OMZ-21009-A Series Technical News Summary V2.0 EN
OMZ-21009-A Series Technical News Summary V2.0 EN
Index
TMZ-22005-WATO EX-55&65&55Pro&65Pro&75&85 ......................................................................................... 1
Release V80 Desflurane Vaporizer .................................................................................................................... 1
TMZ-21062- AM all series ........................................................................................................................................ 4
Temporary change of the flow sensor in the anesthesia machine ...................................................................... 4
TMZ-21059-WATO EX-55&65 Pro&A .................................................................................................................... 7
WATO EX-55Pro(2526)&65 Pro(2525)&A software upgrade .......................................................................... 7
TMZ-21045-WATO&A series ................................................................................................................................... 9
Wato &A service manual update........................................................................................................................ 9
TMZ-21056-AM all series ....................................................................................................................................... 12
Battery upgrade................................................................................................................................................ 12
TMZ-21051-WATO&A series(1.0) .......................................................................................................................... 14
Wato &A material change ................................................................................................................................ 14
TMZ-21038-WATO&A series ................................................................................................................................. 16
Wato &A material change ................................................................................................................................ 16
TMZ-21016-WATO&A series ................................................................................................................................. 18
Operator’s manual and software upgrade ........................................................................................................ 18
TMZ-20012-WATO&A ........................................................................................................................................... 23
New version of manual released ...................................................................................................................... 23
TMZ-21009- Aseries&WATOseries ........................................................................................................................ 26
Aseries&WATOseries released top shelf mounting bracket ............................................................................ 26
TMZ-21006 -All AM&Ventilator ............................................................................................................................ 28
Anesthesia Machine&Ventilator Parts List updated ........................................................................................ 28
TMZ-21001-A5&A7 ............................................................................................................................................... 29
A series’ monitor board optimized ................................................................................................................... 29
TMZ-20054 -All AM&Ventilator ............................................................................................................................ 33
Anesthesia Machine&Ventilator Parts List updated ........................................................................................ 33
TMZ-20053-A7&WATO 55&65 Pro....................................................................................................................... 34
A series and WATO series released new vacuum generator unit ..................................................................... 34
TMZ-20046-A7 ....................................................................................................................................................... 37
A7(Japan) system software V02.23.01.00 released ......................................................................................... 37
TMZ-20045-A5&A7 ............................................................................................................................................... 39
A5&A7 updated Dutch, Italian, Russian, Spanish Operator's Manual ............................................................ 39
TMZ-20039-A7 ....................................................................................................................................................... 41
A7&V60 released in Japan .............................................................................................................................. 41
TMZ-20035-Anesthesia machine&Ventilator ......................................................................................................... 45
Anesthesia machine&Ventilator package label added UDI infromation, module exhaust tube changed ........ 45
TMZ-20034-A series ............................................................................................................................................... 48
A series released new software V02.22.01.00 ................................................................................................. 48
TMZ-20026-A7 ....................................................................................................................................................... 50
Changes for Japan order 21377912.10 ............................................................................................................ 50
TMZ-20022-A5 ....................................................................................................................................................... 53
A5s released in India ....................................................................................................................................... 53
TMZ-20021-A series&WATO series ....................................................................................................................... 59
Technical News
Subject: Release V80 Desflurane Vaporizer
TMZ-22005-WATO EX-55&65&55Pro&65Pro&75&85
RECORD IDENTIFIER
CONTENT
The following part numbers are material only for warranty maintenance.
Item Part number Part name Power cord type
No.: OMZ-21009-A Series Page 1of 428
PMLS Global Technical Support Dept
3、 Update the label on the package and the main unit in Indonesia. The details are
as follows:
Package label:
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Technical News
Subject: Temporary change of the flow sensor in the anesthesia machine
RECORD IDENTIFIER
CONTENT
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Technical News
Subject: WATO EX-55Pro(2526)&65 Pro(2525)&A software upgrade
RECORD IDENTIFIER
CONTENT
3 A3_software(DSP)_V02.24.00.02 02.24.00.02
4 A5_software(DSP)_V02.24.00.02 02.24.00.02
5 A7_software(DSP)_V02.24.00.02 02.24.00.02
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Technical News
Subject: Wato &A service manual update
TMZ-21045-WATO&A series
RECORD IDENTIFIER
CONTENT
1. The service manual is upgraded to optimize the test items after installation. The test
items that have been approved by the manufacturer and do not affect the
performance during transportation are removed from the installation test list. In the
last part of the service manual, test items after installation are optimized for WATO
EX-35&A7&A5, and a test item list after installation and maintenance is added for
WATO EX-55/65/55Pro/65Pro. Take A7 as an example. The test items after
installation are as follows:
If you need download information or have any question, please feel free to contact Mindray regional
service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Technical News
Subject: Battery upgrade
Battery upgrade
RECORD IDENTIFIER
CONTENT
If you need download information or have any question, please feel free to contact Mindray regional
service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Technical News
Subject: Wato &A material change
TMZ-21051-WATO&A series(1.0)
RECORD IDENTIFIER
CONTENT
FAN 12V
Fan&Wire(MF40
40*40*20mm Compati
1 024-000407-00 009-013196-00 201V3-1000C-
6.3CFM 18dB ble
G99)
typeF
If you need download information or have any question, please feel free to contact Mindray regional
service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Technical News
Subject: Wato &A material change
TMZ-21038-WATO&A series
RECORD IDENTIFIER
CONTENT
1. The battery part number is changed due to the battery supplier is changed. The battery has
changed from 022-000008-00 to 022-000382-00 when the system is delivered(except Kore、
Japan、India、Russia、Saudi Arabia).The battery order number is changed in CE area (except
Kore、Japan、India、Russia、Saudi Arabia、Brazil、Peru、Indonesia、Malaysia、Morocco). The
following table shows the substitution relationship of part number.
If you need download information or have any question, please feel free to contact Mindray regional
service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
TMZ-21016-WATO&A series
Technical News
Subject: Operator’s manual and software upgrade
RECORD IDENTIFIER
CONTENT
b) In AG module the auto zeroing will stop when the following alarm trigger:Mixed Agent, Mixed
Agent and MAC ≥ 3;in CO2 module the auto zeroing will stop when the following message
alarms:EtCO2 Too High,EtCO2 Too Low,FiCO2 Too High,EtO2 Too High,EtO2 Too Low,FiO2
Too High,FiO2 Too Low.
c) Add “Null for 30s from AG Zeroing”button in AG module maintenance interface, when its “on”,
Fi/EtO2, Fi/Et CO2, Fi/EtN2O, Fi/EtAA, MAC will display 30 seconds“---”.
d) Add “Null for 30s from CO2 Zeroing”button in CO2 module maintenance interface, when its
“on”, Fi/EtO2, Fi/Et CO2 will display 30 seconds“---”
e) When the “Apnea” alarms, auto zeroing will stop for 5minutes, and the maximum prohibit time
of zeroing is less then 20minutes.
The followings are software name and version,if youneed please contact ITS get the software.
Item Software name Version
6 A3_Software(DSP)_V02.23.00.05_EN 02.23.00.05
7 A5_Software(DSP)_V02.23.00.05_EN 02.23.00.05
8 A7_Software(DSP)_V02.23.00.05_EN 02.23.00.05
:/Anesthesia
WATO EX-30 Operator's
Machine/Ventilator/Anesthesia
9 Manual_EN(Sri 2.0 3.0
Machine/Wato-
Lanka )_V3.0.pdf
30(0617)/Manual/Operation Manual
WATO EX-35 Operator's :/Anesthesia
Manual(With Machine/Ventilator/Anesthesia
10 2.0 3.0
Compressor)_V3.0_EN. Machine/Wato-
pdf 35(0619)/Manual/Operation Manual
:/Anesthesia
WATO EX-65 Pro
Machine/Ventilator/Anesthesia
11 Operator's 2.0 3.0
Machine/Wato-65
Manual_V3.0_EN.pdf
Pro(0627)/Manual/Operation Manual
:/Anesthesia
WATO EX-55 Pro
Machine/Ventilator/Anesthesia
12 Operator's 2.0 3.0
Machine/Wato-55
Manual_V3.0_EN.pdf
Pro(0628)/Manual/Operation Manual
WATO EX-65 Pro :/Anesthesia
Operator's Manual(With Machine/Ventilator/Anesthesia
13 2.0 3.0
Compressor)_V3.0_EN. Machine/Wato-65
pdf Pro(0627)/Manual/Operation Manual
WATO EX-55 Pro :/Anesthesia
Operator's Manual(With Machine/Ventilator/Anesthesia
14 2.0 3.0
Compressor)_V3.0_EN. Machine/Wato-55
pdf Pro(0628)/Manual/Operation Manual
:/Anesthesia
WATO EX-65 Operator's Machine/Ventilator/Anesthesia
15 2.0 3.0
Manual_V3.0_EN.pdf Machine/Wato-65(0625)/Operation
Manual
:/Anesthesia
WATO EX-55 Operator's Machine/Ventilator/Anesthesia
16 2.0 3.0
Manual_V3.0_EN.pdf Machine/Wato-
55(0626)/Manual/Operation Manual
WATO EX-65 Operator's :/Anesthesia
Manual(With Machine/Ventilator/Anesthesia
17 2.0 3.0
Compressor)_V3.0_EN. Machine/Wato-65(0625)/Operation
pdf Manual
WATO EX-55 Operator's :/Anesthesia
Manual(With Machine/Ventilator/Anesthesia
18 2.0 3.0
Compressor)_V3.0_EN. Machine/Wato-
pdf 55(0626)/Manual/Operation Manual
If you need download information or have any question, please feel free to contact Mindray regional
Sincerely yours,
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
TMZ-20012-WATO&A
Technical News
Subject: New version of manual released
RECORD IDENTIFIER
CONTENT
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-21009- Aseries&WATOseries
Technical News
Subject: Aseries&WATOseries released top shelf mounting bracket
RECORD IDENTIFIER
CONTENT
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Technical News
Subject: Anesthesia Machine&Ventilator Parts List updated
RECORD IDENTIFIER
CONTENT
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
TMZ-21001-A5&A7
Technical News
Subject: A series’ monitor board optimized
RECORD IDENTIFIER
CONTENT
1. A series’ monitor board optimized ,boards after change will use new type flow and
pressure sensors for better stability.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Technical News
Subject: Anesthesia Machine&Ventilator Parts List updated
RECORD IDENTIFIER
CONTENT
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Technical News
Subject: A series and WATO series released new vacuum generator unit
RECORD IDENTIFIER
CONTENT
1. A series and WATO series released new vacuum generator unit,new part includes
original vacuum generator unit and muffler assembly.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-20046-A7
Technical News
Subject: A7(Japan) system software V02.23.01.00 released
RECORD IDENTIFIER
CONTENT
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
No.: OMZ-21009-A Series Page 37of 428
PMLS Global Technical Support Dept
TMZ-20045-A5&A7
Technical News
Subject: A5&A7 updated Dutch, Italian, Russian, Spanish Operator's Manual
RECORD IDENTIFIER
CONTENT
Version Version
Operator's Manual Remark
before change after change
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-20039-A7
Technical News
Subject: A7&V60 released in Japan
RECORD IDENTIFIER
CONTENT
1. New labels with JAN code are added on main units and package
Fig4 添付文书
Influences on Pre-sales & Installation
Paper version operation manual and “添付文书”will be put in first drawer(A7) or
package(V60)
Influences on Upgrade
Contact with Mindray engineer to get PDF version
Influences on Maintenance
NA
4. A7(Japan) system software V02.22.01.00 released.
No.: OMZ-21009-A Series Page 43of 428
PMLS Global Technical Support Dept
Influences on Pre-sales & Installation
NA
Influences on Upgrade
Contact with Mindray engineer to confirm before upgrade.
Influences on Maintenance
Contact with corresponding GTS for software if necessary.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-20035-Anesthesia machine&Ventilator
Technical News
Subject: Anesthesia machine&Ventilator package label added UDI infromation, module
exhaust tube changed
RECORD IDENTIFIER
CONTENT
Influences on Upgrade
NA
Influences on Maintenance
NA
2. Module exhaust tube extends longth to 50cm from 37cm to meet customer side
requirement, PN remains the same.
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
NA
If you need download information or have any question, please feel free to contact
No.: OMZ-21009-A Series Page 46of 428
PMLS Global Technical Support Dept
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Thank you for your kind attention and cooperation.
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-20034-A series
Technical News
Subject: A series released new software V02.22.01.00
RECORD IDENTIFIER
CONTENT
Fig1
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-20026-A7
Technical News
Subject: Changes for Japan order 21377912.10
RECORD IDENTIFIER
CONTENT
Fig1
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
NA
Fig2
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
NA
Fig3
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
NA
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-20022-A5
Technical News
Subject: A5s released in India
RECORD IDENTIFIER
CONTENT
Accessories
1. A5s cancelled A5 standard config passive agss tube
Fig1
Influences on Pre-sales & Installation
Please refer to device attached packing list during installation.
Fig2
Influences on Pre-sales & Installation
Please refer to device attached packing list during installation.
Influences on Upgrade
A5s could upgrade O2 sensor, it uses the A5 O2 sensor. Need to purchase O2 sensor
cable and O2 sensor cover at the same time.
Influences on Maintenance
If O2 sensor is not configured, after system software upgrading A5s might have O2 sensor
related alarm, enter Main→System→Code 1234→Other settings→change O2 cell
monitoring to “Off”
Hardware
1. A5s cancelled touch pad
Fig3
Influences on Pre-sales & Installation
NA
Influences on Upgrade
No.: OMZ-21009-A Series Page 54of 428
PMLS Global Technical Support Dept
Touch pad could not be upgraded.
Influences on Maintenance
NA
2. A5s cancelled Auxiliary AC Outlets
Fig4
Influences on Pre-sales & Installation
NA
Influences on Upgrade
Auxiliary sockets could not be upgraded.
Influences on Maintenance
If hardware box rear cover broken please contact with corresponding Mindray engineer.
3. A5s cancelled Auxiliary O2/Air Flowmeters and Auxiliary O2 gas Power Outlet
Fig5
Influences on Pre-sales & Installation
NA
Influences on Upgrade
Auxiliary O2/Air Flowmeters and Auxiliary O2 gas Power Outlet could not be upgraded.
No.: OMZ-21009-A Series Page 55of 428
PMLS Global Technical Support Dept
Influences on Maintenance
NA
4. A5s cancelled module rank
Fig6
Influences on Pre-sales & Installation
NA
Influences on Upgrade
Module rank could not be upgraded. Module is unavailable on A5s.
Influences on Maintenance
NA
Fig7
Influences on Pre-sales & Installation
NA
No.: OMZ-21009-A Series Page 56of 428
PMLS Global Technical Support Dept
Influences on Upgrade
Air cylinder bracket could not be upgraded.
Influences on Maintenance
NA
Fig8
Influences on Pre-sales & Installation
NA
Influences on Upgrade
Center breaker could not be upgraded.
Influences on Maintenance
New wheel is added for A5s
Software
1. A5s cancelled A5 standard config SIMV mode, PS mode, Loop function
Influences on Pre-sales & Installation
Please refer to device PI during installation.
Influences on Upgrade
A5s could upgrade SIMV mode ,PS mode, Loop function, PS&Loop have upgrade
kit,SIMV mode upgrade package is added newly.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
A series&WATO series released new monitor brackets,A series have new manual
Technical News
Subject: A series&WATO series released new monitor brackets,A series have new manual
RECORD IDENTIFIER
CONTENT
Fig1
Influences on Pre-sales & Installation
File Anesthesia System Bracket Installation Instructions is attached with bracket package
as reference.
Influences on Upgrade
WATO EX-55[0622],WATO EX-65[0621],WATO EX-55[0624],WATO EX-65[0623],
WATO EX-35[0618] also could use above bracket to upgrade.
Influences on Maintenance
NA
2. A5 and A7 released new English operation manual V20.0,manual added new monitor
bracket details.
Folder address on TDP(http://intlservice.mindray.com/logon/index.action) as below
View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/
A5 (0631&0634)/Manual/Operation Manual
View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/
A7(0632&0635)/Manual/Operation Manual
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
No.: OMZ-21009-A Series Page 60of 428
PMLS Global Technical Support Dept
NA
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-20006-A5&A7
Technical News
Subject: A5&A7 released in Russia
RECORD IDENTIFIER
CONTENT
Fig1
Influences on Upgrade
NA
Influences on Maintenance
A5&A7 maintenance (include preventive maintenance) needs A Series Service Certificate
from Mindray, the certificate can be got by attending Mindray face-to-face training. Please
inquire with your Mindray service contact for training details.
PS: Branch lecturer could trigger the face-to-face training by process in Portal website
workspace-Overseas Training Application, please get help from HQ training department
Sunny (zhuyunlong@mindray.com) if any doubt.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-20004-WATO&A series
Technical News
Subject: New nylon Breathing Circuits are released for India.
RECORD IDENTIFIER
CONTENT
New nylon Breathing Circuits are released for India. When being wiped by some types of
disinfectant, nylon is expected to be more durable than PPSU.
PPSU and nylon Breathing Circuits’ differences are marked by red arrows shown below:
Influence on installation
N/A
Influence on maintenance
Not all models of WATO series are all concerned by this change. Only the models shown
below are concerned:
Model Factory Code SN
WATO EX-55 0626 KU-**
WATO EX-65 0625 KT-**
WATO EX-55 Pro 0628 F1-**
WATO EX-65 Pro 0627 KY-**
New nylon Breathing Circuits can be used entirely to replace the same type Breathing Circuit
whose material is PPSU. According to the new nylon Breathing Circuits, three FRUs and
five covers are added:
Order Number Part Description Reference
115-067595-00 Breathing system FRU(0634/PA) For A5.
115-067597-00 Breathing Circuit Cell(0635/PA) For A7.
Patient main unit(with bag
115-067598-00 For WATO EX-55,65,55 Pro,65 Pro.
arm/0625/PA)
For nylon Breathing Circuit, not for
043-006704-00 upper cover 2(PA/silk-screen)
PPSU Breathing Circuit.
Influence on upgrade
N/A
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-19036-A5&A7&WATO
A series anesthesia machine knob change and anesthesia machine electronic pressure
sensor change
Technical News
Subject: A series anesthesia machine knob change and anesthesia machine electronic
pressure sensor change
RECORD IDENTIFIER
CONTENT
Change description:
The surface appearance of the knob of A-series is darker than WATO
series, which can affect the feeling of the usage, The R&D changed the
surface appearance of the knob of A-series so that it’s consistent with the
WATO series.
Influence on installation:
NA
Influence on upgrade:
NA
Influence on maintenance:
1) Change description:
Before After
Change Change
Influence on installation:
NA
Influence on upgrade:
NA
Influence on maintenance:
a) For the machine before Jul. 2018,if the pipeline supply electronic pressure sensor
broken, you should apply the corresponding inlet module to maintenance .The
PNs change as follows:
b) For the machine after Jul. 2018, if the pipeline supply electronic pressure sensor
broken, you can apply the sensor to maintenance .The PN changes as follows:
2) Change description:
After change
Before change
Influence on installation:
NA
Influence on upgrade:
NA
If the cylinder supply electronic pressure sensor broken, you should apply the
corresponding yoke assembly to maintenance .The PNs change as follows:
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
No.: OMZ-21009-A Series Page 74of 428
PMLS Global Technical Support Dept
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-19035-A5&A7
Technical News
Subject: A series anesthesia machine software V02.22.00.01 release
RECORD IDENTIFIER
CONTENT
Influence on installation:
NA
Influence on upgrade:
NA
Influence on maintenance:
If the anesthesia machine still has “FiO2 Too High” alarm even the high FiO2 alarm
limit had been set to 100%, please upgrade the machine’s software to V02.22.00.01
or above.
If you want to upgrade software, please contact Mindray regional service engineer.
No.: OMZ-21009-A Series Page 76of 428
PMLS Global Technical Support Dept
Influence on installation:
NA
Influence on upgrade:
If the customer needs to upgrade the anesthesia calculation function of the A5&A7
anesthesia machine, please ensure that the A5&A7 anesthesia machine is equipped
with an external AG module and order activation code to activate the anesthesia
calculation function.
Influence on maintenance:
NA
Change description:
Influence on installation:
NA
Influence on upgrade:
NA
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-19028-V60
Technical News
Subject: V60 vaporizer filler cap’s optimization.
RECORD IDENTIFIER
CONTENT
For Pour Fill and Quik-Fil vaporizers, an issue happens occasionally: Poor sealing of the
filler cap due to thread stuck, and it leads to leakage. Screw threads’ positions as shown
below:
Figure 1.
In order to solve this problem, R&D optimized screw threads of filler port and filler cap.
After the optimization, it’s easy to tighten filler cap.
Influence on maintenance
N/A
Influence on upgrade
N/A
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Technical News
Subject: VT650/VT900 Gas Flow Analyzer release
RECORD IDENTIFIER
CONTENT
Influence on installation:
NA
Influence on upgrade:
NA
Influence on maintenance:
For more information, please refer to the 《 VT650&VT900 Gas Flow Analyzer
Introduction_V1.0_EN》.
You can download the 《VT650&VT900 Gas Flow Analyzer Introduction_V1.0_EN》on TDP:
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Delete the company address from Anesthesia Machine&Ventilator all products’ quick guide
and maintenance guide
Technical News
Subject: Delete the company address from Anesthesia Machine&Ventilator all products’ quick guide and
maintenance guide.
RECORD IDENTIFIER
CONTENT
Change description:
The company's business license address and production address have changed. Now we delete
the address information from all types of Anesthesia Machine&Ventilator’s quick guide and
maintenance guide, to reduce the complexity of HQ product management. New version of
quick guide and maintenance guide are available from TDP.
Impacts to service:
N/A.
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-19015-AllAM&Ventilator
Technical News
Subject: Anesthesia Machine&Ventilator Parts List is released.
RECORD IDENTIFIER
CONTENT
Model Picture(Y/N)
A3(0637) Y
A5(0634) Y
A7(0635) Y
SV600/650(VT60/VT65) Y
SV800/850(VT70/VT75) Y
SV300/350(EV20/EV25) Y
WATO EX-65(0625) Y
No.: OMZ-21009-A Series Page 85of 428
PMLS Global Technical Support Dept
WATO EX-55(0626) Y
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-19014-FPM
Flow Pressure Measurement (FPM) adds pressure sampling port and changes the
pressure sampling mode
Technical News
Subject: Flow Pressure Measurement (FPM) adds pressure sampling port and changes the
pressure sampling mode.
RECORD IDENTIFIER
CONTENT
In order to improve the pressure calibration accuracy of FPM, R&D change the pressure
sampling method.
In terms of appearance, the difference between old type FPM and new type lies in left
view, as shown in the following figure:
The latest version of Operator's Manual have been uploaded to TDP, address lists below:
/Anesthesia Machine/Ventilator/FPM/Operator’s Manual/Flow Pressure Measurement Operator's
Manual_V2.0_EN.pdf
Influence on installation
N/A
Influence on maintenance
When new type FPM is used to calibrate, it must be operated in accordance with the latest
version of Operator's Manual.
Influence on upgrade
N/A
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Technical News
Subject: Anesthesia machine ACGO assembly change
RECORD IDENTIFIER
CONTENT
Influence on installation
N/A
Influence on maintenance
If ACGO need to be replaced, please refer the list above to apply spare parts.
Influence on upgrade
N/A
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-19004-All AM&Ventilator
Technical News
Subject: Anesthesia Machine&Ventilator Parts List is released.
RECORD IDENTIFIER
CONTENT
If you have any questions & suggestions, or you find any mistakes, please feedback
to Robert (haolingguang@mindray.com).Thanks a lot for your support.
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-18036-A7
Technical News
Subject: A7 anesthesia machine hardware change
RECORD IDENTIFIER
CONTENT
⚫ Influence on upgrade
NA
⚫ Influence on maintenance:
Compatible changes,the PN didn’t change.
Change description:
To solve the BFCS stuck problem because of the round end of the solenoid pin and the
ACGO switch is not triggered properly, BFCS module has been upgrade.
⚫ Influence on installation:
NA
⚫ Influence on upgrade
NA
⚫ Influence on maintenance:
Compatible changes,the PN didn’t change.
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-18035-A5
Technical News
Subject: A5 anesthesia machine hardware change
RECORD IDENTIFIER
CONTENT
Change description:
Because M02-001152-01 and M02-001032-01 have stopped production, A5 power board
updated. Use the 051-003469-00 0625 Power Board to replace the 051-000685-00 0631
Power Board PCBA.
⚫ Influence on installation:
NA
⚫ Influence on upgrade
NA
⚫ Influence on maintenance:
If A5 power board need to be replaced. Use the same order number (801-0631-00025-
00) to apply spare part. The order number of the power board is not changed, but the
part number changed.
Change description:
⚫ Influence on upgrade
NA
⚫ Influence on maintenance:
Compatible changes,the PN didn’t change.
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Technical News
Subject: Anesthesia machine hardware upgrade and service manual update
RECORD IDENTIFIER
CONTENT
Because of old flash chips stop production, new flash chips are used for A series main control
board and flow control board. Old and new Main Control board and Flow Control board are
compatible. New software (V02.21.01.00) are released to support new Main Control board
and Flow Control board.
Influence on installation
N/A
Influence on maintenance
Old and new Main Control board and Flow Control board are compatible. If new Old and
new Main Control board and Flow Control board need to be replaced, please upgrade the
latest software.
Influence on upgrade
N/A
To solve auxiliary gas outlet gas leakage problem, new auxiliary gas outlet changed needle
valve design. Old and new auxiliary gas outlet are compatible.
No.: OMZ-21009-A Series Page 98of 428
PMLS Global Technical Support Dept
Influence on installation
N/A
Influence on maintenance
If auxiliary gas outlet has leakage problem, please apply auxiliary gas outlet assembly to
solve the problem.
Influence on upgrade
N/A
Influence on installation
N/A
Influence on maintenance
If gas outlet is much high than Vt setting, and “Tidal volume not reach” “Please check flow
sensor” alarm at the same time. Please apply one way valve to solve the problem.
Influence on upgrade
N/A
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Anesthesia machine’s Ventilation&Control PCBA coating increase and user manual update
Technical News
Subject: Anesthesia machine’s Ventilation&Control PCBA coating increase and user manual
update
RECORD IDENTIFIER
CONTENT
Description
Cleaning and disinfection cards have uploaded to TDP, address lists below:
Influence on installation
During anesthesia machine installation, engineer can use Patient breathing circuit
cleaning and disinfection instruction card to tell customer how to clean and disinfect
patient breathing circuit.
Influence on maintenance
N/A
Influence on upgrade
N/A
No.: OMZ-21009-A Series Page 102of 428
PMLS Global Technical Support Dept
Description
Pre-pak breathing circuit volume: Auto mode 2.85L, Manual mode 1.8L;
Non pre-pak breathing circuit volume: Auto mode 2.6L, Manual mode 1.8L.
Influence on installation
N/A
Influence on maintenance
N/A
Influence on upgrade
N/A
If ventilation&control PCBA’s sensor polluted, tidal volume and pressure measured value
may not accurate. To avoid ventilation&control PCBA polluted, silicon coating
ventilation&control PCBA released.
Other spare parts, which related with PCBA shown above, substitution relationship list
below:
No Old New
. Part description Part number Part description Part number
1 Monitoring board assembly 0621-30-78559 0623 Monitor board assembly 115-050887-00
2 Monitored signal detection 801-0616-00059-00 Monitored signal detection board 115-051193-00
board (FRU) (FRU)
3 0616 Monitor board assembly 115-011624-00 0616 Monitor board assembly 115-050886-00
4 Monitor Board assembly 801-0621-00006-00 Monitor Board assembly 115-051194-00
5 Monitor board assembly 801-0616-00117-00 Monitor board assembly 115-051195-00
6 Ventilator assembly (Air drive) 801-0621-00086-00 Ventilator assembly (Air drive) 115-051197-00
7 Ventilator assembly (O2 drive) 801-0621-00087-00 Ventilator assembly (O2 drive) 115-051198-00
8 ventilator(O2 drive) 115-034998-00 ventilator(O2 drive) 115-051199-00
9 ventilator(AIR drive) 115-034997-00 ventilator(AIR drive) 115-051200-00
10 0625 Monitor board assembly 115-031842-00 0625 Monitor board assembly 115-050879-00
11 0619 Monitor board assembly
115-035472-00 0619 Monitor board assembly 115-050878-00
(DSP)
12 0619 Ventilation kit FRU(O2
115-036952-00 0619 Ventilation kit FRU(O2 Drive) 115-051201-00
Drive)
13 0619 Ventilation kit FRU(Air
115-036953-00 0619 Ventilation kit FRU(Air Drive) 115-051202-00
Drive)
14 Breathing system FRU(Air drive) 115-033003-00 Breathing system FRU(Air drive) 115-051203-00
15 Breathing system FRU(O2 drive) 115-033002-00 Breathing system FRU(O2 drive) 115-051204-00
16 0635 Monitor board assembly
115-036401-00 0635 Monitor board assembly 115-050888-00
(NOVA)
17 Monitoring Board(DSP) FRU 115-036365-00 Monitoring Board(DSP) FRU 115-051205-00
18 Ventilator Control Board(DSP)
115-036366-00 Ventilator Control Board(DSP) FRU 115-051206-00
FRU
19 Ventilation kit(Non-
115-036364-00 Ventilation kit(Non-FDA/DSP)FRU 115-051207-00
FDA/DSP)FRU
20 Ventilation kit(Non-FDA/DSP/Air Ventilation kit(Non-FDA/DSP/Air
115-049798-00 115-051208-00
driven) driven)
Influence on installation
N/A
Influence on maintenance
Use new parts list above to replace old parts. New spare parts and old spare parts are
compatible.
Influence on upgrade
N/A
No.: OMZ-21009-A Series Page 105of 428
PMLS Global Technical Support Dept
Description
The only difference between old and new flowmeter is gear ratio. There is no appearance
difference.
Influence on installation
N/A
Influence on maintenance
Influence on upgrade
N/A
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
New monitor board with Protection used for India market prolong service time
Technical News
Subject: New monitor board with Protection used for India market prolong service time.
RECORD IDENTIFIER
CONTENT
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
Abraham.ma@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-17032-A5&A7
Technical News
Subject: Display change and new software is released.
RECORD IDENTIFIER
CONTENT
Hardware Change:
1, The display supplier of the A series anesthesia has changed. The parts for maintenance
changed as bellow:
0631 0631
A5,A7
051-001519- Backlight 051- Backlight No
/ before
00 Inverter 001519-00 Inverter changed
2015.12.31
PCBA PCBA
Backwards
compatible,
0635 LED 0632 Screen
051-002448- 051- new version A5,A7 after Old is
Backlight Backlight
00 002583-00 is compatible 2015.12.31 deleted
Board PCBA Board PCBA
with old
ones.
Display Display Display
115-036373- Exchange 115- Exchange Fully material
A5,A7
00 Package 036373-00 Package compatible changed,
(0634) (0634) but part
Software Change:
1,To support transmission of data to Philips Monitors with IntelliBridge module, new
software of A series anesthesia with DSP monitor board is released.
Model Bundle version Remark
Others:
1,Add 2 new spare parts for maintenance:
Tank knob
Before
Now
(c) Modify Low Flow Anesthesia from 1L/min to 0.5L/min.
Before
Now
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-17031-All AM&Ventilator
Technical News
Subject: Lithium battery discharge to 30% to meet the requirement of air transport
RECORD IDENTIFIER
CONTENT
To meet the requirement of air transport battery by airline company, ventilators and
anesthesia machine’s lithium batteries need to be discharge to no more than 30%.
Influenced ventilator and anesthesia machine models are: WATO EX-20(0615/061B),
WATO EX-20Vet(0620), WATO EX-25(0616/0629), WATO EX-30(0617), WATO EX-
35(0618/0619), WATO EX-55(0624/0626), WATO EX-55 Pro(0628), WATO EX-
65(0623/0625), WATO EX-65 Pro(0627), A3(0637), A5(0634), A7(0635), SV300(EV20),
E3(VT30), E5(VT50),
Change description:
If the order of lithium battery transported by air, lithium battery need to be discharged to
no more than 30%, and paste “Label for Cargo Aircraft Only”.
Orders for international non-air transport, batteries do not need to be discharged and do
not require the label.
Shipping method and Machine model shown on the top right corner of the label.
Sample of label shown as below:
Influenced materials:
The table below list all influenced materials
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-17030-A5&A7
Technical News
Subject: A5&A7 manual bag arm changes
RECORD IDENTIFIER
CONTENT
1, For solving the sweat marks problem in the case of long-term use, the surface coating
of the manual bag arm assembly is changed from nickel to chrome.
Before After
Difference of
appearance
2, The width of the manual bag arm assembly support tube limit block increases.
Before After
Top view Limit block Wider
gray
Difference of
appearance
Bottom view
Mid-plate
Mid-plate FRU(0631/no
115-021249-00 FRU(0631/no 115-021249-00
ventilation valve)
ventilation valve)
Breathing Circuit Breathing Circuit
115-025564-00 115-025564-00
Cell(0635) Cell(0635)
Spare
Breathing Circuit Breathing Circuit
Parts 115-025565-00 115-025565-00
Cell(0634) Cell(0634)
Breathing system Breathing system
115-025569-00 115-025569-00
FRU(0634) FRU(0634)
801-0631-00105- 801-0631-
Breathing System Breathing System
00 00105-00
Mid-plate failure
Part number is no change. All machines use old part
Maintenance
Breathing Circuit number for maintain and replacement
failure
Upgrade N/A
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
New monitor board with Protection had been released for India
Technical News
Subject: New monitor board with Protection had been released for India.
RECORD IDENTIFIER
CONTENT
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
Abraham.ma@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-17027-All AM&Ventilator
Technical News
Subject: Anesthesia Machine&Ventilator Parts List is released.
RECORD IDENTIFIER
CONTENT
If you have any questions & suggestions, or you find any mistakes, please
feedback to Robert (lingjun@mindray.com).Thanks a lot for your support.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-17021-A7
Technical News
Subject: A7 New Software released & APL valve change.
RECORD IDENTIFIER
CONTENT
◼ Hardware
1. The color of APL valve is changed from black to gray.
Before After
Difference of
Black Gray
appearance
APL valve
Maintenance All of the devices use the same part number for replacement.
failure
Upgrade N/A
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-17020-A5
Technical News
Subject: A5 New Software released & APL valve change.
RECORD IDENTIFIER
CONTENT
◼ Software
1,The new software for A5 with DPS monitor is launched.
No. Package Modules Latest Version
SystemUpdateTo
00 ol Tool 5.8
(V5.7)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)
Bundle Version 02.18.02.00
Host Software 03.18.02.00
FPGA Display 1.4
FPGA Sound 1.6
Bundle
02 Ventilator Protect
(V02.18.02.00) V01.02.00
Module
Ventilator Control
V01.04.00
Module
Language 4.6
flowmeter V1.0
03 Power(V1.7) Power System V1.7
◼ Hardware
1. The color of APL valve is changed from black to gray.
Before After
Black Gray
Difference of
appearance
Upgrade N/A
2. Increase the length of the flow sensor cable, and the ground connection.
Before After
Difference of
appearance
Electronic flow
Maintenance All of the devices use the same part number for replacement.
sensor failure
Upgrade N/A
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-17014-WATO&A series
The measuring mode of Flow and Volume for WATO and A series anestheisa machines is
changed from ATPD to BTPS
Technical News
Subject: The measuring mode of Flow and Volume for WATO and A series anestheisa
machines is changed from ATPD to BTPS.
RECORD IDENTIFIER
CONTENT
1. The measuring mode of Flow and Volume for WATO and A series anesthesia
machines is changed from ATPD to BTPS
Change Description
In order to meet product regulations requirement, the measuring mode of Flow and
Volume for WATO and A series anesthesia machines is changed from ATPD to BTPS.
Because of this change, the description about factory calibration ventilator
parameters test in service manual of WATO EX-20/30/35 need to be update at the same
time. The new manual has been uploaded to TDP:
(1) View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-20(0615)/Manual/Service
Manual
(2) View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-30(0617)/Manual /Service
Manual
(3) View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-35(0618)/Manual /Service
Manual
Influence to installation business
During installation, if want to test the ventilator parameters, please change the
measuring mode of calibration device to BTPS. Please refer to the service manual of
WATO or A series for more details about the test.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-17012-A7
Technical News
Subject: A7 New Software released.
RECORD IDENTIFIER
CONTENT
◼ Software
1,Two new software is launched:
(1)First: Software for devices with EPSON monitor board.
No. Package Modules Latest Version
BIOS
01 BIOS 01.12.00.00
(01.12.00.00)
Language 3.9
Keyboard 01.00/01.01
SystemUpdateTool
00 Tool 5.8
(V5.8)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)
Language 4.6
Keyboard 01.00/01.01
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-17011-A5
Technical News
Subject: A5 New Software released.
RECORD IDENTIFIER
CONTENT
◼ Software
1,Two new software is launched:
(1)First: Software for devices with EPSON monitor board.
No. Package Modules Latest Version
BIOS
01 BIOS 01.12.00.00
(01.12.00.00)
Language 3.9
flowmeter V1.0
NOTE:
1, New software can only be used on the new EPSON monitor board. The DSP monitoring
SystemUpdateTool
00 Tool 5.8
(V5.7)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)
Language 4.6
flowmeter V1.0
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-17010-All AM&Ventilator
Technical News
Subject: Anesthesia Machine&Ventilator Parts List is released.
RECORD IDENTIFIER
CONTENT
If you have any questions & suggestions, or you find any mistakes, please
feedback to Shiling (yangshiling@mindray.com).Thanks a lot for your support.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
Electronic gas supply pressure monitoring released for WATO EX-55 PRO/65 PRO
(0628/0627) and new bundle software released in ECL134
Technical News
Subject: Electronic gas supply pressure monitoring released for WATO EX-55 PRO/65 PRO
(0628/0627) and new bundle software released in ECL134.
RECORD IDENTIFIER
CONTENT
Release new bundle software of new WATO series and A series anesthesia machine and new
electronic pressure monitoring for WATO EX-55 PRO/65 PRO (0628/0627).
1. Hardware Change:
(1) Release electronic gas supply pressure monitoring for WATO EX-55 PRO/65 PRO
(0628/0627). The interface is as below:
(2) New WATO series anesthesia machines add AGSS 3 ways connector assembly.
Newly added spare parts:
Item
P/N Descriptions Applicable models
no.
WATO EX-35/55/65/55Pro/65Pro
1 115-026796-00 AGSS 3 ways connector assembly
(0619/0626/0625/0628/0627)
WATO EX-35/55/65/55Pro/65Pro
2 115-030332-00 AGSS kit (0634/low flow/CE)
(0619/0626/0625/0628/0627)
WATO EX-35/55/65/55Pro/65Pro
3 115-030333-00 AGSS kit (0634/high flow/CE)
(0619/0626/0625/0628/0627)
NOTES:
(1) AGSS kits add the 3 ways connector assembly.
(2) A series anesthesia machines have this connector yet.
E16090001 is the SN of the flow sensor. 134℃ shows it can support 134℃ autoclave.
2. Software change:
Released new WATO series anesthesia machine bundle software V04.05.00.02 and A series
anesthesia machine bundle software V02.18.00.04.
Software version:
Item
Descriptions Old version New version Compatibility
no.
1 New WATO series bundle software V04.04.01.00 V04.05.00.02 One-way
compatible
package
2. A series bundle software package V 02.16.00.03 V02.18.00.04 One-way
compatible
3. Manual
Item
P/N Version Descriptions
no.
1 H-046-008766-00 2.0 WATO EX-35_Operator's Manual_V2.0_Brazilian Portuguese
Portuguese
Portuguese
If you need download information or have any question, please feel free to contact Mindray regional
service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16089-A7
Technical News
Subject: ECL128 Engineering changes.
RECORD IDENTIFIER
CONTENT
◼ Hardware
1,The flow sensor of EFCS is replaced by the new type:
Before after
Same appearance, only support a single gas Support 3 kinds of gas (O2, N2O, and AIR)
Difference of
appearance
0632 Flowmeter O2
051-001414-01 flow sensor board
PCBA ---delete
0632 Flowmeter Air
Spare Flow sensor board
051-001415-01 sensor board PCBA 115-044851-00
Parts PCBA FRU
---delete
0632 Flowmeter N2O
051-001416-01 sensor board PCBA
---delete
Maintenance O2 flow senor failure All of the devices use the modified flow sensor for replacement.
AIR flow senor failure to select software according to the type of monitor board(EPSON or
DSP )
2, For the devices which configured O2/N2O/AIR gas supply, please refer
Total flow senor failure to the service manual for calibrating Total flow sensor, when replacing
the Electronic control board or N2O flow sensor or total flow sensor.
Upgrade N/A
Difference of
appearance Different interface location
ACGO
ACGO
801-0631-00143-00 assembly(A5-CE) 115-043752-00
Spare assembly(0634)
FRU ---delete
Parts
mechanical ACGO
115-017878-00 115-043753-00
ACGO(AG) ---delete assembly(0634/AG)
Mechanical ACGO All of the devices use the modified ACGO assembly for
Maintenance
assembly failure replacement.
Upgrade N/A
◼ Software
1,Two new software is launched:
(1)First: Software for devices with EPSON monitor board.
No. Package Modules Latest Version
Language 3.9
Keyboard 01.00/01.01
SystemUpdateTool
00 Tool 5.7
(V5.7)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)
Language 4.6
Keyboard 01.00/01.01
1, The name of”Silence” button is changed to “Audio Pause”, its icon style is changed from to
.
2, Add a new button “Audio Pause Settings” in the “Other Settings” menu of the system menu. when the
“Audio Pause Settings” is set to ON and meanwhile no mid/high level alarms, Press the “Audio Pause”
button, the low level alarm audio will be turned off; Press the button again, the mid/high level alarm audio
will be turned off 120sec,but the low level alarm still is turned off continuously.
3, Flow Pause function is released. when it is enabled, the ventilation and fresh gas will be paused for
60sec.
4, Fresh Gas Flow Optimizer function can be used with CO2 module
5, Fixed a bug which cause alarm “load configuration failed” under a long-term use.
6, Solved the problem that the CAPNOSTAT CO2 module may stop the communication under the
interference of electric knife.
7,Sloved the problem that the network loop(e.g. local area network,or Multi-port switch)may cause the
device to restart
8, Solved the problem the problem that PCV mode is missing when upgrade the software from EPSON
software to DSP software.
◼ Others
1, O2/AIR cylinder supported pressure increased from 15Mpa to 20Mpa.
2, AGSS 3 ways connector assembly is released
Order Number Description Remark
AGSS 3 ways connector Used between the ACGO exhaust
115-026796-00
assembly vent and AGSS interfaces
4, New main control board spare part is added for the maintenance.
Order Number Description Remark
New, Only used for devices with
115-044211-00 Main control board(Non FDA/EPSON)
EPSON monitor board
0635 main control board FRU(Non Old, Only used for devices with
115-040956-00
FDA/VGA) DSP monitor board
5,Operator’s manual and service manual updated.
6, Dust-proof sponge is added to the 1-year maintenance kit.
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16088-A5
Technical News
Subject: ECL128 Engineering changes.
RECORD IDENTIFIER
CONTENT
◼ Hardware
1,The flow sensor is replaced by the new type:
Before After
O2 flow senor
All of the devices use the modified flow sensor for replacement.
failure
Note:
N2O flow senor 1, For the old flow sensor Maintenance, please refer the guide in the spare
Maintenance
failure part
2, Software Upgrade is necessary after change the flow sensor. you need to
AIR flow senor
select software according to the type of monitor board(EPSON or DSP )
failure
Upgrade N/A
Difference of
appearance Different interface location
ACGO
ACGO
801-0631-00143-00 assembly(A5-CE) 115-043752-00
Spare assembly(0634)
FRU ---delete
Parts
mechanical ACGO
115-017878-00 115-043753-00
ACGO(AG) ---delete assembly(0634/AG)
Mechanical ACGO All of the devices use the modified ACGO assembly for
Maintenance
assembly failure replacement.
Upgrade N/A
◼ Software
1,Two new software is launched:
(1)First: Software for devices with EPSON monitor board.
No. Package Modules Latest Version
BIOS
01 BIOS 01.12.00.00
(01.12.00.00)
Language 3.9
flowmeter V1.0
NOTE:
1, New software can only be used on the new EPSON monitor board. The DSP monitoring
board will be damaged when using the new software.
2, Downgrade the higher version to lower version is not allowed.
SystemUpdateTool
00 Tool 5.7
(V5.7)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)
Language 4.6
flowmeter V1.0
1, The name of “Silence” button is changed to “Audio Pause”, its icon style is changed from
to .
2, Add new button “Audio Pause Settings” in the “Other Settings” menu of the system menu. when the
“Audio Pause Settings” is set to ON and meanwhile no mid/high level alarms, Press the “Audio Pause”
button, the low level alarm audio will be turned off; Press the button again, the mid/high level alarm
audio will be turned off 120sec,but the low level alarm still is turned off continuously.
◼ Others
1 , O2/AIR cylinder supported pressure increased from 15Mpa to 20Mpa.
2, AGSS 3 ways connector assembly is released
Order Number Description Remark
AGSS 3 ways connector Used between the ACGO exhaust
115-026796-00
assembly vent and AGSS interfaces
4, New main control board spare part is added for the maintenance.
Order Number Description Remark
New, Only used for devices with
115-044211-00 Main control board(Non FDA/EPSON)
EPSON monitor board
0635 main control board FRU(Non Old, Only used for devices with
115-040956-00
FDA/VGA) DSP monitor board
5, Operator’s manual and service manual updated.
6, Dust-proof sponge is added to the 1-year maintenance kit.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16068-A7
Technical News
Subject: ECL124F Technical Changes.
RECORD IDENTIFIER
CONTENT
Difference of
appearance
For all A7 machines, Using the new spare part for repair. But for the
Fault of Mother board old machines, VGA function does not work in the case of replacing
only the motherboard.
Maintenance
Fault of Control board For all A7 machines, Using the new spare part for repair.
Upgrade To support VGA function, need to replace the Mother Board and Main Control Board.
2. Some symbols and information had been added on new devices after this technical
change because of the requirement of new regulation.
They are:
⚫ Maximum input flow limit for every inlet
⚫ Maximum output flow information for auxiliary supply
⚫ Maximum inlet flow and pressure information for internal vacuum structure inlet
⚫
This symbol means reading operation manual is necessary before operating the parts
with it.
⚫
This symbol means the parts with it could not be treated as common garbage and
could not throw away directly.
⚫ IPX1
This symbol means the parts with it has below waterproof level: water drop falling
down vertically would not influenced the parts
⚫ Weight information
2.2 Add maximum inlet flow and pressure information for internal vacuum structure inlet.
2.3 Add maximum output flow information and symbol on auxiliary O2 supply port.
If you need download information or have any question, please feel free to contact
No.: OMZ-21009-A Series Page 165of 428
PMLS Global Technical Support Dept
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16067-A5
Technical News
Subject: ECL124F Technical Changes.
RECORD IDENTIFIER
CONTENT
Some symbols and information had been added on new devices after this technical
change because of the requirement of new regulation.
They are:
⚫ Maximum input flow limit for every inlet
⚫ Maximum output flow information for auxiliary supply
⚫
This symbol means reading operation manual is necessary before operating the parts
with it.
⚫
This symbol means the parts with it could not be treated as common garbage and
could not throw away directly.
⚫ IPX1
This symbol means the parts with it has below waterproof level: water drop falling
down vertically would not influenced the parts
2. Add maximum output flow information and symbol on auxiliary O2 supply port.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16060-All AM&Ventilator
Technical News
Subject: Anesthesia Machine&Ventilator Parts List is released.
RECORD IDENTIFIER
CONTENT
Note:
(a)Cause the “WATO EX-55/65Pro”(factory code 0625/0626) which were released
recently had changed their name to “WATO EX-55/65”, the corresponding parts list name
will also be changed from "WATO EX-55Pro(0626)_Parts list" & "WATO EX-
65Pro(0625)_Parts list" to WATO EX-55(0626)_Parts list" & "WATO EX-65(0625)_Parts list" in
this version.
(b)2 new parts list "WATO EX-55 Pro(0628)_Parts List", “WATO EX-65 Pro(0627)_Parts
List" will be added in this version, those new types will be available very soon.
If you have any questions & suggestions, or you find any mistakes, please
feedback to Shiling (yangshiling@mindray.com).Thanks a lot for your support.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16055-Vaporizers(V60&Penlon)
New maintenance packages for V60 and new adapter for vaporizers are released
(ECL120)
Technical News
Subject: New maintenance packages for V60 and new adapter for vaporizers are released
(ECL120).
RECORD IDENTIFIER
CONTENT
To Whom It May Concern,
If want more details about the repair and test procedure, please refer to “H-046-
009797-00 V60 Vaporizer Filler Cap Maintenance Guide”. This file has been uploaded to
TDP platform:
/Anesthesia Machine/Ventilator/Anesthesia Machine/V60 (0602)
Influence to upgrade business:
N./A
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16046-A7 (0635)
Technical News
Subject: A7 (0635) Product Change (ECL117).
RECORD IDENTIFIER
CONTENT
Norway Island
Switzerland Croadia
Bosnia&Herzegovina Montenegro
Serbia Ukraine
Moldoova Belarus
Russia Turkey
Georgria Armenia
Azerbaijan Kazakstan
Kyrgystan Tadzhikistan
Jordan Iraq
Iran Saudi-Arabia
Bahrain Qatar
Yemen /
Table 2
Item Spare Parts Number Spare Parts Description Remark
More details about the Drager Vaporizers,please refer to the operator’s manual.The
operator’s manual has been uploaded to TDP platform:
Link:View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Drager Vaporizers
listed in Table 2 could be applied by Sales from Mindray to meet the comerical
requirement.
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16045-A5(0634)
Technical News
Subject: A5 (0634) Product Change (ECL117).
RECORD IDENTIFIER
CONTENT
Norway Island
Switzerland Croadia
Bosnia&Herzegovina Montenegro
Serbia Ukraine
Moldoova Belarus
Russia Turkey
Georgria Armenia
Azerbaijan Kazakstan
Kyrgystan Tadzhikistan
Jordan Iraq
Iran Saudi-Arabia
Bahrain Qatar
Yemen /
Table 2
Item Spare Parts Number Spare Parts Description Remark
More details about the Drager Vaporizers,please refer to the operator’s manual.The
operator’s manual has been uploaded to TDP platform:
Link:View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Drager Vaporizers
listed in Table 2 could be applied by Sales from Mindray to meet the comerical
requirement.
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16044-A7
Technical News
Subject: A7 new software is launched.
RECORD IDENTIFIER
CONTENT
Software 02.09.00 is not compatible with the devices which have software version 02.05.07 or lower
version, for more details please refer to the Software upgrade instruction in software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.
(a)when “Auto Exit Standby button” is “On”, Auto/Manual switch has to be set to
“Manual” before entering/exiting standby mode, this enter/exit standby mode is safer for
the patient, and was used on A7 with software version 02.05.07 or lower.
No.: OMZ-21009-A Series Page 188of 428
PMLS Global Technical Support Dept
Under this setting, the way to enter/exit standby mode as below.
Exit Standby:Touch the waveform area or turn on the fresh gas flow, or select the
Standby soft key on the screen, then follow the screen prompts to exit Standby mode.
Enter Standby: Select the Standby soft key on the main screen.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16043-A5
Technical News
Subject: A5 new software is launched.
RECORD IDENTIFIER
CONTENT
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.
2, When the devices which have bundle version 02.05.07 or lower need to upgrade to the
V02.09.00 of the Bundle version, please change the Ventilator Control Board(DSP) FRU first
2, Downgrade the higher version to lower version is not allowed.
No.: OMZ-21009-A Series Page 192of 428
PMLS Global Technical Support Dept
The Software Technical Changes include:
1, Add “Auto Exit Standby” button in the “service” menu. A7 with new software now
supports 2 ways of enter/exit standby mode. The factory default setting of this button is
“Off”, details of this function as below.
(a) when “Auto Exit Standby button” is “On”, Auto/Manual switch has to be set to
“Manual” before entering/exiting standby mode, this enter/exit standby mode is safer for
the patient, and was used on A7 with software version 02.05.07 or lower.
Under this setting, the way to enter/exit standby mode as below.
Exit Standby: Touch the waveform area or turn on the fresh gas flow, or select the
Standby soft key on the screen, then follow the screen prompts to exit Standby mode.
Enter Standby: Turn off all fresh gas flows, then select the Standby soft key on the
main screen.
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16038-A7
Technical News
Subject: Anesthesia Prediction Function upgrade kit is released for A7(TCN16-FU002).
RECORD IDENTIFIER
CONTENT
Influence to installation:
N/A
Influence to maintenance:
No.: OMZ-21009-A Series Page 196of 428
PMLS Global Technical Support Dept
N/A
Influence to upgrade:
Choose the corresponding guide depending on the machine ACGO configuration,
then finish the upgrade process and perform tests.
Guide notes:
1, Anesthesia Prediction Function (Electrically Controlled ACGO) Upgrade Guide
2, Anesthesia Prediction Function (Mechanical Controlled ACGO) Upgrade Guide
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.
TMZ-16016-A7
Technical News
Subject: ECL115 Technical Changes
RECORD IDENTIFIER
CONTENT
1,Hardware:
⚫ Change description
Now A series has 2 kinds of monitor board: EPSON platform monitor board and DSP platform
monitor board. In 2015 technical changes R&D had replaced EPSON platform monitor board
to DSP ones both on new devices and in spare parts list .But some Problems were found
when EPSON platform monitor board changed to DSP platform monitor board for
maintenance or upgrade, R&D is finding a solution now, before the final plan to release,
please use EPSON platform monitor board to fix EPSON devices. More details please check
below.
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distinguish 1. Please use the start screen to distinguish(recommended)
the device EPSON Platform DSP Platform device
Difference of
appearance
Original monitor board
For DSP platform device(distinguish way please check above), please use 115-036365-
00 Monitoring Board(DSP) FRU for replacement, and upgrade software, need factory
calibration
Note:
Int-flow 1. Please reference the guide in the package for replacement
sensor issue 2. Software upgrade is necessary, and this monitor board can only supply software
version 02.09.00 or higher
3. Factory calibration is necessary, please refer to service manual. If factory
calibration is unavailable, please use 115-036364-00 Ventilation kit(Non-
FDA/DSP)FRU for replacement.
For DSP platform device(distinguish way please check above), please use 115-
036366-00 Ventilator Control Board(DSP) FRU for replacement, and upgrade
software, need factory calibration
Note:
Monitor 1. Please reference the guide in the package for replacement
Maintenance
board issue 2. Software upgrade is necessary, and this monitor board can only supply software
version 02.09.00 or higher
3. Factory calibration is necessary, please refer to service manual. If
factory calibration is unavailable, please use 115-036364-00
Ventilation kit(Non-FDA/DSP)FRU for replacement.
For DSP platform device(distinguish way please check above), please use 115-
036693-00 Drive Assembly(DSP)FRU for replacement, and upgrade software,
need factory calibration
Note:
Expiration 1. Please reference the guide in the package for replacement
valve issue 2. Software upgrade is necessary, and this monitor board can only
supply software version 02.09.00 or higher
3. Factory calibration is necessary, please refer to service manual. If
factory calibration is unavailable, please use 115-036364-00
Ventilation kit(Non-FDA/DSP)FRU for replacement.
Upgrade N/A
⚫ Change description
For reducing the cost, A7 devices produced after 2015.12.31(Serial number larger than
GC-5D000000) will use new screen assembly and backlight board,
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distinguish 1. If it is possible to make sure the device never replaced any parts
the device influenced by this technical change, engineers could distinguish the
before devices by serial numbers, serial numbers smaller than GC-5D000000
Difference of
appearance
Difference of
appearance
⚫ Change description
For reducing the cost, A7 devices produced after 2015.12.31(Serial number larger than
GC-5D000000) will change the bypass assembly material from metal to plastic
Influence on Installation
CO2 absorber appearance changed, please pay attention during customer training
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts influenced
Distingu by this technical change, engineers could distinguish the devices by serial
ish the numbers, serial numbers smaller than GC-5D000000 are the devices before
device change, serial numbers larger than GC-5D000000 are the devices after
before change.
change 2. If it is impossible to make sure the device never replaced any parts
or after influenced by this technical change, please open the device and use
change Difference of appearance to distinguish(recommended)
Bypass actuator
Differen
ce of
appear
ance
Bypass Assembly
⚫ Change description
For reducing the cost, A7 devices produced after 2015.12.31(Serial number larger than GC-
5D000000) will use new base assembly and castor
Influence on Installation
N/A
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts
influenced by this technical change, engineers could distinguish the
Distinguish
devices by serial numbers, serial numbers smaller than GC-5D000000
the device
are the devices before change, serial numbers larger than GC-5D000000
before
are the devices after change.
change or
2. If it is impossible to make sure the device never replaced any parts
after
influenced by this technical change, please open the device and use
change
Difference of appearance to distinguish(recommended)
Difference
of
appearanc
e
⚫ Change description
To keep consistent with WATO series, R&D also apart the A series 1 year maintenance kit to
3 new ones like WATO series. And a new 3 year maintenance kit is released also. More details
please check below.
Influence on Installation
N/A
Influence on Maintenance & upgrade
Before change After change
Could be used Could be used
on Devices on Devices
Maintain
before&After 115-036367- before&After
Kit(0635,Basic,
change 00 change
1 yr)
Suggest use
new kit
Could be used
Preventative on Devices
115- Maintain
Maintenance 115-036368- before&After
024063- Kit(0635,Exten
kits(1 year) 00 change
00 d,1 yr)
-Still could apply Suggest use
Spare
new kit
Parts
Could be used
on Devices
Maintain
115-036369- before&After
Kit(0635,Advan
00 change
ced,1 yr)
Suggest use
new kit
Could be used Could be used
Preventative
115- on Devices Maintain on Devices
Maintenance 115-036370-
024064- before&After Kit(0635/3 before&After
kits(3 years) 00
00 change years) change
-Still could apply
Suggest use
2,Software:
⚫ Change description
For optimizing the display and solve the software bugs, R&D released new software version.
Specific change points as below
Adding 13 working languages
Turkish, Czech, Polish, Portuguese, Russian, Norwegian, Bulgarian, Danish, Dutch,
Finnish, Romanian, Serbian, Swedish
Original 7 working languages are:English, Chinese, French, German, Italian, Hungarian,
Spanish
New software could support 20 languages in total.
Display optimize for French and Spanish languages
If old software(02.09.00&02.11.00) works in French or Spanish, sometime alarm message
could not show totally can be found, new software could solve this problem
Lung recruitment tool optimize
In old software(02.09.00&02.11.00), if one uses Lung recruitment tool, and during lung
recruitment function changes AUTO/Manual switch from AUTO to Manual, the mode shows
on the top of screen will not change. New software could solve this proble.
NMT module optimize
1. In old software(02.09.00&02.11.00), keep disconnecting and reinstalling NMT module
for several times in short time would make anesthesia device restart, new software
could solve this problem
2. In old software(02.09.00&02.11.00), some messages in NMT form could not show
totally, new software could solve this problem.
SIMV-VC optimize
In old software(02.09.00&02.11.00), set Exp% 70% or higher in Standby mode, after this
mode starting, Exp% will return to default setting(25%)
Anesthesia Prediction optimize
If the drug in air way could not be detected or distinguished, the drug type on screen will
shows ”AA”,and MAC wave form coordinate axis will not show unit. Since unit is
meaningless in this situation
Manual leak test optimize
One of Manual leak test result ,leakage 200-1000ml, has changed its color from red to
white. This color change will not influence the leakage decision condition.
No.: OMZ-21009-A Series Page 207of 428
PMLS Global Technical Support Dept
Bug solution
In old software(02.09.00&02.11.00), sometimes if customer uses 3L manual bag(standard
configuration) for manual leak test, because of the algorithm, even there is no leakage, but
device will give a larger leakage result, and causes manual leak test could not pass.
If you find the similar case, please upgrade the latest software version.
Influence on Installation
Please change the language during installation when necessary
Influence on Maintenance & upgrade
No. Package Modules Latest Version
Bundle Version V02.13.00
Host Software V03.13.02
BIOS V01.14.00
FPGA Display V1.4
Bundle
1 FPGA Sound V1.5
(V02.13.00)
Ventilator Protect Module V01.02.00
Ventilator Control Module V01.03.00
Language V4.2
Keyboard V01.00/01.01
Software 02.09.00 or higher are not compatible with the devices which have software version
02.05.07 or lower version, for more details please refer to the Software upgrade instruction in
software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board
will be damaged when using the new software.
2, When the devices which have bundle version 02.05.07 or lower need to upgrade to
the V02.09.00 or higher, please change the Ventilator Control Board(DSP) FRU first
3, Downgrade the higher version to lower version is not allowed.
4,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.09.00 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test,
please perform steps below to solve this issue:
No.: OMZ-21009-A Series Page 208of 428
PMLS Global Technical Support Dept
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Main】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Main】->【General】->【Zero Flow Meters】,zero flow
meters and restart again;
For those devices which have bundle version 02.05.00 or higher, steps above are
unnecessary during upgrading to 02.09.00
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-16015-A5
Technical News
Subject: ECL115 Technical Changes
RECORD IDENTIFIER
CONTENT
1,Hardware:
⚫ Change description
Now A series has 2 kinds of monitor board: EPSON platform monitor board and DSP platform
monitor board. In 2015 technical changes R&D had replaced EPSON platform monitor board
to DSP ones both on new devices and in spare parts list .But some Problems were found when
EPSON platform monitor board changed to DSP platform monitor board for maintenance or
upgrade, R&D is finding a solution now, before the final plan to release, please use EPSON
platform monitor board to fix EPSON devices. More details please check below.
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distinguish 3. Please use the start screen to distinguish(recommended)
the device EPSON Platform DSP Platform device
⚫ Change description
For reducing the cost,A5 devices produced after 2015.12.31(Serial number larger than FY-
5D000000) will delete sensor interface board (under work surface, beside expiration valve),
the original Int-flow sensor on this board will be changed to new type and moved to monitor
board
Influence on Installation
N/A
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts influenced by this
technical change, engineers could distinguish the devices by serial numbers, serial
Distinguish
numbers smaller than FY-5D000000 are the devices before change, serial numbers
the device
larger than FY-5D000000 are the devices after change.
before
2. If it is impossible to make sure the device never replaced any parts influenced by this
change or
technical change, please open the device and use Difference of appearance to
after change
distinguish(recommended)
Difference of
appearance
Original monitor board
For DSP platform device(distinguish way please check above), please use 115-036365-
00 Monitoring Board(DSP) FRU for replacement, and upgrade software, need factory
calibration
Note:
Int-flow 4. Please reference the guide in the package for replacement
sensor issue 5. Software upgrade is necessary, and this monitor board can only supply software
version 02.09.00 or higher
6. Factory calibration is necessary, please refer to service manual. If factory
calibration is unavailable, please use 115-036364-00 Ventilation kit(Non-
FDA/DSP)FRU for replacement.
For DSP platform device(distinguish way please check above), please use 115-
036366-00 Ventilator Control Board(DSP) FRU for replacement, and upgrade
software, need factory calibration
Note:
Monitor 4. Please reference the guide in the package for replacement
Maintenance
board issue 5. Software upgrade is necessary, and this monitor board can only supply software
version 02.09.00 or higher
6. Factory calibration is necessary, please refer to service manual. If
factory calibration is unavailable, please use 115-036364-00
Ventilation kit(Non-FDA/DSP)FRU for replacement.
For DSP platform device(distinguish way please check above), please use 115-
036693-00 Drive Assembly(DSP)FRU for replacement, and upgrade software,
need factory calibration
Note:
Expiration 4. Please reference the guide in the package for replacement
valve issue 5. Software upgrade is necessary, and this monitor board can only
supply software version 02.09.00 or higher
6. Factory calibration is necessary, please refer to service manual. If
factory calibration is unavailable, please use 115-036364-00
Ventilation kit(Non-FDA/DSP)FRU for replacement.
Upgrade N/A
⚫ Change description
For reducing the cost, A7 devices produced after 2015.12.31(Serial number larger than
GC-5D000000) will use new screen assembly and backlight board,
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distinguish 1. If it is possible to make sure the device never replaced any parts
the device influenced by this technical change, engineers could distinguish the
Difference of
appearance
⚫ Change description
For reducing the cost, A5 devices produced after 2015.12.31(Serial number larger than FY-
5D000000) will use new Pop-off valve
Influence on Installation
N/A
⚫ Influence on Maintenance & upgrade
No.: OMZ-21009-A Series Page 214of 428
PMLS Global Technical Support Dept
1. If it is possible to make sure the device never replaced any parts influenced
Distingui
by this technical change, engineers could distinguish the devices by serial
sh the
numbers, serial numbers smaller than FY-5D000000 are the devices before
device
change, serial numbers larger than FY-5D000000 are the devices after
before
change.
change
2. If it is impossible to make sure the device never replaced any parts
or after
influenced by this technical change, please open the device and use
change
Difference of appearance to distinguish(recommended)
Before change After change
Differenc
e of
appeara
nce
⚫ Change description
For reducing the cost, A5 devices produced after 2015.12.31(Serial number larger than FY-
5D000000) will change the bypass assembly material from metal to plastic
Influence on Installation
CO2 absorber appearance changed, please pay attention during customer training
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts influenced
Distingu
by this technical change, engineers could distinguish the devices by serial
ish the
numbers, serial numbers smaller than FY-5D000000 are the devices before
device
change, serial numbers larger than FY-5D000000 are the devices after
before
change.
change
2. If it is impossible to make sure the device never replaced any parts
or after
influenced by this technical change, please open the device and use
change
Difference of appearance to distinguish(recommended)
Before change After change
Bypass actuator
Differen
ce of
appear
ance
Bypass Assembly
⚫ Change description
For reducing the cost, A5 devices produced after 2015.12.31(Serial number larger than FY-
5D000000) will use new base assembly and castor
Influence on Installation
N/A
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts
influenced by this technical change, engineers could distinguish the
Distinguish
devices by serial numbers, serial numbers smaller than GC-5D000000
the device
are the devices before change, serial numbers larger than GC-5D000000
before
are the devices after change.
change or
2. If it is impossible to make sure the device never replaced any parts
after
influenced by this technical change, please open the device and use
change
Difference of appearance to distinguish(recommended)
Difference
of
appearanc
e
⚫ Change description
To keep consistent with WATO series, R&D also apart the A series 1 year maintenance kit to 3
new ones like WATO series. And a new 3 year maintenance kit is released also. More details
please check below.
Influence on Installation
N/A
Influence on Maintenance & upgrade
Before change After change
Could be used Could be used
on Devices on Devices
Maintain
before&After 115-036367- before&After
Kit(0635,Basic,
change 00 change
1 yr)
Suggest use
new kit
Could be used
Preventative on Devices
115- Maintain
Maintenance 115-036368- before&After
024063- Kit(0635,Exten
kits(1 year) 00 change
00 d,1 yr)
-Still could apply Suggest use
new kit
Could be used
Spare on Devices
Maintain
Parts 115-036369- before&After
Kit(0635,Advan
00 change
ced,1 yr)
Suggest use
new kit
Could be used Could be used
Preventative on Devices on Devices
115- Maintain
Maintenance before&After 115-036370- before&After
024064- Kit(0635/3
kits(3 years) change 00 change
00 years)
-Still could apply Suggest use
new kit
Note:
4. Please pay attention to our part list which will be released on March to learn the specific
difference between old and new maintenance kit.
5. Old maintenance kit will be kept until 2016.07,after July old ones will not be available and
2,Software:
⚫ Change description
For optimizing the display and solve the software bugs, R&D released new software version.
Specific change points as below
Adding 13 working languages
Turkish, Czech, Polish, Portuguese, Russian, Norwegian, Bulgarian, Danish, Dutch, Finnish,
Romanian, Serbian, Swedish
Original 7 working languages are:English, Chinese, French, German, Italian, Hungarian,
Spanish
New software could support 20 languages in total.
Display optimize for French and Spanish languages
If old software(02.09.00&02.11.00) works in French or Spanish, sometime alarm message
could not show totally can be found, new software could solve this problem
Lung recruitment tool optimize
In old software(02.09.00&02.11.00), if one uses Lung recruitment tool, and during lung
recruitment function changes AUTO/Manual switch from AUTO to Manual, the mode shows
on the top of screen will not change. New software could solve this proble.
NMT module optimize
3. In old software(02.09.00&02.11.00), keep disconnecting and reinstalling NMT module for
several times in short time would make anesthesia device restart, new software could
solve this problem
4. In old software(02.09.00&02.11.00), some messages in NMT form could not show totally,
new software could solve this problem.
SIMV-VC optimize
In old software(02.09.00&02.11.00), set Exp% 70% or higher in Standby mode, after this
mode starting, Exp% will return to default setting(25%)
Anesthesia Prediction optimize
If the drug in air way could not be detected or distinguished, the drug type on screen will
shows ”AA”,and MAC wave form coordinate axis will not show unit. Since unit is meaningless
in this situation
Manual leak test optimize
One of Manual leak test result ,leakage 200-1000ml, has changed its color from red to white.
This color change will not influence the leakage decision condition.
Bug solution
In old software(02.09.00&02.11.00), sometimes if customer uses 3L manual bag(standard
configuration) for manual leak test, because of the algorithm, even there is no leakage, but
device will give a larger leakage result, and causes manual leak test could not pass.
If you find the similar case, please upgrade the latest software version.
Influence on Installation
Please change the language during installation when necessary
Influence on Maintenance & upgrade
No. Package Modules Latest Version
Bundle Version V02.13.00
Host Software V03.13.02
BIOS V01.14.00
FPGA Display V1.4
Bundle
1 FPGA Sound V1.5
(V02.13.00)
Ventilator Protect Module V01.02.00
Ventilator Control Module V01.03.00
Language V4.2
Keyboard V01.00/01.01
Software 02.09.00 or higher are not compatible with the devices which have software version
02.05.07 or lower version, for more details please refer to the Software upgrade instruction in
software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board
will be damaged when using the new software.
2, When the devices which have bundle version 02.05.07 or lower need to upgrade
to the V02.09.00 or higher, please change the Ventilator Control Board(DSP) FRU
first
3, Downgrade the higher version to lower version is not allowed.
4,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.09.00
and restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test,
please perform steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-16004-A7
Technical News
Subject: ECL108-F Technical Changes
RECORD IDENTIFIER
CONTENT
1,Hardware:
⚫ Change description
For meet the requirements of new statute(IEC60601-1:2005+A1:2012-7.2.3) , after this
technical change the new device(serial numbers larger than GC-62000000) will use new mark
for “please reading operation manual before using”
Influence on Installation
May be customer could ask
Influence on Maintenance & upgrade
Distingu
ish the
device
before 1.Using Difference of appearance to distinguish
change
or after
change
Before change After change
Differen
ce of
appeara
nce
Spare
N/A
Parts
Mainten
N/A
ance
Upgrade N/A
⚫ Change description
R&D had found because of the Spring-tube of driving gas and Spring-tube of APL valve,
devices could not pass Automatic leak test (manual leak test could pass) after a period of
time. For solving this problem, after this technical change the new device(serial numbers
larger than GC-62000000) will use new material and structure for Spring-tube of driving
gas(115-034448-00) and Spring-tube of APL valve(115-034449-00)
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distingui 1. If it is possible to make sure the device never replaced any parts influenced by
sh the this technical change, engineers could distinguish the devices by serial numbers,
device serial numbers smaller than GC-62000000 are the devices before change, serial
before numbers larger than GC-62000000 are the devices after change.
change 2. If it is impossible to make sure the device never replaced any parts influenced by
or after this technical change, please open the device and use Difference of
change appearance to distinguish(recommended)
Before change After change
Differenc
e of
appeara
nce
Spring-tube of APLvalve
Maintena
Please use 115-034449-00 Spring-tube of APL valve FRU for replacement,
nce
and upgrade software, need factory calibration
Note:
1. Please refer to service manual for replacement
2. In spare part package there are 5 pcs Wire fixed, please fix the tubes as
below
Spring
-tube
of APL
valve
issue
Upgrade N/A
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-16003-A5
Technical News
Subject: ECL108-F Technical Changes
RECORD IDENTIFIER
CONTENT
1,Hardware:
⚫ Change description
For meet the requirements of new statute(IEC60601-1:2005+A1:2012-7.2.3) , after this
technical change the new device(serial numbers larger than FY-62000000) will use new mark
for “please reading operation manual before using”
Influence on Installation
May be customer could ask
Influence on Maintenance & upgrade
Distingu
ish the
device
before 1.Using Difference of appearance to distinguish
change
or after
change
Before change After change
Differen
ce of
appeara
nce
Spare
N/A
Parts
Mainten
N/A
ance
Upgrade N/A
⚫ Change description
R&D had found because of the Spring-tube of driving gas and Spring-tube of APL valve, devices
could not pass Automatic leak test (manual leak test could pass) after a period of time. For
solving this problem, after this technical change the new device(serial numbers larger than FY-
62000000) will use new material and structure for Spring-tube of driving gas(115-034448-00)
and Spring-tube of APL valve(115-034449-00)
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distingui 3. If it is possible to make sure the device never replaced any parts influenced by
sh the this technical change, engineers could distinguish the devices by serial numbers,
device serial numbers smaller than FY-62000000 are the devices before change, serial
before numbers larger than FY-62000000 are the devices after change.
change 4. If it is impossible to make sure the device never replaced any parts influenced by
or after this technical change, please open the device and use Difference of
change appearance to distinguish(recommended)
Before change After change
Differenc
e of
appeara
nce
Spring-tube of APLvalve
Maintena
Please use 115-034449-00 Spring-tube of APL valve FRU for replacement,
nce
and upgrade software, need factory calibration
Note:
3. Please refer to service manual for replacement
4. In spare part package there are 5 pcs Wire fixed, please fix the tubes as
below
Spring
-tube
of APL
valve
issue
Upgrade N/A
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-15050-A7
Technical News
Subject: EFU009 Technical Changes
RECORD IDENTIFIER
CONTENT
Function Upgrades of A7 are released. New hardware and software will be available
from 2015.09.01, most of this changes are not compatible with devices which were
produced before 2015.09, please pay attention to the maintenance and upgrading
ways.
1,Hardware:
⚫ Since the platform of monitoring assembly is changed from EPSON Platform to DSP Platform,
new Drive Board and Monitoring Board are released. Old Drive Board and Monitoring Board
could not be ordered now.
Before After
Cancel capacitance, add a new backup Longer interface between the boards
connector
Difference of
appearance
⚫ New configuration electronic monitoring of gas supply for A7 is launched . The pressure
gauge configuration also is available, but two kinds of configuration can only choose one. For
the new function, the Electronic control board is changed, new gas inlet assembly, new
cylinder assembly, new pressure sensor are added. This new function is not able to
upgrade on-site.
Before After
Difference of
appearance
Fault of Drive Board For all A7 machines, Using the new spare part for repair.
1. For the devices which configured gas supply pressure gauge,
Maintenance Fault of gas inlet please fix as usual
assembly 2. For the device which configured electronic monitoring of gas
supply, please check the form below
Upgrade Not allowed upgrading
New gas inlet assembly and new cylinder assembly as bellow:
Pressure sensor
Pressure sensor
new gas inlet assembly with pressure sensor new gas cylinder assembly with pressure sensor
Device which configured electronic monitoring of gas supply gas inlet assembly maintenance form:
Part of fault Replace SN Part Description reference
N2O origin inlet
N2O inlet assembly(DISS) 115-033733-00 With pressure sensor
module(0635/DISS)
AIR origin inlet
AIR inlet assembly(DISS) 115-033734-00 With pressure sensor
module(0635/DISS)
O2 origin inlet
O2 inlet assembly(DISS) 115-033735-00 With pressure sensor
module(0635/DISS)
Tray
Maintenance form:
Part of fault Replace SN Part Description Reference
2,Software:
New software is launched
No. Package Modules Latest Version
Bundle Version V02.09.00
Host Software V03.09.00
BIOS V01.14.00
FPGA Display V1.4
Bundle
1 FPGA Sound V1.5
(V02.09.00)
Ventilator Protect Module V01.02.00
Ventilator Control Module V01.03.00
Language V3.8
Keyboard V01.00/01.01
Software 02.09.00 is not compatible with the devices which have software version 02.05.07 or lower
version, for more details please refer to the Software upgrade instruction in software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.
Self-checking progress
Help tips
Self-test details
⚫ Alarm visualization: Increase help tips for the High-level alarm only, and add the duration
show about all alarms.
Help tips
Duration
After
Before
New button 1
New button 2
New menu
After
d, Add display about gas supply pressure electric monitoring when the machine is
equipped with electric monitoring of gas supply pressure.
e, Add display about the agent speed when the machine is equipped with the internal AG
module.
Agent speed
Pressure regulated
TV setting
SIMV-VG
PS mode
CPAP mode
10 min 20 min
History Prediction
BISx4 Cable
BISx4 Sensor
2 BIS waves
2 EEG waves
NMT area
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-15049-A5
Technical News
Subject: EFU009 Technical Changes
RECORD IDENTIFIER
CONTENT
Function Upgrades of A5 are released. New hardware and software will be available
from 2015.09.01, most of this changes are not compatible with devices which were
produced before 2015.09, please pay attention to the maintenance and upgrading.
1,Hardware:
⚫ Since the platform of monitoring assembly is changed from EPSON Platform to DSP Platform,
new Drive Board and Monitoring Board are released. Old Drive Board and Monitoring Board
could not be ordered now
Before After
Cancel capacitance, add a new backup connector Longer interface between the boards
Difference of
appearance
Tray
Maintenance form::
Part of
Replace SN Part Description Reference
fault
GCX Tray
Tray 034-000461-00 Tray can be changed independently
GCX 12" x 12" Articulating Arm with Arm problem has to change whole part to
Arm 034-000460-00
Tray solve
2,Software:
New software is launched
No. Package Modules Latest Version
Bundle Version V02.09.00
Host Software V03.09.00
BIOS V01.14.00
FPGA Display V1.4
Bundle
1 FPGA Sound V1.5
(V02.09.00)
Ventilator Protect Module V01.02.00
Ventilator Control Module V01.03.00
Language V3.8
Flowmeter Software V1.9
2 Power(V1.7) Power System V1.7
Software 02.09.00 is not compatible with the devices which have software version 02.05.07 or lower
version, for more details please refer to the Software upgrade instruction in software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.
2, When the devices which have bundle version 02.05.07 or lower need to upgrade to the
V02.09.00 of the Bundle version, please change the Ventilator Control Board(DSP) FRU first
2, Downgrade the higher version to lower version is not allowed.
Self-checking progress
Help tips
Self-test details
⚫ Alarm visualization: Increase help tips for the High-level alarm only, and add the duration
show about all alarms.
Help tips
Duration
Before
After
Before
New button 1
Clear History
switch, the old
discharges setting
New button 2
New menu
After
⚫ Two new ventilation modes are available: SIMV-VG,CPAP/PS. Use the following material to
No.: OMZ-21009-A Series Page 250of 428
PMLS Global Technical Support Dept
activate the new modes through actiavtion code or USB disk
Order number Part description reference
USB Disk with Activation Code (SIMV-
115-034847-00
VG) Need the software versiong is V02.09.00.
115-034846-00 USB Disk with Activation Code (CPAP/PS)
TV setting
PS mode
CPAP mode
BISx4 Cable
BISx4 Sensor
2 BIS waves
2 EEG waves
NMT area
If you need download information or have any question, please feel free to contact
No.: OMZ-21009-A Series Page 254of 428
PMLS Global Technical Support Dept
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-15044-A7
Technical News
Subject: ECL112 Technical Changes
RECORD IDENTIFIER
CONTENT
1, New backup cylinder assembly and copper tubes of high pressure gauge are released, old backup
cylinder assembly and copper tubes of high pressure gauge could not be ordered any more. The new
cylinder assembly and tubes are not compatible with old one. Change details and maintenance plan
please check below
Old backup cylinder assembly and tubes New backup cylinder assembly and tubes
Appearance
differences
O2 cylinder assembly
O2 Yoke Assembly(With
115-014476-01 (0632), without tube 115-034134-00
gauge tube)FRU
(Deleted)
N2O cylinder assembly
Spare Parts N2O Yoke Assembly(With
115-014477-01 (0632), without tube 115-034136-00
changes gauge tube)FRU
(Deleted)
AIR cylinder assembly
AIR Yoke Assembly(With
115-014478-01 (0632) , without tube 115-034135-00
gauge tube)FRU
(Deleted)
2, The structure of electronic ACGO has changed. Old ACGO could not be ordered any more. A new
part No. is created for new ACGO, the new ACGO is not compatible with old one. Change details and
maintenance plan please check below.
Old ACGO Assembly New ACGO Assembly
before
after
Appearance
differences
3, New circuit heater is released, old circuit heater could not be ordered any more, the new circuit
heater is compatible with old one. Change details and maintenance plan please check below
Old circuit heater New circuit heater
Appearance
differences
4, New back pressure valve is released. Old back pressure valve could not be ordered any more, the
new back pressure valve is compatible with old one. Change details and maintenance plan please
check below.
Old back pressure valve New back pressure valve
Appearance
differences
Copper Aluminum
5, New Optical Encoder is released, old Optical Encoder could not be ordered any more,. A new part
No. is created for new Optical Encoder, the new Optical Encoder is compatible with old one. Change
Appearance
differences
Spare Part
old new
6,The material of sample gas return connector changed from materal to plastic, old sample gas return
connector could not be ordered any more. A new part No. is created for new sample gas return
connector, the new sample gas return connector is compatible with old one. Change details and
maintenance plan please check below
Old sample gas return connector New sample gas return connector
115-015457-00
115-034138-00
Appearance
082-001206-00 metal
differences
plastic
M6Q-030093-00
7, The material of drawer changes from aluminum to SGCC, old drawer could not be ordered any
more. A new part No. is created for new material drawer, the new drawer is compatible with old one.
Change details and maintenance plan please check below
Old drawer New drawer
Appearance
differences
8, Tube from Expiration Valve to Breathing Circuit and tube from Breathing Circuit to AGSS
container have changed from Silicone to spring tubes. Old tubes could not be ordered any more,
the new tubes are compatible with old ones. Change details and maintenance plan please check
below
Old tubes New tubes
M6G-020042--- 115-034449-00
Appearance
differences
M6G-020039--- 115-034448-00
9, New Gas Cylinder Wrench is released, old Gas Cylinder Wrench could not be ordered any more. A
new part No. is created for new Gas Cylinder Wrench, the new Gas Cylinder Wrench is compatible
with old one. Change details and maintenance plan please check below
Old Gas Cylinder Wrench Old Gas Cylinder Wrench
Appearance
differences
10, The tubes for O2&N2O&AIR are changed from colorized to transparent, and the material of most
connector changes from metal to plastic, old ones could not be ordered any more. The new tubes and
connectors are compatible with old ones. Change details and maintenance plan please check below.
Old tubes and connectors New tubes and connectors
Appearance
differences
Tube PU (polyether)
082-000517-00 5.5mmX8mm orange
(Deleted)
Tube PU (polyether) Hose,soft PU
082-000521-00 5.5mmX8mm green M6G-020045--- transparent tube
(Deleted) 8mmX5.5mm
Tube PU 5.5mmX8mm
082-000662-00 blue
(Deleted)
tube.PU (polyether)
082-000522-00 4mmX6mm green
(Deleted)
Tube PU (polyether)
Transparent PU Tube
082-000520-00 4mmX6mm orange M6G-020026---
4*6 (20m)
(Deleted)
tube.PU (polyether)
Spare Parts 082-000524-00 4mmX6mm blue
changes (Deleted)
Tube PU (polyether)
082-000523-00 2.5mmX4mm green
(Deleted)
Tube PU (polyether)
082-000518-00 2.5mmX4mm orange M6G-020046--- Tubing.
(Deleted)
Tube PU (polyether)
082-000516-00 2.5mmX4mm blue
(Deleted)
Connector.6560-6
082-000582-00 M6Q-030024--- Fitting 3140-06-00
(Deleted)
Connector.6560-8
082-000583-00 M6Q-030025--- tube to tube.Y type
(Deleted)
connector.cross type
082-001201-00 66006 M6Q-030097--- connector.cross type
(Deleted)
connector.Φ4 to Φ6 4 to 6.straight
082-001285-00 M6Q-030068---
(Deleted) connector
Maintenanc
e For all A7 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A
11, Above new spare parts No. will be released in Part list 20150901 in September, Please pay
attention to our TN and update your Part list in time.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Thank you for your kind attention and cooperation.
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-15043-A5
Technical News
Subject: ECL112 Technical Changes
RECORD IDENTIFIER
CONTENT
1, New backup cylinder assembly and copper tubes of high pressure gauge are released, old backup
cylinder assembly could not be ordered any more, but old copper tubes of high pressure gauge can
be ordered normally. The new cylinder assembly and tubes are not compatible with old one. Change
details and maintenance plan please check below
Old backup cylinder assembly and tubes New backup cylinder assembly and tubes
Appearance
differences
O2 cylinder assembly(for
O2 Yoke Assembly(0634/With
115-015777-00 gas bench), without tube 115-034395-00
tube)FRU
(Deleted)
N2O cylinder
N2O Yoke
Spare Parts assembly(for gas bench),
115-015779-00 115-034397-00 Assembly(0634/With
changes without tube
tube)FRU
(Deleted)
AIR cylinder assembly(for
AIR Yoke Assembly(0634/With
115-015778-00 gas bench), without tube 115-034396-00
tube)FRU
(Deleted)
2, New circuit heater is released, old circuit heater could not be ordered any more, the new circuit
heater is compatible with old one. Change details and maintenance plan please check below
Old circuit heater New circuit heater
Appearance
differences
3, New back pressure valve is released. Old back pressure valve could not be ordered any more, the
new back pressure valve is compatible with old one. Change details and maintenance plan please
check below
Old back pressure valve New back pressure valve
Appearance
differences
Copper Aluminum
4, The material of drawer changes to SGCC from aluminum. A new part No. is created for new material
drawer, old drawer could not be ordered any more, the new drawer is compatible with old one. Change
details and maintenance plan please check below
Old drawer New drawer
Appearance
differences
5, Tube from Expiration Valve to Breathing Circuit and tube from Breathing Circuit to AGSS
container have changed from Silicone to spring tubes. Old tubes could not be ordered any more,
the new tubes are compatible with old ones. Change details and maintenance plan please check
M6G-020042--- 115-034449-00
Appearance
differences
M6G-020039--- 115-034448-00
6, New Gas Cylinder Wrench is released. A new part No. is created for new Gas Cylinder Wrench; old
Gas Cylinder Wrench could not be ordered any more, the new Gas Cylinder Wrench is compatible with
old one. Change details and maintenance plan please check below
Old Gas Cylinder Wrench Old Gas Cylinder Wrench
Appearance
differences
7, The tubes for O2&N2O&AIR are changed from colorized to transparent, and the material of most
connector changes from metal to plastic, the new tubes and connectors are compatible with old ones.
Change details and maintenance plan please check below
Appearance
differences
tube.PU (polyether)
082-000517-00 5.5mmX8mm orange
(Deleted)
Hose,soft PU
tube.PU (polyether)
M6G-020045--- transparent tube
082-000521-00 5.5mmX8mm green
8mmX5.5mm
(Deleted)
tube.PU 5.5mmX8mm blue
082-000662-00
(Deleted)
tube.PU (polyether)
082-000522-00 4mmX6mm green
(Deleted)
tube.PU (polyether)
Transparent PU Tube
082-000520-00 4mmX6mm orange M6G-020026---
4*6 (20m)
(Deleted)
tube.PU (polyether)
082-000524-00 4mmX6mm blue
Spare Parts
(Deleted)
changes
tube.PU (polyether)
082-000523-00 2.5mmX4mm green
(Deleted)
tube.PU (polyether)
082-000518-00 2.5mmX4mm orange M6G-020046--- Tubing.
(Deleted)
tube.PU (polyether)
082-000516-00 2.5mmX4mm blue
(Deleted)
Connector.6560-6
082-000582-00 M6Q-030024--- Fitting 3140-06-00
(Deleted)
Connector.6560-8
082-000583-00 M6Q-030025--- tube to tube.Y type
(Deleted)
connector.cross type 66006
082-001201-00 M6Q-030097--- connector.cross type
(Deleted)
082-001285-00 connector.Φ4 to Φ6 M6Q-030068--- 4 to 6.straight
8, Above new spare parts No. will be released in Part list 20150901 in September, Please pay attention
to our TN and update your Part list in time.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-15016-A7
Technical News
Subject: EFU006 Technical Changes
RECORD IDENTIFIER
CONTENT
Pic 1 【General】menu
b, In【Display】menu(Pic2):
Pressure Display, Plimit Line, CO2 Placement, Gas Scales, Waveform Display, Screen
Brightness, Key Click Volume
c, In【Demographics】menu(Pic 3):
Patient ID, First Name, Last Name, DOB, Bed, Room, Point of Care, Weight, Facility
No.: OMZ-21009-A Series Page 271of 428
PMLS Global Technical Support Dept
Pic 8 Spirometry
⚫ Circuit compliance compensation not available when the Leak test fail or skip after
starting the AM, and Display the Warning when exit the Standby Mode(Pic 9,10) .
NOTE:
1, Downgrade the higher version to lower version is not allowed.
2,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.05.07 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test, please perform
steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Setup】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Setup】->【General】->【Zero Flow Meters】,zero flow meters
and restart again;
3, For those devices which have bundle version 02.05.00 or higher, steps above are unnecessary
during upgrading to 02.05.07.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-15015-A5
Technical News
Subject: EFU006 Technical Changes
RECORD IDENTIFIER
CONTENT
Pic 8 Spirometry
⚫ Circuit compliance compensation not available when the Leak test fail or skip after
starting the AM, and Display the Warning when exit the Standby Mode(Pic 9,10).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14047-A7
Technical News
Subject: ECL096-F Technical Changes
RECORD IDENTIFIER
CONTENT
To Whom It May Concern,
Below content is only for your information.
1. For distinguishing between BFCS(O2&AIR) and BFCS(O2&AIR&N2O).A new
optimizing is released for the configuration A7 BFCS(O2&AIR).This change gives a
better appearance and keeps all the functions offered by old type.Table1 below will
give
The information would be used in installation, for maintenance please check Table2.
More change details please check table1
Before change(Old type) After change(New type)
Appearanc
e
Inside
structure
Compatibility Table
A7(0635) Before After Compatibility
O2 needle valve failure(US
115-029982-00 115-029990-00 NO
Standard)
O2 needle valve
115-029985-00 115-029993-00 NO
failure(Canada Standard)
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14035-A7
Technical News
Subject: EFU004 Technical Changes
RECORD IDENTIFIER
CONTENT
Compatibility Table
A5 (0634) Before After Compatibility
115-
Muffler Assembly 018356- 115-018356-00 YES
00
115-
AGSS passive scavenging tube assembly 002015- 115-002015-00 YES
00
115-
Passive scavenging tube material packet 002342- 115-002342-00 YES
00
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
TMZ-14034-A5
Technical News
Subject: EFU004 Technical Changes
RECORD IDENTIFIER
CONTENT
Compatibility Table
A5 (0634) Before After Compatibility
115-002015-
AGSS passive scavenging tube assembly 115-002015-00 YES
00
115-002342-
Passive scavenging tube material packet 115-002342-00 YES
00
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14033-A7
Technical News
Subject: ECL090-F Technical Changes
RECORD IDENTIFIER
CONTENT
5. The pneumatic system has changed, the previous position of Back Pressure Valve
No.: OMZ-21009-A Series Page 302of 428
PMLS Global Technical Support Dept
is between gas mixer assembly and 3-way valve,but now it has moved to the behind
of main flowmeter assembly,this change can decrease the pressure undulating of
EFCS, make the floater in main flowmeter assembly more stable. P/N of Back
Pressure Valve, main flowmeter assembly, gas mixer assembly has changed.
Change Info as below(Table 2)
Before change After change
Part description
P/N Remark P/N Remark
801-0631-00042- / /
Back Pressure Valve 115-028145-00
00
Assembly has Assembly has new
old silkscreen silkscreen on
on it ,and it ,and contents
main flowmeter assembly 115-014479-00 doesn’t content 115-027534-00 some character
character labels labels and a
and any service service guide
guide
Assembly has Assembly has new
old silkscreen silkscreen on
on it ,and it ,and contents
gas mixer assembly(0632) 115-016781-00 doesn’t content 115-027532-00 some character
character labels labels and a
and any service service guide
guide
Table 2 Change Information
Note: Since the Back Pressure Valve position has changed,,the silkscreen on gas mixer
assembly and total flowmeter assembly can’t match the main pneumatic system tube
connection diagram.
For solving the problem above, we offer the maintenance plans as below(Table 3)
Maintenance Plan
device made after 2014 Sept.
device made before 2014 Sept.
malfunction (Back Pressure Valve is between
(Back Pressure Valve is between gas mixer assembly
main flow sensor assembly and
and 3-way valve)
main flowmeter assembly)
Back Pressure
Valve Change Back Pressure Valve Change Back Pressure Valve
malfunction
gas mixer 《Service Guide for A7 Old Gas Mixer Assembly and
assembly Just change gas mixer assembly
malfunction Old Total Flowmeter Assembly》 which will send
After changing
gas mixer
assembly
silkscreen “122”
in the picture
should change to
“58”
Figure5 gas mixer assembly
“122”
O2 needle valve Change ORC service kit totally, Change O2 needle valve only,do ORC
malfunction calibration is unnecessary
calibration(follow service manual)
N2O needle valve Change ORC service kit totally, Change N2O needle valve only,do ORC
malfunction calibration is unnecessary
calibration(follow service manual)
7. Decrease the output pressure of high-pressure regulator at the cylinder gas supply
from 400 ±40KPa to 360 ±40KPa for solving the malfunction: when cylinder gas supply
and pipeline gas supply were switched on simultaneously, since the cylinder side has
higher pressure, the anesthesia machine used the gas inside the cylinder first. This
spare part do not change the order number, installation method is the same with old
one.
8. Tube color in the Vacuum suction assembly has changed from orange to transparent.
Compatibility Table
A7(0635) Before After Compatibility
0632 Flowmeter O2 flow sensor board
051-001414-01 051-001414-01 YES
PCBA
0632 Flowmeter Air sensor board PCBA 051-001415-01 051-001415-01 YES
0632 Flowmeter N2O sensor board PCBA 051-001416-01 051-001416-01 YES
needle valve assembly(0632) 115-014468-00 115-014468-00 YES
O2 cylinder assembly (0632) 115-014476-00 115-014476-01 YES
N2O cylinder assembly (0632) 115-014477-00 115-014477-01 YES
AIR cylinder assembly (0632) 115-014478-00 115-014478-01 YES
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
No.: OMZ-21009-A Series Page 307of 428
PMLS Global Technical Support Dept
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14032-A5
Technical News
Subject: ECL090-F Technical Changes
RECORD IDENTIFIER
CONTENT
Vacuum suction
tube fixing clamp
5. 3 new ORC service kits are added now to solve a calibration must be down with N2O
after replacement.
Parts Info as below(Table 1)
Part description P/N content
O2 needle valve Change ORC service kit totally, Change O2 needle valve only,do ORC
malfunction calibration is unnecessary
calibration(follow service manual)
N2O needle valve Change ORC service kit totally, Change N2O needle valve only,do ORC
malfunction calibration is unnecessary
calibration(follow service manual)
Compatibility Table
A5 (0634) Before After Compatibility
0631 top lighting board PCBA 801-0631-00039-00 801-0631-00039-00 Yes
gas bench receptacle assembly 115-016480-00 115-016480-00 Yes
0631 Power Board PCBA 801-0631-00025-00 801-0631-00025-00 Yes
No.: OMZ-21009-A Series Page 312of 428
PMLS Global Technical Support Dept
Fresh Flow Sensor Board 801-0631-00040-00 801-0631-00040-00 Yes
Brake Assembly(FRU) 115-016692-00 115-016692-00 Yes
work surface Yes
801-0631-00120-00 801-0631-00120-00
assembly(0631)FRU
ORC Assembly 115-003009-01 115-003009-01 Yes
ORC Maintenance Kit(CE Yes
115-017864-00 115-017864-00
Common) FRU
Mid-plate FRU(0631/no Yes
115-021249-00 115-021249-00
ventilation valve)
Breathing System 801-0631-00105-00 801-0631-00105-00 Yes
Breathing Circuit Unit 801-0631-00103-00 801-0631-00103-00 Yes
Breathing System FRU(0631) 115-017036-00 115-017036-00 Yes
gas mixer assembly(0632) 115-016781-00 115-027532-00 Yes
Drive Assembly(New Yes
115-020937-00 115-020937-00
Diff_pressure)FRU
Ventilation kit(New Yes
115-018148-01 115-018148-01
Diff_pressure)FRU
Rear Door 801-0631-00037-00 115-026809-00 Yes
Check valve dome, A series 801-0631-00061-00 801-0631-00061-00 Yes
upper cover assembly(0631) 115-006544-00 115-006544-00 Yes
Back Pressure Valve 801-0631-00042-00 115-028145-00 Yes
Main control board Yes
115-018146-00 115-026808-00
FRU(SFDA)
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14024-A7
Technical News
Subject: ECL092 Technical Change
RECORD IDENTIFIER
CONTENT
Compatibility Table:
A7(0635) Old Version New Version Compatible
O2 pipeline pressure
115-023193-00 115-024847-00 Yes
gauges(FRU/0632)
N2O pipeline pressure
115-023191-00 115-024848-00 Yes
gauges(FRU/0632)
Air pipeline pressure
115-023192-00 115-024849-00 Yes
gauges(FRU/0632)
O2 tank pressure
115-023194-00 115-024850-00 Yes
gauge(FRU/0632)
N2O tank pressure
115-023195-00 115-024851-00 Yes
gauge(FRU/0632)
Air tank pressure
115-023196-00 115-024852-00 Yes
gauge(FRU/0632)
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14023-A5
Technical News
Subject: ECL092 Technical Change
RECORD IDENTIFIER
CONTENT
Compatibility Table:
A5(0634) Old Version New Version Compatible
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14022-A7
Technical News
Subject: EFU003 Technical Changes
RECORD IDENTIFIER
CONTENT
New Configurations:
1. New configuration for A5 and A7 is launched with only O2 and Air gas inlet.
Also the yoke configuration (with extra gas inlet from a cylinder) is available for A5 and A7.
2. Air-driven A7 is available.
4. Side-stream CO2 modules are available on A5 and A7 with AM software version 02.05.00 or above.
After upgrade, please enter【Setup】>【Service】>558188>【Factory Setup】>Select CO2 module
in this menu.
M02C CO2
Module
11. The correlation between trend table, trend graph and event review is added:
In【History】
✓ Customer can choose the event in trend table, and when switch to event log, this event
will be displayed directly.
✓ Customer can choose the event in trend graphic and it will switch to event log directly for
details.
20. Optimize the PCV waveform in Demo mode,and synchronize the settings and reading values.
21. New notification for ACGO and Monitor mode,and disable the Loops in ACGO and Monitor mode.
22. When External AG module is in use, AA,O2 and N2O waveform can be displayed.
23. Optimizer only indicates for 1-4L/min.
24. Software change to correct the intermittent restart when A5 connects into network.
25. Solve the problem that HL7 protocol in Bundle Version 02.04.00 is not compatible with 01.05.02
26. Optimize the AG zeroing when discharge patient, to correct the intermittent “Internal AG Error 07”
alarm.
27. Optimize software to eliminate the “Bundle Version Time Out ” message when EFCS cannot pass
selftest.
28. Optimize EFCS zeroing.
✓ The range for zero point is changed from[-0.12LPM,0.12LPM] to [-0.20LPM,0.20LPM]
✓ Add zeroing for total flow sensor in EFCS.
29. Optimize the EFCS leakage test in Selftest.
✓ When gas pressure is not enough, there will be no selftest for relevant gas limb in EFCS.
✓ To avoid a zero drift that may cause EFCS leakage test fail:A selftest will first be applied
for EFCS limbs, if one of the limbs has a flow reading out of zero point, then an auto
zeroing will be applied and do the selftest again. This time if it still fails, then it means
there is really a leakage for that limb, and alarms will display.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14021-A5
Technical News
Subject: EFU003 Technical Changes
RECORD IDENTIFIER
CONTENT
New Configurations:
1. New configuration for A5 and A7 is launched with only O2 and Air gas inlet.
Also the yoke configuration (with extra gas inlet from a cylinder) is available for A5 and A7.
Software Changes:
New Software is launched:
M02C CO2
Module
4. Optimize the color display for different gas. Make the color consistent in flowmeter, AG modules
and trend graphic.
5. The default patient type is set to Adult.
6. P-F loop is added.
7. The allowed fresh gas flow leakage is changed to 0.15L/Min during leakage test. In another word,
during leakage test, all the fresh gas should be set to zero, but actually a 0.15L/Min leakge of fresh
gas in total is allowed.
8. The AG module sampling rate is reset as:
Adult water trap: High 200 ml/min(Factory Default), Medium 180 ml/min, Low 150 ml/min;
Neo water trap: High 120 ml/min, Medium 110 ml/min, Low 100 ml/min;
9. The correlation between trend table, trend graph and event review is added:
In【History】
✓ Customer can choose the event in trend table, and when switch to event log, this event
10. O2 sensor monitoring is cancelled when ventilation mode is dependent ACGO, and all the FiO2
alarms are cancelled.
11. When supplied by battery only, there will be message after self-test.
12. When the patient is less than 1 year old,the age will be displayed as“<1”
13. To set the parameters, fresh gas flow, alarm limit or CO2 settings, you can press the UP/DOWN
arrows for long time to have a fast setting.
16. Optimize the PCV waveform in Demo mode,and synchronize the settings and reading values.
17. New notification for ACGO and Monitor mode,and disable the Loops in ACGO and Monitor mode.
18. When External AG module is in use, AA,O2 and N2O waveform can be displayed.
19. Software change to correct the intermittent restart when A5 connects into network.
No.: OMZ-21009-A Series Page 328of 428
PMLS Global Technical Support Dept
20. Solve the problem that HL7 protocol in Bundle Version 02.04.00 is not compatible with 01.05.02
21. Optimize the AG zeroing when discharge patient, to correct the intermittent “Internal AG Error 07”
alarm.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14020-A7
Technical News
RECORD IDENTIFIER
CONTENT
1.Spare Parts:
No. PN Description Quantity
1 TCN13-FB009-001 Tube connection(A7/CE/E_ACGO/AIR DRIVE) 1
2 TCN13-FB009-002 A7 Operator's manual(CE) 1
Tube List
Tube No. PN Description Length(mm)
36 082-000517-00 tube.PU (polyether) 5.5mmX8mm orange 100
61 082-000517-00 tube.PU (polyether) 5.5mmX8mm orange 70
63 082-000517-00 tube.PU (polyether) 5.5mmX8mm orange 330
2.Procedures:
No.: OMZ-21009-A Series Page 330of 428
PMLS Global Technical Support Dept
1. Open the service door, remove the rear panel assembly. Remove the work surface cover plate
2. Unplug Tube 36#、61#、63# and disposal them. Keep the Y-connectors.
3. Connect Tube 60# to the Y-connector that links Tube 38# and 69# (Refer to the notifications in red
below).
Note:Serial mode settings is for calibration. For measurement (flow or pressure), please exit
serial settings:【SETUP】→【Syestem】→【ENTER】→Seral Mode:NONE
2. Confirm Driving Gas Type
For Air-driven, enter setup→service→558188→Accept. Switch Driver Gas to Air
Tube-Checking:Locate the tube to the expiration valve assembly gas input, check the color and
type: Driving gas is air and the tube color is orange.
Preparations:
To ensure the calibration time accuracy, make sure the system time
NOTE: Driving Gas:For air driven AM, the calibration gas should be
compressed air.
Connect gas supplies and regulate the gas inlet pressure to 400kPa±10kPa;
Preparations:Enter “standby”,Set Auto/Manual switch to “Auto” position,Remove bellows
from bellows mount and reinstall bellows cover, remove water trap. Turn off the fresh gas. Connect
the inlet port of VT-PLUS to inspiration port, outlet port the expiration port, set to High flow channel
in advance.
Check Flow Sensor Zero:Enter setup→service→(789789)→Accept→Data Monitors. Ensure
the “Int-Flow Sensor” AD value:100~400,“Insp Flow Sensor” and “Exp Flow Sensor”:200~
1400;
Check Inspiration Valve DA Value:Enter setup→service→(789789)→Accept→Diagnostic
Tests,Set PEEP Valve D/A to “1700”, Safety Valve To “on”,Set Insp Valve D/A to 2300 and
meantime check VT-plus flow reading should be over 90L/min;
Connect the calibration device to the anesthesia machine :Connect VT-plus to AM via
calibration connector board (fig below) to RS232 port in AM or the J7 plug on monitor board.
Keep the tube connection same as flow calibration, just remove the communication cable between
VT-plus and the anesthesia machine.
Enter standby, Setup→service→(789789)→Accept→Diagnostic Tests,Set PEEP Valve D/A to
“1900”, Safety Valve “on”;
Set Insp Valve D/A to 0 ,Make sure that the deviation between the measured data of the
inspiratory flow sensor, expiratory flow sensor and ventilator flow sensor and that of the anesthesia
machine calibration device must not exceed 0.5L/min
Increase Insp Valve Flow or increase Insp Valve D/A , until the VT-plus reading reaches
5±0.5L/min,10±0.5 L/min , 20±0.5L/min , 30±1L/min , 50±1 L/min , 70±2 L/min and 90±2
L/min,
Make sure that the deviation between the measured data of the inspiratory flow sensor, expiratory
flow sensor and ventilator flow sensor and that of the anesthesia machine calibration device must not
exceed 1L/min or 5%of the measured value of the calibration device, whichever is greater.
NOTE:
1)In constant flow test, VT-plus high flow channel is used.
2)Note that AIR must be used for constant flow test. And gas type on VT-plus should be selected
as AIR.
3)In flow test, VT-plus reading and flow sensor readings may be unstable, read the values when
the abbreviation is less than 0.5L/min within 2s.
Enter Standby,
【Service】>【Review logs】>【Service】 ,check there is the record: Flowmeter Zero:
Pass; and PAW and Flow Zero: Pass.
Enter 【Alarms】>【Clear Log】, clear the alarm logs.
Appendix:
错误!链接无效。
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14019-A7
Technical News
Subject: EFU002 Technical Changes
RECORD IDENTIFIER
CONTENT
1. Support arm and cord collector with silica gel piece are available for A series.
PN Description
115-024056-00 cord collector kit
Fix-adaptor
2. A series for pendant are available. THIS IS NOT ABLE TO UPGRADE ON-SITE.
New T-connector
Original T-connector(Still Used on WATO&A series)
(ONLY USED ON PENDENT A SERIES)
PN Description Comments PN Description Comments
115-024109- Reserve Oxygen
NIST
oxygen supply T- 00 Supply(0634/NIST)
115-017723-00 NIST
connection package 115-024163- Reserve Oxygen
DISS
00 Supply(0634/DISS)
4. Improve the tightness of vacuum suction knobs. Add icons for vacuum suction.
6. Upgrade Manuals.
⚫ Upgrade A7 Manuals: Netherland,French,Portugal,Spanish, Turkey, Russian
Before After
Order No. Description Order No. Description
801-0631-00084- Preventative Maintenance
Preventative Maint kits (1 yr) 115-024063-00
00 kits(1 year/Non FDA)
801-0631-00085- Preventative Maintenance
Preventative Maintenance(3 yr) 115-024064-00
00 kits(3 year/Non FDA)
Compatibility Table:
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
No.: OMZ-21009-A Series Page 342of 428
PMLS Global Technical Support Dept
TMZ-14018-A5
Technical News
Subject: EFU002 Technical Changes
RECORD IDENTIFIER
CONTENT
1. Support arm and cord collector with silica gel piece are available for A series.
PN Description
115-024056-00 cord collector kit
Fix-adaptor
2. A series for pendant are available. THIS IS NOT ABLE TO UPGRADE ON-SITE.
New T-connector
Original T-connector(Still Used on WATO&A series)
(ONLY USED ON PENDENT A SERIES)
PN Description Comments PN Description Comments
115-024109- Reserve Oxygen
NIST
oxygen supply T- 00 Supply(0634/NIST)
115-017723-00 NIST
connection package 115-024163- Reserve Oxygen
DISS
00 Supply(0634/DISS)
5. Upgrade Manuals.
a) New release A5 Manuals, check list and reference card: Bulgaria, Denmark, Finland,
Romania
b) Upgrade A5 Manuals: Netherland,French, Portugal ,Spanish, Turkey, Russian, Czech,
German, Italy, Norway, Poland, Serbia, Swedish, Hungary.
Before After
Order No. Description Order No. Description
Compatibility Table:
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14002-A7
Technical News
Subject: ECL089 Technical changes
RECORD IDENTIFIER
CONTENT
Mechanical Changes:
1. Add hook onto the O2 sensor protective bar
Before After
5. Main board material changes to comply with the requirement of RoHS. The order number keeps the
same.
Software:
FDA region ONLY:
Upgrade software to solve the problems:
1. The HL7 protocol in Bundle version 02.04.00(FDA) is not compatible with the HL7 protocol in Bundle
version 01.05.02.
No. Package Content Latest Version
BundleVersion V02.04.01
Host Software V03.04.01
Boot Software V01.11.00
Bundle
1 FPGA V1.4
(V02.04.01)
VCM V01.14.00
VPM V1.1
E flowmeter V1.8
2 Power(V1.7) Power board V1.7
Maintenance kit:
One-year maintenance kit:Add three seals for cylinder yoke
Three-year maintenance kit:Add three seals for cylinder yoke. Add one more O-ring 082-
000673-00 (can be used on bag arm or O2 sensor cover). Add O-ring 082-001504-00 for
water trap.
Others:
1. The lock is replaced with new material to comply with RoHS.( 115-023320-00 )
Compatibility Table:
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-14001-A5
Technical News
Subject: ECL089 Technical changes
RECORD IDENTIFIER
CONTENT
Mechanical Changes:
1. Add hook onto the O2 sensor protective bar
Before After
4. Main board material changes to comply with the requirement of RoHS. The order number keeps the
No.: OMZ-21009-A Series Page 355of 428
PMLS Global Technical Support Dept
same.
Software:
FDA region ONLY:
Upgrade software to solve the problems:
1. The HL7 protocol in Bundle version 02.04.00(FDA) is not compatible with the HL7 protocol in Bundle
version 01.05.02.
2. A5(FDA) unit with module rack can not trigger the alarm of “O2 Sensor Disconnected”
No. Package Content Latest Version
BundleVersion V02.04.01
Host Software V03.04.01
Boot Software V01.11.00
Bundle
1 FPGA V1.4
(V02.04.01)
VCM V01.14.00
VPM V1.1
E flowmeter V1.8
2 Power(V1.7) Power board V1.7
Maintenance kit:
One-year maintenance kit:Add three seals for cylinder yoke
Three-year maintenance kit:Add three seals for cylinder yoke. Add one more O-ring 082-
000673-00 (can be used on bag arm or O2 sensor cover). Add O-ring 082-001504-00 for
water trap.
Others:
1. The lock is replaced with new material to comply with RoHS.( 115-023320-00 )
Compatibility Table:
A5 (0634) Before After Compatibility
Mid-plate FRU(0631/no ventilation
115-021249-00 115-021249-00 Yes
valve)
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-13013-A7
Technical News
Subject: A series Factory Code Change
RECORD IDENTIFIER
CONTENT
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-13012-A5
Technical News
Subject: A series Factory Code Change
RECORD IDENTIFIER
CONTENT
According to the strategy, Factory Code of A5 and A7 in CE region (except Canada) will be changed
to get distinguished from Mahwah.
Anesthesia A7 A5
Product Before EFU001 After EFU001 Before ECL081 After ECL081
International (CE) 0632 0635 0631 0634
Canada (CE) 0632 0632 0631 0631
1)The change time will be 2013.12.25
2)From 2014.1.1, the orders from International(CE, except Canada)will be available for
0634B and 0635B. 0631B and 0632B will be disabled. i, e. Factory code of A7 will be changed to
0635, factory code of A5 will be changed to 0634 (except Canada).
3)From 2014.1.1,0631B and 0632B are only available for Canada, i, e, Factory code of A5
and A7 remain the same for Canada Region.
4)The factory code on SAP will be changed automatically, when you create a note for service,
there is no need to make manual change for factory code.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours
Customer Service Departments
Mindray Co., Ltd.
TMZ-13011-A7
Technical News
Subject: Software, pneumatic circuit, hardware and manuals upgrade for A series.
RECORD IDENTIFIER
CONTENT
1. Software
⚫ New software package of A7, please note the all the software in the package
should be upgraded.
A7:
BundleVersion: 02.04.00
Power:V1.7
EFCS:V01.03.00
FpgaEFCS:3.0
SystemUpdateTool:4.9
⚫ The 02.04.00 bundle version supports all the old HL7 protocols. In [Setup->
System->Set HL7 Compatibility]
Here is the list of Bundle version of software and related HL7 protocol version.
Bundle Version HL7 Protocol Version
01.00.00
01.00.01
01.01.00
01.02.00
01.02.01 01.00.00~01.05.00
01.03.00
01.04.00
01.05.00
01.05.01
01.05.02 01.05.02
02.00.00 02.00.00
02.01.00
02.02.00
02.02.01
Most Recent
02.02.02
02.03.00
02.04.00
⚫ Buzzer will buzz for 10 seconds once the main board has an abnormal reset.
⚫ For FDA standard, A series machines with AG module plus O2 function, the
⚫ When the alarm is silenced, if the current alarm is low class, then the alarm will
only reoccur when there is a new one; if the alarm is middle or high level, then silence
for 120s.
⚫ Apnea CO2 and Apnea Alarm are set to be independent from each other.
⚫ In Service Menu, there newly added FCS Test, including Manual test, Auto Test,
BFCS Test, Solenoid Actuator Test.
Manual Test: When 3-way valve of CPU and FPGA of EFCS monitor board are both
set to H, then it is in EFCS status, otherwise it will be BFCS status.
Manual Test: In EFCS status, you can test the accuracy of each limb (O2, N2O, Air)
by comparing to the total flow reading.
Auto Test: Automatically test the three-way valves and accuracy of flow sensor.
BFCS Test: Manual test for BFCS. Click “Deploy BFCS” to deploy it, and “Retract
BFCS” to retract it. Four position switch status also displays.
Solenoid Actuator Test: Click “Test Solenoid Actuator” button, the solenoid “ON/OFF”
status and relative AD count and Actual value will be checked if they are within range.
⚫ In [Service-> Factory Setup (558188)], there newly added AG module and BIS
module selection.
⚫ Solve the problem of Wrong User Calibration Time. If a wrong user calibration
time is identified on A7, please upgrade to the bundle 02.04.00, and do user
calibration again.
⚫ When the battery is low, the BFCS will be deployed automatically. And can only
be enabled after the AC power is connected.
⚫ When Serial Protocol is set to “None”, the SNTP Protocol (Interval) will be “Off”.
⚫ In the menus, there newly added “To Top” and “To Bottom” icons.
2. Pneumatic Circuit
⚫ There newly added the indication of flow direction and maximum pressure in
ACGO port.
⚫ A7 in CE region has launched air-driven venturi vacuum suction.
⚫ The venture flowmeter in ventilator module has been changed to another shape.
Old
flowmeter
New
flowmeter
⚫ Vaporizer manifold for A5 has been changed to A7, and the sampling of AG
calculation is in ACGO assembly.
⚫ The O2 gas inlet assembly of A5 changes to the same as A7. In other words, A5
and A7 share the same O2 gas inlet assembly. For Air and N2O inlet assembly, A5
share the same DISS connector as A7.
3. Hardware
⚫ A series occasionally has black screen after fast switch off then switch on. The
inverter board was improved to prevent this issue.
⚫ The flow sensors in A series have been upgraded. In A7: three flow sensors of
EFCS and the flow sensor in ventilator have been upgraded.
O2, Air and N2O flow sensors of EFCS have been upgraded to improve the accuracy
from 5% to 3%.
Flow sensors for venturi flowmeter has been upgraded to improve Electro Magnetic
Compatibility
4. Others
⚫ A new hook that can be freely installed in the GCX rail.
⚫ For FDA region, a GCX rail with the gap in the middle.
BEFORE AFTER
⚫ There newly added the tubes indication in service door spare part for replacement.
Item New
P/N Descriptions Old Ver.
No. Ver.
Manual
1 046-004559-00 A7 Operator's Manual(CE) 4.0 5.0
2 046-005034-00 A7 Operator's Manual(Dutch) / 1.0
3 046-005035-00 A7 Operator's Manual(French) / 1.0
4 046-005037-00 A7 Operator's Manual(Portuguese) / 1.0
5 046-005038-00 A7 Operator's Manual(Spanish) / 1.0
6 046-005039-00 A7 Operator's Manual(Turkish) / 1.0
7 046-005040-00 A7 Operator's Manual(Russian) / 1.0
8 046-005049-00 A7 checkout list(Dutch) / 1.0
9 046-005050-00 A7 checkout list(French) / 1.0
10 046-005051-00 A7 checkout list(Portuguese) / 1.0
11 046-005052-00 A7 checkout list(Spanish) / 1.0
12 046-005053-00 A7 checkout list(Turkish) / 1.0
13 046-005054-00 A7 checkout list(Russian) / 1.0
Some changing units are not compatible with the old ones. Below is coordination table
for your reference.
New
A7 (0632) Before Now Compatible
With Old
801-0631-00047-
Drive Assembly(New Diff_pressure)FRU 115-020937-00 Yes
00
801-0631-00137-
Main control board FRU 115-018146-00 Yes
00
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-13010-A5
Technical News
Subject: Software, pneumatic circuit, hardware and manuals upgrade for A series.
RECORD IDENTIFIER
CONTENT
No.: OMZ-21009-A Series Page 377of 428
PMLS Global Technical Support Dept
1. Software
New software package of A5, and please note that during upgrade, you have to
upgrade all the software in the package.
A5:
BundleVersion:02.04.00
Power:V1.7
SystemUpdateTool:4.9
⚫ The 02.04.00 bundle version supports all the old HL7 protocols. In [Setup->
System->Set HL7 Compatibility]
Here is the list of Bundle version of software and related HL7 protocol version.
Bundle Version HL7 Protocol Version
01.00.00
01.00.01
01.01.00
01.02.00
01.02.01 01.00.00~01.05.00
01.03.00
01.04.00
01.05.00
01.05.01
01.05.02 01.05.02
02.00.00 02.00.00
02.01.00
Most Recent
02.02.00
⚫ For FDA standard, A series machines with AG module plus O2 function, the
only reoccur when there is a new one; if the alarm is middle or high level, then silence
for 120s.
⚫ Apnea CO2 and Apnea Alarm are set to be independent from each other.
c) In [Service-> Factory Setup (558188)], there newly added AG module and BIS
module selection.
d) Solve the problem of Wrong User Calibration Time. If a wrong user calibration
time is identified, please upgrade to the bundle 02.04.00, and do user calibration
again.
f) In the menus, there newly added “To Top” and “To Bottom” icons.
2. Pneumatic Circuit
⚫ There newly added the indication of flow direction and maximum pressure in
ACGO port.
⚫ There newly added Ohmeda, P-B and Chemetron US standard pipeline:
⚫ The venture flowmeter in ventilator module has been changed to another shape.
Old
flowmeter
New
flowmeter
⚫ Vaporizer manifold for A5 has been changed to A7, and the sampling of AG
calculation is in ACGO assembly.
For replacement of old manifold, sampling tube or old ACGO assembly, please refer
to :
No.: OMZ-21009-A Series Page 383of 428
PMLS Global Technical Support Dept
Replacement
Old Part
New Spare part Order Number
Vaporizer Mounting Manifold 801-0631-00024-00
vaporizer mounting
anesthetic gas inlet tubes
manifold(AGC)(115- 115-017879-00
module(0632)
015456-00)
mechanical ACGO(AG) 115-017878-00
Anesthetic gas inlet tubes anesthetic gas inlet tubes
115-017879-00
module(115-015829-00) module(0632)
⚫ The O2 gas inlet assembly of A5 changes to the same as A7. In other words, A5
and A7 share the same O2 gas inlet assembly. For Air and N2O inlet assembly, A5
share the same DISS connector as A7.
3. Hardware
⚫ A series occasionally has black screen after fast switch off then switch on. The
inverter board was improved to prevent this issue.
Flow sensors for venturi flowmeter has been upgraded to improve Electro Magnetic
Compatibility
4. Others
⚫ A new hook that can be freely installed in the GCX rail.
⚫ For FDA region, a GCX rail with the gap in the middle.
⚫ For CE region, A5 has changed the color of screen shell, base assembly, and the
wheel to the same as FDA and SFDA.
⚫ There newly added the tubes indication in service door spare part for replacement.
No.: OMZ-21009-A Series Page 386of 428
PMLS Global Technical Support Dept
Item New
P/N Descriptions Old Ver.
No. Ver.
Manual
1 046-005026-00 A5 Operator's Manual(Czech) / 1.0
2 046-005027-00 A5 Operator's Manual(German) / 1.0
3 046-005028-00 A5 Operator's Manual(Italian) / 1.0
4 046-005029-00 A5 Operator's Manual(Norwegian) / 1.0
5 046-005030-00 A5 Operator's Manual(Polish) / 1.0
6 046-005031-00 A5 Operator's Manual(Serbian) / 1.0
7 046-005032-00 A5 Operator's Manual(Swedish) / 1.0
8 046-005033-00 A5 Operator's Manual(Hungarian) / 1.0
9 046-005041-00 A5 checkout list(Czech) / 1.0
10 046-005042-00 A5 checkout list(German) / 1.0
11 046-005043-00 A5 checkout list(Italian) / 1.0
12 046-005044-00 A5 checkout list(Norwegian) / 1.0
13 046-005045-00 A5 checkout list(Polish) / 1.0
14 046-005046-00 A5 checkout list(Serbian) / 1.0
15 046-005047-00 A5 checkout list(Swedish) / 1.0
16 046-005048-00 A5 checkout list(Hungarian) / 1.0
17 046-005055-00 A5 Reference Card(Czech) / 1.0
18 046-005056-00 A5 Reference Card(German) / 1.0
19 046-005057-00 A5 Reference Card(Italian) / 1.0
20 046-005058-00 A5 Reference Card(Norwegian) / 1.0
21 046-005059-00 A5 Reference Card(Polish) / 1.0
22 046-005060-00 A5 Reference Card(Serbian) / 1.0
23 046-005061-00 A5 Reference Card(Swedish) / 1.0
24 046-005062-00 A5 Reference Card(Hungarian) / 1.0
Some changing units are not compatible with the old ones. Below is coordination table
for your reference.
New
A5 (0631) Before Now Compatibl
e With Old
Upper cover 2
043-001133-00 115-021252-00 Yes
FRU(0631/no popoff)
loop low cover 2
043-001135-00 115-022320-00 Yes
assembly FRU
Drive Assembly(New
801-0631-00047-00 115-020937-00 Yes
Diff_pressure)FRU
Ventilation kit(New
115-018148-00 115-018148-01 Yes
Diff_pressure)FRU
0631 Backlight
051-000176-02 051-001519-00 Yes
Inverter PCBA
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-13005-A5
Technical News
Subject: Some changes have been implemented on A5 anesthesia machine.
RECORD IDENTIFIER
CONTENT
List Trends
Event Log
2) Add “Capture Event” function. The ventilation parameters in use will be saved and
can be reviewed in History menu.
3) Adjust “Discharge” function. The user can choose whether save the history logs
and restore default settings when discharging the machine. The Clear History
function of Discharge can be switched on or off in “Setup→System” menu.
Discharge Button
Volume Waveform
2. Optimize the structure of ORC Assembly and Restrictor Assembly. Below are the
pictures of the old and the new. Please apply “Resister block assembly (PN: 801-
0631-00119-00)” for the replacement of the old Restrictor Assembly. Please apply
“ORC Maintenance Kit (CE Common) FRU (PN: 115-017864-00)” for the replacement
of the old ORC Assembly. And there is a “Maintenance Instruction for Old ORC”
contained in the “ORC Maintenance Kit (CE Common) FRU (PN: 115-017864-00)”,
No.: OMZ-21009-A Series Page 395of 428
PMLS Global Technical Support Dept
please follow the Instruction for the replacement.
3. Optimize the design of Pipeline Gas Supply Inlet Assembly and O2 Supply Three-way
Connector to eliminate the abnormal noise during ventilation. The PN doesn’t change.
Internal AG Module
Item New
FRU P/N Descriptions Old Ver.
No. Ver.
Hardware
1 801-0631-00041-00 ORC Assembly 1.0 /
AION 03-31 Multigas
2 115-002504-00 / 1.0
Analyzer
Multigas Analyzer.AION,03-
3 9200-10-10529 1.0 /
31,60-10331-00
4 115-003009-01 ORC Assembly / 1.0
5 115-008789-01 Resister block assembly / 1.0
ORC Maintenance Kit(CE
6 115-017864-00 / 1.0
Common) FRU
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-12037-A5
Technical News
Subject: Some changes have been implemented on A5 anesthesia machine.
RECORD IDENTIFIER
CONTENT
b) When the date of birth inputted is invalid, the machine will display a pop-up
message.
Choose
“Mainboard” here.
d) The alarm limits of PEAK and MV will become “OFF” when the operator choose
“OFF” in “Alarm ON/OFF” button in manual mode.
Choose “OFF”
here, the alarm
limits of “PEAK
and MV” will be
“Off”.
.
f) Change the appearance and function of “Alarm” menu. The alarm limits of
anesthetic agent, CO2 and O2 can be adjusted if an AG module is used on the
machine.
g) Change the appearance and function of “Setup” menu. Add “Gas bench Flow
Rate” in “General” menu. “Gas bench” means “Module slot” on A5. Add CO2
wave configuration in “Display” menu.
The “Pump
Rate” of AG
module can
be set here.
h) When the lever is in “Auto” position and the machine is not in “Standby”, a pop-up
message “Can only discharge in Manual Mode” will displayed if the user press
“Discharge” button.
j) Add “MR-Link” protocol in “Serial Configuration” menu. Now serial port supports
“HL7 protocol”.
k) Add a technical alarm message “Fan Failure 02”, which will be triggered if the fan
speed is less than 3640rpm.
l) If the user switches the machine from “Auto Mode” to “Manual Mode”, status of
“Alarm ON/OFF” button can be restored to the status before last exit from Manual
mode. By comparison, before this software version, the status will be
automatically set to “ON” when the lever switches from “Auto” to “Manual”.
This is “Alarm
ON/OFF” button.
2. The patient monitor bracket is now available on A5. The bracket supports 7000, 8000,
9000, 6800 and 6802 series patient monitors.
3. Module slot, which doesn’t support field upgrade, is added on A5. Mother board and
cylinder yoke are modified to support module slot.
Infrared
Communication
Board
The positions
of connectors
are different.
Cylinder yoke for module slot Cylinder yoke without module slot
4. AG module is available now. When an AG module is plugged in the module slot, it will
work automatically. The user can not turn off the AG module in software. A5 doesn’t
support BIS function. So, it only supports AG module with or without O2. The measured
parameters will be displayed at the upper-left corner of the screen.
5. Add two gas inlet assemblies for the East Europe standard cylinder configuration. The
extra inlets are NIST standard and for O2 and N2O only. It only supports A5 which is
O2 driving.
6. Agent usage calculation function is available on A5. Add internal AG module and
modify vaporizer manifold and mother board for this function. It doesn’t support field
upgrade. The agent usage will be displayed at the upper-left corner of screen in
“Standby Mode”.
Internal AG module
Sampling tube
Waste gas
scavenging
tube
AG transfer
board
Sampling tube
Waste gas
scavenging
tube
7. Negative pressure suction device, which supports field upgrade, is available on A5.
According to the pipeline inlet type and driving gas of the machine, the suction device
has four types: O2 drive/NIST, O2 drive/DISS, Air drive/NIST and Air drive/DISS.
Tee connector
Negative pressure
regulator
Gas supply
tube
Liquid collection
bottle
If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-12024-A5
Technical News
Subject: Some changes have been implemented on A5 anesthesia machine.
RECORD IDENTIFIER
CONTENT
c) Added info about sealing the sampling port and the condensate drain after failed
leak test.
e) Lever in Manual position; turning on fresh gas flow ( ≥0.02 L/min of any
individual gas) takes you out of Standby. Lever in Auto position; turning on fresh
gas flow triggers a popup: Move lever to Manual position.
f) User can only enter Standby screen when fresh gas flow is off.
Must go Auto => Manual => Standby.
g) “Automatic Ventilation Disabled” is displayed as blue technical alarm. If the
reason that it is disabled is a leak test then it will display “Automatic Ventilation
Disabled – Leak Test Failed”.
h) If the user wants to override a startup self-test or leak test failure, a confirmation
dialog has been added that explains that only Manual mode will be available.
No.: OMZ-21009-A Series Page 418of 428
PMLS Global Technical Support Dept
k) Added RS-232 setup screen in System dialog (BAUD rate, parity, stop bits,
etc.).
l) Added date of birth and weight to Demographics tab – display weight, date of
birth and calculated age.
2. Remove the button cell battery and Lithium-ion battery from periodic maintenance kit
(36 months).
Item New
FRU P/N Descriptions Old Ver.
No. Ver.
Hardware
Periodic Maintenance kit (3
1 801-0631-00085-00 1.0 2.0
years)
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
TMZ-12016-A5
Technical News
Subject: Some changes have been implemented on A5 anesthesia machine
RECORD IDENTIFIER
CONTENT
1. Mindray V60 Halothane Vaporizer has been added to the product configuration.
Below is the list of vaporizers suitable for A5 anesthesia machine.
Filling system
Brand Anesthetic agent
Key Filler Pour Fill Quik Fil
Enflurane Yes Yes No
Isoflurane Yes Yes No
Penlon
Sevoflurane Yes Yes Yes
Halothane Yes Yes No
Enflurane Yes Yes No
Mindray Isoflurane Yes Yes No
V60 Sevoflurane Yes Yes Yes
Halothane Yes Yes No
Manuals
1 N/A N/A N/A N/A
Item New
FRU P/N Descriptions Old Ver.
No. Ver.
Hardware
Halothane Vaporizer(Key
1 115-014138-00 N/A 1.0
Filler) bag(int)
Halothane Vaporizer(Pour
2 115-014139-00 N/A 1.0
Fill) bag(int)
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
TMZ-12009-A5
Technical News
Subject: Some software upgrade of A5 anesthesia machine.
RECORD IDENTIFIER
CONTENT
Fig 1
Fig 2
2) Support Multilanguage version. Add five Multilanguage (Dutch, French,
Portuguese, Spanish, Turkish) options of language button. User can choose the
language that required.
Item New
FRU P/N Descriptions Old Ver.
No. Ver.
Hardware
1 N/A N/A N/A N/A
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Sincerely yours,
Customer Service Departments