Professional Documents
Culture Documents
Section 27 - Ranarex Gravitometer
Section 27 - Ranarex Gravitometer
Section 27
Instrumentation – Ranarex
Gravitometer
Table of Contents
Introduction and Objectives……………………………………………………27-2
Specifications……………………………………………………………………27-3
General Information……..………….…………………...………………..Section 1
Installation………………………………………………………………....Section 2
Operation…………………………………………………………………..Section 3
Maintenance and Servicing………………………………………………Section 4
Calibration...………………………………………………………………..Section 5
Introduction
The section contains information on the Ranarex Gas Gravitometer
Objectives
After completing this section, you should be able to
Ranarex Gravitometer
SPECIFICATIONS
PORTABLE TYPE GRAVITOMETER
Case :
Portable with carrying handle
Size: 13%"W x 13h"H x 10%"D, excluding handle and gas hose connections
Weight: Approximately 38 lb
Readings :
Indicating dial
Drive Motor:
115 volt, +lo% 50 or 60 Hz, single-phase
Permanent split-capacitor (no arcing during Start, Run or Stop)
Explosion-proof motor not available
Accuracy :
Within 0.5% of actual value at any reading within range
Portable Gravitometers:
Dual-range: 0.52-1.03 and 0.97-1.90
GENERAL
Materials:
Case and Chassis: Aluminum
Impellers, Impulse Wheels: Molded phenolic
Drive: Stainless steel shafts; sealed ball bearings
Pivots: Stainless steel shafts in jewels
Internal trim: Aluminum and stainless steel
Ambient Temperature Rating:
Saturated gas and air: 40 to 120°F2 (4" - 50°C).
Normal gas and dry air: 0 to 130°F2 (-18" - 54°C).
SECTION I
GENERAL INFORMATION
1.1 RANAREX GAS GRAVITOMETERS
RANAREX gravitometers measure the relative density (specific gravity) of
gases as compared to air at standard temperature and pressure (STP). For all
gases there is a definite relation between the composition of the gas mixture
and its relative density. The instrument is calibrated over two ranges to
indicate relative density.
RANAREX gravitometers excel over more complex and costly analyzers in
total performance, including simplicity, fast response, accuracy, sensitivity,
ruggedness and lowest long-term cost.
Figure 1 illustrates the measuring system for two operating conditions. The
left view shows the position of the cam and reference wheel when the
reference gas is admitted into both chambers. In this case, the impeller wheel
torques will be equal. As a result of this condition and the linear profile of the
cam, the graduated dial is calibrated to indicate 1000.
The right view of Figure 1 shows the position of the cam and reference wheel
when an unknown gas is admitted to the upper chamber and reference air is
admitted into the lower chamber. The lighter gas (in the upper chamber)
creates a smaller torque than the air in the lower chamber. As a result of this
condition, the cam will rotate until the cam radii allows the opposing torques to
balance. The cam rotation is calibrated to indicate the ratio of the torques, or
relative density.
To measure relative densities above 1.000, the unknown gas is admitted to
the lower chamber and the dry reference air is admitted to the upper chamber.
The principle of operation is the same. The graduated dial has two scales.
The inner scale is used for relative densities below 1.000. The outer scale is
used for relative densities above 1.000.
GENERAL INFORMATION
Due to the nature of the design, the pressures and temperatures of the gas
and reference air are equalized. Any changes in pressure, temperature, and
motor speed affect both torques equally. Since the torques produce opposing
forces the effects of varying pressure, temperature and motor speed are
canceled.
An air drier is built into the gravitometer to dry incoming ambient air for use as
the reference gas. The gravitometer is equipped with a selector valve that is
used to direct sample gas and dry reference air to the correct chambers. The
three positions are used for checking the instrument ZERO, for measuring gas
relative densities below 1.000, or for measuring gas relative densities above
1.000.
The motor is designed for operation on 115 VAC + l o % , 60Hz. When the
gravitometer is to be used in an area where AC power is not available, a
DCAC invertor may be used to convert the DC voltage supplied by an
automobile alternator or other DC source to the correct AC voltage.
"The ratio of the density of the gas, under the observed conditions
of pressure and temperature to the density of dry air at the same
pressure and temperature."
1.3.2 Pressure
The gas sampling system must reduce the gas to barometric pressure to
equal the pressure of the reference air. The correct gas pressure corresponds
to a sample flow rate of 10-15 scfh for the portable RANAREX gravitometer.
The flow rate is measured by a rotameter. As an operating convenience, the
flow scale is graduated in relative density units. The operator can adjust the
needle valve so that the flow meter reads the approximate relative density of
the gas. For gas pressures above 20 psig, a suitable pressure reducing
regulator must precede the rotameter.
1.3.3 Temperature
The reference air is admitted to the gravitometer at the ambient temperature
of the instrument. For the gas and air to be measured at the same
temperature, the gas must be heated or cooled to ambient temperature. This
is important if the gas pressure has been reduced more than 100 Psig, or if
the gas flows through a temperature zone more than 10'~ (lZ°C) different from
the ambient temperature of the gravitometer. If the sample gas temperature
must be adjusted, the gas may be passed through a coiled 10 ft. (30m)
section of metal tubing before entering the gravitometer.
1.3.5 Filtering
A commercial filter (25 micron) must be used ahead of the regulator and
flowmeter to protect the instrument. Other precautions must be used to
prevent liquids from entering the instrument. A Balston filter (or equivalent)
may be required.
SECTION 2
INSTALLATION
INSTALLATION
4. Fill the air drier. The air drier must be removed from the upper left side of
the instrument by loosening the two fasteners. Remove the plastic drier
assembly from the gravitometer far enough until the rubber tube can be
removed from the elbow at the rear of the drier. Remove the rubber plug on
the front of the drier by lifting the lever at the center of the plug (but do not
turn the lever).
Hold the drier horizontal with the front face up and pour silica gel using the
funnel into the spout. Tap or shake the drier vigorously and continue to fill the
drier to the base of the spout. Replace the rubber plug, attach the rubber tube
on the elbow at the rear of the drier, install the drier into the gravitometer.
The moisture capacity of the silica gel will vary according to the atmospheric
humidity and it must be inspected at regular intervals. When the upper half of
the narrow compartment shows a change from deep blue color to gray-pink
color, the silica gel must be replaced.
Active silica gel has a deep blue color and must be kept in a tightly closed
container. Expended silica gel has a light grayish-pink color. The silica gel
may be reactivated and used over and over again. To reactive, heat the silica
gel in an open container to approximately 250°F until the dark blue color
returns, then store in a tightly closed container. Additional silica gel may be
obtained from Chandler Engineering as Part
5. Attach the outlet hoses. If the gravitometer is operated in a confined space,
such as inside a vehicle, the gas sample must be discharged outside the
vehicle. This will prevent contaminating the operating space with noxious or
hazardous gas and air mixtures. If measuring only gas below 1.000 relative
density, attach a hose to the port LIGHT GAS OUTLET on upper right side of
gravitometer case. If measuring gases above 1.000 relative density, attach a
hose to the port HEAVY GAS OUTLET. These hoses must remain attached
when checking the instrument "zero point".
6 . Gas filter. If the gas contains suspended particles larger than 25 microns,
or if the gas is "wet", filtering is required. If a commercial filter is not available,
a suitable moisture trap may be made from 1" (2.54cm) pipe 15" (38cm) long
mounted vertically. Cap both ends and install a drain valve in the bottom cap.
Install an inlet connection from the gas line 2" (5cm) from the bottom and
place the outlet to the gravitometer 2" (5cm) from top. Fill the pipe loosely with
commercial No. 0 steel wool. Drain the trap frequently to keep the steel wool
free from condensate.
7. Gas temperature. If the gas sample is at a pressure above 100 Psig, or if
the sample flows through a temperature zone more than 10°F (-12°C)
different from the ambient temperature of the gravitometer, sample
temperature adjustment is required. Install a section of 3/8" (lcm) metal tube
10 ft (3m) long, and locate it near the gravitometer case. The tube may be
coiled or bent to suit the space available. The tube will serve as a heat
exchanger to heat or cool the sample gas.
8. Make the gas sample connection. Attach a pressure reducing regulator to
the sample tap and route the sample line to the gravitometer. The sample line
must be capable of delivering 12 SCFH gas, free of condensed liquids, at a
pressure not exceeding 15 Psig. It is recommended to use proper sample
probes and sampling techniques as defined in ASTM, GPA, and API
standards.
SECTION 3
OPERATION
The "LOCK" position of the pointer is at the upper right corner of the scale,
near the 0.970-1.030 graduations. The position of the locking mechanism is
shown by the arrow of the knob and the legends on the gravitometer front
cover. To check if the pointer is locked, turn the knob to "UNLOCK" and
observe if the pointer drifts downward, then turn the knob to the "LOCK"
position and observe if the pointer snaps into original position.
3.1 PROCEDURE
The correct procedure for operating the RANAREX gravitometer is described
on the instruction plate located over the center of the indicating scale and
reproduced as Figure 5 of this section. It is recommended that the user
observe the sequence of steps as listed. In this way, maximum accuracy will
be achieved.
The following comments are also helpful :
For example, when measuring 0.6 gravity gas, a flow meter setting of about
1.2 will reduce the response time to about 30 seconds. However, the flow
must not exceed the range of the flow meter.
3.1.3 Over-Pressure Protection:
If the flowmeter valve is opened before the selector valve is turned to "LG" or
"HG" , the valve cover will lift off the body to relieve the pressure and will be
reseated by the spring.
SECTION 4
MAINTENANCE AND SERVICING
4 .1 LUBRICATION
The motor bearings, impeller bearings and idler pulley bearings have long-life
lubrication to provide long service. If the impeller bearings bind or become
extremely noisy, they may be lubricated as described in the section titled
Impeller Bearings, or the complete assembly may be replaced. The idler
pulley bearing should be inspected and re-lubricated if it does not turn freely
or becomes noisy.
4 . 2 SELECTOR VALVE
If the sample selector valve is contaminated, it will require cleaning and
lubrication. Should this become repeatedly necessary, a gas filter should be
used. To clean the selector valve, remove the screw at the center of the valve
cover and withdraw the screw, 2 washers, spring and cover. Remove all old
grease and contamination from the surfaces and channels in the cover and
valve body. Lubricate the faces of the cover and body with silicone-based
grease and replace all parts. If the valve cover or valve body become scored,
it is necessary to lap/polish the mating surfaces to eliminate the scoring.
Clean both surfaces, apply fine valve grinding compound to them and
reassemble the valve. Loosen the detent spring which engages the "O”, "LG"
and "HG" notches of the cover. Repeatedly rotate the cover on the valve body
beyond its normal operating range, continuing until the scoring is removed.
After the lapping/polishing operation, thoroughly clean, lubricate and
reassemble all parts.
NOTE: All portable gravitometers with serial number P-259 and later
incorporate the new design b a l l bearing assembly.
4.9 TROUBLESHOOTING
To be accurate the gravitometer must satisfy the following operating
conditions:
* Sufficient flow of gas sample and reference air
* Impellers rotating at correct speed
* Pointer moves without friction
Instrument free of internal leaks
* Pointer statically balanced
* Air drier filled with active silica gel
* Instrument is properly calibrated
If there is reason to believe the gravitometer is reading incorrectly, tests and
remedies should be made in the sequence described below. If necessary to
disassemble the gravitometer, refer to the Gravitometer Disassembly
Procedure found earlier in this section.
1. Check the gas and air flow. Disconnect all inlet and outlet hoses, set
selector valve to "0" and purge the gravitometer. Verify that there is flow at the
discharge ports. If there is no evidence of flow, disconnect the hose from the
rear of the air drier and retest. If flow is established, empty the air drier, clean
the outlet elbow, fill with fresh silica gel (6-16 mesh) and retest.
If there is no evidence of flow, remove the cover of the selector valve, inspect
and clean the ports and channels of the cover and valve body and retest.
Finally, if there still is no evidence of flow, it indicates the passages within the
instrument are obstructed. Remove the rear chamber cover and check the
passages which extend from impeller bearing location to the elbows near the
bottom of the chamber cover. Also check the outlet tubes and nipples from
both chambers.
2. Check the impeller speed. Correct speed is approximately 2800 RPM as
measured with a strobe tachometer. If a tachometer is not available, remove
the drive belt and verify that the motor pulley, impeller pulleys, and the idler
pulley turn freely. If the impeller pulleys do not turn freely, lubricate the
impeller bearings as described earlier in the section titled Impeller Bearings or
replace the assembly. If the idler pulley does not turn freely, the motor should
be replaced.
3. Check for frictionless indicator response. Remove front cover and connect
the instrument to a gas sample. Record the reading. Gently move the
indicator to each side of the true reading and release. Indicator should return
to the same reading from each direction, within -001 relative density within 15
seconds. If it does not, the impulse wheel shafts and bearings should be
cleaned. Remove the front chamber cover of the operating mechanism and
hold it upright by gripping firmly in a vise, or stand it securely on right side
edge. Loosen the set screws in the hub of the cam and reference wheel and
draw the impulse wheels with shafts out of the bearings. Use care that the
cam and reference wheels and the flexible tape are not damaged while
removed from the instrument. Tag the impulse wheels to indicate the chamber
from which they were removed. Polish the shafts with crocus cloth or fine
steel wool. Do not bend the shafts.
If the impulse wheel shafts show a bright polished groove where they rotate in
the jeweled bearings, or if the hubs of the cam and reference wheel show a
groove where they have been drawn against the ends of the impulse wheel
bearings, these grooves should be removed by lapping and polishing. If the
wear is excessive, the shafts must be replaced.
Gently blow out the impulse wheel bearings with clean air or gas and add a
drop of instrument oil to both ends of each bearing. Suitable oil is available
from Chandler Engineering.
Replace the shafts with wheels in their bearings in the correct chamber and
turn them slowly. If the protruding portion of the shaft wobbles or if the wheel
does not turn freely, the shaft should be replaced with a new one. A bent
impulse wheel shaft will cause the instrument to lose accuracy and sensitivity.
Attach the measuring and reference wheels to the impulse wheel shafts being
certain that the set screws bear on the flats of the shafts. A small amount of
shaft end play [.005" to 0.010" (.13-.25mm) maximum] must exist to prevent
binding. Verify that no parts of the indicating mechanism have been bent and
that there is no binding between parts.
4. Check for internal leaks. Operate the instrument connected to a gas sample
and obtain a relative density reading at the far end of the measuring range.
Shut off the gas flow by closing the flowmeter valve. The indicator should
remain steady or should not return toward 1.000 reading at a rate faster than
.007 on the inner scale in 3 minutes. If it does not meet this test, there is an
internal leak. To determine where the leak is located, stop the motor, remove
the front and rear covers and disconnect hose from the inlet port of the
selector valve. Individually pressurize (approximately 10 psig) the upper and
lower chambers with gas and use a gas leak detector to find the leak.
The leak is probably caused by loose screws or a gasket which has become
damaged. Verify that the Tygon7 tubes which join the internal connector to the
selector valve and to the elbows near bottom, of rear chamber cover are not
cracked or damaged.
5. Check the indicator balance. The measuring mechanism was statically
balanced by the manufacturer and does not change with wear or age. The
balancing weights behind the cam should not be disturbed. However, if
required, the balance can be verified and restored as described below.
Stop the motor and remove the front cover. Place the pointer consecutively at
the right side, bottom, and left side of the scale. In each position tap the front
of the gravitometer case to overcome static friction. The pointer should remain
stationary or very nearly so. If it drifts more than two small graduations on the
red scale, make compensating adjustments of the balancing weights on the
threaded rods. When loosening or re-tightening the balancing weights, use
care that the threaded rods, pointer and impulse wheel shaft are not bent or
damaged.
Balancing the cam and mechanism is a trial and error process. The pointer
may be considered balanced if it remains steady or drifts only two scale
divisions. After rebalancing, do not permit the weights to be moved. A drop of
low strength Loctite7, lacquer, or fingernail polish on the weights and threaded
rods will lock the adjustments is place.
6. Check the silica gel in the air drier. If silica gel is exhausted, remove and
empty the air drier. Refill with active silica gel.
7. The gravitometer was accurately calibrated by the manufacturer and the
calibration does not change appreciably with wear or age. The adjustments
should not be changed without first checking every other possible source of
error. Calibration should proceed if all other sources of error are eliminated.
8. Sometimes during calibration, repeatability problems occur due to static
friction
Gently tapping the side of the case will eliminate the static friction.
SECTION 5
CALIBRATION
To calibrate the instrument there are three adjustments:
1. Zero Adjust. This is the only adjustment required in normal operation. The
zero adjust screw operates the chain, sprockets and lower impulse wheel
bearing which positions the lower impulse wheel longitudinally in the lower
chamber to match the torque of the upper chamber when operating on air.
Refer to the section titled Zero Adjustment in Section 3 for instructions for use
of the zero adjust.
2. Sensitivity. The longitudinal position of the upper impulse wheel in the
upper chamber governs sensitivity of the instrument. If the pointer is too
sensitive, it may be reduced by moving the upper impulse wheel closer to the
front of the chamber, by turning the upper impulse wheel bearing
counterclockwise. If the pointer lacks sensitivity (and if there is no evidence of
mechanical binding), it may be increased by moving the upper impulse wheel
closer to the rear of the chamber, by turning the upper impulse wheel bearing
clockwise. This adjustment requires loosening the 7/8" hex lock nut behind the
measuring wheel, turning the 1/2" hex head of the bearing in correct direction,
and tightening the lock nut. Turn only one flat (1/6 turn) at a time and retest.
3. Span Adjustment. The span of the gravitometer is governed by the angular
position of the pointer on the measuring wheel. To permit adjustment, the
pointer is attached to the measuring wheel by two screws which pass through
circular slots in the pointer hub. If span adjustment is necessary, the pointer
should be moved in relation to the measuring wheel in the direction of the
desired reading. For example, if the pointer reads numerically too high on the
red scale (too low on the black scale), the pointer should be moved clockwise.
Similarly, if the pointer reads numerically too low on the red scale (too high on
the black scale), the pointer should be moved counterclockwise.
Before making an adjustment, mark the position of the pointer hub on the face
of the cam, then loosen the screws just enough that the pointer can be turned
against slight friction of the washers under the screw heads, but that it will not
slip freely. Make adjustments in small increments of 2 or 3 small graduations
on the red scale at one time and retest. When either the sensitivity or span
adjustment is changed, it is necessary to reset the zero adjust to 1.000 on air.