Professional Documents
Culture Documents
OVP - OQ Test Protocol: Overall Test Result PASSED
OVP - OQ Test Protocol: Overall Test Result PASSED
OVP - OQ Test Protocol: Overall Test Result PASSED
Resolution Test
Sensitivity Test
Measurem ent Region, Start: 6100.00 cm -1 Measurem ent Region, End: 5600.00 cm -1
Minim um S/N: 8000 Measured S/N: 12145.38
Linearity Test
Reproducibility Test
________________
Page 1 of 8 Page Signature
OVP - OQ Test Protocol
Resolution Test , Spectrum
1.201.15
Single channel
1.10 1.05
________________
Page 2 of 8 Page Signature
OVP - OQ Test Protocol
Sensitivity Test , Spectra
102.0101.0
Transmittance [%]
99.0 100.098.0
________________
Page 3 of 8 Page Signature
OVP - OQ Test Protocol
Energy Distribution Test , Spectrum
0.14
0.10
Single channel
0.06 0.02
________________
Page 4 of 8 Page Signature
OVP - OQ Test Protocol
Wavenumber Accuracy Test , Spectrum
0.090
0.085
Single channel
0.080 0.075
________________
Page 5 of 8 Page Signature
OVP - OQ Test Protocol
Alignment Test , Spectrum
0.2
0.1
0.0
-0.1
-0.2
________________
Page 6 of 8 Page Signature
OVP - OQ Test Protocol
Linearity Test , Spectrum
0.0006
Single channel
0.0004 0.0002
________________
Page 7 of 8 Page Signature
OVP - OQ Test Protocol
Reproducibility Test , Spectra
101.0100.5
Transmittance [%]
99.5 100.099.0
________________
Page 8 of 8 Page Signature