Professional Documents
Culture Documents
Weighing Accuracy: Estimating Measurement Bias & Uncertainty of A Weighing
Weighing Accuracy: Estimating Measurement Bias & Uncertainty of A Weighing
© METTLER TOLEDO
Arthur Reichmuth V2.1 Sept. 2001
Abstract
In many applications, the weighing result needs to be
qualified, i.e., the measurement bias and uncertainty
accompanying the weighing process are required. These
values are usually not readily available, not least because
they are dependent on the application at hand. At other
times, the operator needs to know the minimal amount of
mass (a.k.a. ”minimal weight”) that can be weighed
conforming to a required relative uncertainty and confidence
level.
This paper discusses influences on the balance and the
weighing object, and explains how the weighing uncertainty
and minimal sample mass can be estimated. The assumptions
and restrictions, under which this deduction is valid, and
under which conditions influences may be neglected, are
explained. Two examples with actual data from analytical
balances are given as illustrations.
The theory and examples provided here enables the user to
asses the uncertainty and estimate figures of uncertainty of,
or minimal sample mass for, a weighing.
Contents
Introduction ........................................................... 1
Influences Originating From The Balance ................. 1
Influences Affecting The Weighing Object ................ 2
Influences by the Environment ................................. 3
Modelling The Weighing ........................................ 4
Characterization Of The Individual Influences .......... 6
Combined Weighing Bias ....................................... 17
Combined Weighing Standard Uncertainty.............. 18
Expanded Weighing Uncertainty ............................ 19
Example 1............................................................. 21
Uncertainty Charts ................................................. 25
Minimal Sample Mass ............................................ 31
Example 2............................................................. 33
Chart of Minimal Sample Mass ............................... 33
Conclusions ........................................................... 35
Appendices ........................................................... 37
References............................................................. 42
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
—1—
Introduction
The primary aim of a weighing is to determine the mass of
the weighing object.
Measurement processes are generally subject to influences of
all kinds of origins. The weighing is no exception, as it may
get distorted by many influences which introduce bias and
uncertainty in its result. Such influences may affect the
weighing object, the balance, or both. The distortions may be
caused by properties of the weighing object, the balance, the
environment and the procedure, by which the weighing is
performed.
Influences originating from balance are manifold. They
include effects introduced by the balance’s quantized digital
display, the balance’s limited capability to repeat, its non-
linear characteristic, its sensitivity to eccentric loading, its
deviation of sensitivity from the correct value, as well as its
temperature dependence.
Among the influences that afflict the weighing object, the
most prominent is buoyancy, which is caused by the fact that
usually weighings are carried out in air (instead of an empty
space, the vacuum). Other influences are magnetic and
electrostatic forces.
Ambient conditions like air temperature, humidity, pressure
and air velocity will influence the weighing object and the
balance. Influences caused by vibration, inclination, or other
conditions may be present.
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
—2—
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
—3—
1) see [Niels]
2) see [Clark]
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
—4—
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
—5—
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
—6—
5) Even with no load on the weighing pan, there is always weight acting
on the balance, constitued by the mass of the weighing pan, and other
dead loads, such as mechanical parts of the weighing cell. This
defines the ”no load” condition.
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
—7—
6) a.k.a. ”4/5-rounding”
7) This is valid for a ”4/5-rounding”, with a busy measurement value,
i.e., a weighing value that ”bounces” at least some steps around its
average value.
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
—8—
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
—9—
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 10 —
14) With the exception, when weighing the calibration standard for the
purpose of adjusting the balance’s sensitivity.
15) see [R33]
16) This is not to say that buoyancy has ceased to occur. It means, that the
effect on the weighing result is greatly reduced and can be neglectec
but in a few cases.
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 11 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 12 —
a 2 a 2 sp 2 s 2
sa2 = p s p + – sT = a2 p + T 20 ).
T T
The standard uncertainty of air density is then
sp 2 s 2
sa = a T .
p + T
Buoyancy introduces both a bias and an uncertainty, mainly
stemming from the limited knowledge of the density of the
sample and the air density at the time of weighing. Its
influence is systematic as far as the average buoyancy goes,
and is random to the extent, as the true sample and air
densities are unknown 21).
Sensitivity Temperature Drift
Temperature may also influence sensitivity, as the weighing
transducer of the balance may exhibit a temperature
dependent characteristic. This influence is expressed by the
(sensitivity) temperature coefficient (TC) of the balance and is
defined as the ratio of relative change of sensitivity ∆S/S0
and the temperature difference ∆T causing it, namely
∆S
S
TCS = 0 .
∆T
The TC of sensitivity only states how sensitive the slope of the
balance’s characteristic curve reacts to temperature. To
provoke a slope change, a change of ambient temperature,
with respect to the temperature at the last sensitivity
adjustment, must occur. The sensitivity change can be
expressed as
∆S TC = TCS∆T S0 .
The influence of ambient temperature on sensitivity is of
systematic nature. However, because the actual TC of a
balance are usually unknown to the operator, this influence is
treated as a random contribution 22).
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 13 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 14 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 15 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 16 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 17 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 18 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 19 —
2
sTCS = 13 TCS 2 and sT2 = 13 ∆T 2lim ,
lim
while the variance due to apparent sensitivity, caused by
buoyancy, is
acal 2 2 a 2 1 2
sbuoy = 2 sρ + ρ1 sa + 2 sρ + ρ
2
sa .
rel ρ cal cal cal cal ρ
Eventually the combined variance equals
u2 = u2RD+u2RP+u2NL+m2 s∆µ 2 +sS,TC
2 + sbuoy 2 .
rel rel
Taking the square root of the uncertainty variance yields the
standard uncertainty
u = u2RD+u2RP+u2NL+m2 s∆µ 2 +sS,TC 2 + sbuoy 2 .
rel rel
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 20 —
k=kP
between coverage factor k and coverage probability) P 28).
Sometimes, the missing probability is used, according to
Q = 1–P .
This number states the fraction of those values which are
expected to lie outside the uncertainty interval. Frequently
used coverage factors are 2 and 3, which corresponds to a
coverage probability of about 95% and 99.7%, respectively.
Thus, between a sample of mass m and the result R (single
observation) of its weighing, the following relationship holds
R –U ≤ m ≤ R +U ,
where the expanded uncertainty
U = k⋅u 29),
is obtained by multiplying the standard uncertainty u with
the coverage factor k , the latter being an appropriate choice
(2 or 3, for example), from which the coverage probability
k —> P = f k
follows, or vice versa, the coverage probability is
appropriately chosen, and from it the coverage factor is
determined
P —> k = f P ,
according to the table given.
Eventually, the weighing result can be expressed as
28) Coverage Factor Level of Confidence Missing Probability
(Single Sided) (Coverage Probability)
k P Q = 1–P
——————— ———————–– ———————
1 68.27% 31.73%
1.645 90% 10%
1.960 95% 5%
2 95.45% 4.55%
2.576 99% 1%
3 99.73% 0.27%
4 99.994% 0.006%
5 99.99994% 0.00006%
(A normal probability distribution is assumed for this table.)
29) To keep things simple, we have consequently refrained from deter-
mining, or correcting for, the degree of freedom. Of course, nothing
stands against the notion of correcting for the degree of freedom, if it
is known of all individual contributions. An instruction for how to de-
termine the correction factor can be found in [GUM].
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 21 —
m = R –δ ±U .
We now have all pieces in place to discuss the weighing
result of an example.
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 22 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 23 —
NLlim = 0.03mg
10g
2 = 2s2 = 2 NL2 = 2 0.03mg 2 = 600×10–12g 2
uNL NL 3 lim 3
Calibration weight tolerance
From the specification, we have
∆µ cal = 1.5 ppm
s∆µ 2 = 13 ∆µ 2lim = 13 1.5 ppm 2= 0.75ppm2
rel
Temperature drift
From the balance specifications we know the TC
TCS = 1.5 ppm/K
lim
sTCS = 13
2 TCS 2 = 13 1.5 ppm/K 2 = 0.75 ppm/K 2
lim
and from the environmental conditions the temperature
excursion we have
∆T lim = 2K sT2 = 13 ∆T 2lim = 13 2K 2 = 1.33K 2 .
Hence, the variance of sensitivity temperature drift amounts to
2
sS,TC 2
= sTCS sT2 = 0.75 ppm/K 2⋅1.33K 2= 1ppm2 .
Buoyancy
From the environmental conditions, we have the laboratory
temperature
22°C → T = 295K ,
and we know already the laboratory temperature variance
sT2 = 1.33K 2 .
Considering air pressure, no information is available. We
consider a standard pressure of
p = 1bar .
Assuming that the laboratory is essentially in free pressure
exchange with the outside atmosphere, we could try and find
some data about the latter. From weather data we can
estimate a pressure standard deviation of atmospheric
pressure fluctuations to be in the order of
s p = 10mbar .
We thus get for the air density standard deviation
sp 2 s 2
10mbar 2 1.33K 2
sa = a p + T
T = 1.2 kg/m3 + ≈
1bar 295K 2
≈ 1.2 kg/m3 10–4+1.5×10–5 ≈ 12.9g/m3 .
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 24 —
≈ 2.2×10–9g 2 ,
from which we obtain the standard uncertainty
u ≈ 47µg .
With the required coverage factor
k=2 ,
corresponding to a coverage probability of
P ≈ 95% ,
we get for the expanded uncertainty
U ≈ 0.094mg .
Conclusion Of Example 1
The major contributors to weighing uncertainty in this
example are repeatability and non-linearity, in that order.
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 25 —
Uncertainty Charts
The example discussed just pinpoints one weighing situation,
given by circumstances of the particular weighing object and
container, as delineated in example 1. Instead, we are
interested in the uncertainties for all possible combinations of
sample masses and gross weights applicable to a given
balance.
Uncertainty Chart Of A Semi-Micro Balance
To chart the weighing uncertainty obtained with a balance,
we use the formula of standard uncertainty, derived in the
previous chapter, and multiply it with the coverage factor.
This yields for the expanded uncertainty
U = k u2RD+u2RP+u2NL+m2 s∆µ 2 +sS,TC 2 .
rel
To prevent the cluttering of the chart with too many
parameters, in the above expression the contribution of
buoyancy was dropped. However, the influence of buoyancy
is individually shown in the chart for several sample densities.
Ubuoy = kubuoy = km ubuoy =
rel
32) siehe [Kehl]
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 26 —
acal 2
1 s 2+ a s 2+ 1 s 2
= km s
2 ρ cal + ρ cal acal ρ a
ρ cal ρ2 ρ
As instrument we will use here the AT201 semi-micro balance
introduced in the previous example, hence we will continue to
employ the same specifications.
We have to be aware that sometimes the data sheet may state several
values for a single property, as these may depend on the
circumstances of the weighing. This is the case for the AT201, since its
repeatability and linearity depend on the gross weight (tare plus
sample) and the sample weight, respectively:
Repeatability up to 50g 0.015mg
50-200g 0.04mg
Non-Linearity within 10g 0.03mg
within 200g 0.12mg
This constitutes four operational cases:
I) gross below 50g & sample less than 10g
II) gross 50…200g & sample less than 10g
III) gross below 50 & sample 10…50g
IV) gross 50…200g & sample 10…200g
In cases I) and II), the more favorable non-linearity specification
applies, while in cases I) and III) the more favorable repeatability
specification applies. As it turns out, the numerical values for case III)
are almost identical with those of case IV); case III) is therefore omitted
in the chart.
All other parameters and assumptions in example 1, such as
specifications of the balance and environmental conditions
and assumptions, are used unaltered for this chart (except
sample mass, which is the independent variable, and
container mass, which is a parameter, of the chart).
Relative Uncertainty Chart Of A Semi-Micro Balance
Often it is more convenient, or it may even be required, to
use the relative uncertainty, i.e., the uncertainty normalized to
the sample mass
u
urel = m .
This yields
1
urel = m u2RD+u2RP+u2NL+m2 s∆µ 2 +sS,TC 2 + sbuoy 2 =
rel rel
= 1 u2 +u2 +u2 + s 2 +s2 + s 2
m2 RD RP NL ∆ µ rel S,TC buoy rel
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 27 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 28 —
and
Urel = kurel = k 1 u2 +u2 +u2 + s 2 +s2 + s 2
m2 RD RP NL ∆ µ rel S,TC buoy rel
Again, in this chart we dropped the influence of buoyancy.
Urel = k 1 u2 +u2 +u2 + s 2 +s2
2m RD RP NL ∆µ S,TC
rel
to represent buoyancy separately in the chart
Ubuoy = k ubuoy = k sbuoy .
rel rel rel
These uncertainty charts of the AT201 illustrate that there are
three characteristic behaviors of the uncertainty functions that
are related to three ranges with respect to the sample mass.
First, for small sample masses, below 10g, uncertainty stays
constant. This is attributable to the contributions which
dominate uncertainty and are independent of sample mass.
These are mainly repeatability, non-linearity and readability,
usually in that order of importance. This means that the
weighing uncertainty stays essentially constant and is not a
function of the sample weight. Therefore, as the sample mass
decreases, the relative uncertainty increases in this range.
Second, for large sample masses, larger than half the
maximal weighing capacity, the contributions that are
proportional to sample mass dominate uncertainty. These are
temperature drift 33), initial accuracy (calibration weight
tolerance), long-term drift and buoyancy. Therefore, the
relative uncertainty stays essentially constant in this range. If
the sample density is low, buoyancy may become by far the
largest contributor to uncertainty in this region.
Third, between the two ranges is a transition range where the
constant contributions as well as the ones proportional to
sample mass are of the same order of magnitude.
Although there may be specific differences between balance
models, this behavior is typical for most laboratory balances.
For ”small”samples—i.e., sample masses below 1/10 to
1/20 of the weighing capacity—the (absolute) uncertainty
stays constant, while the relative decreases with increasing
sample mass). For ”large”samples—i.e., sample masses of
1/4 to 1/2 of the weighing capacity and larger—the
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 29 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 30 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 31 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 32 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 33 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 34 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 35 —
Conclusions
Weighings, i.e., the determination of the mass of a weighing
object, are subject to influences from various sources.
Generally, they cause a measurement bias as well as a
measurement uncertainty.
The influences can originate from the balance and from the
weighing object. Thereby, the environment plays a major
role. The weighing result is influenced by non-ideal
characteristics of the balance, such as readability,
repeatability, corner load error, non-linearity and long-term
drift, or by disturbances caused by environmental conditions
such as temperature drift. The weighing sample gets affected
mainly by ambient conditions, such as air temperature, air
humidity, air draft, and substantially so, by air buoyancy.
Often, the mass of the object weighed can be the culprit, as it
may vary. One should be aware that the mass of an object
may not be the unchanging physical quantity, for which it is
often taken.
Whether these influences are of relevance for the weighing
result depends on several conditions, which should be
checked for every application. As a rule of thumb one can
say that the higher the required measurement resolution of
the weighing is, the more likely the influences will manifest
themselves.
Contrary to a maintained balance which as a rule does not
introduce a measurement bias that is to the detriment of the
measurement accuracy, air buoyancy can cause a substantial
measurement bias. There is an international regulation about
conventional mass as how to deal with this influence.
However, it was primarily issued to establish rules for the
dissemination of mass standards. For weighing objects with
distinctly different densities than the conventional density
(8000kg/m3 ) defined is this regulation it is not applicable.
Dependent on the application, the measurement bias caused
by air buoyancy can exceed measurement uncertainty by far.
The measurement uncertainty is determined by the
performance of the balance, the properties of the weighing
object and the state of the environment. The properties of the
balance are partly of random nature (such as readability,
repeatability) and partly of systematic nature (such as non-
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 36 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 37 —
Appendices
Buoyancy
If a weighing takes place in a vacuum on the surface of
earth, a body with mass m produces a weight force of
G = mg .
If the same body is weighed in the atmosphere, it will be
subject to an air buoyancy force which is opposed to the
weight force and diminishes the latter. According to
Archimedes’ principle 39), a body receives a buoyant force
equal to the weight force of the fluid displaced. This buoyant
force can be determined from the density of the air a and the
volume V displaced, according to
GF = aVg .
Since the volume of displacement is equal to the volume of
the body weighed, we can determine it from the density ρ
and the mass m of the body
m
V= ρ
which yields for the buoyant force
m
GF = a ρ g .
For the sum of all forces acting on the buoyant body
(apparent weight) we have
a
GA = G – GF = mg–a m ρ g = g 1– ρ m .
Sometimes, within the above expression the term
a
m A = 1– ρ m
is referred to as ”apparent mass”.
Sensitivity Adjustment (Determination of the calibration factor)
The sensitivity of a balance is established with the aid of a
calibration mass (a.k.a. calibration standard).
Upon loading a mass m onto a balance, it will display a
value
a
R = cg 1– ρ m .
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 38 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 39 —
a
g 1– ρ 1
R= a m .
gcal 1– cal 1+∆µ cal
ρ cal
Dependence Of Air Density On Temperature & Pressure
The density of air depends mainly on air temperature and
pressure; to a lesser extent also on its relative humidity and its
composition. For the sake of uncertainty estimation, we give
here a simplified theory for the dependence of air density on
temperature and pressure.
If we consider air to be an ideal gas—not a severe
simplification for our purposes here—the parameters of an
amount m of air, such as its volume V , absolute
temperature T and pressure p , obey the law of an ideal
gas, according to
pV = nkT ,
where n is the number of molecules in that amount, and k is
Boltzman’s constant. By definition, the density a of this air
equals
m
a= ,
V
and substituting the volume of air from the gas law, we get
p
a= m .
nk T
From this expression we take the partial derivative with
respect to pressure
∂a m 1
∂ p = nk T
and with respect to temperature
∂a –p
= m 2 .
∂T nk T
Substituting the constant term m/ nk in this expressions with
the aid of the original formula for air density, we get
∂a aT 1 a
∂p = p T = p
and
∂a = aT – p = – a 40 ).
∂T p T2 T
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 40 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 41 —
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001
— 42 —
References
[Niels] Henrik S. Nielsen: How long is a piece of string?
NCSL Workshop and Conference, Toronto, Canada, July 17-
20, 2000.
[Clark] John P. Clark: Evaluation Of Methods For Estimating The
Uncertainty Of Electronic Balance Measurements.
Weighing 2000, South Yorkshire Trading Standards Unit,
Sheffield (UK), June 13-14, 2000.
[R33] Conventional value of the result of weighing in air.
International Organization of Legal Metrology (OIML), Paris,
France. International Rrecommendation No 33, 1979.
[VIM] International Vocabulary Of Basic And General Terms In
Metrology
International Organization for Standardization, Geneva,
1993
ISBN 92-67-01075-1
[GUM] Guide To The Expression Of Uncertainty In Measurement
First edition 1995
International Organization for Standardization, Geneva
ISBN 92-67-10188-9
[Rth1] Arthur Reichmuth: Non-Linearity Of High Resolution
Laboratory Balances.
Weighing 2000, South Yorkshire Trading Standards Unit,
Sheffield (UK), June 13-14, 2000.
[Rth2] Arthur Reichmuth: Non-Linearity Of Laboratory Balances And
Its Impact On Uncertainty.
NCSL Workshop & Symposium, Toronto (CAN), July 17-20,
2000.
[Kehl] Karl Kehl, Klaus Weirauch, Samuel Wunderli, Veronika
Meyer: The influence of variations in atmospheric pressure on
the uncertainty budget of weighing results.
The Analyst, 2000, Vol. 125, P. 959-962.
[Giac] P. Giacomo: Equation for the Determination of the Density of
Moist Air (1981). Metrologia [1982], 18, pp. 33-40.
[Davis] R.S. Davis: Equation for the Determination of the Density of
Moist Air (1981/91). Metrologia [1992], 29, pp. 67-70.
© Mettler Toledo, A. Reichmuth Weighing Accuracy 2.1 Prtd.: September 25, 2001