PMLS Global Technical Support Department

Technical News Summary


Version 2.0

Release V80 Desflurane Vaporizer

Version till (TMZ-22005-WATO EX-55&65&55Pro&65Pro&75&85)


PMLS Global Technical Support Dept

Index
TMZ-22005-WATO EX-55&65&55Pro&65Pro&75&85 ......................................................................................... 1
Release V80 Desflurane Vaporizer .................................................................................................................... 1
TMZ-21062- AM all series ........................................................................................................................................ 4
Temporary change of the flow sensor in the anesthesia machine ...................................................................... 4
TMZ-21059-WATO EX-55&65 Pro&A .................................................................................................................... 7
WATO EX-55Pro(2526)&65 Pro(2525)&A software upgrade .......................................................................... 7
TMZ-21045-WATO&A series ................................................................................................................................... 9
Wato &A service manual update........................................................................................................................ 9
TMZ-21056-AM all series ....................................................................................................................................... 12
Battery upgrade................................................................................................................................................ 12
TMZ-21051-WATO&A series(1.0) .......................................................................................................................... 14
Wato &A material change ................................................................................................................................ 14
TMZ-21038-WATO&A series ................................................................................................................................. 16
Wato &A material change ................................................................................................................................ 16
TMZ-21016-WATO&A series ................................................................................................................................. 18
Operator’s manual and software upgrade ........................................................................................................ 18
TMZ-20012-WATO&A ........................................................................................................................................... 23
New version of manual released ...................................................................................................................... 23
TMZ-21009- Aseries&WATOseries ........................................................................................................................ 26
Aseries&WATOseries released top shelf mounting bracket ............................................................................ 26
TMZ-21006 -All AM&Ventilator ............................................................................................................................ 28
Anesthesia Machine&Ventilator Parts List updated ........................................................................................ 28
TMZ-21001-A5&A7 ............................................................................................................................................... 29
A series’ monitor board optimized ................................................................................................................... 29
TMZ-20054 -All AM&Ventilator ............................................................................................................................ 33
Anesthesia Machine&Ventilator Parts List updated ........................................................................................ 33
TMZ-20053-A7&WATO 55&65 Pro....................................................................................................................... 34
A series and WATO series released new vacuum generator unit ..................................................................... 34
TMZ-20046-A7 ....................................................................................................................................................... 37
A7(Japan) system software V02.23.01.00 released ......................................................................................... 37
TMZ-20045-A5&A7 ............................................................................................................................................... 39
A5&A7 updated Dutch, Italian, Russian, Spanish Operator's Manual ............................................................ 39
TMZ-20039-A7 ....................................................................................................................................................... 41
A7&V60 released in Japan .............................................................................................................................. 41
TMZ-20035-Anesthesia machine&Ventilator ......................................................................................................... 45
Anesthesia machine&Ventilator package label added UDI infromation, module exhaust tube changed ........ 45
TMZ-20034-A series ............................................................................................................................................... 48
A series released new software V02.22.01.00 ................................................................................................. 48
TMZ-20026-A7 ....................................................................................................................................................... 50
Changes for Japan order 21377912.10 ............................................................................................................ 50
TMZ-20022-A5 ....................................................................................................................................................... 53
A5s released in India ....................................................................................................................................... 53
TMZ-20021-A series&WATO series ....................................................................................................................... 59

No.: OMZ-21009-A Series Page 1 of 5


PMLS Global Technical Support Dept
A series&WATO series released new monitor brackets,A series have new manual ........................................ 59
TMZ-20006-A5&A7 ............................................................................................................................................... 62
A5&A7 released in Russia............................................................................................................................... 62
TMZ-20004-WATO&A series ................................................................................................................................. 65
New nylon Breathing Circuits are released for India....................................................................................... 65
TMZ-19036-A5&A7&WATO ................................................................................................................................. 69
A series anesthesia machine knob change and anesthesia machine electronic pressure sensor change........... 69
TMZ-19035-A5&A7 ............................................................................................................................................... 76
A series anesthesia machine software V02.22.00.01 release ........................................................................... 76
TMZ-19028-V60 ..................................................................................................................................................... 79
V60 vaporizer filler cap’s optimization ........................................................................................................... 79
TMZ-19020- All AM&Ventilator ............................................................................................................................ 81
VT650/VT900 Gas Flow Analyzer release ...................................................................................................... 81
TMZ-19017- Anesthesia Machine&Ventilator all products..................................................................................... 83
Delete the company address from Anesthesia Machine&Ventilator all products’ quick guide and maintenance
guide ................................................................................................................................................................ 83
TMZ-19015-AllAM&Ventilator .............................................................................................................................. 85
Anesthesia Machine&Ventilator Parts List is released .................................................................................... 85
TMZ-19014-FPM .................................................................................................................................................... 87
Flow Pressure Measurement (FPM) adds pressure sampling port and changes the pressure sampling mode . 87
TMZ-19010-All Anesthesia machine ...................................................................................................................... 90
Anesthesia machine ACGO assembly change ................................................................................................. 90
TMZ-19004-All AM&Ventilator ............................................................................................................................. 93
Anesthesia Machine&Ventilator Parts List is released .................................................................................... 93
TMZ-18036-A7 ....................................................................................................................................................... 94
A7 anesthesia machine hardware change ........................................................................................................ 94
TMZ-18035-A5 ....................................................................................................................................................... 96
A5 anesthesia machine hardware change ........................................................................................................ 96
TMZ-18031 All Anesthesia ..................................................................................................................................... 98
Anesthesia machine hardware upgrade and service manual update ................................................................ 98
TMZ-18004- A series&WATO EX-20&20Vet&25 ............................................................................................... 101
Anesthesia machine’s Ventilation&Control PCBA coating increase and user manual update ...................... 101
TMZ-18002-All Anesthesia Models ...................................................................................................................... 108
New monitor board with Protection used for India market prolong service time .......................................... 108
TMZ-17032-A5&A7 ..............................................................................................................................................110
Display change and new software is released .................................................................................................110
TMZ-17031-All AM&Ventilator ............................................................................................................................114
Lithium battery discharge to 30% to meet the requirement of air transport ...................................................114
TMZ-17030-A5&A7 ..............................................................................................................................................117
A5&A7 manual bag arm changes ...................................................................................................................117
TMZ-17028-All Anesthesia Models ...................................................................................................................... 120
New monitor board with Protection had been released for India .................................................................. 120
TMZ-17027-All AM&Ventilator ........................................................................................................................... 122
Anesthesia Machine&Ventilator Parts List is released .................................................................................. 122
TMZ-17021-A7 ..................................................................................................................................................... 124

No.: OMZ-21009-A Series Page 2 of 5


PMLS Global Technical Support Dept
A7 New Software released & APL valve change .......................................................................................... 124
TMZ-17020-A5 ..................................................................................................................................................... 127
A5 New Software released & APL valve change .......................................................................................... 127
TMZ-17014-WATO&A series ............................................................................................................................... 130
The measuring mode of Flow and Volume for WATO and A series anestheisa machines is changed from ATPD
to BTPS.......................................................................................................................................................... 130
TMZ-17012-A7 ..................................................................................................................................................... 132
A7 New Software released ............................................................................................................................ 132
TMZ-17011-A5 ..................................................................................................................................................... 136
A5 New Software released ............................................................................................................................ 136
TMZ-17010-All AM&Ventilator ........................................................................................................................... 139
Anesthesia Machine&Ventilator Parts List is released .................................................................................. 139
TMZ-17008-WATO EX-35&55&65&55 Pro&65 Pro&A5&A7 .......................................................................... 141
Electronic gas supply pressure monitoring released for WATO EX-55 PRO/65 PRO (0628/0627) and new
bundle software released in ECL134 ............................................................................................................. 141
TMZ-16089-A7 ..................................................................................................................................................... 147
ECL128 Engineering changes ....................................................................................................................... 147
TMZ-16088-A5 ..................................................................................................................................................... 154
ECL128 Engineering changes ....................................................................................................................... 154
TMZ-16068-A7 ..................................................................................................................................................... 160
ECL124F Technical Changes ........................................................................................................................ 160
TMZ-16067-A5 ..................................................................................................................................................... 167
ECL124F Technical Changes ........................................................................................................................ 167
TMZ-16060-All AM&Ventilator ........................................................................................................................... 173
Anesthesia Machine&Ventilator Parts List is released .................................................................................. 173
TMZ-16055-Vaporizers(V60&Penlon) ................................................................................................................. 175
New maintenance packages for V60 and new adapter for vaporizers are released(ECL120) .................. 175
TMZ-16046-A7 (0635).......................................................................................................................................... 179
A7 (0635) Product Change (ECL117) ........................................................................................................... 179
TMZ-16045-A5(0634)........................................................................................................................................... 183
A5 (0634) Product Change (ECL117) ........................................................................................................... 183
TMZ-16044-A7 ..................................................................................................................................................... 187
A7 new software is launched ......................................................................................................................... 187
TMZ-16043-A5 ..................................................................................................................................................... 192
A5 new software is launched ......................................................................................................................... 192
TMZ-16038-A7 ..................................................................................................................................................... 196
Anesthesia Prediction Function upgrade kit is released for A7(TCN16-FU002) .......................................... 196
TMZ-16016-A7 ..................................................................................................................................................... 198
ECL115 Technical Changes ........................................................................................................................... 198
TMZ-16015-A5 ..................................................................................................................................................... 210
ECL115 Technical Changes ........................................................................................................................... 210
TMZ-16004-A7 ..................................................................................................................................................... 222
ECL108-F Technical Changes ....................................................................................................................... 222
TMZ-16003-A5 ..................................................................................................................................................... 226
ECL108-F Technical Changes ....................................................................................................................... 226

No.: OMZ-21009-A Series Page 3 of 5


PMLS Global Technical Support Dept
TMZ-15050-A7 ..................................................................................................................................................... 230
EFU009 Technical Changes........................................................................................................................... 230
TMZ-15049-A5 ..................................................................................................................................................... 245
EFU009 Technical Changes........................................................................................................................... 245
TMZ-15044-A7 ..................................................................................................................................................... 256
ECL112 Technical Changes ........................................................................................................................... 256
TMZ-15043-A5 ..................................................................................................................................................... 264
ECL112 Technical Changes ........................................................................................................................... 264
TMZ-15016-A7 ..................................................................................................................................................... 270
EFU006 Technical Changes........................................................................................................................... 270
TMZ-15015-A5 ..................................................................................................................................................... 278
EFU006 Technical Changes........................................................................................................................... 278
TMZ-14047-A7 ..................................................................................................................................................... 285
ECL096-F Technical Changes ....................................................................................................................... 285
TMZ-14035-A7 ..................................................................................................................................................... 289
EFU004 Technical Changes........................................................................................................................... 289
TMZ-14034-A5 ..................................................................................................................................................... 296
EFU004 Technical Changes........................................................................................................................... 296
TMZ-14033-A7 ..................................................................................................................................................... 301
ECL090-F Technical Changes ....................................................................................................................... 301
TMZ-14032-A5 ..................................................................................................................................................... 309
ECL090-F Technical Changes ....................................................................................................................... 309
TMZ-14024-A7 ..................................................................................................................................................... 315
ECL092 Technical Change ............................................................................................................................ 315
TMZ-14023-A5 ..................................................................................................................................................... 318
ECL092 Technical Change ............................................................................................................................ 318
TMZ-14022-A7 ..................................................................................................................................................... 321
EFU003 Technical Changes........................................................................................................................... 321
TMZ-14021-A5 ..................................................................................................................................................... 326
EFU003 Technical Changes........................................................................................................................... 326
TMZ-14020-A7 ..................................................................................................................................................... 330
A7 in German(FR-43000129)Driving Gas Switch ................................................................................... 330
TMZ-14019-A7 ..................................................................................................................................................... 338
EFU002 Technical Changes........................................................................................................................... 338
TMZ-14018-A5 ..................................................................................................................................................... 344
EFU002 Technical Changes........................................................................................................................... 344
TMZ-14002-A7 ..................................................................................................................................................... 349
ECL089 Technical changes............................................................................................................................ 349
TMZ-14001-A5 ..................................................................................................................................................... 354
ECL089 Technical changes............................................................................................................................ 354
TMZ-13013-A7 ..................................................................................................................................................... 358
A series Factory Code Change ....................................................................................................................... 358
TMZ-13012-A5 ..................................................................................................................................................... 360
A series Factory Code Change ....................................................................................................................... 360
TMZ-13011-A7 ..................................................................................................................................................... 362

No.: OMZ-21009-A Series Page 4 of 5


PMLS Global Technical Support Dept
Software, pneumatic circuit, hardware and manuals upgrade for A series .................................................... 362
TMZ-13010-A5 ..................................................................................................................................................... 377
Software, pneumatic circuit, hardware and manuals upgrade for A series .................................................... 377
TMZ-13005-A5 ..................................................................................................................................................... 390
Some changes have been implemented on A5 anesthesia machine ............................................................... 390
TMZ-12037-A5 ..................................................................................................................................................... 401
Some changes have been implemented on A5 anesthesia machine ............................................................... 401
TMZ-12024-A5 ..................................................................................................................................................... 416
Some changes have been implemented on A5 anesthesia machine ............................................................... 416
TMZ-12016-A5 ..................................................................................................................................................... 423
Some changes have been implemented on A5 anesthesia machine ............................................................... 423
TMZ-12009-A5 ..................................................................................................................................................... 425
Some software upgrade of A5 anesthesia machine ........................................................................................ 425

No.: OMZ-21009-A Series Page 5 of 5


PMLS Global Technical Support Dept

Technical News
Subject: Release V80 Desflurane Vaporizer

TMZ-22005-WATO EX-55&65&55Pro&65Pro&75&85

Release V80 Desflurane Vaporizer

RECORD IDENTIFIER

Ref No.: TMZ-22005-WATO EX- Relate To:


55&65&55Pro&65Pro&75&85
Version:1.0 Software/Manual Upgrade
Author: Tornado Parts Change
Date: Jane. 26th ,2022 Appearance
Model: WATO EX- Stop Production
55&65&55Pro&65Pro&75&85
Comments:ECL281 Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


1、 The V80 Desflurane Vaporizer released in CE area, factory code number is 0607
and the picture as follow:

The following part numbers are material only for warranty maintenance.
Item Part number Part name Power cord type
No.: OMZ-21009-A Series Page 1of 428
PMLS Global Technical Support Dept

1 115-075080-00 Desflurane Vaporizer European standard

2 115-075081-00 Desflurane Vaporizer British Standard

3 115-082436-00 Desflurane Vaporizer European Standard - Turkish

4 115-082875-00 Desflurane Vaporizer Australian Standard

5 115-082882-00 Desflurane Vaporizer South African Standard

6 115-082883-00 Desflurane Vaporizer Brazil - Portuguese

7 115-082884-00 Desflurane Vaporizer Indian Standard

8 115-082885-00 Desflurane Vaporizer Korean standard


American Standard 220 V-
9 115-082887-00 Desflurane Vaporizer
Spanish
10 115-082888-00 Desflurane Vaporizer European Standard - Czech

11 115-082970-00 Desflurane Vaporizer FDA


American Standard 220 V-
12 115-082973-00 Desflurane Vaporizer
English
American Standard 110 V-
13 115-083552-00 Desflurane Vaporizer
Spanish

2、 Update V60 Russian operator’s manual, it has upload to TDS:


Item Part name TDS address
:/Anesthesia Machine-
V60_Anesthetic
Ventilator/Anesthesia Machine/V60
1 Vaporizer_Operator's
(0602)/
Manual_V2.0_RUS

3、 Update the label on the package and the main unit in Indonesia. The details are
as follows:

Main unit label:

No.: OMZ-21009-A Series Page 2of 428


PMLS Global Technical Support Dept

Package label:

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Release V80 Desflurane Vaporizer

Version Effective Date Description


1.0 Jane. 26th ,2022 Document released.

No.: OMZ-21009-A Series Page 3of 428


PMLS Global Technical Support Dept

Technical News
Subject: Temporary change of the flow sensor in the anesthesia machine

TMZ-21062- AM all series

Temporary change of the flow sensor in the anesthesia machine

RECORD IDENTIFIER

Ref No.: TMZ-21062- AM all series Relate To:


Version:1.0 Software/Manual Upgrade
Author: Tornado Parts Change
Date: Dec. 24th ,2021 Appearance
Model: AM all series Stop Production
Comments: TCN21-CL12 Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


1、 Due to the shortage of the supplier of the flow sensor, the material of the
flow sensor is changed temporarily. The picture is as follow. The part number does
not change, and the materials are completely compatible. If the original supplier
provides normally, it will replace to the original material. The materials you will
received may be those before change or those changed by TCN.

No.: OMZ-21009-A Series Page 4of 428


PMLS Global Technical Support Dept

Effect part number:


Item Part number Part description Part n

1 115-041507-00 flow sensor component flow s

2 115-041508-00 expiry flow sensor component Flow

3 115-041519-00 Flow sensor component Breat

4 115-066852-00 Breathing unit cell PM K

5 115-067256-00 PM Kit(A8, Advanced) PM K

6 115-067257-00 PM Kit(A9, Advanced)

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 5of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: Temporary change of the flow sensor in the anesthesia machine

Version Effective Date Description


1.0 Dec. 24th ,2021 Document released.

No.: OMZ-21009-A Series Page 6of 428


PMLS Global Technical Support Dept

Technical News
Subject: WATO EX-55Pro(2526)&65 Pro(2525)&A software upgrade

TMZ-21059-WATO EX-55&65 Pro&A

WATO EX-55Pro(2526)&65 Pro(2525)&A software upgrade

RECORD IDENTIFIER

Ref No.: TMZ-21059-WATO EX- Relate To:


55&65 Pro&A
Version:1.0 Software/Manual Upgrade
Author: Tornado Parts Change
Date: Dec. 3rd ,2021 Appearance
Model: WATO EX-55Pro(2526)&65 Stop Production
Pro(2525)&A
Comments:ELD005&ECL277F Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


1、The software has upgrade to new version, if you need please contact ITS to get the
software.
Item Software name Version
Wato EX-
1 05.05.02.00
55Pro(2528)_Software_V05.05.02.00_EN
Wato EX-
2 05.05.02.00
65Pro(2527)_Software_V05.05.02.00_EN

3 A3_software(DSP)_V02.24.00.02 02.24.00.02

4 A5_software(DSP)_V02.24.00.02 02.24.00.02

5 A7_software(DSP)_V02.24.00.02 02.24.00.02

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

No.: OMZ-21009-A Series Page 7of 428


PMLS Global Technical Support Dept

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: WATO EX-55Pro(2526)&65 Pro(2525)&A software upgrade

Version Effective Date Description


1.0 Dec. 3rd ,2021 Document released.

No.: OMZ-21009-A Series Page 8of 428


PMLS Global Technical Support Dept

Technical News
Subject: Wato &A service manual update

TMZ-21045-WATO&A series

Wato &A service manual update

RECORD IDENTIFIER

Ref No.: TMZ-21045-WATO&A series Relate To:


Version:1.0 Software/Manual Upgrade
Author: Tornado Parts Change
Date: Oct. 21st ,2021 Appearance
Model: WATO EX-25/35/55/65/55Pro/65Pro Stop Production
&A7&A5
Comments: EKG030F Others

CONTENT

To Whom It May Concern,

1. The service manual is upgraded to optimize the test items after installation. The test
items that have been approved by the manufacturer and do not affect the
performance during transportation are removed from the installation test list. In the
last part of the service manual, test items after installation are optimized for WATO
EX-35&A7&A5, and a test item list after installation and maintenance is added for
WATO EX-55/65/55Pro/65Pro. Take A7 as an example. The test items after
installation are as follows:

No.: OMZ-21009-A Series Page 9of 428


PMLS Global Technical Support Dept

The updated post-installation test items of A7 are as follows:

Item Part description TDS Address


A5_Service :/Anesthesia & Ventilator/Anesthesia Machine/A5
1
Manual_V8.0_EN.pdf (0631)/Manual/Service Manual/
A7_Service :/Anesthesia & Ventilator/Anesthesia
2
Manual_V7.0_EN.pdf Machine/A7(0632)/Manual/Service Manual/
:/Anesthesia & Ventilator/Anesthesia Machine/Wato-55
WATO EX-55&65 Service Pro(0626)/Manual/Service Manual/
3
Manual_EN_V6.0.pdf :/Anesthesia & Ventilator/Anesthesia Machine/Wato-65
Pro(0625)/Service Manual/
WATO EX-35&35Vet Service :/Anesthesia & Ventilator/Anesthesia Machine/Wato-
4
Manual_V8.0_EN.pdf 35(0619)/Manual/Service Manual/
:/Anesthesia & Ventilator/Anesthesia Machine/Wato-55
WATO EX-55&65 Pro_Service Pro(2528)/ Manual/Service Manual/
5
Manual_V5.0_EN.pdf :/Anesthesia & Ventilator/Anesthesia Machine/Wato-65
Pro(0627)/Manual/Service Manual/
:/Anesthesia & Ventilator/Anesthesia Machine/Wato-
WATO EX-55&65 Service 55(2526)/Service manual/
6
Manual_EN_V2.0.pdf :/Anesthesia & Ventilator/Anesthesia Machine/Wato-
65(2525)/Service manual/
:/Anesthesia & Ventilator/Anesthesia Machine/Wato-55
WATO EX-55&65 Pro Service Pro(2528)/ Service Manual/
7
Manual_EN_V2.0.pdf :/Anesthesia & Ventilator/Anesthesia Machine/Wato-65
Pro(2527)/ Service Manual/

No.: OMZ-21009-A Series Page 10of 428


PMLS Global Technical Support Dept

If you need download information or have any question, please feel free to contact Mindray regional
service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Technical News Revision


Subject: Wato &A service manual update Ref No: TMZ-21045-WATO&A series

Version Effective Date Description


1.0 Oct. 21st ,2021 Document released.

No.: OMZ-21009-A Series Page 11of 428


PMLS Global Technical Support Dept

Technical News
Subject: Battery upgrade

TMZ-21056-AM all series

Battery upgrade

RECORD IDENTIFIER

Ref No.: TMZ-21056-AM all series Relate To:


Version:1.0 Software/Manual Upgrade
Author: Tornado Parts Change
Date: Nov. 12th ,2021 Appearance
Model: AM all series Stop Production
Comments: ECL273F Others

CONTENT

To Whom It May Concern,


1. The battery add BIS certification label,the BIS label is as follow:

Two part numbers are upgrade:


Item no. Part number Part description

1 115-018011-00 Li-ion Battery Pack

Li-ion Bat Pack (11.1V4500mAh


2 115-018012-00
LI23S002A)

If you need download information or have any question, please feel free to contact Mindray regional
service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

No.: OMZ-21009-A Series Page 12of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: Battery upgrade Ref No: TMZ-21056-AM all series

Version Effective Date Description


1.0 Nov. 12th ,2021 Document released.

No.: OMZ-21009-A Series Page 13of 428


PMLS Global Technical Support Dept

Technical News
Subject: Wato &A material change

TMZ-21051-WATO&A series(1.0)

Wato &A material change

RECORD IDENTIFIER

Ref No.: TMZ-21051-WATO&A series(1.0) Relate To:


Version:1.0 Software/Manual Upgrade
Author: Tornado Parts Change
Date: Oct. 22nd ,2021 Appearance
Model: WATO EX-35(0619) Stop Production
/55/65/55Pro/65Pro/A
Comments: ECL270F Others

CONTENT

To Whom It May Concern,


1. Add a backup fan material in WATO EX-35(0619)/55/65/55Pro/65Pro/A,the two parts will coexist
in part list, The following table shows the relationship of two fans.
Before change After change Compati
Item
Part number Part description Part number Part description bility

FAN 12V
Fan&Wire(MF40
40*40*20mm Compati
1 024-000407-00 009-013196-00 201V3-1000C-
6.3CFM 18dB ble
G99)
typeF

If you need download information or have any question, please feel free to contact Mindray regional
service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Technical News Revision


Subject: Wato &A material change Ref No: TMZ-21051-WATO&A series(1.0)

No.: OMZ-21009-A Series Page 14of 428


PMLS Global Technical Support Dept
Version Effective Date Description
1.0 Oct. 22nd ,2021 Document released.

No.: OMZ-21009-A Series Page 15of 428


PMLS Global Technical Support Dept

Technical News
Subject: Wato &A material change

TMZ-21038-WATO&A series

Wato &A material change

RECORD IDENTIFIER

Ref No.: TMZ-21038-WATO&A series Relate To:


Version:1.0 Software/Manual Upgrade
Author: Tornado Parts Change
Date: Oct. 7th ,2021 Appearance
Model: Anesthesia all series Stop Production
Comments: ECL268F Others

CONTENT

To Whom It May Concern,

1. The battery part number is changed due to the battery supplier is changed. The battery has
changed from 022-000008-00 to 022-000382-00 when the system is delivered(except Kore、
Japan、India、Russia、Saudi Arabia).The battery order number is changed in CE area (except
Kore、Japan、India、Russia、Saudi Arabia、Brazil、Peru、Indonesia、Malaysia、Morocco). The
following table shows the substitution relationship of part number.

Before change After change Compati


Item
Part number Part description Part number Part description bility

Li-ion Bat Pack


Li-ion Bat Pack Compati
1 115-018012-00 (11.1V4500mAh 115-062081-00
(10.95V) ble
LI23S002A)
2. Add a backup fan material in Wato EX-20/25/30/35,the two parts will coexist in part list, The
following table shows the relationship of two fans.
Before change After change Compati
Item
Part number Part description Part number Part description bility

FAN 12V 0616


9.4CFM 20.1dB Fan&Wire(MF50 Compati
1 024-000582-00 009-013068-00
50*50*15*355m 151V3-1000C- ble
m F99)

If you need download information or have any question, please feel free to contact Mindray regional
service engineer or Mindray Service Department (email: service@mindray.com).

No.: OMZ-21009-A Series Page 16of 428


PMLS Global Technical Support Dept

Thank you for your kind attention and cooperation.

Sincerely yours,
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Technical News Revision


Subject: Wato &A material change Ref No: TMZ-21038-WATO&A series

Version Effective Date Description


1.0 Sep. 23th ,2021 Document released.
2.0 Oct. 7th ,2021 Change partnumber.

No.: OMZ-21009-A Series Page 17of 428


PMLS Global Technical Support Dept

TMZ-21016-WATO&A series

Operator’s manual and software upgrade

Technical News
Subject: Operator’s manual and software upgrade

RECORD IDENTIFIER

Ref No.: TMZ-21016-WATO&A series Relate To:


Version:1.0 Software/Manual Upgrade
Author: Tornado Parts Change
Date: Apr. 9th ,2021 Appearance
Model: WATO&A Stop Production
Comments: ECL250 Others

CONTENT

To Whom It May Concern,


1. Software of WATO has upgrade, the follows are change details:
i. Add “21%O2calibration”button in factory maintains,when is “off”, “O2 sensor has not been
calibrated for 72 hours. Would you like to calibrate now?”will not displayed.

ii. Zero optimize in wato and A series:


a) “Zeroing” will display in waveform area while AG and CO2 module zero correction

No.: OMZ-21009-A Series Page 18of 428


PMLS Global Technical Support Dept

b) In AG module the auto zeroing will stop when the following alarm trigger:Mixed Agent, Mixed
Agent and MAC ≥ 3;in CO2 module the auto zeroing will stop when the following message
alarms:EtCO2 Too High,EtCO2 Too Low,FiCO2 Too High,EtO2 Too High,EtO2 Too Low,FiO2
Too High,FiO2 Too Low.
c) Add “Null for 30s from AG Zeroing”button in AG module maintenance interface, when its “on”,
Fi/EtO2, Fi/Et CO2, Fi/EtN2O, Fi/EtAA, MAC will display 30 seconds“---”.
d) Add “Null for 30s from CO2 Zeroing”button in CO2 module maintenance interface, when its
“on”, Fi/EtO2, Fi/Et CO2 will display 30 seconds“---”
e) When the “Apnea” alarms, auto zeroing will stop for 5minutes, and the maximum prohibit time
of zeroing is less then 20minutes.
The followings are software name and version,if youneed please contact ITS get the software.
Item Software name Version

1 Wato EX-35(0619)_Software_V04.09.00.06_EN 04.09.00.06

2 Wato EX-55(0626)_Software_V04.09.00.06_EN 04.09.00.06

3 Wato EX-55Pro(0628)_Software_V04.09.00.06_EN 04.09.00.06

4 Wato EX-65(0625)_Software_V04.09.00.06_EN 04.09.00.06

5 Wato EX-65Pro(0627)_Software_V04.09.00.06_EN 04.09.00.06

6 A3_Software(DSP)_V02.23.00.05_EN 02.23.00.05

7 A5_Software(DSP)_V02.23.00.05_EN 02.23.00.05

8 A7_Software(DSP)_V02.23.00.05_EN 02.23.00.05

2. The marks of minimum change“MIN”to “SP” in APL valves in Wato EX35

No.: OMZ-21009-A Series Page 19of 428


PMLS Global Technical Support Dept

Before change after change


3. WATO&A Series operator’s manual upgrade. Add “Null for 30s from AG/CO2 Zeroing” content in
the operator’s manual. The operator’s manual has been uploaded to TDP, if you need, please
download it. The details manuals that released new version are as follow:
Item Old New
Descriptions TDP address
no. version version
:/Anesthesia
A5 Operator's Machine/Ventilator/Anesthesia
1 21.0 22.0
Manual_V22.0_EN.pdf Machine/A5
(0631&0634)/Manual/Operation Manual
:/Anesthesia
A7 Operator's
2 21.0 22.0 Machine/Ventilator/Anesthesia
Manual_V22.0_EN.pdf
Machine/A7(0632&0635)/Manual/Operat
ion Manual
:/Anesthesia
WATO EX-20 Operator's Machine/Ventilator/Anesthesia
3 7.0 8.0
Manual_V8.0_EN.pdf Machine/Wato-
20(0615)/Manual/Operation Manual
:/Anesthesia
WATO EX-30 Operator's
4 7.0 8.0 Machine/Ventilator/Anesthesia
Manual_V8.0_EN.pdf
Machine/Wato-
30(0617)/Manual/Operation
:/Anesthesia Manual
WATO EX-20 Operator's
Machine/Ventilator/Anesthesia
5 Manual(4yoke)_V6.0_E 5.0 6.0
Machine/Wato-
N.pdf
20(0615)/Manual/Operation Manual
:/Anesthesia
WATO EX-35 Operator's Machine/Ventilator/Anesthesia
6 8.0 9.0
Manual_EN_V9.0.pdf Machine/Wato-
35(0619)/Manual/Operation Manual
:/Anesthesia
WATO EX-30 Operator's
Machine/Ventilator/Anesthesia
7 Manual_EN(India)_V7.0 6.0 7.0
Machine/Wato-
.pdf
30(0617)/Manual/Operation Manual
WATO EX-10 Operator's :/Anesthesia
8 Manual_EN(India)_V5.0 5.0 6.0 Machine/Ventilator/Anesthesia
.pdf Machine/Wato-10(061B)Only For

No.: OMZ-21009-A Series Page 20of 428


PMLS Global Technical Support Dept
India/Manual/Operation Manual

:/Anesthesia
WATO EX-30 Operator's
Machine/Ventilator/Anesthesia
9 Manual_EN(Sri 2.0 3.0
Machine/Wato-
Lanka )_V3.0.pdf
30(0617)/Manual/Operation Manual
WATO EX-35 Operator's :/Anesthesia
Manual(With Machine/Ventilator/Anesthesia
10 2.0 3.0
Compressor)_V3.0_EN. Machine/Wato-
pdf 35(0619)/Manual/Operation Manual
:/Anesthesia
WATO EX-65 Pro
Machine/Ventilator/Anesthesia
11 Operator's 2.0 3.0
Machine/Wato-65
Manual_V3.0_EN.pdf
Pro(0627)/Manual/Operation Manual
:/Anesthesia
WATO EX-55 Pro
Machine/Ventilator/Anesthesia
12 Operator's 2.0 3.0
Machine/Wato-55
Manual_V3.0_EN.pdf
Pro(0628)/Manual/Operation Manual
WATO EX-65 Pro :/Anesthesia
Operator's Manual(With Machine/Ventilator/Anesthesia
13 2.0 3.0
Compressor)_V3.0_EN. Machine/Wato-65
pdf Pro(0627)/Manual/Operation Manual
WATO EX-55 Pro :/Anesthesia
Operator's Manual(With Machine/Ventilator/Anesthesia
14 2.0 3.0
Compressor)_V3.0_EN. Machine/Wato-55
pdf Pro(0628)/Manual/Operation Manual
:/Anesthesia
WATO EX-65 Operator's Machine/Ventilator/Anesthesia
15 2.0 3.0
Manual_V3.0_EN.pdf Machine/Wato-65(0625)/Operation
Manual
:/Anesthesia
WATO EX-55 Operator's Machine/Ventilator/Anesthesia
16 2.0 3.0
Manual_V3.0_EN.pdf Machine/Wato-
55(0626)/Manual/Operation Manual
WATO EX-65 Operator's :/Anesthesia
Manual(With Machine/Ventilator/Anesthesia
17 2.0 3.0
Compressor)_V3.0_EN. Machine/Wato-65(0625)/Operation
pdf Manual
WATO EX-55 Operator's :/Anesthesia
Manual(With Machine/Ventilator/Anesthesia
18 2.0 3.0
Compressor)_V3.0_EN. Machine/Wato-
pdf 55(0626)/Manual/Operation Manual

If you need download information or have any question, please feel free to contact Mindray regional

No.: OMZ-21009-A Series Page 21of 428


PMLS Global Technical Support Dept
service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Technical News Revision


Subject: Operator’s manual and software Ref No: TMZ-21016-WATO&A series
upgrade

Version Effective Date Description


1.0 Apr. 9th ,2021 Document released.

No.: OMZ-21009-A Series Page 22of 428


PMLS Global Technical Support Dept

TMZ-20012-WATO&A

New version of manual released

Technical News
Subject: New version of manual released

RECORD IDENTIFIER

Ref No.: TMZ-20012-WATO&A Relate To:


Series
Version:1.0 Software/Manual Upgrade
Author: Tornado Bian Parts Change
Date: Mar 17th, 2021 Appearance
Model: WATO&A Series Stop Production
Comments: ECL243F Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


1、 WATO&A Series operator’s manual upgrade. All the operator’s manual has
been uploaded to TDP, if you need, please download it. The details manuals that
released new version are as follow:

Item Old New


Descriptions TDP address
no. version version
:/Anesthesia
A5 Operator's Machine/Ventilator/Anesthesia
1 20.0 21.0
Manual_EN_V21.0.pdf Machine/A5
(0631&0634)/Manual/Operation Manual
:/Anesthesia
A7 Operator's
2 20.0 21.0 Machine/Ventilator/Anesthesia
Manual_EN_V21.0.pdf
Machine/A7(0632&0635)/Manual/Operat
ion Manual
:/Anesthesia
WATO EX-20 Operator's Machine/Ventilator/Anesthesia
3 6.0 7.0
Manual_EN_V7.0.pdf Machine/Wato-
20(0615)/Manual/Operation Manual
:/Anesthesia
WATO EX-30 Operator's
4 7.0 8.0 Machine/Ventilator/Anesthesia
Manual_EN_V8.0.pdf
Machine/Wato-
30(0617)/Manual/Operation Manual
No.: OMZ-21009-A Series Page 23of 428
PMLS Global Technical Support Dept
:/Anesthesia
WATO EX-20 Operator's
Machine/Ventilator/Anesthesia
5 Manual(4yoke) 4.0 5.0
Machine/Wato-
_EN_V5.0.pdf
20(0615)/Manual/Operation Manual
:/Anesthesia
WATO EX-30 Operator's
Machine/Ventilator/Anesthesia
6 Manual_EN(India)_V6.0 5.0 6.0
Machine/Wato-
.pdf
30(0617)/Manual/Operation Manual
:/Anesthesia
WATO EX-10 Operator's
Machine/Ventilator/Anesthesia
7 Manual_EN(India)_V4.0 3.0 4.0
Machine/Wato-10(061B) Only For
.pdf
India/Manual/Operation Manual
:/Anesthesia
WATO EX-30 Operator's
Machine/Ventilator/Anesthesia
8 Manual_EN (Sri Lanka) 1.0 2.0
Machine/Wato-
_V2.0.pdf
30(0617)/Manual/Operation Manual
:/Anesthesia
WATO EX-65 Pro
Machine/Ventilator/Anesthesia
9 Operator's 1.0 2.0
Machine/Wato-65 Pro
Manual_EN_V2.0.pdf
(0627)/Manual/Operation Manual
:/Anesthesia
WATO EX-55 Pro
Machine/Ventilator/Anesthesia
10 Operator's 1.0 2.0
Machine/Wato-55 Pro
Manual_EN_V2.0.pdf
(0628)/Manual/Operation Manual
WATO EX-65 Pro :/Anesthesia
Operator's Manual (With Machine/Ventilator/Anesthesia
11 1.0 2.0
Compressor) Machine/Wato-65 Pro
_EN_V2.0.pdf (0627)/Manual/Operation Manual
WATO EX-55 Pro :/Anesthesia
Operator's Manual (With Machine/Ventilator/Anesthesia
12 1.0 2.0
Compressor) Machine/Wato-55 Pro
_EN_V2.0.pdf (0628)/Manual/Operation Manual
:/Anesthesia
Machine/Ventilator/Anesthesia
Machine/Wato-65 Pro
WATO EX-55&65 Pro
(0627)/Manual/A.M. Quick Guide
13 Preoperative Checkout 1.0 2.0
:/Anesthesia
List_EN_V2.0.pdf
Machine/Ventilator/Anesthesia
Machine/Wato-55
Pro(0628)/Manual/A.M. Quick Guide
:/Anesthesia
WATO EX-65 Operator's Machine/Ventilator/Anesthesia
14 1.0 2.0
Manual_EN_V2.0.pdf Machine/Wato-65(0625)/Operation
Manual

No.: OMZ-21009-A Series Page 24of 428


PMLS Global Technical Support Dept
:/Anesthesia
WATO EX-55 Operator's Machine/Ventilator/Anesthesia
15 1.0 2.0
Manual_EN_V2.0.pdf Machine/Wato-
55(0626)/Manual/Operation Manual
:/Anesthesia
WATO EX-65 Operator's
Machine/Ventilator/Anesthesia
16 Manual(With 1.0 2.0
Machine/Wato-65(0625)/Operation
Compressor) _EN_V2.0
Manual
WATO EX-55 Operator's :/Anesthesia
Manual(With Machine/Ventilator/Anesthesia
17 1.0 2.0
Compressor) Machine/Wato-
_EN_V2.0.pdf 55(0626)/Manual/Operation Manual
:/Anesthesia
Machine/Ventilator/Anesthesia
Machine/Wato-55(0626)/Manual/A.M.
WATO EX-55&65
Quick Guide
18 Preoperative Checkout 1.0 2.0
:/Anesthesia
List_EN_V2.0.pdf
Machine/Ventilator/Anesthesia
Machine/Wato-65(0625)/A.M. Quick
Guide

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: New version of manual released Ref No: TMZ-20012-WATO&A Series

Version Effective Date Description


1.0 Mar 17th, 2021 Document released.

No.: OMZ-21009-A Series Page 25of 428


PMLS Global Technical Support Dept

TMZ-21009- Aseries&WATOseries

Aseries&WATOseries released top shelf mounting bracket

Technical News
Subject: Aseries&WATOseries released top shelf mounting bracket

RECORD IDENTIFIER

Ref No.: TMZ-21009- Aseries&WATOseries Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Feb. 4nd ,2021 Appearance
Model: Aseries&WATOseries Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,


Aseries&WATOseries released 2 top shelf mounting brackets

Fig1 top shelf mounting bracket


Part number Part name Part number Part name
Remake
before change before change after change after change
It supports
Top shelf mounting
/ / 115-070794-00 N15/N17/ePM-
for N15/17
c15/ePM-M15
It supports
Top shelf mounting N12/ePM-c10/ePM-
/ / 115-074073-00
for N12 M10/ePMc12/ePM-
M12
Table1 top shelf mounting bracket
Influences on Pre-sales & Installation
Refer to guide Anesthesia System Bracket Installation Instructions_V8.0 for more
No.: OMZ-21009-A Series Page 26of 428
PMLS Global Technical Support Dept
details, it had been uploaded to TDP, path as below
View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Accessories
Influences on Upgrade
Refer to guide Anesthesia System Bracket Installation Instructions_V8.0 for more
details, it had been uploaded to TDP, path as below
View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Accessories
Influences on Maintenance
NA

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Aseries&WATOseries released Ref No: TMZ-21009-Aseries&WATOseries
top shelf mounting bracket

Version Effective Date Description


1.0 Feb. 4nd ,2021 Document released.

No.: OMZ-21009-A Series Page 27of 428


PMLS Global Technical Support Dept

TMZ-21006 -All AM&Ventilator

Anesthesia Machine&Ventilator Parts List updated

Technical News
Subject: Anesthesia Machine&Ventilator Parts List updated

RECORD IDENTIFIER

Ref No.: TMZ-21006 -All AM&Ventilator Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: 2021-02-09 Appearance
Model: All AM&Ventilator models Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,

Anesthesia Machine&Ventilator_Parts List_V210101 was uploaded to TDP:


/Anesthesia Machine/Ventilator/Parts List.
Parts changed in 2020.11&12 ECR were deleted or added from new files.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Technical News Revision


Subject: Anesthesia Machine&Ventilator Parts Ref No: TMZ-21006-All AM&Ventilator
List updated

Version Effective Date Description


1.0 2021-02-09 Document released.

No.: OMZ-21009-A Series Page 28of 428


PMLS Global Technical Support Dept

TMZ-21001-A5&A7

A series’ monitor board optimized

Technical News
Subject: A series’ monitor board optimized

RECORD IDENTIFIER

Ref No.: TMZ-21001-A5&A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Jan. 12nd ,2021 Appearance
Model: A5, A7 Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,

1. A series’ monitor board optimized ,boards after change will use new type flow and
pressure sensors for better stability.

No.: OMZ-21009-A Series Page 29of 428


PMLS Global Technical Support Dept

Fig1 Monitor board change


Influences on Pre-sales & Installation
NA
Influences on Upgrade
New monitor boards just have better stability, no need to upgrade if no failure
Influences on Maintenance
a) New part numers(after change) released for monitor board related parts, they will be
available after 2021.3.15, apply old parts(before change) for repair before this date.
Part number Part name Part number Part name
Remark
before change before change after change after change
1. Parts before
change still can
be used on
Monitoring device after
Monitoring Board change
115-051205-00 115-075153-00
Board(DSP) FRU VCM(CE/DSP/s 2. Refer to guide
ide) attached with
package
3. Software upgrade
is required after
115-051206-00 Ventilator Control 115-075154-00 Monitoring replacement

No.: OMZ-21009-A Series Page 30of 428


PMLS Global Technical Support Dept
Board(DSP) FRU Board 4. Factory
VCM+VPM(CE/ calibration is
DSP/side) required after
replacement
5. In the name after
change, VCM
means single
board,
VCM+VPM
means board
assembly(2
pieces)
1. Parts before
Ventilation Ventilation change still can
115-051207-00 kit(Non- 115-075155-00 kit(CE/DSP/O2 be used on
FDA/DSP)FRU drive/side) device after
change
2. Refer to guide
Ventilation attached with
Ventilation
kit(Non- package
115-051208-00 115-075156-00 kit(CE/DSP/Air
FDA/DSP/Air 3. Software upgrade
drive/side)
driven) is required after
replacement
Table1 Monitor board change
PS: The guide could also been found on TDP folder
-View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/A5
(0631&0634)/Manual/Service Manual
-View Files:/Anesthesia Machine/Ventilator/Anesthesia
Machine/A7(0632&0635)/Manual/Service Manual
b) Since new sensors port diameter is smaller than original ones, short tubes(PN:M6G-
020046---) are used to match the diameter, please confirm those tubes are installed
during replacement. For more details, please read the guide.

Fig2 short tubes on ports


c) Sensor shape change caused tube connection way change, new connection
diagram as below. For more details, please read the guide.
No.: OMZ-21009-A Series Page 31of 428
PMLS Global Technical Support Dept

Fig3 Tube connection diagram after change

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation

Sincerely yours,
Mindray Service Department
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Technical News Revision


Subject: A series’ monitor board optimized Ref No: TMZ-21001-A5&A7

Version Effective Date Description


1.0 Jan. 13nd,2021 Document released.

No.: OMZ-21009-A Series Page 32of 428


PMLS Global Technical Support Dept

TMZ-20054 -All AM&Ventilator

Anesthesia Machine&Ventilator Parts List updated

Technical News
Subject: Anesthesia Machine&Ventilator Parts List updated

RECORD IDENTIFIER

Ref No.: TMZ-20054 -All AM&Ventilator Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: 2020-12-11 Appearance
Model: All AM&Ventilator models Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,

Anesthesia Machine&Ventilator_Parts List_V200501 was uploaded to TDP:


/Anesthesia Machine/Ventilator/Parts List.
Parts changed in 2020.09&10 ECR were deleted or added from new files.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Anesthesia Machine&Ventilator Parts List Ref No: TMZ-20054-All
updated AM&Ventilator

Version Effective Date Description


1.0 2020-12-14 Document released.

No.: OMZ-21009-A Series Page 33of 428


PMLS Global Technical Support Dept

TMZ-20053-A7&WATO 55&65 Pro

A series and WATO series released new vacuum generator unit

Technical News
Subject: A series and WATO series released new vacuum generator unit

RECORD IDENTIFIER

Ref No.: TMZ-20053-A7&WATO 55&65 Pro Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: 2020-12-23 Appearance
Model: A7, WATO 55&65(0625,0626), Stop Production
WATO 55&65 Pro(0627,0628)
Comments:NA Others

CONTENT

To Whom It May Concern,

1. A series and WATO series released new vacuum generator unit,new part includes
original vacuum generator unit and muffler assembly.

Fig1 Vacuum generator unit change


Influences on Pre-sales & Installation
No.: OMZ-21009-A Series Page 34of 428
PMLS Global Technical Support Dept
NA
Influences on Upgrade
New vacuum generator unit keeps the same paramaters comparing with original one, no
need to upgrade.
Influences on Maintenance
New vacuum generator unit kits released to support maintenance, details as table shows

Part number Part number


Part name Remark
before change after change

New kit has different install


Vacuum generator unit steps, please refer to
115-017748-00 115-073514-00
kit(A7) replacement guide attached in
package
1. New kit has different install
steps, please refer to
replacement guide attached
in package
Vacuum generator unit
115-017748-00 115-074405-00 2. This kit could be used on
kit(WATO)
WATO 55&65
(0625,0626),
WATO 55&65 Pro
(0627,0628)
Muffler assembly could not be
replaced independently any
Muffler assembly 115-018356-00 / more, please apply above
vacuum generator unit kit to fix
muffler problem
Table1 vacuum generator unit change
PS: The guide could also been found on TDP folder
View Files:/Anesthesia Machine/Ventilator/Maintenance

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 35of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: A series and WATO series Ref No: TMZ-20053-A7&WATO 55&65 Pro
released new vacuum generator unit

Version Effective Date Description


1.0 2020-12-24 Document released.

No.: OMZ-21009-A Series Page 36of 428


PMLS Global Technical Support Dept

TMZ-20046-A7

A7(Japan) system software V02.23.01.00 released

Technical News
Subject: A7(Japan) system software V02.23.01.00 released

RECORD IDENTIFIER

Ref No.: TMZ-20046-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Nov. 4th ,2020 Appearance
Model: A7 Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,

1. A7(Japan) system software V02.23.01.00 released. It solved an alarm sound problem


during production.
Influences on Pre-sales & Installation
Device after Nov. 4th 2020 will have new system software inside.
Influences on Upgrade
Contact with Mindray engineer to confirm before upgrade.
Influences on Maintenance
If software upgrade is necessary and original version is lower than V02.23.01.00, contact
with corresponding GTS for this version or higher.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
No.: OMZ-21009-A Series Page 37of 428
PMLS Global Technical Support Dept

Mindray Service Department


Mindray Co., Ltd.

Technical News Revision


Subject: A7(Japan) system software Ref No: TMZ-20046-A7
V02.23.01.00 released

Version Effective Date Description


1.0 Nov. 4th ,2020 Document released.

No.: OMZ-21009-A Series Page 38of 428


PMLS Global Technical Support Dept

TMZ-20045-A5&A7

A5&A7 updated Dutch, Italian, Russian, Spanish Operator's Manual

Technical News
Subject: A5&A7 updated Dutch, Italian, Russian, Spanish Operator's Manual

RECORD IDENTIFIER

Ref No.: TMZ-20045-A5&A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Nov. 4th ,2020 Appearance
Model: A5&A7 Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,

1. A5&A7 updated Dutch, Italian,Russian,Spanish Operator's Manual.

Version Version
Operator's Manual Remark
before change after change

A5_Operator's Manual_V3.0_Dutch 2.0 3.0 NA


A5_Operator's Manual_V4.0_Italian 3.0 4.0
A5_Operator's
Manual_V4.0_Russian(For CIS except 3.0 4.0
Russia)
A5_Operator's Manual_V4.0_Spanish 3.0 4.0
A7_Operator's Manual_V3.0_Dutch 2.0 3.0
A7_Operator's Manual_V3.0_Italian 2.0 3.0
A7_Operator's
Manual_V4.0_Russian(For CIS except 3.0 4.0
Russia)
A7_Operator's Manual_V4.0_Spanish 3.0 4.0
Table1 Manual update list
PDF version had been uploaded to TDP(intlservice.mindray.com), path as below:
⚫ View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/A5
(0631&0634)/Manual/Operation Manual
⚫ View Files:/Anesthesia Machine/Ventilator/Anesthesia
No.: OMZ-21009-A Series Page 39of 428
PMLS Global Technical Support Dept
Machine/A7(0632&0635)/Manual/Operation Manual

Influences on Pre-sales & Installation


Device after Nov. 4th 2020 will have new version Operator's Manual in package.
Influences on Upgrade
NA
Influences on Maintenance
NA

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A5&A7 updated Dutch, Ref No: TMZ-20045-A5&A7
Italian,Russian,Spanish Operator's Manual

Version Effective Date Description


1.0 Nov. 4th ,2020 Document released.

No.: OMZ-21009-A Series Page 40of 428


PMLS Global Technical Support Dept

TMZ-20039-A7

A7&V60 released in Japan

Technical News
Subject: A7&V60 released in Japan

RECORD IDENTIFIER

Ref No.: TMZ-20039-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Sup. 8th ,2020 Appearance
Model: A series Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,


A7&V60 released in Japan. New labels with JAN code are added on main units, spare
parts with JAN code are added, Japanese operation manuals and “添付文书” released.
A7(Japan) system software V02.22.01.00 released.

1. New labels with JAN code are added on main units and package

Fig1 A7 main unit label

No.: OMZ-21009-A Series Page 41of 428


PMLS Global Technical Support Dept

Fig2 V60 package label


Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
Contact with Mindray engineer for main units troubleshooting and spare parts.

2. Spare parts with JAN code are added

Part number Part number


Part name Remark
before change after change

Sevoflurane Quik-Fil With JAN code


/ 115-072045-00
Vaporizer(Japan)
AG(O2)(w/o
/ 120-022092-00
accessory)(Japan)
Li-ion Bat Pack
(11.1V4500mAh / 115-072628-00
LI23S002A)
Table1 Spare parts with JAN code

No.: OMZ-21009-A Series Page 42of 428


PMLS Global Technical Support Dept

Fig3 AG JAN code


Influences on Pre-sales & Installation
Other accessories also have JAN code, above parts are mentioned especially because
they used for repairing.
Influences on Upgrade
Please use new PN for upgrade
Influences on Maintenance
Please use new PN for repairing

3. Japanese operation manuals and “添付文书” released

Fig4 添付文书
Influences on Pre-sales & Installation
Paper version operation manual and “添付文书”will be put in first drawer(A7) or
package(V60)
Influences on Upgrade
Contact with Mindray engineer to get PDF version
Influences on Maintenance
NA
4. A7(Japan) system software V02.22.01.00 released.
No.: OMZ-21009-A Series Page 43of 428
PMLS Global Technical Support Dept
Influences on Pre-sales & Installation
NA
Influences on Upgrade
Contact with Mindray engineer to confirm before upgrade.
Influences on Maintenance
Contact with corresponding GTS for software if necessary.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A7&V60 released in Japan Ref No: TMZ-20039-A7

Version Effective Date Description


1.0 Sup. 8th ,2020 Document released.

No.: OMZ-21009-A Series Page 44of 428


PMLS Global Technical Support Dept

TMZ-20035-Anesthesia machine&Ventilator

Anesthesia machine&Ventilator package label added UDI infromation, module exhaust


tube changed

Technical News
Subject: Anesthesia machine&Ventilator package label added UDI infromation, module
exhaust tube changed

RECORD IDENTIFIER

Ref No.: TMZ-20035-Anesthesia Relate To:


machine&Ventilator
Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Anesthesia machine&Ventilator Appearance
Model: A series Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,

1. Anesthesia machine&Ventilator main unit and accessory package label added


UDI(Unique Device Identification) infromation

Fig1 Main unit package label

No.: OMZ-21009-A Series Page 45of 428


PMLS Global Technical Support Dept

Fig2 Accessory package label


Influences on Pre-sales & Installation
Below information for your reference

Fig3 How to distinguish UDI label

Influences on Upgrade
NA
Influences on Maintenance
NA

2. Module exhaust tube extends longth to 50cm from 37cm to meet customer side
requirement, PN remains the same.
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
NA

If you need download information or have any question, please feel free to contact
No.: OMZ-21009-A Series Page 46of 428
PMLS Global Technical Support Dept
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Anesthesia machine&Ventilator Ref No:
package label added UDI infromation, TMZ-20035-Anesthesia
module exhaust tube changed machine&Ventilator

Version Effective Date Description


1.0 Aug. 25th ,2020 Document released.

No.: OMZ-21009-A Series Page 47of 428


PMLS Global Technical Support Dept

TMZ-20034-A series

A series released new software V02.22.01.00

Technical News
Subject: A series released new software V02.22.01.00

RECORD IDENTIFIER

Ref No.: TMZ-20034-A series Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Aug. 25th ,2020 Appearance
Model: A series Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,

1. A series(A5,A7) released new software V02.22.01.00, this version sloved :


Bundle version Time out during system check occasionally

Fig1

No.: OMZ-21009-A Series Page 48of 428


PMLS Global Technical Support Dept
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
1. Please contact with corresponding RTS to get V02.22.01.00 software package
2. V02.22.01.00 software package released new PC upgrade software, uninstall earlier
version and install this one before upgrade
3. V02.22.01.00 software package has different upgrade steps comparing with earlier
version, check the guide in package for more details.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A series released new software Ref No: TMZ-20034-A series
V02.22.01.00

Version Effective Date Description


1.0 Aug. 25th ,2020 Document released.

No.: OMZ-21009-A Series Page 49of 428


PMLS Global Technical Support Dept

TMZ-20026-A7

Changes for Japan order 21377912.10

Technical News
Subject: Changes for Japan order 21377912.10

RECORD IDENTIFIER

Ref No.: TMZ-20026-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: July 6th,2020 Appearance
Model: A7 Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,


Japan order 21377912.10 A7 have following temporarily changes :

1. Temporarily software which supports Japanese is released.


Influences on Pre-sales & Installation
NA
Influences on Upgrade
Temporarily version does not support upgrade.
Influences on Maintenance
Temporarily version does not support upgrade. Contact with corresponding GTS for
software if necessary.

2. New label for A7 main unit is added.

No.: OMZ-21009-A Series Page 50of 428


PMLS Global Technical Support Dept

Fig1
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
NA

3. New label for AG module is added.

Fig2
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
NA

4. Japanese “添付文書”(2 pages) is added in first drawer of A7.

No.: OMZ-21009-A Series Page 51of 428


PMLS Global Technical Support Dept

Fig3
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
NA

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
Changes for Japan order 21377912.10 TMZ-20026-A7

Version Effective Date Description


1.0 July 6th,2020 Document released.

No.: OMZ-21009-A Series Page 52of 428


PMLS Global Technical Support Dept

TMZ-20022-A5

A5s released in India

Technical News
Subject: A5s released in India

RECORD IDENTIFIER

Ref No.: TMZ-20022-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Jun. 28th ,2020 Appearance
Model: A5 Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,


A5s released in India. A5s is just a code name in order system, silk print on unit
surface and label is still A5. Serial number is the same with A5.
A5s is a simple confiuration A5, based on A5 CE standard,Air&O2&N2O 3 gases.It
has less accessories, hardware and software functions compairing with A5. Details
as below

Accessories
1. A5s cancelled A5 standard config passive agss tube

Fig1
Influences on Pre-sales & Installation
Please refer to device attached packing list during installation.

No.: OMZ-21009-A Series Page 53of 428


PMLS Global Technical Support Dept
Influences on Upgrade
A5s could upgrade passive AGSS and Active AGSS, it uses the A5 upgrade kit.
Influences on Maintenance
NA
2. A5s cancelled A5 standard config O2 sensor

Fig2
Influences on Pre-sales & Installation
Please refer to device attached packing list during installation.
Influences on Upgrade
A5s could upgrade O2 sensor, it uses the A5 O2 sensor. Need to purchase O2 sensor
cable and O2 sensor cover at the same time.
Influences on Maintenance
If O2 sensor is not configured, after system software upgrading A5s might have O2 sensor
related alarm, enter Main→System→Code 1234→Other settings→change O2 cell
monitoring to “Off”

Hardware
1. A5s cancelled touch pad

Fig3
Influences on Pre-sales & Installation
NA
Influences on Upgrade
No.: OMZ-21009-A Series Page 54of 428
PMLS Global Technical Support Dept
Touch pad could not be upgraded.
Influences on Maintenance
NA
2. A5s cancelled Auxiliary AC Outlets

Fig4
Influences on Pre-sales & Installation
NA
Influences on Upgrade
Auxiliary sockets could not be upgraded.
Influences on Maintenance
If hardware box rear cover broken please contact with corresponding Mindray engineer.

3. A5s cancelled Auxiliary O2/Air Flowmeters and Auxiliary O2 gas Power Outlet

Fig5
Influences on Pre-sales & Installation
NA
Influences on Upgrade
Auxiliary O2/Air Flowmeters and Auxiliary O2 gas Power Outlet could not be upgraded.
No.: OMZ-21009-A Series Page 55of 428
PMLS Global Technical Support Dept
Influences on Maintenance
NA
4. A5s cancelled module rank

Fig6
Influences on Pre-sales & Installation
NA
Influences on Upgrade
Module rank could not be upgraded. Module is unavailable on A5s.
Influences on Maintenance
NA

5. A5s cancelled Air cylinder bracket

Fig7
Influences on Pre-sales & Installation
NA
No.: OMZ-21009-A Series Page 56of 428
PMLS Global Technical Support Dept

Influences on Upgrade
Air cylinder bracket could not be upgraded.
Influences on Maintenance
NA

6. A5s changed center breaker to independent wheel lock

Fig8
Influences on Pre-sales & Installation
NA
Influences on Upgrade
Center breaker could not be upgraded.
Influences on Maintenance
New wheel is added for A5s

Part number Part number


Part name Remark
before change after change

4 wheels are the


Casters.6inch / 034-000234-00
same
Table1

Software
1. A5s cancelled A5 standard config SIMV mode, PS mode, Loop function
Influences on Pre-sales & Installation
Please refer to device PI during installation.

No.: OMZ-21009-A Series Page 57of 428


PMLS Global Technical Support Dept

Influences on Upgrade
A5s could upgrade SIMV mode ,PS mode, Loop function, PS&Loop have upgrade
kit,SIMV mode upgrade package is added newly.

Part number Part number


Part name Remark
before change after change

Activation Need machine SN


/ 115-070677-00
Code(USB/SIMV-VC) and ID during
application, active
Activation
/ 115-070678-00 process refer to
Code(USB/SIMV-PC)
service manual
Table2
Influences on Maintenance
A5s uses the same software with A5(DSP), please apply A5(DSP) latest software for
maintenance.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A5s released in India Ref No: TMZ-20022-A5

Version Effective Date Description


1.0 Jun. 28th ,2020 Document released.

No.: OMZ-21009-A Series Page 58of 428


PMLS Global Technical Support Dept

TMZ-20021-A series&WATO series

A series&WATO series released new monitor brackets,A series have new manual

Technical News
Subject: A series&WATO series released new monitor brackets,A series have new manual

RECORD IDENTIFIER

Ref No.: TMZ-20021-A series&WATO series Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
th
Date: Jun. 28 ,2020 Appearance
Model: A series&WATO series Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,

1. A series(A5,A7)&WATO series(WATO EX-20,WATO EX-20 Vet, WATO EX-30,


WATO EX-35[0619], WATO EX-55[0626], WATO EX-65[0625], WATO EX-55
Pro[0628], WATO EX-65 Pro[0627]) released 6 new monitor brackets to support new
monitors

Part number Part number


Part name Remark
before change after change

Bracket for N19/N22, GCX M series


/ 115-066025-00
fixed bracket, different
Bracket for types use different
N12/15/17/ePM15, / 115-066028-00 adapter boards
fixed
Bracket for
ePM10/12/uMEC, / 115-070011-00
fixed
Bracket for N19/N22, GCX VHM series
/ 115-066027-00
variable bracket, different
Bracket for types use different
N12/15/17/ePM15, / 115-066029-00 adapter boards
variable

No.: OMZ-21009-A Series Page 59of 428


PMLS Global Technical Support Dept
Bracket for
ePM10/ePM12/uMEC, / 115-070768-00
variable
Table1

Fig1
Influences on Pre-sales & Installation
File Anesthesia System Bracket Installation Instructions is attached with bracket package
as reference.
Influences on Upgrade
WATO EX-55[0622],WATO EX-65[0621],WATO EX-55[0624],WATO EX-65[0623],
WATO EX-35[0618] also could use above bracket to upgrade.
Influences on Maintenance
NA

2. A5 and A7 released new English operation manual V20.0,manual added new monitor
bracket details.
Folder address on TDP(http://intlservice.mindray.com/logon/index.action) as below
View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/
A5 (0631&0634)/Manual/Operation Manual
View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/
A7(0632&0635)/Manual/Operation Manual
Influences on Pre-sales & Installation
NA
Influences on Upgrade
NA
Influences on Maintenance
No.: OMZ-21009-A Series Page 60of 428
PMLS Global Technical Support Dept
NA

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A series&WATO series released Ref No: TMZ-20021-A series&WATO series
new monitor brackets,A series have new
manual

Version Effective Date Description


1.0 Jun. 28th ,2020 Document released.

No.: OMZ-21009-A Series Page 61of 428


PMLS Global Technical Support Dept

TMZ-20006-A5&A7

A5&A7 released in Russia

Technical News
Subject: A5&A7 released in Russia

RECORD IDENTIFIER

Ref No.: TMZ-20006-A5&A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Feb 27th,2020 Appearance
Model: A5&A7 Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,

1. A5&A7 anesthesia machine released in Russia.

Fig1

No.: OMZ-21009-A Series Page 62of 428


PMLS Global Technical Support Dept
Influences on Pre-sales & Installation
New Operator's Manuals have been uploaded to TDP platform, below address for your
reference
Item
Descriptions TDP address
No.
A5_Operator's
Anesthesia Machine/Ventilator/Anesthesia
1 Manual_V1.0_Russian(For
Machine/A5 (0631&0634)/Manual/Operation Manual
Russia)
A7_Operator's
Anesthesia Machine/Ventilator/Anesthesia
2 Manual_V1.0_Russian(For
Machine/A7(0632&0635)/Manual/Operation Manual
Russia)
PS: Above manuals only for Russia, there are the other Russian manuals for other CIS
countries.

Influences on Upgrade
NA

Influences on Maintenance
A5&A7 maintenance (include preventive maintenance) needs A Series Service Certificate
from Mindray, the certificate can be got by attending Mindray face-to-face training. Please
inquire with your Mindray service contact for training details.
PS: Branch lecturer could trigger the face-to-face training by process in Portal website
workspace-Overseas Training Application, please get help from HQ training department
Sunny (zhuyunlong@mindray.com) if any doubt.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 63of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: A5&A7 released in Russia Ref No: TMZ-20006-A5&A7

Version Effective Date Description


1.0 Feb 27th,2020 Document released.

No.: OMZ-21009-A Series Page 64of 428


PMLS Global Technical Support Dept

TMZ-20004-WATO&A series

New nylon Breathing Circuits are released for India

Technical News
Subject: New nylon Breathing Circuits are released for India.

RECORD IDENTIFIER

Ref No.: TMZ-20004-WATO&A series Relate To:


Version:1.0 Software/Manual Upgrade
Author:Dawson Yan Parts Change
Date: Jan, 20th 2020 Appearance
Model: 0625, 0626, 0627, 0628, 0634, Stop Production
0635 Others
Comments:N/A

CONTENT

To Whom It May Concern,


1. New nylon Breathing Circuits are released for India.

New nylon Breathing Circuits are released for India. When being wiped by some types of
disinfectant, nylon is expected to be more durable than PPSU.
PPSU and nylon Breathing Circuits’ differences are marked by red arrows shown below:

No.: OMZ-21009-A Series Page 65of 428


PMLS Global Technical Support Dept

Fig 1. Differences of Upper Cover, Upper Cover 2

Fig 2. Difference of Lower Cover

No.: OMZ-21009-A Series Page 66of 428


PMLS Global Technical Support Dept

Fig 3. Differences of Lower Cover, Lower Cover 2

Influence on installation
N/A

Influence on maintenance
Not all models of WATO series are all concerned by this change. Only the models shown
below are concerned:
Model Factory Code SN
WATO EX-55 0626 KU-**
WATO EX-65 0625 KT-**
WATO EX-55 Pro 0628 F1-**
WATO EX-65 Pro 0627 KY-**
New nylon Breathing Circuits can be used entirely to replace the same type Breathing Circuit
whose material is PPSU. According to the new nylon Breathing Circuits, three FRUs and
five covers are added:
Order Number Part Description Reference
115-067595-00 Breathing system FRU(0634/PA) For A5.
115-067597-00 Breathing Circuit Cell(0635/PA) For A7.
Patient main unit(with bag
115-067598-00 For WATO EX-55,65,55 Pro,65 Pro.
arm/0625/PA)
For nylon Breathing Circuit, not for
043-006704-00 upper cover 2(PA/silk-screen)
PPSU Breathing Circuit.

No.: OMZ-21009-A Series Page 67of 428


PMLS Global Technical Support Dept
Upper Cover. For nylon Breathing
043-007644-00 Top Bezel (PA/0623) Circuit, not for PPSU Breathing
Circuit.
For nylon Breathing Circuit, not for
043-006705-00 lower cover 2(PA/silk-screen)
PPSU Breathing Circuit.
Lower Cover. For nylon Breathing
041-025940-00 bottom cover(PA/0623) Circuit, not for PPSU Breathing
Circuit.
Lower Cover, only for A7. For nylon
043-012859-00 lower cover silk(0635/nylon) Breathing Circuit, not for PPSU
Breathing Circuit.

Influence on upgrade
N/A

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: New nylon Breathing Circuits Ref No: TMZ-20004-WATO&A series
are released for India

Versi Effective Date Description


on
1.0 2020-01-20 Document released.

No.: OMZ-21009-A Series Page 68of 428


PMLS Global Technical Support Dept

TMZ-19036-A5&A7&WATO

A series anesthesia machine knob change and anesthesia machine electronic pressure
sensor change

Technical News
Subject: A series anesthesia machine knob change and anesthesia machine electronic
pressure sensor change

RECORD IDENTIFIER

Ref No.: TMZ-19036-A5&A7&WATO EX- Relate To:


65Pro
Version:1.0 Software/Manual Upgrade
Author: Robert Ling Parts Change
Date: Dec 30th,2019 Appearance
Model: A5&A7&WATO EX-65Pro Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,


1. A series anesthesia machine knob change.

Change description:
The surface appearance of the knob of A-series is darker than WATO
series, which can affect the feeling of the usage, The R&D changed the
surface appearance of the knob of A-series so that it’s consistent with the
WATO series.

Influence on installation:

NA

Influence on upgrade:

NA

Influence on maintenance:

The PNs change as follows:


No.: OMZ-21009-A Series Page 69of 428
PMLS Global Technical Support Dept
Before change(A) After change (B)
NO. Reference
PN Part name PN Part name
encoder knob
1 115-034071-00 assembly
(oxygen0635) Encoder
115-066659- only B
encoder knob knob
00 replace A
assembly package
2 115-034072-00
(balance
gas0635)
O2 knob of
O2 knob of
041-033230- needle only B
3 041-006773-00 needle valve
00 valve replace A
(0632)
(0632)
AIR knob
AIR knob of
041-033231- of needle only B
4 041-006775-00 needle valve
00 valve replace A
(0632)
(0632)
041-033234- select only B
5 041-007598-01 select switch
00 switch replace A
801-0631- Knob of O2
6
00023-00 Needle Valve Flow
801-0631- Knob of N2O 115-066658- adjust only B
7
00021-00 Needle Valve 00 knob replace A
801-0631- Knob of Air package
8
00022-00 Needle Valve

2. Anesthesia machine electronic pressure sensor change.

1) Change description:

In order to reduce the cost of the A7 and WATO EX-65Pro anesthesia


machines which equipped with electronic pipeline pressure monitoring,
R&D changed the electronic pressure sensor.

No.: OMZ-21009-A Series Page 70of 428


PMLS Global Technical Support Dept

Before After
Change Change

Influence on installation:

NA

Influence on upgrade:

NA

Influence on maintenance:

a) For the machine before Jul. 2018,if the pipeline supply electronic pressure sensor
broken, you should apply the corresponding inlet module to maintenance .The
PNs change as follows:

Before change(A) After change (B)


NO. Reference
PN Part name PN Part name
Air origin inlet Air origin inlet
115-044986- 115-056324- only B
1 module(0627/NIS module(0627/
00 00 replace A
T) NIST)
Air origin inlet Air origin inlet
115-044989- 115-056325- only B
2 module(0627/DIS module(0627/
00 00 replace A
S) DISS)
O2 origin inlet O2 origin inlet
115-044984- 115-056320- only B
3 module(0627/NIS module(0627/
00 00 replace A
T) NIST)
O2 origin inlet O2 origin inlet
115-044987- 115-056321- only B
4 module(0627/DIS module(0627/
00 00 replace A
S) DISS)
5 115-045157- O2 supply inlet 115-056326- O2 supply only B

No.: OMZ-21009-A Series Page 71of 428


PMLS Global Technical Support Dept
00 assembly(NIST/0 00 inlet replace A
627) assembly(NI
ST/0627)
N20 origin
N20 origin inlet
115-044985- 115-056322- inlet only B
6 module(0627/NIS
00 00 module(0627/ replace A
T)
NIST)
N20 origin
N20 origin inlet
115-044988- 115-056323- inlet only B
7 module(0627/DIS
00 00 module(0627/ replace A
S)
DISS)
AIR origin
AIR origin inlet
115-033738- 115-056331- inlet only B
8 module(0635/NIS
00 00 module(0635/ replace A
T)
NIST)
AIR origin
AIR origin inlet
115-033734- 115-056332- inlet only B
9 module(0635/DIS
00 00 module(0635/ replace A
S)
DISS)
Reserve
Reserve Oxygen
115-034798- 115-056333- Oxygen only B
10 Supply base
00 00 Supply base replace A
part(0635)
part(0635)
Reserve
Reserve Oxygen
115-034799- 115-056852- Oxygen only B
11 Supply(0635/NIST
00 00 Supply(0635/ replace A
)
NIST)
Reserve
Reserve Oxygen
115-034800- 115-056851- Oxygen only B
12 Supply(0635/DIS
00 00 Supply(0635/ replace A
S)
DISS)
O2 origin inlet O2 origin inlet
115-033736- 115-056327- only B
13 module(0635/NIS module(0635/
00 00 replace A
T) NIST)
O2 origin inlet O2 origin inlet
115-033735- 115-056328- only B
14 module(0635/DIS module(0635/
00 00 replace A
S) DISS)
N2O origin
N2O origin inlet
115-033737- 115-056329- inlet only B
15 module(0635/NIS
00 00 module(0635/ replace A
T)
NIST)
N2O origin inlet N2O origin
115-033733- 115-056330- only B
16 module(0635/DIS inlet
00 00 replace A
S) module(0635/

No.: OMZ-21009-A Series Page 72of 428


PMLS Global Technical Support Dept
DISS)

b) For the machine after Jul. 2018, if the pipeline supply electronic pressure sensor
broken, you can apply the sensor to maintenance .The PN changes as follows:

NO. PN Part name Reference


1 051-002521-00 VT70 air source press sensor PCBA /

2) Change description:

In order to reduce the cost of the A7 and WATO EX-65Pro anesthesia


machines which equipped with electronic cylinder pressure monitoring,
R&D changed the electronic pressure sensor.

After change
Before change

Influence on installation:

NA

Influence on upgrade:

NA

No.: OMZ-21009-A Series Page 73of 428


PMLS Global Technical Support Dept
Influence on maintenance:

If the cylinder supply electronic pressure sensor broken, you should apply the
corresponding yoke assembly to maintenance .The PNs change as follows:

Before change(A) After change (B)


NO. Reference
PN Part name PN Part name
O2 cylinder
115-044990- O2 cylinder yoke 115-056316- yoke only B
1
00 assembly(0627) 00 assembly(06 replace A
27))
N2O cylinder
115-044991- N2O cylinder yoke 115-056317- yoke only B
2
00 assembly(0627) 00 assembly(06 replace A
27)
Air cylinder
115-044992- Air cylinder yoke 115-056318- yoke only B
3
00 assembly(0627) 00 assembly(06 replace A
27)
AIR cylinder
115-033731- AIR cylinder yoke 115-056631- yoke only B
4
00 assembly(0635) 00 assembly(06 replace A
35)
AIR cylinder
AIR cylinder yoke
115-040999- 115-056629- yoke only B
5 assembly(0635/air
00 00 assembly(06 replace A
)
35/air)
O2 cylinder
115-033730- O2 cylinder yoke 115-056630- yoke only B
6
00 assembly(0635) 00 assembly(06 replace A
35)
O2 cylinder
O2 cylinder yoke
115-040997- 115-056628- yoke only B
7 assembly(0635/air
00 00 assembly(06 replace A
)
35/air)
N2O cylinder
115-033732- N2O cylinder yoke 115-056632- yoke only B
8
00 assembly(0635) 00 assembly(06 replace A
35)
This distinguish information will also update to part list, please read remark during
part searching.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
No.: OMZ-21009-A Series Page 74of 428
PMLS Global Technical Support Dept
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A series anesthesia machine knob Ref No: TMZ-19036-A5&A7&WATO EX-
change and anesthesia machine electronic 65Pro
pressure sensor change

Version Effective Date Description


1.0 Dec 30th,2019 Document released.

No.: OMZ-21009-A Series Page 75of 428


PMLS Global Technical Support Dept

TMZ-19035-A5&A7

A series anesthesia machine software V02.22.00.01 release

Technical News
Subject: A series anesthesia machine software V02.22.00.01 release

RECORD IDENTIFIER

Ref No.: TMZ-19035-A5&A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Robert Ling Parts Change
Date: Dec 9th,2019 Appearance
Model: A5&A7 Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,


1. A series anesthesia machine software V02.22.00.01 release.

The software changes as follows:

1. The new software optimizes the “FiO2 Too High” alarm.

2. The new software optimizes the HL7 send BIS data.

Influence on installation:

NA

Influence on upgrade:

NA

Influence on maintenance:

If the anesthesia machine still has “FiO2 Too High” alarm even the high FiO2 alarm
limit had been set to 100%, please upgrade the machine’s software to V02.22.00.01
or above.

If you want to upgrade software, please contact Mindray regional service engineer.
No.: OMZ-21009-A Series Page 76of 428
PMLS Global Technical Support Dept

2. The A5&A7 anesthesia machine canceled the internal AG module and


using the external AG module to realize the function of anesthesia
calculation

Influence on installation:

NA

Influence on upgrade:

If the customer needs to upgrade the anesthesia calculation function of the A5&A7
anesthesia machine, please ensure that the A5&A7 anesthesia machine is equipped
with an external AG module and order activation code to activate the anesthesia
calculation function.

Before change(A) After change (B)


Reference
PN Part name PN Part name
Need to
Activation
provide
/ / 115-044265-00 Code(USB/AA
machine's SN
measurement)
and ID

Influence on maintenance:

NA

3. The internal AG module of A7 anesthesia machine changed

Change description:

The internal AG module of A7 anesthesia machine have changed from 115-


002504-00 to 115-039361-00 for reduce cost.

Influence on installation:

NA

Influence on upgrade:

NA

No.: OMZ-21009-A Series Page 77of 428


PMLS Global Technical Support Dept
Influence on maintenance:

Before change(A) After change (B)


Reference
PN Part name PN Part name
AION 03-31 AION Argentum A and B can't
115-002504-00 Multigas 115-039361-00 AG be replaced
Analyzer Module(M01C) by each other

For the internal AG module failure of A7 anesthesia machine,


If the machine’s SN is earlier than GC-91XXXXXX (91 stands for 2019.01), please
apply 115-002504-00 to maintenance.
If the machine’s SN is after GC-91XXXXXX, please apply 115-039361-00 to
maintenance.
This distinguish information will also update to part list, please read remark during
part searching.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A series anesthesia machine Ref No: TMZ-19035-A5&A7
software V02.22.00.01 release

Version Effective Date Description


1.0 Dec 9th,2019 Document released.

No.: OMZ-21009-A Series Page 78of 428


PMLS Global Technical Support Dept

TMZ-19028-V60

V60 vaporizer filler cap’s optimization

Technical News
Subject: V60 vaporizer filler cap’s optimization.

RECORD IDENTIFIER

Ref No.: TMZ-19028-V60 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Dawson Yan Parts Change
Date: Oct, 18th 2019 Appearance
Model: 0602 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,


V60 vaporizer filler cap’s optimization.

For Pour Fill and Quik-Fil vaporizers, an issue happens occasionally: Poor sealing of the
filler cap due to thread stuck, and it leads to leakage. Screw threads’ positions as shown
below:

Figure 1.
In order to solve this problem, R&D optimized screw threads of filler port and filler cap.
After the optimization, it’s easy to tighten filler cap.

No.: OMZ-21009-A Series Page 79of 428


PMLS Global Technical Support Dept
Influence on installation
N/A

Influence on maintenance
N/A

Influence on upgrade
N/A

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: V60 vaporizer filler cap’s Ref No: TMZ-19028-V60
optimization

Versi Effective Date Description


on
1.0 2019-10-18 Document released.

No.: OMZ-21009-A Series Page 80of 428


PMLS Global Technical Support Dept

TMZ-19020- All AM&Ventilator

VT650/VT900 Gas Flow Analyzer release

Technical News
Subject: VT650/VT900 Gas Flow Analyzer release

RECORD IDENTIFIER

Ref No.:TMZ-19020- All AM&Ventilator Relate To:


Version:1.0 Software/Manual Upgrade
Author: Robert.ling Parts Change
Date: Jul 11th,2019 Appearance
Model: All AM&Ventilator models Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,


Because the Fluke VT Plus Gas Analyzer has stopped production, VT650/VT900
Gas Flow Analyzer replaced VT Plus Gas Analyzer. New calibration kit, which can be used
for all Mindray on marketing anesthesia machine and ventilator, is released to adapt new
VT650/VT900 gas flow analyzer.

VT650/VT900 Gas Flow Analyzer

Influence on installation:

NA

Influence on upgrade:

No.: OMZ-21009-A Series Page 81of 428


PMLS Global Technical Support Dept

NA

Influence on maintenance:

If you want to calibrate anesthesia machine or ventilator with VT650/VT900,


maintenance kit is required.

PN Part Description Reference


The price will take effect one
115-063276-00 maintenance kit
week later

For more information, please refer to the 《 VT650&VT900 Gas Flow Analyzer
Introduction_V1.0_EN》.

You can download the 《VT650&VT900 Gas Flow Analyzer Introduction_V1.0_EN》on TDP:

View Files:/Anesthesia Machine/Ventilator/Maintenance

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: VT650/VT900 Gas Flow Analyzer Ref No: TMZ-19020- All AM&Ventilator
release

Version Effective Date Description


1.0 Jul 11th,2019 Document released.

No.: OMZ-21009-A Series Page 82of 428


PMLS Global Technical Support Dept

TMZ-19017- Anesthesia Machine&Ventilator all products

Delete the company address from Anesthesia Machine&Ventilator all products’ quick guide
and maintenance guide

Technical News
Subject: Delete the company address from Anesthesia Machine&Ventilator all products’ quick guide and
maintenance guide.

RECORD IDENTIFIER

Ref No.: TMZ-19017- Anesthesia Relate To:


Machine&Ventilator all products
Version: 1.0 Software/Manual Upgrade
Author: Dawson Yan Parts Change
Date: Jun 18th , 2019 Appearance
Model: WATO&A Anesthesia Machine, E&SV Stop Production
Ventilator, C3 Compressor Others
Comments: N/A

CONTENT

To Whom It May Concern,

Below content is only for your information.

Change description:
The company's business license address and production address have changed. Now we delete
the address information from all types of Anesthesia Machine&Ventilator’s quick guide and
maintenance guide, to reduce the complexity of HQ product management. New version of
quick guide and maintenance guide are available from TDP.

Impacts to service:
N/A.

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

No.: OMZ-21009-A Series Page 83of 428


PMLS Global Technical Support Dept

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Delete the company address from Anesthesia Ref No: TMZ-19017-Anesthesia
Machine&Ventilator all products’ quick guide and Machine& Ventilator all products
maintenance guide.

Version Effective Date Description


1.0 Jun 18th , 2019 Document released.

No.: OMZ-21009-A Series Page 84of 428


PMLS Global Technical Support Dept

TMZ-19015-AllAM&Ventilator

Anesthesia Machine&Ventilator Parts List is released

Technical News
Subject: Anesthesia Machine&Ventilator Parts List is released.

RECORD IDENTIFIER

Ref No.:TMZ-19015-AllAM&Ventilator Relate To:


Version:1.0 Software/Manual Upgrade
Author: Robert.Ling Parts Change
Date: May.30th,.2019 Appearance
Model: All AM&Ventilator models Stop Production
Comments:NA Others

CONTENT

To Whom It May Concern,


1. Anesthesia Machine&Ventilator_Parts List_ V190501 is released.The Parts List of
SV600/650/800/850&WATO EX-55&WATO EX-65 have added pictures as follows:

2. The models in the following table have added the pictures.

Model Picture(Y/N)
A3(0637) Y
A5(0634) Y
A7(0635) Y
SV600/650(VT60/VT65) Y
SV800/850(VT70/VT75) Y
SV300/350(EV20/EV25) Y
WATO EX-65(0625) Y
No.: OMZ-21009-A Series Page 85of 428
PMLS Global Technical Support Dept
WATO EX-55(0626) Y

Parts list is uploaded to TDP: /Anesthesia Machine/Ventilator/Parts List

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Anesthesia Machine&Ventilator Ref No: TMZ-19015-All AM&Ventilator
Parts List is released.

Version Effective Date Description


1.0 2019-05-30 Document released.

No.: OMZ-21009-A Series Page 86of 428


PMLS Global Technical Support Dept

TMZ-19014-FPM

Flow Pressure Measurement (FPM) adds pressure sampling port and changes the
pressure sampling mode

Technical News
Subject: Flow Pressure Measurement (FPM) adds pressure sampling port and changes the
pressure sampling mode.

RECORD IDENTIFIER

Ref No.: TMZ-19014-FPM Relate To:


Version:1.0 Software/Manual Upgrade
Author: Dawson Yan Parts Change
Date: May, 24th 2019 Appearance
Model: FPM (0604) Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,


Flow Pressure Measurement (FPM) adds pressure sampling port and changes the
pressure sampling method.

In order to improve the pressure calibration accuracy of FPM, R&D change the pressure
sampling method.

The improvement method is as follows:


Add a Pressure nozzle on the sidewall of FPM, this nozzle works as a pressure sampling
port.

In terms of appearance, the difference between old type FPM and new type lies in left
view, as shown in the following figure:

No.: OMZ-21009-A Series Page 87of 428


PMLS Global Technical Support Dept

Old type FPM continues to be used as originally, no need to make changes.


During calibration, the gas pipes connection is different between new type FPM and old
type. Please refer to Chapter 3~7 of the Operator's Manual for details.

The latest version of Operator's Manual have been uploaded to TDP, address lists below:
/Anesthesia Machine/Ventilator/FPM/Operator’s Manual/Flow Pressure Measurement Operator's
Manual_V2.0_EN.pdf

Influence on installation
N/A

Influence on maintenance
When new type FPM is used to calibrate, it must be operated in accordance with the latest
version of Operator's Manual.

Influence on upgrade
N/A

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

No.: OMZ-21009-A Series Page 88of 428


PMLS Global Technical Support Dept

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Flow Pressure Measurement Ref No:TMZ-19014-FPM
(FPM) adds pressure sampling port and
changes the pressure sampling method

Versi Effective Date Description


on
1.0 2019-05-24 Document released.

No.: OMZ-21009-A Series Page 89of 428


PMLS Global Technical Support Dept

TMZ-19010-All Anesthesia machine

Anesthesia machine ACGO assembly change

Technical News
Subject: Anesthesia machine ACGO assembly change

RECORD IDENTIFIER

Ref No.: TMZ-19010-All Anesthesia Relate To:


machine
Version: V1.0 Software/Manual Upgrade
Author: Dave Hao Parts Change
Date: Apr, 15th 2019 Appearance
Model: WATO EX-10&20&30&55&65& Stop Production
55Pro&65Pro&35&A5&A7(061B&0615& Others
0617&0618&0625&0626&0627&0628&0619
&0634&0635&0637)
Comments: N/A

CONTENT

To Whom It May Concern,


1. Anesthesia machine ACGO assembly change
Description
Because ACGO switch stop production, new ACGO assembly replaced old ACGO
assembly. Old and new ACGO switch are not compatible. ACGO switch removed from part
list.

Before change After change


PN Part Description PN Part Description
010-000116-00 SWITCH / /
115-043753-00 ACGO assembly(0634/AG) 115-061229-00 mechanical ACGO
assembly(0634/AG)
115-043752-00 ACGO assembly(0634) 115-061228-00 mechanical ACGO
assembly(0634)
115-032991-00 ACGO assembly(0625) 115-061227-00 mechanical ACGO
assembly(0625)
115-043754-00 mechanical ACGO 115-061216-00 mechanical ACGO
assembly(0615) assembly(0615)

No.: OMZ-21009-A Series Page 90of 428


PMLS Global Technical Support Dept
Old and new ACGO assembly can be distinguish by the switch. All new ACGO assembly
use new ACGO switches:

New ACGO Switch


Old ACGO Switch
(Stop production)

Influence on installation
N/A
Influence on maintenance
If ACGO need to be replaced, please refer the list above to apply spare parts.
Influence on upgrade
N/A

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 91of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: Anesthesia machine ACGO Ref No:TMZ-19010-All Anesthesia machine
assembly change

Version Effective Date Description


1.0 2019-04-15 Document released.

No.: OMZ-21009-A Series Page 92of 428


PMLS Global Technical Support Dept

TMZ-19004-All AM&Ventilator

Anesthesia Machine&Ventilator Parts List is released

Technical News
Subject: Anesthesia Machine&Ventilator Parts List is released.

RECORD IDENTIFIER

Ref No.: TMZ-19004-All AM&Ventilator Relate To:


Version:1.0 Software/Manual Upgrade
Author:Dave Hao Parts Change
Date Jan.31st.2019 Appearance
Model: All AM&Ventilator models Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,


Anesthesia Machine&Ventilator_Parts_List_ V190101 is released.
Parts list is uploaded to TDP: /Anesthesia Machine/Ventilator/Parts List

If you have any questions & suggestions, or you find any mistakes, please feedback
to Robert (haolingguang@mindray.com).Thanks a lot for your support.

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Anesthesia Machine&Ventilator Parts Ref No: TMZ-19004-All AM&Ventilator
List is released.

Version Effective Date Description


1.0 2019-01-31 Document released.

No.: OMZ-21009-A Series Page 93of 428


PMLS Global Technical Support Dept

TMZ-18036-A7

A7 anesthesia machine hardware change

Technical News
Subject: A7 anesthesia machine hardware change

RECORD IDENTIFIER

Ref No.: TMZ-18036-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Robert. Ling Parts Change
Date: 2018/11/30 Appearance
Model: A7 Stop Production
Comments: NA Others

CONTENT

To Whom It May Concern,


ECL173F includes changes as follows:
1. A7 power board update
Change description:
Because M02-001152-01 and M02-001032-01 have stopped production, A7 power board
updated. Use the 051-003468-00 0627 Power Board to replace the 051-001261-00 0632
Power Board PCBA.
⚫ Influence on installation:
NA
⚫ Influence on upgrade
NA
⚫ Influence on maintenance:
If A7 power board need to be replaced. Use the same order number (115-018145-00)
to apply spare part. The order number of the power board is not changed, but the part
number changed.

Before change After change


part part part
order number part number order number
description number description
Power Board Power Board
115-018145-00 051-001261-00 115-018145-00 051-003468-00
PCBA PCBA

2. The material change of conformal coating of the 051-003168-00 0635


Ventilation&Control PCBA.

No.: OMZ-21009-A Series Page 94of 428


PMLS Global Technical Support Dept
Change description:

Change the conformal coating of 051-003168-00 0635 Ventilation&Control PCBA from


silicone 422B to SL1301ECO-FLZ.
.
⚫ Influence on installation:
NA

⚫ Influence on upgrade
NA

⚫ Influence on maintenance:
Compatible changes,the PN didn’t change.

3. The BFCS occationally fails to come back into the machine of A7

Change description:

To solve the BFCS stuck problem because of the round end of the solenoid pin and the
ACGO switch is not triggered properly, BFCS module has been upgrade.
⚫ Influence on installation:
NA

⚫ Influence on upgrade
NA

⚫ Influence on maintenance:
Compatible changes,the PN didn’t change.

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A7 anesthesia machine hardware Ref No: TMZ-18036-A7
change

Version Effective Date Description


1.0 2018/11/30 Document released.

No.: OMZ-21009-A Series Page 95of 428


PMLS Global Technical Support Dept

TMZ-18035-A5

A5 anesthesia machine hardware change

Technical News
Subject: A5 anesthesia machine hardware change

RECORD IDENTIFIER

Ref No.: TMZ-18035-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Robert. Ling Parts Change
Date: 2018/11/30 Appearance
Model: A5 Stop Production
Comments: NA Others

CONTENT

To Whom It May Concern,


ECL173F includes changes as follows:
1. A5 power board update

Change description:
Because M02-001152-01 and M02-001032-01 have stopped production, A5 power board
updated. Use the 051-003469-00 0625 Power Board to replace the 051-000685-00 0631
Power Board PCBA.
⚫ Influence on installation:
NA

⚫ Influence on upgrade
NA

⚫ Influence on maintenance:

If A5 power board need to be replaced. Use the same order number (801-0631-00025-
00) to apply spare part. The order number of the power board is not changed, but the
part number changed.

Before change After change


part part part part
order number order number
number description number description
801-0631-00025-00 051-000685-00 Power Board 801-0631-00025-00 051-003469-00 Power Board

No.: OMZ-21009-A Series Page 96of 428


PMLS Global Technical Support Dept
PCBA PCBA

2. The material change of conformal coating of the 051-003168-00 0635


Ventilation&Control PCBA.

Change description:

Change the conformal coating of 051-003168-00 0635 Ventilation&Control PCBA from


silicone 422B to SL1301ECO-FLZ.
.
⚫ Influence on installation:
NA

⚫ Influence on upgrade
NA

⚫ Influence on maintenance:
Compatible changes,the PN didn’t change.

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A5 anesthesia machine hardware Ref No: TMZ-18035-A5
change

Version Effective Date Description


1.0 2018/11/30 Document released.

No.: OMZ-21009-A Series Page 97of 428


PMLS Global Technical Support Dept

TMZ-18031 All Anesthesia

Anesthesia machine hardware upgrade and service manual update

Technical News
Subject: Anesthesia machine hardware upgrade and service manual update

RECORD IDENTIFIER

Ref No.: TMZ-18031 All Anesthesia Relate To:


Models
Version:1.0 Software/Manual Upgrade
Author: Dave Hao Parts Change
Date: Oct 19th, 2018 Appearance
Model: All anesthesia machine Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,


1. A Series Anesthesia machine Main Control board and Flow Control board change
Description

Because of old flash chips stop production, new flash chips are used for A series main control
board and flow control board. Old and new Main Control board and Flow Control board are
compatible. New software (V02.21.01.00) are released to support new Main Control board
and Flow Control board.

Influence on installation
N/A
Influence on maintenance
Old and new Main Control board and Flow Control board are compatible. If new Old and
new Main Control board and Flow Control board need to be replaced, please upgrade the
latest software.
Influence on upgrade
N/A

2. A Series Auxiliary Gas Outlet change


Description

To solve auxiliary gas outlet gas leakage problem, new auxiliary gas outlet changed needle
valve design. Old and new auxiliary gas outlet are compatible.
No.: OMZ-21009-A Series Page 98of 428
PMLS Global Technical Support Dept

Influence on installation
N/A
Influence on maintenance
If auxiliary gas outlet has leakage problem, please apply auxiliary gas outlet assembly to
solve the problem.
Influence on upgrade
N/A

3. Anesthesia machine one way valve assembly change


Description
To avoid one way valve cannot install in place and get stuck one way valve assembly
changed design. Old and new one way valve assembly are compatible.

Influence on installation
N/A
Influence on maintenance
If gas outlet is much high than Vt setting, and “Tidal volume not reach” “Please check flow
sensor” alarm at the same time. Please apply one way valve to solve the problem.
Influence on upgrade
N/A

4. Anesthesia machine service manual update


Description
A Series, WATO EX-20/25/30/35/55/65/55Pro/65Pro service manual has been updated. You
can download the service manual from TDP.

A5 : /Anesthesia Machine/Ventilator/Anesthesia Machine/A5(0631&0634)/Manual/Service Manual

A7 :/Anesthesia Machine/Ventilator/Anesthesia Machine/A7(0632&0635)/Manual/Service Manual

WATO EX-35 :/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-35(0619)/Manual/Service Manual

WATO EX-55 :/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-55(0626)/Manual/Service Manual

WATO EX-65 :/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-65(0625)/Manual/Service Manual


WATO EX-65 Pro:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-65 Pro(0627)/Manual/Service
Manual
WATO EX-55 Pro:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-55 Pro(0628)/Manual/Service
Manual

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

No.: OMZ-21009-A Series Page 99of 428


PMLS Global Technical Support Dept

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Anesthesia machine hardware Ref No: TMZ-18031-All Anesthesia Models
upgrade and service manual update

Version Effective Date Description


1.0 2018-10-19 Document released.

No.: OMZ-21009-A Series Page 100of 428


PMLS Global Technical Support Dept

TMZ-18004- A series&WATO EX-20&20Vet&25

Anesthesia machine’s Ventilation&Control PCBA coating increase and user manual update

Technical News
Subject: Anesthesia machine’s Ventilation&Control PCBA coating increase and user manual
update

RECORD IDENTIFIER

Ref No.:TMZ-18004- A series&WATO EX- Relate To:


20&20Vet&25
&30&35&55&65&55Pro&65Pro
Version:1.0 Software/Manual Upgrade
Author: Dave Hao Parts Change
Date: Feb 2nd, 2018 Appearance
Model: WATO EX-20/20Vet/25/30/35/55/65/ Stop Production
55Pro/65Pro&A3&A5&A7 Others
(0615/0616/0617/0618/0619/0620/0623/0624/0625/
0626/0627/0628/0634/0635/0637)
Comments:N/A

CONTENT

To Whom It May Concern,

1. Patient breathing circuit cleaning and disinfection instruction card release

Description

Patient breathing circuit cleaning and disinfection instruction card released.


Customers can follow the instruction card to clean and disinfect anesthesia machine’s
patient breathing circuit. Patient breathing circuit cleaning and disinfection instruction card
shown as below:

No.: OMZ-21009-A Series Page 101of 428


PMLS Global Technical Support Dept

WATO EX-20,30, WATO EX-25,35,55 A Series Card


20Vet Card, 65,55Pro,65Pro Card

Cleaning and disinfection cards have uploaded to TDP, address lists below:

1. WATO EX-20,30,20vetcleaning and disinfection card


:/Anesthesia Machine/Ventilator/Anesthesia Machine/wato EX-20Vet(0620)/Manual/Operation guide
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-20(0615)/Manual/A.M. Quick Guide
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-30(0617)/Manual/A.M. Quick Guide
2. WATO EX-35,55,65,55Pro,65Pro cleaning and disinfection card
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-35(0619)/Manual/A.M.Quick Guide
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-55(0626)/Manual/A.M. Quick Guide
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-65(0625)/A.M. Quick Guide
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-55 Pro(0628)/Manual/A.M. Quick Guide
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-65 Pro(0627)/Manual/A.M. Quick Guide
3. A series cleaning and disinfection card
:/Anesthesia Machine/Ventilator/Anesthesia Machine/A5 (0631&0634)/Manual/A.M. Quick Guide
:/Anesthesia Machine/Ventilator/Anesthesia Machine/A7(0632&0635)/Manual/A.M. Quick Guide

Influence on installation

During anesthesia machine installation, engineer can use Patient breathing circuit
cleaning and disinfection instruction card to tell customer how to clean and disinfect
patient breathing circuit.

Influence on maintenance

N/A
Influence on upgrade

N/A
No.: OMZ-21009-A Series Page 102of 428
PMLS Global Technical Support Dept

2. Patient breathing circuit volume declaration changed

Description

Patient breathing circuit volume declaration changed in Operator’s manual as below:

Pre-pak breathing circuit volume: Auto mode 2.85L, Manual mode 1.8L;
Non pre-pak breathing circuit volume: Auto mode 2.6L, Manual mode 1.8L.

Latest operator’s manuals had uploaded to the following address:

1. WATO EX-20 Operation`s manual_V4.0_EN


:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-20(0615)/Manual/Operation Manual
2. WATO EX-20_Opreator's manual(4yoke)_V4.0_EN
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-20(0615)/Manual/Operation Manual
3. WATO EX-20Vet _Opreator's manual_V4.0_EN
:/Anesthesia Machine/Ventilator/Anesthesia Machine/wato EX-20Vet(0620)/Manual/Operation Manual
4. WATO EX-30_Operator`s manual_V3.0_EN
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-30(0617)/Manual/Operation Manual
5. WATO EX-35(0619)_Operator's Manual_V5.0_EN
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-35(0619)/Manual/Operation Manual
6. WATO EX-65_Operator's Manual_V8.0_EN
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-65(0625)/Operation Manual
7. WATO EX-55_Operator's Manual_V9.0_EN
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-55(0626)/Manual/Operation Manual
8. WATO EX-65 Pro_Operator's Manual_V3.0_EN
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-65 Pro(0627)/Manual/Operation Manual
9. WATO EX-55 Pro_Operator’s Manual_V3.0_EN
:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-55 Pro(0628)/Manual/Operation Manual

Influence on installation

N/A

Influence on maintenance

N/A

Influence on upgrade

N/A

3. Silicon coating Ventilation&Control PCBA released

No.: OMZ-21009-A Series Page 103of 428


PMLS Global Technical Support Dept
Description

If ventilation&control PCBA’s sensor polluted, tidal volume and pressure measured value
may not accurate. To avoid ventilation&control PCBA polluted, silicon coating
ventilation&control PCBA released.

Old and new Ventilation&Control PCBA shown as below:

Old(without coating) New(with coating)

Old and new Ventilation&Control PCBA relationship list as below:

No Old PCBA New PCBA


. Part description Part number Part description Part number
1 Monitoring signals detection board 0621-30-78632 0616 signal monitor board PCBA 051-003166-00
2 V01A Ventilater&Control(WATO) PCBA 051-001078-00 0625 Ventilation&Control PCBA 051-003167-00
3 0635 Ventilation&Control PCBA(NOVA) 051-002452-00 0635 Ventilation&Control PCBA 051-003168-00

Other spare parts, which related with PCBA shown above, substitution relationship list
below:

No.: OMZ-21009-A Series Page 104of 428


PMLS Global Technical Support Dept

No Old New
. Part description Part number Part description Part number
1 Monitoring board assembly 0621-30-78559 0623 Monitor board assembly 115-050887-00
2 Monitored signal detection 801-0616-00059-00 Monitored signal detection board 115-051193-00
board (FRU) (FRU)
3 0616 Monitor board assembly 115-011624-00 0616 Monitor board assembly 115-050886-00
4 Monitor Board assembly 801-0621-00006-00 Monitor Board assembly 115-051194-00
5 Monitor board assembly 801-0616-00117-00 Monitor board assembly 115-051195-00
6 Ventilator assembly (Air drive) 801-0621-00086-00 Ventilator assembly (Air drive) 115-051197-00
7 Ventilator assembly (O2 drive) 801-0621-00087-00 Ventilator assembly (O2 drive) 115-051198-00
8 ventilator(O2 drive) 115-034998-00 ventilator(O2 drive) 115-051199-00
9 ventilator(AIR drive) 115-034997-00 ventilator(AIR drive) 115-051200-00
10 0625 Monitor board assembly 115-031842-00 0625 Monitor board assembly 115-050879-00
11 0619 Monitor board assembly
115-035472-00 0619 Monitor board assembly 115-050878-00
(DSP)
12 0619 Ventilation kit FRU(O2
115-036952-00 0619 Ventilation kit FRU(O2 Drive) 115-051201-00
Drive)
13 0619 Ventilation kit FRU(Air
115-036953-00 0619 Ventilation kit FRU(Air Drive) 115-051202-00
Drive)
14 Breathing system FRU(Air drive) 115-033003-00 Breathing system FRU(Air drive) 115-051203-00
15 Breathing system FRU(O2 drive) 115-033002-00 Breathing system FRU(O2 drive) 115-051204-00
16 0635 Monitor board assembly
115-036401-00 0635 Monitor board assembly 115-050888-00
(NOVA)
17 Monitoring Board(DSP) FRU 115-036365-00 Monitoring Board(DSP) FRU 115-051205-00
18 Ventilator Control Board(DSP)
115-036366-00 Ventilator Control Board(DSP) FRU 115-051206-00
FRU
19 Ventilation kit(Non-
115-036364-00 Ventilation kit(Non-FDA/DSP)FRU 115-051207-00
FDA/DSP)FRU
20 Ventilation kit(Non-FDA/DSP/Air Ventilation kit(Non-FDA/DSP/Air
115-049798-00 115-051208-00
driven) driven)

Influence on installation

N/A

Influence on maintenance

Use new parts list above to replace old parts. New spare parts and old spare parts are
compatible.

Influence on upgrade

N/A
No.: OMZ-21009-A Series Page 105of 428
PMLS Global Technical Support Dept

4. Adjust all anesthesia machine minimum O2 concentration to 25%

Description

To meet marketing requirement, all anesthesia machine’s minimum O2 concentration


adjusted to 25%. Spare parts list below and user manuals are updated. New spare parts
and old spare parts are compatible.
No. Part number Part description
1 082-000099-00 6-tube flowmeter EU
2 082-000104-00 4-tube flowmeter EU
3 082-000950-00 6-tube flowmeter.EN color,US order
4 082-000951-00 4-tube flowmeter.EN color,US order
5 082-002115-00 O2&N2O Flow regulator(ANSI/Reverse EN)

The only difference between old and new flowmeter is gear ratio. There is no appearance
difference.

Influence on installation

N/A

Influence on maintenance

New spare parts and old spare parts are compatible.

Influence on upgrade

N/A

No.: OMZ-21009-A Series Page 106of 428


PMLS Global Technical Support Dept

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Anesthesia machine’s Ref No: TMZ-18004-A series&WATO EX-
Ventilation&Control PCBA coating 20&20Vet&25&30&35&55&65&55Pro&65Pro
increase and user manual update

Version Effective Date Description


1.0 2018/2/1 Document released.

No.: OMZ-21009-A Series Page 107of 428


PMLS Global Technical Support Dept

TMZ-18002-All Anesthesia Models

New monitor board with Protection used for India market prolong service time

Technical News

Subject: New monitor board with Protection used for India market prolong service time.

RECORD IDENTIFIER

Ref No.: TMZ-18002-All Anesthesia Models Relate To:


Version:1.0 Software/Manual Upgrade
Author:Dave Hao Parts Change
Date: Jan. 8th, 2018 Appearance
Model: All Anesthesia Models Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


Based on TCN17-CL001, new monitor board with Protection (Silicone conformal
coating) had been released in India for 3 months. Since TCN17-CL001 only has 3
months life time, now we extend the life time of TCN monitor board with CLTCN17-
CL001 to Jan 27th, 2018. Please apply enough boards as inventory in time so that
monitor board maintenance and replacement could be supported.

Part number Name Usable range


TCN17-CL001-003 0616 signal monitor board PCBA 0615 WATO EX-20
(SN: DW-ym000001)
0620 WATO EX-20Vet
(SN: HK-ym000001)
0617 WATO EX-30
(SN: DY-ym000001)
0618 WATO EX-35
(SN: EA-ym000001)
0623 WATO EX-65
(SN: ER-ym000001)
0624 WATO EX-55
(SN: ES-ym000001)

No.: OMZ-21009-A Series Page 108of 428


PMLS Global Technical Support Dept
TCN17-CL001-004 0625 Ventilation&Control PCBA 0619 WATO EX-35
(SN: KG-ym000001)
0625 WATO EX-65
(SN: HX/KT-ym000001)
0626 WATO EX-55
(SN: HY/KU-ym000001)
0627 WATO EX-65Pro
(SN: KY-ym000001)
0628 WATO EX-55Pro
(SN: F1-ym000001)
TCN17-CL001-005 0635 Ventilation&Control PCBA 0631&0634 A5
(SN: EJ/FY-ym000001)
063270635 A7
(SN: FR/GC-ym000001)
Note:
1.Above boards are single board(board on the top), part of monitor assembly, not include
valve drive board
2.Software upgrade and factory calibration are necessary
3.WATO economic types(0615,0620,0617,0618) had one batch of DSP type, but new
board TCN17-CL001-003 is EPSON type, if need to replace monitor board on such
devices, also need to apply latest monitor board assembly(SN please refer to latest
corresponding type part list)
4.A5&A7 old units used EPSON platform monitor board, but new board TCN17-CL001-
005 is DSP type, if need to replace monitor board on such devices, also need to apply
latest monitor board assembly(SN please refer to latest corresponding type part list)

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
Abraham.ma@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: New monitor board with Ref No: TMZ-18002-All Anesthesia Models
Protection used for India market prolong
service time.

Version Effective Date Description


1.0 2018-1-8 Document released.

No.: OMZ-21009-A Series Page 109of 428


PMLS Global Technical Support Dept

TMZ-17032-A5&A7

Display change and new software is released

Technical News
Subject: Display change and new software is released.

RECORD IDENTIFIER

Ref No.: TMZ-17032-A5&A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Yang Parts Change
Date: 21th Nov ,2017 Appearance
Model:A5&A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

Hardware Change:
1, The display supplier of the A series anesthesia has changed. The parts for maintenance
changed as bellow:

Before After Compatibility Application Remark

0631 0631
A5,A7
051-001519- Backlight 051- Backlight No
/ before
00 Inverter 001519-00 Inverter changed
2015.12.31
PCBA PCBA
Backwards
compatible,
0635 LED 0632 Screen
051-002448- 051- new version A5,A7 after Old is
Backlight Backlight
00 002583-00 is compatible 2015.12.31 deleted
Board PCBA Board PCBA
with old
ones.
Display Display Display
115-036373- Exchange 115- Exchange Fully material
A5,A7
00 Package 036373-00 Package compatible changed,
(0634) (0634) but part

No.: OMZ-21009-A Series Page 110of 428


PMLS Global Technical Support Dept

Display Display A5 with number


115-036453- 115- Fully
screen screen O2/AIR/ is not
00 036453-00 compatible
FRU(0634) FRU(0634) N2O changed
Display Display Display
115-036476- screen 115- screen Fully A5 with material
00 FRU(0634/ 036476-00 FRU(0634/O compatible O2/AIR changed,
O2&Air) 2&Air) but part
Display Display number
115-036454- 115- Fully
screen screen A7 is not
00 036454-00 compatible
FRU(0635) FRU(0635) changed

Software Change:
1,To support transmission of data to Philips Monitors with IntelliBridge module, new
software of A series anesthesia with DSP monitor board is released.
Model Bundle version Remark

1,The Philips Monitors need the IntelliBridge


A3 02.19.00.03
module.

A5 02.19.00.03 2,Only support the AM with DSP monitor board,

not support the AM with EPSON monitor board.


A7 02.19.00.03

Others:
1,Add 2 new spare parts for maintenance:

Cylinder lock block

Tank knob

Order number Description


041-002678-00 cylinder lock block
043-001266-00 tank knob

No.: OMZ-21009-A Series Page 111of 428


PMLS Global Technical Support Dept

2,New operator’s manuals for A series anesthesia are released.


(a)Add serial setup about data transmission of Philips Monitors.

(b) Modify the specification about Breathing System Volume.

Before

Now
(c) Modify Low Flow Anesthesia from 1L/min to 0.5L/min.

Before

No.: OMZ-21009-A Series Page 112of 428


PMLS Global Technical Support Dept

Now

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Display change and new software Ref No: TMZ-17032-A5&A7
is released.

Version Effective Date Description


1.0 2017/11/21 Document released.

No.: OMZ-21009-A Series Page 113of 428


PMLS Global Technical Support Dept

TMZ-17031-All AM&Ventilator

Lithium battery discharge to 30% to meet the requirement of air transport

Technical News
Subject: Lithium battery discharge to 30% to meet the requirement of air transport

RECORD IDENTIFIER

Ref No.: TMZ-17031-All AM&Ventilator Relate To:


Version:1.0 Software/Manual Upgrade
Author: Dave Hao Parts Change
Date: Oct 16th 2017 Appearance
Model: All Ventilator and Anesthesia Stop Production
Machine models
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

To meet the requirement of air transport battery by airline company, ventilators and
anesthesia machine’s lithium batteries need to be discharge to no more than 30%.
Influenced ventilator and anesthesia machine models are: WATO EX-20(0615/061B),
WATO EX-20Vet(0620), WATO EX-25(0616/0629), WATO EX-30(0617), WATO EX-
35(0618/0619), WATO EX-55(0624/0626), WATO EX-55 Pro(0628), WATO EX-
65(0623/0625), WATO EX-65 Pro(0627), A3(0637), A5(0634), A7(0635), SV300(EV20),
E3(VT30), E5(VT50),

Change description:
If the order of lithium battery transported by air, lithium battery need to be discharged to
no more than 30%, and paste “Label for Cargo Aircraft Only”.

Orders for international non-air transport, batteries do not need to be discharged and do
not require the label.

Shipping method and Machine model shown on the top right corner of the label.
Sample of label shown as below:

No.: OMZ-21009-A Series Page 114of 428


PMLS Global Technical Support Dept

Influenced materials:
The table below list all influenced materials

Material Number Material Name


115-036370-00 3-year Maintenance package (0635)
801-0623-00016-00 3-year Maintenance package (0623/0625/Pre-
park)
801-0623-00015-00 3-year Maintenance package (0623/0625/old
circuit)
801-VT50-00115-00 3-year Maintenance package
801-0621-00134-00 3-year Maintenance package FRU
801-0611-00103-00 Li-ion 11V 4400mAhLI23S001A
801-0613-00066-00 3-year Maintenance package FRU
115-025022-00 Lithium battery package
115-018011-00 Lithium battery 11.1V 4400mAh LI23S001A
Package
115-018011-00 Lithium battery 11.1V 4500mAh LI23S002A
Package

No.: OMZ-21009-A Series Page 115of 428


PMLS Global Technical Support Dept
Attention:
Because of the discharge of the battery, air-transport battery may run out of electricity
quicker than non air-transport battery.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Lithium battery discharge to 30% Ref No: TMZ-17031-All AM&Ventilator
to meet the requirement of air transport

Version Effective Date Description


1.0 2017/10/16 Document released.

No.: OMZ-21009-A Series Page 116of 428


PMLS Global Technical Support Dept

TMZ-17030-A5&A7

A5&A7 manual bag arm changes

Technical News
Subject: A5&A7 manual bag arm changes

RECORD IDENTIFIER

Ref No.: TMZ-17030-A5&A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Yang Parts Change
Date: 11th Oct ,2017 Appearance
Model:A5&A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

1, For solving the sweat marks problem in the case of long-term use, the surface coating
of the manual bag arm assembly is changed from nickel to chrome.
Before After

Difference of
appearance

Metal Gray Sliver White

manual bag arm


Spare Bag arm assembly
115-017033-00 assembly 115-048600-00
Parts (silvery/spacing)
FRU(0631)—Delete
Manual bag arm All devices use the modified manual bag arm assembly for
Maintenance
failure replacement.
Upgrade N/A

No.: OMZ-21009-A Series Page 117of 428


PMLS Global Technical Support Dept

2, The width of the manual bag arm assembly support tube limit block increases.
Before After
Top view Limit block Wider
gray
Difference of
appearance

Bottom view

Mid-plate
Mid-plate FRU(0631/no
115-021249-00 FRU(0631/no 115-021249-00
ventilation valve)
ventilation valve)
Breathing Circuit Breathing Circuit
115-025564-00 115-025564-00
Cell(0635) Cell(0635)
Spare
Breathing Circuit Breathing Circuit
Parts 115-025565-00 115-025565-00
Cell(0634) Cell(0634)
Breathing system Breathing system
115-025569-00 115-025569-00
FRU(0634) FRU(0634)
801-0631-00105- 801-0631-
Breathing System Breathing System
00 00105-00

Mid-plate failure
Part number is no change. All machines use old part
Maintenance
Breathing Circuit number for maintain and replacement
failure
Upgrade N/A

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 118of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: Ref No:
A5&A7 manual bag arm changes TMZ-17030-A5&A7

Version Effective Date Description


1.0 2017-10-11 Document released.

No.: OMZ-21009-A Series Page 119of 428


PMLS Global Technical Support Dept

TMZ-17028-All Anesthesia Models

New monitor board with Protection had been released for India

Technical News
Subject: New monitor board with Protection had been released for India.

RECORD IDENTIFIER

Ref No.: TMZ-17028-All Anesthesia Models Relate To:


Version:1.0 Software/Manual Upgrade
Author:Abraham Parts Change
Date: Sep. 30th, 2017 Appearance
Model: All Anesthesia Models Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


New monitor board with Protection (Silicone conformal coating) had been released for
India.
Since this technical change is TCN which means the parts released here only have 3
months life time, please apply enough boards as inventory in time so that monitor board
maintenance and replacement could be supported.
Part number Name Usable range
TCN17-CL001-003 0616 signal monitor board PCBA 0615 WATO EX-20
(SN: DW-ym000001)
0620 WATO EX-20Vet
(SN: HK-ym000001)
0617 WATO EX-30
(SN: DY-ym000001)
0618 WATO EX-35
(SN: EA-ym000001)
0623 WATO EX-65
(SN: ER-ym000001)
0624 WATO EX-55
(SN: ES-ym000001)

No.: OMZ-21009-A Series Page 120of 428


PMLS Global Technical Support Dept
TCN17-CL001-004 0625 Ventilation&Control PCBA 0619 WATO EX-35
(SN: KG-ym000001)
0625 WATO EX-65
(SN: HX/KT-ym000001)
0626 WATO EX-55
(SN: HY/KU-ym000001)
0627 WATO EX-65Pro
(SN: KY-ym000001)
0628 WATO EX-55Pro
(SN: F1-ym000001)
TCN17-CL001-005 0635 Ventilation&Control PCBA 0631&0634 A5
(SN: EJ/FY-ym000001)
063270635 A7
(SN: FR/GC-ym000001)
Note:
1.Above boards are single board(board on the top), part of monitor assembly, not include
valve drive board
2.Software upgrade and factory calibration are necessary
3.WATO economic types(0615,0620,0617,0618) had one batch of DSP type, but new
board TCN17-CL001-003 is EPSON type, if need to replace monitor board on such
devices, also need to apply latest monitor board assembly(SN please refer to latest
corresponding type part list)
4.A5&A7 old units used EPSON platform monitor board, but new board TCN17-CL001-
005 is DSP type, if need to replace monitor board on such devices, also need to apply
latest monitor board assembly(SN please refer to latest corresponding type part list)

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
Abraham.ma@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: New monitor board with Ref No: TMZ-17028-All Anesthesia Models
Protection had been released for India.

Version Effective Date Description


1.0 2017-09-30 Document released.

No.: OMZ-21009-A Series Page 121of 428


PMLS Global Technical Support Dept

TMZ-17027-All AM&Ventilator

Anesthesia Machine&Ventilator Parts List is released

Technical News
Subject: Anesthesia Machine&Ventilator Parts List is released.

RECORD IDENTIFIER

Ref No.: TMZ-17027-All AM&Ventilator Relate To:


Version:1.0 Software/Manual Upgrade
Author:Robert.Ling Parts Change
Date Sep 30th,2017 Appearance
Model: All AM&Ventilator models Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

1,Anesthesia Machine&Ventilator_Parts_List_ V170901 is released.


Parts list is uploaded to TDP: /Anesthesia Machine/Ventilator/Parts List

If you have any questions & suggestions, or you find any mistakes, please
feedback to Robert (lingjun@mindray.com).Thanks a lot for your support.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 122of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: Anesthesia Machine&Ventilator Ref No: TMZ-17027-All AM&Ventilator
Parts List is released.

Version Effective Date Description


1.0 2017-09-30 Document released.

No.: OMZ-21009-A Series Page 123of 428


PMLS Global Technical Support Dept

TMZ-17021-A7

A7 New Software released & APL valve change

Technical News
Subject: A7 New Software released & APL valve change.

RECORD IDENTIFIER

Ref No.: TMZ-17021-A7 Relate To:


Version:2.0 Software/Manual Upgrade
Author: Shiling Yang Parts Change
Date: 7th,Jun,2017 Appearance
Model:A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,


◼ Software
1, The new software for A7 with DPS monitor is launched.
No. Package Modules Latest Version
SystemUpdateTool
00 Tool 5.8
(V5.8)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)
Bundle Version 02.18.02.00
Host Software 03.18.02.00
FPGA Display 1.4
FPGA Sound 1.6
Bundle
02 Ventilator Protect
(V02.18.02.00) V01.02.00
Module
Ventilator Control
V01.04.00
Module
Language 4.6
Keyboard 01.00/01.01
03 Power(V1.7) Power System V1.7
EFCS
04 EFCS Software V50.07.00.04
(V50.06.00)

No.: OMZ-21009-A Series Page 124of 428


PMLS Global Technical Support Dept
EFCS_FPGA
05 FPGA EFCS V1.0
(V3.0)
This Software is not compatible with the devices which have software version
02.05.10.03 or lower version, for more details please refer to the Software upgrade
instruction in software package.
NOTE:
1, New software can only be used on the DSP board. The EPSON monitor board will be
damaged when using the new software.
2, When the devices which have bundle version 02.05.10.03 or lower need to upgrade to the
02.18.02.00 of the Bundle version, please upgrade software to 02.05.10.03 first ,then change
the Ventilator Control Board(DSP) FRU and upgrade the software to 02.18.02.00
3, Downgrade the higher version to lower version is not allowed.
4,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.18.02.00 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test, please perform
steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Main】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Main】->【General】->【Zero Flow Meters】,zero flow meters and
restart again;
For those devices which have bundle version 02.05.00 or higher, step a, b, c are unnecessary
during upgrading to 02.18.02.00.

The Software Technical Changes include:


1. Solve a problem: 100% O2 sensor calibration of the A7 occasional unsuccessful. Only the
A7’versions of following table are affected, the other versions do not have this problem.
Mode Monitor type Version of bundle
A7 DSP 02.15.00.00
A7 DSP 02.15.01.00
A7 DSP 02.16.00.03
A7 DSP 02.18.00.04
A7 DSP 02.18.01.02

◼ Hardware
1. The color of APL valve is changed from black to gray.
Before After

Difference of
Black Gray
appearance

No.: OMZ-21009-A Series Page 125of 428


PMLS Global Technical Support Dept

Spare Quick Release APL Valve Quick Release APL


115-012263-00 115-012263-00
Parts Assembly Valve Assembly

APL valve
Maintenance All of the devices use the same part number for replacement.
failure

Upgrade N/A

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A7 New Software released & APL Ref No: TMZ-17021-A7
valve change.

Version Effective Date Description


1.0 2017-06-07 Document released.

No.: OMZ-21009-A Series Page 126of 428


PMLS Global Technical Support Dept

TMZ-17020-A5

A5 New Software released & APL valve change

Technical News
Subject: A5 New Software released & APL valve change.

RECORD IDENTIFIER

Ref No.: TMZ-17020-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Yang Parts Change
Date: 6th,Jun,2017 Appearance
Model:A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

◼ Software
1,The new software for A5 with DPS monitor is launched.
No. Package Modules Latest Version
SystemUpdateTo
00 ol Tool 5.8
(V5.7)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)
Bundle Version 02.18.02.00
Host Software 03.18.02.00
FPGA Display 1.4
FPGA Sound 1.6
Bundle
02 Ventilator Protect
(V02.18.02.00) V01.02.00
Module
Ventilator Control
V01.04.00
Module
Language 4.6
flowmeter V1.0
03 Power(V1.7) Power System V1.7

No.: OMZ-21009-A Series Page 127of 428


PMLS Global Technical Support Dept
This Software is not compatible with the devices which have software version
02.05.10.03 or lower version, for more details please refer to the Software upgrade
instruction in software package.
NOTE:
1, New software can only be used on the DSP board. The EPSON monitor board will be
damaged when using the new software.
2, When the devices which have bundle version 02.05.10.03 or lower need to upgrade to the
02.18.20.00 of the Bundle version, please upgrade software to 02.05.10.03 first ,then change
the Ventilator Control Board(DSP) FRU and upgrade the software to 02.18.02.00
3, Downgrade the higher version to lower version is not allowed.

◼ Hardware
1. The color of APL valve is changed from black to gray.
Before After

Black Gray

Difference of
appearance

Spare APL Valve


801-0631-00062-00 115-046754-00 APL valve FRU
Parts Assembly--delete

All of the devices use the modified APL valve for


Maintenance APL valve failure
replacement.

Upgrade N/A

2. Increase the length of the flow sensor cable, and the ground connection.
Before After

Short, no ground position Longer, has the ground position

Difference of
appearance

No.: OMZ-21009-A Series Page 128of 428


PMLS Global Technical Support Dept
E_Flow assembly E_Flow assembly
Spare
115-044266-00 FRU(0634/O2&N2O& 115-044266-00 FRU(0634/O2&N2O&A
Parts
Air) ir)

Electronic flow
Maintenance All of the devices use the same part number for replacement.
sensor failure

Upgrade N/A

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A5 New Software released & APL Ref No: TMZ-17020-A5
valve change.

Version Effective Date Description


1.0 2017-06-06 Document released.

No.: OMZ-21009-A Series Page 129of 428


PMLS Global Technical Support Dept

TMZ-17014-WATO&A series

The measuring mode of Flow and Volume for WATO and A series anestheisa machines is
changed from ATPD to BTPS

Technical News
Subject: The measuring mode of Flow and Volume for WATO and A series anestheisa
machines is changed from ATPD to BTPS.

RECORD IDENTIFIER

Ref No.: TMZ-17014-WATO&A series Relate To:


Version: V1.0 Software/Manual Upgrade
Author: David Wang Parts Change
Date: 2017.04.17 Appearance
Model: WATO,A series Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

1. The measuring mode of Flow and Volume for WATO and A series anesthesia
machines is changed from ATPD to BTPS
Change Description
In order to meet product regulations requirement, the measuring mode of Flow and
Volume for WATO and A series anesthesia machines is changed from ATPD to BTPS.
Because of this change, the description about factory calibration ventilator
parameters test in service manual of WATO EX-20/30/35 need to be update at the same
time. The new manual has been uploaded to TDP:
(1) View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-20(0615)/Manual/Service
Manual
(2) View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-30(0617)/Manual /Service
Manual
(3) View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Wato-35(0618)/Manual /Service
Manual
Influence to installation business
During installation, if want to test the ventilator parameters, please change the
measuring mode of calibration device to BTPS. Please refer to the service manual of
WATO or A series for more details about the test.

No.: OMZ-21009-A Series Page 130of 428


PMLS Global Technical Support Dept
Influence to maintenance business
During maintenance, if want to do factory calibration or test the ventilator parameters,
please change the measuring mode of calibration device to BTPS. Please refer to the
service manual of WATO or A series for more details about the test.
Influence to upgrading business
N/A

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: The measuring mode of Flow and Ref No: TMZ-17014-WATO&A series
Volume for WATO and A series anestheisa
machines is changed from ATPD to BTPS.

Version Effective Date Description


1.0 2017-04-18 Document released.

No.: OMZ-21009-A Series Page 131of 428


PMLS Global Technical Support Dept

TMZ-17012-A7

A7 New Software released

Technical News
Subject: A7 New Software released.

RECORD IDENTIFIER

Ref No.: TMZ-17012-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Yang Parts Change
Date: 12th,Apr,2017 Appearance
Model:A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

◼ Software
1,Two new software is launched:
(1)First: Software for devices with EPSON monitor board.
No. Package Modules Latest Version

00 SystemUpdateTool Tool 5.8

BIOS
01 BIOS 01.12.00.00
(01.12.00.00)

Bundle Version 02.05.10.03

Host Software 03.05.09.03

FPGA Display 1.4

Bundle FPGA Sound 1.2


02
(V02.05.10.03) Ventilator Protect Module V1.1

Ventilator Control Module V01.15.00

Language 3.9

Keyboard 01.00/01.01

03 Power(V1.7) Power System V1.7


EFCS
04 EFCS Software V50.07.00.04
(V50.07.00.04)

No.: OMZ-21009-A Series Page 132of 428


PMLS Global Technical Support Dept
EFCS_FPGA
05 FPGA EFCS V1.0
(V1.0)
NOTE:
1, New software can only be used on the new EPSON monitor board. The DSP monitoring
board will be damaged when using the new software.
2, Downgrade the higher version to lower version is not allowed.
3,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.05.10.03 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test, please perform
steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Main】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Main】->【General】->【Zero Flow Meters】,zero flow meters and
restart again;
For those devices which have bundle version 02.05.00 or higher, steps a, b, c are unnecessary
during upgrading to 02.05.10.03.

(2)Second: Software for devices with DSP monitor board.


No. Package Modules Latest Version

SystemUpdateTool
00 Tool 5.8
(V5.8)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)

Bundle Version 02.18.01.02

Host Software 03.18.01.02

FPGA Display 1.4

Bundle FPGA Sound 1.6


02
(V02.18.01.02) Ventilator Protect Module V01.02.00

Ventilator Control Module V01.04.00

Language 4.6

Keyboard 01.00/01.01

03 Power(V1.7) Power System V1.7

04 EFCS(V50.06.00) EFCS Software V50.07.00.04


EFCS_FPGA
05 FPGA EFCS V1.0
(V3.0)
This Software is not compatible with the devices which have software version 02.05.09 or lower
version, for more details please refer to the Software upgrade instruction in software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.
2, When the devices which have bundle version 02.05.10.03 or lower need to upgrade to the
02.16.00.03 of the Bundle version, please upgrade software to 02.05.10.03 first ,then change
No.: OMZ-21009-A Series Page 133of 428
PMLS Global Technical Support Dept
the Ventilator Control Board(DSP) FRU and upgrade the software to 02.18.01.02
3, Downgrade the higher version to lower version is not allowed.
4,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.18.01.02 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test, please perform
steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Main】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Main】->【General】->【Zero Flow Meters】,zero flow meters and
restart again;
For those devices which have bundle version 02.05.00 or higher, step a, b, c are unnecessary
during upgrading to 02.18.01.02.

The Software Technical Changes include:


1. Change the "load configuration failure" alarm display method from shows on screen continuously
to only keeps 10 seconds.
2. Solve a problem: the parameters and waveform of Mindray CO2 cannot be displayed and the CO2
module status lights keeps flashing, after the CO2 module stayed in standby mode for more than 10
minutes.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A7 New Software released Ref No: TMZ-17012-A7

Version Effective Date Description


1.0 2017-04-14 Document released.

No.: OMZ-21009-A Series Page 134of 428


PMLS Global Technical Support Dept

No.: OMZ-21009-A Series Page 135of 428


PMLS Global Technical Support Dept

TMZ-17011-A5

A5 New Software released

Technical News
Subject: A5 New Software released.

RECORD IDENTIFIER

Ref No.: TMZ-17011-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Yang Parts Change
Date: 12th,Apr,2017 Appearance
Model:A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

◼ Software
1,Two new software is launched:
(1)First: Software for devices with EPSON monitor board.
No. Package Modules Latest Version

00 SystemUpdateTool Tool 5.8

BIOS
01 BIOS 01.12.00.00
(01.12.00.00)

Bundle Version 02.05.10.03

Host Software 03.05.09.03

FPGA Display 1.4

Bundle FPGA Sound 1.2


02
(V02.05.10.03) Ventilator Protect Module V1.1

Ventilator Control Module V01.15.00

Language 3.9

flowmeter V1.0

03 Power(V1.7) Power System V1.7

NOTE:
1, New software can only be used on the new EPSON monitor board. The DSP monitoring

No.: OMZ-21009-A Series Page 136of 428


PMLS Global Technical Support Dept
board will be damaged when using the new software.
2, Downgrade the higher version to lower version is not allowed.

(2)Second: Software for devices with DSP monitor board.


No. Package Modules Latest Version

SystemUpdateTool
00 Tool 5.8
(V5.7)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)

Bundle Version 02.18.01.02

Host Software 03.18.01.02

FPGA Display 1.4

Bundle FPGA Sound 1.6


02
(V02.18.01.02) Ventilator Protect Module V01.02.00

Ventilator Control Module V01.04.00

Language 4.6

flowmeter V1.0

03 Power(V1.7) Power System V1.7


This Software is not compatible with the devices which have software version 02.05.10.03 or
lower version, for more details please refer to the Software upgrade instruction in software
package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.
2, When the devices which have bundle version 02.05.10.03 or lower need to upgrade to the
02.16.00.03 of the Bundle version, please upgrade software to 02.05.10.03 first ,then change
the Ventilator Control Board(DSP) FRU and upgrade the software to 02.16.00.03
3, Downgrade the higher version to lower version is not allowed.

The Software Technical Changes include:


1. Change the "load configuration failure" alarm display method from shows on screen continuously
to only keeps 10 seconds.
2. Solve a problem: the parameters and waveform of Mindray CO2 cannot be displayed and the CO2
module status lights keeps flashing, after the CO2 module stayed in standby mode for more than 10
minutes.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

No.: OMZ-21009-A Series Page 137of 428


PMLS Global Technical Support Dept

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A5 New Software released. Ref No: TMZ-17011-A5

Version Effective Date Description


1.0 2017-04-14 Document released.

No.: OMZ-21009-A Series Page 138of 428


PMLS Global Technical Support Dept

TMZ-17010-All AM&Ventilator

Anesthesia Machine&Ventilator Parts List is released

Technical News
Subject: Anesthesia Machine&Ventilator Parts List is released.

RECORD IDENTIFIER

Ref No.: TMZ-17010-All AM&Ventilator Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shiling Parts Change
Date Mar.31th.2017 Appearance
Model: All AM&Ventilator models Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

1,Anesthesia Machine&Ventilator_Parts_List_ V170301 is released.


Parts list is uploaded to TDP: /Anesthesia Machine/Ventilator/Parts List

If you have any questions & suggestions, or you find any mistakes, please
feedback to Shiling (yangshiling@mindray.com).Thanks a lot for your support.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 139of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: Anesthesia Machine&Ventilator Ref No: TMZ-17010-All AM&Ventilator
Parts List is released.

Version Effective Date Description


1.0 2017-03-31 Document released.

No.: OMZ-21009-A Series Page 140of 428


PMLS Global Technical Support Dept

TMZ-17008-WATO EX-35&55&65&55 Pro&65 Pro&A5&A7

Electronic gas supply pressure monitoring released for WATO EX-55 PRO/65 PRO
(0628/0627) and new bundle software released in ECL134

Technical News
Subject: Electronic gas supply pressure monitoring released for WATO EX-55 PRO/65 PRO
(0628/0627) and new bundle software released in ECL134.

RECORD IDENTIFIER

Ref No.: TMZ-17008-WATO EX-35&55&65&55 Relate To:


Pro&65 Pro&A5&A7
Version:1.0 Software/Manual Upgrade
Author: Raymond Zheng Parts Change
nd
Date: Jan. 22 , 2017 Appearance
Model: WATO EX-35/55/65/55 Pro/65 Pro Stop Production
(0619/0626/0625/0628/0627), A5/A7(0634/0635)
Comments: N/A Others

CONTENT

To Whom It May Concern,


Below content is only for your information.

Release new bundle software of new WATO series and A series anesthesia machine and new
electronic pressure monitoring for WATO EX-55 PRO/65 PRO (0628/0627).

1. Hardware Change:
(1) Release electronic gas supply pressure monitoring for WATO EX-55 PRO/65 PRO
(0628/0627). The interface is as below:

New spare parts:


Item
P/N Descriptions Applicable models
no.

No.: OMZ-21009-A Series Page 141of 428


PMLS Global Technical Support Dept
1 115-044984-00 O2 origin inlet module(0627/NIST) WATO EX-55PRO/65PRO(0628/0627)
2 115-044985-00 N2O origin inlet module(0627/NIST) WATO EX-55PRO/65PRO(0628/0627)
3 115-044986-00 Air origin inlet module(0627/NIST) WATO EX-55PRO/65PRO(0628/0627)
4 115-044987-00 O2 origin inlet module(0627/DISS) WATO EX-55PRO/65PRO(0628/0627)
5 115-044988-00 N2O origin inlet module(0627/DISS) WATO EX-55PRO/65PRO(0628/0627)
6 115-044989-00 Air origin inlet module(0627/DISS) WATO EX-55PRO/65PRO(0628/0627)
7 115-044990-00 O2 cylinder yoke assembly(0627) WATO EX-55PRO/65PRO(0628/0627)
8 115-044991-00 N2O cylinder yoke assembly(0627) WATO EX-55PRO/65PRO(0628/0627)
9 115-044992-00 Air cylinder yoke assembly(0627) WATO EX-55PRO/65PRO(0628/0627)
10 115-045157-00 O2 supply inlet assembly(NIST/0627) WATO EX-55PRO/65PRO(0628/0627)
NOTES:
(1) This configuration can’t be upgraded on site.
(2) While updating software for machine with this configuration, please pay attention to the bundle
version. The bundle version should be after V04.05.00.02.

(2) New WATO series anesthesia machines add AGSS 3 ways connector assembly.
Newly added spare parts:
Item
P/N Descriptions Applicable models
no.
WATO EX-35/55/65/55Pro/65Pro
1 115-026796-00 AGSS 3 ways connector assembly
(0619/0626/0625/0628/0627)
WATO EX-35/55/65/55Pro/65Pro
2 115-030332-00 AGSS kit (0634/low flow/CE)
(0619/0626/0625/0628/0627)
WATO EX-35/55/65/55Pro/65Pro
3 115-030333-00 AGSS kit (0634/high flow/CE)
(0619/0626/0625/0628/0627)
NOTES:
(1) AGSS kits add the 3 ways connector assembly.
(2) A series anesthesia machines have this connector yet.

(3) Release autoclavable flow sensor as optional accessory.


Newly added spare parts:
Item
P/N Descriptions Applicable models
no.
1 115-041519-00 WATO EX-35/55/65/55Pro/65Pro, A5/A7
Flow sensor component
(0619/0626/0625/0628/0627, 0634/0635)
2 115-041507-00 Inspiration flow sensor WATO EX-35/55/65/55Pro/65Pro, A5/A7
component (0619/0626/0625/0628/0627, 0634/0635)
3 115-041508-00 Expiration flow sensor WATO EX-35/55/65/55Pro/65Pro, A5/A7
component (0619/0626/0625/0628/0627, 0634/0635)
NOTES:
(1) List price will be effective in March.
(2) New flow sensors have the signs as below:

No.: OMZ-21009-A Series Page 142of 428


PMLS Global Technical Support Dept

E16090001 is the SN of the flow sensor. 134℃ shows it can support 134℃ autoclave.

2. Software change:
Released new WATO series anesthesia machine bundle software V04.05.00.02 and A series
anesthesia machine bundle software V02.18.00.04.

Item What shall


Feature Descriptions New version Remark
no. you do?
1 Release new To support electronic gas Please Bundle Affected models:
function: supply pressure monitoring. upgrade: software: WATO EX-
Electronic New software interface: Hardware V04.05.00.02 55PRO/65PRO
pressure Software (0628/0627)
monitoring

2 System Solve the loading Please Bundle Affected models:


software configuration failure upgrade: software: WATO EX-25/35/55/
optimizations problem. Hardware V04.05.00.02 65/55Pro/65Pro(0619
NOTES: When this failure Software /0629/0626/0625/062
happens, please update this 8/0627)
software and no need to
replace the main control
board.
3 System Solve the problem that Please Bundle Affected models:
software “Fresh gas flow too high” upgrade: software: WATO EX-
optimizations alarm can not be triggered Hardware V04.05.00.02 55Pro/65Pro(0628/06
in certain situation. Software 27)
4 Release new New software interface: Please Bundle Affected models:
function: upgrade: software: WATO EX-
Recruitment. Hardware V04.05.00.02 55Pro/65Pro(0628/06
Software 27)

No.: OMZ-21009-A Series Page 143of 428


PMLS Global Technical Support Dept

NOTES: If you want to get


this function, you need to
update the new software
and place an order for
activation code (P/N: 115-
045689-00).
5 System Add flow paused function. Please Bundle Affected models:
software All fresh gas flow, ventilation upgrade: software: WATO EX-
optimizations and physiological alarms will Hardware V04.05.00.02 55Pro/65Pro(0628/06
be paused in this mode. The Software 27)
function interface:

6 System CO2 modules support Please Bundle Affected models:


software Optimizer function. upgrade: software: WATO EX-
optimizations Hardware V04.05.00.02 55Pro/65Pro(0628/06
Software 27)
7 System When the balance gas is air, Please Bundle Affected models:
software the O2 flow must be higher upgrade: software: WATO EX-
optimizations than 0.2L/min. Hardware V04.05.00.02 55Pro/65Pro(0628/06
Software 27)
8 System Add Bulgarian and Dutch Please Bundle Affected models:
software language. upgrade: software: WATO EX-
optimizations Hardware V04.05.00.02 55Pro/65Pro(0628/06
Software 27)
9 System Optimize the following Please Bundle Affected models:
software problem: upgrade: software: WATO EX-25/35/55/
optimizations When connecting to CMS, Hardware WATO 65/55Pro/65Pro(0619
anesthesia machines alarms Software series: /0629/0626/0625/062
remain on CMS screen even V04.05.00.02 8/0627)
though they were eliminated A series: A5/A7(00634/0635)
locally. V02.18.00.04
10 System Optimize the character Please Bundle Affected models:
software string and icon to meet the upgrade: software: WATO EX-25/35/55/

No.: OMZ-21009-A Series Page 144of 428


PMLS Global Technical Support Dept
optimizations user’s using requirement. Hardware V04.05.00.02 65/55Pro/65Pro(0619
Software /0629/0626/0625/062
8/0627)

11 System Add ACGO watermark in Please Bundle Affected models:


software ACGO mode when machine upgrade: software: WATO EX-25/35/55/
optimizations is in standby mode. Hardware WATO 65/55Pro/65Pro(0619
Software series: /0629/0626/0625/062
V04.05.00.02 8/0627)
A series: A5/A7(0634/0635)
V02.18.00.04
12 System New WATO series Please Bundle Affected models:
software anesthesia machine add upgrade: software: WATO EX-25/35/55/
optimizations auto exit standby on/off Hardware V04.05.00.02 65/55Pro/65Pro(0619
options. Software /0629/0626/0625/062
8/0627)

13 System A series anesthesia Please Bundle Affected models:


software machine can display Vt upgrade: software: A5/A7(0634/0635)
optimizations insp/exp value in the loop Hardware V02.18.00.04
interface. Software

Software version:
Item
Descriptions Old version New version Compatibility
no.
1 New WATO series bundle software V04.04.01.00 V04.05.00.02 One-way
compatible
package
2. A series bundle software package V 02.16.00.03 V02.18.00.04 One-way
compatible

3. Manual
Item
P/N Version Descriptions
no.
1 H-046-008766-00 2.0 WATO EX-35_Operator's Manual_V2.0_Brazilian Portuguese

2 H-046-008424-00 2.0 WATO EX-35_Service Manual_V2.0_EN

3 H-046-007082-00 3.0 WATO EX-55,65_Service Manual_V3.0_EN

4 H-046-007098-00 6.0 WATO EX-65_Operator's Manual_V6.0_EN

5 H-046-007113-00 7.0 WATO EX-55_Operator's Manual_V7.0_EN

No.: OMZ-21009-A Series Page 145of 428


PMLS Global Technical Support Dept

6 H-046-008702-00 2.0 WATO EX-65_Operator's Manual_V2.0_Brazilian Portuguese

7 H-046-008703-00 2.0 WATO EX-55_Operator's Manual_V2.0_Brazilian Portuguese

8 H-046-007110-00 2.0 WATO EX-65 Pro_Operator's Manual_V2.0_Brazilian

Portuguese

9 H-046-007125-00 2.0 WATO EX-55 Pro_Operator's Manual_V2.0_Brazilian

Portuguese

10 H-046-009615-00 2.0 WATO EX-65 Pro_Operator's Manual_V2.0_EN

11 H-046-009616-00 2.0 WATO EX-55 Pro_Operator's Manual_V2.0_EN

12 H-046-009688-00 2.0 WATO EX-55,65 Pro_Service Manual_V2.0_EN

If you need download information or have any question, please feel free to contact Mindray regional
service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: New electronic pressure Ref No: TMZ-17008-WATO EX-
monitoring for WATO EX-55 PRO/65 PRO 35&55&65&55 Pro&65 Pro&A5&A7
(0628/0627) and new bundle software
released in ECL134

Version Effective Date Description


1.0 2017-02-06 Document released.

No.: OMZ-21009-A Series Page 146of 428


PMLS Global Technical Support Dept

TMZ-16089-A7

ECL128 Engineering changes

Technical News
Subject: ECL128 Engineering changes.

RECORD IDENTIFIER

Ref No.: TMZ-16089-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Yang Parts Change
Date: 30th,Dec,2016 Appearance
Model:A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

◼ Hardware
1,The flow sensor of EFCS is replaced by the new type:
Before after

Same appearance, only support a single gas Support 3 kinds of gas (O2, N2O, and AIR)

Difference of
appearance

0632 Flowmeter O2
051-001414-01 flow sensor board
PCBA ---delete
0632 Flowmeter Air
Spare Flow sensor board
051-001415-01 sensor board PCBA 115-044851-00
Parts PCBA FRU
---delete
0632 Flowmeter N2O
051-001416-01 sensor board PCBA
---delete

Maintenance O2 flow senor failure All of the devices use the modified flow sensor for replacement.

No.: OMZ-21009-A Series Page 147of 428


PMLS Global Technical Support Dept
Note:
N2O flow senor failure
1, Software Upgrade is necessary after change the flow sensor. you need

AIR flow senor failure to select software according to the type of monitor board(EPSON or
DSP )
2, For the devices which configured O2/N2O/AIR gas supply, please refer
Total flow senor failure to the service manual for calibrating Total flow sensor, when replacing
the Electronic control board or N2O flow sensor or total flow sensor.
Upgrade N/A

2,Structure of Mechanical ACGO assembly changed


before after

Different installation of metal plate

Difference of
appearance Different interface location

ACGO
ACGO
801-0631-00143-00 assembly(A5-CE) 115-043752-00
Spare assembly(0634)
FRU ---delete
Parts
mechanical ACGO
115-017878-00 115-043753-00
ACGO(AG) ---delete assembly(0634/AG)

Mechanical ACGO All of the devices use the modified ACGO assembly for
Maintenance
assembly failure replacement.

Upgrade N/A

◼ Software
1,Two new software is launched:
(1)First: Software for devices with EPSON monitor board.
No. Package Modules Latest Version

00 SystemUpdateTool Tool 5.7

No.: OMZ-21009-A Series Page 148of 428


PMLS Global Technical Support Dept
BIOS
01 BIOS 01.12.00.00
(01.12.00.00)

Bundle Version 02.05.09.00

Host Software 03.05.08.00

FPGA Display 1.4

Bundle FPGA Sound 1.2


02
(V02.05.09.00) Ventilator Protect Module V1.1

Ventilator Control Module V01.15.00

Language 3.9

Keyboard 01.00/01.01

03 Power(V1.7) Power System V1.7


EFCS
04 EFCS Software V50.07.00.04
(V50.07.00.04)
EFCS_FPGA
05 FPGA EFCS V1.0
(V1.0)
NOTE:
1, New software can only be used on the new EPSON monitor board. The DSP monitoring board
will be damaged when using the new software.
2, Downgrade the higher version to lower version is not allowed.
3,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.05.09.00 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test, please perform
steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Main】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Main】->【General】->【Zero Flow Meters】,zero flow meters and
restart again;
For those devices which have bundle version 02.05.00 or higher, steps a, b, c are unnecessary during
upgrading to 02.05.09.00.

(2)Second: Software for devices with DSP monitor board.


No. Package Modules Latest Version

SystemUpdateTool
00 Tool 5.7
(V5.7)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)

Bundle Version 02.16.00.03

Bundle Host Software 03.16.00.03


02
(V02.16.00.03) FPGA Display 1.4

FPGA Sound 1.6

No.: OMZ-21009-A Series Page 149of 428


PMLS Global Technical Support Dept

Ventilator Protect Module V01.02.00

Ventilator Control Module V01.04.00

Language 4.6

Keyboard 01.00/01.01

03 Power(V1.7) Power System V1.7

04 EFCS(V50.06.00) EFCS Software V50.07.00.04


EFCS_FPGA
05 FPGA EFCS V1.0
(V3.0)
This Software is not compatible with the devices which have software version 02.05.09 or lower
version, for more details please refer to the Software upgrade instruction in software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.
2, When the devices which have bundle version 02.05.09.00 or lower need to upgrade to the
02.16.00.03 of the Bundle version, please upgrade software to 02.05.09.00 first ,then change the
Ventilator Control Board(DSP) FRU and upgrade the software to 02.16.00.03
3, Downgrade the higher version to lower version is not allowed.
4,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.16.00.03 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test, please perform
steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Main】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Main】->【General】->【Zero Flow Meters】,zero flow meters and
restart again;
For those devices which have bundle version 02.05.00 or higher, step a, b, c are unnecessary during
upgrading to 02.16.00.03.

The Software Technical Changes include:

1, The name of”Silence” button is changed to “Audio Pause”, its icon style is changed from to

.
2, Add a new button “Audio Pause Settings” in the “Other Settings” menu of the system menu. when the
“Audio Pause Settings” is set to ON and meanwhile no mid/high level alarms, Press the “Audio Pause”
button, the low level alarm audio will be turned off; Press the button again, the mid/high level alarm audio
will be turned off 120sec,but the low level alarm still is turned off continuously.

No.: OMZ-21009-A Series Page 150of 428


PMLS Global Technical Support Dept

Mid/high level alarm, solid ×-line

Low level alarm, dashed ×-line

Low level alarm audio is turned off

3, Flow Pause function is released. when it is enabled, the ventilation and fresh gas will be paused for
60sec.

No.: OMZ-21009-A Series Page 151of 428


PMLS Global Technical Support Dept

4, Fresh Gas Flow Optimizer function can be used with CO2 module
5, Fixed a bug which cause alarm “load configuration failed” under a long-term use.
6, Solved the problem that the CAPNOSTAT CO2 module may stop the communication under the
interference of electric knife.
7,Sloved the problem that the network loop(e.g. local area network,or Multi-port switch)may cause the
device to restart
8, Solved the problem the problem that PCV mode is missing when upgrade the software from EPSON
software to DSP software.

◼ Others
1, O2/AIR cylinder supported pressure increased from 15Mpa to 20Mpa.
2, AGSS 3 ways connector assembly is released
Order Number Description Remark
AGSS 3 ways connector Used between the ACGO exhaust
115-026796-00
assembly vent and AGSS interfaces

3,Activation Code for agent consumption function is released.


Order Number Description Remark
Only used for the devices with
115-044265-00 Activation Code(USB/AA measurement)
internal AG module
No.: OMZ-21009-A Series Page 152of 428
PMLS Global Technical Support Dept

4, New main control board spare part is added for the maintenance.
Order Number Description Remark
New, Only used for devices with
115-044211-00 Main control board(Non FDA/EPSON)
EPSON monitor board
0635 main control board FRU(Non Old, Only used for devices with
115-040956-00
FDA/VGA) DSP monitor board
5,Operator’s manual and service manual updated.
6, Dust-proof sponge is added to the 1-year maintenance kit.

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: ECL128 Engineering changes. Ref No: TMZ-16089-A7

Version Effective Date Description


1.0 2016-12-30 Document released.

No.: OMZ-21009-A Series Page 153of 428


PMLS Global Technical Support Dept

TMZ-16088-A5

ECL128 Engineering changes

Technical News
Subject: ECL128 Engineering changes.

RECORD IDENTIFIER

Ref No.: TMZ-16088-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Yang Parts Change
Date: 30th,Dec,2016 Appearance
Model:A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

◼ Hardware
1,The flow sensor is replaced by the new type:
Before After

Flow sensor assembly Flow sensor assembly

New flow sensor


Difference of
appearance

Old flow sensor New gas mixer

Fresh Flow Sensor


E_Flow assembly
115-027251-00 Board(0634/O2&N2O&AIR) 115-044266-00
FRU(0634/O2&N2O&Air)
--delete
Spare
Fresh Flow Sensor E_Flow assembly
Parts 115-024621-00 115-044267-00
Board(w/o N2O) ---delete FRU(0634/O2&Air)
0634 Flow Control Board
/ / 051-002834-00
PCBA ---new

No.: OMZ-21009-A Series Page 154of 428


PMLS Global Technical Support Dept

O2 flow senor
All of the devices use the modified flow sensor for replacement.
failure
Note:
N2O flow senor 1, For the old flow sensor Maintenance, please refer the guide in the spare
Maintenance
failure part
2, Software Upgrade is necessary after change the flow sensor. you need to
AIR flow senor
select software according to the type of monitor board(EPSON or DSP )
failure

Upgrade N/A

2,Structure of Mechanical ACGO assembly changed


before after

Different installation of metal plate

Difference of
appearance Different interface location

ACGO
ACGO
801-0631-00143-00 assembly(A5-CE) 115-043752-00
Spare assembly(0634)
FRU ---delete
Parts
mechanical ACGO
115-017878-00 115-043753-00
ACGO(AG) ---delete assembly(0634/AG)

Mechanical ACGO All of the devices use the modified ACGO assembly for
Maintenance
assembly failure replacement.

Upgrade N/A

◼ Software
1,Two new software is launched:
(1)First: Software for devices with EPSON monitor board.
No. Package Modules Latest Version

No.: OMZ-21009-A Series Page 155of 428


PMLS Global Technical Support Dept

00 SystemUpdateTool Tool 5.7

BIOS
01 BIOS 01.12.00.00
(01.12.00.00)

Bundle Version 02.05.09.00

Host Software 03.05.08.00

FPGA Display 1.4

Bundle FPGA Sound 1.2


02
(V02.05.09.00) Ventilator Protect Module V1.1

Ventilator Control Module V01.15.00

Language 3.9

flowmeter V1.0

03 Power(V1.7) Power System V1.7

NOTE:
1, New software can only be used on the new EPSON monitor board. The DSP monitoring
board will be damaged when using the new software.
2, Downgrade the higher version to lower version is not allowed.

(2)Second: Software for devices with DSP monitor board.


No. Package Modules Latest Version

SystemUpdateTool
00 Tool 5.7
(V5.7)
BIOS
01 BIOS 01.16.00.00
(V01.16.00.00)

Bundle Version 02.16.00.03

Host Software 03.16.00.03

FPGA Display 1.4

Bundle FPGA Sound 1.6


02
(V02.16.00.03) Ventilator Protect Module V01.02.00

Ventilator Control Module V01.04.00

Language 4.6

flowmeter V1.0

03 Power(V1.7) Power System V1.7


This Software is not compatible with the devices which have software version 02.05.09 or lower
version, for more details please refer to the Software upgrade instruction in software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.
2, When the devices which have bundle version 02.05.09.00 or lower need to upgrade to the

No.: OMZ-21009-A Series Page 156of 428


PMLS Global Technical Support Dept
02.16.00.03 of the Bundle version, please upgrade software to 02.05.09.00 first ,then change
the Ventilator Control Board(DSP) FRU and upgrade the software to 02.16.00.03
3, Downgrade the higher version to lower version is not allowed.

The Software Technical Changes include:

1, The name of “Silence” button is changed to “Audio Pause”, its icon style is changed from

to .
2, Add new button “Audio Pause Settings” in the “Other Settings” menu of the system menu. when the
“Audio Pause Settings” is set to ON and meanwhile no mid/high level alarms, Press the “Audio Pause”
button, the low level alarm audio will be turned off; Press the button again, the mid/high level alarm
audio will be turned off 120sec,but the low level alarm still is turned off continuously.

Mid/high level alarm, solid ×-line

Low level alarm, dashed ×-line

Low level alarm audio is turned off


3, Fixed a bug which cause alarm “load configuration failed” under a long-term use.
4, Solved the problem that the CAPNOSTAT CO2 module may stop the communication under the
interference of electric knife.
5,Sloved the problem that the network loop(e.g. local area network,or Multi-port switch) may cause
the device to restart
No.: OMZ-21009-A Series Page 157of 428
PMLS Global Technical Support Dept
6, Solved the problem the problem that PCV mode is missing when upgrade the software from EPSON
software to DSP software.

◼ Others
1 , O2/AIR cylinder supported pressure increased from 15Mpa to 20Mpa.
2, AGSS 3 ways connector assembly is released
Order Number Description Remark
AGSS 3 ways connector Used between the ACGO exhaust
115-026796-00
assembly vent and AGSS interfaces

3,Activation Code for agent consumption function is released.


Order Number Description Remark
Only used for the devices with
115-044265-00 Activation Code(USB/AA measurement)
internal AG module

4, New main control board spare part is added for the maintenance.
Order Number Description Remark
New, Only used for devices with
115-044211-00 Main control board(Non FDA/EPSON)
EPSON monitor board
0635 main control board FRU(Non Old, Only used for devices with
115-040956-00
FDA/VGA) DSP monitor board
5, Operator’s manual and service manual updated.
6, Dust-proof sponge is added to the 1-year maintenance kit.

No.: OMZ-21009-A Series Page 158of 428


PMLS Global Technical Support Dept

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: ECL128 Engineering changes. Ref No: TMZ-16088-A5

Version Effective Date Description


1.0 2016-12-30 Document released.

No.: OMZ-21009-A Series Page 159of 428


PMLS Global Technical Support Dept

TMZ-16068-A7

ECL124F Technical Changes

Technical News
Subject: ECL124F Technical Changes.

RECORD IDENTIFIER

Ref No.: TMZ-16068-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Sep,16th,2016 Appearance
Model:A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

1. VGA function of A7 anesthesia is released.


before after

Mother board, no VGA interface Mother board has VGA


interface

Difference of
appearance

Control board, no chip Control board has chip

No.: OMZ-21009-A Series Page 160of 428


PMLS Global Technical Support Dept
0632 Mother Board 0635 mother board
051-001259-00 115-040955-00
PCBA--delete FRU(Non FDA/VGA)
Main control board
Spare
FRU(Non-FDA) 0635 main control
Parts
115-035003-00 –only used for 115-040956-00 board FRU(Non
A5,A7 no longer FDA/VGA)
used

For all A7 machines, Using the new spare part for repair. But for the
Fault of Mother board old machines, VGA function does not work in the case of replacing
only the motherboard.
Maintenance

Fault of Control board For all A7 machines, Using the new spare part for repair.

Upgrade To support VGA function, need to replace the Mother Board and Main Control Board.

2. Some symbols and information had been added on new devices after this technical
change because of the requirement of new regulation.
They are:
⚫ Maximum input flow limit for every inlet
⚫ Maximum output flow information for auxiliary supply
⚫ Maximum inlet flow and pressure information for internal vacuum structure inlet

⚫ DEFIBRILLATION-PROOF TYPE BF APPLIED PART.


This symbol means the AC current between the part with it and patient or medic is
smaller than 100uA


This symbol means reading operation manual is necessary before operating the parts
with it.


This symbol means the parts with it could not be treated as common garbage and
could not throw away directly.
⚫ IPX1
This symbol means the parts with it has below waterproof level: water drop falling
down vertically would not influenced the parts
⚫ Weight information

Changing details as below:


2.1 Add maximum input flow limit above gas inlet and gas cylinder
Note: the input flow limits are the same for the same gas inlet and cylinder on same device
Maximum input flow limit
O2(L/Min) AIR(L/Min) N2O(L/Min)
No.: OMZ-21009-A Series Page 161of 428
PMLS Global Technical Support Dept
O2 190 20 20
Drive gas
Air 60 150 20

Fig1 Gas inlet

Fig2 Gas cylinder

No.: OMZ-21009-A Series Page 162of 428


PMLS Global Technical Support Dept

2.2 Add maximum inlet flow and pressure information for internal vacuum structure inlet.

Fig3 Internal vacuum structure inlet

2.3 Add maximum output flow information and symbol on auxiliary O2 supply port.

Fig4 Auxiliary O2 supply port

No.: OMZ-21009-A Series Page 163of 428


PMLS Global Technical Support Dept

2.4 Add symbol above Insp&Exp ports and ACGO port.

Fig5 Insp&Exp ports

Fig6 ACGO port

2.5 Add Symbol above main switch.

No.: OMZ-21009-A Series Page 164of 428


PMLS Global Technical Support Dept

Fig7 Main switch

2.6 Add symbol, IPX1symbol, and weight information on A7 device label.

Fig8 A7 device label

2.7 Add symbol, and weight information on AGSS assembly label.

Fig9 AGSS label

If you need download information or have any question, please feel free to contact
No.: OMZ-21009-A Series Page 165of 428
PMLS Global Technical Support Dept
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: ECL124F Technical Changes. Ref No: TMZ-16068-A7

Version Effective Date Description


1.0 2016-09-19 Document released.

No.: OMZ-21009-A Series Page 166of 428


PMLS Global Technical Support Dept

TMZ-16067-A5

ECL124F Technical Changes

Technical News
Subject: ECL124F Technical Changes.

RECORD IDENTIFIER

Ref No.: TMZ-16067-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Sep,16th,2016 Appearance
Model:A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

Some symbols and information had been added on new devices after this technical
change because of the requirement of new regulation.
They are:
⚫ Maximum input flow limit for every inlet
⚫ Maximum output flow information for auxiliary supply

⚫ DEFIBRILLATION-PROOF TYPE BF APPLIED PART.


This symbol means the AC current between the part with it and patient or medic is
smaller than 100uA


This symbol means reading operation manual is necessary before operating the parts
with it.


This symbol means the parts with it could not be treated as common garbage and
could not throw away directly.
⚫ IPX1
This symbol means the parts with it has below waterproof level: water drop falling
down vertically would not influenced the parts

No.: OMZ-21009-A Series Page 167of 428


PMLS Global Technical Support Dept
⚫ Weight information

Changing details as below:


1. Add maximum input flow limit above gas inlet and gas cylinder
Note: the input flow limits are the same for the same gas inlet and cylinder on same device
Maximum input flow limit
O2(L/Min) AIR(L/Min) N2O(L/Min)
O2 190 20 20
Drive gas
Air 60 150 20

Fig1 Gas inlet

No.: OMZ-21009-A Series Page 168of 428


PMLS Global Technical Support Dept

Fig2 Gas cylinder

2. Add maximum output flow information and symbol on auxiliary O2 supply port.

Fig3 Auxiliary O2 supply port

3. Add symbol above Insp&Exp ports and ACGO port.

No.: OMZ-21009-A Series Page 169of 428


PMLS Global Technical Support Dept

Fig4 Insp&Exp ports

Fig5 ACGO port

4. Add Symbol above main switch.

No.: OMZ-21009-A Series Page 170of 428


PMLS Global Technical Support Dept

Fig6 Main switch

5. Add symbol, IPX1symbol, and weight information on A5 device label.

Fig7 A5 device label

6. Add symbol, and weight information on AGSS assembly label.

No.: OMZ-21009-A Series Page 171of 428


PMLS Global Technical Support Dept

Fig8 AGSS label

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: ECL124F Technical Changes. Ref No: TMZ-16067-A5

Version Effective Date Description


1.0 2016-09-19 Document released.

No.: OMZ-21009-A Series Page 172of 428


PMLS Global Technical Support Dept

TMZ-16060-All AM&Ventilator

Anesthesia Machine&Ventilator Parts List is released

Technical News
Subject: Anesthesia Machine&Ventilator Parts List is released.

RECORD IDENTIFIER

Ref No.: TMZ-16060-All AM&Ventilator Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shiling Parts Change
Date: July.28th.2016 Appearance
Model: All AM&Ventilator models Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

1,Anesthesia Machine&Ventilator_Parts_List_ V160701 is released.


Parts list is uploaded to TDP: /Anesthesia Machine/Ventilator/Parts List

Note:
(a)Cause the “WATO EX-55/65Pro”(factory code 0625/0626) which were released
recently had changed their name to “WATO EX-55/65”, the corresponding parts list name
will also be changed from "WATO EX-55Pro(0626)_Parts list" & "WATO EX-
65Pro(0625)_Parts list" to WATO EX-55(0626)_Parts list" & "WATO EX-65(0625)_Parts list" in
this version.

(b)2 new parts list "WATO EX-55 Pro(0628)_Parts List", “WATO EX-65 Pro(0627)_Parts
List" will be added in this version, those new types will be available very soon.

If you have any questions & suggestions, or you find any mistakes, please
feedback to Shiling (yangshiling@mindray.com).Thanks a lot for your support.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

No.: OMZ-21009-A Series Page 173of 428


PMLS Global Technical Support Dept

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: Anesthesia Machine&Ventilator Ref No: TMZ-16060-All AM&Ventilator
Parts List is released.

Version Effective Date Description


1.0 2016-07-29 Document released.

No.: OMZ-21009-A Series Page 174of 428


PMLS Global Technical Support Dept

TMZ-16055-Vaporizers(V60&Penlon)

New maintenance packages for V60 and new adapter for vaporizers are released
(ECL120)

Technical News
Subject: New maintenance packages for V60 and new adapter for vaporizers are released
(ECL120).

RECORD IDENTIFIER

Ref No.: TMZ-16055- Relate To:


Vaporizers(V60&Penlon)
Version: 1.0 Software/Manual Upgrade
Author:David Wang Parts Change
Date: 2016.07.27 Appearance
Model:Vaporizers(V60、Penlon) Stop Production
Comments:N/A Others

CONTENT
To Whom It May Concern,

Below content is only for your information.


1. New maintenance packages for V60 vaporizers are released
Change Description:
Since some leakage problem caused by the filler happened on V60 vaporizers(Pour
Fill & Quick-Fil) are reported in Chinese domestic market,R&D changed the design for
V60 vaporizers produced after 2015.07 by EDL009. In order to solve the problem
happened on V60 vaporizers produced before 2015.07, new maintenance package for
V60 vaporizers are released by ECL120.
Influence to installation business:
N/A
Influence to maintenance business:
For V60 vaporizers(Pour Fill & Quick-Fil) produced before 2015.07,if the leakage
problem caused by the filler which is hard to be screwed properly, spare parts below could
be used to solve the problem:
Item Part Number Spare Part Description Vaporizers Applicable to
1 115-039569-00 Filler cap of ENF(Pour Fill V60 ENF vaporizers(Pour Fill)
2 115-039571-00 Filler cap of ISO(Pour Fill) V60 ISO vaporizers(Pour Fill)
3 115-039572-00 Filler cap of SEV(Pour Fill V60 SEV vaporizers(Pour Fill)
4 115-039573-00 Filler cap of SEV(Quik-Fil) V60 SEV vaporizers(Quick -Fil)

No.: OMZ-21009-A Series Page 175of 428


PMLS Global Technical Support Dept
5 115-039574-00 Filler cap of Hal(Pour Fill) V60 SEV vaporizers(Pour Fill)

If want more details about the repair and test procedure, please refer to “H-046-
009797-00 V60 Vaporizer Filler Cap Maintenance Guide”. This file has been uploaded to
TDP platform:
/Anesthesia Machine/Ventilator/Anesthesia Machine/V60 (0602)
Influence to upgrade business:
N./A

2. Filler apapter for SEV vaporizers has been changed


Change Description:
In order to reduce the cost, the new filler adapter “115-026747-00, Quik-Fil filling
adapter for sevoflurane” is used to instead of “115-002357-00, Sevoflurane filler
assembly”.
Spare part before the change Spare part after the change compatibility
115-002357-00,Sevoflurane filler 115-026747-00, Quik-Fil filling adapter for Compatible
assembly sevoflurane
The specification, property of new one is same as the old one. There are some difference
in material & appearance between the old and new adapters, as shown in pictures below:

(1)115-002357-00,Sevoflurane filler assembly (2)115-026747-00, Quik-Fil filling adapter for


sevoflurane
Influence to installation business:
N/A
Influence to maintenance business:
For SEV filler adapter, if customers have maintenance requirement, “115-026747-00,
Quik-Fil filling adapter for sevoflurane” could be used instead of “115-002357-00,
Sevoflurane filler assembly”.
Influence to installation business:
For SEV filler adapter, if customers have upgrading requirement, “115-026747-00,
Quik-Fil filling adapter for sevoflurane” could be used instead of “115-002357-00,
Sevoflurane filler assembly”.

No.: OMZ-21009-A Series Page 176of 428


PMLS Global Technical Support Dept

3.The drain adapter is changed to be an optional configuration for V60 SEV


vaporizers(Quick -Fil)
Change Description:
In order to reduce the cost, the drain adapter (040-000067-00,Quik-Fil Drain Funnel
Adaptor 54909) is changed to be an optional configuration for V60 SEV vaporizers(Quick
–Fil).

040-000067-00,Quik-Fil Drain Funnel Adaptor 54909


Influence to installation business:
For V60 SEV vaporizers(Quick -Fil),if the drain adapter hasn’t been selected,
the “040-000067-00,Quik-Fil Drain Funnel Adaptor 54909” will not be contented in the
vaporizer kits.
Influence to maintenance business:
N/A
Influence to upgrade business:
N/A

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 177of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: New maintenance packages for Ref No: TMZ-16055-
V60 and new adapter for vaporizers are Vaporizers(V60&Penlon)
released(ECL120).

Version Effective Date Description


1.0 2016-07-28 Document released.

No.: OMZ-21009-A Series Page 178of 428


PMLS Global Technical Support Dept

TMZ-16046-A7 (0635)

A7 (0635) Product Change (ECL117)

Technical News
Subject: A7 (0635) Product Change (ECL117).

RECORD IDENTIFIER

Ref No.: TMZ-16046-A7 (0635) Relate To:


Version:1.0 Software/Manual Upgrade
Author: David Wang Parts Change
Date: June 13, 2016 Appearance
Model:A7 (0635) Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information


1. Vaporaizers supplied by Drager is released for A7 (0635) in international market
Change Description:
In order to meet commercial requirement, vaporizers (refer to Fig.1) supplied by
Drager are released for A7 (0635) in countries lised in Tabel 1.

No.: OMZ-21009-A Series Page 179of 428


PMLS Global Technical Support Dept

Fig.1 Vaporizers Provided by Drager

The spare parts number and description is listed in Table 3.


Tabel 1
Intercontinental Area Countries

North America Canada United States of America

All countries of the European Union(excluding territories

outside geographical Europe)

Norway Island

Switzerland Croadia

Bosnia&Herzegovina Montenegro

Europe Albania Macedonia(FYROM)

Serbia Ukraine

Moldoova Belarus

Russia Turkey

Africa All countries on the African continent

Georgria Armenia

Azerbaijan Kazakstan

No.: OMZ-21009-A Series Page 180of 428


PMLS Global Technical Support Dept

Kyrgystan Tadzhikistan

Middle East Uzbekistan Syria

&Western Asia Lebanon Israel

Jordan Iraq

Iran Saudi-Arabia

Bahrain Qatar

United Arab Emitates Oman

Yemen /

Table 2
Item Spare Parts Number Spare Parts Description Remark

1 0992-00-0149 Vaporizer,Isoflurane Drager

2 0992-00-0148 Vaporizer,Sevoflurane Drager

3 040-001997-00 Vaporizer,Desflurane Drager,U.S power cord

4 040-002839-00 D-Vapor Desflurane,Euro Drager,Euro power cord

5 040-002840-00 D-Vapor Desflurane,UK Drager,UK power cord

More details about the Drager Vaporizers,please refer to the operator’s manual.The
operator’s manual has been uploaded to TDP platform:
Link:View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Drager Vaporizers

Influence to Installation Business:


If the A7 (0635) anesthesia machine is configuraed with Drager vaporizer,engineer
needs to install the vaporizer and trains the operators.
Influence to Maintenance Business:
If the vaporizer has any failure, vaporizers listed in Table 2 are needed to replace the
faulted vaporizers.
Vaporizer released is defined as the accessary, the warranty policy of these
vaporizers should be according to the accessary warranty policy defined by regoin area.
Influence to Upgrade Business:
If customers from countries listed in Table 1 need Drager vaporizers, spare parts

No.: OMZ-21009-A Series Page 181of 428


PMLS Global Technical Support Dept

listed in Table 2 could be applied by Sales from Mindray to meet the comerical
requirement.

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A7 (0635) Product Change (ECL117). Ref No: TMZ-16046-A7 (0635)

Version Effective Date Description


1.0 2016-06-20 Document released.

No.: OMZ-21009-A Series Page 182of 428


PMLS Global Technical Support Dept

TMZ-16045-A5(0634)

A5 (0634) Product Change (ECL117)

Technical News
Subject: A5 (0634) Product Change (ECL117).

RECORD IDENTIFIER

Ref No.: TMZ-16045-A5(0634) Relate To:


Version:1.0 Software/Manual Upgrade
Author: David Wang Parts Change
Date: June 13, 2016 Appearance
Model:A5(0634) Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information


1. Vaporaizers supplied by Drager is released for A5(0634) in international market
Change Description:
In order to meet commercial requirement, vaporizers (refer to Fig.1) supplied by
Drager are released for A5 (0634) in countries lised in Tabel 1.

No.: OMZ-21009-A Series Page 183of 428


PMLS Global Technical Support Dept

Fig.1 Vaporizers Provided by Drager

The spare parts number and description is listed in Table 3.


Tabel 1
Intercontinental Area Countries

North America Canada United States of America

All countries of the European Union(excluding territories

outside geographical Europe)

Norway Island

Switzerland Croadia

Bosnia&Herzegovina Montenegro

Europe Albania Macedonia(FYROM)

Serbia Ukraine

Moldoova Belarus

Russia Turkey

Africa All countries on the African continent

Georgria Armenia

Azerbaijan Kazakstan

No.: OMZ-21009-A Series Page 184of 428


PMLS Global Technical Support Dept

Kyrgystan Tadzhikistan

Middle East Uzbekistan Syria

&Western Asia Lebanon Israel

Jordan Iraq

Iran Saudi-Arabia

Bahrain Qatar

United Arab Emitates Oman

Yemen /

Table 2
Item Spare Parts Number Spare Parts Description Remark

1 0992-00-0149 Vaporizer,Isoflurane Drager

2 0992-00-0148 Vaporizer,Sevoflurane Drager

3 040-001997-00 Vaporizer,Desflurane Drager,U.S power cord

4 040-002839-00 D-Vapor Desflurane,Euro Drager,Euro power cord

5 040-002840-00 D-Vapor Desflurane,UK Drager,UK power cord

More details about the Drager Vaporizers,please refer to the operator’s manual.The
operator’s manual has been uploaded to TDP platform:
Link:View Files:/Anesthesia Machine/Ventilator/Anesthesia Machine/Drager Vaporizers

Influence to Installation Business:


If the A5 (0634) anesthesia machine is configuraed with Drager vaporizer,engineer
needs to install the vaporizer and trains the operators.
Influence to Maintenance Business:
If the vaporizer has any failure, vaporizers listed in Table 2 are needed to replace the
faulted vaporizers.
Vaporizer released is defined as the accessary, the warranty policy of these
vaporizers should be according to the accessary warranty policy defined by regoin area.
Influence to Upgrade Business:
If customers from countries listed in Table 1 need Drager vaporizers, spare parts

No.: OMZ-21009-A Series Page 185of 428


PMLS Global Technical Support Dept

listed in Table 2 could be applied by Sales from Mindray to meet the comerical
requirement.

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A5 (0634) Product Change (ECL117). Ref No: TMZ-16045-A5(0634)

Version Effective Date Description


1.0 2016-06-20 Document released.

No.: OMZ-21009-A Series Page 186of 428


PMLS Global Technical Support Dept

TMZ-16044-A7

A7 new software is launched

Technical News
Subject: A7 new software is launched.

RECORD IDENTIFIER

Ref No.: TMZ-16044-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Yang Parts Change
Date: 12th,June,2016 Appearance
Model:A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

(1)New software is launched:


No. Package Modules Latest Version
Bundle Version V02.15.00
Host Software V03.15.00
BIOS V01.14.00
FPGA Display V1.4
Bundle
1 FPGA Sound V1.5
(V02.15.00)
Ventilator Protect Module V01.02.00
Ventilator Control Module V01.04.00
Language V4.5
Keyboard V01.00/01.01

2 EFCS (V50.06.00) EFCS Software V50.06.00

3 EFCS_FPGA(V3.0) FPGA EFCS V3.0

4 Power(V1.7) Power System V1.7

Software 02.09.00 is not compatible with the devices which have software version 02.05.07 or lower
version, for more details please refer to the Software upgrade instruction in software package.

NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.

No.: OMZ-21009-A Series Page 187of 428


PMLS Global Technical Support Dept
2, When the devices which have bundle version 02.05.07 or lower need to upgrade to the
V02.09.00 of the Bundle version, please change the Ventilator Control Board(DSP) FRU first
3, Downgrade the higher version to lower version is not allowed.
4,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.09.00 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test, please perform
steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Main】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Main】->【General】->【Zero Flow Meters】,zero flow meters and
restart again;
For those devices which have bundle version 02.05.00 or higher, steps above are unnecessary
during upgrading to 02.09.00.

The Software Technical Changes include:


1, Add “Auto Exit Standby” button in the “service” menu. A7 with new software now
supports 2 ways of enter/exit standby mode. The factory default setting of this button is
“Off”, details of this function as below.

(a)when “Auto Exit Standby button” is “On”, Auto/Manual switch has to be set to
“Manual” before entering/exiting standby mode, this enter/exit standby mode is safer for
the patient, and was used on A7 with software version 02.05.07 or lower.
No.: OMZ-21009-A Series Page 188of 428
PMLS Global Technical Support Dept
Under this setting, the way to enter/exit standby mode as below.
Exit Standby:Touch the waveform area or turn on the fresh gas flow, or select the
Standby soft key on the screen, then follow the screen prompts to exit Standby mode.
Enter Standby: Select the Standby soft key on the main screen.

When auto/manual switch in the


“Auto” position, give tip to enter/exit
standby mode

(b)when “Auto Exit Standby button” is “Off”, it is unnecessary to set Auto/Manual


switch before entering/exiting standby mode, this enter/exit standby mode is more
convenient and simple, and is used on A7 with software version 02.09.00 or higher. And
under this mode ,it supports to preset the fresh gas flow, namely open fresh gas flow
would not entry Standby mode
Under this setting, the way to enter/exit standby mode as below.
Exit Standby:Touch the waveform area or select the Standby soft key on the
screen, then follow the screen prompts to exit Standby mode.
Entry Standby: Select the Standby soft key on the main screen.

No.: OMZ-21009-A Series Page 189of 428


PMLS Global Technical Support Dept

Preset fresh gas flow

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A7 new software is launched. Ref No: TMZ-16044-A7

Version Effective Date Description


1.0 2016-06-14 Document released.

No.: OMZ-21009-A Series Page 190of 428


PMLS Global Technical Support Dept

No.: OMZ-21009-A Series Page 191of 428


PMLS Global Technical Support Dept

TMZ-16043-A5

A5 new software is launched

Technical News
Subject: A5 new software is launched.

RECORD IDENTIFIER

Ref No.: TMZ-16043-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Yang Parts Change
Date: 12th,June,2016 Appearance
Model:A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

(1)New software is launched:

No. Package Modules Latest Version


Bundle Version V02.15.00
Host Software V03.15.00
BIOS V01.14.00
FPGA Display V1.4
Bundle
1 FPGA Sound V1.5
(V02.15.00)
Ventilator Protect Module V01.02.00
Ventilator Control Module V01.04.00
Language V4.5
Flowmeter Software V1.9
2 Power(V1.7) Power System V1.7
Software 02.09.00 is not compatible with the devices which have software version 02.05.07 or lower
version, for more details please refer to the Software upgrade instruction in software package.

NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.
2, When the devices which have bundle version 02.05.07 or lower need to upgrade to the
V02.09.00 of the Bundle version, please change the Ventilator Control Board(DSP) FRU first
2, Downgrade the higher version to lower version is not allowed.
No.: OMZ-21009-A Series Page 192of 428
PMLS Global Technical Support Dept
The Software Technical Changes include:

1, Add “Auto Exit Standby” button in the “service” menu. A7 with new software now
supports 2 ways of enter/exit standby mode. The factory default setting of this button is
“Off”, details of this function as below.

(a) when “Auto Exit Standby button” is “On”, Auto/Manual switch has to be set to
“Manual” before entering/exiting standby mode, this enter/exit standby mode is safer for
the patient, and was used on A7 with software version 02.05.07 or lower.
Under this setting, the way to enter/exit standby mode as below.
Exit Standby: Touch the waveform area or turn on the fresh gas flow, or select the
Standby soft key on the screen, then follow the screen prompts to exit Standby mode.
Enter Standby: Turn off all fresh gas flows, then select the Standby soft key on the
main screen.

No.: OMZ-21009-A Series Page 193of 428


PMLS Global Technical Support Dept

When auto/manual switch in the


“Auto” position, give tip to enter/exit
standby mode

(b)when “Auto Exit Standby button” is “Off”, it is unnecessary to set Auto/Manual


switch before entering/exiting standby mode, this enter/exit standby mode is more
convenient and simple, and is used on A7 with software version 02.09.00 or higher. And
under this mode ,it supports to preset the fresh gas flow, namely open fresh gas flow
would not entry Standby mode
Under this setting, the way to enter/exit standby mode as below.
Exit Standby:Touch the waveform area or select the Standby soft key on the
screen, then follow the screen prompts to exit Standby mode.
Entry Standby: Select the Standby soft key on the main screen.

Preset gas flow

No.: OMZ-21009-A Series Page 194of 428


PMLS Global Technical Support Dept
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject: A5 new software is launched. Ref No: TMZ-16043-A5

Version Effective Date Description


1.0 2016-06-14 Document released.

No.: OMZ-21009-A Series Page 195of 428


PMLS Global Technical Support Dept

TMZ-16038-A7

Anesthesia Prediction Function upgrade kit is released for A7(TCN16-FU002)

Technical News
Subject: Anesthesia Prediction Function upgrade kit is released for A7(TCN16-FU002).

RECORD IDENTIFIER

Ref No.: TMZ-16038-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shiling Yang Parts Change
Date: 2016-5-5 Appearance
Model:A7 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


1, Anesthesia Prediction Function upgrade kit is released for A7
Change Description:
In order to add the anesthesia prediction function, anesthesia prediction upgrade kit
is released. The upgrade kit will be provided by R&D. This kit is only allowed to be used
on the following machines:
SN Configuration of ACGO
GC-63000569
GC-63000570
GC-63000571
GC-63000566
GC-63000567 Electrically Controlled ACGO
GC-61000543
GC-61000544
GC-61000545
GC-5B000509
GC-61000541
Mechanical Controlled ACGO
GC-5B000478

Influence to installation:
N/A
Influence to maintenance:
No.: OMZ-21009-A Series Page 196of 428
PMLS Global Technical Support Dept
N/A
Influence to upgrade:
Choose the corresponding guide depending on the machine ACGO configuration,
then finish the upgrade process and perform tests.
Guide notes:
1, Anesthesia Prediction Function (Electrically Controlled ACGO) Upgrade Guide
2, Anesthesia Prediction Function (Mechanical Controlled ACGO) Upgrade Guide

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Technical News Revision


Subject:Anesthesia Prediction Function upgrade Ref No: TMZ-16038-A7
kit is released for A7(TCN16-FU002).

Version Effective Date Description


1.0 2016-05-11 Document released.

No.: OMZ-21009-A Series Page 197of 428


PMLS Global Technical Support Dept

TMZ-16016-A7

ECL115 Technical Changes

Technical News
Subject: ECL115 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-16016-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Feb.29th,2016 Appearance
Model: A7 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,


Please check the technical news for technical changes ECL115.
This technical news includes 7 hardware changes and 1 software change

1,Hardware:
⚫ Change description
Now A series has 2 kinds of monitor board: EPSON platform monitor board and DSP platform
monitor board. In 2015 technical changes R&D had replaced EPSON platform monitor board
to DSP ones both on new devices and in spare parts list .But some Problems were found
when EPSON platform monitor board changed to DSP platform monitor board for
maintenance or upgrade, R&D is finding a solution now, before the final plan to release,
please use EPSON platform monitor board to fix EPSON devices. More details please check
below.
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distinguish 1. Please use the start screen to distinguish(recommended)
the device EPSON Platform DSP Platform device

No.: OMZ-21009-A Series Page 198of 428


PMLS Global Technical Support Dept

2. Get the serial number and check it on SAP platform


For EPSON platform device, please use 115-018147-00 Monitoring Board FRU for
Monitor replacement
board Note:
issue 1. If customer side could not perform factory calibration please use 115-018148-01
Ventilation kit(New Diff_pressure) FRU for replacement
Monitor For EPSON platform device, please use 801-0631-00027-00 Ventilator Control Board
Maintenance
assembly( for replacement
monitor Note:
board+valv 1. If customer side could not perform factory calibration please use 115-018148-01
e drive Ventilation kit(New Diff_pressure) FRU for replacement
board)
issue
Before the final plan to release, please do not upgrade EPSON platform devices to DSP
Upgrade
platform. That means A series II phase new functions are unavailable for EPSON devices now.
⚫ Change description
For reducing the cost,A7 devices produced after 2015.12.31(Serial number larger than
GC-5D000000) will delete sensor interface board (under work surface, beside expiration
valve), the original Int-flow sensor on this board will be changed to new type and moved
to monitor board
Influence on Installation
N/A
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts influenced by this
technical change, engineers could distinguish the devices by serial numbers, serial
numbers smaller than GC-5D000000 are the devices before change, serial numbers
larger than GC-5D000000 are the devices after change.
Distinguish 2. If it is impossible to make sure the device never replaced any parts influenced by this
the device technical change, please open the device and use Difference of appearance to
before distinguish(recommended)
change or
after change

No.: OMZ-21009-A Series Page 199of 428


PMLS Global Technical Support Dept
Before change After change

Deleted sensor interface board New monitor board

Difference of
appearance
Original monitor board

Added Int-flow sensor on monitor board

Could not Could be used on


Sensor interface
be used on Devices before&
801-0631- board PCBA -
devices After change
00089-00 Could not apply
after
now Monitoring
change
115-036365-00 Board(DSP
Monitoring Could not
) FRU
Board(DSP) be used on
115-035294-
FRU devices
00
-Could not apply after
now change
Ventilator Could not Could be used on
Control be used on Ventilator Devices before&
Spare
115-035296- Board(DSP) devices Control After change
Parts 115-036366-00
00 FRU after Board(DSP
-Could not apply change ) FRU
now
Could not Could be used on
Drive
be used on Devices before&
Assembly(New
devices Drive After change
115-020937- Diff_pressure)F
after 115-036693-00 Assembly(
00 RU
change DSP)FRU
-Could not apply
now

115-035052- Ventilation Could not Ventilation Could be used on


115-036364-00
00 kit(Non- be used on kit(Non- Devices before&

No.: OMZ-21009-A Series Page 200of 428


PMLS Global Technical Support Dept
FDA/DSP)FRU devices FDA/DSP) After change
-Could not apply after FRU
now change

For DSP platform device(distinguish way please check above), please use 115-036365-
00 Monitoring Board(DSP) FRU for replacement, and upgrade software, need factory
calibration
Note:
Int-flow 1. Please reference the guide in the package for replacement
sensor issue 2. Software upgrade is necessary, and this monitor board can only supply software
version 02.09.00 or higher
3. Factory calibration is necessary, please refer to service manual. If factory
calibration is unavailable, please use 115-036364-00 Ventilation kit(Non-
FDA/DSP)FRU for replacement.
For DSP platform device(distinguish way please check above), please use 115-
036366-00 Ventilator Control Board(DSP) FRU for replacement, and upgrade
software, need factory calibration
Note:
Monitor 1. Please reference the guide in the package for replacement
Maintenance
board issue 2. Software upgrade is necessary, and this monitor board can only supply software
version 02.09.00 or higher
3. Factory calibration is necessary, please refer to service manual. If
factory calibration is unavailable, please use 115-036364-00
Ventilation kit(Non-FDA/DSP)FRU for replacement.
For DSP platform device(distinguish way please check above), please use 115-
036693-00 Drive Assembly(DSP)FRU for replacement, and upgrade software,
need factory calibration
Note:
Expiration 1. Please reference the guide in the package for replacement
valve issue 2. Software upgrade is necessary, and this monitor board can only
supply software version 02.09.00 or higher
3. Factory calibration is necessary, please refer to service manual. If
factory calibration is unavailable, please use 115-036364-00
Ventilation kit(Non-FDA/DSP)FRU for replacement.
Upgrade N/A
⚫ Change description
For reducing the cost, A7 devices produced after 2015.12.31(Serial number larger than
GC-5D000000) will use new screen assembly and backlight board,
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distinguish 1. If it is possible to make sure the device never replaced any parts
the device influenced by this technical change, engineers could distinguish the
before devices by serial numbers, serial numbers smaller than GC-5D000000

No.: OMZ-21009-A Series Page 201of 428


PMLS Global Technical Support Dept
change or are the devices before change, serial numbers larger than GC-
after change 5D000000 are the devices after change.
2. If it is impossible to make sure the device never replaced any parts
influenced by this technical change, please open the device and use
Difference of appearance to distinguish(recommended)

Before change After change

Difference of
appearance

0631 Could not be Could be used on


Backlight used on Devices After
0635 LED
051-001519- Inverter devices 051-002448- change
Backlight Board
00 PCBA - after change 00 Could not be used
PCBA
Still could on devices before
apply change
display Could not be Could be used on
Display
screen used on Devices before&
Spare 115-018149- 115-036373- Exchange
FRU of A7 devices After change
Parts 00 00 Package
-Could not after change
(0634)
apply now
display Could not be Could be used on
assembly( used on Devices before&
115-034074- LG devices 115-036454- Display screen After change
00 main/0635) after change 00 FRU(0635)
-Could not
apply now
Please use 115-036373-00 Display Exchange Package(0634)for replacement
Screen issue Note:
1. Please reference the guide in the package for replacement
Maintenance 1. A7 devices produced before 2015.12.17 please use 051-001519-000631
Backlight Backlight Inverter PCBA for replacement
board issue 2. A7 devices produced after 2015.12.17 please use 051-002448-00 0635 LED
Backlight Board PCBA for replacement
Upgrade N/A
⚫ Change description
For reducing the cost, A7 devices produced after 2015.12.31(Serial number larger than GC-
5D000000), if the configuration of BFCS is AIR+O2(2 needle valves),will not have gas mixer
assembly
Influence on Installation
N/A
Influence on Maintenance & upgrade

No.: OMZ-21009-A Series Page 202of 428


PMLS Global Technical Support Dept
1. If it is possible to make sure the device never replaced any parts
influenced by this technical change, engineers could distinguish the
Distinguish devices by serial numbers, serial numbers smaller than GC-5D000000
the device are the devices before change, serial numbers larger than GC-
before 5D000000 are the devices after change.
change or 2. If it is impossible to make sure the device never replaced any parts
after change influenced by this technical change, please open the device and use
Difference of appearance to distinguish(recommended)

Before change After change


Just a connector

Difference of
appearance

Could not be gas Could be used on


115- Gas mixer assembly used on mixer Devices before&
Spare
027533 FRU(0635/O2&Air) - devices after 115-036456-00 service After change
Parts
-00 Could not apply now change kit(0635/
O2&Air)
Gas Please use115-036456-00 gas mixer service kit(0635/O2&Air) for replacement
mixer Note:
Maintenance
assemb 1. Please reference the guide in the package for replacement
ly issue
Upgrade N/A

⚫ Change description
For reducing the cost, A7 devices produced after 2015.12.31(Serial number larger than
GC-5D000000) will change the bypass assembly material from metal to plastic
Influence on Installation
CO2 absorber appearance changed, please pay attention during customer training
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts influenced
Distingu by this technical change, engineers could distinguish the devices by serial
ish the numbers, serial numbers smaller than GC-5D000000 are the devices before
device change, serial numbers larger than GC-5D000000 are the devices after
before change.
change 2. If it is impossible to make sure the device never replaced any parts
or after influenced by this technical change, please open the device and use
change Difference of appearance to distinguish(recommended)

No.: OMZ-21009-A Series Page 203of 428


PMLS Global Technical Support Dept
Before change After change

Bypass actuator

Differen
ce of
appear
ance

Bypass Assembly

Could be used on Could be used on


CO2 Bypass
Devices before& Devices before&
Assembly, A CO2 Bypass
Spare 801-0631- After change 115- After change
series Assembly FRU,
Parts 00099-00 Do not suggest to 036378-00
-Could not apply A series
use old spare
now
part
Mainten Bypass Please use 115-036378-00 CO2 Bypass Assembly FRU, A series for replacement
ance issue
Upgrad N/A
e

⚫ Change description
For reducing the cost, A7 devices produced after 2015.12.31(Serial number larger than GC-
5D000000) will use new base assembly and castor
Influence on Installation
N/A
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts
influenced by this technical change, engineers could distinguish the
Distinguish
devices by serial numbers, serial numbers smaller than GC-5D000000
the device
are the devices before change, serial numbers larger than GC-5D000000
before
are the devices after change.
change or
2. If it is impossible to make sure the device never replaced any parts
after
influenced by this technical change, please open the device and use
change
Difference of appearance to distinguish(recommended)

No.: OMZ-21009-A Series Page 204of 428


PMLS Global Technical Support Dept
Before change After change

Original castor connectors New castor connectors

Difference
of
appearanc
e

Base of Chesis Could not be Could be used


115-
Assembly(FRU) used on on Devices
016691-
-Could not apply devices after before&After
00
now change change
Connector of Could not be
801-
Brake used on
0631-
-Could not apply devices after
00095-00 base frame
now change 115-036459-
FRU(CE&Pend
Principal Axis of Could not be 00
801- ant adapter)
Brake used on
0631-
-Could not apply devices after
00096-00 base frame
now change 115-036460-
FRU(CFDA&big
Brake Could not be 00
115- bottle)
Assembly(FRU) used on
016692-
Spare -Could not apply devices after
00
Parts now change
Base of Chesis Could not be
801-
Assembly - used on
0631-
Could not apply devices after
00097-00
now change
left castor Could not be Could be used
115- assembly(new used on on Devices
016955- color)FRU devices after before&After
00 -Could not apply change change
now 034-000285- casters.6" bule
right castor Could not be 00 and black
115- assembly(new used on
016956- color)FRU - devices after
00 Could not apply change
now

No.: OMZ-21009-A Series Page 205of 428


PMLS Global Technical Support Dept
Base(not 1. For pendent configuration devices, please use 115-036459-00 base frame
include FRU(CE&Pendant adapter) for replacement
Maintenan lock) 2. For big bottle configuration devices, please use 115-036460-00 base frame
ce issue FRU(CFDA&big bottle) for replacement
Castor Please use 034-000285-00 casters.6" bule and black for replacement
issue
Upgrade N/A

⚫ Change description
To keep consistent with WATO series, R&D also apart the A series 1 year maintenance kit to
3 new ones like WATO series. And a new 3 year maintenance kit is released also. More details
please check below.
Influence on Installation
N/A
Influence on Maintenance & upgrade
Before change After change
Could be used Could be used
on Devices on Devices
Maintain
before&After 115-036367- before&After
Kit(0635,Basic,
change 00 change
1 yr)
Suggest use
new kit
Could be used
Preventative on Devices
115- Maintain
Maintenance 115-036368- before&After
024063- Kit(0635,Exten
kits(1 year) 00 change
00 d,1 yr)
-Still could apply Suggest use
Spare
new kit
Parts
Could be used
on Devices
Maintain
115-036369- before&After
Kit(0635,Advan
00 change
ced,1 yr)
Suggest use
new kit
Could be used Could be used
Preventative
115- on Devices Maintain on Devices
Maintenance 115-036370-
024064- before&After Kit(0635/3 before&After
kits(3 years) 00
00 change years) change
-Still could apply
Suggest use

No.: OMZ-21009-A Series Page 206of 428


PMLS Global Technical Support Dept
new kit
Note:
1. Please pay attention to our part list which will be released on March to learn the specific
difference between old and new maintenance kit.
2. Old maintenance kit will be kept until 2016.07,after July old ones will not be available and
will be removed from part list
3. Replacement way is contained in service manual

2,Software:
⚫ Change description
For optimizing the display and solve the software bugs, R&D released new software version.
Specific change points as below
Adding 13 working languages
Turkish, Czech, Polish, Portuguese, Russian, Norwegian, Bulgarian, Danish, Dutch,
Finnish, Romanian, Serbian, Swedish
Original 7 working languages are:English, Chinese, French, German, Italian, Hungarian,
Spanish
New software could support 20 languages in total.
Display optimize for French and Spanish languages
If old software(02.09.00&02.11.00) works in French or Spanish, sometime alarm message
could not show totally can be found, new software could solve this problem
Lung recruitment tool optimize
In old software(02.09.00&02.11.00), if one uses Lung recruitment tool, and during lung
recruitment function changes AUTO/Manual switch from AUTO to Manual, the mode shows
on the top of screen will not change. New software could solve this proble.
NMT module optimize
1. In old software(02.09.00&02.11.00), keep disconnecting and reinstalling NMT module
for several times in short time would make anesthesia device restart, new software
could solve this problem
2. In old software(02.09.00&02.11.00), some messages in NMT form could not show
totally, new software could solve this problem.
SIMV-VC optimize
In old software(02.09.00&02.11.00), set Exp% 70% or higher in Standby mode, after this
mode starting, Exp% will return to default setting(25%)
Anesthesia Prediction optimize
If the drug in air way could not be detected or distinguished, the drug type on screen will
shows ”AA”,and MAC wave form coordinate axis will not show unit. Since unit is
meaningless in this situation
Manual leak test optimize
One of Manual leak test result ,leakage 200-1000ml, has changed its color from red to
white. This color change will not influence the leakage decision condition.
No.: OMZ-21009-A Series Page 207of 428
PMLS Global Technical Support Dept

Bug solution
In old software(02.09.00&02.11.00), sometimes if customer uses 3L manual bag(standard
configuration) for manual leak test, because of the algorithm, even there is no leakage, but
device will give a larger leakage result, and causes manual leak test could not pass.
If you find the similar case, please upgrade the latest software version.
Influence on Installation
Please change the language during installation when necessary
Influence on Maintenance & upgrade
No. Package Modules Latest Version
Bundle Version V02.13.00
Host Software V03.13.02
BIOS V01.14.00
FPGA Display V1.4
Bundle
1 FPGA Sound V1.5
(V02.13.00)
Ventilator Protect Module V01.02.00
Ventilator Control Module V01.03.00
Language V4.2
Keyboard V01.00/01.01

2 EFCS (V50.06.00) EFCS Software V50.06.00

3 EFCS_FPGA(V3.0) FPGA EFCS V3.0

4 Power(V1.7) Power System V1.7

Software 02.09.00 or higher are not compatible with the devices which have software version
02.05.07 or lower version, for more details please refer to the Software upgrade instruction in
software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board
will be damaged when using the new software.
2, When the devices which have bundle version 02.05.07 or lower need to upgrade to
the V02.09.00 or higher, please change the Ventilator Control Board(DSP) FRU first
3, Downgrade the higher version to lower version is not allowed.
4,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.09.00 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test,
please perform steps below to solve this issue:
No.: OMZ-21009-A Series Page 208of 428
PMLS Global Technical Support Dept
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Main】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Main】->【General】->【Zero Flow Meters】,zero flow
meters and restart again;
For those devices which have bundle version 02.05.00 or higher, steps above are
unnecessary during upgrading to 02.09.00

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
ECL115 Technical Changes TMZ-16016-A7

Version Effective Date Description


1.0 2016/3/1 Document released.

No.: OMZ-21009-A Series Page 209of 428


PMLS Global Technical Support Dept

TMZ-16015-A5

ECL115 Technical Changes

Technical News
Subject: ECL115 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-16015-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Feb.29th,2016 Appearance
Model: A5 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,


Please check the technical news for technical changes ECL115.
This technical news includes 7 hardware changes and 1 software change

1,Hardware:
⚫ Change description
Now A series has 2 kinds of monitor board: EPSON platform monitor board and DSP platform
monitor board. In 2015 technical changes R&D had replaced EPSON platform monitor board
to DSP ones both on new devices and in spare parts list .But some Problems were found when
EPSON platform monitor board changed to DSP platform monitor board for maintenance or
upgrade, R&D is finding a solution now, before the final plan to release, please use EPSON
platform monitor board to fix EPSON devices. More details please check below.
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distinguish 3. Please use the start screen to distinguish(recommended)
the device EPSON Platform DSP Platform device

No.: OMZ-21009-A Series Page 210of 428


PMLS Global Technical Support Dept

4. Get the serial number and check it on SAP platform


For EPSON platform device, please use 115-018147-00 Monitoring Board FRU for
Monitor replacement
board Note:
issue 2. If customer side could not perform factory calibration please use 115-018148-01
Ventilation kit(New Diff_pressure) FRU for replacement
Monitor For EPSON platform device, please use 801-0631-00027-00 Ventilator Control Board
Maintenance
assembly( for replacement
monitor Note:
board+valv 2. If customer side could not perform factory calibration please use 115-018148-01
e drive Ventilation kit(New Diff_pressure) FRU for replacement
board)
issue
Before the final plan to release, please do not upgrade EPSON platform devices to DSP
Upgrade
platform. That means A series II phase new functions are unavailable for EPSON devices now.

⚫ Change description
For reducing the cost,A5 devices produced after 2015.12.31(Serial number larger than FY-
5D000000) will delete sensor interface board (under work surface, beside expiration valve),
the original Int-flow sensor on this board will be changed to new type and moved to monitor
board
Influence on Installation
N/A
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts influenced by this
technical change, engineers could distinguish the devices by serial numbers, serial
Distinguish
numbers smaller than FY-5D000000 are the devices before change, serial numbers
the device
larger than FY-5D000000 are the devices after change.
before
2. If it is impossible to make sure the device never replaced any parts influenced by this
change or
technical change, please open the device and use Difference of appearance to
after change
distinguish(recommended)

No.: OMZ-21009-A Series Page 211of 428


PMLS Global Technical Support Dept

Before change After change

Deleted sensor interface board New monitor board

Difference of
appearance
Original monitor board

Added Int-flow sensor on monitor board

Could not Could be used on


Sensor interface
be used on Devices before&
801-0631- board PCBA -
devices After change
00089-00 Could not apply
after
now Monitoring
change
115-036365-00 Board(DSP
Monitoring Could not
) FRU
Board(DSP) be used on
115-035294-
FRU devices
00
-Could not apply after
now change
Ventilator Could not Could be used on
Spare Control be used on Ventilator Devices before&
Parts 115-035296- Board(DSP) devices Control After change
115-036366-00
00 FRU after Board(DSP
-Could not apply change ) FRU
now
Could not Could be used on
Drive
be used on Devices before&
Assembly(New
devices Drive After change
115-020937- Diff_pressure)F
after 115-036693-00 Assembly(
00 RU
change DSP)FRU
-Could not apply
now

115-035052- Ventilation Could not 115-036364-00 Ventilation Could be used on

No.: OMZ-21009-A Series Page 212of 428


PMLS Global Technical Support Dept
00 kit(Non- be used on kit(Non- Devices before&
FDA/DSP)FRU devices FDA/DSP) After change
-Could not apply after FRU
now change

For DSP platform device(distinguish way please check above), please use 115-036365-
00 Monitoring Board(DSP) FRU for replacement, and upgrade software, need factory
calibration
Note:
Int-flow 4. Please reference the guide in the package for replacement
sensor issue 5. Software upgrade is necessary, and this monitor board can only supply software
version 02.09.00 or higher
6. Factory calibration is necessary, please refer to service manual. If factory
calibration is unavailable, please use 115-036364-00 Ventilation kit(Non-
FDA/DSP)FRU for replacement.
For DSP platform device(distinguish way please check above), please use 115-
036366-00 Ventilator Control Board(DSP) FRU for replacement, and upgrade
software, need factory calibration
Note:
Monitor 4. Please reference the guide in the package for replacement
Maintenance
board issue 5. Software upgrade is necessary, and this monitor board can only supply software
version 02.09.00 or higher
6. Factory calibration is necessary, please refer to service manual. If
factory calibration is unavailable, please use 115-036364-00
Ventilation kit(Non-FDA/DSP)FRU for replacement.
For DSP platform device(distinguish way please check above), please use 115-
036693-00 Drive Assembly(DSP)FRU for replacement, and upgrade software,
need factory calibration
Note:
Expiration 4. Please reference the guide in the package for replacement
valve issue 5. Software upgrade is necessary, and this monitor board can only
supply software version 02.09.00 or higher
6. Factory calibration is necessary, please refer to service manual. If
factory calibration is unavailable, please use 115-036364-00
Ventilation kit(Non-FDA/DSP)FRU for replacement.
Upgrade N/A
⚫ Change description
For reducing the cost, A7 devices produced after 2015.12.31(Serial number larger than
GC-5D000000) will use new screen assembly and backlight board,
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distinguish 1. If it is possible to make sure the device never replaced any parts
the device influenced by this technical change, engineers could distinguish the

No.: OMZ-21009-A Series Page 213of 428


PMLS Global Technical Support Dept
before devices by serial numbers, serial numbers smaller than FY-5D000000
change or are the devices before change, serial numbers larger than FY-5D000000
after change are the devices after change.
2. If it is impossible to make sure the device never replaced any parts
influenced by this technical change, please open the device and use
Difference of appearance to distinguish(recommended)

Before change After change

Difference of
appearance

0631 Could not be Could be used on


Backlight used on Devices After
0635 LED
051-001519- Inverter devices 051-002448- change
Backlight Board
00 PCBA - after change 00 Could not be used
PCBA
Still could on devices before
apply change
display Could not be Could be used on
Display
screen used on Devices before&
Spare 115-018149- 115-036373- Exchange
FRU of A7 devices After change
Parts 00 00 Package
-Could not after change
(0634)
apply now
display Could not be Could be used on
assembly( used on Devices before&
115-034074- LG devices 115-036454- Display screen After change
00 main/0635) after change 00 FRU(0635)
-Could not
apply now
Please use 115-036373-00 Display Exchange Package(0634)for replacement
Screen issue Note:
2. Please reference the guide in the package for replacement
Maintenance 3. A7 devices produced before 2015.12.17 please use 051-001519-000631
Backlight Backlight Inverter PCBA for replacement
board issue 4. A7 devices produced after 2015.12.17 please use 051-002448-00 0635 LED
Backlight Board PCBA for replacement
Upgrade N/A

⚫ Change description
For reducing the cost, A5 devices produced after 2015.12.31(Serial number larger than FY-
5D000000) will use new Pop-off valve
Influence on Installation
N/A
⚫ Influence on Maintenance & upgrade
No.: OMZ-21009-A Series Page 214of 428
PMLS Global Technical Support Dept
1. If it is possible to make sure the device never replaced any parts influenced
Distingui
by this technical change, engineers could distinguish the devices by serial
sh the
numbers, serial numbers smaller than FY-5D000000 are the devices before
device
change, serial numbers larger than FY-5D000000 are the devices after
before
change.
change
2. If it is impossible to make sure the device never replaced any parts
or after
influenced by this technical change, please open the device and use
change
Difference of appearance to distinguish(recommended)
Before change After change

Differenc
e of
appeara
nce

Flow Could be used on Could be used on


gas supply
Compensation Devices before& Devices before&
Spare 801-0631- 115- assembly
Valve After change After change
Parts 00116-00 036458-00 service
-Could not Do not suggest to
kit(0634)
apply now use old spare part
Please use 115-036458-00 gas supply assembly service kit(0634) for replacement
Mainten Pop-off valve
Note:
ance issue
1. Please reference the guide in the package for replacement
Upgrade N/A

⚫ Change description
For reducing the cost, A5 devices produced after 2015.12.31(Serial number larger than FY-
5D000000) will change the bypass assembly material from metal to plastic
Influence on Installation
CO2 absorber appearance changed, please pay attention during customer training
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts influenced
Distingu
by this technical change, engineers could distinguish the devices by serial
ish the
numbers, serial numbers smaller than FY-5D000000 are the devices before
device
change, serial numbers larger than FY-5D000000 are the devices after
before
change.
change
2. If it is impossible to make sure the device never replaced any parts
or after
influenced by this technical change, please open the device and use
change
Difference of appearance to distinguish(recommended)
Before change After change

No.: OMZ-21009-A Series Page 215of 428


PMLS Global Technical Support Dept

Bypass actuator

Differen
ce of
appear
ance

Bypass Assembly

Could be used on Could be used on


CO2 Bypass
Devices before& Devices before&
Assembly, A CO2 Bypass
Spare 801-0631- After change 115- After change
series Assembly FRU,
Parts 00099-00 Do not suggest to 036378-00
-Could not apply A series
use old spare
now
part
Mainten Bypass Please use 115-036378-00 CO2 Bypass Assembly FRU, A series for replacement
ance issue
Upgrad N/A
e

⚫ Change description
For reducing the cost, A5 devices produced after 2015.12.31(Serial number larger than FY-
5D000000) will use new base assembly and castor
Influence on Installation
N/A
Influence on Maintenance & upgrade
1. If it is possible to make sure the device never replaced any parts
influenced by this technical change, engineers could distinguish the
Distinguish
devices by serial numbers, serial numbers smaller than GC-5D000000
the device
are the devices before change, serial numbers larger than GC-5D000000
before
are the devices after change.
change or
2. If it is impossible to make sure the device never replaced any parts
after
influenced by this technical change, please open the device and use
change
Difference of appearance to distinguish(recommended)

Before change After change

No.: OMZ-21009-A Series Page 216of 428


PMLS Global Technical Support Dept

Original castor connectors New castor connectors

Difference
of
appearanc
e

Base of Chesis Could not be Could be used


115-
Assembly(FRU) used on on Devices
016691-
-Could not apply devices after before&After
00
now change change
Connector of Could not be
801-
Brake used on
0631-
-Could not apply devices after
00095-00 base frame
now change 115-036459-
FRU(CE&Pend
Principal Axis of Could not be 00
801- ant adapter)
Brake used on
0631-
-Could not apply devices after
00096-00 base frame
now change 115-036460-
FRU(CFDA&big
Brake Could not be 00
115- bottle)
Assembly(FRU) used on
016692-
Spare -Could not apply devices after
00
Parts now change
Base of Chesis Could not be
801-
Assembly - used on
0631-
Could not apply devices after
00097-00
now change
left castor Could not be Could be used
115- assembly(new used on on Devices
016955- color)FRU devices after before&After
00 -Could not apply change change
now 034-000285- casters.6" bule
right castor Could not be 00 and black
115- assembly(new used on
016956- color)FRU - devices after
00 Could not apply change
now

No.: OMZ-21009-A Series Page 217of 428


PMLS Global Technical Support Dept
Base(not 3. For pendent configuration devices, please use 115-036459-00 base frame
include FRU(CE&Pendant adapter) for replacement
Maintenan lock) 4. For big bottle configuration devices, please use 115-036460-00 base frame
ce issue FRU(CFDA&big bottle) for replacement
Castor Please use 034-000285-00 casters.6" bule and black for replacement
issue
Upgrade N/A

⚫ Change description
To keep consistent with WATO series, R&D also apart the A series 1 year maintenance kit to 3
new ones like WATO series. And a new 3 year maintenance kit is released also. More details
please check below.
Influence on Installation
N/A
Influence on Maintenance & upgrade
Before change After change
Could be used Could be used
on Devices on Devices
Maintain
before&After 115-036367- before&After
Kit(0635,Basic,
change 00 change
1 yr)
Suggest use
new kit
Could be used
Preventative on Devices
115- Maintain
Maintenance 115-036368- before&After
024063- Kit(0635,Exten
kits(1 year) 00 change
00 d,1 yr)
-Still could apply Suggest use
new kit
Could be used
Spare on Devices
Maintain
Parts 115-036369- before&After
Kit(0635,Advan
00 change
ced,1 yr)
Suggest use
new kit
Could be used Could be used
Preventative on Devices on Devices
115- Maintain
Maintenance before&After 115-036370- before&After
024064- Kit(0635/3
kits(3 years) change 00 change
00 years)
-Still could apply Suggest use
new kit
Note:
4. Please pay attention to our part list which will be released on March to learn the specific
difference between old and new maintenance kit.
5. Old maintenance kit will be kept until 2016.07,after July old ones will not be available and

No.: OMZ-21009-A Series Page 218of 428


PMLS Global Technical Support Dept
will be removed from part list
6. Replacement way is contained in service manual

2,Software:
⚫ Change description
For optimizing the display and solve the software bugs, R&D released new software version.
Specific change points as below
Adding 13 working languages
Turkish, Czech, Polish, Portuguese, Russian, Norwegian, Bulgarian, Danish, Dutch, Finnish,
Romanian, Serbian, Swedish
Original 7 working languages are:English, Chinese, French, German, Italian, Hungarian,
Spanish
New software could support 20 languages in total.
Display optimize for French and Spanish languages
If old software(02.09.00&02.11.00) works in French or Spanish, sometime alarm message
could not show totally can be found, new software could solve this problem
Lung recruitment tool optimize
In old software(02.09.00&02.11.00), if one uses Lung recruitment tool, and during lung
recruitment function changes AUTO/Manual switch from AUTO to Manual, the mode shows
on the top of screen will not change. New software could solve this proble.
NMT module optimize
3. In old software(02.09.00&02.11.00), keep disconnecting and reinstalling NMT module for
several times in short time would make anesthesia device restart, new software could
solve this problem
4. In old software(02.09.00&02.11.00), some messages in NMT form could not show totally,
new software could solve this problem.
SIMV-VC optimize
In old software(02.09.00&02.11.00), set Exp% 70% or higher in Standby mode, after this
mode starting, Exp% will return to default setting(25%)
Anesthesia Prediction optimize
If the drug in air way could not be detected or distinguished, the drug type on screen will
shows ”AA”,and MAC wave form coordinate axis will not show unit. Since unit is meaningless
in this situation
Manual leak test optimize
One of Manual leak test result ,leakage 200-1000ml, has changed its color from red to white.
This color change will not influence the leakage decision condition.

No.: OMZ-21009-A Series Page 219of 428


PMLS Global Technical Support Dept

Bug solution
In old software(02.09.00&02.11.00), sometimes if customer uses 3L manual bag(standard
configuration) for manual leak test, because of the algorithm, even there is no leakage, but
device will give a larger leakage result, and causes manual leak test could not pass.
If you find the similar case, please upgrade the latest software version.
Influence on Installation
Please change the language during installation when necessary
Influence on Maintenance & upgrade
No. Package Modules Latest Version
Bundle Version V02.13.00
Host Software V03.13.02
BIOS V01.14.00
FPGA Display V1.4
Bundle
1 FPGA Sound V1.5
(V02.13.00)
Ventilator Protect Module V01.02.00
Ventilator Control Module V01.03.00
Language V4.2
Keyboard V01.00/01.01

2 Power(V1.7) Power System V1.7

Software 02.09.00 or higher are not compatible with the devices which have software version
02.05.07 or lower version, for more details please refer to the Software upgrade instruction in
software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board
will be damaged when using the new software.
2, When the devices which have bundle version 02.05.07 or lower need to upgrade
to the V02.09.00 or higher, please change the Ventilator Control Board(DSP) FRU
first
3, Downgrade the higher version to lower version is not allowed.
4,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.09.00
and restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test,
please perform steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";

No.: OMZ-21009-A Series Page 220of 428


PMLS Global Technical Support Dept
b, In standby mode, select 【Main】->【Service】->【Disabled BFCS】, close the
BFCS;
c, In standby mode, select 【Main】->【General】->【Zero Flow Meters】,zero flow
meters and restart again;
For those devices which have bundle version 02.05.00 or higher, steps above are
unnecessary during upgrading to 02.09.00

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
ECL115 Technical Changes TMZ-16015-A5

Version Effective Date Description


1.0 2016/3/1 Document released.

No.: OMZ-21009-A Series Page 221of 428


PMLS Global Technical Support Dept

TMZ-16004-A7

ECL108-F Technical Changes

Technical News
Subject: ECL108-F Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-16004-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Jan.18th,2016 Appearance
Model: A7 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,


Please check the technical news for technical changes ECL108-F.
This technical news includes 2 hardware changes.

1,Hardware:
⚫ Change description
For meet the requirements of new statute(IEC60601-1:2005+A1:2012-7.2.3) , after this
technical change the new device(serial numbers larger than GC-62000000) will use new mark
for “please reading operation manual before using”
Influence on Installation
May be customer could ask
Influence on Maintenance & upgrade
Distingu
ish the
device
before 1.Using Difference of appearance to distinguish
change
or after
change
Before change After change

No.: OMZ-21009-A Series Page 222of 428


PMLS Global Technical Support Dept

Differen
ce of
appeara
nce

Spare
N/A
Parts

Mainten
N/A
ance

Upgrade N/A

⚫ Change description
R&D had found because of the Spring-tube of driving gas and Spring-tube of APL valve,
devices could not pass Automatic leak test (manual leak test could pass) after a period of
time. For solving this problem, after this technical change the new device(serial numbers
larger than GC-62000000) will use new material and structure for Spring-tube of driving
gas(115-034448-00) and Spring-tube of APL valve(115-034449-00)
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distingui 1. If it is possible to make sure the device never replaced any parts influenced by
sh the this technical change, engineers could distinguish the devices by serial numbers,
device serial numbers smaller than GC-62000000 are the devices before change, serial
before numbers larger than GC-62000000 are the devices after change.
change 2. If it is impossible to make sure the device never replaced any parts influenced by
or after this technical change, please open the device and use Difference of
change appearance to distinguish(recommended)
Before change After change

Spring-tube of driving gas

Differenc
e of
appeara
nce

Spring-tube of APLvalve

No.: OMZ-21009-A Series Page 223of 428


PMLS Global Technical Support Dept
Do not
115- Spring-tube of 115- Spring-tube of Could be used on
suggest to
03444 driving gas 03444 driving gas Devices before&
use old
8-00 FRU-Upgraded 8-00 FRU After change
Spare spare part
Parts Do not Spring-tube o
115- Spring-tube of 115- Could be used on
suggest to f APL valve
03444 APLvalve FRU 03444 Devices before&
use old FRU
9-00 -Upgraded 9-00 After change
spare part
Please use 115-034448-00 Spring-tube of driving gas FRU for replacement
Note:
7. Please refer to service manual for replacement
8. In spare part package there are 5 pcs Wire fixed, please fix the tubes as
below
Spring
-tube
of
driving
gas
issue

Maintena
Please use 115-034449-00 Spring-tube of APL valve FRU for replacement,
nce
and upgrade software, need factory calibration
Note:
1. Please refer to service manual for replacement
2. In spare part package there are 5 pcs Wire fixed, please fix the tubes as
below
Spring
-tube
of APL
valve
issue

Upgrade N/A

No.: OMZ-21009-A Series Page 224of 428


PMLS Global Technical Support Dept

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
ECL108-F Technical Changes TMZ-16004-A7

Version Effective Date Description


1.0 2016/1/18 Document released.

No.: OMZ-21009-A Series Page 225of 428


PMLS Global Technical Support Dept

TMZ-16003-A5

ECL108-F Technical Changes

Technical News
Subject: ECL108-F Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-16003-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Parts Change
Date: Jan.18th,2016 Appearance
Model: A5 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,


Please check the technical news for technical changes ECL108-F.
This technical news includes 2 hardware changes.

1,Hardware:
⚫ Change description
For meet the requirements of new statute(IEC60601-1:2005+A1:2012-7.2.3) , after this
technical change the new device(serial numbers larger than FY-62000000) will use new mark
for “please reading operation manual before using”
Influence on Installation
May be customer could ask
Influence on Maintenance & upgrade
Distingu
ish the
device
before 1.Using Difference of appearance to distinguish
change
or after
change
Before change After change

No.: OMZ-21009-A Series Page 226of 428


PMLS Global Technical Support Dept

Differen
ce of
appeara
nce

Spare
N/A
Parts

Mainten
N/A
ance

Upgrade N/A

⚫ Change description
R&D had found because of the Spring-tube of driving gas and Spring-tube of APL valve, devices
could not pass Automatic leak test (manual leak test could pass) after a period of time. For
solving this problem, after this technical change the new device(serial numbers larger than FY-
62000000) will use new material and structure for Spring-tube of driving gas(115-034448-00)
and Spring-tube of APL valve(115-034449-00)
Influence on Installation
N/A
Influence on Maintenance & upgrade
Distingui 3. If it is possible to make sure the device never replaced any parts influenced by
sh the this technical change, engineers could distinguish the devices by serial numbers,
device serial numbers smaller than FY-62000000 are the devices before change, serial
before numbers larger than FY-62000000 are the devices after change.
change 4. If it is impossible to make sure the device never replaced any parts influenced by
or after this technical change, please open the device and use Difference of
change appearance to distinguish(recommended)
Before change After change

Spring-tube of driving gas

Differenc
e of
appeara
nce

Spring-tube of APLvalve

No.: OMZ-21009-A Series Page 227of 428


PMLS Global Technical Support Dept
Do not
115- Spring-tube of 115- Spring-tube of Could be used on
suggest to
03444 driving gas 03444 driving gas Devices before&
use old
8-00 FRU-Upgraded 8-00 FRU After change
Spare spare part
Parts Do not
115- Spring-tube of 115- Spring-tube o Could be used on
suggest to
03444 APLvalve FRU 03444 f APL valve Devices before&
use old
9-00 -Upgraded 9-00 FRU After change
spare part
Please use 115-034448-00 Spring-tube of driving gas FRU for replacement
Note:
9. Please refer to service manual for replacement
10. In spare part package there are 5 pcs Wire fixed, please fix the tubes as
below
Spring
-tube
of
driving
gas
issue

Maintena
Please use 115-034449-00 Spring-tube of APL valve FRU for replacement,
nce
and upgrade software, need factory calibration
Note:
3. Please refer to service manual for replacement
4. In spare part package there are 5 pcs Wire fixed, please fix the tubes as
below
Spring
-tube
of APL
valve
issue

Upgrade N/A

No.: OMZ-21009-A Series Page 228of 428


PMLS Global Technical Support Dept

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
ECL108-F Technical Changes TMZ-16003-A5

Version Effective Date Description


1.0 2016/1/18 Document released.

No.: OMZ-21009-A Series Page 229of 428


PMLS Global Technical Support Dept

TMZ-15050-A7

EFU009 Technical Changes

Technical News
Subject: EFU009 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-15050-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shiling Parts Change
Date: Sep.6th,2015 Appearance
Model: A7 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Function Upgrades of A7 are released. New hardware and software will be available
from 2015.09.01, most of this changes are not compatible with devices which were
produced before 2015.09, please pay attention to the maintenance and upgrading
ways.
1,Hardware:
⚫ Since the platform of monitoring assembly is changed from EPSON Platform to DSP Platform,
new Drive Board and Monitoring Board are released. Old Drive Board and Monitoring Board
could not be ordered now.

Before After

Cancel capacitance, add a new backup Longer interface between the boards
connector

Difference of
appearance

No.: OMZ-21009-A Series Page 230of 428


PMLS Global Technical Support Dept

Cancel the VPM board This connector position changed

Monitoring Board Monitoring Board(DSP)


115-032584-00 115-035294-00
FRU(deleted) FRU
801-0616- Drive Board for Monitoring
115-035295-00 Drive Board(DSP) FRU
00060-00 Valve(deleted)
Spare Ventilator Control Ventilator Control
115-032586-00 115-035296-00
Parts Board(deleted) Board(DSP) FRU
Ventilation kit(New Ventilation kit(Non-
115-032585-00 115-035052-00
Diff_pressure)FRU(deleted) FDA/DSP)FRU
1,The new Monitoring Board(SN:115-035294-00) and new Drive Board(115-
Fault of Drive
035295-00) are not compatible with old devices, but the assembly Ventilator
Board
Control Board(DSP) FRU (SN: 115-035296-00,Monitoring Board+ Drive
Board) is compatible with old device. So if there is any problem about
Maintenance monitoring assembly on device before 2015.09, please use Ventilator Control
Fault of
Board(DSP) FRU(SN:115-035296-00) for replacement ,at the same time, the
Monitoring
software must be upgraded to the V02.09.00.
Board
2, For the devices after 2015.09,please refer to the service manual to do the
maintenance
Notes: When the devices which have bundle version 02.05.07 or lower need to upgrade to the
Upgrade
V02.09.00 , please change the Ventilator Control Board(DSP) FRU first

⚫ New configuration electronic monitoring of gas supply for A7 is launched . The pressure
gauge configuration also is available, but two kinds of configuration can only choose one. For
the new function, the Electronic control board is changed, new gas inlet assembly, new
cylinder assembly, new pressure sensor are added. This new function is not able to
upgrade on-site.

No.: OMZ-21009-A Series Page 231of 428


PMLS Global Technical Support Dept

Before After

Interface for electronic gas supply signal

Difference of
appearance

Fault of Drive Board For all A7 machines, Using the new spare part for repair.
1. For the devices which configured gas supply pressure gauge,
Maintenance Fault of gas inlet please fix as usual
assembly 2. For the device which configured electronic monitoring of gas
supply, please check the form below
Upgrade Not allowed upgrading
New gas inlet assembly and new cylinder assembly as bellow:

Pressure sensor

Pressure sensor

new gas inlet assembly with pressure sensor new gas cylinder assembly with pressure sensor

Device which configured electronic monitoring of gas supply gas inlet assembly maintenance form:
Part of fault Replace SN Part Description reference
N2O origin inlet
N2O inlet assembly(DISS) 115-033733-00 With pressure sensor
module(0635/DISS)
AIR origin inlet
AIR inlet assembly(DISS) 115-033734-00 With pressure sensor
module(0635/DISS)
O2 origin inlet
O2 inlet assembly(DISS) 115-033735-00 With pressure sensor
module(0635/DISS)

No.: OMZ-21009-A Series Page 232of 428


PMLS Global Technical Support Dept
O2 origin inlet
O2 inlet assembly(NIST) 115-033736-00 With pressure sensor
module(0635/NIST)
N2O origin inlet
N2O inlet assembly(NIST) 115-033737-00 With pressure sensor
module(0635/NIST)
AIR origin inlet
AIR inlet assembly(NIST) 115-033738-00 With pressure sensor
module(0635/NIST)
Reserve Oxygen Supply Reserve Oxygen Supply base
115-034798-00 With pressure sensor
base part part(0635)
With pressure sensor,
Reserve Oxygen Reserve Oxygen
115-034799-00 include Supply Connector
Supply(NIST) Supply(0635/NIST)
and Supply base part

With pressure sensor,


Reserve Oxygen Reserve Oxygen
115-034800-00 include Supply Connector
Supply(DISS) Supply(0635/DISS)
and Supply base part

O2 cylinder assembly 115-033730-00 O2 cylinder yoke assembly(0635) With pressure sensor


AIR cylinder yoke
AIR cylinder assembly 115-033731-00 With pressure sensor
assembly(0635)
N2O cylinder yoke
N2O cylinder assembly 115-033732-00 With pressure sensor
assembly(0635)
Pressure sensor on inlet Only use the air inlet
115-035337-00 Sensor Pressure transmitter FRU
assembly assembly
Pressure sensor on Not support Separate
/ /
cylinder assembly maintenance
Sensor cable 009-005916-00 0635 Gas Pressure Sensor Cable /

⚫ New GCX arm with tray is released.


This part is newly added and compatible with devices before 2015.09.

No.: OMZ-21009-A Series Page 233of 428


PMLS Global Technical Support Dept

Tray

Install diagram Physical Pic Arm

Maintenance form:
Part of fault Replace SN Part Description Reference

Tray 034-000461-00 GCX Tray Tray can be changed independently

GCX 12" x 12"


Arm problem has to change whole part to
Arm 034-000460-00 Articulating Arm with
solve
Tray

2,Software:
New software is launched
No. Package Modules Latest Version
Bundle Version V02.09.00
Host Software V03.09.00
BIOS V01.14.00
FPGA Display V1.4
Bundle
1 FPGA Sound V1.5
(V02.09.00)
Ventilator Protect Module V01.02.00
Ventilator Control Module V01.03.00
Language V3.8
Keyboard V01.00/01.01

2 EFCS (V50.06.00) EFCS Software V50.06.00

3 EFCS_FPGA(V3.0) FPGA EFCS V3.0

4 Power(V1.7) Power System V1.7

Software 02.09.00 is not compatible with the devices which have software version 02.05.07 or lower
version, for more details please refer to the Software upgrade instruction in software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.

No.: OMZ-21009-A Series Page 234of 428


PMLS Global Technical Support Dept
2, When the devices which have bundle version 02.05.07 or lower need to upgrade to the
V02.09.00 of the Bundle version, please change the Ventilator Control Board(DSP) FRU first
3, Downgrade the higher version to lower version is not allowed.
4,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.09.00 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test, please perform
steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Main】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Main】->【General】->【Zero Flow Meters】,zero flow meters and
restart again;
5, For those devices which have bundle version 02.05.00 or higher, steps above are unnecessary
during upgrading to 02.09.00.

The Software Technical Changes include:


⚫ New software only support the following 7 kinds of languages:
English, Chinese, French, German, Italian, Spanish, Hungarian.
⚫ Self-checking visualization: layout change and increase help tips when the self-test failure.

Self-checking progress

Help tips

Self-test details

⚫ Alarm visualization: Increase help tips for the High-level alarm only, and add the duration
show about all alarms.

No.: OMZ-21009-A Series Page 235of 428


PMLS Global Technical Support Dept

Help tips

Duration

⚫ System layout changes optimization:


a,The buttons about [setup]&[discharge]&[History] are changed as bellow:
Before

After

b,The layout of password input keyboard is changed.


No.: OMZ-21009-A Series Page 236of 428
PMLS Global Technical Support Dept

c,System menu optimization as below.


Note: add two setting buttons in a new menu “other settings”
1.[Standby settings] button: clear Demographics and Spirometry reference loops
2.[Agent speed]: Displays the consumption speed of the anesthetic agents. This button
only displays when the machine is equipped with internal AG module.

The two buttons


are moved to
new menu
[other settings]

The button is moved


to [Network] menu

Before

No.: OMZ-21009-A Series Page 237of 428


PMLS Global Technical Support Dept

New button 1

Clear History switch,


the old discharges
setting

New button 2

New menu

After

d, Add display about gas supply pressure electric monitoring when the machine is
equipped with electric monitoring of gas supply pressure.

e, Add display about the agent speed when the machine is equipped with the internal AG
module.

Agent speed

No.: OMZ-21009-A Series Page 238of 428


PMLS Global Technical Support Dept
⚫ Two new ventilation modes are available: SIMV-VG,CPAP/PS. Use the following material to
activate the new modes through actiavtion code or USB disk
Order number Part description reference
115-034847-00 USB Disk with Activation Code (SIMV-VG)
Need the software version is V02.09.00.
115-034846-00 USB Disk with Activation Code (CPAP/PS)

SIMV-VG: means to deliver synchronized intermittent mandatory pressure regulated


volumecontrol ventilation to the patient.
CPAP/PS:this mode includes two ventilation modes(CPAP and PS)
Psupp is 0 cmH2O,PS is displayed;Psupp is not 0 cmH2O,CPAP is dispaly.

Pressure regulated

TV setting

SIMV-VG

No.: OMZ-21009-A Series Page 239of 428


PMLS Global Technical Support Dept

PS mode

PS when Psupp is not 0 cmH2O

CPAP mode

CPAP when Psupp is off (0 cmH2O)


⚫ Anesthetic Prediction function is available. Use the following material to activate through
actiavtion code or USB disk
Prediction: Display the concentration curves of Des,Iso,Enf,Sev,Hal,O2
Order number Part description reference

1,Need the software versiong is V02.09.00.


115-034848-00 Activation Code(USB/Prediction) 2,Note:Only available when the internal
AG module and external AG module are

No.: OMZ-21009-A Series Page 240of 428


PMLS Global Technical Support Dept
loaded.

10 min 20 min
History Prediction

⚫ Lung Recruitment Ventilation is available. Use the following material to activate


Recruitment Ventilation:onw lung-protective ventilation strategies, open more collapsed
pulmonary alveoli and prevent the secondary pulmonary atelectasis caused by the low tidal
volumes ventilation,by providing gases of a pressure higher than the mean airway pressure,
and sustains the pressure for a period of time during the mechanical ventilation
Order number Part description reference

115-034849-00 Activation Code(USB/Recruitment) Need the software versiong is V02.09.00.

No.: OMZ-21009-A Series Page 241of 428


PMLS Global Technical Support Dept

⚫ BISx4 module is available


BISx and BISx4 use the same module,only the cable and sensor are different.The
difference between bisx and bisx4 is BISx4 can show 2 waves on screen,but BISx only has 1

BISx4 Cable

BISx4 Sensor

No.: OMZ-21009-A Series Page 242of 428


PMLS Global Technical Support Dept

2 BIS waves

2 EEG waves

⚫ NMT module is available.


Display the NMT area by selecting [Main]->[Display]->[[Waveform Display]->“NMT“.

NMT area

No.: OMZ-21009-A Series Page 243of 428


PMLS Global Technical Support Dept
⚫ Capnostat CO2 module and Oridion CO2 module are available.

Capnostat CO2 module Oridion CO2 module

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
EFU009 Technical Changes TMZ-15050-A7

Version Effective Date Description


1.0 2015/9/6 Document released.

No.: OMZ-21009-A Series Page 244of 428


PMLS Global Technical Support Dept

TMZ-15049-A5

EFU009 Technical Changes

Technical News
Subject: EFU009 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-15049-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shiling Parts Change
Date: Sep.6th,2015 Appearance
Model: A5 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Function Upgrades of A5 are released. New hardware and software will be available
from 2015.09.01, most of this changes are not compatible with devices which were
produced before 2015.09, please pay attention to the maintenance and upgrading.
1,Hardware:
⚫ Since the platform of monitoring assembly is changed from EPSON Platform to DSP Platform,
new Drive Board and Monitoring Board are released. Old Drive Board and Monitoring Board
could not be ordered now

Before After

Cancel capacitance, add a new backup connector Longer interface between the boards

Difference of
appearance

No.: OMZ-21009-A Series Page 245of 428


PMLS Global Technical Support Dept

Cancel the VPM board This connector position changed

Monitoring Board Monitoring Board(DSP)


115-032584-00 115-035294-00
FRU(deleted) FRU
801-0616-00060- Drive Board for Monitoring
115-035295-00 Drive Board(DSP) FRU
00 Valve(deleted)
Spare Ventilator Control Ventilator Control
115-032586-00 115-035296-00
Parts Board(deleted) Board(DSP) FRU
Ventilation kit(New Ventilation kit(Non-
115-032585-00 115-035052-00
Diff_pressure)FRU(deleted) FDA/DSP)FRU
1,The new Monitoring Board(SN: 115-035294-00) and new Drive
Fault of Drive
Board(SN: 115-035295-00) are not compatible with old devices, but the
Board
assembly Ventilator Control Board(DSP) FRU(SN: 115-035296-00)
(Monitoring Board+ Drive Board) is compatible with old device. So if there is
any problem about monitoring assembly on device before 2015.09, please
Maintenance
use Ventilator Control Board(DSP) FRU(SN: 115-035296-00) for
Fault of Monitoring
replacement ,at the same time, the software must be upgraded to the
Board
V02.09.00.
2, For the devices after 2015.09,please refer to the service manual to do the
maintenance
Notes: When the devices which have bundle version 02.05.07 or lower need to upgrade to the
Upgrade
V02.09.00 , please change the Ventilator Control Board(DSP) FRU first

⚫ New GCX arm with tray is released.


This part is newly added and compatible with devices before 2015.09.

No.: OMZ-21009-A Series Page 246of 428


PMLS Global Technical Support Dept

Tray

Install diagram Physical Pic Arm

Maintenance form::
Part of
Replace SN Part Description Reference
fault
GCX Tray
Tray 034-000461-00 Tray can be changed independently

GCX 12" x 12" Articulating Arm with Arm problem has to change whole part to
Arm 034-000460-00
Tray solve

2,Software:
New software is launched
No. Package Modules Latest Version
Bundle Version V02.09.00
Host Software V03.09.00
BIOS V01.14.00
FPGA Display V1.4
Bundle
1 FPGA Sound V1.5
(V02.09.00)
Ventilator Protect Module V01.02.00
Ventilator Control Module V01.03.00
Language V3.8
Flowmeter Software V1.9
2 Power(V1.7) Power System V1.7
Software 02.09.00 is not compatible with the devices which have software version 02.05.07 or lower
version, for more details please refer to the Software upgrade instruction in software package.
NOTE:
1, New software can only be used on the new DSP board. The old monitoring board will be
damaged when using the new software.
2, When the devices which have bundle version 02.05.07 or lower need to upgrade to the
V02.09.00 of the Bundle version, please change the Ventilator Control Board(DSP) FRU first
2, Downgrade the higher version to lower version is not allowed.

No.: OMZ-21009-A Series Page 247of 428


PMLS Global Technical Support Dept
The Software Technical Changes include:
⚫ New software only support the following 7 kinds of languages:
English, Chinese, French, German, Italian, Spanish, Hungarian.
⚫ Self-checking visualization: layout change and increase help tips when the self-test failure.

Self-checking progress

Help tips

Self-test details

⚫ Alarm visualization: Increase help tips for the High-level alarm only, and add the duration
show about all alarms.

Help tips

Duration

⚫ System layout changes optimization:


a,The buttons about [setup]&[discharge]&[History] are changed as bellow:

No.: OMZ-21009-A Series Page 248of 428


PMLS Global Technical Support Dept

Before

After

b, The layout of password input keyboard is changed.

c, System menu optimization as below.


Note: add two setting buttons in a new menu “other settings”
1.[Standby settings] button: clear Demographics and Spirometry reference loops
2.[Agent speed]: Displays the consumption speed of the anesthetic agents. This button
only displays when the machine is equipped with internal AG module..

No.: OMZ-21009-A Series Page 249of 428


PMLS Global Technical Support Dept

The two buttons


are moved to
new menu
[other settings]

The button is moved


to [Network] menu

Before

New button 1

Clear History
switch, the old
discharges setting

New button 2

New menu

After

⚫ Two new ventilation modes are available: SIMV-VG,CPAP/PS. Use the following material to
No.: OMZ-21009-A Series Page 250of 428
PMLS Global Technical Support Dept
activate the new modes through actiavtion code or USB disk
Order number Part description reference
USB Disk with Activation Code (SIMV-
115-034847-00
VG) Need the software versiong is V02.09.00.
115-034846-00 USB Disk with Activation Code (CPAP/PS)

SIMV-VG: means to deliver synchronized intermittent mandatory pressure regulated


volumecontrol ventilation to the patient.
CPAP/PS:this mode includes two ventilation modes(CPAP and PS)
Psupp is 0 cmH2O,PS is displayed;Psupp is not 0 cmH2O,CPAP is dispaly.

SIMV-VG Pressure regulated

TV setting

PS mode

PS when Psupp is not 0 cmH2O

No.: OMZ-21009-A Series Page 251of 428


PMLS Global Technical Support Dept

CPAP mode

CPAP when Psupp is off (0 cmH2O)


⚫ Lung Recruitment Ventilation is available. Use the following material to activate
Recruitment Ventilation:onw lung-protective ventilation strategies, open more collapsed
pulmonary alveoli and prevent the secondary pulmonary atelectasis caused by the low tidal
volumes ventilation,by providing gases of a pressure higher than the mean airway pressure,
and sustains the pressure for a period of time during the mechanical ventilation
Order number Part description reference

115-034849-00 Activation Code(USB/Recruitment) Need the software versiong is V02.09.00.

⚫ BISx4 module is available

No.: OMZ-21009-A Series Page 252of 428


PMLS Global Technical Support Dept
BISx and BISx4 use the same module,only the cable and sensor are different.The difference
between bisx and bisx4 is BISx4 can show 2 waves on screen,but BISx only has 1

BISx4 Cable

BISx4 Sensor

2 BIS waves

2 EEG waves

⚫ NMT module is available.


Display the NMT area by selecting [Main]->[Display]->[[Waveform Display]->“NMT“.

No.: OMZ-21009-A Series Page 253of 428


PMLS Global Technical Support Dept

NMT area

⚫ Capnostat CO2 module and Oridion CO2 module are available.

Capnostat CO2 module Oridion CO2 module

If you need download information or have any question, please feel free to contact
No.: OMZ-21009-A Series Page 254of 428
PMLS Global Technical Support Dept
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
EFU009 Technical Changes TMZ-15049-A5

Version Effective Date Description


1.0 2015/9/6 Document released.

No.: OMZ-21009-A Series Page 255of 428


PMLS Global Technical Support Dept

TMZ-15044-A7

ECL112 Technical Changes

Technical News
Subject: ECL112 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-15044-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Parts Change
Date: Aug.3rd,2015 Appearance
Model: A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

1, New backup cylinder assembly and copper tubes of high pressure gauge are released, old backup
cylinder assembly and copper tubes of high pressure gauge could not be ordered any more. The new
cylinder assembly and tubes are not compatible with old one. Change details and maintenance plan
please check below
Old backup cylinder assembly and tubes New backup cylinder assembly and tubes

Appearance
differences

O2 cylinder assembly
O2 Yoke Assembly(With
115-014476-01 (0632), without tube 115-034134-00
gauge tube)FRU
(Deleted)
N2O cylinder assembly
Spare Parts N2O Yoke Assembly(With
115-014477-01 (0632), without tube 115-034136-00
changes gauge tube)FRU
(Deleted)
AIR cylinder assembly
AIR Yoke Assembly(With
115-014478-01 (0632) , without tube 115-034135-00
gauge tube)FRU
(Deleted)

No.: OMZ-21009-A Series Page 256of 428


PMLS Global Technical Support Dept
Tube of pressure Tube of pressure
115-025396-00 gauge(left)FRU 115-034392-00 gauge(0635/left)FRU,
(Deleted) shorter than before
Tube of pressure Tube of pressure
115-025397-00 gauge(middle)FRU 115-034392-00 gauge(0635/middle)FRU,
(Deleted) shorter than before
Tube of pressure Tube of pressure
115-025398-00 gauge(right)FRU 115-034393-00 gauge(0635/right)FRU,
(Deleted) shorter than before
Fault of Cylinder For all A7 machines, Using the new spare part for repair.
Maintenanc
assembly Note: Copper tubes must be changed at the same time
e
Fault of the High
Plan For all A7 machines, Using the new spare part for repair.
pressure Tube
Upgrade N/A N/A

2, The structure of electronic ACGO has changed. Old ACGO could not be ordered any more. A new
part No. is created for new ACGO, the new ACGO is not compatible with old one. Change details and
maintenance plan please check below.
Old ACGO Assembly New ACGO Assembly

before
after

Appearance
differences

With plug Without plug

E_ACGO service kit(w/o


ACGO assembly(0632)
115-014754-00 115-034440-00 AA),with service guide
(Deleted)
Spare Parts for the AM before
changes E_ACGO service kit(w/
electronic ACGO(AG)
115-017931-00 115-034441-00 AA), with service guide
(Deleted)
for the AM before

No.: OMZ-21009-A Series Page 257of 428


PMLS Global Technical Support Dept
Appearance
For all A7 machines, Using the new spare part for repair.
differences
Note: Repair the history machine According to the service guide in the new spare part kit.

Upgrade N/A N/A

3, New circuit heater is released, old circuit heater could not be ordered any more, the new circuit
heater is compatible with old one. Change details and maintenance plan please check below
Old circuit heater New circuit heater

Appearance
differences

Crimson Light red

Spare Parts Circuit Heater Circuit Heater


801-0631-00069-00 115-034450-00
changes (Deleted) FRU(0634)
Maintenanc
e For all A7 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

4, New back pressure valve is released. Old back pressure valve could not be ordered any more, the
new back pressure valve is compatible with old one. Change details and maintenance plan please
check below.
Old back pressure valve New back pressure valve

Appearance
differences

Copper Aluminum

Spare Parts back pressure valve(0632) back pressure valve


115-028145-00 115-033064-00
changes (Deleted) assembly(0635)
Maintenanc
e For all A7 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

5, New Optical Encoder is released, old Optical Encoder could not be ordered any more,. A new part
No. is created for new Optical Encoder, the new Optical Encoder is compatible with old one. Change

No.: OMZ-21009-A Series Page 258of 428


PMLS Global Technical Support Dept
details and maintenance plan please check below
Old Optical Encoder New Optical Encoder

Appearance
differences

Spare Part

old new

Optical Encoder, only one encoder assembly


010-000150-00 encoder 115-034469-00 service kit(0635),with
Spare Parts (Deleted) two encoders
changes display assembly, with old display assembly(LG
115-014133-00 encoder 115-034074-00 main/0635),with new
(Deleted) encoder
Maintenanc
e For all A7 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

6,The material of sample gas return connector changed from materal to plastic, old sample gas return
connector could not be ordered any more. A new part No. is created for new sample gas return
connector, the new sample gas return connector is compatible with old one. Change details and
maintenance plan please check below
Old sample gas return connector New sample gas return connector

115-015457-00

115-034138-00

Appearance
082-001206-00 metal
differences

plastic
M6Q-030093-00

No.: OMZ-21009-A Series Page 259of 428


PMLS Global Technical Support Dept
connector. copper chrome
Quick Connect
082-001206-00 1/8'ID 801-0621-00160-00
PMCD1601(FRU)
Spare Parts (Deleted)
changes quick connector(0632/
quick connector(0635/
115-015457-00 sampling return) 115-034138-00
sampling return)
(Deleted)
Maintenanc
e For all A7 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

7, The material of drawer changes from aluminum to SGCC, old drawer could not be ordered any
more. A new part No. is created for new material drawer, the new drawer is compatible with old one.
Change details and maintenance plan please check below
Old drawer New drawer

Appearance
differences

Appearance is same as before.


Drawer Assembly
Spare Parts Drawer Module
115-017034-00 FRU(0631) 115-034451-00
changes FRU(0634)
(Deleted)
Maintenanc
e For all A7 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

8, Tube from Expiration Valve to Breathing Circuit and tube from Breathing Circuit to AGSS
container have changed from Silicone to spring tubes. Old tubes could not be ordered any more,
the new tubes are compatible with old ones. Change details and maintenance plan please check
below
Old tubes New tubes

No.: OMZ-21009-A Series Page 260of 428


PMLS Global Technical Support Dept

M6G-020042--- 115-034449-00

Appearance
differences

M6G-020039--- 115-034448-00

Silicone Breathing Hose


Spring-tube of driving
M6G-020039--- Adult 450mm 115-034448-00
gas FRU
Spare Parts (Deleted)
changes Silicone Breathing
Spring-tube of APL
M6G-020042--- Tube(350mm) 115-034449-00
valve FRU
(Deleted)
Maintenanc
e 1,For all A7 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

9, New Gas Cylinder Wrench is released, old Gas Cylinder Wrench could not be ordered any more. A
new part No. is created for new Gas Cylinder Wrench, the new Gas Cylinder Wrench is compatible
with old one. Change details and maintenance plan please check below
Old Gas Cylinder Wrench Old Gas Cylinder Wrench

Appearance
differences

Appearance is same as before.


Gas Cylinder
Spare Parts gas cylinder wrench
041-004480-00 Wrench 115-033063-00
changes assembly
(Deleted)
Maintenanc
e For all A7 machines, Using the new spare part for repair.
Plan
Upgrade Use the new part

10, The tubes for O2&N2O&AIR are changed from colorized to transparent, and the material of most
connector changes from metal to plastic, old ones could not be ordered any more. The new tubes and
connectors are compatible with old ones. Change details and maintenance plan please check below.
Old tubes and connectors New tubes and connectors

No.: OMZ-21009-A Series Page 261of 428


PMLS Global Technical Support Dept

Appearance
differences

Tube PU (polyether)
082-000517-00 5.5mmX8mm orange
(Deleted)
Tube PU (polyether) Hose,soft PU
082-000521-00 5.5mmX8mm green M6G-020045--- transparent tube
(Deleted) 8mmX5.5mm
Tube PU 5.5mmX8mm
082-000662-00 blue
(Deleted)
tube.PU (polyether)
082-000522-00 4mmX6mm green
(Deleted)
Tube PU (polyether)
Transparent PU Tube
082-000520-00 4mmX6mm orange M6G-020026---
4*6 (20m)
(Deleted)
tube.PU (polyether)
Spare Parts 082-000524-00 4mmX6mm blue
changes (Deleted)
Tube PU (polyether)
082-000523-00 2.5mmX4mm green
(Deleted)
Tube PU (polyether)
082-000518-00 2.5mmX4mm orange M6G-020046--- Tubing.
(Deleted)
Tube PU (polyether)
082-000516-00 2.5mmX4mm blue
(Deleted)
Connector.6560-6
082-000582-00 M6Q-030024--- Fitting 3140-06-00
(Deleted)
Connector.6560-8
082-000583-00 M6Q-030025--- tube to tube.Y type
(Deleted)
connector.cross type
082-001201-00 66006 M6Q-030097--- connector.cross type
(Deleted)

No.: OMZ-21009-A Series Page 262of 428


PMLS Global Technical Support Dept

connector.Φ4 to Φ6 4 to 6.straight
082-001285-00 M6Q-030068---
(Deleted) connector

Maintenanc
e For all A7 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

11, Above new spare parts No. will be released in Part list 20150901 in September, Please pay
attention to our TN and update your Part list in time.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
ECL112 Technical Changes TMZ-15044-A7

Version Effective Date Description


1.0 2015-8-3 Document released.

No.: OMZ-21009-A Series Page 263of 428


PMLS Global Technical Support Dept

TMZ-15043-A5

ECL112 Technical Changes

Technical News
Subject: ECL112 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-15043-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Parts Change
Date: Aug.3rd,2015 Appearance
Model: A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

1, New backup cylinder assembly and copper tubes of high pressure gauge are released, old backup
cylinder assembly could not be ordered any more, but old copper tubes of high pressure gauge can
be ordered normally. The new cylinder assembly and tubes are not compatible with old one. Change
details and maintenance plan please check below
Old backup cylinder assembly and tubes New backup cylinder assembly and tubes

Appearance
differences

O2 cylinder assembly(for
O2 Yoke Assembly(0634/With
115-015777-00 gas bench), without tube 115-034395-00
tube)FRU
(Deleted)
N2O cylinder
N2O Yoke
Spare Parts assembly(for gas bench),
115-015779-00 115-034397-00 Assembly(0634/With
changes without tube
tube)FRU
(Deleted)
AIR cylinder assembly(for
AIR Yoke Assembly(0634/With
115-015778-00 gas bench), without tube 115-034396-00
tube)FRU
(Deleted)

No.: OMZ-21009-A Series Page 264of 428


PMLS Global Technical Support Dept
tube of pressure gauge
Tube of pressure
801-0631-00123-00 (left)FRU 115-034389-00
gauge(0634/left)FRU
(Supply normally)
tube of pressure gauge
Tube of pressure
801-0631-00124-00 (middle)FRU 115-034390-00
gauge(0634/middle)FRU
(Supply normally)
tube of pressure
Tube of pressure
801-0631-00125-00 gauge(right)FRU 115-034391-00
gauge(0634/right)FRU
(Supply normally)
1, For A5 machines which use old cylinder assembly , please apply new
Fault of Cylinder cylinder assembly for repair.
assembly 2, For A5 machines which use new cylinder assembly, please apply the new
Maintenance cylinder assembly for repair.
Plan 1, For A5 machines which use old copper tubes, please apply old copper tubes
Fault of the High for repair.
pressure Tube 2. For A5 machines which use new copper tubes, please apply the new copper
tubes for repair.
Upgrade N/A N/A

2, New circuit heater is released, old circuit heater could not be ordered any more, the new circuit
heater is compatible with old one. Change details and maintenance plan please check below
Old circuit heater New circuit heater

Appearance
differences

Crimson Light red

Spare Parts Circuit Heater Circuit Heater


801-0631-00069-00 115-034450-00
changes (Deleted) FRU(0634)
Maintenanc
e For all A5 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

3, New back pressure valve is released. Old back pressure valve could not be ordered any more, the
new back pressure valve is compatible with old one. Change details and maintenance plan please
check below
Old back pressure valve New back pressure valve

No.: OMZ-21009-A Series Page 265of 428


PMLS Global Technical Support Dept

Appearance
differences

Copper Aluminum

Spare Parts Back Pressure Valve back pressure valve


801-0631-00042-00 115-033941-00
changes (Deleted) assembly(0634)
Maintenanc
e For all A5 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

4, The material of drawer changes to SGCC from aluminum. A new part No. is created for new material
drawer, old drawer could not be ordered any more, the new drawer is compatible with old one. Change
details and maintenance plan please check below
Old drawer New drawer

Appearance
differences

Appearance is same as before.


Drawer Assembly
Spare Parts Drawer Module
115-017034-00 FRU(0631) 115-034451-00
changes FRU(0634)
(Deleted)
Maintenanc
e For all A5 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

5, Tube from Expiration Valve to Breathing Circuit and tube from Breathing Circuit to AGSS
container have changed from Silicone to spring tubes. Old tubes could not be ordered any more,
the new tubes are compatible with old ones. Change details and maintenance plan please check

No.: OMZ-21009-A Series Page 266of 428


PMLS Global Technical Support Dept
below
Old tubes New tubes

M6G-020042--- 115-034449-00

Appearance
differences

M6G-020039--- 115-034448-00

Silicone Breathing Hose


Spring-tube of driving
M6G-020039--- Adult 450mm 115-034448-00
gas FRU
Spare Parts (Deleted)
changes Silicone Breathing
Spring-tube of APL
M6G-020042--- Tube(350mm) 115-034449-00
valve FRU
(Deleted)
Maintenanc
e For all A5 machines, Using the new spare part for repair..
Plan
Upgrade N/A N/A

6, New Gas Cylinder Wrench is released. A new part No. is created for new Gas Cylinder Wrench; old
Gas Cylinder Wrench could not be ordered any more, the new Gas Cylinder Wrench is compatible with
old one. Change details and maintenance plan please check below
Old Gas Cylinder Wrench Old Gas Cylinder Wrench

Appearance
differences

Appearance is same as before.


Gas Cylinder
Spare Parts gas cylinder wrench
041-004480-00 Wrench 115-033063-00
changes assembly
(Deleted)
Maintenance
For all A5 machines, Using the new spare part for repair.
Plan
Upgrade Use the new part

7, The tubes for O2&N2O&AIR are changed from colorized to transparent, and the material of most
connector changes from metal to plastic, the new tubes and connectors are compatible with old ones.
Change details and maintenance plan please check below

No.: OMZ-21009-A Series Page 267of 428


PMLS Global Technical Support Dept
Old tubes and connectors New tubes and connectors

Appearance
differences

tube.PU (polyether)
082-000517-00 5.5mmX8mm orange
(Deleted)
Hose,soft PU
tube.PU (polyether)
M6G-020045--- transparent tube
082-000521-00 5.5mmX8mm green
8mmX5.5mm
(Deleted)
tube.PU 5.5mmX8mm blue
082-000662-00
(Deleted)
tube.PU (polyether)
082-000522-00 4mmX6mm green
(Deleted)
tube.PU (polyether)
Transparent PU Tube
082-000520-00 4mmX6mm orange M6G-020026---
4*6 (20m)
(Deleted)
tube.PU (polyether)
082-000524-00 4mmX6mm blue
Spare Parts
(Deleted)
changes
tube.PU (polyether)
082-000523-00 2.5mmX4mm green
(Deleted)
tube.PU (polyether)
082-000518-00 2.5mmX4mm orange M6G-020046--- Tubing.
(Deleted)
tube.PU (polyether)
082-000516-00 2.5mmX4mm blue
(Deleted)
Connector.6560-6
082-000582-00 M6Q-030024--- Fitting 3140-06-00
(Deleted)
Connector.6560-8
082-000583-00 M6Q-030025--- tube to tube.Y type
(Deleted)
connector.cross type 66006
082-001201-00 M6Q-030097--- connector.cross type
(Deleted)
082-001285-00 connector.Φ4 to Φ6 M6Q-030068--- 4 to 6.straight

No.: OMZ-21009-A Series Page 268of 428


PMLS Global Technical Support Dept
(Deleted) connector
Maintenanc
e For all A5 machines, Using the new spare part for repair.
Plan
Upgrade N/A N/A

8, Above new spare parts No. will be released in Part list 20150901 in September, Please pay attention
to our TN and update your Part list in time.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
ECL112 Technical Changes TMZ-15043-A5

Version Effective Date Description


1.0 2015-8-3 Document released.

No.: OMZ-21009-A Series Page 269of 428


PMLS Global Technical Support Dept

TMZ-15016-A7

EFU006 Technical Changes

Technical News
Subject: EFU006 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-15016-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shi ling Parts Change
Date: Apr.10th,2015 Appearance
Model: A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

The Software Technical Changes include:


⚫ After clicking Discharge button, below settings in Setup menu will go back to the O.R
Defaults when choose the “Restore default settings”.
a, In【General】menu(Pic 1):
Gas Bench Flow Rate, Balance Gas, Fresh Gas Control, Breathing System(Warmer)

No.: OMZ-21009-A Series Page 270of 428


PMLS Global Technical Support Dept

Pic 1 【General】menu
b, In【Display】menu(Pic2):
Pressure Display, Plimit Line, CO2 Placement, Gas Scales, Waveform Display, Screen
Brightness, Key Click Volume

图 2 【Display】menu changing buttons after “restore default settings”

c, In【Demographics】menu(Pic 3):
Patient ID, First Name, Last Name, DOB, Bed, Room, Point of Care, Weight, Facility
No.: OMZ-21009-A Series Page 271of 428
PMLS Global Technical Support Dept

Pic 3 【Demographics】 menu changing buttons after “restore default settings”


d, In【Alarms】menu(Pic 4):
Limits, Agents, Audio

Pic 4 【Alarms】menu changing buttons after “restore default settings”


f, In【History】menu(Pic 5):
Display Interval, Display Group, Zoom

No.: OMZ-21009-A Series Page 272of 428


PMLS Global Technical Support Dept

Pic 5 【History】menu changing buttons after “restore default settings”


g, In【CO2 Setup】menu(Pic 6):
Flow Rate, Humidity Compen,O2 Compen,N2O Compen, Des Compen

Pic.6 【CO2 Setup】menu changing buttons after “restore default settings”


h, In【BIS Setup】menu(Pic 7):
EEG Filter, Smoothing Rate

No.: OMZ-21009-A Series Page 273of 428


PMLS Global Technical Support Dept

Pic.7 【BIS Setup】menu changing buttons after “restore default settings”

⚫ Add RAW and Compl display In Spirometry Wave (Pic 8) .

Pic 8 Spirometry

⚫ Circuit compliance compensation not available when the Leak test fail or skip after
starting the AM, and Display the Warning when exit the Standby Mode(Pic 9,10) .

No.: OMZ-21009-A Series Page 274of 428


PMLS Global Technical Support Dept

Pic 9 warning of exit Standby mode(skip Leak test)

Pic 10 warning of exit Standby mode(Leak test Failure)


Note:the warning will not display after the sucussful of Leak test.
⚫ Solve the problem of TV lower than other brands in the same conditions when use
the PCV mode.
⚫ Repair the bug about PCV - VG model can not be achieved in extreme case.
⚫ Repair the bug about AM can not be connected to the Benelink.

New software is launched


No. Package Modules Latest Version
1 Bundle Version V02.05.07

No.: OMZ-21009-A Series Page 275of 428


PMLS Global Technical Support Dept
Host Software V03.05.07
BIOS V01.12.00
FPGA Display V1.4
Bundle FPGA Sound V1.2
(V02.05.07) Ventilator Protect Module V1.1
Ventilator Control Module V01.15.00
Language V3.9
Keyboard V01.00/01.01
EFCS
2 EFCS Software V01.04.00
(V01.04.00)
3 EFCS_FPGA(V3.0) FPGA EFCS V3.0
4 Power(V1.7) Power System V1.7

NOTE:
1, Downgrade the higher version to lower version is not allowed.
2,For devices which have bundle version 02.04.00 or lower, after upgrading to 02.05.07 and
restarting for first time, an alarm “Flowmeter Self Test : Fail" will come during self test, please perform
steps below to solve this issue:
a, Enter the "Service Access" mode by "789789";
b, In standby mode, select 【Setup】->【Service】->【Disabled BFCS】, close the BFCS;
c, In standby mode, select 【Setup】->【General】->【Zero Flow Meters】,zero flow meters
and restart again;
3, For those devices which have bundle version 02.05.00 or higher, steps above are unnecessary
during upgrading to 02.05.07.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 276of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: Ref No:
EFU006 Technical Changes TMZ-15016-A7

Version Effective Date Description


1.0 2015-4-11 Document released.

No.: OMZ-21009-A Series Page 277of 428


PMLS Global Technical Support Dept

TMZ-15015-A5

EFU006 Technical Changes

Technical News
Subject: EFU006 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-15015-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shiling Parts Change
Date: Apr.10th,2015 Appearance
Model: A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

This Technical Changes include:


⚫ After clicking Discharge button, below settings in Setup menu will go back to the O.R
Defaults when choose the “Restore default settings”.
a, In【General】menu(Pic 1):
Gas Bench Flow Rate, Breathing System(Warmer)

No.: OMZ-21009-A Series Page 278of 428


PMLS Global Technical Support Dept

Pic 1 【General】menu changing buttons after “restore default settings”


b, In【Display】menu(Pic2):
Pressure Display, Plimit Line, CO2 Placement, Gas Scales, Waveform Display, Screen
Brightness, Key Click Volume

图 2 【Display】menu changing buttons after “restore default settings”


c, In【Demographics】menu(Pic 3):
Patient ID, First Name, Last Name, DOB, Bed, Room, Point of Care, Weight, Facility
No.: OMZ-21009-A Series Page 279of 428
PMLS Global Technical Support Dept

Pic 3 【Demographics】 menu changing buttons after “restore default settings”


d, In【Alarms】menu(Pic 4):
Limits, Agents, Audio

Pic 4 【Alarms】menu changing buttons after “restore default settings”


f, In【History】menu(Pic 5):
Display Interval, Display Group, Zoom

No.: OMZ-21009-A Series Page 280of 428


PMLS Global Technical Support Dept

Pic 5 【History】menu changing buttons after “restore default settings”


g, In【CO2 Setup】menu(Pic 6):
Flow Rate, Humidity Compen,O2 Compen,N2O Compen, Des Compen

Pic.6 【CO2 Setup】menu changing buttons after “restore default settings”


h, In【BIS Setup】menu(Pic 7):
EEG Filter, Smoothing Rate

No.: OMZ-21009-A Series Page 281of 428


PMLS Global Technical Support Dept

Pic.7 【BIS Setup】menu changing buttons after “restore default settings”

⚫ Add RAW and Compl display In Spirometry Wave (Pic 8) .

Pic 8 Spirometry

⚫ Circuit compliance compensation not available when the Leak test fail or skip after
starting the AM, and Display the Warning when exit the Standby Mode(Pic 9,10).

No.: OMZ-21009-A Series Page 282of 428


PMLS Global Technical Support Dept

Pic 9 warning of exit Standby mode(skip Leak test)

Pic 10 warning of exit Standby mode(Leak test Failure)


Note:the warning will not display after the sucussful of Leak test.
⚫ Solve the problem of TV lower than other brands in the same conditions when use
the PCV mode.
⚫ Repair the bug about PCV - VG model can not be achieved in extreme case.
⚫ Repair the bug about AM can not be connected to the Benelink.

New software is launched


No. Package Modules Latest Version
Bundle Version V02.05.07
Bundle Host Software V03.05.07
1
(V02.05.07) BIOS V01.12.00
FPGA Display V1.4
No.: OMZ-21009-A Series Page 283of 428
PMLS Global Technical Support Dept
FPGA Sound V1.2
Ventilator Protect Module V1.1
Ventilator Control Module V01.15.00
Language V3.9
Flowmeter Software V01.00/01.01
2 Power(V1.7) Power System V1.7
NOTE:Downgrade the higher version to lower version is not allowed
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
EFU006 Technical Changes TMZ-15015-A5

Version Effective Date Description


1.0 2015-4-11 Document released.

No.: OMZ-21009-A Series Page 284of 428


PMLS Global Technical Support Dept

TMZ-14047-A7

ECL096-F Technical Changes

Technical News
Subject: ECL096-F Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-14047-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Ma Parts Change
Date: Dec. 17th 2014 Appearance
Model: A7 Stop Production
Comments: N/A Others

CONTENT
To Whom It May Concern,
Below content is only for your information.
1. For distinguishing between BFCS(O2&AIR) and BFCS(O2&AIR&N2O).A new
optimizing is released for the configuration A7 BFCS(O2&AIR).This change gives a
better appearance and keeps all the functions offered by old type.Table1 below will
give
The information would be used in installation, for maintenance please check Table2.
More change details please check table1
Before change(Old type) After change(New type)

Appearanc
e

No.: OMZ-21009-A Series Page 285of 428


PMLS Global Technical Support Dept

Inside
structure

Block needle valve

Old type A7 BFCS(O2&AIR) used same New type use new


appearance with appearance ,makes it easy to
Description
BFCS(O2&AIR&N2O),this design is distinguish from BFCS(O2&AIR&N2O)
confusing.
Table 1 optimizing details
Maintenance plans for new type please check table2
Before change(Old type) After change(New type)
Malfunction
Order No. Parts name Order No. Parts name
O2 needle 115-029982-00 O2 needle valve 115-029990- O2 needle valve
valve FRU(0635/US/w/gua 00 FRU(0635/US/w/guard)
failure(US rd)
Standard)
O2 needle 115-029983-00 O2 needle valve 115-029991- O2 needle valve
valve FRU(0635/EU/w/gua 00 FRU(0635/EU/w/guard)
failure(EU rd)
Standard)
O2 needle 115-029985-00 O2 needle valve 115-029993- O2 needle valve
valve FRU(0635/Canada/w 00 FRU(0635/Canada/w/g
failure(Canad /guard) uard)
a Standard)
Air needle 115-014468-00 needle valve 115-014468- needle valve
valve assembly(0632) 00 assembly(0632)
failure(for all
standards)
Panel of 115-029974-00 Panel of 115-029976- Panel of
gauge gauge(FRU/0632/US 00 gauge(FRU/0635/US/O
break(US ) 2&AIR)
standard)
Panel of 115-029975-00 Panel of 115-029977- Panel of
gauge gauge(FRU/0632/rev 00 gauge(FRU/0635/Rev_
break(Rev_E erse EU) EU/O2&AIR)

No.: OMZ-21009-A Series Page 286of 428


PMLS Global Technical Support Dept
U standard)

Panel of 115-023186-00 Panel of 115-029978- Panel of


gauge gauge(FRU/0635/EU 00 gauge(FRU/0635/EU/O
break(EU ) 2&AIR)
standard)
Panel of 115-023187-00 Panel of 115-029979- Panel of
gauge gauge(FRU/0635/EU 00 gauge(FRU/0635/EU/O
break(EU /big tank) 2&AIR)
standard/Big
cylinder)

Table 2 maintenance plans

Affected FRU List


NO. Order No. Description Version

1 115-029982-00 O2 needle valve FRU(0635/US/w/guard) 1.0

2 115-029983-00 O2 needle valve FRU(0635/EU/w/guard) 1.0

3 115-029985-00 O2 needle valve FRU(0635/Canada/w/guard) 1.0

4 115-029974-00 Panel of gauge(FRU/0632/US) 1.0

5 115-029975-00 Panel of gauge(FRU/0632/reverse EU) 1.0

6 115-023186-00 Panel of gauge(FRU/0635/EU) 1.0

7 115-023187-00 Panel of gauge(FRU/0635/EU/big tank) 1.0

Compatibility Table
A7(0635) Before After Compatibility
O2 needle valve failure(US
115-029982-00 115-029990-00 NO
Standard)

O2 needle valve failure(EU


115-029983-00 115-029991-00 NO
Standard)

O2 needle valve
115-029985-00 115-029993-00 NO
failure(Canada Standard)

Panel of gauge break(US


115-029974-00 115-029976-00 NO
standard)

Panel of gauge break(Rev_EU


115-029975-00 115-029977-00 NO
standard)

No.: OMZ-21009-A Series Page 287of 428


PMLS Global Technical Support Dept

Panel of gauge break(EU


115-023186-00 115-029978-00 NO
standard)

Panel of gauge break(EU


115-023187-00 115-029979-00 NO
standard/Big cylinder)

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
ECL090-F Technical Changes TMZ-14047-A7

Version Effective Date Description


1.0 2014-12-19 Document released.

No.: OMZ-21009-A Series Page 288of 428


PMLS Global Technical Support Dept

TMZ-14035-A7

EFU004 Technical Changes

Technical News
Subject: EFU004 Technical Changes

RECORD IDENTIFIER

Ref No.:TMZ-14035-A7 Relate To:


Version: 1.0 Software/Manual Upgrade
Author: Abraham Ma Parts Change
Date: Sept. 29th 2014 Appearance
Model: A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,


Below content is only for your information.
1. Operation Manual upgrade:
1) Bellows cleaning way changes (Figure 1)
Content: Maintenance-Cleaning and Disinfection-Cleaning and Disinfecting
Agents / Autoclaving

No.: OMZ-21009-A Series Page 289of 428


PMLS Global Technical Support Dept
Figure 1 Bellows cleaning way changes

2) Breathing system specifications change(Figure 2)


Content: Product Specifications-Breathing System Specifications-Breathing
System Volume

Figure 2 Breathing system specifications change


Structure of Bellows ,Breathing system doesn’t change, P/N numbers are the same,
installation ways are the same with before

2. D-vaporizer assembly(plug-in)( P/N: 115-027692-00) for A7 is released newly.(Figure


3)
Note: 1) Vaporizer offered by Mindray can’t be installed in plug-in chassis
2)Selectatec D-vaporizer assembly don’t support upgrade to Plug-in(Figure 4)

No.: OMZ-21009-A Series Page 290of 428


PMLS Global Technical Support Dept

Figure 3 D-vaporizer assembly(plug-in)

Figure 4 selectatec & Plug-in comparation


3. A7 new device(Air&N2O&O2 3-way gas source) adds optional configuration as below:
BFCS (Air&N2O&O2 3-way gas control) or BFCS(Air&O2 2-way gas control).
Note:1) Configuration [A7(Air&N2O&O2 3-way gas source) + BFCS (Air&N2O&O2
3-way gas control)] can’t upgrade to Configuration [A7(Air&N2O&O2 3-way
gas source) + BFCS (Air&O2 2-way gas control)],opposite doesn’t support
upgrade, either.
2) If a device configuration is [A7(Air&N2O&O2 3-way gas source) + BFCS
(Air&O2 2-way gas control)], please read the corresponding service guides &
spare parts of A7 (Air&O2 2-way gas source) BFCS when the BFCS
(Air&O2 2-way gas control) need maintenance .

No.: OMZ-21009-A Series Page 291of 428


PMLS Global Technical Support Dept
4. A7 Venturi negative pressure Muffler Assembly changes(Figure 5&6)to increase the
negative pressure 。 The Muffler Assembly P/N number doesn’t change(P/N115-
018356-00), installation & maintenance ways are the same with before.(Figure 7)

Figure 5 Muffler Assembly changes1

Figure 6 Muffler Assembly changes2

No.: OMZ-21009-A Series Page 292of 428


PMLS Global Technical Support Dept

Figure 7 Muffler Assembly installation


5. AGSS passive scavenging tube assembly connector external diameter
decrease(Figure 8), since the old one had interference with Quick tube Connect
PMCD.(Figure 9)
AGSS passive scavenging tube assembly P/N number doesn’t change, installation &
maintenance ways are the same with old one

Figure 8 tube assembly connector external diameter comparation

No.: OMZ-21009-A Series Page 293of 428


PMLS Global Technical Support Dept

Figure 9 interference between Quick Connect PMCD & tube connector


Affected FRU List
No. Order No. Description Version
1 115-018356-00 Muffler Assembly 1.0
2 115-002015-00 AGSS passive scavenging tube assembly 2.0
3 115-002342-00 Passive scavenging tube material packet 1.0

Compatibility Table
A5 (0634) Before After Compatibility
115-
Muffler Assembly 018356- 115-018356-00 YES
00
115-
AGSS passive scavenging tube assembly 002015- 115-002015-00 YES
00
115-
Passive scavenging tube material packet 002342- 115-002342-00 YES
00

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments

No.: OMZ-21009-A Series Page 294of 428


PMLS Global Technical Support Dept
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
EFU004 Technical Changes TMZ-14035-A7

Version Effective Date Description


1.0 2014-10-08 Document released.

No.: OMZ-21009-A Series Page 295of 428


PMLS Global Technical Support Dept

TMZ-14034-A5

EFU004 Technical Changes

Technical News
Subject: EFU004 Technical Changes

RECORD IDENTIFIER

Ref No.:TMZ-14034-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Ma Parts Change
Date: Sept. 30th 2014 Appearance
Model: A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,


Below content is only for your information.
1. Operation Manual upgrade:
1) Bellows cleaning way changes (Figure 1)
Content: Maintenance-Cleaning and Disinfection-Cleaning and Disinfecting
Agents / Autoclaving

Figure 1 Bellows cleaning way changes

2) Breathing system specifications change(Figure 2)

No.: OMZ-21009-A Series Page 296of 428


PMLS Global Technical Support Dept
Content: Product Specifications-Breathing System Specifications-Breathing
System Volume

Figure 2 Breathing system specifications change


Structure of Bellows,Breathing system doesn’t change, P/N numbers are the same,
installation ways are the same with before.

2. D-vaporizer assembly(plug-in)( P/N: 115-027692-00) for A5 is released newly.(Figure


3)
Note: 1) Vaporizer offered by Mindray can’t be installed in plug-in chassis
2)Selectatec D-vaporizer assembly don’t support upgrade to Plug-in(Figure 4)

Figure 3 D-vaporizer assembly(plug-in)

No.: OMZ-21009-A Series Page 297of 428


PMLS Global Technical Support Dept

Figure 4 selectatec & Plug-in comparation


3. AGSS passive scavenging tube assembly connector external diameter
decrease(Figure 5), since the old one had interference with Quick tube Connect
PMCD.(Figure 6)
AGSS passive scavenging tube assembly P/N number doesn’t change, installation &
maintenance way are the same with old one.

Figure 5 tube assembly connector external diameter comparation

No.: OMZ-21009-A Series Page 298of 428


PMLS Global Technical Support Dept

Figure 6 interference between Quick Connect PMCD & tube connector


Affected FRU List
No. Order No. Description Version
1 115-002015-00 AGSS passive scavenging tube assembly 2.0
2 115-002342-00 Passive scavenging tube material packet 1.0

Compatibility Table
A5 (0634) Before After Compatibility
115-002015-
AGSS passive scavenging tube assembly 115-002015-00 YES
00
115-002342-
Passive scavenging tube material packet 115-002342-00 YES
00

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: Ref No:
EFU004 Technical Changes TMZ-14034-A5

Version Effective Date Description


No.: OMZ-21009-A Series Page 299of 428
PMLS Global Technical Support Dept
1.0 2014-10-08 Document released.

No.: OMZ-21009-A Series Page 300of 428


PMLS Global Technical Support Dept

TMZ-14033-A7

ECL090-F Technical Changes

Technical News
Subject: ECL090-F Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-14033-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Ma Parts Change
Date: Sept. 22th 2014 Appearance
Model: A7 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


1. Flexible arm(P/N: 115-021470-00) for A5 is released newly.(Figure 1)
Note: This part can be used to upgrade

Figure 1 Flexible arm


2. Three-position vaporizer manifold(P/N: 115-020218-00)for A5 is released newly.
(Figure 2)
Note: This part is for maintenance only, can not be used to upgrade

Figure 2 Three-position vaporizer manifold

No.: OMZ-21009-A Series Page 301of 428


PMLS Global Technical Support Dept
3. the A7 cylinder yokes have been updated for solving the malfunction: leakage occurred
because the client used a cylinder which has higher pressure(20MPa) than the
anesthesia machine maximum cylinder pressure specification (15.5MPa).

Change Info as below(Table 1)


Part description P/N(Before change) P/N(After change)
O2 cylinder yoke assembly
115-014476-00 115-014476-01
(0632)
N2O cylinder yoke
115-014477-00 115-014477-01
assembly (0632)
Air cylinder yoke assembly
115-014478-00 115-014478-01
(0632)
Table 1 Change Information

4. Breath System Update:


1)New mediate plate ((Figure 3) has been used to increase the strength of breathing
circuit.
2)Change the O-ring at the connection between the lower cover 2 and the bottom
shell to star-shaped sealing ring(P/N: 049-000813-00) for solving the leakage caused
by mediate plate gets bent .(Figure 4)

Wider than old

New mediate plate Old mediate plate

Figure 3 Mediate plate

Figure 4 Star-shaped sealing ring(P/N: 049-000813-00)

5. The pneumatic system has changed, the previous position of Back Pressure Valve
No.: OMZ-21009-A Series Page 302of 428
PMLS Global Technical Support Dept
is between gas mixer assembly and 3-way valve,but now it has moved to the behind
of main flowmeter assembly,this change can decrease the pressure undulating of
EFCS, make the floater in main flowmeter assembly more stable. P/N of Back
Pressure Valve, main flowmeter assembly, gas mixer assembly has changed.
Change Info as below(Table 2)
Before change After change
Part description
P/N Remark P/N Remark
801-0631-00042- / /
Back Pressure Valve 115-028145-00
00
Assembly has Assembly has new
old silkscreen silkscreen on
on it ,and it ,and contents
main flowmeter assembly 115-014479-00 doesn’t content 115-027534-00 some character
character labels labels and a
and any service service guide
guide
Assembly has Assembly has new
old silkscreen silkscreen on
on it ,and it ,and contents
gas mixer assembly(0632) 115-016781-00 doesn’t content 115-027532-00 some character
character labels labels and a
and any service service guide
guide
Table 2 Change Information
Note: Since the Back Pressure Valve position has changed,,the silkscreen on gas mixer
assembly and total flowmeter assembly can’t match the main pneumatic system tube
connection diagram.
For solving the problem above, we offer the maintenance plans as below(Table 3)
Maintenance Plan
device made after 2014 Sept.
device made before 2014 Sept.
malfunction (Back Pressure Valve is between
(Back Pressure Valve is between gas mixer assembly
main flow sensor assembly and
and 3-way valve)
main flowmeter assembly)
Back Pressure
Valve Change Back Pressure Valve Change Back Pressure Valve
malfunction

After changing gas mixer assembly,please read

gas mixer 《Service Guide for A7 Old Gas Mixer Assembly and
assembly Just change gas mixer assembly
malfunction Old Total Flowmeter Assembly》 which will send

together with spare part, change the silkscreen using


character labels(Figure 5)

No.: OMZ-21009-A Series Page 303of 428


PMLS Global Technical Support Dept
After changing main flowmeter assembly, please read

《Service Guide for A7 Old Gas Mixer Assembly and


main flowmeter
Just change main flowmeter
assembly
Old Total Flowmeter Assembly》 which will send assembly
malfunction
together with spare part, change the silkscreen using
character labels(Figure 6)
Table 3 maintenance plan

After changing
gas mixer
assembly

silkscreen “122”

in the picture
should change to

“58”
Figure5 gas mixer assembly

After changing main


flowmeter assembly,

silkscreen “26” in the

picture should change to

“122”

Figure6 main flowmeter assembly


6. 3 new ORC service kits are added now to solve a calibration must be down with N2O
after replacement.
Parts Info as below(Table 4)
Part description P/N content

ORC Assembly (0632) 115-016780-01 ORC Assembly 1 unit

Include ORC Assembly 1 unit,


ORC service kit (0635/ Opposite European
115-026814-00
Standard)FRU needle valve assembly 2 units(for
valve change)

No.: OMZ-21009-A Series Page 304of 428


PMLS Global Technical Support Dept

Include ORC Assembly 1 unit,


ORC service kit (0635/ European
115-026815-00
Standard)FRU needle valve assembly 2 units(for
valve change)

Include ORC Assembly 1 unit,


ORC service kit (0635/ American
115-026816-00
Standard)FRU needle valve assembly 2 units(for
valve change)
Table 4 Parts Information
New ORC service kit affords more maintenance plans.(Table 5)
Maintenance plan(After change)
malfunction
No N2O, can’t do calibration Have N2O,calibration is possible

Change ORC service kit totally, Change ORC assembly,do ORC


ORC malfunction
calibration is unnecessary
calibration(follow service manual)

O2 needle valve Change ORC service kit totally, Change O2 needle valve only,do ORC
malfunction calibration is unnecessary
calibration(follow service manual)

N2O needle valve Change ORC service kit totally, Change N2O needle valve only,do ORC
malfunction calibration is unnecessary
calibration(follow service manual)

Table 5 Maintenance plan Form

7. Decrease the output pressure of high-pressure regulator at the cylinder gas supply
from 400 ±40KPa to 360 ±40KPa for solving the malfunction: when cylinder gas supply
and pipeline gas supply were switched on simultaneously, since the cylinder side has
higher pressure, the anesthesia machine used the gas inside the cylinder first. This
spare part do not change the order number, installation method is the same with old
one.

8. Tube color in the Vacuum suction assembly has changed from orange to transparent.

Affected FRU List


NO. Order No. Description Version
0632 Flowmeter O2 flow sensor board
1 051-001414-01 1.0
PCBA
2 051-001415-01 0632 Flowmeter Air sensor board PCBA 1.0
3 051-001416-01 0632 Flowmeter N2O sensor board PCBA 1.0
4 115-014468-00 needle valve assembly(0632) 3.0
5 115-014476-00 O2 cylinder assembly (0632) 1.0
6 115-014477-00 N2O cylinder assembly (0632) 1.0
7 115-014478-00 AIR cylinder assembly (0632) 1.0

No.: OMZ-21009-A Series Page 305of 428


PMLS Global Technical Support Dept
8 115-014479-00 main flowmeter assembly 1.0
9 115-014519-01 backup valves in-out device(0632) 1.0
10 115-015639-00 O2 needle valve (0632/with basic volume) 3.0
11 115-016480-00 gas bench receptacle assembly 1.0
12 115-016692-00 Brake Assembly(FRU) 1.0
13 115-016780-00 ORC assembly(0632) 3.0
14 115-016781-00 gas mixer assembly(0632) 3.0
15 115-018145-00 0632 Power board FRU 1.0
16 115-018146-00 Main control board FRU(0632/FDA) 1.0
17 115-018148-01 Ventilation kit(New Diff_pressure)FRU 1.0
18 115-018420-00 pipeline suction upgrade kit(Br_BS) 3.0
19 115-018426-00 pipeline suction upgrade kit(Au_SIS) 3.0
20 115-018427-00 pipeline suction upgrade kit(Fr_FS) 3.0
21 115-018428-00 pipeline suction upgrade kit(Ge_GS) 3.0
22 115-020937-00 Drive Assembly(New Diff_pressure)FRU 1.0
23 115-021094-00 service door maintenance kit(FRU) 2.0
24 115-021249-00 Mid-plate FRU(0631/no ventilation valve) 1.0
25 115-021254-00 Breathing Circuit Cell FRU(0632) 1.0
26 115-022926-00 N2O needle valve(0632) 1.0
27 115-024065-00 pipeline suction upgrade kit(US_BS/0635) 1.0
28 115-024066-00 pipeline suction upgrade kit(DISS/0635) 1.0
29 115-024067-00 pipeline suction upgrade kit(US_PB/0635) 1.0
30 115-024068-00 pipeline suction upgrade kit(US_OH/0635) 1.0
31 115-024069-00 pipeline suction upgrade kit(US_CH/0635) 1.0
32 115-024629-00 gas mixer assembly(0635/O2/AIR) 1.0
33 115-024630-00 O2 needle valve (0635/with basic volume) 1.0
34 801-0621-00165-00 0621 Infrared Communication Board 1.0
35 801-0631-00039-00 0631 top lighting board PCBA 1.0
36 801-0631-00042-00 Back Pressure Valve 1.0
37 801-0631-00061-00 Check valve dome, A series 4.0
38 801-0631-00120-00 work surface assembly(0631)FRU 2.0

Compatibility Table
A7(0635) Before After Compatibility
0632 Flowmeter O2 flow sensor board
051-001414-01 051-001414-01 YES
PCBA
0632 Flowmeter Air sensor board PCBA 051-001415-01 051-001415-01 YES
0632 Flowmeter N2O sensor board PCBA 051-001416-01 051-001416-01 YES
needle valve assembly(0632) 115-014468-00 115-014468-00 YES
O2 cylinder assembly (0632) 115-014476-00 115-014476-01 YES
N2O cylinder assembly (0632) 115-014477-00 115-014477-01 YES
AIR cylinder assembly (0632) 115-014478-00 115-014478-01 YES

No.: OMZ-21009-A Series Page 306of 428


PMLS Global Technical Support Dept
main flowmeter assembly 115-014479-00 115-027534-00 YES
backup valves in-out device(0632) 115-014519-01 115-014519-01 YES
O2 needle valve (0632/with basic volume) 115-015639-00 115-015639-00 YES
gas bench receptacle assembly 115-016480-00 115-016480-00 YES
Brake Assembly(FRU) 115-016692-00 115-016692-00 YES
115-016780-01 YES
115-026814-00 YES
ORC assembly(0632) 115-016780-00
115-026815-00 YES
115-026816-00 YES
gas mixer assembly(0632) 115-016781-00 115-027532-00 YES
0632 Power board FRU 115-018145-00 115-018145-00 YES
Main control board FRU(0632/FDA) 115-018146-00 115-026808-00 YES
Ventilation kit(New Diff_pressure)FRU 115-018148-01 115-018148-01 YES
pipeline suction upgrade kit(Br_BS) 115-018420-00 115-018420-00 YES
pipeline suction upgrade kit(Au_SIS) 115-018426-00 115-018426-00 YES
pipeline suction upgrade kit(Fr_FS) 115-018427-00 115-018427-00 YES
pipeline suction upgrade kit(Ge_GS) 115-018428-00 115-018428-00 YES
Drive Assembly(New Diff_pressure)FRU 115-020937-00 115-020937-00 YES
service door maintenance kit(FRU) 115-021094-00 115-026810-00 YES
Mid-plate FRU(0631/no ventilation valve) 115-021249-00 115-021249-00 YES
Breathing Circuit Cell FRU(0632) 115-021254-00 115-021254-00 YES
N2O needle valve(0632) 115-022926-00 115-022926-00 YES
pipeline suction upgrade kit(US_BS/0635) 115-024065-00 115-024065-00 YES
pipeline suction upgrade kit(DISS/0635) 115-024066-00 115-024066-00 YES
pipeline suction upgrade kit(US_PB/0635) 115-024067-00 115-024067-00 YES
pipeline suction upgrade kit(US_OH/0635) 115-024068-00 115-024068-00 YES
pipeline suction upgrade kit(US_CH/0635) 115-024069-00 115-024069-00 YES
gas mixer assembly(0635/O2/AIR) 115-024629-00 115-027533-00 YES
O2 needle valve (0635/with basic volume) 115-024630-00 115-024630-00 YES
801-0621-00165- 801-0621-00165-
0621 Infrared Communication Board YES
00 00
801-0631-00039- 801-0631-00039-
0631 top lighting board PCBA YES
00 00
801-0631-00042-
Back Pressure Valve 115-028145-00 YES
00
801-0631-00061- 801-0631-00061-
Check valve dome, A series YES
00 00
801-0631-00120- 801-0631-00120-
work surface assembly(0631)FRU YES
00 00

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
No.: OMZ-21009-A Series Page 307of 428
PMLS Global Technical Support Dept

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: ECL090-F Technical Changes Ref No: TMZ-14033-A7

Version Effective Date Description


1.0 2014-9-24 Document released.

No.: OMZ-21009-A Series Page 308of 428


PMLS Global Technical Support Dept

TMZ-14032-A5

ECL090-F Technical Changes

Technical News
Subject: ECL090-F Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-14032-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Abraham Ma Parts Change
Date: Sept. 22th, Appearance
Model: A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,


Below content is only for your information.
1. Vacuum suction tube fixing clamp (P/N: 043-003125-00) for A5 is released
newly.(Figure 1)
Note: 1) This part is added into pipeline suction upgrade kit, the order number of
pipeline suction upgrade kit do not change.
2)Old devices which have been made before Sept. 2014 can’t upgrade this
part

Vacuum suction
tube fixing clamp

Figure 1 Vacuum suction tube fixing clamp

No.: OMZ-21009-A Series Page 309of 428


PMLS Global Technical Support Dept
2. Flexible arm(P/N: 115-021470-00)for A5 is released newly.(Figure 2)
Note: This part can be used to upgrade

Figure 2 Flexible arm


3. Three-position vaporizer manifold(P/N: 115-020218-00)for A5 is released newly.
(Figure 3)
Note: This part is for maintenance only, can not be used to upgrade

Figure 3 Three-position vaporizer manifold

4. Breath System Update:


1) New mediate plate ((Figure 4) has been used to increase the strength of
breathing circuit.
Note: installation method is the same with old one
2)Change the O-ring at the connection between the lower cover 2 and the bottom
shell to star-shaped sealing ring(P/N: 049-000813-00) for solving the leakage caused
by mediate plate gets bent .(Figure 5)

Wider than old

New mediate plate Old mediate plate

Figure 4 Mediate plate

No.: OMZ-21009-A Series Page 310of 428


PMLS Global Technical Support Dept

Figure 5 Star-shaped sealing ring(P/N: 049-000813-00)

5. 3 new ORC service kits are added now to solve a calibration must be down with N2O
after replacement.
Parts Info as below(Table 1)
Part description P/N content

ORC Assembly (0632) 115-016780-01 ORC Assembly 1 unit

Include ORC Assembly 1 unit,


ORC service kit (0635/ Opposite European
115-026814-00
Standard)FRU needle valve assembly 2 units(for
valve change)

Include ORC Assembly 1 unit,


ORC service kit (0635/ European
115-026815-00
Standard)FRU needle valve assembly 2 units(for
valve change)

Include ORC Assembly 1 unit,


ORC service kit (0635/ American
115-026816-00
Standard)FRU needle valve assembly 2 units(for
valve change)
Table 1 Parts Information
New ORC service kit affords more maintenance plans.(Table 2)
Maintenance plan(After change)
malfunction
No N2O, can’t do calibration Have N2O,calibration is possible

Change ORC service kit totally, Change ORC assembly,do ORC


ORC malfunction
calibration is unnecessary
calibration(follow service manual)

O2 needle valve Change ORC service kit totally, Change O2 needle valve only,do ORC
malfunction calibration is unnecessary
calibration(follow service manual)

N2O needle valve Change ORC service kit totally, Change N2O needle valve only,do ORC
malfunction calibration is unnecessary
calibration(follow service manual)

Table 2 Maintenance plan


6. Decrease the output pressure of high-pressure regulator at the cylinder gas supply
No.: OMZ-21009-A Series Page 311of 428
PMLS Global Technical Support Dept
from 400 ±40KPa to 360 ±40KPa for solving the malfunction: when cylinder gas supply
and pipeline gas supply were switched on simultaneously, since the cylinder side has
higher pressure, the anesthesia machine used the gas inside the cylinder first. This
spare part do not change the order number, installation method is the same with old
one.
Affected FRU List
No. Order No. Description Version
1 801-0631-00039-00 0631 top lighting board PCBA 1.0
2 115-016480-00 gas bench receptacle assembly 1.0
3 801-0631-00025-00 0631 Power Board PCBA 1.0
4 801-0631-00040-00 Fresh Flow Sensor Board 2.0
5 115-016692-00 Brake Assembly(FRU) 1.0
6 801-0631-00120-00 work surface assembly(0631)FRU 2.0
7 115-003009-01 ORC Assembly 2.0
8 115-017864-00 ORC Maintenance Kit(CE Common) FRU 2.0
9 115-021249-00 Mid-plate FRU(0631/no ventilation valve) 1.0
10 801-0631-00105-00 Breathing System 3.0
11 801-0631-00103-00 Breathing Circuit Unit 3.0
12 115-017036-00 Breathing System FRU(0631) 1.0
13 115-016781-00 gas mixer assembly(0632) 3.0
14 115-020937-00 Drive Assembly(New Diff_pressure)FRU 1.0
15 115-018148-01 Ventilation kit(New Diff_pressure)FRU 1.0
16 801-0631-00037-00 Rear Door 3.0
17 801-0631-00061-00 Check valve dome, A series 4.0
18 115-006544-00 upper cover assembly(0631) 6.0
19 801-0631-00042-00 Back Pressure Valve 1.0

20 115-018146-00 Main control board FRU(SFDA) 1.0

21 115-015777-00 O2 cylinder assembly(for gas bench) 1.0


22 115-015779-00 N2O cylinder assembly(for gas bench) 1.0
23 115-015778-00 Air cylinder assembly(for gas bench) 1.0
24 801-0631-00048-00 Oxygen Cylinder Yoke Assembly 1.0
25 801-0631-00049-00 Air Cylinder Yoke Assembly 1.0
26 801-0631-00050-00 Nitrous Oxide Yoke Assembly 1.0
28 115-015266-00 Outer suction Kit (Air drive 1.0
29 115-015267-00 Outer suction Kit (Air drive 1.0
30 115-015268-00 Outer suction Kit (O2 drive 1.0
31 115-015269-00 Outer suction Kit (O2 drive 1.0

Compatibility Table
A5 (0634) Before After Compatibility
0631 top lighting board PCBA 801-0631-00039-00 801-0631-00039-00 Yes
gas bench receptacle assembly 115-016480-00 115-016480-00 Yes
0631 Power Board PCBA 801-0631-00025-00 801-0631-00025-00 Yes
No.: OMZ-21009-A Series Page 312of 428
PMLS Global Technical Support Dept
Fresh Flow Sensor Board 801-0631-00040-00 801-0631-00040-00 Yes
Brake Assembly(FRU) 115-016692-00 115-016692-00 Yes
work surface Yes
801-0631-00120-00 801-0631-00120-00
assembly(0631)FRU
ORC Assembly 115-003009-01 115-003009-01 Yes
ORC Maintenance Kit(CE Yes
115-017864-00 115-017864-00
Common) FRU
Mid-plate FRU(0631/no Yes
115-021249-00 115-021249-00
ventilation valve)
Breathing System 801-0631-00105-00 801-0631-00105-00 Yes
Breathing Circuit Unit 801-0631-00103-00 801-0631-00103-00 Yes
Breathing System FRU(0631) 115-017036-00 115-017036-00 Yes
gas mixer assembly(0632) 115-016781-00 115-027532-00 Yes
Drive Assembly(New Yes
115-020937-00 115-020937-00
Diff_pressure)FRU
Ventilation kit(New Yes
115-018148-01 115-018148-01
Diff_pressure)FRU
Rear Door 801-0631-00037-00 115-026809-00 Yes
Check valve dome, A series 801-0631-00061-00 801-0631-00061-00 Yes
upper cover assembly(0631) 115-006544-00 115-006544-00 Yes
Back Pressure Valve 801-0631-00042-00 115-028145-00 Yes
Main control board Yes
115-018146-00 115-026808-00
FRU(SFDA)

O2 cylinder assembly(for gas Yes


115-015777-00 115-015777-00
bench)
N2O cylinder assembly(for gas Yes
115-015779-00 115-015779-00
bench)
Air cylinder assembly(for gas Yes
115-015778-00 115-015778-00
bench)
Oxygen Cylinder Yoke Yes
801-0631-00048-00 801-0631-00048-00
Assembly
Air Cylinder Yoke Assembly 801-0631-00049-00 801-0631-00049-00 Yes
Nitrous Oxide Yoke Assembly 801-0631-00050-00 801-0631-00050-00 Yes
Outer suction Kit (Air drive 115-015266-00 115-015266-00 Yes
Outer suction Kit (Air drive 115-015267-00 115-015267-00 Yes
Outer suction Kit (O2 drive 115-015268-00 115-015268-00 Yes
Outer suction Kit (O2 drive 115-015269-00 115-015269-00 Yes

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

No.: OMZ-21009-A Series Page 313of 428


PMLS Global Technical Support Dept

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject:ECL090-F engineering Change Ref No: TMZ-14032-A5

Version Effective Date Description


1.0 2014-9-24 Document released.

No.: OMZ-21009-A Series Page 314of 428


PMLS Global Technical Support Dept

TMZ-14024-A7

ECL092 Technical Change

Technical News
Subject: ECL092 Technical Change

RECORD IDENTIFIER

Ref No.: TMZ-14024-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Wang Yongxin Parts Change
Date: May 13th,2014 Appearance
Model:A7 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


To meet RoHS requirement, old version of pipeline pressure gauges and high
pressure gauges will be replaced by new gauges. The P/Ns involved are below:

Old Version New Version

P/N Description P/N Description

O2 pipeline pressure O2 pipeline pressure gauge


115-023193-00 115-024847-00
gauges(FRU/0632) FRU(0635)
N2O pipeline pressure N2O pipeline pressure gauge
115-023191-00 115-024848-00
gauges(FRU/0632) FRU(0635)
Air pipeline pressure Air pipeline pressure gauge
115-023192-00 115-024849-00
gauges(FRU/0632) FRU(0635)
O2 tank pressure O2 High Pressure Gauge
115-023194-00 115-024850-00
gauge(FRU/0632) FRU(0635)
N2O tank pressure N2O High Pressure Gauge
115-023195-00 115-024851-00
gauge(FRU/0632) FRU(0635)
Air tank pressure Air High Pressure Gauge
115-023196-00 115-024852-00
gauge(FRU/0632) FRU(0635)

115-016095-00 suction gauge assembly 115-024874-00 suction gauge assembly(0635)

No.: OMZ-21009-A Series Page 315of 428


PMLS Global Technical Support Dept
Affected FRU List:
NO P/N Description Version
O2 pipeline pressure
1 115-023193-00 1.0
gauges(FRU/0632)
N2O pipeline pressure
2 115-02 191-00 1.0
gauges(FRU/0632)
Air pipeline pressure
3 115-023192-00 1.0
gauges(FRU/0632)

4 115-023194-00 O2 tank pressure gauge(FRU/0632) 1.0

N2O tank pressure


5 115-023195-00 1.0
gauge(FRU/0632)

6 115-023196-00 Air tank pressure gauge(FRU/0632) 1.0

7 115-016095-00 suction gauge assembly 1.0

Compatibility Table:
A7(0635) Old Version New Version Compatible
O2 pipeline pressure
115-023193-00 115-024847-00 Yes
gauges(FRU/0632)
N2O pipeline pressure
115-023191-00 115-024848-00 Yes
gauges(FRU/0632)
Air pipeline pressure
115-023192-00 115-024849-00 Yes
gauges(FRU/0632)
O2 tank pressure
115-023194-00 115-024850-00 Yes
gauge(FRU/0632)
N2O tank pressure
115-023195-00 115-024851-00 Yes
gauge(FRU/0632)
Air tank pressure
115-023196-00 115-024852-00 Yes
gauge(FRU/0632)

suction gauge assembly 115-016095-00 115-024874-00 Yes

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 316of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject:ECL092 Technical Change Ref No: TMZ-14024-A7

Version Effective Date Description


1.0 2014-5-14 Document released.

No.: OMZ-21009-A Series Page 317of 428


PMLS Global Technical Support Dept

TMZ-14023-A5

ECL092 Technical Change

Technical News
Subject: ECL092 Technical Change

RECORD IDENTIFIER

Ref No.: TMZ-14023-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Wang Yongxin Parts Change
Date: May 13th, 2014 Appearance
Model:A5 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


To meet RoHS requirement, old version of pipeline pressure gauges and high
pressure gauges will be replaced by new gauges. The P/Ns involved are below:
Old Version New Version
P/N Description P/N Description
O2 Pressure Gauge O2 pipeline pressure gauge
801-0631-00009-00 115-024841-00
Assembly (0~140psi) FRU(0634)
N2O Pressure Gauge N2O pipeline pressure gauge
801-0631-00011-00 115-024842-00
Assembly (0~140psi) FRU(0634)
Air Pressure Gauge Air pipeline pressure gauge
801-0631-00010-00 115-024843-00
Assembly (0~140psi) FRU(0634)
O2 High Pressure
801-0631-00005-00 115-024844-00 O2 High Pressure Gauge FRU(0634)
Gauge Assembly
N2O High Pressure
801-0631-00007-00 115-024845-00 N2O High Pressure Gauge FRU(0634)
Gauge Assembly
Air High Pressure
801-0631-00006-00 115-024846-00 Air High Pressure Gauge FRU(0634)
Gauge Assembly

No.: OMZ-21009-A Series Page 318of 428


PMLS Global Technical Support Dept
Affected FRU List:
NO P/N Description Version

1 801-0631-00009-00 O2 Pressure Gauge Assembly (0~140psi) 2.0

2 801-0631-00011-00 N2O Pressure Gauge Assembly (0~140psi) 2.0

3 801-0631-00010-00 Air Pressure Gauge Assembly (0~140psi) 2.0

4 801-0631-00005-00 O2 High Pressure Gauge Assembly 4.0

5 801-0631-00007-00 N2O High Pressure Gauge Assembly 4.0

6 801-0631-00006-00 Air High Pressure Gauge Assembly 4.0

Compatibility Table:
A5(0634) Old Version New Version Compatible

O2 pipeline pressure gauge FRU(0634) 801-0631-00009-00 115-024841-00 Yes

N2O pipeline pressure gauge FRU(0634) 801-0631-00011-00 115-024842-00 Yes

Air pipeline pressure gauge FRU(0634) 801-0631-00010-00 115-024843-00 Yes

O2 High Pressure Gauge FRU(0634) 801-0631-00005-00 115-024844-00 Yes

N2O High Pressure Gauge FRU(0634) 801-0631-00007-00 115-024845-00 Yes

Air High Pressure Gauge FRU(0634) 801-0631-00006-00 115-024846-00 Yes

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


No.: OMZ-21009-A Series Page 319of 428
PMLS Global Technical Support Dept
Subject:ECL092 Technical Change Ref No: TMZ-14023-A5

Version Effective Date Description


1.0 2014-5-14 Document released.

No.: OMZ-21009-A Series Page 320of 428


PMLS Global Technical Support Dept

TMZ-14022-A7

EFU003 Technical Changes

Technical News
Subject: EFU003 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-14022-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shangwenyan Gong Parts Change
Date: 2014-4-28 Appearance
Model:A7 Stop Production
Comments:EFU003 Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

New Configurations:

1. New configuration for A5 and A7 is launched with only O2 and Air gas inlet.
Also the yoke configuration (with extra gas inlet from a cylinder) is available for A5 and A7.

2. Air-driven A7 is available.

3. Custom V Hub Arm kit is available for cable management.

Custom V Hub Arm kit

No.: OMZ-21009-A Series Page 321of 428


PMLS Global Technical Support Dept
Software Changes:
New Software is launched:
No. Package Modules Latest Version
BundleVersion V02.05.00
Host Software V03.05.00
BIOS V01.12.00
FPGA Sound V1.2
Bundle
1 FPGA Display V1.4
(V02.05.00)
Ventilator Control Module V01.14.00
Ventilator Protect Module V1.1
Language V3.6
Keyboard V01.00/01.01
EFCS
2 EFCS V01.04.00
(V01.04.00)
EFCS_FPGA
3 EFCS_FPAG V3.0
(V3.0)
4 Power(V1.7) Power System V1.7

NOTE:This bundle version is not compatible with old bundle versions.


After upgrade, there will be error message: “Flowmeter Self Test: Fail” when restart, please
do settings as below:
1, Enter the 【Setup】->【Service】->"789789";
2, In standby mode, select 【Setup】->【Service】->【Disable BFCS】, disable the BFCS;
3, In standby mode, select 【Setup】->【General】->【Zero Flow Meters】,do the zeroing
and restart again;

The technical changes include:

4. Side-stream CO2 modules are available on A5 and A7 with AM software version 02.05.00 or above.
After upgrade, please enter【Setup】>【Service】>558188>【Factory Setup】>Select CO2 module
in this menu.

M02C CO2
Module

M02B Module M02C Module

No.: OMZ-21009-A Series Page 322of 428


PMLS Global Technical Support Dept
5. Optimize the color display for different gas. Make the color consistent in flowmeter, AG modules
and trend graphic.
6. When the total flow control mode is in use, the O2 concentration setting softkey will be right below
the O2 flowmeter.
7. The default patient type is set to Adult.
8. P-F loop is added.
9. The allowed fresh gas flow leakage is changed to 0.15L/Min during leakage test. In another word,
during leakage test, all the fresh gas should be set to zero, but actually a 0.15L/Min leakge of fresh
gas in total is allowed.
10. The AG module sampling rate is reset as:
Adult water trap: High 200 ml/min(Factory Default), Medium 180 ml/min, Low 150 ml/min;
Neo water trap: High 120 ml/min, Medium 110 ml/min, Low 100 ml/min;

11. The correlation between trend table, trend graph and event review is added:
In【History】
✓ Customer can choose the event in trend table, and when switch to event log, this event
will be displayed directly.
✓ Customer can choose the event in trend graphic and it will switch to event log directly for
details.

12. Change the FCS voltage in 【Data Monitors】


13. O2 sensor monitoring is cancelled when ventilation mode is dependent ACGO, and all the FiO2
alarms are cancelled.
14. There will be message on screen when BFCS cannot deploy.
15. When supplied by battery only, there will be message after self-test.
16. When the patient is less than 1 year old,the age will be displayed as“<1”
17. To set the parameters, fresh gas flow, alarm limit or CO2 settings, you can press the UP/DOWN
arrows for long time to have a fast setting.

18. AG Alarms are optimized:


✓ During AG zeroing,all the AG physic alamrs will not display.
✓ Eliminate alarms like” accuracy unspecified”.
✓ The“Data Limit Error“will be judged by AG module to avoid AM error reading.
✓ Elimninate the startup and loading messages for internal AG.

19. CO2 Alarms are optimized:


✓ The “Apean“ Alarm will not display when insp/exp wave is detected.
✓ During zeroing, the CO2 physic alarms will not display.

20. Optimize the PCV waveform in Demo mode,and synchronize the settings and reading values.
21. New notification for ACGO and Monitor mode,and disable the Loops in ACGO and Monitor mode.

No.: OMZ-21009-A Series Page 323of 428


PMLS Global Technical Support Dept

22. When External AG module is in use, AA,O2 and N2O waveform can be displayed.
23. Optimizer only indicates for 1-4L/min.
24. Software change to correct the intermittent restart when A5 connects into network.
25. Solve the problem that HL7 protocol in Bundle Version 02.04.00 is not compatible with 01.05.02
26. Optimize the AG zeroing when discharge patient, to correct the intermittent “Internal AG Error 07”
alarm.
27. Optimize software to eliminate the “Bundle Version Time Out ” message when EFCS cannot pass
selftest.
28. Optimize EFCS zeroing.
✓ The range for zero point is changed from[-0.12LPM,0.12LPM] to [-0.20LPM,0.20LPM]
✓ Add zeroing for total flow sensor in EFCS.
29. Optimize the EFCS leakage test in Selftest.
✓ When gas pressure is not enough, there will be no selftest for relevant gas limb in EFCS.
✓ To avoid a zero drift that may cause EFCS leakage test fail:A selftest will first be applied
for EFCS limbs, if one of the limbs has a flow reading out of zero point, then an auto
zeroing will be applied and do the selftest again. This time if it still fails, then it means
there is really a leakage for that limb, and alarms will display.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 324of 428


PMLS Global Technical Support Dept

Technical News Revision


Subject: EFU003 Technical Changes Ref No: TMZ-14022-A7

Version Effective Date Description


1.0 2014-4-29 Document released.

No.: OMZ-21009-A Series Page 325of 428


PMLS Global Technical Support Dept

TMZ-14021-A5

EFU003 Technical Changes

Technical News
Subject: EFU003 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-14021-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shangwenyan Gong Parts Change
Date: 2014-4-27 Appearance
Model:A5 Stop Production
Comments:EFU003 Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

New Configurations:
1. New configuration for A5 and A7 is launched with only O2 and Air gas inlet.
Also the yoke configuration (with extra gas inlet from a cylinder) is available for A5 and A7.

2. Custom V Hub Arm kit is available for cable management.

Custom V Hub Arm kit

Software Changes:
New Software is launched:

No.: OMZ-21009-A Series Page 326of 428


PMLS Global Technical Support Dept
No. Package Modules Latest Version
Bundle Version V02.05.00
Host Software V03.05.00
BIOS V01.12.00
Bundle FPGA Sound V1.2
1
(V02.05.00) FPGA Display V1.4
Ventilator Control Module V01.14.00
Ventilator Protect Module V1.1
Language V3.6
2 Power(V1.7) Power System V1.7

The technical changes include:


3. Side-stream CO2 modules are available on A5 and A7 with AM software version 02.05.00 or above,
After upgrade, please enter【Setup】>【Service】>558188>【Factory Setup】>Select CO2 module
in this menu.

M02C CO2
Module

M02B Module M02C Module

4. Optimize the color display for different gas. Make the color consistent in flowmeter, AG modules
and trend graphic.
5. The default patient type is set to Adult.
6. P-F loop is added.
7. The allowed fresh gas flow leakage is changed to 0.15L/Min during leakage test. In another word,
during leakage test, all the fresh gas should be set to zero, but actually a 0.15L/Min leakge of fresh
gas in total is allowed.
8. The AG module sampling rate is reset as:
Adult water trap: High 200 ml/min(Factory Default), Medium 180 ml/min, Low 150 ml/min;
Neo water trap: High 120 ml/min, Medium 110 ml/min, Low 100 ml/min;

9. The correlation between trend table, trend graph and event review is added:
In【History】
✓ Customer can choose the event in trend table, and when switch to event log, this event

No.: OMZ-21009-A Series Page 327of 428


PMLS Global Technical Support Dept
will be displayed directly.
✓ Customer can choose the event in trend graphic and it will switch to event log directly for
details.

10. O2 sensor monitoring is cancelled when ventilation mode is dependent ACGO, and all the FiO2
alarms are cancelled.

11. When supplied by battery only, there will be message after self-test.

12. When the patient is less than 1 year old,the age will be displayed as“<1”

13. To set the parameters, fresh gas flow, alarm limit or CO2 settings, you can press the UP/DOWN
arrows for long time to have a fast setting.

14. AG Alarms are optimized:


✓ During AG zeroing,all the AG physic alarms will not display.
✓ Eliminate alarms like” accuracy unspecified”.
✓ The “Data Limit Error“ will be judged by AG module to avoid AM error reading.
✓ Eliminate the startup and loading messages for internal AG.

15. CO2 Alarms are optimized:


✓ The “Apean“ Alarm will not display when insp/exp wave is detected.
✓ During zeroing, the CO2 physic alarms will not display.

16. Optimize the PCV waveform in Demo mode,and synchronize the settings and reading values.

17. New notification for ACGO and Monitor mode,and disable the Loops in ACGO and Monitor mode.

18. When External AG module is in use, AA,O2 and N2O waveform can be displayed.
19. Software change to correct the intermittent restart when A5 connects into network.
No.: OMZ-21009-A Series Page 328of 428
PMLS Global Technical Support Dept
20. Solve the problem that HL7 protocol in Bundle Version 02.04.00 is not compatible with 01.05.02
21. Optimize the AG zeroing when discharge patient, to correct the intermittent “Internal AG Error 07”
alarm.

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: EFU003 Technical Changes Ref No: TMZ-14021-A5

Version Effective Date Description


1.0 2014-4-29 Document released.

No.: OMZ-21009-A Series Page 329of 428


PMLS Global Technical Support Dept

TMZ-14020-A7

A7 in German(FR-43000129)Driving Gas Switch

Technical News

Subject: A7 in German(FR-43000129)Driving Gas Switch

RECORD IDENTIFIER

Ref No.: TMZ-14020-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shangwenyan Gong Parts Change
Date: 2014-4-20 Appearance
Model:A7 Stop Production
Comments:N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


This Technical News is ONLY for FR-43000129 to change driving gas from O2 to Air.

1.Spare Parts:
No. PN Description Quantity
1 TCN13-FB009-001 Tube connection(A7/CE/E_ACGO/AIR DRIVE) 1
2 TCN13-FB009-002 A7 Operator's manual(CE) 1

Tube List
Tube No. PN Description Length(mm)
36 082-000517-00 tube.PU (polyether) 5.5mmX8mm orange 100
61 082-000517-00 tube.PU (polyether) 5.5mmX8mm orange 70
63 082-000517-00 tube.PU (polyether) 5.5mmX8mm orange 330

2.Procedures:
No.: OMZ-21009-A Series Page 330of 428
PMLS Global Technical Support Dept
1. Open the service door, remove the rear panel assembly. Remove the work surface cover plate
2. Unplug Tube 36#、61#、63# and disposal them. Keep the Y-connectors.
3. Connect Tube 60# to the Y-connector that links Tube 38# and 69# (Refer to the notifications in red
below).

No.: OMZ-21009-A Series Page 331of 428


PMLS Global Technical Support Dept
4、Refer to TCN13-FB009-001Tube connection(A7/CE/E_ACGO/AIR DRIVE), to reconnect Tube 36#、
61#、63#. Using the kept Y-connectors. (Refer to the notifications in red below).

5、Stick TCN13-FB009-001Tube connection(A7/CE/E_ACGO/AIR DRIVE) to service door, covering


the old diagram.
6、Re-assemble the rear panel assembly. Close the service door.

No.: OMZ-21009-A Series Page 332of 428


PMLS Global Technical Support Dept
3.Post-tests:
1. System Calibration & Constant Flow Test
Tools:VT-Plus,Mindray Calibration Set.
Check the VT-PLUS settings(Note:pre-cautions to check calibration device settings):
1) Gas Settings: Gas Type, set to AIR.
【SETUP】→【Settings】→【ENTER】→【Gas Settings】→【MODIFY】→AIR (FOR AIR
DRIVEN ONLY);
2) Gas Temperature, Ambient Temp and Rel. Humidity Settings:
【SETUP】→【Settings】→【ENTER】→【Gas Settings】→【MODIFY】→【Gas Temperature】
/【Ambient Temp】 →【MODIFY】. Set Gas Temperature and Ambient Temp according to
the hygrometer & thermometer in the room. Input a round number.
3) Zeroing mode settings:
Press the [Setup] button, and select [Setting] > [ENTER]>[Zero Mode] > [Manual]>
[BACK] > [BACK].
4) Serial mode settings:OTIS Ctrl
Press the [Setup] button, and select [Setting] > [System]>[Enter] > [Serial Mode] > [OTIS
Ctrl] > [BACK] > [BACK].
5) Correction Mode Settings:
【SETUP】→【Settings】→【ENTER】→【Correction Mode】→【MODIFY】→ATP

Note:Serial mode settings is for calibration. For measurement (flow or pressure), please exit
serial settings:【SETUP】→【Syestem】→【ENTER】→Seral Mode:NONE
2. Confirm Driving Gas Type
For Air-driven, enter setup→service→558188→Accept. Switch Driver Gas to Air
Tube-Checking:Locate the tube to the expiration valve assembly gas input, check the color and
type: Driving gas is air and the tube color is orange.

2.1 Flow calibration (service)

Preparations:

To ensure the calibration time accuracy, make sure the system time

on AM is consistent with local time, with error allowance 1 min.

NOTE: Driving Gas:For air driven AM, the calibration gas should be
compressed air.

Cautions:Each time before calibration, there must be a VT-plus zeroing without

No.: OMZ-21009-A Series Page 333of 428


PMLS Global Technical Support Dept
any tube connections to VT-plus.

Connect gas supplies and regulate the gas inlet pressure to 400kPa±10kPa;
 Preparations:Enter “standby”,Set Auto/Manual switch to “Auto” position,Remove bellows
from bellows mount and reinstall bellows cover, remove water trap. Turn off the fresh gas. Connect
the inlet port of VT-PLUS to inspiration port, outlet port the expiration port, set to High flow channel
in advance.
 Check Flow Sensor Zero:Enter setup→service→(789789)→Accept→Data Monitors. Ensure
the “Int-Flow Sensor” AD value:100~400,“Insp Flow Sensor” and “Exp Flow Sensor”:200~
1400;
 Check Inspiration Valve DA Value:Enter setup→service→(789789)→Accept→Diagnostic
Tests,Set PEEP Valve D/A to “1700”, Safety Valve To “on”,Set Insp Valve D/A to 2300 and
meantime check VT-plus flow reading should be over 90L/min;
 Connect the calibration device to the anesthesia machine :Connect VT-plus to AM via
calibration connector board (fig below) to RS232 port in AM or the J7 plug on monitor board.

Fig. 1 calibration connector board

 Enter setup→service→(789789)→enter→Calibration→Flow Sensors;Following the instructions


on screen:
◼ Connect the inspiration port and expiration port of the anesthesia machine. Press Begin
Button.
◼ Following the instructions,connect the anesthesia machine and the calibration device in
the same circuit in serial mode. Connect the low-flow channel of the calibration device first.
◼ Following the instructions,Connect the high-flow channel of the calibration on. Make sure
the calibration successful.

No.: OMZ-21009-A Series Page 334of 428


PMLS Global Technical Support Dept

Fig 2 VT-PLUS Ports


3. Tests after calibration

3.1 Check the Date

Enter 【setup】>【General】>【Calibration Flow Sensors】, check the last calibration date.


Make sure the date is correct.

3.2 Constant flow test

NOTE:perform the following operations after flow calibration.

 Keep the tube connection same as flow calibration, just remove the communication cable between
VT-plus and the anesthesia machine.
 Enter standby, Setup→service→(789789)→Accept→Diagnostic Tests,Set PEEP Valve D/A to
“1900”, Safety Valve “on”;
 Set Insp Valve D/A to 0 ,Make sure that the deviation between the measured data of the
inspiratory flow sensor, expiratory flow sensor and ventilator flow sensor and that of the anesthesia
machine calibration device must not exceed 0.5L/min
 Increase Insp Valve Flow or increase Insp Valve D/A , until the VT-plus reading reaches
5±0.5L/min,10±0.5 L/min , 20±0.5L/min , 30±1L/min , 50±1 L/min , 70±2 L/min and 90±2
L/min,
Make sure that the deviation between the measured data of the inspiratory flow sensor, expiratory
flow sensor and ventilator flow sensor and that of the anesthesia machine calibration device must not
exceed 1L/min or 5%of the measured value of the calibration device, whichever is greater.

NOTE:
1)In constant flow test, VT-plus high flow channel is used.
2)Note that AIR must be used for constant flow test. And gas type on VT-plus should be selected
as AIR.
3)In flow test, VT-plus reading and flow sensor readings may be unstable, read the values when
the abbreviation is less than 0.5L/min within 2s.

3.3Total Flow Sensor Zeroing and Pressure and Flow Zeroing

3.3.1Total Flow Sensor Zeroing

Note:Enter Standby mode before zeroing.


Enter【General】>【Zero Flow Meters】,following the instructions on the screen and ensure the zeroing

No.: OMZ-21009-A Series Page 335of 428


PMLS Global Technical Support Dept
successful.
This is done when zero offset exists in the EFCS flowmeter. The EFCS flowmeter still displays a non-
zero flow reading after all fresh gas supplies are turned off.
3.3.2Pressure and Flow Zeroing

Enter Standby,【Service】>【Calibration】>【Zero sensors】


,following the instructions to do the zeroing
and ensure the zeroing successful.

3.3.3 Check Zero Points and Clear Logs

Enter Standby,
【Service】>【Review logs】>【Service】 ,check there is the record: Flowmeter Zero:
Pass; and PAW and Flow Zero: Pass.
Enter 【Alarms】>【Clear Log】, clear the alarm logs.

3.4 System Ventilation Performance Tests


Following the service manual“4.13 System Ventilation Performance Tests” to perform the tests below
and make a record in the appendix.
⚫ Standby Mode Ventilation Test
⚫ Manual Mode Ventilation Test
⚫ VCV Adult Ventilation Mode Test
⚫ VCV Child Ventilation Mode Test
⚫ PCV Adult Ventilation Mode Test
⚫ PS Ventilation Mode Test

Appendix:

错误!链接无效。

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: A7 in German(FR-43000129) Ref No: TMZ-14020-A7
Driving Gas Switch

No.: OMZ-21009-A Series Page 336of 428


PMLS Global Technical Support Dept

Version Effective Date Description


1.0 2014-4-21 Document released.

No.: OMZ-21009-A Series Page 337of 428


PMLS Global Technical Support Dept

TMZ-14019-A7

EFU002 Technical Changes

Technical News
Subject: EFU002 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-14019-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shangwenyan Gong Parts Change
Date: 2014-4-9 Appearance
Model:A7 Stop Production
Comments:EFU002 Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

1. Support arm and cord collector with silica gel piece are available for A series.

PN Description
115-024056-00 cord collector kit

045-001394-00 0634 support arm

Fix-adaptor

Three cord collectors are


installed already

Fig1 Support Arm


No.: OMZ-21009-A Series Page 338of 428
PMLS Global Technical Support Dept

Fig 2 Cord collector

2. A series for pendant are available. THIS IS NOT ABLE TO UPGRADE ON-SITE.

Fig 3 Pendent A series

No.: OMZ-21009-A Series Page 339of 428


PMLS Global Technical Support Dept
3. New T-connectors are available for PENDENT A SERIES.
NOTE: This connector or so-called Reserve Oxygen Supply assembly is ONLY used on PENDENT A
SERIES.

Reserve Oxygen Supply Reserve Oxygen


connector Supply Base

Fig 4 Reserve Oxygen Supply(New T-connector)

New T-connector
Original T-connector(Still Used on WATO&A series)
(ONLY USED ON PENDENT A SERIES)
PN Description Comments PN Description Comments
115-024109- Reserve Oxygen
NIST
oxygen supply T- 00 Supply(0634/NIST)
115-017723-00 NIST
connection package 115-024163- Reserve Oxygen
DISS
00 Supply(0634/DISS)

4. Improve the tightness of vacuum suction knobs. Add icons for vacuum suction.

Vacuum suction icon

Fig 5 Vacuum suction panel


Old Spare parts New Spare Parts
041-007598- 041-007598-
select switch select switch
00 01

No.: OMZ-21009-A Series Page 340of 428


PMLS Global Technical Support Dept
115-018421- 115-024065- Pipeline suction upgrade
pipeline suction upgrade kit(US_BS)
00 00 kit(US_BS/0635)
115-018422- pipeline suction upgrade 115-024066-
pipeline suction upgrade kit(DISS/0635)
00 kit(US_Diss) 00
115-018423- 115-024067- Pipeline suction upgrade
pipeline suction upgrade kit(US_PB)
00 00 kit(US_PB/0635)
115-018424- 115-024068- Pipeline suction upgrade
pipeline suction upgrade kit(US_OH)
00 00 kit(US_OH/0635)
115-018425- 115-024069- Pipeline suction upgrade
pipeline suction upgrade kit(US_CH)
00 00 kit(US_CH/0635)
115-021255- Negative needle Valve knob&panel 115-024234- Needle Valve knob&panel FRU
00 FRU 00 (0635/CE)

5. Korea Power cord 10A 250V 5M(P/N:009-004316-00) is available for A series.

6. Upgrade Manuals.
⚫ Upgrade A7 Manuals: Netherland,French,Portugal,Spanish, Turkey, Russian

7. Order number of maintenance kit has been changed:


To distinguish from FDA region, the order number of maintenance kit has been changed. The
content remains the same.

Before After
Order No. Description Order No. Description
801-0631-00084- Preventative Maintenance
Preventative Maint kits (1 yr) 115-024063-00
00 kits(1 year/Non FDA)
801-0631-00085- Preventative Maintenance
Preventative Maintenance(3 yr) 115-024064-00
00 kits(3 year/Non FDA)

Affected FRU List:

NO. Order No. Description Version


1 115-024063-00 Preventative Maintenance kits(1 year/Non FDA) 1.0
2 115-024064-00 Preventative Maintenance kits(3 year/Non FDA) 1.0
3 115-024109-00 Reserve O2 Supply Connector 1.0
4 115-024163-00 Reserve Oxygen Supply(0634/NIST) 1.0
5 115-024131-00 Reserve Oxygen Supply base part(0634) 1.0
6 115-024164-00 Reserve Oxygen Supply(0634/DISS) 1.0
7 115-023154-00 Reserve O2 Supply Connect(0634/DISS) 1.0
8 041-007598-01 Select switch 1.0
9 115-024065-00 pipeline suction upgrade kit(US_BS/0635) 1.0
10 115-024066-00 pipeline suction upgrade kit(DISS/0635) 1.0

No.: OMZ-21009-A Series Page 341of 428


PMLS Global Technical Support Dept
11 115-024067-00 pipeline suction upgrade kit(US_PB/0635) 1.0
12 115-024068-00 pipeline suction upgrade kit(US_OH/0635) 1.0
13 115-024069-00 pipeline suction upgrade kit(US_CH/0635) 1.0
14 115-024234-00 Needle Valve knob&panel FRU (0635/CE) 1.0

Compatibility Table:

Old Spare parts New Spare Parts Compatibility


041-007598-00 select switch 041-007598-01 select switch Y
Pipeline suction Y
pipeline suction
115-018421-00 115-024065-00 upgrade
upgrade kit(US_BS)
kit(US_BS/0635)
pipeline suction pipeline suction Y
115-018422-00 upgrade 115-024066-00 upgrade
kit(US_Diss) kit(DISS/0635)
Pipeline suction Y
pipeline suction
115-018423-00 115-024067-00 upgrade
upgrade kit(US_PB)
kit(US_PB/0635)
Pipeline suction Y
pipeline suction
115-018424-00 115-024068-00 upgrade
upgrade kit(US_OH)
kit(US_OH/0635)
Pipeline suction Y
pipeline suction
115-018425-00 115-024069-00 upgrade
upgrade kit(US_CH)
kit(US_CH/0635)
Negative needle Needle Valve Y
115-021255-00 Valve knob&panel 115-024234-00 knob&panel FRU
FRU (0635/CE)
Preventative Y
Preventative Maint
801-0631-00084-00 115-024063-00 Maintenance kits(1
kits (1 yr)
year/Non FDA)
Preventative Y
Preventative
801-0631-00085-00 115-024064-00 Maintenance kits(3
Maintenance(3 yr)
year/Non FDA)

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.
No.: OMZ-21009-A Series Page 342of 428
PMLS Global Technical Support Dept

Technical News Revision


Subject: EFU002 Technical Changes Ref No: TMZ-14019-A7

Version Effective Date Description


1.0 2014-4-10 Document released.

No.: OMZ-21009-A Series Page 343of 428


PMLS Global Technical Support Dept

TMZ-14018-A5

EFU002 Technical Changes

Technical News
Subject: EFU002 Technical Changes

RECORD IDENTIFIER

Ref No.: TMZ-14018-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shangwenyan Gong Parts Change
Date: 2014-4-9 Appearance
Model:A5 Stop Production
Comments:EFU002 Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

1. Support arm and cord collector with silica gel piece are available for A series.

PN Description
115-024056-00 cord collector kit

045-001394-00 0634 support arm

Fix-adaptor

Three cord collectors are


installed already
Fig1 Support Arm

No.: OMZ-21009-A Series Page 344of 428


PMLS Global Technical Support Dept

Fig 2 Cord collector

2. A series for pendant are available. THIS IS NOT ABLE TO UPGRADE ON-SITE.

Fig 3 Pendent A series

3. New T-connectors are available for PENDENT A SERIES.


No.: OMZ-21009-A Series Page 345of 428
PMLS Global Technical Support Dept
NOTE: This connector or so-called Reserve Oxygen Supply assembly is ONLY used on PENDENT A
SERIES.

Reserve Oxygen Supply Reserve Oxygen


connector Supply Base

Fig 4 Reserve Oxygen Supply(New T-connector)

New T-connector
Original T-connector(Still Used on WATO&A series)
(ONLY USED ON PENDENT A SERIES)
PN Description Comments PN Description Comments
115-024109- Reserve Oxygen
NIST
oxygen supply T- 00 Supply(0634/NIST)
115-017723-00 NIST
connection package 115-024163- Reserve Oxygen
DISS
00 Supply(0634/DISS)

4. Korea Power cord 10A 250V 5M(P/N:009-004316-00) is available for A series.

5. Upgrade Manuals.
a) New release A5 Manuals, check list and reference card: Bulgaria, Denmark, Finland,
Romania
b) Upgrade A5 Manuals: Netherland,French, Portugal ,Spanish, Turkey, Russian, Czech,
German, Italy, Norway, Poland, Serbia, Swedish, Hungary.

6. Order number of maintenance kit has been changed:


To distinguish from FDA region, the order number of maintenance kit has been changed. The
content remains the same.

Before After
Order No. Description Order No. Description

No.: OMZ-21009-A Series Page 346of 428


PMLS Global Technical Support Dept
Preventative Maintenance
801-0631-00084-00 Preventative Maint kits (1 yr) 115-024063-00
kits(1 year/Non FDA)
Preventative Maintenance(3 Preventative Maintenance
801-0631-00085-00 115-024064-00
yr) kits(3 year/Non FDA)

Affected FRU List:

NO. Order No. Description Version


1 115-024063-00 Preventative Maintenance kits(1 year/Non FDA) 1.0
2 115-024064-00 Preventative Maintenance kits(3 year/Non FDA) 1.0
3 115-024109-00 Reserve O2 Supply Connector 1.0
4 115-024163-00 Reserve Oxygen Supply(0634/NIST) 1.0
5 115-024131-00 Reserve Oxygen Supply base part(0634) 1.0
6 115-024164-00 Reserve Oxygen Supply(0634/DISS) 1.0
115-023154-00 Reserve O2 Supply Connect(0634/DISS) 1.0

Compatibility Table:

Old Spare parts New Spare Parts Compatibility


Preventative Maintenance 115-024063- Preventative Maintenance Y
801-0631-00084-00
kits (1 yr) 00 kits(1 year/Non FDA)
Preventative 115-024064- Preventative Maintenance Y
801-0631-00085-00
Maintenance(3 yr) 00 kits(3 year/Non FDA)

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

Technical News Revision


Subject: EFU002 Technical Changes Ref No: TMZ-14018-A5

No.: OMZ-21009-A Series Page 347of 428


PMLS Global Technical Support Dept

Version Effective Date Description


1.0 2014-4-10 Document released.

No.: OMZ-21009-A Series Page 348of 428


PMLS Global Technical Support Dept

TMZ-14002-A7

ECL089 Technical changes

Technical News
Subject: ECL089 Technical changes

RECORD IDENTIFIER

Ref No.: TMZ-14002-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shangwenyan Gong Parts Change
Date: 2014-1-9 Appearance
Model:A7 Stop Production
Comments: Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

Mechanical Changes:
1. Add hook onto the O2 sensor protective bar

No.: OMZ-21009-A Series Page 349of 428


PMLS Global Technical Support Dept

2. New N2O needle valve is used.


New N2O needle valve(115-022926-00) for BFCS : enlarge the diameter of N2O inlet base,
distinguish from O2 and Air.

3. O2 sensor plug change.

Before After

4. New assembly of pressure gauge with adhesive tape.

If you need to replace the pressure gauge cover or front panel.


No.: OMZ-21009-A Series Page 350of 428
PMLS Global Technical Support Dept
Stick the tape to the pressure gauge cover

And then stick the cover to the front panel.

5. Main board material changes to comply with the requirement of RoHS. The order number keeps the
same.

Software:
FDA region ONLY:
Upgrade software to solve the problems:
1. The HL7 protocol in Bundle version 02.04.00(FDA) is not compatible with the HL7 protocol in Bundle
version 01.05.02.
No. Package Content Latest Version
BundleVersion V02.04.01
Host Software V03.04.01
Boot Software V01.11.00
Bundle
1 FPGA V1.4
(V02.04.01)
VCM V01.14.00
VPM V1.1
E flowmeter V1.8
2 Power(V1.7) Power board V1.7

Maintenance kit:
One-year maintenance kit:Add three seals for cylinder yoke
Three-year maintenance kit:Add three seals for cylinder yoke. Add one more O-ring 082-
000673-00 (can be used on bag arm or O2 sensor cover). Add O-ring 082-001504-00 for
water trap.

Others:
1. The lock is replaced with new material to comply with RoHS.( 115-023320-00 )

Affected FRU list:


NO. Order No. Description Version

No.: OMZ-21009-A Series Page 351of 428


PMLS Global Technical Support Dept

1 115-021249-00 Mid-plate FRU(0631/no ventilation valve) 1.0


2 801-0631-00103-00 Breathing Circuit Unit 3.0
3 115-021254-00 Breathing Circuit Cell FRU(0632) 1.0
4 115-014468-00 needle valve assembly(0632) 3.0
module of backup needle
5 115-015775-00 1.0
valve(0632/SFDA)
6 801-0631-00077-00 Drawer Security Key, A-series 2.0
N2O tank pressure gauge
7 115-015450-00 1.0
assembly(0632)
8 115-015449-00 O2 tank pressure gauge assembly(0632) 1.0

9 115-015451-00 Air tank pressure gauge assembly(0632) 1.0

10 115-015448-00 Air pipeline pressure gauges(0632) 1.0

11 115-015446-00 O2 pipeline pressure gauges(0632) 1.0

12 115-015447-00 N2O pipeline pressure gauges(0632) 1.0

13 115-018146-00 Main control board FRU 1.0


14 801-0631-00084-00 Preventative Maintenance kits (1 year) 3.0
15 801-0631-00085-00 Preventative Maintenance kits (3 years) 3.0

Compatibility Table:

A7 (0635) Before After Compatibility

Mid-plate FRU(0631/no ventilation valve) 115-021249-00 115-021249-00 Yes

Breathing Circuit Unit 801-0631-00103-00 801-0631-00103-00 Yes

Breathing Circuit Cell FRU(0632) 115-021254-00 115-021254-00 Yes

N2O needle valve(0632) 115-014468-00 115-022926-00 Yes


module of backup needle
115-015775-00 115-015775-00 Yes
valve(0632/SFDA)

Drawer Security Key, A-series 801-0631-00077-00 115-023320-00 Yes

N2O tank pressure gauge


115-015450-00 115-015450-00 Yes
assembly(0632)

O2 tank pressure gauge assembly(0632) 115-015449-00 115-015449-00 Yes

Air tank pressure gauge assembly(0632) 115-015451-00 115-015451-00 Yes

Air pipeline pressure gauges(0632) 115-015448-00 115-015448-00 Yes

O2 pipeline pressure gauges(0632) 115-015446-00 115-015446-00 Yes

N2O pipeline pressure gauges(0632) 115-015447-00 115-015447-00 Yes

No.: OMZ-21009-A Series Page 352of 428


PMLS Global Technical Support Dept

Main control board FRU 115-018146-00 115-018146-00 Yes

Preventative Maintenance kits (1 year) 801-0631-00084-00 801-0631-00084-00 Yes

Preventative Maintenance kits (3 years) 801-0631-00085-00 801-0631-00085-00 Yes

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 353of 428


PMLS Global Technical Support Dept

TMZ-14001-A5

ECL089 Technical changes

Technical News
Subject: ECL089 Technical changes

RECORD IDENTIFIER

Ref No.: TMZ-14001-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shangwenyan Gong Parts Change
Date: 2014-1-9 Appearance
Model:A5 Stop Production
Comments: Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

Mechanical Changes:
1. Add hook onto the O2 sensor protective bar

No.: OMZ-21009-A Series Page 354of 428


PMLS Global Technical Support Dept

2. O2 sensor plug change.

Before After

3. New assembly of pressure gauge with adhesive tape.

If you need to replace the pressure gauge cover or front panel.


Stick the tape to the pressure gauge cover

And then stick the cover to the front panel.

4. Main board material changes to comply with the requirement of RoHS. The order number keeps the
No.: OMZ-21009-A Series Page 355of 428
PMLS Global Technical Support Dept
same.

Software:
FDA region ONLY:
Upgrade software to solve the problems:
1. The HL7 protocol in Bundle version 02.04.00(FDA) is not compatible with the HL7 protocol in Bundle
version 01.05.02.
2. A5(FDA) unit with module rack can not trigger the alarm of “O2 Sensor Disconnected”
No. Package Content Latest Version
BundleVersion V02.04.01
Host Software V03.04.01
Boot Software V01.11.00
Bundle
1 FPGA V1.4
(V02.04.01)
VCM V01.14.00
VPM V1.1
E flowmeter V1.8
2 Power(V1.7) Power board V1.7

Maintenance kit:
One-year maintenance kit:Add three seals for cylinder yoke
Three-year maintenance kit:Add three seals for cylinder yoke. Add one more O-ring 082-
000673-00 (can be used on bag arm or O2 sensor cover). Add O-ring 082-001504-00 for
water trap.

Others:
1. The lock is replaced with new material to comply with RoHS.( 115-023320-00 )

Affected FRU List:


NO. Order No. Description Version
1 115-021249-00 Mid-plate FRU(0631/no ventilation valve) 1.0
2 801-0631-00105-00 Breathing System 3.0
3 115-017036-00 Breathing System FRU(0631) 1.0
4 801-0631-00077-00 Drawer Security Key, A-series 2.0
5 801-0631-00006-00 Air High Pressure Gauge Assembly 3.0
6 801-0631-00005-00 O2 High Pressure Gauge Assembly 3.0
7 801-0631-00007-00 N2O High Pressure Gauge Assembly 3.0
8 801-0631-00009-00 O2 Pressure Gauge Assembly (0~140psi) 1.0
9 801-0631-00010-00 Air Pressure Gauge Assembly (0~140psi) 1.0
10 801-0631-00011-00 N2O Pressure Gauge Assembly (0~140psi) 1.0
11 115-018146-00 Main control board FRU 1.0
12 801-0631-00084-00 Preventative Maintenance kits (1 year) 3.0

No.: OMZ-21009-A Series Page 356of 428


PMLS Global Technical Support Dept
13 801-0631-00085-00 Preventative Maintenance kits (3 years) 3.0

Compatibility Table:
A5 (0634) Before After Compatibility
Mid-plate FRU(0631/no ventilation
115-021249-00 115-021249-00 Yes
valve)

Breathing System 801-0631-00105-00 801-0631-00105-00 Yes

Breathing System FRU(0631) 115-017036-00 115-017036-00 Yes


Drawer Security Key, A-series 801-0631-00077-00 115-023320-00 Yes

Air High Pressure Gauge Assembly 801-0631-00006-00 801-0631-00006-00 Yes

O2 High Pressure Gauge Assembly 801-0631-00005-00 801-0631-00005-00 Yes

N2O High Pressure Gauge Assembly 801-0631-00007-00 801-0631-00007-00 Yes

O2 Pressure Gauge Assembly


801-0631-00009-00 801-0631-00009-00 Yes
(0~140psi)
Air Pressure Gauge Assembly
801-0631-00010-00 801-0631-00010-00 Yes
(0~140psi)
N2O Pressure Gauge Assembly
801-0631-00011-00 801-0631-00011-00 Yes
(0~140psi)
Main control board FRU 115-018146-00 115-018146-00 Yes

Preventative Maintenance kits (1 year) 801-0631-00084-00 801-0631-00084-00 Yes

Preventative Maintenance kits (3 years) 801-0631-00085-00 801-0631-00085-00 Yes

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 357of 428


PMLS Global Technical Support Dept

TMZ-13013-A7

A series Factory Code Change

Technical News
Subject: A series Factory Code Change

RECORD IDENTIFIER

Ref No.:TMZ-13013-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shangwenyan Gong Parts Change
Date: 2013-11-11 Appearance
Model:A7 Stop Production
Comments:ECL081, EFU001 Others

CONTENT

No.: OMZ-21009-A Series Page 358of 428


PMLS Global Technical Support Dept

To Whom It May Concern,

Below content is only for your information.


According to the strategy, Factory Code of A5 and A7 in CE region (except Canada) will be changed
to get distinguished from Mahwah.
Anesthesia A7 A5
Product Before EFU001 After EFU001 Before ECL081 After ECL081
International (CE) 0632 0635 0631 0634
Canada (CE) 0632 0632 0631 0631
1)The change time will be 2013.12.25
2)From 2014.1.1, the orders from International(CE, except Canada)will be available for
0634B and 0635B. 0631B and 0632B will be disabled. i, e. Factory code of A7 will be changed to
0635, factory code of A5 will be changed to 0634 (except Canada).
3)From 2014.1.1,0631B and 0632B are only available for Canada, i, e, Factory code of A5
and A7 remain the same for Canada Region.
4) The factory code on SAP will be changed automatically, when you create a note for service,
there is no need to make manual change for factory code.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 359of 428


PMLS Global Technical Support Dept

TMZ-13012-A5

A series Factory Code Change

Technical News
Subject: A series Factory Code Change

RECORD IDENTIFIER

Ref No.:TMZ-13012-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author:Shangwenyan Gong Parts Change
Date: 2013-11-11 Appearance
Model:A5 Stop Production
Comments:ECL081, EFU001 Others

CONTENT

No.: OMZ-21009-A Series Page 360of 428


PMLS Global Technical Support Dept

To Whom It May Concern,


Below content is only for your information.

According to the strategy, Factory Code of A5 and A7 in CE region (except Canada) will be changed
to get distinguished from Mahwah.
Anesthesia A7 A5
Product Before EFU001 After EFU001 Before ECL081 After ECL081
International (CE) 0632 0635 0631 0634
Canada (CE) 0632 0632 0631 0631
1)The change time will be 2013.12.25
2)From 2014.1.1, the orders from International(CE, except Canada)will be available for
0634B and 0635B. 0631B and 0632B will be disabled. i, e. Factory code of A7 will be changed to
0635, factory code of A5 will be changed to 0634 (except Canada).
3)From 2014.1.1,0631B and 0632B are only available for Canada, i, e, Factory code of A5
and A7 remain the same for Canada Region.
4)The factory code on SAP will be changed automatically, when you create a note for service,
there is no need to make manual change for factory code.
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 361of 428


PMLS Global Technical Support Dept

TMZ-13011-A7

Software, pneumatic circuit, hardware and manuals upgrade for A series

Technical News
Subject: Software, pneumatic circuit, hardware and manuals upgrade for A series.

RECORD IDENTIFIER

Ref No.: TMZ-13011-A7 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shangwenyan Gong Parts Change
Date: 2013-10-11 Appearance
Model:A7 Stop Production
Comments:EDQ034,EDQ033,ECL085 Others

CONTENT

To Whom It May Concern,

Below content is only for your information.


Software, pneumatic circuit, hardware, manuals, and some other changes on A series.

1. Software

1.1 Host Software Change

⚫ New software package of A7, please note the all the software in the package
should be upgraded.
A7:
BundleVersion: 02.04.00
Power:V1.7
EFCS:V01.03.00
FpgaEFCS:3.0
SystemUpdateTool:4.9

⚫ The 02.04.00 bundle version supports all the old HL7 protocols. In [Setup->
System->Set HL7 Compatibility]

No.: OMZ-21009-A Series Page 362of 428


PMLS Global Technical Support Dept

Here is the list of Bundle version of software and related HL7 protocol version.
Bundle Version HL7 Protocol Version
01.00.00
01.00.01
01.01.00
01.02.00
01.02.01 01.00.00~01.05.00
01.03.00
01.04.00
01.05.00
01.05.01
01.05.02 01.05.02
02.00.00 02.00.00
02.01.00
02.02.00
02.02.01
Most Recent
02.02.02
02.03.00
02.04.00

⚫ New languages in software of CE region: Bulgaria, Finland, Denmark, Romania


are released.
⚫ The indicator of Optimizer is automatically set to lay it in the middle of the scale.

⚫ Buzzer will buzz for 10 seconds once the main board has an abnormal reset.

⚫ For FDA standard, A series machines with AG module plus O2 function, the

alarms of O2 Sensor will be prohibited.


⚫ Internal AG warm up message is hidden.

⚫ When the alarm is silenced, if the current alarm is low class, then the alarm will

only reoccur when there is a new one; if the alarm is middle or high level, then silence
for 120s.
⚫ Apnea CO2 and Apnea Alarm are set to be independent from each other.

⚫ Optimize the AG module reading of CO2 measurement, improve the accuracy.

⚫ In [Service->Data Monitors->Component] menu, there newly added FCS option


to display EFCS and BFCS status.

No.: OMZ-21009-A Series Page 363of 428


PMLS Global Technical Support Dept

⚫ In [Service->Data Monitors->Main Board] there newly added module rack fan


speed. (Fan Speed 02)

⚫ In Service Menu, there newly added FCS Test, including Manual test, Auto Test,
BFCS Test, Solenoid Actuator Test.

No.: OMZ-21009-A Series Page 364of 428


PMLS Global Technical Support Dept

Manual Test: When 3-way valve of CPU and FPGA of EFCS monitor board are both
set to H, then it is in EFCS status, otherwise it will be BFCS status.

Manual Test: In EFCS status, you can test the accuracy of each limb (O2, N2O, Air)
by comparing to the total flow reading.

Auto Test: Automatically test the three-way valves and accuracy of flow sensor.

No.: OMZ-21009-A Series Page 365of 428


PMLS Global Technical Support Dept

BFCS Test: Manual test for BFCS. Click “Deploy BFCS” to deploy it, and “Retract
BFCS” to retract it. Four position switch status also displays.

Solenoid Actuator Test: Click “Test Solenoid Actuator” button, the solenoid “ON/OFF”
status and relative AD count and Actual value will be checked if they are within range.

No.: OMZ-21009-A Series Page 366of 428


PMLS Global Technical Support Dept

⚫ In [System->Configuration Info], there newly added “Module Rack” and “AG


Version Limit” Status.

⚫ In [Service-> Factory Setup (558188)], there newly added AG module and BIS
module selection.

⚫ Solve the problem of Wrong User Calibration Time. If a wrong user calibration
time is identified on A7, please upgrade to the bundle 02.04.00, and do user
calibration again.

No.: OMZ-21009-A Series Page 367of 428


PMLS Global Technical Support Dept

⚫ Demo mode supports the external AG and EFCS demonstration.

⚫ When the battery is low, the BFCS will be deployed automatically. And can only
be enabled after the AC power is connected.

⚫ When Serial Protocol is set to “None”, the SNTP Protocol (Interval) will be “Off”.

No.: OMZ-21009-A Series Page 368of 428


PMLS Global Technical Support Dept

⚫ In the menus, there newly added “To Top” and “To Bottom” icons.

⚫ BIS function is available in A series in CE and SFDA region.

⚫ In [History->Graphic Trends] are added.

No.: OMZ-21009-A Series Page 369of 428


PMLS Global Technical Support Dept

⚫ CO2 Alarm icon is automatically added when AG module is recognized. In Manual


mode, the CO2 alarm can be set to OFF.

When AG module is plugged in:

When AG module is taken off:

1.2 Ventilator Module Software Change


⚫ The calibration characteristic curve in ventilator module software is changed due

to the new ventilator module is in use. (Please refer to 2. Pneumatic Circuit)


⚫ In manual mode, the refreshing time of parameters are set the same as Apnea

alarm time when there is no ventilation.


No.: OMZ-21009-A Series Page 370of 428
PMLS Global Technical Support Dept
⚫ The O2 sensor calibration is allowed the lowest atmospheric pressure to be
50Kpa.
(3600-4000 Altitude)

1.3 iChart-OR software has been upgraded.


⚫ There newly added two language in the iChart-OR system: Spanish and
Portuguese
⚫ There newly added Preoperative interview and Preoperative management
options.
Improve the fluid calculation: when the fluid unit is set to L instead of ML, the fluid
value will be corrected automatically.

2. Pneumatic Circuit
⚫ There newly added the indication of flow direction and maximum pressure in

ACGO port.
⚫ A7 in CE region has launched air-driven venturi vacuum suction.

⚫ A7 in CE region has launched pipeline vacuum suction.

⚫ There newly added Ohmeda, P-B and Chemetron US standard pipeline:

⚫ The venture flowmeter in ventilator module has been changed to another shape.

Old
flowmeter

New
flowmeter

⚫ Vaporizer manifold for A5 has been changed to A7, and the sampling of AG
calculation is in ACGO assembly.

No.: OMZ-21009-A Series Page 371of 428


PMLS Global Technical Support Dept

⚫ The O2 gas inlet assembly of A5 changes to the same as A7. In other words, A5
and A7 share the same O2 gas inlet assembly. For Air and N2O inlet assembly, A5
share the same DISS connector as A7.

3. Hardware

⚫ A series occasionally has black screen after fast switch off then switch on. The
inverter board was improved to prevent this issue.

⚫ Auxiliary flowmeter backup light change.

Before:Low brightness,Square Shape After:High brightness,Fan Shape

⚫ The flow sensors in A series have been upgraded. In A7: three flow sensors of
EFCS and the flow sensor in ventilator have been upgraded.

O2, Air and N2O flow sensors of EFCS have been upgraded to improve the accuracy
from 5% to 3%.

No.: OMZ-21009-A Series Page 372of 428


PMLS Global Technical Support Dept

Flow sensors for venturi flowmeter has been upgraded to improve Electro Magnetic
Compatibility

4. Others
⚫ A new hook that can be freely installed in the GCX rail.

⚫ For FDA region, a GCX rail with the gap in the middle.

⚫ A new support pole is used to backup A7 front panel

No.: OMZ-21009-A Series Page 373of 428


PMLS Global Technical Support Dept

BEFORE AFTER

⚫ A7 manuals and checklist in new languages are launched.


A7: Dutch, French, Portugal, Spanish,Turkish,Russian.

⚫ Li-ion battery package has been changed in CE and SFDA region.


Due to the new delivery requirement of international flight, the Li-ion batteries are
packaged as accessories and need to get installed by engineer when unpack the
machine.

⚫ There newly added the tubes indication in service door spare part for replacement.

Below is the list of the affected P/Ns:

Item New
P/N Descriptions Old Ver.
No. Ver.
Manual
1 046-004559-00 A7 Operator's Manual(CE) 4.0 5.0
2 046-005034-00 A7 Operator's Manual(Dutch) / 1.0
3 046-005035-00 A7 Operator's Manual(French) / 1.0
4 046-005037-00 A7 Operator's Manual(Portuguese) / 1.0
5 046-005038-00 A7 Operator's Manual(Spanish) / 1.0
6 046-005039-00 A7 Operator's Manual(Turkish) / 1.0
7 046-005040-00 A7 Operator's Manual(Russian) / 1.0
8 046-005049-00 A7 checkout list(Dutch) / 1.0
9 046-005050-00 A7 checkout list(French) / 1.0
10 046-005051-00 A7 checkout list(Portuguese) / 1.0
11 046-005052-00 A7 checkout list(Spanish) / 1.0
12 046-005053-00 A7 checkout list(Turkish) / 1.0
13 046-005054-00 A7 checkout list(Russian) / 1.0

No.: OMZ-21009-A Series Page 374of 428


PMLS Global Technical Support Dept
14 046-005063-00 A7 Reference Card(Dutch) / 1.0
15 046-005064-00 A7 Reference Card(French) / 1.0
16 046-005065-00 A7 Reference Card(Portuguese) / 1.0
17 046-005066-00 A7 Reference Card(Spanish) / 1.0
18 046-005067-00 A7 Reference Card(Turkish) / 1.0
19 046-005068-00 A7 Reference Card(Russian) / 1.0

Affected FRU list:


Item
FRU P/N Descriptions Ver.
No.
1 115-020937-00 Drive Assembly(New Diff_pressure)FRU 1.0
2 115-018148-01 Ventilation kit(New Diff_pressure)FRU 1.0
3 801-0631-00088-00 Main body of Drive Gas Assenbly 1.0
4 801-0631-00002-00 Flow Meter Lighting Board PCBA 2.0
5 115-018165-00 Auxiliary Gas Outlet Assembly FRU(US) 2.0
6 115-018166-00 Auxiliary Gas Outlet Assembly FRU(Euro) 2.0
7 115-018167-00 Auxiliary Gas Outlet Assembly FRU(SFDA) 2.0
8 115-014827-00 oxygen origin inlet module(0632/DISS) 1.0
9 115-015435-00 O2 supply inlet assembly(0632/NIST) 1.0
10 801-0631-00089-00 sensor interface board PCBA 2.0
11 115-017879-00 anesthetic gas inlet tubes module(0632) 1.0
12 801-0631-00024-00 Vaporizer Mounting Manifold 1.0
13 051-001519-00 0631 Backlight Inverter PCBA 1.0
14 051-001286-01 0632 BFCS Position Switch Board PCBA 1.0
15 115-014519-01 Backup valves in-out device(0632) 1.0
16 051-001414-01 0632 Flowmeter O2 flow sensor board PCBA 1.0
17 051-001416-01 0632 Flowmeter N2O sensor board PCBA 1.0
18 051-001415-01 0632 Flowmeter Air sensor board PCBA 1.0
19 115-018146-00 Main control board FRU 1.0
20 115-021094-00 Service door maintenance kit(FRU) 1.0
21 041-007598-00 select switch 1.0
22 115-015247-00 Suction needle valve assembly 1.0
23 115-016095-00 Suction gauge assembly 1.0
24 115-017748-00 Vacuum generator unit 1.0
25 115-017947-00 Suction bottle bracket assembly 1.0
26 115-018132-00 Vacuum pipeline supply assembly(NIST) 1.0
27 115-018133-00 Vacuum pipeline supply assembly(DISS) 1.0
28 115-018356-00 Muffler Asembly 1.0
29 115-021249-00 Mid-plate FRU(0631/no ventilation valve) 1.0
30 115-021251-00 Lower cover FRU(0632/no flowsenser) 1.0
31 115-021254-00 Breathing Circuit Cell FRU(0632) 1.0
32 115-021255-00 Negative needle Valve knob&panel FRU 1.0

No.: OMZ-21009-A Series Page 375of 428


PMLS Global Technical Support Dept
33 115-021256-00 negative control main body FRU 1.0
34 040-001532-00 Vacuum flask 1.0
35 040-001533-00 Vacuum flask(no protection ) 1.0
36 082-001341-00 Hose Assembly.Vacuum BS 1.0
37 115-018420-00 Pipeline suction upgrade kit(Br_BS) 1.0
38 115-021297-00 Suction upgrade kit(pipeline to Venturi) 1.0
39 115-022320-00 loop low cover 2 assembly FRU 1.0
40 115-021252-00 Upper cover 2 FRU(0631/no popoff) 1.0
41 115-021250-00 Lower cover FRU(0631/no flowsenser) 1.0

Some changing units are not compatible with the old ones. Below is coordination table
for your reference.

New
A7 (0632) Before Now Compatible
With Old

gas bench receptacle assembly 115-012781-00 115-016480-00 Yes

Upper cover 2 FRU(0631/no popoff) 043-001133-00 115-021252-00 Yes

loop low cover 2 assembly FRU 043-001135-00 115-022320-00 Yes

Lower cover FRU(0631/no flowsenser) 115-009046-00 115-021250-00 Yes

801-0631-00047-
Drive Assembly(New Diff_pressure)FRU 115-020937-00 Yes
00

Ventilation kit(New Diff_pressure)FRU 115-018148-00 115-018148-01 Yes

0631 Backlight Inverter PCBA 051-000176-02 051-001519-00 Yes

0632 BFCS Position Switch Board PCBA 051-001286-00 051-001286-01 Yes

backup valves in-out device(0632) 115-014519-00 115-014519-01 Yes

0632 Flowmeter O2 flow sensor board PCBA 051-001414-00 051-001414-01 Yes

No.: OMZ-21009-A Series Page 376of 428


PMLS Global Technical Support Dept

0632 Flowmeter N2O sensor board PCBA 051-001416-00 051-001416-01 Yes

0632 Flowmeter Air sensor board PCBA 051-001415-00 051-001415-01 Yes

service door maintenance kit(FRU) 115-017948-00 115-021094-00 Yes

801-0631-00137-
Main control board FRU 115-018146-00 Yes
00

Li-ion Battery Pack 022-000008-00 115-018012-00 Yes

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

TMZ-13010-A5

Software, pneumatic circuit, hardware and manuals upgrade for A series

Technical News
Subject: Software, pneumatic circuit, hardware and manuals upgrade for A series.

RECORD IDENTIFIER

Ref No.: TMZ-13010-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Shangwenyan Gong Parts Change
Date: 2013-10-11 Appearance
Model:A5 Stop Production
Comments:EDQ034,EDQ033,ECL085 Others

CONTENT
No.: OMZ-21009-A Series Page 377of 428
PMLS Global Technical Support Dept

To Whom It May Concern,

Below content is only for your information.


Software, pneumatic circuit, hardware, manuals, and some other changes on A series.

1. Software

1.1 Host Software Change

New software package of A5, and please note that during upgrade, you have to
upgrade all the software in the package.

A5:
BundleVersion:02.04.00
Power:V1.7
SystemUpdateTool:4.9

⚫ The 02.04.00 bundle version supports all the old HL7 protocols. In [Setup->
System->Set HL7 Compatibility]

Here is the list of Bundle version of software and related HL7 protocol version.
Bundle Version HL7 Protocol Version
01.00.00
01.00.01
01.01.00
01.02.00
01.02.01 01.00.00~01.05.00
01.03.00
01.04.00
01.05.00
01.05.01
01.05.02 01.05.02
02.00.00 02.00.00
02.01.00
Most Recent
02.02.00

No.: OMZ-21009-A Series Page 378of 428


PMLS Global Technical Support Dept
02.02.01
02.02.02
02.03.00
02.04.00

⚫ New languages in software of CE region: Bulgaria, Finland, Denmark, Romania


are released.
⚫ Buzzer will buzz for 10 seconds once the main board has an abnormal reset.

⚫ For FDA standard, A series machines with AG module plus O2 function, the

alarms of O2 Sensor will be prohibited.

⚫ Internal AG warm up message is hidden.


⚫ When the alarm is silenced, if the current alarm is low class, then the alarm will

only reoccur when there is a new one; if the alarm is middle or high level, then silence
for 120s.
⚫ Apnea CO2 and Apnea Alarm are set to be independent from each other.

⚫ Optimize the AG module reading of CO2 measurement, improve the accuracy.

a) In [Service->Data Monitors->Main Board] there newly added module rack fan


speed. (Fan Speed 02)

b) In [System->Configuration Info], there newly added “Module Rack” and “AG


Version Limit” Status.

No.: OMZ-21009-A Series Page 379of 428


PMLS Global Technical Support Dept

c) In [Service-> Factory Setup (558188)], there newly added AG module and BIS
module selection.

d) Solve the problem of Wrong User Calibration Time. If a wrong user calibration
time is identified, please upgrade to the bundle 02.04.00, and do user calibration
again.

No.: OMZ-21009-A Series Page 380of 428


PMLS Global Technical Support Dept
e) When Serial Protocol is set to “None”, the SNTP Protocol (Interval) will be “Off”.

f) In the menus, there newly added “To Top” and “To Bottom” icons.

g) BIS function is available in A series in CE and SFDA region.

h) In [History->Graphic Trends] are added.

No.: OMZ-21009-A Series Page 381of 428


PMLS Global Technical Support Dept

i) CO2 Alarm icon is automatically added when AG module is recognized. In Manual


mode, the CO2 alarm can be set to OFF.

When AG module is plugged in:

When AG module is taken off:

1.2 Ventilator Module Software Change


⚫ The calibration characteristic curve in ventilator module software is changed due
to the new ventilator module is in use. (Please refer to 2. Pneumatic Circuit)
⚫ In manual mode, the refreshing time of parameters are set the same as Apnea

No.: OMZ-21009-A Series Page 382of 428


PMLS Global Technical Support Dept
alarm time when there is no ventilation.
⚫ The O2 sensor calibration is allowed the lowest atmospheric pressure to be

50Kpa. (3600-4000 Altitude)

1.3 iChart-OR software has been upgraded.


⚫ There newly added two language in the iChart-OR system: Spanish and
Portuguese
⚫ There newly added Preoperative interview and Preoperative management
options.
⚫ Improve the fluid calculation: when the fluid unit is set to L instead of ML, the fluid
value will be corrected automatically.

2. Pneumatic Circuit
⚫ There newly added the indication of flow direction and maximum pressure in

ACGO port.
⚫ There newly added Ohmeda, P-B and Chemetron US standard pipeline:

⚫ The venture flowmeter in ventilator module has been changed to another shape.

Old
flowmeter

New
flowmeter

⚫ Vaporizer manifold for A5 has been changed to A7, and the sampling of AG
calculation is in ACGO assembly.

For replacement of old manifold, sampling tube or old ACGO assembly, please refer
to :
No.: OMZ-21009-A Series Page 383of 428
PMLS Global Technical Support Dept

Replacement
Old Part
New Spare part Order Number
Vaporizer Mounting Manifold 801-0631-00024-00
vaporizer mounting
anesthetic gas inlet tubes
manifold(AGC)(115- 115-017879-00
module(0632)
015456-00)
mechanical ACGO(AG) 115-017878-00
Anesthetic gas inlet tubes anesthetic gas inlet tubes
115-017879-00
module(115-015829-00) module(0632)

⚫ The O2 gas inlet assembly of A5 changes to the same as A7. In other words, A5
and A7 share the same O2 gas inlet assembly. For Air and N2O inlet assembly, A5
share the same DISS connector as A7.

3. Hardware

⚫ A series occasionally has black screen after fast switch off then switch on. The
inverter board was improved to prevent this issue.

⚫ Auxiliary flowmeter backup light change.

No.: OMZ-21009-A Series Page 384of 428


PMLS Global Technical Support Dept

Before:Low brightness,Square Shape After:High brightness,Fan Shape

⚫ The flow sensors in A series have been upgraded.


In A5, the O2, Air and N2O flow sensors of fresh gas have been upgraded to improve
the accuracy from 5% to 3%.

Flow sensors for venturi flowmeter has been upgraded to improve Electro Magnetic
Compatibility

No.: OMZ-21009-A Series Page 385of 428


PMLS Global Technical Support Dept

4. Others
⚫ A new hook that can be freely installed in the GCX rail.

⚫ For FDA region, a GCX rail with the gap in the middle.

⚫ A5 manuals and checklist in new languages are launched.


A5: Zech,German,Italian,Norwegian,Poland,Serbia,Swedish ,Hungary.

⚫ Li-ion battery package has been changed in CE and SFDA region.


Due to the new delivery requirement of international flight, the Li-ion batteries are
packaged as accessories and need to get installed by engineer when unpack the
machine.

⚫ Two external suction for A5:


Outer suction Kit (Air drive,NIST)
Outer suction Kit (O2 drive,NIST)

⚫ For CE region, A5 has changed the color of screen shell, base assembly, and the
wheel to the same as FDA and SFDA.

⚫ There newly added the tubes indication in service door spare part for replacement.
No.: OMZ-21009-A Series Page 386of 428
PMLS Global Technical Support Dept

Below is the list of the affected P/Ns:

Item New
P/N Descriptions Old Ver.
No. Ver.
Manual
1 046-005026-00 A5 Operator's Manual(Czech) / 1.0
2 046-005027-00 A5 Operator's Manual(German) / 1.0
3 046-005028-00 A5 Operator's Manual(Italian) / 1.0
4 046-005029-00 A5 Operator's Manual(Norwegian) / 1.0
5 046-005030-00 A5 Operator's Manual(Polish) / 1.0
6 046-005031-00 A5 Operator's Manual(Serbian) / 1.0
7 046-005032-00 A5 Operator's Manual(Swedish) / 1.0
8 046-005033-00 A5 Operator's Manual(Hungarian) / 1.0
9 046-005041-00 A5 checkout list(Czech) / 1.0
10 046-005042-00 A5 checkout list(German) / 1.0
11 046-005043-00 A5 checkout list(Italian) / 1.0
12 046-005044-00 A5 checkout list(Norwegian) / 1.0
13 046-005045-00 A5 checkout list(Polish) / 1.0
14 046-005046-00 A5 checkout list(Serbian) / 1.0
15 046-005047-00 A5 checkout list(Swedish) / 1.0
16 046-005048-00 A5 checkout list(Hungarian) / 1.0
17 046-005055-00 A5 Reference Card(Czech) / 1.0
18 046-005056-00 A5 Reference Card(German) / 1.0
19 046-005057-00 A5 Reference Card(Italian) / 1.0
20 046-005058-00 A5 Reference Card(Norwegian) / 1.0
21 046-005059-00 A5 Reference Card(Polish) / 1.0
22 046-005060-00 A5 Reference Card(Serbian) / 1.0
23 046-005061-00 A5 Reference Card(Swedish) / 1.0
24 046-005062-00 A5 Reference Card(Hungarian) / 1.0

Affected FRU list:


Item No. FRU P/N Descriptions Ver.
1 115-020937-00 Drive Assembly(New Diff_pressure)FRU 1.0
2 115-018148-01 Ventilation kit(New Diff_pressure)FRU 1.0
801-0631-00088-
3 Main body of Drive Gas Assenbly 1.0
00
801-0631-00002-
4 Flow Meter Lighting Board PCBA 2.0
00
5 115-018165-00 Auxiliary Gas Outlet Assembly FRU(US) 2.0
6 115-018166-00 Auxiliary Gas Outlet Assembly FRU(Euro) 2.0

No.: OMZ-21009-A Series Page 387of 428


PMLS Global Technical Support Dept
7 115-018167-00 Auxiliary Gas Outlet Assembly FRU(SFDA) 2.0
8 115-014827-00 Oxygen origin inlet module(0632/DISS) 1.0
9 115-015435-00 O2 supply inlet assembly(0632/NIST) 1.0
801-0631-00040-
10 Fresh Flow Sensor Board 2.0
00
801-0631-00089-
11 sensor interface board PCBA 2.0
00
12 115-017879-00 anesthetic gas inlet tubes module(0632) 1.0
801-0631-00024-
13 Vaporizer Mounting Manifold 1.0
00
14 051-001519-00 0631 Backlight Inverter PCBA 1.0
15 115-018146-00 Main control board FRU 1.0
801-0631-00037-
16 Rear Door 2.0
00
801-0631-00084-
17 Preventative Maintenance kits (1 year) 3.0
00
801-0631-00085-
18 Preventative Maintenance kits (3 years) 3.0
00
801-0631-00141-
19 Maintenance kit of O-rings(FRU) 2.0
00
20 115-021249-00 Mid-plate FRU(0631/no ventilation valve) 1.0
21 040-001532-00 Vacuum flask 1.0
22 040-001533-00 Vacuum flask(no protection ) 1.0
23 115-022320-00 loop low cover 2 assembly FRU 1.0
24 115-021252-00 Upper cover 2 FRU(0631/no popoff) 1.0

Some changing units are not compatible with the old ones. Below is coordination table
for your reference.

New
A5 (0631) Before Now Compatibl
e With Old
Upper cover 2
043-001133-00 115-021252-00 Yes
FRU(0631/no popoff)
loop low cover 2
043-001135-00 115-022320-00 Yes
assembly FRU
Drive Assembly(New
801-0631-00047-00 115-020937-00 Yes
Diff_pressure)FRU
Ventilation kit(New
115-018148-00 115-018148-01 Yes
Diff_pressure)FRU
0631 Backlight
051-000176-02 051-001519-00 Yes
Inverter PCBA

No.: OMZ-21009-A Series Page 388of 428


PMLS Global Technical Support Dept
Main control board
801-0631-00137-00 115-018146-00 Yes
FRU
Li-ion Battery Pack 022-000008-00 115-018012-00 Yes
Caster Assembly
801-0631-00073-00 115-016956-00 Yes
(right)
Caster Assembly (left) 801-0631-00071-00 115-016955-00 Yes
Brake Assembly 801-0631-00094-00 115-016692-00 Yes
Base of Chesis
801-0631-00097-00 115-016691-00 Yes
Assembly
Rear shell of chassis 043-000781-00 043-003030-00 Yes
front shell of chassis 043-000782-00 043-003031-00 Yes
Left pedal of brake 043-000783-00 043-003032-00 Yes
801-0631-00024-00 Vaporizer
Mounting Manifold;
vaporizer mounting 115-017879-00,anesthetic gas
115-015456-00 Yes
manifold(AGC) inlet tubes module(0632);
115-017878-00 mechanical
ACGO(AG);
Anesthetic gas inlet anesthetic gas inlet tubes
115-015829-00 Yes
tubes module module(0632)115-017879-00

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 389of 428


PMLS Global Technical Support Dept

TMZ-13005-A5

Some changes have been implemented on A5 anesthesia machine

Technical News
Subject: Some changes have been implemented on A5 anesthesia machine.

RECORD IDENTIFIER

Ref No.:TMZ-13005-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Rusty Song Parts Change
Date: Feb 21th, 2013 Appearance
Model: A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

Some changes have been implemented on A5 anesthesia machine:

1. Release new Software Bundle V02.02.00


1) Add “History” function. The List Trends and Event Log can be displayed in History
menu.

No.: OMZ-21009-A Series Page 390of 428


PMLS Global Technical Support Dept

List Trends

No.: OMZ-21009-A Series Page 391of 428


PMLS Global Technical Support Dept

Event Log

2) Add “Capture Event” function. The ventilation parameters in use will be saved and
can be reviewed in History menu.

No.: OMZ-21009-A Series Page 392of 428


PMLS Global Technical Support Dept

Captured Event reviewed in History

3) Adjust “Discharge” function. The user can choose whether save the history logs
and restore default settings when discharging the machine. The Clear History
function of Discharge can be switched on or off in “Setup→System” menu.

Discharge Button

No.: OMZ-21009-A Series Page 393of 428


PMLS Global Technical Support Dept

Pop-up menu after pressing Discharge button

Discharge Settings button in System tab of Setup menu

Discharge Settings menu

4) O2 sensor monitoring function and alarms can be switched on or off in “Setup→


No.: OMZ-21009-A Series Page 394of 428
PMLS Global Technical Support Dept
System” menu.

O2 Cell Monitoring button in System tab of Setup menu

5) Add Volume waveform, which can be selected in “Setup→Display” menu. The


user can choose either Volume or Flow waveform.

Waveform Display button in Display tab of Setup menu

Volume Waveform

2. Optimize the structure of ORC Assembly and Restrictor Assembly. Below are the
pictures of the old and the new. Please apply “Resister block assembly (PN: 801-
0631-00119-00)” for the replacement of the old Restrictor Assembly. Please apply
“ORC Maintenance Kit (CE Common) FRU (PN: 115-017864-00)” for the replacement
of the old ORC Assembly. And there is a “Maintenance Instruction for Old ORC”
contained in the “ORC Maintenance Kit (CE Common) FRU (PN: 115-017864-00)”,
No.: OMZ-21009-A Series Page 395of 428
PMLS Global Technical Support Dept
please follow the Instruction for the replacement.

ORC Assembly Restrictor Assembly

The new ORC Assembly(PN: 115-003009-01)

The old ORC Assembly(PN: 801-0631-00041-00, not available)

No.: OMZ-21009-A Series Page 396of 428


PMLS Global Technical Support Dept

The new Restrictor Assembly(PN: 115-008789-01)

The old Restrictor Assembly(PN: 801-0631-00119-00, available)

3. Optimize the design of Pipeline Gas Supply Inlet Assembly and O2 Supply Three-way
Connector to eliminate the abnormal noise during ventilation. The PN doesn’t change.

Pipeline Gas Supply Inlet Assembly

No.: OMZ-21009-A Series Page 397of 428


PMLS Global Technical Support Dept

O2 Supply Three-way Connector


4. Add “Patient Circuit Blocking Kit (PN: 115-016612-00)” for the A5 without O2 Sensor.
It contains “Cap of oxygen battery joint” and “O2 sensor cover assembly”.

No.: OMZ-21009-A Series Page 398of 428


PMLS Global Technical Support Dept
5. Change the PN of Internal AG Module. The new PN: 115-002504-00 is available.

Internal AG Module

Below is the list of the affected P/Ns:


Item
P/N Descriptions Old Ver. New Ver.
No.
Software
1 110-001996-00 Bundle Version 02.01.00 02.02.00
Manuals
1 N/A N/A N/A N/A

Affected FRU list:

Item New
FRU P/N Descriptions Old Ver.
No. Ver.
Hardware
1 801-0631-00041-00 ORC Assembly 1.0 /
AION 03-31 Multigas
2 115-002504-00 / 1.0
Analyzer
Multigas Analyzer.AION,03-
3 9200-10-10529 1.0 /
31,60-10331-00
4 115-003009-01 ORC Assembly / 1.0
5 115-008789-01 Resister block assembly / 1.0
ORC Maintenance Kit(CE
6 115-017864-00 / 1.0
Common) FRU

No.: OMZ-21009-A Series Page 399of 428


PMLS Global Technical Support Dept
If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 400of 428


PMLS Global Technical Support Dept

TMZ-12037-A5

Some changes have been implemented on A5 anesthesia machine

Technical News
Subject: Some changes have been implemented on A5 anesthesia machine.

RECORD IDENTIFIER

Ref No.: TMZ-12037-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Rusty Song Parts Change
Date: Sep 20th, 2012 Appearance
Model: A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

Some changes have been implemented on A5 anesthesia machine:

1. Release new Software Bundle V02.00.00


a) Software upgrade records will be saved in “Review Logs”→“Service Event”.

b) When the date of birth inputted is invalid, the machine will display a pop-up
message.

No.: OMZ-21009-A Series Page 401of 428


PMLS Global Technical Support Dept

c) Fan speed can be checked in “Data Monitors” menu.

Choose
“Mainboard” here.

d) The alarm limits of PEAK and MV will become “OFF” when the operator choose
“OFF” in “Alarm ON/OFF” button in manual mode.

Choose “OFF”
here, the alarm
limits of “PEAK
and MV” will be
“Off”.

e) Add an alarm message “Auto Ventilation is Non-functional” in “Only Manual


No.: OMZ-21009-A Series Page 402of 428
PMLS Global Technical Support Dept
Mode”. “Only Manual Mode” will be activated if the user override the system self-
test or automatic circuit leak test fails.

.
f) Change the appearance and function of “Alarm” menu. The alarm limits of
anesthetic agent, CO2 and O2 can be adjusted if an AG module is used on the
machine.

Alarm menu without AG module

No.: OMZ-21009-A Series Page 403of 428


PMLS Global Technical Support Dept

Alarm menu with AG module

g) Change the appearance and function of “Setup” menu. Add “Gas bench Flow
Rate” in “General” menu. “Gas bench” means “Module slot” on A5. Add CO2
wave configuration in “Display” menu.

The “Pump
Rate” of AG
module can
be set here.

No.: OMZ-21009-A Series Page 404of 428


PMLS Global Technical Support Dept

The user can


configure CO2
wave here.

h) When the lever is in “Auto” position and the machine is not in “Standby”, a pop-up
message “Can only discharge in Manual Mode” will displayed if the user press
“Discharge” button.

i) Add “MR-WATO” protocol in “Serial Configuration” menu. “BeneLink” function is


now available on A5. “Benelink” allows the information (patient date, alarms and
etc) of A5 to be displayed, saved and printed through a BeneView patient
monitor. When this protocol is chosen, the configuration items will become
unchangeable.

No.: OMZ-21009-A Series Page 405of 428


PMLS Global Technical Support Dept

These items will


be unchangeable
if “MR-WATO” is
selected.

j) Add “MR-Link” protocol in “Serial Configuration” menu. Now serial port supports
“HL7 protocol”.

k) Add a technical alarm message “Fan Failure 02”, which will be triggered if the fan
speed is less than 3640rpm.

l) If the user switches the machine from “Auto Mode” to “Manual Mode”, status of
“Alarm ON/OFF” button can be restored to the status before last exit from Manual
mode. By comparison, before this software version, the status will be
automatically set to “ON” when the lever switches from “Auto” to “Manual”.

No.: OMZ-21009-A Series Page 406of 428


PMLS Global Technical Support Dept

This is “Alarm
ON/OFF” button.

2. The patient monitor bracket is now available on A5. The bracket supports 7000, 8000,
9000, 6800 and 6802 series patient monitors.

3. Module slot, which doesn’t support field upgrade, is added on A5. Mother board and
cylinder yoke are modified to support module slot.

No.: OMZ-21009-A Series Page 407of 428


PMLS Global Technical Support Dept

Infrared
Communication
Board

The positions
of connectors
are different.

Cylinder yoke for module slot Cylinder yoke without module slot

4. AG module is available now. When an AG module is plugged in the module slot, it will
work automatically. The user can not turn off the AG module in software. A5 doesn’t
support BIS function. So, it only supports AG module with or without O2. The measured
parameters will be displayed at the upper-left corner of the screen.

No.: OMZ-21009-A Series Page 408of 428


PMLS Global Technical Support Dept

Displayed screen with AG module

5. Add two gas inlet assemblies for the East Europe standard cylinder configuration. The
extra inlets are NIST standard and for O2 and N2O only. It only supports A5 which is
O2 driving.

No.: OMZ-21009-A Series Page 409of 428


PMLS Global Technical Support Dept

Inlets for east They are


Europe standard NIST
cylinders. standard
only.

6. Agent usage calculation function is available on A5. Add internal AG module and
modify vaporizer manifold and mother board for this function. It doesn’t support field
upgrade. The agent usage will be displayed at the upper-left corner of screen in
“Standby Mode”.

No.: OMZ-21009-A Series Page 410of 428


PMLS Global Technical Support Dept

Sampling tube from


the outlet of
vaporizer manifold.

Internal AG module

Sampling tube

Waste gas
scavenging
tube

No.: OMZ-21009-A Series Page 411of 428


PMLS Global Technical Support Dept

AG transfer
board
Sampling tube

Waste gas
scavenging
tube

7. Negative pressure suction device, which supports field upgrade, is available on A5.
According to the pipeline inlet type and driving gas of the machine, the suction device
has four types: O2 drive/NIST, O2 drive/DISS, Air drive/NIST and Air drive/DISS.

No.: OMZ-21009-A Series Page 412of 428


PMLS Global Technical Support Dept

Tee connector

Negative pressure
regulator

Gas supply
tube

Liquid collection
bottle

Below is the list of the affected P/Ns:


Item
P/N Descriptions Old Ver. New Ver.
No.
Software
1 110-001996-00 Bundle Version 01.05.00 02.00.00
2 110-000949-00 Boot Software 01.08.00 01.09.00
3 110-001090-00 Host Software 02.05.00 03.00.00
Manuals
1 046-002298-00 A5 Operator's Manual(CE) 3.0 4.0
2 046-004265-00 Bracket Installation Instruction(A5) N/A 1.0

Affected FRU list:

Item Old New


FRU P/N Descriptions
No. Ver. Ver.
Hardware
1 801-0631-00108-00 Mother Board PCBA 1.0 2.0

No.: OMZ-21009-A Series Page 413of 428


PMLS Global Technical Support Dept
2 115-015036-00 vaporizer manifold(AGC) N/A 1.0
3 115-016021-00 L-shape bracket assembly(FRU) N/A 1.0
4 115-015765-00 Prepak Handle FRU N/A 1.0
O2 cylinder yoke assembly(for
5 115-015777-00 N/A 1.0
module slot)
Air cylinder yoke assembly(for
6 115-015778-00 N/A 1.0
module slot)
N2O cylinder yoke assembly(for
7 115-015779-00 N/A 1.0
module slot)
FAN 12V 40*40*20mm 6.3CFM
8 024-000407-00 N/A 1.0
18dB type F
9 115-012781-00 Module Slot Assembly N/A 1.0
Tubes for agent usage calculation
10 115-015829-00 N/A 1.0
module
MULTIGAS ANLYZR AION3-
11 9200-10-10529 N/A 1.0
3,PM9E,9VET
12 801-0613-00033-00 AG Transfer Board PCBA N/A 1.0
O2 supply inlet assembly(East
13 115-015135-00 N/A 1.0
EUR 0631)
N2O supply inlet assembly(East
14 115-015136-00 N/A 1.0
EUR 0631)
Negative pressure suction Kit (Air
15 115-015266-00 N/A 1.0
drive, NIST)
Negative pressure suction Kit (Air
16 115-015267-00 N/A 1.0
drive, DISS)
Negative pressure suction Kit(O2
17 115-015268-00 N/A 1.0
drive, NIST)
Negative pressure suction Kit (O2
18 115-015269-00 N/A 1.0
drive, DISS)
GCX bracket Kit for monitor 7000
19 115-015769-00 N/A 1.0
8000
GCX bracket Kit for monitor 6802
20 115-015770-00 N/A 1.0
9000
21 115-015783-00 GCX bracket Kit for monitor 6800 N/A 1.0
AG module(with O2/with
22 801-0621-00142-00 N/A 1.0
accessory kit)
AG module(without O2/with
23 801-0621-00144-00 N/A 1.0
accessory kit)

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.


No.: OMZ-21009-A Series Page 414of 428
PMLS Global Technical Support Dept

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 415of 428


PMLS Global Technical Support Dept

TMZ-12024-A5

Some changes have been implemented on A5 anesthesia machine

Technical News
Subject: Some changes have been implemented on A5 anesthesia machine.

RECORD IDENTIFIER

Ref No.: TMZ-12024-A5 Relate To:


Version:1.0 Software/Manual Upgrade
Author: Rusty Song Parts Change
Date: July 2nd, 2012 Appearance
Model: A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

Some changes have been implemented on A5 anesthesia machine:

1. Release new Software Bundle V01.05.00


a) Changed the order of start-up self-test to do Auto mode first followed by Manual
mode so machine ends up with lever in Manual position.
b) Added graphics to initial leak test screen to clarify what we mean by “seal the Y-
piece”, set “APL to SP”, etc.

No.: OMZ-21009-A Series Page 416of 428


PMLS Global Technical Support Dept

c) Added info about sealing the sampling port and the condensate drain after failed
leak test.

d) Disabled “Continue” button if lever is in wrong position or Fresh Gas Flow is on


and display pop-up message.

No.: OMZ-21009-A Series Page 417of 428


PMLS Global Technical Support Dept

e) Lever in Manual position; turning on fresh gas flow ( ≥0.02 L/min of any
individual gas) takes you out of Standby. Lever in Auto position; turning on fresh
gas flow triggers a popup: Move lever to Manual position.

f) User can only enter Standby screen when fresh gas flow is off.
Must go Auto => Manual => Standby.
g) “Automatic Ventilation Disabled” is displayed as blue technical alarm. If the
reason that it is disabled is a leak test then it will display “Automatic Ventilation
Disabled – Leak Test Failed”.

h) If the user wants to override a startup self-test or leak test failure, a confirmation
dialog has been added that explains that only Manual mode will be available.
No.: OMZ-21009-A Series Page 418of 428
PMLS Global Technical Support Dept

i) Changed the results next to the Leak/Compliance Test button. Currently it


displays the last successful leak test. Changing it to the last test result
regardless of whether it passed, failed or was skipped.

j) Eliminated the Standby button. Only use Discharge to get to Standby.

k) Added RS-232 setup screen in System dialog (BAUD rate, parity, stop bits,
etc.).

No.: OMZ-21009-A Series Page 419of 428


PMLS Global Technical Support Dept

l) Added date of birth and weight to Demographics tab – display weight, date of
birth and calculated age.

m) Changed patient size nomenclature to “Adult”, “Pediatric” and “Infant”.

n) Added Default Gateway settings in Network Setup to facilitate more complex


network installations.

No.: OMZ-21009-A Series Page 420of 428


PMLS Global Technical Support Dept

2. Remove the button cell battery and Lithium-ion battery from periodic maintenance kit
(36 months).

Below is the list of the affected P/Ns:


Item
P/N Descriptions Old Ver. New Ver.
No.
Software
1 110-001996-00 Bundle Version 01.04.00 01.05.00
2 110-001103-00 Ventilator 1.11 1.12
3 110-001104-00 Electronic flow-meter 1.7 1.8
4 110-001090-00 Host Software 02.04.00 02.05.00
Manuals
1 046-003697-00 A5 Operator's Manual(Russian) N/A 1.0

Affected FRU list:

Item New
FRU P/N Descriptions Old Ver.
No. Ver.
Hardware
Periodic Maintenance kit (3
1 801-0631-00085-00 1.0 2.0
years)

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,

No.: OMZ-21009-A Series Page 421of 428


PMLS Global Technical Support Dept

Customer Service Departments


Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 422of 428


PMLS Global Technical Support Dept

TMZ-12016-A5

Some changes have been implemented on A5 anesthesia machine

Technical News
Subject: Some changes have been implemented on A5 anesthesia machine

RECORD IDENTIFIER

Ref No.: TMZ-12016-A5 Relate To:


Version:V1.0 Software/Manual Upgrade
Author: Rusty Song Parts Change
Date: 2012-6-12 Appearance
Model: A5 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

Some changes have been implemented on A5 anesthesia machine:

1. Mindray V60 Halothane Vaporizer has been added to the product configuration.
Below is the list of vaporizers suitable for A5 anesthesia machine.

Filling system
Brand Anesthetic agent
Key Filler Pour Fill Quik Fil
Enflurane Yes Yes No
Isoflurane Yes Yes No
Penlon
Sevoflurane Yes Yes Yes
Halothane Yes Yes No
Enflurane Yes Yes No
Mindray Isoflurane Yes Yes No
V60 Sevoflurane Yes Yes Yes
Halothane Yes Yes No

Below is the list of the affected P/Ns:


Item New
P/N Descriptions Old Ver.
No. Ver.
No.: OMZ-21009-A Series Page 423of 428
PMLS Global Technical Support Dept
Software
1 N/A N/A N/A N/A

Manuals
1 N/A N/A N/A N/A

Affected FRU list:

Item New
FRU P/N Descriptions Old Ver.
No. Ver.
Hardware
Halothane Vaporizer(Key
1 115-014138-00 N/A 1.0
Filler) bag(int)
Halothane Vaporizer(Pour
2 115-014139-00 N/A 1.0
Fill) bag(int)

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 424of 428


PMLS Global Technical Support Dept

TMZ-12009-A5

Some software upgrade of A5 anesthesia machine

Technical News
Subject: Some software upgrade of A5 anesthesia machine.

RECORD IDENTIFIER

Relate To: Some software upgrade of


Ref No.: TMZ-12009-A5
A5 anesthesia machine.
Author: Daisy Gong Software/Manual Upgrade
Date: Mar 26th, 2012 Parts Change
Model: A5 Appearance
Comments:NA Stop Production
Others

CONTENT

To Whom It May Concern,

Below content is only for your information.

Software update on A5 anesthesia machine:


1. Release new host software V02.04.00.
1) Add Flowmeter standard selection in “Factory Setup” menu. In ‘Setup’ menu,
select “Service”. Enter Service Password: 558188. (Fig 1) In “Factory Setup”
menu, different “Flowmeter Standard” can be selected. (Fig 2)

No.: OMZ-21009-A Series Page 425of 428


PMLS Global Technical Support Dept

Fig 1

Fig 2
2) Support Multilanguage version. Add five Multilanguage (Dutch, French,
Portuguese, Spanish, Turkish) options of language button. User can choose the
language that required.

No.: OMZ-21009-A Series Page 426of 428


PMLS Global Technical Support Dept

2. Release new ventilator module software V1.11.

Below is the list of the affected P/Ns:


Item
P/N Descriptions Old Ver. New Ver.
No.
Software
1 110-001090-00 Host Software 02.03.01 02.04.00
2 110-001103-00 Ventilator Module Software V1.10 V1.11
Manuals
1 046-003360-00 A5 Operator's Manual(Dutch) N/A 1.0
2 046-003361-00 A5 Operator's Manual(French) N/A 1.0
3 046-003362-00 A5 Operator's Manual(Portuguese) N/A 1.0
4 046-003363-00 A5 Operator's Manual(Spanish) N/A 1.0
5 046-003364-00 A5 Operator's Manual(Turkish) N/A 1.0

Affected FRU list:2

Item New
FRU P/N Descriptions Old Ver.
No. Ver.
Hardware
1 N/A N/A N/A N/A

No.: OMZ-21009-A Series Page 427of 428


PMLS Global Technical Support Dept

If you need download information or have any question, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments
Mindray Co., Ltd.

No.: OMZ-21009-A Series Page 428of 428