CAPPO 100 User Manual: The Cutting Edge Portable Multicalibrator For Temperature, Pressure and Electrical Signals
CAPPO 100 User Manual: The Cutting Edge Portable Multicalibrator For Temperature, Pressure and Electrical Signals
USER MANUAL
The cutting edge Portable Multicalibrator for
Temperature, Pressure and Electrical Signals
temperatura industrial
With more than 91 years of history, Ecil is the Brazilian leader on industrial temperature
measurement and control. We are specialists in offering the most complete solutions in this area for
the most demanding industry segments such as: Oil & Gas, Steelmaking, Metalworking, Chemical,
Cement, Food, Pharmaceutical, Plastic, Glass, Pulp & Paper etc.
We offer to the market a wide range of products that consists of temperature sensors, such
as thermocouples, RTs (Resistance Thermometers), infrared thermometers and other instruments
for temperature measurement, transmission, insulation, control, recording and calibration. We also
provide solutions in pressure calibration and flue gas analysis.
Our Metrology Laboratory was the first industrial lab to be accredited by Cgcre, the Brazilian
General Accreditation Coordination of INMETRO (Brazilian National Metrology Laboratories). It offers
to our clients services such as Temperature Calibration for all kinds of temperature sensors, Electrical
Signals Calibration and Flue Gas Analyzers Calibration. The laboratory is also used in our in-house
made temperature sensor’s raw material testing and instrument’s tests and calibration.
Our team of specialized and proven in many years of market service engineers is ready to assist
you with the best value for the money choice for the needs of your company.
temperatura industrial
Ecil Produtos - ALL RIGHTS RESERVED -
This document’s total or partial reproduction, regardless of its nature (including photocopies e
digital storage) and transmission to third parties, no matter what medium, even electronically, is
strictly forbidden, unless explicitly authorized by Ecil Produtos.
TABLE OF
CONTENTS
1. IMPORTANT INFORMATION 6
1.1 Information about this manual 6
2. OPERATING PRINCIPLES 7
2.1 General Description 7
2.2 Component’s Description 7
2.2.1 Component’s Description> Screen 7
2.2.2 Component’s Description> Connections 8
3. TECHNICAL SPECIFICATIONS 9
4. PRESENTATION 12
4.1 Layout 12
4.2 Display 13
4.3 Keyboard 15
5. START 16
5.1 Connection Schemes - Input 16
5.2 Connection Schemes - Output 17
5.3 On / Off 18
5.4 Built-In Help 20
6. CONFIGURATION 22
6.1 Colors 23
6.2 Header 25
6.3 Probe 26
6.4 Auto-Ramp 32
6.5 SCALE X 36
6.6 Pressure Units 42
6.7 Reference Junction 43
6.8 Leak Test Time 45
6.9 Battery Saving 46
6.10 Protection 48
Read this operation manual before utilizing it. The user must be familiarized with the instrument and
follow the instructions carefully.
This user manual is subject to changes due to technical improvements; The manufacturer does not
assume any liability for any printing error or flaw.
The calibrator is light and compact, powered by 3.7Vdc @ 5,200mA Li-Ion batteries, the plastic outer
casing has a modern design, with a touchscreen, 65,536 colors, 272x480 pixels graphic display. It also has a 5
buttons tactile keyboard, giving the user an alternative way to navigate the menus.
In order to better utilize the resistive touchscreen display, we recommend the use of an object
with a non-piercing soft and thin tip.
The instrument has two simultaneous and independent universal channels that are isolated from each
other. The Input channel is used for measurement, while the Output channel is used for simulation. Both
channels accept the following types of signals; Thermocouples, RTs, mV, V, mA and Ohmic Resistance. Each
channel has its own 24V @ 30mA power source, also isolated from each other.
As an optional feature, the instrument can also have two more configurable channels for pressure
measurement, one internal (CP1) and the other external (CP2).
All measurements and simulations are made using high-resolution integrated circuits (24bits), with
sophisticated conditioning, filtering and amplification systems, thereby giving the product excellent precision
and thermal stability over time.
There are two options for pressure modules, one internal and one input for external modules, CP1 and
CP2 respectively, located on the bottom part of the instrument.
On the right-hand side there is an USB type B input for the battery recharging.
Input/Output – Current mA
Range Resolution Accuracy Observation
0 to 51mA 0.1µA ± 0.02% of reading + 0.4µA Active or Passive loop
Passive loop: Considers that the Calibrator is being used in a non-energized loop, meaning that it must
supply power to the circuit itself, therefore the 24V source of the input channel will be turned on and the max
current of the measurement will be 51mA.
Active loop: Considers the Calibrator is being used in an energized loop, meaning that the circuit is
powered by an external power supply, therefore the 24V source of the input channel will be turned off and the
max current of the measurement will be 51mA.
Input – Resistance Ω
Range Resolution Accuracy Observation
0 to 500 Ω 1mΩ ± 0.02% of reading + 20mΩ 2 Wires Output
0 to 4,000 Ω 10mΩ ± 0.02% of reading + 200mΩ 2 Wires Output
Input/Output – Thermocouples
TC Type Range Resolution Accuracy
J -210ºC (-346ºF) to 1,200ºC (2,192ºF) 0.01 ºC/F ± 0.02% of reading + 0.1ºC (0.18ºF)
K -270ºC (-454ºF) to 1,372ºC (2,501.60ºF) 0.01 ºC/F ± 0.02% of reading + 0.1ºC (0.18ºF)
N -270ºC (-454ºF) to 1,300ºC (2,372ºF) 0.01 ºC/F ± 0.02% of reading + 0.1ºC (0.18ºF)
E -270ºC (-454ºF) to 1,000ºC (1,832ºF) 0.01 ºC/F ± 0.02% of reading + 0.1ºC (0.18ºF)
T -210ºC (-346ºF) to 400ºC (752ºF) 0.01 ºC/F ± 0.02% of reading + 0.1ºC (0.18ºF)
R -50ºC (-58ºF) to 1,768ºC (3,214.40ºF) 0.10 ºC/F ± 0.02% of reading + 0.2ºC (0.36ºF)
S -50ºC (-58ºF) to 1,768ºC (3,214.40ºF) 0.10 ºC/F ± 0.02% of reading + 0.2ºC (0.36ºF)
B 0ºC (32ºF) to 1,820ºC (3,308ºF) 0.10 ºC/F ± 0.02% of reading + 0.3ºC (0.54ºF)
Obs: types J, K, N, E and T thermocouples accuracy, in temperatures below -200°C (-328°F), will be ±
0,02% of the reading +1.0°C (+ 1.8°F). The error limits stand for 360 days for operation between 18°C (64.4°F)
to 28°C (82.4°F), out of this range the thermal stability is of 0.001 % FS / °C/F.
4.1 Layout
Pt100 - Input Channel’s
Pt100 - Output Channel’s Reference Junction
Reference Junction
Input Channel’s Golden
Plated Terminals
Output Channel’s Golden
Plated Terminals
480x272 pixels
Touchscreen Color Display
Battery Charging
Input
ON/OFF
Button
Tactile Keyboard
Pressure Module
(optional)
CAPPO 100
Input
TC K Rj External
Upper Section
234.56
Output
200mV Lower Section
-1234.56
3.6V
On the lower right-hand corner, right below the Lower Section and over the Config icon, an icon in-
dicating the battery level is dynamically displayed. The letter P in the icon indicates that the instrument is
connected to the Power Supply.
Whenever the battery reaches a critical level (red color), a message ‘BATTERY LOW’ is displayed and
the instrument is automatically shut down.
A battery recharger (110/220Vac input and 5V@1A output) is supplied with the calibrator, the usual re-
charging time with the instrument turned off is 6 hours. Other types of recharging methods not supplied with
the instrument are not recommended.
The battery life is approximately 6 hours for Thermocouples, RTs, mV and V Input/Output and 3 hours
for 20mA current simulation in passive mode.
CAPPO 100
Input Input
TC K Rj External TC K Rj Externa
234.56 234.56
Output Output
200mV 200mV
-1234.56 -1234.56
3.6V 3.6V
The configuration and operation of the Cappo 100 is very simple and all parameters can be accessed via
the 8-key tactile keyboard, with the following functions:
Button Function
ON/OFF Turning the Instrument On and Off
Arrow Keys Navigate through Instrument Icons and Parameters
ENTER Check to choice or the parameter value
INPUT Go to Input Menu
OUTPUT Go to Output Menu
CONFIG Go to Config Menu
And the boot screen will be displayed Just tap anywhere and the
Home Screen will be displayed:
CAPPO 100
Input
TC K Rj External
234.56
Output
200mV
-1234.56
3.6V
Curiosity: The aerial photo on the Home Page is of our new headquarters, made up of three buildings
inaugurated in early 2017. The facilities are very modern, which allows us to produce cleaner and more ef-
ficiently, with greater safety for our employees and guaranteeing the satisfaction of our customers, our top
priority. Come visit us!
After pressing the Info button the New Cappo 100 Introduction screen will appear:
Many of the screens, as shown above, have more than one page, to scroll between them just use your
fingers and swipe up and down, making a screen scroll. To exit any help screen, simply tap on the Home icon
in the right-hand corner of the header at the top of the screen.
On the Home Screen, at the bottom right-hand corner, tap on the Config Icon or press the CONFIG but-
ton on the physical keyboard and the following screen will be displayed:
Config
Ch1
Ch2
a--b
x--y
c--d
AUTORAMPA
AUTORAMP ESCALA
X SCALEX PRESSURE
UNIT
mbar
15s
PROTECTION Sair
Exit
Sair
Exit
This screen allows you to select the colors of the header, input channel and output channel.
CAPPO 100
Input
200mV
1234.56
EXIT DEFAULT EXIT DEFAULT
FRAME NUMBER
10
15
63
56
0
21
Through 3 colored bars, red, green and blue, up to 65,000 color combinations can be adjusted, note
that each bar can be adjusted independently of one another.
Just tap on the bar of one of the colors and drag the white ball to the desired value, checking the result
in the model shown above the bars.
As for the header, the text and background colors can be adjusted.
For the input and output channels, the text colors, background, frame and number of each channel can
all be adjusted independently.
The DEFAULT option recovers the original colors defined at the factory and the EXIT option finishes the
configuration, then you can choose to save or not the modifications made.
CAPPO 100
When tapping on the YES option, a horizontal keyboard will be displayed on the screen so the header
text can be edited with the following settings:
An alphanumeric keyboard drawn horizontally on the display must be used to input the desired text.
As the keys are of reduced dimensions, in order to type the text it is recommended to use an object
with a non-piercing fine and soft tip.
The EXIT key finishes the configuration and allows the user to whether or not save the changes made.
X1
X3
X2
X4
CAPPO 100 _ EXIT
1 2 3 4 5 6 7 8 9 0 @
q w e r t y u i o p
a s d f g h j k l ç ^
z x c v b n m \ / % ´
o
_ ( ) \ . - +
On the Config screen, tap on the PROBE icon and the following screen will be displayed:
Sair
Exit
mV mA TIC mV mA
A B A B TIC A B A B
V Tc FIC V Tc FIC
Pt100 NOME? C RO C RO Pt100 NOME? C RO C RO
A B A B 2+2 A B 2+2
2x2 2x2
C RO C RO =4 C RO =4
Up to 3 (three) Probes can be configured, and for each one parameters bellow it must be defined:
Tag: Through the horizontal alphanumeric keyboard, you can define a tag (name) to identify the Probe with up
to 8 characters.
Coefficient A: As stated in the thermocouple or Pt100 sensor calibration certificate, being a number in scien-
tific notation format (mantissa + exponent) within the range from +9.99999E+33 to -9.99999E-33.
Coefficient RO (Resistance at 0ºC): Same as the coefficient A, but it is only shown if the type is Pt100.
Pt 100 TC R TC S
IEC
Exit
Just tap on the desired sensor type; If you choose a thermocouple, you will have to choose the type of
reference junction, internal or external, through a dialog box like the one below:
CAPPO 100
SELECT T H E
REFEREN C E J U N C TION ?
INTERNAL EXTERNAL
_
probe:
AXXXXXXX EXIT
1 2 3 4 5 6 7 8 9 0 @
q w e r t y u i o p
a s d f g h j k l ç ^
z x c v b n m \ / % ´
o
_ ( ) \ . - +
After finishing, just tap on EXIT and a dialog box will appear so you can choose to save or not:
CAPPO 100
S AV E THE
A LTERA TI O NS? ?
YES NO
Limits
Max : +9.99999E+33
M i n : -9.99999E-33
R e s o l : 0.00001
-6.62280E-02
7 8 9 E
4 5 6 C
1 2 3
0 . +
-
ESC
Use the keyboard to set the value, E to confirm, C to clear, the arrow key to navigate between digits and
ESC to cancel.
0 0
7 8 9 E 7 8 9 E
4 5 6 C 4 5 6 C
1 2 3 1 2 3
0 . +
-
ESC
0 . +
-
ESC
The Auto-Ramp function allows the generation of an output that will automatically follow a curve pre-
defined by the user.
On the Settings screen, tap on the AUTORAMP icon and the following screen will be displayed, with
icons that give access to the parameters’ configuration at the top part of the screen and a table containing the
parameters set at the bottom:
mV mA
V Tc
Pt100
CYCLES Exit
Sair
TYPE: TC K Rj External
R A´ M P S T A R T : -270.00
STEP: 1.00
STEP UP TIME: 1
STEP DOWN TIME: 1
SOAK UP TIME: 10
SOAK DOWN TIME: 20
N U´ M B E R O F C Y C L E S : 1
THERMOCOUPLE RT CURRENT
Just browse the icons and choose the output signal; However, if you choose Thermocouple or Current,
you will need to choose the type of reference junction for the type of thermocouple chosen and between pas-
sive or active loop for current, all should be set through the dialog boxes below:
SE L E CT T H E SELECT THE TY PE O F
RE F E RE NCE J U N C T I ON ? CURRENT LO O P ?
INTERNAL EXTERNAL PASSIVE ACTIVE
Limits Limits
Max : 1372.00 Max : 51.0000
M i n : - 270.00 Min : 0.0000
Resol : 0.01 Resol : 0.0001
0 0
7 8 9 E 7 8 9 E
4 5 6 C 4 5 6 C
1 2 3 1 2 3
0 . +
-
ESC
0 . +
-
ESC
Use the numeric keypad to set the value, E to confirm, C to clear, the arrow key to navigate between
digits and ESC to cancel. Below is the description of the selectable parameters:
Parameter Function
START Ramp’s starting point value
END Ramp’s ending point value
STEP Value for each step of the upward or downward ramp
TSTPUP Time for each step of the upward Ramp
TSTPDWN Time for each step of the downward Ramp
TSOAKUP Time for the upper level
TSOAKDWN Time for the lower level
CYCLES Numbers of iterations of the curve
CAPPO 100
Input
TC K Rj External
134.05
UPWARD RAMP
Output
TC J Rj Internal
234.56
3.6V
Through this function, the user can display a linear input or simulate a linear output such as Voltage
or Current, scaled as needed, through the following equation:
The decimal point position of the value to be displayed is freely configurable, also, an 8-character
unit can be assigned to it.
On the Config screen, tap on the SCALE X icon and the following screen will be displayed, with icons
that give access to the parameters’ configuration at the top part of the screen and a table containing the
parameters set at the bottom:
mV mA HI HI
V Tc
Pt100 LO LO
T IPO
TYPE RANGE DISPLAY
m 3 /s ax+b
Kg/h ax +b
2
Type: 200mV
200 mV 2000 mV 20 V
CURRENT
CORRENTE Sair
Exit
Just browse the icons and choose the output signal; However, if you choose Current, you will need to
choose between passive or active loop via the dialog box below:
CAPPO 100
S ELEC T THE TY PE O F
C U R R EN T LO O P ?
PASSIVE ACTIVE
CAPPO 100
AD J U S T T H E R A N G E ?
LOW HIGH
Having chosen the start (LOW button) or the end (HIGH button) of the input range, you must use the
numeric keypad as displayed below:
Limits
Max : 2100.00
Min : -210.00
Resol : 0.01
7 8 9 E
4 5 6 C
1 2 3
0 . +
-
ESC
Use the numeric keypad to set the value, E to confirm, C to clear, the arrow key to navigate between
digits and ESC to cancel.
CAPPO 100
Having chosen the initial (LOW) and final (HIGH) values for the Display, you must use the numeric
keypad as displayed below:
Limits
Max : 999.999
Min : -999.999
Resol : 0.001
7 8 9 E
4 5 6 C
0 . +
-
ESC
digits and ESC to cancel.
m/s _ EXIT
1 2 3 4 5 6 7 8 9 0 @
q w e r t y u i o p
a s d f g h j k l ç ^
z x c v b n m \ / % ´
o
_ ( ) \ . - +
After finishing, just tap on EXIT and a dialog box will appear so you can choose to save or not:
CAPPO 100
S A V E T HE
A L T E R ATIO NS? ?
YES NO
By tapping on the FUNCTION icon, a dialog box will be displayed to choose the type of scale function:
CAPPO 100
USER SCALE
C H OOS E T H E F U N C T I O N ?
LINEAR QUADR.
CAPPO 100
Input
X SCALE m/s
90.961
Output
200mV
0.00
3.6V
To choose the pressure unit, just tap on the PRESSURE UNIT icon and the following screen will be dis-
played:
The Reference Junction function allows the temperatures measured on the Pt100 sensors installed in
the terminal 2 of the Input and Output channels to be viewed.
The temperature measured by these sensors will be used as compensation of the reference junction in
case the channel (Input and/or Output) is configured as thermocouple and Rji (Internal Reference Junction).
When the Rje (External Reference Junction) option is used, the temperature used as compensation is
set in this menu.
See chapters 7.1. Thermocouple Input and 8.1. Thermocouple Output for more information on selecting
the Internal or External Reference Junctions.
To adjust the temperatures to be used as compensation for the Input and/or Output channel’s ther-
mocouple reference junction when the Rje option is selected and to view the Input and/or Output channel’s
temperature readings at terminal 2 when the option Rji is selected, just tap on the REFER JUNCRT icon, and
the following screen will be displayed:
CH1 CH2
EXT EXT
CH1 CH2
INT INT
27.40 27.52
Limits
Max : 50.00
Min : - 50.00
Resol : 0.01
7 8 9 E
4 5 6 C
1 2 3
0 . +
-
ESC
Use the numeric keypad to set the value, E to confirm, C to clear, the arrow key to navigate between
digits and ESC to cancel.
This function allows you to enter a number from 1 to 1800 seconds that will be the duration of the Leak
Test.
To set this time, just tap on the LEAK TEST TIME icon and a screen with a numeric keypad will be dis-
played like the one below:
Limits
Max : 1800
Min : 1
Resol : 1
7 8 9 E
4 5 6 C
1 2 3
0 . +
-
ESC
To start the test, just go to the Input Menu and tap on the LEAK TEST icon and you will be be redirected
to the Start Screen, where the test will be performed in the Input section with the time that was set here.
You will find more information about the test in chapter 7.8 Leak Test, in the Input section.
To maximize battery usage, there are two icons, BRIGHTNESS and POWER OFF, which allow a time from
0 to 20 minutes to be entered for each of them. To set this time, just tap on the BATTERY SAVING and you will
be directed to the screen below:
BRIGHTNESS: After the set time elapses and if no keyboard key or display function is activated, the
brightness will be reduced to 10% of the original value.
By typing on the tactile keyboard or touching the display, the original brightness will be restored.
POWER OFF: After the set time elapses and if no keyboard key or display function is activated, the in-
strument will be turned off.
The charger and USB cable are provided with the instrument and due to the high charging current, it
is not recommended to charge the instrument through a computer, i.e., for best performance, always use the
charger and cable provided for this function.
The battery icon shown on the main screen indicates the charge level and the voltage value, when the
level reaches 3.2V the on/off button led starts to flash and when the voltage reaches 3.1V, the instrument will
be automatically turned off. To turn it on again, plug in the battery charger.
Limits
Max : 20
Min : 0
Resol : 1
7 8 9 E
4 5 6 C
1 2 3
0 . +
-
ESC
Use the numeric keypad to set the value, E to confirm, C to clear, the arrow key to navigate between
digits and ESC to cancel.
Through this function, it is possible to define which icons of the Config Menu will have its access locked.
To use this functionality it is necessary to know the input password (Default Password: 2112); Just tap on the
PROTECTION icon and you will be directed to the screen below:
Limits
Max : 999999
Min : 0
Resol : 1
7 8 9 E
4 5 6 C
1 2 3
0 . +
-
ESC
Use the numeric keypad to set the password, E to confirm, C to clear, the arrow key to navigate be-
tween digits and ESC to cancel.
Colors, Header, Probe, Auto-Ramp, SCALE X, Pressure Unit, Reference Junction, Leak Test Time, and
Battery Saving.
Locked: Red color and closed padlock; It will not allow parameters changes in the Config Menu.
Unlocked: Green color and open padlock; It will allow parameters changes in the Config Menu.
By touching the icon or using the keyboard’s ENTER key, you can change the selected icon’s status from
Locked to Unlocked and vice versa.
Config Config
CHANGE Exit
Sair CHANGE Exit
Sair
PASSWORD PASSWORD
Limits
Max : 999999
Min : 0
Resol : 1
7 8 9 E
4 5 6 C
1 2 3
0 . +
-
ESC
Use the numeric keypad to set the password, E to confirm, C to clear, the arrow key to navigate between
digits and ESC to cancel.
After two valid sequences, a confirmation will be asked to save the new password, if the option YES is
selected, the new password will be saved.
Write down the new password, because if you forget it, access to the Config Menu may be limited.
On the Start Screen, at the bottom left corner, tap on the Input Icon or press the Input button on the
tactile keyboard and the following screen will appear:
Input
THERMOCOUPLE
TERMOPAR RT
TERMO CURRENT
CORRENTE
^
RESISTENCIA
a--b
x--y
c--d
~ ^
TENSAO
VOLTAGE RESISTANCE
RESISTENCIA X SCALE
Exit
Sair
By pressing and holding any for approximately 3 seconds, The Built-In Help will be displayed with useful
information about the input signal selected.
If there is no instrument connected generating any input signal, up or down arrows will be displayed
indicating that the input is open.
CAPPO 100
Input
TC K Rj External
Output
200mV
0.00
3.6V
To configure the instrument’s input for Thermocouple sensors, just tap on the Thermocouple icon and
the following screen will be displayed:
TC J TC K TC N
TC E TC T TC R
TC S TC B Exit
Sair
To choose the thermocouple type, just tap on one of the icons or navigate through the tactile key-
board’s cursor keys and press ENTER to confirm. After doing so, a dialog box will give the option to choose the
Reference Junction, internal or external.
CAPPO 100
SELECT THE
REFERENCE JUNCTION ?
INTERNAL EXTERNAL
When the option is External RJ, the compensation will be performed using the value defined by the user
in the range of -50.00ºC (-58.00ºF) to +50,00ºC (+122.00ºF) in the Config Menu, option Reference Junction,
Ext item, see chapter 8.5 Reference Junction for more information.
To configure the instrument’s input for RT sensors, just tap on the RT icon and the following screen will
be displayed:
Pt 1000 Cu 10 Ni 120
IEC
Sair
Exit
To choose the type of RT, just tap on one of the icons or navigate through the tactile keyboard’s cursor
keys and press ENTER to confirm. After doing so, a dialog box will give the option to choose the number of
wires of the connection. For more information on connecting RTs, see chapter 5.1 Connection Diagrams.
CAPPO 100
To configure the instrument’s input for Current, just tap on the Current icon and the following screen
will be displayed:
CAPPO 100
S ELEC T THE TY PE O F
C U R R EN T LO O P ?
PASSIVE ACTIVE
Passive loop: considers that the Calibrator is being used in a non-energized circuit, that is, the Calibra-
tor will power the circuit, so the 24V source at the input channel’s terminals 1 and 2 will be turned on and the
maximum current in the measurement will be 51mA.
Active loop: considers that the Calibrator is being used in an energized circuit, that is, the circuit is
powered by a source external to the Calibrator, so the 24V source at the input channel’s terminals 1 and 2 will
be turned off and the maximum current in the measurement will be 51mA.
Input Entrada
mA Active Loop mA Passive Loop
or
24.86 14.84
Output Saída STORE RECALL
200mV 200mV
-1234.56 -1234.56
3.6V 3.6V
To configure the instrument’s input for Voltage, just tap on the Voltage icon and the following screen
will be displayed:
200 mV 2000 mV 20 V
Sair
Exit
To choose the voltage range, just tap on one of the icons or navigate through the tactile keyboard’s
cursor keys and press ENTER to confirm.
To configure the instrument’s input for Resistance, just tap on the Resistance icon and the following
screen will be displayed:
To choose the resistance range, just tap on one of the icons or navigate through the tactile keyboard’s
cursor keys and press ENTER to confirm. After doing so, a dialog box will give the option to choose the number
of wires of the connection. For more information on connecting Resistance and RTs, see chapter 5.1 Connec-
tion Diagrams.
CAPPO 100
The SCALE X function allows you to display an input or simulate an output in an user-defined engi-
neering unit using correlation (linear interpolation).
Through this function, the user can display a linear input such as Voltage or Current, scaled as need-
ed, through the following equation:
The Scale must be configured as instructed in Chapter 6.5 Scale X Configuration. To define Scale X as
instrument input just tap on the SCALE X icon in the Input Menu.
CAPPO 100
Input
X SCALE m/s
90.961
Output
200mV
0.00
3.6V
Configures the channel for reading a pressure module, which can be internal (CP1) or external (CP2),
note that the pressure measurement is an optional that must be defined at the time of purchase.
The pressure unit to be displayed can be configured in the Config Menu, with 14 different units.
On the display there is a button called ZERO that can be used to zero the pressure at a certain point,
thus generating an offset on the indicated pressure.
When the ZERO function is activated, the left side of the button is displayed (Z), by pressing the ZERO
button again, the offset is turned off and the indication (Z) is removed.
To configure the instrument’s pressure input, just tap on the Pressure icon and a dialog box will give the
option to choose the pressure module:
CAPPO 100
For an active pressure module (CP1) or (CP2) connected to the input channel, this function shows the
initial pressure, final pressure and the differential (initial – final), after the Leak Test time (configured in the
Config Menu) has elapsed.
The time starts to be counted by pressing the ENTER key on the tactile keyboard or tapping the START
button on the display.
During the time counting, pressing the ENTER key on the keyboard or tapping the RESET symbol on the
display resets the time.
If the chosen pressure module CP1 or CP2 is inactive, this function is not executed.
The Leak Test function is turned off when one of the Input, Output or Config menus is accessed.
Note: During the Leak Test, the Auto Power Off function will be disabled so that the instrument is not
automatically turned off before the end of the test.
To configure the instrument’s input for Leak Test, just tap on the LEAK TEST icon and if the Input is not
yet configured for any of the pressure modules, a dialog box will be displayed so that a module can be chosen:
CAPPO 100
F O R T H E L EA K TEST, CH1
M U S T B E CP1 O R CP2 ?
CP1 CP2
Ecil Metrology Laboratory can provide calibrated sensors, informing the coefficients A, B, C, and R0 for
automatic error correction through the Probe function.
Next, the user must select Probe 1, Probe 2, or Probe 3, which must be previously configured in the
Config Menu.
Note that to increase accuracy, the connection of the Pt100 type Probe is always made with 4 wires.
To configure the instrument’s input for Probes, just tap on the PROBE icon and the following screen will
be displayed:
Sair
Exit
On the Start Screen, at the bottom, tap on the Output Icon or press the Output button on the tactile
keyboard and the following screen will appear:
Output
THERMOCOUPLE
TERMOPAR RT
TERMO CURRENT
CORRENTE
^
RESISTENCIA
a--b
x--y
c--d
~ ^
TENSAO
VOLTAGE RESISTANCE
RESISTENCIA X SCALE
M M
STORE RECALL Exit
Sair
Tap on the desired option or navigate through the cursor keys and press the ENTER KEY to choose be-
tween one of the 11 options for the output channel.
By pressing and holding any for approximately 3 seconds, The Built-In Help will be displayed with useful
information about the input signal selected.
1 – Tapping the touchscreen in any space inside the output box, at the bottom section of the Start Scre-
en. When doing so, just use the numeric keypad that will be displayed, where E is to confirm, C to clear, the
arrow key is to navigate between digits, and ESC to cancel;
2 - Using the tactile keyboard, move the cursor to the desired digit and using the up and down arrows,
and enter the desired value for the output.
CAPPO 100
Input Limits
TC J Rj Internal
Max : 1372.00
234.56
M i n : - 270.00
Resol : 0.01
Output 0
TC K Rj External
7 8 9 E
-12.87
4 5 6 C
3.6V
1 2 3
To configure the instrument’s output for Thermocouple sensors, just tap on the Thermocouple icon and
the following screen will be displayed:
TC J TC K TC N
TC E TC T TC R
TC S TC B Exit
Sair
To choose the thermocouple type, just tap on one of the icons or navigate through the tactile keybo-
ard’s cursor keys and press ENTER to confirm. After doing so, a dialog box will give the option to choose the
Reference Junction, internal or external.
CAPPO 100
SELECT THE
REFERENCE JUNCTION ?
INTERNAL EXTERNAL
When the option is External RJ, the compensation will be performed using the value defined by the user
in the range of -50.00ºC (-58.00ºF) to +50,00ºC (+122.00ºF) in the Config Menu, option Reference Junction,
Ext item, see chapter 8.5 Reference Junction for more information.
To configure the instrument’s output for RT sensors, just tap on the RT icon and the following screen
will be displayed:
Pt 1000 Cu 10 Ni 120
IEC
Sair
Exit
To choose the type of RT, just tap on one of the icons or navigate through the tactile keyboard’s cursor
keys and press ENTER to confirm.
To configure the instrument’s output for Current, just tap on the Current icon and the following screen
will be displayed:
CAPPO 100
S ELEC T THE TY PE O F
C U R R EN T LO O P ?
PASSIVE ACTIVE
Passive loop: considers that the Calibrator is being used in a non-energized circuit, that is, the Calibra-
tor will power the circuit, so the 24V source at the input channel’s terminals 1 and 2 will be turned on and the
maximum current in the measurement will be 51mA.
Active loop: considers that the Calibrator is being used in an energized circuit, that is, the circuit is
powered by a source external to the Calibrator, so the 24V source at the input channel’s terminals 1 and 2 will
be turned off and the maximum current in the measurement will be 51mA.
Input Entrada
TC K Rj External TC K Rj Externa
234.56 234.56
Output Saída
mA Active Loop mA Passive Loop
or
26.416 26.416
3.6V
3.6V
To configure the instrument’s output for Voltage, just tap on the Voltage icon and the following screen
will be displayed:
200 mV 2000 mV 20 V
Sair
Exit
To choose the voltage range, just tap on one of the icons or navigate through the tactile keyboard’s
cursor keys and press ENTER to confirm.
To configure the instrument’s output for Resistance, just tap on the Resistance icon and the following
screen will be displayed:
To choose the resistance range, just tap on one of the icons or navigate through the tactile keyboard’s
cursor keys and press ENTER to confirm.
The SCALE X function allows you to display an input or simulate an output in an user-defined engi-
neering unit using correlation (linear interpolation).
Through this function, the user can simulate a linear output such as Voltage or Current, scaled as
needed, through the following equation:
The Scale must be configured as instructed in Chapter 6.5 Scale X Configuration. To define Scale X as
instrument input just tap on the SCALE X icon in the Input Menu.
CAPPO 100
Input
TC K Rj External
234.56
Output
X SCALE m/s
90.961
3.6V
With the help of pressure generation accessories, configures an internal (CP1) or external (CP2) pres-
sure module’s output channel, note that the pressure measurement is an optional that must be defined at the
time of purchase.
The pressure unit to be displayed can be configured in the Config Menu, with 14 different units.
On the display there is a button called ZERO that can be used to zero the pressure at a certain point,
thus generating an offset on the indicated pressure.
When the ZERO function is activated, the left side of the button is displayed (Z), by pressing the ZERO
button again, the offset is turned off and the indication (Z) is removed.
To configure the instrument’s pressure input, just tap on the Pressure icon and a dialog box will give the
option to choose the pressure module:
CAPPO 100
The Auto-Ramp function allows the generation of an output that will automatically follow a predefined
curve.
The user can define the type of output signal, the step-up and step-down ratios, the superior and infe-
rior levels, and the number of iterations for this cycle.
The Auto-Ramp function will be automatically cancelled when either the Input, the Output, or the Con-
fig Menus’ is accessed.
Note: During the execution of the Auto-Ramp, the Auto Power Off function will be disabled so that the
instrument is not automatically turned off before the end of the ramp.
CAPPO 100
Input
TC K Rj External
134.05
UPWARD RAMP
Output
TC J Rj Internal
234.56
3.6V
This function is used for thermostats and pressure switches’ calibrations, showing the status (Open or
Closed) of a dry contact or an energized one with maximum voltage of 24Vdc.
On the display it will be shown the contact’s status (Open or Closed) and the read value will be recorded
on the Input channel at the instant that the contact was opened or closed, displaying the difference between
the values (Hysteresis = Closed - Open).
CAPPO 100
Input
TC K Rj External
234.56
SWITCH TEST
CONTACT STATUS: OPEN
OPEN: ________
CLOSED: ________
HYSTERESIS: ________
3.6V
The Switch test function will be automatically cancelled when either the Input, the Output or the Config
Menus’ is accessed.
This function is divided into two icons allocated in the Output Menu, Store and Recall, allowing up to 5
output configurations to be stored (Store) or recalled (Recall).
The output signal type and its respective value can be stored or recalled in 5 memory slots defined as
M1, M2, M3, M4, and M5.
These slots are useful to facilitate the adjustment in the instruments’ calibration, which have interac-
tive zero and span, in addition to providing other intermediate points for checking the calibration.
To store the current output value in a memory space, just tap on the STORE icon, and on the following
screen tap on the memory slot icon to be used, and to retrieve and use a stored value as an output signal, just
tap on the RECALL icon and tap on the memory slot icon where the desired value is allocated:
M M M M M M
M1 M2 M3 M1 M2 M3
M M M M
M4 M5 Sair
Exit M4 M5 Sair
Exit
10.2 Assistance
TECHNICAL ASSISTANCE CONTACT US
Willing to always offer Ecil Produtos e Sistemas de Medição e Controle
the most complete solutions to Ltda
our clients, Ecil has an in-house Rodovia Raimundo Antunes Soares, 1315
technical assistance structure that Paulas e Mendes - Piedade - SP
performs maintenance services in all the equipment CEP: 18170-000
and instruments we manufacture and represent, Tel. +55 15 3244-8003
within or already out of the warranty term. Fax. +55 15 3244-1672
E-mail: seast@ecil.com.br
http://www.ecil.com.br
User Interfac
JUNCT. INTRO
LEAK INFO
TYPES TEST
RESISTANCE
2
TYPES INPUT
8 MAIN
VOLTS
3
THERMOCOUPLE RT CURRENT VOLTAGE
SCALE
MODULE X
TYPES
6 ACTIVE/ PRESSURE
PASSIVE
PROBE
COEF. X
3 SAVE? UNITY
PROBE X RESISTANCE FUNCT.
3 TEST
TAG HEADER
TYPE SCALE X
EMF
TEST TYPE
Pt 100 TC R TC S
VOLTS
TEXT RANGE
3
INPUT
CURRENT HIGH/LOW
2
JUNCTION SAVE? ACTIVE/
PASSIVE PRESSU
UNIT
DISPLAY
DECIMAL UNITS
14
SCREEN
PROTECTION J
ENABLE CH X EXT
ICON 9 CHANGE 2
PASSWORD
DIALOG BOX
ENABLE/DISABLE
INDICATOR
78 manual_c100_jvscf_v4 temperatura industrial
ce
STORE
MODULE
SWITCH RECALL
TEST AUT O PRESSURE
RAMP
SLOTS
2
JUNCT.
OUTPUT
TYPES
8
VOLTS
3
INPUT EXIT CURRENT
VOLTAGE RT THERMOCOUPLE
TYPES
2 RESISTANCE
A CTIVE/
SCALE PASSIVE
TYPES
X
6
CYCLES
COLORS
LEAK TYPES
TEST AUT O
TIME 6 RT RAMP
HEADER
CH1 CH2
START/END TIMES
KEYBOARD TYPE STEP
BATTERY 2 4
SAVING ?
REFER.
JUNCTION
T CH X
INT