Professional Documents
Culture Documents
Zeus Ie Sw2n Ifu 9510797 en
Zeus Ie Sw2n Ifu 9510797 en
Zeus Ie Sw2n Ifu 9510797 en
Screen reproductions
Use of terms
WARNING
A WARNING statement provides important in-
formation about a potentially hazardous situa-
tion which, if not avoided, could result in
death or serious injury.
CAUTION
A CAUTION statement provides important infor-
mation about a potentially hazardous situation
which, if not avoided, may result in minor or mod-
erate injury to the user or patient or in damage to
the medical device or other property.
NOTE
A NOTE provides additional information intended
to avoid inconvenience during operation.
The tasks described in this document specify the The target groups may only perform the following
requirements that have to be met by each tasks if they meet the corresponding requirements.
respective target group.
The operating organization of this product must Users
ensure the following:
Task Requirement
– The target group has the required qualifications
(e.g., has undergone specialist training or Use of the product in Specialist medical
acquired specialist knowledge through accordance with the knowledge in anes-
experience). intended use thesia
Service personnel
Task Requirement
Installation Specialist knowledge
Basic service work in electrical engi-
(inspection, mainte- neering and me-
nance according to chanics, experience
the "Maintenance" servicing medical de-
chapter) vices
Contents
Neuromuscular
Transmission Monitoring NMT . . . . . . . . . . . . . 437
Anesthesia depth monitoring BIS . . . . . . . . . . . 438
Essential performance data . . . . . . . . . . . . . . . 439
EMC Declaration . . . . . . . . . . . . . . . . . . . . . . . 440
Device combinations . . . . . . . . . . . . . . . . . . . . 442
Connections to IT networks . . . . . . . . . . . . . . . 443
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . 445
Gas flow diagram . . . . . . . . . . . . . . . . . . . . . . . 446
Gas control loops . . . . . . . . . . . . . . . . . . . . . . . 449
Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 454
Ventilation modes . . . . . . . . . . . . . . . . . . . . . . . 454
Appendix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 459
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 461
WARNING WARNING
Risk due to incompatible accessories Risk of electric shock and device malfunction
Dräger has only tested the compatibility of ac- Any connected devices or device combina-
cessories listed in the current list of accesso- tions not complying with the requirements
ries. If other, incompatible accessories are mentioned in these instructions for use can
used, there is a risk of patient injury due to compromise the correct functioning of the
medical device failure. medical device and lead to an electric shock.
Before using the medical device, refer to and
Dräger recommends that the medical device is
strictly comply with the instructions for use of
only used together with accessories listed in
all connected devices and device combina-
the current list of accessories.
tions.
WARNING
WARNING
Risk of operating errors and incorrect use
Risk of device malfunction
Strictly observe the instructions for use of all
The medical device may be operated in combi-
accessory parts, e.g.:
nation with other Dräger devices or with de-
– Water traps
vices from third party manufacturers. If a de-
– Flow sensors
vice combination is not approved by Dräger,
– CLIC adapter the safety and the functionality of the individ-
– CLIC absorber ual devices may be compromised.
– Soda lime – The operating organization must ensure
– Breathing hoses that the device combination complies with
– Masks the applicable editions of the relevant stan-
– Filters dards for medical devices.
– Endotracheal suction system – Strictly observe assembly instructions and
– Vaporizer instructions for use for each networked de-
– Manual resuscitator vice.
– AGSS terminal unit
WARNING
Risk of explosion and fire
This medical device is neither approved nor
certified for use in areas where oxygen con-
centrations greater than 25 Vol.%, or combus-
tible or explosive gas mixtures, are likely to
occur.
Safe connection with other electrical Patient safety can be achieved by a wide variety of
equipment means ranging from electronic surveillance of
medical device performance and patient condition
to direct observation of clinical signs.
CAUTION
Risk of patient injury The responsibility for selecting the best level of
patient monitoring lies solely with the user of the
Connections to electrical devices not listed in medical device.
these instructions for use may only be made when
approved by each respective device manufactur- NOTE
er. Only use the medical device for therapy on one pa-
tient at a time.
Patient safety
Information on electromagnetic
The design of the medical device, the compatibility
accompanying documentation, and the labeling on
the medical device are based on the assumption General information on electromagnetic
that the purchase and the use of the medical device compatibility (EMC) according to international
are restricted to persons familiar with the inherent EMC standard IEC 60601-1-2:
characteristics of the medical device.
Electrically powered medical devices are subject to
Instructions and WARNING and CAUTION special precautionary measures regarding
statements are therefore largely limited to the electromagnetic compatibility (EMC) and must be
specifics of the Dräger medical device. installed and put into operation in accordance with
The instructions for use do not contain any the provided EMC information (see page 440).
information on the following points: Portable and mobile radio frequency
– Dangers that are obvious to the user communication equipment can affect medical
electrical equipment.
– Consequences of obvious improper use of the
medical device
– Possible negative impacts on patients with
various underlying diseases
Medical device modification or misuse can be
dangerous.
Patient monitoring
WARNING WARNING
Risk due to electrostatic discharge Danger of electric shock
If no measures are taken to protect against The connection of devices to auxiliary power
electrostatic discharge in the following situa- sockets can lead to an increased leakage cur-
tions, malfunctions may occur which put the rent. If the protective ground of one of these
patient at risk: devices fails, the leakage current may rise
– When touching the pins of connectors that above the permissible values.
display the ESD warning symbol. – Only connect with the approval of the re-
– When establishing connections with these spective device manufacturer.
connectors. – Have the leakage current checked by ser-
vice personnel.
In order to avoid malfunctions, observe the
– If the permissible value is exceeded, use a
following measures and instruct those per-
mains power socket on a wall instead of
sons directly involved accordingly:
the auxiliary power socket of the device.
– Observe ESD protective measures. These
protective measures can include wearing
antistatic clothing and shoes, touching a
potential equalization pin before and Sterile accessories
during connection of the pins, or using
electrically insulating and antistatic CAUTION
gloves. Risk of medical device failure and patient injury
– Comply with the requirements of the elec-
trostatic environment. Observe the follow- Do not use accessories in sterile packaging if the
ing section: "Electromagnetic environ- packaging has been opened, damaged, or if there
ment" on page 440 are other signs that the accessories are not
sterile.
WARNING Reuse, reprocessing, and sterilization of dispos-
Risk due to electromagnetic disturbances able products is not permitted.
WARNING WARNING
Risk of misinterpretation Risk of device failures
Misdiagnosis or misinterpretation of the mea- The device may fail if the power supply is in-
sured values or other parameters may endan- terrupted.
ger the patient.
Always connect the device on an uninterrupt-
Do not make therapeutic decisions based ible power supply.
solely on individual measured values and
monitoring parameters. Therapeutic deci- WARNING
sions must be made solely by the user.
Risk of not hearing the alarm tone
WARNING WARNING
Risk of burns Risk of device catching fire
Conductive breathing hoses or face masks In the event of a patient-side fire, there is a risk
may cause burns during HF surgery. that the device may also ignite.
Do not use this type of hose and mask com- In the event of a fire, separate any connections
bined with HF surgery. carrying oxygen from the device.
WARNING WARNING
Risk of burns Risk due to the accumulation of acetone in the
patient
Do not use humidifiers or heated hoses!
When low-flow anesthesia is used (especially
WARNING in the uptake mode), the risk of accumulation
of acetone in patients with ketoacidosis or
Fire hazard!
who are under the influence of alcohol in-
Ignition sources such as electrosurgery and creases as the duration of the therapy increas-
laser surgery equipment could cause a fire es.
when in contact with oxygen or nitrous oxide.
With patients with ketoacidosis or who are un-
– Avoid leaks, e.g., in the endotracheal
der the influence of alcohol, use a fresh-gas
tubes, laryngeal masks, breathing masks,
flow of at least 1 L/min.
y-piece, breathing system including hos-
es, filters and breathing bag, at the exter-
nal fresh-gas outlet, and at the outlet for O2 WARNING
insufflation. Fire hazard!
– Only use hoses on the outlet for O2 insuf-
The flow sensor may ignite medications or
flation that are intact and leakproof. other substances that are easily flammable.
– Before starting electrosurgery or laser sur- – Do not nebulize medications or other sub-
gery, sufficient flushing with air (<25 % O2) stances that are easily flammable or spray
must be carried out under drapes as well. them into the device.
– Close the flow control valve of the O2 flow- – Do not use substances containing alcohol.
meter all the way. – Do not allow flammable or explosive sub-
– If O2 outlets are used (e.g., for insufflation), stances to enter the breathing system or
do not use any ignition sources in the im- the breathing circuit.
mediate proximity. – Do not use cyclopropane or ether.
– Do not place sources of oxygen in close
proximity to ignition sources, e.g., electri-
WARNING
cal connections.
– Electrical connections must be kept at Risk due to flow measurement failure
least 200 mm (7.9 in) from any parts carry- Deposits which are not removed during repro-
ing oxygen and/or nitrous oxide. cessing may damage the measuring wires in
– Cables and connections must be suffi- the flow sensor or cause fire.
ciently insulated and may not be damaged. – Before inserting the flow sensor check for
Check the cables for damage daily. visible damage, soiling, and particles. Re-
peat this check regularly.
– Replace flow sensors when damaged,
soiled, or not particlefree.
WARNING WARNING
Risk of strangulation Risk of patient injury
Make connection to the patient carefully. The following operating states are prohibited
when the patient is connected:
WARNING – Standby
Zeus IE must not be used with magnetic reso- – Self test
nance imaging (MRI, NMR, NMI). Device oper- – Leak test
ation may be affected, thus placing the patient – Leakage assistant
at risk. – Pretest
– Service mode.
WARNING No monitoring is possible in these operating
Device contamination possible. states. During the test, pressures and concen-
trations may occur in the system which are
Only operate Zeus IE with airway filters.
hazardous to the patient!
WARNING
CAUTION
Risk of patient injury
Risk of crushing
When using near-patient filters do not scav-
Movable device parts or attached components
enge at the patient end! In the event of steno-
may cause crushing due to clamping. Pay special
sis or increased filter resistance the patient
attention to edges, movable parts, and corners
may be injured by vacuum.
when working with the following components:
– Column cover
WARNING – Breathing system cover
Risk of equipment malfunction – Drawers
Only have network functions configured by – Extensible writing tray
authorized, trained personnel. – Swivel arms for mounted devices
– Accessories such as gas cylinders, vaporiz-
Errors in configuration can impair operation ers, CLIC absorber, and CLIC adapter
of the device and the network.
CAUTION
Risk of crushing
If the writing tray is not correctly locked in place,
objects can fall down or fingers and breathing
hoses, for example, can be pinched.
Make sure that the writing tray is correctly locked
when folding down or sliding into the device.
CAUTION NOTE
Risk of tipping over during transport Do not cover or close off the vents on the device.
Make sure there is an adequate supply of air. Oth-
The medical device may tip over if handled incor-
erwise, the device may overheat.
rectly. Observe the following points when trans-
porting medical devices:
NOTE
– The medical device may only be moved by
people who have the physical ability to do so. Risk of injury
– To improve the maneuverability, transport the Handle the DIVA metering module with care and
device with 2 persons. prevent it from tilting and falling down.
– When transporting over inclines, around cor-
ners, or over thresholds (e.g., through doors
NOTE
or in elevators), make sure that the medical
device does not bump against anything. Operate in sufficiently ventilated rooms or the CO2
– Clear the writing tray and slide it completely measurement may be impaired.
into the device.
– Do not pull the medical device over hoses, ca- NOTE
bles, or other obstacles lying on the floor. The ambient conditions specified in the Technical
– Do not activate the brake while the medical Data must be observed during operation and stor-
device is being moved. age of the device.
– Use only the handles provided to push or pull
the device. NOTE
There may be a restriction of ambient conditions
CAUTION owing to mounted accessories or specifications is-
Health risk sued by the anesthetic agent manufacturer.
Only operate Zeus IE with connected anesthetic
gas scavenging system (AGSS). NOTE
In a closed system, foreign gases can accumulate
in the breathing gas with longer surgery times,
CAUTION
e.g., by degradation of the breathing gas, etc.
Do not use silicone spray! Silicone spray may
cause breathing system valves to stick. Increased concentrations of foreign gases can be
prevented by regularly flushing the breathing sys-
tem.
CAUTION
Do not use DIVA metering modules which have
been dropped or damaged. Any damage may re-
sult in faulty metering.
CAUTION
Risk of tipping over
When transporting the device and all the hinged
arms are swung inward to the center of the device
and locked, ensure that the slope of the floor does
not exceed ±10°.
If the transport position is not observed, the max-
imum slope of the floor is reduced to ±5°.
Application
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Ventilation modes . . . . . . . . . . . . . . . . . . . . . . . 22
Device monitoring. . . . . . . . . . . . . . . . . . . . . . . 22
Patient monitoring (optional). . . . . . . . . . . . . . . 23
Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Gas supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Gas disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Breathing system . . . . . . . . . . . . . . . . . . . . . . . 24
Indications/Contraindications . . . . . . . . . . . . 25
Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Contraindication . . . . . . . . . . . . . . . . . . . . . . . . 25
Intended use
Additionally there are chronological trends of – Inspiratory O2 and N2O concentrations FiO2,
measured values (trends) as well as a log book inN2O
available. – Inspiratory and expiratory CO2 concentrations
inCO2, etCO2
Monitoring
Gas supply
The monitoring of the patient is ensured by
adjustable alarm limits. The plug formats of the Zeus IE anesthesia
workstation are selectable and conform to the
Device monitoring standard formats of the particular country.
– Airway pressure Paw Zeus IE has connections for O2, Air, or N2O backup
– Expiratory minute volume MVe gas cylinders, which must be equipped with
– Apnea pressure reducers and pressure sensors.
– Tidal volume, inspiratory VTi
– Inspiratory anesthetic gas concentration
– Recognition of anesthesia gas mixtures
The anesthetic gas receiving system (AGS) must Interfaces to the patient are the 22 mm diameter
be used as a required accessory. The AGS is sockets on the breathing system for the connection
connected to the correspondingly marked port on of EN 12342-compliant breathing hoses.
Zeus IE.
For breathing systems without rebreathing, e.g.,
Bain systems, an external fresh-gas outlet is
available (optional).
USB ports
The USB ports enable data exchange with Dräger
USB storage media.
Indications/Contraindications
Indications
Contraindication
NOTE
The device administers medical gases such as
O2, N2O, or volatile anesthetic agents. For contra-
indications to the applied medical gases, strictly
observe the instructions for use of the medical
gas.
Overview
Components . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Breathing system view . . . . . . . . . . . . . . . . . . . 29
Connection for oxygen therapy. . . . . . . . . . . . . 29
Connection field for the patient monitoring
system and the IV system (optional) . . . . . . . . 30
Rear with doors open . . . . . . . . . . . . . . . . . . . . 32
Connections to the gas supply block . . . . . . . . 33
Connections on the rear of Zeus IE . . . . . . . . . 34
Connections to the back of the screen . . . . . . . 35
Connections, power supply
and additional sockets . . . . . . . . . . . . . . . . . . . 36
DIVA metering modules . . . . . . . . . . . . . . . . . . 37
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . 38
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Product labels. . . . . . . . . . . . . . . . . . . . . . . . . 45
Components
Front
O
N B
M
C
L
K
D
J
I E
H
F
G
Q 001
L K
211
Connection for oxygen therapy
A O2 flush button O2+
B Mechanical pressure gauge for airway pressure
C Safety rotary knob Anesth. Ventilator Off
D Safety rotary knob for emergency O2 delivery
O2
A
E APL valve
F Self-test adapter
G O2 flow meter for regional anesthesia
O2
H O2 flow meter outlet
I External fresh-gas outlet (ext. FG outlet),
optional
212
J Expiratory port The additional oxygen flow meter (A) delivers a set
flow of pure oxygen, e.g., for oxygen metering
K Breathing bag nozzle
through a nasal cannula. The supplemental oxygen
L CO2 absorber can be used in each ventilation mode:
– Standby
– Zeus IE switched off
as long as the central O2 supply is connected.
Connection field for the patient NIBP Connection for accessories to mea-
monitoring system and the IV system sure non-invasive blood pressure
(optional) (NIBP)
SpO2 Connection for accessories to mea-
SpO2 measurement with SpO2 SmartPod sure oxygen saturation (SpO2) ac-
cording to the technology marked
(Masimo SET oder Nellcor OxiMax)
Sync. Connection for synchronization of a
defibrillator
IV system Connection for Dräger IV system
(IVenus)
HEMOMED1 MULTIMED
Aux/HEMO2 Aux/HEMO3
NIBP SpO2
IV-System Sync.
009
SpO2 measurement without SpO2 SmartPod MULTIMED Connection for accessories to mea-
(optional) sure oxygen saturation (SpO2),
electrocardiography (ECG
HR/PLS), and body temperature
(Temp)
HEMOMED 1 Connection for accessories to mea-
sure invasive blood pressures (IBP)
and cardiac output (C.O.)
Aux/Hemo 2/3 Connection for accessories to mea-
sure more invasive blood pressures
(More IBP) and cardiac output
146
(C.O./ Calculat.)
Connection for anesthesia effect
HEMOMED1 MULTIMED monitoring accessories (Trident
pod, BISx pod)
NIBP Connection for accessories to mea-
sure non-invasive blood pressure
(NIBP)
Aux/HEMO2 Aux/HEMO3 Sync. Connection for synchronization of a
defibrillator
IV system Connection for Dräger IV system
NIBP Sync. (IVenus)
I B
C
H
E D
E E
F
G
173
A B C D E F
084
A Connection for central N2O supply
B Connection for N2O backup gas cylinder
C Connection for central Air supply
D Connection for Air backup gas cylinder
E Connection for central O2 supply
F Connection for O2 backup gas cylinder
The Air outlet and the O2 outlet options are not
shown in this illustration.
Ports relevant to the user: Data transmitted over the MEDIBUS interface
are for information purposes only and are not
A COM 1, RS 232 interface, MEDIBUS intended as the sole basis for diagnostic or
B COM 2, RS-232 interface, MEDIBUS therapeutic decisions. The data accessible via
this interface is not intended for use with a
C COM 3, RS 232 interface distributed alarm system conforming to
D Ethernet 1 and Ethernet 2, hospital network IEC 60601-1-8:2012 (in the context of remote
port, optional monitoring).
CAUTION
Unsupported USB devices
USB devices that are not supported or have too
high a current consumption can impair correct
equipment operation.
A B A
Only use supported USB devices with a maximum
current consumption of 500 mA.
CAUTION
USB devices with their own mains supply
A C
020
USB devices with their own power supply (e.g.,
USB printers) may endanger patients and users. A USB ports
Only USB devices with no power supply of their B LAN ports
own should be connected to the Zeus IE. C RS232 ports
A B A
C C A C
017
A USB ports
A B
025
B LAN ports
A COM 1, COM 2
C RS232 ports
B USB ports
C one USB port on each side
C
B
A
D
170
NOTE
Auxiliary sockets are not supplied by the uninter-
ruptible power supply UPS and are not supplied
with power in the event of a mains power failure!
WARNING
Do not connect HF surgery devices to the aux-
iliary power sockets, as any interference can
impair Zeus IE functionality!
B C
A
063
A DIVA metering module with Quik Fil filling
device or Dräger Fill
B DIVA metering module with desflurane filling
system (Safe-Fill)
C DIVA metering module with safety filling device
Sevoflurane
L D
K
J E
I
Desflurane
F
Isoflurane
G
H
064
D Filling port
E Locking cap
F Release for desflurane lock
G Lever to release the locking bolt
H Locking bolt for safety filling device
I Drain plug (recessed, 2.5 mm hexagon socket)
J Filling level, inspection glass window
Abbreviations
FG DES DES concentration in the fresh gas i/e VA Ratio of inspiratory to expiratory
anesthetic agent concentration
FG ENF ENF concentration in the fresh gas
IBF Integrated bacterial filter
FG flow Fresh-gas flow
IBP Invasively measured blood
FG HAL HAL concentration in the fresh gas pressure
FG ISO ISO concentration in the fresh gas IBW Ideal body weight
FG O2 O2 concentration in the fresh gas ICP Intracranial pressure
FG SEV SEV concentration in the fresh gas ICP M Mean intracranial pressure
FGF Fresh-gas flow IEC International Electrotechnical
FiO2 Inspiratory O2 concentration Commission
Flow Fresh-gas flow in Inspiratory
GP1 D, Local non-specific blood pressure inch Inch, unit of length
GP2 D (General Pressure), diastole inCO2 Inspiratory CO2 concentration
GP 1 corresponds to the 1st
measurement channel inDes Inspiratory desflurane
GP 2 corresponds to the 2nd concentration
measurement channel inDESmax Maximum insp. desflurane
GP1 M, Local non-specific blood pressure limitation
GP2 M (General Pressure), medium inEnf Inspiratory enflurane concentration
pressure InfR 1 Infusion rate 1
GP 1 corresponds to the 1st
measurement channel inHal Inspiratory halothane
GP 2 corresponds to the 2nd concentration
measurement channel inIso Inspiratory isoflurane concentration
GP1 S, Local non-specific blood pressure inISOmax Maximum insp. isoflurane limitation
GP2 S (General Pressure), systole
inN2O Inspiratory N2O concentration in
GP 1 corresponds to the 1st
the breathing air
measurement channel
GP 2 corresponds to the 2nd inSev Inspiratory sevoflurane
measurement channel concentration
HAL Halothane inSEVmax Maximum insp. sevoflurane
limitation
HF surgery High-frequency surgery
insp. Inspiratory
HME filter Heat and moisture exchange filter
Abbreviation Explanation
TOFR TOF ratio
UPS Uninterruptible power supply
USB Universal Serial Bus (computer
interface)
V.tach Ventricular tachycardia
VAC Volts, alternating current
VDC Volts, direct current
Vol. Ctrl. Volume Control AutoFlow
AutoFlow ventilation mode
Vol% Percentage gas component
relative to the total volume
Vol%SL Vol% at a standard atmospheric
pressure of 1013 hPa (sea level)
VT Tidal volume
VTe Expiratory tidal volume
VTi Inspiratory tidal volume
VTmand Mandatory tidal volume
VTspon Spontaneously breathed tidal
volume
xMAC Multiple of MAC
Symbols
DIVA, unlocking
O2
Safety rotary knob O2 for
emergency O2 delivery
Biomed service...
Product labels
max.800mm
Label on right side of door
When transporting the device, move all hinged arms into the transport
position. To do this, swing the hinged arms to the center of the device and
lock them in place.
max.800mm
Observe the weight of the nominal configuration and the total permissible
nom. 185 kg weight, refer to "Technical data".
max. 300 kg
Operating concept
Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Variant 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Variant 2 and Variant 3 . . . . . . . . . . . . . . . . . . . 49
Screen layout . . . . . . . . . . . . . . . . . . . . . . . . . 50
Main menu bar . . . . . . . . . . . . . . . . . . . . . . . . . 51
Buttons in system standby . . . . . . . . . . . . . . . . 52
Color concept . . . . . . . . . . . . . . . . . . . . . . . . . 54
Therapy controls. . . . . . . . . . . . . . . . . . . . . . . . 54
Waveforms and parameters . . . . . . . . . . . . . . . 54
Color coding for anesthetic agents
and medical gases . . . . . . . . . . . . . . . . . . . . . . 54
Screen
CAUTION
Touch and hold the On/Off button for 4 sec. to turn
the screen off. Ventilation is continued. A fresh-
gas flow of 0.5 L/min with 100 % O2 is delivered.
Anesthetic agent metering is switched off. Use a
manual pressure gauge to check ventilation and
ventilate manually if necessary.
Touch the On/Off button again to turn the screen
back on.
Variant 2 and Variant 3 G LED status display for the gas supply from the
backup gas cylinders (O2, Air, N2O)
– Green = pressure sufficient
I – Green, flashing = pressure slightly below
level
– Off = pressure not sufficient
H Alarm pause function
A
Press the button to suppress the audible
alert of all active alarms for 2 minutes. For
additional information, see page 205.
I The alarm bar on the front and back indicates
D alarm conditions:
O2 Air N2O O2 Air N2O
Screen layout
B
C
G
100
B
completed/ operable
C
D
E
Standby
To start, touch "Start..." and confirm.
F
300
A 1 D Therapy bar
B The therapy bar on the main screen contains the
therapy controls for the active Ventilation settings
C and the MixGas settings.
A B
305 C D
301
Using the dialog windows A Button to open the MixGas settings dialog
window. The active setting for the mixed gas is
Dialog windows consist of one or several pages displayed.
which are displayed by touching the corresponding
B Button to open the Ventilation settings dialog
horizontal or vertical tab. Dialog windows contains
window. The active ventilation settings are
elements for operating the device and informing the
displayed.
user on current settings. Dialog windows can be
opened by touching a button in the main menu bar C Therapy controls for the mixed gas setting
or by touching the respective parameter field on the
D Therapy controls for the ventilation settings
screen.
A Dialog window title
B The horizontal tabs indicate primary topics.
Touch the relevant tab to open a page.
C The vertical tabs open secondary topics. Touch
the relevant tab to open a page.
D Touch the button to close the dialog window.
If the button is yellow, the dialog window can
also be closed by pressing the rotary knob.
Color concept
Screen setup
306
420
Dark gray Currently not operable, function
activated
Yellow Selected function, not yet confirmed Color coding for anesthetic agents and
with rotary knob. medical gases
Light green Operable element, function not
The color coding standardized in accordance with
activated.
ISO 5359/ISO 32/ISO 5360 is used for anesthetic
Dark green Operable element, function agents and medical gases.
activated.
The colors for O2, Air, and N2O are adapted to
locally applicable standards.
Set values or operating states
1 Touch the therapy control. The color turns
yellow.
2 Turn the rotary knob to set the value.
3 Press the rotary knob to confirm the value. The
color of the therapy control turns dark green or
bright green.
The following chapters of the Instructions for Use
use a simplified explanation for these instructions:
"Use the rotary knob to set and confirm the value."
– Disable options
System setup
Screen setup
– Screen layout – General
– Waveforms
– Data
– Web access
Sensors/Parameters
– ECG – ECG HR/PLS
– ECG ST
– ECG C.O.
– ST segm.
– More ECG
– SpO2 – SpO2
– NIBP – NIBP
– IBP – Config.
– IBP sites/ Channels
– Cal.
– Wedge
– Temp – Temp
Sensors/Parameters (Cont.)
– BIS/NMT – NMT
– BIS
– Imped. check
Infusion settings
– Infusion 1 – General
– Advanced
– Infusion 2 – General
– Advanced
– Infusion 3 – General
– Advanced
– Infusion 4 – General
– Advanced
Trends/Data
– Overview – Graph. trend
– Tab. trend
Trends/Data (Cont.)
– BIS/NMT – Graph. trend
– Tab. trend
– All EEG values
Alarms
– Alarm info Detailed alarm information
Test details
– Checklist – Supplies/ Safety
– Breathing system
This page was intentionally
– left blank. Y-piece
Suction/
– Test info
– Test history
– Leakage assistant
– System info
Ventilation settings
– Pause Set parameters
MixGas settings
– Fresh-gas control Settings for metering mode
Patient setup
Change patient data
Instructions for
mounting the accessories . . . . . . . . . . . . . . . 84
The battery is automatically recharged if the mains Residual vapors of highly flammable disinfec-
plug has been inserted. tants (e.g., alcohols) and deposits that were
not removed during reprocessing may ignite
when the flow sensor is in use.
– Ensure particle-free cleaning and disinfec-
tion.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor check for
visible damage and soiling, such as resid-
ual mucus, medication aerosols, and parti-
A B cles.
– Replace flow sensors when damaged,
soiled, or not particlefree.
O2 Air N2O O2 Air N2O
050
A B
039
Remove the breathing system, see page 385.
1 Unscrew inspiratory and expiratory connecting
nozzles (A) for breathing hoses.
2 Insert both SpiroLife flow sensors (B).
3 Screw connecting nozzles (A) back on again
(use only blue, non-rotatable nozzles).
Fit the breathing system again and lock it in
place.
Features
CAUTION
Observe the instructions for use for the ceiling
supply unit.
The mains voltage must correspond to the voltage 3 Check whether the circuit breaker (B) is
range indicated on the rating plate. pressed. Pressed is indicated by a white circle
on a black background.
CAUTION
Only press the circuit breaker if the screen has
completely failed
A
The circuit breaker disconnects Zeus IE, except
for the screen, from the battery and from the ex-
ternal power supply.
To disconnect the entire device from the power
supply, unplug the mains plug.
174
1 Plug the IEC line socket (A) into the device. Establishing potential equalization
Secure with the screws.
2 Insert the mains plug into a wall socket. The A
LED with the symbol lights up on the
screen.
WARNING
Danger of electric shock
If the device is connected to a power socket
with incorrect mains voltage or without a pro-
tective ground, persons may be injured and
the device may be damaged.
Only connect the device to power sockets with
015
the permissible mains voltage and a protec- Differences in electrical potential between devices
tive ground. can be reduced by potential equalization.
Potential equalization does not replace the
protective ground connection.
During operation, the potential equalization
connectors must be readily accessible and the
B connection must be able to be disconnected
without the use of tools.
1 Connect the potential equalization cable to the
potential equalization pin (A) on the device.
2 Connect the potential equalization cable to a
potential equalization connector of the hospital
264
A A
015
WARNING WARNING
Risk due to gas supply failure Risk through use of O2 concentrators
All gas supplies (central gas supply, gas cylin- The following may occur:
ders) must be correctly connected since oth- – Deviations between the set value and the
erwise the backup system (gas cylinders) will actual value for the fresh-gas flow and O2
not be available if gas supply fails. concentration in the fresh gas
– Make sure that all compressed gas hoses – Inaccurate measured values for volume
are correctly connected to the rear side of – Inaccurate volume dosage in volume-con-
the device. trolled ventilation modes
– After connecting the gas supply, check for – Uptake of argon in low-flow operation and
correct function. minimum-flow operation
– Even when the anesthesia machine is con-
Do not use O2 concentrators.
nected to the central gas supply, the gas
cylinders should remain at the device with
valves closed as backup. NOTE
A failure of the gas supply can result in failure of
WARNING the connected devices.
Risk due to poor quality gas
Devices connected to the gas supply may be-
come damaged due to poor quality gas. The
use of non-medical gas may lead to gas com-
positions that impair the functional integrity of
the device.
– Only use medical gases.
– Take note of the national and international
standards for the use of medical gases.
WARNING
Risk through use of prohibited gas supply
If using an oxygen supply that delivers less
than 99.5 % oxygen, this may lead to errors in
the composition of the gas.
Only use oxygen supplies that deliver at least
99.5 % oxygen.
Connecting the compressed gas hose Connecting backup gas cylinders for O2,
Air, or N2O
1 Screw on the compressed-gas hoses of the
central gas supply (CS) for O2, Air, and N2O to Caution when handling compressed gas cylinders:
the front connections of the gas supply block.
The three ports on the back are intended for the WARNING
backup gas cylinders. Risk of explosion
One Air outlet and one O2 outlet are optionally Do not oil or grease the O2 cylinder valves or
available. O2 pressure reducers and do not handle them
2 Connect the compressed-gas hoses to the gas with greasy fingers.
supply block.
WARNING
3 Insert the plugs of the compressed-gas hoses
into the wall supply valves. Risk of explosion
Only use pressure reducers listed in the List
of Accessories Follow the instructions for use
of the pressure reducers used.
CAUTION
Risk of injury
A
O2 Air N2O O2 Air N2O
CAUTION
Risk of explosion
If the cylinder valves are open, gas can be re-
moved unnoticeably from the gas cylinders (in-
stead of from the central gas supply) during CS
operation. Always keep the cylinder valves closed
during CS operation.
179
B C
180
CAUTION
Increased occupational exposure
Do not deliver any anesthetic agent if the AGS is
not connected!
CAUTION
Immediately after replacing the absorber during
operation, proceed directly with a check for leak-
tightness by performing manual ventilation and
conduct a leak test at the next opportunity to elim-
inate possible leaks.
Filling the reusable CO2 absorber 6 Insert the CO2 absorber into the breathing
system from below and tilt it to the right until it
Recommendation: Use only Drägersorb 800 Plus engages.
or Drägersorb FREE.
A A
B 005
018
Pulling out the used absorber Installing the new CO2 absorber
1 Shake the disposable CO2 absorber before
inserting, e.g., by inverting it several times, to
A loosen the soda lime.
2 Remove the seal from the new disposable CO2
absorber.
3 Slide the new disposable CO2 absorber (C) into
the holder (B) as far as possible until you hear it
C snap into place in the CLIC adapter.
068
B
011
NOTE
Depending on the date of manufacture, it is possi-
ble that the breathing system does not have
threaded bushings for mounting the flexible
breathing bag holder. In this case, assembly is
only possible if a new breathing system with
threaded bushings is used instead of the existing
breathing system.
WARNING WARNING
Risk of burns to the patient's airways/mucous Risk due to particles and dust
membranes
In order to protect the patient from particles
A defect in the ventilator or in the DIVA meter- and dust, a filter must be used between the in-
ing module can lead to increased breathing spiratory limb of the breathing system and the
gas temperatures. patient.
To prevent the risk of burns, the length of the Use a Y-piece filter or a filter at the inspiratory
breathing hose may not be less than 110 cm. port.
WARNING WARNING
Risk of patient injury Risk of infection
When using near-patient filters do not scav- If no microbial filter is used at the Y-piece or at
enge at the patient end! the expiratory port, contamination of the
breathing system with infectious agents may
In the event of stenosis or increased filter re-
result.
sistance the patient may be injured by vacu-
um. In this case, hygienically reprocess the
breathing system after each patient.
WARNING
Risk due to leakage in the coaxial breathing NOTE
circuit Zeus IE is manufactured without natural rubber
latex.
Leakage in the inner hose of a coaxial breath-
ing circuit may result in CO2 rebreathing or in- To minimize the risk of exposure to latex, use
sufficient gas exchange. The device can de- breathing bags and breathing hoses manufactured
tect leakages only if a separate test is per- without natural rubber latex.
formed with a coaxial test adapter.
– Use a coaxial test adapter to check the in-
ner hose for leaks, see page 121. After that,
perform a leakage test for the entire
breathing circuit.
– Monitor the gas concentrations measured
during ventilation.
A G
A
C
D
149
F F E
6 Connect the breathing bag (G) with the
connecting nozzle to the breathing bag hose
and attach to the angled nozzle. Insert the
breathing bag downward, hanging in the
B direction of the floor, into the mount.
176
WARNING
Risk of burns when using HF surgery equip-
ment
Do not use any antistatic or conductive
breathing hoses!
NOTE
Only scavenge from the near-patient filter, other-
wise the measured values may be incorrect.
A
A
G DE
F C
F
272
B 1 Connect Y-piece with filter (G) to the self-test
adapter firmly.
267
H
K
L
K J
I
280
insp.
insp.
exsp.
exsp.
Or
Connections on the side for the
Filter at inspiratory port and expiratory port, connection for sample line support the CO2
sample line at Y-piece: measurement and help to flush the
dead space in the Y-piece and the
insp. hose adapter.
If it is not possible to use the filter on
the expiratory port (e.g., in the event of
intrinsic PEEP due to air trapping),
reprocess the device after this patient,
exsp.
see "Cleaning, disinfection, and
sterilization".
An intrinsic PEEP due to air trapping can be prevented by the
following measures:
– Filters on the inspiratory port and the expiratory port instead
of on the Y-piece
– No expiratory filter
In this case, reprocess the device after this patient, see
"Cleaning, disinfection, and sterilization".
WARNING
Risk of tipping and risk of injury
If the maximum permissible weight is exceed-
ed or if monitors and other auxiliary devices
are placed on the medical device, the device
can fall. Especially if the medical device is
rolled over door thresholds and similar obsta-
cles.
Before moving the anesthesia machine, re-
move all the devices that are placed on it.
1) At the maximum specified distance to the upper housing cover (see picture)
The device can be operated with one or two DIVA Inserting and removing the DIVA
metering modules. The selection of the slot is metering module
arbitrary.
Fill the DIVA metering module with the
appropriate anesthetic agent.
– Filling can be done with the DIVA metering
module plugged in or outside of the device.
– Filling is also possible during anesthesia.
– Before the internal self test of the device, the
DIVA metering modules must be filled and
connected in order to ensure a complete
test. A A
WARNING
Storing the DIVA metering module
– Only when empty (except if supplied for
imminent clinical use)
040
– In well-ventilated rooms
The DIVA metering module can be exchanged in all
– Not over 40 °C
operating states (switched off, on standby or during
Empty the DIVA metering module for regular anesthesia).
maintenance and for transport!
Plugging in the DIVA metering module
Insert the DIVA metering module into the slot
until it can be felt and heard to have engaged.
Removing the DIVA metering module Fill the DIVA metering modules
Keep the unlock button (A) pressed and remove
the DIVA metering module. CAUTION
Health risk
NOTE
Pressing the unlock button during operation re- Only fill DIVA metering modules in areas with ad-
sults in immediate shutdown of the DIVA metering equate ventilation.
module! This can lead to a rapid drop in anesthetic Do not spill any anesthetic agents and do not in-
agent concentration in the breathing gas. hale anesthetic vapor.
CAUTION CAUTION
Risk of injury Health risk
Place DIVA metering modules only on firm, hori- The DIVA metering module can be filled too full.
zontal surfaces. The DIVA metering module can Anesthetic agent vapors and anesthetic agent
be damaged when dropped. may escape.
Keep an eye on the sight glass. Fill the DIVA me-
Sight glasses tering module to the maximum mark at most.
NOTE
Observe the expiration date and storage tempera-
ture of the anesthetic agent.
The DIVA metering modules have a sight glass with Anesthetic agents from different manufacturers
marks for minimum and maximum filling levels. A with different brand names can be metered in the
third mark in the lower section indicates that an DIVA metering module from a technical viewpoint
entire anesthetic agent bottle (250 mL) can be individually or mixed if they are equal in
refilled. composition as well as their physical and chemical
features and if they are approved for
pharmaceutical use.
Filling devices
The following filling devices are possible for the
different anesthetic agents:
WARNING
Transportation in filled state
Risk of patient injury
– Only within the scope of regular clinical
Do not use anesthetic agents that were emp- operation, not for storage and dispatch
tied from a DIVA metering module, but dispose
of in a proper manner. Otherwise, the anes- – Only with ambient conditions in accordance
thetic agents may be contaminated, mixed with the technical data
with others or used by mistake. – The DIVA metering module may remain filled
provided operating interruptions do not exceed
CAUTION 6 months; otherwise, empty the DIVA metering
Increased occupational exposure module.
When emptying the DIVA metering module, do The anesthesia equipment can be moved at the
not fill the anesthetic agent bottle up to the brim! place of work when the DIVA metering module is
This can cause anesthetic agent to be released plugged in and locked.
into the atmosphere.
CAUTION
Health risk
Only fill the DIVA metering modules at tempera-
tures of up to 40 °C, otherwise anesthetic agent
can escape.
D
A
B C
B
067
1 Only use anesthetic agent bottle with
053
CAUTION
332
061
5 Screw the locking cap (E) off the filling device
slowly so that any pressure which has built up Finishing the filling process
in the DIVA metering module can escape 1 Reduce the pressure in the anesthetic agent
slowly. bottle and slowly pull out the anesthetic agent
6 Introduce the anesthetic agent bottle with the bottle.
filling adapter (F) into the filling opening. When 2 Check the filling level in the sight glass (H) – the
doing so, turn the anesthetic agent bottle DIVA metering module must hang or stand
clockwise until the coded area has engaged in vertically during the check.
the filling opening.
3 Screw on the locking cap (J).
CAUTION Screw the Dräger Fill filling adapter off the
Health risk anesthetic agent bottle and
If turned in the wrong direction, the screw-on fill- 4 use the screw cap (K) to close the anesthetic
ing adapter may come loose from the anesthetic agent bottle.
agent bottle and anesthetic agent may escape.
CAUTION
Do not turn the anesthetic agent bottle counter-
clockwise. Increased occupational exposure
Firmly screw on the locking cap. Otherwise, anes-
thetic agent can escape.
NOTE
G Anesthetic agent bottles which have a Dräger Fill
filling adapter fitted may be kept in storage.
H I K
J
062
B
C
A
263
WARNING
Risk of patient injury
Do not use anesthetic agents that were emp-
tied from a DIVA metering module, but dispose
of in a proper manner. Otherwise, the anes-
thetic agents may be contaminated, mixed
with others or used by mistake.
MixGas settings
A
A
B
C
013
332
Safe-T-Seal filling system with permanently Touch the Ventilate reservoir (A) button.
installed adapter The DIVA metering module is vented for approx.
The Safe-T-Seal filling system consists of the 30 seconds. No anesthetic agent vapor escapes
following components: into the environment during filling.
B
030
E
3 Firmly attach the screw-on Safe-T-Seal filling
adapter (F) to the anesthetic agent bottle.
D
034
K
H
M
G
037
J L
041
4 Remove the locking cap (G) from the filling
6 Press the anesthetic agent bottle (I) into the
device slowly so that any pressure buildup in
filling opening as far as it will go and keep it
the vapor can escape slowly.
pressed while still holding the anesthetic agent
5 Insert the anesthetic agent bottle with the filling bottle firmly. Avoid the use of excessive force
adapter (F) into the filling opening (H). When and jamming of the anesthetic agent bottle.
doing so, turn the anesthetic agent bottle
7 Watch the filling level in the sight glass (J). The
clockwise until the coded area has engaged in
supply will be interrupted automatically when
the filling opening.
the maximum mark is reached.
CAUTION 8 Fill the DIVA metering module to the maximum
Health risk mark at most.
If turned in the wrong direction, the screw-on fill- 9 If the DIVA metering module is filled beyond the
ing adapter may come loose from the anesthetic maximum mark, the anesthetic agent will flow
agent bottle and anesthetic agent may escape. out through an overflow hole (K).
Do not turn the anesthetic agent bottle counter-
clockwise.
CAUTION
Increased occupational exposure.
Firmly screw on the locking cap. Otherwise, anes-
thetic agent can escape.
NOTE
Anesthetic agent bottles which have a Dräger Fill
filling adapter fitted may be kept in storage.
B MixGas settings
A
A
C
051
for isoflurane
332
The safety filling system consists of the following
components: 1 Press the Ventilate reservoir (A) button before
A Anesthetic-specific filling device on the DIVA filling.
metering module The DIVA metering module is vented for
B Anesthetic-specific Dräger filling adapter approx. 30 seconds. No anesthetic agent vapor
escapes into the environment during filling.
C Anesthetic-specific collar and thread on the
neck of the bottle
E
Recommendations
– Use only anesthetic agent bottles with an
anesthetic-specific collar on the bottle neck. D
– Use Dräger filling devices only.
– Use filling adapters with a non-return valve only.
– Use undamaged filling adapters and anesthetic
052
CAUTION
Increased occupational exposure
I
With a leaky connection between the bottle adapt-
er and anesthetic agent bottle, anesthetic agent
vapor can escape.
057
9 Swivel the anesthetic agent bottle (I) slowly to
an overhead position and keep it in this position.
10 Watch the filling level (J) in the sight glass.
The supply will be interrupted automatically
055
G
056
A NOTE
Do not store any anesthetic agent bottles with the
B filling adapter unscrewed! Anesthetic agent will be
4
D released!
058
2 Check the filling level with the sight glass (B). NOTE
The DIVA metering module must be in an Close the anesthetic agent bottle, even if it has
upright position. been emptied. Allow remaining anesthetic agent in
3 Swivel the lever (C). the filling adapter and anesthetic agent bottle to
evaporate under an extractor hood, otherwise an-
4 Pull out the filling adapter (D). esthetic agent vapor will be released.
E
059
A
265
A
B
053
060
metering module
2 Unscrew the cap (B) from the bottle adapter.
B Anesthetic-specific adapter on the anesthetic
agent bottle. The bottle adapter must be positioned firmly
and tightly on the anesthetic agent bottle and
must be undamaged.
Filling during operation
CAUTION
Increased occupational exposure
MixGas settings
With a leaky connection between the bottle adapt-
er and anesthetic agent bottles, anesthetic agent
A vapor can escape.
332
D
A
C
B
061
062
slowly so that any pressure which has built up in
1 Reduce the pressure in the anesthetic agent
the DIVA metering module can escape slowly.
bottle (A) and slowly pull out the anesthetic
4 Insert the anesthetic agent bottle with the bars agent bottle.
into the respective grooves (D) of the filling
2 Check the filling level in the sight glass (B) – the
opening. Only use anesthetic agent bottles with
DIVA metering module must stand vertically
suitable bars!
during the check.
Press the anesthetic agent bottle up to the stop
into the filling opening and keep it pressed. CAUTION
Avoid excessive force and tilting of the Increased occupational exposure
anesthetic agent bottle! Firmly screw on the locking cap. Otherwise, anes-
Watch the filling level in the sight glass. thetic agent can escape.
The supply will be interrupted automatically Screw the cap onto the bottle adapter.
when the maximum mark is reached.
Always keep the anesthetic agent bottle closed!
A
263
A B
C
A
B
065
C
The Safe-Fil filling system consists of the following
components:
A Anesthetic-specific filling device on the DIVA
metering module
054
B Unlock button on the DIVA metering module 1 Touch the unlock button (A).
C Anesthetic-specific adapter on the anesthetic 2 Pull out the locking cover (B) of the DIVA
agent bottle metering module.
3 Open the desflurane anesthetic agent bottle
and insert the bottle neck (C) into the filling
opening until it engages.
4 Keep the anesthetic agent bottle pressed.
The anesthetic agent bottle is equipped with an
internal valve to prevent spilling.
5 Watch the filling level in the sight glass. The
DIVA metering module must be in a vertical
position when checking.
NOTE
Increased occupational exposure
Fill the DIVA metering module to the maximum
mark only!
If filled beyond the maximum mark, excess desflu-
rane may splash out of the filling device when re-
moving the anesthetic agent bottle.
066
1 Attach an empty anesthetic agent bottle.
2 Turn the DIVA metering module with the
anesthetic agent bottle attached upside down
and push.
3 After emptying in this position, touch the unlock
button and remove the anesthetic agent bottle.
4 Turn the DIVA metering module upright and
close it with the locking cover.
5 Close the anesthetic agent bottle.
Startup
Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Supply & safety. . . . . . . . . . . . . . . . . . . . . . . . . 113
Breathing system . . . . . . . . . . . . . . . . . . . . . . . 113
Suction & Y-piece . . . . . . . . . . . . . . . . . . . . . . . 113
Pretest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Device check
179
Immediately change over to cylinder supply in the
event of a central gas supply failure:
1 Open the cylinder valves (A) slowly.
A
O2 Air N2O O2 Air N2O
050
NOTE
If the valves are open during central supply opera-
tion: Risk of gas being drawn from the backup gas
cylinders!
A A
038
The O2 flush is provided for flushing and filling the
breathing system and the breathing bag with
oxygen quickly (approx. 35 L/min), bypassing
anesthetic agent delivery.
047
There is a knob (A) on each side of the breathing 1 It is complete and engaged. The hoses are
system with identical functions. firmly connected.
1 Manually close the filter on the Y-piece or 2 New soda lime, not discolored violet.
connect the test lung.
2 Press the O2+ button (A). CAUTION
Risk of device failure
3 The breathing bag fills; the filling process can
be heard. To avoid a backflow of condensed water into the
breathing system, the hoses between the patient
and the breathing system must have sufficient
Checking the emergency O2 delivery slack.
The condensate in the hose must be continuously
A monitored and, if necessary, disposed of or dis-
charged to the hose water traps. Standing water
impairs device operation!
042
checked.
1 Transfer hose (A) is connected to the nozzle on
the back of the device and to the receiving
system (AGS).
2 The scavenging hose is connected to the
receiving system, the plug (B) is in the Dräger
wall socket, whose indicator is green.
3 The float in the AGS hangs between the two
markings.
Follow the Instructions for Use of the anesthetic
gas receiving systems (AGS) and the
anesthetic gas scavenging system (AGSS).
O2
212
Switching on Zeus IE
Self test
Test details
A
Self test
O2 Air N2O O2 Air N2O
050
310
Manual ventilation and emergency O2 delivery are
possible at all times. The device switches to the self test mode
automatically. The Test details > Checklist dialog
Anesthetic agent is not delivered yet! window is opened.
WARNING
Perform the self test once a day in order to
check the operational reliability of the device
and hence minimize the risk of exposing the
patient to a hazard!
618
Checklist
Use the checklist (A) to make sure the device is Breathing system
ready for operation.
The checklist consists of the following test steps: 1 Touch the Breathing system tab (C).
– Supplies & Safety (B) 2 Perform the following test steps:
– Breathing system (C)
– Suction unit & Y-piece (D) – Anesthetic gas scavenging system
– Breathing system
Self test – Manual breathing bag
Test details – APL valve
– Soda lime
A B
C
D
Suction & Y-piece
– Y-piece
– Sample line
CAUTION
– Watertraps
The checklist is part of the self test and must be
applied each time before operation. Otherwise The checklist can also be called up during
device operation may be impaired. operation.
1 Switch to the system standby screen.
Call up the test steps one after the other and
work through the checklist. 2 Touch the Device test... button.
The Test details dialog window appears.
Self test
WARNING The self test can only be started from the system
standby screen.
Risk of patient injury
The self test must never be performed when a
patient is connected!
CAUTION
Increased occupational exposure
When filling during the warm-up phase of the
DIVA metering module, anesthetic agent may es-
cape. A
Only fill the DIVA metering module after the warm-
up phase.
302
1 Touch the Device test... button (A).
A The Test details dialog window appears.
Self test
Test details
Self test
C D
048
2 Connect sample line to the filter. 2 Touch the Self test (B) button.
3 Set the APL valve to 20 mbar. 3 Touch the Name of tester button (C).
4 Switch off the manual O2 supply. 4 Enter a new name or select a name from an
existing list of the last six testers. Names can
also be deleted from the list.
5 The Start (D) button automatically turns yellow.
Confirm with rotary knob.
The self test starts. The Start button (D) is dark After the pretest, the following options are available
green. to continue the self test:
Self test – Pretest passed:
Test details
– Self test continues automatically.
– At the end of the self test, the current test
result is displayed.
– Pretest failed:
Call up the cause of the error via the Test
info tab.
Self test
Perform the pretest again or start the self
E test directly to obtain accurate error
analysis.
A failed pretest does not change the last self test
436
result.
The total duration of the test is approx. 8 minutes.
Two progress indicators (E) show the elapsed time The pretest can be activated as a separate test in
in %. preparation for the automatic self test,
see page 118.
The top bar displays the Pretest. The pretest takes
approx. 1 minute. The device is tested for the The bottom progress bar displays the progress of
following most frequent errors: the self test.
– Large leak
A tone sequence is played from the primary
– Y-piece not connected properly
loudspeaker after completion of the pretest. Next 2
– APL valve setting
tone sequences, each with 3 constant tones, are
– Central supply status of O2 and Air
played from the backup loudspeaker.
– Flow sensors inoperable
If these tones cannot be heard, service personnel
If a problem is determined in the pretest,
must be contacted.
information about the cause and how to remedy the
problem appears above the progress bar.
When the problem has been remedied, touch
the Repeat pretest button.
Self test
Test details
F
I
Self test
G
H
501
437
A
B
500
Pretest
Pretest
Test details
Pretest
B
C
435
ISO SEV
C D
316
319
The test result is displayed in color: All faults from the self test or leak test are listed.
Green test successful The user receives the following information about
Red test unsuccessful troubleshooting:
– Fault (C)
Gray not tested – Cause and remedy (D)
– graphic presentation of the system components
In addition to the image, important system data is
displayed:
– Central supply pressure and cylinder pressure
– Leaks and compliance with the time of the last
measurement
– Battery capacity
– Anesthetic agent quantity in the DIVA metering
module
Touch the Auto self test tab (A). Calling up the test history:
316
general basic settings, see page 360.
The results of previous tests are displayed with the
The default values can be changed for the next test: following information:
– Time-controlled self test starts at (B)
– Date and Time
– Time-controlled self test (C) On/Off
– Test
The new values apply to the next automatic self – Name of tester
test. After the next automatic self test is completed, – Number of test loops
the values are reset to the default values. – Result
Factory setting: Off If a test did not detect any faults but is incomplete,
the result is indicated with canceled.
When activating On, the following messages
appear:
– Auto self test is enabled. Conduct Pretest.
Check start time.
When activating Off, the following message
appears:
– Auto self test is disabled.
Leakage assistant
A G
B
C
107
315
Prerequisites:
Starting the leakage assistant:
– The emergency O2 delivery is not active.
Touch the Leakage assistant button (A). The
Perform test:
Start button (B) turns dark green automatically.
1 Connect the connector (D) to the inspiratory
The leakage assistant starts.
port.
The test pressure is 25 to 30 mbar and is indicated
2 Connect the coaxial test adapter (E) to the
in the form of dotted lines on the pressure display.
expiratory port.
If the part of the system closest to the patient is the
3 Remove the elbow (F) from the hose. Connect
cause of the leak, a leak value above 100 mL/min
the connector (G) to the coaxial test adapter
is displayed in red. For other causes, a leak value
(E).
above 500 mL/min is displayed in red.
4 Touch the Leakage assistant button.
A message indicates where the cause of the leak
– Wait until the leakage value is stable.
can be found and provides information about how
to remedy the problem. If the leakage value exceeds 500 mL/min,
use another breathing circuit.
Leaving the leakage assistant:
5 To return to the selection of the available tests,
Touch the Cancel button (C).
touch the Cancel button.
6 Connect the breathing circuit completely.
Testing a coaxial breathing hose for leakage
– Remove the coaxial test adapter.
The leakage assistant can be used to find leakage – Attach the elbow again.
in the inner hose of coaxial breathing circuits. When – Connect the connector (H) (expiration) to
Dräger coaxial breathing circuits are used, the the expiratory port.
inner hose is connected directly to the expiratory – Perform the system test or leakage test.
port using a test adapter for this purpose.
Test details
A
Absorber Blower Flow sensor, insp.
Bag
Pressure sensor
Leak test 2
Leak test 1
246
316
Leaks of up to 100 mL/min are tolerated.
– The leak value is relevant to automatic
ventilation. Service
Leak test 2: An access code is required to display the following
– Leak test in the system as a whole. Leaks of up service-related information:
to 500 mL/min are tolerated. – System info
– The leak value is relevant to manual and – Info log
spontaneous breathing. – Service info
– More Info
– System data
– DIVA data
– GMZ Data
– Access codes
– Apollo Software.
320
Startup in emergencies
WARNING
Use only in urgent cases if there is no time for
the self test!
B
If a self test is not completed, malfunctions
can go undiscovered!
Pay special attention during operation!
A
071
Range: 0 to 12 L/min
1 Touch the button (A) on the screen.
3 Start manual ventilation.
Wait for the software to launch and for electronics
test to end. The device only performs a minimum
test.
NOTE
During this phase, no anesthetic gas delivery is
possible!
Test details
MAN/SPON. F
Self test
323
C Touch the MAN/SPON field (F).
316
Startup dialog
The leak and compliance test is not performed.
The levels of accuracy (e.g., for flow measure-
ment) indicated under "Technical data" cannot be
guaranteed.
WARNING
In order to ensure device features in an emer-
gency operation, check APL valve and breath-
D ing system operation by ventilating manually
in MAN/SPON ventilation mode.
312
Operation
3
Operation
Start settings
G F E
A C
502
The case data and the associated time stamps are
displayed in the list (B). Information for the export is
displayed in field (C), e.g., USB stick is ready.
Prerequisite: USB storage medium is installed and 2 Select a case with the rotary knob or with the
is connected to the USB interface. Detailed arrow keys (D).
information about connecting and installing can be
The patient data for the selected case are
found in the instructions for use.
displayed in field (E). This line may be empty if
The data are stored in the "Draeger\ExportData" there are no patient data available for the case.
directory. Each exported case is contained in a
3 Touch the Export selected button (F).
separate compressed archive in .tar.gz format.
Each archive contains files with data in csv format,
depending on the options that have been enabled. Exporting all case data
The archive name includes the following
Touch the Export all button (G).
information:
– Case ID
– Date and time of the start of the case
472
Alternative: When connected to an Infinity network, it is
possible to transfer patient data (trend data,
Touch the Pause button (L) and confirm with the demographic data).
rotary knob to close the start dialog window and
to switch to the Pause mode. Demographic data is sent to Zeus IE directly.
Trend data is only available for subsequent transfer
Changing IV data (optional) (to monitors or other Zeus IE systems) or for
printout (via Infinity network).
Standby
The transferred patient is automatically always a
New patient.... Zeus IE suggests the most suitable
profile.
Startup dialog
1 Touch the Patient transfer button (A). The
button turns dark green.
The following information is displayed (B):
– Bed
M – Name
– Care unit (only with Multiview workstation in the
Infinity network).
2 The selected bed is shown in yellow. Confirm
with rotary knob.
312
Touch the More IV settings button (M). All buttons are gray and the Transferring patient
The 4 blocks of the IV configuration containing all data, please wait! message is displayed during the
relevant pump data are displayed. Drug data and patient transfer. After the transfer, the patient data
assignments can be changed. is updated and the page closes.
Open the Patient setup dialog window during Setting the Gender (G):
operation:
Touch the Female or Male button.
Touch the field for patient data (A).
Print case (I) (optional):
Patient setup Touch the Graphic or Tab. button.
The graphic or tabular trends of the entire patient
B E are printed.
C F
G
3
D H
I
468
Dräger 20 6 100
Ped. 1
Dräger 40 12 150
Ped. 2
Dräger 80 60 180
Adult 1
Dräger 100 60 180
Adult 2
WARNING
By selecting a patient category, values for pa-
tient weight and age, which have an effect on
patient and ventilation settings, are preselect-
ed.
The selected patient category should always
match the patient.
WARNING
When acknowledging the settings selected,
the user accepts responsibility for their suit-
ability.
SmartPilot View* is only available in the adult Integrated SmartPilot View is optionally displayed
(Adult) patient category. or hidden in the left-hand area (B) of the screen.
1:35
The view can be saved in a profile (Profiles).
Detailed information can be found in the SmartPilot
B View instructions for use.
A
503
The Smart Ventilation Control software supports Detailed information can be found in the Smart
the user during anesthesia from intubation to Ventilation Control instructions for use (9053869,
extubation. 2nd edition and higher).
Gas delivery
WARNING
Risk due to residual concentrations of anes-
thetic agent
Volatile anesthetic agent may trigger malig-
nant hyperthermia. As a result, the patient
could be put at risk.
– For patients suspected of malignant hyper-
thermia: Do not use any volatile anesthetic A
agent or devices with residual concentra-
329
Fresh-gas control mode When selecting the therapy control for the
anesthetic gas concentration for the first time, the
initial value is preset automatically, see page 376.
MixGas settings
Button (G) opens the Ventilation settings
A dialog window
Button (H) opens the Alarms dialog window
Fresh-gas control
J Volume Control - CMV -AF
K
B C I
338
D E F G After closing the MixGas settings dialog window,
the most important therapy controls (I) are shown in
H 330
4 The color changes to yellow, confirm with rotary Use the rotary knob to set and confirm the
knob. value.
Default settings:
Parameter Setting range Default settings:
Neo Ped1 Ped2, Adult1,
Adult2
Carrier gas Air or N2O N2O N2O N2O
Fresh-gas flow 0 to 18 L/min 2 L/min 4 L/min 6 L/min
Fresh-gas flow for external 3 to 18 L/min 6 L/min 6 L/min 6 L/min
fresh-gas outlet
With Air as the carrier gas: 21 to 100 Vol% 90 Vol% 100 Vol% 100 Vol%
FG O2
With N2O as the carrier gas 25 to 100 Vol% 90 Vol% 100 Vol% 100 Vol%
FG O2
FG ISO 0 to 5 Vol% 0 Vol% 0 Vol% 0 Vol%
FG SEV 0 to 8 Vol% 0 Vol% 0 Vol% 0 Vol%
FG DES 0 to 18 Vol% 0 Vol% 0 Vol% 0 Vol%
K
M
L
N 338
Default settings:
Parameter Setting range Default settings:
Neo Ped1 Ped2, Adult1,
Adult2
Carrier gas 0 to 100 Vol% Air N2O N2O N2O
1)
Air or N2O 0 to 75 Vol% N2O
FiO2 25 to 100 Vol% 100 Vol% 100 Vol% 100 Vol%
MinFG flow ; 0.25 to 18 L/min
Flush flow 3 to 18 L/min 4 L/min 6 L/min 6 L/min
etISO 0 to 5 Vol% 0 Vol% 0 Vol% 0 Vol%
inISOmax 0 to 5 Vol% 4 Vol% 4 Vol% 4 Vol%
etSEV 0 to 8 Vol% 0 Vol% 0 Vol% 0 Vol%
inSEVmax 0 to 8 Vol% 8 Vol% 8 Vol% 8 Vol%
etDES 0 to 18 Vol% 0 Vol% 0 Vol% 0 Vol%
inDESmax 0 to 18 Vol% 12 Vol% 12 Vol% 18 Vol%
1) A nitrous oxide-free device system can be configured under Basic settings.
Uptake mode
Ventilation
Ventilation settings
A
C
329
335
Touch the (A) button. Setting ventilation parameters:
The following ventilation modes are available: 1 Touch the corresponding therapy control.
– Pause 2 Use the rotary knob to set and confirm the
– MAN/SPON value.
– Vol. Ctrl. AutoFlow
3 Touch the button to close the dialog window.
– Volume Control
– Pressure Control Direct switching to the MixGas sett. dialog window
– Pressure Support (C) and Alarms dialog window (D) is possible from
– Smart Ventilation Control (optional) all ventilation modes.
– External FG outlet (optional)
Ventilation settings
B
334
337
The top progress indicator shows the current
inspiratory and expiratory tidal volume VT. The
expiratory tidal volume is also displayed
numerically. The progress indicator tracks the
inspiratory and expiratory tidal volume VT. At the
end of inspiration, the tidal volume delivered is
represented by a bar. The leak in tidal volume is
displayed at the end of the expiration phase.
The bottom progress indicator shows the measured
336
After closing the Ventilation settings dialog 1 Touch the therapy control (A).
window, the most important therapy controls are 2 Use the rotary knob to set and confirm the
displayed in the therapy bar of the main page and value.
can be set there directly.
3 Touch screen in the area of the progress
MAN/SPON B indicator.
600
1200 On the progress indicator, the individual breaths
C 5.2
45 s
are separated by segments. After 60 seconds, the
volumeter will stop automatically. The measured
337
The (B) button shows the last selected ventilation values are displayed for 4 minutes and then
mode or, initially, the ventilation mode preselected cleared.
in the basic settings (see page 376). If the volumeter is pressed again before the
Switching to the previous ventilation mode: 60 seconds have elapsed, the volumeter will stop
and the values will be cleared. If pressed again, the
Touch the (B) button and confirm with rotary volumeter is restarted.
knob.
The volumeter function (C) is used for monitoring
and evaluating ventilation with spontaneous
breathing and manual ventilation in the
MAN/SPON ventilation mode.
10 20 30
002
012
Turn the valve head of the APL valve left as far
Set the valve head to the desired maximum
as it will go.
airway pressure.
The dots coincide and the valve head is lifted.
Settings between the grid marks are also possible.
The pressure limitation is cancelled, the valve is
CAUTION open for free spontaneous breathing and the set
Risk of patient injury CPAP level is used.
Even with automatic ventilation, the APL valve
must be adjusted to a pressure that is safe for the
patient!
Quick release
003
PEEP
t
A B C D
Trigger window for insp. and
339
Ti
exp. synchronization
1 Adjust volume-controlled ventilation with autoflow,
RR
Flow synchronization, synchronization/Psupp:
A Maximum airway pressure Pmax
t
B Tidal volume VT
Fixed mandatory minute volume MV adjusted with E Switching Sync. (synchronization) on and off
tidal volume VT and respiratory rate RR. Switching this on activates synchronization
For patients with and without spontaneous and, if set, activation of the set pressure
breathing. support.
The patient can breathe spontaneously at any time F Flow trigger threshold for Trigger
(free deep breathing ability). Spontaneous (synchronization of mandatory breaths)
breathing can be supported with Pressure An exceeded trigger threshold is represented in
Support. the pressure curve by triangles.
G Pressure support ∆Psupp
WARNING
Risk of patient injury H Inspiratory time Ti or I:E
Do not use Volume Control AutoFlow ventila- In the configuration (see page 375), specify
tion mode with intrinsic PEEP (alarm mes- whether Ti or the I:E ratio is set.
sage: Exp. time too short). Expiratory flow
does not return to the base line. This can I Pressure rise time Slope
cause an excessive pressure rise in ventila-
tion pressure.
CAUTION
The administered volume may be lower than the
set volume if there is a leak!
Paw
% Tplat
E F
Pmax
Pplat
(configu- PEEP
rable)
Ti Te
t
A B C D
340
1
RR
Flow Adjust ventilation pattern with ventilation
Insp. Flow parameters:
A Maximum airway pressure Pmax
t B Tidal volume VT
C Respiratory rate RR
259
CAUTION
The administered volume may be lower than the
set volume if there is a leak!
Pinsp fast rise time slow rise time C Positive end-expiratory pressure PEEP
Ventilation settings
D E F G H
A B C
341
PEEP modifications
Ventilation settings
J I
342
Time [s]
B Positive end-expiratory pressure PEEP
Slope
1/RRmin
C Flow trigger threshold for Trigger
1/RRmin (synchronization of mandatory breaths)
Flow An exceeded trigger threshold is represented in
the pressure curve by triangles.
D Pressure support ∆Psupp
Time [s]
E Pressure rise time Slope
Flow trigger Flow trigger No trigger
260
A B
343
Set point Vol. Ctrl. AutoFlow / Pressure Control Pressure Support MAN/SPON
Volume Control
VT VT (measured) ––– ––– –––
Pinsp ––– Pplat ––– –––
Ti X X ––– –––
Te X X ––– –––
PEEP X X ––– –––
∆Psupp X X X –––
1)
Slope X X X –––
CPAP ––– ––– X/PEEP X
1)
Trigger X X X –––
1) Only applies to Vol. Ctrl. AutoFlow.
CAUTION
Risk of gas contamination
In A-cone operation, the sample gas withdrawn is
returned to the internal breathing system. When
changing patients or anesthetic agent, flush the
breathing system!
B
C
A
611
3 Remove the sealing plug from the inlet nozzle
269
NOTE
If the Ext. fresh-gas outlet is switched to when in
344
Auto control control mode, the Zeus IE activates After closing the Ventilation settings dialog
Fresh-gas control control mode automatically. window, the External fresh-gas outlet mode is
displayed in the lower part of the screen.
Ventilation settings
A Pause mode
Ext. fresh-gas outlet
Overview
The Pause mode stops ventilation and gas delivery
for an indefinite time. Gas measurement and
Check fresh-gas settings.
MixGas will be patient monitoring (optional) remain activated. This
switched to Fresh-
gas control mode is suitable, for example, when intubating or
B C D automatically.
disconnecting the patient.
470
604
Dräger recommends monitoring the patient’s
hemodynamics when carrying out a maneuver. 1 Use the therapy controls to set Pressure (C)
and Duration (D).
Available maneuvers:
2 If necessary, touch the More button (E) and set
– One-step recruitment
a reminder with the therapy control (G).
– Multi-step recruitment
3 Touch the Start button (F) and confirm.
1 Open the Procedures dialog window.
The rise in pressure from the PEEP level to the set
2 Touch the tab of the corresponding maneuver. pressure level and the pressure drop at the end of
the maneuver occur with 20 hPa/s.
Procedures
The maneuver automatically comes to an end
when the time is up. To cancel the maneuver, touch
the Cancel button (H) and confirm. If necessary,
A the PEEP may be adjusted before confirming.
When starting the maneuver, the alarm limit Paw is
checked. If the alarm limit is too low, it will be
B changed to a value 5 hPa below the set pressure.
After the maneuver, the alarm limit is reset to the
original value.
604
Alarm Maneuver
One-step recruitment Multi-step lung recruitment
Pressure alarms:
AW pressure high Alarm limit increased to Alarm limit increased to Pinsp
Pressure +5. If the alarm limit max +5. If the alarm limit is al-
is already at pressure +5 or ready at Pinsp max +5 or high-
higher, the alarm limit remains er, the alarm limit remains un-
unchanged. changed.
Airway pressure not reached Suppressed ...
Volume alarms:
Tidal volume high Suppressed ...
Minute volume low Shown 90 seconds after starting the maneuver at the earliest.
Apnea alarms:
Apnea (no pressure)
Apnea (no flow)
Apnea (no CO2) Suppressed ...
Apnea
AW pressure low
Changing the patient from ventilation mode Changing the patient during the Pause mode
1:35 1:35
Standby
A C
469
473
1 Touch the Standby... button (A). 1 Switch to the Pause mode.
All settings previously modified in operation are Prerequisite: the New patient... button has been
preserved. configured, see page 357.
The Standby dialog window appears. 2 Touch the New patient... button (C).
2 Confirm Standby button (B) with rotary knob. The startup dialog opens.
Or continue ventilation: 3 Perform configuration, see page 127.
Close dialog window.
Pressing the rotary knob automatically switches the
device to the system standby screen.
3 Call up the Startup dialog and perform
configuration, see page 127.
Startup dialog
312
NOTE
Observe hospital regulations for infectious pa-
tients! Dispose of the old filter according to hospital
hygiene regulations.
CAUTION
After each change of patient, a leak test must be
performed to rule out the possibility of major
leaks. A
Standby
300
1 Touch the Device test... button (A).
048
Test details
D C 310
CAUTION
If the leak test has been canceled, it can lead to
leaks not being detected.
Pay special attention during operation.
B D
C
A
486
IV system - IVenus
Information and safety instructions on Observe the Module DPS instructions for use:
the IV application 1 Activate configuration group PAr.
Operation of the Module DPS syringe pump and 2 Set syringe selection PAr 4.
other safety instructions are explained in separate 3 Set at least one syringe type to SEL.
instructions for use.
Attach the IVDock and Module DPS to Zeus IE,
WARNING
see page 85. Note the respective instructions for
use before using these modules. Risk of patient injury
The Module DPS syringe pump can be remotely When starting the Module DPS syringe pump,
controlled with the IVDock on Zeus IE. Module DPS the syringe size, syringe type and drug name
may also be operated independently. must be confirmed directly on the Module DPS
syringe pump. To eliminate the risk of incor-
WARNING rect settings and ensure the patient's safety,
Risk of patient injury the user must check on the syringe pump and
on Zeus IE that the correct drug is being used
Synchronous or parallel and overlapping op- in the correct dosage.
erations of the pump and Zeus IE may lead to
operating errors.
WARNING
The current operations must be completed or Risk of patient injury
interrupted before new operations are under-
Only use weight-related metering at a patient
taken with other devices (Module DPS or
weight of 5 kg and upwards.
Zeus IE).
Correct adjustment of patient weight is re-
Remote control via Zeus IE is not possible until quired because otherwise in the event of
operation of the Module DPS syringe pump is weight-based metering the drug may be me-
terminated. The control panels for the IV system on tered incorrectly.
the screen remain gray for that period of time (i.e.
they are not operable). If any malfunctions should occur on Zeus IE
(IVenus), the syringe pumps can be controlled
If there is an interruption in the data link between
independently of Zeus IE:
Zeus IE and the IV dock, the pumps cannot be
remotely controlled. The syringe pumps then Remove the IVDock plug from the IV socket on
continue running at the set delivery rate and can be Zeus IE. Observe the instructions for use of the
controlled independently of Zeus IE. Module DPS syringe pump!
If remote control of a syringe pump is not possible, NOTE
the syringe pumps must be reconfigured.
When the IVDock plug is disconnected from the IV
socket on Zeus IE, the base (IVDock) is no longer
supplied with power. Each Module DPS is pow-
ered by its own battery. Pay attention to the battery
warning! Infusion is automatically stopped when
the battery is empty.
Infusion settings
1 1 2 3 4
A B
476
345
When a pump is operated by Zeus IE for the first If the IVenus IV system is connected to Zeus IE,
time or after a change of drug, a dialog window (A) operation of the Module DPS syringe pumps is
appears showing the parameters which determine controlled via the screen. Zeus IE can control up to
the metering. 4 Module DPS.
If the application time of a bolus is >180 seconds, Module DPS syringe pumps are controlled from:
this dialog window will also appear. The value of the
A The Infusion settings dialog window for all
bolus rate appears in yellow.
settings
WARNING Or
Risk of patient injury B The field for IV control in the lower part of the
To eliminate the risk of incorrect settings and screen for direct access to key functions
ensure the patient's safety, the patient data The gas supply is then displayed in the data
and/or IV settings must be checked on the sy- modules area.
ringe pump and on Zeus IE when the pump
settings are displayed. During operation, the Module DPS syringe pumps
are controlled via the IV control field.
1 ATRAC C
3.00
Inf mg/kg/h
D
Start bolus
463
WARNING C G
Risk of patient injury
Before starting infusion, the short drug name D E F H
and the dose in mL or mL/h on the pump dis-
play must be compared with those on the I
347
Zeus IE screen in order to prevent administra-
A Drug Select: Selecting a drug from the drug
tion of an incorrect drug.
library
D Depending on the state of the Module DPS For a selected drug, Name, Abbreviation,
syringe pump connected, information, buttons Concentration, Infusion rate limit, and Bolus
or therapy controls are displayed. rate limit are displayed. For "Selecting a drug",
see page 171.
In the example: Button with display of infusion
rate currently set, user-defined unit of B Pump On/Off: Switching the Module DPS
measurement, button for starting the bolus. syringe pump on or off.
C On/Off: Starting or stopping a Bolus.
D Bolus rate: Therapy control for the bolus rate in
Infusion dialog window
a user-defined unit of measurement and display
of the set bolus rate in mL/h.
Opening the Infusion settings dialog window:
E Bolus: Therapy control for the bolus volume in
Touch the header bar in the IV control field.
a user-defined unit of measurement and display
The horizontal tabs stand for Module DPS syringe of the set bolus volume in mL. This is displayed
pumps 1 to 4. The pages are identical. in red for application of an initial bolus.
The vertical tabs open the settings for General and F Time: Display of the bolus application time in
Advanced. seconds.
G On/Off: Starting or stopping an infusion.
H Infusion: Therapy control for the infusion in a
user-defined unit of measurement and display
of the set infusion rate in mL/h.
I Status: Display of a running or waiting function
(e.g., Bolus running ...).
N Max.inf.rate: Limits the infusion rate saved in If a Module DPS syringe pump which is switched off
the profile. is connected to Zeus IE:
Therapy control for the maximum infusion rate
Touch the Start pump button (A).
in a user-defined unit of measurement and
display of the set maximum infusion rate in Or
mL/h
In the Infusion settings (see page 165) dialog
O Infusion: Therapy control for the infusion in a window, touch the Pump On button.
user-defined unit of measurement and display
Confirm with the rotary knob.
of the set infusion rate in mL/h
The Module DPS syringe pump is switched on.
Zeus IE checks information from the syringe pump
being used. The default values are set.
270
Zeus IE prompts the user to confirm the syringe
being used on the Module DPS syringe pump (size Starting the infusion
and make):
1 Confirm syringe at pump . 1 ATRAC
The LED in the (B) button on the syringe pump
flashes in green. A
2 Press the (B) key.
The LED continues to flash in green.
463
1 Confirm drug at pump . Touch the Start infusion button (A).
The LED in the (B) button on the syringe pump Or
flashes in green.
In the Infusion settings (see page 165) dialog
2 Press the (B) key. window, touch the On button for the Infusion.
The LED continues to flash in green. The LED in the (B) button on the syringe pump
An infusion or a bolus application can be started goes out and the infusion starts.
only after the syringe size and the drug have been
confirmed on the syringe pump and after visual
inspection of the assigned Module DPS syringe
pump on Zeus IE.
Checking patient data/weight For each syringe pump, the delivery can be limited
to mL/h (B), if there is no appropriate drug in the
Before starting the first infusion, check the patient
drug library:
weight entered and correct if necessary. The
individual IV deliveries are related to that weight. 1 Touch the respective button.
4
B
A B
3
467
2
1 Touch the Start infusion button (A).
1 The therapy control (B) for the infusion rate is
displayed with the preset value.
464
Extreme settings on the syringe pump which lead to If an infusion is restarted, the infusion rate value
a delivery rate of >1200 mL/h may under certain last set is displayed on the therapy control.
circumstances be a departure from the set range of
the infusion therapy control or bolus therapy
control. This is indicated on the therapy control by Administering a bolus
> max.
Check the concentration and weight setting and A bolus can be started after switching on a Module
adjust if necessary. Select suitable metering. DPS syringe pump or during an infusion. The initial
bolus is displayed in red.
Syringe change After the second bolus and after transport of the
Module DPS to Zeus IE, the maintenance bolus is
If the content of a syringe is used up completely loaded. If the bolus is changed for the maintenance
during the course of an infusion, it can be replaced bolus, that dose is loaded and administered for the
with another syringe filled with the same drug when following maintenance boluses.
the syringe pump is on. The simultaneous
replacement of syringe and drug is described on
page 171. Checking patient data/weight
Before starting the initial bolus, the correct patient
Stopping the infusion weight must be entered as when starting an initial
infusion, see page 167.
1 Touch the Start infusion button (A).
The therapy control (B) for the infusion rate is Starting a bolus
displayed with the preset value.
2 Set 0 with the rotary knob and confirm.
1 ATRAC
Infusion is stopped.
Infusion settings
A 463
1 ATRAC 1 ATRAC
B
C
465
466
1 Touch the Start bolus button (A). Touch the Stop bolus button (C) and confirm.
The therapy control (B) for the bolus volume is Bolus administration stops. Infusion stops.
displayed with the preset value.
2 Use the rotary knob to set and confirm the Stopping the infusion in the Infusion settings
value. dialog window
The bolus starts. Infusion settings
The set value for the bolus is displayed. The pump
symbol indicates the Bolus running ... function by
the display of the plunger moving in the syringe.
If a bolus is started during an infusion, infusion
continues after the end of the bolus.
D
347
Select drug The settings (C) for the drug selected are displayed
alongside the drug list:
The selectable drugs are stored in the drug library, – Concentration
see page 172. – Bolus rate
In the Infusion settings > General dialog window: – Maximum bolus rate
– Bolus rate limit
Infusion settings – Infusion rate
– Maximum infusion rate
– Infusion rate limit
– Initial bolus
A – Maintenance bolus
3 Press rotary knob.
The drug selected is applied. The dialog window
closes.
and drug
1 Touch the Select button (A) and confirm with
rotary knob. If an empty syringe is replaced with a syringe filled
with another drug during the course of an infusion:
The Select drug dialog window appears.
1 Remove the empty syringe from the pump.
Select drug
2 Select the drug on Zeus IE.
3 Insert new syringe.
C
4 Confirm the new drug on the syringe pump.
B
422
The drug list (B) shows all the drugs stored with the
following information:
– Name
– Abbreviation
2 Select a drug from the drug list (B) using the
rotary knob.
Drug library C The settings for the drug selected are displayed
alongside the drug list:
The drug data is managed in the drug library. – Concentration
Access is password-protected, see "Drug library" – Bolus rate
on page 369. – Maximum bolus rate
– Bolus rate limit
Basic settings
– Infusion rate
A – Maximum infusion rate
– Infusion rate limit
– Initial bolus
– Maintenance bolus
– Last modified
D The Delete, Edit, and Add buttons are used to
change the settings for drugs.
E The Import and Export buttons are used to
save drug library data to USB storage media,
from which the data can then be read into
423
C
B
D E
425
F N P
O Q
G
K
H R
L S
I
427
J Touch the Dose settings tab (M).
The infusion and bolus delivery settings are
426
displayed:
Touch the General settings tab (F).
N Infusion rate (mL/h)
The basic drug settings are displayed: O Maximum infusion rate (mL/h)
G Name P Initial bolus (mL)
H Abbreviation Q Maintenance bolus (mL)
I Concentration R Bolus rate (mL/h)
J Concentration unit: µg/mL, mg/mL S Maximum bolus rate (mL/h)
K Quantity unit: mL, mg, µg/kg, mg/kg
Units can be defined by the user and may vary.
L Time unit: min, h
NOTE
If a drug is stored in a profile, the abbreviation can-
not be changed. The drug must first be removed
from the profile.
T W
U V
428
429
Touch the Limit settings tab (T). Touch the Color tab (W).
The infusion and bolus limit settings are displayed: The color coding settings of the respective drug are
U Infusion rate limit (mL/h) displayed.
V Bolus rate limit (mL/h)
Units can be defined by the user and may vary.
NOTE
These limits cannot be changed during operation
Delivery beyond these set limits is possible. A vio-
lation of these limits is announced.
Drug editor
Z
M
426
427
If the name entered is not accepted, there is
already a drug with this name in the drug library. Touch the Dose settings tab (M).
Choose a new name.
The infusion and bolus delivery settings are
Zeus IE allows abbreviation (Y) of the drug name. displayed.
The abbreviation for the drug name must be
1 Touch the buttons of the respective parameters.
unique. 5 characters (letters, numbers) must be
entered. Do not use special characters! 2 Use the rotary knob to set and confirm the
Examples: value.
– PROP1 = Propofol 1 %
– PROP2 = Propofol 2 %
Entering a new abbreviation:
4 Touch the Abbreviation button (Y). A keyboard
window opens.
5 Use the BackSp button in the keyboard window
to delete the placeholder for the abbreviation.
6 Enter name and confirm with the rotary knob.
CAUTION
Risk of patient injury
After transferring a drug library, check the drug
configurations to eliminate the risk of incorrect
settings.
1200
3
1200
2 C G
0
D
1200
1 E
0
F
351
353
The following graphic trends for syringe pumps 1 to Open the tabular display of the data from syringe
4 are displayed from top to bottom: pumps 1 to 4:
– InfR 1 to InfR 4
– administered volume Vtot Touch the Tab. trend button (G).
The scaling limits are: Select Interval [min] for creating a new row in the
table:
– Maximum infusion rate
– Maximum bolus rate Touch the corresponding button (H) (1, 2, 3, 5,
10, 15, 30 min).
For each Module DPS syringe pump, the current
data is, relating to the position of the cursor,
indicated as a numerical value next to the graphic
trend displays:
– Name and Concentration
– Selected infusion
– Delivery rate
– Volume actually administered Vtot
System setup
431
End of operation
C
O2 Air N2O O2 Air N2O
050
1 Touch the On/Off button (C).
The Shut down? dialog window appears.
Standby
300
NOTE
WARNING
Charge the battery at least every 4 weeks, other-
Remove gas supply hoses from the wall termi-
wise there will be no uninterruptible power supply
nal units (central supply).
available when the device is put into operation,
If the gas supply hoses remain connected to see page 64.
the wall terminal units, even minor internal
leaks may lead to contamination of the supply
gases.
NOTE
As the DIVA metering modules cool down slowly
(when using desflurane), metering accuracy is
temporarily reduced if Zeus IE is shut down and
put back into operation immediately.
Alarms
Displaying alarms
B A 1:35
Alarms of highest priority
– The alarm message requires immediate action.
– Text on a red background
– The corresponding parameter flashes
C
– Continuous tone sequence
Medium-priority alarms
– The alarm message requires immediate action.
301
High Yes
Depending on the overall alarm situation, this tone sequence
may be played only as a 5-tone sequence due to the timing
of the individual alarms.
Agent Yes
Low No
When Signal
– Start of therapy
– Test tone during self test
– Dialog could not be opened
(Pitch of the individual tones differs de-
pending on the cause)
Timeout imminent, as a setting has not
been confirmed yet
Alarm info
Touch alarm message field. All current alarms are listed in chronological order
with the following information:
The Alarms >Alarm info dialog window (A) will
B Duration
open with all currently active alarm messages.
C Prio
Alarms D Alarms
E Alarm limits
A F Cause
B C D E G Remedy
!!!
! See also the alarm messages from page 287.
!
!
!!
F G
356
Alarm log
The Alarm log records all alarm messages that All alarms that occur are listed in chronological
have occurred, including those that are no longer order with the following information:
active, in chronological order (maximum 200). The B Date
alarm log deletes the oldest entries as soon as new C Time
entries are recorded in an already full alarm log. D Priority
E Patient alarms
Alarms E Technical alarms
A See also the alarm messages from page 287.
B C D E F The Graph.trend (G) and Tab. trend (H) buttons
!!!
! open the dialog windows for trend display as a
!
!
curve or table at the time of the selected alarm.
!!
! The alarm log is cleared:
!
! – At the start of a new case
!!
!!
– When switching on the system
!!! – In the event of a total power failure of more than
G H 5 min (after the batteries are empty)
358
Alarm limits
message field
Alarms
359
The Alarms > Alarm limits dialog window will
open for the parameter selected. The following user configurations (E) with the
corresponding upper (F) and lower (G) alarm limits
3 Touch the button of the alarm limit to be can be displayed:
modified. The button turns yellow. – Noninvasive (optional)
4 Set the required value by turning the rotary – Invasive (optional)
knob and push to confirm. – Ventilation
– Anesth.
Or – Infusion (optional)
1 Touch the Alarms... button (C).
The Alarms > Alarm limits dialog window will
open. The alarm-related limit is preselected, the
button is yellow.
2 Set the required value by turning the rotary
knob and push to confirm.
359
1 Touch the vertical tab of the user configuration
of the alarm limit to be modified. The following upper and lower alarm limits can be
2 Touch the button of the alarm limit to be configured:
modified. The button turns yellow. B HR (heart rate)
C SpO2 (blood oxygen saturation)
3 Set the required value by turning the rotary D NIBP S/NIBP D/NIBP M
knob and push to confirm. Blood pressure depending on the set
parameter: S, D or M
The graphic displays a 30-minute trend of the
E T1 (temperature, measuring point 1)
alarm-related parameter. The broken lines (H) in
F T2 (temperature, measuring point 2)
the graphic represent the alarm limits and are
G TBlood (blood temperature)
moved accordingly.
monitored for cardiac output measurement
If the alarm limit of the parameter is being H NMT (TOF count)
monitored, the color of the button is dark green - I BIS (BIS index)
otherwise it is bright green.
A A
B C D E F G H
B
361
360
Touch the Anesth. button (A). Touch the Infusion button (A).
Alarms Alarms
A
B C 1 A
B C
362
479
The following upper and lower alarm limits can be The upper alarm limits are only displayed if the
configured: syringe pumps are connected to Zeus IE.
B Inspiratory N2O content of the breathing air
(inN2O) These buttons only appear if N2O is The following upper alarm limits can be configured:
selected as the carrier gas. B Infusion pressure
C Inspiratory anesthetic agent content of the C Infusion volume
breathing air (depending on the selected
anesthetic agent, e.g., inIso). KVO function*
If the KVO function is activated in the syringe pump,
the syringe pump does not shut down if the upper
alarm limit for the infusion volume is exceeded.
KVO appears on the pump display. The infusion
continues at a low infusion rate. The syringe pump
cannot be remote-controlled.
1 Switch off syringe pump.
2 Increase the volume alarm limit on Zeus IE.
3 Switch syringe pump back on.
Suspending alarms
Alarms
A
B G
C
D
E
F
363
Pause no no
MAN/SPON no yes
Vol. Ctrl. AutoFlow yes yes
Volume Control yes yes
Pressure Control yes yes
Pressure Support yes yes
Smart Ventilation Control (optional) yes yes
External FG outlet (optional) no yes
Technical alarms
Technical alarms can be acknowledged and then
set to a low priority.
CAUTION
Suspending technical alarms does not eliminate
the cause of the alarm. Eliminate the cause of the
alarm immediately, see page 287!
WARNING
National and international standards specify
minimum monitoring with alarm functions.
These regulations may not be complied with if
the alarm function of monitoring parameter et-
CO2 is disabled!
Clearing alarms
Alarms can no longer be cleared by means of the The following alarm messages can be completely
Audio paused key, but the alarm priority of the cleared:
following active alarms can be reduced from
medium to low: Alarm text
ART nonstatic
Alarm text
ART transducer unplugged
NIBP D high
BISX pod disconnected
NIBP D low
BISX sensor disconnected
NIBP M high
C.O. complete
NIBP M low
C.O. out of range
NIBP S high
C.O. unplugged
NIBP S low
C.O. unstable baseline
TOF count high
Check NIBP cuff size
TOF count low
Check patient category
A
ALARM
RESET
1:35
CVP nonstatic
CVP transducer unplugged
DIVA off! Module removed
Drug data read error
Drug data write error
Event from web appl.
GP1 nonstatic
GP1 transducer unplugged
GP2 nonstatic
612
In the header bar, touch the ALARM RESET GP2 transducer unplugged
button (A) and confirm. HemoMed pod 1 disconnected
All the alarms displayed in the alarm message field HemoMed pod 2 disconnected
that can be cleared are cleared.
HemoMed pod 3 disconnected
ICP nonstatic
ICP transducer unplugged
Infusion 1 stopped
Infusion 2 stopped
Infusion 3 stopped
Infusion 4 stopped
IV remote control failed
Select alarms for which the alarm recorder Calling up Recorder config.:
automatically starts when they occur.
1 Touch the Recorder config. tab (C).
Touch the Alarm recorder tab (A).
Alarms
Alarms
A
A G
B H C
C
D E I
J
K
L
M N
F
364
357
All limits
B
K L M N O
365
365
BP Other BPs
Touch the IBP tab (A). Touch the More IBP tab (A).
Alarms Alarms
B C D E F G H A B C D E F G H I
A
I J K J K L M N
366
367
The alarm limits of the following parameters can be The alarm limits of the following parameters can be
configured: configured:
B ART S (systolic arterial blood pressure) B LV S (systolic left ventricular blood pressure)
C ART D (diastolic arterial blood pressure) C LV D (diastolic left ventricular blood pressure)
D ART M (mean arterial blood pressure) D LV M (mean left ventricular blood pressure)
E PA S (systolic pulmonary-arterial blood E RV S (systolic right ventricular blood pressure)
pressure) F RV D (diastolic right ventricular blood pressure)
F PA D (diastolic pulmonary-arterial blood G RV M (mean right ventricular blood pressure)
pressure) H LA M (mean left atrial pressure)
G PA M (mean pulmonary-arterial blood pressure) I RA M (mean right atrial pressure)
H CVP M (mean central venous pressure) J GP2 S (other non-specific systolic blood
I GP1 S (non-specific systolic blood pressure) pressure) (general purpose)
(General Purpose) K GP2 D (other non-specific diastolic blood
J GP1 D (non-specific diastolic blood pressure) pressure) (general purpose)
(General Purpose) L GP2 M (other non-specific mean blood
K GP1 M (non-specific mean blood pressure) pressure) (general purpose)
(General Purpose) M ICP M (mean intracranial pressure)
N CPP M (cerebral perfusion pressure)
Arrhythmia Anesthesia
Touch the Arrhythmia tab (A). Touch the Anesth. tab (A).
Alarms Alarms
A
H I J K L M N A
B
C
368
369
The following parameters can be configured The alarm limits of the following parameters can be
depending on the option used: configured:
B Dual lead processing: Off/On. B inN2O (inspiratory N2O content of the breathing
C Arrhythmia monitoring: air)
– Adv. (optional) C inHal (inspiratory anesthetic agent
– Basic concentration of the breathing air)
– Off D inEnf (inspiratory anesthetic agent
concentration of the breathing air)
Specify the settings for Rate and Count when an E inIso (inspiratory anesthetic agent
arrhythmia event is reported. concentration of the breathing air)
F inSev (inspiratory anesthetic agent
concentration of the breathing air)
Setting the Rate
G inDes (inspiratory anesthetic agent
Touch button, set the required value and concentration of the breathing air)
confirm. H etCO2 (endexpiratory CO2 content of the
breathing air)
For ASY, VF and ARTF the rate cannot be I inCO2 (inspiratory CO2 content of the breathing
changed. air)
J FiO2 (inspiratory O2 content of the breathing
Setting the Count air)
K MVe (expiratory minute volume)
Touch button, set the required value and L VTi (inspiratory tidal volume)
confirm. M Paw (airway pressure)
For setting ranges, see "Set ranges of the patient N Brody Factor
monitoring alarm limits" on page 207
WARNING
Risk of patient injury
The user must adjust maximum inspiratory VA
concentration to suit the patient or else there
is a risk of concentration levels being too high
or too low.
Infusion
Alarms
1 2 3 4
B C A
350
Autoset limits
Alarms
D
E
A F
359
WARNING
The Autoset function always refers only to the
alarm limits highlighted in yellow on the par-
371
ticular page! The Autoset difference (X) can be set for the
This resets the previously active alarm limits! following areas:
E Hemodyn.
Disabled alarms are not automatically en- F IBP
abled. G Anesth./ Ventilation
Autoset can be performed for the following areas:
– Noninvasive
– Invasive
– Ventilation
– Anesth.
1 Touch the corresponding tab, e.g., (C).
2 Touch the Autoset button (A). The Autoset
button is preselected in yellow. Alarm limits that
will be recalculated by the Autoset function are
highlighted in yellow on the respective page.
3 Touch the Autoset (A) button. The alarm limits
highlighted in yellow will be recalculated.
The limit is adjusted based on the formula:
Upper limit = Current measurement + X %
Lower limit = Current measurement – X %
Setting Autoset limits for the Hemodyn. The following parameters can be configured:
area B ART S, ART D, ART M (arterial blood pressure
values)
Touch the Hemodyn. tab (A). C PA S, PA D, PA M (pulmonary-arterial blood
pressure values)
D CVP M (mean central venous pressure)
Alarms
E GP1 S, GP1 D, GP1 M (non-specific (general
purpose) blood pressure values)
F LV S, LV D, LV M (left ventricular blood
A pressure values)
B C D E F G H I G RV S, RV D, RV M (right ventricular blood
pressure values)
H LA M (mean left atrial pressure)
I RA M (mean right atrial pressure)
J GP2 S, GP2 D, GP2 M (non-specific (general
purpose) blood pressure values)
K ICP M (mean intracranial blood pressure)
371
The following parameters can be configured: Setting Autoset limits for the Anesth./
B HR (heart rate) Ventilation area
C SpO2 (blood oxygen saturation)
D NIBP S (systolic blood pressure) Touch the Anesth./ Ventilation tab (A).
E NIBP D (diastolic blood pressure)
F NIBP M (mean blood pressure)
Alarms
G T1 (temperature, measuring point 1)
H T2 (temperature, measuring point 2)
I TBlood (blood temperature, monitored for
cardiac output measurement) B C
A
Setting Autoset limits for the IBP area
The alarm tone can be suppressed for a maximum Switching the audible alarm on again
of 2 minutes.
Press the Audio paused key (A) again.
The information field in the header bar is empty.
Exceptions
The following alarms immediately end the silence
time:
Alarm message
Air supply fail
Air+N2O delivery cutoff
O2 Air N2O O2 Air N2O
In the header bar, the symbol (B) with the DIVA protect fail
remaining silence time (minutes:seconds) will
DIVA+mixer failure
appear.
Fresh gas intern/extern?
If the Audio paused key is pressed a second time
during this time, the All symbol will be shown. Infusion line 1 blocked
All audible alarms which appear in the next Infusion line 2 blocked
2 minutes will be suppressed. The visual signal on
Infusion line 3 blocked
the alarm bar disappears.
Infusion line 4 blocked
If there are no alarms active before pressing the
Audio paused key, then all alarms which occur in Insp. desflurane high
the next 2 minutes will be suppressed (preventive Insp. enflurane high
silence function). Insp. halothane high
Insp. isoflurane high
Insp. oxygen low
Insp. sevoflurane high
Alarm message
Insp. vol. agent high
Internal safety fan failed
N2O supply fail
Oxygen supply fail
Power supply temp high
Pump 1 battery empty
Pump 1 failure
Pump 1 operation failed
Pump 2 battery empty
Pump 2 failure
Pump 2 operation failed
Pump 3 battery empty
Pump 3 failure
Pump 3 operation failed
Pump 4 battery empty
Pump 4 failure
Pump 4 operation failed
Remove DIVA module 1
Remove DIVA module 2
Safety O2 opened
Safety settings active
Syringe 1 empty
Syringe 2 empty
Syringe 3 empty
Syringe 4 empty
VA delivery failure
Ventilator failure
Device-related
Patient-related
Arrhythmia detection
1:35 1:35
Trends/Data
B C D E F G H
A
469
374
can be printed (with Infinity network option).
Horizontal tabs open the diagnosis windows.
Vertical tabs show graphic, tabular displays and
Printing graphic trend display other figures.
The graphic trend always starts with the current The following applications can be selected:
time and prints the entire trend from the time the B Overview
patient case started. Trends in: HR, NIBP, ART, CVP, etCO2, i/e O2,
SpO2, i/e VA, Paw, MV
C Hemodyn.
Printing tabular trend display Trends in: HR, ART, CVP, PLS, NIBP, PA, RA,
The entire trend with the selected interval time from SpO2, T1, T2, Pmean, C.O.
the start of the patient case is printed. Cardiac calculat.: C.O., HR, ART M, CVP, PA
M, PWP, CI, SV, SVI, SVR, SVRI, PVR, PVRI,
Diagnosis windows LVSWI, RVSWI, BSA
ST view: inferior, lateral, anterior
Measured values in different applications, arranged D BIS/NMT
according to different parameters are saved and Trends in: BIS, SQI, EMG, BSR, BCT, SEF,
displayed in the Trends/Data dialog window. PWR, SNGL, TOF, TOFR, PTC
E Infusion
Touch the Trends/Data... button (A).
Trends in: Infusion data (optional,
The Trends/Data dialog window appears. see page 177)
F Ventilat./ Oxygenat.
Trends in: etCO2, MV, i/e O2, SpO2, Paw, C, R,
RRmand, RRspon, VT
Resp. mech.:
– Pressure: PIP, Plat, Mean, PEEP,
– Lung: Cdyn, R
– Ventilator: Circle flow
– Vol: MVe, VTi, VTe, RR, VTmand, RRmand,
MVespon, VTspon, RRspon
G Gases/ Anesth.
Trends in: xMAC, i/e VA, ART, NIBP, HR, N2O,
O2 upt., RRspon, T NMT, TOF, SQI, BIS
Current flows:
– Fresh-gas control: System closed
– Current flows: O2, Air, N2O, Total
– Injected anesth. agent (fl.)
Total consumpt.:
– Gas amount since start: O2, Air, N2O
– Volatile agent amount since start
– Gas costs since start: O2, Air, N2O
– Volatile agent costs since start
H System
Automatically opens the Test details > Test
history page.
Monitoring
1200
B
A
389
0
0 20
mL/mbar mbar/L/s
For hemodynamic patient monitoring (optional), the
C following pages can be configured:
D ECG
E SpO2
F NIBP
G IBP
H Temp
309
I C.O./ Calculat.
Touch the parameter field, e.g., (A). J BIS/NMT
The corresponding page opens in the For device monitoring, the following page can be
Sensors/Parameters dialog window. configured:
K Anesth./ Ventilation
Or
The parameter fields are configured in the Screen
Touch the area of the loop display (B).
setup > Screen layout > Datadialog window,
The Sensors/Parameters > Anesth./ Ventilation see page 356.
> PV & FV Loops dialog window will appear.
Or Parameter module
If the Sensors/ Parameters... button is configured If "---" is displayed in a parameter module, that
(see page 357): measured value is invalid or out of date.
Touch the Sensors/ Parameters... button (C). If an apnea alarm or an asystole alarm occurs, no
measurements will be displayed in the CO2 scale
The Sensors/Parameters dialog window appears.
and HR/PLS parameter modules. The length of
time the alarm has already been present is
displayed in the relevant parameter module.
Device monitoring
A
B 73 53
23
in
7
225
C
D The CO2 scale parameter module displays:
E – The inspiratory (in) CO2 concentration in the
breathing gas
F – The expiratory (et) CO2 concentration in the
G breathing gas
– The respiratory rate RR from the CO2 signal
Touching the Display all button (requirement: the
404
Setting the amplitude of the CO2 scale Setting the amplitude of the O2 scale
Sensors/Parameters Sensors/Parameters
A B
B A
C C
D D
404
405
Touch the CO2 scale button (A). Touch the Oxygen button (A).
The CO2 scale page displays: The Oxygen page displays:
B The current CO2 curve B The current O2 curve
C CO2 scale (scaling of the CO2 curve in relation C O2 scale (scaling of the O2 curve in relation to
to the amplitude) the amplitude)
D The button opens the Alarms > Alarm D The button opens the Alarms > Alarm
limits > Ventilation dialog window for alarm limits > Ventilation dialog window for alarm
limit setting. limit setting.
Oxygen Flow
O2 % MV L/min
MV VTe
31
6.0
in
R
500
27 12.0
et Off
Off
20 4.0
226
227
406
Touch the Flow button (A).
The Flow page displays:
B The current Flow curve
C Flow scale (scaling of the Flow curve in relation
to the amplitude for the flow)
D Vol. scale (scaling of the Flow curve in relation
to the amplitude for the volume)
E Waveform selection: represent Flow or Vol.
as a curve
F The button opens the Alarms > Alarm
limits > Ventilation dialog window for alarm
limit setting.
Sensors/Parameters
Paw mbar
PIP Pmean
13
30
B
PEEP
14 C A
The Paw parameter box displays: 228
407
Touch the Paw button (A).
Paw mbar
The Paw page displays:
30
PIP Pmean
Pplat
13 B The current Paw curve
23
PEEP
40
Auto 5 C Scale (scaling of the Paw curve in relation to
the amplitude)
229
Display 1
B
– Inspiratory (in) and expiratory (et) values of the
anesthetic agent, of N2O, and O2
C
A
Iso N 2O O2%
2.2 60 37
in
1.2 50 27
et
D
408
Touch the Volatile agent button (A).
230
Pressure/volume loops
0 30 Pressure/volume loops show changes in the
compliance, resistance and respiratory work.
613
The trend of the CO2 production, calculated as the The loop of a mechanically generated breath is
product of MV and (etCO2–inCO2), is displayed in plotted in an counterclockwise direction. The loop
the MVxCO2 parameter module. of a spontaneous breath is plotted in a clockwise
direction. The inspiration starts at a point which is
defined via the base pressure and the volume at the
R/C trend beginning of inspiration. The loops show the
appropriate values for dynamic compliance (Cdyn).
R/C
30 100 Flow/volume loops
R
Flow/volume loops plot mechanical and
C
spontaneous breaths in the same manner.
Inspiration starts at the starting point of the loop and
is then plotted upwards and from the left-hand side
0 30
to the right-hand side (clockwise). Expiration is
plotted below the horizontal axis from right to left,
614
The trend of the resistance and compliance values back to the starting point.
is displayed in the R/C parameter module.
∆VT (optional)
∆VT mL
VTi
335
15 VTe
320
237
0 Paw
0 20
0 Paw
0 20 Cdyn 24 R 2
mL/mbar mbar/L/s
234
A B
Displaying the Save and Recall buttons again:
Touch the screen in the area of the values for
232
min
Touch the PV & FV Loops button (C).
The PV & FV Loops page displays:
477
D Paw/Vol, Flow/Vol (possible loop displays) Touch parameter field.
E Auto, Volume, Customized (scaling of the
loops) The current alarm limit setting appears.
F Volume (scaling of the volume)
G Paw (scaling of the pressure axis)
H Flow (scaling of the flow axis)
I Loop selection: Paw/Vol, Flow/Vol (display of
the loop transferred to the screen)
J The button opens the Alarms > Alarm
limits > Ventilation dialog window for alarm
limit setting.
A The current flows of the anesthetic agent, e.g., and can be adjusted.
Iso and the gases O2, Air, and N2O are Touch the display of the flow tubes.
displayed.
The Trends/Data > Current flows dialog window
B The status display indicates the status of gas will appear.
delivery (fresh-gas control mode, closed
system, current fresh-gas flows, etc.).
C The O2 uptake display is used for leakage
detection when operating in a closed system. If
the device is in Uptake mode, the complete O2
uptake of the patient is displayed.
Patient monitoring
148
389
* optional
SpO2 measurement with the SpO2 SmartPod SpO2 measurement without SpO2 SmartPod
(optional)
The connection field has the following connections:
The connection field has the following connections:
HEMOMED1 MULTIMED
A B HEMOMED1 MULTIMED
H I
Aux/HEMO2
C Aux/HEMO3
Aux/HEMO2 J Aux/HEMO3
NIBP SpO2
D E
NIBP Sync.
IV-System Sync.
K L
F G
IV-System
M
009
A HEMOMED 1
Connection for accessories to measure 1 to 4
010
invasive blood pressures (IBP), and C.O.
H HEMOMED 1
B MULTIMED Connection for accessories to measure 1 to 4
Connection for accessories to measure 3 to 6 invasive blood pressures (IBP), and C.O.
ECG leads, and 1 or 2 temperatures
I MULTIMED
C Aux/Hemo 2/3 Connection for accessories to measure SpO2, 3
Connections for accessories to measure 1 to 4 to 6 ECG leads, and 1 or 2 temperatures
invasive blood pressures and C.O., as well as
for anesthesia effect monitoring (Trident pod, J Aux/Hemo 2/3
BISx pod) Connections for accessories to measure 1 to 4
invasive blood pressures and C.O., as well as
D NIBP for anesthesia effect monitoring (Trident pod,
Connection to measure the non-invasive blood BISx pod)
pressure
K NIBP
E SpO2 Connection to measure the non-invasive blood
Connection to measure oxygen saturation pressure
(sensor technology must be identified)
L Sync
F IV system Synchronization connection for defibrillators
Connection of the Module DPS for intravenous
anesthesia (optional) M IV system
Connection of the Module DPS for intravenous
G Sync anesthesia (optional)
Synchronization connection for defibrillators
Zeus IE connec- Accesso- ECG SpO2 NIBP BP Temp. C.O. NMT/ BIS
tion field ries TOF
MULTIMED MULTIMED 3 to 5 1 sen- 1 or 2
5 cable elec- sor1) sen-
trodes sors2)
MULTIMED 3 to 6 1 Sen- 1 or 2
6 cable elec- sor1) sen-
trodes sors2)
MULTIMED 3 to 6 1 Sen- 1 or 2
PLUS OR elec- sor1) sen-
cable trodes sors2)
HEMOMED 1 Y-cable 1 or 2
pressure
transduc-
er(s)2)
HemoMed 4 pres- x
Pod sure
transduc-
ers
Aux/Hemo 2 or Hemo2 Pod 2 pres- 2 sen- x
Aux/Hemo 3 sure sors
transduc-
ers
Hemo4 Pod 4 pres- 2 sen- x
sure sors
transduc-
ers
Quad Hemo 4 pres- 2 sen- x
Pod sure sors
transduc-
ers
Trident pod 1 NMT
sensor
BISx pod 1 BIS
sen-
sor
Zeus IE connec- Accesso- ECG SpO2 NIBP BP Temp. C.O. NMT/ BIS
tion field ries TOF
SpO23 Masimo SET 1 Masi-
cable mo sen-
sor3)
Nellcor Oxi- 1 Nellcor
Max cable sensor3)
NIBP NIBP hose 1 sen-
sor
1) optional for measurement without SpO2 SmartPod
2) depends on the use of a Y-cable
3) omitted for measurement without SpO2 SmartPod
185
Safety with high frequency and defibrillation
A For the display of 3, 5 or 6 ECG leads, different
Safety measures when using a defibrillator:
cables can be connected to the MULTIMED
– The monitor is protected from defibrillator
connection on the Zeus IE.
voltages, high-frequency interference from
– MULTIMED 5 cable
electro-surgery devices, and 50/60 Hz
– MULTIMED 6 cable
interference.
– MULTIMED PLUS OR cable
– If the correct electrodes are used and these
electrodes are attached in accordance with the
manufacturer's instructions, the monitor will
display waveforms again within 5 seconds of
the defibrillation.
WARNING
Risk of burns
To achieve accurate ECG data and for the pro-
tection of the patient during defibrillation, only
use ECG electrodes and cables specified by
Dräger. To prevent sensor faults and the dis-
persion of energy, applied parts such as dis-
posable SpO2 sensors that are not classified
as defibrillation-safe must be removed.
CAUTION
Risk of burns and danger of electric shock
Electrodes and sensors can divert the electric cur-
rent. This can result in burns and cause an electric
shock.
– Only perform defibrillation above the rib cage.
– Do not position the defibrillator in the vicinity of
electrodes or sensors.
C B G
D H
F E J I
142
143
B MULTIMED cable G MULTIMED PLUS OR cable with integrated
ESU block
C ESU block for protection from interference due
H ECG cable (3, 5 or 6 leads) for connection to the
to high-frequency surgery electrodes applied to the patient
D ECG cable (3, 5 or 6 leads) for connection to the I Connection socket for the temperature cable
electrodes applied to the patient Two temperature sensors may be connected
via a Y-cable.
E Connection socket for the temperature cable J Connection socket for the SpO2 cable*
Two temperature sensors may be connected
Insert white ECG standard lead wire set (D) into
via a Y-cable.
the MULTIMED PLUS OR Pod (C).
F Connection socket for the SpO2 cable*
1 Insert white ECG standard lead wire set (D) into
the ESU block (C). If a MULTIMED 5 cable is
used, leave the V+ connection vacant.
2 Connect ESU block (C) to the MULTIMED cable
(B).
5-pin and 6-pin ESU blocks are available.
3 All electrode inputs must be covered.
Use ESU block during HF surgery only.
optional without
NIBP NIBP SpO2 SmartPod
A A
SpO2 Sync.
E E
185
185
Connection for IBP accessories Hemodynamic monitoring modules which perform
the following measurements with only one monitor
For invasive blood pressure measurement, four connection:
alternatives are available on Zeus IE: – Cardiac output
– Y-cable – 2 invasive blood pressure measurements
– Hemo2 Pod (Hemo2) or
– Hemo4 Pod – 4 invasive blood pressure measurements
– HemoMed Pod (Hemo4, HemoMed and Quad Hemo)
– Quad Hemo Pod Hemo2, Hemo4 and Quad Hemo must be
connected to an Aux/Hemo 2/3 connection (F) on
Y-cable Zeus IE.
The Y-cable can be used to connect two IBP The HemoMed Pod must be connected to the
pressure transducers to measure two invasive HEMOMED 1 connection (E) on the Zeus IE.
blood pressures. The Y-cable is connected to the In contrast to the Hemo2 and Hemo4 Pods, the
HEMOMED 1 connection on the Zeus IE HemoMed Pod has no LCD displays on the module
connection field, see page 226. and no temperature connections.
C CAUTION
D Do not autoclave hemo modules.
136
Hemo2 Pod, Hemo4 Pod and HemoMed Pod Quad Hemo Pod
G H I
J K L
135
The modules have three buttons:
103
G The SmartZero key performs the zeroing
of all pressure transducers connected to the J The SmartZero key performs the zeroing
module which are open to the ambient of all pressure transducers connected to the
pressure. module which are open to the ambient
H The C.O. Start key starts cardiac output pressure.
measurement. K The C.O. Start key starts cardiac output
I The Wedge key starts pulmonary wedge measurement.
pressure measurement. L The Wedge key starts pulmonary wedge
pressure measurement.
M
135
N 133
102
Q Connection for C.O.
R Connection for Zeus IE
2 Connect the red end of the Pod communicatons
cable (T) to connnection (R). Connect the other
end (S) to the Aux/Hemo2 or Aux/Hemo3
connection on Zeus IE.
O S
160
U U U
T
P
V
101
B
138
A MULTIMED Pod
B Temperature sensor
A Y-cable has to be used for the measurement
of 2 temperatures. C
optional without
SpO2 SmartPod
B D
150
C Connecting cable
NIBP NIBP
D SpO2 sensor
A
SpO2 Sync.
E F
A
I H G
E
C B
J
F D
K
137
A Injectate
B Injectate temperature sensor
C Proximal lumen
G D Thermistor T-piece
E Three-way tap
F C.O. Y-cable
152
* optional
G
A F
B
276
F Pod Com cable to the Zeus IE connection field
or to the Trident pod
G Sensor cable for connecting the BIS sensor
C
SpO2 SmartPod
A Covered connection, currently no function
HEMOMED1 MULTIMED HEMOMED1 MULTIMED
A
185
60
by Dräger.
120
ST To prevent sensor faults and the distribution
45 of energy, applied parts that are not classified
as defibrillation-safe must be removed.
224
Amplitude
WARNING
– 0.2 to 5.0 mV (for channel amplitudes of x 4 and Evoked potential equipment can cause faults
x 8) in ECG monitoring.
– 0.4 to 5.0 mV (for all other channel amplitudes)
QRS width
– 70 to 120 ms for adults
– 40 to 120 ms for neonates. Further information
regarding QRS detection can be found in "Fault
- Cause - Remedy".
WARNING
With dual-chamber pacemakers, Zeus IE may
record the second pacemaker artifact as a val-
id QRS complex, particularly when the AV in-
terval is less than 150 ms. A cardiac arrest will
thus not be recognized in unfavorable circum-
stances.
WARNING
During cardiac arrest or certain arrhythmias,
the heart rate generated by the pacemaker
may still need to be recorded and counted.
Therefore, do not relay solely on the heart rate
measurement. Special attention must be paid
to pacemaker patients.
WARNING
Do not rely entirely on the displayed heart rate
or respiratory rate for evaluating the status of
a pacemaker patient. Such patients must al-
ways be monitored closely and their vital
signs checked carefully.
WARNING
ECG leads, temperature sensors, pressure
measuring accessories, SpO2 sensors, and in-
termediate cables must be kept away from the
grounding, as well as from the electrosurgical
device and its cables.
WARNING
When using HF surgery devices, only use blue
233
Patients
218
Only use sterile electrodes.
The electrodes must be fastened with a
watertight bandage (approx. 5 cm wide) or with
sterile adhesive tape and protected against
3-lead standard / 5-lead standard
liquids.
217
Electrode positioning
ECG connection cable Lead wire set Display channels Leads available
MULTIMED 5, 6, 3-lead ECG 1 or ECG 2 or ECG 3 I, II, III
MULTIMED PLUS OR
MULTIMED 5, 6 5-lead ECG 1, ECG 2, ECG 3 I, II, III, aVR, aVL, aVF, V
MULTIMED PLUS OR
MULTIMED 6, 6-lead ECG 1, ECG 2, ECG 3 I, II, III, aVR, aVL, aVF, V, V+
MULTIMED PLUS OR
WARNING
Electrical artifacts arising from a cause other
than heart disease, e.g., attacks/spasms, may
prevent the detection of certain arrhythmias.
In the case of patients prone to attacks, the
user should not solely rely on the ECG.
247
priority.
Zeus IE can detect and display deviations in the ST
segment of the QRS complex in relation to an
Automatic learning and relearning isoelectric baseline.
If the ECG cables are connected to the patient, This is done by measuring voltage values and
Zeus IE will start a learning phase in the following comparing them at two measurement points, the
situations: isoelectric point and ST measurement point. The
– After switching on Zeus IE isoelectric point defines the zero voltage (no
– When leaving the system standby mode electrical activity = 0 mm) and occurs on the
– After switching on arrhythmia detection horizontal time axis as the basic setting 28 ms
– When changing the lead in the ECG channels before the beginning of the QRS complex. The ST
ECG-HF (ECG1) or ECG-ST (ECG2) while measurement point occurs in the ST segment
processing one or two ECG signals between the end of the QRS complex (J point) and
the T segment as the basic setting 80 ms after the
end of the QRS complex.
Sensors/Parameters
A ST segment analysis
The dialog window displays the screens for setting Setting and evaluating the ST segment analysis:
ECG functions: Touch the ST segm. tab (A).
B ECG HR/PLS (first ECG channel)
C ECG ST (second ECG channel) Sensors/Parameters
D ECG C.O. (third ECG channel)
E ST segm. (ST segment analysis)
F More ECG
B
The possible settings of the three ECGs (ECG
HR/PLS, ECG ST, ECG C.O.) are identical.
G Waveform display of the respective ECG lead
C
D A
H ECG lead
I ECG amplitude E
F
The normal QRS detection threshold lies at
approx. 0.4 mV, depending on the QRS width. G
This threshold is used for channel amplitudes of
H I J 391
223
In the display of the advanced ST option, all ST view
available leads are displayed.
inferior lateral anterior
Measuring points II
The start point and end point of the QRS complex -8 30 min
459
Touch the respective button. Use the rotary
knob to put the cursor on the required place and The example for ECG lead II shows the display of
confirm. the current QRS complexes (green) and the one-
time measurement of the reference complexes
The following settings are possible:
(red) above each other. In addition, a short-time
C ST iso point: Setting the measuring position for
trend of 30 minutes of the determined ST rise/fall is
the isoelectric point
applied.
D ST meas. point: Setting the measuring position
of the ST measuring point
E ST trend scale: Amplitude setting for the ST
trends
F ST event duration: Setting the time after which
there will be an alarm if there has been an
uninterrupted limit transgression
G ST monitoring: Switching On/Off the ST
segment analysis
H Save reference
I ST view, opens ST view in the Trends/Data
dialog window
J The button opens the Alarms >All limits >
Hemodyn. dialog window for alarm limit setting
of the measured values ST LAT, ST INF, and
ST ANT.
CAUTION
Do not irradiate or autoclave sensors or cables or
immerse them in water or other liquids. It can
cause malfunctions.
99
– Using Nellcor OxiMax: 8 seconds on
PLS
100
90 67 average
– Using MultiMed*: 90 % of the SpO2 change
222
is displayed within 30 seconds (this setting
Zeus IE displays SpO2 values and the pulse rate in is less susceptible to artifacts, but alarm
the SpO2 parameter module. Additionally, the messages occur more slowly).
alarm limits for the SpO2 value are always
displayed. The corresponding waveform field Fast:
represents the SpO2 waveform (pulse – Using Masimo SET: 2 to 4 seconds on
plethysmogram). average
– Using Nellcor OxiMax: 2 to 4 seconds on
NOTE average
The SpO2 pulse plethysmogram is normalized and – Using MultiMed*: 90 % of the SpO2 change
is not directly proportional to the pulse amplitude. is displayed within 15 seconds (this setting
The waveform amplitude is automatically maxi- is more susceptible to artefacts, however
mized. Consequently the size is only reduced in alarm messages occur more quickly).
the event of poor signal quality. To enlarge the F Pulse tone modulation (depending on the
waveform, tap on it and adjust the scale. saturation rate): On, Off
G NIBP interlock: On, Off
Switches SpO2 alarms off during an NIBP
measurement. The SpO2 value is frozen and
Settings on Zeus IE displayed in gray for the duration of the NIBP
measurement.
Touch the SpO2 tab (A). H The button opens the Alarms > Alarm
limits > Noninvasive dialog window for alarm
The dialog window for SpO2 monitoring appears. limit setting.
Sensors/Parameters
Notes on testing and on accuracy of the blood
A oxygen meter (oximeter)
WARNING NOTE
Risk of incorrect measured values and high The accuracy of the oscillometric blood pressure
cuff pressure signal may be impaired under the following condi-
tions (to the point of loss of the measurement).
Incorrectly selected cuffs and patient catego-
– Weak pulse
ries may cause cuff pressures that are too
– Incorrectly selected measurement point
high for children and neonates.
– Irregular pulse
Before monitoring neonates and children: – Artifacts caused by movement of the patient
– Select the correct cuff size and hose. – Artifacts caused by convulsions of the patient
– Select the neonatal or the pediatric patient – Artifacts caused by breathing of the patient
category. – General condition of the patient
WARNING NOTE
Risk of patient injury Position the cuff in such a way that no pressure is
exerted on the joints.
The blood pressure measurement must not be
carried out on the upper arm on the side on In the case of patients suffering from spasms, vio-
which a mastectomy has been performed. lent tremors or certain arrhythmias, it may not be
possible to obtain reliable NIBP measurements.
WARNING
Risk of failure of measurements Selecting and positioning the cuff
If a blood pressure measurement and the mea- WARNING
surement of another parameter are carried out
The reliability of NIBP measurements is most-
at the same time on the same limb, the mea-
ly dependent on choosing the appropriate
surement of the other parameter (e.g., IBP,
NIBP cuff. An incorrect cuff size or use of a
SpO2, temperature) may be interrupted tem-
cuff which does not fall within the size range
porarily.
of cuffs manufactured by Dräger can lead to
incorrect measurements. Only use cuffs ap-
NOTE proved by Dräger. The circumference of the
The blood pressure measurement may be com- limbs must always match the marks on the
promised by arrhythmias (e.g., atrial and prema- cuff.
ture ventricular contractions), atrial fibrillation,
compromised tissue perfusion, diabetes, kidney WARNING
disease, or tremors. In the event of implausible
Risk of patient injury
measurements, check the above-noted points
and carry out a further blood pressure measure- If the cuff is placed over damaged tissue and
ment if necessary. If possible, wait a few minutes then inflated, further damage may be caused
before taking a second measurement at the same to the tissue.
measurement point.
Do not place the cuff over damaged tissue.
221
ed or in which a vascular prosthesis is pres-
ent. Zeus IE displays the non-invasive blood pressure
as a numerical value and as a trend. The NIBP
NOTE parameter box shows the status during and after
Position the cuff in such a way that no pressure is measurement. The following is displayed:
exerted on the joints. In the case of patients suf- – Systolic blood pressure
fering from spasms, violent tremors or certain ar- – Diastolic blood pressure
rhythmias, it may not be possible to obtain reliable – Mean blood pressure
NIBP measurements. – Time of measurement
– Bar graph (progress indicator) for the time until
Select cuff according to the circumference of the next measurement
the limbs. The cuff sizes are shown on the cuff Touching the Display all button (requirement: the
(see list of accessories). Display all button has been configured,
If SpO2 and NIBP are measured on the same see page 357) will also display:
limb and the interlock function (see page 249) is – Currently set alarm limits
on, make sure that SpO2 monitoring suspends
the alarm during an NIBP measurement. Extended display of NIBP measurements
Position the cuff as close as possible to the Other displays of the NIBP parameter module can
height of the heart. If not possible, the be configured in the Screen setup > Screen
measured values must be corrected: For every layout > Data dialog window, see page 356:
10 cm that the cuff is positioned above (or – NIBP double size (display twice the width)
below) the height of the heart, 8 mmHg must be – NIBP trend (display as graphic trend over
added or deducted respectively. 30 min)
Do not kink the hose. The extension hose must
not be kinked, squeezed or constricted.
Sensors/Parameters
B C
D I
E J
F
G
301
Or
H
393
145/82 S
160
90
(105)
measurements in progress, the interval elapsed
time meter is reset. A bar graph indicates the time
remaining until the next measurement.
118/81
the restriction of blood circulation, the next
160 Cont.
18 measurement is started after 30 seconds at the
earliest.
S 90
Aborting a running intermittent measurement:
Auto measurement (5 min)
10:12 Touch the NIBP stop button.
NIBP mmHg
Zeus IE cancels the current measurement and
118/81 S
160
90
(94)
5 min
continues the measurement interval with the next
measurement.
118/81 S 160
90
(94)
15 min
Stasis measurement
NIBP mmHg 10:12
Cuff:
118/81 S 160
90
Stasis
86
216
IBP parameters
Par. Pressure Measured value Measuring range
ART Arterial Systolic, diastolic, mean –50 to 400 [mmHg]
LV Left ventricular Systolic, diastolic, mean
PA Pulmonary arterial Systolic, diastolic, mean –50 to 400 [mmHg]
RV Right ventricular Systolic, diastolic, mean
CVP Central venous Mean
RA Right atrial Mean
LA Left atrial Mean
ICP Intracranial Mean
GP1 Unspecific, 1 Systolic, diastolic, mean –50 to 400 [mmHg]
GP2 Unspecific, 2 Systolic, diastolic, mean
CAUTION
If an IBP parameter which is still to be measured
is physiologically almost static (flat), do not use
the SmartZero function. The channels to be ze-
roed must be zeroed individually with the Zero
button instead. Otherwise faulty zeroing of the
corresponding channel will result.
167/69 S
160
90
(106) change in catheter position or pressure
designation.
Sensors/Parameters
CVP mmHg
18 25
5
200 ART
0
220
395
(transport case with patient). Zeus IE adopts The Config. page displays:
the measuring point setting of the pod and the B The measurement curve for the current catheter
previous setting on Zeus IE is overwritten. position
– The pod has no pressure transducer cable C IBP meas. site (ART, LV, PA, RV, CVP, RA, LA,
ICP, GP1, GP2)
connected to the viewed channel (new
D IBP scale: Amplitude setting for the respective
configuration). Preselection of the measuring
measuring point (Auto: Display optimized for
point is adopted by Zeus IE. The pod saves this maximum and minimum values, ±25 mmHg
setting. See also: Designation conflicts. setting only selectable for ICP)
If a pressure transducer cable is connected to a E Parameter with alarm: S, D, M (only for
channel, the appropriate button turns dark green pulsating measuring points)
(active). Open inputs are shown in light green F IBP filter [Hz]: 8, 16, 32 (filter default for the
(inactive). The buttons in the matrix are never gray blood pressure channels)
since the preselection of a measuring point is G The button opens the Alarms > Alarm
limits > Invasive dialog window for alarm limit
possible on any channel at any time.
setting.
Designation conflicts
The pressure designations are saved both in the
Hemo2 and Hemo4 pods and in Zeus IE.
Designation conflicts between a module and
Zeus IE occur when a module with a previously
saved catheter position is connected to Zeus IE.
The pressure designation stored in the module has
priority if a pressure transducer is connected.
Sensors/Parameters Sensors/Parameters
J
H
I
K
L
549
394
Touch the IBP sites/ Channels tab (H). Touch the Cal. tab (I).
Calibration of individual catheter positions
The assignment of measuring points to IBP
J IBP meas. site: (ART, LV, PA, RV, CVP, RA,
channels and connection positions is determined. LA, ICP, GP1, GP2),
Each measuring point can be assigned only once! K Zero individual blood pressure (only
activated when a pressure gauge is connected)
If a measuring point already assigned is then L Calibration factor (only activated when a
assigned to a different channel, the previous pressure gauge is connected)
channel loses that measuring point and is "de-
designated".
CPP measurement
ICP mmHg
99 50
20
CPP
60
Sensors/Parameters PA Auto T
PA Auto
N M
O
R
M
P R U V W X Y
Q
401
S 2 In the Sensors/Parameters > IBP dialog
window, touch the Wedge (M) button.
401
4 Confirm with the rotary knob. If the patient is allergic to latex, use dispos-
able temperature sensors.
The button turns bright green. The value can still be
changed.
CAUTION
The following is displayed in the note field: Press Risk due to using incorrect disinfectants
"Save" to save value in cardiac calculations
and to finish. The sheath of the cable can absorb some disin-
fectants and consequently experience damage.
Saving measured values: – Do not use any disinfectants containing phe-
5 Touch the Save button (W). nols.
– Do not use any highly aromatic agents or any
The PWP value is saved* and the dialog window compounds containing chloroacetone, ether
closed. or ,esters.
If there is no confirmation within 4 minutes, the – Do not immerse the sensor cables for long pe-
PWP value will be saved. riods in alcohol, mild organic solvents, or
strongly alkaline solutions.
Open the Trends/Data > Hemodyn. > Cardiac
calculat. dialog window:
CAUTION
6 Touch the Calculations button (Y). Risk due to excessive temperatures
The PWP value is displayed with the time of Vinyl can withstand temperatures of up to 100 °C,
measurement. but starts to soften above approx. 90 °C.
7 Touch the Save button. – Do not boil or steam sterilize the cables.
– Handle the sensor cable carefully when it is
The PWP value is saved. heated. Wipe from the sensor tip towards the
cable end.
NOTE
Before the reusable temperature sensor is used,
the protective sheath must be pulled over the sen-
sor.
NOTE
The temperature measurement function and the
temperature sensors must be calibrated every two
years by qualified personnel so that a measuring
* Alternatively, the Wedge switch on the hemodynamic accuracy of ±0.1 °C is maintained.
module can also be pressed.
37,7 D
34,0
397
Nasal 39,0
34,0
236
CAUTION
Only for adults and pediatric patients at least 1
year of age! Measurements on younger patients
lead to inaccurate results.
Overview
The measurement of cardiac output is based on the
thermodilution method. A solution whose quantity
and temperature are known is injected in the right
atrium into the blood circulation. The solution mixes
with the blood and cools the blood down. Blood
temperature reaches a low point and then
increases again slowly up to the initial temperature.
The cooling down of the blood is inversely If C.O. measurements are performed
proportional to the blood output of the heart: the successively at different times in the respiration
lower the output, the more the injectate cools down cycle, then the results may deviate from each
the blood temperature. A thermistor in the C.O. other, particularly in ventilated patients.
catheter tip measures the blood temperature after
Measured values which deviate from the
the blood leaves the heart.
general trend or correspond to a very irregular
If the C.O. Start button is pressed on the Hemo curve segment are not considered when
pod after changing the patient data or switching to determining the average.
system standby, the C.O. dialog window will open
If an injectate at room temperature is used,
to check the settings.
10 cc must be injected. Only deviate from this
Starting the measurement: volume when there are important clinical
reasons.
Touch the Manual start button.
Blood output is measured in liters per minute.
C.O. screen display
Zeus IE takes the following factors into account for
the C.O. measurement:
– Volume, temperature, density and specific heat C.O. L/min 11:26
214
mixture
– Area underneath the temperature curve The C.O. parameter box displays:
– C.O. (cardiac output): The saved average value
of the last measuring series (in L/min)
Measuring accuracy – TBlood (blood temperature)
Take the following measures to ensure optimal – Time of the creation of the average value
measuring accuracy: – CI (cardiac index)
Calculation: CI = C.O. / Body surface area
Observe the attached instructions provided by
the catheter manufacturer. Filled syringes or the Touching the Display all button (requirement: the
closed autoinjector should be placed in an ice Display all button has been configured,
bath. see page 357) will also display:
– Currently set alarm limits
Check the temperature of the ice bath. It should
be lie between 0 °C and 5 °C. The accuracy of WARNING
the measurements increases as the injectate Always ensure before a C.O. measurement
temperature approaches 0 °C. that the settings are appropriate for the cathe-
Check the volume of the injectate. ters actually used. Otherwise, measurements
may be unreliable, causing a wrong assess-
Check catheter constant. ment of the patient's status.
If the syringes are filled manually, the same
volume should always be used (normally 10 cc In automatic measuring mode, you must wait for
for adults and 5 cc for pediatric patients). the Inject now message for the injection of the
injectate. The message appears if the initial
Administer the injectate quickly at the end of a temperature of the blood is stable.
respiration cycle.
If all free storage fields are allocated, the oldest In the Sensors/Parameters > C.O./ Calculat.
C.O. curve is deleted or a storage field must be dialog window, touch the C.O. tab (A).
deleted manually:
Sensors/Parameters
Touch the Delete C.O. trial button.
A
C
E
B
F G H D I
457
The C.O. page displays:
B The largest field with the temperature
progression and the calculated value of the
measurement that has just been performed
C The previous measurements are displayed in
the 4 smaller fields
D Manual start: Start of C.O. measurement
E Delete C.O. trial: Entry is deleted
F C.O.average: Display of the continuous mean
value
G C.O. complete: Ending the C.O. measurement.
H Cancel: Cancellation of a measuring series
without saving a new measured value. The
dialog window closes.
I The button opens the Alarms > Alarm
limits > Noninvasive dialog window for alarm
limit setting.
Sensors/Parameters Sensors/Parameters
K P 1 2 3 4 AVG
L J
M Q
N
O
R
398
399
The settings on the Config. page can only be The Cardiac calculat. page contains the results of
modified at the beginning of a new measuring series. four C.O. measurements, derived hemodynamic
The Config. page displays: parameters and the mean values calculated from
K Catheter type: Baxter, Spectra, Ohmeda, them.
Arrow, Other
Hemodynamic calculations
WARNING
The input of a wrong catheter constant can During C.O. measurements, Zeus IE automatically
lead to inaccurate measured results and pos- calculates various derived hemodynamic
sibly to a wrong medical diagnosis. If the cath- parameters, assigns time markers to them, and
eter constant was entered manually, make stores them in a special database.
sure an appropriate catheter is used. The values in the dark fields may be entered and
modified manually.
If catheters from Baxter, Spectra, Ohmeda, or
Arrow are used, the catheter constant is Touch the Save button (R).
automatically specified. The catheter constant The table data is ultimately saved in the calculation
can be changed as needed, e.g., if the volume trends, see page 212. At least the C.O. or PWP
or temperature of the injectate is changed. value must be measured. The time in the
If catheters from other manufacturers (P) are calculation trends is the time when the Save button
used, the selection fields for Catheter size and was touched.
Injection volume are shown in gray. A
computation constant (O) must first be
entered.
NOTE
Neuromuscular Transmission
Monitoring (NMT) Do not use the NMT temperature sensor for mea-
suring core body temperature.
The pulses are sent via two stimulation electrodes
over the nervus ulnaris. The NMT sensor consists NOTE
of an accelerometer (acceleration measuring The NMT module produces high-frequency puls-
transducer) and measures muscle response es. Make sure they do not interfere with other sen-
(muscular blockade) to the pulses. Skin sitive equipment.
temperature is also measured using a temperature
sensor. NOTE
At the start of measurement, the non-relaxed Do not use the NMT pod in the presence of com-
patient is exposed to test pulses in order to bustible anesthetic gases.
determine the supramaximal current. Four pulses
are applied with the determined supramaximal CAUTION
current in order to establish a reference for muscle Only use accessories approved by Dräger. Other-
response in non-relaxed patients. wise, correct functioning may be impaired.
Operating modes: TOF, Single, PTC
WARNING
Single Twitch
Take special care when using electrodes with
A single pulse is applied. The response of the pulse a current density of >2 mA eff./cm2! Only use
is compared to the reference pulse. electrodes which are suitable for use with pe-
ripheral nerves stimulators!
WARNING WARNING
Simultaneous use of HF surgery and the NMT To avoid electric shock all the cables must be
pod on the same extremity of the patient can checked carefully before use. No liquids must
lead to burns caused by the electrodes of the be allowed to drip onto connectors, elec-
pod and to a fault in the pod. trodes or electrical leads.
WARNING WARNING
Do not touch active NMT electrodes. Active Check cables and sensors for damage before
stimulation is indicated on the NMT pod. Oth- use. Otherwise there is a danger of electric
erwise there is a risk that measuring results shock and the possibility of distorting mea-
will be distorted. surement results.
WARNING WARNING
Only attach the electrodes as described. In- Risk of burns
correct positioning of the electrodes can seri-
When fitting the module, make sure there is
ously endanger the patient! Do not attach
sufficient air ventilation/heat dissipation and
electrodes to places where inflammation or in-
that there can be no direct contact between
jury is apparent. Electrodes must not be laid
the patient and the exterior of the module.
across the chest.
WARNING
WARNING
Make sure that there is no contact between the
Risk of patient injury
electrodes and other conducting parts.
NMT modules must not be used on patients
Cover stimulation electrodes with insulating
with heart pacemakers unless this is instruct-
material so that, e.g., catheters are not sub-
ed by a medical specialist. Before using the
jected to stimulation.
NMT monitoring, check whether the stimula-
tion has an effect on the heart pacemaker.
WARNING
When using the NMT pod, maintain a distance
of 1 meter from shortwave and microwave
equipment. Otherwise, correct measurements
will not be possible.
WARNING
With patients who have nervous or neuromus-
cular problems, it may occur that they re-
spond incorrectly to nerve stimulation. Un-
usual patterns can be displayed in the NMT
module while monitoring relaxation.
Trig
In the Sensors/Parameters > BIS/NMT dialog
window, touch the NMT tab (A).
F Sensors/Parameters
C
NMT
EEG
B A
! Monitor
C
D
E
E
D F
A G
B
H I
278
400
1 Attach the black stimulation electrode (A) and The NMT page displays:
the red stimulation electrode (B) to the arm B Stimulation: The NMT start button
(above the nervus ulnaris). C Mode of the measurement: Single, TOF, PTC
2 Connect the pod connecting cable (C) to the D Interval: Time intervals for applying a pulse
Trident pod and to the Aux/Hemo 2/3 In PTC mode, no time intervals can be
selected. In TOF mode, the interval of 1s
connection on Zeus IE.
cannot be selected (grayed out).
3 Attach the NMT sensor (D) to the thumb and E Pulse width [µs]: Duration of the pulse applied
index finger. Secure with a rubber band. in µs
F Current [mA]: When selecting the current
4 Connect the electrodes and NMT sensor to the mode Auto, the button remains disabled (gray).
NMT accessory cable (E). G Current mode: Selection for setting stimulation
5 Connect the NMT accessory cable to the current:
Trident pod (F). Auto: A suitable stimulation current between 5
and 60 mA is determined in conjunction with
each new referencing procedure, ensuring
reliable measurement and minimizing the load
on the patient.
Man: Pulse value which can be set with the
rotary knob from 5 to 60 mA in 5 mA increments
If the value is changed in current mode Man, a
new reference value will be determined!
H Delete reference: Clear button to delete the
saved reference pulse and the supramaximal
current determined.
I The button opens the Alarms > Alarm
limits > Noninvasive dialog window for alarm
limit setting.
In the Sensors/Parameters > BIS/NMT > NMT A single pulse is applied. Muscle response to this
dialog window, touch the NMT start button (J). pulse is measured and compared with the
reference pulse.
Or, if the NMT start button is configured
(see page 357): The NMT pod repeats stimulation at the set time
interval.
In the main menu bar, touch the NMT start
button. If the value for stimulation current or pulse width is
changed, a new reference value will be determined!
Sensors/Parameters
Settings on Zeus IE for Single Twitch
J Select Single measuring mode (K) and confirm
with NMT start button (J).
K
NMT screen display
NMT %
Ratio Temp °C
100
90 Single
11:34
421
4
0 0
At the start of measurement, the NMT pod emits 213
Touching the Display all button (requirement: the TOF ratio screen display
Display all button has been configured,
see page 357) will also display:
– Currently set alarm limits NMT %
100 Ratio Temp °C
48
TOF
NMT %
100 Temp °C
Single
0
4
0 11:38
209
4
15 min
0 0 Muscle response to each individual pulse is
measured and displayed as a bar graph.
210
Skin temperature is displayed. The ratio of the fourth pulse to the first pulse, T4/T1,
is displayed as a TOF ratio in %, e.g., 48.
60 seconds after a successful single or TOF
measurement, the measured value is deleted.
NMT %
Temp °C
100
TOF TOF
Train of four – A series of 4 pulses is applied. The 4 5 min
number of pulses answered indicates the degree of 0 0
208
muscle relaxation. The NMT pod repeats
stimulation at the set time interval. After 15 seconds, the last measured value is
deleted. The time until the next measurement is
displayed as a bar graph.
TOF settings on Zeus IE
J NMT %
Temp C
100 Count
L
1
TOF
0
4
0 11:39 207
Select TOF measuring mode (L) and confirm For TOF count, an upper and a lower alarm limit
with NMT start button (J). can be set, see page 186.
1
PTC
WARNING
Only perform on a patient under sufficient se- 4
0
dation when there are no further TOF respons-
205
es! PTC stimulation on a conscious patient is
very painful. The number of pulses answered is displayed. While
measurement is taking place the measured value is
PTC settings on Zeus IE grayed out.
Sensors/Parameters
NMT %
Count Temp °C
10
J PTC
M 4
0 11:41
206
The measured value remains up to the next
measurement (Auto mode is not possible).
BIS monitoring
488
403
This table reflects a general association between
clinical state and BIS values. The BIS page displays:
B EEG waveform display
The areas are based on a multi-center study. C EEG scale: Selection of the scale for the EEG
waveform in µV
For the BIS values and BIS areas, it must be
D Smoothing rate [s]: Selection of the smoothing
assumed that they are free of artefacts that could
rate used to calculate the BIS value average
influence the measurement.
– 15 s (basic setting): Increased
BISx technology is the latest generation of BIS responsiveness to changes (e.g., induction
technology, which features enhanced screening or waking).
against artefacts from external HF surgery – 30 s: A smoother trend (with reduced
equipment and filters out any EMG influences. variability and sensitivity to artefact)
E Filter:
The BIS index provides insight to the cerebral
– On: Filter out interference from the raw EEG
status and is not a measurement of drug
signal
concentration. The cerebral cortex responds very
– Off: Display of the raw EEG without filter
sensitively to external influences, drugs and
F The button opens the Alarms > Alarm
pathological processes. The BIS index decreases limits > Noninvasive dialog window for alarm
during normal sleep and also during administration limit setting.
of hypnotic drugs.
However, the values during normal sleep do not fall Displaying all BISx parameters
as low as those during anesthesia.
Touch the All EEG values button (G).
The BIS index is determined as an average over a
period of 15 seconds and generated every second.
WARNING NOTE
If a BISx pod and a defibrillator are being used Artefacts and inferior signal quality can lead to in-
simultaneously, do not position the BIS sen- correct BIS values. Possible artefacts can be
sor between the defibrillator and the patient! caused by inadequate skin contact (high imped-
Risk of electric shock! ance), muscular activity or rigidity, head and body
movements, repeated eye movements, incorrect
WARNING sensor positioning and unusual or excessive elec-
trical interference.
When fitting the module, make sure there is
sufficient heat dissipation and that there can
be no direct contact between the patient and NOTE
the exterior of the module. Risk of burns! Carefully interpret BIS values for:
– Patients with known neurological diseases
– Patients taking other psychoactive drugs
WARNING
– Children under one year of age
Operation of the BISx module during magnetic
resonance imaging (MRI) is not supported at
present.
WARNING
To prevent strangulation risks the BIS sensor
cable must be positioned and attached care-
fully.
CAUTION
A correctly positioned BISx module is attached to
the infusion apparatus stand, bars of the bed or
the bed sheet.
CAUTION
In the event of a continuous reduction in BIS sig-
nal, skin irritations or intense perspiration replace
the sensor more frequently.
NOTE
When interpreting the BIS together with other clin-
ical signs available, a clinical assessment must al-
ways be used in addition. Solely relying on the BIS
when administering intraoperative anesthetic
agents and sedatives is not recommended.
1
2
4
X
BIS
284
275
CAUTION
5 Connect the patient connecting cable (A) to the
Only use the BISx Quatro sensor (for adults) or
sensor. Lay and secure the cable carefully.
the BISx Pediatric sensor (pediatric patients).
Otherwise, incorrect measuring results will be ob- 6 Connect the pod connecting cable (B) to the
tained. Aux/Hemo 2/3 connection (see page 226) or to
the Trident NMT pod (see page 268).
1 Observe the instructions and numbering of
7 Attach the BISx pod to a bracket or infusion
electrodes on the sensor!
apparatus stand using the clip on the back of
2 Position the adult sensor at an oblique angle the pod.
across the forehead:
– Electrode 1 approx. 5 cm above the nose
– Electrode 4 above the eyebrow
– Electrode 3 between the corner of the eye
and the hairline
3 For good contact on the skin, carefully press all
electrodes onto the forehead with a fingertip for
at least 5 seconds.
4 Fix the sensor in place with adhesive tape or
plaster.
NOTE
B An automatic sensor impedance check is per-
formed whenever a BIS sensor is connected to the
C A sensor cable.
D NOTE
E When the sensor impedance check has been com-
pleted, all signal status messages and alarms are
F suspended for 60 seconds. The time is required in
G order to restore signal quality.
402
A B 1:35
In the event of an alarm and device-related failures,
the corresponding alarm message appears in the
alarm message field (A).
301
If the mains power supply fails, Zeus IE 1 Press the O2 rotary safety knob to unlock it for
automatically switches over to the integrated emergency O2 delivery and adjust it to the
uninterruptible power supply (UPS). required O2 flow. Range 0 to 12 L/min. Use
pressure gauge to monitor the pressure.
With a fully charged battery, operation is
maintained for at least 30 minutes. During this phase, anesthetic agent delivery is not
possible!
The Mains supply failure message appears in the
alarm message field. 2 Ventilate the patient manually!
The remaining battery capacity is displayed in 3 Recharge the empty battery by plugging in the
percent in the system field (B). power cable for at least 10 hours.
Example: 82 % 4 If the power supply is restored within 5 minutes,
the device will behave as follows after switching
If the battery capacity is low, one of the following
on:
messages appears:
– Battery low: Battery capacity <20 % – Ventilation is continued with the previous
settings.
Or
– Battery very low: Battery capacity <10 %. The – The gas delivery system administers an O2
device may switch off at any time! flow of 0.5 L/min, but with no anesthetic
agent.
In the event of a power failure and discharged
batteries, Zeus IE enables manual ventilation and – The Safety settings active alarm is
spontaneous breathing with 100% O2. displayed.
– The ventilator, gas and anesthetic agent
delivery as well as monitoring are
inoperable.
– The DIVA metering modules are locked.
CAUTION
Risk of patient injury
The Basis (IVDock) is no longer supplied with
power. Each Module DPS is powered by its own
battery. Pay attention to the battery warning! Infu-
sion is automatically stopped when the battery is
empty. Charge the battery as soon as possible.
Gas failure
WARNING
Zeus IE displays the following messages in the
event of a gas failure: Risk of contamination of the gas supply
Ventilator failure
If gas delivery is switched off due to internal device Zeus IE displays the following alarm message in
errors, the FG ISO button (example) see page 133 the event of anesthetic gas delivery failure:
or etISO button (example) see page 135 will no
VA delivery failure
longer be available. If necessary, select the FG ctrl.
mode and set a low fresh-gas flow. Continue anesthesia with IV agents
Ventilation is still operational. Contact DrägerService!
If gas delivery and gas mixing are switched off due Anesthetic agent is not delivered!
to internal device errors, ventilation remains
After switching off the device, return the rotary
operational.
safety know for O2 to the original position and push
Zeus IE indicates the failure with: it in.
DIVA+mixer failure
Press the rotary safety knob O2 to release it and
turn it to adjust the O2 flow.
The breathing bag fills; the filling process can
be heard.
097
1 Press the rotary safety knob O2 (B) to move it
If ventilation and delivery have to be immediately
out.
canceled due to internal device errors:
2 Turn the O2 (B) rotary safety knob and adjust
1 Press the rotary safety knob Anesth.
the O2 flow.
Ventilator Off to move it out.
The breathing bag fills; the filling process can
2 Turn the rotary safety knob Anesth. Ventilator
be heard.
Off (A) fully counterclockwise.
Anesthetic agent is not delivered!
– Ventilation stops.
– Anesthetic agent delivery is locked. After switching off the device, return the Anesth.
– Fresh-gas delivery is stopped. However, Ventilator Off (A) and O2 (B) rotary safety knobs
0.5 L/min of O2 will continue to be delivered. to their original positions and push them in.
– Patient monitoring remains activated.
Zeus IE indicates the failure of the following Reboot: This takes approx. 30 seconds to
components: 3 minutes. Monitoring is unavailable during this
time.
– DIVA+mixer failure
– Ventilation continues (fresh-gas mode at
– Patient monitoring fail
0.5 L/min; with optional ext. FG outlet: 3 L/min).
A – Check ventilation with manual pressure gauge.
Perform manual ventilation if necessary.
326
CAUTION
Touch the Software reset? button (A) in the header Risk of patient injury
bar if necessary.
Anesthetic gas is not delivered.
The device performs a warm start.
If the device no longer responds or if the screen Anesth. Ventilator Off means:
fails, the following must be observed: – Ventilation: Manual/Spontaneous
– Gas delivery: 0.5 L/min O2
CAUTION – Anesthetic agent delivery: Off
Based on the mechanical airway manometer, 1 Press the rotary safety knob O2 (B) to move it
check whether ventilation continues. Manually out.
ventilate the patient if necessary!
2 Turn the O2 rotary safety knob (B) and adjust
If necessary, press the rotary safety knob Anesth.
the O2 flow.
Ventilator Off to move it out and activate by turn-
ing clockwise to switch the ventilation and gas de- 3 If necessary, fill the breathing bag via the O2+
livery to manual emergency operation. O2 flush button.
Ensure anesthesia. No anesthetic gases are de- Contact DrägerService and use a substitute
livered. Patient risk! anesthesia machine.
If Zeus IE is unable to operate a Module DPS – Incorrect configuration, see page 163.
syringe pump, the reasons can be as follows:
Observe Module DPS Instructions for Use.
– Module DPS is not a Dräger syringe pump and
A syringe pump, which has stopped and is already
has a software version which is not compatible.
or subsequently connected to Zeus IE, generates
Zeus IE cannot operate it. The syringe pump
the alarm message Infusion 1 stopped on Zeus IE
continues to run.
after one minute.
Zeus IE messages
The alarm messages are displayed in the message Alarms are displayed according to priority,
field of the header bar in hierarchical order. downwards from top to bottom and from left to right.
Different background colors indicate the priority Low-priority alarms are only displayed in the right-
levels of the alarms. hand field of the alarm message field, the maximum
being four alarms.
Zeus IE divides alarm messages into three
categories by priority:
Warning (red) High priority Special note on DIVA filling level alarms
Caution (yellow) Medium priority When DIVA module 1 is active, the DIVA module 1
Note (cyan) Low priority level low alarm is issued with medium priority.
When DIVA module 2 is active, the DIVA module 1
To classify the alarms within an alarm category, level low alarm is issued with low priority.
internal priority figures are indicated in the following
When DIVA module 1 is active, the DIVA module 1
table. The most critical alarm is given the number
empty alarm is issued with high priority. When
255. Lower numbers indicate a lower alarm priority.
DIVA module 2 is active, the DIVA module 1
In the tables for Alarm info and Alarm log, empty alarm is issued with medium priority.
see page 184, the priority of the alarm messages is
The same behavior also applies correspondingly
also indicated by exclamation points.
for the DIVA module 2 level low and DIVA module
Up to four high or medium-priority alarms from the 2 empty alarms.
following four alarm areas can be displayed
simultaneously:
– Anesthesia
– Hemodynamics (optional)
– Ventilation
– Infusion (optional)
From each alarm area at least one high-priority
alarm is displayed, given that there is an active
alarm.
In operation
Configuration
Configuration information
Calling up during
operation or in
system standby Zeus IE settings
Load/Save Load/Save
Profiles View
(5 factory profiles, up to (1 factory view, up to
16 user profiles) 16 user views)
248
System setup
1200
G H I J K L B
C
D
0
0 20
2
mL/mbar mbar/L/s
1
4
A E
2
3
F 3
3 4
410
The Manage setups page displays two vertical
311
tabs:
In the main menu bar, touch the System
– Profiles (B)
setup... (A)button.
– Views (C)
The System setup dialog window appears.
Touch the Profiles tab (B).
The following settings can be configured in the
The page for setting the profiles with the following
System setup dialog window:
options appears:
– Manage setups
D Button for deleting the assigned order of the
– Sensors/ Parameters
profile keys
– IBP
E Five buttons (profile keys) for determining the
– MixGas/ Ventilation order of the profiles for the start dialog
– Infusion F Buttons for preset management
– Alarms
– Screen layout The following is displayed for each profile:
– Basic settings G Profile name: Dräger Neo., Dräger Ped. 1,
– Autoset limits Dräger Ped. 2, Dräger Adult 1, Dräger Adult
2 or user-defined profiles
H Categ.: Symbols for Adult , Pediatric ,
Neonatal
Manage setups for Profiles and Views I Key: The assigned profile key
J Last change: Date and time
In the main menu bar, touch the System K Default: Indication whether the profile is preset
setup... button. (tick)
The System setup > Manage setups dialog L Protected: Indication whether the profile is
window will appear. saved or access code protected (lock icon)
M
1
313
2
Touch the Views tab (C). 3
4
The page for view settings with the following
options appears: A 5
410
M Buttons for preset management
The following is displayed for each view: 1 Select the desired profiles/views from the
N View name Profile name list or View name list using the
O Last change: Date and time rotary knob.
P Protected: Indication whether the profile is 2 Touch the Load button (A) and confirm with
saved or access code protected (lock icon) rotary knob.
The saved settings of the profiles or views selected
overwrite all current settings. Profiles can only be
changed in standby mode. Views can be adapted
during operation. The screen shows the name of
the selected profile in the patient data field (below
the patient name), see page 50.
CAUTION
The applicability of the data from the loaded con-
figuration must be checked individually by the
user for each patient to eliminate the risk of incor-
rect settings and ensure the patient's safety.
The Manage setups > Profiles > or Views page is The Manage setups > Profiles > or Views page is
open. open.
1 1
2 2
3 3
4 4
A 5 A 5
410
410
If the profiles/views have not been access code 1 Select the desired profiles or views from the
protected, profiles/views that already exist can be Profile name list or the View name list with the
overwritten using Save. rotary knob.
Touch the Save button (A) and confirm with 2 Touch the Rename button (A).
rotary knob.
A keyboard with the current name opens.
3 Change the name and confirm with rotary knob.
The required name is thus activated.
The Dräger default Profiles/Dräger default
views cannot be changed.
The Manage setups > Profiles > or Views page is System setup
open.
System setup
B
1
2
1
A 3
4
2
5
3
410
4
The Manage setups > Profiles page is open.
A 5
410
1 All settings saved to the new profile or to the Assigning profile keys
new view must be checked before saving and
Use the vertical buttons (A) to assign the profiles to
corrected, if necessary.
the 5 profile keys in the Start dialog
2 Touch the Profiles or Views tab. (see page 127).
3 Touch the New button (A). 1 Use the rotary knob to select the required profile
from the list.
A keyboard field appears.
2 Select button 1 to 5 and confirm with rotary
4 Give the profile or view a meaningful name
knob.
(e.g., Adult) and confirm it with Enter or by
pressing the rotary knob.
Deleting assigned profile keys
The name appears as the current profile in the
patient data field (below the profile name), The button (B) can be used to delete the assigned
see page 50. profile key. In this case, the corresponding profile
key will not appear in the Start dialog
(see page 127). A profile marked with Default
cannot be deleted from the startup dialog.
1 Select the required profile with the rotary knob.
2 Touch the (C) button and confirm with rotary
knob.
The profile key does not appear in the startup
dialog.
The Manage setups > Profiles > or Views page is The Default property defines the default setting
open. which is loaded when the device is switched on (not
with Views).
System setup
Only profiles that are assigned to a profile key can
be defined as the default.
The Manage setups > Profiles page is open.
System setup
1
2
3
4
A 5
1
415
2
1 Touch the Profiles or Views tab.
3
2 Use the rotary knob to select the profile or view 4
to be deleted from the Profile name list or View
name list. A 5
410
3 Touch the Delete button (A) and confirm with If a specific start setting is required:
rotary knob.
1 Use the rotary knob to select the required profile
The required setting is then deleted. Access code from the Profile name list.
protected profiles or views must be unlocked
2 Touch the Default button (A).
beforehand!
3 Enter access code and confirm with rotary
Dräger default profiles/Dräger default views and
knob.
the loaded profile or view currently active cannot be
deleted. The Default button is now dark green. This profile
is preset after switching on the device and when
calling up new patient data (default setting).
With the Protect property, the profile or view setting System setup
cannot be renamed, deleted or saved.
A B C D E F G H
The Manage setups > Profiles > or Views page is
open.
System setup
1
I
412
2
The other horizontal tabs show the parameters and
3 presets of:
4
A Sensors/ Parameters (see page 226)
A 5
410
1:35
Screen layout
Screen setup
A
A B
C
D
E
301
432
The Screen setup > Screen layout dialog window
will appear. The Screen layout page (A) shows the settings for:
The Screen layout > General page (A) is open. 1 Touch the On button for Display mini trends
(G).
Screen setup
If the Off button under Additional monitoring field
has not been activated:
B A – The parameters currently set are displayed as a
C mini-trend between the current curve and the
loop display and flow tubes.
D
E If the Off button under Additional monitoring field
has been activated:
F
G – The parameters current set are displayed as a
mini-trend on the left next to the current
H
parameter curves.
432
433
2 Touch the Off button for Additional monitoring A maximum of 7 curves can be selected.
field. Exception: With the IBP comb. module, 4 IBP
curves are displayed in one field double the height.
Loops, mini-trends and flow tubes are not
displayed. 1 Touch the Waveforms tab (A).
If the Display mini trends option has been 2 The arrow keys ( ) can be used to add
activated, the parameters current set are displayed curves to the configuration from the Available
as a mini-trend on the left next to the current list (B), or to remove curves from the
parameter curves. configuration via the Displayed list (C).
For the display of the ventilation loops, The Up or Down buttons (D) can be used to
see page 223 and for that of the flow tubes, configure the position of the curves on the
see page 225. screen.
The loops are always displayed in combination with 3 Touch the Apply changes button (E) to apply
the flow tubes. the changes.
Screen setup
A B
B C
A
D
474
E
Web access is activated on the screen in the
447
C
Repositioning buttons in the main menu bar
B D
The associated buttons are repositioned in the
respective group in the main menu bar.
1 Touch the Change button position button (D).
E
All buttons in the main menu bar are displayed in
418
yellow.
Touch the Customize buttons tab (A).
Select the button to be moved from the main menu
A list (B) of all configurable buttons opens. The
bar.
following buttons can be selected:
– Open dialog: Screen setup The color of the selected button remains yellow and
– Open dialog: Brightness/Volume all other buttons turn green.
– Open dialog: Help
2 Move the selected button and confirm with the
– Open dialog: New patient
rotary knob.
– Open dialog: Patient setup
– Open dialog: Sensors/Parameters
– Open dialog: System setup Removing buttons from the main menu bar
– Open dialog: Web access
1 Touch the Remove button button (D).
– Open dialog: Procedures
– Direct action: Zero all BPs All buttons that can be deleted are displayed in the
– Direct action: Stop all pumps main menu bar in yellow.
– Direct action: Display all
2 Touch the respective button.
– Direct action: Print screen
– Direct action: Apnea alarms off The button is removed from the main menu bar.
– Direct action: CO2 alarms off
– Direct action: Print case
– Direct action: CBM mode on
– Direct action: Freeze waveforms
– Direct action: CV alarms off
– Direct action: Neuro alarms off
– Direct action: NMT start
– Direct action: Switch to next web application
– Direct action: SmartPilot
– Direct action: Recorder start
– Direct action: Confirm technical alarms
– Web access 1 to 4
Screen setup
Assigning or changing the colors for individual
parameters.
A
Screen setup
B A
B C
C
D D
434
420
Touch the Brightn./ Volume tab (A).
1 Touch the Colors tab (A).
The following settings can be configured:
B Screen brightness [%] 2 Select the respective parameters from the list
C Alarm tone volume [%] (B).
D Pulse tone volume [%] 3 Touch the required color (C).
WARNING All parameters with the same color are listed under
Risk of not hearing the alarm tone Assigned params (D).
A
Access: Basic settings
300
304
1 Touch the Biomed service... button (A) in
system standby.
Entering the access code
The Biomed service dialog window opens.
1 Touch the numeric keys.
2 Then press .
Biomed service
B
C
D
E
F
303
423
423
The horizontal tabs (A) for the basic settings The General settings page (A) displays the
contain the settings for: following vertical tabs:
– General settings B General
– Infusion C Units
– Event recorder D Printer
– Web access E USB devices
– NIBP/SpO2 F Alarm signaling
– Infinity config. G Date/Time
– MediBus config. H Trend order
– Disable options I Access codes
– More settings
More settings are possible on the vertical tabs (B).
CAUTION
Basic settings
Computer viruses
Mass storage devices can contain computer vi-
ruses which can impair the operation of Zeus IE.
A C – Check mass storage devices for computer vi-
D ruses before using them with Zeus IE.
E – Do not use any mass storage device contain-
ing computer viruses with Zeus IE.
F
Touch the Export/Import of setups button (F).
G
B The Export and import of setups dialog window
423
appears.
The General settings > General page is open. Export and import of setups
The following settings are possible:
G
A Language: User interface language selection
H I
NOTE
Select only languages which can be understood!
448
These buttons are used to define the coordinate
axes for the flow/volume loop display. When 1 Touch the Export/ Import tab (G).
ISO is selected, the display is the inverse of the
A message with corresponding instructions is
Dräger display.
displayed.
E Access to colors/ customize buttons:
2 Insert the USB stick into the USB port.
Allowed, Denied
F Access to config. profile keys 1-5: Allowed,
Exporting setups:
Denied
If Denied is active, the profile key assignment Prerequisite: The USB stick has been installed,
cannot be changed and the profile keys cannot see page 365
be seen in the System setup dialog window,
Touch the Export setup to USB stick button
see page 347.
(H) and confirm with rotary knob.
G Export/Import of setups: A USB stick can be
All profile and view setups are exported to a USB
used to export and import profile and view
stick.
configurations, as well as basic settings.
327
J 1 Touch the Basic settings tab (N).
options.
After selecting the Install views or Install profiles Touch the Press to reboot button (P) and
tab (J), all available profile and view setups are confirm with rotary knob.
displayed.
1 Select the respective views or profiles from the
Imported views or Imported profiles (K) and
Installed views or Installed profiles (L) lists.
2 Touch the Install button (M) and confirm it with
the rotary knob.
The selected views or profiles applied to the
system.
Export and import of setups The page General settings > Units is open.
Q Basic settings
R
A F
B G
C H
D I
449
E J
A complete configuration is applied to the system
446
from the USB stick without having to individually
select views or profiles. The existing profiles or The following settings are possible:
views are deleted.
A N2O: %, kPa
1 Touch the Install all tab (Q).
B CO2: %, mmHg, kPa
A message featuring information regarding the
software version to be installed or information about C Volatile anesthetic agent: %SL, kPa
which configurations are installed appears. D Gas measurem. correction: BTPS, normal
2 Touch the Install all button (R) and confirm with E Gas supply: kPa, kPa x 100, MPa, bar, psi
rotary knob.
F Temperature: °C, °F
The process is cancelled if the menu changes to
the startup dialog or the device test is activated. G NIBP, IBP: mmHg, kPa
The drug library is not part of the profiles and must H Weight: kg, lbs.
be imported separately, see page 172. I Height: cm, inch
J Airway pressure: hPa, kPa, mbar, cmH2O
Printer CAUTION
If office equipment or components such as a print-
The General settings > Printer page is open.
er or a network switch are used in conjunction with
Zeus IE, these must conform to protection class 1
Basic settings
and meet the requirements of IEC 60950-1.
Such devices or components must not be con-
nected to the auxiliary Zeus IE power sockets.
They may only be connected via Ethernet.
These devices or components must be located
outside of the patient environment (1.5 m from the
patient bed or operating table) and must not be
touched by the user at the same time the patient
is also in contact.
If ignored, it can put the patient and user at risk.
443
The list of available printers (defined in the With regard to the following printers, Dräger has
operating system) appears. The settings are set by given approval confirming correct data exchange:
DrägerService. – Zeus IE with HP Laser Jet 4250N
The printer selected here only prints the current – Zeus IE with HP Laser Jet 4015N
screen.
– Zeus IE with HP Laser Jet 600
The network printer, which is connected via the
Infinity network, can print out all graphic and tabular CAUTION
trends. Do not touch electrical contacts of connected data
cables and patients simultaneously. The electric
CAUTION
data cables could place the patient at risk or mea-
Danger resulting from incorrect connection of the surement signals could be affected.
printer
If incorrect printer cables are used, these cables CAUTION
may cause the leakage current to exceed the per- If additional medical devices are plugged into the
missible requirement for medical devices and auxiliary power sockets, the user must ensure
may present a hazard if the patient comes into that the requirements of IEC 60601-1-1 are met.
contact with them.
Plugging medical devices into the auxiliary sock-
If the printer is connected to the LAN port, use ets can bring about a considerable increase in dis-
printer cable MK04638. Connect the printer cable charge current above permitted levels.
corresponding to the labeling for the cable on the
printer. If ignored, it can put the patient and user at risk.
CAUTION
Unsupported USB devices
USB devices that are not supported or have too
high a current consumption can impair correct
equipment operation.
Only use supported USB devices with a maximum
314
current consumption of 500 mA. Starting the installation:
CAUTION
Computer viruses
Mass storage devices can contain computer vi-
ruses which can impair the operation of Zeus IE.
– Check mass storage devices for computer vi-
ruses before using them with Zeus IE.
– Do not use any mass storage device contain-
ing computer viruses with Zeus IE.
L
E Sorting trends
A F
B G The General settings > Trend order page is open.
C Basic settings
H
D I
K J
A A A A A
441
493
adjusted to these settings and the settings are
transferred to the Infinity Network (if available). On the Trend order page, each column (A)
corresponds to a certain tabular trend. The order in
C Date format: Default, M-D-Y, D-M-Y which the parameters in the respective trend are
D Time format: Default, 12, 24 displayed can be configured with the buttons (B).
E Constant readiness for operation: On, Off 1 Touch the Trend order tab (C).
When the function is activated, a continuous 2 Touch the parameter or select it with the rotary
working day is defined. The gas measurement knob.
is then permanently ready for use. The buttons
for the start and end of the working day are not 3 Set the desired order with the buttons (B).
active.
F Operating time begins at: Enter the time
G Operating time ends at: Enter the time
H Time-controlled self test starts at: Enter the
time
I Time-controlled self test: On, Off
J Autostart self test after power up: On, Off
The General settings > Access codes page is The Infusion > General page (A) is open.
open.
Basic settings
Basic settings
A
1 2 3
B
A 4 5 6
B 7 8 9
C 0
.
D
600
For all syringe pumps, the standard alarm limit can
610
1 Select the button for which the access code is be configured for each syringe size:
to be changed: B Pressure
– Profiles, views (A) Touch the corresponding button and set the
– Basic settings (B) desired value.
– Service mode (C)
– Drug library (D) The standard alarm limits are automatically
activated if the drug is changed during operation.
2 Enter the new access code and confirm with the For setting ranges, see "Alarms > All limits >
rotary knob. Infusion" on page 199.
3 Repeat the new access code and confirm with
the rotary knob.
A green check mark will be displayed if the
password change was successful.
If the password could not be changed (e.g.,
because a different password was entered as the
repeated password), a red "X" will be displayed.
Contact service personnel if the access code for the
basic settings is lost.
Basic settings
A B C
440
All setting changes and events are documented.
601
The Access code dialog window opens. The 4- Use the buttons (A) and (B) to scroll up and down
digit access code can be entered. in the log book.
For the full description, see "Drug library" Use the Refresh button (C) to display the last event
on page 172. in the log book.
Configuring web applications E Display scroll bar: Turn the side scrollbar on
and off in a web application
The options for web access make it possible to F All web applications are displayed on the
display and run applications from the hospital selected screen
intranet. An option is required for each application.
G Tab with the available web applications (Web
CAUTION acc. 1 - Web acc. 4). The pages have an
Only have network functions configured by identical layout.
trained personnel! Errors in configuration can im- 1 Touch the (A), (B), (C), (D) buttons.
pair operation of the device and the network.
The keyboard field to enter names opens.
CAUTION 2 Enter the data for all other web applications.
The Zeus IE includes a firewall for operation in a
hospital network in order to provide the maximum Configuring buttons to call up web applications
level of network security. In addition, the user is directly
required to implement and maintain normal state-
of-the-art security and precautions in the network. The following buttons must be configured to call up
If not, device performance may be affected. web access:
– Web access... to call up the web application
The Web access > Web acc. 1 page is open. settings during operation
– Next web application to toggle between the
Basic settings active application windows directly
– Buttons (name freely selectable) to call up a
specific web application directly
To open the dialog window and configure the
A G Screen setup > Web access buttons,
B see page 357.
C
D
E
F
413
The Basic settings > NIBP/SpO2 page is open. The Infinity network connects monitors and other
Dräger medical devices to the network center and
Basic settings to each other, offering a variety of different
monitoring functions.
Depending on the software version number and the
A monitors connected, incompatibilities may occur in
B the network. The range 192.168.0 is not
C permissible when configuring the IP address. Not
D all alarms and functions of the Infinity network are
supported.
Further details are available from your Dräger
product specialist.
E
439
NOTE
The sensor type can only be switched over if the
sensor is not plugged in.
A F L
B G M
C H N
D I O
E J
K
L P
489
490
The following settings can be configured: The following settings can be configured:
A Bed name L Recorder 1
B Care unit M Recorder 2
C Hospital name N Primary recorder
Configuring MediBus The Set default button (I) resets MediBus to the
default values:
The MediBus config. > COM 1 page is open. – Port: Off
– Baud rate: 9600
Basic settings – Parity: Even
– Stop bits: 1
– MIB/IDS filter: On
A G – Commands to MediBus device: On (COM 1)
– MediBus version: SW 4.n
B Check the following device settings in the
Device identification list:
C
– Device number
D – Device name
E – Device version
F – MediBus version
I H
438
The Basic settings > Disable options page is The following settings can be configured in the
open. Basic settings > More settings dialog window:
– Gas prices
Basic settings – Ventilation settings (ventilation settings)
– Ventilation modes (ventilation modes)
– VA/N2O (volatile anesthetic agent/N2O)
A F – Calculat. settings (calculation settings)
B G – Total consumpt.
C H
D I Gas prices
E J
The More settings > Gas prices page is open.
K L Basic settings
452
An accurate overview for calculating the anesthetic C MV alarms during cardiac bypass mode:
agent costs incurred per operation is displayed in Yes, No
the Trends/Data > Gases/ Anesth. > Total
Configuration option to specify whether in CBM
consumpt. dialog window (see page 212).
mode you want MV monitoring to remain
activated or not.
Ventilation settings D Alarm priority for alarm "Apnea ventilation":
Medium, Low
The More settings > Ventilation settings page is Definition of the priority of the Apnea
open. ventilation alarm in the Pressure Support and
Smart Ventilation Control ventilation modes
Basic settings
E CO2 auto wake-up (breaths) Set the number
of breaths (CO2 Auto Wakeup) for activating the
A CO2 alarm.
B
C
D
E
444
The More settings > Ventilation modes page is The More settings > VA/N2O page is open.
open.
Basic settings
Basic settings
A
A B
B C
C
D
E
454
453
The More settings > Calculat. settings page is C The All button: All waveforms are displayed.
open. The various waveform colors are assigned in a
legend.
Basic settings
D The Active button: Only the enabled
waveforms are displayed. The various
VT R waveform colors are assigned in a legend.
1500
1250
1000
80
60
C
D
750 40
500
20
250
0 0
2.0 15.0 65.0 100.0
Weight
A
B E
VT
f
455
The More settings > Total consumpt. page is The More settings > Proced. page is open.
open.
Basic settings
A B A
B
C
D
E
C D
602
456
Default maneuver One-step recruitment, Multi-step recruit- Determines the standard maneuver.
ment
Displayed parame- Cdyn, VT Determines which additional parame-
ter ters in the One-step recruitment and
Multi-step recruitment dialogs will be
shown.
Pressure rise per 2 to 10 2 to 10 2 to 10 Determines the pressure increment by
step which PEEP and Pinsp are gradually in-
5 4 3
[mbar]; [hPa]; creased.
[cmH2O]
Pressure de- 2 to 10 2 to 10 2 to 10 If PEEP >15: Determines the pressure
crease per step increment by which PEEP and Pinsp are
5 4 3
when PEEP > 15 gradually reduced.
[mbar]; [hPa];
[cmH2O]
Pressure de- 1 to 10 1 to 10 1 to 10 If PEEP 15: Determines the pressure
crease per step increment by which PEEP and Pinsp are
2 2 2
when PEEP ≤ 15 gradually reduced.
386
375
375
?Help
Touch the button in the header bar (A) for:
A B C D E – Date
– Time
– Location and stopwatch
The Stopwatch dialog window appears.
Stopwatch
B E F D
C G
387
388
one.
C Forward: The pages can be turned forward one The following settings can be configured:
by one. Countdown:
D Page up: The displayed page is moved down. B Countdown timer
C Countdown time: Format: H:MM:SS
E Page down: The displayed page is moved up.
F Reset: The start time is entered into the timer.
G Start: The countdown is started.
H Stop: The countdown is stopped.
I Silence button: The alarm tone can be switched
on and off when 00:00:00 is reached. The
button (F) can be used to turn off the alarm tone.
The alarm is also displayed visually in the
header bar.
Triggered alarms can be deleted again via
Reset.
Stopwatch:
151
J Stopwatch
K Start: The clock runs forward from 00:00:00
after starting.
L Stop: The clock is stopped.
M Reset: The clock is reset to 00:00:00.
The stopwatch, the timer function, and the
status of its alarm are displayed in the header
bar after the function has been started. The
display disappears again after the function is
reset.
N Clock size: Toggle the display size between
Large and Small
Large size: Standard display of the time and
the duration of the surgery. If the countdown
counter or the stopwatch is running, the
duration of the surgery is replaced by the
corresponding value.
Small size: Standard display of:
– Location
– Date
– Time
– Duration of the surgery
If the countdown counter is running, the date is
replaced by the countdown counter. If the
stopwatch is running, the time of day is replaced
by the stopwatch.
If the countdown and stopwatch have been
activated at the same time, the display switches to
Small regardless of the preset display.
Service functions
Service call dialog window C Time of error: Touching the button opens a
dialog window to enter the time of the event.
Use the Service Call function to communicate with D Description: Touching the button opens a
the Remote Call Server at DrägerService. dialog window to enter a problem description.
Prerequisites: E Remote Device Check / Remote Inspection:
Touching the button establishes the connection
– Remote service is enabled. to the Remote Call Server for agreed, regular
– Zeus IE is provided with a service contract data transfers without any current problem.
covering use of the Remote Service System F Help ticket: Touching the button establishes
(Service call). the connection to the Remote Call Server for
current events with acute need for action.
– The settings for external communication have
been configured by DrägerService.
– Zeus IE is connected to a data network Service mode and service access dialog
(Ethernet), see page 34. windows
WARNING
Service call
Calling up the service mode is prohibited
C D A when a patient is connected. In this operating
B state, it is not possible to monitor the patient.
E F During the process, pressures and concentra-
tions may occur in the system which are haz-
ardous to the patient.
460
Service
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 388
Safety checks . . . . . . . . . . . . . . . . . . . . . . . . . . 388
Metrological checks . . . . . . . . . . . . . . . . . . . . . 389
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . 390
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 392
Safety information
WARNING WARNING
Risk due to inappropriately reprocessed prod- Risk if maintenance work is not performed
ucts correctly
The product may be contaminated with infec- If the device is connected to the power or gas
tious agents. supply during maintenance, there is a risk of
personal injury and property damage.
Reprocess the product before any service ac-
tivities are carried out or the product is re- Before performing any maintenance work, dis-
turned for repair. Reprocessing must be car- connect all electrical connections from the
ried out in accordance with the reprocessing power supply and all gas connections from
instructions that are supplied with the prod- the gas supply.
uct.
WARNING
WARNING Risk when opening the housing
Risk if service activities are not carried out
There are live electrical components under the
regularly
housing which could lead to an electric shock.
Wear and material fatigue of the components
The housing may only be opened by those
may lead to device failures and malfunctions.
people who have been designated for this
Perform the service activities at the specified task.
intervals.
WARNING
WARNING Risk if the battery is not replaced properly
Risk if service activities are not performed
If the battery is not replaced properly, short-
correctly
circuits and high temperatures may occur
If service activities are not performed correct- which could lead to an explosion or a fire.
ly, personal injury and property damage may
The battery must be replaced by the persons
occur.
designated for this task.
Service activities must be performed by those
people who have been designated for the re-
spective activity.
Term Definition
Service All measures (inspection, maintenance, repair) intended to maintain or re-
store the functional integrity of a product
Inspection Measures intended to determine and assess the actual state of a product
Maintenance Recurrent specified measures intended to maintain the functional integrity
of a product
Repair Measures intended to restore the functional integrity of a product after a de-
vice malfunction
Inspection
Perform the safety checks in the specified in- – Check flow measurement.
tervals. – Check pressure measurement.
– Check battery backup for function.
– Check the concentration delivery of the
metering module.
– Check ECG measurement.
– Check SpO2 measurement.
– Check NIBP measurement.
– Check IBP measurement.
– Check body temperature measurement.
– Check the BIS measurement.
The permissible tolerances for the measurement
parameters can be found in the appropriate
instructions for use.
3 Check the accuracies of the following gas 6 Check the electrical safety in accordance with
measurement functions (PGA and SGA): the IEC 62353 standard.
– Check O2 measurement. 7 Check the safety features:
– Function of the acoustic alarm device
Ambient air Accuracy ±3 Vol%
21 Vol%: 8 Check the internal O2 hoses for leaks:
100 Vol%: Accuracy –5 Vol% – Leakage <1000 mL/min
SGA:
Desflurane Accuracy ±0.5 Vol%
2 Vol%
N2O 60 Vol% Accuracy ±6.8 Vol%
Maintenance
Repair
Disposal
Disposal of accessories
Technical data
Gas delivery, FG ctrl. control mode . . . . . . . 406 Invasive blood pressure (IBP) . . . . . . . . . . . 428
Anesthetic gas metering modules . . . . . . . . 416 Anesthesia depth monitoring BIS . . . . . . . . 438
General information
Identification
The serial number and the article number of Zeus IE and the [CE logo] mark are located on the rating plate
(type plate) attached to the rear of the device.
Classification
Classification (Cont.)
Dimensions
Dimensions of the device; W x H x D approx. 700 x 1650 x 1100 mm
Dimensions of the breathing system; W x H x D approx. 270 x 400 x 350 mm
Dimensions of a DIVA metering module; W x H x D approx. 80 x 280 x 180 mm
Latex-free
Screen
Variant 1 - C500
Type Resistive touch screen
Size (H x W x D) 284 mm x 414 mm x 95 mm
Weight (without mounting): 7 kg
Resolution 1440 x 900
Contrast (minimum) 500:1
Viewing angle Horizontal: 130°, vertical: 100°
Variant 2 - C500
Type Resistive touch screen
Size (H x W x D) 321 mm x 416 mm x 100 mm
Weight (without mounting): 7 kg
Resolution 1440 x 900
Contrast (minimum) 500:1
Viewing angle Horizontal: 140°, Vertical: 130°
Variant 3 - C700
Type Resistive touch screen
Size (H x W x D) 376 mm x 490 mm x 136 mm
Weight (without mounting): 9.5 kg
Resolution 1680 x 1050
Contrast (minimum) 800:1
Viewing angle Horizontal: 160°, Vertical: 160°
Power supply
Ambient conditions
Air pressure
Operation 700 to 1060 hPa
Storage / transportation 500 to 1060 hPa
CO2 content of the ambient air
Operation 300 to 800 ppm
When transporting filled DIVA metering modules,
the specifications of the anesthetic agent manu-
facturer must be observed.
Only store and transport DIVA metering modules if
they are empty! (page 89).
O2 concentration in fresh gas (carrier gas N2O), 25 to 100 Vol% (measured under ATPD conditions)
FG ctrl. control mode
Accuracy depends on the carrier gas and the set
nominal O2 concentration:
Nominal O2 concen- Carrier gas Carrier gas
tration N2O Air
Below 25 Vol% Not applicable ±0.8 Vol%
25 Vol% ±3.75 Vol% ±0.8 Vol%
30 Vol% ±4.2 Vol% ±1.6 Vol%
40 Vol% ±4.8 Vol% ±2.9 Vol%
50 Vol% ±5.0 Vol% ±3.7 Vol%
60 Vol% ±4.8 Vol% ±4.0 Vol%
70 Vol% ±4.2 Vol% ±3.7 Vol%
80 Vol% ±3.2 Vol% ±3.0 Vol%
90 Vol% ±1.8 Vol% ±1.8 Vol%
100 Vol% ±1.5 Vol% ±1.5 Vol%
Resolution 1 Vol%
Inspiratory O2 concentration in the fresh gas, 25 to 100 Vol% (maximum achievable O2 concen-
Auto control control mode tration due to 100 % relative air humidity in the
breathing system = 93 Vol%. Corrected to BTPS
conditions)
Accuracy: ±(2.5 Vol% + 2.5 % of the measured val-
ue) in accordance with ISO 80601-2-55
Resolution:
25 % to 40 % = 1 Vol%
40 % to 100 % = 5 Vol%
ORC function O2 delivery of 250 mL/min
O2 flush >35 L/min
O2 safety flow 0 to 12 L/min
Accuracy: ±35 % or ±2 L/min
Maximum insp. concentration of volatile anesthetic (inspiratory limit with automatic control)
agents
Isoflurane 0 to 5 Vol%SL
Accuracy: ±20 % or 0.2 Vol%SL
Resolution: 0.1 Vol%SL
Sevoflurane 0 to 8 Vol%SL
Accuracy: ±20 % or 0.2 Vol%SL
Resolution: 0.1 Vol%SL
Desflurane 0 to 18 Vol%SL
Accuracy: ±20 % or 0.2 Vol%SL
Resolution: 0.2 Vol%SL
Ventilation
Ventilation (Cont.)
Ventilation (Cont.)
Breathing system
Airway resistances
Flow (L/min)
289
Measuring systems
CO2 measurement
Concentration in [Vol%]: Measuring range: 0 to 10 Vol%
Resolution: 0.1 Vol%
Accuracy: ±(0.43 Vol% + 8 % of the measured val-
ue) in accordance with ISO 80601-2-55
Partial pressure in [mmHg]: Measuring range: 0 to 78 mmHg
Resolution: 1 mmHg
Accuracy: ±(3 mmHg + 8 % of the measured value)
in accordance with ISO 80601-2-55
Partial pressure in [kPa]: Measuring range: 0 to 10 kPa
Resolution: 0.1 kPa
Accuracy: ±(0.4 kPa + 8 % of the measured value)
in accordance with ISO 80601-2-55
Data communication
Relevant standards
In addition to the standards listed here, this medi-
cal device meets various other standards, e.g.,
standards concerning special national require-
ments.
Functional:
IEC 60601-1 3rd ed. Part 1:
Medical electrical equipment General requirements for basic safety and essen-
tial performance
IEC 60601-1-2 Part 1-2:
Medical electrical equipment Electromagnetic compatibility - Requirements and
tests
IEC 60601-1-8 Part 1-8:
Alarm systems - General requirements, tests and
guidance for alarm systems in medical electrical
equipment and medical systems
ISO 80601-2-13 Part 2-13:
Medical electrical equipment Particular requirements for basic safety and essen-
tial performance of anesthesia workstations
ISO 80601-2-55 Part 2-55:
Particular requirements for basic safety and essen-
tial performance of respiratory gas monitors
IEC 80601-2-30 Part 2-30:
Medical electrical equipment Particular requirements for basic safety and essen-
tial performance of automated non-invasive sphyg-
momanometers
Observe the conditions of use for auxiliary devices. The application of the whole system may be restricted
as a result.
Observe the instructions for use for auxiliary devices and the technical data specified therein.
The anesthetic agent used can restrict the application range of the device with respect to the temperature
range as well as the maximum fresh-gas flow.
Open-source software
Hemodynamic modules
BISx pod
Monitoring specifications
ECG
Display Up to 3 leads
Leads for adults and pediatric I, II, III, aVR, aVL, aVF, V, V+, (aVR/aVL/aVF/V
only with five-lead wire set.
V+ only with six-lead wire set)
Leads for neonates I, II, III, aVR, aVL, aVF, V, V+, (aVR/aVL/aVF/V
only with five-lead wire set.
V+ only with six-lead wire set)
Measuring range 15 to 300 beats/min
Accuracy ±2 beats/min or ±1 %
QRS detection Amplitude: 0.2 to 5.0 mV (for channel amplitudes of
x 4 and x 8)
Amplitude: 0.4 to 5.0 mV (for all other channel am-
plitudes)
Duration: 70 to 120 ms (adults and children); 40 to
120 ms (neonates)
Frequency ranges Filter = monitor: 0.5 to 40 Hz
Filter = ESU: 0.5 to 16 Hz
Filter = OFF: 0.05 to 40 Hz
Electrical specifications Type CF in accordance with IEC 60601-1
Defibrillation protection in accordance with
IEC 60601-2-27
Arrhythmia detection Only for adults and children; not for neonates
Pacemaker detection For adults and children on leads I, II and III, not for
neonates
Pacemakers with the following features can be de- Amplitude ±2 to ±700 mV
tected
Width (dp) 0.2 to 2.0 ms
Rise/fall (min) 0.1 dp, 100 ms
Overshoot (min) 0.025 ap, 2 mV
Recharge constant 4 to 100 ms
ECG (Cont.)
ECG/Arrhythmia/ST supplementary
information
Breathing measurement current Square wave signal, 50 A, 39.896 KHz
Noise suppression Does not apply
Time until alarm for tachycardia Ventricular tachycardia 1 mVpp, 206 beats/min
Gain 0.5, range 3.0 to 3.5 seconds, average
3.3 seconds
Gain 1.0, range 2.9 to 3.3 seconds, average
3.2 seconds
Gain 2.0, range 2.8 to 3.5 seconds, average
3.0 seconds
ECG (Cont.)
Accuracy of the response time of the heart rate Ventricular bigeminy 80 beats/min
measuring equipment to an irregular heartbeat
Slowly changing ventricular bigeminy: 60 beats/min
Rapidly changing ventricular bigeminy:
120 beats/min
Bidirectional systoles: 90 beats/min
Accuracy of the reproduction of the input signal Methods A, B, C, and D according to AAMI EC11
have been used to set up the overall system fault
and frequency response.
ST segment analysis
Leads for the analysis 3-lead wire set: I, II, III (one lead selectable)
5-wire lead set: I, II, III, aVR, aVL, aVF, or V (three
leads selectable without ST option, all leads with
ST option)
6-wire lead set: I, II, III, aVR, aVL, aVF, V, or V+
(three leads selectable without ST option, all leads
with ST option)
Isoelectric point Set range: Start of the QRS complex up to refer-
ence point
Basic setting: Start of the QRS –28 ms
ST measuring point Reference point up to end of the QRS complex
Basic setting: End of the QRS +80 ms
ST complex Length: 892 ms (225 units)
Frequency reaction: 0.05 to 40 Hz
Update interval 15 seconds, with at least one normal beat
Trend time (ST display) 30 minutes
Trend resolution One data point every 30 seconds
Duration of the ST alarm OFF, 15, 30, 45, 60 seconds
Classification Caution
Alarm On/Off Off: No alarms generated
Automatic alarm limits Current measured value ±2 mm
ECG (Cont.)
Wedge measurement
Interfering substances:
Carboxyhemoglobin may erroneously increase the readings. The level of increase corresponds
roughly to the quantity of carboxyhemoglobin present. Dyes, nail varnish, and other substances can
absorb an abnormal amount of red light, thereby compromising the accuracy of the measurement.
Make sure that the sensor is applied at a site where there are no artificial pigments.
SpO2 alarms
Upper limit: adjustable, 71 to 100 %
Lower limit: adjustable, 70 to 99 %
Nominal wavelength:
Dräger: Red: 660 nm IR: 905 nm
Masimo: Red: 660 nm IR: 905 nm
Nellcor: Red: 660 nm IR: 910 nm
Note:
If a further radiation source (e.g., for photodynamic therapy) is used on the same patient in addition to the
SpO2 measurement, there may be interactions, depending on the wavelengths used. In this case, the
wavelength for the SpO2 measurement may have a special clinical significance.
Power:
Dräger: Red: 1.8 mW (max.) IR: 2.0 mW (max.)
Masimo: Red: 0.9 mW (max.) IR: 0.9 mW (max.)
Nellcor: Red: 3 mW (max.) IR: 4 mW (max.)
Note: LED drive is voltage-limited by a hardware mechanism.
Level of protection from electric shock: Type CF in accordance with IEC 60601-1
Defibrillation protection: In accordance with IEC 60601-1A2
Note:
1) Since pulse oximeter measurements are statistically distributed, only two thirds of these measurements
are expected to fall within ±1 standard deviation of the value measured by a CO oximeter.
2) The pulse rate accuracy has been tested by means of blood samples from healthy adult test subjects
in studies with simulated hypoxia in the range of 70-100 % SpO2 in comparison with a laboratory CO ox-
imeter and an ECG monitor. This variation corresponds to ±1 standard deviation of the pulse rate mea-
sured with the ECG monitor.
3) Functional test equipment cannot be used to evaluate the accuracy of a pulse oximeter probe or of a
pulse oximeter monitor.
Masimo:
LNOP-DCI, LNOP-DCIP, NR125 ±2
EAR ±3.5
Pulse rate: ±3 bpm or ±3 %
(whichever is greater)
Measurement accuracy (neonatal):
Saturation (% SpO2):
0 to 69 % Not specified
70 to 100 % Sensor-specific as fol-
lows:
Nellcor:
N-25, OxiMAX MAX-N ±3
Masimo:
LNOPNEO, LNOPNEO SS, LNOP-YI ±3
Description of value
Pulse rate measurement range: 25 to 239 beats/min
Measuring accuracy:
Valid for all Masimo LNCS and LNOP sensors.
Saturation (% SpO2) under motionless conditions:
0 to 69 % Not specified
70 to 100 %
Adult, pediatric ±2
Neonates ±3
Saturation (% SpO2) if patient movement occurs:
0 to 69 % Not specified
70 to 100 %
Adult, pediatric ±3
Neonates ±3
Pulse rate (beats/min) – if no patient movement
occurs:
Adults, pediatric, neonates ±3
Temperature
Disposable sensors
At 45 to 50 °C (113 to 122 °F) ±0.2 °C (±0.4 °F)
At 25 to 45 °C (77 to 113 °F) ±0.1 °C (±0.2 °F)
At –5 to 25 °C (23 to 77 °F) ±0.2 °C (±0.4 °F)
WARNING
The following parameters are not monitored
for neonates: Arrhythmia, cardiac output and
ST segment analysis. No alarm will be activat-
ed.
EMC Declaration
This device has been tested for electromagnetic This device may only be operated in the
compatibility using accessories from the list of environments listed under "Application" in the
accessories. Other accessories may only be used instructions for use.
if they do not impair the electromagnetic
compatibility. The use of non-compliant Transmissions Compliance
accessories may result in increased Radiated emissions Class A, Group 1
electromagnetic emissions or the decreased (30 MHz to 1 GHz)
electromagnetic immunity of the device.
Conducted emis- Class A, Group 1
This device may only be used in immediate sions (150 kHz to 30 MHz)
proximity to other devices if Dräger has authorized
this arrangement of the equipment. Without NOTE
authorization from Dräger, whether the device
The emission-specific properties of this device
functions without interference in the desired
permit its use in the industrial sector and in hospi-
arrangement must be checked before use. The
tals (CISPR 11, Class A). If using this device in a
instructions for use for the other devices must be
domestic establishment (for which Class B is gen-
observed.
erally required as per CISPR 11), it may not pro-
vide sufficient protection for wireless services.
The user may need to take remedial action, such
as resetting or repositioning the device.
Device combinations
Connections to IT networks
During operation, this device can exchange This device must only be connected to the network
information with other devices by means of IT by service personnel. The IT administrator of the
networks and supports the following functions: hospital must be consulted in advance.
– Display of waveforms and parameter data The following documents must be followed:
– This device exchanges data with other medical – Data are sent incomplete, sent to the wrong
devices over the LAN network. The network device, or not sent at all.
must support the following transmissions and
– Patient data are intercepted, falsified, or
protocols:
damaged.
– TCP/IP
– Unicast (static or dynamic addressing with – Data have incorrect time stamps.
the ARP or RARP network protocols)
– An overload of the device due to very high
– Multicast
network loads (e.g., caused by denial-of-
– Broadcast
service attacks) can lead to deactivation of the
– IGMP (Version 2)
interface. The interface will only be available
This device can join or leave an IP multicast again after the device has been restarted. In
group by using the IGMP network protocol. rare cases, a warm boot may take place and
may occur repeatedly.
Serial ports
Requirements for the electrical characteristics
This device supports the following ports:
of connected devices and networks
– RS-232 ports conforming to EIA RS-232
The LAN ports and the serial ports are only suitable
(CCITT V.24/V.28) for the following
for connection of devices or networks that have a
applications:
nominal voltage of 24 V DC on the network side at
– MEDIBUS, MEDIBUS.X
maximum, and that meet the requirements of one of
– Paging systems
the following standards:
– Connections to medical devices from other
manufacturers – IEC 60950-1: Ungrounded SELV circuits
– Ports based on USB 1.1 or USB 2.0 (e.g., – IEC 60601-1 (as of 2nd edition): Touchable
mouse, keyboard, hard disk, USB stick, CD secondary circuits
drive)
Description
Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 454
Flow measurement and volume delivery . . . . . 454
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 459
Units and standards . . . . . . . . . . . . . . . . . . . . . 459
V
W
E
P
A Man
Spont
B X C D F
Fresh
gas/Anesthetic gas
O2 flush/
O
emergency O2 G
P N
Q J K
P H
I
R U
S M P L
M
T
system (AGSS)
N Pressure gauge
B APL valve
O Additional air valve
C PEEP/PMAX valve
P TurboVent 2 blower module
D Expiratory valve
Q Pressure sensor – breathing bag
E Pressure sensor – airway pressure
R Breathing bag
F SpiroLife expiration flow sensor
S Microbial filter sample gas return line
G Patient's lung
T Gas analysis
H SpiroLife inspiration flow sensor
U CO2 water trap
I Inspiratory valve
V External area for the leak test: for manual
J CO2 absorber ventilation
K Optional CO2 Clic absorber, unlocked position W Closed circulatory system valve
L Safety valve in breathing system (excess X Man/Spont changeover valve
pressure)
200 mL/min
Filter (F1b)
A B R1
Filter (F1a) Valve (V2)
Pump
(Pu)
170 mL/min ILCA3
dual R2 R3 R4
sensor PIA
sensor C1 C2 C3
Gas
P P
E E outlet
Valve
(V1)
30 mL/min R5
Filter (F2)
200 mL/min
sample gas
from Water trap
Pneumatic diagram of the gas measurement system
patient
PCB Flower Power P
P
E
Pambient
VCC RS232
12 V
A Diaphragm for
sample gas flow
B Diaphragm for
bypass flow
During anesthesia, the user must ensure an Volume control loop (Auto control
adequate oxygen supply and sufficient depth of control mode, MinFG flow = "Uptake")
anesthesia for the patient. The former is ensured by
setting an inspiratory O2 concentration. The latter,
the assurance of sufficient depth of anesthesia, is Closed system
achieved by setting the expiratory anesthetic agent
concentration. Zeus IE can be run in normal
operation (Auto control mode), in a closed circuit
(Uptake mode) or in a semi-closed circuit (FGmin Pressure
0.5 L/min). In both operating modes, the gas sensor Gas mixing
concentration is controlled in the breathing system system
(inspiratory O2 control loop) and in the gas
concentration exhaled by the patient (expiratory P
Deficit Excess
Concentration control loops (Auto
251
control control mode) The level in the breathing bag is determined by a
pressure sensor. The volume control loop keeps
Set points (target values for the control loops) are
the level in the breathing bag constant. The uptake
the inspiratory oxygen concentration and the
of the patient as well as possible leakages can be
expiratory anesthetic agent concentration.
compensated in conjunction with the concentration
A scavenging concentration measurement at the Y- control loop for oxygen. Thus a patient uptake
piece (patient gas measurement) and a second measurement for oxygen is available. If there is too
scavenging concentration measurement in the much gas in the breathing system, the gas is
inspiratory limb (supervisor gas measurement) drained into the anesthetic gas scavenging system
provide measured values for O2, N2O, CO2, and via the excess gas outlet.
anesthetic gas. In MAN/SPON mode, the volume control loop is not
Patient gas analysis is subjected to further active.
processing for gas monitoring. Secondly, this
sensor supplies the actual inspiratory and
expiratory values for the concentration control
loops.
In MAN/SPON mode, the gas control loop is not
active.
Inspiratory oxygen concentration If the set O2 value is lowered, the fresh-gas flow is
control not increased beyond the set minimum flow. In
Uptake mode, the O2 concentration drops solely
due to the O2 uptake flow.
Patient gas analysis To set a lower O2 concentration, carrier gas is
O2, anesthetic gas, CO2
added; the oxygen concentration cannot fall below
FiO2 = 30 %
21 Vol%SL in the fresh gas, however. The control
Exp. valve behavior of the carrier gas N2O is therefore
FG mixer identical to that of the carrier gas Air.
Circle flow
Oxygen metering by means of emergency O2
Patient delivery, O2 flush, or additional O2 delivery leads to
Target value an increase in the O2 concentration.
FiO2 = 30 % TurboVent 2 blower module
Due to the release of nitrogen (N2) into the system
252
End tidal CO2 value measured in Insp. O2 set value Maximum deviation
terms of released breaths at the Y-
35 Vol%SL –3 Vol%SL
piece
>35 Vol%SL 60 Vol%SL –5 Vol%SL
Minute volume
>60 Vol%SL –7 Vol%SL
Control val- Control of mixed gas metering with the
ue controlled variables The oxygen control loop takes a humidity of
Volume of fresh-gas flow [L/min] 7 Vol%SL in the breathing system into account. The
maximum achievable inspiratory oxygen
O2 concentration of the fresh-gas flow concentration in control mode is therefore only
[Vol%SL] 93 Vol%SL, additionally reduced by the anesthetic
agent concentration in the circuit system.
The controller determines the volume and
concentration of the fresh-gas flow to be metered.
The latter cannot fall below the fresh-gas flow set
by the user (MinFG flow).
To achieve the new set value quickly (T90 typically
<60 s), the fresh-gas flow is automatically
increased with an increase in the set O2 value. A
higher O2 concentration is achieved by metering
pure oxygen (taking into account the set anesthetic
gas concentration).
breathing system.
If the patient has already been subjected to
Measured value larger than set val-
induction, the Prime Dose is prevented from
ue of mixed gas flow [L/min]
becoming active at the same time.
The anesthetic agent is supplied as saturated The second stage is a fixed limit for the Steady
steam to the breathing system. Circle flow and State Dose. The level of the Steady State Dose is
minute volume provide for an even mixing of dependent on the anesthetic agent used.
breathing gas. To achieve the new set value quickly
(T90 AgasINSP typically <120 s), the flow is
automatically increased with the increase in the set
value of expiratory anesthetic agent. The T90 time
of expiratory anesthetic agent concentration
Duration Prime Dose Criteria for starting a flush phase with carrier gas
3 min for isoflurane, sevoflurane and des- Air
flurane Start inN2O or etN2O >6 Vol%SL
Level of the Steady State Dose End after 60 s
1.0 Vol%SL for isoflurane, sevoflurane and
inN2O and etN2O <1 Vol%SL
2.0 Vol%SL for desflurane
Criteria for starting a flush phase with carrier gas
N2O
Flush phases
Safety concept (HighFlow operation)
If an anesthetic agent mixture consisting of several
different anesthetic gases is detected by the patient There is a second gas analyzer (SGA) in the
gas analyzer, normal operation is stopped. If the inspiration to a pathway in addition to the gas
concentration of the anesthetic gas metered first is sensor in the Y-piece, in order to safeguard the
so high that it can still be identified, the device control loop function. This gas analyzer checks the
switches to HighFlow operation (flow >12 L/min). control loop function by monitoring the inspiratory
Regular flush phases during normal operation must limit values. The gas analyzers are also checked
prevent the uptake of foreign gases. In contrast to for plausibility by mutual comparison of the
HighFlow operation, the control loops (O2 and measured values.
anesthetic agent) remain active during the flush The safety concept increases the fresh-gas flow
phase. Only the rate of mixed gas flow is increased. (High Flow operation) in the case of error. The
The flush flow rate is 12 L/min. adjusted concentration values then apply to the
increased fresh-gas flow. The control loops are
thus switched off.
O2 emergency delivery
O2 flush
Ventilation
Flow measurement and volume delivery ventilation. Depending on airway pressure, Zeus IE
increases the tidal volume by the amount remaining
Depending on the compliance of the lung to the in the breathing hoses.
compliance of the breathing circuit, the volume The expiratory minute volume and tidal volume are
supplied by the ventilation device is distributed measured with the aid of two internal flow sensors.
between the lung of the patient and the breathing Measurement is adjusted for hose compliance. The
circuit inserted between the device and the patient. volume of the measured gas and the volume of the
Deviations resulting for the delivered volume are delivered gas depend, however, on the ambient
low for adult patients. This is due to their relatively conditions of temperature and pressure in the
high lung compliance in relation to the much lower breathing system. In lung physiology, the minute
compliance of the breathing hoses. Only the volume and tidal volume are related to the
volume attained and surrendered by the lung is conditions in the lung: 37 °C body temperature,
relevant to the efficiency of ventilation. Larger lung pressure, 100 % rel. humidity (BTPS). The
differences are possible during pediatric ventilation. measured values for MV and VT calculated by
Zeus IE basically compensates for the influence of Zeus IE, as well as the set value for VT are always
hose compliance on volume-controlled ventilation. in relation to BTPS conditions. Leakages which
Zeus IE determines the compliance of the occur at the patient end are not taken into
breathing hoses during the device check before consideration by Zeus IE during volume
measurement, nor during volume delivery.
Ventilation modes
Pressure Control (Pressure-controlled The change-over from the expiratory pressure level
ventilation) to the inspiratory pressure level and also the
change from the inspiratory pressure level to the
This ventilation mode is known as pressure- expiratory pressure level are synchronized with the
controlled or time-controlled ventilation, during patient's spontaneous breathing. Good adaptation
which the patient can always breathe to the spontaneous breathing behavior of the
spontaneously. Pressure Control is thus often patient is thus achieved.
described as a time-controlled switch between two The respiratory rate of changes is also kept
CPAP levels. constant during synchronization via the "trigger
Controlled ventilation is achieved via the time- window" with a fixed position in time.
controlled pressure change which corresponds to
the pressure-controlled ventilation (Pressure
Control). The constant possibility of spontaneous
breathing allows, however, a fluent transition of
controlled ventilation up to complete spontaneous
breathing without changing the ventilation mode.
The trigger window is 25 % of the set inspiratory inspiration. The transition from controlled
time at the upper pressure level (Pinsp), but a ventilation over the weaning phase towards
maximum of 5 seconds or 25 % of Te (Te = complete spontaneous breathing occurs via a
expiratory time). However, at the PEEP level, the gradual reduction of the inspiratory pressure Pinsp
maximum is 5 seconds. The trigger window can be and/or the respiratory rate RR.
switched off.
The good adaptation to the spontaneous breathing
of the patient results in a low sedation requirement Volume Control AutoFlow
and thus a quicker return of the patient to
spontaneous breathing. Recent clinical experience The Vol. Ctrl. AutoFlow volume mode ensures the
validates this. application of an set tidal volume by gradual
adaptation of the mechanical inspiratory pressure
As with all pressure-controlled ventilation modes, a to lung conditions.
non-fixed tidal volume VT is specified for the
patient. The tidal volume is primarily the result of Spontaneous breathing is possible at any time. The
the pressure difference between the set values for pressure patterns and flow patterns of the
PEEP and Pinsp. mechanical inspiratory breaths correspond to those
of pressure-controlled ventilation.
Modifications of lung compliance and the airways
as well as active counter-breathing of the patient The calculated inspiratory pressure Pinsp adapts to
can modify the tidal volume. This is a desired effect the current lung parameters (resistance R,
in this ventilation mode. compliance Cdyn) and the spontaneous breathing
requirements of the patient. Pinsp is always the
Knowing the non-constant tidal volume and thus minimum ventilation pressure. This ventilation
also the minute volume, an adequate setting of the pressure is required to fill the compliance of the
alarm limits for minute volume must be observed. lung with the required tidal volume VT.
To set the necessary difference between the two With modifications of lung compliance, the plateau
pressure levels, the display of the measured pressure changes accordingly, as is always the
expiratory tidal volume VTe is used. An increase in case with volume-controlled ventilation forms. With
the difference causes a higher pressure-controlled AutoFlow, the ventilation pressure modifications
ventilation breath. are performed in up to 3 mbar steps from ventilation
As with synchronization with Vol. Ctrl. AutoFlow, breath to ventilation breath.
the time pattern is adjusted via the basic setting If the selected mechanical inspiratory time Ti is
parameters for respiratory rate RR and inspiratory greater than or equal to the lung filling time, Pinsp is
time Ti. The lower pressure level is adjusted with exactly equal to the plateau pressure. If the
the parameter PEEP, the upper pressure level with inspiration time is shorter than the fill time of the
the parameter Pinsp. lung, the inspiration flow at the end of the
During synchronization of pressure-controlled inspiration time does not drop to zero. This can be
ventilation with Vol. Ctrl. AutoFlow, only the Pinsp determined from the flow curve. In this case as well,
setting needs to be modified while the time pattern AutoFlow sets the minimum Pinsp under these
is maintained. conditions, which is, however, above the plateau
pressure. Here the user must decide whether an
The steepness of the increase from the lower extension of inspiration time is useful in order to
pressure level to the upper pressure level is further decrease the ventilation pressure.
controlled by the Slope setting. The effective time
for the pressure rise cannot go above the set If the set inspiratory time Ti is longer than the fill
inspiratory time Ti. This ensures that the upper time of the lung, then the patient can breathe in and
pressure level Pinsp is achieved safely during out spontaneously within the remaining time at the
Pressure Support ventilation mode for the patient's breathing requirement, up to the
preselected Psupp. pressure. The time for the
For pressure support of insufficient spontaneous pressure increase is can be set to 0 to 2 seconds.
breathing In the event of a rapid increase in pressure, Zeus IE
Paw
supports the insufficient spontaneous breathing of
the patient with high peak flow.
Phase = Specification of In the case of a slow increase in pressure, Zeus IE
ideal airway starts gently with an even inspiratory flow. The
Psupp pressure patient must do more breathing work, and the
breathing musculature is trained.
With an adapted pressure rise and Psupp pressure,
CPAP
the patient determines the required inspiration flow
with breathing activity.
t
Start of inspiration End of inspiration If the inspiratory flow decreases to 25 % of the
previously supplied maximum value, and after
4 seconds at the latest (1.5 seconds for pediatric
ventilation), the pressure support Pressure Support
Flow 4 s max.
is terminated.
spontaneous triggering.
In a similar way to how the anesthetist senses the
Upon cancellation of the Pressure Support cycle,
restarting of spontaneous breathing in the patient at
the device triggers the Apnea ventilation alarm.
the breathing bag and supports it manually, the
Check whether a lower flow trigger is set or there is
device can support insufficient spontaneous
a switchover to mandatory breathing.
breathing.
The device takes over the inspiration work partially,
but the patient, however, retains control of
spontaneous breathing.
The CPAP system supplies the spontaneously
breathing patient with breathing gas, even with low
inspiration efforts.
When the spontaneous inspiration flow reaches the
set value of the flow trigger, the pressure support
Pressure Support is started. The device then
generates a pressure rise, which can be adjusted
Synchronization with Vol. Ctrl. AutoFlow Since synchronization of the mandatory breath
(Sync. = ON) shortens the effective Sync. time and thus would
inadvertently increase the effective respiratory rate,
Synchronized intermittent mandatory ventilation Zeus IE extends the following spontaneous
breathing time by the missing time difference T.
Pressure support Exceeding the Sync. respiratory rate is thus
Paw
∆Psupp avoided. The factor RR responsible for minimum
fast rise time slow rise time ventilation in addition to tidal volume VT remains
Pinsp fast rise time slow rise time constant. If the patient has inspired a substantial
volume at the beginning of the trigger window, the
PEEP
t
device reduces the following mandatory ventilation
Ti
Trigger window for insp. and exp. breath. The time for the inspiratory flow phase and
synchronization
1
inspiration time is shortened. Thus the tidal volume
Flow
RR VT remains constant and overfilling of the lungs is
avoided.
without spontaneous with spontaneous In the course of further weaning, the respiratory
breathing breathing
rate RR is further reduced at the device and thus
258
Appendix
CO2 concentration
The concentration setting for CO2 on Zeus IE can
be either BTPS or normal (without correction). The
appropriate correction then applies both to
concentration setting and to concentration
measurement.
2
Sevoflurane
Enflurane
1
Isoflurane
0
Halothane
0 10 20 30 40 50 60 70 80
Age (years)
253
Example:
etSEV = 1.5 Vol%SL; et N2O = 60 %; age = 10 years
MACage-corrected of SEV: MAC10 = 2.5 Vol%SL
MACage-corrected of N2O: MAC10 = 126 Vol%SL
xMAC = 0.6 + 0.5 = 1.1
Mixture detection
Zeus IE automatically detects the anesthetic gas
used and switches the anesthetic gas
concentration measurement and monitoring to the
gas detected.
If a mixture of two volatile anesthetic agents is
detected, the concentration of the secondary agent
is displayed as of 0.4 hPa or 0.4 Vol%SL.
Index
S T
Safe-Fil filling system for desflurane . . . . . . . . 104 Technical data . . . . . . . . . . . . . . . . . . . . . . . . 395
Safety concept (HighFlow operation) . . . . . . . 452 Airway resistances . . . . . . . . . . . . . . . . . . 411
Saving profiles/views . . . . . . . . . . . . . . . . . . . . 349 Ambient conditions . . . . . . . . . . . . . . . . . . 401
Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 Anesthetic gas measurement . . . . . . . . . . 415
Screen layout . . . . . . . . . . . . . . . . . . . . . . . . . 353 Anesthetic gas metering modules . . . . . . 416
Select profile . . . . . . . . . . . . . . . . . . . . . . . . . . 127 Auto Control control mode . . . . . . . . . . . . 406
Selecting a drug . . . . . . . . . . . . . . . . . . . . . . . 171 BIS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 438
Serial ports . . . . . . . . . . . . . . . . . . . . . . . . . . . 417 BISx pod . . . . . . . . . . . . . . . . . . . . . . . . . . 422
Service Call . . . . . . . . . . . . . . . . . . . . . . . . . . . 383 Breathing system . . . . . . . . . . . . . . . . . . . 410
Service functions . . . . . . . . . . . . . . . . . . . . . . . 383 C.O. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 429
Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . 383 Central gas supply . . . . . . . . . . . . . . . . . . 403
Set and select loops . . . . . . . . . . . . . . . . . . . . 224 Classification . . . . . . . . . . . . . . . . . . . . . . 398
Set APL valve . . . . . . . . . . . . . . . . . . . . . . . . . 142 CO2 measurement . . . . . . . . . . . . . . . . . . 415
Set details . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 Data communication . . . . . . . . . . . . . . . . . 417
Set drug dose . . . . . . . . . . . . . . . . . . . . . . . . . 173 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . 399
Setting ranges of the monitoring alarms . . . . . 207 ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 423
Sevoflurane filling system . . . . . . . . . . . . . . . . 101 Essential performance data . . . . . . . . . . . 439
Single measurements . . . . . . . . . . . . . . . . . . . 254 FG met. control mode . . . . . . . . . . . . . . . . 406
Single Twitch . . . . . . . . . . . . . . . . . . . . . . . . . . 272 Flow measurement . . . . . . . . . . . . . . . . . . 412
SpO2 measurement connection . . . . . . . . . . . 235 Fresh-gas outlet for
SpO2 safety instructions . . . . . . . . . . . . . . . . . 248 non-rebreathing systems . . . . . . . . . . . . . 410
SpO2 settings on Zeus IE . . . . . . . . . . . . . . . . 249 Hemodynamic modules . . . . . . . . . . . . . . 420
ST measuring points . . . . . . . . . . . . . . . . . . . . 246 IBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 428
ST screen display . . . . . . . . . . . . . . . . . . . . . . 246 Identification . . . . . . . . . . . . . . . . . . . . . . . 398
ST segment analysis . . . . . . . . . . . . . . . . . . . . 245 Latex-free . . . . . . . . . . . . . . . . . . . . . . . . . 399
Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128 Measuring systems . . . . . . . . . . . . . . . . . . 412
Start dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 NMT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 437
Start NIBP measurement . . . . . . . . . . . . . . . . . 253 NMT Pod . . . . . . . . . . . . . . . . . . . . . . . . . 421
Start settings . . . . . . . . . . . . . . . . . . . . . . . . . . 126 O2 measurement . . . . . . . . . . . . . . . . . . . 414
Starting a bolus . . . . . . . . . . . . . . . . . . . . . . . . 169 O2, CO2, and anesthetic gas . . . . . . . . . . 414
Starting in emergencies . . . . . . . . . . . . . . . . . . 123 Power supply . . . . . . . . . . . . . . . . . . . . . . 401
Starting NMT measurement . . . . . . . . . . . . . . 272 Pressure measurement . . . . . . . . . . . . . . 412
Starting the infusion . . . . . . . . . . . . . . . . . . . . . 167 Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . 400
Starting the self test manually . . . . . . . . . . . . . 114 SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 429
Stopping a bolus . . . . . . . . . . . . . . . . . . . . . . . 170 ST segment analysis . . . . . . . . . . . . . . . . 425
Stopping the infusion . . . . . . . . . . . . . . . . . . . . 169 Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . 407
Stopwatch dialog window . . . . . . . . . . . . . . . . 381 Wedge measurement . . . . . . . . . . . . . . . . 428
Suppress alarm tone . . . . . . . . . . . . . . . . . . . . 205 Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . 398
Suspending alarms . . . . . . . . . . . . . . . . . . . . . 189 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 263
Switch audible alarm on again . . . . . . . . . . . . 205 Temperature measurement connection . . . . . 235
Syringe change . . . . . . . . . . . . . . . . . . . . . . . . 169 Temperature screen display . . . . . . . . . . . . . . 264
System info . . . . . . . . . . . . . . . . . . . . . . . . . . . 122 Temperature settings on Zeus IE . . . . . . . . . . 264
System setup . . . . . . . . . . . . . . . . . . . . . . . . . . 347 Test history . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Test info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Therapy bar . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Therapy controls . . . . . . . . . . . . . . . . . . . . . . . . 54
TOF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 273
Total consumption . . . . . . . . . . . . . . . . . . . . . 378
Touch screen . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Transportation of
filled DIVA metering modules . . . . . . . . . . . . . . 89
U
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 363
Unlocking profiles/views . . . . . . . . . . . . . . . . . 352
USB devices . . . . . . . . . . . . . . . . . . . . . . . . . . 365
USB ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . 417
Use port saver . . . . . . . . . . . . . . . . . . . . . . . . . . 86
V
VA/N2O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 376
Venostasis . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255
Ventilation . . . . . . . . . . . . . . . . . . . . 139, 217, 454
Ventilation modes . . . . . . . . . . . . . . . 22, 376, 454
Ventilation settings . . . . . . . . . . . . . . . . . . . . . . 375
Ventilator failure . . . . . . . . . . . . . . . . . . . . . . . . 284
View settings . . . . . . . . . . . . . . . . . . . . . . . . . . 346
Volume control loop . . . . . . . . . . . . . . . . . . . . . 449
Volumeter function . . . . . . . . . . . . . . . . . . . . . . 141
W
Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Wedge pressure measurement . . . . . . . . . . . . 262
Z
Zeroing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257
Zeus IE messages . . . . . . . . . . . . . . . . . . . . . . 287
Directive 93/42/EEC
concerning medical devices
Manufacturer
9510797 – GA 5133.035 en
© Drägerwerk AG & Co. KGaA
Edition: 1 – 2019-10
Dräger reserves the right to make modifications
to the device without prior notice.