Professional Documents
Culture Documents
Appendix Ii B - BP 2020
Appendix Ii B - BP 2020
Appendix Ii B - BP 2020
Verification of photometric Determine the plrowmetric noise at a relevant photometric region of the spectrum using a suitable reflectance standard, for example,
noisi 1) white reflective cernmIc tiles or carbon-doped polymer standards. Follow the manufacturer's methodology and specifications.
Bench/mobile instrument Scan the reflectance low flux standard (e.g, 5 or 10 per cent, carbon-doped polymer Scan the transmittance high flux standard
standard) over a suitable wavelength range in accordance with the manufacturer's (e.g. 90 or 99 per cent, carbon-doped
recommendation and calculate the photometric noise as peak-to-peak noise. polymer standard) over a suitable
wavelength/wavenumber range in
accordance with the manufacturer's
recommendation and calculate the
photometric noise as peak-to-peak noise.
Process instrument As above, or if not pmctic:illy possible, use the standard built into the instrument for As above, or if not practically possible, use
noise testing and manufucwrer specifications. the standard built into the instrument for
noise testing and manufacturer
specifications.
(1) Verification of photometric l£nearity and Verifii:aliun ofphorometric noise are not required for instruments using methods to perform simple identifications which do
not use the photometric absorbances as part of model strategy (for example, simple correlation with absorbing wavelengths).
models. Wavelength compression (wavelength averaging) The expression Ai~: cent representing the specific absorbance
techniques may be applied to the data. of a dissolved substance refers to the absorbance of a 10 gIL
Model validation parameters solution in a 1 em cell and measured at a defined wavelength
Analytical performance characteristics to be considered for so that:
demonstrating the validation of NIR methods are similar to
Al per cent _ 108
those required for any analytical procedure. Specific I em - M
r
acceptance criteria for each validation parameter must be
consistent with the intended use of the method..Validation Unless otherwise prescribed, measure the absorbance at the
parameters for quantitative methods are accuracy, linearity, prescribed wavelength using a path length of 1 em, Unless
precision (repeatability and intermediate precision), otherwise prescribed, the measurements are carried out with
robustness and specificity. reference to the same solvent or the same mixture of
ONGOING MODEL EVALUATION solvents. The absorbance of the solvent measured against air
NIR models validated for use are subjected to ongoing and at the prescribed wavelength shall not exceed 0.4 and is
performance evaluation and monitoring of validation preferably less than 0.2. Plot the absorption spectrum with
parameters. absorbance or function of absorbance as ordinate against
wavelength or function of wavelength as abscissa.
TRANSFER OF DATABASES
Where a monograph gives a single value for the position of
When databases are transferred to another instrument,
an absorption maximum, it is understood that the value
spectral range, number of data points, spectral resolution and
obtained may differ by not more than ± 2: nm.
other parameters have to be taken into consideration. Further
procedures and criteria must be applied to demonstrate that Apparatus
the model remains valid with the new database or new Speettophotometers suitable for measuring in the ultraviolet
instrument. and visible range of the spectrum consist of an optical system
capable of producing monochromatic radiation in the range
of 200-800 nm and a device suitable for measuring the
absorbance.
B. Ultraviolet and Visible Absorption Control of wavelengths
Verify the wavelength scale using the absorption maxima of
Spectrophotometry holmium perchlorate solution R, the line of a hydrogen or
(Ph. Bur. method 2.2.25) deuterium discharge lamp or the lines of a mercury vapour
arc shown in Table 2.2.25.-1. The permitted tolerance is .
Determination of absorbance ± 1 nm for the ultraviolet range and ± 3 nm for the visible
The absorbance (A) of a solution is defined as the logarithm range. Suitable certified reference materials may also be used.
to base 10 of the reciprocal of the transmittance (1) for
monochromatic radiation: Table 2.2.25.-1. - Absorption maximafor control of wavelength
scale
241.15 nm (Ho) 404.66 nm (Hg)
253.7 nm (Hg) 435.83 nm (Hg)
T 1/10 ; 287.15 nm (Ho) 486.0 nm (D~)
10 intensity of incident monochromatic radiation; 302.25 nm (Hg) 486.1 nm (H~)
1 intensity of transmitted monochromatic radiation.
313.16 nm (Hg) 536.3 nm (Ho)
334.15 nm (Hg) 546.07 nm (Hg)
In the absence of other physico-chemical factors, the
361.5 nm (Ho) 576.96 nm (Hg)
absorbance (A) is proportional to the path length (b) through
365.48 nm (Hg) 579.07 nm (Hg)
which the radiation passes and to the concentration (c) of the
substance in solution in accordance with the equation:
Control of absorbance
A=ecb Check the absorbance using suitable filters or a solution of
molar absorptivity, if b is expressed in centimetres and c in potassium dichromate R at the wavelengths indicated in
moles per litre. Table 2.2.25.-2:, which gives for each wavelength the exact
www.webofpharma.com
n B V-A191
value and the permitted limits of the specific absorbance. The second-order-derivative spectrum at any wavelength ).., is
The table is based on a tolerance for the absorbance of related to concentration by the following equation..
± O.OI.
For the control of absorbance, use solutions of potassium
dichromate R that has been previously dried to constant mass
at 130 "C. For the control of absorbance at 235 urn,
257 urn, 313 urn and 350 urn, dissolve 57.0-63.0 mg of c' concentration of the absorbing solute, in grams per litre.
potassium dichromate R in O. 005 M sulfuric acid and dilute to
1000.0 mL with the same acid. For the control of Apparatus
absorbance at 430 nm, dissolve 57.0-63.0 mg of potassium Use a spectrophotometer complying with the requirements
dichromate R in 0.005 M sulfun'c acid and dilute to 100.0 mL prescribed above and equipped with an analogue resistance-
with the same acid. Suitable certified reference materials may capacitance differentiation module or a digital differentiator
also be used. or other means of producing derivative spectra. Some
methods of producing second-order-derivative. spectra
Table 2.2.25.-2 produce a.wavelength shift relative to the zero-order
Wavelength Specific absorbance Maximum spectrum and this is to be taken into account where
(nm) Al percent tolerance applicable.
1 em
www.webofpharma.com