Content
2
Endoscope Structure & Theory
◆Endoscope Structure
◆Endoscope Working Theory
◆550 Series Endoscope Difference
◆Endoscope model comparison
3
Endoscopy System Composition
4
Endoscope Structure
Endoscope
1 2 3 4
5
Endoscope Structure
Connector Section
6
Endoscope Structure
Connector Section
No. Part Name Description
Auxiliary water-feeding Used for connecting to water pump or syringe to make water
3
port or solution jetted from the distal end
Electrosurgical equipment
6 Used for connecting the electrosurgical equipment
connector
7
Endoscope Structure
Control Section & Insertion Section
8
Endoscope Structure
Control Section & Insertion Section
9
Endoscope Structure
Control Section & Insertion Section
No. Part Name Description
The functions of these buttons can be set through the
12 Remote buttons (0-3) image processor used with the endoscope. For details,
refer to the user manual of relevant image processor.
This channel should be used with the biopsy valve and
the functions are as follows.
● Used to feed the liquid to the distal end of the
13 Instrument channel
endoscope.
● Used for the endo-therapy accessory.
● Used as a suction channel.
Indicates the maximum length of the insertion section
14 Insertion limit mark
that can be inserted into the body
15 Suction valve Press it to extract the liquid, debris or gas from the body
● Cover the hole on the valve with finger to feed air and
press the valve to feed water.
16 Air/water valve
● The air and water can remove the blood, debris or
mucous membrane adhering to the objective lens.
10
Endoscope Structure
Control Section
Knobs
Gear, Chain
Bracket
Buttons
Valves
The gear, chain and steel wire are controlled by knobs, then
drive the snake bone of bending section to realize bending.
11
Endoscope Structure
Insertion tube inner structure
1.4 电子内窥镜结构——插入部
Spring coil
Angle steel wire Metal helix tube Metal knitmesh Synthetic resin tube
Light fiber
Instrument tube
Light fiber
CMOS cable
Air/Water tube
12
Endoscope Structure
Bending section
Snake bone
Metal knitmesh
Bending rubber
14
Endoscope Structure
Distal end
15
Endoscope Structure
Distal
1.4end
电子内窥镜结构——头端
16
Endoscope system block diagram
Connector Control
Endoscope Distal end
section section
17
Endoscope Structure & Theory
◆Endoscope Structure
◆Endoscope Working Theory
◆550 Series Endoscope Difference
◆Endoscope model comparison
18
Endoscope Theory
Air/water feeding & suction working flow
19
Endoscope Theory
Air/water feeding working flow
Air
20
Endoscope Theory
Air/water feeding working flow
21
Endoscope Theory
Air/water feeding working flow
P
Water
A/W
three-way
22
Endoscope Theory
Air/water feeding working flow
1.4 电子内窥镜结构——操作部水气阀 Blue mark
Base
Rubber bowl
Air hole
Sealing ring
Distal
Connector
end
side
side
Air
Air
Water
Water
23
Endoscope Theory
Air/water feeding working flow
1.4 电子内窥镜结构——操作部水气阀 Blue mark
Base
Rubber bowl
Air hole
Sealing ring
Distal
Connector
end
side
side
Air
Air
Water
Water
24
Endoscope Theory
Air/water feeding working flow Blue mark
Block
Base
Rubber bowl
Air hole
Sealing ring
Distal
Connector
end
side
side
Air
Air
Water
Water
25
Endoscope Theory
Air/water feeding working flow
1.4 电子内窥镜结构——操作部水气阀 Blue mark
Push Base
Rubber bowl
Air hole
Sealing ring
Distal
Connector
end
side
side
Air
Air
Water
Water
26
Endoscope Theory
Suction working flow
Connector
section
Suction
valve
Air
Instrument
channel
Suction
pump
Suction tube
Instrument tube
Instrument channel outlet
27
Endoscope Theory
Suction working flow
Connector
section
Suction
valve
Instrument
channel
Suction
pump
Suction tube
Instrument tube
Instrument channel outlet
28
Endoscope Theory
Suction working flow
Push
Air
Distal end
instrument
outlet
29
Endoscope Theory
Suction working flow
1.4 电子内窥镜结构——操作部吸引阀
Red mark
Groove
Base Hole
30
Endoscope Theory
Biopsy valve
slit
Biopsy valve
Valve cap
Hole
Base
Instrument
channel port
31
Endoscope Theory
Light transmission route Light
Objective
image
Light
Distal end
32
Endoscope Theory
Image signal flow
Image signal
Image processor
Objective
image
Light
Distal end
33
Endoscope Structure & Theory
◆Endoscope Structure
◆Endoscope Working Theory
◆550 Series Endoscope Difference
◆Endoscope model comparison
34
550 Endoscope Difference
Connector section
Electrical contacts
1. Cancel the ODU head and endoscope cable, waterproof design electric pin
2. Cancel the waterproof cap, the leakage detection port is sealed.
35
ODU Head of Connector Section of 500
36
550 Endoscope Difference
Connector section
No. Part Name Description
Image transmission
1 Used for transmitting image signals.
optical fiber connector
2 Electrical contacts Used for passing electrical signals.
High frequency Used for connecting the high frequency electrosurgical
3
feedback earthing rod generator
Used for connecting to the water bottle to feed water/air to
4 Air/water feeding ports
the distal end.
37
550 Endoscope Difference
Control section
550 500
The control section of 550 is slimmer than 500, more easily for handling
38
550 Endoscope Difference
Control section
550 500
Remote button layout is different, 550 is more easily for handling.
39
550 Endoscope block diagram
Auxiliary Auxiliary Auxiliary Water
Auxiliary water feeding tube
water pump water port water outlet
Suction tube
Suction Suction Instrument tube Instrument
Suction port
pump valve outlet
Grounding
Endotherm
knife Instrument
channel
Remote
Cable
buttons
Light
Light guide
Light source Light fiber
lens
Image
processor
In 550 endoscope, convert the electrical signal to optical signal could
enhance the capacity of anti-interference.
40
550 Endoscope Difference
Image signal flow DVI cable
Monitor
Image optical signal
Electrical signal
Objective
Image signal image
Light
Endoscope Distal end
41
Endoscope Structure & Theory
◆Endoscope Structure
◆Endoscope Working Theory
◆550 Series Endoscope Difference
◆Endoscope model comparison
42
Endoscope model comparison
Instrument Bending
Work Max insertion
Series Type Model channel section
length section diameter
diameter diameter
EG-550 2.8 1050 9.3 10.5
Gastroscope
EG-550L 3.2 1050 9.8 11.5
EC-550 3.8 1350 12.5 13.8
550
EC-550L 4.2 1350 12.9 14.5
Colonoscope
EC-550T 3.8 1700 12.5 13.8
EC-550L/T 4.2 1700 12.9 14.5
EG-500 2.8 1050 9.3 10.5
Gastroscope
EG-500L 3.2 1050 9.8 11.5
EC-500 3.8 1350 12.5 13.8
500
EC-500L 4.2 1350 12.9 14.5
Colonoscope
EC-500T 3.8 1700 12.5 13.8
EC-500L/T 4.2 1700 12.9 14.5
43
Endoscope model comparison
Instrument Bending
Work Max insertion
Series Type Model channel section
length section diameter
diameter diameter
Gastroscope EG-330 2.8 1050 9.3 10.0
The model label of 320 and 330 scopes are same, both mark 330
on the label, distinguish from color ring.
320 Ring:Green
330 Ring:Purple
44
Endoscope model comparison
Image
Auxiliary Waterproof Imaging Pre emphasis A/W
Series Ring color transmission
water cap sensor board three-way
optical fiber
45
Content
46
Spaulding Classification
47
Basic Washing Accessories
48
Basic Washing Accessories
Channel plug
Non-return valve
Injection tube
49
Basic Washing Accessories
50
CDS procedure
Exam room 1 Pre-Cleaning
Contaminated Area
Enzyme
3 Manual Cleaning
solution
51
1. Pre-Cleaning
Step1 – Clean insertion tube
Do pre-cleaning as soon
as finishing the exam.
Wipe the insertion tube with a detergent dampened lint-free cloth after
withdrawing the endoscope from the patient's body.
52
1. Pre-Cleaning
Step2 – Flush suction channel
Enzyme solution
Suck water and air alternately, until the water and air could be seen
go into the suction bottle. Suck detergent for 30s, suck air for 10s,
repeat at least 3 times
53
1. Pre-Cleaning
Step3 – Flush air/water channel and nozzle
Press Block
54
1. Pre-Cleaning
Step4 – Flush Auxiliary water feeding channel
55
1. Pre-Cleaning
Step5 – Shut down the light source and processor
Shut down
56
1. Pre-Cleaning
Step6 – Take down the accessories
57
1. Pre-Cleaning
Step7 – Unplug all connectors and tubes
Remove suction
tube.
Unplug endoscope cable
Unplug suction pump tube
Unplug water bottle connector
58
1. Pre-Cleaning
Step8 – Put on waterproof cap
59
1. Pre-Cleaning
Caution!
The light guide will be very hot after using, do not touch
as soon as unplugging from light source.
60
2. Leakage Test
The area between the blue border and the green tube is
the air inflating space while doing leakage test.
61
2. Leakage Test
Dry test
Put the endoscope on the table, connect leakage detector and add
pressure to 22Kpa(180mmHg);
Maintain the air pressure for 3 mins, if the pointer does not decrease
continually, the test is pass.
62
2. Leakage Test
Wet test
Immerse the scope which already add pressure into the water for 3 mins;
Turn the U/D and L/R knobs to adjust the angle of bending section;
If there are bubbles coming out, the leakage test is fail.
63
2. Leakage Test
Wet test
Take out the scope from the water, open the vent valve of the detector to release the
pressure until the pointer goes back to zero;
Disconnect the leakage detector from the waterproof cap.
64
2. Leakage Test
Caution!
Do dry test before wet test, make sure there are no serious leakage,
otherwise the water will flow into the scope to cause further damage;
If the wet test is failed, stop doing CDS, otherwise it will cause further
damage;
65
3. Manual Cleaning
Caution:
Do not make too
much foam of the
detergent solution
66
3. Manual Cleaning
3. Brush the suction channel from suction valve cylinder to distal end.
4. Repeat several times until there are no residue on the brush head
67
3. Manual Cleaning
5. Brush the suction channel from the suction valve cylinder to the suction port;
6. Brush the instrument channel from the instrument channel port to the distal end.
7. Repeat several times until there are no residue on the brush head
68
3. Manual Cleaning
8. Brush the suction port, instrument channel port, air water valve cylinder
and suction valve cylinder.
9. Brush the air/water valve, suction valve and biopsy valve.
10. Connect the channel plug and injection tube to the endoscope, inject
detergent solution into the suction and A/W channel until no bubbles
comes out of the distal end.
70
3. Manual Cleaning
11. Disconnect the channel plug and injection tube, immerse the endoscope
in the detergent solution.
12. Wipe the outer surface and all the ports with a lint-free cloth
71
4. Rinsing
72
4. Rinsing
73
4. Rinsing
74
5. Disinfection
1. Completely immerse the endoscope and accessories into the disinfectant solution.
2. Connect channel plug and injection tube, inject disinfectant solution into the A/W
and suction channel until no bubbles comes out of the distal end.
75
5. Disinfection
76
5. Disinfection
Recommended Disinfectant
High-Level
Contact Contact
Disinfectant Disinfection Precautions
Period Type
Parameter
Active chlorine
level: (60±10)
1. When organic matters exist, the disinfection effect decreases
mg/L, pH
sharply.The endoscope should be
Acidic value: 2.0-3.0,
cleaned completely before the disinfection.
electrolyzed chlorination
≥ 10min Immersion 2. This disinfectant has certain corrosion on the endoscope.
oxidizing water reduction
3. Flowing immersion method should be adopted for disinfection.
(AEOW) potential:
4. The endoscope should be rinsed with sterile or filtered water for
1100mV,
30s after disinfection.
Residual chloride
ion level:<1000mg/L
77
5. Disinfection
5. Take the endoscope and accessories out of the disinfectant solution, connect the
channel plug and injection tube.
6. Inject air to A/W and suction channels to drain the disinfectant solution,then
disconnect the channel plug and injection tube.
78
6. Terminal Rinsing
79
6. Terminal Rinsing
80
7. Drying
81
7. Drying
82
8. Storage
1. Free the angulation locks, adjust the bending section to straight status.
2. Hang the endoscope in the endoscope cabinet and ensure the insertion section is
vertically hanging.
3. Place the accessories dispersedly, ensure each accessories no touching.
83
Content
84
Spot inspection significance
85
Spot inspection process
Start
End
86
Spot Inspection
◆Leakage test
◆Appearance inspection
◆Function inspection
87
Leakage test
The basic step is same with CDS, below extra operations are needed while doing wet test for spot
inspection.
1. Lift knobs lightly, confirm if the knobs is leaking or not, especially the sealing ring
2. Swing the insertion tube and universal cable segment by segment, especially check the bite mark.
3. Press remote buttons from different directions, confirm the buttons no leaking.
4. Use blowing dust ball to feed water into A/W, suction, auxiliary water channels, confirm there are
no bubbles come out of the channel.
88
Leakage test
If there is leakage problem, use waterproof tape to seal the leak point
temporarily, then find out other leak points.
89
Leakage test
Ensure instrument channel is unblocked
90
Spot Inspection
◆Leakage test
◆Appearance inspection
◆Function inspection
91
Distal end
Objective lens:
Nozzle: loose, sharp, block
scratch, crack, glue breakage, fog
92
Bending rubber
93
Insertion tube
Insertion tube:
dents, bulge, scratches, breakage, deformation,
Foreign body attached, protein coagulation,
hydrophilic coating peeling.
94
Control section
Knobs:
scratches, deformation, slack
95
Universal cable
Universal cable:
dents, bulge, scratches, breakage, deformation,
Foreign body attached, protein coagulation,
hydrophilic coating peeling.
96
Connector section
97
Spot Inspection
◆Leakage test
◆Appearance inspection
◆Function inspection
98
Function inspection
Bending angle
Measure the max angle, confirm the knobs tightness, ensure could locked tightly.
99
Function inspection
Let the distal end face to normal lamp, disassemble the light guide
cap, use magnifying glass to inspect how much light guide fibers are
broken at the light guide end. It needs to be repair if the fiber silks
are broken over 30%.
100
Function inspection
Instrument channel
101
Function inspection
Image check-not turn on lamp
1. Connect the scope to processor and light source, switch on the processor.
2. Check the image in white balance cap, confirm there is no color spot, noise and
other abnormal image.
102
Function inspection
Image check-not turn on lamp
103
Function inspection
Image check-turn on lamp
1. Use alcohol wet cotton swab to clean the objective lens surface
2. Do white balance then confirm the image color and no dark corner.
3. Cover the objective lens with finger and move the finger on the lens ,
confirm there is no any not moving shadows on the image.
4. Clean the objective lens with cotton swab again.
104
Function inspection
Light control cable mark Move close to distal end Move away from distal end
105
Function inspection
1. Install the air/water valve, suction valve, connect the suction pump and water bottle.
2. Switch on the air pump, immerse the distal end into the water, there should not be
bubbles come out in natural status.
3. Feeding air in water, there should be amounts of air bubbles come out of the water
4. Feeding water in water, after the air in the tube been exhausted, confirm there is no
bubble comes out of the water.
106
Function inspection
1. Feeding water inclining 45°, confirm the water jet out of the nozzle as a
straight line and cover the entire objective lens.
2. Feeding air, confirm the objective lens could be blow dry within the
required time.
Blow dry time: 10s; Total length colonoscope 15s (Air pump set to H level)
320/330 series have no blow dry time limit.
107
Function inspection
108
Function inspection
1. Connect the suction pump to endoscope and switch on, set the
suction level, gastroscopy set to 40KPa, colonoscope set to 70KPa.
2. Aspirate the water to confirm the suction function,
aspirate amount: over 400ml/min
109
Function inspection
1. Connect the auxiliary water feeding connector tube, use syringe to feed water
2. The water should eject out from the auxiliary water feeding outlet with straight
line status。
3. If the auxiliary water feeding tube is blocked, Use converted cell brush to clean
and unblock the tube.
1. Check the remote buttons, both the hand feeling and function.
111
Function inspection
Processor and light source function check
113
Example of Faulty Endoscopes
Image
Faulty Image
Foggy No image
11
深圳市开立科技有限公司PPT模板
Example of Faulty Endoscopes
Angulation
Angulation is failed if:
- Free Play (Bending Section not responsive to Angulation knob)
- Angulation angle do not meet the specification angle
Water Feeding
Water Feeding Failure will cause objective lens cannot be clean properly
Point to check:
- Water bottle connection
- Air pump is on low mode
- Air/water valve broken
- Endoscope didn't connect well to the light source
- Existence of foreign material in the air/water tube
Air Feeding
Air Feeding Failure will cause endoscope unable to blow air properly
Point to check:
- Air pump is on the low mode
- Air/water valve is broken
- Endoscope didn't connect well to the light source
- Existence of foreign material in the air/water tube
11
深圳市开立科技有限公司PPT模板
Example of Faulty Endoscopes
Suction
Suction Failure will cause endoscope’s aspirated amount decrease
Point to check:
- Suction valve faulty
- Suction port didn't connect well to the suction pump
- Blockage of Instrument channel or Suction channel
- The biopsy valve didn't cap on the instrument channel
- Suction pump faulty
11
深圳市开立科技有限公司PPT模板
Thank you