Patient Monitoring Solutions

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 24

Patient Monitoring Solutions

www.axcentmedical.com
PAVO Vital Sign Monitor
Features
• 8” color TFT LCD Screen (touch screen optional)
• Portable, lightweight and sturdy design
• Flexible parameters configuration for different clinical environments
• Rechargeable Li-ion Battery (up to 12 hours uninterruptable work)
• Big font and font color display setting
• Spot-check and continuous monitoring mode
• Selectable for Adult, Pediatric and Neonatal patients
• Wired/Wireless CMS, support HL7 protocol to HIS
• Barcode scanner support
• Thermal recorder support
• Graphical & tabular trend review
• 48 hours holographic wave review for each patient (stored in SD card)

www.axcentmedical.com
For Out-Patient Department, Spot-check, Transport,
Ward and other Basic Monitoring.

Configuration Optional

Masimo/Nellcor SpO2, Quick Temp,


SpO2 + NIBP, Li-ion battery
Barcode scanner

SpO2+NIBP+ECG+TEMP, Masimo/Nellcor SpO2, EtCO2, Quick Temp,


Li-ion battery Barcode scanner, Thermal Recorder

Technical Specifications

Display RESP
8” color TFT LCD Screen, resolution: 800 x 600 Measurement method: Thoracic electrical
bioimpedance
ECG Rate: 0 – 150 bpm
Lead type Measuring lead: Lead I, II
3-lead: I, II, III Wave gain: ×0.25, ×0.5, ×1, ×2
5-lead: I, II, III, aVR, aVL, aVF, V Respiratory impedance range: 0.5-5 Ω
Display sensitivity: Baseline impedance: 500-4000 Ω
2.5 mm/mV (×0.25), 5 mm/mV (×0.5), Gain: 10 grades
10 mm/mV (×1.0), 20 mm/mV (×2.0) Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
Wave sweep speed: 6.25 mm/s, 12.5 mm/s,
25 mm/s, 50 mm/s TEMP
Bandwidth Measurement method: Thermistor
Diagnostic mode: 0.05Hz~100Hz Measuring range: 5~50 °C (41~122 °F)
Monitor mode: 0.5Hz~40Hz Resolution: 0.1 °C
Surgery mode: 1Hz~20Hz Measurement accuracy: ±0.1 °C
Strong filter mode: 5Hz~20Hz CMRR
>100dB Recorder (optional)
Notch: 50/60 Hz notch filter can be set to Built-in, Thermal dot array
on or off Horizontal resolution: 16 dots/mm (25 mm/s
Differential input impedance >5 MΩ paper speed)
Electrode polarization voltage range: ±400 mV Vertical resolution: 8 dots/mm
Baseline recovery time <3 s after defibrillation Paper speed: 25 mm/s, 50 mm/s
(in monitor and surgery mode) Number of waveform channels: 3
Calibration signal: 1 mV (peak - peak),
accuracy ±3%
PAVO Vital Sign Monitor
Technical Specifications

NIBP Interval: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120,


Measurement method: Automatic 180, 240, 480 minutes
oscillometric method Overpressure protection: Software and
Operating mode: Manual, automatic, hardware, double safety protection
continuous Cuff pressure range: 0-280 mmHg
Measurement unit: mmHg/kPa selectable
Typical measurement time: 20~40 s Standard SpO2
Measurement type: Systolic, Diastolic, Mean Measurement range: 0-100%
Measurement range (mmHg) Resolution: 1%
Range of Systolic pressure: Adult 40-270 Accuracy: ±2% (70-100%, Adult/Pediatric);
Pediatric 40-230 ±3% (70-100%, Neonate);
Neonatal 40-135 0-69%, unspecified
Range of Diastolic pressure: Adult 10-210 Refreshing Rate: 1s
Pediatric 10-150
Masimo SpO2 (optional)
Neonatal 10-100
Measurement range: 0-100%
Range of Mean pressure: Adult 20-230
Resolution: 1%
Pediatric 20-165
Accuracy: ±2% (70-100%, Adult/Pediatric),
Neonatal 20-110
non-motion, low
Measurement accuracy
±3% (70-100%, Neonate,
Maximum average error: ±5 mmHg
non-motion);
Maximum standard deviation: 8 mmHg
±3% (70-100%, motion);
Resolution: 1 mmHg
0-69%, unspecified
Refreshing Rate: 1s

Portable Design Touch Screen Quick Temp


(Optional) (Infrared Ear Thermometer)

www.axcentmedical.com
Infrared Ear Thermometer (optional) Phasein IRMA™ Sidestream CO2 (optional)
Displayed range: 34~42.2 °C (93.2~108 F°) Warm-up time: Full accuracy within
Operation ambient temperature range: 10 seconds
10~40 °C (50~104 °F) Sampling flow rate: 50 ml/min (+/-10/min)
Accuracy for displayed temperature range: Accuracy: ± (0.2% +2% of the reading)
≥35 °C (95.9 °F) ~ ≤42.2 °C (107.6 °F) range Measurement Range: 0 -15%
±0.2 °C (0.4 °F) Rise time: 200 ms, typical at 50 ml/min
<35 °C (95.9 °F) ~ ≥34 °C (93.2 °F) range flow rate
±0.3 °C (0.5 °F) Total response time: within 3 seconds
(with 2m Momoline sampling line)
AWRR Range: 0-150 bpm
AWRR Accuracy: ±1 breath

Phasein IRMA™ Mainstream CO2 (optional)


Measurement Range: 0-15%
Warm-up time: Full accuracy within 10 seconds
Accuracy: ± (0.2% +2% of the reading)
AWRR Range: 0-150 bpm
AWRR Accuracy: ±1 breath

Operation Environment
Power: AC 100-250 V, 50/60 Hz
Temperature: 0-40 °C
Humidity: 15-85%
Patient Range: Adult, Pediatric, Neonate
Battery backup: Standard 4-5 hrs (2.600 mAh),
optional 8-10 hrs (5.200 mAh) or 12-15 hrs
(7.800 mAh)
www.axcentmedical.com
CETUS x12 Patient Monitor
Features
• 12.1” color TFT LCD screen
• 8 waveform display, up to 12-lead ECG analysis
• Useful calculation
(Hemodynamic, Drug Dose, Oxygenation, Ventilation)
• Pacemaker detection
• ST & arrhythmia analysis
• OxyCRGs screen
• Wired/Wireless CMS, support HL7 protocol to HIS
• SpO2 pulse-tone modulation (Pitch Tone)
• MEWS (Modified Early Warning Score)
• Graphical & tabular trend review (120 hours)
• Rechargeable Lithium-Ion Battery (2600 mAh)
12.1”color TFT LCD screen, wide and flat screen design,ecnomic and reliable

Configuration: ECG+SpO2+NIBP+2TEMP+PR+RESP, Li-ion battery


Optional: Touch-Screen, 12-lead ECG, Masimo SpO2, 2/4/6 IBP, C.O., EtCO2, Multi-Gas, BIS, NMT,
VGA, Thermal Recorder, Wired/Wireless CMS

Technical Specifications

Display RESP
12.1” TFT (touch screen optional) Measurement method: Thoracic electrical
Resolution: 800 x 600 bioimpedance
Number of traces: 8 waveforms Rate: 0 – 150 bpm
Measuring lead: Lead I, II
ECG Wave gain: ×0.25, ×0.5, ×1, ×2
Lead type: 3-lead, 5-lead, 12-lead Respiratory impedance range: 0.5-5Ω
ECG waveform: 2 channels, 7 channels, Baseline impedance: 500-4000Ω
12 channels Gain: 10 grades
Display sensitivity: 2.5 mm/mV (×0.25), Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
5 mm/mV (×0.5), 10 mm/mV (×1.0),
20 mm/mV (×2.0) Pulse Rate
Wave sweep speed: 6.25 mm/s, 12.5 mm/s, Range: 30~254 bpm
25 mm/s, 50 mm/s Resolution: 1bpm
Bandwidth Accuracy: ±2bpm (non-motion)
Diagnostic mode: 0.05 Hz~100 Hz ±5bpm (motion)
Monitor mode: 0.5 Hz~40 Hz Refreshing rate: 1s
Surgery mode: 1Hz~20Hz
Strong filter mode: 5 Hz~20 Hz TEMP
CMRR >100 dB Accuracy: ±0.1 °C or ±0.2 °C °F (without probe)
Notch: 50/60 Hz notch filter can be set to Measurement range: 5~50 °C (41~122 °F)
on or off Channel: Two channels
Differential input impedance >5MΩ Resolution: 0.1 °C
Electrode polarization voltage range: ±400 mV Parameters: T1,T2 and TD
Baseline recovery time <3s after defibrillation
(in monitor and surgery mode)
Calibration signal: 1mV (peak - peak),
accuracy ±3%

7-lead ECG

www.axcentmedical.com
CETUS x12 Patient Monitor
Technical Specifications

NIBP SpO2
Measurement method: Automatic oscillome- Measurement range: 0-100%
tric method Resolution: 1%
Operating mode: Manual, automatic, Accuracy: ±2% (70-100%, Adult/Pediatric);
continuous ±3% (70-100%, Neonate);
Measurement unit: mmHg/kPa selectable 0-69%, unspecified
Typical measurement time: 20~40 s Refreshing Rate: 1s
Measurement type: Systolic, Diastolic, Mean
Measurement range (mmHg) Masimo SET® SpO2 (Optional)
Range of Systolic pressure: Adult 40-270 Measurement range: 0-100%
Pediatric 40-200 Resolution: 1%
Neonatal 40-135 Accuracy: ±2% (70-100%, Adult/Pediatric,
Range of Diastolic pressure: Adult 10-210 non-motion, low prefusion);
Pediatric 10-150 ±3% (70-100%, Neonate,
Neonatal 10-95 non-motion);
Range of Mean pressure: Adult 20-230 ±3% (70-100%, motion);
Pediatric 20-165 0-69%, unspecified
Neonatal 20-105 Refreshing Rate: 1s
Measurement accuracy
Recorder (Optional)
Maximum average error: ±5 mmHg
Built-in, Thermal dot array
Maximum standard deviation: 8 mmHg
Horizontal resolution: 16 dots/mm (25 mm/s
Resolution: 1 mmHg
paper speed)
Interval: 1, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180,
Vertical resolution: 8 dots/mm
240, 480 minutes
Paper speed: 25 mm/s, 50 mm/s
Overpressure protection: Software and
Number of waveform channels: 3
hardware, double safety protection
Cuff pressure range: 0-280 mmHg
Operation Environment
Power: AC 100-250V, 50/60Hz
Temperature: 5-40 °C
Humidity: <80%
Patient Range: Adult, Pediatric, Neonate
Battery backup: Standard 2-3 hrs (2.600 mAh),
optional 3-5 hrs (4.800 mAh)
www.axcentmedical.com
CETUS x15
Critical Care Patient Monitor
Features
• 15.6” High resolution TFT LCD Touch screen
• 10 waveform display, up to 12-lead ECG analysis
• Useful calculation (Hemodynamic, Drug Dose, Oxygenation, Ventilation)
• Pacemaker detection
• ST & arrhythmia analysis
• SpO2 support PVI and PI, low perfusion 0.2%
• Aspect BISx module, NMT module optional
• Wired/Wireless CMS, support HL7 protocol to HIS
• SpO2 pulse-tone modulation (Pitch Tone)
• VGA support external display
• Graphical & tabular trend review (120 hours)
• 48 hours full disclosure wave review for each patient
CETUS x15 Critical Care Patient Monitor
Multiple parameter options satisfy the need for ICU, CCU, NICU.

Configuration: ECG, SpO2, NIBP, TEMP, Resp, PR; Li-ion battery


Optional: Touch-Screen, 12-lead ECG, Masimo SpO2, 2/4/6 IBP, C.O., EtCO2, Multi-Gas, BIS, NMT,
VGA, Thermal Recorder, Wired/Wireless CMS

Masimo SET® SpO2 Bispectrial Index™ by Aspect


Provides anti-motion and anti-low Monitor the level of consciousness
perfusion SpO2 measurement. of the patient under general anesthesia
or sedation. provides BIS, SQI, EMG, SR,
SEF, TP, PC value and EEG wave.

Masimo Phasein IRMATM/ISA NMT


Sidestream/Mainstream EtCO2 Intergrade Organon TOF-Watch® SX
Allows selection of the modality best
suited to the application, monitoring
with infrared absorption technique.

IBP C.O.
2-4 Channel, support IBP waveform Cardiac Output
overlapping display

Technical Specifications

Display ECG
15.6” TFT (touch screen optional) Lead type: 3-lead, 5-lead,12-lead
Resolution: 1366 x 768 ECG waveform: 2 channels, 7 channels,
Number of traces: 10 waveforms 12 channels
Display sensitivity: 2.5 mm/mV (×0.25),
I/O 5 mm/mV (×0.5), 10 mm/mV (×1.0),
LAN: 1 standard RJ45 port 20 mm/mV (×2.0)
WLAN: IEEE 802.11b/g/n Wave sweep speed: 6.25 mm/s, 12.5 mm/s,
USB: 2 USB connectors 25 mm/s, 50 mm/s
SD: 1 SD card socket Bandwidth
VGA: 1 VGA monitor connector Diagnostic mode: 0.05 Hz~100 Hz
Output: 1 connector for Nurse call, Monitor mode: 0.5 Hz~40 Hz
Defib Sync Analog Output Surgery mode: 1 Hz~20 Hz
Strong filter mode: 5Hz~20 Hz
CMRR>100 dB

www.axcentmedical.com
Technical Specifications

Notch: 50/60 Hz notch filter can be set Masimo SET® SpO2(Optional)


to on or off Measurement range: 0-100%
Differential input impedance >5MΩ Resolution: 1%
Electrode polarization voltage range: ±400mV Accuracy: ±2% (70-100%, Adult/Pediatric,
Baseline recovery time <3s after defibrillation non-motion, low prefusion);
(in monitor and surgery mode) ±3% (70-100%, Neonate,
Calibration signal: 1 mV (peak - peak), non-motion); 
accuracy ±3% ±3% (70-100%, motion); 0-69%,
unspecified
RESP Refreshing Rate: 1s
Measurement method: Thoracic electrical
bioimpedance Pulse Rate
Rate: 0 – 150 bpm Range: 30~254 bpm
Measuring lead: Lead I, II Resolution: 1 bpm
Wave gain: ×0.25, ×0.5, ×1, ×2 Accuracy: ±2bpm (non-motion)
Respiratory impedance range: 0.5-5 Ω ±5bpm (motion)
Baseline impedance: 500-4000 Ω Refreshing rate: 1s
Gain: 10 grades
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s NIBP
Measurement method: Automatic
TEMP oscillometric method
Accuracy: ±0.1 or ±0.2 °F (without probe) Operating mode: Manual, automatic,
Measurement range: 5~50℃ (41~122 °F) continuous
Channel: Two channels Measurement unit: mmHg/kPa selectable
Resolution: 0.1℃ Typical measurement time: 20~40 s
Parameters::T1, T2 and TD Measurement type: Systolic, Diastolic,
Mean Measurement range (mmHg)
SpO2 Range of Systolic pressure: Adult 40-270
Measurement range: 0-100% Pediatric 40-200
Resolution: 1% Neonatal 40-135
Accuracy: ±2% (70-100%, Adult/Pediatric); Range of Diastolic pressure: Adult 10-210
±3% (70-100%, Neonate); Pediatric 10-150
0-69%, unspecified Neonatal 10-95
Refreshing Rate: 1s
CETUS x15 Critical Care Patient Monitor
Technical Specifications

Range of Mean pressure: Adult 20-230 Standard Mainstream CO2 (Optional)


Pediatric 20-165 Measurement range: 0-19.7%, 150 mmHg,
Neonatal 20-105 or 0-20 kPa
Measurement accuracy Resolution: 0.1 mmHg
Maximum average error: ±5 mmHg Measurement accuracy
Maximum standard deviation: 8 mmHg 0-40 mmHg: ± 2 mmHg
Resolution: 1 mmHg 41-70 mmHg: ± 5% of reading
Interval: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 71-100 mmHg: ± 8% of reading
180, 240, 480 minutes 101-150 mmHg: ± 10% of reading
Overpressure protection: Software and Respiration rate: 3-150 bpm
hardware, double safety protection Respiration rate accuracy: 1 ±1bpm
Cuff pressure range: 0-280 mmHg Warm-up time: 97% within 8 s,
full accuracy within 20 s
IBP (Optional)
Channel: 2, 4 or 6-channel Standard Sidestream CO2 (Optional)
ART: 0 to 300 mmHg Measurement rage: 0-20% (0-150 mmHg)
PA: -6 to 120 mmHg Accuracy: < 5.0% CO 2: ± 2 mmHg
CVP/RAP/LAP/ICP: -10 to 40 mmHg > 5.0% CO 2: < 6% of reading
Measurement range: P1/P2 -50 to 300 mmHg Respiration rate: 2~150 BPM
Resolution: 1mmHg Respiration rate accuracy: 1% ±1BPM
Accuracy: ±2% or ±1mmHg, Warm-up time: 97% within 45 s, full accuracy
whichever is greater (without sensor) within 10 min.
Sensitivity: 5uV/mmHg/V Rise times (t 10-90%): About 100 ms,
Impedance range: 300 to 3000 Ω when flow is 100 ml/min, adult water trap,
1.5 m sampling tube
C.O. (Optional) Delay time: <3 sec when flow is 100 ml/min,
Method: Thermodilution adult water trap, 1.5 m sampling tube
Range: C.O.: 0.2 to 20 L/min
TB: 23 to 45 ℃ Recorder (Optional)
T1: -1 to 27℃ Built-in, Thermal dot array
Accuracy: C.O.: ±5% or ±0.1L/min, Horizontal resolution: 16 dots/mm
whichever is greater (25 mm/s paper speed)
TB, T1:±0.5℃ (without sensor) Vertical resolution: 8 dots/mm
Paper speed: 25 mm/s, 50 mm/s
Number of waveform channels: 3

www.axcentmedical.com
12-lead ECG 4 channel IBP

OxyCRG screen Dynamic trends

Phasein ISA Sidestream CO2 (Optional) Phasein IRMA™ AX+ Mainstream Multi-gas
Warm-up time: Full accuracy within (Optional)
10 seconds Gas: CO2, N2O, HAL, ISO, ENF, SEV, DES with
Sampling flow rate: 50ml/min(+/-10/min) automatic identification
Measurement Range: 0-25% Warm-up time: Full accuracy within
Accuracy: 0~15% (±0.2% of the reading) 20 seconds for IRMA
15~25%, unspecified AX+ CO2 Accuracy:
Rise time: 200 ms, typical at 50 ml/min 0-10%: ± (0.2%+2% of the reading)
flow rate 0-15%: ± (0.3%+2% of the reading)
Total response time: within 3 seconds N2O Accuracy:
(with 2 m Nomoline sampling line) 0-100%: ± (2%+2% of the reading)
AWRR Range: 0-150 bpm HAL, ISO, ENF:
AWRR Accuracy: ±1 breath 0-8%: ± (0.15%+5% of the reading)
SEV:0-10%: ± (0.15%+5% of the reading)
Phasein IRMA™ Mainstream CO2 (Optional) DES:0-22%: ± (0.15%+5% of the reading)
Measurement Range: 0-25% Agent identification time: <20 s (typical <10 s)
Accuracy: 0~15% (±0.2% of the reading) AWRR range: 0-150 bpm
15~25%, unspecified AWRR accuracy: +/-1bpm
Warm-up time: Full accuracy within ‚ Apnea time: 20~60 s
10 seconds
AWRR Range: 0-150 bpm
AWRR Accuracy: ±1 breath
CETUS x15 Critical Care Patient Monitor
Technical Specifications

Aspect BISx module (Optional) Acceleration transducer: Accuracy ±5% of full


Parameter Measurement: scale value
BC: 0~30 (Only limited to the combined use Temperature sensor: Range 20.0-41.5°C
of an external sensor with a BIS module) (accuracy ±5 °C)
EMG: 30~55 dB (bar chart) with intensity
between 30 dB and 80 dB (tendency chart) Operation Environment
BIS: 0~100 Power: AC 100-250 V, 50/60 Hz
SQI: 0%~100% Temperature: 5-40 °C
SR: 0%~100% Humidity: <80%
SEF: 0.5 Hz~30 Hz Patient Range: Adult, Pediatric, Neonate
TP: 40~100 Db Battery backup: Standard 2-3 hrs (2.600 mAh),
EEG Measurement: optional 3-5 hrs (4.800 mAh)
Input impedance >5 MΩ
Noise (RTI) <2 μV (0.25~50 Hz)
Input signal range: ±1 Mv
EEG bandwidth between: 0.25 Hz~110 Hz

NMT Tof-Watch® SX (Optional)


Microprocessor-controlled
Stimulation Mode: TOF, TOFS, PTC,
1 Hz Twitch, 0.1 Hz Twitch, DBS DBS3.3 and
3.2 (Double Burst), Tetanic Stimulation (Burst),
5s – 50 Hz or 100 Hz
Output (accuracy ±5% of full scale value)
Surface electrodes:
Constant current, 0-60 mA (0-12/18μC)
up to 5 KOhm.
Monophasic, 200 μs or 300 μs pulse width
Needle electrodes:
Constant current,0-6 mA (0-0.24 μC)
up to 5 KOhm.
Monophasic, 40 μs pulse width

www.axcentmedical.com
CETUS xl Advanced Patient Monitor
Features
• 15.6”/17/19” switchable TFT LCD Touch Screen
• Aluminium material shell
• Fanless design suitable for quite care environment
• 10 waveform display, up to 12-lead ECG analysis
• Useful calculation (Hemodynamic, Drug Dose, Oxygenation, Ventilation)
• SpO2 support PVI and PI, low perfusion 0.2%
• Aspect BISx module, NMT module optional
• Wired/Wireless CMS, support HL7 protocol to HIS
• SpO2 pulse-tone modulation (Pitch Tone)
• VGA support external display
• Graphical & tabular trend review (120 hours)
• 48 h full disclosure wave review for each patient

www.axcentmedical.com
CETUS xl Advanced Patient Monitor
Multiple parameter options satisfy the needs of ICU, CCU, NICU

Configuration: ECG, SpO2, NIBP, Resp, PR; Li-ion battery


Optional: 12-lead ECG, Masimo SpO2, 2/4/6 IBP, C.O., EtCO2, Multi-gas, BIS, NMT;
VGA, Thermal Recorder, Wired/Wireless CMS

Masimo SET® SpO2 Bispectrial Index™ by Aspect


Provides anti-motion and anti-low Monitor the level of consciousness
perfusion SpO2 measurement. of the patient under general anesthesia
or sedation. provides BIS, SQI, EMG, SR,
SEF, TP, PC value and EEG wave.

Masimo Phasein IRMA™/ISA NMT


Sidestream/Mainstream EtCO2 Intergrade Organon TOF-Watch® SX
Allows selection of the modality best
suited to the application, monitoring
with infrared absorption technique.

IBP C.O.
2-4 Channel, support IBP waveform Cardiac Output
overlapping display

Technical Specifications

Display ECG
15.6” TFT Touch screen Lead type: 3-lead, 5-lead,12-lead
Resolution: 1366 x 768 ECG waveform: 2 channels, 7 channels,
Number of traces: 10 waveforms 12 channels
Display sensitivity: 2.5 mm/mV (×0.25),
I/O 5 mm/mV (×0.5), 10 mm/mV (×1.0),
LAN: 1 standard RJ45 port 20 mm/mV (×2.0)
WLAN: IEEE 802.11b/g/n Wave sweep speed: 6.25mm/s, 12.5 mm/s,
USB: 2 USB connectors 25 mm/s, 50 mm/s
SD: 1 SD card socket Bandwidth
VGA: 1 VGA monitor connector Diagnostic mode: 0.05 Hz~100 Hz
Output: 1 connector for Nurse call, Monitor mode: 0.5 Hz~40 Hz
Defib Sync Analog Output Surgery mode: 1 Hz~20 Hz
Strong filter mode: 5Hz~20 Hz
CETUS xl Advanced Patient Monitor
Technical Specifications

CMRR>100dB SpO2
Notch: 50/60Hz notch filter can be set to Measurement range: 0-100%
on or off Resolution: 1%
Differential input impedance >5 MΩ Accuracy: ±2% (70-100%, Adult/Pediatric);
Electrode polarization voltage range: ±400 mV ±3% (70-100%, Neonate);
Baseline recovery time <3s after defibrillation 0-69%, unspecified
(in monitor and surgery mode) Refreshing Rate: 1s
Calibration signal: 1 mV (peak - peak),
accuracy ±3% Masimo SET® SpO2(Optional)
Measurement range: 0-100%
RESP Resolution: 1%
Measurement method: Thoracic electrical Accuracy: ±2% (70-100%, Adult/Pediatric,
bioimpedance non-motion, low prefusion);
Rate: 0 – 150 bpm ±3% (70-100%, Neonate,
Measuring lead: Lead I, II non-motion);
Wave gain: ×0.25, ×0.5, ×1, ×2 ±3% (70-100%, motion);
Respiratory impedance range: 0.5-5 Ω 0-69%, unspecified
Baseline impedance: 500-4000 Ω Refreshing Rate: 1s
Gain: 10 grades
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s Pulse Rate
Range: 30~254 bpm
TEMP Resolution: 1bpm
Accuracy: ±0.1 °C or ±0.2 °F (without probe) Accuracy: ±2bpm (non-motion)
Measurement range: 5~50 °C (41~122 °F) ±5bpm (motion)
Channel: Two channels Refreshing rate: 1s
Resolution: 0.1 °C
Parameters: T1, T2 and TD

www.axcentmedical.com
12-lead ECG OxyCRG screen

4 channel IBP Dynamic trends

NIBP IBP (Optional)


Measurement method: Automatic Channel: 2, 4 or 6-channel
oscillometric method ART: 0 to 300 mmHg
Operating mode: Manual, automatic, PA: -6 to 120 mmHg
continuous CVP/RAP/LAP/ICP: -10 to 40 mmHg
Measurement unit: mmHg/kPa selectable Measurement range: P1/P2 -50 to 300 mmHg
Typical measurement time: 20~40s Resolution: 1 mmHg
Measurement type: Systolic, Diastolic, Mean Accuracy: ±2% or ±1 mmHg,
Measurement range (mmHg) whichever is greater (without sensor)
Range of Systolic pressure: Adult 40-270 Sensitivity: 5uV/mmHg/V
Pediatric 40-200 Impedance range: 300 to 3000 Ω
Neonatal 40-135
Range of Diastolic pressure: Adult 10-210 C.O. (Optional)
Pediatric 10-150 Method: Thermodilution
Neonatal 10-95 Range: C.O.: 0.2 to 20 L/min
Range of Mean pressure: Adult 20-230 TB: 23 to 45 °C
Pediatric 20-165 T1: -1 to 27 °C
Neonatal 20-105 Accuracy: C.O.: ±5% or ±0.1L/min,
Measurement accuracy whichever is greater TB, T1: ±0.5°C
Maximum average error: ±5 mmHg (without sensor)
Maximum standard deviation: 8 mmHg
Resolution: 1 mmHg
Interval: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180,
240, 480 minutes
Overpressure protection: Software and
hardware, double safety protection
Cuff pressure range: 0-280mmHg
CETUS xl Advanced Patient Monitor
Technical Specifications

Standard Mainstream CO2 (Optional) Phasein ISA Sidestream CO2 (Optional)


Measurement range: 0-19.7%, Warm-up time: Full accuracy within 10 se-
150 mmHg, or 0-20kPa conds
Resolution: 0.1 mmHg Sampling flow rate: 50ml/min(+/-10/min)
Measurement accuracy Measurement Range: 0 -25%
0 - 40 mmHg: ± 2 mmHg Accuracy: 0~15% (±0.2% of the reading)
41 - 70 mmHg: ± 5% of reading 15~25%, unspecified
71 - 100 mmHg: ± 8% of reading Rise time: 200 ms, typical at 50 ml/min
101 - 150 mmHg: ± 10% of reading flow rate
Respiration rate: 3-150 bpm Total response time: within 3 seconds
Respiration rate accuracy: 1% ±1 bpm (with 2 m Nomoline sampling line)
Warm-up time: 97% within 8s, full accuracy AWRR Range: 0-150 bpm
within 20s AWRR Accuracy: ±1 breath

Standard Sidestream CO2 (Optional) Phasein IRMA™ Mainstream CO2 (Optional)


Measurement rage: 0-20% (0-150 mmHg) Measurement Range: 0 -25%
Accuracy: < 5.0% CO 2: ± 2 mmHg Accuracy: 0~15% (±0.2% of the reading)
> 5.0% CO 2: < 6% of reading 15~25%, unspecified
Respiration rate: 2~150 BPM Warm-up time: Full accuracy within
Respiration rate accuracy: 1% ±1BPM 10 seconds
Warm-up time: 97% within 45s, full accuracy AWRR Range: 0-150 bpm
within 10 min AWRR Accuracy: ±1 breath
Rise times (t10-90%): About 100 ms, when
flow is 100 ml/min, adult water trap,1.5m Phasein IRMA™ AX+ Mainstream Multi-gas
sampling tube (Optional)
Delay time: <3sec when flow is 100 ml/min, Gas: CO2, N2O, HAL, ISO, ENF, SEV, DES with
adult water trap,1.5 m sampling tube automatic identification
Warm-up time: Full accuracy within
Recorder (Optional) 20 seconds for IRMA AX+
Built-in, Thermal dot array CO2 Accuracy: 
Horizontal resolution: 16 dots/mm (25 mm/s 0-10%: ± (0.2%+2% of the reading)
paper speed) 0-15%: ± (0.3%+2% of the reading)
Vertical resolution: 8 dots/mm N2O Accuracy:
Paper speed: 25 mm/s, 50 mm/s 0-100%: ± (2%+2% of the reading)
Number of waveform channels: 3 HAL, ISO, ENF:
0-8%: ± (0.15%+5% of the reading)

www.axcentmedical.com
Vivid visualized icons ... Engineered for the most impressive operation

SEV:0-10%: ± (0.15%+5% of the reading) NMT Tof-Watch® SX (Optional)


DES:0-22%: ± (0.15%+5% of the reading) Microprocessor-controlled
Agent identification time: <20s(typical <10s) Stimulation Mode: TOF, TOFS, PTC, 1Hz
AWRR range: 0-150 bpm Twitch, 0.1Hz
AWRR accuracy: +/-1 bpm Twitch, DBS DBS3.3 and 3.2 (Double Burst),
Apnea time: 20~60s Tetanic Stimulation (Burst), 5s – 50 Hz or
100 Hz
Aspect BISx module (Optional) Output (accuracy ±5% of full scale value)
Parameter Measurement: Surface electrodes:
BC: 0~30 (Only limited to the combined use of Constant current,0-60mA(0-12/18μC) up to
an external sensor with a BIS module) 5 KOhm.
EMG: 30~55dB (bar chart) with intensity Monophasic, 200 μs or 300 μs pulse width
between 30dB and 80dB (tendency chart) Needle electrodes:
BIS: 0~100 Constant current, 0-6 mA (0-0.24 μC) up to
SQI: 0%~100% 5 KOhm.
SR: 0%~100% Monophasic, 40μs pulse width
SEF: 0.5 Hz~30Hz Acceleration transducer: Accuracy ±5% of full
TP:40~100 Db scale value
EEG Measurement: Input impedance >5 MΩ Temperature sensor: Range 20.0-41.5 °C
Noise (RTI) <2μV (0.25~50 Hz) (accuracy ±5 °C)
Input signal range: ±1 Mv
EEG bandwidth between: 0.25 Hz~110 Hz Operation Environment
Power: AC 100-250 V, 50/60 Hz
Temperature: 5-40 °C
Humidity: <80%
Patient Range: Adult, Pediatric, Neonate
Battery backup: Standard 2-3 hrs (2.600 mAh),
optional 3-5 hrs (4.800 mAh)
Patient Monitoring Solutions
For more information, please contact us.

aXcent medical GmbH


Josef-Görres-Platz 2
56068 Koblenz
Germany
Phone: +49 261 30 11 117
Fax: +49 261 30 11 111
Mail: info@axcentmedical.com

www.axcentmedical.com