PAT E N T P E N D I N G
BR-0238-M MEPS Brochure.qxd 12/4/07 9:54 AM Page 2
MEPS
FRITS PACKED BED
END PLUG
NEEDLE
BR-0238-M MEPS Brochure.qxd 12/4/07 9:54 AM Page 3
M E P S : M I C R O E X T R A C T I O N B Y PA C K E D S O R B E N T
BR-0238-M MEPS Brochure.qxd 12/4/07 9:54 AM Page 4
1 2 3 4 5 6 7 8 9 10
R E T E N T I O N T I M E ( M IN U T ES )
BR-0238-M MEPS Brochure.qxd 12/4/07 9:54 AM Page 5
How Does
MEPS compare?
Accuracy and Precision
Table 1 summarizes the precision TABLE 1 : Comparison of accuracy and precision between MEPS
and accuracy results obtained and other methods for ropivacaine (local anesthetics)
analyzing Ropivacaine by four Method Ropivacaine Accuracy Precision (RSD%) Handling time
LOD (nM) (%) (Inter-assay)
sample preparation methods:
MEPS, Liquid-Liquid Extraction [1] MEPS / GC-MS 2 105 5.0 1 min
(LLE), conventional SPE and Solid [2] LLE / GC-MS 2 101 3.8 20 min
Phase Microextraction (SPME). [3] SPE / LC-UV 100 101 3.0 20 min
Compared to SPME, also a µSPE [4] SPME / GC-MS 5 110 6.3 40 min
SPME and SBSE (Stirring Bar Sorbent MEPS SPME SBSE MEPS SPME SBSE MEPS SPME SBSE MEPS SPME SBSE
results from the MEPS and SBSE Fluoranthene 100 84 100 9 4 4 5 100 1.2 2 30 200
techniques are significantly better Fluorene 103 96 97 5 5 4 1 40 0.7 2 30 200
than SPME but MEPS processes each Pyrene 115 86 100 7 3 3 1 40 0.7 2 30 200
sample 100 times faster than SBSE. M. Abdel-Rehim / J. Chromatog. A 1114 (2006) 234-238
M E P S : M I C R O E X T R A C T I O N B Y PA C K E D S O R B E N T
BR-0238-M MEPS Brochure.qxd 12/4/07 9:54 AM Page 6
Is there significant FIGURE 5 Phenols at 25ppb in alkane contaminated water. 10 x 100µL cycles
on C18 MEPS cartridge (conditioned with methanol and water). Sample was
carry over using eluted with methanol 10µL, 2µL injected onto a BPX5 column and analyzed
by GC-MS.
MEPS?
Devices used for SPE are 2.4-dichlorophenol 4-chloro-3-methylphenol
traditionally considered to be single
use. The precision engineering used TIC 2.4-dimethyphenol 2.4.6-trichlorophenol
220000
in the design and manufacture of
MEPS allows for simple wash steps 2-chlorophenol
and consequently the re-use of the
device. To demonstrate this we phenol
extracted phenols from waste
water and measured the carry over Phenols 25ppb
between experiments. Extraction 2
Figure 5 shows the chromatograms Extraction 3
for the three 10µL extractions.
The first is labeled “phenols 3 5 7 9 11 13
25ppb,” the second is “extraction R ETEN TI O N TI ME ( MIN U T ES )
2,” and the third “extraction 3”.
Clearly all of the phenols were
eluted in the first 10µL TABLE 3 : Comparison of carryover and wash regime.
of Methanol used.
Wash volume (uL) # of washes and Carry-over Source
wash solution
One of the most often stated
PAH's in water 50µL 4X methanol, 5X water 0.2% - 1% M. Abdel-Rehim / J.
reasons for the use of disposable
Chromatog. A 1114
SPE cartridges is the issue of carry (2006) 234-238
over. Results from 5 studies are Anesthetics in 50µL 4X methanol, 4X water ~ 0.2% (I.S.) M. Abdel-Rehim / J.
summarized in Table 3. Utilizing a Human Serum of Chromatography B,
series of washes using eluent and 801 (2004) 317-321
then wash solution the carry over Roscovitine in 50µL 5X methanol/water (95:5, v/v), <0.1% M. Abdel-Rehim / J.
was virtually eliminated. With a plasma and urine 5X water/methanol (90:10, v/v) of Chromatography B,
817 (2005) 303-307\
cycle time measured in seconds
Roscovitine in 50µL 4X methanol, 4X water <0.01% M. Abdel-Rehim / J.
for MEPS this washing takes less Human plasma Mass Spectrom.
than 5 minutes. To accomplish 204;39:1488-1493
the same for conventional SPE Olomoucine in 50µL 5X methanol/water (95:5, v/v), <0.1% M. Abdel-Rehim et al. /
takes an hour or more and uses Human plasma 5X water/methanol (90:10, v/v) Analytica Chimica
significant amounts of solvents. Acta 2005
Summary
Table 4 summarizes the
comparison of MEPS, SPME and
conventional SPE. MEPS requires TABLE 4 : Comparison of MEPS, SPME and conventional SPE.
much less time than either SPE or Factor MEPS SPE SPME
SPME, and shows much better Amount sorbent 0.5-2 mg 50-2000 mg thickness 150 mm
recovery and sensitivity than SPME. Sample prep. time 1-2 min 10-15 min 10-40 min
MEPS also eliminates any Cartridge use 40 to 100 extractions once 50-70 extractions
intermediate steps between the Recoveries good good low
sample preparation steps and the Sensitivity good good low
injection into a GC or LC system. M. Abdel-Rehim / J. of Chromatography B, 801 (2004) 317-321
BR-0238-M MEPS Brochure.qxd 12/4/07 9:54 AM Page 7
C18 MEPS Cartridge for CTC Analytics systems using 250µL syringes 006292 5 2900301
Silica MEPS Cartridge for CTC Analytics systems using 250µL syringes 006292 5 2900302
C8+SCX MEPS Cartridge for CTC Analytics systems using 250µL syringes 006292 5 2900303
C2 MEPS Cartridge for CTC Analytics systems using 250µL syringes 006292 5 2900304
C8 MEPS Cartridge for CTC Analytics systems using 250µL syringes 006292 5 2900306
MEPS Development kit for CTC Analytics systems using 250µL syringes 006292 5 2900305
(contains 1 each of C18, C8, C2, SILICA and C8+SCX)
C18 MEPS Cartridge for CTC Analytics systems using 250uL syringe 006292 5 2900501
Silica MEPS Cartridge for CTC Analytics systems using 250uL syringe 006292 5 2900502
C8+SCX MEPS Cartridge for CTC Analytics systems using 250uL syringe 006292 5 2900503
C2 MEPS Cartridge for CTC Analytics systems using 250uL syringe 006292 5 2900504
C8 MEPS Cartridge for Thermo Scientific and CTC Analytics systems 006292 5 2900506
MEPS Development kit for CTC Analytics systems using 250µL syringes 006292 5 2900505
(contains 1 each of C18, C8, C2, SILICA and C8+SCX)
Base material is silica with mean particle size of 45µm and pore size of 60Å.
M E P S : M I C R O E X T R A C T I O N B Y PA C K E D S O R B E N T
BR-0238-M MEPS Brochure.qxd 12/4/07 9:54 AM Page 8
BR-0238-M
AUSTRALIA & PACIFIC REGION FRANCE INDIA
SGE Analytical Science Pty Ltd SGE Europe Ltd (France) SGE Laboratory Accessories Pty Ltd
Tel: +61 (0) 3 9837 4200 Tel: +33-1-69 29 80 90 Tel: +91-22-24715895
Fax: +61 (0) 3 9874 5672 Fax: +33-1-69 29 09 25 Fax: +91-22-24716592
Email: info@sge.com E-mail : france@sge.com E-mail : sgeindia@sge.com