Professional Documents
Culture Documents
Vent User Manual Iss1
Vent User Manual Iss1
User
Manual
MODIFICATIONS LABEL
0120
Contents
How to use this manual ......................................................................................8
SECTION 1 - INTRODUCTION............................................... 9
1.1 Symbols & Abbreviations ......................................................................... 10
1.2 Identify your Ventilator ............................................................................. 12
1.3 Description of the Ventilator ................................................................... 13
1.3.1 Overview ........................................................................................... 13
1.3.2 Features ............................................................................................. 14
1.3.2.1 Controls & Indicators ........................................................ 14
1.3.2.2 Microprocessor Control ................................................... 14
1.3.2.3 Dual Ventilation Mode .................................................... 14
1.3.2.4 Pressure Limit Control ...................................................... 14
1.3.2.5 Built-In Ventilator Pre-Use Test ...................................... 14
1.3.2.6 Comprehensive Pressure Monitoring and Alarms ..... 14
1.3.2.7 Spirometry ........................................................................... 15
1.3.2.8 Battery Backup ................................................................... 15
1.3.2.9 Calibration Connector ...................................................... 15
1.3.2.10 Data Output....................................................................... 15
1.3.3 Operating Modes ............................................................................. 16
1.3.3.1 Volume Control ................................................................ 17
1.3.3.2 Pressure Control ............................................................... 17
1.3.3.3 Pressure Support .............................................................. 18
1.3.3.4 SIMV + PSV........................................................................ 18
1.3.3.5 Inspiratory Pause .............................................................. 18
1.3.3.6 Sigh ...................................................................................... 19
1.3.3.7 Pressure Support .............................................................. 19
1.3.4 Principles of Operation ................................................................. 20
1.3.5 Parameters ...................................................................................... 22
1.3.5.1 PEEP.................................................................................... 22
1.3.5.2 Trigger ................................................................................. 23
1.3.5.3 Support Pressure .............................................................. 23
1.3.5.4 Fresh Gas Compensation ............................................... 23
1.3.5.5 Oxygen ................................................................................ 23
1.3.5.6 Tidal Volume ...................................................................... 23
1.3.5.7 Minute Volume ................................................................. 23
1.3.5.8 Expired Tidal Volume ....................................................... 23
List of Figures
Figure 1 700/900 Series Ventilator ................................................................13
Figure 2 Pneumatic System D iagram ............................................................21
Figure 3 Front Panel ............................................................................................25
Figure 4 Airway Sensor Placement .................................................................47
Figure 5 Exchanging Bellows Units ................................................................80
Figure 6 Exchanging Fuses ................................................................................85
Figure 7 Exchanging Fuses ................................................................................85
Figure 8 Exchanging Fuses ................................................................................85
Figure 9 Removing the pop-off valve ............................................................87
1. Introduction
Welcome to the User and Routine Maintenance Manual for the Blease 700/900
Series Ventilators.
This handbook is intended for users of the ventilator and contains detailed
installation and operating instructions. It also includes cleaning and sterilization
methods, basic troubleshooting and routine maintenance procedures.
Personnel operating the ventilator must become thoroughly familiar with the
instructions contained in this handbook prior to using the ventilator with patients.
The Blease 700/900 Series ventilators are intended for use by a qualified
practitioner under the direction of a qualified anesthetist. When the ventilator is
connected to a patient, it is recommended that a licensed medical practitioner be
in attendance at all times to react to an alarm or other indication of a problem.
In communication with Blease Medical, quote the model and serial number of the
equipment, with the approximate date of purchase.
If the unit is being returned for repair, indicate the nature of the fault or the work
you require to be carried out. Please obtain clearance from Blease Technical
Support before dispatch.
Power off
Power on
Dangerous voltage
ml Millilitres
O2 Oxygen
psi Pounds per square inch
l Litres
E of C End of Case
Volume Control
Pressure Control
SIMV + PSV
Pressure Support
Vent monitoring
Oxygen
Exhaled volume
monitoring
Airway Pressure
Pressure waveform
Flow waveform
Spirometry Loops
Data Output
1.3.1 Overview
The 700/900 series ventilator is an advanced, microprocessor-controlled,
pneumatically driven ventilator which has been designed specifically for the
mechanical ventilation of adult and pediatric patients under general anesthesia.
The control unit houses the electronic and pneumatic control systems. It sets the
desired ventilatory parameters and generates alarm, warning and status messages.
The control unit also calculates and implements the functions necessary to drive
the bellows pneumatically.
The 700/900 series ventilator is designed for use with a circle-type absorber and is
ideally suited to both low-flow and high-flow applications. The ventilator may also
be used with either a Bain-type coaxial circuit or a Jackson-Rees/Ayres T-piece
type pediatric circuit by using an adaptor.
There are variations based on the drive gas used (air or oxygen) and the drive gas
connector type (depending on the eventual country of use).
See 6.6 Associated Equipment and Accessories in the Appendices for ordering
information for the various types.
1.3.2 Features
Alarms are activated to guard reliably against disconnection, circuit leakage and
unachieved set volumes. The alarms are audible and visible. When activated, the
alarms are listed on the display.
The mains and battery alarms also have light emitting diode (LED) indication.
1.3.2.7 Spirometry
The 700/900 series ventilator offers a simple, accurate and user- selectable display
of set and monitored tidal volumes.
The user also has a choice of one of the following on-screen waveforms:
1) Airway Pressure
2) Pressure/Volume
3) Flow/Volume
4) Airway Flow
There are four ventilation modes; Volume Control, Pressure Control, Pressure
Support and SIMV + PSV.
The differences between the Adult and Pediatric modes are:
• The ventilator sustained pressure requires a greater pressure 'movement' in
adult mode to avoid being activated.
• Setting alarms and limits are, in general, optimized for lower pressures and
volumes to be used in pediatric mode.
When the tidal volume limit is reached, the ventilator immediately cycles into
expiration.
The ventilator delivers volume until the set pressure limit is reached; then
maintains the pressure until the end of the inspiratory time.
The effect of Volume and Pressure modes on the pressure and flow waveforms is
shown below:
Insp
Flow
Exp
The user must set two parameters: the Trigger threshold (5 to 15 L/min), and the
Support Pressure which is referenced to the PEEP level (5 to 30 cm H2 O).
To prevent hyperventilation, any changes in the breathing rate are only allowed to
take place slowly. If the patient fails to make any effort to breathe, the SIMV
function automatically defaults to the ventilation function, identical to Volume
Control Mode at the set SIMV rate.
1.3.3.6 Sigh
When Sigh is activated, the delivered tidal volume in Volume Control Mode is
increased by the set percentage of the duration of one breath at the set
frequency. The default percentage is 10% and default frequency is once every
tenth breath.
In Pressure mode bpm, tidal volume and I:E ratio are all monitored values.
The flow and pressure are measured and monitored by the microprocessor
feedback system.
Expiration occurs when the flow control valve is closed and the expiratory
solenoid opens and releases the gas from the bellows assembly.
In Pressure Control Mode the set pressure is achieved during inspiration and
maintained at that level by allowing a controlled bypass through the expiratory
valve. This allows the required pressure level to be maintained whilst
compensating for any fresh gas flow into the patient circuit.
For expiration the expiratory solenoid is opened which releases the gas from the
bellows assembly.
PEEP
VALVE INSP
SOCKET CONTROL
BOARD
75cmH2O
EXHAUST
RELIEF
PORT
VALVE EXPIRATORY
SOLENOID
VALVE
34.5 psi
FLOW CONTROL VALVE
PNEUMOTACHOGRAPH
DRIVE
GAS GAS
SUPPLY OUTLET
INLET PORT
36-101 psi
FILTER
2.5-7 BAR
Section 1 - 700/900 Series Ventilators
21
1. Description Section 1 - 700/900 Series Ventilators
1.3.5 Parameters
The following parameters can be set and/or monitored when the ventilator is in
operation.
1.3.5.1 PEEP
The 700/900 series ventilator
PEEP
continuously monitors and P(cmH2O)
displays the Positive End
Expiratory Pressure (PEEP). By
default, a PEEP of 2 - 4 cm H2O is
introduced to the circuit, caused PEEP OFF
(Resident PEEP
2-4cmH2O) O
by the bellows assembly, which is
shown as OFF. A PEEP value T
between 3 and 20 cm H2O can be
P(cmH2O)
set from the panel, with an
automatic alarm of more than Pressure Control
When the user changes a value such that it does not comply with the required
difference the box surrounding the feature being adjusted and the value of the
feature causing limiting to be enforced is highlighted red. The value being
adjusted could either be the PEEP level or the relevant pressure control.
1.3.5.2 Trigger
The level of negative flow caused by the patient's attempt to breathe before the
ventilator initiates flow to assist the patient with the spontaneous breath.
1.3.5.5 Oxygen
A fuel cell sensor can be connected to allow monitoring of 0% to 100% O2, with
alarms of 18% to 110%. The sensor can be calibrated
1.3.5.14 Compliance
System compliance in ml/ cm H2 O
Alarms
Activity
Parameters
Status Line
Graphical
Display
Peak
Marker
If a parameter has a pre-set level, as well as a monitored value, the pre-set level is
indicated at the bottom left of the parameter window and the monitored value is
shown in the larger font to the right of the window.
Parameter
Monitored
Value
Pre-set
Level
Editing a Parameter
1. When a parameter is selected, turn the Trak Wheel clockwise to increase
the numerical value or anticlockwise to decrease the value.
2. Press the Trak Wheel when the required value is displayed to confirm
selection.
Selecting a Menu I tem
1. Within a menu, turn the Trak Wheel clockwise to descend the menu or
anticlockwise to ascend the menu. If a menu item is displayed in white, it is
not selectable at this time.
2. Press the Trak Wheel to confirm selection.
User adjustable parameters can be set from the normal screen:
1. Turn the Trak Wheel until the desired parameter is highlighted. When the
cursor reaches the end of one row it will move to the other.
2. Press the Trak Wheel. The parameter is highlighted, indicating that it can be
edited.
3. Rotate the Trak Wheel until the required value is indicated.
1.4.1.5.2 Touchscreen
Setting Parameter
Be sure to use your fingertip to activate the touchscreen, do not use pens or
pencils or any other sharp instruments as these will damage the touchscreen.
To change settings.
• Select parameter
Note If confirm button is not pressed the old value will be restored.
SETUP
1.4.3 Menu Tree
MEASUREMENTS
WAVEFORM SELECT
SYSTEM INFORMATION
MODE
DATE/TIME SETUP
SERVICE MODE
CONFIGURATION
BACKLIGHT
DEFAULTS
ALARM LIMITS
ALARM SETUP
1.4.3.3 Alarm Setup
MUTE TIME
APNEA TIME
WARNING: Pressure limiting is a safety feature and is also used in adult and
pediatric mode to provide ventilation (Pressure Control). It should not be used
for any other purpose, such as creation of micro-breaths or emulation of
specialized high frequency/low volume ventilation systems, otherwise injury to
the patient may result.
WARNING: Pressure limiting is a safety feature and is also used in adult and
pediatric mode to provide ventilation (Pressure Control). It should not be used
for any other purpose, such as creation of micro-breaths or emulation of
specialized high frequency/low volume ventilation systems, otherwise injury to
the patient may result.
SIR
The patient airway flow sensor head must be in the patient circuit in
order to carry out compliance compensation.
The option to run a Pre-Use Test for Compliance Compensation is built into the
Startup menu. (model numbers shown on screens would be different dependant
on configuration).
The opening screen at switch on. The warning screen if the pre-use is
cancelled.
Following YES this is the start screen for the pre-use tests. Follow all on screen
instructions.
At this stage 1 of 2 screens may appear depending on how the gas analyzer is
set-up. The left screen is if the analyzer is scavenged or the right screen if the
analyzer exhaust is returned to the circle.
The instructions given are the preparation for performing 3 ventilator tests.
1. PEEP Valve calibration, 2. Calculation of system dead space (compliance), 3. Leak Test.
Having calibrated the PEEP valve and measured the dead space the leak test starts
on the left and concludes with the right hand screen after 20 seconds.
Having successfully passed the leak test the option is given to test the Manual or
Bag system of ventilation. Making a no decision at this point will enter the
normal run screen. Yes will follow on to the preparation for system test.
NB. This is just a set of instructions it is not an active test. Set the controls as instructed and
carry out the manual test as instructed making the necessary observations that all is working as
correctly.
Having proved that the manual system works the ventilator can be tested at
nominal settings of TV 500 mL, 12 BPM, I:E 1:2.0 Set up the system as described
Press the track wheel when ready and set the Bag/Vent switch to Vent. Ensure
that the Bag and Bellows are full. Observe the displayed wave form and expired
volume are consistent with the expected result. Other tests can be carried out at
the users discretion.
When this test is completed satisfactorily the system is ready for use.
If the optional external PEEP valve is tested during the Pre-Use test. If the valve is
replaced or fitted during use the Pre-Use test must be carried out to ensure
continued accurate operation.
If errors are encountered during the pre-use test then one of the following
screens will be displayed. The screens are designed to give an indication of what
may have caused the error and possible solution.
This screen suggests that the patient pressure is too high (>4cmH2O) to begin the
test.
If the absorber is not set to Vent when starting the test the system will give a
reminder.
Pull forwards a couple of cm, Check the gauges for correct gas
check no foreign objects are supply.
between the absorber and main
frame, push absorber fully back.
Failure to fill the bellows at the If the system has a leak that is
start of the tests will be too large or if the circuit is not
occluded correctly the test will
fail at this point.
NB. It has been noted that this can be caused by fitting a Re-Breathing bag to the “Y”
connector instead of occluding the circuit on the plug provided on the absorber.
If a leak is detected the above causes are the possible places to look. It should be
noted that these are not the only leak possibilities, small holes in the sensor
tubing, Cracks in Breathing circuits are sometimes possibilities too.
Below are the screens that set-up the conditions affecting the opening Pre-use
screens regarding gas analyzer exhaust positioning.
The gas analyzer exhaust gas can be dealt with in two ways a) it can be simply
exhausted to atmosphere (scavenged) or b) it may be returned to the circle.
These choices need to made by the hospital and the selection made at the time
of commissioning the system. The selection can have marked effects on he leak
tests carried out during the pre-use testing.
If the sample line is disconnected but the exhaust is returned to the circle then it
may mask leaks by providing an additional gas input. If the reverse situation I.e.
the sample line is connected but the exhaust is not then it would constitute a
leak.
Fresh Gas Flow Compensation has several effects on the operation of the
ventilator. In normal conditions the volume added to each breath due to fresh
gas flow / time is deducted from the driving gas volume that is used to compress
the bellows, this gives a volume at the "Y" piece that is constant.
2.2.4 Standby/Bag/
Run
By turning the absorber to VENT
the ventilator will automatically
run.
The ‘normal’ situation at mid-high tidal volumes is that there will be the option to
change either I:E or fresh gas. The I:E is highlighted in red, and the value to set
the fresh gas to is displayed in the alarm panel (i.e. in this case 10 LPM).
If the tidal volumes are low, the allowable fresh gas flow will be low, and where
the suggested calculated fresh gas flow would be less than 2 LPM, the ventilator
would say “Reduce Fresh Gas”.
In some cases, at high fresh gas flow, the calculated I:E would be outside the
range of I:E settable on the ventilator. The maximum I:E settable on the ventilator
is 1:5.0. i.e. if calculated I:E should be 1:7 which is not possible. Then the only
option is to reduce the fresh gas flow. Thus the I:E will NOT be forced red, and
will be left at it’s set value.
Now rotate to It. It IMMEDIATELY shows the correct value, just CLICK and
accept.
If nothing happens during the first 12 seconds of the alarm the suggested I:E will
be forced. The alarm will continue to sound.
3. Turn the Trak Wheel to highlight a feature on the menu, for example Fresh
Gas.
4. Press the Trak Wheel. The left-hand button on the second row of displayed
parameters is highlighted.
5. Turn the Trak Wheel to highlight each button in turn until a new one is
selected.
6. Press the Trak Wheel again to assign Fresh Gas to the highlighted button and
return to the Features Selection menu. Follow the same process to un-assign
a button.
7. Select Return to close Measurements Selection Menu.
8. Select Return to close the Main menu and go back to the normal screen.
4. Select either “Sensor At Patient” or “Sensor At Absorber” and press the Trak
Wheel to confirm.
When the sensor is at the absorber expiratory port the ventilator can only be run
in Volume Control and Pressure Control modes.
Placing the sensor at the absorber has several implications on the running of the
ventilator.
2. The user cannot select to view spirometry graphs. Waveform Select is grayed
out and inaccessible in the menu. If the graph was previously set to “Flow/
Pressure” or “Flow/Volume” when “Sensor at Absorber” is chosen, the graph is
forced to be pressure.
P P
Patient Airway
Patient Airway
Flow Sensor
Flow Sensor
Connector
Connector
O2 O2
% %
Spirometry
Spirometry
Sensor Head
Sensor head
YYCircuit
Circuit
YYCircuit
Circuit
Patient Airway
Flow Sensor
Patient Airway
Flow Sensor
Connector
Connector
Spirometry
Spirometry
Sensor Head
Sensor Head
At Patient At Absorber
In all modes the airway pressure bar is shown to the left of the trace area. This
shows the real time airway pressure. The peak value of the last breath is also
displayed.
Previous Peak
2.6 Spirometry
Once either of the spirometry waveforms has been selected from the waveform
menu, an extra box (’Store Loop’) becomes available on screen for selection.
Once the user is satisfied that the patient is in a stable pre-operative condition,
the next loop to be displayed can be saved.
To save the loop, highlight the loop button directly below the loop and confirm.
A message that states saving, will be displayed. The next loop will be saved and
displayed in red. Up to 6 loops can be saved.
Any one of the stored loops can be recalled by highlighting the stored loop
button, then use the trakwheel up/down arrows to select the required loop, then
confirm.
By using the highlighted button, you can select which loop defined by time to
recall and display or turn the saved function off. As shown below.
Recall a
Loop
(by saved
time) Turn
saved
Loop OFF
Save a
Loop
(Time of
saving is
displayed
next to
button)
2.6.2 Pressure/Volume
2.6.3 Flow/Volume
Case Evolution
When the pressure is greater than or equal to zero, the trace is blue and when it
is less than zero, it is shown in green.
Zero Graticule
Inspired to patient
This waveform represents the airway flow. The Trace autoscales so that the
amplitude is 80% of the available height. The scale to the side still shows airway
pressure. The zero graticule is shown as a black dotted line.
The flow to the patient is shown as green and the expired flow is shown as blue.
4. Each mode enables different options - the available functions are those which
are displayed in black text.
This selection represents the same standby function as End Case, except the case
timer does not stop and measurement of flow in the patient circuit does not
bring the ventilator back out of standby.
To exit Cardiac Bypass turn the ventilator ON by activating the bag to vent switch
or end the selection of Cardiac Bypass in the ventilator control menu.
2. The selected parameter window is highlighted. Turn the Trak Wheel clockwise
to increase the set value or anti-clockwise to decrease the set value.
The value will not increase above the high pre-set limit, or decrease below the
low pre-set limit.
2.9 Alarms
The 700/900 series ventilator contains a number of pre-set and user-defined
alarms for patient monitoring and monitoring of the ventilator itself.
• the alarm is added to the list of active alarms in the alarm zone on the
screen.
The active alarm is displayed on the list in the alarm zone for a minimum Of
30 seconds. If the alarm condition is removed during this time, the alarm is
automatically cleared from the list. If more than one alarm is activated, each new
alarm is added to the top of the list.
ALARM RANGE/SETTING
Apnea (Default) 30 seconds User Defined 15-60 secs.
As a safety precaution while this alarm is active, the ventilator flow delivery
system is prevented from trying to compensate for the lack of supply gas by
opening the flow control valves further. This stops the large surge of flow that
may otherwise occur if the supply gas is suddenly re-applied, particularly when a
bottled supply is changed.
• The ventilator settings (BPM, I:E, TV) require a flow greater than 100 Ipm or
less than 1.5 Ipm.
A few seconds after mains failure occurs, the Power Fail indicator is illuminated. If
the condition persists for more than one minute, the alarm is activated and the
LED remains illuminated RED.
To conserve power, the display is switched off while the Power Fail alarm is
activated. To restore power to display press a key or turn trak wheel.
The ventilator continues to function normally, powered by its internal battery for
a minimum of 1 hour.
When the internal battery is nearly exhausted and has less than 15 minutes of
running capacity left, the Battery Low alarm is activated and the Battery Low
indicator is illuminated AMBER.
When mains power is restored, the Battery Low indicator may remain lit for a few
minutes as the battery recharges.
If the Battery Low indicator illuminates without a mains failure, there may be a
battery or charging fault. The ventilator will then have no battery backup and
must not be used until the fault has been repaired.
2.9.9 No Charge
Mains supply ON, battery not charging.
In pressure control mode the high pressure alarm is 25% above set pressure. In
Volume Control mode, SIMV and Pressure Support modes the pressure high
alarm limit is set via pressure limit setting on main screen. It will be activated if
one of the following occurs:
• The patient breathing system pressure rises significantly above that set
limit.
• The patient breathing system pressure has not fallen after a pressure-
terminated breath has cycled into expiration.
• Pressure cycling occurs in adult ventilation mode.
Activated when the monitored expired volume exceeds the set limit. When the
expired volume falls below the alarm limit, the alarm immediately clears.
Activated when the monitored expired volume falls below the set limit.
Activated when the monitored BPM falls below the set level.
Activated when the monitored oxygen level rises above the set level.
Activated when the monitored oxygen level falls below the set level.
2) during Ventilation if fresh gas is more than 65% of required tidal volume.
The set values are ± 25% of the current monitored value. Except in AUTO SET
the following cases:- OFF
If the user goes into the Set Alarms Menu whilst the auto
alarms are set the previous values set will be restored
without warning.
The user can restore the previous set of alarm limits at any
time by either selecting Off in the auto box, or going into the
Alarm Limits Menu and making no changes.
High and low alarms are linked intelligently so that a low alarm
value greater than a high alarm value cannot be set.
4. Alarm levels can be set by turning the Trak Wheel clockwise (increase) or
anti-clockwise (decrease).
5. When the required value is displayed, press the Trak Wheel to set the
parameter.
6. Select Return to close the screen and return to the alarm options.
The first three items can be stored in Adult or Pediatric mode. Ensure
that the Patient Selection Switch is in the required position before
selecting these items.
3. Select one of the names displayed using the Trak Wheel. The first character of
the name is displayed in Red when the Track wheel is pressed.
4. Turn the Trak Wheel clockwise to move through the alphabet from A to Z
and then through the numbers 1 to 0. To insert a space, use the under-score
character ( _ ).
5. Press the Trak Wheel when the desired character is displayed to enter the
character and move the cursor on to the next letter.
7. Press Default Menu again to save the name and exit the menu.
This menu option allows the change of default start up settings from the factory
set ones and is password protected. Contact the distributor for further details.
4. Press the Trak Wheel to store the ventilator's current settings as the start up
settings.
The configuration of the ventilator can be saved and recalled. This includes all
user-settable features, alarm limits etc.
2. The Default Menu opens. Using the Trak Wheel, select Save Current
Settings?
4. Press the Trak Wheel to initiate the save. A message will be displayed played
stating the setting (Adult or Ped), and the saved name.
2. The Default Menu opens. Using the Trak Wheel, select Recall Settings?
3. The Recall Settings menu opens. Select a name from the menu using the Trak
Wheel
2.13.6 Configuration
By selecting the Configuration option, the following menu is displayed:
2.13.7 Set Up
The Set up menu allows the user to:
2. Select configuration.
2. Enter the Setup Menu and choose Configuration and then Service Mode.
3. Disconnect the gas supply and turn off cylinders to ensure that there is no gas
flow when zero is set. Having set zero reconnect the gas supply and turn on
the cylinders, you will be prompted for each required flow, after each flow has
been set there will be an audible beep to confirm.
The calibration can be aborted at any time - doing so will cause the last set of
saved calibration values to be used. If the user inadvertently sets the same
flow for two readings, the calibration will automatically be aborted.
3. The volume of the sigh breath can be set to the volume plus 5% - 60% of the
set volume in increments of 5%. (Default settings are 1-10 and 10%).
3. Installation
3.1 Cautions
Cautions Cont.
To preserve battery life, never store the ventilator with its battery
discharged. Do not store or use the vent ilator in close proximity to
heat sources of any kind.
The driving gas is discharged through the port on the rear panel of the
ventilator. This port m ust be completely free of any obstruction and
should have nothing connected to it. The exhaust gas is oxygen or air
and does not contaminate the environment.
Notes
Notes
A fully charged battery (8 hours) in serviceable condition will allow the ventilator
to run for a minimum of one hour.
3. Push the adult bellows housing down over the bellows, then twist it slightly
clockwise to engage the bayonet tabs.
For example:
2. Disconnect the mains supply and check that the MAINS FAIL alarm activates.
3. Reconnect the mains electrical power and check that the alarm resets.
5. Disconnect the O2 and Air hose and turn off the O2 cylinder. The SUPP GAS
LOW alarm should activate.
4. There is an O-ring located in the bellows base, (see opposite) which provides a
seal with the pop-off valve. After cleaning, check that the O-ring is in place, as
the ventilator cannot function correctly without it.
If the valve seat is damaged, the pop-off valve will leak and may
cause serious malfunction. Take care not to damage the
precision-molded surface of the valve seat while cleaning. Never
use a hard object or abrasive detergent. Use only a soft, lint-free
cloth.
G
D
To clean the flow sensor, pour distilled water slowly and gently into the outlet
port. Water must not be directed into this port under pressure.
4.5.4 Bellows
Only the bellows base and the parts inside the bellows require
sterilization.
2. Loosen the two thumb screws retaining the bellows base and remove the
bellows base and housing from the control unit.
3. Remove the pop-off valve and pediatric bellows adapter (if fitted) and clean
and sterilize them separately.
ITEM METHOD
Notes:
Notes
5. Troubleshooting
The table below describes faults, their probable causes and recommended
corrective actions. If a fault persists, do not use the ventilator. Contact your Blease
service representative.
Exchange hose if
damaged.
Unwanted PEEP and Defective or poorly Adjust or exchange
overfull bellows. regulated scavenging scavenging system.
system.
6. Appendices
6.1 Performance Data
CONTROL/PARAMETER RANGE/VALUE
Frequency 50/60 Hz
Power consumption 50 VA
6.2 Dimensions
Height 185 mm
Width 310 mm
Weight 10kg
VOLUME
CONTROL
Set Meas/set Set Pressure
Limit
PRESSURE
SUPPORT
ALWAYS ALWAYS Meas Meas Meas Pressure
Limit
SIMV +
PSV
ALWAYS ALWAYS Meas/
set
Meas/set Meas/
set
Pressure
Limit
Set
PRESSURE
CONTROL
Set IFLOW Set Pressure
RS232 Setup
Data Bits 8
Stop Bits 1
Parity None
Baud Rate 19200
6.7.2 Protocol
Message Function
Set User set value
Measured Monitored measured value
Event Monitor events
Alarm Alarm status and message
[MPPxxx ’\r’ Measured breath pressure peak xxx=3 digit value in cmH2O*10
The relevant bits are set if an alarm has occurred unused bits should not ever be
set.
7. Notices
7.1 Product Improvement
The 700/900 Series Ventilator conforms with the specifications and operating
procedures described in this manual and on any accompanying notices and labels
only if it has been installed, used and maintained in accordance with the
instructions. The safe function of the ventilator can only be guaranteed if it is
regularly checked and serviced at or in excess of the standards specified in this
manual.
If the ventilator is suspected of being worn, defective or otherwise unfit for use, it
should under no circumstances be used.
The manufacturer accepts responsibility for the effects on safety, reliability and
performance of the equipment only if:
NB
If during the warranty period the equipment is serviced by an unauthorized party,
the warranty will be void.
7.4 Disclaimer
Opening of the control unit by unauthorized personnel automatically voids all
warranties and specifications. The prevention of tampering is solely the user's
responsibility; the manufacturer assumes no liability for any malfunction or failure
of the ventilator if the control unit is opened.
Technical data, calibration co-efficients and all intellectual property rights of this
product remain the property of Blease. To ensure correct functioning, the
equipment must be serviced at regular intervals.
The contents of this manual are not binding. If any significant difference is found
between the product and this manual please contact Blease Medical Equipment
Limited for further information.
In communication with Blease Medical Equipment Limited, quote the model and
serial number of the equipment, with the approximate date of purchase. If the
equipment is being returned for repair, indicate the nature of the fault or the
work you require to be carried out.
www.blease.com
7.7 Copyright
All rights reserved. The information contained in this publication may not be used
for any other purpose than that for which it was originally supplied.
This publication may not be reproduced in part or in whole without the written
consent of Blease Medical Equipment Limited.
7.8 CE Marking
0120
The product is labeled with the CE mark.
Warning Notices
Caution Notices
Warnings
1. USE OF EQUIPMENT
Incorrect use of the equipment described herein may result in injury to
the patient. Read this manual before operating the machine. You must
be familiar with the machine and its functions before using it on a
patient.
3. PRESSURE LIMITING
Pressure limiting is a safety feature and is also used in adult and
pediatric mode to provide ventilation (Pressure Control). It should not
be used for any other purpose, such as creation of micro-breaths or
emulation of specialized high frequency/low volume ventilation
systems, otherwise injury to the patient may result.
5. AUDIBLE ALARM
An audible alarm indicates an anoma lous condition that may result in
damage to the equipment or injury to the patient. The cause of each
alarm should be investigated and any necessary measures taken to
remove the alarm condition.
8. EXPLOSION HAZARD
9. DRIVING GAS
Use of a driving gas other than oxygen or air may damage the
ventilator and will cause inaccurate operation, resulting in potential
injur y to the patient. Do not use any other driving gas.
This equipment may be damaged if the gas supply pressure is too high.
The ventilator must only be connected to gas pipeline supply lines that
are fitted with pressure relief valves that limit the supply pressure to
less than 7 bar.
2. BATTERY LIFE
To preserve battery life, never store the ventilator with its battery
discharged. Do not store or use the ventilator in close proximity to heat
sources of any kind.
3. POWER FAILURE
4. ELECTROMAGNETIC INTERFERENCE
5. EXHAUST GAS
The driving gas is discharged through the port on the rear panel
of the ventilator. This port must be completely free of any
obstruction and should have nothing connected to it. The exhaust gas
is oxygen or air and does not contaminate the environment.
6. BELLOWS EXHAUST
8. CLEANING AGENTS
9. VALVE SEAT
If the bellows valve seat is damaged, the pop-off valve will leak and
may cause serious malfunction. Take care not to damage the precision-
molde d surface of the valve seat while cleaning. Never use a hard
object or abrasive detergent. Use only a soft, lint-free cloth.
Cautions continued.
Check flow sensor tubing regularly for trapped water as it could affect
ventilator readings.
8. Glossary
I:E Ratio The ratio of the inspiratory time to the expiratory time.
TV Tidal Volume.
9. Index
A B
Absorber Setting................................ 71 BAG .......................................................40
Activity Display Area......................... 26 Battery Backup ...................................15
Adult and Pediatric ........................... 45 Battery Status LEDs ...........................30
Airway Flow ........................................ 52 Bellows Adult (installing) .................81
Airway pressure................................. 52 Bellows Units (exchanging) .............80
Alarms ................................................. 55 BPM (Frequency Control) .................23
apnea ............................................... 57 Breathing Control ..............................101
auto .................................................. 62 Breathing Modes................................101
display Area .................................... 27 Built-in ventilator Pre-use test........14
fresh Gas Too High ........................ 61
high BPM ......................................... 61 C
high Oxygen .................................... 61 Calibration Connector.......................15
low BPM .......................................... 61 Cardiac Bypass ...................................53
low Oxygen ..................................... 61 Cautionary Notices ............................112
low Supply Gas Pressure.............. 57 Cleaning and sterilization ................86
MV high............................................ 60 bellows ............................................88
MV low ............................................. 60 patient airway flow sensor..........88
menu ................................................ 32 pop-off valve seat .........................86
messages ......................................... 103 Ventilator Surfaces .......................86
mute ................................................. 64 CE Marking ..........................................107
PEEP.................................................. 59 Change Names ...................................66
power fail & Low Battery ............. 59 Copyright .............................................107
pressure high .................................. 60 Configuration......................................70
pressure low ................................... 60 Configuration menu ..........................66
pre-set .............................................. 57 Control Unit ........................................84
setting .............................................. 63 Compliance ..................................... 24/82
setting Apnea Time ....................... 65 Compliance test skipped ..................39
setting Error .................................... 58 Controls & Indicators ........................14
setting Mute Time.......................... 64 Comms fail ..........................................59
setup..............................................32/63 Comprehensive pressure monitoring
summary.......................................... 56 ...............................................................14
sustained pressure alarm ............ 58
under Pressure ............................... 59
user Defined ................................... 60
vent Inop ......................................... 57
Appendices ......................................... 91
Associated equip & Accessories 98
D G
Data Transmission Settings............. 100 Glossary .............................................. 116
Data Output ........................................ 15
Defaults Menu.................................... 32 H
Description of the Ventilator .......... 13 Hazard information .......................... 110
Discharged Battery ........................... 79 Hazard Notices………………………108
Disclaimer ........................................... 106
Dimensions ......................................... 94 I
Dual mode ventilation ..................... 14 Identify you ventilator ..................... 12
I:E Ratio ............................................... 24
E I:E Setting ............................................ 42
End of Case ......................................... 33 Inspiratory Pause .............................. 19
Event Codes ........................................ 104 Installation ......................................... 73
Event Message.................................... 102 Introduction ....................................... 9
Exchanging fuses ............................... 85
Expired Tidal Volume ....................... 23 M
External Communication Spec........ 100 Maintenance (other) ......................... 85
Mean Pressure ................................... 24
F Measurements Menu ........................ 44
Factory Default Settings................... 95 Measured Value Messages .............. 102
Features ............................................... 14 Menu Tree ........................................... 31
Fixed Features .................................... 96 Methods of Sterilization .................. 89
Flow/Volume ...................................... 50 Microprocessor Control ................... 14
Fresh Gas............................................. 39 Minute Volume .................................. 23
calibration ....................................... 71 Mode Dependant Features .............. 97
compensation ................................. 23 Mode Selection Menu....................... 53
Front panel ......................................... 25
N
Normal Screen ................................... 26
Note to Service Personnel ............... 106
No Charge ........................................... 59
O
Operating Modes............................... 16
Overview ............................................. 13
Oxygen................................................. 23
Oxygen Calibration ........................... 43
R U
Recalling Loops ................................. 49 User Interface .................................... 25
Recall previous settings ................... 69 Using the Ventilator ......................... 33
Removing Pop-off valve .................. 87
Responsibilities of the user ............ 105 V
Responsibilities of the Manufacturer .... 102 Ventilator (using) .............................. 33
Routine user maintenance .............. 84 Powering up ........................... 33
Run ....................................................... 40 Powering down...................... 33
Starting .................................... 33
S Stopping .................................. 33
Save Current Settings ....................... 68 Volume Control ................................. 17
Save Hospital Defaults ..................... 68 Volume Ventilation........................... 17
Saving and recalling Settings ......... 68
Sensor placement ............................. 46 W
Waveform Select Menu.................... 48
Weekly Checks................................... 84