Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

CERTIFICATE OF VERIFICATION ROCKWELL

DETAILS

Product type: Rockwell Certificate number: 2020212142431


Model: FH-10-0100 Date of inspection: 12-2-2020
Serial number: FH-1010200205 Ambient temp.: 19,9 °C
Date of manufacturing: 11-2-2020 Date of issuing: 12-2-2020
Verified by: Randy Simons Traceable to: ASTM-E18 / ISO 6508-2

DIRECT VERIFICATION
Force calibration
Load cells

Nr. Serial Range Type Manufacturer


L205 545128 3-150kg LCF455 Futek

Data

Target MEASURED ERROR RANGE


Run 1 Run 2 Run 3 Found Limit Found Limit
(Kg)
(kg) (kg) (kg) (%) (%) (%) (%)
3 3,000 3,000 3,000 0,000 2,000 0,000 1,500
15 15,000 15,000 15,000 0,000 0,667 0,000 0,500
3 3,020 3,020 3,010 0,667 2,000 0,333 1,500
3 3,000 3,000 3,000 0,000 2,000 0,000 1,500
30 30,000 30,000 30,000 0,000 0,667 0,000 0,500
3 3,010 3,010 3,010 0,333 2,000 0,000 1,500
3 3,000 3,000 3,000 0,000 2,000 0,000 1,500
45 45,000 45,000 45,000 0,000 0,667 0,000 0,500
3 3,030 3,030 3,030 1,000 2,000 0,000 1,500
10 10,000 10,000 10,000 0,000 2,000 0,000 1,500
60 59,990 59,990 59,990 -0,017 0,750 0,000 0,563
10 10,030 10,030 10,030 0,300 2,000 0,000 1,500
10 10,000 10,000 10,000 0,000 2,000 0,000 1,500
100 100,010 100,020 100,020 0,020 0,650 0,010 0,488
10 10,030 10,030 10,030 0,300 2,000 0,000 1,500
10 10,000 10,000 10,000 0,000 2,000 0,000 1,500
150 150,010 150,010 150,010 0,007 0,600 0,000 0,450
10 10,050 10,040 10,030 0,500 2,000 0,200 1,500

Page 1/3
Depth measuring system
LINEAR GAUGE USED FOR DISPLACEMENT VERIFICATION

nr Serial Range Type Manufacturer


P101 68318829 12mm MT1271 Heidenhain

Data

Target Standard Measured Error Limit


(µm) (µm) (µm) (µm) (µm)
10 11,30 11,30 0,00 0,50
90 92,20 92,30 0,10 0,50
170 174,40 174,60 0,20 1,00
250 252,10 252,50 0,40 1,00

Test cycle
Timing device

nr Serial Range Type Manufacturer


T103 181261014 Sec/24hr ±0.01 06-662-56 Control Company

Data

Apply time preload (tPA) Dwell time preload (tPF) Time to mainforce (tTA) Dwell time mainload (tTF) Dwell time recovery (tR)
(s) (s) (s) (s) (s)
1,74 2,61 3,75 4,11 3,02
<2 1 to 4 1 to 8 2 to 6 1 to 5

HRT & HRN


Apply time preload (tPA) Dwell time preload (tPF) Time to mainforce (tTA) Dwell time mainload (tTF) Dwell time recovery (tR)
(s) (s) (s) (s) (s)
1,32 2,14 1,75 3,55 2,71
<2 1 to 4 0 to 4 2 to 6 1 to 5

Machine Hysteresis
Data

Measured values Error


Target
1. 2. 3. 4. 5. 6. 7. 8. 9. 10. AVG Limit
(HR) (HR) (HR) (HR) (HR) (HR) (HR) (HR) (HR) (HR) (HR) (HR) (HR)

100 99,12 99,15 99,18 99,20 99,20 99,20 99,22 99,20 99,22 99,20 -0,79 1,00

Page 2/3
Indirect verification
Test Blocks
Indenters used

nr serial type Certificate nr


FH-300-0005 6722 HR Dia 6722
IN/7506 157822 1/16"Ball 298276

Data

Hardness block Measured Average Error Repeatability

Value 1 2 3 4 5 AVG Meas. Allowed Meas. Allowed


Serial Scale
(HR) (HR) (HR) (HR) (HR) (HR) (HR) (HR) (HR) (HR) (HR)
EP18213389 HRA 31,1 31,25 31,29 31,33 31,47 31,34 31,34 0,24 1,00 0,22 1,38
EP17200193 HRA 61,0 61,11 61,03 61,03 61,08 61,18 61,09 0,09 1,00 0,15 0,80
EP18211404 HRA 83,1 83,13 83,18 83,18 83,13 83,06 83,14 0,04 0,50 0,12 0,80
EP18213244 HRC 27,0 27,21 27,10 27,05 27,05 27,21 27,12 0,12 1,00 0,16 1,46
EP18211345 HRC 47,1 47,01 46,91 46,96 46,86 46,96 46,94 -0,16 1,00 0,15 1,06
EP18213231 HRC 62,4 62,51 62,43 62,43 62,57 62,47 62,48 0,08 0,50 0,14 0,80
EP17198933 HR15N 74,5 74,62 74,47 74,57 74,57 74,67 74,58 0,08 1,00 0,20 1,20
EP18211325 HR15N 85,0 85,01 85,06 84,96 84,96 85,01 85,00 0,00 1,00 0,10 1,20
EP18211359 HR15N 91,2 91,05 91,05 91,10 91,20 91,25 91,13 -0,07 0,70 0,20 1,00
EP18209170 HR45N 26,3 26,49 26,44 26,29 26,47 26,25 26,39 0,09 1,00 0,24 2,00
EP18211314 HR15N 50,4 50,44 50,44 50,51 50,51 50,44 50,47 0,07 1,00 0,07 1,98
EP18211306 HR45N 69,4 69,51 69,51 69,51 69,41 69,41 69,47 0,07 0,70 0,10 1,00
EP18213326 HRB 28,7 28,92 29,13 28,99 29,23 29,37 29,13 0,43 1,50 0,45 2,00
EP18213274 HRB 64,3 64,53 64,53 64,39 64,71 64,66 64,56 0,26 1,00 0,32 1,50
EP18213365 HRB 93,4 93,58 93,58 93,62 93,49 93,49 93,55 0,15 1,00 0,13 1,46
EP18211180 HR30T 46,4 46,66 46,81 46,77 46,77 46,92 46,79 0,39 1,50 0,26 2,00
EP18211369 HR30T 58,2 57,77 57,81 57,88 57,71 57,92 57,82 -0,38 1,00 0,21 1,50
EP18211175 HR30T 72,6 72,90 72,95 73,02 73,02 72,84 72,95 0,35 1,00 0,18 1,50
-
-
-
-
-
-
-

Randy Simons

Instrument verification Passed

*The proving devices and testblocks that are used are all traceable to ASTM / ISO: certificates on file.
Page 3/3

You might also like