CAPPO 100

USER MANUAL
The cutting edge Portable Multicalibrator for
Temperature, Pressure and Electrical Signals

temperatura industrial
With more than 91 years of history, Ecil is the Brazilian leader on industrial temperature
measurement and control. We are specialists in offering the most complete solutions in this area for
the most demanding industry segments such as: Oil & Gas, Steelmaking, Metalworking, Chemical,
Cement, Food, Pharmaceutical, Plastic, Glass, Pulp & Paper etc.

We offer to the market a wide range of products that consists of temperature sensors, such
as thermocouples, RTs (Resistance Thermometers), infrared thermometers and other instruments
for temperature measurement, transmission, insulation, control, recording and calibration. We also
provide solutions in pressure calibration and flue gas analysis.

Our Metrology Laboratory was the first industrial lab to be accredited by Cgcre, the Brazilian
General Accreditation Coordination of INMETRO (Brazilian National Metrology Laboratories). It offers
to our clients services such as Temperature Calibration for all kinds of temperature sensors, Electrical
Signals Calibration and Flue Gas Analyzers Calibration. The laboratory is also used in our in-house
made temperature sensor’s raw material testing and instrument’s tests and calibration.

Our team of specialized and proven in many years of market service engineers is ready to assist
you with the best value for the money choice for the needs of your company.

For more information access


www.ecil.com.br/en

temperatura industrial
Ecil Produtos - ALL RIGHTS RESERVED -
This document’s total or partial reproduction, regardless of its nature (including photocopies e
digital storage) and transmission to third parties, no matter what medium, even electronically, is
strictly forbidden, unless explicitly authorized by Ecil Produtos.
TABLE OF
CONTENTS
1. IMPORTANT INFORMATION 6
1.1 Information about this manual 6
2. OPERATING PRINCIPLES 7
2.1 General Description 7
2.2 Component’s Description 7
2.2.1 Component’s Description> Screen 7
2.2.2 Component’s Description> Connections 8

3. TECHNICAL SPECIFICATIONS 9
4. PRESENTATION 12
4.1 Layout 12
4.2 Display 13
4.3 Keyboard 15
5. START 16
5.1 Connection Schemes - Input 16
5.2 Connection Schemes - Output 17
5.3 On / Off 18
5.4 Built-In Help 20
6. CONFIGURATION 22
6.1 Colors 23
6.2 Header 25
6.3 Probe 26
6.4 Auto-Ramp 32
6.5 SCALE X 36
6.6 Pressure Units 42
6.7 Reference Junction 43
6.8 Leak Test Time 45
6.9 Battery Saving 46
6.10 Protection 48

4 manual_c100_jvscf_v4 temperatura industrial


7. Input 51
7.1 Thermocouple Input 53
7.2 RT Input 55
7.3 Current Input 56
7.4 Voltage Input 57
7.5 Electrical Resistance Input 58
7.6 SCALE X Input 59
7.7 Pressure Input 60
7.8 Leak Test 61
7.9 Probe Input 62
8. Output 63
8.1 Thermocouple Output 65
8.2 RT Output 67
8.3 Current Output 68
8.4 Voltage Output 69
8.5 Electrical Resistance Output 70
8.6 SCALE X Output 71
8.7 Pressure Output 72
8.8 Auto-Ramp Output 73
8.9 Switch Test 75
9. MEMORY 76
10. SERVICES AND ASSISTANCE 77
10.1 Services 77
10.2 Assistance 77
Annex 1 - User Interface 78

5 manual_c100_jvscf_v4 temperatura industrial


1. IMPORTANT
INFORMATION
1.1 Information about this manual
This manual describes the operation and characteristics of Ecil’s Cappo 100 Multicalibrator.

Read this operation manual before utilizing it. The user must be familiarized with the instrument and
follow the instructions carefully.

This user manual is subject to changes due to technical improvements; The manufacturer does not
assume any liability for any printing error or flaw.

6 manual_c100_jvscf_v4 temperatura industrial


2. OPERATING
PRINCIPLES
2.1 General Description
Ecil Produtos introduces the Cappo 100 portable calibrator, totally developed with in-house technology
and 100% locally manufactured.

The calibrator is light and compact, powered by 3.7Vdc @ 5,200mA Li-Ion batteries, the plastic outer
casing has a modern design, with a touchscreen, 65,536 colors, 272x480 pixels graphic display. It also has a 5
buttons tactile keyboard, giving the user an alternative way to navigate the menus.

In order to better utilize the resistive touchscreen display, we recommend the use of an object
with a non-piercing soft and thin tip.

The instrument has two simultaneous and independent universal channels that are isolated from each
other. The Input channel is used for measurement, while the Output channel is used for simulation. Both
channels accept the following types of signals; Thermocouples, RTs, mV, V, mA and Ohmic Resistance. Each
channel has its own 24V @ 30mA power source, also isolated from each other.

As an optional feature, the instrument can also have two more configurable channels for pressure
measurement, one internal (CP1) and the other external (CP2).

All measurements and simulations are made using high-resolution integrated circuits (24bits), with
sophisticated conditioning, filtering and amplification systems, thereby giving the product excellent precision
and thermal stability over time.

Cappo 100 is supplied with a calibration certificate accredited by Cgcre.

2.2 Component’s Description


2.2.1 Component’s Description> Screen
All parameters are visualized on a backlit color 480x272 touchscreen TFT display, allowing lots of visual
space to visualize and configure all operations’ parameters.

7 manual_c100_jvscf_v4 temperatura industrial


2.2.2 Component’s Description> Connections
The input and output connections are made through the 4 golden plated terminals located on the top
part of the instrument, as illustrated below:

There are two options for pressure modules, one internal and one input for external modules, CP1 and
CP2 respectively, located on the bottom part of the instrument.

On the right-hand side there is an USB type B input for the battery recharging.

8 manual_c100_jvscf_v4 temperatura industrial


3. TECHNICAL
SPECIFICATIONS
Input/Output – Voltage mV and V
Range Resolution Accuracy
-21 to 210mV 1µV ± 0.02% of reading + 3µV
-210 to 2,100mV 10µV ± 0.02% of reading + 10µV
-2,1 to 21V 100µV ± 0.02% of reading + 0.1mV

Input/Output – Current mA
Range Resolution Accuracy Observation
0 to 51mA 0.1µA ± 0.02% of reading + 0.4µA Active or Passive loop

Passive loop: Considers that the Calibrator is being used in a non-energized loop, meaning that it must
supply power to the circuit itself, therefore the 24V source of the input channel will be turned on and the max
current of the measurement will be 51mA.

Active loop: Considers the Calibrator is being used in an energized loop, meaning that the circuit is
powered by an external power supply, therefore the 24V source of the input channel will be turned off and the
max current of the measurement will be 51mA.

Input – Resistance Ω
Range Resolution Accuracy Observation
0 to 500 Ω 1mΩ ± 0.02% of reading + 20mΩ 2 Wires Output
0 to 4,000 Ω 10mΩ ± 0.02% of reading + 200mΩ 2 Wires Output

9 manual_c100_jvscf_v4 temperatura industrial


Input/Output – RTs
RT Type Range Resolution Accuracy
Pt100 IEC -200ºC (-328ºF) to 850ºC (1,562ºF) 0.01ºC/F ± 0.02% of reading + 0.05ºC (0.09ºF)
Pt100 JIS -200ºC (-328ºF) to 600ºC (1,112ºF) 0.01ºC/F ± 0.02% of reading + 0.05ºC (0.09ºF)
Pt100 SAMA -200ºC (-328ºF) to 600ºC (1,112ºF) 0.01ºC/F ± 0.02% of reading + 0.05ºC (0.09ºF)
Pt1000 IEC -200ºC (-328ºF) to 600ºC (1,112ºF) 0.01ºC/F ± 0.02% of reading + 0.05ºC (0.09ºF)
Cu10 -100ºC (-148ºF) to 260ºC (500ºF) 0.10ºC/F ± 0.02% of reading + 0.40ºC (0.72ºF)
Ni120 -80ºC (-112ºF) to 260ºC (500ºF) 0.01ºC/F ± 0.02% of reading + 0.05ºC (0.09ºF)

Input/Output – Thermocouples
TC Type Range Resolution Accuracy
J -210ºC (-346ºF) to 1,200ºC (2,192ºF) 0.01 ºC/F ± 0.02% of reading + 0.1ºC (0.18ºF)
K -270ºC (-454ºF) to 1,372ºC (2,501.60ºF) 0.01 ºC/F ± 0.02% of reading + 0.1ºC (0.18ºF)
N -270ºC (-454ºF) to 1,300ºC (2,372ºF) 0.01 ºC/F ± 0.02% of reading + 0.1ºC (0.18ºF)
E -270ºC (-454ºF) to 1,000ºC (1,832ºF) 0.01 ºC/F ± 0.02% of reading + 0.1ºC (0.18ºF)
T -210ºC (-346ºF) to 400ºC (752ºF) 0.01 ºC/F ± 0.02% of reading + 0.1ºC (0.18ºF)
R -50ºC (-58ºF) to 1,768ºC (3,214.40ºF) 0.10 ºC/F ± 0.02% of reading + 0.2ºC (0.36ºF)
S -50ºC (-58ºF) to 1,768ºC (3,214.40ºF) 0.10 ºC/F ± 0.02% of reading + 0.2ºC (0.36ºF)
B 0ºC (32ºF) to 1,820ºC (3,308ºF) 0.10 ºC/F ± 0.02% of reading + 0.3ºC (0.54ºF)

Obs: types J, K, N, E and T thermocouples accuracy, in temperatures below -200°C (-328°F), will be ±
0,02% of the reading +1.0°C (+ 1.8°F). The error limits stand for 360 days for operation between 18°C (64.4°F)
to 28°C (82.4°F), out of this range the thermal stability is of 0.001 % FS / °C/F.

10 manual_c100_jvscf_v4 temperatura industrial


Pressure Modules
Range Resolution Accuracy
-100 to 100 mbar 0.001mbar ±0.025 % of full scale
-300 to 300 mbar 0.01mbar ±0.025 % of full scale
-500 to 500 mbar 0.01mbar ±0.025 % of full scale
-1 to 1 bar 0.01mbar ±0.025 % of full scale
-1 to 2 bar 0.01mbar ±0.025 % of full scale
-1 to 3 bar 0.01mbar ±0.025 % of full scale
0 to 2 bar (ABS) 0.01mbar ±0.025 % of full scale
-1 to 10 bar 0.1mbar ±0.025 % of full scale
-1 to 20 bar 0.1mbar ±0.025 % of full scale
0 to 20 bar (ABS) 0.1mbar ±0.025 % of full scale
-1 to 25 bar 1 mbar ±0.025 % of full scale
-1 to 30 bar 1 mbar ±0.025 % of full scale
-1 to 40 bar 1 mbar ±0.025 % of full scale
0 to 70 bar 1 mbar ±0.025 % of full scale
0 to 100 bar 1 mbar ±0.025 % of full scale
0 to 150 bar 1 mbar ±0.025 % of full scale
0 to 300 bar 10 mbar ±0.025 % of full scale
0 to 350 bar 10 mbar ±0.025 % of full scale
0 to 350 bar (ABS) 10 mbar ±0.025 % of full scale
0 to 700 bar 10 mbar ±0.025 % of full scale
0 to 700 bar (ABS) 10 mbar ±0.025 % of full scale
0 to 1,000 bar 10 mbar ±0.025 % of full scale

Max pressure supported: 125% of full scale

11 manual_c100_jvscf_v4 temperatura industrial


4. PRESENTATION

4.1 Layout
Pt100 - Input Channel’s
Pt100 - Output Channel’s Reference Junction
Reference Junction
Input Channel’s Golden
Plated Terminals
Output Channel’s Golden
Plated Terminals

480x272 pixels
Touchscreen Color Display

Battery Charging
Input

ON/OFF
Button
Tactile Keyboard

Pressure Module
(optional)

12 manual_c100_jvscf_v4 temperatura industrial


4.2 Display
All parameters are visualized on a 480x272 pixels backlit display, providing a large viewing area to fa-
cilitate the visualization of the operations’ parameters measurements and configurations.
The display is split in two sections, Upper and Lower, that allows two variables to be displayed simulta-
neously.

CAPPO 100

Input
TC K Rj External
Upper Section

234.56
Output
200mV Lower Section

-1234.56
3.6V

3.8V 3.7V 3.5V 3.3V 3.2V 3.2V

Input Output Config P


The display’s Upper and Lower Sections’ first line identifies the variable type that is being displayed or
simulated, working range, measurement unit and other details relating to the variable type.

On the lower right-hand corner, right below the Lower Section and over the Config icon, an icon in-
dicating the battery level is dynamically displayed. The letter P in the icon indicates that the instrument is
connected to the Power Supply.

Whenever the battery reaches a critical level (red color), a message ‘BATTERY LOW’ is displayed and
the instrument is automatically shut down.

A battery recharger (110/220Vac input and 5V@1A output) is supplied with the calibrator, the usual re-
charging time with the instrument turned off is 6 hours. Other types of recharging methods not supplied with
the instrument are not recommended.

The battery life is approximately 6 hours for Thermocouples, RTs, mV and V Input/Output and 3 hours
for 20mA current simulation in passive mode.

13 manual_c100_jvscf_v4 temperatura industrial


To configure the screen brightness and the instrument’s beep volume, just tap on the small triangle
located in the lower left corner of the Home Screen header, that will display both the brightness’ and beep
volume’s increase and decrease icons.

CAPPO 100

Input Input
TC K Rj External TC K Rj Externa

234.56 234.56
Output Output
200mV 200mV

-1234.56 -1234.56
3.6V 3.6V

Input Output Config Input Output Config

14 manual_c100_jvscf_v4 temperatura industrial


4.3 Keyboard

The configuration and operation of the Cappo 100 is very simple and all parameters can be accessed via
the 8-key tactile keyboard, with the following functions:

Button Function
ON/OFF Turning the Instrument On and Off
Arrow Keys Navigate through Instrument Icons and Parameters
ENTER Check to choice or the parameter value
INPUT Go to Input Menu
OUTPUT Go to Output Menu
CONFIG Go to Config Menu

15 manual_c100_jvscf_v4 temperatura industrial


5. START

5.1 Connection Schemes - Input

16 manual_c100_jvscf_v4 temperatura industrial


5.2 Connection Schemes - Output

17 manual_c100_jvscf_v4 temperatura industrial


5.3 Turning On and Off
To turn on the instrument, just press the On/Off button:

And the boot screen will be displayed Just tap anywhere and the
Home Screen will be displayed:

CAPPO 100

Input
TC K Rj External

234.56
Output
200mV

-1234.56
3.6V

Input Output Config

Curiosity: The aerial photo on the Home Page is of our new headquarters, made up of three buildings
inaugurated in early 2017. The facilities are very modern, which allows us to produce cleaner and more ef-
ficiently, with greater safety for our employees and guaranteeing the satisfaction of our customers, our top
priority. Come visit us!

18 manual_c100_jvscf_v4 temperatura industrial


To turn off the instrument, just press the On/Off button:

And the shutdown screen will be displayed:

ON/OFF KEY PRESSED


TURNING THE INSTRUMENT OFF

19 manual_c100_jvscf_v4 temperatura industrial


5.4 Built-In Help
The Cappo 100 comes with Help screens built into the system, through which it is possible to obtain
technical and operational information about the instrument. The first of these is introductory and can be ac-
cessed via the boot screen; Just press the Info button at the center of the lower part of the screen.

After pressing the Info button the New Cappo 100 Introduction screen will appear:

20 manual_c100_jvscf_v4 temperatura industrial


The other help screens can be accessed by holding down any of the icons on the Input, Output and
Config screens for three seconds. Below is an example of the RT Output icon help screen:

Many of the screens, as shown above, have more than one page, to scroll between them just use your
fingers and swipe up and down, making a screen scroll. To exit any help screen, simply tap on the Home icon
in the right-hand corner of the header at the top of the screen.

21 manual_c100_jvscf_v4 temperatura industrial


6. SETTINGS

On the Home Screen, at the bottom right-hand corner, tap on the Config Icon or press the CONFIG but-
ton on the physical keyboard and the following screen will be displayed:

Config
Ch1
Ch2

COLORS HEADER PROBE

a--b
x--y
c--d

AUTORAMPA
AUTORAMP ESCALA
X SCALEX PRESSURE
UNIT
mbar

REFER. LEAK TEST BATTERY


JUNCT. TIME SAVING

15s

PROTECTION Sair
Exit

22 manual_c100_jvscf_v4 temperatura industrial


6.1 Colors
On the Config screen, tap on the COLOR icon and the following screen will be displayed:

Ch1 Ch1 Ch1


Ch2 Ch2 Ch2

HEADER CH1 CH2

Sair
Exit

This screen allows you to select the colors of the header, input channel and output channel.

23 manual_c100_jvscf_v4 temperatura industrial


Header Input or Output Channel

CAPPO 100
Input
200mV

1234.56
EXIT DEFAULT EXIT DEFAULT

TEXT BACKGR. TEXT BACKGR.

FRAME NUMBER

10
15

63
56

0
21

Through 3 colored bars, red, green and blue, up to 65,000 color combinations can be adjusted, note
that each bar can be adjusted independently of one another.

Just tap on the bar of one of the colors and drag the white ball to the desired value, checking the result
in the model shown above the bars.

As for the header, the text and background colors can be adjusted.

For the input and output channels, the text colors, background, frame and number of each channel can
all be adjusted independently.

The DEFAULT option recovers the original colors defined at the factory and the EXIT option finishes the
configuration, then you can choose to save or not the modifications made.

24 manual_c100_jvscf_v4 temperatura industrial


6.2 Header
On the Config screen, tap on the HEADER icon and the following dialog box will be displayed:

CAPPO 100

DO YOU WISH TO USE THE


HORIZONTAL KEYBOARD? ?
YES NO

When tapping on the YES option, a horizontal keyboard will be displayed on the screen so the header
text can be edited with the following settings:

X1 Size: Two lines with 30 characters each


X2 Size: Two lines with 15 characters each
X3 Size: One line with 11 characters
X4 Size: One line with 8 characters

An alphanumeric keyboard drawn horizontally on the display must be used to input the desired text.

As the keys are of reduced dimensions, in order to type the text it is recommended to use an object
with a non-piercing fine and soft tip.

The EXIT key finishes the configuration and allows the user to whether or not save the changes made.

X1

X3
X2

X4
CAPPO 100 _ EXIT

1 2 3 4 5 6 7 8 9 0 @

q w e r t y u i o p

a s d f g h j k l ç ^

z x c v b n m \ / % ´
o
_ ( ) \ . - +

25 manual_c100_jvscf_v4 temperatura industrial


6.3 Probe

On the Config screen, tap on the PROBE icon and the following screen will be displayed:

PROBE 1 PROBE 2 PROBE 3

Sair
Exit

26 manual_c100_jvscf_v4 temperatura industrial


By tapping on one of the three icons the following screen will be displayed with the parameters that
must be defined for the chosen probe:

mV mA TIC mV mA
A B A B TIC A B A B
V Tc FIC V Tc FIC
Pt100 NOME? C RO C RO Pt100 NOME? C RO C RO

TYPE TAG A COEF. B COEF. TYPE TAG A COEF. B COEF.

A B A B 2+2 A B 2+2
2x2 2x2
C RO C RO =4 C RO =4

C COEF. RO COEF. TEST Exit C COEF. TEST Exit

Type: Pt 100 4 Wires

TAG: AXXXXXXX Type: TC R Rj External

A COEFFICIENT: -6.62280E-02 TAG: AXXXXXXX

B COEFFICIENT: +7.91540E-05 A COEFFICIENT: -6.62280E-02

C COEFFICIENT: -1.65120E-08 B COEFFICIENT: +7.91540E-05

RO COEFFICIENT:+0.00000E-00 C COEFFICIENT: -1.65120E-08

Up to 3 (three) Probes can be configured, and for each one parameters bellow it must be defined:

Type: Thermocouple type R or S or Pt100 IEC.

Tag: Through the horizontal alphanumeric keyboard, you can define a tag (name) to identify the Probe with up
to 8 characters.

Coefficient A: As stated in the thermocouple or Pt100 sensor calibration certificate, being a number in scien-
tific notation format (mantissa + exponent) within the range from +9.99999E+33 to -9.99999E-33.

Coefficient B: Same as coefficient A

Coefficient C: Same as coefficient A

Coefficient RO (Resistance at 0ºC): Same as the coefficient A, but it is only shown if the type is Pt100.

27 manual_c100_jvscf_v4 temperatura industrial


By tapping on the TYPE icon, the following screen will be displayed:

Pt 100 TC R TC S
IEC

Exit

Just tap on the desired sensor type; If you choose a thermocouple, you will have to choose the type of
reference junction, internal or external, through a dialog box like the one below:

CAPPO 100

SELECT T H E
REFEREN C E J U N C TION ?
INTERNAL EXTERNAL

28 manual_c100_jvscf_v4 temperatura industrial


By tapping on the TAG icon, a horizontal keyboard will be displayed where you can input a tag for the

_
probe:

AXXXXXXX EXIT

1 2 3 4 5 6 7 8 9 0 @

q w e r t y u i o p

a s d f g h j k l ç ^

z x c v b n m \ / % ´
o
_ ( ) \ . - +

After finishing, just tap on EXIT and a dialog box will appear so you can choose to save or not:

CAPPO 100

S AV E THE
A LTERA TI O NS? ?
YES NO

29 manual_c100_jvscf_v4 temperatura industrial


The next step is to define the coefficients A, B, C and RO (if the sensor type is Pt100), just tap on one of
the icons and the following screen will be displayed:

Limits
Max : +9.99999E+33
M i n : -9.99999E-33
R e s o l : 0.00001

-6.62280E-02

7 8 9 E

4 5 6 C

1 2 3

0 . +
-
ESC

Use the keyboard to set the value, E to confirm, C to clear, the arrow key to navigate between digits and
ESC to cancel.

30 manual_c100_jvscf_v4 temperatura industrial


Once the coefficients are configured, the TEST icon allows a quick test, by typing the EMF (Electro-Mo-
tive Force) of the thermocouple sensor or the Electrical Resistance of the Pt100 and comparing the calculated
with the expected temperature.

Enter the EMF in uV Enter the Resistance

0 0

7 8 9 E 7 8 9 E

4 5 6 C 4 5 6 C

1 2 3 1 2 3

0 . +
-
ESC
0 . +
-
ESC

31 manual_c100_jvscf_v4 temperatura industrial


6.4 Auto-Ramp

The Auto-Ramp function allows the generation of an output that will automatically follow a curve pre-
defined by the user.

On the Settings screen, tap on the AUTORAMP icon and the following screen will be displayed, with
icons that give access to the parameters’ configuration at the top part of the screen and a table containing the
parameters set at the bottom:

mV mA
V Tc
Pt100

TYPE START END STEP

TSTPUP TSTPDWN TSOAKUP TSOAKDWN

CYCLES Exit
Sair

TYPE: TC K Rj External

R A´ M P S T A R T : -270.00

RAMP END: 1372.00

STEP: 1.00
STEP UP TIME: 1
STEP DOWN TIME: 1

SOAK UP TIME: 10
SOAK DOWN TIME: 20
N U´ M B E R O F C Y C L E S : 1

32 manual_c100_jvscf_v4 temperatura industrial


At first, it is necessary to choose the Auto-Ramp’s type of output signal; Just tap on the TYPE icon, and
the following screen will be displayed:

THERMOCOUPLE RT CURRENT

VOLTAGE RESISTANCE Sair


Exit

Just browse the icons and choose the output signal; However, if you choose Thermocouple or Current,
you will need to choose the type of reference junction for the type of thermocouple chosen and between pas-
sive or active loop for current, all should be set through the dialog boxes below:

CAPPO 100 CAPPO 100

SE L E CT T H E SELECT THE TY PE O F
RE F E RE NCE J U N C T I ON ? CURRENT LO O P ?
INTERNAL EXTERNAL PASSIVE ACTIVE

33 manual_c100_jvscf_v4 temperatura industrial


To configure the other parameters just tap on their respective icons and use the numeric keypad as
displayed below:

Limits Limits
Max : 1372.00 Max : 51.0000
M i n : - 270.00 Min : 0.0000
Resol : 0.01 Resol : 0.0001

0 0

7 8 9 E 7 8 9 E

4 5 6 C 4 5 6 C

1 2 3 1 2 3

0 . +
-
ESC
0 . +
-
ESC

Use the numeric keypad to set the value, E to confirm, C to clear, the arrow key to navigate between
digits and ESC to cancel. Below is the description of the selectable parameters:

Parameter Function
START Ramp’s starting point value
END Ramp’s ending point value
STEP Value for each step of the upward or downward ramp
TSTPUP Time for each step of the upward Ramp
TSTPDWN Time for each step of the downward Ramp
TSOAKUP Time for the upper level
TSOAKDWN Time for the lower level
CYCLES Numbers of iterations of the curve

All time values are in seconds.

34 manual_c100_jvscf_v4 temperatura industrial


To trigger the Auto-Ramp, go to the Output Menu and tap on the AUTORAMP icon and you will be re-
directed to the Start Screen, where in the upper right-hand corner of the output part, an indication that the
Ramp is being executed will appear, as displayed below:

CAPPO 100

Input
TC K Rj External

134.05
UPWARD RAMP
Output
TC J Rj Internal

234.56
3.6V

Input Output Config

35 manual_c100_jvscf_v4 temperatura industrial


6.5 SCALE X
The SCALE X function allows you to display an input or simulate an output in an user-defined engi-
neering unit using correlation (linear interpolation).

Through this function, the user can display a linear input or simulate a linear output such as Voltage
or Current, scaled as needed, through the following equation:

Input: Y = X% . (DispHi - DispLo) + DispLo

Start of Range (RangeLo) Initial value to display (DispLo)


Value measured/simulated (X) Adjusted Output Value (Y)
End of Range (RangeHi) Final value to display (DispHi)

Output= RangeLo = 0% Output start of range


Output =X= X% Value between O to 100%
Output= RangeHi = 100% Output end of range

The decimal point position of the value to be displayed is freely configurable, also, an 8-character
unit can be assigned to it.

On the Config screen, tap on the SCALE X icon and the following screen will be displayed, with icons
that give access to the parameters’ configuration at the top part of the screen and a table containing the
parameters set at the bottom:

mV mA HI HI
V Tc
Pt100 LO LO

T IPO
TYPE RANGE DISPLAY

m 3 /s ax+b
Kg/h ax +b
2

UNITY FUNCTION Sair


Exit

Type: 200mV

Range low: -21.00

Range high: 210.00


Display low: 0.000
Display high: 100.000
Unity: m/s
Function: Quadratic

36 manual_c100_jvscf_v4 temperatura industrial


At first, it is necessary to choose the type of output signal for SCALE X; Just tap on the TYPE icon, and
the following screen will be displayed:

200 mV 2000 mV 20 V

CURRENT
CORRENTE Sair
Exit

Just browse the icons and choose the output signal; However, if you choose Current, you will need to
choose between passive or active loop via the dialog box below:

CAPPO 100

S ELEC T THE TY PE O F
C U R R EN T LO O P ?
PASSIVE ACTIVE

37 manual_c100_jvscf_v4 temperatura industrial


To configure the start and end of the input range, just tap on the RANGE icon, and a dialog box will be
displayed:

CAPPO 100

AD J U S T T H E R A N G E ?
LOW HIGH

Having chosen the start (LOW button) or the end (HIGH button) of the input range, you must use the
numeric keypad as displayed below:

Limits
Max : 2100.00
Min : -210.00
Resol : 0.01

7 8 9 E

4 5 6 C

1 2 3

0 . +
-
ESC

Use the numeric keypad to set the value, E to confirm, C to clear, the arrow key to navigate between
digits and ESC to cancel.

38 manual_c100_jvscf_v4 temperatura industrial


To configure the initial and final values for the Display, just tap on the DISPLAY icon, and a dialog box
will be displayed asking if you want to adjust the decimal point position beforehand.

CAPPO 100

ADJUST THE DECIMAL


POINT? ?
YES NO

CAPPO 100 CAPPO 100

DECIMAL POINT ADJUSTMENT


? ADJUST THE DISPLAY ?
0.000

<- - - END LOW HIGH

Having chosen the initial (LOW) and final (HIGH) values for the Display, you must use the numeric
keypad as displayed below:

Limits
Max : 999.999
Min : -999.999
Resol : 0.001

7 8 9 E

4 5 6 C

1 2 3 Use the numeric keypad to set the value, E to


confirm, C to clear, the arrow key to navigate between

0 . +
-
ESC
digits and ESC to cancel.

39 manual_c100_jvscf_v4 temperatura industrial


By tapping on the UNIT icon, a horizontal keyboard will be displayed and you may type the scale unit
using up to 9 characters:

m/s _ EXIT

1 2 3 4 5 6 7 8 9 0 @

q w e r t y u i o p

a s d f g h j k l ç ^

z x c v b n m \ / % ´
o
_ ( ) \ . - +

After finishing, just tap on EXIT and a dialog box will appear so you can choose to save or not:

CAPPO 100

S A V E T HE
A L T E R ATIO NS? ?
YES NO

By tapping on the FUNCTION icon, a dialog box will be displayed to choose the type of scale function:

CAPPO 100

USER SCALE
C H OOS E T H E F U N C T I O N ?
LINEAR QUADR.

40 manual_c100_jvscf_v4 temperatura industrial


In order to use the created scale, just access the Input or the Output Menu and tap on the SCALE X icon
to be redirected to the Start Screen, where either on Input, Output or both, your scale will be displayed:

CAPPO 100

Input
X SCALE m/s

90.961
Output
200mV

0.00
3.6V

Input Output Config

41 manual_c100_jvscf_v4 temperatura industrial


6.6 Pressure Unit

To choose the pressure unit, just tap on the PRESSURE UNIT icon and the following screen will be dis-
played:

mbar bar KPa Kg/cm2

mmHg cmHg mmH2O cmH2O

mH2O Torr atm psi

inHg inH2O Sair


Exit

Then just browse the icons and choose the unit.

42 manual_c100_jvscf_v4 temperatura industrial


6.7 Reference Junction

The Reference Junction function allows the temperatures measured on the Pt100 sensors installed in
the terminal 2 of the Input and Output channels to be viewed.

The temperature measured by these sensors will be used as compensation of the reference junction in
case the channel (Input and/or Output) is configured as thermocouple and Rji (Internal Reference Junction).

When the Rje (External Reference Junction) option is used, the temperature used as compensation is
set in this menu.

See chapters 7.1. Thermocouple Input and 8.1. Thermocouple Output for more information on selecting
the Internal or External Reference Junctions.

To adjust the temperatures to be used as compensation for the Input and/or Output channel’s ther-
mocouple reference junction when the Rje option is selected and to view the Input and/or Output channel’s
temperature readings at terminal 2 when the option Rji is selected, just tap on the REFER JUNCRT icon, and
the following screen will be displayed:

CH1 CH2
EXT EXT

0.00 0.00 Sair


Exit

CH1 CH2
INT INT

27.40 27.52

43 manual_c100_jvscf_v4 temperatura industrial


To adjust the temperatures to be used as compensation when choosing Rje, just tap on one of the first
two icons in the top row, CH1 EXT or CH2 EXT, and then enter the value using a numeric keypad like the one
below:

Limits
Max : 50.00
Min : - 50.00
Resol : 0.01

7 8 9 E

4 5 6 C

1 2 3

0 . +
-
ESC

Use the numeric keypad to set the value, E to confirm, C to clear, the arrow key to navigate between
digits and ESC to cancel.

44 manual_c100_jvscf_v4 temperatura industrial


6.8 Leak Test Time

This function allows you to enter a number from 1 to 1800 seconds that will be the duration of the Leak
Test.

To set this time, just tap on the LEAK TEST TIME icon and a screen with a numeric keypad will be dis-
played like the one below:

Limits
Max : 1800
Min : 1
Resol : 1

7 8 9 E

4 5 6 C

1 2 3

0 . +
-
ESC

To start the test, just go to the Input Menu and tap on the LEAK TEST icon and you will be be redirected
to the Start Screen, where the test will be performed in the Input section with the time that was set here.

You will find more information about the test in chapter 7.8 Leak Test, in the Input section.

45 manual_c100_jvscf_v4 temperatura industrial


6.9 Battery Saving

To maximize battery usage, there are two icons, BRIGHTNESS and POWER OFF, which allow a time from
0 to 20 minutes to be entered for each of them. To set this time, just tap on the BATTERY SAVING and you will
be directed to the screen below:

BRIGHTNESS POWER OFF Sair


Exit
OFF OFF

To maximize battery usage:


BRIGHTNESS function, allows a
time to be set, during which, if no
key or display function is activat-
ed, the brightness will be reduced
by 90%.
POWER OFF function, similar to the
BRIGHTNESS function, however,
after the time has elapsed, the in-
strument is turned off.
Adjustable time from 0 to 20 min,
where zero = function disabled.

BRIGHTNESS: After the set time elapses and if no keyboard key or display function is activated, the
brightness will be reduced to 10% of the original value.
By typing on the tactile keyboard or touching the display, the original brightness will be restored.

POWER OFF: After the set time elapses and if no keyboard key or display function is activated, the in-
strument will be turned off.

The charger and USB cable are provided with the instrument and due to the high charging current, it
is not recommended to charge the instrument through a computer, i.e., for best performance, always use the
charger and cable provided for this function.

The battery icon shown on the main screen indicates the charge level and the voltage value, when the
level reaches 3.2V the on/off button led starts to flash and when the voltage reaches 3.1V, the instrument will
be automatically turned off. To turn it on again, plug in the battery charger.

46 manual_c100_jvscf_v4 temperatura industrial


To set the times, just tap on one of the two available icons, BRIGHTNESS and POWER OFF, and a screen
will be displayed with a numeric keypad like the one below:

Limits
Max : 20
Min : 0
Resol : 1

7 8 9 E

4 5 6 C

1 2 3

0 . +
-
ESC

Use the numeric keypad to set the value, E to confirm, C to clear, the arrow key to navigate between
digits and ESC to cancel.

47 manual_c100_jvscf_v4 temperatura industrial


6.10 Protection

Through this function, it is possible to define which icons of the Config Menu will have its access locked.
To use this functionality it is necessary to know the input password (Default Password: 2112); Just tap on the
PROTECTION icon and you will be directed to the screen below:

ENTER THE PASSWORD

Limits
Max : 999999
Min : 0
Resol : 1

7 8 9 E

4 5 6 C

1 2 3

0 . +
-
ESC

Use the numeric keypad to set the password, E to confirm, C to clear, the arrow key to navigate be-
tween digits and ESC to cancel.

48 manual_c100_jvscf_v4 temperatura industrial


If the password is entered correctly, the nine Config Menu icons will be available, namely:

Colors, Header, Probe, Auto-Ramp, SCALE X, Pressure Unit, Reference Junction, Leak Test Time, and
Battery Saving.

Each icon will be displayed according to its status:

Locked: Red color and closed padlock; It will not allow parameters changes in the Config Menu.

Unlocked: Green color and open padlock; It will allow parameters changes in the Config Menu.

By touching the icon or using the keyboard’s ENTER key, you can change the selected icon’s status from
Locked to Unlocked and vice versa.

Config Config

COLORS HEADER PROBE COLORS HEADER PROBE

AUTORAMP X SCALE PRESSURE AUTORAMP X SCALE PRESSURE


UNIT UNIT
mmHg mmHg

REFER. LEAK TEST BATTERY REFER. LEAK TEST BATTERY


JUNCT. TIME SAVING JUNCT. TIME SAVING
7s 7s

CHANGE Exit
Sair CHANGE Exit
Sair
PASSWORD PASSWORD

49 manual_c100_jvscf_v4 temperatura industrial


A tenth icon, Change Password, will also be displayed and allows the password to be changed, to do so,
it will be necessary to type twice the same sequence of numbers from 1 to 6 digits, without decimal point or
polarity sign (±).

ENTER NEW PASSWORD

Limits
Max : 999999
Min : 0
Resol : 1

7 8 9 E

4 5 6 C

1 2 3

0 . +
-
ESC

Use the numeric keypad to set the password, E to confirm, C to clear, the arrow key to navigate between
digits and ESC to cancel.

After two valid sequences, a confirmation will be asked to save the new password, if the option YES is
selected, the new password will be saved.

Write down the new password, because if you forget it, access to the Config Menu may be limited.

50 manual_c100_jvscf_v4 temperatura industrial


7. Input

On the Start Screen, at the bottom left corner, tap on the Input Icon or press the Input button on the
tactile keyboard and the following screen will appear:

Input

THERMOCOUPLE
TERMOPAR RT
TERMO CURRENT
CORRENTE
^
RESISTENCIA

a--b
x--y
c--d
~ ^
TENSAO
VOLTAGE RESISTANCE
RESISTENCIA X SCALE

PRESSURE LEAK TEST PROBE

Exit
Sair

By pressing and holding any for approximately 3 seconds, The Built-In Help will be displayed with useful
information about the input signal selected.

51 manual_c100_jvscf_v4 temperatura industrial


Tap on the desired option or navigate through the tactile keyboard’s cursor keys and press ENTER to
choose between the 9 input channel options.

If there is no instrument connected generating any input signal, up or down arrows will be displayed
indicating that the input is open.

CAPPO 100

Input
TC K Rj External

Output
200mV

0.00
3.6V

Input Output Config

52 manual_c100_jvscf_v4 temperatura industrial


7.1 Thermocouple Input

To configure the instrument’s input for Thermocouple sensors, just tap on the Thermocouple icon and
the following screen will be displayed:

TC J TC K TC N

TC E TC T TC R

TC S TC B Exit
Sair

To choose the thermocouple type, just tap on one of the icons or navigate through the tactile key-
board’s cursor keys and press ENTER to confirm. After doing so, a dialog box will give the option to choose the
Reference Junction, internal or external.

CAPPO 100

SELECT THE
REFERENCE JUNCTION ?
INTERNAL EXTERNAL

53 manual_c100_jvscf_v4 temperatura industrial


When the Internal RJ option is selected, the ambient temperature will be measured through a Pt100
installed on the input channel’s terminal 2 (see Chapter 4.1 Layout), and will be used to compensate for refer-
ence junction of the thermocouple that will be measured.

When the option is External RJ, the compensation will be performed using the value defined by the user
in the range of -50.00ºC (-58.00ºF) to +50,00ºC (+122.00ºF) in the Config Menu, option Reference Junction,
Ext item, see chapter 8.5 Reference Junction for more information.

54 manual_c100_jvscf_v4 temperatura industrial


7.2 RT Input

To configure the instrument’s input for RT sensors, just tap on the RT icon and the following screen will
be displayed:

Pt 100 Pt 100 Pt 100


IEC JIS SAMA

Pt 1000 Cu 10 Ni 120
IEC

Sair
Exit

To choose the type of RT, just tap on one of the icons or navigate through the tactile keyboard’s cursor
keys and press ENTER to confirm. After doing so, a dialog box will give the option to choose the number of
wires of the connection. For more information on connecting RTs, see chapter 5.1 Connection Diagrams.

CAPPO 100

SELECT THE NUMBER OF


C ON N E C TI O N W I R E S ?
3 WIRES 4 WIRES

55 manual_c100_jvscf_v4 temperatura industrial


7.3 Current Input

To configure the instrument’s input for Current, just tap on the Current icon and the following screen
will be displayed:

CAPPO 100

S ELEC T THE TY PE O F
C U R R EN T LO O P ?
PASSIVE ACTIVE

Passive loop: considers that the Calibrator is being used in a non-energized circuit, that is, the Calibra-
tor will power the circuit, so the 24V source at the input channel’s terminals 1 and 2 will be turned on and the
maximum current in the measurement will be 51mA.

Active loop: considers that the Calibrator is being used in an energized circuit, that is, the circuit is
powered by a source external to the Calibrator, so the 24V source at the input channel’s terminals 1 and 2 will
be turned off and the maximum current in the measurement will be 51mA.

CAPPO 100 CAPPO 100

Input Entrada
mA Active Loop mA Passive Loop
or

24.86 14.84
Output Saída STORE RECALL

200mV 200mV

-1234.56 -1234.56
3.6V 3.6V

Input Output Config Entrada Saída Config

56 manual_c100_jvscf_v4 temperatura industrial


7.4 Voltage Input

To configure the instrument’s input for Voltage, just tap on the Voltage icon and the following screen
will be displayed:

200 mV 2000 mV 20 V

Sair
Exit

To choose the voltage range, just tap on one of the icons or navigate through the tactile keyboard’s
cursor keys and press ENTER to confirm.

57 manual_c100_jvscf_v4 temperatura industrial


7.5 Electrical Resistance Input

To configure the instrument’s input for Resistance, just tap on the Resistance icon and the following
screen will be displayed:

500 Ohms 4 KOhms Sair


Exit

To choose the resistance range, just tap on one of the icons or navigate through the tactile keyboard’s
cursor keys and press ENTER to confirm. After doing so, a dialog box will give the option to choose the number
of wires of the connection. For more information on connecting Resistance and RTs, see chapter 5.1 Connec-
tion Diagrams.

CAPPO 100

SELECT THE NUMBER OF


C ON N E C TI O N WI R E S ?
3 WIRES 4 WIRES

58 manual_c100_jvscf_v4 temperatura industrial


7.6 SCALE X Input

The SCALE X function allows you to display an input or simulate an output in an user-defined engi-
neering unit using correlation (linear interpolation).

Through this function, the user can display a linear input such as Voltage or Current, scaled as need-
ed, through the following equation:

Y = X% . (DispHi - DispLo) + DispLo


Start of Range (RangeLo) Initial value to display (DispLo)
Value displayed (X) Adjusted Output Value (Y)
End of Range (RangeHi) Final value to display (DispHi)

Output= RangeLo = 0% Output start of range


Output =X= X% Value between O to 100%
Output= RangeHi = 100% Output end of range

The Scale must be configured as instructed in Chapter 6.5 Scale X Configuration. To define Scale X as
instrument input just tap on the SCALE X icon in the Input Menu.

CAPPO 100

Input
X SCALE m/s

90.961
Output
200mV

0.00
3.6V

Input Output Config

59 manual_c100_jvscf_v4 temperatura industrial


7.7 Pressure Input

Configures the channel for reading a pressure module, which can be internal (CP1) or external (CP2),
note that the pressure measurement is an optional that must be defined at the time of purchase.

The pressure unit to be displayed can be configured in the Config Menu, with 14 different units.

On the display there is a button called ZERO that can be used to zero the pressure at a certain point,
thus generating an offset on the indicated pressure.

When the ZERO function is activated, the left side of the button is displayed (Z), by pressing the ZERO
button again, the offset is turned off and the indication (Z) is removed.

To configure the instrument’s pressure input, just tap on the Pressure icon and a dialog box will give the
option to choose the pressure module:

CAPPO 100

S ELEC T THE PRESSURE


M O D U LE ?
CP1 CP2

60 manual_c100_jvscf_v4 temperatura industrial


7.8 Leak Test

For an active pressure module (CP1) or (CP2) connected to the input channel, this function shows the
initial pressure, final pressure and the differential (initial – final), after the Leak Test time (configured in the
Config Menu) has elapsed.

The time starts to be counted by pressing the ENTER key on the tactile keyboard or tapping the START
button on the display.

During the time counting, pressing the ENTER key on the keyboard or tapping the RESET symbol on the
display resets the time.

If the chosen pressure module CP1 or CP2 is inactive, this function is not executed.

The Leak Test function is turned off when one of the Input, Output or Config menus is accessed.

Note: During the Leak Test, the Auto Power Off function will be disabled so that the instrument is not
automatically turned off before the end of the test.

To configure the instrument’s input for Leak Test, just tap on the LEAK TEST icon and if the Input is not
yet configured for any of the pressure modules, a dialog box will be displayed so that a module can be chosen:

CAPPO 100

F O R T H E L EA K TEST, CH1
M U S T B E CP1 O R CP2 ?
CP1 CP2

61 manual_c100_jvscf_v4 temperatura industrial


7.9 Probe Input
Through the coefficients A, B, C, and R0, a Standard Thermocouple (R, S) probe, or a Stantard PRT probe
will be linearized according to its own characteristic curve, eliminating calibration errors.

Ecil Metrology Laboratory can provide calibrated sensors, informing the coefficients A, B, C, and R0 for
automatic error correction through the Probe function.

Next, the user must select Probe 1, Probe 2, or Probe 3, which must be previously configured in the
Config Menu.

Note that to increase accuracy, the connection of the Pt100 type Probe is always made with 4 wires.

To configure the instrument’s input for Probes, just tap on the PROBE icon and the following screen will
be displayed:

PROBE 1 PROBE 2 PROBE 3

Sair
Exit

62 manual_c100_jvscf_v4 temperatura industrial


8. Output

On the Start Screen, at the bottom, tap on the Output Icon or press the Output button on the tactile
keyboard and the following screen will appear:

Output

THERMOCOUPLE
TERMOPAR RT
TERMO CURRENT
CORRENTE
^
RESISTENCIA

a--b
x--y
c--d
~ ^
TENSAO
VOLTAGE RESISTANCE
RESISTENCIA X SCALE

PRESSURE AUTORAMP SWITCH


TEST

M M
STORE RECALL Exit
Sair

Tap on the desired option or navigate through the cursor keys and press the ENTER KEY to choose be-
tween one of the 11 options for the output channel.

By pressing and holding any for approximately 3 seconds, The Built-In Help will be displayed with useful
information about the input signal selected.

63 manual_c100_jvscf_v4 temperatura industrial


There are two ways to enter the desired output value:

1 – Tapping the touchscreen in any space inside the output box, at the bottom section of the Start Scre-
en. When doing so, just use the numeric keypad that will be displayed, where E is to confirm, C to clear, the
arrow key is to navigate between digits, and ESC to cancel;

2 - Using the tactile keyboard, move the cursor to the desired digit and using the up and down arrows,
and enter the desired value for the output.

CAPPO 100

Input Limits
TC J Rj Internal
Max : 1372.00

234.56
M i n : - 270.00
Resol : 0.01

Output 0
TC K Rj External

7 8 9 E
-12.87
4 5 6 C
3.6V

1 2 3

Input Output Config


0 . +
-
ESC

64 manual_c100_jvscf_v4 temperatura industrial


8.1 Thermocouple Output

To configure the instrument’s output for Thermocouple sensors, just tap on the Thermocouple icon and
the following screen will be displayed:

TC J TC K TC N

TC E TC T TC R

TC S TC B Exit
Sair

To choose the thermocouple type, just tap on one of the icons or navigate through the tactile keybo-
ard’s cursor keys and press ENTER to confirm. After doing so, a dialog box will give the option to choose the
Reference Junction, internal or external.

CAPPO 100

SELECT THE
REFERENCE JUNCTION ?
INTERNAL EXTERNAL

65 manual_c100_jvscf_v4 temperatura industrial


When the Internal RJ option is selected, the ambient temperature will be measured through a Pt100
sensor installed on the input channel’s terminal 2 (see Chapter 4.1 Layout), and will be used to compensate for
reference junction of the thermocouple that will be measured.

When the option is External RJ, the compensation will be performed using the value defined by the user
in the range of -50.00ºC (-58.00ºF) to +50,00ºC (+122.00ºF) in the Config Menu, option Reference Junction,
Ext item, see chapter 8.5 Reference Junction for more information.

66 manual_c100_jvscf_v4 temperatura industrial


8.2 RT Output

To configure the instrument’s output for RT sensors, just tap on the RT icon and the following screen
will be displayed:

Pt 100 Pt 100 Pt 100


IEC JIS SAMA

Pt 1000 Cu 10 Ni 120
IEC

Sair
Exit

To choose the type of RT, just tap on one of the icons or navigate through the tactile keyboard’s cursor
keys and press ENTER to confirm.

67 manual_c100_jvscf_v4 temperatura industrial


8.3 Current Output

To configure the instrument’s output for Current, just tap on the Current icon and the following screen
will be displayed:

CAPPO 100

S ELEC T THE TY PE O F
C U R R EN T LO O P ?
PASSIVE ACTIVE

Passive loop: considers that the Calibrator is being used in a non-energized circuit, that is, the Calibra-
tor will power the circuit, so the 24V source at the input channel’s terminals 1 and 2 will be turned on and the
maximum current in the measurement will be 51mA.

Active loop: considers that the Calibrator is being used in an energized circuit, that is, the circuit is
powered by a source external to the Calibrator, so the 24V source at the input channel’s terminals 1 and 2 will
be turned off and the maximum current in the measurement will be 51mA.

CAPPO 100 CAPPO 100

Input Entrada
TC K Rj External TC K Rj Externa

234.56 234.56
Output Saída
mA Active Loop mA Passive Loop
or

26.416 26.416
3.6V
3.6V

Input Output Config Entrada Saída Config

68 manual_c100_jvscf_v4 temperatura industrial


8.4 Voltage Output

To configure the instrument’s output for Voltage, just tap on the Voltage icon and the following screen
will be displayed:

200 mV 2000 mV 20 V

Sair
Exit

To choose the voltage range, just tap on one of the icons or navigate through the tactile keyboard’s
cursor keys and press ENTER to confirm.

69 manual_c100_jvscf_v4 temperatura industrial


8.5 Electrical Resistance Output

To configure the instrument’s output for Resistance, just tap on the Resistance icon and the following
screen will be displayed:

500 Ohms 4 KOhms Sair


Exit

To choose the resistance range, just tap on one of the icons or navigate through the tactile keyboard’s
cursor keys and press ENTER to confirm.

70 manual_c100_jvscf_v4 temperatura industrial


8.6 Output Escala X

The SCALE X function allows you to display an input or simulate an output in an user-defined engi-
neering unit using correlation (linear interpolation).

Through this function, the user can simulate a linear output such as Voltage or Current, scaled as
needed, through the following equation:

X = Y% . (RangeHi - RangeLo) + RangeLo

Start of Range (RangeLo) Initial value to display (DispLo)


Value simulated (X) Adjusted Output Value (Y)
End of Range (RangeHi) Final value to display (DispHi)

Input= DisplayLo = 0% Input initial value to display


Input =Y= Y% Value between O to 100%
Input= DisplayHi = 100% Input final value to display

The Scale must be configured as instructed in Chapter 6.5 Scale X Configuration. To define Scale X as
instrument input just tap on the SCALE X icon in the Input Menu.

CAPPO 100

Input
TC K Rj External

234.56
Output
X SCALE m/s

90.961
3.6V

Input Output Config

71 manual_c100_jvscf_v4 temperatura industrial


8.7 Pressure Output

With the help of pressure generation accessories, configures an internal (CP1) or external (CP2) pres-
sure module’s output channel, note that the pressure measurement is an optional that must be defined at the
time of purchase.

The pressure unit to be displayed can be configured in the Config Menu, with 14 different units.

On the display there is a button called ZERO that can be used to zero the pressure at a certain point,
thus generating an offset on the indicated pressure.

When the ZERO function is activated, the left side of the button is displayed (Z), by pressing the ZERO
button again, the offset is turned off and the indication (Z) is removed.

To configure the instrument’s pressure input, just tap on the Pressure icon and a dialog box will give the
option to choose the pressure module:

CAPPO 100

S ELEC T THE PRESSURE


M O D U LE ?
CP1 CP2

72 manual_c100_jvscf_v4 temperatura industrial


8.8 Auto-Ramp Output

The Auto-Ramp function allows the generation of an output that will automatically follow a predefined
curve.

The user can define the type of output signal, the step-up and step-down ratios, the superior and infe-
rior levels, and the number of iterations for this cycle.

Such parameters can be configured in Chapter 6.4 Auto-Ramp Configuration.

The Auto-Ramp function will be automatically cancelled when either the Input, the Output, or the Con-
fig Menus’ is accessed.

Note: During the execution of the Auto-Ramp, the Auto Power Off function will be disabled so that the
instrument is not automatically turned off before the end of the ramp.

73 manual_c100_jvscf_v4 temperatura industrial


To trigger the Auto-Ramp, go to the Output Menu and tap on the AUTORAMP icon and you will be re-
directed to the Start Screen, where in the upper right-hand corner of the output part, an indication that the
Ramp is being executed will appear, as displayed below:

CAPPO 100

Input
TC K Rj External

134.05
UPWARD RAMP
Output
TC J Rj Internal

234.56
3.6V

Input Output Config

74 manual_c100_jvscf_v4 temperatura industrial


8.9 Switch Test

This function is used for thermostats and pressure switches’ calibrations, showing the status (Open or
Closed) of a dry contact or an energized one with maximum voltage of 24Vdc.

On the display it will be shown the contact’s status (Open or Closed) and the read value will be recorded
on the Input channel at the instant that the contact was opened or closed, displaying the difference between
the values (Hysteresis = Closed - Open).

CAPPO 100

Input
TC K Rj External

234.56
SWITCH TEST
CONTACT STATUS: OPEN

OPEN: ________
CLOSED: ________
HYSTERESIS: ________

3.6V

Input Output Config

The Switch test function will be automatically cancelled when either the Input, the Output or the Config
Menus’ is accessed.

75 manual_c100_jvscf_v4 temperatura industrial


9. MEMORY

This function is divided into two icons allocated in the Output Menu, Store and Recall, allowing up to 5
output configurations to be stored (Store) or recalled (Recall).

The output signal type and its respective value can be stored or recalled in 5 memory slots defined as
M1, M2, M3, M4, and M5.

These slots are useful to facilitate the adjustment in the instruments’ calibration, which have interac-
tive zero and span, in addition to providing other intermediate points for checking the calibration.

Stored values are retained in non-volatile flash memory.

To store the current output value in a memory space, just tap on the STORE icon, and on the following
screen tap on the memory slot icon to be used, and to retrieve and use a stored value as an output signal, just
tap on the RECALL icon and tap on the memory slot icon where the desired value is allocated:

M M M M M M
M1 M2 M3 M1 M2 M3

M M M M
M4 M5 Sair
Exit M4 M5 Sair
Exit

76 manual_c100_jvscf_v4 temperatura industrial


10. SERVICES
AND ASSISTANCE
10.1 Services
CALIBRATION LABORATORY GAS LABORATORY
Ecil Produtos’s Ecil Gas Analysis
Metrology Laboratory is Laboratory provides calibration
accredited by Cgcre (Brazilian services on portable flue gas
Accredited Calibration Labs analyzers. Since 2018, we’ve
System) for Temperature and been accredited by Cgcre for the
Electrical Signals calibration: calibration of gas analysers.

Temperature: RTs, thermocouples, analog/ In our structure, we have primary


digital thermometers with thermocouples standards, in some concentrations of: O2, CO, NO,
or resistance thermometers, extension/ NO2, SO2, HC (CxHy) e CO2.
compensation cables, liquid in glass
thermometers, infrared pyrometers, blackbody
sources, simulators, analog/digital indicators
or controllers for resistance thermometers and
thermocouples and temperature transmitters.

Electrical Signals: DC power source/


voltage meters, DC power source/Current meters,
resistance decade boxes and resistance meters.

10.2 Assistance
TECHNICAL ASSISTANCE CONTACT US
Willing to always offer Ecil Produtos e Sistemas de Medição e Controle
the most complete solutions to Ltda
our clients, Ecil has an in-house Rodovia Raimundo Antunes Soares, 1315
technical assistance structure that Paulas e Mendes - Piedade - SP
performs maintenance services in all the equipment CEP: 18170-000
and instruments we manufacture and represent, Tel. +55 15 3244-8003
within or already out of the warranty term. Fax. +55 15 3244-1672
E-mail: seast@ecil.com.br
http://www.ecil.com.br

77 manual_c100_jvscf_v4 temperatura industrial


Annex 1 - User Interface

User Interfac

JUNCT. INTRO
LEAK INFO
TYPES TEST
RESISTANCE
2

TYPES INPUT
8 MAIN

VOLTS
3
THERMOCOUPLE RT CURRENT VOLTAGE
SCALE
MODULE X
TYPES
6 ACTIVE/ PRESSURE
PASSIVE
PROBE

WIRES NBR. CONFIG


3

COEF. X
3 SAVE? UNITY
PROBE X RESISTANCE FUNCT.
3 TEST

TAG HEADER
TYPE SCALE X

EMF
TEST TYPE
Pt 100 TC R TC S
VOLTS
TEXT RANGE
3
INPUT
CURRENT HIGH/LOW
2
JUNCTION SAVE? ACTIVE/
PASSIVE PRESSU
UNIT
DISPLAY

DECIMAL UNITS
14
SCREEN
PROTECTION J

ENABLE CH X EXT
ICON 9 CHANGE 2
PASSWORD

DIALOG BOX

ENABLE/DISABLE

INDICATOR
78 manual_c100_jvscf_v4 temperatura industrial
ce
STORE
MODULE

SWITCH RECALL
TEST AUT O PRESSURE
RAMP
SLOTS
2

JUNCT.
OUTPUT
TYPES
8
VOLTS
3
INPUT EXIT CURRENT
VOLTAGE RT THERMOCOUPLE

TYPES
2 RESISTANCE
A CTIVE/
SCALE PASSIVE
TYPES
X
6
CYCLES

COLORS
LEAK TYPES
TEST AUT O
TIME 6 RT RAMP
HEADER

CH1 CH2
START/END TIMES
KEYBOARD TYPE STEP
BATTERY 2 4
SAVING ?

BRIGHTNESS THERMOCOUPLE RESISTANCE


SAVE? VOLTAGE
OFF POWER6 OFF CURRENT
OFF
TYPES
URE TYPES VOLTS 2
TEXT 8 3
INPUT ACTIVE/
PASSIVE
JUNCTS.

REFER.
JUNCTION

T CH X
INT
2

79 manual_c100_jvscf_v4 temperatura industrial


Ecil Produtos - ALL RIGHTS RESERVED -
This document’s total or partial reproduction, regardless of its nature (including photocopies e
digital storage) and transmission to third parties, no matter what medium, even electronically, is
strictly forbidden, unless explicitly authorized by Ecil Produtos.