Sample tube assy YZ-0341 T444C
nozzles)
Pinch valve tube Pharmed tube 1 × 3 - 260 6114-924208 —
Pinch valve tube Pharmed tube 100 6114-094961 —
Rinse O ring O ring AS568-003 925538 —
Pump tube Pump tube assy YS-001B1 T462
Filter (10 filters) Hemoglobin filter assy YS-002B2 T802
Solenoid valve — — —
Preparation
According to Section 2 “Preparation” of the operator’s manual, turn on the power
and check the obtained data from the hematology analyzer before the periodic
maintenance check by counting the diluent and MEK-5D hematology control.
The obtained data must be printed out with the optional printer or written for
the record of the maintenance check. Also check that the “Priming” operation is
6
normal after turning on the power. Prepare a container, syringe and CLEANAC·3
detergent to clean fluid path components such as aperture cap. When calibration
in capillary mode is required, prepare the T857 sample cup and T812 or T813
micro cap.
Appearance
Check that there is no damage or fluid leakage on the hematology analyzer.
Safety
Checking the power cord
• Check that there is no damaged AC plug and exposed wire on the power cord.
• Check that the 3-pin plug type power cord is used and the 3 pins and plug
housing are not deformed.
• Check that the resistance of the protective ground line of the power cord is 0.1
Ω or less by using an earth tester or check the continuity with a multimeter.
Check that the earth leakage current is 0.5 mA or less under normal condition.
Check that the earth leakage current is 1.0 mA or less under each single fault
condition.
Reagents
Check that Nihon Kohden recommended diluent, detergent and hemolysing
reagent are used. Check that these reagents are used before the expiration date.
Cleaning/Replacing
1. Press the Strong clean key on the Operation screen to start strong cleaning.
2. Disconnect the diluent, detergent and reagent tubes other than the waste fluid
tube from the right side panel of the hematology analyzer.
3. Press the Drain all key on the Operation screen to drain all the fluid inside
the hematology analyzer.
5. Open the front cover according to the procedure described in “Opening the
Front Cover” of Section 4 “Disassembly and Assembly”.
2. While pressing the Reset key on the front panel, press the Power key on the 6
front panel. The power turns on without priming. The Ready screen appears.
3. Press the System key on the screen. The System screen appears.
4. Press the Service key on the System screen. The Service screen appears.
5. Press the Sensor Monitor key on the Service screen. The Sensor Monitor
screen which displays each sensor output voltage appears.
6. Check that the sensor output voltages of WBC Up, WBC Lo 360, WBC Lo
500, RBC Up, RBC Lo 360, RBC Lo 500, Diluent 1, Diluent 2, Ly Hemo3
and Ly Hemo5 are 3.5 V or more under no fluid condition. Write down each
voltage on the check sheet.
7. Connect the diluent, detergent and reagent tubes to the right side panel of the
hematology analyzer.
8. Press the Operation key on the screen. The Operation screen appears.
9. Press the Refill all key on the Operation screen. The screen returns to the
Ready screen after priming operation is completed.
10. Check that the following sensor output voltages are displayed on the Sensor
Monitor screen. Write down each voltage on the check sheet. If there is a
sensor output voltage out of the acceptable range, adjust the sensor output
voltage for the acceptable range according to Section 5 “Adjustment”.
WBC Up, WBC Lo 360, WBC Lo 500, RBC Up, RBC Lo 360, RBC Lo
500, Diluent 1, Diluent 2, Ly Hemo3 and Ly Hemo5: 1.5 V or less
ELECTRODE: 17.7 V to 18.3 V
HGB LED ON: 1.5 V to 4.5 V (Adjust this voltage for 2.9 V to 3.1 V if it is
out of 1.5 V to 4.5 V.)
HGB LED OFF: 0.5 V or less
2. Press the Circuit Check key on the Operation screen. The Circuit check
screen which displays the check result of specified measurement parameters
appears.
3. Check that the following values are displayed on the Circuit Check screen.
WBC: 8.0±5%
RBC: 1.6±5%
HGB ON: 1.5 V to 4.5 V
HGB OFF: 0.5 V or less
MCV: 100±15%
PLT: 160±5%
Sensitivity and threshold
WBC: 5 (sensitivity), 4 (threshold)
RBC: 5 (sensitivity), AT (AUTO) or 5 (threshold)
PLT: 5 (threshold)
Background noise
1. Press the count switch to count the diluent so that the background noise is
checked. Refer to “Measuring Background Noise” in Section 2 “Preparation”
of the operator’s manual. The measurement result appears.
2. Check that the CV of FS and FL are 7% or less on the Adj. Flowcell screen.
If the CV is more than 7%, clean the flow cell, measure the polymer
microsphere suspensions again, and adjust the flow cell position.
3. Check that the obtained results are within the acceptable range. If the
obtained data is out of the range, go to “New Calibration Coefficients”.
NOTE
Before entering the new calibration coefficient or measured data, ask
the customer if you can change the setting. Some customers calibrate
the hematology analyzer for data consistency with other manufacturers’
hematology analyzers.
Software Version
1. Press the System key on the screen. The System screen appears.
2. Press the User Maint. key on the System screen. The User Maintenance
screen appears.
Fluid leakage
Check that there is no fluid leakage on the hematology analyzer, especially after
each automatic cleaning operation.
Dispense operation
Check that the constant volume of the diluent is properly dispensed when
pressing the Dispense key on the screen in pre-dilution mode.
Bar Code Reader Operation (When the bar code reader is installed)
1. Clean the light emitter and detector parts with a cotton swab moistened with
80% alcohol.
Others
Write down any other points on the check sheet.
General
You can display the total operating time, total number of counts, and number
of counts that filters and pump tubes are used to determine the maintenance
schedule.
When filters or tubes are used more than the following number of sample counts,
6
the messages appears on the Ready screen to prompt you to check and replace
them.
Filters: 1000 counts
Pump tubes: 3000 counts
Cap pierce needle: 1000 counts
The screen shows the number of counts that filters, pump tubes, cap pierce
needle, sampling nozzle and rinse unit are used.
After checking and replacing filters, pump tubes, cap pierce needle, sampling
nozzle and rinse unit, reset the counts to zero by pressing the Reset key.
Replacing Filters
Materials Required
• Powder-free gloves, lab coat, safety glasses
• Phillips screwdriver
• Tweezers
6
Procedure
Replace the filters when they are clogged, dirty and/or after every 1,000 sample
counts.
Replace the filter packing when they are dirty, deformed or damaged. For
replacing the filter packing, contact your Nihon Kohden representative.
1. Press the Strong clean key on the Operation screen to perform strong
cleaning to disinfect inside the analyzer. Refer to “Strong Cleaning” in
Section 5 of the operator’s manual.
2. Press the Drain baths key on the Operation screen to drain the measurement
baths and sub baths. Refer to “Draining Measurement Baths and Sub Baths”
in Section 5 of the operator’s manual.
Main power switch 3. Turn off the main power by pressing the main power switch on the rear panel
of the analyzer.
4. Remove the two screws on each side of the front cover of the hematology
analyzer.
5. Open the front cover by pulling it from the right side. Check that the tube
holder is closed before opening the front cover.
Filter joint
Filter joint
Filter 7. Remove the filter from each assembly. Use tweezers to remove any dust
from the filter. If it is still dirty, replace it with a new one.
Replace the filter packing when they are dirty, deformed or damaged. For
replacing the filter packing, contact your Nihon Kohden representative.
8. Reattach the filter joint assemblies to the bottom of the RBC measurement
Filter packing bath and air trap. Make sure that the tube with the same number as the
number label on the attaching part is connected back to the original position.
Only finger tighten the filter joint.
NOTE
• When attaching the filter joint assembly, be careful not to bend or
damage the filter packing at the bottom of the measurement bath.
• If there is leakage noted after installment of the filter, check that there
are no scratches or damage around the filter. Damage may occur if a
component is overtightened.
9. Reattach the front cover and fasten it with the two screws on each side of the
front cover.
10. Press the [Power] key to turn on the power. The analyzer starts priming the
fluid pathway.
11. If filters were replaced, reset the filter counter. Before resetting the counter,
the measurement baths and sub baths should be cleaned. Refer to the
“Checking and Cleaning Measurement Baths, Sub Baths and MC Tray”
section.
To reset the counter, check <Filters> and press the Reset key on the User
Maintenance screen to reset the counts to zero.
Check the measurement baths, sub baths and MC tray every day.
Clean the measurement baths, sub baths and MC tray when there is any blood or
dust on them. (Once a month or every 1,000 sample counts)
Materials Required
• Powder-free gloves, lab coat, safety glasses
• Phillips and flat-blade screwdrivers
• CLEANAC•3 detergent
• Dry lint-free cloth
Procedure
NOTE
• The sub bath and measurement bath are made of special plastic. When
cleaning the sub bath and measurement bath, do not damage the
surface (inside). After cleaning, do not touch the surface (inside) with
bare hands. It may be the cause of dirt.
• Do not use alcohol to clean the sub bath and measurement bath.
1. Press the Strong clean key on the Operation screen to perform strong
cleaning to disinfect inside the analyzer. Refer to “Strong Cleaning” in
Section 5.
2. Press the Drain baths key on the Operation screen to drain the measurement
baths and sub baths. Refer to “Draining Measurement Baths and Sub Baths”
in Section 5.
Main power switch 3. Turn off the main power by pressing the main power switch on the rear panel
of the analyzer.
4. Remove the two screws on each side of the front cover of the hematology
analyzer.
5. Open the front cover by pulling it from the right side. Check that the tube
holder is closed before opening the front cover.
6. Pull up the four tabs of the MC tray until they click and release the cable
clamp on the bottom of the tray.
Cable clamp
7. Remove the MC tray by pulling up.
8. Rinse the MC tray with water and wipe it with a dry lint-free cloth.
10. Loosen the screw on the HGB cover and remove the HGB cover.
HGB cover
11. Check the WBC and RBC measurement baths and sub baths. If there is any
blood or dust on them, remove and clean them taking the following steps.
Step 16 Step 12 12. Remove the tube joint connected to the WBC measurement
Sub baths WBC meas bath bath by turning the knurl joint.
13. Remove filter joints on the RBC and WBC measurement bath
assemblies by turning the tube connectors. The joint with filter
is connected to the RBC measurement bath. Remove the filter
packing together with the filter joints.
Tube joint Tube 14. Loosen the screws fastening the measurement baths. (The
screws cannot be removed from the measurement baths.)
Filter Filter joints 18. Rinse the measurement baths and sub baths with water and
wipe them with a dry lint-free cloth.
20. Reattach the measurement baths so that the sub bath is in the measurement
bath, the shaft of the sub bath is in the tab of the measurement bath, and the
round indent of the measurement bath fits the aperture.
21. Tighten the screws which were loosened in step 9 to fasten the measurement
baths.
NOTE
RBC measurement bath Before tightening the screws, check and remove any dirt or rust on and
around the screws. If dirt or rust is present, noise alarm may occur during
measurement.
22. Reconnect the filter joints to the RBC and WBC measurement bath
Filter packing
assemblies by turning the tube connectors. Attach the filter packing to the
RBC measurement bath.
NOTE
When connecting the filter joints to the RBC measurement bath, check
the following:
• The filter packing is not bent or damaged.
• The filter joint is fixed firmly.
• If there is leakage after installing the filter, check that there are no
scratches or damage around the filter and connect the joint again.
23. Reattach the tube joint to the WBC measurement bath by turning the knurl
joint.
24. Reattach the HGB cover and fasten it with three screws.
25. Put the MC tray so that the fixing tabs and holes of the clump match.
26. Push the fixing tab to fix the MC tray and fix the cable with the cable clamp
6
on the bottom of the tray.
Cable clamp
27. Reattach the front cover and fasten it with two screws on each side.
28. Press the [Power] key to turn on the power. The analyzer starts priming the
fluid pathway.
Check and clean the rinse unit and cap pierce nozzle once a month or every
1,000 sample counts whichever comes first.
WARNING
The cap pierce nozzle is sharp and potentially contaminated with
infectious materials. Be careful when handling the cap pierce nozzle
and performing this procedure.
Materials Required
• Powder-free gloves, lab coat, safety glasses
• Phillips and flat-blade screwdrivers
• Cotton swabs
• CLEANAC•3 detergent
• Lint-free pad
• New cap pierce nozzle (when required)
Procedure
1. Press the Strong clean key on the Operation screen to perform strong
cleaning to disinfect inside the analyzer. Refer to “Strong Cleaning” in
Section 5.
2. Press the Drain baths key on the Operation screen to drain the measurement
baths and sub baths. Refer to “Draining Measurement Baths and Sub Baths”
in Section 5.
3. Press the [Power] key while holding down the [Reset] key to turn the power
off. Check that the power lamp is off.
4. Remove the two screws on each side of the front cover of the hematology
analyzer.
5. Open the front cover by pulling it from the right side. Check that the tube
holder is closed before opening the front cover.
6. Check the following parts for dirt or blood clot. Remove blood or salt
crystals on the rinse unit and the tip of the cap pierce nozzle and sampling
nozzles with a cotton swab or lint-free pad moistened with CLEANAC•3
detergent.
Rinse unit
(bottom view)
Cap pierce nozzle
(bottom view with the tube
holder opened)
If the cap pierce nozzle is damaged or dirt or blood cannot be removed, replace
Cap pierce nozzle
the cap pierce nozzle with a new one.
v) Insert the cap pierce nozzle into the cap pierce nozzle guide and turn the cap
pierce nozzle 90º counterclockwise.
7. Turn the rinse unit cap counterclockwise to remove the rinse unit.
8. Loosen the joint assemblies to remove the tubes. Be careful not to lose the
O-ring from the rinse unit.
9. Insert a cotton swab into the rinse unit from the bottom and push out the
O-ring to remove it from the rinse unit.
10. Wipe the inside of the rinse unit and rinse unit cap with a cotton swab
moistened with CLEANAC•3 detergent. If they are still dirty, soak them in
CLEANAC•3 for about 10 minutes.
NOTE
Do not use alcohol to clean the rinse unit.
11. Rinse the rinse unit, rinse unit cap and O-ring with water and dry thoroughly
with a dry cloth.
12. Reattach the O-ring to the rinse unit and return the rinse unit and rinse unit
cap to the original position.
When replacing the O-ring, clean the rinse unit and rinse unit cap and attach
the new O-ring.
Joint Tube
18. Rinse the sample cup with water and dry thoroughly with a dry cloth.
19. Reattach the joint which was removed in step 14 to the bottom of
the sample cup. Attach the packing inside the sample cup.
Sample cup
NOTE
Screw
• Do not bend the packing inside the sample cup.
Filter • Attach the joint firmly.
packing • If there is a leak, check that the filter is not damaged or
cracked and attach the joint again.
Filter
Tube 20. Reattach the transparent joint.
Joint
21. Reattach the sample cup to the original position and fix the cup with screw.
23. Reattach the front cover and fasten it with two screws on each side.
24. Press the main power switch on the rear panel to turn on the main power.
25. Press the [Power] key to turn on the power. The analyzer starts priming the
fluid pathway.
26. If the rinse unit and cap pierce nozzle were checked and cleaned, the rinse
unit and cap pierce nozzle counter will have to be reset. To reset the counter,
check <Rinse unit> and/or <Cap Pierce> and press the Reset key on the User
Maintenance screen to reset the counts to zero.
Check and clean the sampling nozzles once every four months or every 3,000
sample counts whichever comes first.
When PLT background count increases or the sampling nozzle is bent, replace
the sampling nozzles with a new one.
6
WARNING
The sampling nozzles are sharp and potentially contaminated with
infectious materials. Be careful when handling the sampling nozzles
and performing this procedure.
Materials Required
• Powder-free gloves, lab coat, safety glasses
• Phillips and flat-blade screwdrivers
• Cotton swabs
• CLEANAC•3 detergent
• Lint-free pad
• New sampling nozzle(s) (when required)
Procedure
1. Press the Strong clean key on the Operation screen to perform strong
cleaning to disinfect inside the analyzer. Refer to “Strong Cleaning” in
Section 5.
2. Press the Drain baths key on the Operation screen to drain the measurement
baths and sub baths. Refer to “Draining Measurement Baths and Sub Baths”
in Section 5.
Main power switch 3. Turn off the main power by pressing the main power switch on the rear panel
of the analyzer.
4. Remove the two screws on each side of the front cover of the hematology
analyzer.
5. Open the front cover by pulling it from the right side. Check that the tube
holder is closed before opening the front cover.
6. Check the sampling nozzles for dirt or blood clot. Remove blood or salt
crystals on the tip of the sampling nozzles with a cotton swab or lint-free pad
moistened with CLEANAC•3 detergent.
NOTE
Be careful not to drop the screws into the analyzer.
7. Reattach the front cover and fasten it with the two screws on each side.
8. Press the main power switch on the rear panel to turn on the main power.
9. Press the [Power] key to turn on the power. The analyzer starts priming the
fluid pathway.
10. If the sampling nozzles were checked and cleaned, the sampling nozzle
counter will have to be reset. To reset the counter, check <Sampling nozzle>
and press the Reset key on the User Maintenance screen to reset the counts
to zero.
Check the pump tube for water droplets and leaks every day.
Replace the pump tube when there are water droplets or leaks. (Once every 4
months or every 3,000 sample counts whichever comes first.)
NOTE
Do not leave the pump tube with water droplets or leaks on it. 6
Materials Required
• Powder-free gloves, lab coat, safety glasses
• Phillips screwdriver
Procedure
1. Press the Strong clean key on the Operation screen to perform strong
cleaning to disinfect inside the analyzer. Refer to “Strong Cleaning” in
Section 5 of the operator’s manual.
2. Press the Drain baths key on the Operation screen to drain the measurement
baths and sub baths. Refer to “Draining Measurement Baths and Sub Baths”
Main power switch in Section 5 of the operator’s manual.
3. Turn off the main power by pressing the main power switch on the rear panel
of the analyzer.
4. Remove the two screws on each side of the front cover of the hematology
analyzer.
5. Open the front cover by pulling it from the right side. Check that the tube
holder is closed before opening the front cover.
6. Check the pump tube for water droplets and leaks. If any droplet or leak is
found, replace the tube with a new one by using the following steps.
Pump tube
Pump cover
Tube holder
8. Pull out the white tube joint from the tube holder and pull out the pump tube
Black by turning the pump rotator counterclockwise. Then pull the black tube joint
out of the tube holder.
White
Pump rotator
6.28 Service Manual MEK-7300K
6. MAINTENANCE
9. Remove the white and black tube joints and replace the pump tube.
Tube holder 10. Return the white tube joint to the original position and push the pump tube
into the tube guide by turning the rotator counterclockwise.
6
Pump rotator
White
Tube guide
Pump tube
Tube holder 11. Return the black tube joint to the original position.
White NOTE
Black
• Be careful not to pinch the new pump tube between the tube guide and
housing. This may damage the pump tube.
• Do not attach the black tube joint to the tube holder before the white
tube joint because internal compressed air may disconnect the tube.
• Put back the pump tube properly. If the pump tube has slack, remove
the slack by turning the rotator clockwise. If the pump tube has slack, it
will be damaged by the tube guide.
• Make sure the joints are held properly by the tube holder as shown
below. Otherwise, the pump tube may be damaged or the life of the
Pump rotator pump tube will be shortened.
13. Reattach the front cover and fasten it with the two screws on each side.
14. Press the main power switch on the rear panel to turn on the main power.
15. Press the [Power] key to turn on the power. The analyzer starts priming the
fluid pathway.
16. If the pump tube was replaced, the pump tube counter will have to be reset.
To reset the counter, check <Pump tube> and press the Reset key on the User
Maintenance screen to reset the counts to zero.
When the “CLOG” alarm occurs, remove the clog by the following procedure.
1. Press the Operation key on the screen to display the Operation screen.
2. Press the Remove clog key on the Operation screen. The confirmation
message appears.
3. Press the Yes key to remove the clog from the aperture. The analyzer starts
removing the clog and the “Removing clog” message appears on the screen.
Press the No key to cancel the procedure.
After removing the clog, the screen returns to the Operation screen.
Materials Required
• Powder-free gloves, lab coat, safety glasses
• Phillips and flat-blade screwdrivers
• Dry cloth or tissue paper
• CLEANAC•3 detergent
• Microscope
6
Procedures
For daily cleaning of the aperture caps, press the [ Clean] key on the front
panel.
NOTE
The aperture caps are behind the measurement baths.
1. Press the Strong clean key on the Operation screen to perform strong
cleaning to disinfect inside the analyzer. Refer to “Strong Cleaning” in
Section 5 of the operator’s manual.
2. Remove the diluent tube from the ISO3 inlet, the hemolysiing reagent tubes
from the HEMO3 and HEMO5 inlets, and the detergent tubes from the
CLN3 inlets on the right side panel.
NOTE
Waste comes out from the CLN inlet when Drain all is performed.
Waste
outlet
3. Press the Drain all key on the Operation screen.
Waste container
NOTE
Be sure all reagent has drained into the container. Failure to do so may
result in a liquid spill.
Main power switch 5. After draining, turn off the main power by pressing the main power switch
on the rear panel of the analyzer.
6. Remove the two screws on each side of the front cover of the hematology
analyzer.
7. Open the front cover by pulling it from the right side. Check that the tube
holder is closed before opening the front cover.
8. Pull up the four tabs of the MC tray until they click and release the cable
clamp on the bottom of the tray.
Cable clamp
11. Remove the screw on the HGB cover and remove the HGB cover.
HGB cover
WBC meas bath 12. Remove the tube joint connected to the WBC measurement
bath by turning the knurl joint.
13. Remove the filter joints on the RBC and WBC measurement
bath assemblies by turning the tube connectors.
Aperture cap
Steps 15,16 Joint
Filter
Step 14
Service Manual MEK-7300K 6.35
6. MAINTENANCE
16. Place a cloth or tissue paper under your hand and remove the aperture cap by
Aperture cap
pulling it toward you. If it is not easy to pull the aperture cap, move it slowly
left and right to remove it.
17. Carefully rinse the aperture cap. Remove all protein build-up, especially
from the inside. The condition of the aperture cap can be checked with a
100× microscope.
18. If a clog or dust still remains in the aperture caps, soak the aperture caps in
CLEANAC•3 detergent for about an hour.
CAUTION
Handle the aperture caps with care. They can be damaged if a sharp
object is used to clean them.
19. Rinse the aperture caps with water and replace them in
Aperture cap the original positions. Make sure that the black O-ring is
Concave facing the hole (analyzer side).
Analyzer NOTE
When replacing the aperture cap, do not push the
Do not push the center. aperture cap with your ball of finger. The aperture cap
may be broken.
Aperture cap
Measurement bath Sub bath 20. Reattach the measurement baths so that the sub bath is in the
measurement bath, the shaft of the sub bath is in the tab of the
measurement bath, and the round indent of the measurement
bath fits the aperture.
Screw
21. Tighten the screws which were loosened in step 15 to fasten
the measurement baths.
NOTE
Aperture cap Before tightening the screws, check and remove any
Step 21, 22 dirt or rust on and around the screws. If dirt or rust is 6
present, noise alarm may occur during measurement.
22. Reconnect the filter joints to the RBC and WBC measurement
bath assemblies by turning the tube connectors. Attach the
filter packing to the RBC measurement bath.
NOTE
• Do not bend the packing inside the sample cup.
WBC meas bath
• Attach the joint firmly.
RBC meas bath • If there is a leak, check that the filter is not damaged or
cracked and attach the joint again.
25. Put the MC tray so that the fixing tabs and holes of the clump
match.
Joint
26. Push the fixing tab to fix the MC tray and fix the cable with the
Filter cable clamp on the bottom of the tray.
Step 23, 24
Cable clamp
27. Reattach the front cover and fasten it with the two screws on
each side.
28. Press the [Power] key to turn on the power. The analyzer starts
priming the fluid pathway.
Calibrate the touch screen when the pressed position and operating position do
not match.
1. Hold down the [ Reset] key until the emergency stop message disappears.
Then press the [ Eject] key while holding down the [ Reset] key. The
Touch panel calibration screen appears.
NOTE
Do not use a sharp object to press the mark. Use your finger.
Customer:
Customer Address:
Service Personnel: Service Company:
Instrument Name: Hematology Analyzer Instrument Model: MEK-7300K
Instrument Serial Number: Hardware Revision:
Software Version: 6
Appearance
There is no damage on the hematology analyzer. Yes No
There is no fluid leakage on the hematology analyzer. Yes No
Safety
There is no damaged AC plug and exposed wire on the power cord. Yes No
3-pin plug type power cord is used and the 3 pins and plug housing are not deformed. Yes No
Resistance of the protective ground line of the power cord is 0.1 Ω or less. Yes No
Earth leakage current is 0.5 mA or less under normal condition. Yes ( mA) No
Earth leakage current is 1.0 mA or less under each single fault condition. Yes ( mA) No
Reagents
Nihon Kohden recommended diluent, detergent and hemolysing reagent are used. Yes No
ISOTONAC·3 diluent is not past the expiration date. Yes No
CLEANAC detergent is not past the expiration date. Yes No
CLEANAC·3 detergent is not past the expiration date. Yes No
HEMOLYNAC·3N hemolysing reagent is not past the expiration date. Yes No
HEMOLYNAC·5 hemolysing reagent is not past the expiration date. Yes No
Cleaning/Replacing
Rinse unit is cleaned. Yes No
Cap pierce needle is cleaned. Yes No
Four filters are replaced with new ones. Yes No
Two pump tubes are replaced with new ones. Yes No
Sampling cup (Bath for WBC 5 part differential measurement) is cleaned. Yes No
Two sub baths and two measurement baths are cleaned. Yes No
Two aperture caps are cleaned. Yes No
Touch screen is cleaned with a soft cloth moistened with 80% alcohol. Yes No
Two sampling nozzles are cleaned or replaced with new ones if they are damaged or
PLT background noise increases. Yes No
Software Version
Write down the software version on the User Maintenance screen:
Software version: ___________
Operations
Touch screen function is checked. Yes No
Date and time settings are checked. Yes No
Automatic cleaning operations are normal during this maintenance check. Yes No
There is no fluid leakage on the hematology analyzer. Yes No
Constant volume of the diluent is properly dispensed when pressing the
Dispense key on the screen in pre-dilution mode. Yes No
Cap pierce unit properly works and the Open mode or Closed mode is selected. Yes No
Bar Code Reader Operation (When the bar code reader is connected)
Light emitter and detector parts are cleaned with a cotton swab moistened with
80% alcohol. Yes No
Bar code reading operation properly works. Yes No
Others
Write down any other points on this sheet.
General................................................................................................................................................................. 7.2
Displaying the Service Screen.............................................................................................................................. 7.3
Sensor Monitor Screen......................................................................................................................................... 7.5
DP Calibration............................................................................................................................................ 7.5
MV Check Screen................................................................................................................................................. 7.6
Unit Check Screen................................................................................................................................................ 7.9 7
MC Unit Noise Check Screen..................................................................................................................... 7.9
MP-730V (Diluter) Check Screen............................................................................................................. 7.10
MP-730V (Sheath) Check Screen............................................................................................................ 7.10
MP-731V (Lyse Pump) Check Screen...................................................................................................... 7.11
Sample Pump (MP-731V) Check Screen . .............................................................................................. 7.11
Sub Bath Check Screen........................................................................................................................... 7.12
Sampler (MS-730V) Check screen........................................................................................................... 7.12
Cap Pierce (MS-731V) Check Screen...................................................................................................... 7.13
Continuous Measurement Screen...................................................................................................................... 7.14
SD Card Screen................................................................................................................................................. 7.15
Printer Settings Screen...................................................................................................................................... 7.16
Morpho. Flag Settings Screen............................................................................................................................ 7.17
Language Settings Screen................................................................................................................................. 7.19
Display Settings Screen..................................................................................................................................... 7.20
Advanced Settings Screen................................................................................................................................. 7.21
Sens/Threshold Screen...................................................................................................................................... 7.23
Instrument Settings Screen................................................................................................................................ 7.25
Serial Number Setting Screen............................................................................................................................ 7.26
XD-CV Screen.................................................................................................................................................... 7.27
X10-CV Screen................................................................................................................................................... 7.28
X31-CV Screen................................................................................................................................................... 7.29
Calibration Limit Setting...................................................................................................................................... 7.30
Unit Correction Coefficient Screen..................................................................................................................... 7.31
Initial Menu Screen............................................................................................................................................. 7.32
Dummy Data Screen.......................................................................................................................................... 7.33
Serial Setting Screen.......................................................................................................................................... 7.34
General
Provided on the Service screen has useful functions for servicing the MEK-7300.
These functions are not used by the user; therefore, they are not mentioned on
the Operator’s Manual.
They are intended for use by appropriately skilled service personnel. Improper
use can cause leaks inside the equipment.
The Service screen includes functions that can only be used with dedicated tools
at a factory. Without the specific tools, such functions do not operate. Do not use
these functions, which are not explained in this Service Manual.
Never use the functions without learning how to use them. There is a risk of
damaging the analyzer if the functions are used without proper knowledge.
1. To access the Service screen, the User setting must be set to [Factory].
i) Select [System] on the right side of the window and select [Users].
ii) Select [Factory] and press [OK].
iii) When you are prompted to enter the password, enter [4321] and press
the [ENTER] key.
iv) The bar at the top of the window turns orange when the operator
changes.
7
2. Select [System] on the right side of the window and select [Service].
Press the [DP Calib.] key at the bottom of the window to change sensor voltage
settings.
DP Calibration
This screen is for adjusting the offset voltages for the manometer sensors and
fluid sensors. Select the sensor for which to change the voltage setting and press
[Enter] after entering a value.
To check sensor voltages, press the [Sens.Moni.] key at the bottom right.
Increasing the setting value raises the sensor voltage while decreasing the setting
value lowers the sensor voltage (0 to 255).
WBC manometer sensor
(lower)
The voltage of the sensor is displayed under each item. Press this key to return to the
sensor monitoring screen.
MV Check Screen
CD Valve:
MC Valve: Solenoid valves in JQ-731V
Solenoid valves in MC-730V.
Involved with CBC
measurement Cap Valve:
Valves that control the rinse
part of the cap pierce unit (MS-
JQ Valve: 731V). Included in JQ-732V
Solenoid valves in unit
priming/draining
unit of JQ-730V
Sample Pump Valve:
Fluid control valves for
sample pump (MJ-730V)
Sheath Pump Valve:
Fluid control valves for sheath
pump (MJ-730V)
There is no limit to the number of solenoid valves you can turn on at the same
time on this screen.
To go to the next solenoid valve control screen, press the right arrow key at the
bottom of the window.
24 25
MJ-730V UT-7236 Sensor:2
JQ-730V 9 7
4
1 1
ISOTONAC.3
Thermistor 2 2
1 <10> 10 CLEANAC
3 3
MC-730V
<6>
<8> 21 Sample pump
30 valves 6 Sensor:1
11
HEMOLYNAC.3N
12
HEMOLYNAC.5
<5> <7>
4 1 Sensor:3
5 2 27
R W Sensor:4
2 3 4
17
Filter 1 9 MS-731V 1 12 13 22
13
WASTE
6 3 Warmer
8 10
26
5 7 8 9
28
2 Filter
3 7
8 7
11
Filter
4 23
JQ valves
14 4
31
2 5 3 4
13 14
15 6
Case 5 6
9
MO-820V 1 2
5 19 Pressure
Sensor 1 2
13 12
18 19 20 18
9
CD valves
Manifold
16 17 18
12 10
7 ISO3 HEMO3N HEMO5
11
JQ-731V JQ-732V 15 8
MP-731V
10 2
1 1
11
1 2 3 4 5 6
29 3
3
<11>
MP-640V MP-640V 14 15 16 17
1
20
1 2
16
<4>
21 22 23 4
24
<1>
<2> <3> MP-730V
Cap valves
Front Side
Diluter Valve:
Sheath Unit Valve: (XP-513V)
Valves in MJ-730V. Diluter solenoid valve in
Involved with MP-730V.
5-part differential
measurement. Lyse Valve:
(XP-513V)
Sampler Valve: Valves that control the
(XP-513V) hemolyzing reagent pump
Valve that switches of the triple pump unit
between WBC and RBC (MP-731V).
nozzles in MS-730V.
Pinch Valve:
(Pinch valves) 7
Blood sample
control valves for
5-part differential
measurement.
In addition to XP-612V, XP-513V and pinch valves are controlled on this screen.
Up to five solenoid valves (XP-513V and pinch valves) can be turned on at the
same time on this screen. If more than five solenoid valves are turned on, it may
damage the power supply unit.
Pressing the left arrow key at the bottom of the window returns to the previous
solenoid valve control screen.
MP-640V MP-640V 1
14 15 16 17 20
1 2 Diluter valve
16
<4>
21 22 23 4
24
<1>
<2> <3> MP-730V
Front Side
Rotary pumps have two operating directions: forward and reverse. (The ■ button
stops the pump.)
Operating the pumps without taking into consideration the flow of fluid can
cause leaks, as well as damage to tubes and solenoid valves.
This screen allows individual control of units. It is useful for performing checks
after replacing a unit.
If operations have been stopped using the RESET key, press [All Initial] to
initialize all units before further maintenance.
Do not use
Do not use
If the analyzer is started by pressing RESET and Power ON, the [A: 068 HGB
INITIALIZE ERROR] error message appears. This error can be canceled by
pressing the [HGB Blank] key when the WBC measurement bath is filled with
diluent.
Do not use
Do not use
MP-730V initialization
Do not use
This screen checks the operation of the MP-730V sheath syringe. Always press
[Initial] before using this function.
Press [Aspirate] to draw in diluent from the diluent port. Press [Dispense] to
release the diluent to the waste chamber.
Do not use
MP-731V initialization
Do not use
This screen checks the operation of the MP-731V five-part differentiation sample
syringe. Press [Aspirate] to draw in diluent, and press [Dispense] to release the
diluent.
MS-730V initialization
The nozzle will move; take due care to prevent it from bending or breaking.
Do not use
MS-731V initialization
Do not use
7
The sampling needle will move; take due care to prevent accidental puncturing.
Open
Performs continuous measurement
in open mode
Close
Performs continuous measurement
in closed mode
Run continuous measurement (aging) only after checking that there are sufficient
amounts of the reagents and that there is no accumulated waste fluid.
SD Card Screen
NOTE
• Only the current settings can be saved and multiple conditions cannot
be saved.
• Data is deleted upon restoration.
Save button:
Outputs operation logs stored in the analyzer to
the SD card.
Dump button:
Display operation logs.
IG% and IG# cannot be printed from an external printer or card printer.
Threshold values can be set to a value in the range of 1 to 5, with 5 causing the
easiest trigger for flags to be set.
On the STEP screen, detailed settings are available for flag thresholds.
NOTE
The initial values to be applied when settings are reset depends on the
language set here.
Japanese:
Unit system that is used in Japan.
Date/Time: YY/MM/DD
English:
Unit system that is used in the U.S.A.
Date/Time: DD/MMM/YY
The screen sets the display of scattergrams and histograms on the result screen.
Sets how the histogram is displayed
Histogram
Max value of the vertical axis is fixed. The
higher the cell count, the taller the height of
the histogram becomes.
Distribution
The vertical direction is fixed to full scale
regardless of the overall count values. This
makes it easy to read the distribution pattern,
but the overall count values are not shown.
The ratio of the vertical axis of the histogram and the position of the scale of
histogram for RBC can be adjusted.
Use this function when the position of the RBC histogram scale does not match
the MCV.
This screen changes various settings for measurement. For normal use, use the
default settings.
Selects whether to use MEK-CAL or
human blood for optical adjustment.
(Default: MEK-CAL)
Fan status
ON: Fan is running
OFF: Fan is stopped
(Default: OFF)
Highland mode
Turn this mode on when using the
analyzer in a location more than 1,000
m above sea level. (Default: OFF)
Settings for CBC and optical sensitivity are different for control blood and
human blood.
Sens/Threshold Screen
This screen sets the impedance sensitivity and thresholds for human blood
measurement.
The threshold settings for RBC and PLT are generally set to [Auto], but when
using a fixed threshold, remove the check from the AUTO checkbox and enter a
value.
You can check the history of changes applied to the sensitivity and threshold
settings.
You can set the CBC sensitivity and thresholds for MEK-5DL/N/H and the
offsets for optical sensitivity.
This screen sets individual default values for the analyzer and checks the average
and CV.
• Serial Number
• DP Calib.
• Unit Calib.
• XD-CV 7
• X10-CV
• X31-CV
• CAL Limit Set
The measurement data includes a serial number. Always compare it with the
serial number on the analyzer’s label and confirm that they match.
If the MAIN BD (UT-7253) is replaced, enter the new serial number on this
screen.
XD-CV Screen
Based on all stored normal value data, this screen displays the daily average and
center values for the past three days.
Use as a way to acquire the center values (MCV, MCH, MCHC) for X̅B
management
The periodic average of X̅D data is used as the center value for each item and as
the initial value for X̅B management.
Samples with measured values that are out of pre-set range are not included in
the calculation.
X10-CV Screen
Based on all stored normal value data, this screen displays the average and center
values for every 10 samples.
Use this screen when continuously measuring hematology control blood and
checking the reproducibility of the machine.
The screen displays the average and center values for 3 sets of 10 counts, in other
words the average and center values from 30 samples, as well as the average
mean and CV mean for every 10 counts.
The last 30 samples are used for calculation regardless of the standard value
range.
X31-CV Screen
Based on all stored normal value data, this screen displays the average and center
values for the last 31 samples measured.
If there are more than 30 measurement data sets, the last 31 data sets are used to
calculate the average and CV%.
The last 31 samples are used for calculation regardless of the standard value
range.
7
X31-CV data is not calculated when there are less than 31 data sets.
This screen sets the correction coefficients for the dilution ratio and the
aspiration volume in the manometer.
If the correct correction coefficients are not set, accurate calculation and
measurement cannot be ensured.
This screen allows for the creation of the following demonstration data.
This function creates a DUMMY_DATA (1-byte, all capital letters) folder on the
SD card for MEK-7300. Data saved in the folder can be read into the analyzer.
Do not use
• Calibration data
• QC data
When demonstration data is created, the relevant stored data will be deleted.
After demonstration, delete all data via the [Initial] screen. Performing
measurements with the demonstration data existing can cause problems, such as
not being able to save data.
Sales
USA Asia
NIHON KOHDEN AMERICA, INC. SHANGHAI KOHDEN MEDICAL ELECTRONIC
90 Icon Street, Foothill Ranch, CA 92610, USA INSTRUMENT CORPORATION
Phone +1 (949) 580-1555 7th Floor, Dawning Centre Tower A No.500
Fax +1 (949) 580-1550 Hongbaoshi Road, Changning District
Shanghai 201103, China
Europe Phone +86 (21) 6270-0909
Fax +86 (21) 6270-9700
European Representative
NIHON KOHDEN EUROPE GmbH Beijing Branch
Room 1701, East Ocean Centre
Raiffeisenstrasse 10
No. 24A JianGuoMenWai Street, Beijing, 100004
D-61191 Rosbach v.d.H., Germany
Phone 010-6515-5750
Phone +49 6003 827-0 Fax 010-6515-5758
Fax +49 6003 827-599
Guangzhou Branch
NIHON KOHDEN ITALIA S.r.l. Room 2514, Yian Plaza
Via Fratelli Bronzetti 28 No. 33 Jian She Liu Ma Road, Guangzhou, 510060
I-24124 Bergamo, Italy Phone 020-8363-3737
Phone +39 035 219 543 Fax 020-8363-3807
Fax +39 035 232 546
NIHON KOHDEN SINGAPORE PTE LTD
NIHON KOHDEN FRANCE SARL 1 Maritime Square, #10-34 (Lobby C), Harbour Front Centre
8, rue Francois Delage, 94230 Cachan, France Singapore 099253
Phone +33 1 49 08 05 50 Phone +65 6376-2210
Fax +33 1 49 08 93 32 Fax +65 6376-2264
Contact information is accurate as of May 2013. Visit www.nihonkohden.com for the latest information.
The model and serial number of your instrument are identified on the rear or bottom of the unit.
Write the model and serial number in the spaces provided below. Whenever you call your representative concerning
this instrument, mention these two pieces of information for quick and accurate service.
Your Representative