1, Connect the cable to the laptop.
2, To start “BAlarm.exe” software.
3, Select COM port, where cable is connected. “Communications ‐> COM port…”
4, In the PPM reading screen press “MODE” key two times and following screen will be show
5, Press ‘ENTER” key, countdown timer starting: “200‐>199‐>198‐>197‐>196… ‐>0”
To start data transfer: select “communications‐>download”, if monitor is correctly connected,
transfer should start automatically, and the internal storage dates (from around 25 month)
will be transfer to the laptop.
Screen “communication window” will be show in next page.
Program will now transfer all data stored on memory card. This process will take minutes, pls
wait.
If an error occurs, program will finish communication and inform to user, in this case, pls
check cable connection, configure of COM port, or if monitor is transmission mode or not.
Transmission mode:
You can see the picture of item 4 above, Of course, count down timer must start through
pushing “ENTER”.
If transmission ends correctly, communication window will disappear, receive data process
will start, this process also takes some minutes, pls wait.
6, Data presentation
State
Table
Estado
Graphic
Event
Table
Description in the next page ‐>
State Table Estado:
COLUMN DESCRIPTION
Num Number of record
Date Date in which record was saved
Time Time in which record was saved
S1 PPM on cell 1
S2 PPM on cell 2
LM Manual cleaning
LA Auto. cleaning/oil discharge
SBS Separator pump overload
SBC Circulating pump overload (option)
RS Recirculation to bilge
DM Overboard discharge
BS Separator on
BC Circulating pump on (option)
LQ Chemical cleaning
A1D Alarm 1 delayed
A2I Alarm 2 instantaneous
A2D Alarm 2 delayed
Event Table
COLUMN DESCRIPTION
Num Number of record
Date Date in which record was saved
Time Time in which record was saved
Event Type of event
Parameters Parameters depending on type of event
Column “Event” of events table can contain following values:
z POWER OFF: Saves date and time in which equipment was disconnected, disconnection
instant is saved on column “PARAMETERS”.
z OFFSET S1/S2: Shows current value of measurements offset on cell1/2. This value will be
shown in column “PARAMETERS”.
z “NEW OFFSET S1/S2”: Indicates that there was a charge in offset value on cell 1/2. Old and
new offset value will be shown on column “PARAMETERS”.
z DELAY AL1/AL2: Indicates actual set‐point of alarm1/2 on column “PARAMETERS”.
z NEW SETPOINT AL1/AL2: Indicates that there was an adjustment on alarm 1/2 set point.
Set‐point value will be shown on column “PARAMETERS” before and after adjustment.
z ERROR: An error is indicated through default code (Type” and “Num”) on column
“PARAMETERS”, this code corresponds with Alarm Codes listed on next page.
7, Data filter
You can filter option that you want
Bilge Water Separator and 15ppm Bilge Alarm Monitor Instruction Manual
2.5 TROUBLE-SHOOTING
MESSAGE ON
FAULTY ACTION
DISPLAY
Intensity of direct light is over It is possible that unit needs technical support.
A03
maximum level on sensor 1.
Intensity of direct light is over It is possible that unit needs technical support.
A04
maximum level on sensor 2.
Intensity of scattered light is below Cell 1 can need to be cleaned (see section 2.4.1.)
A05
minimum level on sensor 1.
Intensity of scattered light is below Cell 2 can need to be cleaned (see section 2.4.1.)
A06
minimum level on sensor 2.
Intensity of scattered light is over It is possible that unit needs technical support.
A07
maximum level on sensor 1.
Intensity of scattered light is over It is possible that unit needs technical support.
A08
maximum level on sensor 2.
Inserted FLASH card does not belong Each monitor operates only with included card
A13
to the unit and non with others
A14 Hardware error in FLASH card It is possible that unit needs technical support.
A15 Hardware error in A/D converter It is possible that unit needs technical support.
Change desiccator
Desiccator used up
>5ppm Clean water Clean simples tube (See section
Sample tube is dirty
2.4.1.)