Professional Documents
Culture Documents
Operation Manual
Operation Manual
Version: 1.1
MA-6000
Real-Time Quantitative Thermal Cycler
Safety Introduction......................................................................................................................... - 5 -
Chapter 1 Introduction..................................................................................................................- 10 -
Chapter 10 Troubleshooting......................................................................................................... - 67 -
Update History..............................................................................................................................- 68 -
Notes
Hardware:
Molarray Biotech guarantees that the MA-6000 Real-Time Quantitative Thermal Cycler that you
have received has been thoroughly tested and meets its published specifications.
This warrantee is void if the guidelines in this User Instruction Manual regarding the system and
Included Software:
Any software distributed with this product is provided free of charge as a service to the customer.
The software is intended to be a tool for development and as an example of one possible method
of code implementation..
Software Copyright:
Molarray Biotech maintains the copyright on this material, but grants the customer rights to use
the software described herein without obtaining Molarray Biotech’s permission and without the
Software Changes:
Molarray Biotech reserves the right, without prior or future notice, to make changes to any of its
Liability:
No liability is accepted for loss or incidental or consequential damages arising from the incorrect
use of the MA-6000 system. Molarray Biotech’s liability is limited to the repair or replacement
of the unit. Molarray Biotech is not liable for any consequential damages.
Read this Instruction Manual carefully before using the MA-6000 to ensure that you obtain
-4-
Safety Introduction
1. Definition of symbols
The following symbols will appear in the instruction manual.
Symbol Title Description
Warning This symbol is used to indicate the following
information: if it does not follow the prescribed
procedure or instruction, it may result in physical injury
or damage to the instrument.
-5-
Crushing Labeled on the surface close to the thermo block.
2. Operating Requirements
The instrument can only be operated by skilled personnel.
►The instrument is mechanical and electrical equipment, if it is not strictly
used according to the manual, it may cause electric shock or physical damage and
other potential dangers to the user;
►Operation should be in strict accordance with the instrument safety tips;
►The user can replace the fuse according to the manual. Be sure to cut off the
power before replacing. Parameter of the fuse is F8AL 250V. However, the user can
not open the instrument or replace other accessories, and the damage to the
instrument is not included in the scope of warranty;
►Only the professional personnel of this instrument manufacturer can repair
the instrument;
►Do not disassemble when the instrument is running to avoid damage to the
instrument. Be sure to cut off the power before disassembly.
►The track of heated-lid is mechanical moving component. Touch is strictly
prohibited when the instrument is running to avoid hand crushed.
►The instrument must be installed in clean indoor ventilated, avoid corrosive
gas and strong magnetic field interference, avoid sunlight and direct strong light, and
used in the specified temperature and relative humidity conditions;
►According to the product technical specifications, working conditions of the
instrument for indoor air temperature is between 18℃ and 35℃; relative humidity
should be ≤85%.
►Safety goggles and gloves must be taken when dealing with toxic, corrosive
or infectious materials;
► Be careful to guard against the potential hazards of all biological materials
although the nucleic acids are highly purified. The disposal or discard of these
-6-
wastes must comply with local safety regulations.
If the spatter or leak occurs carelessly, it should be disinfected immediately with
appropriate disinfectant to prevent the contamination of laboratory personnel and
equipment.
►The damaged instrument must be returned to the manufacturer for repair; the
surface of the instrument must be disinfected before return.
► It is strictly forbidden to touch thermo block to avoid scald when the
instrument is running and when it is finished.
3. Electrical safety
►The electrical safety design protection level of the MA-6000 Real-Time
Quantitative Thermal Cycler is Class I (IEC);
► In order to prevent the shock hazard, the instrument must be connected to a
three-core grounding socket that meets the standard of the safety standard. The
voltage is 220V (50Hz).
►Before the instrument is connected to the power line, it is necessary to ensure
that the voltage, frequency is consistent with the requirements of the instrument.
When the power cord is connected, the power supply must be cut off.
►Before the instrument is connected to the computer, it is necessary to ensure
that the power supply of instrument and computer is cut off to prevent any
unfortunate damage to the instrument;
► Do not touch the power switch and power cord with wet hand;
► Do not remove the power cord when the instrument is not cut off;
► Do not clean the instrument when the instrument is not cut off;
► Do not replace the fuse when the instrument is not cut off;
►Please turn off the power when the instrument is not in use.
4. Electromagnetic Compatibility
Real-Time Quantitative Thermal Cycler should comply with the emission and
noise immunity requirements of GB/T 18268.1-2010 Electrical Equipment for
Measurement, Control and Laboratory Use-EMC Requirements-Part1: General
-7-
requirements and Part 26: Special requirements in vitro diagnosis (IVD) medical
equipment.
Manufacturer is responsible to provide electromagnetic compatibility
information of Real-Time Quantitative Thermal Cycler.
User should ensure to use the instrument in proper electromagnetic compatibility
environment.
Real-Time Quantitative Thermal Cycler complies with GB/T 18268 noise
immunity requirements:
-8-
Prohibit using the instrument near strong radiation source, or the instrument
will be interrupted.
-9-
Chapter 1 Introduction
- 10 -
products. A broad spectrum excitation source provides high excitation energies over
a wavelength range from 380 to 780nm. All sample positions are illuminated and the
resulting fluorescent emissions are transmitted through optical fibers to a CCD
camera. Excitation and emission wavelengths are selected from within this range
using interference filters. Five filter sets are provided, which are optimized for the
detection of the most commonly used fluorescence. The filter sets for other new
fluorescence are also available and can be supplied according to customer
requirements.
- 11 -
Chapter 2 Overview of MA-6000 Instrument
2.1.3 Contraindication
No.
- 12 -
Gradient temperature range over 32℃ in 12 columns;
Dynamic range: linear dynamic range of sample concentration up to 10 orders
of magnitude;
Rapid data acquisition;
Capacity to simultaneously track 96 samples;
Accessibility of raw data files by the user;
Ability to re-analyze data;
User-friendly, intuitive software to speed up setup and data analysis;
- 13 -
2.1.7 Fluorescence detection system
The fluorescence system of MA-6000 consists of two optical parts: an
excitation system and a detection system. The excitation system, comprised of a
halogen lamp and a set of optical filters, allows simultaneous illumination of the 96
wells through 96 optical fibers. The detection system includes a set of optical fibers,
a wheel with five emission filters, a set of lenses, and a cooled CCD camera. The
fluorescence signal transmits through the optical fibers, emission filters and the
lenses and is then detected by the CCD camera.
Specifications of optical system:
Light source: halogen lamp.
Fluorescence detection range: 380 to 780nm.
Excitation wavelength: up to 5 channels (380~780nm)
Emission wavelength: up to 5 channels (380~780nm).
2.1.8 Software
- 14 -
The MA-6000 software should be run in Win7 operation system and above. It
can monitor all the process of gene amplification via computer. User can setup PCR
protocol, setup sample parameters, set file location, collect and analyze the data
signal quickly and easily in the software.
Processor Inter i3
RAM 4G
USB 3.0
USB
Location: Lower left on back of the instrument.
USB cable connects the instrument and computer via their USB 3.0 ports.
Power port
Location: Lower left on back of the instrument.
- 15 -
Three core power cord connects to the port in above picture, another end
connects to 220V AC power. Ground wire of power should be properly grounded.
The length of power line should be less than 1.5m. The socket should be installed in
where user can pull and plug power line conveniently.
2.2 Cautions
Some sample maybe contain infection source, must be sure deal with samples
following relative safe operation rules. User can wear safe glasses, latex gloves and
lab coat. Please operate in corresponding environment if needed.
Person in charge must take prevention measures to train for whom will operate
this instrument and avoid touching samples containing infection source directly.
Follow applicable local bylaws and safety regulations during handling or
disposal of the waste. In the accident of spills or leaks, appropriate disinfection
procedures must be followed to prevent and eliminate personnel or instrument
contamination.
- 16 -
Chapter 3 Unpacking & Installation
- 17 -
Voltage: 220V±22 V, 50Hz±1Hz.
Power: 1000VA
Click “Next” to install step by step. Software running platform will be installed
successfully after clicking “Finsh”.
- 18 -
3.3.2 MA-6000 workspace installation (Step2)
Click “MA-6000-Workspace”, enter password “molarray”, advise to install to
non-system disk. Please install following below instructions.
- 19 -
Click “CCD”.
- 20 -
After software installation, connect the instrument and computer using the USB
cable. Turn on the instrument, check the icons of USB and CCD in “My
Computer”-“Device Manager”. The instrument is ready to run if the icons are
displayed correctly.
- 21 -
(3) Click main menu-Instrument, and then select Run From Template.
(4) Select “Testing” template to run for a self-test routine. The routine will
check the heating elements, light source, cooling fans, heated lid, filter wheel, lenses
and CCD camera. This approximately takes 30 minutes.
Notes:
You can carry out qPCR experiment only when the test results are passed.
The MA-6000 is very easy to operate. For detailed operation procedures, please see
- 22 -
Chapter 4 Introduction of MA-6000 Software
MA-6000 software can be used for setting, running, data collection, analysis and
management. Graphic display of the amplification process is shown in real time. The
extensive preferences menu allows the user to easily re-analyze the results.
- 23 -
Make sure that MA-6000 software is properly installed on the computer.
Fig 4.3.2 a
- 24 -
Fig 4.3.2 b
Click Add Stage to add a Holding or Cycle, click Add Step to add different
steps in one stage. Set temperature in T column, set time in Hold column, check
Signal Acquisition to collect signal.
Fig 4.3.2 c
For example:
37℃ 2min;1 Cycle;
94℃ 2min;1 Cycle;
94℃ 20s;
55℃ 30s; 40 Cycles; 55℃ Signal Acquisition
Run Melt Curve experiment after amplification completed, Signal Acquisition
of Melt Curve is default. The protocol is as Fig 4.3.2 d.
- 25 -
Fig 4.3.2 d
Turn to Plate page to set sample parameters after protocol is completed. Select
well and double click or right click on it, below page will display as Fig 4.3.2 e.
Click on well setting.
Fig 4.3.2 e
Select different SubSet ID to define different groups. Enter sample name and
gene name, select sample type, select reporter from the left table. Select all intended
reporters if need, click Add and Apply to complete the sample parameters setting for
the selected well. Set sample information for other wells as above steps. (Fig 4.3.2 f)
- 26 -
Fig 4.3.2 f
to run, or click Save As Template from File to save the current file as
template. User can select Run From Template from Instrument to directly run
experiment next time.
4. After experiment run, click File, select Open to open the experiment file just
completes run.
- 27 -
Chapter 5 Document
5.1 Glossary
- 28 -
Check it in certain step to collect signal.
Module
The heating block can be divided into six independent heating modules.
- 29 -
Sample Type
A sample can be designated as one of the following five sample types. They
are Unknown Sample, Negative Control, Positive Control, Weak Positive Control,
and Standard Sample. For Standard Samples, you need to enter the quantity of the
template.
Quantity
Enter the concentration value of each standard sample dilution gradient.
Add
Add sample information to select wells.
Apply
Apply the new sample information to the selected wells.
Cancel
Cancel the sample information setting.
Fig 5.2.1 a
- 30 -
Name: a file name will be automatically created for the new experiment, or user
can edit it.
Date: date of experiment run. The default value is the date of computer system.
Operator: it is login ID by default; user can edit it to technician name.
Template: click the folder icon of template to import only protocol of the
template, the sample setting information will also be imported if the checkbox (Load
the plate settings) is checked.
Location: the file path to save the experiment file.
Finish: click the Finish to set up a new Thermal Cycling Protocol. (Fig 5.2.1 b)
Fig 5.2.1 b
Click Add Stage to add a Holding or Cycle, click Add Step to add different
steps in one stage. Set temperature in T column, set time in Hold column, check
Signal Acquisition to collect signal.
For example:
37℃ 2min; 1 Cycle;
94℃ 2min; 1 Cycle;
94℃ 20s;
55℃ 30s; 40 Cycles; 55℃ Signal Acquisition
Run Melt Curve experiment after amplification completed, Signal Acquisition
of Melt Curve is default. The protocol is as Fig 5.2.1 c.
- 31 -
Fig 5.2.1 c
- 32 -
Fig 5.2.1 d
Fig 5.2.1 e
Multiple Block
Select the annealing step; click the button of Multiple Temp. Select Set and
click Multiple Block.
Enter a number of the temperature for each module, such as 55℃ for Module 1,
56℃ for Module 2,
58℃ for Module 3, 60℃ for Module 4, 59℃ for Module 5 and 57℃ for
Module 6.
Note:
Temperature interference can occur in the wells between modules when the multiple
It is suggested that the difference in temperature should be smaller than 5℃, and do
- 33 -
Fig 5.2.1 f
Fig 5.2.1 g
Fig 5.2.1 h
Cancel the Multiple Temp. function: Click Multiple Temp, select Cancel. The default
- 34 -
5.2.2 Plate Setup
The Plate Setup window allows you to edit sample information for each well,
such as the sample ID, primer names, sample types, and reporters.
After protocol setting, click Plate (Fig 5.2.2 a), edit sample parameters for
selected wells. Click on a well and drag the mouse to select the desired well or wells
from Well Selection Panel. The color of the selected area will change from white to
gray. Right click of the mouse then click well setting to enter the screen of ‘Well
Settings’ (Fig. 5.2.2 b) and to edit Subset ID, Sample name, Gene name, Sample type
and Reporters.
1) Subset ID: The software allows you to define seven different samples in one
PCR run, (SI to SVII);
2) Sample Name: Enter a Sample Name, such as H-100;
3) Gene Name: Typing a Gene Name, such as HBV;
4) Sample Type: A sample can be designated as one of the following seven
sample types. There are five types, such as Unknown Sample, Negative Control,
Weak Positive Control, Positive Control, and Standard Sample. For Standard
Samples, you can easily enter the quantity of the template;
5) Reporter Selection: After the wells have been defined, the channel can be
selected. Click on the desired reporter name, such as FAM, and then click the button
of Add button to confirm;
6) Click Apply button to finish the well setting;
7) The results of Well Setting will be displayed.
- 35 -
Fig 5.2.2 a
Fig 5.2.2 b
- 36 -
Fig 5.2.2 c
Fig 5.2.2 d
- 37 -
Channel (filter)# Excitation (nm) Emission (nm)
1 470±10 520±5
2 525±5 570±5
3 570±5 620±5
4 620±5 670±5
5 670±5 710±5
Extension User customized User customized
Fig 5.2.3
- 38 -
Protocol: click it to check the protocol of this experiment.
Plate: click Plate to check the sample parameters, user can edit the sample
parameters after the run; ensure to save the change after editing.
Amplification Plot: this page displays curves of Rn (Rn) vs cycle number, Ct
value, Copy number, standard curve.
Dissociation: check the Melt Curve experiment result.
HRM Analysis: check the HRM experiment result.
ΔΔCt: this analysis method is widely used in gene expression analysis. It
requires same amplification efficiency of reference gene and target gene. 2-ΔΔCt
method is based on assumption that amplification efficiency is 100%.
Quality Control: it can monitor same type experiment in long term, compare
their Ct values and amplification efficiency.
Results: it lists the data of each well.
Fig 5.3 a
Select well and its reporter to check the data in the certain Reporter, or select all
to check the data of all channels.
Data includes: Delta Rn, Rn and Std Curve.
Delta Rn: the normalization of Rn value and plot.
Rn: intensity of fluorescence (Fig 5.3 b), the max Rn value can be detected is
60000, advise to optimize the max Rn value in experiment is 45000. Rn value can be
increase or decrease by adjusting exposure time.
- 39 -
Fig 5.3 b
Fig 5.3 c
Fig 5.3 d
- 40 -
Line Color: Line Color includes Auto color and Sys color. User can select any
color type to define a color for the amplification plot of one Reporter, or
amplification plot of each well has its own color.
Threshold: Threshold Setting includes Auto Threshold and Manual Threshold.
Baseline: it is the fluorescence value which does not increase in early several
cycles of qPCR experiment. Default values of Baseline Start and End are 3 and 12. If
Ct value is less than 12, user needs to adjust the Baseline End to a cycle less than the
Ct.
Select data: right click on the blank near the amplification plot (Fig 5.3 e).
View Data Table: user can check the well position, group, reporter, sample
name, sample type, Ct value, Mean Ct, StdDev Ct. etc. Right click again to select
Return to return to amplification plot.
Graph Style: Line and Log (Fig 5.3 f).
Curve Setting: set the thickness of plot (Fig 5.3 g).
Auto Scale: auto fit the amplification plot to the window (Fig 5.3 h).
Copy: copy the date to excel.
Save as picture: save the selected amplification plot as picture.
Fig 5.3 e
- 41 -
Fig 5.3 f
Fig 5.3 g
Fig 5.3 h
Click Protocol to check the reaction program settings information (Fig 5.3 i).
- 42 -
Fig 5.3 i
Click Plate to check the sample parameters (Fig 5.3 j), user can edit the sample
information after experiment run, click Save after editing.
Fig 5.3 j
Dissociation: check the Melt Curve. It has two data types: Rn and Di/Dt (Fig
5.3 k, Fig 5.3 l). User can see the Tm value in Results table, or get the Tm value by
right clicking on the blank near the melt curve and selecting View Data Table.
- 43 -
Fig 5.3 k
Fig 5.3 l
Delta Delta Ct: before ΔΔCt analysis, user must set parameters for intended
wells, each well only has one Subset ID, Sample name and Gene name.
Control: define the sample which will be used for control.
Reference: usually a housekeeping gene used as a reference to which compare
the expression of target genes.
Target gene: The gene whose expression is compared to the reference gene
Click Gene Expression, Select RQ from Data Mode dropdown list, Click
Define Reference, select a Subset ID, set a group as Control Group, set a
housekeeping gene as Reference Gene, click Add or Remove, the setting is
- 44 -
completed. Close the setting page; the gene expression plot will be displayed (Fig
5.3 m).
Fig 5.3 m
Quality Control
The MA-6000 software offers a tool for quality management. To measure any
variability between runs, you can fix a well to carry out the quality control
experiment with the same conditions (same reagent and template) in every qPCR
run.
Channel or reporter Selection;
Well Selection;
Sample Type Selection;
Set the starting and end time for analysis;
Select a display mode: The curve can be displayed as Copy number vs. Days or
Ct value vs. Days;
Click Query to display the result.
Results
Data types: There are two Data types of result, Research and Medical (Fig 5.3
n).
Mean Ct: Mean Ct is the average Ct value of wells which have the same
sample parameters setting.
- 45 -
Stdev Ct: Stdev Ct is the standard deviation of Ct values which have the same
sample parameters setting.
Quantity: Quantity is the concentration of standard sample and unknown
sample.
Tm: Tm is the temperature at which 50% of the DNA is double-stranded and
50% of the DNA is dissociated into single-stranded DNA. The Tm is displayed in the
melt curve.
Fig 5.3 n
- 46 -
Fig 5.5 a
Fig 5.5 b
The exported report is as Fig 5.5 c, click the different worksheet to check
Protocol&Plate, Data&Graph etc.
Fig 5.5 c
Fig 5.6
- 48 -
Chapter 6 Settings
This chapter lists some settings used before experiment running and when
analyses experiment, such as Static Fluorescence and History Data. The settings
before experiment running includes Reporter Management, Common Reporters,
CCD Setting, etc., analysis setting includes threshold setting and baseline setting,
etc.
Fig 6.1
- 49 -
Put the PCR plate into the instrument, turn on the instrument and launch the
software. Click Tools- Static Fluorescence, the page of Fig 6.2 a will display, click
Run, the page will display as Fig 6.2 b after running.
The table data is consistent with the PCR plate.
Exposure time: user can customize it, the default value is 1s.
Fig 6.2 a
Fig 6.2 b
- 50 -
6.3 History Data
All the experiment data have run will be saved in database. User can search the
experiment in History Data.
Click Tools, select History Data, and enter a keyword such as sample name,
date and click Query to search. The listed experiment can be ordered by clicking the
column name. Click Export to export the report. The exported reports will be saved
in the reports folder in the software installation path (Fig 6.3).
Fig 6.3
the software.
- 51 -
Fig 6.4.1
Fig 6.4.2
- 52 -
on the sample background and the dynamic range of fluorescence intensity. The
Exposure Time can be adjusted channel by channel as the different reporters have
different quantum efficiency. The default value of Exposure Time is set for all
channels. The value of Exposure Time can be adjusted to optimize the background of
baseline and the fluorescence intensity of the Plateau. The dynamic range of
fluorescence intensity for the MA-6000 instrument is recommended to be between 0
and 50,000. Thus, the maximum fluorescence intensity of fluorescence (at the PCR
plateau phase) must be below 50,000 (arbitrary units). It is recommended that
Exposure Time value be set between 0.5 and 3).
Fig 6.4.3
- 53 -
Fig 6.4.4
Fig 6.4.5 a
User also can set different judgment threshold for different gene name. Click
Add to go to Fig 6.4.5 b. Enter Gene Name, select Test Mode and corresponding
judgment threshold, click Create button. After an experiment run, if the gene name
- 54 -
set in the experiment is same with a gene name in the Negative judgment which user
has created, the judgment will be automatically matched and used for the negative or
positive judgment. User also can remove the selected judgment.
Fig 6.4.5 b
- 55 -
Chapter 7 Software Strategy
Fig 7.1 a
2. Following Fig 7.1 b, select Report Mode (Lis), select File Type (Excel), and
click Export.
Fig 7.1 b
3. Open the exported file in the exported location (Fig 7.1 c).
- 56 -
Fig 7.1 c
Fig 7.2 a
- 57 -
Fig 7.2 b
Fig 7.2 c
User also can edit the color for well and amplification plot for experiment after
run. Open an experiment file, go to Plate window, edit the color of selected well,
click Save after editing (Fig 7.2 d).
- 58 -
Fig 7.2 d
The amplification plot color is consistent with the well color (Fig 7.2 e).
Fig 7.2 e
- 59 -
Fig 7.3
Select wells user want to export, go to File-Export, select Medical report mode
to export. Go the exported location to check the exported medical report and print it.
- 60 -
Fig 7.4 a
Fig 7.4 b
Fig 7.4 c
- 61 -
Fig 7.4 d
After experiment run, go to Gene Expression page. Select RQ in Data Mode
dropdown list (Fig 7.4 e), click Define Reference button to go to Control Group and
Reference Gene Setup page (Fig 7.4 f), click Add after setting.
Fig 7.4 e
If RQ<1, it indicates that target gene expression of treatment group is less than
it in control group. If RQ>1, It should be the opposite.
Fig 7.4 f
- 62 -
Fig 7.4 g
Steps:
Remove MA-6000.exe from installation location;
Remove the MA-6000 icon from desktop;
Copy new MA-6000.exe file to the installation location folder;
Right click on the new MA-6000.exe file and create shortcut to desktop;
If needs, copy new files to the Data folder and replace the existing file;
Launch software from the MA-6000 icon on desktop;
Create a new experiment and run it. Software updates successfully if system
does not prompt any error.
- 63 -
Chapter 8 Maintenance
8.1 Cleaning the Instrument
Cleaning the surface of the instrument
►The surface of the instrument should be regularly cleaned using a soft cloth
with small amounts of clean water, dry the surface after cleaning. Any spillage of
reagents on the instrument surface should be cleaned with 70% isopropyl alcohol.
Reaction well cleaning
►Dust or impurities in the reaction wells can affect PCR amplification and
fluorescence detection. Routine cleaning must be conducted every 3 months. A blow
bulb can be used to carefully blow away dust.
►When the machine is not in use, keep the cover closed to avoid dust entering
the reaction wells.
►Any leakage in the reaction well should be cleaned with 70% isopropyl
alcohol using a soft cloth.
Do not pour liquid into the reaction module(s) or inside the instrument. Do not
use corrosive solvents or organic solvents to wipe the machine.
- 64 -
►This instrument will be transported with carton and wooden box for long
distance.
►Inspect packaging integrity before unpacking, please contact us if any
abnormity.
►Unpacking the package and get the instrument. Inspect the machine and other
accessories according to the packing list. Please contact us if any abnormity.
- 65 -
Chapter 9 Information
MA-6000 Nameplate
- 66 -
Chapter 10 Troubleshooting
- 67 -
Update History
Effective
Version Update description Editor
Date
V1.0 Draft. Sun Chin 2016/3/1
Add product expected lifetime and update
V1.1 history, modify the contents in nameplate, add Sun Chin 2017/6/1
intended use.
- 68 -