1. Introduction
This document describes the optional DICOM WORKLIST installation procedure for the
WHA-200 Pleno.
With DICOM worklist option, it is able to obtain the patient information from DICOM devices
which is specified as DICOM WorklistSCP. Installation involves only connecting the LAN
cable and making settings in the service application.
1 M516-E324
<This page is intentionally left blank.>
M516-E324 2
2. Environment Information
To connect the WHA-200 Pleno to DICOM device, it must be connected to the hospital
network. Obtain the following network information from the hospital network administrator
and set them for the WHA-200 Pleno. Section 3 describes the setting method.
IP address :
Subnet mask :
Default gateway :
2) If the net mask default value is required for the IP address, set the net mask as
follows:
(1) IP address begins with 1 to 127, set the net mask to 255.0.0.0
(2) IP address begins with 128 to 191, set the net mask to 255.255.0.0
(3) IP address begins with 192 to 254, set the net mask to 255.255.255.0
In some cases, (1) is called Class A, (2) is called Class B, (3) is called Class C.
3 M516-E324
2.2. DICOM Environment Information
The following information must be exchanged with the connected DICOM device. Provide
the WHA-200 Pleno information to the connected DICOM device and then acquire the
required information of the connected DICOM device. Enter this information in the table
below. This information is required to make the settings in Section 3.
Item Description
WHA-200 Pleno AE title WHA_WorklistSCU
information IP address
Connected DICOM AE title
device information IP address
Port number
M516-E324 4
3. Setting Procedure
Insert the end of a network cable connected to the hospital network into the LAN cable inlet
at the rear of the monitor cart. (The user disconnects this cable when moving the monitor
cart.)
Follow the procedure below to set the network settings for the PC inside the monitor cart.
4) Right-click the Local Area Connection icon in the window. Select Properties in the
pop-up menu.
5 M516-E324
5) The following window opens. Select [Internet Protocol (TCP/IP)] and click
[Properties].
6) The Internet Protocol (TCP/IP) properties window opens. Select [Use the following
IP address] and enter “IP address”, “Subnet mask”, and “Default gateway”
information. Click [OK].
M516-E324 6
3.3. Environment Information Settings
File to be changed
[Worklist_Service_List]
SERVICES_SUPPORTED =3
SERVICE_1 = MODALITY_WORKLIST_FIND
SERVICE_2 = PERFORMED_PROCEDURE_STEP
SERVICE_3 = STANDARD_ECHO
[Worklist_Service_List]
SERVICES_SUPPORTED =2
SERVICE_1 = MODALITY_WORKLIST_FIND
SERVICE_2 =STANDARD_ECHO
# SERVICE_2 = PERFORMED_PROCEDURE_STEP
# SERVICE_3 = STANDARD_ECHO
7 M516-E324
3.3.2. Set Service application
Follow the procedure below to set the DICOM Environment Information from the service
application.
1) Click the [Worklist Devices] button in service application (Configuration window). The
following screen opens.
2) Enter an appropriate name for the connected device in [Name]. (This name is
displayed as the connected device name in clinical application.) Correctly enter in [IP
Addr], [AE Title], and [Port #] for the connected device. [Local AE Title] should contain
“WHA_WorklistSCU” by default. If not, enter “WHA_WorklistSCU” in this box as
follows.
It is not allowed to newly set “Empty” or the character string including “Empty”
at [Name] field in this screen. Be sure to set the name other than “Empty”.
M516-E324 8
3) When complete entering the data, click [Next]. The settings of connected device are
saved and this screen is closed to return to the service application screen.
Click [OK] to close the dialog and click again [Delete]. The following dialog
opens.
Click [OK] to close the dialog, then click [Next]. The deletion applies to the
settings.
When [Delete] is clicked and on deleting the setting on the screen, this
deletion actually does not apply to the settings yet unless [Next] is clicked. Be
sure to click [Next] to apply the deletion and close this screen.
9 M516-E324
<This page is intentionally left blank.>
M516-E324 10
4. Checking Worklist
1) Confirm that the connected DICOM device is ready to send patient information.
2) Run the clinical application to open [Worklist] tab of [Patient List] window and
confirm that the patient information can be queried properly. (For the details see
“M516-E013 OPESCOPE PLENO Image Processing Section Installation
Manual, 2.6 Import Patient Data Using DICOM Worklist Feature (Option)”)
11 M516-E324
<This page is intentionally left blank.>
M516-E324 12
5. Applying the Label
After checking worklist, stick the label on the monitor unit to indicate that the DICOM
WORKLIST option is installed. (If other labels have already been applied, stick the label at
any suitable position inside the frame.)
13 M516-E324
Revision history
First
Feb/22/2008 Newly created. Andoh, (Ueno)
printing