HEALTHCARE Quick Installation Guide

Imaging Services Installing the Digitizer at NX


A604993.4

Safety Connections (1) Perform connections at workstation. (2) Perform connections at digitizer.

WARNING:
Improper operation or service activities
may cause damage or injuries.

INSTRUCTIONS:
The following steps describe:
(1) Read the "Generic Safety Directions" Direct
document (See Agfa HealthCare Library) License network
Dongle
prior to attempting any operation, repair • Connections at NX workstation connection
or maintenance task on the equipment.
(2) Strictly observe all safety directions within • Connections at digitizer
the "Generic Safety Directions" and on the Mains HDMI to
product. DVI converter Mains
(3) Strictly observe all safety directions in (optional)
Mains
safety.cdr chapter 2 of the digitizer service manual. install_poster_02a.cdr
install_poster_003.cdr

Workstation Setup (3) Check NX SW version and switch on devices. (4) Load license. (5) Copy the NX Config Wizard to NX D:\drive.

A Copy the application license file (ALF) to the NX directory: A Insert the digitizer USB flash drive at the NX.
The following steps describe: C:\AGFA\Healthcare\NX\Configuration\Licensing
• Checking the NX software
• Loading the license file at NX
B Double-click the Licensing Manager 5
icon on the desktop.
License
• Running the NX config wizard on NX * Manager
C Select <Add License> and browse to the folder
where the ALF file is stored. Click: <open>. B Copy the folder \NX_Config_Wizard
from USB flash drive to drive D:\ on the NX.
C Re-insert the USB flash drive at the digitizer.
D Wait a few seconds until the message appears:
* config wizards for previous NX versions are The licenses have been updated. Press <OK>.
available from the Agfa HealthCare Library.
• Synchronizing the configuration database on NX E Close the License Manager. install_poster_05.cdr install_poster_06.cdr

(6) Run the Wizard: Set language, date and time. (7) Set keyboard and input language. (8) Define regional settings for new users. (9) Finalize the NX Config Wizard.

Region and Language


Keyboards and Languages
A Select Tab: Keyboards and Languages
A Double-click file: A Select Tab: Administrative A Wait approx. 5 minutes while NX is activating.
Region and Language
Administrative

D:\NX_Config_Wizard\ActivateNXxxxxxx.exe
Change keyboards...

B Change keyboards... Copy settings...

Warning B Next > Change system locale


B Next
OK Cancel Apply

OK Cancel Apply

C Choose Language for the NX GUI D Next > Welcome screen and new user accounts settings
Reboot now
Text Services and Input Languages
Current user

Default input Language


Main internal steps during
B Copy settings...
Welcome screen
English

Installed Services

Finish activation:
Ready to be activated E Activate C
New user accounts

DE German (Germany)
Keyboard
German
Add
C Add keyboard (if required) Welcome screen and system accounts
- Digitizer is configured on NX
- NX is set to
New user accounts

D Change Input Language (if required)


Date and Time OK Cancel

OK
Date and Time

Date and Time E IP Address 192.192.192.193


F Change time zone... OK Cancel Apply

(if changes have Welcome screen


Change date and time...

E Apply C D OK - User "Agfa" is configured


been made) New user accounts
Time zone

Change time zone...

G Change date and time... - Digitizer Error Viewer is installed


OK
Notify me when the clock changes

H OK Cancel Apply

install_poster_08.cdr
F OK install poster 09.cdr
install_poster_10.cdr install_poster_10.cdr

Page 1 of 2 Type 5151 / 100 / 110 / 200 / 210 / 300 / 310


03-2015 Document ID 34000041

Agfa Company Confidential Copyright © 2015 Agfa HealthCare N.V. Agfa HealthCare N.V., Septestraat 27, B-2640 Mortsel, Belgium
Quick Installation Guide
Installing the Digitizer at NX

(10) Login and stop NX. (11) Synchronize NX database. Installation Verification (12) Erase cassette.

A Login
User name Automatic A Double-click: "HostName Synchronizer" A Press the "Erase" button.
as service user Password
reboot The following steps describe: *
HostName
B Click: "Synchronize" Synchronizer
B Login User name
• Erasing a cassette
as service user Password

• Creating a flatfield image


C Double-click "Stop NX" C Wait a few seconds for message:
NX <<<Synchronization of hostname succeeded ! >>> • Evaluating the image quality
Stop NX B Insert largest available cassette
D Follow Instructions. Ignore message about "Image D Close the HostName Synchronizer. with black side facing up.
processing SW is not available". * For types 5151/300 & 310:
If NIP image plates are used, verification steps C Wait approx. 3 minutes until
NX
Stop NX

E Double-click: "Restart NX"


C:\Agfa\Healthcare\N\Bin\Service>echo off

*************************************************
ARE YOU SURE?
*************************************************
12 – 14 must be performed with both, NIP and status LED turns green.
Press any key to continue ... and press any key to continue. Restart NX
PIP cassettes. D Remove cassette.
After restart of NX, digitizer status LED turns green.
install_poster_22.cdr install poster 23.cdr install_poster_24.cdr

(13) Expose flatfield. (14) Insert and identify cassette. (15) Evaluate flatfield image quality. Installation Finalization
A Insert cassette with B Identify cassette.
A Insert CU Filter If image quality is not OK:
black side facing up.
B Expose Cassette: See digitizer service manual
System chapter "Troubleshooting".
Large Focus
Diagnosis
12 mAs Genrad The following steps describe:
75 kVp
Flat • Performing the digitizer installation wizard
C Rotate cassette by 180°
1,3 m D Expose again
Field • Creating a backup
200
E Remove CU Filter • Finalizing the installation
Collimated
Area
Black side
facing up C Wait approx. 3 minutes until image transfer is complete. install_poster_15.cdr
install_poster_013.cdr
install poster 25.cdr

(16) Connect to digitizer service GUI. (17) Perform digitizer installation wizard. (18) Create a backup. (19) Finalize Installation.

192.192.192.192
A Select Service Menu "Backup".
Digitizer Installation Wizards starts up automatically: The mandatory connection and configuration steps for a
Backup standalone system are finished.
B Click: "Save to USB"
Authentication Required

Refer to digitizer service manual chapter 4 for:


Next
User Name:

Password: Digitizer Settings Adapt local settings Complete Backup


OK Cancel
Backup single files only • Info about current system configuration status and for
• Following additional steps:
Start Destination Settings Do not change Next Select singe files (active only if "Backup single files
only" selected):
o Mandatory:
A Double-click the Internet Explorer icon. Start
Infocounter Logfiles - Perform tests according to national regulations
Site Specific Data Adapt Next Calibration Files Service Activity Log
- Train customer
B Enter "192.192.192.192" in the address window. o Optional, depending on the local situation:
Authentication Required
Installation Date Confirmation Confirm installation - Configure NX, e.g. change Autocropping setting
C Enter user name and password - Add additional components
Logout
User Name:

Password:
C Click:
obtained in the Web based Training. - Put NX and digitizer in the local network
OK Cancel install_poster_17.cdr
install_poster_16.cdr install_poster_18.cdr

Page 2 of 2 Type 5151 / 100 / 110 / 200 / 210 / 300 / 310


03-2015 Document ID 34000041