System Setting Version 1.

00

Ziehm Vision R

Change History:
Version Date Major changes / Remarks Author
Initial Version of this document, based on the document
“MTS-V2_Organ_V2.16”. This document is split into single
documents, one for each system.
Correction of typos and appearance in order to achieve a
better handing of the document.
New APRs and modifiers with according settings:
• Bone Extremities
• Bone Trunk
• Vascular Extremities
• Vascular Trunk
• Bolus Chase
• Low Dose
• DSA/RSA/MSA Extremities
• DSA/RSA/MSA Trunk
• DSA/RSA/MSA Extremities CO2
• DSA/RSA/MSA Trunk CO2
• Auto LPK
Adapting several dose settings for US
New Gamma settings with Offset at 50 kV and 100 kV
1.00 16.03.2011 SW
Settings for Auto LPK and Auto Gain
Adapting US kV/mA-curves of system without boost caps in
bone APRs
Adapting Black Level in several APRs
Removing settings from the document which are now fixed
programmed.
New RSA parameters:
• Noise Cut: 512
• Trace Threshold: 1023
• Trace Offset: 800
Adapting the RSA settings:
• dose values
• filter settings
• Max Frms/sec
• Grid off
• Gamma 5
Introduction of Max Frms/sec and Start Frms/sec

SYS_ZiehmVisionR_V1.00.doc Page 1/106


System Setting Version 1.00

Ziehm Vision R

Table of Content

1. Overview ................................................................................................. 3
2. General Requirements and Information................................................... 4
2.1 Dose and Dose Rate Measurements .............................................................. 4
3. Specification for Vision R ......................................................................... 5
3.1 Overall Settings.......................................................................................... 5
3.2 BONE Extremities ....................................................................................... 7
3.3 BONE Trunk ..............................................................................................13
3.4 HEART......................................................................................................20
3.5 ABDOMEN.................................................................................................27
3.6 Soft .........................................................................................................34
3.7 Vascular Trunk ..........................................................................................40
3.8 Vascular Extremities ..................................................................................46
3.9 Bolus Chase ..............................................................................................51
3.10 DSA Extremities ........................................................................................56
3.11 RSA Extremities ........................................................................................59
3.12 DSA Extremities (CO2)................................................................................63
3.13 RSA Extremities (CO2)................................................................................66
3.14 DSA Trunk ................................................................................................70
3.15 RSA Trunk ................................................................................................73
3.16 DSA Trunk (CO2) .......................................................................................77
3.17 RSA Trunk (CO2) .......................................................................................80
3.18 ENDO.......................................................................................................84
3.19 U R O (MTS 2.16) ......................................................................................90
3.20 URO-Storz (MTS 2.16) ...............................................................................97
4. Risk Prevention ................................................................................... 102
5. Installation Requirements ................................................................... 103
6. Requirements from Homologations and Standards .............................. 104
7. Production Requirements and Limitations ........................................... 105
8. Appendix ............................................................................................. 106
8.1 Scope and Readership ..............................................................................106
8.2 Definitions and Abbreviations ....................................................................106

SYS_ZiehmVisionR_V1.00.doc Page 2/106


System Setting Version 1.00

Ziehm Vision R

1. Overview
This document describes all the parameters that have to be adjusted at the system Ziehm Vision R.
These values are set in the tab cards “Dose 1”, “Dose 2”, “Dose 3” in the service settings; please
see graphics below as an example. All values are defined and may not be changed except those
values, which are denoted with tba (to be adjusted).

SYS_ZiehmVisionR_V1.00.doc Page 3/106


System Setting Version 1.00

Ziehm Vision R

2. General Requirements and Information


An empty field means this field is not applicable (NA).
Rest of world ROW includes EU.

The settings described in this document are valid ONLY for new-build Ziehm Vision R systems or
installed systems updated with the according update kit for 5.30.0.
This SYS is not valid in general for the installed base (please use the settings described in the
document “MTS-V2OrganAdjustment_Vision-V2_2 16.doc” instead).
The MINIMUM REQUIREMENTS to enable the settings and get the best results regarding image
quality, described in that document, are the following:
• Software Version: ≥ 5.30.0 SV
• PPC Boot Software: ≥ 2.00
• Firmware Version (U622): = K13
• Firmware Version (U544): ≥ K24 (InMux: 4, Noisefilter: 22, ORCA3: 6, ORCA4: 24)
• Firmware Version (U664): ≥ K51 (InMux: 5, Noisefilter: 22, ORCA3: 6, ORCA4: 51)

2.1 Dose and Dose Rate Measurements

Dose and dose rate measurements are done by 25 pulses/sec and with grid functionality disabled.

For US:
Dose and dose rate measurements are done by 30 pulses/sec and with grid functionality disabled.

SYS_ZiehmVisionR_V1.00.doc Page 4/106


System Setting Version 1.00

Ziehm Vision R

3. Specification for Vision R


3.1 Overall Settings
Max Dose:
For all Fluoro and LPK modes: < 88mG/min (ROW)
For all Fluoro and LPK modes: < 10 R/min (US, non HLC)
For US HLC mode: < 20 R/min (US, HLC)

3.1.1 Overview Dose Settings


Dose and dose rate measurements are done by 25 pulses/sec (US: 30 pulses/sec) and with grid
functionality disabled.

3.1.2 Default Modifier RSA


The default modifier for RSA is all RSA modes “Fluoro”.

3.1.3 Table „Grid“


Grid 31 23 15 10 Res
Grid Regulation Parameters
Motion Level Min 4 4 4 4
Motion Level Max 100 100 100 100
Motion Level Step 2 2 2 2
Motion Count 7500 7500 7500 7500
Motion Hysteresis 500 500 500 500

Auto Dose Reduction Max Pulse tba tba tba tba tba 200 - 240V
pause
Auto Dose Reduction Max Pulse tba tba tba tba tba USA
pause
Screen 15 15 15 15 0…15

kV reduction tba tba tba tba % (max 2%)


mA reduction / addition tba tba tba tba % (max 2%)

Cooling Water Pump


On Temp. 32 32 32 32 °C
Off Temp. 30 30 30 30 °C

Remark:
The kV reduction and the mA reduction have to be adjusted with every single system. During a
pulse down sequence the image has to keep the same impression regarding brightness and
contrast.

SYS_ZiehmVisionR_V1.00.doc Page 5/106


System Setting Version 1.00

Ziehm Vision R

3.1.4 Table „Collimator“


Collimator 12” II 9” II Res

Vignetting 50 30 0-99 %
HEAL 50 50 0-99 %

Arc Protection
Arc Protection on Off Off
Arc Protection kV na na 40-120
Time1 na na 0-3000 s
Step1 na na 1(00)-30(00) V
Time2 na na 0-3000 s
Step2 na na 1(00)-30(00) V

3.1.5 Table „Radiography“


Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V w/o
Cap’s Cap’s

kV mA kV mA kV mA kV mA
1 40 64 40 18 40 50 40 18
2 50 56 50 18 50 44 50 18
3 60 52 60 20 60 40 60 20
4 70 48 120 17 70 37 120 13,7
5 80 46 0 0 80 35 0 0
6 90 44 0 0 90 33 0 0
7 100 42 0 0 100 32 0 0
8 110 41 0 0 110 32 0 0
9 120 40 0 0 120 31 0 0

Short shot < 100ms


kV range 40 110 40 110 40 100 40 100

mA 65 19 60 15

SYS_ZiehmVisionR_V1.00.doc Page 6/106


System Setting Version 1.00

Ziehm Vision R

3.2 BONE Extremities


3.2.1 Adjustments

3.2.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,43 0,43 0,43 0,43 0,43 µG/sec
Dose rate US 0,8 1,1 1,6 1,1 1,6 2,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 5 5 5 5 5 5 %
White Level On On On On On On On/Off
White Level 95 95 95 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Metall Peak tba tba tba tba tba tba

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 8 8 8 8 8 8 0-9
Offs. Lev. (50kV) On On On On On On On/Off
Offs. Lev. (50kV) 5 5 5 5 5 5 0-100%
Offs. Lev.(100kV) On On On On On On On/Off
Offs. Lev.(100kV) 25 25 25 5 5 5 0-100%
Dynamic 10 10 10 10 10 10 5-100%

SYS_ZiehmVisionR_V1.00.doc Page 7/106


System Setting Version 1.00

Ziehm Vision R

Filter M0 M1 M2
Nr. 1 2 3 1 2 3 1 2 3
2 5 10 2 5 10 2 5 10
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 1 1 2 1 1 2 1 1
Fluoro 3 1 1 3 1 1 3 1 1
LPK 3 0 1 3 0 1 3 0 1
SSHT 2 1 1 2 1 1 2 1 1
Low D 3 1 1 3 1 1 3 1 1
REP 1 0 1 1 0 1 1 0 1

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 3,0
3 50 2,5 50 4,0
4 60 5,0 60 10,0
5 70 9,0 70 13,7
6 80 8,8 80 12,5
7 120 6,6 100 11,0
8 0 0 120 10,0 Max Dose!
9 0 0 0 0
Start (kV/mA) 60 5,0 60 10,0
Max Pulse Rate 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 8/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain Off Off
Threshold kV
Max Gain
Auto LPK off off
off kV
on kV

Grid
White Clipping
Offset 700 650 650 700 650 650 0…4095
Difference 1700 1600 1600 1700 1600 1600 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection On On On On On On On/Off
Auto Dose Red. On On On On On On On/Off
Motion Sensi. tba tba tba tba tba tba 0…4095

3.2.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,0 2,0 2,0 2,0 2,0 µG/sec
Dose rate US 1,4 1,5 1,8 1,7 2,0 2,2
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Dose Gain US 0 0 0 0 0 0 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 6 7 7 6 7 7 %
White Level 95 95 80 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V1.00.doc Page 9/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
120 45,0
120 45 46 48 50 52 54 56
120 62,5 62,5 58 60 62,
5

Start (kV/mA) 80 60 80 60

Max Frms/s 25 25 15 30
Pulse Length 20 10 16 8 10, 16, 20, 30
in ms
Start Frms/s 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 10/106


System Setting Version 1.00

Ziehm Vision R

3.2.1.3 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 6 7 7 6 7 7 %
White Level 95 95 80 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8
9
Start 60 60 60 60
Max Frms/s 25 25 30 30 1-25 / 30
Pulse Length in ms 20 16 10 8 10, 16, 20, 30
Start Frms/s 25 25 30 30 1-25 / 30

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity na na na na

SYS_ZiehmVisionR_V1.00.doc Page 11/106


System Setting Version 1.00

Ziehm Vision R

3.2.1.4 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,27 0,27 0,27 0,27 0,27 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 20 25 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 5 5 5 5 5 5 %
White Level on on on on on on 0-200
White Level 95 95 95 95 95 95 0-200
Metall Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0,1 40 0,1
2 80 0,1 80 0,1
3 120 10 120 10
4
5
6
7
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 10 10 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 12/106


System Setting Version 1.00

Ziehm Vision R

3.3 BONE Trunk


3.3.1 Adjustments

3.3.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,52 0,60 1,0 µG/sec
Dose rate USA 0,8 1,1 1,6 1,1 1,6 2,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 5 5 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 3 7 7 7 7 7 %
White Level On On On On On On On/Off
White Level 100 95 95 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 7 7 7 7 7 7 0-9
Level On On On On On On On/Off
Offs. lev. (50kV) 5 5 5 5 5 5 0-100%
Level On On On On On On On/Off
Offs. lev.(100kV) 25 25 25 25 25 25 0-100%
Dynamic 10 10 10 10 10 10 5-100%

Filter M0 M1 M2
Nr. 1 2 3 1 2 3 1 2 3
4 8 12 4 8 12 4 8 12
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 1 1 2 1 1 2 1 1
Fluoro 3 1 1 3 1 1 3 1 1
LPK 3 0 1 3 0 1 3 0 1
SSHT 2 1 1 2 1 1 2 1 1
REP 1 0 1 1 0 1 1 0 1

SYS_ZiehmVisionR_V1.00.doc Page 13/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 3,0
3 50 2,5 50 4,0
4 60 5,0 60 10,0
5 70 9,0 70 13,7
6 80 8,8 80 12,5
7 120 6,6 100 11,0
6 0 0 120 10,0
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length in ms 30 20 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain on On On/off
Threshold 100 100 kV
Max Gain 4 4
Auto LPK On On
off 80 80
on 100 100

Grid
White Clipping
Offset 700 700 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection off off off off off off On/Off
Auto Dose Red. off off off off off off On/Off
Motion Sensi. 500 600 660 480 600 600 0…4095

SYS_ZiehmVisionR_V1.00.doc Page 14/106


System Setting Version 1.00

Ziehm Vision R

3.3.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,0 2,0 2,0 2,0 2,0 µG/sec
Dose rate USA 1,4 1,5 1,8 1,7 2,0 2,2 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 6 7 7 6 7 7 %
White Level 95 95 80 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60 50,0


Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length 20 10 16 8 10, 16, 20, 30
in ms
Start Frms/s 25 25 15 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pause Burst Pause Bursts Pause

SYS_ZiehmVisionR_V1.00.doc Page 15/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.3.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Iris Camera 100 100 100 100 100 100 %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 6 8 7 8 7 7 %
White Level 95 95 80 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8
9
Start 60 60
Max Frms/sec 25 30 10, 16, 20,
30
Pulse Length in ms 30 30 10, 16, 20,
30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 16/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.3.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 6 7 7 6 7 7 %
White Level 95 95 80 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8
9
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 16 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V1.00.doc Page 17/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0 0

3.3.1.5 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,27 0,27 0,27 0,27 0,27 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 20 25 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 5 5 5 5 5 5 %
White Level 95 95 95 95 95 95 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0,1 40 0,1
2 70 0,1 70 0,1
3 120 6,5 120 6,5
4
5
6
7
8
9
Start 60 60
Max Frms/sec 25 30 1-25 / 30
Pulse Length in ms 16 10 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 18/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 19/106


System Setting Version 1.00

Ziehm Vision R

3.4 HEART
3.4.1 Adjustments

3.4.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 25 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 15 15 15 15 15 15 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 200 - 11 11 11 11 11 11 %
240V
Black Level US 15 15 15 15 15 15 %
White Level On On On On On On On/Off
White Level 90 90 90 90 90 90 0-200
Metal Black Level tba tba tba tba tba tba
Metall Peak tba tba tba tba tba tba
Windowing
Window 90 90 90 90 90 90 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Level On On On On On On
Offs. lev. (50kV) 2 2 2 2 2 2
Level Off Off Off Off Off Off
Offs. lev.(100kV)
Dynamic 10 10 10 10 10 10 5-100%

Filter M0 M1 M2
Nr. 1 2 3 1 2 3 1 2 3
2 3 4 2 3 4 2 3 4 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 0 0 1 0 0 1 0 0 1
Fluoro 1 0 1 1 0 1 1 0 1
LPK 1 0 1 1 0 1 1 0 1
SSHT 0 0 1 0 0 1 0 0 1
REP 0 0 1 0 0 1 0 0 1

SYS_ZiehmVisionR_V1.00.doc Page 20/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 50 3,0 55 2,0
3 70 6,0 65 8,7
4 85 15,0 90 8,0
5 100 14,0 120 6,8
6 120 13,0 0 0
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

Auto Gain/Auto LPK


Auto Gain off off
Threshold kV
Max Gain
Auto LPK on on
off 80 80
on 110 110

SYS_ZiehmVisionR_V1.00.doc Page 21/106


System Setting Version 1.00

Ziehm Vision R

3.4.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose gain 2,0 2,6 3,0 2,0 2,6 3,0 µG/sec
Dose gain USA 1,2 1,6 2,0 1,4 1,8 2,3 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 25 %
Dose Gain tba tba tba tba tba tba 0-9
Dynamic 100 100 100 100 100 100 %
Peak Part 15 15 15 15 15 15 %

Video
Black Level 11 11 11 11 11 11 %
White Level 95 95 95 95 95 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/sec 25 25 15 30 1-25 / 30
Pulse Length in 20 10 16 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 15 30 1-25 / 30
Burstmode

SYS_ZiehmVisionR_V1.00.doc Page 22/106


System Setting Version 1.00

Ziehm Vision R

Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity na na na na

3.4.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 25 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 15 15 15 15 15 15 %

Video
Black Level on on on on on on %
Black Level 200 - 11 11 11 11 11 11
240V
Black Level US 15 15 15 15 15 15 %
White Level on on on on on on 0-200
White Level 95 95 95 95 95 75
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V1.00.doc Page 23/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 3,0 50 2,5
4 70 6,0 60 5,0
5 85 15,0 70 9,0
6 100 14,0 80 9,0
7 120 13,0 120 11,2 -
8,0
8 0 0 0 0
9 0 0 0 0
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 0 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res


Auto Gain / Auto Dose
Auto Gain Off Off Off Off On/Off
Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.4.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 25 25 25 25 25 25 %

Video
Black Level 200 - 11 11 11 11 11 11 %
240V
Black Level US 17 17 17 17 17 17 %
White Level 95 95 95 95 95 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V1.00.doc Page 24/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V w/o
Cap’s Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 69,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/s 25 25 30 30 1-25 / 30
Pulse Length in ms 20 16 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.4.1.5 Low Dose (new APR)


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,27 0,27 0,27 0,27 0,27 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 20 30 30 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 15 15 15 15 15 15 %

Video
Black Level 11 11 11 11 11 11 %
White Level 95 95 95 95 95 95 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V1.00.doc Page 25/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0,1 40 0,1
2 70 0,1 70 0,1
3 120 8 120 8
4
5
6
7
8
9
Start 60 60
Max Frms/s 25 30 1-25 / 30
Pulse Length in ms 16 10 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 26/106


System Setting Version 1.00

Ziehm Vision R

3.5 ABDOMEN
3.5.1 Adjustments

3.5.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µGy/s
Dose rate USA 0,8 1,0 1,2 1,0 1,2 1,4 µGy/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 9 9 9 9 9 9
Black Level US 3 4 4 3 6 6 %
White Level On On On On On On On/Off
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba
Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Gamma USA 7 7 7 7 7 7 0-9
Level On On On On On On On/Off
Offs. lev. (50kV) 5 5 5 5 5 5 0-100%
Offs. lev. (50kV) 8 8 8 8 8 8 0-100%
USA
Level Off Off Off Off Off Off On/Off
Offs. lev.(100kV l 0-100%
Dynamic 10 10 10 10 10 10 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 4 8 2 4 8 2 4 8 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 2 2 0 2 2 0 2
Fluoro 2 0 2 2 0 2 2 0 2

SYS_ZiehmVisionR_V1.00.doc Page 27/106


System Setting Version 1.00

Ziehm Vision R

LPK 3 0 1 3 0 1 3 0 1
SSHT 2 0 1 2 0 1 2 0 1
Low D. 2 0 8 2 0 8 2 0 8
REP 1 0 1 1 0 1 1 0 1

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,0
2 50 5,0 50 5,0
3 65 8,7 65 8,7
4 90 8,0 80 7,5
5 120 6,6 100 6,3
6 0 0 120 5,5
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/s 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. 400 400 400 400 400 400 0…4095

Auto Gain/Auto LPK


Auto Gain on on
Threshold 100 100 kV
Max Gain 3 3
Auto LPK on on
off 80 80
on 100 100

SYS_ZiehmVisionR_V1.00.doc Page 28/106


System Setting Version 1.00

Ziehm Vision R

3.5.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,6 3,0 2,0 2,6 3,0 µGy/S
Dose rate USA 1,8 2,2 2,8 2,2 2,6 3,0 µGy/S
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 4 4 4 4 4 4 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60 50,0

Max Frms/s 25 25 30 30 1-25 / 30


Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 29/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.5.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µGy/S
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on
Black Level 9 9 9 9 9 9
Black Level US 3 7 6 4 6 4 %
White Level on on on on on on 0-200
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8 0 0 0 0
9 0 0 0 0
Start 60 60
Max Frms/s 25 30 1-25 / 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

SYS_ZiehmVisionR_V1.00.doc Page 30/106


System Setting Version 1.00

Ziehm Vision R

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.5.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba Tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba Tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 6 9 8 6 6 6 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba Tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 16 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V1.00.doc Page 31/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.5.1.5 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,27 0,27 0,27 0,27 0,27 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 20 25 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 9 9 9 9 9 9 %
White Level on on on on on on 0-200
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0,1 40 0,1
2 80 0,1 80 0,1
3 120 6,5 120 6,5
4
5
6
7
8
9
Start 60 60
Max Frms/sec 25 30
Pulse Length in ms 16 10 10, 16, 20,
30
Start Frms/sec s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 32/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 33/106


System Setting Version 1.00

Ziehm Vision R

3.6 Soft
3.6.1 Adjustments

3.6.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Dose rate USA 0,5 0,6 0,9 0,6 0,9 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 15 20 20 15 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 8 8 8 8 8 8 %
White Level On On On On On On On/Off
White Level 95 95 95 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Metall Peak tba tba tba tba tba tba

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 7 7 7 7 7 7 0-9
Level Off Off Off Off Off Off On/Off
Level 0 0 0 0 0 0 0-100%
Dynamic 10 10 10 10 10 10 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 4 6 2 4 6 2 4 6 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 0 4 1 0 4 1 0 4
Fluoro 1 0 4 1 0 4 1 0 4
LPK 2 0 8 2 0 8 2 0 8
SSHT 1 0 4 1 0 4 1 0 4
Low D. 2 0 4 2 0 4 2 0 4
REP 1 0 1 1 0 1 1 0 1

SYS_ZiehmVisionR_V1.00.doc Page 34/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,5 40 1,0
2 60 1,5 50 1,2
3 120 6,6 58 1,5
4 0 0 90 8,0
5 0 0 120 6,8
6 0 0 0 0
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 58
Max Frms/sec 25 30 1-25 / 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

Auto Gain/Auto LPK


Auto Gain on on
Threshold 100 100 kV
Max Gain 3 3
Auto LPK on on
off 80 80
on 100 100

SYS_ZiehmVisionR_V1.00.doc Page 35/106


System Setting Version 1.00

Ziehm Vision R

3.6.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,6 3,0 2,0 2,6 3,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 15 20 20 15 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 9 9 9 9 9 9 %
White Level on on on on on on 0-200
White Level 95 95 95 95 95 95 0-200
Metall Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,5

Start (kV/mA) 60 60 60 60

Max Frms/sec 25 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 30 30 1-25 / 30

Burstmode
SYS_ZiehmVisionR_V1.00.doc Page 36/106
System Setting Version 1.00

Ziehm Vision R

Frms/sec Burs Pause Burst Pauses Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity na na na na

3.6.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 15 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 8 8 8 8 8 8 %
White Level on on on on on on 0-200
White Level 95 95 95 95 95 95 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8
9
Start 60 60
Max Frms/s 25 30
Pulse Length in ms 30 30 10, 16, 20, 30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 37/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.6.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 15 20 20 15 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9 %
White Level 95 95 95 95 95 95 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 16 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off


Auto Dose Reduction on Off Off Off Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 38/106


System Setting Version 1.00

Ziehm Vision R

3.6.1.5 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,27 0,27 0,27 0,27 0,27 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 15 20 20 15 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on
Black Level 8 8 8 8 8 8 %
White Level on on on on on on
White Level 95 95 95 95 95 95 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0,1 40 0,1
2 80 0,1 80 0,1
3 120 4 120 4
4
5
6
7
8
9
Start 60 60
Max Frms/sec 25 30
Pulse Length in ms 16 10 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0
SYS_ZiehmVisionR_V1.00.doc Page 39/106
System Setting Version 1.00

Ziehm Vision R

3.7 Vascular Trunk


3.7.1 Adjustments

3.7.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Dose rate USA 0,8 1,0 1,2 1,0 1,2 1,4 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 9 9 9 9 9 9
Black Level US 3 4 4 3 6 6 %
White Level On On On On On On On/Off
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba
Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Offs. Lev. (50kV) On On On On On On
Offs. Lev. (50kV) 5 5 5 5 5 5
Offs. Lev. On On On On On On
(100kV)
Offs. Lev. 25 25 25 25 25 25
(100kV)
Dynamik 10 10 10 10 10 10

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 3 6 2 3 6 2 3 6 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 1 2 0 1 2 0 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 3 0 1 3 0 1 3 0 1

SYS_ZiehmVisionR_V1.00.doc Page 40/106


System Setting Version 1.00

Ziehm Vision R

SSHT 2 0 1 2 0 1 2 0 1
Low D
REP

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 50 5,0 50 5,0
3 65 8,7 65 8,7
4 90 8,0 80 7,5
5 120 6,6 100 6,3
6 0 0 120 5,5
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/s 25 30
Pulse Length in ms 0 0 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain on on
Threshold 100 100 kV
Max Gain 2 2
Auto LPK On on
off 80 80
on 100 100

Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

SYS_ZiehmVisionR_V1.00.doc Page 41/106


System Setting Version 1.00

Ziehm Vision R

3.7.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,6 3,0 2,0 2,6 3,0 µG/sec
Dose rate USA 1,8 2,2 2,8 2,2 2,6 3,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 4 4 4 4 4 4 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60

Max Frms/sec 25 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 42/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.7.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Dose rate USA 0,8 1,0 1,2 1,0 1,2 1,4 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on
Black Level 9 9 9 9 9 9
Black Level US 3 7 6 4 6 4 %
White Level on on on on on on 0-200
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8 0 0 0 0
9 0 0 0 0
Start 60 60
Max Frms/sec 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

SYS_ZiehmVisionR_V1.00.doc Page 43/106


System Setting Version 1.00

Ziehm Vision R

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain on on on on on On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.7.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba Tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba Tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 6 9 8 6 6 6 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba Tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 16 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V1.00.doc Page 44/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 45/106


System Setting Version 1.00

Ziehm Vision R

3.8 Vascular Extremities


3.8.1 Adjustments

3.8.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,43 0,43 0,43 0,43 0,43 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off
Video
Black Level On On On On On On On/Off
Black Level 9 9 9 9 9 9
Black Level US 3 4 4 3 6 6 %
White Level On On On On On On On/Off
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba
Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Offs. Lev. (50kV) On On On On On On
Offs. Lev. (50kV) 5 5 5 5 5 5
Offs. Lev. 100kV) On On On On On On
Offs. Lev. 100kV) 25 25 25 25 25 25
Dynamik 10 10 10 10 10 10

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 3 5 2 3 5 2 3 5 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 2 2 0 2 2 0 2
Fluoro 2 0 2 2 0 2 2 0 2
LPK 3 0 1 3 0 1 3 0 1
SSHT 2 0 1 2 0 1 2 0 1
Low D na na na na na na na na na
REP na na na na na na na na na

SYS_ZiehmVisionR_V1.00.doc Page 46/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 50 5,0 50 5,0
4 90 8,0 80 7,5
5 120 6,6 100 6,3
6 0 0 120 5,5
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain off off
Threshold kV
Max Gain
Auto LPK off off
off
on

Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

SYS_ZiehmVisionR_V1.00.doc Page 47/106


System Setting Version 1.00

Ziehm Vision R

3.8.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,6 3,0 2,0 2,6 3,0 µG/sec
Dose rate USA 1,8 2,2 2,8 2,2 2,6 3,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 4 4 4 4 4 4 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V1.00.doc Page 48/106


System Setting Version 1.00

Ziehm Vision R

Burstmode
Frms/sec Burst Pause Burst Pauses Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.8.1.3 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba Tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba Tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 6 9 8 6 6 6 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba Tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 16 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V1.00.doc Page 49/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 50/106


System Setting Version 1.00

Ziehm Vision R

3.9 Bolus Chase


3.9.1 Adjustments

3.9.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,43 0,43 0,43 0,43 0,43 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off
Video
Black Level On On On On On On On/Off
Black Level 9 9 9 9 9 9
Black Level US 3 4 4 3 6 6 %
White Level On On On On On On On/Off
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba
Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Offs. Lev. (50kV) On On On On On On
Offs. Lev. (50kV) 5 5 5 5 5 5
Offs. Lev. 100kV) On On On On On On
Offs. Lev. 100kV) 25 25 25 25 25 25
Dynamik 10 10 10 10 10 10

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 3 5 2 3 5 2 3 5 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 0 0 0 0 0 0 0 0 0
Fluoro 0 0 0 0 0 0 0 0 0
LPK 0 0 0 0 0 0 0 0 0
SSHT 0 0 0 0 0 0 0 0 0
Low D na na na na na na na na na
REP na na na na na na na na na

SYS_ZiehmVisionR_V1.00.doc Page 51/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 50 5,0 50 5,0
4 90 8,0 80 7,5
5 120 6,6 100 6,3
6 0 0 120 5,5
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/sec s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain off off
Threshold kV
Max Gain
Auto LPK off off
off
on

Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

SYS_ZiehmVisionR_V1.00.doc Page 52/106


System Setting Version 1.00

Ziehm Vision R

3.9.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,6 3,0 2,0 2,6 3,0 µG/sec
Dose rate USA 1,8 2,2 2,8 2,2 2,6 3,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 4 4 4 4 4 4 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60 50,0


Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V1.00.doc Page 53/106


System Setting Version 1.00

Ziehm Vision R

Burstmode
Frms/sec Burst Pause Burst Pauses Burst Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.9.1.3 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba Tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba Tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 9 9 9 9 9 9
Black Level US 6 9 8 6 6 6 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba Tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 60 60 60 60
Max Frms/sec 25 25 30 30 1-25 / 30
Pulse Length in ms 20 16 10 8 10, 16, 20, 30
Start Frms/sec 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V1.00.doc Page 54/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 55/106


System Setting Version 1.00

Ziehm Vision R

3.10 DSA Extremities


3.10.1 Adjustments

3.10.1.1 Fluoro
Dose 1 12” II 9” II
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
White Level On On On On On On On/Off
Windowing
Window 45 45 45 45 45 45 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 3 3 3 3 3 3 0-9
Off. 50kV On On On On On On On/Off
Off. Lev. 50kV 1 1 1 1 1 1 0-100%
Off. 100kV On On On On On On On/Off
Off. Lev. 100kV 7 7 7 7 7 7 0-100%
Dynamic 100 100 100 100 100 100 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
7 8 12 7 8 12 7 8 12 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 3 0 1 3 0 1 3 0 1
Fluoro na na na na na na na na na
LPK na na na na na na na na na
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

SYS_ZiehmVisionR_V1.00.doc Page 56/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 100 100 0…100
Black Level Reduction
low 50 50 kV
Dose low 30 30 0…100%
Video low 15 15 0…100%
high 100 100 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%

Grid
Offset 750 750 750 750 750 750 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off

3.10.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 3,0 3,6 4,0 3,0 3,6 4,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 17 17 17 17 17 %
White Level 90 90 90 90 90 90 0-200
Metal Black Level 17 17 17 17 17 17 %
Peak 95 95 95 95 95 95 %

SYS_ZiehmVisionR_V1.00.doc Page 57/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,0 40 1,0
2 50 4,0 50 4,0
3 60 16,0 60 16,0
4 70 30,0 70 30,0
5 80 42,0 80 42,0
6 90 47,0 90 47,0
5 100 55,0 100 55,0
7 110 56,0 110 56,0
7 120 62,5 120 62,5

Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length 20 216 10, 16, 20, 30
in ms
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Bursts Pauses
s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction Off Off Off Off On/Off

SYS_ZiehmVisionR_V1.00.doc Page 58/106


System Setting Version 1.00

Ziehm Vision R

3.11 RSA Extremities


3.11.1 Adjustments

3.11.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,0 1,2 1,6 1,2 1,6 1,6 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 17 17 17 17 17 17
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Metall Black 17 17 17 17 17 17
Peak 95 95 95 95 95 95
Windowing
Window 65 65 65 65 65 65 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Off. 50kV Off Off Off Off Off Off On/Off
Off. Lev. 50kV 0 0 0 0 0 0 0-100%
Off. 100kV Off Off Off Off Off Off On/off
Off. Lev. 100kV 50 50 50 50 50 50 0-100%
Dynamic 100 100 100 100 100 100 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
3 4 5 3 4 5 3 4 5 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 2 1 1 2 1 1 1 1
Fluoro 2 1 1 2 1 1 2 1 1
LPK 0 0 0 0 0 0 0 0 1
SSHT 0 0 0 0 0 0 0 0 1
Low D 0 0 0 0 0 0 0 0 1
REP 0 0 0 0 0 0 0 0 1

SYS_ZiehmVisionR_V1.00.doc Page 59/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 60 10,0 60 10,0
3 80 15,0 80 15,0
4 100 15,0 100 15,0
5 120 13,0 120 13,0

Start (kV/mA) 60 60
Max Frms/sec 25 30
Pulse Length in ms 20 20 10, 16, 20,
30
Start Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 0 0 0…100%
Video low 0 0 0…100%
high 100 100 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%

Grid
Offset 700 700 700 700 700 700 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off

RSA Parameter
Noise Cut 512 512 512 512 512 512 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 800 800 800 800 800 800 0…1023

SYS_ZiehmVisionR_V1.00.doc Page 60/106


System Setting Version 1.00

Ziehm Vision R

3.11.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 3,0 3,6 4,0 3,0 3,6 4,0 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95a 95 95 95 95 95 %

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction Off Off Off Off On/Off
RSA Default Modifier Fluoro Fluoro Fluoro Fluoro

SYS_ZiehmVisionR_V1.00.doc Page 61/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/sec 12 12 15 30 1-25 / 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 12 12 15 30 1-25 / 30

Burstmode
Frms/sec Burs Pause Burst Pauses Burst Pauses Bursts Pauses
ts s s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 62/106


System Setting Version 1.00

Ziehm Vision R

3.12 DSA Extremities (CO2)


3.12.1 Adjustments

3.12.1.1 Fluoro
Dose 1 12” II 9” II
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
White Level On On On On On On On/Off
Windowing
Window 16 16 16 16 16 16 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 3 3 3 3 3 3 0-9
Off. 50kV On On On On On On On/Off
Off. Lev. 1 1 1 1 1 1 0-
50kV 100%
Off. 100kV On On On On On On On/Off
Off. Lev. 7 7 7 7 7 7 0-
100kV 100%
Dynamic 100 100 100 100 100 100 5-
100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
7 8 12 7 8 12 7 8 12 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 1 2 0 1 2 0 1
Fluoro na na na na na na na na na
LPK na na na na na na na na na
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

SYS_ZiehmVisionR_V1.00.doc Page 63/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 0 0 0…100%
Video low 0 0 0…100%
high 100 100 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%

Grid
Offset 750 750 750 750 750 750 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off

3.12.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 4,5 4,5 4,5 4,5 4,5 4,5 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95a 95 95 95 95 95 %

SYS_ZiehmVisionR_V1.00.doc Page 64/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 2,0 40 2,0
2 50 8,0 50 8,0
3 60 16,0 60 16,0
4 70 30,0 70 30,0
5 80 42,0 80 42,0
6 90 47,0 90 47,0
5 100 55,0 100 55,0
7 110 56,0 110 56,0
7 120 62,5 120 62,5

Start (kV/mA) 60 60
Max Pulse 25 30
Rate
Pulse Length 20 16 10, 16, 20, 30
in ms
Bilder/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Bursts Pauses
s s
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction Off Off Off Off On/Off

SYS_ZiehmVisionR_V1.00.doc Page 65/106


System Setting Version 1.00

Ziehm Vision R

3.13 RSA Extremities (CO2)


3.13.1 Adjustments

3.13.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,0 1,2 1,6 1,2 1,6 1,6 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting Off Off Off Off Off Off On/Off
Video
Black Level On On On On On On On/Off
Black Level 17 17 17 17 17 17
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Metall Black 17 17 17 17 17 17
Peak 95 95 95 95 95 95
Windowing
Window 22 22 22 22 22 22 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Off. 50kV Off Off Off Off Off Off On/Off
Off. Lev. 50kV 1 1 1 1 1 1 0-100%
Off. 100kV Off Off Off Off Off Off On/off
Off. Lev. 100kV 7 7 7 7 7 7 0-100%
Dynamic 100 100 100 100 100 100 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
3 4 5 3 4 5 3 4 5 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 1 1 1 1 1 1 1 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 0 0 0 0 0 0 0 0 1
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

SYS_ZiehmVisionR_V1.00.doc Page 66/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 60 10,0 60 10,0
3 80 15,0 80 15,0
4 100 15,0 100 15,0
5 120 13,0 120 13,0
6 0 0 0 0
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max pulse rate 25 30 1-25 / 30
Pulse Length in ms 20 20 10, 16, 20,
30
Max Frms/sec 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 0 0 0…100%
Video low 0 0 0…100%
high 100 100 kV
Dose high 0 0 0…100%
Video high 0 0 0…100%

Grid
Offset 700 700 700 700 700 700 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
RSA Parameter
Noise Cut 516 516 516 516 516 516 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 800 800 800 800 800 800 0…1023

SYS_ZiehmVisionR_V1.00.doc Page 67/106


System Setting Version 1.00

Ziehm Vision R

3.13.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 3,0 3,6 4,0 3,0 3,6 4,0 µGy/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95a 95 95 95 95 95 %

SYS_ZiehmVisionR_V1.00.doc Page 68/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/s 12 12 15 30 1-25 / 30
Pulse Length in 20 10 10 8 10, 16, 20, 30
ms
Start Frms/sec 12 12 15 30 1-25 / 30

Burstmode
Frms/sec Burst Pause Burst Pauses Burst Pauses Bursts Pauses
s s s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 69/106


System Setting Version 1.00

Ziehm Vision R

3.14 DSA Trunk


3.14.1 Adjustments

3.14.1.1 Fluoro
Dose 1 12” II 9” II
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
White Level On On On On On On On/Off
Windowing
Window 45 45 45 45 45 45 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 3 3 3 3 3 3 0-9
Off. 50kV On On On On On On On/Off
Off. Lev. 1 1 1 1 1 1 0-
50kV 100%
Off. 100kV On On On On On On On/Off
Off. Lev. 7 7 7 7 7 7 0-
100kV 100%
Dynamic 100 100 100 100 100 100 5-
100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
7 8 12 7 8 12 7 8 12 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 3 0 1 3 0 1 3 0 1
Fluoro na na na na na na na na na
LPK na na na na na na na na na
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

SYS_ZiehmVisionR_V1.00.doc Page 70/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 30 30 0…100%
Video low 15 15 0…100%
high 100 100 kV
Dose high 30 30 0…100%
Video high 15 15 0…100%

Grid
Offset 750 750 750 750 750 750 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off

3.14.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 3,0 3,6 4,0 3,0 3,6 4,0 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95a 95 95 95 95 95 %

SYS_ZiehmVisionR_V1.00.doc Page 71/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 5,0 40 5,0
2 50 25,0 50 25,0
3 60 50,0 60 50,0
4 80 75,0 70 75,0
5 90 75,0 80 75,0
6 100 73,0 90 73,0
5 120 62,5 100 62,5

Start (kV/mA) 60 60
Max Frms/s 25 30
Pulse Length 20 16 10, 16, 20, 30
in ms
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Bursts Pauses
s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction Off Off Off Off On/Off

SYS_ZiehmVisionR_V1.00.doc Page 72/106


System Setting Version 1.00

Ziehm Vision R

3.15 RSA Trunk


3.15.1 Adjustments

3.15.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,0 1,2 1,6 1,2 1,6 1,6 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 17 17 17 17 17 17
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Metall Black 17 17 17 17 17 17
Peak 95 95 95 95 95 95
Windowing
Window 70 70 70 70 70 70 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Off. 50kV Off Off Off Off Off Off On/Off
Off. Lev. 50kV 0 0 0 0 0 0 0-100%
Off. 100kV Off Off Off Off Off Off On/off
Off. Lev. 100kV 50 50 50 50 50 50 0-100%
Dynamic 100 100 100 100 100 100 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
3 4 6 3 4 6 3 4 6 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 2 1 1 2 1 1 1 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 0 0 0 0 0 0 0 0 1
SSHT 0 0 0 0 0 0 0 0 1
Low D 0 0 0 0 0 0 0 0 1
REP 0 0 0 0 0 0 0 0 1

SYS_ZiehmVisionR_V1.00.doc Page 73/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 60 10,0 60 10,0
3 80 15,0 80 15,0
4 100 18,0 100 18,0
5 120 17,2 120 17,2

Start (kV/mA) 60 60
Max Frms/s 25 30 1-25 / 30
Pulse Length in ms 30 20 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 5 5 0…100%
Video low 5 5 0…100%
high 100 100 kV
Dose high 5 5 0…100%
Video high 5 5 0…100%

Grid
Offset 700 700 700 700 700 700 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off

RSA Parameter
Noise Cut 512 512 512 512 512 512 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 800 800 800 800 800 800 0…1023

SYS_ZiehmVisionR_V1.00.doc Page 74/106


System Setting Version 1.00

Ziehm Vision R

3.15.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 3,0 3,6 4,0 3,0 3,6 4,0 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95a 95 95 95 95 95 %

SYS_ZiehmVisionR_V1.00.doc Page 75/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 80 60
Max Frms/s 12 12 15 15 1-25 / 30
Pulse Length 20 10 10 8 10, 16, 20, 30
in ms
Start Frms/s 12 12 15 15 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses
s s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 76/106


System Setting Version 1.00

Ziehm Vision R

3.16 DSA Trunk (CO2)


3.16.1 Adjustments

3.16.1.1 Fluoro
Dose 1 12” II 9” II
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Vignetting Off Off Off Off Off Off On/Off

Video
Black Level On On On On On On On/Off
White Level On On On On On On On/Off
Windowing
Window 16 16 16 16 16 16 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 3 3 3 3 3 3 0-9
Off. 50kV On On On On On On On/Off
Off. Lev. 1 1 1 1 1 1 0-
50kV 100%
Off. 100kV On On On On On On On/Off
Off. Lev. 7 7 7 7 7 7 0-
100kV 100%
Dynamic 100 100 100 100 100 100 5-
100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
7 8 12 7 8 12 7 8 12 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 1 2 0 1 2 0 1
Fluoro na na na na na na na na na
LPK na na na na na na na na na
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

SYS_ZiehmVisionR_V1.00.doc Page 77/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 30 30 0…100%
Video low 15 15 0…100%
high 100 100 kV
Dose high 5 5 0…100%
Video high 5 5 0…100%

Grid
Offset 750 750 750 750 750 750 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off

3.16.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 4,5 4,5 4,5 4,5 4,5 4,5 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95 95 95 95 95 95 %

SYS_ZiehmVisionR_V1.00.doc Page 78/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 5,0 40 5,0
2 50 20,0 50 20,0
3 60 50,0 60 50,0
4 80 75,0 70 75,0
5 90 75,0 80 75,0
6 100 73,0 90 73,0
7 120 62,5 100 62,5

Start (kV/mA) 60 60
Max Frms/s 25 30
Pulse Length 20 16 10, 16, 20, 30
in ms
Start Frms/s 12,5 15 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Bursts Pauses
s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction Off Off Off Off On/Off

SYS_ZiehmVisionR_V1.00.doc Page 79/106


System Setting Version 1.00

Ziehm Vision R

3.17 RSA Trunk (CO2)


3.17.1 Adjustments

3.17.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,0 1,2 1,6 1,2 1,6 1,6 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 25 20 30 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting Off Off Off Off Off Off On/Off
Video
Black Level On On On On On On On/Off
Black Level 17 17 17 17 17 17
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Metall Black 17 17 17 17 17 17
Peak 95 95 95 95 95 95
Windowing
Window 22 22 22 22 22 22 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Off. 50kV Off Off Off Off Off Off On/Off
Off. Lev. 50kV 1 1 1 1 1 1 0-100%
Off. 100kV Off Off Off Off Off Off On/off
Off. Lev. 100kV 7 7 7 7 7 7 0-100%
Dynamic 100 100 100 100 100 100 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
3 4 5 3 4 5 3 4 5 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 1 1 1 1 1 1 1 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 0 0 0 0 0 0 0 0 1
SSHT na na na na na na na na na
Low D na na na na na na na na na
REP na na na na na na na na na

SYS_ZiehmVisionR_V1.00.doc Page 80/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 60 10,0 60 10,0
3 80 15,0 80 15,0
4 100 18,0 100 18,0
5 120 17,2 120 17,2
6 0 0 0 0
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/s 25 30
Pulse Length in ms 30 20 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II
31 23 15 23 15 10 cm
Auto Gain off off On/Off
Threshold 100 100 kV
Max Gain 5 5 0…100
Black Level Reduction
low 50 50 kV
Dose low 5 5 0…100%
Video low 5 5 0…100%
high 100 100 kV
Dose high 5 5 0…100%
Video high 5 5 0…100%

Grid
Offset 700 700 700 700 700 700 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
RSA Parameter
Noise Cut 516 516 516 516 516 516 0…1023
Trace Threshold 1023 1023 1023 1023 1023 1023 0…1023
Trace Offset 800 800 800 800 800 800 0…1023

SYS_ZiehmVisionR_V1.00.doc Page 81/106


System Setting Version 1.00

Ziehm Vision R

3.17.1.2 HQ
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 3,0 3,6 4,0 3,0 3,6 4,0 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 30 30 30 30 30 30 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 17 15 15 17 15 15 %
White Level 90 90 90 90 90 90 0-200
Metall Black 17 15 15 17 15 15 %
Peak 95 95 95 95 95 95 %

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction Off Off Off Off On/Off
RSA Default Modifier Fluoro Fluoro Fluoro Fluoro

SYS_ZiehmVisionR_V1.00.doc Page 82/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,
5

Start (kV/mA) 60 60 60 60
Max Frms/s 12 12 15 15 1-25 / 30
Pulse Length 20 10 10 8 10, 16, 20, 30
in ms
Start Frms/s 12 12 15 15 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses
s s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 83/106


System Setting Version 1.00

Ziehm Vision R

3.18 ENDO
3.18.1 Adjustments

3.18.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 15 0 15 15 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 5 5 5 5 5 5 %
White Level On On On On On On On/Off
White Level 95 85 75 95 95 80 0-200
Metal Black Level tba tba tba tba tba tba
Metall Peak tba tba tba tba tba tba

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Level On On On On On On On/Off
Offs. lev. (50kV) 5 5 5 5 5 5 0-100%
Level Off Off Off Off Off Off On/Off
Offs. lev.(100kV)
Dynamic 10 10 10 10 10 10 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
2 3 4 2 3 4 2 4 6 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 1 0 1 1 0 1 1 0 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 2 0 1 2 4 1 2 4 1
SSHT 1 0 1 1 0 1 1 0 1
Low D. 2 0 1 2 4 1 2 4 1
REP 1 0 1 1 4 1 1 4 1

SYS_ZiehmVisionR_V1.00.doc Page 84/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 – 120V

kV mA kV mA
1 40 1,5 40 1,5
2 55 6,0 55 6,0
3 60 11,3 60 11,3
4 70 18,5 70 15,3
5 90 15,5 90 12,5
6 100 14,5 100 12
7 120 13,0 120 10,3
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/s 25 30 1-25 / 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. 0 0 0 0 0 0 0…4095

Auto Gain/Auto LPK


Auto Gain on on
Threshold 100 100 kV
Max Gain 3 3
Auto LPK on on
off 80 80
on 100 100

SYS_ZiehmVisionR_V1.00.doc Page 85/106


System Setting Version 1.00

Ziehm Vision R

3.18.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,6 3,0 2,0 2,6 3,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 15 0 15 15 %

Video
Black Level 5 5 5 5 5 5 %
White Level 95 85 75 95 85 75 0-200
Metall Black tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 1,7 40 1,7 40 5 40 5,0
2 50 5,0 50 5,0 50 20 50 20,0
3 60 60 60 50 60 50,0
4 75 38,0 75 38,0
5 80 80 80 40 44 46 50 51 54 58 80 67,0
80 80 60 62 65 67 71 73 75
80 80 75 75 75
6 90 90 90 37 40 42 46 47 50 53
90 90 55 57 60 62 65 68 70
90 90 72 75 75
7 100 100 100 34 37 39 42 44 46 49 100 58,0
100 100 51 53 56 58 60 63 65
100 100 67 70 73
8 120 28,0 120 28,0 120 30 32 34 36 38 40 42 120 50,0 < 10 R/min !!!
120 120 45 46 48 50 52 54 56
120 120 58 60 62,
5

Start (kV/mA) 75 75 80 60
Max Frms/s 25 25 30 30 1-25 / 30
Pulse Length 20 20 10 8 10, 16, 20, 30
in ms
Start Frms/s 25 25 30 30 1-25 / 30

Burstmode

SYS_ZiehmVisionR_V1.00.doc Page 86/106


System Setting Version 1.00

Ziehm Vision R

Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses


s s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.18.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 5 5 5 5 5 5 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,5 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8 0 0 0 0
9 0 0 0 0
Start 80 80 %
Max Frms/s 25 30 1-25 / 30
Pulse Length in ms 30 30 10, 16, 20, 30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 87/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

3.18.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 1,6 1,6 1,6 1,6 1,6 1,6 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 25 25 25 25 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 15 15 %

Video
Black Level 5 5 5 5 5 5 %
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V
Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 80 60 80 60
Max Frms/s 25 25 30 30 1-25 / 30
Pulse Length in ms 20 16 10 8 10, 16, 20, 30
Start Frms/s 25 25 30 30 1-25 / 30

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 88/106


System Setting Version 1.00

Ziehm Vision R

3.18.1.5 Low Dose


Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 0,27 0,27 0,27 0,27 0,27 0,27 µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 25 25 20 25 25 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level on on on on on on %
Black Level 9 9 9 9 9 9 %
White Level on on on on on on 0-200
White Level 95 85 75 95 85 75 0-200
Metal Black Level tba tba tba tba tba tba
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 0,1 40 0,1
2 80 0,1 80 0,1
3 120 6,5 120 6,5
4
5
6
7
8
9
Start 60 60
Max Frms/sec 25 30
Pulse Length in ms 16 10 10, 16, 20,
30
Start Frms/sec s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 89/106


System Setting Version 1.00

Ziehm Vision R

3.19 U R O (MTS 2.16)


3.19.1 Adjustments

3.19.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Dynamic 100 100 100 100 100 100 %
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 5 5 5 5 5 4 %
Peak 95 95 95 95 95 95 %
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Peak 95 95 95 95 95 95 %
Metall Black tba tba tba tba tba tba
Level
Metall Peak tba tba tba tba tba tba

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Level On On On On On On On/Off
Level 15 15 15 15 15 15 0-
100%
Dynamic 10 10 10 10 10 10 5-
100%

SYS_ZiehmVisionR_V1.00.doc Page 90/106


System Setting Version 1.00

Ziehm Vision R

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
5 8 16 5 8 16 5 8 16 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 1 2 0 1 2 0 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 2 0 1 2 0 1 2 0 1
SSHT 2 0 1 2 0 1 2 0 1
REP 1 0 1 1 4 1 1 4 1

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,5 40 1,5
2 55 6,0 55 6,0
3 90 13,0 90 13,0
4 100 13,0 100 13,0
5 120 13,0 120 11,2 – < 10 R/min!
8,0
6 0 0 0 0
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 90 90
Pulse Length in ms 30 30 10, 16, 20,
30
Bilder/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 91/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Grid
Offset 700 650 650 700 650 650 0…4095
Difference 4095 4095 4095 4095 4095 4095 0…4095
Grid Regulation On On On On On On On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

3.19.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 2,0 2,6 3,0 2,0 2,6 3,0 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Dynamic 100 100 100 100 100 100 %
Peak Part 0 0 0 0 0 0 %

Video
Black Level 5 5 5 5 5 5 %
Peak 95 95 95 95 95 95 %
White Level 95 95 95 95 95 95 0-200
Peak 95 95 95 95 95 95 %
Metall Black tba tba tba tba tba tba
Peak tba tba tba tba tba tba

SYS_ZiehmVisionR_V1.00.doc Page 92/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap’s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5 40 5,0
2 50 20,0 50 20,0 50 20 50 20,0
3 60 50,0 60 50,0 60 50 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 67,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 58,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 50,0
120 45 46 48 50 52 54 56
120 58 60 62,5

Start (kV/mA) 80 60 80 60
Pulse Length 20 10 10 8 10, 16, 20, 30
in ms
Bilder/s 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses
1

Dose 3 31cm 23cm 15cm 10cm Res

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity 0 0 0 0

SYS_ZiehmVisionR_V1.00.doc Page 93/106


System Setting Version 1.00

Ziehm Vision R

3.19.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate 0,43 0,56 0,85 0,56 0,85 1,1 µG/sec
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Dynamic 100 100 100 100 100 100 %
Peak Part 0 0 0 0 0 0 %

Video
Black Level 3 3 3 3 3 4 %
Peak 95 95 95 95 95 95 %
White Level 95 95 95 95 95 95 0-200
Peak 95 95 95 95 95 95 %
Metall Black tba tba tba Tba tba tba
Level
Peak tba tba tba Tba tba tba

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,7 40 1,7
2 45 2,0 45 2,0
3 50 2,5 50 2,5
4 60 5,0 60 5,0
5 70 9,0 70 9,0
6 80 9,0 80 9,0
7 120 13,0 120 11,2 –
8,0
8
9
Start 80 80
Pulse Length in ms 30 30 10, 16, 20,
30
Bilder/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 94/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity 0 0 0 0

3.19.1.4 SSHT
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose 1,6 1,6 1,6 1,6 1,6 1,6 µG
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Dynamic 100 100 100 100 100 100 %
Peak Part 0 0 0 0 0 0 %

Video
Black Level 5 5 5 5 5 5 %
Peak 95 95 95 95 95 95 %
White Level 95 95 95 95 95 95 0-200
Peak 95 95 95 95 95 95 %
Metall Black tba tba tba tba tba tba
Level
Peak tba tba tba tba tba tba

Dose 2 200 - 240V 200 - 240V w/o 100 - 120V 100 - 120V w/o
Cap’s Cap´s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5,0 40 5,0
2 50 20,0 50 20,0 50 20,0 50 20,0
3 60 50,0 60 50,0 60 50,0 60 50,0
4 80 75,0 80 62,5 80 75,0 80 62,5
5 90 75,0 90 62,5 90 75,0 90 62,5
6 100 73,0 100 62,5 100 73,0 100 62,5
7 120 62,5 120 62,5 120 62,5 120 62,5
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
Start 80 60 80 60
Pulse Length in ms 20 16 10 8 10, 16, 20, 30
Bilder/s 25 25 30 30 1-25 / 30

SYS_ZiehmVisionR_V1.00.doc Page 95/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 31cm 23cm 15cm 10cm Res

Auto Dose Reduction on Off Off Off Off On/Off


Motion Sensitivity

SYS_ZiehmVisionR_V1.00.doc Page 96/106


System Setting Version 1.00

Ziehm Vision R

3.20 URO-Storz (MTS 2.16)


3.20.1 Adjustments

3.20.1.1 Fluoro
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba 0 0 0 0-9
Peak Part 0 0 0 0 0 0 %
Vignetting On Off Off On Off Off On/Off

Video
Black Level On On On On On On On/Off
Black Level 5 5 5 5 5 4 %
White Level On On On On On On On/Off
White Level 95 95 95 95 95 95 0-200
Metal Black Level tba tba tba 0 0 0 %
Peak tba tba tba 85 90 90 %

Windowing
Window 100 100 100 100 100 100 0-100
Level 50 50 50 50 50 50 0-100

Gamma
Gamma On On On On On On On/Off
Gamma 5 5 5 5 5 5 0-9
Level On On On On On On On/Off
Level 15 15 15 15 15 15 0-100%
Dynamic 10 10 10 10 10 10 5-100%

Filter MAG 0 MAG 1 MAG 2


Nr. 1 2 3 1 2 3 1 2 3
5 8 16 5 8 16 5 8 16 1-16
NR RTE LIH NR RTE LIH NR RTE LIH
HQ 2 0 1 2 0 1 2 0 1
Fluoro 2 0 1 2 0 1 2 0 1
LPK 2 0 1 2 0 1 2 0 1
SSHT 2 0 1 2 0 1 2 0 1
REP 1 0 1 1 4 1 1 4 1

SYS_ZiehmVisionR_V1.00.doc Page 97/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,5 40 1,5,0
2 55 6,0 55 6,0
3 90 13,0 90 13,0-
10,0
4 100 13,0 100 13,0-
10,0
5 120 13,0 120 13,0-
10,0
6 0 0 0 0
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start (kV/mA) 60 60
Max Frms/s 25 30 1-25 / 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

SYS_ZiehmVisionR_V1.00.doc Page 98/106


System Setting Version 1.00

Ziehm Vision R

Dose 3 12” II 9” II Res


31 23 15 23 15 10 cm
Auto Gain/Auto LPK
Auto Gain on on
Threshold 100 100 kV
Max Gain 2 2
Auto LPK On on
off 80 80
on 100 100

Grid
Offset 700 650 650 700 650 650 0…4095
Difference 1700 1600 1600 1700 1600 1600 0…4095
Grid Regulation Off Off Off Off Off Off On/Off
Motion Detection Off Off Off Off Off Off On/Off
Auto Dose Red. Off Off Off Off Off Off On/Off
Motion Sensi. na na na na na na 0…4095

3.20.1.2 HQ CINE
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dose rate µG/s
Iris Camera tba tba tba Tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain 0 0 0 0 0 0 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 5 5 5 5 5 5 %
White Level 95 95 95 95 95 95 0-200
Metal Black Level tba tba tba Tba tba tba
Peak tba tba tba Tba tba tba

SYS_ZiehmVisionR_V1.00.doc Page 99/106


System Setting Version 1.00

Ziehm Vision R

Dose 2 200 - 240V 200 - 240V 100 - 120V 100 - 120V


w/o Cap’s w/o Cap´s

kV mA kV mA kV mA kV mA
1 40 5,0 40 5,0 40 5 40 5,0
2 50 20,0 50 20,0 50 20 50 20,0
3 60 50,0 60 50,0 60 50 60 50,0
4 80 75,0 80 75,0 80 40 44 46 50 51 54 58 80 75,0
80 60 62 65 67 71 73 75
80 75 75 75
5 90 75,0 90 75,0 90 37 40 42 46 47 50 53 90 75,0
90 55 57 60 62 65 68 70
90 72 75 75
6 100 73,0 100 73,0 100 34 37 39 42 44 46 49 100 71,0
100 51 53 56 58 60 63 65
100 67 70 73
7 120 62,5 120 62,5 120 30 32 34 36 38 40 42 120 62,5
120 45 46 48 50 52 54 56
120 58 60 62,5

Start (kV/mA) 60 60 60 60
Max Frms/s 25 25 30 30 1-25 / 30
Pulse Length 20 10 10 8 10, 16, 20, 30
in ms
Start Frms/s 25 25 30 30 1-25 / 30

Burstmode
Frms/sec Bursts Pause Burst Pauses Burst Pauses Bursts Pauses
s s s

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
Motion Sensitivity na na na na

SYS_ZiehmVisionR_V1.00.doc Page 100/106


System Setting Version 1.00

Ziehm Vision R

3.20.1.3 LPK
Dose 1 12” II 9” II Res
Dose 31cm 23cm 15cm 23cm 15cm 10cm
Dosis rate µG/s
Iris Camera tba tba tba tba tba tba %
CCD Fill 20 20 20 20 20 20 %
Dose Gain tba tba tba tba tba tba 0-9
Peak Part 0 0 0 0 0 0 %

Video
Black Level 3 3 3 3 3 4 %
White Level 95 95 95 95 95 95 0-200
Metal Black Level tba tba tba 0 0 0 %
Peak tba tba tba 85 80 80 %

Dose 2 200 - 240V 100 - 120V

kV mA kV mA
1 40 1,5 40 1,5
2 55 6,0 55 6,0
3 90 13,0 90 13,0-
10,0
4 100 13,0 100 13,0-
10,0
5 120 13,0 120 13,0
6 0 0 0 0
7 0 0 0 0
8 0 0 0 0
9 0 0 0 0
Start 60 60
Max Frms/s 25 30 1-25 / 30
Pulse Length in ms 30 30 10, 16, 20,
30
Start Frms/s 25 30 1-25 / 30

Burstmode
Frms/sec Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses Bursts Pauses

Dose 3 31cm 23cm 15cm 10cm Res

Auto Gain Off Off Off Off On/Off


Auto Dose Reduction on Off Off Off Off On/Off
na na na na na na na

SYS_ZiehmVisionR_V1.00.doc Page 101/106


System Setting Version 1.00

Ziehm Vision R

4. Risk Prevention
The definition of the dose rate is done with respect to the DIN EN 60601-1 in every APR including
the modifiers (HQ, LPK, SSHT). The dose value must be adjusted for each single APR,
magnifications and modifiers separately.

SYS_ZiehmVisionR_V1.00.doc Page 102/106


System Setting Version 1.00

Ziehm Vision R

5. Installation Requirements

SYS_ZiehmVisionR_V1.00.doc Page 103/106


System Setting Version 1.00

Ziehm Vision R

6. Requirements from Homologations and Standards


NA

SYS_ZiehmVisionR_V1.00.doc Page 104/106


System Setting Version 1.00

Ziehm Vision R

7. Production Requirements and Limitations


NA

SYS_ZiehmVisionR_V1.00.doc Page 105/106


System Setting Version 1.00

Ziehm Vision R

8. Appendix

8.1 Scope and Readership

Additionally, the documentation is read by:


- R&D designers,
- Company Management,
- Quality Assurance Management,
- Production,
- Product Management,
- System Test Engineers,
- Test Engineers,
- Service Engineers.

8.2 Definitions and Abbreviations

Concept /
Definition Reference / Remarks
Abbreviations
tba to be adjusted ---
NA not applicable ---
APR anatomical program ---

SYS_ZiehmVisionR_V1.00.doc Page 106/106