Manuscript received November 10, 2020; revised March 3, 2021; accepted April 21, 2021; date of publication April 28, 2021;
Digital Object Identifier (DOI): 10.1109/TEKNOKES.v1.i1.1
ABSTRACT A manual sphygmomanometer is an instrument used to measure blood pressure, and consists
of an inflatable cuff, a mercury manometer (or aneroid gauge) and an inflation ball and gauge. To assess the
condition, accuracy and safety of mercury and anaeroid sphygmomanometers in use in general practice and
to pilot a scheme for sphyg- momanometer maintenance within the district. Therefore, it must be calibrated
periodically. This Using the MPX 5050GP sensor as a positive pressure sensor. Requires a maximum
pressure of 300 mmHg. This tool is also equipped with a SD Card as external storage. The display used in
this module is TFT Nextion 2.8”. After conductings measurements of the three comparisons consisting of
Multifunction, DPM and mercury tensimeter to 6 times, the smallest result 0 mmHg and the largest results
251.52 mmHg. While the error in mercury tensimeter’s of leak test to module and rigel is 0.56% and
0.404%.
Positive Pressure". This tool has been equipped with Bluetooth Module HC-05 then displayed on the PC using
negative pressure measurements and unit selection at PLX DAQ.
positive pressure (Kusumadewi, 2020). However, this tool is
not equipped with data storage.
studies indicate that substantial proportions of
sphygmomanometers in general practices and hospitals
exhibit clinically significant (>3 mmHg) systematic pressure
errors and other faults (Turner, 2007).
Based on the above studies, the author will create a Digital
Pressure Meter tool using a power bank as a power supply
and equipped with external storage using an SD Card which
then the results of the measurement data will be displayed on
the tft Nextion 2,8”. This studies aims to add storage features
and display features on TFT screen. FIGURE 1. Diagram Block
A. EXPERIMENTAL SETUP
This study uses a measurement of a mercury
sphygmomanometer with a positive pressure setting;
0mmHg, 50mmHg, 100mmHg, 150mmHg, 200mmHg,
250mmHg. Data collection was repeated 6 times for
calibration mode.
1) Materials and tool
This study uses the MPX 5050 GP sensor to function as
blow pressure (‘Datasheet MPX 5050.pdf’, no date). This
Sensor has a pressure limit 7.25 psi (Yuwono et al., no date).
Different amplifier circuit using IC LM358 as a amplifier
module (‘datasheet LM358.pdf’, no date).Arduino Nano
microcontroller as a data processor. The Arduino Nano has a
number of facilities for communicating with a computer,
another Arduino, or other microcontrollers(‘Arduino Nano
V2.3 User Manual’, 2008). Data results will be saved via SD
Card V-Gen 4GB and SD Card Module . TFT Nextion as a
display (Wiki, 2018). A mercury Sphygmomanometer (GEA
Medical MI-2001) and Rigel Medical UNI-SiM are used as FIGURE 2. The Flowchart of Flowrate
module comparisons.
2) Experiment C. THE FLOWCHART
In this study, after the design was finished, the Digital The flowrate process is based on the flow chart as shown
Pressure Meter (Positive Pressure) output was tested with in Fig. 2. After initialization, a flow rate or occlusion will be
pressures of 0mmHg, 50mmHg, 100mmHg, 150mmHg, selection and the sensor will read the frequency of each
200mmHg, 250mmHg, then measured with a Tensimeter. droplet. After that Arduino will convert into flow rate value,
Each setting, was calculated to validate the results of this then the final result of the data will be sent to pc using
study. The leak test is carried out by pumping the Bluetooth HC05.
sphygmomanometer at a certain pressure (200 or 250) and
waiting for 1 minute, with a pressure drop of 15 mmHg D. ANALOG CIRCUIT
tolerance. 1) SENSOR CIRCUIT
A.
III. RESULT
In this study, the Digital Pressure Meter has been tested
using an Rigel Medical UNI-SiM and GEA medical MI- 4) THE LISTING PROGRAM ON DISPLAY
2001 Mercury Sphygmomanometer. The initial display when the tool is turned on, there are
2 selection modes, namely calibration mode and leak test
1) RESULT OF DIGITAL PRESSURE METER mode. List program for display in tft nextion was shown in
In this study, figure 5 . shows a series of Digital Pressure the Pseucode 3.
Meter final schematics/circuit Pseucode: 3 Program On Display
1. n4.val++ 6. tm1.en=0
2. if(n4.val>59) // n4=detik, n5=menit
7. }else
3. {
4. n5.val=1 8. {
5. n4.val=0 9. n5.val=0
6. tm1.en=0 10 }
7. }else
8. {
.
9. n5.val=0
10. } 7) MEASUREMENT RESULT
The measurement results on the module are compared
5) THE LISTING PROGRAM CALIBRATION AND LEAK with Rigel Medical UNI-SiM and GEA medical MI-2001
TEST Mercury Sphygmomanometer.
The listing program for calibration and leak test was a) Measurement with Rigel
shown in the Pseucode 4. This program determines the The module measurement using DPM Rigel Medical
sphygmomanometer being tested is suitable for use or not, UNI-SiM was performed in a Poltekkes Kemenkes
namely by providing good/leak information on the leak test Surabaya laboratory. These measurements were made to
Pseucode: 4 Program Calibration and Leak Test. know the truth of the DPM module. The measurement
1. if (mmhg<0) result was showed in Table I.
TABLE I MEASUREMENT WITH RIGEL
2. if (mmhg>0
3. if (mmhg>20 STDEV.
AVG UA
4. if (mmhg>60 P
5. String command2 ="dika.txt= 0 0.00 0
6. \""+String(mmhg2,2)+"\"";
50.4 0.16 0.07
7. Serial.print(command2);
8. Serial.write(0xff); 101.0 0.10 0.04
9. Serial.write(0xff); 149.9 0.19 0.08
10. Serial.write(0xff);
201.0 0.24 0.10
11. delay(200);
12 pin_value=mmhg2; 251.3 0.16 0.07
13. if(progress_value !=pin_value){
14. progress_value=pin_value; Table I. shows the results of the comparison of values
15. if(progress_value>165 between the module and the Rigel Medical UNI-SIM 6
16. Serial.print("t2.txt=\"BOCOR\""); times consisting of 3 measurements of rising pressure and 3
17. Serial.write(0xff); measurements of falling pressure.
18. Serial.write(0xff);
19. Serial.write(0xff); b) Measurement with Mercury Sphygmomanometer
20. } The module measurement using DPM Rigel Medical
21. else if(progress_value>185 UNI-SiM was performed in a Poltekkes Kemenkes
22. Serial.print("t2.txt=\"BAIK\""); Surabaya laboratory. These measurements were made to
23. Serial.write(0xff); know the truth of the DPM module. The measurement
24. Serial.write(0xff); result was showed in Table II.
25. Serial.write(0xff); TABLE II MEASUREMENT WITH MERCURY
26. } SPHYGMOMANOMETER
AVG STDEV.P UA
6) THE LISTING PROGRAM TIMER
The listing program for calibration and leak test was 0.00 0.00 0.00
shown in the Pseucode 5. This program uses milliseconds 50.64 0.26 0.11
as timer. The time it takes is 60 seconds.
Pseucode: 5 Timer program 100.68 0.24 0.10
Medical MI-2001 for 6 times consisting of 3 measurements 0.68% (100), 0.9% (150), 0.26% (200), 0.23% (250). While
of rising pressure and 3 measurements of falling pressure. for the measurement results of leak test error using Rigel
Medical UNI-SiM produce a value of 0.404% and using A
c) The Leakage Test mercury Sphygmomanometer (GEA Medical MI-
TABLE III THE LEAKAGE TEST 2001)produce a value of 0.56% eror.
200 mmHg/60s Rata-
Pengukur V. CONCLUSION
Rata
X1 X2 X3 X4 X5 X6
Overall, this research concludes that, a Digital Pressure
Rigel
Medical 196.67 197.14 197.3 196.6 196.9 196.87 196.9 Meter can be made to monitor the performance of the
UNI-SiM Sphygmomanometer, namely the output in the form of
198.2
Modul 197.47 197.96
3
197.26 197.5 197,66 197.7 Positive pressure. This research is built based on Arduino
Tensimeter 197 198 197 197 198 197 197.3 Nano, MPX5050GP sensor, differential circuit, 16x4
195.7 character LCD. After taking data using the Rigel tool, it can
Modul 195.66 196.23 195.7 197.5 197,66 196.2
4 be concluded that the connecting hose between the module
In Table III. Shows the measurement results of the output and the comparison hose greatly influences the
GEA medical MI-2001 Mercury Sphygmomanometer leak measurement results. The results of data storage on the SD
test on the module and GEA medical MI-2001 Mercury Card will be displayed in the form of a Notepad (txt) file.
Sphygmomanometer against the Rigel Medical UNI-SIM From the results of calibration measurements using Rigel
for 6 measurements.
Medical UNI-SiM, the smallest error is 0.00% and the largest
error is 1% and the error from the leak test is 0.404%. While
d) MPX5050GP Output Sensor Measurement
the measurement results using the GEA medical MI-2001
TABLE I OUTPUT SENSOR MEASUREMENT Mercury Sphygmomanometer, the smallest error is 0.00%
V. Supply Tekanan V. Out Sensor (volt) and the largest error is 1.28% and the error value from the
leak test is compared to the Rigel Medical UNI-SIM which is
(Volt) (mmHg) Naik Turun
0.56%.
5.00 0 0.17 0.17
REFERENCES
5.00 50 0.76 0.76 [1] Anna, N. L. (2017) ‘Kalibrator Tensimeter
5.00 100 1.35 1.36 Dilengkapai Dengan Thermohygrometer Berbasis
PC’, Kalibrator Tensimeter Dilengkapai Dengan
5.00 150 1.96 1.97 Thermohygrometer Berbasis PC Novella, p. 2.
5.00 200 2.56 2.57 [2] ‘Arduino Nano V2.3 User Manual’ (2008) Arduino,
pp. 1–5. Available at:
5.00 250 3.15 3.15 https://www.arduino.cc/en/Main/ArduinoBoardUno.
In TABLE IV. the output voltage produced by the [3] Cristina Silva, P. et al. (2018) ‘Analysis of Mercury
mpx5050gp sensor with a supply of 5 volts is between 0.17 Sphygmomanometers in A Hospital School-Analysis
volts with a pressure of 0 mmHg to 3.15 volts with a of Mercury Sphygmomanometers’, Journal of
pressure of 250 mmHg. Cardiology and Therapy, 5(1), pp. 697–700. doi:
10.17554/j.issn.2309-6861.2018.05.138.
IV. DISCUSSION [4] ‘datasheet LM358.pdf’ (no date).
Based on the pressure measurement the sensor circuit [5] ‘Datasheet MPX 5050.pdf’ (no date).
with DPM produces output according to the sensor [6] Ghareeb, S. et al. (2020) ‘Results of a project to
specifications with data collection of 6 sets. That is 0, 50, calibrate mercury sphygmomanometer blood pressure-
100, 150, 200, and 250 for positive pressure in units of measuring devices in Egypt’, Journal of Human
mmHg. Hypertension. doi: 10.1038/s41371-020-00424-0.
Results The output pressure sensor produces a variable [7] De Jong, W., Hartemann, P. and Thomsen, M. (2009)
and linear voltage each increase and decrease ranging from ‘Mercury Sphygmomanometers in Healthcare and the
0.15-3.15 V. The results of the DPM tool are also compared Feasibility of Alternatives’, SCENIHR (Scientific
with Rigel Medical UNI-SiM and GEA medical MI-2001 Committee on Emerging and Newly Identified Health
Mercury Sphygmomanometer tools for detecting Pressure. Risks), (September). Available at:
The value of the error obtained from measurements between http://ec.europa.eu/health/ph_risk/risk_en.htm.
Module and Rigel UNI-SiM starting from 0.00% (0), 0,8% [8] Knight, T. et al. (2001) ‘Sphygmomanometers in use
(50), 1% (100), 0.0007% (150), 0.5% (200), 0.52% (250). in general practice: An overlooked aspect of quality in
The value of the error obtained from measurements between patient care’, Journal of Human Hypertension, 15(10),
pp. 681–684. doi: 10.1038/sj.jhh.1001251.
Module and GEA medical MI-2001 Mercury
[9] Kusumadewi, K. D. (2020) ‘DPM Dua Mode
Sphygmomanometer starting from 0.00% (0), 1.28% (50),