Professional Documents
Culture Documents
Diapo Lod - Loq
Diapo Lod - Loq
1
Internal and external QC
Internal quality control
¾ calibration
¾ blank control charts
¾ mean control charts
¾ LOD & LOQ
¾ result validation using ion balance andconductivity check
2
Why to estimate LOD and LOQ?
3
Aim of this presentation
4
IUPAC definition
xL = xbi + ksbi
where xbi is the mean of the blank measures, sbi is the standard
deviation of the blank measures, and k is a numerical factor chosen
according to the confidence level desired
5
IUPAC definition
2002
6
ISO 13530 1997
Water quality - Guidance on analytical quality control for
chemical and physicochemical water analysis
1997
7
Eurachem definition
1998
8
EPA Document 815-R-05-006
2004
9
EPA Document 815-R-05-006 2004
10
10
EPA Document 815-R-05-006 2004
11
11
EPA Document 815-R-05-006
LCMRL
2004 Lowest Concentration Minimum Reporting Level
12
Standard Methods 21° Ed. 2005: definitions
2005
13
13
Standard Methods 21° Ed. 2005: definitions
2005
14
14
Standard Methods 21° Ed. 2005: definitions
LOD LOQ
15
15
Detection Limit in analysis
NORDTEST definition
2007
16
16
SD of blank samples
The Limit of Detection (LOD) expressed as the concentration is derived from the
smallest measure, xL, that can be detected with reasonable certainty for a given
analytical procedure.
xL = xbi + ksbi
where xbi is the mean of the blank measures, sbi is the standard deviation of the
blank measures, and k is a numerical factor chosen according to the
confidence level desired (2 < k < 3)
To obtain LOD,
signals are transformed in
concentrations using the
calibration curve
signal 17
17
Calibration curve
N-NH4 spectrophotometry indophenol blue
5 cm optical path
0.9
0.5
0.3
0.1
18
18
Calibration curve
N-NH4 spectrophotometry indophenol blue
5 cm optical path
0.05
0.9
0.04
0.7
0.03
signal (A)
signal (A)
0.5
0.02
0.3
0.01
0.1 0.00
0.000 0.002 0.004 0.006 0.008 0.010
concentration
concentration (mg N-NH4 /L)(mg N-NH4 /L)
19
19
Calibration curve
Chloride by Ion Chromatography determination
Dionex AG19-AS19-AAES 100µL sample loop
2.0
1.5
signal (peak area)
1.0
0.5
0.0
0 0.5 1 1.5 2
-0.5
20
20
Calibration curve
Chloride by Ion Chromatography determination
Dionex AG19-AS19-AAES 100µL sample loop
2.0 0.20
1.5 0.15
1.0 0.10
0.5 0.05
0.0 0.00
0 0.5 0.00 1 0.05 1.5 0.10 2 0.15 0.20
-0.5 -0.05
concentrationconcentration
(mg Cl /L) (mg Cl /L)
21
21
In practice…
If you have mean control charts at different concentration,
use them to evaluate LOD and LOQ
else, if calibration slope is very stable (i.e. spectrophotometry)
use SD of the blanks (if possible)
or of a control sample of low concentration
or of the lowest standard
else,
apply Hubax-Vos method on as many calibrations possible,
or use the variability of the calibration standards
22
22
SD of blank samples
N-NH4 spectrophotometry indophenol blue
5 cm optical path
0.06
Mean
Mean = 0.025 mA
0.05 data
SD = 0.007 mA
Absorbance
0.04
xL = xbi + ksbi
0.03
xL = 0.025 + 3 x 0.007 = 0.046 mA
0.02
Calibration
0.01
mg N/L = 0.196 A + 0.0006
0.00
05
05
5
06
06
ot -07
7
9
06
08
t-0
t-0
-0
t-0
b-
n-
u-
r-
o-
r-
o
ar
se
ot
ap
ap
gi
ag
ag
ge
fe
23
Variability of the lowest standard
Chloride by Ion Chromatograpy determination
Dionex AG19-AS19-AAES 100µL sample loop
0.09
Lowest standard
0.08
0.05 mg Cl/L
0.07
Peak area mean = 0.057
Peak area
fe 6
fe 7
no 6
m 07
no 7
ap 6
ag 7
gi 8
08
m 9
09
ag 6
ge 8
gi 9
0
0
-0
0
0
0
t-0
0
0
-0
v-
v-
v-
o-
b-
o-
b-
u-
n-
u-
b-
r-
ag
ar
ot
no
24
Variability of calibration standards
Chloride by Ion Chromatography determination
Dionex AG19-AS19-AAES 100µL sample loop
15
Standard Peak RSD
mg/L area
0.05 0.058 13.8
RSD peak area
25
Using control chart
Single control chart in use
-1
0.60
from Nov 2006 to Aug 2007 Statistics mg L
UCL
0.58 Average 0.508
0.56 UWL
Standard Deviation 0.013
0.54
Standard Deviation recommended 0.025
0.52
-1
RSD % 2.7
mg L
0.50
0.48 RSD % recommended 5.0
0.46 Minimum 0.47
LWL
0.44 Maximum 0.54
0.42 LCL
N. data 115
0.40
101
106
111
1
6
11
16
21
26
31
36
41
46
51
56
61
66
71
76
81
86
91
96
25
20
IC determinations (in 20 years)
Chloride
15
Ion Chromatography
10
0
0 1 2 3 4 5 6 7 8 9 10 11 26
-1
mg L
26
LOD from control charts
Chloride by Ion Chromatography determination (100µL sample loop)
25
0
0 1 2 3 4 5 6 7 8 9 10 11
-1
mg L
27
Hubaux-Vos
28
28
LOD from daily calibration (Hubaux-Vos)
29
29
LOD from daily calibration (Hubaux-Vos)
30
LOD 0.027 mg Cl/L
30
LOD from average calibrations (Hubaux-Vos)
Chloride by Ion Chromatography determination
Dionex AG19-AS19-AAES 100µL sample loop
0.07
Range 0.05 – 10 mg/L
7 standards
0.05
More than 300 calibrations
0.03 in 2-3 years
Data
Signal
31
31
Values obtained:
Chloride by Ion Chromatography determination
Dionex AG19-AS19-AAES 100µL sample loop
32
32
In practice…
If you have mean control charts at different concentration,
use them to evaluate LOD and LOQ
else, if calibration slope is very stable (i.e. spectrophotometry)
use SD of the blanks (if possible)
or of a control sample of low concentration
or of the lowest standard used for calibration
else,
apply Hubax-Vos method,
or use the variability of the calibration standards
Don’t calculate the LOD and LOQ for determinations of:
pH, Conductivity
Alkalinity: measured through 2 end points or Gran’s method
(with Gran’s the negative values of alkalinity indicate acidity)
33
33
Conclusions
Quality Control is an important tool to criticize analytical
activity, to improve it and to assure that it is suitable for
the purpose of monitoring.
The estimation of LOD and LOQ is an important part of
Quality Control.
LOD and LOQ can be simply evaluated using results of
control charts, blank samples or calibration curves.
If you do not store these data yet, please store them
anywhere and in a few years you will have years of data...
34
34
LOD & LOQ on the web
http://goldbook.iupac.org/
http://www.eurachem.org/guides/valid.pdf
http://www.chemometry.com/Research/LOD.html
http://www.chemometry.com/Research/References/ReferencesLOD.html
http://www.epa.gov/OGWDW/methods/pdfs/methods/methods_lcmrl.pdf
http://www.nordicinnovation.net/nordtestfiler/tr569_ed_3.pdf
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2556583/
35
35
References for LOD & LOQ
A.P.H.A., A.W.W.A., W.E.F. 2005. Standard Methods for the Examination of Water and Wastewater. (Method 1030 C). Amer.
Publ. Health Ass., Washington.
Currie. L.A. 1997. Detection: international update, and same emerging di-lemmas involving calibration, the blank, and multiple
detection decisions. Chemometrics and Intelligent Laboratory Systems 37, 151-181.
Currie. L.A. 1999. Nomenclature in evaluation of analytical methods including detection and quantifcation capabilities. (IUPAC
Recommendations 1995). Analytica Chimica Acta 391, 105-126.
Currie. L.A. 1999. Detection and quantifcation limits: origins and historical overview. Analytica Chimica Acta 391, 127-134.
De Vos B., S. Lettens, B. Muys & J. A. Deckers. 2007. Walkley–Black analysis of forest soil organic carbon: recovery, limitations
and uncertainty. Soil Use and Management 23, 221–229.
EURACHEM Guide. 1998. The Fitness for Purpose of Analytical Methods. A Laboratory Guide to Method Validation and Related
Topics
Gibbons R.D. and D.E. Coleman. 2001. Statistical Methods for Detection and Quantification of Environmental Contamination.
New York: Wiley, 384 pp.
Munch D. and P. Branson. 2004. Statistical Protocol for the Determination of the Single-Laboratory Lowest Concentration
Minimum Reporting Level (LCMRL) and Validation of Laboratory Performance at or Below the Minimum Reporting Level (MRL).
EPA Document 815-R-05-006.
Mocak J., A. S. Bond, M. Mitchell, G. Schollary. 1997. A Statistical Overview of Standard (IUPAC and ACS) and New Procedures
for Determining the Limits of Detection and Quantification: Application to Voltammetric and Stripping Techniques. Pure Appl.
Chem., 69, 297-328.
Thompson M., S. L. R. Ellison and R. Wood. 2002. Harmonized guidelines for single laboratory validation of methods of analysis
(IUPAC Technical Report). Pure Appl. Chem., Vol. 74, No. 5, pp. 835–855.
Thomsen V., D. Schatzlein and D. Mercuro. 2003. Limits of Detection in Spectroscopy. Spectroscopy 18(12): 112-114.
Vogelgesang J. J. Hädrich. 1998. Limits of detection, identification and determination: a statistical approach for practitioners.
Accred Qual Assur 3: 242–255.
Voigtman E. 2008. Limits of detection and decision. Part 1, 2, 3, 4 Spectrochimica Acta Part B 63: 115–128, 128-141, 142-153,
154-165.
Janiga I., J. Mocak, I. Garaj. 2008. Comparison of Minimum Detectable Concentration with the IUPAC Detection Limit. 36
Measurement Science Review, Vol. 8, Sec. 1, No. 5, 108-110.
36