Scratchpad sequences - special test sequence ............ to test different components of
the VJ 1000 Series. Using mnemonics see the list below: SYSTEM MENU.
Examples:
To activate the VG Valve for 2 seconds and immediately deactivate it, the sequence
would be:
VG1,WM2000,VG0
To activate the VG Valve for 2 seconds and deactivate it immediately and repeat this
for 3x the sequence would be:
J03,VG1,WM2000,VG0,WM2000,J
1
To turn on the ink micropump and go to a pressure of 3.00 bar, leave it on for 10
seconds and repeat this for 2x, the sequence would be:
J02,DP0300,WM10000,DP0001,WM10000,J
To start the vacuum pump at 100%, energize the Vc Valve for 2 seconds,
leave it on for 10 seconds and repeat this for 2x, the sequence would be:
J02,DG1,WM5000,VC1,WM2000,VC0,WM2000,DG0,WM1000,J
To activate the 4 Valves of the VF, VL, VG and VP cylinder head for 1 sec.
and repeat this for 4x then, the sequence would be:
J04,VF1,WM1000,VF0,WM1000,VL1,WM1000,VL0,WM1000,VG1,WM1000,VG
0,WM1000,VP1,WM1000,VP0,WM1000,J
To measure the modulation signal in Bico / Nozzle for 20 seconds, the sequence
would be:
J05,DM0050,WM20000, DM0080,WM20000, DM0150,WM20000,J
Hold the pump at 2.00 bar, wait 1 second, turn off the pump, wait 1 sec, the
sequence would be:
J10,DP0200,WM1000,DP0001,WM1000,J
Attach the vacuum pump and click on the Vc valve 3x then the sequence would be:
DG1,J03, VC1,WM1000,VC0,WM1000,J,DG0
Fill inkcore with make-up:
VA0,DG1,VC1,VT1,VM1
Fill inkcore with ink :
VA0,DG1,VC1,VT1,VI1
Umbilical purge using empty cartridges:
VA0,DP200,VV1,VR1,VI1
Then whichever valve you working on (as typed in the sequence) , while the continuous
clicking sound is being heard, spray continuously, cleaning solution inside the holes
where the gasket is.
You can do for all four valves one after the other.