Professional Documents
Culture Documents
Linac QA Form
Linac QA Form
1. Identification
1.1 Name of the institution :
1.2 Address with Pin code :
2
2. Mechanical Tests
2.1.1 Couch
2.1.2 Collimator
Note: Check for complete gantry rotation, at the gantry positions: 0°, 90°, 180°, & 270°.
3
2.1.2.19 Abutting of leaves on and :
off the field central axis
2.1.2.20 Reproducibility of leaf : , mm Tolerance: 1 mm
positions
2.1.2.21 Interlocks for leaf : Yes / No Tolerance: functional
2.1.2.22 Jaw position interlocks : Yes / No Tolerance: functional
2.1.2.23 Alignment of MLC axes and : Yes / No Tolerance: 1 mm
secondary collimator axes
2.1.2.24 Collimator centring
Use a field size of 10 cm x 10 cm and place a graph paper at isocentre with collimator at
0°. Record the maximum deviations of the upper and lower jaws for combinations of gantry and
collimator angles at 0°, 90°, 180°, and 270°. Repeat the observations for MLC leaf ends and
sides.
Maximum difference
(a) between jaws : , mm
(b) leaves : , mm
Tolerance: 2 mm
2.1.2.25 Correspondence between : Acceptable / Not acceptable
MLC irregular fields & shapes
and BEV target cross-section
Tolerance: 2 mm
The centring of the light source with collimator rotation shall be checked by using an
object having a right angle at one corner, suspended close to cross-wires. The radial and
transverse shifts in shadow of the object measured at the collimator angles of 90° and 270° can
be used to calculate the centring of the light source. Tolerance: I mm shift
Angle
Test 90° 270°
Radial shift, mm
Transverse shift, mm
The projection of the cross-wires is marked at collimator angles 0°, 90°, 180° and 270°
at 100 cm NTD. Measure the diameter of the smallest circle that includes all the cross-wire
projections at the isocentre.
The projection of the cross-wires with respect to the horizontal front pointer shall be
recorded at various gantry angles through 360° by placing a graph sheet just behind the
pointer.
Cross-wire projections at : , mm Tolerance: 2 mm φ Circle
different gantry positions
4
2.3 Gantry
2.4 Isocentre
(a) Mechanical
(b) Radiation
Stated Measured
Use a solid phantom and adjust its top surface at 100 cm. Raise the phantom by 30 cm
in steps of 1 cm and record ODI readings. Bring back the phantom to NTD and lower it by 20
cm in steps of 1 cm and record the corresponding ODI readings. All the readings must lie within
the given tolerance. This procedure will enable testing of ODI Scale in the range of 80 – 130
cm.
Optical distance indicator scale verification
Couch Position Optical distance indicator, readings Tolerance
NTD
Note: NTD (normal treatment distance) corresponds to 0 cm position of the couch, i.e.,
SAD / SSD = 100 cm.
5
2.6.4 Laser beam indicator
Check positions of 100 cm lasers with respect to mechanical isocentre. Check the
levelness (tolerance: 1 mm) and plumb (tolerance: 0.3°) using a level (horizontal & vertical
lasers) or a plumb (vertical lasers). Tolerance is met if displacement 20 cm from the isocentre is
less than 1 mm. Co-linearity of the lasers in the transverse direction should also be within 1
mm. Sagittal laser should pass through the mechanical isocentre. Ceiling laser position is
measured with respect to the mechanical isocentre. Check the co-linearity with respect to the
lateral and sagittal lasers. Enclose the results of the measurements for evaluation.
Align graph paper with cross-wires at collimator angle 0° and SAD = 100 cm. Adjust the
collimators to match several field sizes covering the clinical range. Verify that the field sizes
defined by upper and lower jaws respectively matches with the readings on the graph paper.
Record the observations for both symmetric and asymmetric motion of the jaws. Analyse the
data to ensure that the results are within the tolerance limits.
For MLC, use the manufacturer provided leaf position table in auto-cycle mode. Record
the leaf positions observed on the graph paper and verify that the measured leaf position
matches with the specifications within the tolerance limits. Tabulate and submit the
observations for evaluation and record.
Tolerance: 1 mm
6
2.8 Wedge position reproducibility : , mm Tolerance: 2 mm
3. Electrical Tests
3.2.1 Emergency stop push buttons to cut-off all electrical supply to the area in which the
accelerator is situated
Provided? : Yes / No
Working? : Yes / No
3.2.2 Emergency stop button to interrupt the irradiation
Provided? : Yes / No
Working? : Yes / No
7
3.6.2 Are all operating parameter push buttons functional? : Yes / No
3.6.3 Are safety switches operable from control console? : Yes / No
3.6.4 Is arc / rotation switch start / terminate the rotation at the
pre-assigned angle? : Yes / No
3.6.5 Is mode selection switch provided? : Yes / No
3.6.6 Does alteration of the mode during the irradiation cuts - off
the beam? : Yes / No
3.6.7 Are the range selection switches adjacent to each other? : Yes / No
3.6.8 Does changing of the range switch terminate irradiation? : Yes / No
3.6.9 Are dose monitor switches adjacent to each other and
close to the system used for preselection of monitor units?: Yes / No
3.6.10 Do the readings of the two monitor systems agree to
within 10% ? : Yes / No
3.6.11 Is irradiation possible only after having pre-selected
monitor units on the control console ? : Yes / No
3.6.12 Is the number of monitor units continued to be displayed
until reset to zero ? : Yes / No
3.6.13 Are residual monitor units displayed (at least for 20 mins.)
even after accidental or intentional power cut off ? : Yes / No
3.6.14 Do dose monitor systems function independently ? : Yes / No
3.6.15 Is each monitor system capable of terminating the
irradiation independently ? : Yes / No
3.6.16 Is independent timer provided? : Yes / No
3.6.17 Does the timer start and stop simultaneously with the
irradiation ? : Yes / No
3.6.18 Is wedge filter integrated into unit head with only one
possible mode of insertion and interlocked with console? : Yes / No
3.6.19 Are characteristics of the wedge filter displayed during use?: Yes / No
3.6.20 Is compensator interlocked with the console? : Yes / No
3.6.21 Is shielding tray interlock provided? : Yes / No
3.6.22 Is shielding tray interlock operational? : Yes / No
8
4. Photon Beam Characteristics
For accelerators capable of emitting photons of different energies, the following checks
shall be carried out, for each nominal beam energy:
The congruence between optical and radiation fields must be assessed at the depth of
the reference plane - which is usually at SSD + dmax or with the reference plane placed at the
isocentre of the accelerator - for square fields of 5 cm x 5 cm, 10 cm x 10 cm and 30 cm x 30
cm or for a field of maximum dimension, available on the unit, at the level of the reference plane
and for gantry positions of 0°, 90°, 180° and 270° for each nominal photon energy (MV). In all
cases, the separation between optical field edge and the 50% isodose line shall not
exceed 2 mm.
Place film at 100 cm SFD on a levelled surface. Set fields using specified collimator
settings. Measure 50% edge and distance between corresponding cross-wires. Insert graticule
tray. Deliver 70 MU (no build-up). Set 30 cm x 30 cm field, give 10 MU. Compare the
congruence between set optical and radiation fields.
Set field size to 10 x 10 cm, SSD to 100 cm. Score field edges and cross-wires. Deliver
70 MU (no build-up). Repeat for collimator angles of 90° and 270°.
Set field size to 10 x 10 cm, SSD to 130 cm. Score field edges and cross-wires. Deliver
100 MU (no build-up). Establish that the differences between optical and radiation fields with
changing SSD are within the tolerance limits.
Set jaws to 26 x 40 cm. Place film at 100 cm SFD on a levelled surface. Use 4 L – R
fields from the MLC “QA” file. Measure 50% edge and with between corresponding cross-wires.
Deliver 70 MU (no build-up) for each field. Compare the results. Retract MLC jaws. Set jaw to
30 x 30 cm and give 10 MU. Compare the deviations.
9
4.2 Field Flatness and Symmetry ++
Beam flatness and symmetry can be measured with any of the following methods:
Record the average unflatness and asymmetry over 80% of the field. Compare the
results with the data obtained at the time of commissioning.
It is important that field flatness and symmetry are tested over a range of available field
sizes and beam energies at the clinically relevant depths (e.g., dmax, & 10 cm) for at least four
cardinal gantry orientations of 0°, 90°, 180° and 270°. However, at least three field sizes,
namely, 5 cm x 5 cm, 10 cm x 10 cm and 30 cm x 30 cm should be covered during each QA
test.
4.2.1 Flatness
Along X - axis :
Along Y - axis : Tolerance: ± 3% **
Along Diagonal :
** To determine field flatness, obtain transverse beam profiles, at the depths of dmax and 10
cm, along the two orthogonal axes and diagonal axis of the beam. Obtain maximum (Vmax ) and
minimum (Vmin ) values of dose in the central 80% region of the respective beam profiles.
Calculate the required field flatness parameter, V, using the following equation:
for the respective beam axes. V values should lie within ± 3%. In addition, field flatness should
be evaluated at all other clinically relevant depths and for each nominal photon beam energy.
*** Deviation between the doses at lateral points and the radiation beam axis must
remain within .97 and 1.03. It should be ensured that the stability of the flatness with gantry
rotation is checked at the reference depth. This check should be repeated for a range of field
sizes and all nominal beam energies.
10
4.2.3 Symmetry
With regard to symmetry, the ratio between measured values for each pair of
symmetrical points (with respect to beam axis) for a range of field sizes and gantry orientations
must lie between .98 and 1.02 within the central 80% flattened beam area.
Place a Kodak X-Omat V film at a depth of dmax with SAD: 100 cm, field size 25 x 25 cm
and collimator at 0°. Irradiate the films at 0°, 90°, 180°, 270° gantry angles and compare the
asymmetry at different gantry angles. Alternatively, mount the Profiler to the head of the
machine and determine the asymmetry at different gantry angles.
Tolerance: 2%
++ Please attach transverse beam profiles and check films for evaluation of optical and
radiation beam congruence, field flatness and symmetry of the radiation field. The data
will be retained by the Division for records.
Following a standard dosimetry formalism (e.g., IAEA Protocol 398), check the output
calibration of the accelerator for all available beam energies – photon as well as electron
beams. Apply the end correction factor, if relevant. Record information about the equipment
used and any current calibration factors. Take al least three output readings for each beam
energy at dmax. Record the MU1, MU2, time, and mechanical backup counter. Calculate the
dose per MU.
Place the output checker in the beam as per morning set up procedure. Verify the
calibration by normalising the readings with the reference calibration field size (10 x 10 cm) and
compare with the tolerance values for each beam.
Tolerance: 1%
11
4.4(c) Output vs Gantry angle
Measure the output at isocentre for gantry angles of 0°, 90°, 180°, and 270° for each
beam. Calculate the ratio of the outputs at 90°, 180°, and 270° to that at 0°. Compare with the
baseline value.
Tolerance: 2%
4.4(d) Output factor constancy
Measure the output of the machine for different field sizes by placing an ion chamber on
the central axis in a phantom at the depth of SSD + dmax. Normalise the values with respect to
the reference field size of 10 x 10 cm. Compare the measured output factors with the baseline
values. Output factor constancy should be checked for each beam energy. Use a small ion
chamber for field sizes less than 6 x 6 cm. Use gantry angle: 0° and collimator angle: 0° for all
measurements.
Tolerance: 2%
Set the phantom at normal SSD and place an ion chamber at SSD + dmax. Measure the
outputs for different field sizes defined by exchanging collimator jaws (e.g., 10 x 15 cm & 15 x
10 cm). If the difference between output values obtained from exchanged jaw fields is 2% or
less then take the mean value for output factor calculations. However, for differences greater
than 2% calculate output factors for each jaw separately for use in treatment planning
calculations. Establish the constancy of output factors as discussed above.
4.4(g) Output for field sizes defined by asymmetric jaws, MLC, and DMLC
Set the phantom at normal SSD and place the ion chamber at SSD + dmax along axis of
the beam. Measure the output for different field sizes defined by the asymmetric jaws, MLC and
DMLC jaws and compare the results with 3D TPS calculated values and baseline values
established at the time of the commissioning.
Tolerance: 2%
12
TABLE 1
Specifications for photon beam energy determination
Energy, MV Dm, cm %DD, at 10 cm depth
The ratio of the ionisation (J20 / J10) measured at 20 cm and 10 cm depths respectively
for a field size of 10 cm x 10 cm at the detector level and with a constant source - detector
distance (e.g., 100 cm) is called Quality index (QI). Quality index is also termed as Tissue -
phantom Ratio (TPR). QI is always less than 1 and tends to 1 with increasing photon beam
energy. Hence, QI is defined as
J20
QI = ⎯⎯
J10
Under condition of infinite SSD, ionisation and absorbed dose ratios are identical. Table
2 presents QI values for a few selected beam energies. Values for intermediate energies can be
obtained by interpolation.
TABLE - 2
Quality Index for a few Photon Beam Energies
Nominal Energy Quality Index
4 MV 0.629 ± 0.015
6 MV 0.676 ± 0.009
9 MV 0.716 ± 0.012
12 MV 0.740 ± 0.001
18 MV 0.771 ± 0.005
23 MV 0.781 ± 0.001
Measuring the quality index periodically and comparing it with the corresponding base
line value check energy stability. This check assures that the beam energy has not changed
over the test interval.
13
Variation in quality index (J20 / J10) :
with time Tolerance: ± 1%
Since there is considerable variation in the values of dm and Ds with field size, it should
be carefully measured and documented for planning calculations.
%Ds
Note: The variation of Ds and dm with Fs must be determined for each nominal beam energy.
Please enclose central axis depth dose profiles for above fields for our records.
Wedge factor measurement should be done at the depth of dose maximum along the
radiation beam axis at 0° & 180° collimator position for all wedge fields and wedge angles
supplied with the accelerator. However, if motorised / dynamic / or auto-wedge filter is being
used for the treatment, then respective wedge factor should be measured at the depth of dose
maximum. Wherever the aforementioned wedge angles are not appropriate, the wedge angles
in use should be used for wedge factor determination.
Measured
Specified, by the manufacturer
Note: Wedge factors shall also be measured for “LEFT” and “RIGHT” orientations for 270° and
at 0° and 90° for “IN” orientation to observe any change with collimator orientation (chambers
shall be accurately centred for these measurements). Wedge factor also depends on the depth
of measurement and depth dependence must be correctly determined for planning calculations.
The transmission of the blocking tray is measured using an ion chamber in a phantom
under the condition listed below with and without the tray in place. Compare with baseline value.
14
5. Electron Beam Characteristics
The congruence between optical and radiation field should be assessed at the depth of
the reference plane with a distance between source and the entry surface on the phantom
corresponding to the normal treatment distance. These checks should at least be carried out for
square fields: 5 cm x 5 cm, 10 cm x 10 cm and 30 cm x 30 cm (or for a field of maximum
dimensions) defined at the level of the reference plane in normal conditions of use; for gantry
positions of 0°, 90°, 180° and 270° and for each nominal electron energy. In all cases, the
separation between optical field edge and the 50% isodose line must be ≤ 2 mm for all
field dimensions.
Flatness and symmetry should be checked at the depth of the reference plane for
normal conditions of use. The distance between source and the entry surface on the phantom
should correspond to the normal treatment distance. All measurements should be performed for
large fields (≥ 15 cm x 15 cm) and at least one measurement with the field of maximum
dimensions at the level of reference plane. Measurements should be repeated for gantry
positions of 0°, 90°, 180° and 270°, for each nominal beam-energy.
5.2.1 Flatness
Along X - axis :
Along Y - axis : Tolerance: R ≥ 0.85
Along diagonal axis :
Note: Check films and transverse beam profiles for all the fields and beam energies, available
on the unit, should accompany this data sheet for our records. Some alternative methods for
checking field flatness and symmetry of electron beam profiles are given in Appendix 3 for
reference.
15
5.2.4 Stability of Flatness
The stability of flatness during rotation of gantry has to be verified (1) at gantry positions
of 0°, 90°, 180° and 270°, (2) at the reference depth in a phantom, and (3) for maximum field
size available. The points of measurements are chosen simultaneously on the radiation beam
axis and laterally on the major axes of the field at 2/3 of the distance between the centre and
edge of the light field.
It is lateral distance between 80% and 20% isodose lines. All measurements are done at
the reference depth, which is usually at dmax .Penumbra width should be measured for all beam
energies.
Penumbra width :
The nominal energies of electron beams are tested using "range-energy" relationship.
The energy distribution at the surface of the medium is characterised by model energy, Ep,0
(MeV) which is related to the extrapolated practical range, Rp,(cm) in water by the formula:
Ep,0 = C1 + C2 * Rp + C3 * Rp 2
Where,
C1 = 0.22 MeV, C2 = 1.98 MeV . cm-1 & C3 = 0.0025 MeV . cm-2
The mean energy, E0 , is related to 50% range, R50, (cm) by the formula:
_
E0 = 2.33 * R50
Note: Rp and R50 should be determined under reference conditions, using sufficiently large
field size to ensure electron diffusion conditions.
16
5.6 Variation of surface dose and depth of dose maximum with field sizes
Field size, cm x cm
Test 5x5 10x10 15x15 20x20 30x30
Percent surface dose(Ds)
(with respect to the dose maximum)
Depth of dose maximum(dm), cm
Air-gap correction factors are measured for all electron energies using an ion chamber in
a polystyrene phantom placed at the isocentre for a 10 x 10 cm field at dmax. Measure air-gap
corrections for the clinical range of applicators and compare with the baseline data obtained at
commissioning.
Tolerance: 2%
Check a range of MUs and arc lengths, and compare stop angles and MUs delivered.
Also measure the doses delivered for the arc and compare to the doses delivered to that for the
static reference field.
Position a cylindrical chamber in air at isocentre. For each gantry angle, measure the
open field and the scanned field exposure. At gantry angles of 90° and 270°, also measure
exposure for collimator angles of 90°. Calculate the ratio of the scanned field to the reference
field outputs. (Note: check gantry clearance before commencement of arc mode tests).
Tolerance: 3% or 1 MU, whichever is greater, for arc angles between 45° and 89°
2% or 1 MU, whichever is greater, for arc angles 90° or greater
(Note: Measured dose for an arc should be within 2% of the static dose measured at 0°).
It should be ensured that for a given nominal beam energy, the monitor response is
independent of dose rate, the direction of the radiation beam, temperature and pressure. The
variation of monitor response with dose rate is related to the ion recombination in monitor
chamber due to inadequate voltage supply to the chamber. Hence, monitor should be calibrated
for each dose rate in clinical use.
6.1 Reproducibility
C = 100 * Rσ / Rav
Where, Rσ is standard deviation and Rav is the mean of observations R. If Ri is the ratio between
the ith "monitor units" and "the units indicated by the dosimeter" and n is the number of
measurements (at least 10 measurements should be made for one nominal energy in photon
17
mode and one nominal energy in electron mode within the range of dose rates commonly used
in actual clinical situations), then
n
Rσ = √ {Σ (Ri - Rav)2 / n}
i
n
Rav = Σ (Ri / n)
i
Note : These measurements should be made with both gantry and collimator at 0°.
6.2 Linearity
S=L/U
S factor :
(at least 10 measurements should be
taken to verify monitor linearity response) Tolerance: ±1%
With gantry at 0°, 90°, 180° and 270° a series of measurements are made with a dose
rate within the range commonly used in clinical practice. Calculate the mean of the
measurements corresponding to each gantry position and the mean of the sum, Rm. If Rmax
and Rmin are the maximum and minimum values of the mean respectively, then the ratio H is
defined by:
Rmax - Rmin
H = ---------------------- * 100
Rav
18
6.4 Dependence of the monitor response on field shape
The jaws of the collimator can influence the monitor response. In order to quantify this
effect, measurements are made with two fields of dimension: 5 cm x 30 cm and 30 cm x 5 cm,
normal to each other. The detector is maintained in the same position relative to the short side
of the field. Give a dose, say, of 100 MU and take at least five readings for each field shape.
Detector Response
Field size, cm x cm 1 2 3 4 5 Ave.
5 x 30
30 x 5
Test Readings
1 2 3 4 5 6
Monitor response, at different times of
the day
Note: This test shall be carried out for each nominal beam energy, in use.
7. Treatment Table
The treatment table should be carefully evaluated for it’s various motions and
movements so that it could be used for patient treatment with confidence. Besides performing
the tests described herein under, all linear and angular scale markings should also be checked.
Fix a sheet of paper on the top of the table, which is at the height of isocentre and with
table half extended longitudinally, rotate throughout its circular course. Observe the shift in
cross-wire position during the rotation. Repeat this procedure at least 3 times and record the
observations.
1 2 3
Diameter of the circle described :
by the projection of the cross-wires
Tolerance: ≤ 2 mm
19
7.3 Longitudinal movement
Measurements of radiation transmission through beam limiting jaws shall be made with
any residual aperture shielded by at least two TVL of x-ray shielding material. For non-
overlapping jaws, measurements shall be made at minimum radiation field size.
8(a) Each beam limiting jaw (excluding MLC jaws) shall attenuate x-radiation such that kerma
rate due to transmitted radiation does not exceed 2% of the maximum kerma rate
measured along the central axis of the beam at NTD in a 10 x 10 cm radiation beam.
8(b) For radiation fields of any size, the average kerma rate due to transmission through the
beam limiting jaws, including MLC jaws, shall not exceed 0.75% of the maximum kerma
rate on the central axis at NTD in a 10 x 10 cm radiation field.
If this limit is exceeded when areas of greater than 300 cm2 at NTD are protected by
MLC jaws, the conditions under which the limit is exceeded and the extent to which the limit is
exceeded shall be stated by the vendor of the equipment.
8(c) When a MLC is provided that by itself does not comply with the requirements of (a) & (b)
above, and consequently requires adjustable of interchangeable jaws in order to comply, these
shall be adjusted automatically to provide the minimum size radiation field surrounding the
radiation fields defined by MLC jaws.
The kerma rate due to transmitted radiation through the parts of a MLC jaws that project
into the rectangular field formed by the automatically adjustable jaws, referred to above,
shall not exceed 5% of the maximum kerma rate measured along the central axis at
NTD in a 10 x 10 cm radiation field.
20
9. Leakage Radiation Measurements
Measurements Tolerance
9.2 Kerma rate due to leakage radiations : ≤ 0.5% of the air kerma rate on
(excluding neutrons) at 1 m from the path : the central axis of the beam
of electrons between their origin and the :
target or electron window :
9.3 Kerma rate of leakage radiation through : Imparts less energy to patient
adjustable beam limiting devices : than a treatment field of 10 cm2
9.4.1 Inside the treatment area : < 0.1% of the photon dose rate
Note: State clearly the experimental set-up and detectors used for photon and neutron
leakage radiation levels measurement.
Draw a neat diagram (on a separate sheet of paper) of radiation installation indicating
clearly areas occupied by the radiation and non - radiation workers, including members of the
public. Earmark these locations as A, B, C,......, etc. Measure air kerma rate (of photons and
neutrons separately) at these locations under maximum scattering conditions for four cardinal
gantry positions of 0°, 90°, 180° and 270°. These measurements are essential for evaluation of
the safety integrity of radiation installation. Appropriate build-up caps should be used wherever
necessary. Tabulate the kerma rate values for analysis and evaluation.
21
10.1 Measured kerma rate levels for-------------------------------------- MV accelerator
11. Miscellaneous
2.
3.
4.
5.
6.
7.
Signature: Signature:
Name : Name:
Designation: Designation:
Date : Date:
22
APPENDIX 1
Reference Conditions
For the different measurements, it is preferable to use as the depth of the reference
plane, the recommended values established as function of the type and energy of the radiation
beam (NACP, 1980, AAPM, 1983; CFMRI No.2. 1986).
For measurements within photon beams, the reference plane should pass through the
isocentre of the electron accelerator. If not, the entry surface on the phantom is placed at the
normal treatment distance.
For measurements within electron beams, the entry surface on the phantom is placed at
the normal treatment distance. The phantom should be large enough to allow conditions of
maximal electron diffusion.
Measurements are made either on the radiation beam axis or in a plane perpendicular to
the radiation beam axis, at a depth within a "water- equivalent" phantom chosen from the table
below.
Photon Beam
Nominal energy, MV Depth of reference plane, in water phantom Field size, at NTD
E < 10 5 cm 10 cm x 10 cm
E ≥ 10 10 cm 10 cm x 10 cm
However, it is prudent to use a single depth of 10 cm for reference plane for all beam energies.
Electron Beam
Energy, Ep, MeV Depth of reference plane, in water Field size, at NTD
1≤E<5 Maximum absorbed dose on central axis 10 cm x 10 cm
5 ≤ E < 10 Dose maximum or 1 cm 10 cm x 10 cm
10 ≤ E < 20 Dose maximum or 2 cm 10 cm x 10 cm
20 ≤ E < 30 Dose maximum or 3 cm 10 cm x 10 cm
30 ≤ E < 50 Dose maximum or 3 cm 15 cm x 15 cm
23
APPENDIX 2
Essential Equipment Required for Dosimetric Measurements
1.1 Two ionisation chambers: One cylindrical and one flat (preferably with no polarity effect);
1.2 Associated electrometer - with preferably two opposite polarities and dual voltages;
1.3 Solid phantom - an assembly of tissue equivalent plates of different thickness;
1.4 Water phantom;
1.5 Radiographic films;
1.6 Photo-densitometer with manual or automatic scanning system;
1.7 Barometer;
1.8 Thermometer;
1.9 Chronometer
Other accessories are also required to carry out mechanical checks (e.g., precision spirit
level, plumb line, etc).
2. Dosimetric measurements
Several types of detectors can be used for dosimetry. The most commonly employed
detectors are:
Air cavity ionisation chambers are the detectors of choice for most measurements. They
shall meet requirement that: (a) their response is relatively independent of the influence of the
beam characteristics (beam quality, dose and dose rate) and (b) they possess good
reproducibility and repeatability. Cylindrical chambers which can be used for photon beams or
for electron beams with an energy of 10 MeV or more should preferably be small (internal
diameter < 6mm, length < 25 mm). The leakage and parasite currents produced in the chamber
and associated electrometer assembly shall be negligible fraction of the current measured
during irradiation of the chamber - particularly for measurements of low doses made at large
depths in the medium as well as at the edges and outside of the beam. For electron energies of
less than 10 MeV, flat chambers are recommended. When a polarity effect (particularly strong in
24
some flat chambers) is present, it is necessary to use the average of the measurements
obtained with negative polarity and positive polarity for all conditions of measurement.
Films give practical and rapid indications of the dose distribution in a given plane. The
advantage of films lies in their high spatial resolution, which is particularly useful in regions of
high dose gradients. The sensitometric curve depends on type of the film used and can even
vary between different batches of the same emulsion. It is therefore necessary to verify the
sensitometric curve for each batch. If the curve is not linear, the necessary corrections for non-
linearity need to be applied for correct dose evaluation. The film development process also
plays a vital role in dosimetry and needs to standardised with care.
Semiconductor diodes should only be used for relative measurements. Because of their
small size, they are particularly suitable for measuring dose distributions in regions of high dose
gradients. The use of these dosimeters requires some particular precautions since their
response exhibit dependence on energy, beam direction and temperature. In addition, large
variations in sensitivity can occur as they age.
• materials which present small variations with beam quality should only be used,
• calibration should be carried out with a suitable reference dosimeter in conditions close to
the actual use conditions,
• validity of the calibration limits should preferably be determined by testing (with a reference
dosimeter) the behaviour of the thermoluminiscent dosimeters when for example, the
energy, the depth and the dimensions of the field are varied.
3. Phantoms
All measurements should preferably be carried out in a water phantom but practical
considerations may necessitate the use of solid tissue-equivalent phantoms. However, it should
be mandatory to study the characteristics of the solid phantom before its use in practical
dosimetry studies. This will facilitate inter-comparison of the measured values in different
phantoms and thereby lead to proper interpretation of the experimental data.
25
APPENDIX 3
Where, Imax and Imin are maximum and minimum intensities within the region 1.5 cm inside
field borders (50% intensity level). Tolerance limits are 5% for low energy electron beams upto 9
MeV and 3% beyond 9 MeV.
Field Symmetry: Field symmetry (Ss) is defined by these manufacturers by the following
equation:
Where, AL and AR are areas under the left and right halves of the profile. Tolerance limit is 2%.
whichever is greater. The maximum and minimum signals are determined over the range of 2
cm inside the field borders. Ica is the signal on the central axis. Tolerance limit is 3% while 5%
is considered acceptable.
Sa = 100 * |Il / Ir - 1|
Where, Il and Ir are signals at two most asymmetric points within the region 2 cm inside the
geometric field borders. The tolerance limit is 2%.
Any of these definitions could be used for testing flatness and symmetry of a clinical
electron beam because definitions given in the main text and above give clinically acceptable
results.
26
APPENDIX 4
Leakage Radiation Measurement on a Medical Linear Accelerator
Leakage radiation is the radiation that emerges from the source housing through its
protective barriers. Besides, in an electron accelerator, some of the electrons being accelerated
may hit wave-guide tube and thus produce off-target radiations. Accelerators operating at
potentials 10 MV or above produce unwanted neutrons with a range of energies. Contributions
of each component must be evaluated separately. It is mandatory to keep leakage radiation
levels within the limit prescribed by the national competent authority.
The equipment shall be provided with radiation shielding which shall be so designed that
the kerma rate due to leakage radiation (excluding neutrons) at any point outside the maximum
useful beam, but inside a plane circular area of radius 2 m centred around, and perpendicular
to, the central axis of the beam at the normal treatment distance, shall not exceed 0.2% of air-
kerma rate on the axis at the same distance. The air-kerma rate should be measured with the
useful beam blocked with thick absorbing plug. Except in area defined above, the kerma rate of
leakage radiation (excluding neutrons) at 1 m from the path of electrons between their origin
and the target or electron window shall not exceed 0.5% of air-kerma rate on the central axis of
the beam at the normal treatment distance.
The contribution of neutrons to the dose inside and outside the treated area should be
kept as low as practicable. Inside the field the neutron tissue-kerma rate should be kept below
1% of the x-ray kerma rate. Measurements made in primary beam from electron accelerator
have indicated that dose equivalent rate of neutrons does not exceed 1% of the dose equivalent
rate of photons (NBS Special Publication 554). This implies that the neutron absorbed dose rate
is never greater than 0.1% of photon dose rate and makes only a small addition to the
therapeutic dose. Neutrons outside the primary x-ray beam make only marginal contribution to
the integral dose received by the patient. Outside the treated area the neutron kerma rate shall
be reduced as far as practicable.
The adjustable beam limiting diaphragms shall be so constructed that leakage radiation
imparts less energy to the patient than is imparted by a treatment field of 10 cm2. This implies
that for a field of 10 cm x 10 cm the leakage radiation shall contribute less than 10% of the
energy imparted by the useful beam. Similarly, if the maximum field size is 35 cm x 40 cm, the
transmission of radiation through the diaphragms shall not exceed 0.8 %.
X-ray and electron accelerators working above 10 MeV may produce a significant
amount of photo-disintegration of the nucleus. For most elements, the threshold for photo-
disintegration occurs between 10-20 MeV. A variety of methods - like activation foils, P2O5,
track registration in organic foils, BF3 and other counters in moderating hydrogenous spheres, -
could be used to measure the neutron contribution to the patient dose. More than one
measurement technique should be used to get reliable results.
27
Measurements outside the treatment room have in some instances shown a significant
dose equivalent rate of neutrons outside the door to the labyrinth. Care should therefore be
taken both at design stage and during protection surveys to ensure that there is no significant
hazard from neutrons outside the treatment room.
The presence of neutrons in the primary beam may also result in measurable
radioactivity in components of diaphragm, target, etc. The amount of radioactivity shall be
evaluated and adequate precautions should be taken to avoid unnecessary exposures during
patient set-up.
The leakage radiation through the source housing in beam "ON" position should be
measured as follows:
28