Acoustic Microstreaming Detection and Measurement Around Ultrasonic Scalers

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

PERI 7006.

print 10/5/00 3:32 PM Page 626

Volume 70 • Number 6

Acoustic Microstreaming: Detection and


Measurement Around Ultrasonic Scalers*
B.S. Khambay and A.D. Walmsley

Background: Acoustic microstreaming (AMS) may be useful to


the clinician when using the ultrasonic scaler to remove particulate
matter from the teeth. The aim of this study was to detect and mea-
sure the effects of AMS produced by ultrasonic scalers.
Methods: For the study, an ultrasonic generator was selected with
4 differently shaped scaling tip inserts (TFI-3, TFI-9, TFI-1, and P-
12). A plaque substitute (0.2 mm thick soft cream cheese) was
coated onto a microscope slide and immersed in water. The ultra-

T
sonic scaler tip was placed in the water and orientated either per- he ultrasonic scaler removes plaque
pendicular or parallel to the slide. The instrument was operated both by either the chipping action of the
contacting the slide under a load of 0.3 N and non-contacting at tip or the biophysical action of the
various distances from the slide surface. This was repeated with the ultrasound within the associated cooling
tip parallel to the slide. The area of medium removed was quantified water. 1,2 The oscillatory action of the
by digital image analysis. ultrasonic scaling tip within water pro-
Results: It was found that AMS removed the plaque substitute duces acoustic microstreaming. There is a
from around the tip. The TFI-9 insert significantly removed more rapid rate of change of streaming velocity
material with increasing displacement amplitude (P <0.05). with distance, from the oscillating tip.3-5
Significantly larger areas of plaque substitute were removed when These velocity gradients occur over a
the tips of the TFI-3, TFI-9, and P-12 inserts were orientated per- small distance and produce large hydro-
pendicularly to the slide compared to the parallel orientation (P dynamic shear stresses even though the
<0.05). Of the 4 inserts used, the TFI-9 insert removed the most velocities themselves are only in the order
material while the straight tip produced no apparent removal. of a few cm sec-1. These forces are large
Removal by AMS required the presence of a water medium and enough to disrupt or damage most biolog-
such forces were found to decrease with distance from the scaling ical cells and tissues.5,6 Dental plaque is a
tip. No plaque substitute removal was seen at a distance of 7 mm biofilm made of organisms and surround-
for the TFI-9 insert at 37.5 µm displacement with the tip orientation ing extracellular matrix and will therefore
parallel to the slide. be susceptible to the physical effects of
Conclusions: It is concluded that AMS occurs around ultrasonic ultrasound.7 The forces generated by the
scalers and this depends on the displacement amplitude, tip orienta- streaming should be of sufficient magni-
tion, and presence of a water medium. AMS may play a role in dis- tude to shear the plaque away from itself
ruption of subgingival biofilms associated with periodontal disease. and from the tooth surface. Such forces
J Periodontol 1999;70:626-631. will assist the mechanical and cavitational
role that occurs during plaque removal by
KEY WORDS the ultrasonic scaler. High energy levels
Acoustic microstreaming; periodontal diseases/therapy; generated by the ultrasonic scaler may
scaling/instrumentation; ultrasonics/instrumentation. disrupt or disaggregate bacteria within the
dental plaque and periodontal bacteria
such as Gram-negative organisms have
* School of Dentistry, University of Birmingham, Birmingham, UK. been shown to be sensitive to ultra-
sound.7-9 The relative resistance of Gram-
positive bacteria, spores and viruses10,11
can be explained by differences in size,
rigidity, and strength of their cell walls.10 It
is postulated that the lack of such features
in Gram-negative species render them

626
PERI 7006.print 10/5/00 3:32 PM Page 627

J Periodontol • June 1999 Khambay, Walmsley

susceptible to the shearing forces of ultrasonic


acoustic phenomena,7,8 and thereby contribute to
the post-ultrasonic instrumental changes in the
microbial composition of dental plaque.12-14 In a
deep periodontal pocket not all the bacteria will be
exposed to the same magnitude of shear forces,
thus possibly producing a further clinical variability.
There have been few studies that have measured the
amount of acoustic microstreaming that occurs dur-
ing plaque removal by the ultrasonic scaler. Previous
work with a model system composed of particulate
matter removed from a slide showed that activity in
the associated cooling water cleaned the slide close
to the probe.2 This removal was demonstrated on
the plaque covered surface of the tooth but no direct
measurement of the amount removed was made.15
Similar areas of erosion or “cleaning” were observed
on root surfaces subjected in vitro to a scaling tip Figure 1.
The scaling tips used in the study (from left to right) TFI-3,TFI-9,TFI-1, and P-12.
with water flow.16 The development of a suitable
model system would assist in determining the rela- Photomicrographs were taken of the observed motion
tive contributions of both the mechanical and the of the particles.
streaming forces around the ultrasonic scaler. This pre-
vious work has demonstrated the presence of acoustic Development of Model System
streaming; 15,16 however, it is not known how such The model system consisted of soft cheese‡ applied
forces are affected by different scaling tip designs and to a microscope slide which was used as a suitable
the magnitude of their oscillation. plaque substitute. An even thickness of the plaque
The aims of this investigation were as follows: 1) to substitute of 0.2 mm was obtained by placing the
develop an in vitro model to detect and measure the slide on a perspex block designed with 2 stops on
hydrodynamic shear stresses produced by acoustic either side of the slide. Excess material was removed
microstreaming; and 2) to determine the factors which by drawing a further slide across the top surface. The
effect the amount of acoustic microstreaming that slide was positioned on one pan of a weighing scale
occurs around the ultrasonic scaler. and weights were placed on the opposite pan. A sin-
gle nylon bristle was removed from a toothbrush and
MATERIALS AND METHODS allowed to contact the surface of the slide 1 minute
Instrumentation after the cheese was first taken out of a refrigerator at
Four designs of ultrasonic scaler tip were selected 4°C. The force required to remove the plaque substi-
(Fig. 1): 1) straight (TFI-1); 2) broad base with taper tute in a single pass was recorded by the gradual
(TFI-3); 3) broad shaped (TFI-9); and 4) sickle addition of weights to the opposite pan. This was
shaped (P-12). All scaling tips were powered by an repeated 10 times.
ultrasonic generator† working at a frequency of 25
Effect of Tip Operation on Medium Removal
kHz.
A coated microscope slide (cheese thickness 0.2
Detection of Acoustic Microstreaming mm) was submerged in a water filled petri dish. The
A glass petri dish was filled with distilled water and scaling tip was positioned at a distance of 1 mm by
zinc stearate powder was directly sieved onto the the use of a calibrated guide. The scaling tip was
water. The tip of an ultrasonic scaler (P-12) was sub- powered for 30 seconds at 2 displacement ampli-
merged 2 mm beneath the surface with the tip paral- tudes corresponding to mid-range power settings.
lel to the surface of the water. All cooling water was The longitudinal oscillations of the tip are along the
directed away from the tip but allowed to cool the long axis of the instrument17 and the scaling tip was
main part of the working insert. The scaling tip was orientated either parallel or perpendicular to the
viewed under a stereomicroscope to produce an coated slide. Therefore, the oscillations were directed
image of the zinc stearate and the water surface. The perpendicular into and away from the slide and par-
scaling tip displacement amplitudes ranged from 10.5 allel where the oscillations moved across the coated
(low power) to 47.5 µm (high power). The displace-
ment amplitude of the scaling tip was measured using † Dentsply, York Division, Long Island, NY.
a stereomicroscope and calibrated graticule. 17 ‡ Kites Cheese Spread, Portsmouth, UK.

627
PERI 7006.print 10/5/00 3:32 PM Page 628

Acoustic Microstreaming Around Ultrasonic Scalers Volume 70 • Number 6

slide was removed and replaced with a new


medium covered slide and the tip reactivated to
ensure the medium was starting to be removed.
This distance was recorded and taken as the fur-
thest extent of the streaming forces. This was
repeated 5 times with the tip orientated either par-
allel or perpendicular and driven at the 2 displace-
ment amplitudes for all 4 designs.
Effect of Water on Medium Removal
A coated slide was placed in the petri dish with no
water added. The TFI-9 scaling tip was driven at a
displacement amplitude of 37.5 µm and orientated
with the scaling tip oscillations parallel to the slide.
The tip was lowered on to the coated slide to contact
the plaque substitute medium under a load of 0.3 N
and activated for 30 seconds. The slide was
removed and the medium was allowed to bench dry
for future measurements. The experiment was
repeated 5 times. The arrangement was repeated
but with water added to the petri dish.
Determination of the Area of Medium Removed
The dried medium covered slides were assessed to
calculate the area of medium which had been
removed during the experiments. This was under-
taken using a computerized imaging system consist-
ing of a personal computer,§ a frame grabber with
768 x 512, 8 bit grey scale square pixels,|| and an
image analysis program.¶ The areas of removal
were captured with a grey scale scanner# at a reso-
lution of 1 pixel = 42.55 mm corresponding to the
area of 1 pixel = 1810.50 mm2. A preliminary scan
of the slides compared the digitized image to the
actual outline on the slide and the contrast and
Figure 2. brightness were adjusted until they displayed the
Diagram showing the experimental apparatus for the detection of acoustic
microstreaming. A (top).The probe tip is set at a parallel orientation with the long
area of interest accurately. Finally the images were
axis of the probe tip parallel to the long axis of the slide; i.e., broken lines are thresholded to binary at an initially determined value
parallel to each other. B (bottom).The probe tip is set at a perpendicular orientation image so that foreground corresponded to the
with the long axis of the probe tip perpendicular to the long axis of the slide, i.e., removed areas and background to the unremoved
broken lines intersect at right angles. cheese. All the slides were scanned using the same
settings for contrast, brightness, and thresholding
slide (Fig. 2). The slide was removed from the water levels. The area of removal was calculated by counting
and allowed to bench dry for future measurements. the number of pixels that belonged to the foreground in
The experiment was repeated for all 4 scaling tip the image.
designs using both orientations. All of the scaling tips
were adjusted so that there was a difference of 7.5 Statistics Analysis
µm between the 2 mid-range power settings. A 2-sample t-test was the statistical analysis of choice
and a confidence level of 95% (P <0.05) was considered
Effect of Tip Distance From Medium to be statistically significant.
and Its Removal
A prepared slide was placed in the water filled petri RESULTS
dish. The scaling tip was lowered into the water The use of zinc stearate particles on the water surface
directly above the medium and activated at different showed that acoustic microstreaming did occur around
displacement amplitudes. Measurements were made
of when the medium began to be removed from the § Pentium 60MHz, Viglin Computers, Alperton, Middlesex, UK.
|| Vision plus-AT OFG, Imaging Technology Inc., Bedford, NY.
slide during lowering of the tip. The instrument was ¶ Optimas Corporation, version 4.0, Bothell, WA.
then deactivated and maintained in this position. The # Hewlett-Packard Scanjet Plus, Tokyo, Japan.

628
PERI 7006.print 10/5/00 3:32 PM Page 629

J Periodontol • June 1999 Khambay, Walmsley

Figure 3.
Video captured digitized image of a P-12 insert showing the moving zinc stearate particle vortices A (left) at 10.5 µm displacement amplitude
and B (right) at 47.5 µm displacement amplitude.

the tip, these were seen as distinct vortices around the Water is required for removal to occur. The removal
tip. At low powers there was an ordered arrangement to was greatest when the slide was immersed in water
these vortices (Fig. 3A). While with increasing power (Fig. 4). Allowing the tip to contact a nonimmersed
these vortices became more erratic in nature and coated slide resulted in a smaller amount of removal.
decreased in number (Fig. 3B).
DISCUSSION
The force required to remove the cheese medium
from the slide in a single pass of the nylon toothbrush The detection of acoustic microstreaming may be
filament was 1.8 grams ± 0.16 grams (1 S.D.; n = 10). demonstrated by the movement of zinc stearate parti-
The effect of tip displacement produced different cles floating on the water surface into which the scaler
amounts of removal of the plaque substitute (Table
Table 1.
1). The TFI-3, TFI-9, and P-12 inserts all removed
more medium from the slide at the higher tip dis- Mean Area Removed From Slide for Different
placement. When the tip displacement was Inserts, Tips With a Parallel or Perpendicular
increased by 7.5 µm, the area of medium did not
Orientation
increase equally for each probe when orientated
perpendicularly to the medium. The TFI-1 insert did Insert Tip Displacement Area removed 1 S.D.
not remove any medium regardless of tip displace- orientation amplitude (µm) (mm2) (n = 5)
ment and there was no evidence of surface distur-
bance. Where removal occurred, increasing tip dis- TFI-3 Parallel 30.0 1.4 0.0
placement by 7.5 µm did not produce similar linear Perpendicular 30.0 0.0 -
increases in removal of the medium. Furthermore, Parallel 37.5 32.2 11.8
where the tip displacement was 37.5 µm, the 3 tips Perpendicular 37.5 0.0 -
differed in the amount of medium removed with the
TFI-9 insert removing the most medium and the P- TFI-9 Parallel 30.0 34.1 0.7
12 tip the least. Perpendicular 30.0 13.2 0.7
There was a marked reduction of medium Parallel 37.5 56.7 14.9
removal with the TFI-9, TFI-3, and P-12 inserts Perpendicular 37.5 31.0 0.8
when the orientation changes from parallel to per-
pendicular (Table 1). All the settings of the probes TFI-1 Parallel 22.5 0.0 -
are the same. The TFI-1 insert still resulted in no Perpendicular 22.5 0.0 -
medium being removed. Parallel 30.0 0.0 -
The effectiveness of the streaming was depen- Perpendicular 30.0 0.0 -
dent on distance. The TFI-9 insert operated at a tip
P-12 Parallel 37.5 0.0 -
displacement of 37.5 µm will produce medium
Perpendicular 37.5 0.3 0.4
removal up to 6.6 mm away from the slide.
However at a tip displacement of 30 µm this was Parallel 45.0 0.9 0.3
reduced to 4.2 mm. Perpendicular 45.0 0.2 0.4

629
PERI 7006.print 10/5/00 3:32 PM Page 630

Acoustic Microstreaming Around Ultrasonic Scalers Volume 70 • Number 6

is placed. Although this provides evidence of the


180
streaming patterns, it does not provide a measurement
of any forces generated. It is these forces which are of
160
clinical significance when using the ultrasonic scaler.

Area of medium removed (sq mm)


It has been possible to develop an in vitro model that 140
begins to place a value on these forces and also assess
the many variables that appear to effect the size of the 120
force. The chosen medium of soft cream cheese has a
tenacity to the glass slide that is comparable to plaque 100
and requires 2 to 4 grams for its removal from teeth.18
Hence the soft cream cheese medium closely resem- 80
bles the in vivo situation and was chosen as a plaque
substitute. Further research is being undertaken to 60
develop an artificial plaque system to determine if there
are any differences in the effects of streaming. 40
The ultrasonic scaler removes the cheese from the
slide due to streaming forces. However this removal 20
pattern was dependent on the different operating vari-
ables of the instrument which were: displacement 0
amplitude of the tip, design, and orientation. As the dis- Non-immersed Immersed
placement amplitude of the inserts increased, there was
a correlated increase in the area of medium removed. Figure 4.
The exception was the TFI-1 insert, which removed no The area of medium removed from the glass slide with the TFI-9 insert at
medium. The displacement amplitude of the tip is pro- 37.5 µm displacement amplitude.The tip was touching the slide and
portional to the power output of the scaler,6,17 and with operated either immersed in or without an associated water medium
(error bars, n = 5, 1 SD).
increasing power output, the magnitude of the shear
forces will be increased and therefore in a larger area of
medium removal. The TFI-1 insert produced acoustic acterized by the production of large shear forces. When
microstreaming that was not of significant strength to the TFI-9 insert was placed on the non-immersed
remove cheese. When operated at the same displace- medium, only the area in contact with the face of the
ment amplitude, the TFI-3 and TFI-9 tips did remove insert was removed (9.2 mm2). When the procedure
material. Therefore, the geometric design appears to be was repeated, with both the medium and tip submerged
a major influence on the production of streaming forces. in water, a much larger area of medium was removed
Of the 4 inserts tested, the TFI-9 insert removed the (138.3 mm2). The difference of 129.1 mm2, can be
most medium, followed by the TFI-3 insert and then the attributed to the effects of acoustic microstreaming. It
P-12 insert. The TFI-9 insert is a rigid design with a was also shown that the magnitude of the shear forces
groove along its front face and may behave as a “pad- decreases with increasing distance from the probe tip. A
dle,” disturbing a larger area of water with greater force, decrease in the displacement amplitude from 37.5 µm
therefore increasing the shear forces. The opposite to 30 µm leads to reduction in the effective distance of
holds true for the TFI-1 insert which has a slimmer the shear forces from 7.6 mm to 4.4 mm. The acoustic
design and is unable to disturb the water to the same microstreaming is not only dependent upon displace-
extent producing little or no streaming forces. ment amplitude but also the shear forces decrease with
The removal of the plaque substitute may also occur increasing distance from the ultrasound source.
by cavitation.2 Small air bubbles in the cheese may act Clinically the shear stresses will be highest in the imme-
as nuclei for cavitation and undergo transient cavitation diate vicinity of the tip and will be most active especially
with subsequent disruption of the cheese from within its if the scaler is working within a periodontal pocket.
structure. The streaming forces then clear the medium
away and this same process may be occurring within CONCLUSION
plaque in the mouth. Both cavitation and acoustic In this in vitro study acoustic microstreaming can be
microstreaming are related physical consequences of indirectly detected by its effect on the disruption and
ultrasound and are synergistic in their actions in clearance of particulate matter in the presence of a
medium removal and their effects merit further investi- water medium. Acoustic microstreaming increased
gation. with increasing displacement amplitude, although it
Acoustic microstreaming is a phenomenon that was dependent upon tip geometry, tip orientation, and
exists in a fluid environment such as water and is char- distance from the oscillating tip.

630
PERI 7006.print 10/5/00 3:32 PM Page 631

J Periodontol • June 1999 Khambay, Walmsley

ACKNOWLEDGMENTS 15. Walmsley AD, Laird WRE, Williams AR. Dental plaque
The authors would like to thank Dr. G. Landini for his removal by cavitational activity during ultrasonic scaling.
J Clin Periodontol 1988;15:539-543.
help with the image analysis and the Photographic 16. Walmsley AD, Walsh TF, Laird WRE, et al. Effects of
Department of the School of Dentistry, University of cavitational activity on the root surface of the teeth
Birmingham. during ultrasonic scaling. J Clin Periodontol 1990;17:
306-312.
REFERENCES 17. Walmsley AD, Laird WRE, Williams AR. Inherent
1. Johnson WN, Wilson JR. Application of the ultrasonic variability of the performance of the ultrasonic descaler.
dental unit to scaling procedures. J Periodontol J Dent 1986;14:121-125.
1957;28:264-271. 18. Volpenhein DW, Walsh ME, Dellerman PA, et al. A new
2. Walmsley AD, Laird WRE, Williams AR. A model system method for in vitro evaluation of the interproximal
to demonstrate the role of cavitational activity in penetration of manual toothbrushes. J Clin Dent
ultrasonic scaling. J Dent Res 1984;63:1162-1165. 1994;1(Pt 1):27-33.
3. Nyborg WL. Acoustic streaming. In: Physical Acoustics.
New York: Academic Press; 1965:265-383. Send reprint requests to: Dr. A.D. Walmsley, School of
4. Holtzmark J, Johnsen I, Sikkeland T, Skavlem S. Dentistr y, The University of Birmingham, St. Chad’s
Boundary layer flow near a cylindrical obstacle in an Queensway, Birmingham, B6 4NN, UK. E-mail: a.d.walmsley
oscillating incompressional fluid. J Acoust Soc Am @bham.ac.uk
1954;26:26-39.
5. Nyborg WL. Physical Mechanisms for Biological Effects of Accepted for publication October 20, 1998.
Ultrasound. HEW Publication (FDA); 1977; publication
no 78-8026. Bethesda, MD. 1997
6. Williams AR, Chater BV. Mammalian platelet damage in
vitro by an ultrasonic therapeutic device. Arch Oral Biol
1980;25:175-179.
7. Thilo BE, Baehni PC. Effect of ultrasonic instrumentation
on dental plaque microflora in vitro. J Periodont Res
1987;22:518-521.
8. Robrish SA, Grove SB, Bernstein RS, et al. Effect of
sonic treatment on pure cultures and aggregates of
bacteria. J Clin Microbiol 1976;3:474-479.
9. Loesche WJ, Laughon BE. Role of spirochaetes in
periodontal disease. In: Genco RJ, Mergenhagen SE,
eds. Host-Parasite Interactions in Periodontal Disease.
Washington: American Society for Microbiology; 1982;
62-75.
10. Thacker J. An approach to the mechanism of killing
cells in suspension by ultrasound. Biochemica
Biophysica Acta 1973;304:240-248
11. Hughes DE. The disintegration of bacteria and other
micro-organisms by the M.S.E.P. Mullard ultrasonic
disintegrator. J Biochem Microbiol Technol Eng
1961;3:405-433.
12. Oosterwaal PJM, Matee MI, Mikx FHM et al. The effect of
subgingival debridement with hand and ultrasonic
instruments on the subgingival microflora. J Clin
Periodontol 1987;14:528-533
13. Loos B, Kiger R, Egelberg J. An evaluation of basic
periodontal therapy using sonic and ultrasonic scalers. J
Clin Periodontol 1987;14:29-33.
14. Baehni P, Thilo B, Chapuis B, et al. Effects of ultrasonic
and sonic scalers on dental plaque microflora in vitro
and in vivo. J Clin Periodontol 1992;19:455-459.

631

You might also like