Gastrointestinal Endoscope Training Material

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 118

Gastrointestinal Endoscope

Training Material
Endoscope Technical Support Team
No.: XXX
Version: V1.0

赵喜波
3680
29.Sep.2018

1
Content

⚫ Endoscope Structure & Theory


⚫ Cleaning, Disinfection, Sterilization
⚫ Spot Inspection

2
Endoscope Structure & Theory
◆Endoscope Structure
◆Endoscope Working Theory
◆550 Series Endoscope Difference
◆Endoscope model comparison

3
Endoscopy System Composition

4
Endoscope Structure

Endoscope

1 2 3 4

Connector Section Control section Insertion Section Distal End

5
Endoscope Structure
Connector Section

Integrating the connectors of light, water feeding, air


feeding, electrical signal, suction, auxiliary water feeding.

Note:330 series endoscope do not have auxiliary water feeding function.

6
Endoscope Structure
Connector Section
No. Part Name Description

Used for connecting the endoscope cable to acquire video


1 Electrical connector
signal of image

2 Suction port Used for connecting to the suction pump

Auxiliary water-feeding Used for connecting to water pump or syringe to make water
3
port or solution jetted from the distal end

Used for protecting the auxiliary water-feeding port from


Cap for auxiliary
4 being blocked by foreign objects. Cover the auxiliary water-
water-feeding port
feeding port with the cap when the port is not in use

Connect it to the electrical connector when immersing the


5 Waterproof cap
endoscope for cleaning and disinfection

Electrosurgical equipment
6 Used for connecting the electrosurgical equipment
connector

Used for connecting to the water bottle to feed water/air to


7 Water bottle port
the distal end

7
Endoscope Structure
Control Section & Insertion Section

8
Endoscope Structure
Control Section & Insertion Section

No. Part Name Description

● Rotating it to F direction (clockwise) frees angulation in the


Left/right angulation left/right direction.
8
lock ● Rotating it in the opposite direction (anticlockwise) locks the
bending section at any desired angle

Left/right angulation ● Rotate it to R direction, the bending section moves right.


9
control knob ● Rotate it to L direction, the bending section moves left

● Rotating it to F direction (clockwise) frees angulation in the


Up/down angulation up/down direction.
10
lock ● Rotating it in the opposite direction (anticlockwise) locks the
bending section at any desired angle.

Up/down angulation ● Rotate it to U direction, the bending section moves up.


11
control knob ● Rotate it to D direction, the bending section moves down.

9
Endoscope Structure
Control Section & Insertion Section
No. Part Name Description
The functions of these buttons can be set through the
12 Remote buttons (0-3) image processor used with the endoscope. For details,
refer to the user manual of relevant image processor.
This channel should be used with the biopsy valve and
the functions are as follows.
● Used to feed the liquid to the distal end of the
13 Instrument channel
endoscope.
● Used for the endo-therapy accessory.
● Used as a suction channel.
Indicates the maximum length of the insertion section
14 Insertion limit mark
that can be inserted into the body
15 Suction valve Press it to extract the liquid, debris or gas from the body

● Cover the hole on the valve with finger to feed air and
press the valve to feed water.
16 Air/water valve
● The air and water can remove the blood, debris or
mucous membrane adhering to the objective lens.

10
Endoscope Structure

Control Section

Knobs

Gear, Chain
Bracket

Buttons

Valves

Control Section Shell

The gear, chain and steel wire are controlled by knobs, then
drive the snake bone of bending section to realize bending.
11
Endoscope Structure
Insertion tube inner structure
1.4 电子内窥镜结构——插入部
Spring coil
Angle steel wire Metal helix tube Metal knitmesh Synthetic resin tube
Light fiber

Instrument tube

Light fiber

CMOS cable

Air/Water tube

The outermost layer of insertion tube is resin


tube; second layer is metal knitmesh; third layer
is helix tube; The interior contains CMOS cable,
light fiber, instrument tube, air water tube and
the steel wire which control the bending.

12
Endoscope Structure
Bending section

Snake bone

Metal knitmesh

Bending rubber

The bending function is realized by snake bone which


has four bending directions (gap between joints ), the
metal knitmesh and bending rubber are packed outside
the snake bone.
13
Endoscope Structure
Distal end

21. Instrument channel outlet

20. Auxiliary water 18. Light guide lens


feeding outlet

17. Objective lens

19. Air/water nozzle

14
Endoscope Structure
Distal end

No. Part Name Description

Used to transmit the optical signals of the observed


17 Objective lens area to the CMOS sensor.

Used as an outlet for feeding air or water to clean the


18 Air/water nozzle objective lens. Feed air to inflate the stomach.

Light is transmitted through these lenses to assist


19 Light guide lens endoscopic image observation.

Instrument channel Used as an instrument (such as biopsy forceps) outlet ,


20 outlet liquid feeding outlet or suction inlet

Auxiliary water feeding


21 outlet Water or cleaning solution eject out from this outlet.

15
Endoscope Structure
Distal
1.4end
电子内窥镜结构——头端

Objective lens Distal end cover


CMOS Sensor Light fiber Light guide
lens
Air/water Auxiliary water
Nozzle feeding tube
tube

Light guide lens Instrument


tube connector
Distal end base
Instrument tube
Light fiber

Including imaging module, light guide lens assembly,


air/water feeding tube, auxiliary water feeding tube.

16
Endoscope system block diagram

Connector Control
Endoscope Distal end
section section

17
Endoscope Structure & Theory
◆Endoscope Structure
◆Endoscope Working Theory
◆550 Series Endoscope Difference
◆Endoscope model comparison

18
Endoscope Theory
Air/water feeding & suction working flow

19
Endoscope Theory
Air/water feeding working flow

Air

20
Endoscope Theory
Air/water feeding working flow

Block the hole Air

21
Endoscope Theory
Air/water feeding working flow

Push the A/W valve

P
Water
A/W
three-way

22
Endoscope Theory
Air/water feeding working flow
1.4 电子内窥镜结构——操作部水气阀 Blue mark

Base

Rubber bowl
Air hole
Sealing ring

Distal
Connector
end
side
side
Air

Air

Water

Water

23
Endoscope Theory
Air/water feeding working flow
1.4 电子内窥镜结构——操作部水气阀 Blue mark

Base

Rubber bowl
Air hole
Sealing ring

Distal
Connector
end
side
side
Air

Air

Water

Water

24
Endoscope Theory
Air/water feeding working flow Blue mark

Block
Base

Rubber bowl
Air hole
Sealing ring

Distal
Connector
end
side
side
Air

Air

Water

Water

25
Endoscope Theory
Air/water feeding working flow
1.4 电子内窥镜结构——操作部水气阀 Blue mark

Push Base

Rubber bowl
Air hole
Sealing ring

Distal
Connector
end
side
side
Air

Air

Water

Water

26
Endoscope Theory
Suction working flow
Connector
section

Suction
valve

Air

Instrument
channel

Suction
pump

Suction tube

Instrument tube
Instrument channel outlet

27
Endoscope Theory
Suction working flow
Connector
section

Suction
valve

Instrument
channel

Suction
pump

Suction tube

Instrument tube
Instrument channel outlet

28
Endoscope Theory
Suction working flow

Push

Air
Distal end
instrument
outlet

To Suction pump To Suction pump

29
Endoscope Theory
Suction working flow
1.4 电子内窥镜结构——操作部吸引阀

Red mark

Groove
Base Hole

Front view Bottom view

30
Endoscope Theory
Biopsy valve

slit
Biopsy valve
Valve cap
Hole

Base
Instrument
channel port

Through the rubber elasticity to guarantee the slit keep


sealing while using instrument.

31
Endoscope Theory
Light transmission route Light

Light control signal

Light source Image processor


Light control cable

Video signal CMOS Objective lens

Objective
image

Light
Distal end

32
Endoscope Theory
Image signal flow

Monitor DVI cable

Image signal

Image processor

Electrical signal CMOS Objective lens

Objective
image

Light
Distal end
33
Endoscope Structure & Theory
◆Endoscope Structure
◆Endoscope Working Theory
◆550 Series Endoscope Difference
◆Endoscope model comparison

34
550 Endoscope Difference
Connector section

Electrical contacts

Image transmission fiber


cable connector

1. Cancel the ODU head and endoscope cable, waterproof design electric pin
2. Cancel the waterproof cap, the leakage detection port is sealed.

35
ODU Head of Connector Section of 500

36
550 Endoscope Difference
Connector section
No. Part Name Description
Image transmission
1 Used for transmitting image signals.
optical fiber connector
2 Electrical contacts Used for passing electrical signals.
High frequency Used for connecting the high frequency electrosurgical
3
feedback earthing rod generator
Used for connecting to the water bottle to feed water/air to
4 Air/water feeding ports
the distal end.

● Used for protecting the auxiliary water-feeding port.


● Used for preventing the auxiliary water-feeding channel
Cap for auxiliary from being blocked by foreign objects.
5
water-feeding port ● Used for preventing gas or liquid in patient’s body from
blowing or flowing back during examination. Cover the cap
when the auxiliary water-feeding tube is not in use.

6 Auxiliary waterfeeding port Used for connecting to water pump or syringe

7 Suction port Used for connecting to the suction pump.


8 Leakage detection port Used for connecting the leakage detector

37
550 Endoscope Difference
Control section

550 500

The control section of 550 is slimmer than 500, more easily for handling

38
550 Endoscope Difference
Control section

550 500
Remote button layout is different, 550 is more easily for handling.

39
550 Endoscope block diagram
Auxiliary Auxiliary Auxiliary Water
Auxiliary water feeding tube
water pump water port water outlet

Suction tube
Suction Suction Instrument tube Instrument
Suction port
pump valve outlet

Grounding

Endotherm
knife Instrument
channel

Knobs Steel wire Snake bone


Instruments Patient body

Remote
Cable
buttons

Electrical signal image


Connector Calbe Optical- Objective
Cable CMOS
electrical lens
Image optical signal Fiber
Optical port converter
A/W
Water bottle Water port Water
A/W
A/W valve Nozzle
Three-way
Air port Air

Light
Light guide
Light source Light fiber
lens

Image
processor
In 550 endoscope, convert the electrical signal to optical signal could
enhance the capacity of anti-interference.
40
550 Endoscope Difference
Image signal flow DVI cable

Monitor
Image optical signal

Electrical signal

Light control cable

Light source Image processor

Armoured optical fiber

CMOS Objective lens

Objective
Image signal image

Light
Endoscope Distal end

41
Endoscope Structure & Theory
◆Endoscope Structure
◆Endoscope Working Theory
◆550 Series Endoscope Difference
◆Endoscope model comparison

42
Endoscope model comparison
Instrument Bending
Work Max insertion
Series Type Model channel section
length section diameter
diameter diameter
EG-550 2.8 1050 9.3 10.5
Gastroscope
EG-550L 3.2 1050 9.8 11.5
EC-550 3.8 1350 12.5 13.8
550
EC-550L 4.2 1350 12.9 14.5
Colonoscope
EC-550T 3.8 1700 12.5 13.8
EC-550L/T 4.2 1700 12.9 14.5
EG-500 2.8 1050 9.3 10.5
Gastroscope
EG-500L 3.2 1050 9.8 11.5
EC-500 3.8 1350 12.5 13.8
500
EC-500L 4.2 1350 12.9 14.5
Colonoscope
EC-500T 3.8 1700 12.5 13.8
EC-500L/T 4.2 1700 12.9 14.5

L:Large instrument channel


T:Total length

43
Endoscope model comparison

Instrument Bending
Work Max insertion
Series Type Model channel section
length section diameter
diameter diameter
Gastroscope EG-330 2.8 1050 9.3 10.0

320/330 EC-330 3.8 1600 12.9 13.8


Colonoscope
EC-330T 3.8 1600 12.9 13.8

The model label of 320 and 330 scopes are same, both mark 330
on the label, distinguish from color ring.

320 Ring:Green
330 Ring:Purple

44
Endoscope model comparison

Image
Auxiliary Waterproof Imaging Pre emphasis A/W
Series Ring color transmission
water cap sensor board three-way
optical fiber

320 Green No Yes CCD No No No

330 Purple No Yes CCD No No No

430 Yellow Yes Yes CMOS No Yes Yes

500 Yellow Yes Yes CMOS No Yes Yes

550 Yellow Yes No CMOS Yes Yes Yes

Note:The waterproof cap of 330 and 500 scope is different.

45
Content

⚫ Endoscope Structure & Theory


⚫ Cleaning, Disinfection, Sterilization
⚫ Spot Inspection

46
Spaulding Classification

Classification Definition Level of Disinfection Application

Biopsy forceps, snare,


A device that enters injection needle, cytology
Critical normally sterile tissue or Sterilization brush, incision knife, guide
the vascular system wire, dilating balloon, urinary
catheter, sphincterotomes

A device that comes in


contact with intact
Semi-Critical mucous membrane. Not High (Disinfection) Endoscope
contact sterile tissue,
and broken mucosa.

A device that doesn't


ordinarily touch the Stethoscope, blood pressure
Noncritical Medium or Low
patient or touch only meter
intact skin.

47
Basic Washing Accessories

Waterproof cap Leakage detector

48
Basic Washing Accessories

Channel plug

Non-return valve

Injection tube

49
Basic Washing Accessories

50
CDS procedure
Exam room 1 Pre-Cleaning

Contaminated Area

2 Leakage Test Filtered water

Enzyme
3 Manual Cleaning
solution

4 Rinsing Filtered water


CDS room
Cleaning area
Auto cleaning
Disinfection or machine
5 Disinfectant
Sterilization

6 Terminal Rinsing Sterile water

7 Drying 70% Alcohol

Storage room 8 Storage

51
1. Pre-Cleaning
Step1 – Clean insertion tube

Do pre-cleaning as soon
as finishing the exam.

Wipe the insertion tube with a detergent dampened lint-free cloth after
withdrawing the endoscope from the patient's body.

52
1. Pre-Cleaning
Step2 – Flush suction channel

Press suction valve

Enzyme solution

Suck water and air alternately, until the water and air could be seen
go into the suction bottle. Suck detergent for 30s, suck air for 10s,
repeat at least 3 times

53
1. Pre-Cleaning
Step3 – Flush air/water channel and nozzle

Press Block

Press A/W valve to feed water for 10-15s;


Block A/W valve air hole to feed air for 10-15s;
Feed air and water alternately at least 3 times.

54
1. Pre-Cleaning
Step4 – Flush Auxiliary water feeding channel

Flush water into auxiliary water feeding port;


Flush air into auxiliary water feeding port;
Repeat feeding water and air at least 3 times

55
1. Pre-Cleaning
Step5 – Shut down the light source and processor

Shut down

HD-500, HD-550, VLS-50 and HDL-500X support endoscope


hot plugging, could shut down after the last exam of the day.

56
1. Pre-Cleaning
Step6 – Take down the accessories

Take down air/water, suction and biopsy valve then


immerse them in detergent solution.

57
1. Pre-Cleaning
Step7 – Unplug all connectors and tubes

Remove suction
tube.
Unplug endoscope cable
Unplug suction pump tube
Unplug water bottle connector

58
1. Pre-Cleaning
Step8 – Put on waterproof cap

Put on waterproof cap;


Unplug the scope from the light source;
Move the scope to CDS room to do further cleaning.

59
1. Pre-Cleaning

Caution!

The light guide will be very hot after using, do not touch
as soon as unplugging from light source.

60
2. Leakage Test

Leakage test theory

The area between the blue border and the green tube is
the air inflating space while doing leakage test.

61
2. Leakage Test

Dry test

The bending rubber will be expanded


after being inflated. It is normal.

Put the endoscope on the table, connect leakage detector and add
pressure to 22Kpa(180mmHg);
Maintain the air pressure for 3 mins, if the pointer does not decrease
continually, the test is pass.

62
2. Leakage Test

Wet test

Caution: Do not put the leakage detector into water

Immerse the scope which already add pressure into the water for 3 mins;
Turn the U/D and L/R knobs to adjust the angle of bending section;
If there are bubbles coming out, the leakage test is fail.

63
2. Leakage Test

Wet test

Take out the scope from the water, open the vent valve of the detector to release the
pressure until the pointer goes back to zero;
Disconnect the leakage detector from the waterproof cap.

64
2. Leakage Test

Caution!

Do dry test before wet test, make sure there are no serious leakage,
otherwise the water will flow into the scope to cause further damage;
If the wet test is failed, stop doing CDS, otherwise it will cause further
damage;

65
3. Manual Cleaning

Caution:
Do not make too
much foam of the
detergent solution

1. Immerse the endoscope into detergent solution(Enzyme cleaning fluid)


2. Wipe the outer surface of the endoscope.

66
3. Manual Cleaning

3. Brush the suction channel from suction valve cylinder to distal end.
4. Repeat several times until there are no residue on the brush head

67
3. Manual Cleaning

5. Brush the suction channel from the suction valve cylinder to the suction port;
6. Brush the instrument channel from the instrument channel port to the distal end.
7. Repeat several times until there are no residue on the brush head

68
3. Manual Cleaning

8. Brush the suction port, instrument channel port, air water valve cylinder
and suction valve cylinder.
9. Brush the air/water valve, suction valve and biopsy valve.

Hospital could connect the auto cleaning


machine after this step
69
3. Manual Cleaning

10. Connect the channel plug and injection tube to the endoscope, inject
detergent solution into the suction and A/W channel until no bubbles
comes out of the distal end.

70
3. Manual Cleaning

Caution: Change enzyme solution for each scope

11. Disconnect the channel plug and injection tube, immerse the endoscope
in the detergent solution.
12. Wipe the outer surface and all the ports with a lint-free cloth

71
4. Rinsing

1. Rinse the outer surface and all valves of the endoscope


with flowing water.

72
4. Rinsing

2. Immerse the endoscope in the clean filtered water.


3. Connect channel plug and injection tube, inject filtered water into the
A/W and suction channel.
4. Inject air to A/W and suction channel to drain the water.

73
4. Rinsing

Wipe dry the endoscope, otherwise the remaining water


will decrease the concentration of the disinfectant

5. Take the endoscope and accessories out of the rinsing tank.


6. Wipe dry the outer surface and accessories of the endoscope with a lint-free cloth

74
5. Disinfection

1. Completely immerse the endoscope and accessories into the disinfectant solution.
2. Connect channel plug and injection tube, inject disinfectant solution into the A/W
and suction channel until no bubbles comes out of the distal end.

75
5. Disinfection

3. Disconnect channel plug and injection tube, immerse them in


disinfectant solution together with the endoscope.
4. Immerse the endoscope and accessories according to the requirement
of the disinfectant solution.

76
5. Disinfection
Recommended Disinfectant
High-Level
Contact Contact
Disinfectant Disinfection Precautions
Period Type
Parameter

Phthalic 1. The cloth, skin, and instrument are prone to dyeing.


Level: 0.55%
dicarboxaldehyde ≥ 5min Immersion 2. Steam of this disinfectant may stimulate the respiratory tract and
(0.5%-0.6%)
(OPA) eyes.

1. This disinfectant has sensitization and irritation on the skin, eye,


and respiratory
tract. In addition, it can cause dermatitis, conjunctivitis, nasal cavity
Glutaraldehyde Level: ≥ 2% inflammation,
≥ 10min Immersion
(GA) (alkaline) and occupational asthma. It is suitable for use in the automatic
disinfection machine.
2. This disinfectant is easy to condense on the endoscope and
cleaning and disinfection devices.

1. This disinfectant has irritation on the skin, eye, and respiratory


Peroxyacetic acid Level: 0.2%-
≥ 5min Immersion tract.
(PAA) 0.35% (W/V)
2. This disinfectant has strong corrosion on metal.

Active chlorine
level: (60±10)
1. When organic matters exist, the disinfection effect decreases
mg/L, pH
sharply.The endoscope should be
Acidic value: 2.0-3.0,
cleaned completely before the disinfection.
electrolyzed chlorination
≥ 10min Immersion 2. This disinfectant has certain corrosion on the endoscope.
oxidizing water reduction
3. Flowing immersion method should be adopted for disinfection.
(AEOW) potential:
4. The endoscope should be rinsed with sterile or filtered water for
1100mV,
30s after disinfection.
Residual chloride
ion level:<1000mg/L

77
5. Disinfection

5. Take the endoscope and accessories out of the disinfectant solution, connect the
channel plug and injection tube.
6. Inject air to A/W and suction channels to drain the disinfectant solution,then
disconnect the channel plug and injection tube.

78
6. Terminal Rinsing

2. Connect the channel plug and injection tube.


3. Inject sterile water into the A/W and suction channels until no bubbles
come out of the distal end (To ensure no disinfectant remaining )

79
6. Terminal Rinsing

4. Take out the endoscope of the rinsing tank.


5. Inject air into A/W and suction channels to drain the water inside the channels.
6. Disconnect the channel plug and injection tube, wipe dry the outer surface and
accessories with lint-free cloth.

80
7. Drying

1. Place the endoscope on a sterile mat which is placed on a drying table.


2. Connect channel plug and injection tube, put the filter of the injection
tube into 70% alcohol.
3. Inject alcohol into the A/W and suction channel.
4. Inject air into the A/W and suction channel to blow-dry the channel wall.

81
7. Drying

5. Wipe dry the outer surface and accessories with alcohol


moistened lint-free cloth.

82
8. Storage

Do not store the endoscope in suitcase

1. Free the angulation locks, adjust the bending section to straight status.
2. Hang the endoscope in the endoscope cabinet and ensure the insertion section is
vertically hanging.
3. Place the accessories dispersedly, ensure each accessories no touching.

83
Content

⚫ Endoscope Structure & Theory


⚫ Cleaning, Disinfection, Sterilization
⚫ Spot Inspection

84
Spot inspection significance

Spot inspection is checking the endoscope system according to


corresponding standard every 3-4 months, in order to find the
hidden trouble timely, and solve it actively, to enhance the
endoscope using experience and life span.

85
Spot inspection process

Start

Confirm no leakage of each


Leakage
section, and no block of the tube
test

Confirm the appearance of each


section, distal end-bending
Appearance section-insertion section-control
inspection section-connector section.

Confirm angle adjust, light, image,


Function air/water feeding, suction, buttons,
inspection image processor, light source

End

86
Spot Inspection
◆Leakage test
◆Appearance inspection
◆Function inspection

87
Leakage test

Inflate to 22Kpa Swing the tube Adjust angle to test


bending rubber and knobs

The basic step is same with CDS, below extra operations are needed while doing wet test for spot
inspection.

1. Lift knobs lightly, confirm if the knobs is leaking or not, especially the sealing ring

2. Swing the insertion tube and universal cable segment by segment, especially check the bite mark.

3. Press remote buttons from different directions, confirm the buttons no leaking.

4. Use blowing dust ball to feed water into A/W, suction, auxiliary water channels, confirm there are
no bubbles come out of the channel.

88
Leakage test

If there is leakage problem, use waterproof tape to seal the leak point
temporarily, then find out other leak points.

89
Leakage test
Ensure instrument channel is unblocked

Use brush to go through below channels:


1. Suction valve cylinder to instrument channel outlet
2. Suction valve cylinder to suction port
3. Instrument channel port to instrument channel outlet

90
Spot Inspection
◆Leakage test
◆Appearance inspection
◆Function inspection

91
Distal end

Distal end cover:


scratch, breakage, especially the
instrument channel outlet nearby.
Auxiliary water outlet: block

Light guide lens:


scratch, crack, glue breakage

Objective lens:
Nozzle: loose, sharp, block
scratch, crack, glue breakage, fog

92
Bending rubber

Snake bone pile up, Rubber scratch, breakage

93
Insertion tube

Scale mark disappear

Boot cone: breakage

Insertion tube:
dents, bulge, scratches, breakage, deformation,
Foreign body attached, protein coagulation,
hydrophilic coating peeling.

94
Control section

Knobs:
scratches, deformation, slack

Instrument channel port:


wear out

A/W valve: sealing ring missing, breakage


Suction valve cylinder: ware out Remote buttons:
breakage, aging, scratch

95
Universal cable

Universal cable:
dents, bulge, scratches, breakage, deformation,
Foreign body attached, protein coagulation,
hydrophilic coating peeling.

Boot core: breakage

96
Connector section

Light guide cap: loose


Color mark:missing

Electrical Pin: Plug base pin: missing Image fiber connector:stain


deformation, rust, loose

97
Spot Inspection
◆Leakage test
◆Appearance inspection
◆Function inspection

98
Function inspection

Bending angle

Measure the max angle, confirm the knobs tightness, ensure could locked tightly.

Gastroscope: U 210° D 90° L 100° R 100°


Colonoscope: U 180° D 180° L 160° R 160°
Tolerance: -10%

99
Function inspection

Light guide fiber

Let the distal end face to normal lamp, disassemble the light guide
cap, use magnifying glass to inspect how much light guide fibers are
broken at the light guide end. It needs to be repair if the fiber silks
are broken over 30%.

100
Function inspection

Instrument channel

Ensure the instrument channel is not blocked.

Adjust angle for four directions(gastroscope up angle adjust to 180°), other


angles adjust to max, use biopsy forceps to check the channel is blocked or not.

Caution: Do not open the forceps inside the channel.

101
Function inspection
Image check-not turn on lamp

1. Connect the scope to processor and light source, switch on the processor.
2. Check the image in white balance cap, confirm there is no color spot, noise and
other abnormal image.

102
Function inspection
Image check-not turn on lamp

1. Shake the endoscope cable, confirm the image is OK.


2. Adjust angle to each directions(8 directions), confirm the image is OK.

103
Function inspection
Image check-turn on lamp

1. Use alcohol wet cotton swab to clean the objective lens surface
2. Do white balance then confirm the image color and no dark corner.
3. Cover the objective lens with finger and move the finger on the lens ,
confirm there is no any not moving shadows on the image.
4. Clean the objective lens with cotton swab again.

104
Function inspection

Light control function check

Light control cable mark Move close to distal end Move away from distal end

1. Check the connection of the light control cable between the


light source and processor.
2. Move hand close to and away from the distal end repeatedly,
confirm the auto light control is ok.

105
Function inspection

Air/water feeding function check

1. Install the air/water valve, suction valve, connect the suction pump and water bottle.
2. Switch on the air pump, immerse the distal end into the water, there should not be
bubbles come out in natural status.
3. Feeding air in water, there should be amounts of air bubbles come out of the water
4. Feeding water in water, after the air in the tube been exhausted, confirm there is no
bubble comes out of the water.

106
Function inspection

Air/water feeding function check

Water feeding: over 45ml/min

Air feeding: over 800ml/min

1. Feeding water inclining 45°, confirm the water jet out of the nozzle as a
straight line and cover the entire objective lens.
2. Feeding air, confirm the objective lens could be blow dry within the
required time.

Blow dry time: 10s; Total length colonoscope 15s (Air pump set to H level)
320/330 series have no blow dry time limit.

107
Function inspection

Suction function check

Cover the instrument channel


valve and auxiliary water feeding
cap while check suction function,

otherwise the air in the glove will


leak from these two ports。

1. Sealing the distal end with rubber glove.


2. Block air/water valve to inflate the glove.
3. Shut down the air pump, press the glove and confirm there is no air
leakage at the suction valve.
4. If the suction valve is leaking, replace a new valve first, if still leaking,
the valve cylinder is worn out.

108
Function inspection

Suction function check

1. Connect the suction pump to endoscope and switch on, set the
suction level, gastroscopy set to 40KPa, colonoscope set to 70KPa.
2. Aspirate the water to confirm the suction function,
aspirate amount: over 400ml/min

109
Function inspection

Auxiliary water feeding function check

1. Connect the auxiliary water feeding connector tube, use syringe to feed water
2. The water should eject out from the auxiliary water feeding outlet with straight
line status。
3. If the auxiliary water feeding tube is blocked, Use converted cell brush to clean
and unblock the tube.

Note: 320/330 series no auxiliary water feeding function


110
Function inspection

Remote buttons function check

1. Check the remote buttons, both the hand feeling and function.

111
Function inspection
Processor and light source function check

500 system 550 system

1. Check the front panel every indicators.


2. Check the front panel every buttons
3. Check the status of the xenon lamp, confirm if it is close to life span.

Note: Xenon lamp life span is 500h, suggest hospital to replace


if over life span. LED lamp: 10000h
112
End the inspection

1. Shut down the light source, image processor and monitor。


2. Disconnect suction pump, water bottle, auxiliary water feeding
connector tube and endoscope cable。
3. Cover the waterproof cap, disconnect the scope from light source。
4. Take out all valves, place the scope upside down to pour-out the water
from the control section.
5. Connect the injection tube and channel plug, feed air with syringe to
drain the remaining water.
6. Wipe dry the outer surface of the scope, hand over to nurse for CDS。
7. Fill in the spot-inspection check list then give it to hospital.

113
Example of Faulty Endoscopes
Image
Faulty Image

Foggy No image

White screen Mosaic screen


11
深圳市开立科技有限公司PPT模板
Example of Faulty Endoscopes
Light Quality
Light Guide from Distal End Light Guide from Connector Side

If light quality is decreased, it could be due to the following reasons :


- The optical fiber has some problem
- The endoscope is not connected well on the light source
- The xenon lamp is faulty
- The light control become manual

Bending Section Remote Button


Example of Faulty Bending Section Example of Faulty Remote Button

11
深圳市开立科技有限公司PPT模板
Example of Faulty Endoscopes
Angulation
Angulation is failed if:
- Free Play (Bending Section not responsive to Angulation knob)
- Angulation angle do not meet the specification angle

➢ Reason: Bending wire degrades after usage for a long period

Water Feeding
Water Feeding Failure will cause objective lens cannot be clean properly
Point to check:
- Water bottle connection
- Air pump is on low mode
- Air/water valve broken
- Endoscope didn't connect well to the light source
- Existence of foreign material in the air/water tube

Air Feeding
Air Feeding Failure will cause endoscope unable to blow air properly
Point to check:
- Air pump is on the low mode
- Air/water valve is broken
- Endoscope didn't connect well to the light source
- Existence of foreign material in the air/water tube
11
深圳市开立科技有限公司PPT模板
Example of Faulty Endoscopes
Suction
Suction Failure will cause endoscope’s aspirated amount decrease
Point to check:
- Suction valve faulty
- Suction port didn't connect well to the suction pump
- Blockage of Instrument channel or Suction channel
- The biopsy valve didn't cap on the instrument channel
- Suction pump faulty

11
深圳市开立科技有限公司PPT模板
Thank you

118