Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

STCAL

Spark Test Calibration System

>>Accurate calibration of AC, DC and high frequency spark testers

>>Lightweight modular components

>>Built-in spark tester profiles

>>Calibration certificates with data can be printed or downloaded


for storage

>>Optional SM Sensitivity Module checks fault detection sensitivity


to EN50356, IEC/CEI 62230 and NEMA specifications STCAL Calibration System
with HF-15A Spark Tester

C linton Instrument Company introduces a modular


calibration system for AC mains, high frequency,
and DC spark testers. The system includes separate
high- and low-frequency voltmeters and a case
containing a Windows-based touch-screen tablet and
connecting cables.

The calibration procedure is a series of simple steps:

>> Select the appropriate voltmeter and cables


>> Connect the voltmeter to the spark tester and Optional STCAL-SM Sensitivity Module
tablet
>> Choose a spark tester profile from the touch
screen menu The system calibrates AC mains spark testers to 30kV,
high frequency units to 15kV, and DC spark testers to
>> Follow the on-screen directions for completing 40kV.
the calibration
The system generates a certificate of calibration The optional SM Sensitivity Module has a rotating
suitable for download, printing and storage in the spark gap that controls arc duration and gap distance.
customer’s computer. Data is traceable to NIST The operator simply selects the type of test to be
standards when the STCAL is incorporated into an performed; the SM displays the detection current and
effective quality assurance program. the required number of controlled faults the spark
tester must identify to meet the norm. After the test,
STCAL components are light enough to be carried to the operator compares the SM fault count to the faults
the hard-to-reach locations where spark testers are counted by the spark tester to determine its conformity
sometimes situated. with the specification.
MODEL STCAL
SPECIFICATIONS
High Frequency Calibrator Low Frequency Calibrator
For voltage calibration of high frequency spark testers. For voltage calibration of DC spark testers to 40kV and 50/60 Hz AC spark testers
to 30kV.

Components..........................................High Frequency Voltmeter (HFVM); Tablet Components..........................................Low Frequency Voltmeter (LFVM); Tablet
and connecting cables in soft case. and connecting cables in soft case.

Input Voltage Range..............................500 volts to 15kV AC. Input Voltage Range..............................100 volts to 40kV DC and 1kV to 30kV AC.

Input Frequency Range..........................1 kHz to 5 kHz. Input Frequency Range.......................... 50-60 Hz.

Output Voltage Reading Accuracy............+/- .5%. Output Voltage Reading Accuracy............ +/- .5%.

HFVM Weight........................................8.9 lbs. (4.1 kg). LFVM Weight......................................... 5.3 lbs. (2.4 kg.)

HFVM Dimensions.................................5.9” H x 9.8” W x 10.2” D LFVM Dimensions..................................5.9” H x 9.8” W x 10.2” D


(150mm H x 249mm W x 259mm). (150mm H x 249mm W x 259mm).

Power...................................................from tablet via USB. Power...................................................from tablet via USB.

Tablet...................................................10.1” Windows 8 with charger, 1.5 lbs. Tablet...................................................10.1” Windows 8 with charger, 1.5 lbs.
(approx dimensions and weight). (approx dimensions and weight).

Connecting Cables..................................HV Cables 1.5’ and 3’; USB Cable 6’; Connecting Cables..................................HV Cables 1.5’ and 3’; USB Cable 6’;
RS485 Cable 10’; Ground Cable. RS485 Cable 10’; Ground Cable.

Spark Testers that can be calibrated with the STCAL (partial list) Certificate of Calibration
Certificate No. A418-0000@4/4/2013
Voltmeter Calibration Available
Model Number
Equipment Owner: Calibration Service Provider:
Name: Clinton Instrument Company Name: Joe's Calibration Service
Required AutoCal Assisted Manual Address: 295 East Main Street Address: 10 Main Street
Clinton, CT 06413 Colchester, CT 06111
Clinton AC-30A & AC-30AS 50/60 Hz 30kV AC USA USA

Spark Testers LFVM


Equipment Manufacturer: Calibration performed at:
Name: Clinton Instrument Company Name: Clinton Instrument Company

HFVM
Model: AC-30AR Address: 295 East Main Street
Clinton HF-15A High Frequency Spark Testers Description: 30kV AC Spark Tester Clinton, CT 06413
Electrode Model: BD-A418 USA
Unit Serial Number: A418-0000
Clinton HF-20E, HF-20H, HF-20GA and HF-
20GB High Frequency Spark Testers HFVM
Date of Calibration: 4/4/2013 2:21:04 PM Temperature: N/A Humidity: N/A

Clinton FL-8A and FL-20A Cable Fault Locators LFVM As Found Readings
Calibration Data
As Left Readings
Actual Nominal On Actual
Displayed Voltage Voltage Range Acceptable Acceptable Voltage
Clinton DC-10A, DC-20A, and PDC-20A DC Spark Voltage (kV) (kV) Result (kV) (KV) Min (kV) Max (kV) (kV) Result

Testers LFVM 5.00


10.00
5.03
10.07
PASS
PASS
05.00
10.00
30.00
30.00
04.90
09.80
05.10
10.20
5.03
10.08
PASS
PASS
15.00 15.22 PASS 15.00 30.00 14.70 15.30 15.16 PASS
Clinton STB-10G, STB-20G and PDC-20 DC 20.00 20.13 PASS 20.00 30.00 19.60 20.40 20.11 PASS

Spark Testers LFVM 25.00


30.00
24.96
29.33
PASS
FAIL
25.00
30.00
30.00
30.00
24.50
29.40
25.50
30.60
25.02
29.93
PASS
PASS

Clinton STA-30A and STA-30GB 50/60 30kV AC Condition as Received: Condition as Left:

Spark Testers LFVM Out of Calibration In Calibration

Standards Used

DC Spark Testers by other manufacturers, to 40kV LFVM Model No.


LFVM
Serial No.
LFVM-0003
Calibration Date
01/15/2013
Due Date
01/15/2014
Traceability No.
L01E865C9C

High Frequency Spark Testers by other manufac-


turers, to 15kV HFVM The calibrations of the equipment identified above have been checked against Measurement Standards (Reference)
maintained and owned by The Clinton Instrument Company. The accuracy of these standards is traceable to the national
standards at the National Institute of Standards and Technology in Washington, D.C. and Boulder, CO.

50/60 Hz AC Spark Testers by other manufactur-


ers, to 30kV LFVM Calibration Technician: N/A

Sample Calibration Certificate


Specifications subject to change without notice. 03/14 EN

295 East Main St. • Clinton, CT 06413 USA • Tel: 860.669.7548 • Fax: 860.669.3825 • www.clintoninstrument.com

You might also like