Professional Documents
Culture Documents
MAN-00792 Rev 002 Selenia Detector Replacement Checklist
MAN-00792 Rev 002 Selenia Detector Replacement Checklist
PURPOSE
Provide a guide to replace an image receptor in a Selenia™ system.
OTHER DOCUMENTS
MAN-00745 Selenia Installation and Hardware Maintenance Manual
MAN-00746 Calibration and Software Maintenance Manual
TB-00106 AEC Calibration, AWS versions earlier than 3.2.3.0.
MAN-00799 AEC Calibration, AWS versions 3.2.3.0 to 3.3.1.1
MAN-01334 AEC Calibration, AWS versions 3.4.0.9 and above
DRC_FF_0017 Grid Troubleshooting Guide
DRC_FF_0018 Operating Temperature Guidelines for Detectors
TB-00146 Proper Alignment/Adjustment of Array Rotation Switch
TB-00200 Image Quality Issues
TB-00213 Detector Power Control / EL Frequency On Limit
SW-Compatibility Software Compatibility
PROCEDURE
Record system information before changing the image receptor.
o Click on Info and About the Acquisition Station and record the following
information:
AWS Version ________________________
Readout Sequence ________________________
Detector Firmware ________________________
Installation Reference Document
Title: Detector Replacement Checklist
Bedford, MA 01730-1401 Subject: Selenia Image Receptor Replacement
USA
Number: MAN-00792 Revision: 002 Date: October 2009 Page: 2 of 3
Phone +1-800-760-8342 Release date: October 2009
Verify the array software version is compatible with the AWS software version.
o See SW-Compatibility.pdf
Verify that the detector C-arm rotation switch is adjusted properly per
TB-00146, ‘‘Proper Alignment Adjustment of Array Rotation Switch’’.