Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

An American National Standard

Designation: D 2035 – 80 (Reapproved 2003)

Standard Practice for


Coagulation-Flocculation Jar Test of Water1
This standard is issued under the fixed designation D 2035; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope 4. Summary of Practice


1.1 This practice covers a general procedure for the evalu- 4.1 The coagulation-flocculation test is carried out to deter-
ation of a treatment to reduce dissolved, suspended, colloidal, mine the chemicals, dosages, and conditions required to
and nonsettleable matter from water by chemical coagulation- achieve optimum results. The primary variables to be investi-
flocculation, followed by gravity settling. The procedure may gated using the recommended practice include, but are not
be used to evaluate color, turbidity, and hardness reduction. limited to:
1.2 The practice provides a systematic evaluation of the 4.1.1 Chemical additives,
variables normally encountered in the coagulation-flocculation 4.1.2 pH,
process. 4.1.3 Temperature, and
1.3 This standard does not purport to address the safety 4.1.4 Order of addition and mixing conditions.
concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and 5. Significance and Use
health practices and determine the applicability of regulatory 5.1 This practice permits the evaluation of various coagu-
limitations prior to use. lants and coagulant aids used in the treatment of water and
waste water for the same water and the same experimental
2. Referenced Documents conditions.
2.1 ASTM Standards: 2 5.2 The effects of concentration of the coagulants and
D 1129 Terminology Relating to Water coagulant aids and their order of addition can also be evaluated
D 1192 Specification for Equipment for Sampling Water by this practice.
and Steam in Closed Conduits3
D 1193 Specification for Reagent Water 6. Interferences
D 1293 Test Methods for pH of Water 6.1 There are some possible interferences that may make the
D 1889 Test Method for Turbidity of Water determination of optimum jar test conditions difficult. These
D 3370 Practices for Sampling Water from Closed Conduits include the following:
6.1.1 Temperature Change (During Test)—Thermal or con-
3. Terminology vection currents may occur, interfering with the settling of
3.1 Definitions—For definitions of terms used in this prac- coagulated particles. This can be prevented by temperature
tice, refer to Terminology D 1129. control.
6.1.2 Gas Release (During Test)—Flotation of coagulated
1
floc may occur due to gas bubble formation caused by
This practice is under the jurisdiction of ASTM Committee D19 on Water and mechanical agitator, temperature increase or chemical reaction.
is the direct responsibility of Subcommittee D19.03 on Sampling of Water and
Water-Formed Deposits, Analysis of Water for Power Generation and Process Use, 6.1.3 Testing-Period— Biological activity or other factors
On-Line Water Analysis, and Surveillance of Water. may alter the coagulation characteristics of water upon pro-
Current edition approved July 3, 1980. Published October 1980. Originally longed standing. For this reason the period between sampling
approved in 1964. Last previous edition approved in 1980 as D 2035–80.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
and testing should be kept to a minimum, with the time being
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM recorded.
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
3
Withdrawn.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

1
D 2035 – 80 (2003)

FIG. 1 Reagent Rack for Multiple Stirrer Jar Test Apparatus

7. Apparatus 8.3 The following chemicals and additives are typical of


7.1 Multiple Stirrer— A multiposition stirrer with continu- those used for test solutions and suspensions. The latter, with
ous speed variation from about 20 to 150 rpm should be used. the exception of coagulant aids, may be prepared daily by
The stirring paddles should be of light gage corrosion-resistant mixing chemicals with water to a concentration of 10 (60.1)
material all of the same configuration and size. An illuminated g/L (1.0 mL of test solution or suspension when added to 1 L
base is useful to observe the floc formation. Precautionary of sample is equivalent to 10 mg/L):
measures should be taken to avoid heat being imparted by the Prime Coagulants
illumination system which may counteract normal settling. Alum[Al2(SO4)3·18H2O]
Ferric sulfate [Fe2(SO4)3·xH2O]
7.2 Jars (or Beakers), all of the same size and shape; Ferric chloride (FeCl3·6H2O)
1500-mL Griffin beakers may be used (1000-mL recommended Ferrous sulfate (FeSO4·7H2O)
minimum size). Magnesium carbonate (MgCO3·3H2O)
Sodium aluminate (NaAlO2)
7.3 Reagent Racks— A means of introducing each test Coagulant Aids
solution to all jars simultaneously. There should be at least one Activated silica
rack for each test solution or suspension. The racks should be Anionic (polyelectrolyte)
Cationic (polyelectrolyte)
similar to that shown in Fig. 1. Nonionic Polymer
Oxidizing Agents
Chlorine (Cl2)
8. Reagents
Chlorine dioxide (ClO2)
8.1 Purity of Reagents—Reagent grade chemicals shall be Potassium permanganate (KMnO4)
Calcium hypochlorite [CaCl(ClO)·4H2O]
used in all tests. Unless otherwise indicated, it is intended that Sodium hypochlorite (NaClO)
all reagents shall conform to the specifications of the Commit- Alkalis
tee on Analytical Reagents of the American Chemical Society, Calcium carbonate (CaCO3)
Dolomitic lime
where such specifications are available.4 Other grades may be (58 % CaO, 40 % MgO)
used, provided it is first ascertained that the reagent is of Lime, hydrated [Ca(OH)2]
sufficiently high purity to permit its use without lessening the Magnesium oxide (MgO)
Sodium carbonate (Na2CO3)
accuracy of the determination. Sodium hydroxide (NaOH)
8.2 Purity of Water— Unless otherwise indicated, reference Weighting Agents
to water for reagent preparation shall be understood to mean Bentonite
Kaolin
Type IV reagent water conforming to Specification D 1193. Other clays and minerals
Miscellaneous
Activated carbon (powdered)

4
Reagent Chemicals, American Chemical Society Specifications, American 8.4 Coagulant Aids— There are numerous commercially
Chemical Society, Washington, DC. For suggestions on the testing of reagents not available coagulant aids or polyelectrolytes. All polyelectro-
listed by the American Chemical Society, see Analar Standards for Laboratory lytes are classified anionic, cationic or nonionic, depending
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, upon their composition. These aids may have the ability to
MD. produce large, tough, easily-settled floc when used alone or in

2
D 2035 – 80 (2003)
conjunction with inorganic coagulants. A small dosage (under Flash mix for approximately 1 min after the additions of
1 mg/L) may permit a reduction in the dosage of, or complete chemicals. Record the flash mix time and speed (rpm).
elimination of, the coagulant. In the latter case, the polyelec- 10.4 Reduce the speed as necessary to the minimum re-
trolyte would be considered the prime coagulant rather than a quired to keep floc particles uniformly suspended throughout
coagulant aid. Aids come in powdered and liquid form. the “slow mix” period. Slow mix for 20 min. Record the time
Powdered aids should be prepared as 0.1 % solutions with for the first visible floc formation. Every 5 min (during the
appropriate aliquots to provide proper dosage. Always add slow mix period), record relative floc size and mixer speed
powdered aids to the dissolving water rather than the reverse, (rpm). If coagulant aids are used, mixing speed is critical
and add slowly to the shoulder of a vortex created by stirring. because excessive stirring tends to break up early floc forma-
If a vortex is not formed, the dry powder will merely collect on tion and may redisperse the aid.
the surface of the water in gummy masses and become very 10.5 After the slow mix period, withdraw the paddles and
difficult to dissolve. Dissolving time may vary from several observe settling of floc particles. Record the time required for
minutes to several hours. Suggested manufacturers’ procedures the bulk of the particles to settle. In most cases this time will
for wetting, dissolving, and storing should be followed when be that required for the particles to settle to the bottom of the
available. Liquid forms can be readily prepared to the above beaker; however, in some cases there may be interfering
strength without difficulty.5 convection currents. If so, the recorded settling time should be
that at which the unsettled or residual particles appear to be
9. Sampling moving equally upward and downward.
9.1 Collect the water sample under test in accordance with 10.6 After 15 min of settling, record the appearance of floc
the applicable Specification D 1192 and Practices D 3370. on the beaker bottom. Record the sample temperature. By
means of a pipet or siphon, withdraw an adequate sample
10. Procedure volume of supernatant liquor from the jar at a point one half of
10.1 Measure equal volumes (1000 mL) of sample into the depth of the sample, to conduct color,6 turbidity, pH and
each of the jars or 1500-mL Griffin beakers. As many sample other required analyses, (Note 1) determined in accordance
portions may be used as there are positions on the multiple with Test Methods D 1889 and D 1293. A suggested form for
stirrer. Locate beakers so that the paddles are off-center, but recording results is appended (see Fig. 2).
clear the beaker wall by about 6.4 mm (1⁄4 in.). Record the NOTE 1—Tests for residual chemicals should be included, for example,
sample temperature at the start of the test. alum; residual Al2O3; copperas; residual Fe2O 3; etc.
10.2 Load the test chemicals in the reagent racks. Use one
10.7 Repeat steps 10.1-10.6 until all pertinent variables
rack for each series of chemical additions. Make up each tube
have been evaluated.
in the rack to a final volume of 10 mL, with water, before
10.8 The times given in 10.3, 10.4, and 10.6 are only
using. There may be a situation where a larger volume of
suggestions.
reagent will be required. Should this condition prevail, fill all
tubes with water to a volume equal to the largest volume of 11. Reproducibility
reagent in the reagent rack. When adding slurries, it may be 11.1 It is recognized that reproducibility of results is impor-
necessary to shake the rack to produce a swirling motion just tant. To demonstrate reproducibility, the so-called 3 and 3
prior to transfer. procedure is suggested. In this procedure, duplicate sets of 3
10.3 Start the multiple stirrer operating at the “flash mix” jars each are treated simultaneously with the same chemical
speed of approximately 120 rpm. Add the test solution or dosages in jars 1 and 4, 2 and 5, and 3 and 6.
suspensions, at predetermined dosage levels and sequence.
12. Keywords
12.1 coagulation; flocculation; jar tests
5
A periodically updated “Report on Coagulant Aids for Water Treatment” is
published by the Environmental Protection Agency Office of Water Supply,
6
Cincinnati, Ohio 45268, listing coagulant aids that may be used in water treatment For the color determination, reference is made to Standard Methods for the
without adverse physiological effects on those using the water, based on information Examination of Water and Waste Water, Fourteenth edition, American Public Health
submitted by the manufacturers or distributors, or both. Association, Inc., New York, NY, 1975, pp. 64–71.

3
D 2035 – 80 (2003)

FIG. 2 Jar Test Data

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

You might also like