External Quality Assurance Scheme - Print Monthly All Result Summary-1

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

9/6/22, 5:34 PM External Quality Assurance Scheme - Print Monthly All Result Summary

    
  CHRISTIAN MEDICAL COLLEGE  
  DEPARTMENT OF CLINICAL BIOCHEMISTRY

  CMC EXTERNAL QUALITY ASSURANCE SCHEME

  ALL ANALYSER SUMMARY REPORT - AUGUST 2022

   
Lab Name AROGYA PATHOLOGY LABORATORY Lab No 16259

Constituent
Thyroid Hormones & Cortisol Date of Result Entered 22/08/2022
Group

PT item Lyophilized Serum Date of Report Published 05/09/2022

Constituent Method / Principle No of Participants Your


Sl.No Analyzer Name DV SDI U
Name Name Participants CV SD Value
 CLIA-  Beckman
 TOTAL
1 Chemiluminescence Acces2, Acces2I, 173 8.97 10.47 0.94   0.14
THYROXINE
immunoassay / CMIA DXI600, DXI800
 Abbott Architect
    87 6.34 7.92 0.50   0.11
( i1000 / i2000 )
 Siemens Advia
    centaur, Immulite, 187 7.07 9.13 0.65   0.09
Atellica Series
     Any Analyser 285 7.59 15.41 1.17 6.05 ug/dL -1.32  0.14
 CLIA-  Beckman
 FREE
2 Chemiluminescence Acces2, Acces2I, 140 1.70 9.82 0.17   0.03
THYROXINE
immunoassay / CMIA DXI600, DXI800
 Abbott Architect
    64 1.19 5.89 0.07   0.02
( i1000 / i2000 )
 Siemens Advia
    centaur, Immulite, 109 1.57 9.64 0.15   0.03
Atellica Series
     Any Analyser 183 1.70 17.97 0.30 2.46 ng/dL 2.49  0.05
 CLIA-  Beckman
3  TSH Chemiluminescence Acces2, Acces2I, 191 3.76 6.22 0.23   0.03
immunoassay / CMIA DXI600, DXI800
 Abbott Architect
    95 3.58 7.25 0.26   0.05
( i1000 / i2000 )
 Siemens Advia
    centaur, Immulite, 196 3.97 10.23 0.41   0.06
Atellica Series
4.47
     Any Analyser 341 4.61 12.14 0.56
uIU/mL -0.25  0.06
 CLIA-  Beckman
4  TOTAL T3 Chemiluminescence Acces2, Acces2I, 159 160.96 10.38 16.71   2.65
immunoassay / CMIA DXI600, DXI800
 Abbott Architect
    83 155.44 7.13 11.09   2.43
( i1000 / i2000 )
 Siemens Advia
    centaur, Immulite, 176 192.57 7.88 15.18   2.29
Atellica Series
     Any Analyser 278 176.45 13.41 23.67 166 ng/dL -0.44  2.84

SDI Range Interpretation


Within -1.00 to +1.00    Excellent.
Within ±1.01 to ±2.00    Good.
Within ±2.01 to ±2.99    Accept with caution. Warning Signal.
Beyond ±3.0    Unacceptable performance. Action Signal.

 
LAB ADDRESS:
AROGYA PATHOLOGY LABORATORY
NO.3, SAI GANGA COMPLEX, KALUBHA ROAD, KALANALA, NEAR RASOI DINING HALL
BHAVNAGAR
GUJARAT,364001
         
home.cmcvellore.ac.in/clinqc/ViewMonthlySummary/printAllResult_new.aspx?resDetId=256&noSamples=1&month=8&year=2022&groupId=2&ye… 1/2
9/6/22, 5:34 PM External Quality Assurance Scheme - Print Monthly All Result Summary

Coordinator Contact details: Dr. Pamela Christudoss


Email:clinqc@cmcvellore.ac.in CMC EQAS Coordinator
Contact Number: 0416-2283102 Christian Medical College, Vellore
 
Homogeneity and Stability of the sample is passed.
Data in CMC EQAS reports is confidential
CMC EQAS does not sub contract any components
******* End of Report ******
 
    