Libra Technical Training 2015
Libra Technical Training 2015
- Product Description
- Tool Requirements
- Principles of Operations
- Testing/Repair
- PM Requirements
3 | Covidien | November 8, 2016 | Confidential
“Nellcor Bedside SpO2 Patient Monitoring
System” (aka “LIBRA”)
USB Ports
For data
download and
SpO2 Connector upgrades
Insert sensor here
Jog Dial
Use to control system functions
AC Power Connector
Nurse Call Port
Battery Cover
AC Power Indicator
Interference Indicator
Nurse Call
Board
NELL1SR Main
Board
Power Supply PCBA
Converts AC power to DC
• Accepts 100/240VAC, 50/60Hz
Distributes Power to Charger Board
AC Power
DC Power (In)
(Out)
Fuses F1,
F2
Nurse Call Board
Provides Nurse Call Functionality
DC Power
Fuse F2 Fuse F1 (In)
RJ-11 Connector
Charger Board
Charges the Battery
Distributes Power to the other Boards
To Nurse
Fuse F1
Call Board
To Main
Board DC In (from
Power Supply)
NELL1SR PCBA
Calculates SpO2 and PR Values
RoHS compliant Patient Signals
In (connects to
Main Board)
Power In, SpO2 &
PR out (connects
to Main Board)
RJ-11
Connector
Speaker
SpO2
Power Wire
Nurse Call
Wire
Bi-Weekly RMS Program Review l <Date> I <Project Name> I PROPRIETARY & CONFIDENTIAL
Sub Assemblies
Front
Jog Dial
Back
High Voltage
Spare Parts Illustration
# #
2
Add Torque wrench!!!!!! 1
SRC-MAX Doc 10
DS-
100A
Stop Watch
- Read and consider all warnings and notes prior to performing any disassembly.
- Follow disassembly sequence (Fig 12-1).
- NOTE: On page 167 & 176 in the “Tools Required” sections of the manual calls for
a ¼ “ driver…The ACTUAL size that is required is a 13/64 driver.