Professional Documents
Culture Documents
Phased Array - Annex B
Phased Array - Annex B
Phased Array - Annex B
Encoder calibration.
PA calibrations - Linear Scan
SECTORIAL SCAN
Sensitivity / Velocity / Wedge Delay Calibrations
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
The velocity calibration will fine tune the velocity to that of the calibration
block. It will require two reflectors of the same size at two different known
depths;
In many cases the default material velocity set by the operator in the group
wizard is sufficient and this calibration can be skipped because wedge delay
will adjust accordingly to allow accurate measurements;
Select A-Scan
Put the probe on your block over the
first reflector;
Choose an angle near the middle of
your angle sweep using the angle
function;
Adjust range as needed to ensure you
will see the 2 reflectors you have
chosen.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
Accept or Restart
View the velocity determination;
If it is sensible accept the calibration,
if not restart.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
Wedge Delay will enter compensation for the travel of the beams through
the wedge medium and account for the various exit points;
Wedge delay needs only 1 distinct reflector - a known depth reflector will
be used here;
Accept or Restart
After hitting calibrate - rerun the probe through the angular range verifying
you are within the set tolerance;
Accept or Restart.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
1.3) Sensitivity
This will insure no matter what angle the reflector is seen at the % FSH is
the same for rejection or detection purposes as well as for the amplitude
based color coded imaging selections.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
1.3) Sensitivity
1.3) Sensitivity
Set Section
1.3) Sensitivity
Set Gate A on Echo A
Set the gate to encompass the SDH reflection. Adjust the
start/width/threshold of the gate;
Move the probe forward and backward to ensure the gate covers the
reflector through all the angles and is not on any other reflectors that may
inhibit the calibration.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
1.3) Sensitivity
Set Compensation Gain
Allows you to adjust the gain as needed to ensure the reflector is seen
properly throughout all the angles;
In most cases no adjustment or an adjustment of the overall gain (if over
reference level FSH%) is needed.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
1.3) Sensitivity
Calibrate and Accept
Applying even pressure and not skewing right or left, move the probe
forward and backwards to build the green line as the reflector is seen by
the angles;
Ensure the green line is on-screen across the whole screen-not to low or
over the tolerance lines.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
1.3) Sensitivity
Calibrate and Accept
Press Calibrate;
Verify the calibration by repeating the forward and backward movement
of the probe;
The green line should fall within the defined tolerance lines across all
the angles;
Accept or Recalibrate as needed.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
1.3) Sensitivity
To Verify Sensitivity Calibration
To verify sensitivity go to Display menu;
Select properties;
Change source to Highest;
Run the probe back and forth and observe FSH in the A% reading field as
the data cursor tracks the highest amplitude signal.
PA calibrations - Linear Scan
LINEAR SCAN
Sensitivity / Velocity / Wedge Delay Calibrations
PA calibrations - Linear Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
The velocity calibration will fine tune the velocity to that of the calibration
block. It will require two reflectors of the same size at two different known
depths;
In many cases the default material velocity set by the operator in the group
wizard is sufficient and this calibration can be skipped because wedge delay
will adjust accordingly to allow accurate measurements;
Select A-Scan
Put the probe on your block over the
first reflector;
Choose a VPA near the middle of your
probe;
Adjust range as needed to ensure you
will see the 2 reflectors you have
chosen.
PA calibrations - Linear Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
Accept or Restart
View the velocity determination;
If it is sensible accept the calibration,
if not restart.
PA calibrations - Linear Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
Wedge Delay will enter compensation for the travel of the beams through
the wedge medium and account for the various exit points;
Wedge delay needs only 1 distinct reflector - a known depth reflector will
be used here;
Move the probe forward and backward applying even pressure and
avoiding skewing right or left;
Move in both directions until the green line is built across the whole
range. Adjust gain as needed;
Hit calibrate.
PA calibrations - Linear Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
Accept or Restart
After hitting calibrate - rerun the probe through the VPAs range verifying
you are within the set tolerance;
Accept or Restart.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
2.3) Sensitivity
This will insure no matter what VPA the reflector is seen at the % FSH is
the same for rejection or detection purposes as well as for the amplitude
based color coded imaging selections.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
2.3) Sensitivity
2.3) Sensitivity
Choose a SDH and couple your probe above it;
Use a guide system to keep the probe from skewing side to side.
Set Section
Choose whether to calibrate the whole VPA range in one step or break it up
into separate calibrations - if your calibration block has adequate spacing
between SDHs you can usually do the whole range.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
2.3) Sensitivity
Set Gate A on Echo A
Set the gate to encompass the SDH reflection. Adjust the
start/width/threshold of the gate;
Move the probe forward and backward to ensure the gate covers the
reflector through all the VPAs and is not on any other reflectors that may
inhibit the calibration.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
2.3) Sensitivity
Set Compensation Gain
Allows you to adjust the gain as needed to ensure the reflector is seen
properly throughout all the VPAs;
In most cases no adjustment or an adjustment of the overall gain (if over
reference level FSH%) is needed.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
2.3) Sensitivity
Calibrate and Accept
Applying even pressure and not skewing right or left, move the probe
forward and backwards to build the green line as the reflector is seen by
the VPAs;
Ensure the green line is on-screen across the whole screen-not to low or
over the tolerance lines.
PA calibrations - Sectorial Scan
Performing Sensitivity / Velocity / Wedge Delay Calibrations
2.3) Sensitivity
Calibrate and Accept
Press Calibrate;
Verify the calibration by repeating the forward and backward movement
of the probe;
The green line should fall within the defined tolerance lines across all
VPAs;
Accept or Recalibrate as needed.
PA calibrations - Linear Scan
SECTORIAL SCAN
TCG Calibration
PA calibrations - Sectorial Scan
Performing TCG Calibration
3.1) TCG
TCG equalizes the sensitivity for a reference reflector through time/depth
to compensate for attenuation during beam travel;
In phased array on the OmniScan we perform TCG calibration across all the
angles;
When allowed TCG is almost always a better choice than DAC in phased
array.
PA calibrations - Sectorial Scan
Performing TCG Calibration
3.1) TCG
Using Wizard for TCG
Enter calibration menu, select TCG-start;
Set reference amplitude and tolerance.
PA calibrations - Sectorial Scan
Performing TCG Calibration
3.1) TCG
Set Section
Set Section - choose to add TCG to the whole angular range or to break the
calibration up into sections- in most cases you can do the whole range in
one step.
PA calibrations - Sectorial Scan
Performing TCG Calibration
3.1) TCG
3.1) TCG
Using Wizard for TCG
Run the probe over first reflector building a green line across the screen;
Keep constant pressure and even coupling and avoid skewing right or left;
Adjust gain as needed to keep line on screen and below red tolerance lines;
Add point.
PA calibrations - Sectorial Scan
Performing TCG Calibration
3.1) TCG
3.1) TCG
Using Wizard for TCG
3.1) TCG
Repeat the aforementioned steps until you have built the TCG through
your inspection range;
When you complete the last point Accept the TCG.
PA calibrations - Sectorial Scan
Performing TCG Calibration
3.1) TCG
Verify TCG
Set A% and DA readings;
Set display-properties-overlay-Source
to Highest to track highest reflector;
Slowly move the probe forward and
backwards over reflectors while
observing these readings to ensure
within tolerances;
For the tracking Highest feature to
work the reflector must fall within the
A-gate.
PA calibrations - Encoder
ENCODER
Calibration
PA calibrations - Encoder
Performing Encoder Calibration
4.1) ENCODER
Almost all the parameters to set up the encoder can be found under two
menus - Scan Menu, and Wizard - Calibration - encoder Menu;
The encoder should be attached to the wedge using the brackets. Align
the wheel in the direction you want to encoder - the scan axis-direction
probe will move;
Make sure there is some tension on the wheel when you press down to
ensure adequate pressure to avoid slipping;
Insert the encoder cable into the encoder slot on top of the OmniScan
being conscious of removed pin and its location.
PA calibrations - Encoder
Performing Encoder Calibration
4.1) ENCODER
4.1) ENCODER
4.1) ENCODER
4.1) ENCODER
Set Overwrite Parameters - Scan/Data Menu
What would you like the OmniScan to record when the encoder goes over an
area it has already covered?
Can be last seen data, highest amplitude in the A gate, or the minimum or
maximum thickness when doing thickness inspections.
PA calibrations - Encoder
Performing Encoder Calibration
4.1) ENCODER
Calibrating the Encoder
After the scan parameters are set the encoder will need to be calibrated if
the encoder resolution is unknown;
Needed items- small rule or distance reference, good surface to calibrate
on.
Wizard - Calibration Menu - choose Type - Encoder -Start
PA calibrations - Encoder
Performing Encoder Calibration
4.1) ENCODER
Setting Origin x axis 0.00
4.1) ENCODER
Set Distance to Travel
Input the distance you plan on moving the encoder.
4.1) ENCODER
Move this Distance Push Calibrate and Verify
When you push calibrate the axis
value will change to your set
parameter;
Carefully move the probe back
to the 0 point and verify x axis
reading is at 0;
Accept or Restart.