PF300 SERIES

DIGITAL POWER METER


USER’S MANUAL
V 2.03
P166681CJ7371124

EVERFINE Corporation (Stock Code: 300306)


ADD:Bldg.1 #669 Binkang Rd., Binjiang Hi-Tech
Zone, Hangzhou(310053), China
Tel :86-571-86698333
Fax :86-571-86696433
E-mail:Global@everfine.net
Globalservice@everfine.net
http://www.everfine.net

Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Foreword

Thank you for purchasing the EVERFINE PF300 Digital Power Meter. This user’s
Manual contains useful information involving the instrument’s functions and
operating procedures as well as precautions that should be noticed during operation.
In order to use the instrument correctly, please read the manual carefully first, then put
it in a right place for quick references.
Notes:

 EVERFINE pursues a continuing improvement of the performance and functions


of its products, therefore, the contents of this manual may be changed without
prior notice.
 Great efforts have been made in preparation of this manual to ensure the accuracy
of its contents. If you have any questions or find any errors, please contact your
P166681CJ7371124

dealer or EVERFINE sales office.


 If you have different understanding to this manual, please refer to the Technical
Service Department of EVERFINE.

Checking package contents

Please check the instrument carefully when you unpack the box for the first time. If
the instrument and related accessories are missing or appear abnormal, please contact
the dealer immediately.

Warm notice to valued customers of EVERFINE


"Ensure the quality, insist on continuous improvement and make every customer
more satisfied" is the quality policy of EVERFINE. Therefore, the quality of
products and services provided by EVERFINE should be better than those have
been promised. If you have further suggestions or advices on our products and
services, please provide your feedback to our quality supervision department.
E-mail: qc@everfine.cn; Tel: +86-571-86698333
Your supervision is the motivation for us to move forward!

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 1


PF300 series User’s Manual

Copyright Statement

The copyright of this manual and the related information belongs to EVERFINE,

and it is protected by the copyright law of Peoples Republic of China and other

relevant international treaties. Copying, modifying, spreading, excerpting,

backing up or translating the whole or part contents of this manual by any

company or personnel without the written permission of EVERFINE is

prohibited. Otherwise it will be treated as infringement and the infringer will

assume law responsibility and all loss of EVERFINE. Any infringement related

above can be traced back to the responsible user by the unique product number

printed in the manual.

If EVERFINE has signed a written agreement with user and the contents in

the agreement are in conflict with above terms, the contents in the written
P166681CJ7371124

agreement have preferential force effect.

2 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Precautions

To ensure the safety of the persons and instruments, please notice the following

general safety precautions:

1. Operation precautions

(1) The rated power supply of the instrument is 100VAC-240V AC, 50Hz /60Hz.

(2) Do not dismantle the instrument without permission as its precision device

inside and partial high voltage. Please contact with sale and technical service

center if internal inspection or debug is needed.

(3) Please do not touch the wiring terminal or testing line to prevent the operator

from electric shock during testing. Inserting or pulling out communication port

with electric is forbidden.


P166681CJ7371124

(4) It is dangerous to touch the input interface when electrical circuit works for

the electric communicating terminals inside.

(5) No pressure on the instrument and cables. Keep the instrument far away from

heat source.

(6) Please make sure that the ventilation on both upper and down panel are

unblocked. Otherwise, the internal temperature may rise to damage the

instrument.

(7) It is extremely dangerous for the instrument comes in connect with liquid like

water or oil, or other metal scraps, and it is forbidden to place any items on the

instrument.

(8) Please cut off the power supply immediately if bad smell or smoke is

perceived and contact with sale and technical service center of EVERFINE.

(9) Please keep distance and pull out the power cord plug from the power outlet if

the instrument is not going to be used for some time.

(10) Ensure all cables are disconnected especially the power cord before delivery.

Carry the instrument by the handles on both sides of the instrument gently

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 3


PF300 series User’s Manual

without collisions.

2. Storage precautions

(1) The instrument shall be deposited in the condition with temperature between

0℃ and 45℃, and relative humidity below 80% with no condensation. The

drying environment at 20℃ is recommended.

(2) Please keep the product packaging materials including carton, cushion layer

and plastic bags to protect the instrument from temperature change, shock and

vibration during transportation.

(3) Please do not deposit the instrument in the environment with dust, smoke or

chemical gas.

(4) Direct sunlight is forbidden.

3. Calibration precautions

(1) In order to ensure measurement accuracy, regular calibration is recommended.


P166681CJ7371124

(2) Calibration period depends on the practical use of the instrument.

Recommended calibration period is at least once a year.

(3) Please connect EVERFINE Test and Calibration center for more information

about calibration. The email is lab@everfine.net.

4 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Contents

Foreword........................................................................................................................ 1

Copyright Statement.......................................................................................................2

Precautions..................................................................................................................... 3

Contents..........................................................................................................................5

Chapter 1 Overview....................................................................................................... 9

Chapter 2 Specifications...............................................................................................11

2.1 Main specifications.........................................................................................11

2.2 General technical requirements...................................................................... 17

Chapter 3 Description of Instrument panel.................................................................. 19

3.1 Front Panel..................................................................................................... 19

3.1.1 Indicative function on the panel.......................................................... 20


P166681CJ7371124

3.1.2 Function of panel key.......................................................................... 22

Save or load measurement configuration parameters...................................24

Quit remote-control...................................................................................... 24

3.1.3 Digital and character display...............................................................25

3.2 Rear Panel...................................................................................................... 25

Chapter 4 Measurement Condition Settings.................................................................28

4.1 Measurement Mode........................................................................................28

4.2 Connection mode (just for PF330)................................................................. 29

4.3 Measurement Range....................................................................................... 34

4.3.1 Range of voltage and current...............................................................34

4.3.2 Range Settings.....................................................................................34

4.3.3 Condition for automatic range switch................................................. 35

4.3.4 Range Switch.......................................................................................36

4.3.5 Range Check....................................................................................... 37

4.3.6 Power Range....................................................................................... 37

4.4 External current sensor input..........................................................................42

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 5


PF300 series User’s Manual

4.5 Mutual inductor transformer ratio.................................................................. 44

4.5.1 Transformer ratio settings....................................................................44

4.5.2 Open or close function of ratio............................................................45

4.5.3 Connection method of mutual inductor...............................................46

4.6 Wave Crest Ratio Setting................................................................................46

4.7 Synchronization source measurement............................................................ 47

4.8 Filter............................................................................................................... 48

4.9 Data refresh rate............................................................................................. 49

4.10 Average......................................................................................................... 50

4.10.1 Average method................................................................................. 50

4.10.2 Average function settings.................................................................. 50

4.11 Data Hold......................................................................................................52

4.11.1 Data Hold...........................................................................................52


P166681CJ7371124

4.11.2 Single Measurement.......................................................................... 52

4.11.3 Peak Hold.......................................................................................... 52

Chapter 5 Measurement Function................................................................................ 53

5.1 Normal Measurement Function......................................................................53

5.1.1 Normal Measurement parameters....................................................... 53

5.1.2 Display window...................................................................................54

5.1.3 Using Method...................................................................................... 55

5.2 Function of Integral Measurement................................................................. 58

5.2.1 Integral Mode...................................................................................... 58

5.2.2 Measuring Parameters of Integration.................................................. 59

5.2.3 Display in windows............................................................................. 59

5.2.4 Using Method...................................................................................... 61

5.3 Function of harmonic measurement............................................................... 64

5.3.1 Harmonic measurement parameters.................................................... 64

5.3.2 Display................................................................................................ 67

6 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

5.3.3 Using method...................................................................................... 69

5.4 Storage function............................................................................................. 73

5.4.1 Measurement data storage................................................................... 73

5.4.2 Measurement configuration storage.................................................... 74

Chapter 6 Other Function............................................................................................. 76

6.1 View the system information..........................................................................76

6.2 Initialization settings...................................................................................... 76

6.3 Zero-location Compensation.......................................................................... 77

6.4 Function of key lock.......................................................................................78

6.5 Function of calculation...................................................................................78

6.5.1 Calculation of wave crest ratio............................................................79

6.5.2 Efficiency calculation(Only for PF330).........................................79

Chapter 7 Software Instruction.....................................................................................80


P166681CJ7371124

7.1 System requirements...................................................................................... 80

7.2 Software installation.......................................................................................80

7.3 Software overview..........................................................................................81

7.4 Measurement settings..................................................................................... 83

7.5 Measurement display......................................................................................86

Chapter 8 Common Failures and Resolution............................................................... 91

8.1 Common failures and resolution.................................................................... 91

8.2 Error Code Query........................................................................................... 92

Chapter 9 Calibration and Inspection...........................................................................95

9.1 Conditions for calibration and inspection...................................................... 95

9.2 Technical documents of calibration and test based........................................ 95

9.3 Equipments for calibration and test................................................................95

9.4 Inspection items and methods........................................................................ 96

9.4.1 Items of calibration and test................................................................ 96

9.4.2 Method of calibration and test.............................................................96

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 7


PF300 series User’s Manual

Appendix...................................................................................................................... 98

Appendix I Definition and symbol of measuring parameter................................98

Appendix II Process of Menu settings............................................................... 101

Overview.................................................................................................................... 109

Specifications..............................................................................................................111
P166681CJ7371124

8 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Chapter 1 Overview

PF300 series digital power meter is a compact and versatile digital power meter,
which is widely used in the fields of production, test, evaluation and research.
Product Standard: Q/HYG 003-2014 Intelligent power meter.
Functional features of PF300 series digital power meter:
(1) Measurement Unit
PF310/PF310A: single channel .
PF330: three channels
(2) Wide Rang Input
Voltage:75mV~600V.
Current: PF310/PF310A:0.025mA~20A;
PF330:2.5mA~20A.
P166681CJ7371124

Frequency: DC,0.5Hz~100kHz.
This instrument also has the function of transformation ratio to measure parameters
through an external voltage or current transformer, or to measure large current
through a current sensor of voltage output type.
(3) Simultaneous measurement of multi-parameter
Conventional measurement, harmonic measurement and integral
measurement can be synchronously carried out.
(4) Function of automatic range swithcing
Automatic switching function for conventional measurement at specified
range.
Automatic range switching for integral measurement.
(5) Function of peak value holding
Hold multiple peak values including maximum, minimum and true effective
values of voltage or current, maximum value of active power, reactive power and
apparent power.
(6) Function of efficiency calculation (Only for PF330)
Efficiency calculation can be realized with single instrument.
(7) Fast display and data update

Copyright © EVERFINE, Copy or spread without authorization is prohibited. 9


PF300 series User’s Manual

Rate of data update can be customized, and the fastest rate can be up to 0.1s.
(8) PC software
The special application software is provided to set up the measurement
parameters quickly and obtain measurement data and waveform.
(9) Communication interface
USB for default, RS-232, RS-485,CAN , LAN or GP-IB(optional).
P166681CJ7371124

10 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Chapter 2 Specifications

2.1 Main specifications

(1) Input specifications


Table 2.1 Input specifications
Items Voltage Current
Direct input:bigger binding post
Input terminals Smaller binding post External current sensor input:
insulation BNC interface
Security isolation floating input
Input type
Resistor divider Shunt input
PF310/PF310A direct input:
About 500mΩ+10μH(range of
200mA and below)
About 5mΩ+10μH(range of 0.25A
P166681CJ7371124

input
About 2MΩ+13pF and above)
impedance
PF330direct input: about
5mΩ+10μH
PF310/PF310A/330 external input:
about 100kΩ
PF310/PF310A direct input:
CF=3:
5mA/10mA/20mA/50mA/100mA/
200mA/0.5A/1A/2A/5A/10A/20A
CF=6:
2.5mA/5mA/10mA/25mA/50mA/10
Measuring PF310/PF310A/PF330direct input: 0mA/0.25A/0.5A/1A/2.5A/5A/10A
range CF=3:15V/30V/60V/150V/300V/600V
(RMS) CF=6:7.5V/15V/30V/75V/150V/300V PF330direct input:
CF=3:0.5A/1A/2A/5A/10A/20A
CF=6:0.25A/0.5A/1A/2.5A/5A/10A
PF310/PF310A/PF330 external
input(optional):
CF=3:2.5V/5V/10V
CF=6:1.25V/2.5V/5V
Frequency
DC, AC 0.5Hz ~ 100kHz
range

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 11


PF300 series User’s Manual

Instantaneous
maximum
The smaller one between 2kVpk and The smaller one between 90Apk and
allowable input
1.5kVrms 30Arms
(peak time ≤
1s)
Continuous
The smaller one between 1.5kVpk and The smaller one between 60Apk and
maximum
1kVrms 22Arms
allowable input
Line filter: OFF, 500Hz or 5kHz
Filter
Frequency filter: OFF, 500Hz or 5kHz
(2) Measurement accuracy
Table 2.2 Measurement accuracy of voltage, current and active power
Temperature: 23℃±5℃

Humidity: 30%RH~75%RH

Input waveform: stable sine wave


Measurement
Input range: 10%~100% of the range
conditions
P166681CJ7371124

Common mode voltage: 0 V

Power factor: COSΦ=1

Wave crest ratio:CF=3

Items Voltage(V)/Current(A) Active power(W)

DC ±(0.1%rdg+0.1%rg+1digit) ±(0.1%rdg+0.2%rg+1digit)

10Hz≤f<45Hz ±(0.1%rdg+0.2%rg+1digit) ±(0.3%rdg+0.2%rg+1digit)

45Hz≤f≤65Hz ±(0.05%rdg+0.1%rg+1digit) ±(0.05%rdg+0.1%rg+1digit)

65Hz<f≤1kHz ±(0.1%rdg+0.2%rg+1digit) ±(0.2%rdg+0.2%rg+1digit)


±[(0.1%rdg+0.3%rg)
1kHz<f≤10kHz ±(0.07×f %rdg+0.3%rg+1digit)
+(0.067×(f-1)%rdg)+1digit]
±[(0.5%rdg+0.5%rg) ±[(0.5%rdg+0.5%rg)
10kHz<f≤100kHz
+(0.04×(f-10)%rdg)+1digit] +(0.09×(f-10)%rdg)+1digit]
Frequency
f in the above formula is the frequency of the input signal, with unit of kHz.
(f)
Measurement can be performed with low pass filters and the cut -off
Note Influence
frequency is 5kHz or 500Hz.
from line
The measurement accuracy is increased ±0.5% range error when the line
filter
filter is on and frequency is between 45Hz and 65Hz.

12 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Influence
The error when the wave crest ratio is set as 6 is two times as the one when
from wave
wave crest ratio is set as 3.
crest ratio
Temperatu
The measurement accuracy is increased ±0.02%rg/℃ as the temperature
re
is 5℃~18℃ or 28℃~40℃.
influence
100%~130
The measurement accuracy is increased ±0.5%rdg.
% range
0.5Hz~10Hz: all the measurement values displayed are for reference.
Frequency
DC,10Hz~45Hz,400Hz~100kHz: the accuracy of power is for reference
influence
when the current exceeds 20A.
For range of direct input current, the measurement accuracy of current is
increased ±10μA and the measurement accuracy of power is increased ±
Other 10μA/range ×100%.
influence For range of external input current, the measurement accuracy of current is
increased ±100μA and the measurement accuracy of power is increased
±100μA/range ×100%.

(3) Measurement of voltage, current and active power


P166681CJ7371124

Table 2.3 Measuring parameter setting


Measurement
RMS, MEAN, DC
model
Connection Single phase 2 wires system(1P2W), Single phase 3 wires system(1P3W),
mode Three-phase 3 wires system(3P3W), Three-phase 4 wires system(3P4W),
(just for PF330) 3 voltmeters 3 ammeters (3V3A)
Wave crest ratio 3 or 6
Range switch Manual or automatic
Range change up (The range will be automatic changed up when any of
the following cases should occur):
(1) The value of Urms or Irms is larger than 130% of the current
measuring range.
(2) When CF=3, the value of Upk or Ipk is more than 300% of the current
measuring range.
When CF= 6, the value of Upk or Ipk is more than 600% of the current
Automatic range
measuring range.
Range change down (The range will be automatic changed down when
any of the following cases should occur):
(1) The value of Urms or Irms is not more than 30% of the current
measuring range.
(2) The value of Urms or Irms is not more than 125% of the next
measuring range.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 13


PF300 series User’s Manual

(3) When CF = 3, the value of Upk or Ipk is less than 300% of the next
measuring range.
When CF = 6, the value of Upk or Ipk is less than 600% of the next
measuring range.
Synchronous
Take the whole interval of update cycle of voltage, current or data as
source of
synchronous source of measurement.
measurement
Line filter OFF, 500Hz or 5kHz
Peak value Maximum value of voltage, minimum value of voltage, maximum value of
measurement current, minimum value of current
Zero-location
Compensation for zero electric level
compensation

(4) Measurement of frequency

Table 2.4 The relationship between the measuring range of frequency and refresh rate of

the data

Measure object The voltage and current of each measuring unit


P166681CJ7371124

Method of measurement Counting numbers in a given time


Refresh rate of the data The measuring range of frequency
0.1s DC,25Hz≤f<100kHz
0.25s DC,10Hz≤f<100kHz
Range of measurement 0.5s DC, 5Hz≤f<100kHz
1s DC,2.5Hz≤f<100kHz
2s DC,1.5Hz≤f<100kHz
5s DC,0.5Hz≤f<100kHz
Accuracy of measurement ±0.06% rdg

Note-------------------------------------------------------------------------------------------------
● In order to ensure the accuracy of measurement, when CF = 3 (or 6,), the voltage or current

should be more than 30% (60%) of the range.

● Frequency filter: OFF, 5kHz or 500Hz.

14 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

(5) Operation (Only for PF330)


Table 2.5 Σ Operational formula(Only for PF330)
Single phase 3 Three-phase 3 Three-phase 4
3 voltmeters 3
wires wires wires
ammeters(3V3A)
system(1P3W) system(3P3W) system(3P4W)
UΣ[V] (U1+U3)/2 (U1+U2+U3)/3
IΣ[A] (I1+I3)/2 (I1+I2+I3)/3
PΣ[W] P1+P3 P1+P2+P3 P1+P3
3 3
SΣ[VA] Si=Ui×Ii S1+S3 (S1+S3) S1+S2+S3 (S1+S2+S3)
2 3

QΣ[var] Q i  Si 2  Pi 2 Q1+Q3 Q1+Q2+Q3 Q1+Q3

λΣ λi=Pi/Si PΣ/SΣ
ΦΣ[deg] Φi=cos-1(Pi/Si) cos-1(PΣ/SΣ)
i is the number of input unit.
S、Q、λ and Φ are calculated by the measuring value of voltage, current and active
Note power. Therefore, these values may be slightly different from the measuring results of
other measuring instruments based on different measuring principle when the distortion
P166681CJ7371124

signal is input.

Table 2.6 Other operation


The phase lead and lag of voltage and current can be measured below
following conditions:
• Sine wave
Phase lead and
• The measuring value is not less than 50% (100%) of the measuring range
lag
when the CF = 3 (or 6).
• Frequency: 20Hz ~ 2kHz
• Phase difference: ±( 5° ~ 175° )
The instrument provides the transformation ratio function to measure the high
voltage (>600V)and large current(>20A). The power coefficient and the
transformation ratio of VT (Voltage transformer) and CT (Current
transformer) should be set when the output of VT and CT of transformer
Transformation
input to the instrument.
ratio
• Default settings: V=1.000, C=1.000, F=1.000
• Effective bits: can be set automatically according to the effective bits of the
range of voltage and current.
• Setting range: 0.001 ~ 9999
Select two methods as followed:
• Arithmetic moving average(LIN)
Average • Exponential average(EXP)
Select attenuation constant of exponential average or constant of moving
average in 8, 16, 32 and 64.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 15


PF300 series User’s Manual

Efficiency
The efficiency is calculated with the measuring unit of power in PF330.
(just for PF330)
Calculate the wave crest ratio of voltage and current(peak/true effective
Wave crest ratio
value)

(6) Integral measurement


Table 2.7 Function of integral measurement
Integral model Manual integral, standard integral or continuous repeated integral
Time benchmark is provided by special real-time clock chip. The precision:
Time reference
±0.005%
Integral time 0 hour 00 minute 00 second to 9999 hours 59 minutes 59 seconds
Integral project Wh,Wh+,Wh-,Ah,Ah+,Ah-
The accuracy is increased ±0.1%rdg than the accuracy of power or
current(fixed range).
In the case of automatic range:
Integral precision
Do not perform the measurement at the time of switching range. And the first
measuring value after range switch is appended, as well as during the time of
no measurement.
Range setting Automatic range or fixed range
P166681CJ7371124

(7) Harmonic measurement

Table 2.8 Function of harmonic measurement


Method Phase-locked loop synchronization method(PLL)
The effective value of each harmonic component and relative value of
voltage, current and active power
Items The total harmonic voltage, the total harmonic current, the total harmonic
power, the total harmonic distortion of voltage, the total harmonic distortion
of current.
Frequency range The fundamental frequency of PLL source is between 20Hz and 2.56kHz.
When CF = 3(or 6), the input signal of voltage or current should be more
than 50% (or 100%)of the range.
PLL Source
The frequency filter of 500Hz is open when the fundamental frequency is not
more than 200Hz.
Data length of
2048
FFT
Window function Rectangular
Accuracy Increase ±0.2% rg than the normal

16 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Table 2.9 Relationship of fundamental frequency, sampling rate and the maximum

number of harmonic analysis


Fundamental the maximum number
Sampling rate (Hz) Width of the window
frequency f(Hz) of harmonic analysis
20Hz≤f<40Hz f×2048 1 50
40Hz≤f<80Hz f×1024 2 50
80Hz≤f<160Hz f×512 4 50
160Hz≤f<320Hz f×256 8 50
320Hz≤f<640Hz f×128 16 50
640Hz≤f<1280Hz f×64 32 32
1280Hz≤f<2560Hz f×32 64 16

(8) Display
Table 2.10 Display and refresh
Display window 4 windows
Display type 7 LEDs
Data update cycle 0.1s, 0.25s, 0.5s, 1s, 2s, 5s
Hold Hold the measuring value
P166681CJ7371124

Single
measurement Press the SINGLE key to update a display value when hold.
update
Peak hold Hold the peak of U, I, P, S, Q, Upk± and Ipk±

(9) Communication
Table 2.11 Communication settings
Communication USB for default, RS-232 、 RS-485 、 CAN 、 LAN or GP-IB ( optional )
interface (instrument of CAN add CAN communication port as standard)
RS-232 baud rate 9600, 19200, 38400, 57600, 115200
GP-IB address 0, 1, 2……31

2.2 General technical requirements

Table 2.12 General technical requirements


Preheating time ≥30 minutes
Operating Temperature: 5℃~40℃
environment Humidity: 20%R.H.~80%R.H.(without condensation)
Insulation Signal input end, shell and power supplying input end should be insulated
resistance from each other, with insulation resistance over 50MΩ.
AC2000V should last 1 minute between signal input end and the either of
Withstand voltage shell and power supply input end. DC2200V should last 1 minute between
shell and power supply input end.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 17


PF300 series User’s Manual

Power supply 100VAC~240VAC,50Hz/60Hz


Box size 248mm×147mm×384mm(W×H×D)
PF310: 5.3kg
Weight PF310A: 3.4kg
PF330: 6.6kg
P166681CJ7371124

18 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Chapter 3 Description of Instrument panel

3.1 Front Panel

The front panel of PF310 is shown in Figure 3.1.


P166681CJ7371124

Figure 3.1 The Front Panel of PF310

The front panel of PF330 is shown in Figure 3.2.

Figure 3.2 The Front Panel of PF330

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 19


PF300 series User’s Manual

The front panel of PF310A is shown in Figure 3.3.

Figure 3.3 The Front Pannel of PF310A


P166681CJ7371124

3.1.1 Indicative function on the panel

Table 3.1 Indicative function on the panel of PF310/PF330

Indicator Light Function Description


Indicator light of data refresh: the indicator light is flickering when data Iis
UPDATE refresh.ing
The Indicator light is off in the state of HOLD.
CHECK RANGE The indicator light is on when the value of voltage or current is more than
VOLTAGE
130% of the range.
CURRENT
MODE Measurement mode indicator of voltage and current measurement mode:
RMS RMS: true effective value
MEAN MEAN: correction value of rectification
DC DC: arithmetic mean value
Indicator light of peak hold: the light is on when the function of peak hold
PEAK HOLD
opens.
RANGE(window Indicator light of range switch: the light is on when the function of range
s of A and B) switch works.
MATH(window Indicator light of calculation: the light is on when display calculation function
of C) is selected in C widow.
THD(window of Indication light of total harmonic distortion: the light is on when total
A) harmonic distortion of voltage or current is selected in D window.
LINE Indicator light of line filter: the light is on when line filter function works.

20 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Indicator light of frequency filter: the light is on when frequency filter


FREQ
function works.
Indicator light of average function: the light is on when average function
AVG
works.
Indicator light of transformer ratio: the light is on when transformation ratio
SCALING
function works.
Indicator light of harmonic measurement: the light is on when harmonic
HARM
measurement function works.
Indicator light of measurement data storage: the indicator light flashes when
STORE
data is saved according to time interval preset.
LOCK Indicator light of key lock: the light is on when key lock function works.
Indicator light of remote mode: the light is on when the instrument is in
REMOTE
remote control mode.

Table 3.2 Indicative function on the panel of PF310A

Indicator Light Function Description


Indicator light of data refresh: the indicator light is flickering when data is
UPDATE
refresh.ing;The Indicator light is off in the state of HOLD.
Indicator light of range switch: the light is on when the function of range
P166681CJ7371124

RANGE
switch works.
Measurement mode indicator of voltage and current measurement mode:
M RMS
RMS: true effective value
M MEAN
MEAN: correction value of rectification
M DC
DC: arithmetic mean value
Indicator light of the maximum value hold: the light is on when the function
MAX HOLD
of peak hold works.
LINE Indicator light of line filter: the light is on when line filter function works.
Indicator light of frequency filter: the light is on when frequency filter
FREQ
function works.
Indicator light of average function: the light is on when average function
AVG
works
Indicator light of transformer ratio: the light is on when transformation ratio
SCALING
function is works.
Indicator light of measurement data storage: the indicator light flashes when
STORE
data is saved according to time interval preset.
The indicator light is on when the value of voltage is more than 130% of the
V OVER
range.
The indicator light is on when the value of current is more than 130% of the
A OVER
range.
Indicator light of calculation: the light is on when display calculation function
MATH
is selected in C widow.
Indication lamp of total harmonic distortion: the light is on when total
THD
harmonic distortion of voltage or current is selected in D window.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 21


PF300 series User’s Manual

3.1.2 Function of panel key

Table 3.3 Function of PF310/330 panel key

Classification Key Function Description


FUNCTION Select the parameters to display
Display
ELEMENT
switching Select the units to display: 1, 2, 3, Σ
(only for PF330)
Display menu of voltage range setting. The light
V RANGE indicator of ‘AUTO’ of V RANGE is on when the
automatic range of voltage range is selected.
Range and mode Display menu of current range setting. The light
of measurement A RANGE indicator of ‘AUTO’ of V RANGE is on when the
automatic range of current range is selected.
SHIFT+V RANGE
Switch measuring mode: RMS, MEAN, DC.
(MODE)
Wiring method Select connection mode of voltage or current
WIRING
(Only for PF330) terminal in rear panel.
P166681CJ7371124

To open or close peak hold function. The indicator


SHIFT+A RANGE
light of ‘PEAK HOLD’ is on when the function
(PEAK HOLD)
works,or the light is off .
Data hold switch. To hold the current data when the
Display
switch is on. And the indicator light of ‘UPDATE’ is
Maintenance HOLD
off. Press the key again to refresh data. And the
indicator light of ‘UPDATE’ flashes normally.
SHIFT+HOLD Press the SINGLE key to update display value when
(SINGLE) in hold.
▲、▼ Select parameters and value to set

SHIFT+▲(▶) To move current data bits in right circulation.


Common settings
SHIFT+▼(•) To move current decimal point in right circulation.
ENTER To enter the menu or confirm the parameters to set
Zero-location SHIFT+ENTER
Zero calibration function.
compensation (CAL)
START Start integration.
STOP Stop integration.
SHIFT+START Menu of integral settings: parameters such as integral
Integral
(INTEG SET) mode or integral time can be set.
SHIFT+STOP
To reset integral value and integral time.
(INTEG RESET)
Communication INTERFACE Display communication information and menu to set

22 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

interface communication interface.


Turn on or turn off the function of key lock. The
SHIFT+INTERFACE
Key lock indicator light is on when the function works. And
(LOCK)
press the key of LOCK again to close the function.
ESC Exit the menu
Set the parameters such as synchronous source, filter,
SETUP ratio, external sensor (optional), average, calculation
and time interval of data refresh.
Display system information (such as instrument
SHIFT+SETUP model and software version), initialization settings,
(UTILITY) wave crest ratio settings, measurement data storage
Others and turn on the function of range switch.,etc
SAVE/LOAD Save or load measurement configuration parameters.
Turn on or turn off the function of harmonic analysis
SHIFT+SAVE/LOAD (channel selection is included in PF330),
(HARM) synchronous source, the number of analysis,
calculation method.
SHIFT+SHIFT
Quit remote-control
(LOCAL)
P166681CJ7371124

Table 3.4 Function of PF310A panel key

Classification Key Function Description


FUNCTION
Display (No silk-screen, key Select the parameters to display, Note:converse
switching on the right of display choice through shift+ FUNCTION button
window)
Display menu of voltage range setting. The light
V RANGE indicator of V RANGE is on when the automatic
range of voltage range is selected..
Range and mode Display menu of current range setting. The light
of measurement A RANGE indicator of A RANGE is on when the automatic
range of current range is selected.
SHIFT+V RANGE
Switch measuring mode: RMS, MEAN, DC.
(MODE)
Open or close peak hold function. The indicator light
SHIFT+A RANGE
of ‘MAX HOLD’ is on when the function works or
(PEAK HOLD)
the light is off.
Display Data hold switch. To hold the current data when the
Maintenance switch is on. And the indicator light of ‘UPDATE’ is
HOLD
off. Press the key again to refresh data. And the
indicator light of ‘UPDATE’ flashes normally.
SHIFT+HOLD Each time press the SINGLE key to update a display

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 23


PF300 series User’s Manual

(SINGLE) value when in hold.


▲、▼ Select parameters and value to set
左右 Move current data bits in left or right circulation.
Common settings Move current decimal point in left or right
Shift+所有
circulation.
ENTER Enter the menu or confirm the parameters to set
Zero-location SHIFT+ENTER
Zero calibration function.
compensation (CAL)
START Start integration.
STOP Stop integration.
SHIFT+START Menu of integral settings: parameters such as integral
Integral
(INTEG SET) mode or integral time can be set.
SHIFT+STOP
Reset integral value and integral time.
(INTEG RESET)
Open or close the function of key lock. The indicator
SHIFT+INTERFACE
Key lock light is on when the function is open. And press the
(LOCK)
key of LOCK again to close the function.
Communication Display communication information and menu to set
INTERFACE
P166681CJ7371124

interface communication interface.


ESC Exit the menu
Set the parameters such as synchronous source, filter,
SETUP ratio, external sensor (optional), average, calculation
and time interval of data refresh.
Display system information (such as instrument
SHIFT+SETUP model and software version), initialization settings,
(UTILITY) wave crest ratio settings, measurement data storage
and open the function of range switch.
Others Open or close function of harmonic analysis
synchronous source, the number of analysis,
HARM
calculation method. The indicator of “HARM” key is
on when the function of harmonic analysis works.
SHIFT+HARM(SAVE
Save or load measurement configuration parameters.
/LOAD)

SHIFT+SHIFT(LOCA
Quit remote-control
L)

24 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

3.1.3 Digital and character display

Table 3.5 Comparison table of digital and character display

0→ A→  K→ U→ →


1→ B→ L→ V→ - →
2→ C→ M→ W→ (→
3→ D→ N→ X→ (→
4→ E→ O→ Y→
5→ F→ P→ Z→
6→ G→ Q→ c→
7→ H→ R→ h→
8→ I→ S→ i→
9→ J→ T→ u→

3.2 Rear Panel


P166681CJ7371124

The rear panels of PF310 and PF310A are shown in Figure 3.4 and Figure 3.5.

1 2 3 4 5 6

Figure 3.4 Rear Panel of PF310

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 25


PF300 series User’s Manual

Figure 3.5 Rear Panel of PF310A


P166681CJ7371124

Table 3.6 Rear Component of PF310/PF310A


Numb
Component Function description
ers
1 binding post of voltage input ± terminal is black and V terminal is red.
2 binding post of current input ± terminal is black and A terminal is red.
3 Insulating BNC interface Input terminal of external sensor of current.
To ensure safety, please make sure that the
4 Ground terminal
instrument is reliable grounding in use.
Input interface of instrument power, including 3A or
5 Power supply socket
2A fuse.
PF310: USB for default, RS232、RS485、LAN or
GP-IB(optional)(instrument of CAN add CAN
6 Communication interface communication port as standard)
PF310A: USB for default, RS232、RS-485、CAN、
LAN or GP-IB(optional)
Note: Simultaneously using current input terminal and insulating BNC interface is
forbidden.
The rear panel of PF330 is shown in Figure 3.6.

26 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual
P166681CJ7371124

Figure 3.6 Rear panel of PF330

Table 3.7 Rear Component of PF330


Numb
Component Function description
ers
1 Insulating BNC interface Input terminal of external sensor of current.
2 binding post of voltage input ± terminal is black and V terminal is red.
3 binding post of current input ± terminal is black and A terminal is red.
identifier of power Element 1 to 3 is used to distinguish 3 power
4
measurement unit measurement units.
5 Communication interface USB for default, RS232、RS-485、CAN
To ensure safety, please make sure that the
6 Ground terminal
instrument is reliable grounding in use.
Input interface of instrument power, including 3A
7 Power supply socket
fuse
8 GP-IB (optional) GP-IB,LAN or GP-IB,

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 27


PF300 series User’s Manual

Chapter 4 Measurement Condition Settings

4.1 Measurement Mode

There are 3 measurement modes to select in the measurement of voltage and current .

Table 4.1 Measurement Mode


Indicator
Mode Voltage Formula Current Formula
Light

true effective 1 1
f t  dt f t  dt
T T
 
2 2
RMS
value T 0 T 0

correction 1
π 1
π
f t  dt f t  dt
T T
value
rectification
of MEAN
2 2 T
 0
2 2 T
 0

1 1
f t  dt f t  dt
linear mean T T
DC
 
P166681CJ7371124

value T 0 T 0

Correction value of rectification: calibrate to the average value of effective value of

rectifier.

Linear mean value: the average value of voltage or current in a period.

Press the keys of SHIFT+V RANGE (MODE) to select measurement mode.

Switching sequence is shown in Figure 4.1.

Figure 4.1 Switch of Measurement Mode

Note---------------------------------------------------------------------------------------------

● Switch of measurement mode will affect the measured value such as voltage,

current and apparent power, except the measured value of active power.

● The measurement mode will switch to RMS mode automatically when the

function of harmonic measurement is activated.

● Users are not allowed to switch the measurement mode while the indicator light

“HOLD” is on.

28 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

● Before integration is activated, users can select any one measurement mode, once

the integration is activated, there is no permission to switch the measurement

mode.

4.2 Connection mode

There are different connection modes in available to measure voltage and current in

PF330. The relationships between connection mode and calculation results of

measurement are shown in table 2.5.


Table 4.2 Connection mode

Connection mode Measurement unit


Single phase 3 wires system
unit1, unit3 (unit2 is in 1P2W mode)
(1P3W)
Three-phase 3 wires system
unit1, unit3 (unit2 is in 1P2W mode)
(3P3W)
P166681CJ7371124

Three-phase 4 wires system


unit1, unit2, unit3
(3P4W)
3 voltmeters 3 ammeters
unit1, unit2, unit3
(3V3A)
If the external current transformer is connected,the equivalent is shown in Figure 4.2.

Figure 4.2 The equivalent of the external current transformer connection


(1) Connection mode of 1P2W

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 29


PF300 series User’s Manual

(a) the current is relatively larger (b) the voltage is relatively larger
P166681CJ7371124

(c) the external current transformer is connected


Figure 4.3 Connection mode of 1P2W

(2) Connection mode of 1P3W

(a)The internal current range is connected

30 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

(b)The external current transformer is connected

Figure 4.4 Connection mode of 1P3W

(3) Connection mode of 3P3W


P166681CJ7371124

(a)The internal current range is connected

(b)The external current transformer is connected

Figure 4.5 Connection mode of 3P3W

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 31


PF300 series User’s Manual

Connection instruction: Three are no neutral line in 3P3W system. Here the S is

regarded as virtual neutral N.

(4) Connection mode of 3P4W

(a)The internal current range is connected


P166681CJ7371124

(b)The external current transformer is connected

Figure 4.6 Connection mode of 3P4W

Connection instruction: 3P4W system includes neutral line, using the N line as the

reference terminal.

32 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

(5) Connection mode of 3V3A

(a)The internal current range is connected


P166681CJ7371124

(b)The external current transformer is connected

Figure 4.7 Connection mode of 3V3A

Note---------------------------------------------------------------------------------------------
● PF310/PF310A: Single phase 2 wires system only(1P2W)。
● PF330: please select a connection mode which is consistent with the
measurement circuit, otherwise it may lead to erroneous measurement results.
Unit 1, 2 and 3 is optional when the measuring signal is single phase 2 wires.
● The heavy lines above represent the current circuit and the thin lines indicate
voltage circuit.
● Generally, it is more simply and conveniently to connect current circuit before
connecting the voltage circuit.
-------------------------------------------------------------------------------------------------------

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 33


PF300 series User’s Manual

4.3 Measurement Range

4.3.1 Range of voltage and current


(1) Range of voltage (direct input):

When CF= 3: 15V, 30V, 60V, 150V, 300V, 600V.

When CF =6: 7.5V, 15V, 30V, 75V, 150V, 300V.

(2) Range of current (direct input):

1) PF310/PF310A:

When CF = 3: 5mA, 10mA, 20mA, 50mA, 100mA, 200mA,

0.5A, 1A, 2A, 5A, 10A, 20A.

When CF = 6: 2.5mA, 5mA, 10mA, 25mA, 50mA, 100mA,

0.25A, 0.5A, 1A, 2.5A, 5A, 10A.


P166681CJ7371124

2) PF330:

When CF = 3: 0.5A, 1A, 2A, 5A, 10A, 20A.

When CF= 6: 0.25A, 0.5A, 1A, 2.5A, 5A, 10A.

(3) Range of current (external sensor input):

When CF= 3: 2.5V, 5V, 10V.

When CF= 6: 1.25V, 2.5V, 5V.

The range can be set manually or automatically, it also can carry out the function of

automatical range jumping through the PC software to configure range.

4.3.2 Range Settings


The method of voltage range setting is shown in Figure 4.8.

Figure 4.8 Voltage Range Setting

34 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

The method of current range setting in PF310/PF310A and PF330 are shown in Figure

4.9and 4.10.
(Window D)
A RANGE ENTER

If CF=6 ,
A RANGE:10、5、2.5、1、0.5、0.25

Unit:A

If CF=6 ,
A RANGE:100、50、25、10、5、2.5
Unit:mA

If CF=6 ,
A RANGE:5、2.5、1.25
Unit:V
P166681CJ7371124

Figure 4.9 Current range setting

Figure 4.10 Current range setting

4.3.3 Condition for automatic range switch


(1) Range change up (The range will be automatically changed up when meeting

any of the following conditions):

1) The Value of Urms or Irms is more than 130% of the current

measurement range.

2) When CF = 3, the value of Upk or Ipk is more than 300% of the current

measurement range.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 35


PF300 series User’s Manual

When CF = 6, the value of Upk or Ipk is more than 600% of the current

measurement range.

(2) Range change down (The range will be automatically changed down when

meeting any of the following conditions):

1) The value of Urms or Irms is less than 30% of the current measurement

range.

2) The value of Urms or Irms isless than 125% of the next measurement

range.

3) When CF= 3, the value of Upk or Ipk is less than 300% of the next

measurement range.

When CF= 6, the value of Upk or Ipk is less than 600% of the next measurement

range.
P166681CJ7371124

4.3.4 Range Switch


According to the measurement range needed for the PC software configuration, user

can turn on or turn off the function of range switch though the key of instrument in all

direct input range provided by PF310/PF310A or PF330. The setting method is shown

in Figure 4.11.

Figure 4.11 Settings of Range switch

(OFF): close the function of range switch

 (ON): open the function of range switch

The indicator light of range in window A and B, as well as the indicator light of

AUTO will be on when the function of range switch is opened.

36 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

4.3.5 Range Check


The following indicator light will be on when the value of voltage or current is more

than 130% of the range in manual range measurement:

PF310/PF330:CHECK RANGE indicator light is “VOLTAGE” or “CURRENT”

PF310A: V OVER or A OVER

4.3.6 Power Range


The power range including active power, apparent power and reactive power depends
on the voltage range, the current range and connection mode, as table 4.3 showed.
Table 4.3 Range of power
Connection mode Range of power

1P2W (Single phase 2 wires system) Range of voltage× range of current

1P3W (Single phase 3 wires system)

3P3W (Three-phase 3 wires system) Range of voltage× range of current×2


P166681CJ7371124

3V3A (3 voltmeters 3 ammeters)

3P4W (Three-phase 4 wires system) Range of voltage× range of current×3

When CF= 3(or 6), the range of active power is shown in table 4.4(or 4.5). The
range of reactive power and apparent power is similar to active power.
Table 4.4 Range of active power in PF310/PF310A(CF=3)

Range of
Range of current
voltage

5.0000 mA 10.000 mA 20.000 mA 50.000 mA 100.00 mA 200.00 mA

15.000 V 75.000 mW 150.00 mW 300.00 mW 750.00 mW 1.5000 W 3.0000 W

30.000 V 150.00 mW 300.00 mW 600.00 mW 1.5000 W 3.0000 W 6.0000 W

60.000 V 300.00 mW 600.00 mW 1.2000 W 3.0000 W 6.0000 W 12.000 W

150.00 V 750.00 mW 1.5000 W 3.0000 W 7.5000 W 15.000 W 30.000 W

300.00 V 1.5000 W 3.0000 W 6.0000 W 15.000 W 30.000 W 60.000 W

600.00 V 3.0000 W 6.0000 W 12.000 W 30.000 W 60.000 W 120.00 W

Range of Range of current

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 37


PF300 series User’s Manual

voltage

500.00 mA 1.0000 A 2.0000 A 5.0000 A 10.000 A 20.000 A

15.000 V 7.5000 W 15.000 W 30.000 W 75.000 W 150.00 W 300.00 W

30.000 V 15.000 W 30.000 W 60.000 W 150.00 W 300.00 W 600.00 W

60.000 V 30.000 W 60.000 W 120.00 W 300.00 W 600.00 W 1.2000 kW

150.00 V 75.000 W 150.00 W 300.00 W 750.00 W 1.5000 kW 3.0000 kW

300.00 V 150.00 W 300.00 W 600.00 W 1.5000 kW 3.0000 kW 6.0000 kW

600.00 V 300.00 W 600.00 W 1.2000 kW 3.0000 kW 6.0000 kW 12.000 kW

Table 4.5 Range of active power in PF310/PF310A(CF=6)

Range of
Range of current
voltage

2.5000 mA 5.0000 mA 10.000 mA 25.000 mA 50.000 mA 100.00 mA


P166681CJ7371124

7.5000 V 18.750 mW 37.500 mW 75.000 mW 187.50 mW 375.00 mW 750.00 mW

15.000 V 37.500 mW 75.000 mW 150.00 mW 375.00 mW 750.00 mW 1.5000 W

30.000 V 75.000 mW 150.00 mW 300.00 mW 750.00 mW 1.5000 W 3.0000 W

75.000 V 187.50 mW 375.00 mW 750.00 mW 1.8750 W 3.7500 W 7.5000 W

150.00 V 375.00 mW 750.00 mW 1.5000 W 3.7500 W 7.5000 W 15.000 W

300.00 V 750.00 mW 1.5000 W 3.0000 W 7.5000 W 15.000 W 30.000 W

Range of
Range of current
voltage

250.00 mA 500.00 mA 1.0000 A 2.5000 A 5.0000 A 10.000 A

7.5000 V 1.8750 W 3.7500 W 7.5000 W 18.750 W 37.500 W 75.000 W

15.000 V 3.7500 W 7.5000 W 15.000 W 37.500 W 75.000 W 150.00 W

30.000 V 7.5000 W 15.000 W 30.000 W 75.000 W 150.00 W 300.00 W

75.000 V 18.750 W 37.500 W 75.000 W 187.50 W 375.00 W 750.00 W

150.00 V 37.500 W 75.000 W 150.00 W 375.00 W 750.00 W 1.5000 kW

38 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

300.00 V 75.000 W 150.00 W 300.00 W 750.00 W 1.5000 kW 3.0000 kW

Table 4.6 Range of active power(CF=3)

syste Range of
Range of current
m voltage

500.00 mA 1.0000 A 2.0000 A 5.0000 A 10.000 A 20.000 A

15.000 V 7.5000 W 15.000 W 30.000 W 75.000 W 150.00 W 300.00 W

30.000 V 15.000 W 30.000 W 60.000 W 150.00 W 300.00 W 600.00 W

1P2W 60.000 V 30.000 W 60.000 W 120.00 W 300.00 W 600.00 W 1.2000 kW

150.00 V 75.000 W 150.00 W 300.00 W 750.00 W 1.5000 kW 3.0000 kW

300.00 V 150.00 W 300.00 W 600.00 W 1.5000 kW 3.0000 kW 6.0000 kW

600.0 0V 300.00 W 600.00 W 1.2000 kW 3.0000 kW 6.0000 kW 12.000 kW

15.000 V 15.000 W 30.000 W 60.000 W 150.00 W 300.00 W 600.00 W


P166681CJ7371124

30.000 V 30.000 W 60.000 W 120.00 W 300.00 W 600.00 W 1.2000 kW


1P3W
60.000 V 60.000 W 120.00 W 240.00 W 600.00 W 1.2000 kW 2.4000 kW
3P3W
150.00 V 150.00 W 300.00 W 600.00 W 1.5000 kW 3.0000 kW 6.0000 kW
3V3A
300.00 V 300.00 W 600.00 W 1.2000 kW 3.0000 kW 6.0000 kW 12.000 kW

600.0 0V 600.00 W 1.2000 kW 2.4000 kW 6.0000 kW 12.000 kW 24.000 kW

15.000 V 22.500 W 45.000 W 90.000 W 225.00 W 450.00 W 900.00 W

30.000 V 45.000 W 90.000 W 180.00 W 450.00 W 900.00 W 1.8000 kW

60.000 V 90.000 W 180.00 W 360.00 W 900.00 W 1.8000 kW 3.6000 kW


3P4W
150.00 V 225.00 W 450.00 W 900.00 W 2.2500 kW 4.5000 kW 9.0000 kW

300.00 V 450.00 W 900.00 W 1.8000 kW 4.5000 kW 9.0000 kW 18.000 kW

600.0 0V 900.00 W 1.8000 kW 3.6000 kW 9.0000 kW 18.000 kW 36.000 kW

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 39


PF300 series User’s Manual

Table 4.7 Rang of active power(CF=6)

syste Range of
Range of current
m voltage

250.00 mA 500.00 mA 1.0000 A 2.5000 A 5.0000 A 10.000 A

7.5000 V 1.8750 W 3.7500 W 7.5000 W 18.750 W 37.500 W 75.000 W

15.000 V 3.7500 W 7.5000 W 15.000 W 37.500 W 75.000 W 150.00 W

1P2W 30.000 V 7.5000 W 15.000 W 30.000 W 75.000 W 150.00 W 300.00 W

75.000 V 18.750 W 37.500 W 75.000 W 187.50 W 375.00 W 750.00 W

150.00 V 37.500 W 75.000 W 150.00 W 375.00 W 750.00 W 1.5000 kW

300.00 V 75.000 W 150.00 W 300.00 W 750.00 W 1.5000 kW 3.0000 kW

7.5000 V 3.7500 W 7.5000 W 15.000 W 37.500 W 75.000 W 150.00 W

15.000 V 7.5000 W 15.000 W 30.000 W 75.000 W 150.00 W 300.00 kW


1P3W
P166681CJ7371124

30.000 V 15.000 W 30.000 W 60.000 W 150.00 W 300.00 kW 600.00 kW


3P3W
75.000 V 37.500 W 75.000 W 150.00 W 375.00 kW 750.00 kW 1.5000 kW
3V3A
150.00 V 75.000 W 150.00 W 300.00 W 750.00 kW 1.5000 kW 3.0000 kW

300.00 V 150.00 W 300.00 W 600.00 W 1.5000 kW 3.0000 kW 6.0000 kW

7.5000 V 5.6250 W 11.250 W 22.500 W 56.250 W 112.50 W 225.00 W

15.000 V 11.250 W 22.500 W 45.000 W 112.50 W 225.00 W 450.00 W

30.000 V 22.500 W 45.000 W 90.000 W 225.00 W 450.00 W 900.00 W


3P4W
75.000 V 56.250 W 112.50 W 225.00 W 562.50 W 1.1250 kW 2.2500 kW

150.00 V 112.50 W 225.00 W 450.00 W 1.1250 kW 2.2500 kW 4.5000 kW

300.00 V 225.00 W 450.00 W 900.00 W 2.2500 kW 4.5000 kW 9.0000 kW

Caution---------------------------------------------------------------------------------------------

● PF330: In a fixed range, the range of Element1, Element2 and Element3 are

consistent. In an automatic range, the range of Element1, Element2 and Element3

are switched independently according to the actual signal and the maximum range

of voltage or current is displayed in panel.

40 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

● In a fixed range, the measurement range will be not stable if the pulse signal input

is irregular. Please select the appropriate fixed range profile.

● In an automatic range, voltage and current can be adjusted independently

according to the condition of range lifting. Therefore, the same power values will

have different power ranges.

● After the range switch is activated, voltage (or current) will exit the function of

range switch if manual or automatic range is set. The original range opening

range switch will switch to automatic range if the function of range switch is

closed.

● If the voltage terminal is hanged, there might be a voltage result displayed ( <

0.3V) caused by the interference. The measured value will return to 0V if the

voltage terminals are short connected.

-------------------------------------------------------------------------------------------------------
P166681CJ7371124

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 41


PF300 series User’s Manual

4.4 External current sensor input

The maximum measurement range of current is 20A. User can input a voltage signal

of current sensor with external voltage-output type to the instrument insulating BNC

interface to test lager current which is calculated through transformer ratio.


Table 4.8 External current sensor transformer relationship

Transformer
Measuring Results after
ratio Input signal
items transformation
coefficient

Current E A A/E

Active power E W W/E

Reactive power E var var / E

Apparent power E VA VA / E
P166681CJ7371124

The default transformer ratio coefficient E is 10.00, where E=10.00 means the output

voltage will be 10 mV after 1A current pass through the current sensor. Please set the

proportion coefficient as 2.500 if voltage-output type current sensor with a ratio of

2.5mV/A is used.

(1) Setting Method of PF310/PF310A

Figure 4.12 Current sensor transformer ratio setting in PF310/PF310A

(2) Setting method of PF330: the transformer ratio coefficient of the current sensor of

each unit can be set synchronously (select ALL mode) or separately (select EACH

mode).

42 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Figure 4.13 Settings of current sensor transformer ratio setting in PF330

(RATIO): transformer ratio of external sensor

 (ALL): set all units

 (EL1): set measurement unit1

 (EL2): set measurement unit2

 (EL3): set measurement unit3

 .(E 10.00):current sensor transformer ratio coefficient

Note---------------------------------------------------------------------------------------------

The measurement accuracy is influenced by the characteristics of amplitude and


P166681CJ7371124

phase frequency when current sensor is connected to the insulating BNC

interface.

● Please ensure that the polarity has connected correctly.

-------------------------------------------------------------------------------------------------------

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 43


PF300 series User’s Manual

4.5 Mutual inductor transformer ratio

4.5.1 Transformer ratio settings


The ratio coefficient and power coefficient F of transformer VT or CT should be set

when the transformer output end is connected to the input end of the instrument. It

will be applied to voltage, current, active power, apparent power and reactive power.
Table 4.9 Transformer ratio relationship of voltage and current

Results after
Measuring items Input signal Note
transformation

Voltage U U×V

Current I I×C

Active power P P×V×C×F V:VT transfer ratio

Reactive power Q Q×V×C×F C:CT transfer ratio


P166681CJ7371124

Apparent power S S×V×C×F F:power coefficient

Voltage peak Upk Upk×V

Current peak Ipk Ipk×C

Transformer ratio coefficient settings of PF310/PF310A is shown in Figure 4.14.

Figure 4.14 Transformer ratio settings of PF310/PF310A

Transformer ratio settings of PF330 is shown in Figure 4.15, all unites can be set

synchronously (select ALL mode) or independently (select ELEMENT 1, ELEMENT

2, ELEMENT 3).

44 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Figure 4.15 Transformer ratio settings of PF330

(SCALE): settings of transformer ratio function

 (DATA): digital of ratio

 (ALL): set all units synchronously

 (EL1): set unit1

 (EL2): set unit2

 (EL3): set unit3


P166681CJ7371124

 .(V 1.000): voltage transformer ratio coefficient

 .(C 1.000): current transformer ratio coefficient

 .(F 1.000):power coefficient

4.5.2 Open or close function of transformer ratio

Figure 4.16 Open or close transformer ratio

(SCALE): set function of transformer ratio

 (OFF): close function of transformer ratio

 (ON): open function of transformer ratio

 (DATA): ratio coefficient

The indicator light of ‘SCALING’ will be on when the function of ratio is opened.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 45


PF300 series User’s Manual

4.5.3 Connection method of mutual inductor


Connection method of voltage and current mutual inductor is shown in Figure 4.17.

Figure 4.17 Connection method of voltage and current

4.6 Wave Crest Ratio Setting


The setting method of wave crest ratio is shown in Figure 4.18.
P166681CJ7371124

Figure 4.18 Wave crest ratio settings

(CF):wave crest ratio settings, select 3 or 6

Default setting:CF=3

The voltage range, current range, valid input range and measuring accuracy will be

changed with CF, see chapter 4.3 for more details.

Note---------------------------------------------------------------------------------------------

● In a fixed range, the measurement range will be set to the maximum range profile

after setting wave crest ratio.

● Voltage range, current range and valid input range will be different with wave

crest ratio.

-------------------------------------------------------------------------------------------------------

46 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

4.7 Synchronization source measurement


Synchronization source measurement is used to determine the signal sampling period.

The setting method is shown in Figure 4.19. To ensure the measurement accuracy,

please select the appropriate synchronization source.

Figure 4.19 Synchronization source settings

(SYNC): synchronization source measurement settings

(VOLT): voltage synchronization

(CURR): current synchronization


P166681CJ7371124

 (OFF): close the synchronization source and take total duration of data

refresh as sampling period.

Default setting:VOLT

Note---------------------------------------------------------------------------------------------

● Please select the signal with stable amplitude and frequency as a synchronization source.

And only when the synchronization source is detected accurately, the measurement accuracy

can be guaranteed.

● Voltage synchronization: Take signal cycle of voltage as synchronization signal to sample

the voltage and current. And take the total duration of data refresh as sampling period when

the cycle of voltage is not detected.

● Current synchronization: Take signal cycle of current as synchronization signal to sample

the voltage and current. And take the total duration of data refresh as sampling period when

the cycle of current is not detected.

● Close synchronization source: take the total duration of data refresh as sampling cycle to

sample.

-------------------------------------------------------------------------------------------------------

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 47


PF300 series User’s Manual

4.8 Filter

Please set the cutoff frequency of the low pass filter first and then open or close the

filter when the line filter and frequency filter is set, as shown in Figure 4.20.

Figure 4.20 Filter settings


P166681CJ7371124

(FILTER): filter settings

. (C.FRE): cutoff frequency

. (L.FILE): line filter

. (F.FILE): frequency filter

 (5k): 5kHz cutoff frequency

 (500): 500Hz cutoff frequency

 (OFF): close filter

 (ON): open filter

Note---------------------------------------------------------------------------------------------

● OFF, 500Hz and 5kHz are available in low pass filter.

● Line filter and frequency filter can be opened or closed independently.

-------------------------------------------------------------------------------------------------------

48 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

4.9 Data refresh rate

The data refresh rate of instrument refers to the time cycle of signal sampling,

processing and finally displayed on the interface. There are 6 refresh rates for the user

to choose in the instrument as followed: 0.1s, 0.25s, 0.5s, 1s, 2s and 5s. The data

refreshing is synchronous with “Update” indicator flashing. Setting method is shown

in Figure 4.21.
P166681CJ7371124

Figure 4.21 Display refresh rate settings

.(U.RATE):data refresh rate, 0.1s, 0.25s, 0.5s, 1s, 2s, 5s are optional.

The frequency range of measured signal with different data refresh rate is shown in

table 2.4.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 49


PF300 series User’s Manual

4.10 Average

It is available to use average function when the measured signal amplitude is smaller,

the fluctuation is larger or the frequency is too low to read .

4.10.1 Average method


There are two kinds of average method in the instrument as followed: Arithmetic

moving average (LIN) and Exponential average(EXP).

(1) Arithmetic moving average:


M(n - m  1)    M (n - 1)  Mn
D(n) 
m
Where, D(n): the nth value displayed. That is when n = 1, D(1) = M(a).

M(n): the nth measuring value

m: average constant
P166681CJ7371124

(2) Exponential average:


M (n) - D(n - 1)
D(n)  D(n - 1) 
m
Where, D(n): the nth value displayed

M(n): the nth measuring value

m: attenuation constant

4.10.2 Average function setting


The setting method is shown in Figure 4.22.

Figure 4.22 Average function setting

 (AVG): set average function

50 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

 (OFF): close average function

 (ON): open average function

(TYPE): selection of average function type

 (LIN): liner average(Arithmetic moving average)

 (EXP): Exponential average

8, 16, 32, 64 are optional for average constant of arithmetic moving average or

attenuation constant of exponential average.

Note---------------------------------------------------------------------------------------------

● The average function is only valid for routine measurement parameters. The

average function will be closed automatically if integral measurement or

harmonic analysis activated.

-------------------------------------------------------------------------------------------------------
P166681CJ7371124

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 51


PF300 series User’s Manual

4.11 Data Hold

4.11.1 Data Hold


Press the key of HOLD to enter or exit the state of HOLD.

When the indicator light is on, the instrument will stop sampling and refreshing data,

but the FUNCTION key below each display window is still valid in the state of data

hold.

4.11.2 Single Measurement


Press key of SINGLE (SHIFT + HOLD/ESC) to perform the function of single

measurement in state of data hold. Each press the key of SINGLE the data refresh for

once. And stay into the state of data hold again.

4.11.3 Peak Hold


P166681CJ7371124

In normal measurement mode, press the key of PEAK HOLD (SHIFT+A RANGE)

to enter or exit the state of peak hold. The holding parameters are as followed:

Voltage (rms, mn, dc), current (rms, mn, dc), the maximum value of voltage, The

maximum value of voltage, active power, reactive power and apparent power.

The indicator light (PEAK HOLD or MAX HOLD) is on when being in the state of

peak hold.

Only the peak value of conventional data will be held in the state of harmonic analysis,

while the peak value of harmonic measurement will not.

52 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Chapter 5 Measurement Function

5.1 Normal Measurement Function

5.1.1 Normal Measurement parameters

The measurement parameters calculated in normal measurement state is shown in

Table 5.1.

Table 5.1 Measurement parameters in normal measurement state


parameter symbol Indicator light Display window
true effective value Urms V A、B、C、D
correction value of Umn
Voltage rectification Udc
arithmetic mean
value
P166681CJ7371124

true effective value Irms A


correction value of Imn
Current rectification Idc
arithmetic mean
value
Active power P W
Reactive power Q var A
Power Apparent power S VA
Power factor λ PF B、D
Phase angle Φ deg B
Voltage frequency fU V Hz D
Frequency
Current frequency fI A Hz
The maximum Upk+ V pk+ C
value of voltage
The minimum Upk- V pk-
value of current
Peak
The maximum Ipk+ A pk+
value of current
The minimum Ipk- A pk-
value of current
Wave crest ratio of CFU MATH
Wave crest
voltage
ratio
Wave crest ratio of CFI MATH

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 53


PF300 series User’s Manual

current
Efficiency η MATH
(Only for ——
PF330)

5.1.2 Display window

There are 4 display windows in this instrument: A, B, C and D, which can display

four same or different measurement results simultaneously. The content of display

windows can be selected by the corresponding FUNCTION key.

There are 2 keys of FUNCTION for parameters selection with bidirectional

circulation in PF310.

There is 1 FUNCTION key for parameters selection with circulation in

PF330/PF310A, while the key of SHIFT+FUNCTION is for reverse cycle, and the

key of ENTER is used for measurement unit selection.


P166681CJ7371124

V: voltage;

A: current;

W: active power;

VA: apparent power;

var: reactive power;

V: voltage;

A: current;

W: active power;

PF: power factor;

deg: phase angle;

54 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

V: voltage;

A: current;

W: active power;

V pk+: the maximum value of voltage;

V pk-: the minimum value of voltage;

A pk+: the maximum value of current;

A pk-: the minimum value of current;

MATH: calculation(wave crest ratio of voltage, wave crest ratio or

efficiency of current)
P166681CJ7371124

V: voltage;

A: current;

W: active power;

PF: power factor;

V Hz: voltage frequency;

A Hz: current frequency.

ELEMENT:circulating selection of measurement unit is as followed.

Figure 5.1 Selection of measurement unit

5.1.3 Using Method

(1) Select the measurement unit to display: use the key of ELEMENT to

switch.

(2) Select the appropriate measurement mode: RMS, MEAN OR DC.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 55


PF300 series User’s Manual

(3) Select the appropriate connection mode: using the key of WIRING to

switch.

(4) Select the measurement parameters to display: using the key of

FUNCTION to switch.

(5) Select appropriate wave crest ratio: 3 or 6.

(6) Select appropriate range of voltage and current: manual or automatic.

(7) Select appropriate measurement synchronization source: Volt, Curr or OFF;

(8) Select the required filter: OFF, ON (5kHz or 500Hz);

(9) Select the required function of average: OFF, ON (8, 16, 32, 64);

(10) Select the required function of transformer ratio: OFF, ON (V, C or F);

(11) Select the function of HOLD, SINGLE and PEAK HOLD as needed.

Note:

It shows ‘’ when the range of voltage or current is overflowed.


P166681CJ7371124

● It shows ‘’ of power factor, phase angle , the maximum or

minimum value of voltage and current and wave crest ratio if the measuring mode

is DC mode.

● The voltage(current) will show 0 when the voltage(current) signal is less than

0.5% of range (when CF = 6, the value is less than 1% of range).

● It will show 0 of active power P(W), reactive power Q(var) or apparent power

S(VA) and ‘’ of power factor, phase angle, wave crest ratio of

voltage or current and the maximum or minimum of voltage or current if the

measuring value of voltage or current is 0.

● It shows ‘Error’ of frequency of voltage or current and ‘’ of power

factor or phase angle when the AC component of voltage or current is less than

3% range.

● It shows ‘’ of phase angle when the voltage or current is

overflowed, range switching or frequency exceeds the measuring range.

● The absolute value of λ(PF) calculated may be lager than 1 in the special

56 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

situations such as average function (AVG) in the state of PEAK HOLD or the

mode of MEAN or DC. It will show ﹢1.000 or ﹣1.000 of λ(PF) at this time.

● The phase angel is obtained by the calculated power factor: Φ = arccos(PF). The

result is the angle between 0° and 180°, which has additional error compared with

the calculate value from the power factor showed.

● D means that the phase of current leads the one of voltage. G means the phase of

current is lag behind the one of voltage. The accuracy of phase angle can be

ensured under the condition that the voltage and current are steady sine wave.

● The sampling processes of voltage and current are synchronized with the input

signals,and it can measure frequency synchronously. In order to ensure the

measuring accuracy of frequency, voltage and current, App or Vpp of the signal

should be over than 30% of the range.

When the frequency is less than 0.5Hz, window will show ‘’.
P166681CJ7371124

When the frequency is more than 100kHz, window will show ‘’.

-------------------------------------------------------------------------------------------------------

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 57


PF300 series User’s Manual

5.2 Function of Integral Measurement

The integration means the accumulation of active power(W) or current(A) according

to the time to obtain the consumption of power or energy. During the intergration, the

positive and negative parts are processed respectively at the same time. The integral

results are showed by positive or negative values. Besides the integration time and

integral value, numbers of conventional measured value can be displayed at the same

time.

5.2.1 Integral Mode


There are several integral modes as follows:

Table 5.2 Integral Mode

Integral Mode Start Stop Integral Time


P166681CJ7371124

The time from pressing the key

Manual Integral Key of START Key of STOP of START to pressing the key of

STOP.

The end of
Standard Integral Key of START The set Integral time
Integral time

Continuous Repeat according to the period


Key of START Key of STOP
repeated integral of integral time

Note: Press the key of RESET(SHIFT+STOP)to exit the integral mode.

Table 5.3 Indicator light of integral mode

Indicator Light Indicator Light Applicable


State description
of START of STOP integral mode

Not the integral mode(or exited All integral


off off
already) modes

Manual integral
on off Ongoing of integral
mode

on flicker Ongoing of integral, not to the Standard

58 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

ending time integral

Ongoing of continuous repeated Continuous


flicker flicker
integral. repeated integral

Finish integral correlative or


All integral
off on press the key of STOP to stop
mode
compulsively

5.2.2 Measurement Parameters of Integration

Table 5.4 Definition of integral parameters

Integral term Power integral Current integral

Measurement Mode RMS / MEAN / DC RMS / MEAN DC

n N n
Formula Definition  ui  ii
i 1
I 1
II i
i 1
i
P166681CJ7371124

Ui and Ii stand for instantaneous voltage and current respectively. n is

Note refreshed number.

II is refresh data each time. N is the times of data refreshing.

5.2.3 Display in windows

V: voltage;

A: current;

W: active power;

VA: apparent power;

var: reactive power;

TIME: Accumulated integral time.

V: voltage;

A: current;

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 59


PF300 series User’s Manual

W: active power;

PF: power factor;

deg: phase angle.

V: voltage;

A: current;

W: active power;

V pk+: the maximum value of voltage;

V pk-: the minimum value of voltage;

A pk+: the maximum value of current;

A pk-: the minimum value of current;

Wh:total power integral;


P166681CJ7371124

Wh+: positive power integral;

Wh-: negative power integral;

Ah: integral of total electric quantity;

Ah+: integral of positive electric quantity;

Ah-: integral of positive electric quantity;

MATH: calculation (wave crest ratio of voltage, wave crest ratio or

efficiency of current).

V: voltage;

A: current;

W: active power;

PF: power factor;

V Hz: frequency of voltage;

A Hz: frequency of current.

60 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

5.2.4 Using Method

(1) Basic Parameter Settings

Before activating the integral function, these parameters should be set firstly:

measurement mode, connection mode, display refresh rate, synchronous source, wave

crest ratio, measurement range, transformer ratio and filter. These parameters can not

be changed in the process of integral.

(2) Integral mode setting

(MANUAL): Manual integral;

 (NOR): standard integral;


P166681CJ7371124

 (CONT): continuous repeated integral.

(3) Integral time settings

(H): hour;

(M):minute;

(S): second.

(4) Power unit selection

Use the key of FUNCTION in window C to switch unit.

(5) Integral operation such as START, STOP, INTEG RST

1) Press key of START to start integral in normal measurement. And the

indicator light of START is on at this time. The integral time is showed in

window A and the result is showed in window C.

2) Press key of START to exit the state of HOLD or PEAK HOLD and start

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 61


PF300 series User’s Manual

integral in the state of HOLD or PEAK HOLD of normal measurement.

3) Press the key of STOP to stop integral and press the key of START to

continue.

4) The indicator light of STOP will be on after the integral. Then pressing the

key of STRAT is invalid.

5) Pressing the key of INTEG RST is invalid in the state of START.

6) Press the key of INTEG RST to reset the integral, clear the accumulated

integral time and exit the integral state in the state of STOP.
Caution---------------------------------------------------------------------------------------------
Table 5.5 Limit table of integral state settings

Integral state INTEG RST State START State STOP State

Setting

Measurement Mode Yes No No


P166681CJ7371124

Connection Mode Yes No No

Refresh Rate of Display Yes No No

Synchronous Source Yes No No

Wave Crest Ratio Yes No No

Measurement Range Yes No1 No1

Transformer Ratio Yes No No

Function

Filter Yes No No

Average Function Yes No2 No2

Peak Hold Yes No No

Hold Yes No No

Single Measurement Yes No No

Uint selection Yes Yes Yes

Integral Mode Yes No No

Integral Time Yes No No

62 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

START of integral Yes No Yes

STOP of integral No Yes No

RESET of integral No No Yes

Open or close harmonic Yes Yes Yes

PLL Source Yes No No

the highest number of Yes No No

harmonic analysis

Calculation method of Yes No No

harmonic

Storage of data Yes Yes Yes

Zero calibration Yes No No

● Note 1: When the automatic range or range switch is activated, the automatic
P166681CJ7371124

range or range switch is still effective but manual setting is forbidden when the

function of integral is activated.

● Note 2: The average function is closed automatically when the integral function is

opened, which is still closed after exiting the integration.

● There will show the symbol of ‘’ at the left side of the window C when it

shows WH- or AH-.

● Standard integral and continuous repeated integral can not be started when the

integral time is set as 0000:00:00.

● It will switch to voltage automatically if the parameter of TIME is original

displayed in window A and it will be switched to active power if any parameter of

Wh, Wh+, Wh-, Ah, Ah+ and Ah-is original displayed in window C when the

integral function is exited.

● The integral function will be closed after the instrument is reactivated.

-------------------------------------------------------------------------------------------------------

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 63


PF300 series User’s Manual

5.3 Function of harmonic measurement

5.3.1 Harmonic measurement parameters

Harmonic components of voltage and current are mainly analyzed in harmonic

measurement. The related parameters and some normal parameters can be measured

in the harmonic measurement state at the same time, as shown in table 5.6.

Table 5.6 Display content of each window in harmonic measurement state


parameter symbol Indicator light Display window
Total harmonic voltage ΣU V A、D
Total harmonic current ΣI A A、D
Total harmonic power ΣP W A、D
Harmonic order k Or.(LED display) A
True effective value of
U(k) V
single harmonic voltage
True effective value of
P166681CJ7371124

I(k) A
single harmonic current
Har
Active power of single
mo P(k) W
harmonic
nic
Relative value of single B
mea U(k)% V%
harmonic voltage
sure
Relative value of single
me I(k)% A%
harmonic current
nt
Relative value of single
par P(k)% W%
harmonic active power
ame
Phase angle Φ deg
ters
Fundamental frequency of
fU V Hz
voltage
Fundamental frequency of
fI A Hz
current
D
Total harmonic voltage
Uthd THD V %
distortion
Total harmonic current
Ithd THD A %
distortion
Par True effective value of
Urms V
ame voltage
ters True effective value of C
Irms A
of voltage
nor Active power P W

64 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

mal The maximum value of


Upk+ V pk
and voltage
inte The minimum value of
Upk- V pk
gral voltage
mea The maximum value of
Ipk+ A pk
sure current
me The minimum value of
Ipk- A pk
nts current
Total power
Wh Wh
integral
Positive power
Wh+ W h+
integral
Negative
Wh- W h-
power integral
Total electric
quantity Ah Ah
Integral
integral
function
Positive
P166681CJ7371124

electric
Ah+ A h+
quantity
integral
Negative
electric
Ah- A h-
quantity
integral
Wave crest
CFU MATH
ratio of voltage
Wave crest
MATH CFI MATH
ratio of current
efficiency
(Only for η MATH
PF330)
The measurement function fully meets the requirements of IEC61000-3-2 standard in

harmonic measurement state. The normal measurement and integral measurement can

be realized at the same time in harmonic measurement function.

(1) Two calculation methods of total harmonic distortion and each harmonic

distortion

IEC method():

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 65


PF300 series User’s Manual

CSA method():
max max

 Uk   Uk 
2 2

k 2 k 2
Uthd  100 Ithd  100
U I
U k  Ik 
Uhdf  100 Ihdf  100
U I
Where Uthd or Ithd: the total harmonic distortion of voltage or current

U(1)、I(1): the effective value of fundamental

frequency (the first harmonic) of voltage、

current;

U(k)、I(k): the effective value of kth harmonic of

voltage、current;
P166681CJ7371124

Uhdf 或 Ihdf: the relative value of kth harmonic of

voltage、current;

k: harmonic order;

max: the maximum harmonic order set by user.

(2) Calculation method of total harmonic voltage, current and power

max
Total harmonic voltage: U  Uk
2

k1

max

I  Ik
2
Total harmonic current:
k 1

max
Total harmonic power: P  Pk
k1

In the formula, max: the maximum harmonic order set by user;

k: harmonic order;

U(K): The effective value of single harmonic voltage;

I(k): The effective value of single harmonic current;

P(k): Active power of single harmonic.

66 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Note--------------------------------------------------------------------------------------------

● .The counting process has removed the Direct Current component and the value

over maximum harmonic in total harmonic voltage, current and power calculation.

Thus the calculated values may be different from the values in normal

measurement.

● In order to facilitate users to do a comparison analysis, harmonic analysis and

normal measurement parameters are measured at the same time in this

instrument.

-------------------------------------------------------------------------------------------------------

5.3.2 Display
There are 4 display windows in this instrument: A, B, C and D. Four same or different
P166681CJ7371124

measurement values can be displayed at the same time. The content of display

window can be selected in circulation by the corresponding FUNCTION and

ELEMENT keys.

or.01:harmonic order, 1~n(n is the maximum order to be set);

V: total harmonic voltage;

A: total harmonic current;

W: total harmonic active power.

V: True effective value of single harmonic voltage (corresponding the

order shown in window A);

A: True effective value of single harmonic current (corresponding the

order shown in window A);

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 67


PF300 series User’s Manual

W: Active power of single harmonic (corresponding the order shown in

window A);

V %: Relative value of single harmonic voltage (corresponding the order

shown in window A);

A %: Relative value of single harmonic current (corresponding the order

shown in window A);

W %:Relative value of single harmonic active power (corresponding the

order shown in window A);

deg: phase difference between fundamental current and voltage.

Only the normal measurement parameters are shown in window C in harmonic


P166681CJ7371124

analysis.

V: the true effective value of voltage;

A: the true effective value of current;

W: active power;

V pk+: the maximum value of voltage;

V pk-: the minimum value of voltage;

A pk+: the maximum value of current;

A pk-: the minimum value of current;

MATH: calculation (wave crest ratio or efficiency).

V: total harmonic voltage;

A: total harmonic current;

W: total harmonic active power;

PF: power factor of fundamental wave;

V Hz: fundamental frequency of voltage;

68 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

A Hz: fundamental frequency of current;

THD V %: total harmonic distortion of voltage;

THD A %: total harmonic distortion of current;

Note: the indicator light of THD is on the left side of window D.

When the harmonic and integral is measured simultaneously, each window is

shown as follows:
P166681CJ7371124

5.3.3 Using method

(1) Basic Parameter Setting

Before activating the integral function, the parameters should be set as follows:

display refresh rate, wave crest ratio, filter.

(2) Select synchronous source of harmonic measurement: voltage or current.

.(PLL.SRC): synchronous source;

 (VOLT): voltage synchronization;

 (CURR): current synchronization.

(3) Select the maximum harmonic order: 1 to 50

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 69


PF300 series User’s Manual

ORDER: the maximum harmonic order,1~50.

(4) Select the calculation method of total harmonic distortion: IEC, CSA

(THD): harmonic calculation method;

(IEC): use IEC method;


P166681CJ7371124

(CSA): use CSA method.

(5) Activate harmonic measurement:

PF310/PF310A:select OFF or ON to open or close harmonic measurement.

PF330:OFF, Element1, Elemen2, Element3 or ALL.

(ELEM): selection of unit;

 (OFF): close the harmonic function;

 (EL1): unit 1;

 (EL2): unit 2;

70 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

 (EL3): unit 3;

 (ALL): all units.

(6) Use the FUNCTION key in each window to display the results of analysis.

(7) Use HOLD or SINGLE function as needed.

(8) Close the function of harmonic measurement.

Note---------------------------------------------------------------------------------------------

● In normal measurement state, RMS mode will be set automatically once enter

into the state of harmonic analysis whichever the measurement mode of voltage

or current is.

● Harmonic analysis of PF330 is only for Element1, Element2 and Element3.

● The selection of the maximum number of harmonic analysis will affect the

calculation results of total harmonic voltage, total harmonic current, total

harmonic power and the total harmonic distortion of voltage or current.


P166681CJ7371124

● The range of fundamental frequency is 20Hz to 2.56kHz in harmonic analysis. If

the frequency detected is not in this range or the signal is too small to detect, the

indicator light of UPDATE will be on and it will show  in window

D (‘' for small signal) and show ‘’ for each

measurement parameters. The harmonic analysis will start again if the state of

harmonic analysis is exited or the frequency detected is in the requreid range.

● The measured value will not be accurate if the fundamental frequency of PLL

synchronous source can not be measured accurately because of signal

interference or noise. Taking voltage as PLL synchronous source is suggested in

general.

● The synchronization of PLL source may be failed when the signal of PLL

synchronous source is less than 50% of range (CF=3) or 100% of range (CF=6).

● 500Hz frequency filter is suggested when the fundamental frequency is less than

200Hz and include high frequency components.

● The relationship of fundamental frequency, sampling rate and the maximum

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 71


PF300 series User’s Manual

number of harmonic analysis is shown in table 2.9. The excess part of single

harmonic parameters will show '' if the maximum number of

harmonic analysis is over the actual analysis number.

-------------------------------------------------------------------------------------------------------
P166681CJ7371124

72 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

5.4 Storage function

5.4.1 Measurement data storage


The measurement data can be recorded and saved to the internal memory in this

instrument. And the stored data can be read by the PC software. The stored data items:

data of normal measurement, data of integral measurement and data of harmonic

measurement. The measured data is saved as a set of single update data. 100 sets of

data can be saved in PF310/PF310A and PF330.

(1) Set the interval of data storage

Set the interval of data storage. The measured data can be saved according to the time

interval after the function of data storage is opened.


P166681CJ7371124

Figure 5.2 Data storage interval setting

(INTVL):Data storage interval

(2) Open or close the function of data storage

Figure 5.3 Open or close the function of data storage

(OFF): close the function of measured data storage. The indicator light of

STORE is off after the function is closed.

 (ON): open the function of measured data storage. The indicator light of

STORE is on after the function is opened.

Stop saving the measured data if it meets any of the following conditions:

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 73


PF300 series User’s Manual

1) The 100 groups is full of storage space in storage.

2) Close the function of data storage.


Note---------------------------------------------------------------------------------------------
● The saved data can not be read by the local end of the instrument. The function of
data storage will be closed automatically after restart of the instrument.
● If the time interval is set as 00∶00∶00 and the function of data storage is
opened, each refreshing data will be saved in this instrument.
● The original data will be overwritten if the function of data storage is restarted.
● The stored data will be cleared if the configuration parameters is reloaded or the
instrument is turned off or initialized.
● The measurement conditions can not be set in the process of data storage except
for the switch of measurement range.
● Press the key of HOLD to hold the measuring data and stop data storage in the
process of data storage.
● To store the data in the state of PEAK HOLD, peak value is shown and stored.
P166681CJ7371124

The function of peak hold can not be opened or closed in the process of data
storage.
-------------------------------------------------------------------------------------------------------

5.4.2 Measurement configuration storage

Four sets (File1 to File 4) of measurement configuration parameters can be pre-saved

according to users’ different measurement requirements. The measurement

configuration can be called quickly.

The measurement configuration parameters to save or load include: measurement

mode, measurement range, measurement synchronous source, transformer ratio

coefficient, average setting, filter setting, peak hold setting, calculation setting, data

refresh rate, wave crest ratio, integral measurement setting, harmonic measurement

setting and data storage setting.

(1) Save measurement configuration

74 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Figure 5.4 Save measurement configuration

The document status will be shown in window D , which is selected in window C.


(SAVED): saved
 (FREE): free
(2) Load measurement configuration

Figure 5.5 Load measurement configuration

Load the measurement configuration in the saved documents from FILE1 to FILE 4.
P166681CJ7371124

Then measure the parameters.


Note---------------------------------------------------------------------------------------------
● The original data will be overwritten if the configuration parameters are saved in
saved files (File1 to File 4).
-------------------------------------------------------------------------------------------------------

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 75


PF300 series User’s Manual

Chapter 6 Other Functions

6.1 View the system information

(MODE):instrument type, the specific type is shown in window D.

 (VER): Version, the specific version is shown in window D.

6.2 Initialization settings


P166681CJ7371124

 (NO): no initialization;

 (YES):initialization.

The measurement configuration is restored to the factory default value after

initialization, as shown in table 6.1.


Table 6.1. Initialize default configuration
Item default
display:V ( ELEMENT1 as the default in
Window A ——
PF330)
display:A (ELEMENT1 as the default in
Window B ——
PF330)
display:W ( ELEMENT1 as the default in
Window C ——
PF330)
display:V Hz ( ELEMENT1 as the
Window D ——
default in PF330)
Measurement
MODE RMS
mode
Connection WIRING 1P3W

76 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

mode
( Only for
PF330)
Data refresh ratio Update 0.25s
Synchronous
source of SYN Volt
measurement
Calculation MATH CFU
Hold HOLD Function of data hold: OFF
Peak hold PEAK HOLD Function of peak hold: OFF
Wave crest ratio CF 3
Measurement
RANGE Auto
range
Line filter LINE OFF
Frequency filter FREQ OFF
Average function: OFF
Average function AVG Average type: Arithmetic moving average
Average constant: 8
P166681CJ7371124

Function of
SKIPE OFF
range switch
Ratio function:OFF
Transformer ratio SCALING
V:1.000 C:1.000 F:1.000
Ratio of external
sensor E 10.00
(optional)
Integral Integral function: off
measurement INTEG Integral mode: manual
settings Integral time: 0000∶00∶00
Harmonic analysis function: OFF
Harmonic PLL source: Volt
measurement HARM Calculation method of total harmonic
setting distortion: IEC
The highest number of analysis: 50
Function of data storage: OFF
Data storage STORE
Data storage interval: 00∶00∶00
Communication
INTERFACE USB
interface

6.3 Zero-location Compensation

Press CAL (SHIFT+ENTER) to execute the function of zero-location compensation

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 77


PF300 series User’s Manual

(zero calibration). The zero position is calibrated automatically.

The zero-location compensation is executed automatically when the measurement

range and input filter is changed.

Note---------------------------------------------------------------------------------------------

● To ensure the measurement accuracy, zero compensation manually is suggested at

30 minutes after start.

● Zero compensation manually is suggested when the zero position is changed with

environmental temperature and stay in the same measurement range for a long

time.

-------------------------------------------------------------------------------------------------------

6.4 Function of key lock

Press LOCK (SHIFT+INTERFACE) to execute the function of key lock. As a result,


P166681CJ7371124

the key is in the state of invalid. Press the LOCK key again or reboot to release the

lock state.

6.5 Function of calculation

Use the key of FUNCTION of window C to switch to MATH function and the

indicator light at left side of window C will be on.

(1) PF310/PF310A:CFU or CFI.

(2) PF330:CFU, CFI or EFFI.

78 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

 (CFU): wave crest ratio of voltage;

 (CFI): wave crest ratio of current;

(EFFI): efficiency.

6.5.1 Calculation of wave crest ratio


Use the calculation function of wave crest ratio to calculate the wave crest ratio of

voltage or current in each unit. The formula is as follows:


P166681CJ7371124

Xpk
Wave crest ratio : CF 
Xrms
Where, Xpk:peak

Xrms:true effective value

Note:Xpk and Xrms are the same as the voltage or current.

6.5.2 Efficiency calculation(Only for PF330)


The input active power is measured in unit 2 and the output active power is measured

in unit 1 and unit 3. The formula is given by:

P1  P3
Efficiency    100%
P2

P2:input active power

P1, P3:output active power

Note---------------------------------------------------------------------------------------------

● The efficiency unit is %,’%’ is not shown in the panel.

● The efficiency value is shown as ‘’ when η<0.0001% or P2=0.

-------------------------------------------------------------------------------------------------------

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 79


PF300 series User’s Manual

Chapter 7 Software Instruction

7.1 System requirements

(1) The operation system must be Windows XP or Win7 (Chinese or English);

(2) More than 1G dis space.

(3) The display resolution is 1280x1024 or above. 1280x1024 is advised.

(4) One serial port of USB.

7.2 Software installation

(1) Installation in .net4.0

dotNetFx40_Full_x86_x64.exe should be installed before installing the application

software. The file path is CD\dotNetFramework\dotNetFx40_Full_x86_x64.exe and


P166681CJ7371124

double click to start the installation. The application software can be used after

installation according to the prompts. Please pay special attention that incorrect

installation will cause the software useless.

Figure 7.1 Interface to begin the installation of .NET

Figure 7.2 Interface of completing the installation of .NET

80 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

(2) Installation of application software

File path of installed software: CD\PF300 Setup.exe. Double click to start the

installation according to the prompts.

(3) Installation of USB drive process

It needs to install drive process before Using USB mode to communicate with PC,

File path of installed software: CD\CP210x_VCP_Win2K_XP_WIN7.exe, double

knock installation key and finish it according to the prompts, otherwise it could not

use USB terminal to communicate well.

See USB terminal: The instrument is connected to computer using USB line, use the

right key of mouse to knock “My computer” and select “property” , and then knock

“equipment management” in the page of “hardware”, knock the “+” symbol of

“terminal(COM and LPT)”,at that time, all terminal information will be shown on the

interface, “CP210x USB to UART bridge Controller(COM X)” in it is the


P166681CJ7371124

communication terminal X USB. Select COM X when the normal software is assorted

with normal instrument.

7.3 Software overview

The PC software of PF300 series is used to obtain the measured value and the data of

harmonic wave, which is transmitted to the PC through the interface of USB(or

RS-232、RS-485、CAN、LAN 、GP-IB). These data are displayed on PC and stored

in PC.(instrument of CAN leave USB and CAN communication ports as standard,

only one port can be used while communicating)

Open the PF300 software to show the login interface. The serial port, instrument type

and baud rate can be set, as figure 7.3 shown. The default baud rate is 115200(Please

choose PF310 if the instrument type is PF310A)

Offline mode is convenient for the user to use the file reading function when it is not

connected to the instrument.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 81


PF300 series User’s Manual

Figure 7.3 Login interface of application software

Click ‘login’ to enter the main interface of software after selection of serial port, baud

rate and instrument type, as figure 7.4 shown.

Function modules of software interface are as follows:

(1) Menu: Open, Save, Save as, Export general data, Export harmonic data,
P166681CJ7371124

Export integral data, ect.

(2) Shortcut icon: Open, Save, Enter remote, Exit remote, Measurement setup,

Start testing, Start integrating,Open harmonics, ect.

(3) Document operation area: Add to, Delete, Clear.

(4) Measurement display area: Normal measurement, Harmonic measurement,

Integral measurement and Data record.

(5) Status

82 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Figure 7.4 Main interface of software


P166681CJ7371124

7.4 Measurement setting

The interface of measurement parameters setting could be set in figure7.5 and 7.6

when the instrument goes into remote state. The function of related measurement

parameters is detailed in chapter 4 .

Figure 7.5 Measurement setting interface of PF310/PF310A

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 83


PF300 series User’s Manual

Figure 7.6 Measurement setting interface of PF330

Table 7.1 Parameters setting


Initialization
P166681CJ7371124

Item Parameters to select


settings
Measurement
RMS、MEAN、DC RMS
mode
Connection mode
(Only for 1P3W、3P3W、3P4W、3V3A 1P3W
PF330)
Display refresh
0.1s、0.25s、0.5s、1s、2s、5s 0.25s
rate
Synchronous
Voltage、Current、OFF Voltage
source
Calculation CFU, CFI, η(PF310/PF310A contains CFU and CFI only) CFU
Peak hold OFF、ON OFF
Wave crest ratio CF=3、CF=6 CF=3
Direct input:
Voltage range CF=3:15V、30V、60V、150V、300V、600V、Auto Auto
CF=6:7.5V、15V、30V、75V、150V、300V、Auto
Direct input:
• PF310/PF310A:
CF=3::5mA、10mA、20mA、50mA、100mA、200mA
0.5A、1A、2A、5A、10A、20A、Auto
CF=6::2.5mA、5mA、10mA、25mA、50mA、100mA Direct input
Current range
0.25A、0.5A、1A、2.5A、5A、10A、Auto Auto
• PF330:
CF=3:0.5A、1A、2A、5A、10A、20A、Auto
CF=6:0.25A、0.5A、1A、2.5A、5A、10A、Auto
External sensor input (optional):

84 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

CF=3:2.5V、5V、10V
CF=6:1.25V、2.5V、5V
Average function OFF、ON OFF
Average mode LIN、EXP LIN
Average constant 8、16、32、64 8
Cutoff frequency 5kHz、500Hz 500Hz
Line filter OFF、ON OFF
Frequency filter OFF、ON OFF
Transformer ratio
OFF、ON OFF
function
Voltage
transformer ratio 1.000 1.000
VT
Current
transformer ratio 1.000 1.000
CT
Power
transformer ratio 1.000 1.000
F
External sensor
ratio
P166681CJ7371124

Transformer ratio
10.00 10.00
coefficient E
Harmonic
measurement
Harmonic PF310/PF310A:OFF、ON
OFF
measurement unit PF330:OFF、Element 1、Element 2、Element 3、ALL
The maximum
50 50
harmonic order
PLL Synchronous
Volt、Curr Volt
source
Calculation
IEC、CSA IEC
method
Integral
measurement
Integral mode Manual、Standard、Continuous repititive Manual
Integral time 00:00:00 00:00:00
Data storage OFF、ON OFF
Data storage
00:00:00 00:00:00
interval
Measurement
configuration File1, File2, File3, File4
storage
Zero calibration Zero-location calibration
System Measurement configuration parameters restore to factory
initialization default values

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 85


PF300 series User’s Manual

7.5 Measurement display

(1) Normal measurement

Select the appropriate setting parameters of parameters in measurement interface such

as measurement mode, measurement range, connection mode and measurement

synchronous source. See chapter 7.4 for more details.

Click ‘start to measure(F2)’ in shortcut icon to open normal measurement function.

Click ‘stop to measure(F3)’ in shortcut icon to stop measurement and hold the

measuring data.

Click ‘single measurement(F4)’ in shortcut icon to single measure and refresh the

measuring data.

The measurement parameters are shown in figure 7.7 and 7.8, normal measurement

parameters and symbols are shown in table 7.2


P166681CJ7371124

Figure 7.7 Display window of normal measurement of PF310/PF310A

Figure 7.8 Display window of normal measurement of PF330

Table 7.2 Parameters of normal measurement


Element 1, 2, 3(PF310/PF310A is the measurement unit) Σ(Only for PF330)

parameters symbol parameters symbol parameters symbol parameters symbol parameters symbol
True True True
effective effective effective
value of value of value of True
Urms Irms UrmsΣ
voltage current Active voltage effective
Umn Imn P UmnΣ IrmsΣ
Rectifying Rectifying power Rectifying value of
Udc Idc UdcΣ
correction correction correction current
value of value of value of
voltage current voltage

86 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Arithmetic Arithmetic Arithmetic


average average average
value of value of value of
voltage current voltage
The The
maximum maximum Reactive Active Power
Upk+ Ipk+ Q PΣ λΣ
value of value of power power factor
voltage current
The The
minimum minimum Apparent Reactive Phase
Upk- Ipk- S QΣ ΦΣ
value of value of power power angle
voltage current
Voltage Current Power Apparent
fU fI λ SΣ
frequency frequency factor power
Wave crest Wave crest
ratio of CFU ratio of CFI Phase angle Φ efficiency η
voltage current
Note:the voltage true effective values of Element 1, 2, 3 and Σ is indicated by Urms1,
Urms2, Urms3 and UrmsΣ. And other parameters can be done in the same manner.
P166681CJ7371124

(2) Harmonic measurement

Select the appropriate parameters of normal measurement parameters and harmonic

measurement associated setting in measurement interface such as harmonic

measurement unit, the highest number of harmonic analysis, PLL synchronous source

and calculation method. See chapter 7.4 for more details.

Click ‘start harmonic(F8)’ in shortcut icon to open harmonic measurement function.

Click ‘stop harmonic(F9)’ in shortcut icon to stop harmonic measurement function.

The measurement parameters are shown in figure 7.9 and 7.10.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 87


PF300 series User’s Manual

Figure 7.9 Display window of harmonic measurement (total harmonic)


P166681CJ7371124

Figure 7.10 Display window of harmonic measurement (single harmonic)

88 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Table 7.3 Harmonic measurement parameters


Total Total
Fundamenta Total Total
harmonic harmonic Total harmonic
parameters l frequency harmonic harmonic
distortion of distortion of power
PLL source voltage current
voltage current
symbol f (Hz) Uthd (%) Ithd (%) ΣU (V) ΣI (A) ΣP (W)
True Relative True Relative
Active Relative value
effective value of effective value of
power of of single
parameters value single single value single single
single harmonic
harmonic harmonic harmonic harmonic
harmonic active power
voltage voltage current current
symbol Urms (V) Uhdf (%) Irms (A) Ihdf (%) P (W) P (%)

(3) Integral measurement

Select the appropriate parameters of normal measurement parameters and integral

measurement associated setting in measurement interface such as integral

measurement unit, integral mode and integral time. See chapter 7.4 for more details.
P166681CJ7371124

Click ‘start integral(F5)’ in shortcut icon to open integral measurement function.

Click ‘stop integral(F6)’ in shortcut icon to stop integral measurement function.

Click ‘reset integral(F7)’ in shortcut icon to reset the integral state, clear the integral

time and measuring data.

The integral measurement is shown in figure 7.11, corresponding integral

measurement parameters areshown in figure 7.4.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 89


PF300 series User’s Manual

Figure 7.11 Display window of integral measurement


P166681CJ7371124

Table 7.4 Parameters of integral measurement


Integral Positive Negative Positive Negative
Total power Total current
parameter accumulated power power current current
integral integral
time integral integral integral integral
symbol TIME Wh Wh+ Wh- Ah Ah+ Ah-

90 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Chapter 8 Common Faults and Solutions

8.1 Common faults and solutions

Serial
Phenomenon Solutions
number
(1) Ensure the instrument power line connected
The instrument window
correctly.
1 display nothing after
(2) Ensure that the power supply is in the allowed
operation.
range.
(1) Ensure that the working environment temperature
and humidity is in the allowed range.
(2) Ensure that the display is not affected by noise
interference.
(3) Check the test line connected correctly or not.
The measured value of
2 (4) Check the connection mode settings correct or
display is not accurate.
P166681CJ7371124

not.
(5) Check the setting of filter switch.
(6) Check synchronous source of measurement and
data refresh ratio.
(7) reboot.
(1) Check the indicator light of REMOTE. Please
press LOCAL to exit remote if the indicator light is
The key operation is
3 on.
invalid
(2) Check the indicator light of LOCK. Please exit the
function of key lock first if the indicator light is on.
(1) Check the setting of PLL synchronous source.
Harmonic measurement is (2) Check whether the input signal of PLL
4
irregular. synchronous source meets the measurement
conditions of harmonic analysis.
(1) Check whether the communication line is
connected properly or not.
(2) Check the communication setting of RS-232 if use
5 Remote control is fail.
the communication mode RS232
(3) Check the communication setting of GB-IB if use
the communication mode RS232.

Please see the notes in each chapter for other problems .

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 91


PF300 series User’s Manual

8.2 Error Code Query

The error code is shown in window D of the instrument if the user operates illegally.

The error operation can be eliminated according to table 8.1.


Table 8.1 Query table of error code
Error
code Cause analysis Solutions
(Err-xx)
Set the measurement mode in the state of Exit the state of integral measurement,
01 integral measurement, harmonic harmonic measurement, HOLD or STORE and
measurement, HOLD or STORE. then set the measurement mode.
Switch connection mode in the state of Please exit the state of integral measurement or
02
integral or STORE. STORE first and then set the connection mode.
Press STOP of integral in the state of Illegal operation is forbidden. It is valid to
03
START of non integral. press STOP only in the state of START.
Press RESET of integral in the state of Please press STOP to stop integral and then
04
STOP of non integral. press RESET to reset.
P166681CJ7371124

Press START at the end of integral in the Please press RESET to reset and then press
05
standard integral mode. START to start standard integral.
The integral time is set as 0 and press Please set the appropriate integral time and then
06 START in the mode of standard integral or press START to start standard integral or
continuous repeated integral. continuous repeated integral.
Set HOLD in the state of integral Please exit integral measurement and then use
07
measurement. the HOLD function.
Please press HOLD to enter the state of data
08 Single measure in non- HOLD state.
hold and then execute single measurement.
Set the transformer ratio coefficient in the Please exit the state of integral measurement,
11 state of integral measurement, harmonic harmonic measurement or STORE and then set
measurement or STORE. the transformer ratio coefficient.
Set the transformer ratio of external sensor Please exit the state of integral measurement,
12 in the state of integral measurement, harmonic measurement or STORE and then set
harmonic measurement or STORE. the sensor transformer ratio.
Set Synchronous source in the state of Please exit the state of integral measurement,
13 integral measurement, harmonic harmonic measurement or STORE and then set
measurement or STORE. the synchronous source.
Set line or frequency filter in the state of Please exit the state of integral measurement,
14 integral measurement, harmonic harmonic measurement or STORE and then set
measurement or STORE. the line or frequency filter.
Set data refresh ratio in the state of integral Please exit the state of integral measurement,
15
measurement, harmonic measurement or harmonic measurement or STORE and then set

92 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

STORE. the data refresh ratio.

Set average function in the state of integral Please exit the state of integral measurement,
16 measurement, harmonic measurement or harmonic measurement or STORE and then set
STORE. the average function.
Please exit the state of integral measurement,
Set calculation function in the state of
17 harmonic measurement or STORE and then set
integral or STORE.
the calculation function.
Please exit the state of integral measurement or
Set the initialization of instrument in the
21 STORE and then set the initialization of
state of integral or STORE.
instrument.
Set measurement wave crest ratio in the Please exit the state of integral measurement or
22
state of integral measurement or STORE. STORE and then set the wave crest ratio.
Set the range switch function in the state Please exit the state of integral measurement,
23 of integral measurement, harmonic harmonic measurement or STORE and then set
measurement or STORE the range switch function.
Change the settings of storge interval in Please exit the state of integral measurement,
24 the state of integral measurement or harmonic measurement or STORE and then set
harmonic measurement. the storage interval.
P166681CJ7371124

Execute zero-location compensation


Please exit the integral measurement and then
25 function in the state of integral
execute zero-location compensation.
measurement.
Set the Peak hold function in the state of Please exit the state of integral measurement or
26
integral measurement or STORE. STORE and then open the Peak hold function.
Set the PLL source in the state of integral Please exit the state of integral measurement or
32
measurement or STORE. STORE and then set the PLL source.
Set the highest number of harmonic Please exit the state of integral measurement or
33 analysis in the state of integral STORE and then set the highest number of
measurement or STORE. harmonic analysis.
Set the calculation method of harmonic in Please exit the state of integral measurement or
34 the state of integral measurement or STORE and then set the calculation method
STORE. of harmonic.
Set integral mode in the stage of integral Please exit the integral measurement and then
41
measurement. set the integral mode.
Set integral time in the stage of integral Please exit the integral measurement and then
42
measurement. set the integral time.
Please exit the integral measurement and then
Set the range of voltage in the state of
51 set the range of voltage or open the function of
integral measurement or STORE.
range switch.
Please exit the integral measurement and then
Set the range of current in the state of
52 set the range of current or open the function of
integral measurement or STORE.
range switch.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 93


PF300 series User’s Manual

The transformer ratio coefficient can not The effective range of transformation ratio
61
be set as 0. setting: 0.001~9999。
Save or load the measurement Please exit the integral measurement and then
71 configuration parameters in the state of save or load the measurement configuration
integral measurement or STORE. parameters.

Note:press LOCAL(SHIFT+SHIFT) to exit remote control in the state of remote .


Other keys functions are not provided and the .measured data are not displayed.
P166681CJ7371124

94 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Chapter 9 Calibration and Test

9.1 Conditions for calibration and test

(1) Ambient temperature: 23℃  2℃

(2) Ambient humidity: 35% ~ 55%

(3) Voltage of power supply: 220V±11V

(4) Frequency of power supply: 50Hz ±1Hz

(5) Preheating time: ≥30 minutes

(6) No mechanical vibration and interference aroundaffect the normal work.

9.2 Technical documents of calibration and test


P166681CJ7371124

(1) JJG(Zhe)90-2007 Digital multimeter calibrator

(2) JJG 780-1992 AC digital power meter

(3) JJG(Zhe)89-2007 Electric parameters test meter

9.3 Equipments for calibration and test

(1) Standard meter(0V~600V,0A~20A)

The precision between 45Hz and 65Hz should be better than 0.05%. The

uncertainty of other frequency should be less than 1/3 to 1/5 of the maximum

permissible error of the tested instrument.

PF2010 digital power meter from EVERFINE is recommended to calibrate

PF300 series instrument.

(2) Signal source(0V~600V,0A~20A)

The precision between 45Hz and 65Hz should be better than 0.02%/3min. The

stability of other frequency should be better than 1/5 to 1/10 of the accuracy of

the tested instrument.

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 95


PF300 series User’s Manual

Recommend: YS6100 standard power from EVERFINE ( test between 45Hz

and 65Hz)

Note:the test equipments should be qualified regular metrological verification

specification.

9.4 Test items and methods

9.4.1 Items of calibration and test

Test the measurement accuracy of voltage, current, active power and active power in

PF300 series instrument. The accuracy of other electric parameters is calculated such

as wave crest ratio and power factor.

9.4.2 Method of calibration and test


P166681CJ7371124

(1) The wiring methods of test are shown in figure 9.1 and 9.2. Please connect

the instrument, signal source and standard meter correctly.

(2) Ensure that the output of voltage and current of signal is in the input range of

the instrument. Read and record the data of voltage, current, active power

and frequency.

The error is calculated according to the formula (1) and the results should meets the

requirements of measurement accuracy in chapter 2.2.

R=BX-BN (1)

where:

BN:reading data of standard meter

BX:reading data of tested instrument

R: instrument measurement error.

96 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Figure 9.1 Wiring diagram of calibration and test of PF310/PF310A


P166681CJ7371124

Figure 9.2 Wiring diagram of calibration and test of PF330

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 97


PF300 series User’s Manual

Appendix

Appendix I Definition and symbol of measurement parameter

(1) Routine measurement and integral measurement


Symbols
Measurement of
Formula Definitions
Parameters parameter
s
Urms Urms Umn Udc
Voltage Umn
AVG  u n  
π
AVG [ u ( n ) 2 ] AVG [ u ( n )]
Udc 2 2
Irms Irms Imn Idc
Current Imn
AVG  i n  
π
AVG [ i ( n ) 2 ] AVG [ i ( n )]
Idc 2 2
Active Power P[W] AVG u n   i n 
P166681CJ7371124

Apparent
S[VA] Urms·Irms Umn·Imn Udc·Idc
Power
Reactive
Q[var] S 2 - P2
Ro Power
uti Power Factor λ P/S
ne
Phase Angle Φ[deg] COS-1(P/S)
Voltage
me fU[Hz] Voltage Frequency
Frequency
asu
Current
re fI[Hz] Current Frequency
Frequency
me
The
nt
Maximum The maximum value of instantaneous
Upk+[V]
value of voltage
voltage
The
Minimum The minimum value of instantaneous
Upk-[V]
value of voltage
voltage
The
Maximum The maximum value of instantaneous
Ipk+[A]
value of current
current

98 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

The
Minimum The minimum value of instantaneous
Ipk-[A]
value of current
current
Wave Crest
Upk / Urms
Ratio of CFU
Upk= |Upk+| or |Upk-|,take the larger
Voltage
Wave Crest
Ipk / Irms
Ratio of CFI
Ipk= |Ipk+| or |Ipk-|,take the larger
Current
Efficiency(PF
—— η
330)
Int Wh
1 N

egr Power
Wh+ N

 u n   i n   Time
n 1
al integral
Wh- Time:integral time (unit: hour)
me Ah
asu 1 N

re Current
Ah+ 1 N

 I n   Time N
 i n   Time
P166681CJ7371124

n 1
N n 1
me integral i(n):the nth
Ah- I(n):the nth measuring value
nt sampling value

(2) Hormonic measurement


Relative values of
Parameters RMS Relative values of CSA
IEC
max

 Uk 
max 2
max
 Uk  2

U  Uk
2
Voltage k 2
Uthd  k 2
100 Uthd  100
U1
k1
U
max
Total
 Uk 
max 2
max
 Uk
2

I  Ik
2
hormonic Current k 2
Ithd  k 2
100 Ithd  100
k1 I1 I
max

P  Pk
Active
power k1

U k  U k 
Voltage Uk  Urk  Ujk
2 2 Uhdf  100 Uhdf  100
Single U 1 U
hormonic
Ik  Ik 
Ik  Irk  Ijk Ihdf  100 Ihdf  100
2 2
Current
I1 I

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 99


PF300 series User’s Manual

Active P(k) = Ur(k) • Ir(k) +


power Uj(k) • Ij(k)

Power factor λ=P(1)/S(1)

Phase angle The phase angle between fundamental voltage and current

max:The max times of harmonic analysis


Note
k: harmonic times(1~max)
P166681CJ7371124

100 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Appendix II Process of Menu settings

(1) SETUP menu

Menu setting of PF310/PF310A


P166681CJ7371124

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 101


PF300 series User’s Manual

Menu setting of PF310 /PF310A(continued)


P166681CJ7371124

102 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Menu settings of PF330


P166681CJ7371124

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 103


PF300 series User’s Manual

Menu setting of PF330(continued)


P166681CJ7371124

104 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

(2) UTILITY menu

UTILITY menu of PF310/PF310A


P166681CJ7371124

UTILITY menu of PF330

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 105


PF300 series User’s Manual

(3) HARMONICS menu

HARMONICS menu of PF310/ PF310A


P166681CJ7371124

HARMONICS menu of PF330

106 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

(4) INTEGRATION Menu


P166681CJ7371124

(5) SAVE/LOAD Menu

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 107


PF300 series User’s Manual

(6) INTERFACE Menu


P166681CJ7371124

108 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Overview

PF300 series digital power meter is a type of compact multi-channel and

multi-functions digital power meter, it's applicable for requirements of production,

measurement, evaluation and research etc.

PF300’s characteristics are as follows:

(1) Test element

PF310/PF310A: single channel test element;

PF330: multi-channel test element.

(2) Wide voltage input

Voltage:PF310/PF310A/330:75mV~600V;

Current:PF310/PF310A:0.025mA~20A;PF330:2.5mA~20A;
P166681CJ7371124

Frequency:DC、0.5Hz~100kHz。

In addition, PF300 can provide scaling function, test with external voltage or

current mutual inductor, it also provides external output-voltage type of output

current mutual inductor to test large current.

(3) Simultaneous measurement of multi-function

General parameters, harmonic analysis and integration capabilities can be

simultaneously measured.

(4) Auto range function

Specify range skips automatically in general measurement.

Auto range in the integrating function.

(5) Peak value holding function

Can hold the real value and peak value of voltage and current, the maximum

value of active power, reactive power, apparent power.

(6) Efficiency calculation function (PF330 only)

Single instrument can realize the efficiency calculation function.

(7) Fast display and data update

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 109


PF300 series User’s Manual

Data update rate is optional, speeds up to 0.1s.

(8) PC Software

Provide application software to set measurement parameters quickly and

obtain measurement data and wave.

(9) Communication Interface

USB for default, RS-232, RS-485、CAN、LAN and GP-IB for optional..


P166681CJ7371124

110 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

Specifications

Input Specifications

Items Voltage Current


Direct input: Big terminal
Input terminal
Small terminal External current sensor input(option):
type
Isolated BNC connector
Floating inputs with safety isolation
Input format
Resistive voltage divider Shunt input
Direct input:
PF310/PF310A:
Approx. 500mΩ+10μH
(range ≤ 200mA)
Input impedance Approx. 2MΩ+13pF Approx. 5mΩ+10μH
(range ≥0.25A)
P166681CJ7371124

PF330: Approx. 5mΩ+10μH


External input:
PF310/330 Approx. 100kΩ
PF310/PF310A Direct input:
CF=3:
5mA/10mA/20mA/50mA/100mA/
200mA/0.5A/1A/2A/5A/10A/20A
CF=6:
Measurement PF310/PF330 Direct input: 2.5mA/5mA/10mA/25mA/50mA/
Range CF=3:15V/30V/60V/150V/300V/600V 100mA/0.25A/0.5A/1A/2.5A/5A/10A
(RMS) CF=6:7.5V/15V/30V/75V/150/V300V PF330 Direct input:
CF=3: 0.5A/1A/2A/5A/10A/20A
CF=6: 0.25A/0.5A/1A/2.5A/5A/10A
External input(option):
CF=3: 2.5V/5V/10V
CF=6: 1.25V/2.5V/5V
Frequency
DC,AC 0.5Hz ~ 100kHz
Range
Instantaneous
maximum
2kVpk or 1.5kVrms, whichever is less 90Apk or 30Arms, whichever is less
allowable
input(≤1s)
Continuous
maximum 1.5kVpk or 1kVrms, whichever is less 60Apk or 22Arms, whichever is less
allowable input

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 111


PF300 series User’s Manual

Line filter:Select OFF, 500Hz or 5kHz


Filter
Frequency filter:Select OFF, 500Hz or 5kHz

Measurement Accuracy

Temperature: 18℃ ~ 28℃

Humidity: 30%R.H.~75%R.H.

Input Wave: Stable sine wave


Measurement
Input Range: 10~100% of the fixed range
Condition
Common-mode voltage: 0V

PF: COSΦ=1

Crest Factor: CF=3

Items Voltage(V)/Current(A) Active Power(W)


DC ±(0.1%Reading+0.1%Range+1digit) ±(0.1%Reading+0.2%Range+1digit)
P166681CJ7371124

10Hz≤f<45Hz ±(0.1%Reading+0.2%Range+1digit) ±(0.3%Reading+0.2%Range+1digit)

45Hz≤f≤65Hz ±(0.05%Reading+0.1%Range+1digit) ±(0.05%Reading+0.1%Range+1digit)

65Hz<f≤1kHz ±(0.1%Reading+0.2%Range+1digit) ±(0.2%Reading+0.2%Range+1digit)


±(0.07×f %Reading+0.3%Range ±[(0.1%Reading+0.3%Range)
1kHz<f≤10kHz
+1digit) +(0.067×(f-1)%Reading)+1digit]
±[(0.5%Reading+0.5%Range) ±[(0.5%Reading+0.5%Range)
10kHz<f≤100kHz
+(0.04×(f-10)%Reading)+1digit] +(0.09×(f-10)%Reading)+1digit]
f f in the read error equation is the input signal frequency in kHz.
Line filter Low pass filter is available during test. Cut-off frequency: 5kHz or 500Hz.
influence When the line filters are opened(45Hz ≤ f ≤ 65Hz), add ±0.5% of range.
Crest factor When crest ratio is set as 6, doubling the range error of that crest factor is set
influence to 3.

Note Temperatur
Original accuracy ±0.02%Range/℃ in 5℃ ~ 18℃ and 28℃ ~ 40℃.
e influence
100%~130
Original accuracy ±0.5%Reading.
% Range
0.5Hz~10Hz:Measurement data is reference value.
Frequency
DC,10Hz~45Hz,400Hz~100kHz:
influence
Measurement data is reference value when the current is over 20A.

112 Copyright © EVERFINE, Copy or spread without authorization is prohibited.


PF300 series User’s Manual

General Specifications
Preheating Time Approx. 30 minutes or above
Operating Temperature: 5℃~40℃
environment Humidity: 20%R.H.~80%R.H. (no condensation)
Insulation
>50M( (each one between the signal input, power input and shell)
Resistance

AC2000V (between the signal input and shell,between the signal


Withstand Voltage input and power input in 1 minute), DC2200V (between the power
input and shell in 1 minute)

Power Supply 100VAC~240VAC,50Hz/60Hz


External
248mm×147mm×384mm (W×H×D)
Dimensions
PF310: 5.3kg
Weight PF310A: 3.4kg
PF330: 6.6kg
P166681CJ7371124

Copyright © EVERFINE, Copy or spread without authorization is prohibited.. 113