ENG-0034-T33, Rev. 002
Installation Reference Document
Title: Detector Replacement Checklist
Bedford, MA 01730-1401 Subject: Selenia Image Receptor Replacement
USA
Number: MAN-00792 Revision: 002 Date: October 2009 Page: 1 of 3
Phone +1-800-760-8342 Release date: October 2009

PURPOSE
Provide a guide to replace an image receptor in a Selenia™ system.
OTHER DOCUMENTS
MAN-00745 Selenia Installation and Hardware Maintenance Manual
MAN-00746 Calibration and Software Maintenance Manual
TB-00106 AEC Calibration, AWS versions earlier than 3.2.3.0.
MAN-00799 AEC Calibration, AWS versions 3.2.3.0 to 3.3.1.1
MAN-01334 AEC Calibration, AWS versions 3.4.0.9 and above
DRC_FF_0017 Grid Troubleshooting Guide
DRC_FF_0018 Operating Temperature Guidelines for Detectors
TB-00146 Proper Alignment/Adjustment of Array Rotation Switch
TB-00200 Image Quality Issues
TB-00213 Detector Power Control / EL Frequency On Limit
SW-Compatibility Software Compatibility

TOOLS, EQUIPMENT, AND MATERIALS


Standard tool kit
ERGONOMIC AND SAFETY CONCERNS
Follow all building safety procedures.

Always observe Electrostatic Discharge (ESD) precautions when


working with electronics and electronic components.

PROCEDURE
‰ Record system information before changing the image receptor.
o Click on Info and About the Acquisition Station and record the following
information:
ƒ AWS Version ________________________
ƒ Readout Sequence ________________________
ƒ Detector Firmware ________________________
Installation Reference Document
Title: Detector Replacement Checklist
Bedford, MA 01730-1401 Subject: Selenia Image Receptor Replacement
USA
Number: MAN-00792 Revision: 002 Date: October 2009 Page: 2 of 3
Phone +1-800-760-8342 Release date: October 2009

‰ Gather and save the following information as described in TB-00200.


o m4_tube.cfg
o img_proc.cfg or img_proc.w.cfg or img_proc.mo.cfg
o Flat field images with large and small focal spot, and with two filters
o Gain calibration mode flat field image
o ACR QC phantom image with acrylic disk and record reported kV, mAs and
AGD at 4.5cm.
o SNR and CNR worksheets
o ‘‘For Processing’’ and ‘‘For Presentation’’ images

‰ Remove existing image receptor from gantry.


o See Maintenance Manual MAN-00745, Chapter 9, Section 5.2 for detailed
image receptor removal/replacement/shipping instructions.

‰ Verify the array software version is compatible with the AWS software version.
o See SW-Compatibility.pdf

‰ Preserve a copy of the Drapi.cfg file.


o Older versions of the install.sh file may possibly overwrite the present
Drapi.cfg file with an invalid version.

‰ Run detector installation software.


o Minimize the application.
o Open a terminal window.
o Type su <return>
o Type <password> <return>
o Place the Array Configuration CD that came with the detector in the CD drive.
o Type cd<space>/cdrom/cdrom1<return>
o Type ./install.sh<return>

‰ Verify that the detector C-arm rotation switch is adjusted properly per
TB-00146, ‘‘Proper Alignment Adjustment of Array Rotation Switch’’.

‰ Allow image detector temperature to stabilize one hour.


o See Field Flash DRC_FF_0018 for temperature recommendations.

‰ Perform gain calibration.


Installation Reference Document
Title: Detector Replacement Checklist
Bedford, MA 01730-1401 Subject: Selenia Image Receptor Replacement
USA
Number: MAN-00792 Revision: 002 Date: October 2009 Page: 3 of 3
Phone +1-800-760-8342 Release date: October 2009

‰ Perform AEC calibration.


o If AWS version is earlier than 3.2.3.0 use TB-00106.
o If AWS version is 3.2.3.0 to 3.3.1.1 use MAN-00799.
o If AWS version is 3.4.0.9 and above use MAN-01334.

‰ Perform grid calibration.


o See Field Flash DRC_FF_0017 for Grid Troubleshooting Guide.

‰ Configure proper dose and match image quality.


o See Technical Bulletin TB-00200.
o Phantom image quality evaluation --- Selenia Quality Control Manual
o Artifact evaluation --- Selenia Quality Control Manual

‰ Configure Detector Power and EL Frequency On Limit.


o See Technical Bulletin TB-00213.

‰ Arrange for a physicist evaluation of the system.

This completes the image receptor installation.