Professional Documents
Culture Documents
Bennett 840
Bennett 840
The trigger mechanism may be set as a pressure or flow trigger. Baseline flow is
automatically 1.5 LPM above the flow trigger setting which is adjustable from 1 -
20 LPM. The 840 also has automatic tube compensation (TC) to provide an
adjustable level of pressure support to overcome the resistance of breathing
through the endotracheal tube.
After connecting circuit and mains,
leave patient wye open
and power ON. If patient wye is
blocked during power ON,
ventilator will alarm and switch to
safety ventilation. After the 840 does
a power on self test, the bottom
screen will appear as seen on the
left. Same patient, new patient, or
short self-test (SST) may be selected
by touching screen and highlighting
item.
If SST is selected, operator has 4 seconds to press Test button on left side of
ventilation to activate SST. The 840 SST only requires the operator to 1) block
patient wye, 2) disconnect inspiratory limb of circuit and 3) disconnect
expiratory limb of circuit. During SST, the type of humidifier may be selected.
After touching New Patient, the screen goes to IBW. Touch IBW and adjust
body weight with knob, then press "Accept" located above knob and to the right
of CLEAR. IBW is given in pounds and kilograms. The 840 will automatically
set tidal volume to equal 7.25 ml/kg. All parameters can be adjusted.
Alarm limits are set by touching alarm box and adjusting knob. After all alarms
are adjusted, press Accept. Actual patient values are indicated by flags on the
left of the bar.
effective for all but large leaks. There is also a Disconnect % which adjusts the
% loss of the returned flow signaling a disconnect alarm. A setting of 50% will
mean the expiratory tidal volume must fall to half of the inspiratory volume
before an alarm is activated.